1 A Highly Stable Alkyl Amide Silica-Based Column Packing for Reversed-Phase HPLC of Polar and Ionizable Compounds Researchers have designed a new polar-linked, reversed-phase column for separating polar, ionizable, and especially highly basic compounds with excellent column efficiency and peak shapes. The column combines an ultrapure, low-acidity porous silica support with a unique stationary phase that has a polar, embedded amide group linking the sterically protecting diisopropyl groups with a C14 n-alkyl functionality. This new column demonstrates stability in low-ph mobile phases because of its resistance to hydrolysis and stationary-phase loss. Dense triple-endcapping minimizes unwanted silanol interactions and allows excellent peak shapes and stability at intermediate ph. The column can be used with totally aqueous mobile phases without the phase collapse typical of alkyl stationary phases. J.J. Kirkland, J.W. Henderson, J.D. Martosella, B.A. Bidlingmeyer, J. Vasta-Russell*, and J.B. Adams Jr. Hewlett-Packard Co., Little Falls Analytical Division, Newport Site, 538 First State Boulevard, Newport, Delaware * Mac-Mod Analytical, 127 Commons Court, Chadds Ford, Pennsylvania Adams Research, 759 Morris Road, Hockessin, Delaware Address correspondence to J.J. Kirkland. C olumn packings with alkyl stationary phases containing polar-linked groups or embedded polar groups are becoming increasingly popular for reversedphase high performance liquid chromatography (HPLC) separations (1 4). Columns of these packings often exhibit good peak shapes and efficiencies for polar solutes, especially ionizable compounds that can be particularly troublesome in separations (5,6). Compared with hydrocarbon-only stationary phases, the stationary phases with polar groups linked in an alkyl matrix are less retentive and typically require a lower concentration of organic mobile-phase modifier (1,2,7). In some cases, chromatographers can separate certain ionizable compounds in purely aqueous mobile phases without using ion-pair reagents (8). Although column packings with polar groups provide advantages for certain separations, these materials have a potential disadvantage that has not been investigated carefully. Bonded-phase packings that have shorter chains or polar groups exhibit poorer stability than long-chain purely alkyl functionalities in low and high ph environments (9 12). Therefore, packings with shorter stationary-phase chains or more-polar functional groups can provide performance improvements for separating highly polar compounds but come with the price of shorter column life. Users wanted a better approach for increasing column stability that maintained the desirable features for separating polar compounds. We have developed a new silane-based, embedded polar group stationary phase that uses sterically protecting groups. This structure greatly enhances the stability of the bonded phase in low-ph applications by significantly decreasing the loss of stationary phase through hydrolysis (9,13,14). To enhance performance and stability in intermediate-ph applications, this stationary phase is densely triple-endcapped. This feature results in a packing material that resists dissolution of the silica support, which is the degradation process usually responsible for failure of bonded-phase columns in intermediate- and high-ph applications (11,15). This new packing material is based on Hewlett-Packard s Zorbax Rx-Sil ultrapure, low-acidity Type B porous silica support (Wilmington, Delaware), which forms the basis for a large family of widely used, high performance reversed-phase columns (16). The combination of this silica support with the new alkyl amide stationary phase results in a robust column packing that exhibits superior characteristics for separating difficult polar and ionizable compounds. EXPERIMENTAL Apparatus and reagents: We performed chromatographic tests using HP 1050, HP 1090, and HP 1100 chromatographs (Hewlett-Packard). We calculated plate numbers using the halfpeak-height method described by equation 2.8a in reference 17. Peak tailing values were determined at 5% of the peak height using the conventional approach (17). Samples were injected with a Rheodyne model 7125 sampling valve (Rohnert Park, California). The 15 cm 0.46 cm columns were prepared at the Hewlett-Packard Newport Site using a conventional slurry-packing method (18). The porous silica support used in this work is a Type B silica formed by aggregating ultrapure silica sols (16). Type B silicas are the newer highly purified, less acidic chromatographic supports that provide superior separations, especially for ionizable compounds (19). The physical and surface properties of this silica have been reported elsewhere (16). We performed the diisopropyl C14 amide (alkyl amide) polar-linked stationary-phase bonding and triple-endcapping procedures
2 FIGURE 1: Structure of the diisopropyl alkyl amide stationary phase. FIGURE 2: Diffuse reflectance FTIR spectra of alkyl amide packing (solid line) and starting silica (dashed line). on a 5- m d p, 80-Å, 180-m 2 /g Zorbax Rx-Sil silica support (Hewlett-Packard) by proprietary processes. The carbon values for this column packing typically were 10% for this support. The other experimental columns used in this study also were prepared at the Newport site. Columns of the endcapped alkyl amide packing are now available from Hewlett-Packard under the trade name, Zorbax Bonus-RP. We obtained HPLC-grade solvents from EM Science (Gibbstown, New Jersey). Test solutes from Sigma Chemical Co. (St. Louis, Missouri) were used as received. Stationary-phase characteristics: Figure 1 depicts the sterically protected alkyl amide stationary phase of the new column. The 10% carbon content for the alkyl amide packing represents approximately 2.1 mol/m 2 of the bonded phase, which is typical for dense coverage by sterically protected bonded silanes (20). Figure 2 shows reflectance infrared spectra of the bonded alkyl amide silica material and the starting silica of this packing. The spectrum for the starting silica (dashed line) suggests unbonded (free) silanols at 3739 cm 1, associated (bonded) silanols at 3645 cm 1, and silanols and adsorbed water at 3477 cm 1. The bonded material (solid line) shows no silanol adsorption at 3739 cm 1, which suggests that the starting unbonded silanols are fully reacted. Characteristic carbon hydrogen stretching bands for CH 2 and CH 3 groups are present at 2853 and 2925 cm 1 and 2890 and 2970 cm 1, respectively. Figure 2 also indicates an overtone band for an amide II structure at 3083 cm 1 and a secondary amide (unassociated) N H stretching band at 3450 cm 1. The broad 3314 cm 1 band likely represents N H stretching from the amide group and associated silanols and adsorbed water. The characteristic 1649-cm 1 amide I band for the alkyl amide stationary phase clearly is visible over the silica absorption in this spectrum, as are other secondary amide bands at 1542 and 1650 cm 1. In short, the infrared spectrum for the alkyl amide material is in qualitative agreement with the proposed structure of Figure 1. Column aging studies: Individual columns were purged continuously with either a low- or intermediate-ph mobile phase. Low-pH aging was performed with a 1.0-mL/min flow of 50:50 (v/v) methanol 0.1% trifluoroacetic acid at 60 C. Intermediate-pH aging studies were conducted by continuously purging the columns with a 1.5-mL/min flow of 40:60 (v/v) acetonitrile M sodium phosphate buffer (ph 7.0) at 24 C. Periodically, we tested these columns chromatographically at 40 C using a mixture of amitriptyline, trimipramine, and toluene with a mobile phase of 60:40 (v/v) acetonitrile 0.01 M sodium phosphate buffer (ph 7.0) at a 1.5- ml/min flow rate. The mobile phase ph values refer to the ph of the buffer used in the experiments. We periodically tested the columns with solutes to determine the level of changes that occurred during use. Each column first was flushed with at least 20 volumes of chromatographic mobile phase before the chromatographic test. Other apparatus and reagents: Data for the basic drug separations were obtained with an HP model 1090 chromatograph with UV detection at 254 nm. The mobile phases were methanol 10 mm sodium citrate buffers. We chose the citrate buffer because of its wide buffering range (ph ). We collected and analyzed data using HP ChemStation software (Hewlett-Packard). The captions for each figure provide the experimental details for each separation shown in this study. We measured diffuse reflectance infrared spectra using a Nicolet Avatar model 380 FTIR spectrometer (Madison, Wisconsin). We made no special attempt to maintain the dryness of the samples used in these tests. RESULTS AND DISCUSSION Low-pH studies: Previous studies with sterically protecting bulky groups on the silane silicon atom showed that stationary-phase stability at low ph is greatly enhanced by the much-decreased hydrolysis rate of the silane-attaching siloxane bond (9,13,14). This characteristic also is found for the alkyl amide packing of this study, because the diisopropyl groups of the structure (Figure 1) afford the usual steric protection against low-ph hydrolysis. Figure 3 shows that the rate of change for the alkyl amide column when continuously purged with a methanol 0.1% trifluoroacetic acid mobile phase (approximately ph 1.9) at 60 C is much less than that for a nonsterically protected double-endcapped dimethyl C18 amide column prepared with the same silica support and by a comparable reaction. In Figure 3a, the diisopropyl C14 amide column shows only a slight increase in retention factor (k) for
3 amitriptyline, a strongly basic drug (pk a 9.5). Retention values increase steeply with aging for the comparable dimethyl C18 amide column. Presumably, the k value increase is the result of a loss in silane stationary phase by hydrolysis with subsequent formation of silanol groups that bind the basic drug. Values of toluene k for the dimethyl C18 amide column show a decrease of approximately 40% for the period of testing. Figure 3b shows that the amitriptyline plate height for the diisopropyl C14 amide column is essentially unchanged (within experimental error) after purging with approximately 37,000 column volumes, when the test arbitrarily was terminated. This result equates to almost four months of 8-h/day operation. In contrast, the comparable nonsterically protected dimethyl C18 amide column showed significant degradation in column efficiency for both amitriptyline and toluene after purging with approximately 2000 column volumes under these conditions. Low-pH separations with the diisopropyl C14 amide column produce excellent peak shapes and column efficiencies for both acidic and basic compounds. Figure 4 shows the separation of a mixture of acidic fruit acids of interest to the food industry using a methanol trifluoroacetic acid mobile phase. This column also exhibits useful characteristics for basic compounds, as illustrated in Figure 5 for the separation of basic cardiac drugs using a methanol sodium dihydrogen phosphate (ph 3.0) mobile phase. In both cases, column efficiency and peak shapes generally are superior to those found for columns with simple alkyl stationary phases (for example, C18). Intermediate-pH studies: Columns with polar-linked groups often are used in the ph 4 8 range for separating polar or ionizable compounds. Under some separating conditions, particularly at intermediate ph, columns with conventional alkyl-bonded phases can show less-than-desirable peak shapes and column efficiencies for some compounds. In these instances, O Gara and co-workers (2) speculated that the superior performance of polar-linked group columns may be the result of the association of the embedded polar groups with unreacted silanol groups on the surface of the silica support. An alternative postulation is that the linked-polar group reduces the hydrophobicity of the stationary phase near the silica support surface, allowing water molecules to approach and deactivate unreacted silanol groups. These effects would minimize unwanted interactions that cause poorer column performance, and they would result in better peak shapes and column efficiency for polar and ionizable compounds. Compared with conventional alkyl-bondedphase columns, the alkyl amide column of our FIGURE 3: Plots of (a) retention factor and (b) plate height versus column volumes of mobile-phase purge for alkyl amide columns at ph 2. Columns: 150 mm 4.6 mm; aging mobile phase: 50:50 (v/v) methanol aqueous (0.1%) trifluoroacetic acid (ph 1.9); flow rate: 1.0 ml/min; temperature: 60 C. Mobile phase for the amitriptyline test: 60:40 acetonitrile 0.01 M sodium phosphate buffer (ph 7.0); flow rate: 1.5 ml/min; temperature: 40 C. Mobile phase for the toluene test: 80:20 methanol water; flow rate: 1.0 ml/min; temperature: 23 C. dimethyl C18 amide, amitriptyline; diisopropyl C14 amide, amitriptyline; dimethyl C18 amide; diisopropyl C14 amide, toluene. study usually produces superior peak shapes and column efficiencies for difficult ionizable compounds when operated with intermediateph mobile phases. Figure 6 shows the separation of a mixture of highly basic drugs on our alkyl amide column, an endcapped alkyl C18 column, and a nonendcapped alkyl C8 column, all of which were prepared with the same silica support. Solute retention is lower on the alkyl amide column, but the peak shapes and column efficiency clearly are superior under the same ph 6 conditions. Figure 7 illustrates the improved performance of the alkyl amide stationary phase compared with conventional alkyl phases. In this figure the separation of the strongly basic drug amitriptyline (pk a 9.4) is performed on three columns an endcapped alkyl amide column, an endcapped dimethyl C18 column, and a nonendcapped C8 column at ph 5. At this ph, amitriptyline should be largely protonated. All columns were prepared with the same silica support. Two important features arise from this comparison. First, the alkyl amide column is less retentive, as in Figure 6, presumably because of the reduced hydrophobicity caused by the polar group in the stationary phase. However, another reason may be the greater interaction of the solute with surface silanol groups for the two pure alkyl columns. The second feature of this comparison is the broader, tailing peaks for the alkyl columns; the nonendcapped column shows the largest effect. Clearly, the alkyl amide column provides superior performance under these separation conditions. Figure 8 shows the greatly reduced interaction of surface silanols when using the alkyl amide column for protonated or partially protonated basic compounds at intermediate ph. Peak-tailing factors for a series of basic compounds remain essentially constant throughout
4 the ph 3 6 range for a methanol citric acid buffer mobile phase. In contrast, peak-tailing values for an endcapped C18 column and a nonendcapped C8 column generally are higher and increase with ph increases. These results suggest less interaction of the alkyl amide column for these protonated basic compounds with increasingly ionized silanol groups as the ph increases. We found that the alkyl amide column produced better peak shapes and column efficiencies for polar ionizable compounds in the intermediate ph range where ionizable solutes and unreacted silanol groups on the silica support often are at least partially ionized. Previous studies have shown that column packings that were exhaustively endcapped show superior stability in intermediate- and high-ph mobile phases (10,20). This effect apparently is the result of increased protection of the silica support surface from dissolution, the typical cause of silica-based column failure at intermediate and high ph. To test the stability of the diisopropyl C14 amide column at intermediate ph, we continuously purged a column of this material with 40:60 (v/v) acetonitrile 25 mm sodium phosphate buffer (ph 7.0) at ambient temperature. Periodically, we tested this column chromatographically, with the results shown in Figure 9. For both toluene (Figure 9a) and amitriptyline (Figure 9b), plate height values show little change for the diisopropyl C14 amide column after purging with more than 36,000 column volumes, after which the experiment was terminated arbitrarily. This result suggests that the column underwent little change after the use equivalent of more than three months of 8-h workdays. We believe that the exhaustive endcapping method used in preparing the diisopropyl C14 amide column provided superior protection for the silica support from dissolution by the ph 7 mobile phase. Retention values for the basic solute amitriptyline increased approximately 35% after the purging study, indicating the formation of more acidic silanol groups on the silica support surface caused by slow silica hydrolysis. The alkyl amide column can be operated routinely at intermediate ph for separating a wide range of polar and ionizable compounds. Figure 10 shows the separation of a mixture of benzodiazepine drugs using a ph 4.6 buffer with good peak shapes and column efficiencies for all solutes. Bromazepam (pk a 2.9, 11.0), clobazam (pk a unavailable), lorazepam (pk a 1.3, 11.5), and diazepam (pk a 3.3) compose this mixture, which has a wide range of protonated (ionized) and nonionized structures at the separation ph. Figure 11 shows the separation of cephalosporin drugs at ph 7.0, again with superior peak shapes and column efficiency. These compounds have pk a values ranging from 2.2 to FIGURE 4: Separation of fruit acids at low ph. Column: 150 mm 4.6 mm diisopropyl C14 amide; mobile phase: 70:30 (v/v) aqueous (0.1%) trifluoroacetic acid methanol; flow rate: 1.0 ml/min; temperature: 24 C; detection: UV absorbance at 254 nm; sample volume: 5 L. Peaks: 1 gallic acid, 2 protocatechuic acid, 3 syringic acid, 4 vanillic acid (N 7500, tailing factor 1.05), 5 gentisic acid. FIGURE 5: Separation of basic cardiac drugs at low ph. Mobile phase: 55:45 (v/v) methanol M sodium phosphate buffer (ph 3.0); sample volume: 3 L. Other conditions were the same as in Figure 4. Peaks: 1 diltiazem, 2 dipyrimidole, 3 nifedipine, 4 lidoflazine, 5 flunarizine (N 4900, tailing factor 1.08). 7.3, and some compounds, such as cefalexin, have multiple pk a values. As a result of the linked polar amide group of the alkyl amide structure, this stationary phase is readily wetted by totally aqueous mobile phases. Consequently, analysts can separate compounds that require little or no organic modifiers for proper retention. Figure 12 shows an example of this effect in which a mixture of nucleic acid bases and related compounds were separated with a simple aqueous sodium acetate mobile phase. Studies showed no change in retention and column efficiency properties for this column when the mobile phase was adjusted from partially organic to totally aqueous mobile phases and back again; equilibration also is very rapid. These results indicate that this stationary phase does not collapse when used with low-organic-content or totally aqueous mobile phases.
5 FIGURE 7: Peak characteristics of a basic drug at ph 5 for different columns. Same conditions as for Figure 6, except mobile phase: citrate buffer (ph 5.0); solute: amitriptyline. Traces: 1 endcapped alkyl amide (N 6500, tailing factor 1.04), 2 endcapped C18 (N 4800, tailing factor 2.32), 3 nonendcapped C8 (N 3100, tailing factor 4.05). FIGURE 6: Separation of basic drugs at ph 6 for (a) endcapped diisopropyl C14 amide, (b) endcapped C18, and (c) nonendcapped C8 columns. Column dimensions: 150 mm 4.6 mm; mobile phase: 62:38 (v/v) methanol 0.01 M citrate buffer (ph 6.0); flow rate: 1.0 ml/min; temperature: 30 C; detection: UV absorbance at 254 nm. Peaks: 1 uracil, 2 propranolol, 3 nortriptyline, 4 doxepin, 5 amitriptyline, 6 trimipramine. CONCLUSION A silica-based column with a sterically protected, triple-endcapped C14 stationary phase with a linked polar amide group showed significant promise for separating a wide range of polar and ionizable compounds with good peak shapes and column efficiency. Because of the presence of protecting diisopropyl groups on the silane silicon atom, this column demonstrates outstanding stability at low ph. The column also exhibits excellent stability with intermediate-ph mobile phases because of the exhaustive endcapping that inhibits degradation of the column, typically caused by dissolution of the silica support. Based on previous experiences, monofunctional attachment of the silanebased stationary phase should provide good preparation reproducibility. The excellent kinetic properties (column efficiency and peak shape) we found for the bidentate stationary phase is typical of those of monofunctional structures. ACKNOWLEDGMENTS We thank C.H. Dilks Jr. of the Hewlett-Packard Co. (Newport, Delaware) for experimental data from column aging studies. REFERENCES (1) T. Czajkowska and M. Jaroniec, J. Chromatogr. A 762, (1997). FIGURE 8: Effect of ph on peak tailing factors for basic drugs with (a) alkyl amide, (b) endcapped C18, and (c) nonendcapped C8 columns. Conditions were the same as in Figure 6 except for citrate buffers with various ph values. Analytes: propranolol, nortriptyline, doxepin, amitriptyline, trimipramine. FIGURE 9: Plots showing the aging of an alkyl amide column at ph 7.0 with (a) toluene and (b) amitriptyline. Column: 150 mm 4.6 mm diisopropyl C14 amide; mobile phase: 40:60 (v/v) acetonitrile M sodium phosphate buffer (ph 7.0); flow rate: 1.5 ml/min; temperature: 24 C. FIGURE 10: Separation of benzodiazepine drugs at intermediate ph. Mobile phase: 55:45 (v/v) methanol 0.05 M sodium acetate buffer (ph 4.6); sample volume: 1 L. Other conditions were the same as in Figure 4. Peaks: 1 bromozepam, 2 clobazam, 3 lorazepam, 4 diazepam (N 7700, tailing factor 0.99).
6 FIGURE 11: Separation of basic cephalosporins at ph 7.0. Mobile phase: 20:80 acetonitrile M sodium phosphate buffer (ph 7.0). Other conditions were the same as in Figure 4. Peaks: 1 cefalexin, 2 cefaclor, 3 cefoxitin (N 7000, tailing factor 1.07). (2) J.E. O Gara, B.A. Alden, T.H. Walter, J.S. Petersen, C.L. Niederlaender, and U.D. Neue, Anal. Chem. 67, (1995). (3) T.L. Ascah and B. Feibush, J. Chromatogr. 506, 357 (1990). (4) T.L. Ascah, K.M.R. Kallury, C.A. Szafranski, S.D. Scott, and F. Liu, J. Liq. Chromatogr. Relat. Technol. 19, (1996). (5) T. Czajkowska, I. Hrabovsky, B. Buszewski, R.K. Gilpin, and M. Jaroniec, J. Chromatogr. A 691, (1995). (6) N. Fukunaga and T. Kaneko, Chromatography 18, (1997) (Japanese). (7) D.V. McCalley, J. Chromatogr. A 738, (1996). (8) M. Przybyciel and M.A. Santangelo, The Development of Hydrocarbon-Based Stationary Phase for HPLC with Highly Aqueous Mobile Phases, poster presented at the Chromatography Forum of the Delaware Valley, Media, Pennsylvania, June (9) J.J. Kirkland, J.L. Glajch, and R.D. Farlee, Anal. FIGURE 12: Separation of nucleic acid bases with a totally aqueous mobile phase. Column: 150 mm 4.6 mm alkyl amide; mobile phase: M sodium acetate (ph 4.6); flow rate: 1.0 ml/min; temperature: 24 C; detection: UV absorbance at 254 nm. Peaks: 1 cytosine, 2 5-fluorocytosine, 3 uracil, 4 5-fluorouracil, 5 thymine (N 10,700, tailing factor 1.07), 6 adenine. Chem. 61, 2 11 (1989). (10) J.J. Kirkland, J.W. Henderson, J.J. DeStefano, M.A. van Straten, and H.A. Claessens, J. Chromatogr. A 762, (1997). (11) J.J. Kirkland, M.A. van Straten, and H.A. Claessens, J. Chromatogr. A 691, 3 19 (1995). (12) J.J. Kirkland, M.A. van Straten, and H.A. Claessens, J. Chromatogr. A 797, (1998). (13) J.L. Glajch and J.J. Kirkland, LC GC 8(2), (1990). (14) L.R. Snyder, J.J. Kirkland, and J.L. Glajch, Practical HPLC Method Development (John Wiley & Sons, New York, 2nd ed., 1997), pp (15) H.A. Claessens, M.A. van Straten, and J.J. Kirkland, J. Chromatogr. A 728, (1996). (16) J.J. Kirkland, C.H. Dilks Jr., and J.J. DeStefano, J. Chromatogr. 635, (1993). (17) L.R. Snyder, J.J. Kirkland, and J.L. Glajch, Practical HPLC Method Development (John Wiley & Sons, New York, 2nd ed., 1997), p. 42. (18) L.R. Snyder and J.J. Kirkland, Introduction to Modern Liquid Chromatography (John Wiley & Sons, New York, 2nd ed., 1979), pp (19) J. Köhler and J.J. Kirkland, J. Chromatogr. 385, (1987). (20) J.J. Kirkland and J.W. Henderson, J. Chromatogr. Sci. 32, (1994).. Copyright Notice Copyright by Advanstar Communications Inc. Advanstar Communications Inc. retains all rights to this article. This article may only be viewed or printed (1) for personal use. User may not actively save any text or graphics/photos to local hard drives or duplicate this article in whole or in part, in any medium. Advanstar Communications Inc. home page is located at
Stability of Silica-Based, Monofunctional C 18 Bonded- Phase Column Packing for HPLC at High ph
Stability of Silica-Based, Monofunctional C 18 Bonded- Phase Column Packing for HPLC at High ph J.J. Kirkland Rockland Technologies, Inc., 538 First State Blvd., Newport, Delaware 19804 Abstract Previous
More informationProntoSIL C18-EPS Reversed-Phase HPLC Columns
ProntoSIL C8-EPS Reversed-Phase HPLC Columns Provides excellent separation of polar compounds Better peak shape for acids and bases Stabilized bonded phase for rugged, robust HPLC methods More retentive
More informationLC and LC/MS Column Selection Flow Chart
LC and LC/MS Column Selection Flow Chart To use the column selection diagram below, simply follow the path for your analyte and mobile phase. At the far right, follow your final column selection to the
More informationpenta-hilic UHPLC COLUMNS
penta-hilic UHPLC COLUMNS penta-hilic Highly retentive, proprietary penta-hydroxy-ligand Excellent peak shape for polar compounds with a variety of functional groups: acids, bases, zwitterions strong and
More informationThe Retention Behavior of Reversed-Phase HPLC Columns with 100% Aqueous Mobile Phase
9 LCGC NORTH AMERICA VOLUME NUMBER OCTOBER www.chromatographyonline.com The Retention Behavior of Reversed-Phase HPLC Columns with % Aqueous Mobile Phase The authors studied the retention behavior of reversed-phase
More informationWilliam E. Barber, Ph.D. Applications Chemist October
William E. Barber, Ph.D. Applications Chemist ctober 0 00 Secrets of Good Peak Shape in HPLC Time: :00 p.m. CET Telephone Number: + 44 0 76 088 Chairperson: John Vis The Secrets of Good Peak Shape in HPLC
More informationHypersil BDS Columns TG 01-05
TG 0-0 Hypersil BDS Columns Introduction Hypersil BDS columns have gained a reputation over the years as one of the most robust, reproducible and reliable HPLC column brands available. This Technical Guide
More informationChoosing Columns and Conditions for the Best Peak Shape
Choosing Columns and Conditions for the Best Peak Shape What is Good Peak Shape and Why is it Important? Good peak shape can be defined as a symmetrical or gaussian peak and poor peak shape can include
More informationToo Polar for Reversed Phase What Do You Do?
Too Polar for Reversed Phase What Do You Do? June 20, 2013 Mark Powell Columns and Consumables Technical Support Page 1 C8 or C18 Doesn t Always Do the Job Typical reversed phase conditions involve water/buffer
More informationPackings for HPLC. Packings for HPLC
Summary of packings for HPLC In analytical HPLC, packings with particle sizes of 3 to 10 µm are preferred. For preparative separation tasks, also particles with diameters larger than 10 µm are applied.
More informationpenta-hilic UHPLC COLUMNS
penta-hilic UHPLC COLUMNS Highly retentive, proprietary penta-hydroxy-ligand penta-hilic Excellent peak shape for polar compounds with a variety of functional groups: acids, bases, zwitterions strong and
More informationC18 Column. Care & Use Sheet
C18 Column Care & Use Sheet HALO Description HALO C18 is a high-speed, high-performance liquid chromatography column based on a new Fused-CoreTM particle design. The Fused-Core particle provides a thin
More informationProntoSIL HPLC Columns
ProntoSIL HPLC Columns For Technical Assistance Call -800--708 analytical inc. To Order and For More Information: Call FAX -800--708-0-8-99 -0-8-99 We ship nearly 00% of your orders the same day they are
More informationProduct Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns
Product Bulletin HPLC Columns ACE LC/MS and Rapid Analysis HPLC Columns 0 mm, 30 mm, 35 mm and 50 mm column lengths.0,., 3.0, 4.0 and 4.6 mm column diameters Configured for High Sample Throughput Specially
More informationCOMPARISON GUIDE. to C18 Reversed Phase HPLC Columns. Comparison Data on 60 Commonly Used C18 Phases. Stationary Phase Specifications
COMPARISON GUIDE to C18 Reversed Phase HPLC Columns Comparison Data on 60 Commonly Used C18 Phases Stationary Phase Specifications Phases Compared According to Relative Hydrophobicity Phases Compared According
More informationChemistry Instrumental Analysis Lecture 28. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 28 High Performance Liquid Chromatography () Instrumentation Normal Phase Chromatography Normal Phase - a polar stationary phase with a less polar mobile phase.
More informationP R O D U C T B U L L E T I N. Fused-Core particle technology for hyper-fast and super-rugged HPLC columns
P R O D U C T B U L L E T I N Fused-Core particle technology for hyper-fast and super-rugged HPLC columns HALO column packings are not made the typical way. Instead, the particles packed into HALO columns
More informationcolumns Acclaim Mixed-Mode WCX-1 for Separating Basic Molecules
columns Acclaim Mixed-Mode WCX- for Separating Basic Molecules The Acclaim Mixed-Mode WCX- is a novel, high-efficiency, silica-based column specially designed for separating various basic analytes. This
More informationImprove Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application
Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application Pharmaceuticals Authors John W. Henderson Jr., Nona Martone, and Cliff Woodward
More informationHybrid Organic Inorganic Particle Technology: Breaking Through Traditional Barriers of HPLC Separations
6 LCGC VLUME 8 NUMBER NVEMBER 000 www.chromatographyonline.com Hybrid rganic Inorganic Particle Technology: Breaking Through Traditional Barriers of HPLC Separations Hybrid organic inorganic particles
More informationTips and tricks to increase HPLC column lifetime. Dr. Frank Michel
Tips and tricks to increase HPLC column lifetime Dr. Frank Michel Overview How to influence HPLC column life time? HPLC method Helpful accessories Best practice 2 HPLC Method Mobile Phase Mobile phase
More informationThe Effect of Contact Angle and Wetting on Performance in Solid Phase Extraction
The Effect of Contact Angle and Wetting on Performance in Solid Phase Extraction Edouard S. P. Bouvier, Randy E. Meirowitz and Uwe D. Neue Waters Corporation, 34 Maple Street, Milford, MA 01757 USA Presented
More informationLuna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency
Luna 2.5 µm C18(2)-HST Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency Table of Contents Part 1 Theory 1.1 Abstract...3 1.2 Introduction...3 Part 2 Set Up 2.1
More informationAcclaim Mixed-Mode WCX-1
Acclaim Mixed-Mode WCX-1 Product Manual for the Acclaim Mixed-Mode WCX-1 Column Page 1 of 28 PRODUCT MANUAL for the Acclaim Mixed-Mode WCX-1 Columns 4.6 x 150 mm, P/N (068353) 4.6 x 250 mm, P/N (068352)
More informationHigh Performance Liquid Chromatography
Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN
More informationHigh Performance Liquid Chromatography
Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 1 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN
More informationHPLC Column Classification
HPLC Column Classification Introduction 1978 USP XIX Fourth supplement : L1 designation for C18 column 1980 USP XX : 7 columns were classified -Classified HPLC column to 56 descriptions -More than 220
More informationNew Product Update. HPLC Columns
New Product Update HPLC Columns ZORBAX MicroBore HPLC Columns ZORBAX Extend-C18 Columns ZORBAX Bonus-RP Columns Inertsil ODS-2 Cartridge Columns ZORBAX Carbohydrate Analysis Columns ZORBAX Eclipse dsdna
More informationExceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens
Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-exyl Columns to Separate Estrogens Application Note Pharmaceutical, Environmental Authors John W. enderson Jr. and William J. Long Agilent
More informationThe Role of Pore Size and Stationary Phase Composition in Preventing Aqueous-Induced Retention Time Loss in Reversed-Phase HPLC
The Role of Pore Size and Stationary Phase Composition in Preventing Aqueous-Induced Retention Time Loss in Reversed-Phase HPLC Brian A. Bidlingmeyer and Alan D. Broske Agilent Technologies, Inc., 2850
More informationShodex TM ODP2 HP series columns
HPLC Columns Shodex TM ODP2 HP series columns Better retention of highly polar substances Technical notebook No. 6 Contents 1. Introduction 1-1. Specifications 1-2. Eluent Compatibility of ODP2 HP Series
More informationSeparation selectivity of some phenolic acids in RP HPLC systems with binary mobile phase comprised various modifiers
Adsorption (2010) 16: 287 294 DOI 10.1007/s10450-010-9241-2 Separation selectivity of some phenolic acids in RP HPLC systems with binary mobile phase comprised various modifiers Anna Klimek-Turek Tadeusz
More informationComparison Guide to C18 Reversed Phase HPLC Columns
Comparison Guide to C18 Reversed Phase HPLC Columns Comparison Data on Commonly Used C18 Phases Stationary Phase Specifications Phases Compared According to Relative Hydrophobicity Phases Compared According
More informationHigh Performance Liquid Chromatography
High Performance Liquid Chromatography What is HPLC? It is a separation technique that involves: Injection of small volume of liquid sample Into a tube packed with a tiny particles (stationary phase).
More informationAlternatives to Sub-2 µm UHPLC Columns
Alternatives to Sub-2 µm UHPLC Columns J. J. DeStefano, B. E. Boyes, S. Schuster, W. L. Miles, and J. J. Kirkland Advanced Materials Technology, Inc. 3521 Silverside Rd., Quillen Bld., Ste. 1-K Wilmington,
More informationNew 5-micron HALO-5 columns based on Fused- Core particle technology boost the performance of HPLC. Compared to other HPLC columns, HALO-5 columns
New 5-micron HALO-5 columns based on Fused- Core particle technology boost the performance of HPLC. Compared to other HPLC columns, HALO-5 columns have: the highest plate number versus any other 5-micron
More informationInertsil ODS-EP Technical Information
Technical Information Advantages of The selectivity is completely different from those of conventional columns such as ODS column due to its specific polar group in the stationary phase. Durability can
More information( )( ) Selectivity Choices in Reversed-Phase Fast LC. Introduction. R s = 1 a 1 k 4 a 1 + k
Selectivity Choices in Reversed-Phase Fast LC Luisa Pereira, Monica Dolci, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20543 Key Words Accucore, Hypersil GOLD, Syncronis, Column Chemistry,
More informationLiquid Chromatography
Liquid Chromatography 1. Introduction and Column Packing Material 2. Retention Mechanisms in Liquid Chromatography 3. Method Development 4. Column Preparation 5. General Instrumental aspects 6. Detectors
More informationHICHROM. Chromatography Columns and Supplies. LC COLUMNS Zorbax. Catalogue 9. Hichrom Limited
HICHROM Chromatography Columns and Supplies LC COLUMNS Zorbax Catalogue 9 1 The Markham Centre, Station Road Theale, Reading, Berks, RG7 4PE, UK Tel: +44 (0)118 930 3660 Fax: +44 (0)118 932 3484 Email:
More informationfor Acclaim Mixed-Mode HILIC-1 Column
for Acclaim Mixed-Mode HILIC-1 Column Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 1 of 17 Product Manual for ACCLAIM Mixed-Mode HILIC-1 Column 5µm, 4.6 x 250mm, P/N 066844 5µm, 4.6 x 150mm, P/N
More informationValue of Eclipse Plus C18 and ph Selectivity to Analyze Amine-Containing Drugs
Value of Eclipse Plus C18 and ph Selectivity to Analyze Ae-Containing Drugs Application Pharmaceuticals Authors John W. Henderson Jr., William J. Long, and Cliff Woodward Agilent Technologies 285 Centerville
More informationChromatographic Analysis
Chromatographic Analysis Distribution of Analytes between Phases An analyte is in equilibrium between the two phases [S 1 ] [S 2 ] (in phase 1) (in phase 2) AS [S2 ] K 2 A S [S1 ] 1 AS, A 1 S Activity
More informationSilanol. Groups Liquid Chromatography
Abstract umber: 0-8P Study of Secondary Interaction Based on Residual lanol Groups for Reversed-Phase Liquid Chromatography II orikazu agae Ph.D, Kouji Yamamoto, Chiaki Kadota Chromaik Technologies Inc.
More informationWelcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster
Welcome to our E-Sear: Choosing HPLC Columns for Faster Analysis Smaller and Faster High Throughput/Fast LC Requires. Short columns 0 mm or shorter Small particle sizes. µm Rapid Resolution or new.8 µm
More informationHPLC Background Chem 250 F 2008 Page 1 of 24
HPLC Background Chem 250 F 2008 Page 1 of 24 Outline: General and descriptive aspects of chromatographic retention and separation: phenomenological k, efficiency, selectivity. Quantitative description
More informationfor Acclaim Mixed-Mode WCX-1
for Acclaim Mixed-Mode WCX- Product Manual for the Acclaim Mixed-Mode WCX- Column Page of 8 Product Manual for Acclaim Mixed-Mode WCX- Columns Acclaim Mixed-Mode WCX-, 3µm, Analytical column, 3.0 x 50mm
More informationA C18 Silica Column With Exceptional Temperature and ph Stability
A C18 lica Column With Exceptional Temperature and ph Stability Brian A. Jones 1, Stephanie J. Marin 1, Jody Clark 1, Nathan L. Porter 1, J. Andreas Lippert 2, and Todd M. Johnson 2 1. Selerity Technologies,
More informationAn Efficient Approach to Column Selection in HPLC Method Development
464 LCGC NT AMEICA VLUME 20 NUMBE 5 MAY 2002 www.chromatographyonline.com An Efficient Approach to Column Selection in PLC Method Development The base-particle characteristics and bonded-phase chemistries
More informationReversed Phase Solvents
Part 1. General Chromatographic Theory Part 2. verview of HPLC Media Part 3. The Role of the Mobile Phase in Selectivity Part 4. Column Care and Use Reversed Phase Solvents 2 Solvents for RP Chromatography
More informationSunniest C18 Sunniest RP-AQUA Sunniest C8 Patent pending
Patent pending Website : www.chromtech.net.au E-mail : info@chromtech.net.au telno : 0 97 0... in AUSTRALIA A Novel Bonding Technique (patent pending) An unique trifunctional silyl-reagent was developed
More informationCertificate of Analysis
National Institute of Standards & Technology Certificate of Analysis Standard Reference Material 870 Column Performance Test Mixture for Liquid Chromatography This Standard Reference Material (SRM) is
More informationChapter content. Reference
Chapter 7 HPLC Instrumental Analysis Rezaul Karim Environmental Science and Technology Jessore University of Science and Technology Chapter content Liquid Chromatography (LC); Scope; Principles Instrumentation;
More informationState-of-the-art C18 HPLC Columns
An HPLC GL Sciences Newest and Most Advanced ODS Phase-New For 00 State-of-the-art C HPLC s Improved Peak Shapes and Heights Enhancing Sensitivity High Resolution Fast Equilibration Compatible with 00%
More informationFast Separation of Vastly Different Compounds by Isocratic HPLC
Fast Separation of Vastly Different Compounds by Isocratic HPLC Aimee N. Heyrman and Yury Zelechonok Resolution Systems, Inc. 590 E. 32nd St. Holland. MI, 49423 SIELC Technologies, 15 E. Palatine Rd, Suite
More informationFast Screening Methods for Steroids by HPLC with Agilent Poroshell 120 Columns
Fast Screening Methods for Steroids by HPLC with Agilent Poroshell 2 Columns Application Note Pharma, BioPharma, and Clinical Research Author William Long Agilent Technologies, Inc. Introduction Steroids
More informationComparison of different aqueous mobile phase HPLC techniques
June 009 ewsletter Pharmaceutical Analysis: What is Your Problem? SIELC Technologies, Inc., Prospect Heights, IL 0070 Pharmaceutical analysis involves liquid chromatography of various compounds, from active
More informationLC Column Troubleshooting. Do you have an equivalent column for my LC column?
LC Column Troubleshooting Do you have an equivalent column for my LC column? LC columns from different suppliers may have very different retention properties even if the bonding is the same. Although there
More informationLogical Method Development for Peptides and Proteins Using RP-HPLC and SEC
Logical Method Development for Peptides and Proteins Using RP-HPLC and SEC Slide 4 001284P1.PPT Seminar Outline 1. Gel-filtration separations by SEC 2. Developing methods for protein and peptide separations
More informationACE Excel SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability
ACE Excel SuperC8 TM Ultra-Inert UHPLC and HPLC s with Extended ph Stability Exploit selectivity at low, intermediate and high ph Use with LC/MS-compatible buffers between ph.. High efficiency µm, µm,
More informationLiquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.
High performance liquid chromatography (HPLC) is a much more sensitive and useful technique than paper and thin layer chromatography. The instrument used for HPLC is called a high performance liquid chromatograph.
More informationChromatography. Gas Chromatography
Chromatography Chromatography is essentially the separation of a mixture into its component parts for qualitative and quantitative analysis. The basis of separation is the partitioning of the analyte mixture
More informationBarry E. Boyes, Ph.D. Consumables and Accessories Business Unit May 10, 2000
Barry E. Boyes, Ph.D. Consumables and Accessories Business Unit May 0, 000 HPLC Column Troubleshooting What Every HPLC User Should Know :00 a.m. EST Telephone Number: 86-650-06 Chair Person: Tim Spaeder
More informationChemistry Instrumental Analysis Lecture 28. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 28 Two types in general use: -packed (stationary phase) -open tubular or capillary determine selectivity and efficiency of the sample. Column Materials Column
More informationKromasil. for your analytical HPLC. Kromasil. The way to peak performance in liquid chromatography
Kromasil for your analytical HPLC Kromasil The way to peak performance in liquid chromatography Kromasil is known, worldwide, for its high performance and excellent total economy in preparative and industrial
More informationACE Ultra Inert Base Deactivated HPLC Columns
ACE Ultra Inert Base Deactivated HPLC Columns Contents Page ACE HPLC Columns...- Independent Comparison of HPLC Columns #... Independent Comparison of HPLC Columns #... ACE 00Å HPLC Columns Specifications...-
More informationMethod Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples
Technical Note 101 Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples This technical note includes by specific information on the extraction
More informationPreparative Chromatography with Improved Loadability
05 LCGC RTH AMERICA VLUME 0 UMBER VEMBER 00 www.chromatographyonline.com Preparative Chromatography with Improved Loadability In combinatorial chemistry and drug development, as well as in the full-scale
More informationUPLC Method Development and Validation
UPLC Method Development and Validation Rev. 2 2008 Waters Corporation Challenges of Method Development Methods are developed throughout the drug development process Samples vary in complexity Redundancy
More informationOverview Applications of NUCLEODUR HPLC Phases
verview Applications of HPLC Phases Phase Specification Characteristics* Stability Structure A B C octadecyl phase, high density coating multi-endcapping 8% C USP L ph stability -, suitable for LC/MS (Si-
More informationPeptide and protein analysis by capillary HPLC Optimization of chromatographic and instrument parameters. Application. Angelika Gratzfeld-Huesgen
Peptide and protein analysis by capillary HPLC Optimization of chromatographic and instrument parameters Application Angelika Gratzfeld-Huesgen Abstract This application has been verified using an Agilent
More informationHPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application
HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application Pharmaceuticals Author Cliff Woodward and Ronald Majors Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808
More informationThermo Scientific Accucore XL HPLC Columns. Technical Manual
Thermo Scientific Accucore XL HPLC Columns Technical Manual Thermo Scientific Accucore XL HPLC Columns Based on Core Enhanced Technology using µm solid core particles, Accucore XL HPLC columns allow users
More information(HILIC) Bill Champion Agilent Technologies, Inc opt 3/opt3/opt 2 HILIC - Agilent Restricted
Hydrophilic Interaction Chromatography (HILIC) Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com Page 1 HILIC A method of recent attention Bill Champion
More informationANALYSIS OF BASIC COMPOUNDS
COMPARISON OF Fully AND Superficially POROUS PARTICLE COLUMNS FOR THE ANALYSIS OF BASIC COMPOUNDS Kenneth J. Fountain, Jane Xu, Zhe Yin, Pamela C. Iraneta, Diane M. Diehl Waters Corporation GOAL Compare
More informationOpen Column Chromatography, GC, TLC, and HPLC
Open Column Chromatography, GC, TLC, and HPLC Murphy, B. (2017). Introduction to Chromatography: Lecture 1. Lecture presented at PHAR 423 Lecture in UIC College of Pharmacy, Chicago. USES OF CHROMATOGRAPHY
More informationExperiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks
Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks bjective: The purpose of this experiment is to quantify the caffeine content of a diet soda sample using Ultra-High Performance
More informationRP C18 column with feature of a silanol group
RP C8 column with feature of a silanol group lanol Activity Controlled C8 Column Sunrise ctacosyl C8) Sunrise ctadecyl C8) Sunrise ctadecyl-sac C8-SAC has an interaction of silanol groups ChromaNik Sunrise
More informationImportant Guidelines for Optimizing Speed and Sensitivity in Small-Molecule LC UV and LC MS
Página 1 de 6 Important Guidelines for Optimizing Speed and Sensitivity in Small-Molecule LC UV and LC MS May 1, 2005 By: Richard A. Henry, David S. Bell LCGC North America The task of developing a new
More informationInertsil Technical Library
HPLC COLUMN INERTSIL TECHNICAL LIBRARY H P L C C o l u m n Inertsil Technical Library Inertsil ODS-4, C8-4 Comparison of Performance List of Compared Columns Experimental Explanation, Analytical Conditions
More informationACE SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability
ACE SuperC8 TM Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability Exploit selectivity at low, intermediate and high ph Use with LC/MS compatible buffers between ph.. High efficiency µm, µm,
More informationBringing uhplc Performance to the Separation of Peptides
P e p t i d e E S - C 1 8 U H P L C C o l u m n s Bringing uhplc Performance to the Separation of Peptides Compatible with UHPLC and conventional HPLC equipment. Sumpfstr. 3, CH-6300 Zug; Fax: 041 748
More informationCh.28 HPLC. Basic types of Liquid Chromatography Partition (LLC) Adsorption (LSC) Ion Exchange (IC) Size Exclusion (SEC or Gel Chromatography)
Ch.28 HPLC 28.1 Basic types of Liquid Chromatography Partition (LLC) Adsorption (LSC) Ion Exchange (IC) Size Exclusion (SEC or Gel Chromatography) High Performance (Pressure) LC Glass column st.steel (high
More informationLC Technical Information
LC Technical Information Method Transfer to Accucore.6 μm Columns Containing solid core particles, which are engineered to a diameter of.6μm and a very narrow particle size distribution; Accucore HPLC
More informationHydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity
Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity Monica Dolci, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20544 Key Words Hydrophilic, HILIC,
More informationAlltech Alltima HP Introduction
Alltech Introduction High-Stability, High-Purity, High-Performance, Low-Bleed Columns for Demanding Applications Better Peak Symmetry high-purity silica eliminates peak tailing problems Long Column Life
More informationTechnical Guide Thermo Scientific Syncronis HPLC Columns
Technical Guide Thermo Scientific Syncronis HPLC Columns Consistent, predictable separations, Column after column, time after time Thermo Scientific Syncronis Columns When developing a new method, one
More informationIntroduction. Chapter 1. Learning Objectives
Chapter 1 Introduction Learning Objectives To understand the need to interface liquid chromatography and mass spectrometry. To understand the requirements of an interface between liquid chromatography
More informationAn Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.
An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols. Jaap de Zeeuw, Ron Stricek and Gary Stidsen Restek Corp Bellefonte, USA To analyze basic compounds at
More informationQUALITY HPLC COLUMNS PARTICLE PORE PAGE SIZE SIZE. C18 Bidentate C18 Extremely long column life and rugged 4um 100Å 3
Cogent HPLC Columns Manufactured by Table of Contents QUALITY HPLC COLUMNS PARTICLE PORE PAGE SIZE SIZE C18 Bidentate C18 Extremely long column life and rugged 4um 100Å 3 C8 Bidentate C8 Extremely long
More informationAn HPLC column offering moderate retentivity with superb peak shape Inertsil ODS Sprint
An HPLC column offering moderate retentivity with superb peak shape Inertsil ODS Sprint An important new column in the Inertsil series Bonded-Phase Structure Perfect balance of retention for both polar
More informationAnalytical Advantages of Highly Stable Stationary Phases for Reversed-Phase LC
Analytical Advantages of Highly Stable Stationary Phases for Reversed-Phase LC During the past five years, many manufacturers of high performance liquid chromatography (HPLC) columns have focused on improving
More informationzorbax Packed by Phenomenex Material Characteristics High efficiency Guaranteed quality Phenomenex columns packed with Zorbax
Packed by Phenomenex High efficiency Guaranteed quality Phenomenex columns packed with 7 μm materials offer the advantages of lower cost as well as lower operating backpressures, which result in long column
More informationReverse Phase High Performance Liquid Chromatography method for determination of Lercanidipine hydrochloride in bulk and tablet dosage form
Research Article ISSN: 0974-6943 M.V.Kumudhavalli et al. / Journal of Pharmacy Research 2014,8(11), Available online through http://jprsolutions.info Reverse Phase High Performance Liquid Chromatography
More informationJournal of Pharmaceutical and Biomedical Analysis
Journal of Pharmaceutical and Biomedical Analysis 50 (2009) 426 431 Contents lists available at ScienceDirect Journal of Pharmaceutical and Biomedical Analysis journal homepage: www.elsevier.com/locate/jpba
More informationChapter 27: Gas Chromatography
Chapter 27: Gas Chromatography Gas Chromatography Mobile phase (carrier gas): gas (He, N 2, H 2 ) - do not interact with analytes - only transport the analyte through the column Analyte: volatile liquid
More informationImpSilHPLC Columns. The solution for your chromatography IMPTECH SCIENTIFIC
ImpSilHPLC Columns The solution for your chromatography IMPTECH SCIENTIFIC 7-1-60,203-Veena Dharam Karam Road,Ameerpet,Hyderabad-500 016 Tel:040-23740443/483 Ph:9848051071 imptechscientific@gmail.com www.impscience.com
More informationAn Overview of the the Development of Stationary Phases for Reversed-Phase Liquid Chromatography
An Overview of the the Development of Stationary Phases for Reversed-Phase Liquid Chromatography Analytical Potential of Stable Phases for Reversed-Phase Liquid Chromatography by Jacek Nawrocki, Jon Thompson,
More informationSeparation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.
Separation of Explosives in EPA 833: Column Choices Optimize Speed, Resolution, and Solvent Use Application Environmental Authors John W. Henderson Jr. and William J. Long Agilent Technologies, Inc. 28
More informationReversed Phase Chromatography: Mobile phase Considerations
Reversed Phase: Mobile Phase Considerations Dr. www.forumsci.co.il/hplc Example of a Reversed Phase Method with UV Detection 1 Chromatographic Process B+A Mobile phase Elution through the Column-movie
More information