Vet Photometer DP 700 Operating Manual. Version V5.9

Size: px
Start display at page:

Download "Vet Photometer DP 700 Operating Manual. Version V5.9"

Transcription

1 1 Vet Photometer DP 700 Operating Manual Version V5.9

2 2 Dear customer, Thank you for selecting Diaglobal s Vet Photometer and for the confidence you have placed in us. The Vet Photometer belongs to a new generation of small mobile instruments which are being developed by Diaglobal GmbH and are specially designed for on-site analysis. An automatic operational device check has been additionally integrated with the actual software version V5.4. The Vet Photometer can be used to determine four clinical-chemical parameters from serum/plasma, blood: NEFA (free fatty acids), Lactate, Calcium and Magnesium. The kits and accessories which are required to carry out the tests and measurements are also available from Diaglobal GmbH. All the best for your work with the new Vet Photometer! Yours Diaglobal GmbH

3 3 Table of contents Page 1. General information regarding the Photometer 4 2. Installation 5 3. Description of the instrument Power supply Mains power operation Network-independent operation Measuring system 6 4. Service Calibration Maintenance Disturbances Disposal 7 5. Required reagents and laboratory accessories Reagents / parameter list Control materials Disposables and accessories 8 6. Measuring process Multipoint measurement and recognition of the end point End point measurement taking standard 8 7. Measurement Switching the instrument on Self-testing when switching on Test selection Exit the program / Switching the instrument off Integrated operational device checks 9 8. Technical data General guidelines and notes Appendix: Step-by-step measurement 12ff

4 4 1. General information regarding the Photometer Device name: Vet Photometer Model: DP 700 Features: In-vitro diagnostics, measuring instrument for the determination of NEFA (free fatty acids), Calcium and Magnesium from serum/plasma and Lactate from plasma/blood The Vet Photometer fulfils the basic requirements of Low Voltage Directive 2006/95/EG. The conformity of the instrument with Directive 2006/95/EG is confirmed by the use of the CE marking. The Vet Photometer was specially developed for analyses in veterinary medicine, however it is universally applicable. Manufacturer: Diaglobal GmbH Innovationspark Wuhlheide Köpenicker Str. 325 D Berlin Tel: Fax: info@diaglobal.de

5 5 2. Installation Please observe the following instructions for use: Insert the rechargeable battery if the instrument is to be used networkindependent or connect the photometer to the power supply unit. Press the key </> (Fig. 1) to activate the internal instrument check which is automatically carried out by the instrument. Following this, the instrument is immediately ready for measuring. 3. Description of the instrument Cuvette shaft Display Function keys Fig. 1

6 Power supply The Vet Photometer can be operated as desired using a power supply, a (9V block) battery or (model 6F22 or PP3) rechargeable battery Mains power operation The power supply connector plug is connected to the instrument s socket. The Vet Photometer is provided with a power supply unit (6-9V=) for operation at 230 V/50 Hz Network-independent operation Insertion of the rechargeable battery or the normal battery: Unscrew the knurled screws on the bottom side and remove the cap. Connect the rechargeable battery or the normal battery using the push-button contact and insert it. Place the cap back on top and fasten the knurled screws. The rechargeable battery cannot be charged whilst it is installed. A separate battery charger is required for this purpose. Please note: The Vet Photometer can be operated using a power supply and it is not necessary to remove the rechargeable battery or the normal battery for this purpose. 3.2 Measuring system The optical section is shown in Fig. 2. Focus Cuvette Fig. 2 The light emitted by an LED is first selected by an interference filter IF (HBW ~ 5 nm) in its spectral ranges (520 nm) and then guided concentrated to the cuvette in the shaft. After it has passed through the cuvette, a broadband photo sensor converts the light falling onto its sensor surface to current, proportional to its intensity.

7 4. Service Calibration The instrument is works calibrated when supplied - it is not necessary for the customer to calibrate it. 4.2 Maintenance After the expiration of warranty we recommend regular yearly maintenance of the instrument. 4.3 Disturbances If any disturbances or problems occur, simply call us. We can answer most questions on the telephone. Faulty instruments must be sent to our Berlin address. We can lend our customers an instrument whilst theirs is being repaired. 4.4 Disposal We take back instruments free of charge which are no longer needed or cannot be repaired and dispose of them. 5. Required reagents and laboratory accessories 5.1 Reagents / parameter list The following tests can be measured by the Vet Photometer: Parameter Sample material Blood Serum Plasma Tests/Pckg. Art. No. NEFA NEFA 013 Lactate LAC 142 Calcium CA 015 Magnesium MG Control materials Art. No. Description Contents LAC QS Lactate Control Set 2.00 / 4.00 / 10.0 mmol/l 3 x 4 ml CA QS Calcium-/Magnesium-Control 3 x 1.5 ml NEFA QS NEFA Control 3 x 1.0 ml

8 5.3 Disposables and accessories 8 Art. No. Description Contents LH 004 Capillaries 10 µl, with ringmark (Micropipettes) 250 LH 006 Cuvette rack 1 LH 007 Micropipettor (pipetting aid) 1 LH 009 Cellulose swabs 500 LH 010 Cellulose swab box 1 LH 011 Alcohol pads, non-sterile 100 LH 012 Latex gloves, size 6-7 small or 7-8 medium 100 LH 053 Pipette tips ml, clear 200 LH 054 Pipette tips µl, yellow 500 LH 055 Pipette tips µl, blue 500 LH 503 Pipette fix 1500 µl 1 LH 504 Pipette variabel ml 1 LH 550 Pipette fix 50 µl 1 All reagent kits, control materials and other materials can be stored and transported together with the Vet Photometer in a practical case. 6. Measuring process 6.1 Multipoint measurement, taking the sample s blank point and recognition of the end point into account Parameter: Lactate (LAC 142) After measurement of the sample blank value (= measurement 1), the colour reaction is started in the cuvette. The reaction process is monitored by the instrument (= measurement 2). The measuring procedure is stopped as soon as the end point is reached. The time needed to reach the end point is temperature-dependent. It is normally 2-6 minutes for the lactate test. If temperatures are close to freezing point, measuring times of up to 20 minutes may result, dependent on the parameters. Both single measurements and series measurements can be selected. If single measurements are made, the samples are measured one after another. For series measurements, all A1 values are measured first. Calculation of concentration = Absorbance Difference x Factor Calculation is made using a complex special program by the photometer 6.2 End point measurement taking standard Parameter: NEFA (free fatty acids), Calcium and Magnesium Measurement of absorbance after reaching the end point in relation to the reagent blank value and taking standard. Calculation: Concentration = Factor x (EAnalysis - EBlank) / (EStandard - EBlank)

9 7. Measurement Switching the instrument on Press the < / > key 7.2 Self-testing when switching on A self-testing of the digital and analogue circuit occurs when switching on the instrument. The operational device check proceeds automatically after the switching-on. It lasts ca. 5 seconds, after this the device is ready for measuring. Note: In case during the testing it should emerge that one of the device functions does not satisfy the required settings, the following display appears: <SERVICE> In this case please switch off the instrument. Please call the Service of Diaglobal GmbH (tel / 30 / ) or contact your speciality retailer. 7.3 Test selection Press the < / > key Select the desired method using from the menu using the right or left arrow key: NEFA - LAC - CA - MG - ABS520 Press the right arrow key to mark the respective next test and the left arrow key to return to the previous test. The test currently selected is shown in the upper right-hand corner of the display. Confirm test selection by pressing the < / > key 7.4 Switching the instrument off To switch the instrument off, press the keys together. 7.5 Integrated operational device checks Self-testing when switching on The testing of the device s digital and analogue circuit occurs automatically already when switching on the instrument. Please see chapter 8, point 2. Differential measurements All measurements are based on differential measurements, i.e. after having chosen the desired test the instrument requests a reference measurement with a blank value cuvette. Thus, a reference basis to the measured value is established so that minor deviations may be compensated.

10 10 Measurement range controls The measurement ranges of all measurement results indicated in the display become verified by an integrated measurement range control. An error display appears when exceeding the measurement range. All measurement ranges, which are separately defined for each parameter, are documented in the respective package inserts as well as in the operating manual, chapter 9, technical data. Plausibility controls When performing multiple point measurements the extinction, which was measured first, represents the reference basis. The programme tests the individual measured values for plausibility. In case particular requirements (e.g. E2 > E1 for rising reactions) are not met, an error message is displayed.

11 11 8. Technical data Storage temperature: Operating temperature: Dimensions: Weight: Measuring principle: Projection: Spectroscopic apparatus: Measuring wavelengths: Spectral half-width value: Extraneous light influence: - 20 C...70 C 0 C...40 C 200 x 100 x 50 mm 450 g Absorption metering with a single-beam photometer (fig. 2), choppered operation LED Interference filter 520 nm ~ 5 nm Negligible Interface: V24 (9600, 8, n, 2) Power supply: Warm-up time: 9V battery or rechargeable battery block 6F22 or PP3 optional plug-in power pack 0 min Interference suppression: According to DIN VDE 0871 and DIN VDE 0875 Electromagnetic Compatibility: According to EN Measuring ranges: NEFA mmol/l LAC CA MG mmol/l mmol/l mmol/l ABS 520nm E = Inaccuracy: < 0.5 % at E = Relative photometric short-time standard deviation: < 0.1 %

12 12 9. General Guidelines and Notes EC Directives 1. Low Voltage Directive 2006/95/EG EN / ISO standards 2. EN ISO 9001:1994, Qualitäty Management Systems, Model for quality assurance in design, development, production, installation and servicing 3. EN , Safety requirements for electrical equipment for measurement, control and laboratory use - Part 1: General requirements 4. EN , Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements Note on electromagnetical compatibility a) This photometer meets the requirements of the IEC series of standards regarding electromagnetic radiation and interference immunity. b) Do not use this unit near to sources of intense electromagnetic radiation as these may disturb its correct functioning. During the measurement, a distance of at least 1 m should be kept between an operative (turned on) mobile telephone and this photometer. Note on the unit's internal quality control This unit checks its functionality when it is turned on. Moreover, it performs electronically controlled individual tests during the measurement causing an error message when defined conditions are not met. 10. Appendix: Step-by-step measurement Please refer to the illustrations of the following pages.

13 Step by step instructions Device manual 1. Switch on: 2. Select test: Press / Press arrow key until required test appears Wait for device check and confirm with / 3. Confirm required test: 4. Switch on: Press / Press both arrow keys at the same time Note: If SERVICE appears after the device check, the device has a fault. In this case, please contact our customer service at +49 (0) Diaglobal GmbH Köpenicker Straße Berlin +49 (0) info@diaglobal.de

14 Step by step instructions NEFA (free fatty acids) Example: 2 samples Additionally required: Free fatty acids standard (NEFA ST), Diaglobal-round cuvettes (LH 075) 1. According to instruction prepare the reagent mixtures Thereafter, pipette 1000 µl of the bottle R1a in each cuvette 2. Pipette samples: cuvette 1: nothing (= blank) cuvette 2: 50 µl standard cuvette 3: 50 µl sample 1 cuvette 4: 50 µl sample 2 Close cuvettes, mix and leave for 10 minutes 3. After 10 minutes of waiting time pipette 500 µl R2a in each cuvette Close and mix again and leave again at room temperature for 10 minutes Insert NEFA Insert standard NEFA NEFA Standard Abs blank 4. Switch on the photometer with / Wait for device check, confirm with / Select NEFA, confirm with / 5. Blank adjustment: Insert cuvette 1 (= blank) into photometer, blank is stored in memory Remove after signal tone 6. Insert cuvette 2 with standard into photometer Photometer shows Absorption of cuvette with standard Hint: If necessary, note this value for documentation purposes. Sample 01 insert NEFA NEFA Sample mmol/l 7. Remove cuvette 2 with standard 8. Insert cuvette 3 (sample 1) and thereafter cuvette 4 (sample 2) into photometer Read results successively In regard to serial measurement After blank setting any number of additional samples can be measured Hint: To verify the measured values, we recommend carrying a control: NEFAcontrol (Art. No. NEFA QS) Diaglobal GmbH Köpenicker Straße Berlin +49 (0) info@diaglobal.de

15 Step by step instructions CA 015 / MG 013 Example: 2 samples Additionally Required: Calcium-Magnesium-Standard (CA ST), Diaglobal-round cuvettes (LH 075) CA Insert blank 1. According to the instructions, pipette the prescribed amounts buffer and color solution into cuvettes Thereafter, pipette the standard and the two samples (fig. 2) 2. Pipette into cuvettes, mix well: cuvette 1: nothing (= blank) cuvette 2: standard cuvette 3: sample 1 cuvette 4: sample 2 CA 015: 50 µl / without waiting time MG 013: 10 µl / wait 5 minutes 3. Switch on the photometer with / Wait for device check, confirm with / Select the required test, confirm with / Insert standard CA Standard Abs CA Sample 01 insert CA 4. Blank adjustment: Insert cuvette 1 (= blank) into photometer, blank is stored in memory Remove after signal tone 5. Insert cuvette 2 with standard into photometer Photometer shows Absorption of cuvette with standard Hint: If necessary, note this value for documentation purposes. 6. Remove cuvette 2 with standard CA Sample mmol/l 7. Insert cuvette 3 (sample 1) and thereafter cuvette 4 (sample 2) into photometer Read results successively In regard to serial measurement After blank setting any number of additional samples can be measured In regard to quality control To verify the measured values, we recommend carrying a control: Calcium Magnesium Control (Art. No. CA QS) Diaglobal GmbH Köpenicker Straße Berlin +49 (0) info@diaglobal.de

16 Step by step instructions LAC 142 / LAC 342 / GLU 142 / CHO 142 / TRI 142 Single measurement 1. Insert capillary with sample into cuvette 2. Wash out sample with micropipetter 3. Screw on cap Turn cuvette upside down several times M1 LAC M1 LAC Sample 01 insert Sample 02 insert or 4. Switch on photometer with / Wait for device check, confirm with / Select the required test, confirm with / 5. Blank adjustment: Insert cuvette with sample (fig. 3) into photometer, blank is stored in memory Remove after signal tone 6. Replace screw cap against starter cap M2 LAC Sample 01 00:30 Measuring 7. Turn cuvette upside down several times 8. First press / Thereafter insert cuvette in photometer 9. Timing appears Wait for result Diaglobal GmbH Köpenicker Straße Berlin +49 (0) info@diaglobal.de

17 Step by step instructions LAC 142 / CHO 142 / TRI 142 Serial measurement M1 Sample 01 insert LAC 1. Wash out samples from capillaries into cuvettes with micropipetter 2. Screw on cap Turn cuvette upside down several times 3. Switch on photometer with / Wait for device check, confirm with / Select the required test, confirm with / M1 Sample 19 insert or LAC 4. Blank adjustment: Insert cuvette with samples (fig. 2) into photometer, blank is stored in memory Ensure correct order of samples! 5. After blank adjustment of the last cuvette replace all screw caps against starter caps 6. Turn cuvettes upside down several times at the same time M2 LAC Sample 01 00:30 Measuring M2 LAC Sample mmol/l 7. First press / Thereafter insert cuvette 1 in photometer Timing appears Wait for result 8. Read the measured value of cuvette 1, remove cuvette Insert cuvette 2, read the measuring value, remove cuvette etc. 9. Repeat the process until the measured value of the last cuvette is displayed Pay attention to correct assignment and sequence of samples! Diaglobal GmbH Köpenicker Straße Berlin +49 (0) info@diaglobal.de

Vario Photometer DP 300 Operating Manual. Version 5.11 / 5.11 SI

Vario Photometer DP 300 Operating Manual. Version 5.11 / 5.11 SI 1 Vario Photometer DP 300 Operating Manual Version 5.11 / 5.11 SI 2 Dear customer, Thank you for selecting Diaglobal s Vario Photometer and for the confidence you have placed in us. The Vario Photometer

More information

Technology that counts

Technology that counts Application Note No. 100 Version 1.3 Determination of Sperm Density in Ejaculates from Boars Application This Application Note describes how sperm cell density in ejaculates from boars can be determined

More information

Technology that counts

Technology that counts Application Note No. 103 Version 1.2 Determination of Sperm Density in Bull Semen Doses for AI Application This Application Note describes how sperm cell density in bull semen doses for Artificial Insemination

More information

Technology that counts

Technology that counts Application Note No. 102 Version 1.3 Determination of Sperm Density in Ejaculates from Bulls Application This Application Note describes how sperm cell density in ejaculates from bulls can be determined

More information

Operating Instructions for Digital Manometer. Model: MAN-SD

Operating Instructions for Digital Manometer. Model: MAN-SD Operating Instructions for Digital Manometer Model: MAN-SD 1. Contents 1. Contents...2 2. Note...3 3. Instrument Inspection...3 4. Regulation Use...4 5. Operating Principle...4 6. Mechanical Connection...5

More information

SC400 Colorimeter. Silica. Photometer-System. Instruction Manual. SiO 2. Bedienungsanleitung Seite Instruction Manual Page 19 33

SC400 Colorimeter. Silica. Photometer-System. Instruction Manual. SiO 2. Bedienungsanleitung Seite Instruction Manual Page 19 33 Photometer-System SC400 Colorimeter SiO 2 Silica DE GB FR IT ES Bedienungsanleitung Seite 3 17 Instruction Manual Page 19 33 d'emploi Page 35 49 Instruction Manual Istruzioni d'uso Pagina 51 65 Instrucciones

More information

DS 300. Electronic Pressure Switch. with IO-Link interface. Stainless Steel Sensor. accuracy according to IEC 60770: 0.35 % FSO.

DS 300. Electronic Pressure Switch. with IO-Link interface. Stainless Steel Sensor. accuracy according to IEC 60770: 0.35 % FSO. DS 00 Electronic Pressure Switch with IO-Link interface Stainless Steel Sensor accuracy according to IEC 60770: 0.5 % FSO Nominal pressure from 0... 00 mbar up to 0... 600 bar Digital output signal IO-Link

More information

FLT5A FLT5A. abc. abc. Thermometer User Guide. Thermometer User Guide. Internet:

FLT5A FLT5A. abc. abc. Thermometer User Guide. Thermometer User Guide. Internet: abc instruments international FLT5A Thermometer Land Instruments International Dronfield, S18 1DJ England Telephone: (1246) 417691 Facsimile: (1246) 41585 Email: infrared.sales@landinst.com Internet: www.landinst.com

More information

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials Nitrate, HR DOC316.53.01068 Chromotropic Acid Method Method 10020 0.2 to 30.0 mg/l NO 3 N (HR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific information

More information

Semi-automatic titrators PH-Burette. A great ph-meter with integrated burette. Incredible price! Infinite possibilities

Semi-automatic titrators PH-Burette. A great ph-meter with integrated burette. Incredible price! Infinite possibilities Semi-automatic titrators PH-Burette. A great ph-meter with integrated burette. Incredible price! Infinite possibilities instruments 5 YEARS G U A R A N T E E Quality at a reasonable price The optimum combination

More information

Instruction Manual HI 96759C Maple Syrup Transmittance Analyzer

Instruction Manual HI 96759C Maple Syrup Transmittance Analyzer Instruction Manual HI 96759C Maple Syrup Transmittance Analyzer www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary information for

More information

TPH (Total Petroleum Hydrocarbons)

TPH (Total Petroleum Hydrocarbons) TPH (Total Petroleum Hydrocarbons) DOC316.53.01475 Immunoassay 1 Method 10050 Scope and application: For water. 1 This test is semi-quantitative. Results are shown as more or less than the threshold value

More information

Please use only the valid version of the package insert provided with the kit. This kit is intended for Research Use Only.

Please use only the valid version of the package insert provided with the kit. This kit is intended for Research Use Only. Please use only the valid version of the package insert provided with the kit This kit is intended for Research Use Only. Not for use in diagnostic procedures. INTENDED USE The Vitamin Niacin test is a

More information

Acceleration/Velocity/Displacement VIBRATION METER

Acceleration/Velocity/Displacement VIBRATION METER Acceleration/Velocity/Displacement VIBRATION METER Model : VB-8220 Your purchase of this VIBRATION METER marks a step forward for you into the field of precision measurement. Although this METER is a complex

More information

NECi Nitrate Kits FAQs

NECi Nitrate Kits FAQs Why use enzymes? Enzymes are catalysts that drive complex biological reactions. They happen to be excellent reagents for analytical chemistry because they are reliable, accurate, sensitive, selective,

More information

ISOCON-6. 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER

ISOCON-6. 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER ISOCON-6 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER Whilst every effort has been taken to ensure the accuracy of this document, we accept no responsibility for damage, injury, loss or expense resulting

More information

Color Compensation Set 1

Color Compensation Set 1 Color Compensation Set 1 Version 2, March 2012 Solutions for the generation of color compensation objects for the LightCycler System. Order No. A 500 08 For 5 calibration runs. Store in dark at -15 to

More information

HI REDUCING SUGARS ISM for wine analysis

HI REDUCING SUGARS ISM for wine analysis Instruction Manual HI 83746 REDUCING SUGARS ISM for wine analysis www.hannainst.com This Instrument is in Compliance with the CE Directives 1 Dear Customer, Thank you for choosing a Hanna product. This

More information

Instructions for use. Chromogranin A ELISA TM E-9000

Instructions for use. Chromogranin A ELISA TM E-9000 Instructions for use Chromogranin A ELISA TM E-9000 Chromogranin A ELISA 1. Intended use and principle of the test Enzyme Immunoassay for the quantitative determination of Chromogranin A in serum and plasma.

More information

Human Mullerian Inhibiting Substance/Anti-Mullerian hormone (MIS/AMH)Elisa Kit

Human Mullerian Inhibiting Substance/Anti-Mullerian hormone (MIS/AMH)Elisa Kit Human Mullerian Inhibiting Substance/Anti-Mullerian hormone (MIS/AMH)Elisa Kit Catalog No. CSB-E12756h (96T) This immunoassay kit allows for the in vitro quantitative determination of human MIS/AMH concentrations

More information

Instructions for use. Glycine ELISA BA E-2100

Instructions for use. Glycine ELISA BA E-2100 Instructions for use Glycine ELISA BA E-2100 Glycine Urine ELISA 1. Intended use and principle of the test Enzyme Immunoassay for the quantitative determination of Glycine in urine. After derivatization

More information

Mouse Platelet Factor 4 (PF4) ELISA

Mouse Platelet Factor 4 (PF4) ELISA Mouse Platelet Factor 4 (PF4) ELISA For the quantitative determination of mouse PF4 in plasma Cat. No. KT-25527 For Research Use Only. Not for use in diagnostic procedures. Page 1 of 7 Rev. 2010-02-17

More information

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832 , TNTplus 832, 10205 DOC316.53.01083 Salicylate Method Method 10205 HR (2 to 47 mg/l NH 3 N) TNTplus 832 Scope and Application: For surface waters, municipal and industrial wastewaters. Test preparation

More information

Latest generation testo air velocity & IAQ probes with cable handle. Instruction manual

Latest generation testo air velocity & IAQ probes with cable handle. Instruction manual Latest generation testo air velocity & IAQ probes with cable handle Instruction manual Contents Contents 1 About this document... 5 2 Safety and disposal... 5 3 System description... 6 4 Description of

More information

Mouse Glutathione S Transferase Alpha 1 (GSTa1)

Mouse Glutathione S Transferase Alpha 1 (GSTa1) Mouse Glutathione S Transferase Alpha 1 (GSTa1) For the quantitative determination of mouse (GSTa1) in serum, plasma and other biological fluids Cat. No. KT-16414 For Research Use Only. Not for use in

More information

Ustekinumab Drug Level ELISA

Ustekinumab Drug Level ELISA Ustekinumab Drug Level ELISA For the quantitative determination of free Stelara concentration in human serum and EDTA plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number:

More information

Mouse Cholecystokinin (CCK) ELISA Kit

Mouse Cholecystokinin (CCK) ELISA Kit Mouse Cholecystokinin (CCK) ELISA Kit Catalog No. CSB-E08115m (96T) This immunoassay kit allows for the in vitro quantitative determination of mouse CCK concentrations in serum, plasma and other biological

More information

VIBRATION METER. Instruction Manual.

VIBRATION METER. Instruction Manual. VIBRATION METER 840063 Instruction Manual CONTENTS Title Page I. Introduction... 2 II. Panel Description... 3 III. Operating Instructions A. General Measurement Procedures... 3~4 B. Peak... 5 C. Data Hold...

More information

Pig Vascular Endothelial Cell Growth Factor A (VEGF-A) ELISA

Pig Vascular Endothelial Cell Growth Factor A (VEGF-A) ELISA Pig Vascular Endothelial Cell Growth Factor A (VEGF-A) ELISA For the quantitative determination of pig VEGF-A in serum, plasma and other biological fluids. Cat. No. KT-32008 For Research Use Only. Not

More information

Instructor s Advance Preparation

Instructor s Advance Preparation INSTRUCTOR'S MANUAL Instructor s Advance Preparation This protocol is designed for 80 workstations of 4 students. Each group will prepare a set of standards, a blank, and 2 milk samples (can be a blind

More information

RIDASCREEN. 17ß-Östradiol. Enzymimmunoassay zur quantitativen Bestimmung von 17ß-Östradiol

RIDASCREEN. 17ß-Östradiol. Enzymimmunoassay zur quantitativen Bestimmung von 17ß-Östradiol RIDASCREEN 17ß-Östradiol Enzymimmunoassay zur quantitativen Bestimmung von 17ß-Östradiol Enzyme immunoassay for the quantitative analysis of 17ß-estradiol Art. No.: R2301 In vitro Test Lagerung bei 2-8

More information

Mouse Anti-OVA-IgE ELISA KIT

Mouse Anti-OVA-IgE ELISA KIT Mouse Anti-OVA-IgE ELISA KIT Research Reagent Please, read this instruction carefully before use. This is an ELISA (Enzyme Linked ImmunoSorbent Assay) kit for measurement of mouse anti-ova (ovalbumin)-ige

More information

Operation Manual. SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

Operation Manual. SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Operation Manual SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC. 4.18 Foreword

More information

PHS-25CW/3BW Benchtop ph Meter Instruction Manual

PHS-25CW/3BW Benchtop ph Meter Instruction Manual PHS-25CW/3BW Benchtop ph Meter Instruction Manual BANTE INSTRUMENTS CO., LTD PHS-W Series Benchtop ph Meter 1 Introduction Thank you for selecting the PHS-W series benchtop ph meter. This manual provides

More information

RIDASCREEN. Testosteron. Enzymimmunoassay zur quantitativen Bestimmung von Testosteron

RIDASCREEN. Testosteron. Enzymimmunoassay zur quantitativen Bestimmung von Testosteron RIDASCREEN Testosteron Enzymimmunoassay zur quantitativen Bestimmung von Testosteron Enzyme immunoassay for the quantitative analysis of testosterone Art. No.: R2401 In vitro Test Lagerung bei 2-8 C Storage

More information

CVS Calibration System CVS-KAL. Short Instruction Manual. System Configuration Description of the Equipment

CVS Calibration System CVS-KAL. Short Instruction Manual. System Configuration Description of the Equipment CVS Calibration System Short Instruction Manual System Configuration Description of the Equipment State: 11/21/2005 Table of contents 1. Documentation and Appropriate Usage 2. EC Conformity Declaration

More information

Golimumab Free Anti-Drug Antibody ELISA

Golimumab Free Anti-Drug Antibody ELISA Golimumab Free Anti-Drug Antibody ELISA For the semi-quantitative determination of free anti-drug antibodies against SIMPONI in human serum and EDTA plasma. For Research Use Only. Not For Use In Diagnostic

More information

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows Nitrate DOC316.53.01067 Cadmium Reduction Method Method 8192 0.01 to 0.50 mg/l NO 3 N (LR) Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Human anti-gliadin antibody (IgA)ELISA Kit

Human anti-gliadin antibody (IgA)ELISA Kit Human anti-gliadin antibody (IgA)ELISA Kit Catalog No. MBS701727 (96 tests) This immunoassay kit allows for the in vitro semi-quantitative determination of human anti-gliadin antibody(iga) concentrations

More information

Acceleration, Velocity, Separate probe VIBRATION METER Model : VB-8202

Acceleration, Velocity, Separate probe VIBRATION METER Model : VB-8202 Acceleration, Velocity, Separate probe VIBRATION METER Model : VB-8202 Your purchase of this VIBRATION METER marks a step forward for you into the field of precision measurement. Although this METER is

More information

Instruction Manual HI Refractometer for Invert Sugar Measurements

Instruction Manual HI Refractometer for Invert Sugar Measurements Instruction Manual HI 96804 Refractometer for Invert Sugar Measurements www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary information

More information

Mouse Insulin ELISA KIT S-Type

Mouse Insulin ELISA KIT S-Type Mouse Insulin ELISA KIT S-Type Research Reagent Please, read this instruction carefully before use. This is an ELISA (Enzyme Linked ImmunoSorbent Assay) kit for measurement of mouse insulin with high specificity

More information

Micro Volume QuEChERS kit

Micro Volume QuEChERS kit 225-37872 Sep. 2018 Small Capacity Pretreatment Kit Micro Volume QuEChERS kit Instruction Manual Read this manual thoroughly before you use the product. Keep this manual for future reference. This page

More information

Human anti-ige receptor antibody ELISA Kit

Human anti-ige receptor antibody ELISA Kit Human anti-ige receptor antibody ELISA Kit Catalog No. MBS702743 (96 T) This immunoassay kit allows for the in vitro semi-quantitative determination of human anti-ige receptor antibody concentrations in

More information

Rat Coagulation Factor V (F5) ELISA

Rat Coagulation Factor V (F5) ELISA Rat Coagulation Factor V (F5) ELISA For the quantitative determination of rat F5 in plasma Cat. No. KT-11028 For Research Use Only. Not for use in diagnostic procedures. Page 1 of 7 Rev. 1295911028 INTENDED

More information

PiCUS Calliper Version 3

PiCUS Calliper Version 3 Technology PiCUS Calliper Version 3 Digital calliper for precise geometry detection argus electronic GmbH Erich-Schlesinger-Straße 49d 18059 Rostock Germany Tel.: +49 (0) 381/49 68 14 40 www.argus-electronic.de

More information

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD PHS-25CW/3BW Benchtop ph Meter Instruction Manual BANTE INSTRUMENTS CO., LTD PHS-W Series Benchtop ph Meter 1 Introduction Thank you for selecting the PHS-W series benchtop ph meter. This manual provides

More information

Canine Erythropoietin,EPO ELISA Kit

Canine Erythropoietin,EPO ELISA Kit Canine Erythropoietin,EPO ELISA Kit Catalog No: E0028c 96 Tests Operating instructions www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH ENTIRE PROCEDURE

More information

Human papillomavirus,hpv ELISA Kit

Human papillomavirus,hpv ELISA Kit Human papillomavirus,hpv ELISA Kit Catalog No: E0787h 96 Tests Operating instructions www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH ENTIRE PROCEDURE

More information

Human Papillomavirus Antibody (IgG) ELISA Kit

Human Papillomavirus Antibody (IgG) ELISA Kit Human Papillomavirus Antibody (IgG) ELISA Kit Catalog No. CSB-E08782h (96 tests) This immunoassay kit allows for the in vitro semi-quantitative determination of human papillomavirus antibody(igg) concentrations

More information

EASYLYTE Na/K/Cl/Ca/Li SOLUTIONS PACK REF mL

EASYLYTE Na/K/Cl/Ca/Li SOLUTIONS PACK REF mL EASYLYTE Na/K/Cl/Ca/Li SOLUTIONS PACK REF 2124 800mL INTENDED USE The EasyLyte Na/K/Cl/Ca/Li Solutions Pack is intended for the quantitative determination of sodium (Na + ), potassium (K + ), chloride

More information

Polychlorinated Biphenyls (PCB) in Soil

Polychlorinated Biphenyls (PCB) in Soil Polychlorinated Biphenyls (PCB) in Soil DOC316.53.01107 Immunoassay 1 Method 10050 Scope and application: For soil. 1 This test is semi-quantitative. Results are shown as more or less than the threshold

More information

CHGA (Human) ELISA Kit

CHGA (Human) ELISA Kit CHGA (Human) ELISA Kit Catalog Number KA1884 96 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay... 3 General

More information

OPERATION MANUAL. 3 in 1, IR, Type K/J/R/E/T, Pt 100 ohm THERMOMETER. Model : TM-2000

OPERATION MANUAL. 3 in 1, IR, Type K/J/R/E/T, Pt 100 ohm THERMOMETER. Model : TM-2000 3 in 1, IR, Type K/J/R/E/T, Pt 100 ohm THERMOMETER Model : TM-2000 Your purchase of this THERMOMETER marks a step forward for you into the field of precision measurement. Although this THERMOMETER is a

More information

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual Page 1 / 11 VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual ZV-3030-0200-20 200 2-8 C Page 2 / 11 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST PRINCIPLE... 3

More information

Anti-Gliadin Antibody ELISA

Anti-Gliadin Antibody ELISA Anti-Gliadin Antibody ELISA For the quantitative determination of anti-gliadin antibodies in human stool. For Research Use Only. Not For Use In Diagnostic Procedures. Please read carefully due to critical

More information

Porcine Immunoglobulin E (IgE)ELISA Kit

Porcine Immunoglobulin E (IgE)ELISA Kit Porcine Immunoglobulin E (IgE)ELISA Kit Catalog No. MBS703293 (96T) This immunoassay kit allows for the in vitro quantitative determination of porcine IgE concentrations in serum and plasma,. Expiration

More information

Instructions for use 5-HIAA ELISA BA E-1900

Instructions for use 5-HIAA ELISA BA E-1900 Instructions for use ELISA BA E-1900 ELISA 1. Intended use and principle of the test Enzyme Immunoassay for the quantitative determination of 5-Hydroxy-3-Indole Acetic Acid () in urine. First, is quantitatively

More information

HBeAg and HBeAg Ab ELISA Kit

HBeAg and HBeAg Ab ELISA Kit HBeAg and HBeAg Ab ELISA Kit Cat. No.:DEIA3532 Pkg.Size:96T Intended use This kit is designed for detection of HBeAg and HBeAg Ab. General Description The solid phase of HBEAG AND HBEAG ANTIBODY ELISA

More information

Measurement method for the proficiency testing program

Measurement method for the proficiency testing program APLAC T088 Appendix Measurement method for the proficiency testing program Introductions This measurement method is prepared for use by the APLAC Proficiency Testing Program Photometric measurement of

More information

Cortisol (Horse) ELISA Kit

Cortisol (Horse) ELISA Kit Cortisol (Horse) ELISA Kit Catalog Number KA2298 96 assays Version: 05 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay... 3 General

More information

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows Potassium DOC316.53.01127 Tetraphenylborate Method Method 8049 0.1 to 7.0 mg/l K Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific information

More information

aa + bb cc + dd Equation 1

aa + bb cc + dd Equation 1 Experiment: The Determination of K eq for FeSCN 2+ Introduction For any reversible chemical reaction at equilibrium, the concentrations of all reactants and products are constant or stable. There is no

More information

BIOO FOOD AND FEED SAFETY. Histamine Enzymatic Assay Kit Manual. Catalog #: Reference #:

BIOO FOOD AND FEED SAFETY. Histamine Enzymatic Assay Kit Manual. Catalog #: Reference #: BIOO FOOD AND FEED SAFETY Histamine Enzymatic Assay Kit Manual Catalog #: 1032-05 Reference #: 1032-05 BIOO Scientific Corp. 2010 TABLE OF CONTENTS GENERAL INFORMATION... 1 Product Description... 1 Procedure

More information

HIPPURIC ACID and o-m-p- METHYLHIPPURIC ACIDS IN URINE BY UV CODE Z06010

HIPPURIC ACID and o-m-p- METHYLHIPPURIC ACIDS IN URINE BY UV CODE Z06010 HIPPURIC ACID and o-m-p- METHYLHIPPURIC ACIDS IN URINE BY UV CODE Z06010 Biochemistry The determination of Hippuric Acid ( HPA ) in urine is important to determine the risk due to the exposition to Toluene.

More information

INSTRUCTION MANUAL Milwaukee Refractometer MA881 Refractometer for Invert Sugar Measurements

INSTRUCTION MANUAL Milwaukee Refractometer MA881 Refractometer for Invert Sugar Measurements www.milwaukeeinst.com INSTRUCTION MANUAL Milwaukee Refractometer MA881 Refractometer for Invert Sugar Measurements www.milwaukeeinst.com 1 Instruction Manual MA881 Refractometer for Invert Sugar Measurements

More information

Porcine Factor-related Apoptosis(FAS) ELISA Kit

Porcine Factor-related Apoptosis(FAS) ELISA Kit Porcine Factor-related Apoptosis(FAS) ELISA Kit Catalog No. CSB-E09369p (96T) This immunoassay kit allows for the in vitro quantitative determination of porcine FAS concentrations in serum, plasma and

More information

Human anti-ganglioside IgG antibody (GM1-IgG) ELISA Kit

Human anti-ganglioside IgG antibody (GM1-IgG) ELISA Kit Human anti-ganglioside IgG antibody (GM1-IgG) ELISA Kit Catalog No. CSB-E09694h (96 tests) This immunoassay kit allows for the in vitro semi-quantitative determination of human GM1-IgG concentrations in

More information

Eppendorf UVette. Instructions for use

Eppendorf UVette. Instructions for use Instructions for use Instructions Eppendorf see Fig. Tab. p. p. UVette for use Copyright 2014 Eppendorf AG, Hamburg, Germany. All rights reserved, including graphics and images. No part of this publication

More information

Anti-Transglutaminase Antibody ELISA

Anti-Transglutaminase Antibody ELISA Anti-Transglutaminase Antibody ELISA For the quantitative determination of anti-transglutaminase antibodies in human stool. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number:

More information

Instruction Manual HI Refractometer for Sodium Chloride Measurements

Instruction Manual HI Refractometer for Sodium Chloride Measurements Instruction Manual HI 96821 Refractometer for Sodium Chloride Measurements www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully

More information

ISOCON-3 MAINS POWERED ISOLATING SIGNAL CONVERTER

ISOCON-3 MAINS POWERED ISOLATING SIGNAL CONVERTER ISOCON-3 MAINS POWERED ISOLATING SIGNAL CONVERTER CAUTION: This equipment is designed for connection to mains voltages and must be used in accordance with this guide. If it is not, the safety protection

More information

Hepatitis B virus IgM ELISA Kit

Hepatitis B virus IgM ELISA Kit Hepatitis B virus IgM ELISA Kit Catalog Number KA0289 96 assays Version: 05 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay...

More information

Monkey Kidney injury molecule 1,Kim-1 ELISA Kit

Monkey Kidney injury molecule 1,Kim-1 ELISA Kit Monkey Kidney injury molecule 1,Kim-1 ELISA Kit Catalog No: E0785Mo 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

Human Anti-Ovary Antibody (IgG)ELISA Kit

Human Anti-Ovary Antibody (IgG)ELISA Kit Human Anti-Ovary Antibody (IgG)ELISA Kit Catalog No. MBS703636 (96 tests) This immunoassay kit allows for the in vitro semi-quantitative determination of human Anti-Ovary Antibody(IgG) concentrations in

More information

Human placenta lactogen (HPL)ELISA Kit. MyBioSource.com. Catalog No. MBS (96T)

Human placenta lactogen (HPL)ELISA Kit. MyBioSource.com. Catalog No. MBS (96T) \ Human placenta lactogen (HPL)ELISA Kit Catalog No. MBS703864 (96T) This immunoassay kit allows for the in vitro quantitative determination of human HPL concentrations in serum, plasma. Expiration date

More information

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV Iron, Total DOC316.53.01338 FerroZine Method 1 Method 10264 1 to 100 µg/l Fe (10-cm cell) Reagent Solution Scope and application: For ultrapure water. 1 Adapted from Stookey, L.L., Anal. Chem., 42(7),

More information

Human hepatitis B virus surface antigen(hbsag) ELISA Kit

Human hepatitis B virus surface antigen(hbsag) ELISA Kit Human hepatitis B virus surface antigen(hbsag) ELISA Kit Catalog Number. CSB-E10089h For the qualitative determination of human hepatitis B virus surface antigen (HBsAg) concentrations in serum, plasma.

More information

Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM / / 2014

Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM / / 2014 Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM8004 80003223 / 00 05 / 2014 Contents 1 Preliminary note...3 1.1 Symbols used...3 2 Safety instructions...4 3 Functions and features...5

More information

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO Nitrate DOC316.53.01070 Dimethylphenol Method Method 10206 0.23 to 13. NO 3 N or 1.00 to 60.00 mg/l NO 3 (LR) TNTplus 835 Scope and application: For wastewater, drinking water, surface water and process

More information

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials Nitrogen, Total DOC316.53.01086 Persulfate Digestion Method Method 10071 0.5 to 25.0 mg/l N (LR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific

More information

HI Refractometer for Sodium Chloride Measurements INSTRUCTION MANUAL

HI Refractometer for Sodium Chloride Measurements INSTRUCTION MANUAL HI96821 Refractometer for Sodium Chloride Measurements INSTRUCTION MANUAL Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using this

More information

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials Nitrogen, Total DOC316.53.01085 Persulfate Digestion Method Method 10072 2 to 150 mg/l N (HR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific information

More information

INSTITUT FÜR ANGEWANDTE LABORANALYSEN GMBH. First-Beer Magnetic DNA Kit. Extraktion von Hefe- und Bakterien-DNA aus Bier und anderen Getränken

INSTITUT FÜR ANGEWANDTE LABORANALYSEN GMBH. First-Beer Magnetic DNA Kit. Extraktion von Hefe- und Bakterien-DNA aus Bier und anderen Getränken First-Beer INSTITUT FÜR ANGEWANDTE LABORANALYSEN GMBH First-Beer Magnetic DNA Kit Extraktion von Hefe- und Bakterien-DNA aus Bier und anderen Getränken Extraction of bacteria and yeast DNA from beer and

More information

MAGMA NOTE! 21+ USER MANUAL

MAGMA NOTE! 21+ USER MANUAL 21+ MAGMA USER MANUAL NOTE! Please do not put this product away or leave it for extended periods of time with low charge because if the battery falls below 2% then it will no longer accept a charge. Thank

More information

Sheep myeloperoxidase(mpo) ELISA kit

Sheep myeloperoxidase(mpo) ELISA kit Sheep myeloperoxidase(mpo) ELISA kit Catalog No. CSB-EL014757SH (96T) This immunoassay kit allows for the in vitro quantitative determination of sheep MPO concentrations in serum and plasma. Expiration

More information

Instructions for use. Tryptophan ELISA BA E-2700

Instructions for use. Tryptophan ELISA BA E-2700 Instructions for use ELISA BA E-2700 ELISA 1. Intended use and principle of the test Enzyme Immunoassay for the quantitative determination of in urine, serum, and plasma samples. After extraction and derivatization

More information

Beta Defensin 2 ELISA

Beta Defensin 2 ELISA Beta Defensin 2 ELISA For the quantitative determination of beta-defensin 2 in human stool. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 30-BD2HU-E01 Size: 96 Wells Version:

More information

IgD ELISA. For the quantitative determination of IgD in Human serum and plasma. Please See Appendix A for Reference Serum Information

IgD ELISA. For the quantitative determination of IgD in Human serum and plasma. Please See Appendix A for Reference Serum Information IgD ELISA For the quantitative determination of IgD in Human serum and plasma. Please See Appendix A for Reference Serum Information For Research Use Only. Not for Use in Diagnostic Procedures. Catalog

More information

Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats

Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats Application Note 24 Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats INTRODUCTION Most spectrophotometric

More information

IgD ELISA. For the quantitative determination of IgD in Human Sera. Please See Appendix A for Reference Serum Information

IgD ELISA. For the quantitative determination of IgD in Human Sera. Please See Appendix A for Reference Serum Information IgD ELISA For the quantitative determination of IgD in Human Sera. Please See Appendix A for Reference Serum Information For Research Use Only. Not for Use in Diagnostic Procedures Catalog Number: 41-IGDHU-E01

More information

Human placenta lactogen, HPL ELISA kit

Human placenta lactogen, HPL ELISA kit Human placenta lactogen, HPL ELISA kit Catalog No. E1442h (96 tests) Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH ENTIRE

More information

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Cadion Method Method to 0.30 mg/l Cd TNTplus 852 Cadmium DOC316.53.01013 Cadion Method Method 10217 0.02 to 0.30 mg/l Cd TNTplus 852 Scope and application: For wastewater and process control. Test preparation Instrument-specific information Table 1 shows

More information

Human rheumatoid factor (RF) antibody (IgM) ELISA Kit

Human rheumatoid factor (RF) antibody (IgM) ELISA Kit Human rheumatoid factor (RF) antibody (IgM) ELISA Kit Catalog Number. CSB-EQ027654HU For the qualitative determination of human rheumatoid factor (RF) antibody (IgM) concentrations in serum, plasma. This

More information

Human von Willebrand Factor cleavingprotease(vwf-cp) ELISA Kit. Catalog No. MBS (96T)

Human von Willebrand Factor cleavingprotease(vwf-cp) ELISA Kit. Catalog No. MBS (96T) Human von Willebrand Factor cleavingprotease(vwf-cp) ELISA Kit Catalog No. MBS705447 (96T) This immunoassay kit allows for the in vitro quantitative determination of human vwf-cp concentrations in serum,

More information

INSTRUCTION MANUAL. Lowry Assay Kit. Kit for Protein Quantification. (Cat. No )

INSTRUCTION MANUAL. Lowry Assay Kit. Kit for Protein Quantification. (Cat. No ) INSTRUCTION MANUAL Lowry Assay Kit Kit for Protein Quantification (Cat. No. 39236) SERVA Electrophoresis GmbH Carl-Benz-Str. 7 D-69115 Heidelberg Phone +49-6221-138400, Fax +49-6221-1384010 e-mail: info@serva.de

More information

Human Paraneoplastic pemphigus (PNP)antibody ELISA Kit. Catalog No. MBS (96 T)

Human Paraneoplastic pemphigus (PNP)antibody ELISA Kit. Catalog No. MBS (96 T) Human Paraneoplastic pemphigus (PNP)antibody ELISA Kit Catalog No. MBS704248 (96 T) This immunoassay kit allows for the in vitro quantitative determination of human Paraneoplastic pemphigus(pnp)antibody

More information

Automatic titrators, TitroMatic

Automatic titrators, TitroMatic 112 TitroMatic These are the fourth generation of CRISON titrators. CRISON has over 25 years experience in this field and knows that, for certain applications, the performance and cost of large titrator

More information

Application Note for Gentian Cystatin C Immunoassay on. Cobas c501 1 (Optimised Volumes II), Roche Diagnostics 1 v06-september 2010

Application Note for Gentian Cystatin C Immunoassay on. Cobas c501 1 (Optimised Volumes II), Roche Diagnostics 1 v06-september 2010 Page of 6 Application Note for Gentian Cystatin C Immunoassay on For in vitro diagnostic use only Cobas c50 (Optimised Volumes II), Roche Diagnostics v06-september 200 General Information Intended Use

More information

Mouse Retinol Binding Protein 4 ELISA

Mouse Retinol Binding Protein 4 ELISA Mouse Retinol Binding Protein 4 ELISA For the quantitative determination of RBP4 in mouse serum and plasma. Please see Appendix A for Reference Serum Information. For Research Use Only. Not For Use In

More information