1 CVS Calibration System Short Instruction Manual System Configuration Description of the Equipment State: 11/21/2005

2 Table of contents 1. Documentation and Appropriate Usage 2. EC Conformity Declaration 3. Technical Data Sheet 4. Device Setup 5. Basics of Operation 6. Measurement Ranges and Program Change 7. Integrated Sensors and Flow-Elements 8. Configuration of the Controller S Electrical Wiring Scheme 10. Measurement Uncertainty

3 1. Documentation and Appropriate Usage Structure of the Documentation: Partl Denomination Content A _man_e Short Instruction Manual for CVS Calibration System: system configuration and device description. B LMF_man_e Reference Manual for LMF LAMINARMASTER Flow-System LMF: parameter systematics, basics of the measurement procedure. C References and Denomination see _man_e Chapter Sensors and Flow- Elements in this manual Documentation and Manuals for the used sensors and flow elements D CVS_appendix_e Device Specific Documentation for CVS- KAL: sensor and flow element data, calibration and test certificates Appropriate Usage: The CVS Calibration System CVS_KAL is member of the system family of the series Laminar Master Flow LMF. The instruments of the series LAMINARMASTER are exclusively made for measurement and control of air volume flow and mass-flow and the correlating sensor values. Used as measuring instrument in complex devices, in a machine compound, in a production line or production station, the signal outputs may be used only to inform a superior control system, e.g. a PLC. Used as a stand-alone laboratory measuring device with control function, pay attention to the rules and hints for emergency shut-down and return of the power supply after a power failure. For basic security guidelines see the reference manual LMF LaminarMasterFlow. Other applications than mentioned above are not appropriate. Damages caused by impropriate usage TetraTec Instruments GmbH is not liable for. Appropriate usage also implies: to follow all instructions from the operating manual. to do the inspection and maintenance work.

4 2. EC Conformity Declaration Herewith we declare that the devices delivered by us, concerning design and construction as well as the model put into circulation, comply with the EC-Directives listed below. This declaration is no longer valid, if the delivered device is modified without our agreement. Appropriate EC-Directives: EC-Machinary Directive (98 / 37 / EC)* EC-Low Voltage Directive (73 / 23 / EEC)* EC-Electromagnetic Compatibility Directive (89 / 336 / EEC) Applied harmonized standards, especially: DIN EN ISO , Basic Concepts* DIN EN 294, DIN EN 811 Safety Distances* DIN EN 983 Fluid Power Systems - Pneumatics* DIN EN (VDE ) Safety Requirements* DIN EN (VDE ) Emission Interference Immunity DIN EN (VDE ) Interference Immunity Applied national standards and technical specifications, especiallly: BGV A1 Principles of Prevention* BGV A2 Electrical Plants and Equipment * BGV B6 Gases* (German) Pressure Vessels Ordinance* *as appropriate with the particular design of the device. Place, Date: Steinenbronn, Signature: W. Höhn, Manager of the Company

5 3. Technical Data Sheet Power Supply: Gas Type: Flow range of measurement section: Ambient conditions during operation: 230VAC, 50/60Hz Air: dust-free (5 µ m filtered), oil-free, non-condensing Standard volume See flow:* Docu Part D Actual volume flow: See Docu Part D Temperature: C Pressure: atmospheric pressure Humidity: %, noncondensing Operating conditions for measurement section: Absolute pressure: Differential pressure: Temperature: 0,8-1,2 bar abs mbar C Overload strength: * Standard operations: Pressure: 2 bar abs. Temperature: C pressure: 1013,25 mbar Temperature: 0 C Rel. humidity: 0 %

6 4. Device Setup Front View: Rear View Flow Element

7 5. Basics of Operation Switch-On Press POWER key. The POWER key is illuminated. The LMF starts with a self test. Measuring There are two modes of measurement. After power-on the device is in continuous measuring mode. In this mode the current measurement values are regularly displayed. In addition there is an averaging measurement mode. This is composed of two phases controlled by the keys START and STOP. In the first phase data is collected and written to a buffer. In the second phase the mean values are displayed. Press START key. Starts an averaging measurement. Data collection takes the time defined in parameter Px701 but can be terminated sooner pressing the STOP key. Afterwards the results (mean values) are displayed. Press the F1 key repeatedly to browse the results. The results are stepped through with each keypress. The last page is followed by the first page again (infinite loop). Press STOP key. 1. During an averaging measurement: stops the averaging measurement. The device displays the mean values. 2. If the results of an averaging measurement are displayed: the device returns to the continuous measuring mode. Note: The results are only valid, if a sufficient leak tightness of the system is ensured regularly and the zero adjustment of the gauge and differential pressure sensors is performed regularly. To do this follow the instructions described below.

8 Performing the Zero Adjustment of the Pressure Sensors The zero point of differential and gauge pressure sensors changes with time. Especially after big pressure changes there can be an offset. The zero adjustment is applied to all analogue and serial sensor inputs, that are indicated as zero adjustable with the parameters S2x32 and S3x32. (value 0: zero adjustment not applicable, value 1: zero adjustment applicable, value 2: zero adjustment applicable but results in an error message, if the difference exceeds 5%) Note: The Operator is responsible to ensure the pressure balance. If the pressure sensors are zero adjusted with pressure differences applied, faulty measurements are the consequence! Press Zero key until it is illuminated. As soon as the key is no longer illuminated, the zero adjustment has finished. If pressure sensors have to be adjusted, that are cannot be adjusted with the Zero key (e. g. gauge pressure sensors), this can be done using the test menu, provided they are displayed there (see LMF operating manual). Pay attention to the necessary pressure balance with the ambient air. Absolute pressure sensors, temperature sensors and sensors for relative humidity must not be zero adjusted and cannot be zero adjusted. Switch-Off Press POWER key.

9 6. Measurement Ranges and Program Change The device supports depending on the firmware up to 3 measurement sections at the same time. The measurement sections are numbered from 0 to 2. To the measurement sections configurations can be assigned which are stored in 10 available programs. The programs are numbered from 0 to 9. In the following special settings for the device were described. Assignment of the Programs to the Measurement Sections Number of available measurement sections: 1 The programs are assigned as follows to the measurement sections and configured /assigned: Section -Number Program -Number Configuration / Assignment 0 0 Assignement see Device Specific Appendix _appendix_e 0 1 Assignement see Device Specific Appendix _appendix_e 0 2 Assignement see Device Specific Appendix _appendix_e 0 3 Assignement see Device Specific Appendix _appendix_e 0 4 Assignement see Device Specific Appendix _appendix_e 0 5 Assignement see Device Specific Appendix _appendix_e 0 6 Assignement see Device Specific Appendix _appendix_e 0 7 Assignement see Device Specific Appendix _appendix_e 0 8 Assignement see Device Specific Appendix _appendix_e 0 9 Assignement see Device Specific Appendix _appendix_e Switching of Programs: The program switching can be done via front panel key pad as follows: - Hold F2 function key for 3 seconds - The measurement section and program menu appears: Upper line: Maximum possible program number for the measurement section Middle line: Active Program number for the measurement section Lower line: Minimum possible program number for the measurement section - Upon need toggle by pressing the F1 function key to reach the next section - Press right arrow -key to increase the program number - Press left arrow -key to lower the program number - Hold F2 function key for 3 seconds to store a program change - Hold F1 and F3 function key for 3 seconds to leave the menu without any changes. The same can be achieved by pressing the Stop button if available.

10 7. Integrated Sensors and Flow-Elements Sensors: Type: Differential Pressure- Connection Type: Sensor Article-No.: 1500-DN0020 Measurement Range: mbar Signal: RS485 Sensor Data Set No.: 0 Display-Name: Pdiff Plug connector name: RS485-Bus Device Documentation: 1500_man_d Type: Absolute Pressure- Connection Type: Sensor Article-No.: 1500-AI0900 Measurement Range: mbar Signal: RS485 Sensor Data Set No.: 1 Display-Name: Pabs Plug connector name: RS485-Bus Device Documentation: 1500_man_d Type: Temperature-Sensor Connection Type: Article-No.: WIT-PT G Connection S320 (Slot): 0 Measurement Range: C Connection Port: 0 Signal: Pt100 Sensor Data Set No.: 2 Display-Name: Temp Plug connector name: X10 Device Documentation: WIT-PT-11_man_d Type: Relativ Humidity Connection Type: Sensor Article-No.: HUM-50U-MS Connection S320 (Slot): 0 Measurement Range: % Connection Port: 1 Signal: 0...1V Sensor Data Set No.: 3 Display-Name: Hum Plug connector name: X11 Device Documentation: HUM_man_d Flow Elements Example of a Standard Flow-Element: Type: Laminar Flow Element Connection Type: (LFE) Article-No.: 50MC02-06-F Connection Pdiff: ¼ NPT Measurement Range: L/min Connection Flow: 6 Pdiff: mbar Flow Data Set No.: 0 Device Documentation: 50M_operation_man_e Available Flow-Elements and Program Assignement: See Device Specific Appendix _appendix_e

11 8. Configuration of the Controller S320 General description: Supply: VAC 50/60Hz 24V DC 12V DC Cabinet Design: 19 rack 280mm NT table 19 rack 380mm NT table IP54 cabinet IP54 compact S320 OEM, encod. connector Special cabinet: 84TE/3HE/280mm depth Connector Options: RS link in front PLC connection: 40-pin Special connector Cable length: 2,5m Software: Software: Flow Version: OS-version: s320a501 Documentation: english Standard evaluations: LFE calculation classical LFE calculation universal ACCUTUBE calculation ORIFICE calculation Gas counter evaluation Sonic nozzle evaluation Mass flow sensor evaluation CONTROLLER S320: S/No: see appendix 512 kb RAM 2048 kb RAM SLOT 0: type: 100 port0: Temp0 supply: 1mA range: Pt100 disp. span: C port 1: Hum0 supply: 24V range: 0..1 V disp. span: 0 100% SLOT 1: type: port0: supply: range: disp. span: port 1: supply: range: disp. span: Special features: see section "options" Integrated Digital Inputs Used SLOT 2: type: port0: supply: range: disp. span: port1: supply: range: disp. span: SLOT 3: type: port0: supply: range: disp. span: port 1: supply: range: disp. span: SLOT 4: type: port0: supply: range: disp. span: port 1: supply: range: disp. span Yes Options: Single section Double section Triple section Leak check Test pressure limit control Pressure control Flow control PLC-mode No Integrated Digital Outputs Used Yes No Assignment of integrated digital signals see section 0. Digital Expansion Modules Yes No If yes: Internal supply (pin 24V not connected)

12 Integrated Digital Interface, Push Button Assignement Digital Input ID Function Remark DI0 Start Key 0, Label START DI1 Stop Key 1, Label STOP DI2 Leak Test Key 2, Label LEAK TEST DI3 DI4 DI5 DI6 DI7 Digital Output Name Function Remark DO0 Start LED Key 0, Label START DO1 Stop LED Key 1, Label STOP DO2 Leak Test LED Key 2, Label LEAK TEST DO3 DO4 DO5 DO6 DO7

13 :25:13 f=0.66 T:\Projekte\EagleProjektPlaene41\.sch (Sheet: 1/4)

14 :25:13 f=0.66 T:\Projekte\EagleProjektPlaene41\.sch (Sheet: 2/4)

15 :25:13 f=0.66 T:\Projekte\EagleProjektPlaene41\.sch (Sheet: 3/4)

16 :25:13 f=0.66 T:\Projekte\EagleProjektPlaene41\.sch (Sheet: 4/4)

17 10. Measurement Uncertainty Basic considerations Qv, Qm, ρ(p, T, xv) The determination of the actual volume flow Q v of the unit under test (uut) is generally done by measurement of the actual volume flow of the calibration-standard (master) and conversion with the density ratio (density ρ) to the conditions of the unit under test. Q v,uut = Q v,master * ρ Master / ρ uut The measured quantity mass flow (Q m ) is calculated as the product of the actual volume flow and the density and has the same value at each point of the measuring system. Q m,uut = Q m,master = Q v,master * ρ Master The effect of error propagation of the relative uncertainty of measurement from the different measured quantities is determined according to ISO/TR 5168 as the standard deviation. u ges = u, std i 2 i The extended uncertainty of measurement u ges, resulting from the relative standard uncertainty of measurement u ges,std by multiplication with the extension factor k = 2, corresponds to the interval in which the measurement value lies with a probability of 95%. The smallest assignable extended uncertainty of measurement of the reference measurement (calibration) is identical with this extended standard uncertainty of measurement. In the standard uncertainty of measurement of the unit under test, another contribution has to be taken into account which describes the spread of the unit under test resp. the calibration results. Decisive point for the uncertainty of measurement of the reference measurement is the uncertainty of the determination of the actual volume flow through the calibration standard. Furthermore there s an uncertainty when determining the density ratio between master and the unit under test (for quantity actual volume flow), resp. in the determination of the density of the calibration standard (for quantity mass flow) from the measurement quantities like humidity, pressure and temperature of the calibration standard resp. the unit under test. Measurement uncertainty caused by leakages in the measurement section In the run-up of any reference measurement, one has to make sure by a leak test of the system (pressure-drop method) that the maximum error by leaks will stay below a fixed value. Let the volume of the measurement section be V, the operating pressure for leak test p and the smallest flow to be calibrated Q min, so the maximum pressure drop allowed in the measurement section for an uncertainty u L is dp/dt u L. Q min. p / V

18 Uncertainty of measurement during calibration with Laminar Flow Elements: The extended standard uncertainty of measurement of the reference standards is fixed by the calibration within a continuous measurement chain related to the PTB Physikalisch- Technische Bundesanstalt. The calculation of the actual volume flow at the unit under test with reference measurement by Laminar Flow Elements is done according to the following relation (Hagen-Poiseuille Equation and conservation of mass / continuity equation): Q vol,uut = Q cal,lfe (dp). η cal /η uut. ρ LFE /ρ uut The uncertainty of measurement for calibration with Laminar Flow Elements consists of the following factors: uncertainty of measurement u cal of the reference standard calibration, typically u Kal = 0,325%v.M. (Half of the extended measurement uncertainty of typically 0,65%) uncertainty of measurement u dp for measurement of the differential pressure at the LFE For measurement of the differential pressure of the LFE, at works-calibration as well as at external reference measurement, the same differential pressure sensor is used, so that not the absolute precision of the sensor is relevant but only the repeatability of the measurement values. In addition the thermal and long-time drift of the sensor has to be taken into account. Typical values in the dp-interval 2 25 hpa: relative measurement uncertainty u dp = 0,15%v.M. thermal uncertainty: u L = 0,02% v.m./ C. zero point of the sensor: u N = 0,05% v.e. uncertainty of measurement u η for the viscosity ratio by calculation from calibration conditions to actual conditions during reference measurement, typically u η = 0,056% uncertainty of measurement u ρ for the density ratio. Mainly the accuracy of the measurement of temperature and pressure as well as for air the humidity is significant for the conversion from conditions of the calibration standard to conditions of the unit under test, typically u ρ = 0,14% for mass flow u ρ = 0,12% for volume flow uncertainty of measurement u LFE for reference measurement with Laminar Flow Elements. This part of uncertainty includes the standard deviation of the calibration points in regard to polynomial linearisation, as well as an estimation of the short- and long-time drift behaviour between reference measurements. The value in first instance is fixed and will later be adjusted by historical data. u LFE = 0,15% For the extended uncertainty of measurement it follows: u ges = 2. ( u Kal 2 + u dp 2 + u η 2 + u ρ 2 + u L 2 + u LFE 2 ) 1/ u N For the example volume flow follows: u ges = 2. ( 0, , , , , ,15 2 ) 1/2 = 0,82%v.M. + 0,1% v.e. and finally the worst case for the mass flow of humid air is: u ges = 2. ( 0, , , , , ,15 2 ) 1/2 = 0,84%v.M. + 0,1% v.e.

19 Uncertainty of measurement during calibration with Orifices: The extended standard uncertainty of measurement of the reference standards is fixed by the calibration within a continuous measurement chain related to the PTB Physikalisch- Technische Bundesanstalt. The calculation of the actual volume flow at the unit under test with reference measurement by Orifices is done according to the following relation (Bernoulli Equation and conservation of mass / continuity equation): Q vol,uut = (dp. ρ uut ) 0,5. C cal (Re). /ρ uut The uncertainty of measurement for calibration with orifices consists of the following factors: uncertainty of measurement u cal of the reference standard calibration, typically u Cal = 0,325%v.M. (Half of the extended measurement uncertainty of typically 0,65%) uncertainty of measurement u dp for measurement of the differential pressure at the orifices. For measurement of the differential pressure of the orifices, at works-calibration as well as at external reference measurement, the same differential pressure sensor is used, so that not the absolute precision of the sensor is relevant but only the repeatability of the measurement values. In addition the thermal and long-time drift of the sensor has to be taken into account. Typical values in the dp-interval 2 25 hpa: relative measurement uncertainty u dp = 0,15%v.M. thermal uncertainty: u L = 0,02% v.m./ C. zero point of the sensor: u N = 0,05% v.e. uncertainty of measurement u Re for the Reynolds number influence by calculation of the flow coefficient C cal (Re), typically u Re = 0,06% uncertainty of measurement u ρ for the density ratio. Mainly the accuracy of the measurement of temperature and pressure as well as for air the humidity is significant for the conversion from conditions of the calibration standard to conditions of the unit under test, typically u ρ = 0,14% for mass and volume flow uncertainty of measurement u OR for reference measurement with orifices. This part of uncertainty includes the standard deviation of the calibration points in regard to polynomial linearisation, as well as an estimation of the short- and long-time drift behaviour between reference measurements. The value in first instance is fixed and will later be adjusted by historical data. u OR = 0,15% For the extended uncertainty of measurement it follows: u ges = 2. ( u Kal 2 + 0,5. u dp 2 + u Re 2 +0,5. u ρ 2 + u L 2 + u OR 2 ) 1/ u N For the mass and volume flow follows: u ges = 2. ( 0, ,5 = 0,76%v.M. + 0,1% v.e. 0, , ,5. 0, , ,15 2 ) 1/2

20 Uncertainty of measurement during calibration with critical nozzles: The extended standard uncertainty of measurement of the reference standards is fixed by the calibration within a continuous measurement chain related to the PTB Physikalisch- Technische Bundesanstalt. The calculation of the actual volume flow at the unit under test with reference measurement by critical nozzles / critical flow orifices (CFO) is done according to the following relation (equation for the velocity of sound and conservation of mass / continuity equation): Q vol,uut = Q vol,cfo. ρ CFO /ρ uut = F(c(T)). ρ CFO /ρ Prüfling The uncertainty of measurement for calibration with critical nozzles consists of the following factors: uncertainty of measurement u cal of the reference standard calibration, typically u Kal = 0,325%v.M. (Half of the extended measurement uncertainty of typically 0,65%) uncertainty of measurement u c for the dependence of the sonic velocity on temperature typically u c = 0,06% uncertainty of measurement u ρ of density ratio. Mainly the accuracies of pressure and temperature measurement, and furthermore for air the humidity is significant for conversion from conditions at the reference standard to conditions at the unit under test, typically u ρ = 0,14% for mass-flow u ρ = 0,12% for volume-flow uncertainty of measurement u CFO for reference measurement with Laminar Flow Elements. This part of uncertainty includes the standard deviation of the calibration points in regard to polynomial linearisation, as well as an estimation of the short- and long-time drift behaviour between reference measurements. The value in first instance is fixed and will later be adjusted by historical data. u CFO = 0,15% For the extended measurement uncertainty for an orifice follows: u ges = 2. ( u Kal 2 + u c 2 + u ρ 2 + u CFO 2 ) 1/2 For the example volume flow follows: u ges = 2. ( 0, , , ,15 2 ) 1/2 = 0,77%v.M. and finally the worst case for the mass flow of humid air is: u ges = 2. ( 0, , , ,15 2 ) 1/2 = 0,78%v.M.

PMU-DLS Calibration System PMU-DLS-KAL. Short Instruction Manual. System Configuration Description of the Equipment

PMU-DLS Calibration System PMU-DLS-KAL. Short Instruction Manual. System Configuration Description of the Equipment PMU-DLS Calibration System Short Instruction Manual System Configuration Description of the Equipment State: 04/21/2011 Table of contents 1. Documentation and Appropriate Usage 2. EC Conformity Declaration

More information

DS 300. Electronic Pressure Switch. with IO-Link interface. Stainless Steel Sensor. accuracy according to IEC 60770: 0.35 % FSO.

DS 300. Electronic Pressure Switch. with IO-Link interface. Stainless Steel Sensor. accuracy according to IEC 60770: 0.35 % FSO. DS 00 Electronic Pressure Switch with IO-Link interface Stainless Steel Sensor accuracy according to IEC 60770: 0.5 % FSO Nominal pressure from 0... 00 mbar up to 0... 600 bar Digital output signal IO-Link

More information

ED 701 General Industry Pressure Transmitter

ED 701 General Industry Pressure Transmitter ED 701 General Industry Pressure Transmitter Standard industrial process connections Complete range of electrical connections 4... 20 ma and Voltage outputs Accuracy: 0.1%, 0.2% and 0.4% FS Quick response

More information

Operating Instructions for Digital Manometer. Model: MAN-SD

Operating Instructions for Digital Manometer. Model: MAN-SD Operating Instructions for Digital Manometer Model: MAN-SD 1. Contents 1. Contents...2 2. Note...3 3. Instrument Inspection...3 4. Regulation Use...4 5. Operating Principle...4 6. Mechanical Connection...5

More information

ISOCON-6. 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER

ISOCON-6. 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER ISOCON-6 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER Whilst every effort has been taken to ensure the accuracy of this document, we accept no responsibility for damage, injury, loss or expense resulting

More information

CALYS 80P. Field pressure calibrator with high accuracy 0.01%

CALYS 80P. Field pressure calibrator with high accuracy 0.01% Field pressure calibrator with high accuracy 0.01% CALYS 80P is a field pressure calibrator within CALYS range with high accuracy of 0.01%. Description CALYS 80P is a field pressure calibrator within CALYS

More information

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure DATASHEET Pascal ET Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure - electrical signals (ma, mv, V, ) - temperature (TC and

More information

FLT5A FLT5A. abc. abc. Thermometer User Guide. Thermometer User Guide. Internet:

FLT5A FLT5A. abc. abc. Thermometer User Guide. Thermometer User Guide. Internet: abc instruments international FLT5A Thermometer Land Instruments International Dronfield, S18 1DJ England Telephone: (1246) 417691 Facsimile: (1246) 41585 Email: infrared.sales@landinst.com Internet: www.landinst.com

More information

Vet Photometer DP 700 Operating Manual. Version V5.9

Vet Photometer DP 700 Operating Manual. Version V5.9 1 Vet Photometer DP 700 Operating Manual Version V5.9 2 Dear customer, Thank you for selecting Diaglobal s Vet Photometer and for the confidence you have placed in us. The Vet Photometer belongs to a new

More information

THM80X Series Feature Application. Temperature & Humidity / Flow Measuring Specialist

THM80X Series  Feature Application. Temperature & Humidity / Flow Measuring Specialist THM80X Series Feature Application IP67 protection degree, rugged aluminum case, fit in variety harsh environment Capable of temperature compensation Linear adjustment temperature & humidity by computer,

More information

E2T0 DVP

E2T0 DVP 29-5-26 5116868-E2T DVP-11363-1 ENGLISH Thank you for choosing Delta s DVP series PLC. DVP4TC-E2 temperature measurement module receives 4 points of external thermocouple temperature sensors (J-type, K-type,

More information

Pressure Transmitter ED 701

Pressure Transmitter ED 701 Pressure Transmitter ED 701 Hygienic and industrial process connections Zero setting function Accuracy < 0.1% FS Quick response time < 25 ms High long-term stability Excellent repeatability Active compensation

More information

User Manual VX4750 Function Generator Module

User Manual VX4750 Function Generator Module User Manual VX4750 Function Generator Module 070-9151-05 This document applies for firmware version 2.60 and above. Copyright Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S.

More information

Acceleration/Velocity/Displacement VIBRATION METER

Acceleration/Velocity/Displacement VIBRATION METER Acceleration/Velocity/Displacement VIBRATION METER Model : VB-8220 Your purchase of this VIBRATION METER marks a step forward for you into the field of precision measurement. Although this METER is a complex

More information

Acceleration, Velocity, Separate probe VIBRATION METER Model : VB-8202

Acceleration, Velocity, Separate probe VIBRATION METER Model : VB-8202 Acceleration, Velocity, Separate probe VIBRATION METER Model : VB-8202 Your purchase of this VIBRATION METER marks a step forward for you into the field of precision measurement. Although this METER is

More information

v = m / s Thermal Flow Sensors TA with thin-film sensor element Probes Probe Guides Measuring Tubes

v = m / s Thermal Flow Sensors TA with thin-film sensor element Probes Probe Guides Measuring Tubes Thermal Flow Sensors TA with thin-film sensor element Probes Probe Guides Measuring Tubes Examples of Application Flow measurements also in hazardous locations X zone 0 or zone 1 in clean rooms in air

More information

FATEK. FBs-1HLC. Precision Load Cell Module. Operation manual V1.1 06/07/2017 FATEK AUTOMATION CORP.

FATEK. FBs-1HLC. Precision Load Cell Module. Operation manual V1.1 06/07/2017 FATEK AUTOMATION CORP. FATEK FBs-1HLC Precision Load Cell Module Operation manual V1.1 06/07/2017 FATEK AUTOMATION CORP. 1 Contents Chapter 1 1HLC module introduction... 3 1.1 Module specification... 3 1.2 Module appearance

More information

473-SHX Dew Point Mirror

473-SHX Dew Point Mirror Humidity and Temperature Reference Hygrometer For Temperature up to 125 C Precise and stable chilled mirror dew point mirror technology High temperature optical components High temperature sample fan Cable

More information

ISOCON-3 MAINS POWERED ISOLATING SIGNAL CONVERTER

ISOCON-3 MAINS POWERED ISOLATING SIGNAL CONVERTER ISOCON-3 MAINS POWERED ISOLATING SIGNAL CONVERTER CAUTION: This equipment is designed for connection to mains voltages and must be used in accordance with this guide. If it is not, the safety protection

More information

CALYS 120 IS. ATEX documenting multifunction calibrator with high accuracy 0.006%

CALYS 120 IS. ATEX documenting multifunction calibrator with high accuracy 0.006% ATEX documenting multifunction calibrator with high accuracy 0.006% is the top instrument of our range of ATEX documenting multifunction calibrators. Description is the top instrument of our range of ATEX

More information

DMP 331. Industrial Pressure Transmitter for Low Pressure. Stainless Steel Sensor

DMP 331. Industrial Pressure Transmitter for Low Pressure. Stainless Steel Sensor DMP Industrial Pressure Transmitter for Low Pressure Stainless Steel Sensor accuracy according to IEC 60770: : 0.5 % FSO option: 0.5 / 0. % FSO Nominal pressure from 0... 00 mbar up to 0... 60 bar Output

More information

OPERATION MANUAL. 3 in 1, IR, Type K/J/R/E/T, Pt 100 ohm THERMOMETER. Model : TM-2000

OPERATION MANUAL. 3 in 1, IR, Type K/J/R/E/T, Pt 100 ohm THERMOMETER. Model : TM-2000 3 in 1, IR, Type K/J/R/E/T, Pt 100 ohm THERMOMETER Model : TM-2000 Your purchase of this THERMOMETER marks a step forward for you into the field of precision measurement. Although this THERMOMETER is a

More information

Field IT 50T Series Transmitters

Field IT 50T Series Transmitters Data Sheet SS/G/A_ Field IT 0T Series Transmitters gauge/absolute pressure transmitter Ranges: -00 to 0000kPa - to 00bar -. to 6000psi Base accuracy : 0.% (BFSL) Chemical Vapour Deposition (CVD) technology

More information

v = 00,20 m/s Thermal Flow Sensors TA with thin-film sensor element Probes Probe guides Measuring tubes

v = 00,20 m/s Thermal Flow Sensors TA with thin-film sensor element Probes Probe guides Measuring tubes with thin-film sensor element Probes Probe guides Measuring tubes Examples of Application Flow measurements also in hazardous locations Ex zone 0 or zone 1 in clean rooms in air conditioning and ventilating

More information

Certificate: / 29 April 2014

Certificate: / 29 April 2014 Test report: 936/21218896/A of 20 September 2013 Initial certification: 01 April 2014 Date of expiry: 31 March 2019 Publication: BAnz AT 01 April 2014 B12, chapter IV, No. 5.1 Approved application The

More information

Air velocity & IAQ measuring instruments

Air velocity & IAQ measuring instruments Data sheet dp Air velocity & IAQ measuring instruments - air velocity & IAQ measuring instrument - dp air velocity & IAQ measuring instrument incl. differential pressure C CO Intuitive: clearly structured

More information

SINEAX VK 636 Programmable Temperature Transmitter with PROFIBUS-PA protocol

SINEAX VK 636 Programmable Temperature Transmitter with PROFIBUS-PA protocol with PROFIBUS-PA procol for installation in the terminal head of a temperature sensor DIN 79, shape B Application 00 II () G FISCO The SINEAX VK 66 is a temperature transducer for use in aumation systems

More information

Certificate: / 30. April 2015

Certificate: / 30. April 2015 Test report: 936/21226418/A of 29 September 2014 Initial certification: 2. April 2015 Date of expiry: 1. April 2020 Publication: BAnz AT 02 April 2015 B5, chapter III number 3.1 Approved application The

More information

Turbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas

Turbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas Turbine Meter TRZ 0 PRODUCT INFORMATION Reliable Measurement of Gas TURBINE METER TRZ 0 Method of operation, Construction Method of operation The TRZ 0 turbine meter is a flow meter suitable for gas measurement

More information

DMP 331Pi. Precision Pressure Transmitter. pressure ports and process connections with flush welded stainless steel diaphragm

DMP 331Pi. Precision Pressure Transmitter. pressure ports and process connections with flush welded stainless steel diaphragm DMP Pi Precision Pressure Transmitter pressure ports and process connections with flush welded stainless steel diaphragm accuracy according to IEC 600: 0, % FSO Nominal pressure from 0... 00 mbar up to

More information

compact ejector, Series ECS with release valve with silencer with non-return valve, With filter vacuum switch: electronic 1x PNP and 1x analog

compact ejector, Series ECS with release valve with silencer with non-return valve, With filter vacuum switch: electronic 1x PNP and 1x analog with release valve with silencer with non-return valve, With filter vacuum switch: electronic x PNP and x analog Ambient temperature min./max. +0 C / +0 C Medium temperature min./max. +0 C / + C Working

More information

4/4-way servo solenoid directional control valves, directly operated, with electrical position feedback and on-board electronics (OBE)

4/4-way servo solenoid directional control valves, directly operated, with electrical position feedback and on-board electronics (OBE) 4/4-way servo solenoid directional control valves, directly operated, with electrical position feedback and on-board electronics (OBE) RE 9035/. Replaces: 05. 1/1 Type 4WRPEH 6 Size 6 Unit series X Maximum

More information

Data Sheet / Instructions. Dewpoint MeterO

Data Sheet / Instructions. Dewpoint MeterO Data Sheet / Instructions Dewpoint MeterO Dewpoint Meter ISO 8502-4: Preparation of steel substrates before application of paints and related products. Tests for the assessment of surface cleanliness.

More information

POINTS OF CAUTION BEFORE USE...

POINTS OF CAUTION BEFORE USE... INSTRUCTION MANUAL NPN DISCRETE INPUT & NPN TRANSISTOR OUTPUT MODULE, 6 points each (-I/-II, e-con connector) MODEL BEFORE USE... Thank you for choosing M-System. Before use, check the contents of the

More information

DS 200P. Electronic Pressure Switch. Pressure Ports And Process Connections With Flush Welded Stainless Steel Diaphragm

DS 200P. Electronic Pressure Switch. Pressure Ports And Process Connections With Flush Welded Stainless Steel Diaphragm Pressure Ports And Process Connections With Flush Welded Stainless Steel Diaphragm accuracy according to IEC 600: standard: 0. % FSO option: 0. % FSO Nominal pressure from 0... 00 mbar up to 0... 0 bar

More information

Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM / / 2014

Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM / / 2014 Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM8004 80003223 / 00 05 / 2014 Contents 1 Preliminary note...3 1.1 Symbols used...3 2 Safety instructions...4 3 Functions and features...5

More information

Lamp and Control Panel (Lamp Panel)

Lamp and Control Panel (Lamp Panel) www.reinhausen.com Lamp and Control Panel (Lamp Panel) Operating Instructions BA 47/0 EN NOTE! Changes may have been made to a product after going to press with this documentation. We expressly reserve

More information

DMP 331Pi. Precision Pressure Transmitter. pressure ports and process connections with flush welded stainless steel diaphragm

DMP 331Pi. Precision Pressure Transmitter. pressure ports and process connections with flush welded stainless steel diaphragm DMP Pi Precision Pressure Transmitter pressure ports and process connections with flush welded stainless steel diaphragm accuracy according to IEC 600: 0, % FSO Nominal pressure from 0... 00 mbar up to

More information

DMP 331P. Industrial Pressure Transmitter. Process Connections With Flush Welded Stainless Steel Diaphragm

DMP 331P. Industrial Pressure Transmitter. Process Connections With Flush Welded Stainless Steel Diaphragm DMP P Industrial Pressure Transmitter Process Connections With Flush Welded Stainless Steel accuracy according to IEC 60770: standard: 0.5 % FSO option: 0.5 % FSO Nominal pressure from 0... 00 mbar up

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer gh Performance Chilled Hygrometer With Cable Mounted Measuring Heads ghly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct insertion

More information

Air velocity & IAQ measuring instruments

Air velocity & IAQ measuring instruments Data sheet dp Air velocity & IAQ measuring instruments - air velocity & IAQ measuring instrument - dp air velocity & IAQ measuring instrument incl. differential pressure C CO Intuitive: clearly structured

More information

Semi-automatic titrators PH-Burette. A great ph-meter with integrated burette. Incredible price! Infinite possibilities

Semi-automatic titrators PH-Burette. A great ph-meter with integrated burette. Incredible price! Infinite possibilities Semi-automatic titrators PH-Burette. A great ph-meter with integrated burette. Incredible price! Infinite possibilities instruments 5 YEARS G U A R A N T E E Quality at a reasonable price The optimum combination

More information

- compressed air plants (refrigerating/adsorption dryers) - granulate dryers - medical gases - non-corrosive gases, e. g. nitrogen

- compressed air plants (refrigerating/adsorption dryers) - granulate dryers - medical gases - non-corrosive gases, e. g. nitrogen INSTRUCTION MANUAL FA 510 The CS dew point sensor FA 510 with 3-wire technology 4 20 ma and RS 485 output enables a reliable and long-term stable monitoring of the dew point in industrial applications

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer High Performance Chilled Mirror Hygrometer With Cable Mounted Measuring Heads Highly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct

More information

BTE5000 / PTE5000 / PTU5000 Series Precision differential pressure transmitters

BTE5000 / PTE5000 / PTU5000 Series Precision differential pressure transmitters FEATURES 5 mbar to 0 bar, 0. to 50 psi differential pressure...6 or...0 ma Precision temperature compensated and calibrated Rugged aluminium housing Female /8" BSP and /8" NPT fittings MEDIA COMPATIBILITY

More information

Latest generation testo air velocity & IAQ probes with cable handle. Instruction manual

Latest generation testo air velocity & IAQ probes with cable handle. Instruction manual Latest generation testo air velocity & IAQ probes with cable handle Instruction manual Contents Contents 1 About this document... 5 2 Safety and disposal... 5 3 System description... 6 4 Description of

More information

Dewpoint Meter Data Sheet / Instructions

Dewpoint Meter Data Sheet / Instructions Paint Test Equipment Dewpoint Meter Data Sheet / Instructions Dewpoint Meter Information ISO 8502-4: Preparation of steel substrates before application of paints and related products. Tests for the assessment

More information

TECHNICAL MANUAL 820 LX / 910 LX / 1300 LX

TECHNICAL MANUAL 820 LX / 910 LX / 1300 LX TECHNICAL MANUAL 820 LX / 910 LX / 1300 LX LANCER reserves the right to modify, constantly its documentation for its improvement. The values of adjustments indicated in the displays of this manual are

More information

Conductivity sensor for hygienic applications

Conductivity sensor for hygienic applications Conductivity sensor for hygienic applications Type 8221 can be combined with Perfect for demanding applications in the hygienic industry (CIP and SIP compatible) Variants available for usage over a wide

More information

Direct-operated 2/2-way compact solenoid valves Type EV210A

Direct-operated 2/2-way compact solenoid valves Type EV210A Data sheet Direct-operated 2/2-way compact solenoid valves Type EV210A EV210A covers a wide range of small, directoperated 2/2-way solenoid valves for use in industrial machinery. The compact design together

More information

Heat Meter WR 200-D. 2 temperature inputs for Pt100 in four-wire circuit. Exact calculation of the heat parameters according IAPWS-IF97

Heat Meter WR 200-D. 2 temperature inputs for Pt100 in four-wire circuit. Exact calculation of the heat parameters according IAPWS-IF97 Heat Meter WR 200-D electronic computing device for heat quantity measuring Heat transfer medium water steam Any flow sensor with analog signal connectable 2 temperature inputs for Pt100 in four-wire circuit

More information

Pressure transmitters for industrial applications MBS 32 and MBS 33

Pressure transmitters for industrial applications MBS 32 and MBS 33 Data sheet Pressure transmitters for industrial applications MBS 32 and MBS 33 The standard pressure transmitters MBS 32 and MBS 33 are designed for use in almost all industrial applications, and offer

More information

PHOENIX CONTACT - 04/2016. Features

PHOENIX CONTACT - 04/2016. Features Signal conditioner Data sheet 100238_de_06 1 Description PHOENIX CONTACT - 04/2016 Features The MCR-C-UI-UI(-450)-DCI(-NC) 3-way isolation amplifier is used to electrically isolate and convert analog signals.

More information

Operating instructions Magnetic-inductive flow meter SM6x04 SM7x04 SM8x04

Operating instructions Magnetic-inductive flow meter SM6x04 SM7x04 SM8x04 Operating instructions Magnetic-inductive flow meter SM6x04 SM7x04 SM8x04 UK 80223776 / 00 11 / 2018 Contents 1 Preliminary note...3 2 Safety instructions...4 3 Functions and features...4 4 Function...5

More information

PRODUCT CATALOGUE PREssURE switch

PRODUCT CATALOGUE PREssURE switch PRODUCT CATALOGUE PREssURE switch w w w. p a a b. c o m Pressure AT THe HIGHesT LeVeL Successful mediumsized companies are not successful because they are active in many areas, but rather because they

More information

VK2 volume flow controller Overview

VK2 volume flow controller Overview New option with sound attenuator Maintenance-free VK2 volume flow controller that operates without an auxiliary power supply, for ventilation and air conditioning systems. Adjustable on site. Outstanding

More information

O P E R A T I N G M A N U A L

O P E R A T I N G M A N U A L OPERATING MANUAL WeatherJack OPERATING MANUAL 1-800-645-1061 The baud rate is 2400 ( 8 bits, 1 stop bit, no parity. Flow control = none) To make sure the unit is on line, send an X. the machine will respond

More information

Safety Precautions WARNING If critical situations that could lead to user s death or serious injury is assumed by mishandling of the product.

Safety Precautions WARNING If critical situations that could lead to user s death or serious injury is assumed by mishandling of the product. Safety Precautions Observe the following notices to ensure personal safety or to prevent accidents. To ensure that you use this product correctly, read this User s Manual thoroughly before use. Make sure

More information

SECTION 1 - WHAT IS A BTU METER? BTU's = Flow x ΔT Any ISTEC BTU Meter System consists of the following main components:

SECTION 1 - WHAT IS A BTU METER? BTU's = Flow x ΔT Any ISTEC BTU Meter System consists of the following main components: SECTION 1 - WHAT IS A BTU METER? ISTEC BTU Meters measure energy usage by multiplying flow rate and temperature difference. As the water (or other liquid) passes through these lines, the multi-wing turbine

More information

Circular Motion and Centripetal Force

Circular Motion and Centripetal Force [For International Campus Lab ONLY] Objective Measure the centripetal force with the radius, mass, and speed of a particle in uniform circular motion. Theory ----------------------------- Reference --------------------------

More information

Solenoid valves 2/2-way direct-operated type EV210A

Solenoid valves 2/2-way direct-operated type EV210A MAKING MODERN LIVING POSSIBLE Data sheet Solenoid valves 2/2-way direct-operated type EV210A EV210A covers a wide range of small, direct-operated 2/2-way solenoid valves for use in industrial machinery.

More information

QUANTUM CONCEPT. Swimming User s Manual

QUANTUM CONCEPT. Swimming User s Manual QUANTUM CONCEPT Swimming User s Manual 480.508.0 Version.4 Edition July 05 Documentation Updates Swiss Timing Ltd. reserves the right to make improvements in the products described in this documentation

More information

air and solenoid air operated 5/2-5/3 ISO 5599/01 - Size 2

air and solenoid air operated 5/2-5/3 ISO 5599/01 - Size 2 Series 4/PH MULTIFUNCTION SPOOL VALV air and solenoid air /-/ ISO 99/0 - Size ports / positions SPCIFICATIONS FLUID : or neutral gas, filtered, lubricated or not PRSSUR : 0,90 to + bar (can be used with

More information

Overload Relays. SIRIUS 3RU1 Thermal Overload Relays. 3RU11 for standard applications. 5/46 Siemens LV 1 AO 2011

Overload Relays. SIRIUS 3RU1 Thermal Overload Relays. 3RU11 for standard applications. 5/46 Siemens LV 1 AO 2011 SIRIUS 3RU1 Thermal Overview 1 2 7 3 4 6 "Increased safety" type of EEx e according to ATEX directive 94/9/EC The 3RU11 thermal overload relays are suitable for the overload of explosion-proof motors with

More information

Proximity Sensor Ideal for High Temperatures and Cleaning Processes. Applicable to 120 C

Proximity Sensor Ideal for High Temperatures and Cleaning Processes. Applicable to 120 C ASH & ALAIN INDIA PVT LTD S-00, F.I.E.E., Okhla Industrial Area, Phase-ii, New Delhi-000(India) Tel : 0-797575 Fax : 0-79757 E-mail : sales@ashalain.com Proximity Ideal for High Temperatures and Cleaning

More information

Its modular concept allows to build up a ball valve for any degree of automation and any kind of application

Its modular concept allows to build up a ball valve for any degree of automation and any kind of application Type 2657...59 DN 10 up to 80 mm; Plastic; 2/2 way dvantages/benefits ntegrated pilot valves, position feed-back and limit switches available Customized system solutions with Burkert valves & sensors With

More information

PTF30 Series. Features. Ordering information. G 6 N N F8 (0 to 5 kgf/cm 2 ) PTF30

PTF30 Series. Features. Ordering information. G 6 N N F8 (0 to 5 kgf/cm 2 ) PTF30 eries Features Best pressure transmission solution for small and middle sized project such as indoor heating and air conditioning, water and sewer, or incinerator, etc. Improves noise resistance and minimize

More information

TF 212 / TF 212-Ex. Field mounted temperature transmitters, PROFIBUS PA, Pt 100 (RTD), thermocouples, 1 or 2 independent channels 10/11-8.

TF 212 / TF 212-Ex. Field mounted temperature transmitters, PROFIBUS PA, Pt 100 (RTD), thermocouples, 1 or 2 independent channels 10/11-8. TF / TF -Ex Field mounted temperature transmitters, PROFIBUS PA, Pt 00 (RTD), thermocouples, or independent channels 0/-8.70 EN P R O F I PROCESS FIELD BUS B U S Input Resistance thermometer (,, -wire

More information

PiCUS Calliper Version 3

PiCUS Calliper Version 3 Technology PiCUS Calliper Version 3 Digital calliper for precise geometry detection argus electronic GmbH Erich-Schlesinger-Straße 49d 18059 Rostock Germany Tel.: +49 (0) 381/49 68 14 40 www.argus-electronic.de

More information

The Ordering Code for various standard model Recorders with an AC supply and without any additional options are as follows:

The Ordering Code for various standard model Recorders with an AC supply and without any additional options are as follows: The basic PC software is supplied free with the recorder. There is an additional charge for the extensive Data Acquisition Software supplied with communication of RS-232/422/485 or Ethernet. The Ordering

More information

16 Input AC Block I/O Module

16 Input AC Block I/O Module 6 Input AC Module Cat. No. 79-6A0 Series B Installation Mount the block I/O module in a vertical (recommended) or horizontal position. Allow sufficient room around the block for cooling air flow through

More information

Double Inverted Pendulum (DBIP)

Double Inverted Pendulum (DBIP) Linear Motion Servo Plant: IP01_2 Linear Experiment #15: LQR Control Double Inverted Pendulum (DBIP) All of Quanser s systems have an inherent open architecture design. It should be noted that the following

More information

RoHS. Versatile. X g. X 2 e

RoHS. Versatile. X g. X 2 e s IN8, - and The inclinometers of the IN8 series allow measuring inclinations in the range of ±85 or -dimensional inclinations up to 0. With their high robustness, their protection level up to max. IP9k

More information

BEFORE USE... POINTS OF CAUTION COMPONENT IDENTIFICATION INSTRUCTION MANUAL MODEL R7K4DH-1-DAC32D R7K4DH-1-DAC32D

BEFORE USE... POINTS OF CAUTION COMPONENT IDENTIFICATION INSTRUCTION MANUAL MODEL R7K4DH-1-DAC32D R7K4DH-1-DAC32D INSTRUCTION MANUAL PNP DISCRETE INPUT & PNP TRANSISTOR OUTPUT MODULE, 6 points each (High-speed Link System, e-con connector) R7KDH--DACD MODEL R7KDH--DACD BEFORE USE... Thank you for choosing M-System.

More information

Automatic titrators, TitroMatic

Automatic titrators, TitroMatic 112 TitroMatic These are the fourth generation of CRISON titrators. CRISON has over 25 years experience in this field and knows that, for certain applications, the performance and cost of large titrator

More information

Commenting template CEN/TC 102 Date : Document: pren 13060

Commenting template CEN/TC 102 Date : Document: pren 13060 Commenting template CEN/TC 0 Date : 006-09-3 Document: pren 3060 MB Clause / 3.) ) FR general Ge There have been so many discussions during the CEN/TC 0/WG5 meetings these years. The committee has recognised

More information

E8EB-N0C2B E8EB-N0B2B

E8EB-N0C2B E8EB-N0B2B Slim Sensor CSM DS_E 1 Ideal for Workpiece Position and Original Checking The to 1 kpa model can be used for workpiece position checking. The to 1MPa model is ideal for original pressure checking. Degree

More information

Appendix B1. Reports of SMU

Appendix B1. Reports of SMU Appendix B1 Reports of SMU Euromet Project 600 Comparison of Surface Roughness Standards /13 EUROMET SUPPLEMENTARY COMPARISON SURFACE TEXTURE Project No. 600 Final Report Elaborated by: M. Szmicskova Bratislava,

More information

/2 way valves (solenoid actuated diaphragm with forced lifting) 1/4... 2

/2 way valves (solenoid actuated diaphragm with forced lifting) 1/4... 2 / way valves (solenoid actuated diaphragm with forced lifting) /4... Damped operation Operation without differential Suitable for vacuum igh flow rate For systems with low or fluctuating Brass body Wide

More information

3B SCIENTIFIC PHYSICS

3B SCIENTIFIC PHYSICS 3B SCIENTIFIC PHYSICS 30 V, 50/60 Hz: 1018884 / U07001-30 115 V, 50/60 Hz: 101888 / U07001-115 Millikan s Apparatus Instruction sheet 07/16 UD/ALF Safety instructions Millikan s apparatus conforms to the

More information

AIR PREPARATION. ELECTRONICALLY CONTROLLED PRESSURE REGULATING VALVE (PROPORTIONAL PRESSURE REGULATING VALVE) SERIES control

AIR PREPARATION. ELECTRONICALLY CONTROLLED PRESSURE REGULATING VALVE (PROPORTIONAL PRESSURE REGULATING VALVE) SERIES control AIR PREPARATION ELECTRONICALLY CONTROLLED PRESSRE REGLATING ALE (PROPORTIONAL PRESSRE REGLATING ALE) SERIES control Description Abb. Port size Recommended Type Data Sheet No. flow (l/min)* airfit tecno

More information

Remote Display Unit. Installers Handbook Copyright 2001 AirSense Technology Ltd. LM Remote Display Unit Installers Handbook Issue 1.

Remote Display Unit. Installers Handbook Copyright 2001 AirSense Technology Ltd. LM Remote Display Unit Installers Handbook Issue 1. Remote Display Unit Installers Handbook Copyright 2001 AirSense Technology Ltd AirSense, ClassiFire, FastLearn PipeCAD, SenseNET, Stratos-HSSD and Stratos-Quadra are trademarks. HSSD is a Registered Trademark.

More information

WEATHER STATION FOR SOLAR FARM MONITORING

WEATHER STATION FOR SOLAR FARM MONITORING WEATHER STATION FOR SOLAR FARM MONITORING SOLAR FARM MONITORING SYSTEM: Measures global, horizontal, & background irradiance. Measures wind speed, wind direction, ambient temperature, and relative humidity.

More information

VIBRATION METER. Instruction Manual.

VIBRATION METER. Instruction Manual. VIBRATION METER 840063 Instruction Manual CONTENTS Title Page I. Introduction... 2 II. Panel Description... 3 III. Operating Instructions A. General Measurement Procedures... 3~4 B. Peak... 5 C. Data Hold...

More information

32 Input AC Block I/O Module

32 Input AC Block I/O Module Input AC Module Cat. No. 79-A0 Series B Installation Mount the block I/O module in a vertical (recommended) or horizontal position. Allow sufficient room around the block for cooling air to flow through

More information

Operator's Manual. optris BR 400. Calibration source

Operator's Manual. optris BR 400. Calibration source Operator's Manual optris BR 400 Calibration source Optris GmbH Ferdinand-Buisson-Str. 14 13127 Berlin Germany Tel.: +49 30 500 197-0 Fax: +49 30 500 197-10 E-mail: info@optris.de Internet: www.optris.de

More information

by A.Tonina*, R.Iuzzolino*, M.Bierzychudek* and M.Real* S. Solve + R. Chayramy + and M. Stock +

by A.Tonina*, R.Iuzzolino*, M.Bierzychudek* and M.Real* S. Solve + R. Chayramy + and M. Stock + Bilateral Comparison of 1.018 V and 10 V Standards between the INTI (Argentina) and the BIPM, August to October 2009 (part of the ongoing BIPM key comparison BIPM.EM-K11.a and b) by A.Tonina*, R.Iuzzolino*,

More information

TI 226H Calibration of Flow Check Devices using Positive Displacement Flow Meter

TI 226H Calibration of Flow Check Devices using Positive Displacement Flow Meter Page 1 of 9 TI 226H Calibration of Flow Check Devices using Positive Displacement Flow Meter TABLE OF CONTENTS 1. PURPOSE AND APPLICABILITY... 2 2. RESPONSIBILITIES... 2 2.1. Field Specialist... 2 2.2.

More information

Reflex-Radar BM 100 A

Reflex-Radar BM 100 A KROHNE 09/2001 7.02317.22.00 GR/PRINTO Reflex-Radar The universal TDR level gauge for liquids, solids and interface Variable area flowmeters Vortex flowmeters Flow controllers Electromagnetic flowmeters

More information

Installation guide 862 MIT / MIR

Installation guide 862 MIT / MIR Installation guide 862 MIT / MIR in combination with: 864 MTT or 863 MRT or (dual) spot element November 2005 Part no. 4416.232_Rev3 Enraf BV PO Box 812 2600 AV Delft Netherlands Tel. : +31 15 2701 100

More information

CERTIFICATE of Product Conformity (QAL1 )

CERTIFICATE of Product Conformity (QAL1 ) ATÜVRheinlancl Precisely Right. CERTIFICATE of Product Conformity (QAL1 ) Certificate No.: 0000040212 01 Certified AMS: Manufacturer: Test lnstitute: Fidas 200 S respectively Fidas 200 for particulate

More information

DM70. Hand-held Dewpoint Meter. Truly Fast Response - Superior Long-term Stability

DM70. Hand-held Dewpoint Meter. Truly Fast Response - Superior Long-term Stability DM70 Hand-held Dewpoint Meter Truly Fast Response - Superior Long-term Stability DM70 - When Dry Air Is Important The Vaisala DM70 Hand-held Dewpoint Meter offers stable and fast measurement for industrial

More information

ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES

ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES (RPB-01) Approved by: Date: SECTION PAGE # 1.0 Scope 2.0 Definitions 3.0 References 4.0 Inspection

More information

xlogic SuperRelay is a compact and expandable CPU replacing mini PLCs, multiple timers, relays and counters.

xlogic SuperRelay is a compact and expandable CPU replacing mini PLCs, multiple timers, relays and counters. SuperRelay S U P E R R E L A Y, T H E P E R F E C T A L T E R N A T I V E T O L O W C O S T P L C s A N D B A S I C R E L A Y S I N T H I S B R O C H U R E : ELC 18 Standard Models ELC 18 Economy Models

More information

SINEAX VK 626 Programmable Temperature Transmitter for RTD and TC inputs, with HART protocol

SINEAX VK 626 Programmable Temperature Transmitter for RTD and TC inputs, with HART protocol SINEAX VK 66 for RTD and TC inputs, with HART procol for installation in the terminal head of a temperature sensor DIN 79, shape B Application II () G SINEAX VK 66 is a two-wire head-mounted transmitter.

More information

Non-Acoustical Inputs

Non-Acoustical Inputs CHAPTER 18 Non-Acoustical Inputs This chapter discusses the use of external transducers and devices to provide non-acoustical data to the Model 831. Included are the following: 831-INT 831-INT Interface

More information

Comparison of V and 10 V DC Voltage References

Comparison of V and 10 V DC Voltage References Comparison of 1.018 V and 10 V DC Voltage References Technical protocol for BIPM.EM-K11.a and.b comparisons 1. Introduction The Mutual Recognition Arrangement among National Metrology Institutes (NMIs)

More information

Analog - Digital multimeters

Analog - Digital multimeters CAT-IV -3, MI, Satpur, Nashik-4 7,India. Tel.: +9 53 6, ax : +9 53 3564 Page of / 5 / I I / 7 / B Automatic Terminal Blocking System (ABS) The automatic Terminal blocking system prevents incorrect connection

More information

Direct-operated 2/2-way solenoid valves Type EV210B

Direct-operated 2/2-way solenoid valves Type EV210B Data sheet Direct-operated 2/2-way solenoid valves Type EV20B EV20B covers a wide range of direct-operated 2/2-way solenoid valves for universal use. EV20B is a very robust valve program with high performance

More information

Experiment A4 Sensor Calibration Procedure

Experiment A4 Sensor Calibration Procedure Experiment A4 Sensor Calibration Procedure Deliverables: Checked lab notebook, Brief technical memo Safety Note: Heat gloves and lab coats must be worn when dealing with boiling water. Open-toed shoes

More information