DOG-1/-3. Oscillation Flowmeter. for gases. Measuring ranges: m 3 /h air : 120 C. p max. : PN 40; t max
|
|
- Marshall Patterson
- 6 years ago
- Views:
Transcription
1 Oscillation Flowmeter for gases measuring monitoring analysing DOG1/3 Measuring ranges: m 3 /h air p max : PN 40; t max : 120 C Connection: flange DN DN 400 Material: cast iron, steel or stainless steel Accuracy: ±1.5 % of measured value No moving parts Longterm stability S5 1/ KOBOLD companies worldwide: ARGENTINA, AUSTRIA, BELGIUM, BULGARIA, CANADA, CHILE, CHINA, COLoMBIA, CZECHIA, DOMINICAN REPUBLIC, EGYPT, FRANCE, GERMANY, GREAT BRITAIN, HUNGARY, INDIA, INDONESIA, ITALY, MALAYSIA, MEXICO, NETHERLANDS, PERU, POLAND, ROMANIA, SINGAPORE, SOUTH KOREA, SPAIN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, turkey, USA, VIETNAM KOBOLD Messring GmbH Nordring 2224 D65719 Hofheim/Ts. Head Office: +49(0) (0) info.de@kobold.com 1
2 Oscillation Flowmeter Model DOG1, DOG3 Description The KOBOLD flowmeter DOG1 and DOG3 are used for noncontact flow measurement of gases. The medium flows through an orifice in a tube. Bypass bores are located at the sides. The dynamic pressure at the orifice causes part of the gas volumetric flow to flow into the bypass. The division ratio remains constant over the whole measuring range. The bypass channel contains the Oscillator the measuring cell itself. When the gas flows through the measuring cell, a gas column oscillates in a Ushaped channel mounted to the left and right. This oscillation frequency is proportional to the flow velocity and thus to the total volume flow. The oscillation frequency is sensed with a hot wire sensor. An electrical alternating signal is generated that is displayed in the seriesconnected. Application The inner, connected flow channels are generously dimensioned. The constant changes of direction of the flow in the channels have a selfcleaning effect. The devices are therefore extremely dirt resistant and have no consumables. The mounting position can be chosen at will. When condensate forms in the gas, the horizontal mounting position with the sensing element pointing upwards is recommended. The gas flow velocity anywhere in the pipework upstream of the flowmeter should not exceed the sound velocity. Pressure drops above critical and pulsating streams must be avoided. The recommended inlet pipe section is 10xDN and the outlet pipe section 5xDN. Areas of Application Compressed air Natural gas, biogas, fermentation gas Propane Hydrogen gas Nitrogen Argon Technical Data Measuring accuracy: ±1.5% of meas. value (at Qt 100%*) ±5% of measured value (at 1%Qt *) *The lower limit Q t depends on the density Qt= 8% at density 1 kg/m 3 Qt= 4% at density 2 kg/m 3 Qt= 2% at density 4 kg/m 3 Qt= 1% at density 8 kg/m 3 Repeatability: 0.1% of measured value Temperature span: C C (ExVersion) option C (Exversion) Ambient temperature: max. 80 C C (ExVersion) Operating pressure: DOG11/12.., DOG31/32..: PN 16 DOG12/13.., DOG32/33..: PN 40 Span: DOG1..: 1:100 DOG3..: 1: 50 Sensor: hotwire, RDC Protection: IP 65 Materials Housing: DOG11..: cast iron DOG13..: steel DOG33..: steel DOG12/32..: st. steel / Orifice: stainless steel Sensing element: polyphenylene sulfide (PPS) Sensor: platinum Gaskets: silicone, nitrile or FPM Electronics DOG X Electrical connection: conduit thread Protection: IP 65 Display: 4digit LCD display, counter Power supply: 230 V AC 10%, +12%/ Hz Input: hot wire sensor Pulse output 1: 12 V DC, max.100 ma, open collector Pulse output 2: potentialfree contact decadic 250 V AC,max. 3 A Analogue output: ma max. 500 Ω Ambient temperature: C Permissible distance: max. 50 m to DOG1, DOG3 max m to DOG2 Connection cable: minimum 0.5 mm 2, screening recommended ExVersion: ` II 1G EEx ia IIC T4 2 1/
3 Oscillation Flowmeter Model DOG1, DOG3 1/ Electronics DOG W/Z/Q for DOGflowmeters and all meters with pulse detection by means of a proximity switch (model NAMUR) Electrical connection: terminal Protection: IP 65 Mounting type: wall mounting Display: 2 x digits LCD with backlit display line 1: flow rate (Nm 3 /h, Nm 3 /min, m 3 /h, m 3 /min, kg/h, kg/min), 7 digits, floating decimal point current pressure (barg), only available for option Q Default pressure for all models line 2: totaliser (Nm 3, m 3, kg), 12 digits, floating decimal point current temperature ( C, F, K), only available for option Q Default temperature for all models Engineering units are configured at the factory and should thus be mentioned in the P. O. while ordering. DOG W Display indication: flow rate/accumulated volume, resettable Output: pulse, 1224 V open collector or 12 V DC active DOG Z Display indication: Output: DOG Q Flow calculator input signal from: flow rate/accumulated volume, resettable 1/ pulse, 1224 V open collector or 12 V DC active 2/ ma, galvanic isolated (a) gas flowmeter pulse, (b) pressure transmitter ma (1x bar g), (c) temperature sensor Pt 100 Builtin calculation function for PT correction Display indication of all process parameters Output signal: (i) pulse, resettable, for selected units, e.g. m 3, Nm 3 or kg (ii) ma, galvanic isolated, with selected volume or mass units, such as m 3 /h, Nm 3 /h or kg/h All models have a builtin linearisation function of the flow meter signal Input signal (pulse train) Flowmeter DOG: directly or Proximity switch: max 8 V DC (high level) Speed range: Hz Over voltage protection: 24 V Input signal (analogue) Temperature sensor: Pt 100 (four wire lead comp.) Pressure transducer: ma Basic measuring resolution: 12 bit Accuracy: 0.05 % at 20 C Update rate: 1 update/sec Reverse polarity: no ill effects Over current limit: 12 V DC, 100 ma (fault protected) Output signal (i) pulse train assignable to uncompensated or PT compensated volume total, or mass total 12 V DC active (voltage pulse) alt. passive (open collector, max 24 V DC load) pulse width adjustable (ii) isolated analogue output signal ma assignable to uncompensated or PT compensated flow rate Adjustable damping: time constant sec Accuracy: 0.1% of full scale at 20 C Update rate: 5 updates/sec Maximum load: 500 Ω 3
4 Oscillation Flowmeter Model DOG1, DOG3 Order Details for DOG1 with flange (Example: DOG1101L F25N S Q) Measuring range air [m³/h] Material cast iron DOG1101L.. DOG1102L.. DOG1103L.. DOG1107L.. DOG1108L.. DOG1109L.. DOG1110L.. DOG1111L.. DOG1112L.. DOG1116L.. DOG1117L.. DOG1118L.. DOG1119L.. DOG1120L.. DOG1121L.. Model Connection flange Gasket Remote Material Material Standard PN 16 Special PN 40 steel st. steel only GG, VA only steel, st. steel DOG1304L.. DOG1305L.. DOG1306L.. DOG1313L.. DOG1314L.. DOG1315L.. DOG1322L.. DOG1323L.. DOG1324L.. DOG1325L.. DOG1326L.. DOG1327L.. DOG1328L.. DOG1329L.. DOG1330L.. DOG1331L.. DOG1332L.. DOG1333L.. DOG1334L.. DOG1335L.. DOG1336L.. DOG1337L.. DOG1338L.. DOG1339L.. DOG1340L.. DOG1341L.. DOG1342L.. DOG1201L.. DOG1202L.. DOG1203L.. DOG1207L.. DOG1208L.. DOG1209L.. DOG1210L.. DOG1211L.. DOG1212L.. DOG1213L.. DOG1214L.. DOG1215L.. DOG1216L.. DOG1217L.. DOG1218L.. DOG1219L.. DOG1220L.. DOG1221L.. DOG1222L.. DOG1223L.. DOG1224L.. DOG1225L.. DOG1226L.. DOG1227L.. DOG1228L.. DOG1229L.. DOG1230L.. DOG1231L.. DOG1232L.. DOG1233L.. F25N = DN 25 F25S = DN 25 F32N = DN 32 F32S = DN 32 F40N = DN 40 F40S = DN 40 F50N = DN 50 F50S = DN 50 F65N = DN 65 F65S = DN 65 F80N = DN 80 F80S = DN 80 F1HN = DN 100 F1HS = DN 100 F1ZN = DN 125 F1ZS = DN 125 F1FN = DN 150 F1FS = DN 150 F2HN = DN 200 F2HS = DN 200 F2FN = DN 250 F2FS = DN 250* F3HN = DN 300 F3HS = DN 300 F3FN = DN 350 F4HN = DN 400 * not for DOG12 (stainless steel) ** Engineering units are configured at the factory and should thus be mentioned in the P.O. while ordering Please state the exact operating conditions (gas types, flow volume, pressure, temperature, installation position etc.) when ordering. S = silicone N = nitrile V = FPM with external...e RDC input, without display, with EX protection, with analogue and pulse output...x RDC input, with display for instantaneous value and total, with EX protection, with analogue and pulse output...w** resettable pulse output...z** resettable, pulse output, ma...q** flow calculator, input: pulse, Pt 100, ma (pressure) output: pulse and ma 4 1/
5 Oscillation Flowmeter Model DOG1, DOG3 Order Details for DOG3 with flange (Example: DOG3301L F25N S Q) Measuring range air [m³/h] Material steel DOG3301L.. DOG3302L.. DOG3303L.. DOG3307L.. DOG3308L.. DOG3309L.. DOG3310L.. DOG3311L.. DOG3312L.. DOG3313L.. DOG3314L.. DOG3315L.. DOG3316L.. DOG3317L.. DOG3318L.. DOG3319L.. DOG3320L.. DOG3321L.. DOG3322L.. DOG3323L.. DOG3324L.. DOG3325L.. DOG3326L.. DOG3327L.. DOG3328L.. DOG3329L.. DOG3330L.. DOG3331L.. DOG3332L.. DOG3333L.. DOG3334L.. DOG3335L.. DOG3336L.. DOG3337L.. DOG3338L.. DOG3339L.. DOG3340L.. DOG3341L.. DOG3342L.. Model Connection flange Gasket Remote Material Standard PN 16 Special PN 40 st. steel only GG, VA only steel, st. steel DOG3201L.. DOG3202L.. DOG3203L.. DOG3207L.. DOG3208L.. DOG3209L.. DOG3210L.. DOG3211L.. DOG3212L.. DOG3213L.. DOG3214L.. DOG3215L.. DOG3216L.. DOG3217L.. DOG3218L.. DOG3219L.. DOG3220L.. DOG3221L.. DOG3222L.. DOG3223L.. DOG3224L.. DOG3225L.. DOG3226L.. DOG3227L.. DOG3228L.. DOG3229L.. DOG3230L.. DOG3231L.. DOG3232L.. DOG3233L.. DOG3234L.. DOG3235L.. DOG3236L.. DOG3237L.. DOG3238L.. DOG3239L.. DOG3240L.. DOG3241L.. DOG3242L.. F25N = DN 25 F25S = DN 25 F40N = DN 40 F40S = DN 40 F50N = DN 50 F50S = DN 50 F65N = DN 65 F65S = DN 65 F80N = DN 80 F80S = DN 80 F1HN = DN 100 F1HS = DN 100 F1ZN = DN 125 F1ZS = DN 125 F1FN = DN 150 F1FS = DN 150 F2HN = DN 200 F2HS = DN 200 F2FN = DN 250 F2FS = DN 250 F3HN = DN 300 F3HS = DN 300 F3FN = DN 350 F3FS = DN 350 F4HN = DN 400 F4HS = DN 400 * Engineering units are configured at the factory and should thus be mentioned in the P.O. while ordering S = silicone N = nitrile V = FPM with external...x RDC input, with display for instantaneous value and total, with EX protection, with analogue and pulse output...w* resettable pulse output...z* resettable, pulse output, ma...q* flow calculator, input: pulse, Pt 100, ma (pressure) outputs: pulse and ma Please state the exact operating conditions (gas types, flow volume, pressure, temperature, installation position etc.) when ordering. Order Details Spare Sensor for gas flowmeter max. medium temperature: +60 C Order Code: ERSDOG5193 1/
6 Oscillation Flowmeter Model DOG1, DOG3 Pressure Loss/Flow The diagram applies for gases with a density of air at NPT (0 C and 1000 mbar). The pressure loss is always proportional to the density of the gas. For example, the pressure loss doubles at 100% higher operating pressure. Calculating the Actual Density The actual density can be calculated with the following formula: D = D 0 *P *T 0 T D 0 = density at 1 bar abs. and 0 C (= 273 K) T = temperature in K (= C for example 20 C = = 293 K) T 0 = 273 K P = operating pressure in bar (absolute pressure) Dimensions and Weight DOG1 L H Calculating the Norm Flow P 273 Q N = Q T Q N = norm flow at bar abs. and 0 C Q = operating flow P = operating pressure in bar (absolute pressure) T = operating temperature in K D Dh Ds DN L D Dh Ds 6 Number of holes H H with AVF Weight [kg] 1/
7 Oscillation Flowmeter Model DOG1, DOG3 Dimensions and Weight DOG3 Dimensions of Electronics DOG E/X DN DN D Weight [kg] Dimensions of Electronics DOG...W/Z/Q D=5 194 Accessory Valve in the bypass (between measuring tube and measuring cell) for easy sensor changing and for protection of the sensor when starting the installation. 1/
Oscillation Flow Meter for Gases
Oscillation Flow Meter for Gases measuring monitoring analysing Measuring ranges: 0.2 to 000 m 3 /h air pmax: PN 0, tmax: 0 C Connection: flange DN 25 DN 00 : Cast Iron, steel or stainless steel Accuracy:
More informationThermal Mass Flow Meter for Gases
Thermal Mass Flow Meter for Gases measuring monitoring analysing Direct mass flow rate measurement of gases Measuring accuracy: ±1.0% f.s. +0.5% m.v. pmax PN 16, tmax 175 C Fast response time No moving
More informationDVZ. Vortex Flowmeter Compact. for low viscosity liquids. Range: l/min. Accuracy: ±2.5% of full scale. p max : 20 bar; t max : 80 C
Vortex Flowmeter Compact for low viscosity liquids measuring monitoring analysing DVZ Range: 0.4. 000 l/min Accuracy: ±.% of full scale p max : 0 bar; t max : 80 C Connections: G ¼... G, ¼" NPT..." NPT
More informationTSK. Flap-Type Flowmeter. Range: to GPM Accuracy: ± 2.5% of Full Scale P max. : 580 PSI T max
Flap-Type Flowmeter measuring monitoring analyzing TSK Range: 6.6...26.4 to 880...6600 GPM Accuracy: ± 2.5% of Full Scale P max : 580 PSI T max : -40 572 F Connection: Wafer Flange 1.5" 20" ASME Material:,
More informationMIK. Electromagnetic Flowmeter Compact. for conductivity liquids. Range from liquids, acids and caustic solutions:
Electromagnetic Flowmeter Compact for conductivity liquids measuring monitoring analysing MIK Range from liquids, acids and caustic solutions: 0.0-0. 3-700 l/min frequency-, switching-, analogue output
More informationIndustrial Resistance Thermometers according to DIN Model MWD
Industrial Resistance Thermometers according to DIN measuring monitoring analysing MWD OMeasuring O range: -98... +600 ºC O O Pt00 sensor: class A, class B, class /3, class /0, or O O Output: resistance
More informationMIK. Electromagnetic Flowmeter Compact. for conductivity liquids
Electromagnetic Flowmeter Compact for conductivity liquids measuring monitoring analysing MIK MIK with frequency-, switching-, analogue output MIKwith digital plug on display Range from liquids, acids
More informationFull Metal Variable Area Flow Meter and Counter
Full Metal Variable rea Flow Meter and Counter for horizontal and vertical mounting measuring monitoring analysing Special versions up to 600 bar Measuring range: 10-100...00-000 L/h water 0.3-3.0...-0
More informationInfrared Hand-Held Thermometers
Infrared Hand-Held Thermometers measuring monitoring analysing Model: TIR-H... Measuring range from -2... 5 C to -32... 9 C Measuring accuracy: ±1% to ± 2% of measured value Option: analogue output, RS
More informationUltrasonic Gas Flow Meter
P r o d u c t i n f o r m at i o n FLOWSIC600 Ultrasonic Gas Flow Meter For custody transfer and process applications Analyzers and Process Instrumentation FLOWSIC600 For Custody Transfer and Process Applications
More informationPressure Measurement. Transmitters for basic requirements SITRANS P MPS (submersible sensor) Transmitter for hydrostatic level 2/27
Overview Siemens AG 011 Function s are for measuring the liquid levels in wells, tanks, channels and dams. Diaphragm p Sensor U const. U + (brown) I EM I -1 (blue) 0, U B s are submersible sensors for
More informationOperating Instructions for Digital Manometer. Model: MAN-SD
Operating Instructions for Digital Manometer Model: MAN-SD 1. Contents 1. Contents...2 2. Note...3 3. Instrument Inspection...3 4. Regulation Use...4 5. Operating Principle...4 6. Mechanical Connection...5
More informationTurbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas
Turbine Meter TRZ 0 PRODUCT INFORMATION Reliable Measurement of Gas TURBINE METER TRZ 0 Method of operation, Construction Method of operation The TRZ 0 turbine meter is a flow meter suitable for gas measurement
More informationv = m / s Thermal Flow Sensors TA with thin-film sensor element Probes Probe Guides Measuring Tubes
Thermal Flow Sensors TA with thin-film sensor element Probes Probe Guides Measuring Tubes Examples of Application Flow measurements also in hazardous locations X zone 0 or zone 1 in clean rooms in air
More informationMass Flow Meter and Controller
ass Flow eter and Controller for Gases measuring monitoring analysing asuring range: 0-10 m N /min...0-500 N /min ccuracy: ±1.5% f.s. pmax 35 bar; tmax 50 C nalogue output: 0-5 V or 4-20 m Digital display
More informationInfrared Hand-Held Thermometers
Infrared Hand-Held Thermometers measuring monitoring analysing Model: TIR-H... Measuring range from -2 to 5 C to -32 to 9 C Measuring accuracy: ±1% to ± 2% of measured value Option: analogue output, RS
More informationElectromagnetic Flowmeter
Overview Flokal B.V. smart electromagnetic flow meter is hallmarked by its high performance and reliability that are based on successful, field-proven technology. It s being widely used in industries such
More informationED 701 General Industry Pressure Transmitter
ED 701 General Industry Pressure Transmitter Standard industrial process connections Complete range of electrical connections 4... 20 ma and Voltage outputs Accuracy: 0.1%, 0.2% and 0.4% FS Quick response
More informationMX Series COMPACT AND USER FRIENDLY. Metal Tube Type Variable Area Flowmeter OUTLINE STANDARD SPECIFICATION MODEL CODE MX-400, MX-71 MX-52E
OMPT ND USER FRIENDLY MX Series Metal Tube Type Variable rea Flowmeter OUTLINE MX Series flowmeter is the metal tube type variable area flowmeter having the unified face-to-face dimension. The compact
More informationSECTION 1 - WHAT IS A BTU METER? BTU's = Flow x ΔT Any ISTEC BTU Meter System consists of the following main components:
SECTION 1 - WHAT IS A BTU METER? ISTEC BTU Meters measure energy usage by multiplying flow rate and temperature difference. As the water (or other liquid) passes through these lines, the multi-wing turbine
More informationTransmitters for basic requirements SITRANS P MPS (submersible sensor) Transmitter for hydrostatic level
Overview Function s are for measuring the liquid levels in wells, tanks, channels and dams. Sensor Diaphragm p U const. U +2 (brown) I EM I - (blue) 0, U B s are submersible sensors for hydrostatic level
More informationPro-V Multivariable Flowmeter Model M22 In-line Vortex
Pro-V Multivariable Flowmeter Model M22 In-line Vortex Pro-V TM Advantage: Volumetric or mass flow monitoring of most liquids, gases, and steam Multivariable meter delivers mass flow, temperature, pressure,
More informationv = 00,20 m/s Thermal Flow Sensors TA with thin-film sensor element Probes Probe guides Measuring tubes
with thin-film sensor element Probes Probe guides Measuring tubes Examples of Application Flow measurements also in hazardous locations Ex zone 0 or zone 1 in clean rooms in air conditioning and ventilating
More information2017 Source of Foreign Income Earned By Fund
2017 Source of Foreign Income Earned By Fund Putnam Emerging Markets Equity Fund EIN: 26-2670607 FYE: 08/31/2017 Statement Pursuant to 1.853-4: The fund is hereby electing to apply code section 853 for
More informationDS 200P. Electronic Pressure Switch. Pressure Ports And Process Connections With Flush Welded Stainless Steel Diaphragm
Pressure Ports And Process Connections With Flush Welded Stainless Steel Diaphragm accuracy according to IEC 600: standard: 0. % FSO option: 0. % FSO Nominal pressure from 0... 00 mbar up to 0... 0 bar
More informationCoriolis Mass Flowmeter. (MCM Series) 100% Customer Satisfaction MAXIFLO TM. Coriolis Mass Flowmeter Series MCM
MAXIFO TM Features MaxiFlo Coriolis Mass Flowmeter series MCM is a mass flow meter that measures mass flow, density and temperature of liquids, slurries and gases. And it even measures oil content in water.
More informationPressure Measurement Single-range transmitters for general applications
Siemens AG 07 Overview Function SITRANS LH00 Transmitter for hydrostatic level U const. p U I EM Sensor Connection for auxiliary power supply Vent pipe with humidity filter Protective conductor connection/
More informationTF 212 / TF 212-Ex. Field mounted temperature transmitters, PROFIBUS PA, Pt 100 (RTD), thermocouples, 1 or 2 independent channels 10/11-8.
TF / TF -Ex Field mounted temperature transmitters, PROFIBUS PA, Pt 00 (RTD), thermocouples, or independent channels 0/-8.70 EN P R O F I PROCESS FIELD BUS B U S Input Resistance thermometer (,, -wire
More informationGear Wheel Flowmeter for viscous liquids Model DZR. (option 3):
Gear Wheel Flowmeter for viscous liquids measuring monitoring analysing DZR easuring range:. -... 3-7 l/min Accuracy: ±.3... ±1% of reading p max : bar t max : 1 C Orocess O connection: G ⅛, G ⅜, G ½,
More informationCanadian Imports of Honey
of 0409000029 - Honey, natural, in containers of a weight > 5 kg, nes (Kilogram) Argentina 236,716 663,087 2,160,216 761,990 35.27% 202.09% /0 76,819 212,038 717,834 257,569 35.88% 205.69% /0 United States
More informationVolumetric Flow Measurement
Volumetric Flow Measurement TVF Series Vortex Flow Meter Xi an Tosilon Automation Co., Ltd No.299, Daqing Rd, Lianhu District, Xi'an Shaanxi, China Tel: +86-29-8823 8550 info@tosilon.com; www.tosilon.com
More informationVersaFlow Ultrasonic clamp-on Flowmeter Specifications
VersaFlow Ultrasonic clamp-on Flowmeter Specifications 34-VF-03-14 September 2009 Clamp-On Technology HONEYWELL S VersaFlow stands for continuity and long term reliability. Flow measurement can be done
More informationPro-V Multivariable Flowmeter Model M23 Insertion Vortex
Pro-V Multivariable Flowmeter Model M23 Insertion Vortex Pro-V TM Advantage: Volumetric or mass flow monitoring of most liquids, gases, and steam Multivariable meter delivers mass flow, temperature, pressure,
More informationSITRANS F flowmeters. SITRANS F US System information SITRANS F US Ultrasonic flowmeters 4/181
Siemens AG 09 SITRANS F flowmeters System information Overview Siemens offers two types of ultrasonic flowmeters, in-line flowmeters and clamp-on flowmeters. This offers the end user the maximum flexibility
More informationIts modular concept allows to build up a ball valve for any degree of automation and any kind of application
Type 2657...59 DN 10 up to 80 mm; Plastic; 2/2 way dvantages/benefits ntegrated pilot valves, position feed-back and limit switches available Customized system solutions with Burkert valves & sensors With
More informationFacts and figures MULTILIFT SELECT 15
Facts and figures MULTILIFT SELECT 15 www.arburg.com MULTILIFT SELECT 15 Technical data min. 180 Stroke Z-axis Stroke X-axis Clamping face fixed mould platen 220 E A D K Stroke Y axis 195 B 2) C 157 P
More informationDMP 331Pi. Precision Pressure Transmitter. pressure ports and process connections with flush welded stainless steel diaphragm
DMP Pi Precision Pressure Transmitter pressure ports and process connections with flush welded stainless steel diaphragm accuracy according to IEC 600: 0, % FSO Nominal pressure from 0... 00 mbar up to
More informationISEmax CAM40/CAS40. Technical Information
Technical Information Online measurement of nutrient parameters Ion-selective electrode system for the continuous measurement of ammonium and nitrate Application The ion-selective electrode system works
More informationDMP 331P. Industrial Pressure Transmitter. Process Connections With Flush Welded Stainless Steel Diaphragm
DMP P Industrial Pressure Transmitter Process Connections With Flush Welded Stainless Steel accuracy according to IEC 60770: standard: 0.5 % FSO option: 0.5 % FSO Nominal pressure from 0... 00 mbar up
More informationDMP 331Pi. Precision Pressure Transmitter. pressure ports and process connections with flush welded stainless steel diaphragm
DMP Pi Precision Pressure Transmitter pressure ports and process connections with flush welded stainless steel diaphragm accuracy according to IEC 600: 0, % FSO Nominal pressure from 0... 00 mbar up to
More informationHeat Meter WR 200-D. 2 temperature inputs for Pt100 in four-wire circuit. Exact calculation of the heat parameters according IAPWS-IF97
Heat Meter WR 200-D electronic computing device for heat quantity measuring Heat transfer medium water steam Any flow sensor with analog signal connectable 2 temperature inputs for Pt100 in four-wire circuit
More informationDirect-operated 2/2-way solenoid valves Type EV210B
Data sheet Direct-operated 2/2-way solenoid valves Type EV20B EV20B covers a wide range of direct-operated 2/2-way solenoid valves for universal use. EV20B is a very robust valve program with high performance
More informationGRUNDFOS DATA BOOKLET MAGNA, UPE. Series 2000 circulator pumps
GRUNDFOS DATA BOOKLET MAGNA, UPE Series circulator pumps Contents Product data MAGNA/UPE 3 Duty range 3 Characteristic features 3 Benefits 3 Applications 3 eating systems 3 Pumped liquids 3 Type key Performance
More informationTechnical specification FLUXUS F502TE
Technical specification FLUXUS F502TE Permanently installed and non-invasive ultrasonic flowmeter for the measurement of thermal energy and volumetric flow rate Precise and intelligent energy measuring
More informationPro-V Multivariable Flowmeter Model M23 Insertion Vortex
Pro-V Multivariable Flowmeter Model M23 Insertion Vortex Pro-V TM Advantage: Volumetric or mass flow monitoring of most liquids, gases, and steam Multivariable meter delivers mass flow, temperature, pressure,
More informationGRUNDFOS DATA BOOKLET MAGNA, UPE. Series 2000 circulator pumps
GRUNDFOS DATA BOOKLET Series circulator pumps Contents Product data MAGNA and UPE 3 Duty range 3 Characteristic features 3 Benefits 3 Applications 3 eating systems 3 Pumped liquids 3 Type key Performance
More informationGRUNDFOS DATA BOOKLET MAGNA, UPE. Series 2000 circulator pumps
GRUNDFOS DATA BOOKLET MAGNA, UPE Series circulator pumps Contents Product data MAGNA and UPE 3 Duty range 3 Characteristic features 3 Benefits 3 Applications 3 eating systems 3 Pumped liquids Type key
More informationApplicable Fluid. Accuracy. Flow Rate Range. Fluid Temperature. Fluid Viscosity. Connection Size. Flange Rating. Housing. Material. Blade.
TOKICO TURINE METER GS-F211E- Overview Turbine meter has an excellent performance with its distinctive structure. Comparing with the flow meter using other method, it has compact and lighter design and
More informationclose to performances of the turbine gas meters designed for custody transfer measurements.
QUANTOMETER DESIGN AND FUNCTION CPT Quantometers have been designed in order to provide our customers with reliable and inexpensive measuring instruments for secondary flow measurements. Our great experience
More informationDirect-operated 2/2-way compact solenoid valves Type EV210A
Data sheet Direct-operated 2/2-way compact solenoid valves Type EV210A EV210A covers a wide range of small, directoperated 2/2-way solenoid valves for use in industrial machinery. The compact design together
More informationFlow Measurement. SITRANS F M System information SITRANS F M electromagnetic flowmeters. 4/22 Siemens FI
Function All are based on Faraday s law of induction: U M = B v d k U M = Measured voltage induced in the medium perpendicular to the magnetic field and the flow direction. The voltage is tapped at two
More informationBSH Description. 2. Features. 3. Applications. 4. Pinning information. N-channel enhancement mode field-effect transistor
M3D88 Rev. 25 August 2 Product specification. Description in a plastic package using TrenchMOS technology. Product availability: in SOT23. 2. Features TrenchMOS technology Very fast switching Logic level
More informationTHM80X Series Feature Application. Temperature & Humidity / Flow Measuring Specialist
THM80X Series Feature Application IP67 protection degree, rugged aluminum case, fit in variety harsh environment Capable of temperature compensation Linear adjustment temperature & humidity by computer,
More informationTF 202 / TF 202-Ex. Field mounted temperature transmitters, FOUNDATION Fieldbus (H1), Pt 100 (RTD), thermocouples, 1 or 2 independent channels
TF 0 / TF 0-Ex Field mounted temperature transmitters, FOUNDATION Fieldbus (H), Pt 00 (RTD), thermocouples, or independent channels 0/-8.69 EN USA Input Resistance thermometer (,, wire circuit) Thermocouples
More informationPTF30 Series. Features. Ordering information. G 6 N N F8 (0 to 5 kgf/cm 2 ) PTF30
eries Features Best pressure transmission solution for small and middle sized project such as indoor heating and air conditioning, water and sewer, or incinerator, etc. Improves noise resistance and minimize
More informationDS 300. Electronic Pressure Switch. with IO-Link interface. Stainless Steel Sensor. accuracy according to IEC 60770: 0.35 % FSO.
DS 00 Electronic Pressure Switch with IO-Link interface Stainless Steel Sensor accuracy according to IEC 60770: 0.5 % FSO Nominal pressure from 0... 00 mbar up to 0... 600 bar Digital output signal IO-Link
More informationDMP 331. Industrial Pressure Transmitter for Low Pressure. Stainless Steel Sensor
DMP Industrial Pressure Transmitter for Low Pressure Stainless Steel Sensor accuracy according to IEC 60770: : 0.5 % FSO option: 0.5 / 0. % FSO Nominal pressure from 0... 00 mbar up to 0... 60 bar Output
More informationCyclone Click to go to the page. Cyclone
Cyclone.0311.2 Click to go to the page Cyclone The Cyclone (1,000-8,000 CFM) Suitable for collection of virtually any type of dust Applications: Heavy Moulding Grinding Planing Cabinet Shops Dust Transfer
More informationIntegral V-Cone Flowmeter VD series
TRANSMITTER DIRECTLY MOUNTED! SIMPLE AND EASY INSTALLATION Integral V-Cone Flowmeter VD series OUTLINE VD series Integral V-Cone Flowmeter is a differential pressure type flowmeter. Unlikely to existing
More informationSolenoid valves 2/2-way direct-operated type EV210A
MAKING MODERN LIVING POSSIBLE Data sheet Solenoid valves 2/2-way direct-operated type EV210A EV210A covers a wide range of small, direct-operated 2/2-way solenoid valves for use in industrial machinery.
More informationPressure transmitters for industrial applications MBS 32 and MBS 33
Data sheet Pressure transmitters for industrial applications MBS 32 and MBS 33 The standard pressure transmitters MBS 32 and MBS 33 are designed for use in almost all industrial applications, and offer
More informationGRUNDFOS DATA BOOKLET MAGNA, UPE. Series 2000 circulator pumps
GRUNDFOS DATA BOOKLET Series circulator pumps Contents Product data MAGNA and UPE 3 Duty range 3 Characteristic features 3 Benefits 3 Applications 3 eating systems 3 Pumped liquids Type key 5 Performance
More informationMODEL: CR002, CR003, CR004. ADC12 Display shield. Liquid (Density range: 0.3 to 2.0g/mL) Operating temperature range
Liquid Inexpensive Microflow Coriolis Flowmeter CoriMate Ⅱ MODEL: CR002, CR003, CR004 GENERAL Based on the knowledge and experience on Coriolis technologies OVAL has gained over many years, the Corimate
More informationBy-Pass. This voltage is proportional to the liquid level (threewire potentiometer circuit). The resistance reading can
" " ' " ' / The magnetic field which is in the ball or cylindrical floats actuates very small reed contacts through the wall of a guide tube and these pick up an uninterrupted measuring-circuit voltage
More informationCALYS 120 IS. ATEX documenting multifunction calibrator with high accuracy 0.006%
ATEX documenting multifunction calibrator with high accuracy 0.006% is the top instrument of our range of ATEX documenting multifunction calibrators. Description is the top instrument of our range of ATEX
More informationCALEFFI. Modulating digital controller for heating and cooling. 161 series 01122/08 GB. replaces dp 01122/05
Modulating digital controller for heating and cooling series REGI ERED SI EN ISO 00:000 Cert. n FM UNI EN ISO 00:000 Cert. n 000 0/0 G replaces dp 0/0 Function The controller can be used with three-point
More informationDELTA FLOWMETER FOR GAS SERVICE (INSERTION TYPE) MODEL VF SERIES
DETA FOWMETER FOR GAS SERVICE (INSERTION TYPE) MODE VF SERIES GENERA SPECIFICATION GS. No. GBD0-0-E GENERA In the thermistor type Delta meter, a probe inserted in the pipeline detects the representative
More informationHYGRASGARD KFF KFTF HYGRASGARD AFF AFTF HYGRASGARD AFF - LC AFTF - LC
HYGRASGARD AFF - LC AFTF - LC sensors, calibrateable, with and active passive output Quality product for HVAC sector, accuracy % r.h. The calibrateable room humidity and temperature sensor HYGRASGARD KFF
More informationReflex-Radar BM 100 A
KROHNE 09/2001 7.02317.22.00 GR/PRINTO Reflex-Radar The universal TDR level gauge for liquids, solids and interface Variable area flowmeters Vortex flowmeters Flow controllers Electromagnetic flowmeters
More informationCoriolis Mass Flow Measurement System promass 60 (HART)
Technical Information TI 029D/06/en No. 5006899 Coriolis Mass Flow Measurement System promass 60 (HART) Cost-effective mass flow measurement for liquids and gases Flexible System The system can be customised
More informationPermanently installed and non-invasive ultrasonic flowmeter for the measurement of thermal energy and volumetric flow rate
Technical specification FLUXUS F502TE Permanently installed and non-invasive ultrasonic flowmeter for the measurement of thermal energy and volumetric flow rate Stationary ultrasonic clamp-on system for
More informationCoriolisMaster FCB130, FCB150, FCH130, FCH150 Coriolis mass flowmeter
Data Sheet DS/FCB100/FCH100-EN Rev. D CoriolisMaster FCB130, FCB150, FCH130, FCH150 Coriolis mass flowmeter High precision measurement of mass and volume flow, density, temperature and concentration with
More informationHiPace 1200, with TC 1200, DN 200 CF-F
HiPace 1200, with TC 1200, DN 200 CF-F HiPace 1200, with TC 1200, DN 200 CF-F Similar Image Powerful turbopump with a pumping speed of up to 1250 l/s for N 2 DN 200 CF-F flange for demanding UHV applications
More informationFXP DATASHEET JUNHO 2013
FXP DATASHEET JUNHO 01 Data Sheet D1S0U0 Rev. 0. FXP000 (PARTI-MAG II) Electromagnetic Flowmeter Model DP1F and DPF Measurement made easy For full and partially full pipelines (free and surface pipelines)
More informationVersaFlow Ultrasonic clamp-on Flowmeter Technical Datasheet. Specification. Applications. Highlights. Industries. 34-VF September 2009
VersaFlow Ultrasonic clamp-on Flowmeter Technical Datasheet 34-VF-03-14 September 2009 Specification Clamp-on technology HONEYWELL S VERSAFLOW stands for continuity and long term reliability. Flow measurement
More informationPressure Transmitter ED 701
Pressure Transmitter ED 701 Hygienic and industrial process connections Zero setting function Accuracy < 0.1% FS Quick response time < 25 ms High long-term stability Excellent repeatability Active compensation
More informationCapacitance Level Measurement Electronic Inserts EC 37 Z, EC 47 Z
Technical Information TI 271F/00/en Operating Instructions 017182-1000 Capacitance Level Measurement Electronic Inserts EC 37 Z, EC 47 Z Transmitters for capacitance probes Application The electronic inserts
More informationPAA-ANALYZERS PRINCIPLE OF OPERATION
PAA-ANALYZERS PRINCIPLE OF OPERATION The PAA process analyzer comprises a control unit and an analyzing section which includes the reaction chamber and measuring cell, reagent pumps and associated valves
More informationHiPace 300 with TC 110 and power supply pack OPS 100, DN 100 CF-F
HiPace 300 with TC 110 and power supply pack OPS 100, DN 100 CF-F HiPace 300 with TC 110 and power supply pack OPS 100, DN 100 CF-F Similar Image Rugged, powerful turbopump with a pumping speed of up to
More informationDS 200 DS 200 DRUCK & TEMPERATUR LEITENBERGER GMBH. Electronic Pressure Switch. Stainless Steel Sensor. Electronic Pressure Switch
DRUCK & TEMPERATUR LEITENBERGER GMBH DS 00 Electronic Pressure Switch Stainless Steel Sensor accuracy according to IEC 00: standard: 0. % FSO option: 0. % FSO Nominal pressure: from 0... 00 mbar up to
More informationCALYS 80P. Field pressure calibrator with high accuracy 0.01%
Field pressure calibrator with high accuracy 0.01% CALYS 80P is a field pressure calibrator within CALYS range with high accuracy of 0.01%. Description CALYS 80P is a field pressure calibrator within CALYS
More informationCoriolisMaster FCB430, FCB450, FCH430, FCH450 Coriolis mass flowmeter
Data Sheet DS/FCB400/FCH400-EN Rev. D CoriolisMaster FCB430, FCB450, FCH430, FCH450 Coriolis mass flowmeter High precision measurement of mass and volume flow, density, temperature and concentration with
More informationPRODUCT CATALOGUE PREssURE switch
PRODUCT CATALOGUE PREssURE switch w w w. p a a b. c o m Pressure AT THe HIGHesT LeVeL Successful mediumsized companies are not successful because they are active in many areas, but rather because they
More informationSolenoid valves Type EVR 2 g 40 NC/ NO New EVR 6 with Steel cover design
MAKING MODERN LIVING POSSIBLE Technical brochure Solenoid valves EVR 2 g 40 NC/ NO New with Steel cover design EVR is a direct or servo operated solenoid valve for liquid, suction, and hot gas lines with
More informationSINEAX VK 636 Programmable Temperature Transmitter with PROFIBUS-PA protocol
with PROFIBUS-PA procol for installation in the terminal head of a temperature sensor DIN 79, shape B Application 00 II () G FISCO The SINEAX VK 66 is a temperature transducer for use in aumation systems
More informationtekcor TKC Multivariable Coriolis Flow Sensors
Sensing the pulse of industry Technical Data Sheet & General Specifications tekcor TKC Multivariable Coriolis Flow Sensors The tekcor TKC Series principle was initiated by tekflo sensors engineers, who
More informationS95.3U01 / S95.3U11. Head mounted temperature transmitters, programmable, Pt 100 (RTD), thermocouples, electrical isolation
S95.U0 / S95.U Head mounted temperature transmitters, programmable, Pt 00 (RTD), thermocouples, electrical isolation Resistance thermometer (,, wire circuit) Thermocouples Resistance remote signalling
More informationPHT4NQ10LT. 1. Description. 2. Features. 3. Applications. 4. Pinning information. N-channel enhancement mode field-effect transistor
MD87 Rev. September Product specification. Description in a plastic package using TrenchMOS technology. Product availability: in SOT.. Features TrenchMOS technology Fast switching Low on-state resistance
More informationFitting PS P - B - R2
INTERNAL STRUCTURE PS86 (G 3/8) 2 3 FEATURES Solidstate pressure switch with no mechanical moving parts Isolated output via a photo-coupler Switching pressure adjustable and hysteresis band adjustable
More informationSVH with HYDROVAR (HVL range)
SVH with HYDROVAR (HVL range) SVH-HVL Rev. B Ed.9/1 (EN) SVH SERIES (e-sv WITH HYDROVAR) Background and context For all pumping needs in commercial or residential building and in industry applications,
More informationBilateral Labour Agreements, 2004
Guest Austria Canada Turkey ( 64) Canada, Czech Republic, Hungary ( 98), Belgium Italy ( 46, 54), Turkey ( 64) Bulgaria ( 99), Pol (02) Germany ( 91) Bulgaria ( 99), Mongolia ( 99), Pol ( 92), Russia (
More informationWorking pressure. 420 bar. Style T. Style B. Style D Style V Style Z. Style S B B D.I. D.I. D.I. D.I. D.I. D.I.
5 FP SRIS FP Working pressure 40 bar Style S Style Style T Style D Style V Style Z 5 Technical data Filter body (Materials) ead: Cast iron (chemical heat treatment) ousing: Steel (chemical heat treatment)
More informationCooling Water Flow Regulator
Temperature controlled, indirectly and directly actuated diaphragm valves / seat valves Port size G 3/8 to G DN 0 to DN 0 ontinuous regulation High accuracy No auxiliary energy required Technical data
More information04 June Dim A W V Total. Total Laser Met
4 June 218 Member State State as on 4 June 218 Acronyms are listed in the last page of this document. AUV Mass and Related Quantities Length PR T TF EM Mass Dens Pres F Torq Visc H Grav FF Dim A W V Total
More informationREADY TO SCRAP: HOW MANY VESSELS AT DEMOLITION VALUE?
READY TO SCRAP: HOW MANY VESSELS AT DEMOLITION VALUE? August 206 VesselsValue Global number of vessels at demolition value At scrap value 7,27 6 Above scrap value,8 84 Number of vessels at demolition value
More informationIBV Series C Carbon Steel Inverted Bucket Vertical Steam Trap
Local regulations may restrict the use of this product to below the conditions quoted. In the interests of development and improvement of the product, we reserve the right to change the specification without
More informationFLOW SENSOR FOR LIQUID MEDIA Serie 210
FLOW SENSOR FOR LIQUID MEDIA Serie 0 Flow range 0.... 0 l/min Nominal diameters DN 6 / 8 / 0 / / 0 / Temperature measurement -0... + ÀC In comparison to the OEM flow sensor (type 00), the type 0 is available
More informationIBV Series Z Alloy Steel Inverted Bucket Vertical Steam Trap
Local regulations may restrict the use of this product to below the conditions quoted. In the interests of development and improvement of the product, we reserve the right to change the specification without
More informationVK2 volume flow controller Overview
New option with sound attenuator Maintenance-free VK2 volume flow controller that operates without an auxiliary power supply, for ventilation and air conditioning systems. Adjustable on site. Outstanding
More informationField mounted Temperature Transmitter TF212/TF212-Ex
Data Sheet Field mounted Temperature Transmitter TF/TF-Ex PROFIBUS PA, Pt 00 (RTD), Thermocouples, or independent channels Input Resistance thermometer (,, -wire circuit) Thermocouples Resistance remote
More information