1 Página 1 de 6 Important Guidelines for Optimizing Speed and Sensitivity in Small-Molecule LC UV and LC MS May 1, 2005 By: Richard A. Henry, David S. Bell LCGC North America The task of developing a new analytical method usually falls to someone who has at least several years of high performance liquid chromatography (HPLC) experience. Using this experience, the analyst selects appropriate instruments and columns that are likely to succeed for the analysis. A simple analysis frequently can be accomplished with C18 columns, isocratic mobile phases, and a wide variety of HPLC instrument designs. However, difficult analyses might require consideration of columns having special selectivity, use of mobile phase gradients, and very careful instrument selection. To optimize methods for maximum sensitivity (1), one should start with a detector that exhibits a high signal-to-noise ratio (S/N) for each component. When LC UV is not adequate, sensitive techniques such as LC-mass spectrometry (MS) must be considered, or the sample must be concentrated. It is also important to keep retention factors small (k = 1 5) as long as peak resolution can be maintained. Because modern 5-μm packings routinely exhibit about 750 plates/cm (75,000 plates/m), a 5-cm column can create a rugged separation system having several thousand total plates. Such short columns often can provide adequate resolution, especially with LC MS, if selectivity has been optimized. Small internal diameter columns can provide additional sensitivity by minimizing sample dilution when limited sample amount is available; however, excessive instrument dispersion volume can reduce efficiency and offset any sensitivity gains achieved. Finally, detector signal can be maximized by introducing the largest practical sample volume. Often, >10% of the column volume can be injected before loss of resolution is noticed (even more can be injected under gradient conditions). Column volume can be estimated from the solvent front using the relationship V M = t M F, where t M is retention time of an unretained peak and F is flow rate. Column selectivity influences efficiency and retention requirements: The column and mobile phase are both very important and can be considered the two primary variables in HPLC method development (2). While not the main emphasis of this article, it is worthwhile to briefly examine the resolution equation (3) and the importance of the selectivity term. Although efficiency (N) and retention factor (k) are essential elements of column performance, increased selectivity (α) allows the same resolution to be achieved with smaller N values (shorter columns) and smaller k values. Short columns and small retention factors increase separation speed and improve sensitivity. Figure 1a shows a good result with a 15-cm, C18 column. However, marginal selectivity for critical pairs 3,2 and 4,7 precludes the use of shorter columns and retention factors to improve speed and sensitivity. Figure 1: Separation of polar compounds on (a) C18 and (b) reversedphase amide columns. The same separation under identical mobile phase conditions is shown with a 15-cm, amide-type column in Figure 1b. Note that the embedded polar group has increased both retention and selectivity for this class of polar compounds. Because there is greater selectivity, fewer plates (shorter
2 Página 2 de 6 Column: 15 cm 4.6 mm, 5-µm Ascentis RP-Amide or C18; mobile phase: 75:25 (v/v) 20 mm phosphoric acid (ph 2.0)-acetonitrile; temperature: 30 C; detection: UV absorbance at 270 nm; injection volume: 25 ml; sample: all components 50 µg/ml in 20 mm phosphoric acid (ph 2.0). Peaks: 1 = resorcinol, 2 = catechol, 3 =2-methyl resorcinol, 4 = 4-methyl catechol, 5 = 2,5-dimethyl resorcinol, 6 = 3-methyl catechol, 7 = 4-nitro catechol. columns) would be sufficient to achieve baseline separation. The relationship between plates and selectivity can be seen by substituting typical values of R S = 1 and k = 2 into equation 1. The dramatic impact of improving selectivity is shown in Table I. Selectivity values for the critical group of peaks 4-6 with the amide phase (calculated by the ratios k 5 /k 4 and k 6 /k 5 ) would exceed the α value of 1.1 shown in the table and require only a few thousand plates for fast, sensitive separation on a short column. This option is made possible due to greater selectivity of the amide group for the critical pairs. A complete discussion of column and mobile phase selectivity is beyond the scope of this article. However, it should be pointed out that embedded polar group phases containing amide, carbamate, urea, and other polar groups are extremely complementary to traditional alkyl phases and often are more selective and retentive for polar compounds than alkyl, cyano, or phenyl phases. Phases showing different selectivity and elution order are said to be orthogonal for that class of compound. Neue (4) has pointed out the importance of these new embedded polar group phases and proposed a new selectivity term s 2, which is derived from the correlation coefficient r 2, and makes it convenient to quantitatively compare column orthogonal behavior for different classes of compounds. Selectivity difference s 2 would vary from a value of zero when two columns (or other variables such as ph) show no selectivity difference (perfect straight-line correlation or r 2 equal to unity) to a value of one when two columns show completely different selectivity (no correlation or r 2 equal to zero). Figure 2 confirms that the amide embedded polar group phase can be highly orthogonal to C18 (s 2 = 0.55 for this group of polar compounds). Simply interchanging C18 and embedded polar group columns under the same mobile phase conditions can often improve resolution and allow the use of shorter columns. Newer versions of embedded polar group phases have been designed specifically to show very low bleed and ionic background under LC MS conditions, allowing them to be used interchangeably with C18 or C8 during method Figure 2: Typical selectivity difference plot for development. polar compounds with Ascentis C18 and RP- Amide phases. Table I: Impact of selectivity on column requirement for Rs = 1 and k = 2 Another column type that is showing increasing popularity for alternate selectivity, is the fluorinated phase, especially the perfuorophenylpropyl version. It is interesting to note that embedded polar group, perfuorophenylpropyl, and alkyl phases are often orthogonal to each other, making them a powerful method development trio for difficult separations. The early use of these three column types can ensure optimum retention and band spacing, and also confirm that all analytes and interferences have been identified. Because detectors can often limit mobile phase options, especially in LC MS, changing columns to alter selectivity is becoming more attractive than attempting extensive changes to the mobile phase. The bottom line is that one should always use the most selective column to optimize band spacing, allow shorter columns, and produce faster, more sensitive assays. Gradient conditions: Gradient elution can be advantageous for sensitive analysis, especially when using MS and other detectors or conditions that do not show higher baseline noise and drift with changing mobile phase. Gradients avoid peak dilution that occurs with increasing k values under isocratic conditions; an optimum gradient elutes multiple components in narrow zones as though all have the same short k value. Gradients also can allow very large sample volumes to be introduced in a weak mobile phase, a sample concentration technique that has been referred to as on-column trace enrichment. When the sample contains diverse components that cannot be spaced optimally by careful column choice and isocratic conditions, programming solvent strength can optimize retention and reduce overall analysis time. Finally, gradients can keep the column and other components free of sample carryover
3 Página 3 de 6 due to trace adsorption on active surfaces. Figure 3: Plot of efficiency versus retention factor for several different columns using an eight-component mixture (9). Of course, there are several distinct downsides to gradient elution, such as a requirement for more complex and costly instrumentation, excessive baseline drift and noise with certain detectors and conditions, and additional equilibration time between sample runs. Differences in gradient dwell volume and delay time also can make results difficult to reproduce between instruments and labs, especially at low flow rates. Figure 4: Illustration of how dwell volume can create a lag time between composition change at the pump and arrival at the column inlet; the actual gradient profile can be obtained by adding a UV-absorber such as acetone to the B reservoir (10). small internal volume. Instrument Guidelines for Optimizing Speed and Sensitivity Much of the effort during method development is spent trying to optimize columns and mobile phases, which are well documented in validated HPLC methods. HPLC instruments, however, are considered interchangeable, although it is well known that the same column and mobile phase conditions can produce very different results when used with different HPLC components. This is especially true when reduced-bore columns are employed to optimize sensitivity. This article will attempt to alert users to some important HPLC instrument variables that are not always considered or documented during development of an HPLC method using columns having Instrument design: Majors (5) has provided a summary of the various HPLC instrument volume elements and how they can affect chromatographic performance. One important element is the volume that the sample experiences external to the column bed. It is well known that sample components spread and dilute as they travel along the column (band dispersion). However, it is not so well known that they also spread as they travel through the open channels that make up the injector and detector flow paths before and after the column. According to Bakalyar (6-8), the equations that describe these two kinds of zone spreading are Figure 5: Step gradients with UV detection and acetone in the B reservoir to show actual gradient performance in creating solvent blends for delivery to the column (9). The chromatograms were generated with 20% step gradients to 100% after a 5-min hold. where s is the peak width standard deviation of a Gaussian distribution in volume units (estimated from width at baseline, which is about 4s). Band spreading and dilution inside the column are proportional to the volume of the column V M retention factor k, and inversely related to column plate count N. Sensitivity, therefore, is optimized with short, small internal diameter columns, low k values, and high column efficiency because they all reduce s and increase sample concentration. Band spreading and dilution in connections are influenced greatly by the diameter of the connector opening d tube, as well as length L and flow rate F. Band spreading is also inversely related to analyte diffusion coefficient D M, which means that bands related to large molecules, such as proteins, spread more rapidly than small ones under laminar flow in open channels outside of the column. This is due to slow diffusion out of the stagnant layer near the channel walls, and in this respect, causes band-broadening similar to the C-term or resistance to mass-transfer term of the Van Deemter equation inside the column. Although the two types of band spreading have different origins, the overall effects are additive according to a variance equation Ideally, instrument volume should be negligible. However, when the volume of the instrument becomes comparable to the volume occupied by the narrowest peak, resolution degrades rapidly. Table II shows how peak volume and other important
4 Página 4 de 6 Figure 6: Blow-up of the 80% step region in Figure 5 to show effect of dwell volume when its value becomes comparable to the volumetric flow rate in ml/min (9). operating parameters change as column internal diameter becomes smaller. The loss of efficiency caused by several different instrument internal volumes on a peak that is 80 μl in base width (σ = 20 μl) is shown in Table III. Peak widths in this range should be observed routinely for short columns in the 2 3 mm i.d. range with 5-μm particles. Columns having 3-μm and smaller packings will require instruments having even lower internal volumes in order to preserve column efficiency. While instrument manufacturers rarely supply data on dispersion volume, and it is challenging to measure, there is a simple experiment that can be used to calibrate and compare instruments (9). Select two or three short columns having the same phase and length, but different internal diameters (that is, 50 mm 4.6 mm, 50 mm 3 mm, and 50 mm 2 mm C18 columns). Inject 20 μl of a multicomponent mixture into the 4.6-mm i.d. column at 1.25 ml/min and plot efficiency (N or N/m) versus k value, as shown in Figure 3. Use a minimum of four or five components with k values between 1 and 20 to best illustrate the effect. Repeat the experiment for the 50 mm 3 mm column at 0.6 ml/min with a 10-μL injection, the 50 mm 2 mm column at 0.2 ml/min with a 5-μL injection, and even smaller internal diameter columns and sample sizes if desired. Modern instruments should show very little loss of efficiency for the smaller internal diameter columns (plots should nearly overlap) down to about 50 mm 1 mm; however, older instruments with larger internal volume will show a plot similar to that shown. Clearly, 50 mm 2 mm columns should not be used in this particular HPLC instrument for optimum performance. This technique also can be used to maintain instruments in suitable operating condition because changes can occur in injectors, filters, and other components that can increase instrument dispersion over time. As one drops to ever smaller column internal diameters (at lower flows and smaller injection volumes), every HPLC instrument eventually will depart from the ideal of a rapid rise to constant efficiency (equal to true column efficiency) even for very short retention factors. Solvent delivery systems: A critical aspect of pump design is the dwell volume, which, along with flow rate, determines the time required for a gradient to arrive at the column inlet. Figure 4 illustrates the effect of excessive dwell volume, which can introduce a significant isocratic hold time when its magnitude becomes comparable to the flow rate in milliliters per minute (10). Dwell volume can be estimated from the actual gradient profile and the equation Figure 7: Using delayed sample injection to address problems caused by excessive delay time due to dwell volume (11). Shown are (a) the standard approach, (b) delayed injection alone, and(c) combined equilibration and delayed injection. where F is flow rate, t G is gradient duration time, and t 1/2 is the time at which detector response reaches 50% amplitude. A step gradient also can be employed wheret G = 0 and the equation simplifies to F D = t 1/2 F; if the column is left in place for this measurement, column volume V M must be subtracted to arrive at V D. Dwell volume also impacts the equilibration time between gradient runs. Large differences in dwell volume can cause methods to translate poorly between instruments, especially at the low flows required with small internal diameter columns. An instrument with large dwell volume can take much longer to equilibrate after an isocratic change or a gradient run. Dwell volume cannot be completely eliminated in any design because solvents must be well mixed before reaching the column. However, low-pressure designs that feature prepump blending tend to have more dwell volume than high pressure designs, which use postpump blending at a mixing tee. The step gradients at several flow rates in Figures 5 and 6 show the increasing importance of dwell volume at low flows (9). Note that delay time due to dwell volume becomes noticeable in this design at 100 μl/min flow rate, which is becoming comparable to the total dwell volume of 200 μl. Although the system chosen for illustration has only 80 μl of dwell volume attributed to the pump, total dwell volume rises to about200 μl when other components such as autosampler and injector are included. Even with this rather small dwell volume, it can take a very long time at low flows for gradient arrival at the column or to change isocratic conditions during method development. Although flow splitters often are inconvenient for other reasons, the use of split flow at constant pump flow rate can eliminate excessive delay time for changes in mobile phase composition due to dwell volume. Another technique to consider when faced with a significant dwell volume effect is to invoke an injection delay that is nearly equal to the delay time due to dwell volume, as shown in Figure 7 (11). This eliminates variable retention times created by HPLC instruments having different dwell volumes. By starting a new gradient run with delayed injection
5 Página 5 de 6 immediately after completion of the previous run, as shown in Figure 7c, equilibration time and dwell volume delay time can be overlapped, cutting total delay between runs in half. These delayed injection approaches require an accurate knowledge of instrument dwell volume. Methods that involve the use of a gradient or other programmed change in solvent composition should include instrument dwell volume as part of their documentation. Detectors: The use of UV detectors that feature wide wavelength range and spectral output for qualitative analysis has enhanced their popularity greatly. Method development typically is carried out with UV detection in the nm wavelength range, where most organic compounds Table II: Operating parameters that diminish as HPLC column i.d. becomes smaller; estimated values for 150 mm columns and 5 µm particles at k = 1 show enough molar absorptivity for detection. UV flow cells must approach 1 cm in length in order for Beer's law to produce enough change in absorbance to allow low level detection. This is a serious limitation when reducedbore columns are employed to minimize dilution of mass-limited samples. Flow cells can be miniaturized adequately to allow use with 1 2 mm i.d. columns, but UV detectors begin to destroy resolution when flow cell volume becomes comparable to sample zone volume. LC MS is much more sensitive than LC UV when interface conditions can be optimized for mobile phase evaporation and analyte ionization. The absence of anything analogous to a UV flow cell can be a significant advantage for interfacing LC MS to small-diameter columns. For LC MS method development, it is especially desirable to select a few proven, volatile mobile phases and change stationary phase chemistries to optimize selectivity. A complete discussion of detector requirements for rapid, sensitive analysis is beyond the scope of this article. However, fast detector response and faster integrator sampling rate than normal is required for accurate area determination of narrow peaks. Table III: Examples of how extracolumn dispersion can increase sample band spreading and reduce efficiency when instrument volume becomes comparable in size to column peak volume (peak volume usually is estimated by assuming base width of 4s). Conclusions HPLC with sensitive detectors is solving new problems in high speed analysis. There is still a tendency to specify C18 columns, however, even when marginal resolution prevents the use of short columns to improve speed and sensitivity. Substitution of more selective columns, containing amide or other embedded polar group phases, without changing the mobile phase can improve resolution enough to allow the use of shorter columns. When reduced-bore columns and low flow rates are used to further optimize sensitivity, HPLC instrument design can become very important. Two important aspects of an optimized instrument system are low instrument dispersion volume from injector to detector and low dwell volume in the solvent delivery system and autosampler. Even when manufacturers provide values for dispersion and dwell volume, it is useful to learn how to conduct experiments that confirm values and compare instruments. Ideally, both dispersion and dwell volumes should be documented for every HPLC instrument and referenced in method specifications when appropriate. This is becoming increasingly important as HPLC applications move toward smaller column and particle diameters. Richard A. Henry is a consultant. He is retired from Penn State University, State College, Pennsylvania, where he taught instrumental analysis to chemistry majors for more than 15 years. His address is rhenry@psualum.com. David S. Bell is currently applications and technical support manager with Supelco, Bellefonte, Pennsylvania. References (1) L.R. Snyder, J.J. Kirkland, and J.L. Glajch, Practical HPLC Method Development, 2nd ed. (John Wiley & Sons, Hoboken, New Jersey, 1997). (2) U.D. Neue, K.V. Tran, P.S. Iraneta, and B.A. Alden, J. Sep. Sci. 26, 1-6 (2003). (3) R.E. Majors and P.W. Carr, LCGC 19(2), (2001). (4) U.D. Neue, E.S. Grumbach, J.R. Mazzeo, K.V. Tran, and D.M. Wagrowski-Diehl, "Bioanalytical Separations," in Handbook of Analytical Separations, I.D. Wilson, ed., (Elsevier Science B.V., New York, 2003).
6 Página 6 de 6 (5) R.E. Majors, LCGC 21(12), (2003). (6) S.R. Bakalyar and K. Olsen, Rheodyne Tech. Notes 9, Rheodyne, Rohnert Park, California. (7) S.R. Bakalyar, C. Phipps, B. Spruce, and K. Olsen, J. Chromatogr., A 762, (1997). (8) M.D. Foster, M.A. Arnold, J.A. Nichols, and S.R. Bakalyar, J. Chromatogr., A 869, (2000). (9) Thermo Electron User Training Seminars, Thermo Electron Corp., Waltham, Massacusetts. (10) Supelco User Training Seminars, Supleco, Bellefonte, Pennsylvania. (11) Uwe D. Neue, Waters Corporation, Milford, Massachusetts, private communication.
LC Technical Information
LC Technical Information Method Transfer to Accucore.6 μm Columns Containing solid core particles, which are engineered to a diameter of.6μm and a very narrow particle size distribution; Accucore HPLC
More informationImpact of Instrument Dispersion on Performance of HPLC Capillary Columns
Impact of Instrument Dispersion on Performance of HPLC Capillary Columns Wendy Roe, Richard A. Henry, and Hillel Brandes Supelco, Div. of Sigma-Aldrich Bellefonte, PA 16823 USA T413131 sigma-aldrich.com/analytical
More informationpenta-hilic UHPLC COLUMNS
penta-hilic UHPLC COLUMNS penta-hilic Highly retentive, proprietary penta-hydroxy-ligand Excellent peak shape for polar compounds with a variety of functional groups: acids, bases, zwitterions strong and
More informationMethod Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions
Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions Today, ultra-high-performance liquid chromatography (UHPLC) has taken a firm foothold in the analytical laboratory.
More informationHPLC Praktikum Skript
HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques,
More informationAssay Transfer from HPLC to UPLC for Higher Analysis Throughput
MAY 2005 SEPARATION SCIENCE REDEFINED 31 Assay Transfer from HPLC to UPLC for Higher Analysis Throughput A typical HPLC assay was transferred and optimized for a Waters ACQUITY UPLC system to achieve both
More informationIntroduction. Chapter 1. Learning Objectives
Chapter 1 Introduction Learning Objectives To understand the need to interface liquid chromatography and mass spectrometry. To understand the requirements of an interface between liquid chromatography
More informationIntroduction to Chromatographic Separations
Introduction to Chromatographic Separations Analysis of complex samples usually involves previous separation prior to compound determination. Two main separation methods instrumentation are available:
More informationChromatography and other Separation Methods
Chromatography and other Separation Methods Probably the most powerful class of modern analytical methods for analyzing mixture of components---and even for detecting a single component in a complex mixture!
More informationHow to Troubleshoot a Failed System Suitability Test
Volume 34 Number 12, 881 932 Volume 34 Number 12 December 216 www.chromatographyonline.com LCGC NORTH AMERICA How to Troubleshoot a Failed System Suitability Test December 216 Preventing Carryover in GC
More informationChromatography. Gas Chromatography
Chromatography Chromatography is essentially the separation of a mixture into its component parts for qualitative and quantitative analysis. The basis of separation is the partitioning of the analyte mixture
More informationLiquid Chromatography
Liquid Chromatography 1. Introduction and Column Packing Material 2. Retention Mechanisms in Liquid Chromatography 3. Method Development 4. Column Preparation 5. General Instrumental aspects 6. Detectors
More informationHPLC Background Chem 250 F 2008 Page 1 of 24
HPLC Background Chem 250 F 2008 Page 1 of 24 Outline: General and descriptive aspects of chromatographic retention and separation: phenomenological k, efficiency, selectivity. Quantitative description
More informationAbstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.
Chromatography Primer Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments. At its heart, chromatography is a technique
More informationC18 Column. Care & Use Sheet
C18 Column Care & Use Sheet HALO Description HALO C18 is a high-speed, high-performance liquid chromatography column based on a new Fused-CoreTM particle design. The Fused-Core particle provides a thin
More informationSuccessfully Scaling and Transferring HPLC and UPLC Methods
Successfully Scaling and Transferring HPLC and UPLC Methods Esa Lehtorinne Esa_Lehtorinne@waters.com Tel: +358-9-5659 6288 Fax: +358-9-5659 6282 Waters Finland Kutomotie 16 00380 Helsinki 2013 Waters Corporation
More informationAnalytical Chemistry
Analytical Chemistry Chromatographic Separations KAM021 2016 Dr. A. Jesorka, 6112, aldo@chalmers.se Introduction to Chromatographic Separations Theory of Separations -Chromatography Terms Summary: Chromatography
More informationIntroduction to Chromatographic Separations (Chapter 1) Many determinations involve separation followed by analysis chromatography electrophoresis
Introduction to Chromatographic Separations (Chapter 1) Many determinations involve separation followed by analysis chromatography electrophoresis Chromatography: sample transported by mobile phase electrostatic
More informationpenta-hilic UHPLC COLUMNS
penta-hilic UHPLC COLUMNS Highly retentive, proprietary penta-hydroxy-ligand penta-hilic Excellent peak shape for polar compounds with a variety of functional groups: acids, bases, zwitterions strong and
More informationHigh Performance Liquid Chromatography
High Performance Liquid Chromatography What is HPLC? It is a separation technique that involves: Injection of small volume of liquid sample Into a tube packed with a tiny particles (stationary phase).
More informationChromatographic Analysis
Chromatographic Analysis Distribution of Analytes between Phases An analyte is in equilibrium between the two phases [S 1 ] [S 2 ] (in phase 1) (in phase 2) AS [S2 ] K 2 A S [S1 ] 1 AS, A 1 S Activity
More informationLuna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency
Luna 2.5 µm C18(2)-HST Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency Table of Contents Part 1 Theory 1.1 Abstract...3 1.2 Introduction...3 Part 2 Set Up 2.1
More informationImproving HPLC Column Selection and System Performance
Improving HPLC Column Selection and System Performance Richard A. Henry Penn State University T410170 sigma-aldrich.com LC Particle Innovation Leads the Way* Particle Size and Architecture (porosity) More
More informationOpen Column Chromatography, GC, TLC, and HPLC
Open Column Chromatography, GC, TLC, and HPLC Murphy, B. (2017). Introduction to Chromatography: Lecture 1. Lecture presented at PHAR 423 Lecture in UIC College of Pharmacy, Chicago. USES OF CHROMATOGRAPHY
More informationLC and LC/MS Column Selection Flow Chart
LC and LC/MS Column Selection Flow Chart To use the column selection diagram below, simply follow the path for your analyte and mobile phase. At the far right, follow your final column selection to the
More informationHigh Pressure/Performance Liquid Chromatography (HPLC)
High Pressure/Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) is a form of column chromatography that pumps a sample mixture or analyte in a solvent (known as the
More informationChapter 26. An Introduction to Chromatographic Separations. Chromatography
Chapter 26 An Introduction to Chromatographic Separations Chromatography 1 Chromatography-Model as Extraction Chromatography-Model as Extraction 2 Chromatography Planar Chromatography-Types paper chromatography
More informationTransferring Yesterdays Methods to Tomorrows Technology: The Evolution from HPLC to UPLC TM
Transferring Yesterdays Methods to Tomorrows Technology: The Evolution from HPLC to UPLC TM Eric S. Grumbach Thomas E. Wheat Chuck Phoebe Joe Arsenault Jeffrey R. Mazzeo Diane M. Diehl UPLC TECHNOLOGY
More informationCHEM 429 / 529 Chemical Separation Techniques
CHEM 429 / 529 Chemical Separation Techniques Robert E. Synovec, Professor Department of Chemistry University of Washington Lecture 1 Course Introduction Goal Chromatography and Related Techniques Obtain
More informationPDG.pdf G-20 CHROMATOGRAPHY 3 4 INTRODUCTION
1 2 3 4 5 INTRODUCTION G-20 CHROMATOGRAPHY 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 Chromatographic separation techniques are multi-stage separation methods in which the components
More informationIntroductory Separations
Introductory Separations General Figure Acknowledgements Colin F. Poole s- The Essence of Chromatography, Elsevier Science, 2003. C.F. Poole and S. A. Schuette, Contemporary Practice of Chromatography
More informationChem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS
Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS 1. List two advantages of temperature programming in GC. a) Allows separation of solutes with widely varying retention factors in a reasonable
More informationM > ACN > > THF
Method Development in HPLC Dr. Amitha Hewavitharana School of Pharmacy University of Queensland Method Development in HPLC References: D A Skoog Principles of instrumental analysis, 3 rd Edition Chapters
More informationLECTURE 2. Advanced Separation Science Techniques Present and Future Separation Tools
LECTURE 2 Advanced Separation Science Techniques Present and Future Separation Tools Jack Henion, Ph.D. Emeritus Professor, Analytical Toxicology Cornell University Ithaca, NY 14850 Lecture 2, Page 1 Contents
More informationChapter content. Reference
Chapter 7 HPLC Instrumental Analysis Rezaul Karim Environmental Science and Technology Jessore University of Science and Technology Chapter content Liquid Chromatography (LC); Scope; Principles Instrumentation;
More informationCHAPTER 6 GAS CHROMATOGRAPHY
CHAPTER 6 GAS CHROMATOGRAPHY Expected Outcomes Explain the principles of gas chromatography Able to state the function of each components of GC instrumentation Able to state the applications of GC 6.1
More informationChromatographic Separation
What is? is the ability to separate molecules using partitioning characteristics of molecule to remain in a stationary phase versus a mobile phase. Once a molecule is separated from the mixture, it can
More informationChemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets
Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets Liquid chromatography was developed by Tswett in early 1900 s and was shown to be a powerful separation
More informationSeparation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.
Separation of Explosives in EPA 833: Column Choices Optimize Speed, Resolution, and Solvent Use Application Environmental Authors John W. Henderson Jr. and William J. Long Agilent Technologies, Inc. 28
More informationPrep 150 LC System: Considerations for Analytical to Preparative Scaling
Andrew Aubin and Jo-Ann Jablonski Waters Corporation, Milford, MA, USA APPLICATION BENEFITS The Prep 150 LC System is an affordable, highly reliable system for preparative chromatography and is suitable
More informationTheory and Instrumentation of GC. Chromatographic Parameters
Theory and Instrumentation of GC Chromatographic Parameters i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this
More informationLEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT. Thomas Wenzel, Bates College. In-class Problem Set Extraction.
LEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT Thomas Wenzel, Bates College In-class Problem Set Extraction Problem #1 1. Devise a scheme to be able to isolate organic acids, bases
More informationGas Chromatography. Introduction
Gas Chromatography Introduction 1.) Gas Chromatography Mobile phase (carrier gas) is a gas - Usually N 2, He, Ar and maybe H 2 - Mobile phase in liquid chromatography is a liquid Requires analyte to be
More informationRapid Screening and Analysis of Components in Nonalcoholic Drinks Application
Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application Food Author Michael Woodman Agilent Technologies, Inc. 285 Centerville Road Wilgton, DE 1988-161 USA Abstract Soft drinks,
More informationChemistry Gas Chromatography: Separation of Volatile Organics
Chemistry 3200 Gas chromatography (GC) is an instrumental method for separating volatile compounds in a mixture. A small sample of the mixture is injected onto one end of a column housed in an oven. The
More informationChapter 26: An Introduction to Chromatographic Separations
Chapter 26: An Introduction to Chromatographic Separations Column Chromatography Migration Rates Distribution Contstants Retention Times Selectivity Factor Zone Broadening & Column Efficiency Optimizing
More informationIntroduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011
?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention
More informationIntroduction to Chromatography
Introduction to Chromatography Dr. Sana Mustafa Assistant Professor Department of Chemistry, Federal Urdu University of Arts, Science & Technology, Karachi. What is Chromatography? Derived from the Greek
More informationEasy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator
Technical Note 75 Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator Introduction The goal of every chromatographic optimization is a method that sufficiently resolves all
More informationChromatography Outline
Chem 2001 Summer 2004 Outline What is? The Chromatogram Optimization of Column Performance Why Do Bands Spread? Gas High-Performance Liquid Ion-Exchange 2 What is? In chromatography, separation is achieved
More informationSeparation Methods Based on Distributions in Discrete Stages (02/04/15)
Separation Methods Based on Distributions in Discrete Stages (02/04/15) 1. Chemical Separations: The Big Picture Classification and comparison of methods 2. Fundamentals of Distribution Separations 3.
More informationReverse Phase High Performance Liquid Chromatography method for determination of Lercanidipine hydrochloride in bulk and tablet dosage form
Research Article ISSN: 0974-6943 M.V.Kumudhavalli et al. / Journal of Pharmacy Research 2014,8(11), Available online through http://jprsolutions.info Reverse Phase High Performance Liquid Chromatography
More informationHigh Resolution Fast LC
High Resolution Fast LC Easier Than You Think Rita Steed LC Columns Application Engineer May 9, 2013 What is High Resolution Fast LC? Maintain Resolution with Faster Run Time Increased Resolution with
More informationPreparative Chromatography with Improved Loadability
05 LCGC RTH AMERICA VLUME 0 UMBER VEMBER 00 www.chromatographyonline.com Preparative Chromatography with Improved Loadability In combinatorial chemistry and drug development, as well as in the full-scale
More informationChromatography- Separation of mixtures CHEM 212. What is solvent extraction and what is it commonly used for?
Chromatography- Separation of mixtures CHEM 212 What is solvent extraction and what is it commonly used for? How does solvent extraction work? Write the partitioning coefficient for the following reaction:
More informationHigh Speed 2D-HPLC Through the Use of Ultra-Fast High Temperature HPLC as the Second Dimension
High Speed 2D-HPLC Through the Use of Ultra-Fast High Temperature HPLC as the Second Dimension Minnesota Chromatography Forum Spring Symposium Dwight Stoll and Peter W. Carr Department of Chemistry University
More informationThe Theory of HPLC. Quantitative and Qualitative HPLC
The Theory of HPLC Quantitative and Qualitative HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference
More informationFall 2012 Due In Class Friday, Oct. 19. Complete the following on separate paper. Show your work and clearly identify your answers.
CHEM 322 Name Fall 2012 Due In Class Friday, Oct. 19 Complete the following on separate paper. Show your work and clearly identify your answers. General Separations 1. Describe the relative contributions
More informationIntroduction to Capillary GC
Introduction to Capillary GC LC Columns and Consumables Simon Jones Chromatography Applications Engineer February 20, 2008 Page 1 Introduction to Capillary GC t r K c?? Kβ k = - tr t m? t m R s Page 2
More informationOMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (11) 04 QUALIFICATION OF EQUIPMENT ANNEX 1: QUALIFICATION OF HPLC EQUIPMENT Full document title and reference Document type
More informationBasic Principles for Purification Using Supercritical Fluid Chromatography
Basic Principles for Purification Using Supercritical Fluid Chromatography Jo-Ann M. Jablonski, Christopher J. Hudalla, Kenneth J. Fountain, Steven M. Collier, and Damian Morrison Waters Corporation, Milford,
More informationGas Chromatography. 1. Experiment Category: 2. Experiment Name: 3. Date and Issue number: 4. Instructor Name: 5. Institution: Ain Shams University
Project Title: e-laboratories for Gas chromatography 1. Experiment Category: Chemistry >> chromatography 2. Experiment Name: Gas Chromatography 3. Date and Issue number: 4. Instructor Name: 5. Institution:
More information637. Thiamethoxam. HPLC method
637. Thiamethoxam HPLC method CIPAC Collaborative Trial according to CIPAC Information Sheet N o 293 Dr. Sven Adolph Syngenta Crop Protection Münchwilen AG CH-4333 Münchwilen Switzerland May 212 page 1
More informationAlternatives to Sub-2 µm UHPLC Columns
Alternatives to Sub-2 µm UHPLC Columns J. J. DeStefano, B. E. Boyes, S. Schuster, W. L. Miles, and J. J. Kirkland Advanced Materials Technology, Inc. 3521 Silverside Rd., Quillen Bld., Ste. 1-K Wilmington,
More informationLiquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.
High performance liquid chromatography (HPLC) is a much more sensitive and useful technique than paper and thin layer chromatography. The instrument used for HPLC is called a high performance liquid chromatograph.
More informationInstrumental Chemical Analysis
L2 Page1 Instrumental Chemical Analysis Chromatography (General aspects of chromatography) Dr. Ahmad Najjar Philadelphia University Faculty of Pharmacy Department of Pharmaceutical Sciences 2 nd semester,
More informationHPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application
HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application Pharmaceuticals Author Cliff Woodward and Ronald Majors Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808
More informationCHROMATOGRAPHIC SEPARATION TECHNIQUES SUPERCRITICAL FLUID CHROMATOGRAPHY
2.2.45. Supercritical fluid chromatography EUROPEAN PHARMACOPOEIA 7.0 Control solutions. In addition to the TOC water control, prepare suitable blank solutions or other solutions needed for establishing
More informationUPLC Method Development and Validation
UPLC Method Development and Validation Rev. 2 2008 Waters Corporation Challenges of Method Development Methods are developed throughout the drug development process Samples vary in complexity Redundancy
More informationHigh Performance Liquid Chromatography
STANDARDBASE techniques: High Performance Liquid Chromatography Drenthe College, The Netherlands 1. Introduction HPLC. High Performance Liquid Chromatography High Performance Liquid Chromatography (HPLC)
More informationHigh Performance Liquid Chromatography
Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN
More informationHigh Performance Liquid Chromatography
Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 1 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN
More informationP R O D U C T B U L L E T I N. Fused-Core particle technology for hyper-fast and super-rugged HPLC columns
P R O D U C T B U L L E T I N Fused-Core particle technology for hyper-fast and super-rugged HPLC columns HALO column packings are not made the typical way. Instead, the particles packed into HALO columns
More informationfor Acclaim Mixed-Mode HILIC-1 Column
for Acclaim Mixed-Mode HILIC-1 Column Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 1 of 17 Product Manual for ACCLAIM Mixed-Mode HILIC-1 Column 5µm, 4.6 x 250mm, P/N 066844 5µm, 4.6 x 150mm, P/N
More informationRoutine MS Detection for USP Chromatographic Methods
Daniel S. Root, Thomas E. Wheat, and Patricia McConville Waters Corporation, Milford, MA, USA APPLICATION BENEFITS This approach enables mass spectral analysis of peaks directly from unmodified USP HPLC
More informationBiochemistry. Biochemical Techniques HPLC
Description of Module Subject Name Paper Name 12 Module Name/Title 13 1. Objectives 1.1. To understand the basic concept and principle of 1.2. To understand the components and techniques of 1.3. To know
More informationChapter 31 Gas Chromatography. Carrier Gas System
Chapter 31 Gas Chromatography GAS-LIQUID CHROMATOGRAPHY In gas chromatography, the components of a vaporized sample are fractionated as a consequence of being partitioned between a mobile gaseous phase
More informationH 3 CO H 3 CO S CH 3
FENITROTHION 35 H 3 CO P H 3 CO S O CH 3 NO 2 ISO common name Chemical name Fenitrothion O,O-Dimethyl O-4-nitro-m-tolyl phosphorothioate (IUPAC) O,O-Dimethyl O-(3-methyl-4-nitrophenyl)- phosphorothioate
More informationThe Use of ACQUITY UPLC in Pharmaceutical Development
The Use of ACQUITY UPLC in Pharmaceutical Development MAY 2005 SEPARATION SCIENCE REDEFINED 15 Now more than ever, the demands of the pharmaceutical industry require companies to look for new ways to cut
More informationChemistry Instrumental Analysis Lecture 31. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 31 High Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) Solvent Delivery
More information"Theory and Practice of High Speed Chromatography for Bioanalysis" Stuart Coleman March 20, 2007
"Theory and Practice of High Speed Chromatography for Bioanalysis" Stuart Coleman March 20, 2007 General Observations about HPLC Separations A good separation is necessary for good bioanalytical quantitation.
More informationUtilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification
Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography Tips and Techniques to Optimize ELSD and MS based Purification Utilizing ELSD and MS as Secondary Detectors for Prep
More informationSIMULTANEOUS RP HPLC DETERMINATION OF CAMYLOFIN DIHYDROCHLORIDE AND PARACETAMOL IN PHARMACEUTICAL PREPARATIONS.
Ind. J. Anal. Chem Vol. 7 11. 2008 SIMULTANEOUS RP HPLC DETERMINATION OF CAMYLOFIN DIHYDROCHLORIDE AND PARACETAMOL IN PHARMACEUTICAL PREPARATIONS. Authors for correspondence : R. R. Singh1*, M. V. Rathnam,
More informationDETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY
CHAPTER 9 DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 9 Determination of drug release during
More informationIntroduction to Capillary GC
?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention
More informationDilution(*) Chromatography
WA20264 Poster # 184, HPLC 2002, Montreal, 4-5 June 2002 At-Column Column-Dilution for Preparative Dilution(*) Chromatography Cecilia Mazza, Jie Cavanaugh, Ziling Lu,Tom Sirard,Tom Wheat and Uwe Neue Waters
More informationLC III: HPLC. Originally referred to as High-Pressure Liquid Chromatography. Now more commonly called High Performance Liquid Chromatography
LC III: HPLC What is HPLC? Originally referred to as High-Pressure Liquid Chromatography Now more commonly called High Performance Liquid Chromatography In general: The instrument controlled version of
More informationCourse goals: Course goals: Lecture 1 A brief introduction to chromatography. AM Quality parameters and optimization in Chromatography
Emqal module: M0925 - Quality parameters and optimization in is a separation technique used for quantification of mixtures of analytes Svein.mjos@kj.uib.no Exercises and lectures can be found at www.chrombox.org/emq
More informationProduct Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns
Product Bulletin HPLC Columns ACE LC/MS and Rapid Analysis HPLC Columns 0 mm, 30 mm, 35 mm and 50 mm column lengths.0,., 3.0, 4.0 and 4.6 mm column diameters Configured for High Sample Throughput Specially
More informationVolatile organic compounds (VOCs):
Volatile organic compounds (VOCs): Organic chemicals with a high vapour pressure at room temperature. High vapour pressure results from a low boiling point. The World Health Organization (WHO) defined
More information[S016. CHROMATOGRAPHY]
Phyto-Analysis Sheet Number : 16 Prof. Dr. Talal Aburjai Page 1 of 9 How to read the chromatogram? Comes from any automated chromatography. The chromatograms show the 0 t (t m ) which indicates the solvent
More informationDetermination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection
Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection Mark Tracy, Xiaodong Liu, and Ian Acworth, Thermo Fisher Scientific, Sunnyvale, CA,
More informationDATES: LAB: Liquid Chromatography Separation of Grape Kool-Aid
NAME: AP CHEMISTRY DATES: LAB: Liquid Chromatography Separation of Grape Kool-Aid PURPOSE There are a number of analytical techniques used to separate components of a mixture, or solution. They include
More informationDetermination of Polymer Modifier in Asphalt
Standard Method of Test for Determination of Polymer Modifier in Asphalt AASHTO Designation: T xxx-xx (2005) 1. SCOPE 1.1. This method of test is used to determine the polymer content of an asphalt sample.
More informationHPLC COLUMNS WILEY-VCH. Theory, Technology, and Practice. Uwe D. Neue with a contribution from M. Zoubair El Fallah
HPLC COLUMNS Theory, Technology, and Practice Uwe D. Neue with a contribution from M. Zoubair El Fallah WILEY-VCH New York Chichester Weinheim Brisbane Singapore Toronto CONTENTS Preface ix 1 Introduction
More informationDetermination of Retention Factors of Aromatic Compounds by Gradient-Elution Reverse-Phase High Performance Liquid Chromatography
Korean J. Chem. Eng., 19(6), 978-985 (22) Determination of Retention Factors of Aromatic Compounds by Gradient-Elution Reverse-Phase High Performance Liquid Chromatography Ju Weon Lee and Kyung Ho Row
More informationThe Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times
The Secrets of Rapid HPLC Method Development Choosing Columns for Rapid Method Development and Short Analysis Times Rapid Analysis Is More Than Run Time It is developing a method to meet a goal and developing
More informationApplication Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China
Rapid Deteration of Eight Related Aromatic Acids in the p-phthalic Acid Mother Liquid Using an Agilent 126 Infinity LC System and an Agilent Poroshell 12 SB-C18 Column Application Note Author Xinlei Yang
More informationLC Column Troubleshooting. Do you have an equivalent column for my LC column?
LC Column Troubleshooting Do you have an equivalent column for my LC column? LC columns from different suppliers may have very different retention properties even if the bonding is the same. Although there
More informationWhat is Chromatography?
What is Chromatography? Chromatography is a physico-chemical process that belongs to fractionation methods same as distillation, crystallization or fractionated extraction. It is believed that the separation
More information