SPECIFICATION SHEET Model 2A
|
|
- Peregrine Blake
- 5 years ago
- Views:
Transcription
1 SPECIFICATION SHEET Model 2A MEASUREMENT UNITS CENTIGRADE ( C), FARENHEIT ( F) & PPM (V) SENSOR TYPE Aluminium Oxide Ultra High Capacitance Sensor RANGE -80 C to -20 C dew point AUTOMATIC CALIBRATION Via potentiometer on front panel. POWER SUPPLY Mains: 110/240v AC & 9V DC - Six C type disposable batteries. BATTERY LIFE In excess of 150 hours continuous operation. WARM UP TIME Less Than 60 Seconds SENSOR CALIBRATION ACCURACY ±2 C dewpoint. - All units supplied with certificates documenting factory calibration against known moisture levels traceable to National & International humidity Standards. REPEATABILITY ± 0.5 C dewpoint. OPERATING CONDITIONS Temperature: 0 C to +50 C Humidity: 0-98% RH, Non-condensing Storage Temperature: -30 C to +70 C Operating Pressure for Measurement Head: Atmospheric to 0.5 barg. Max. Inlet Pressure: 28bar max. ELECTROMAGNETIC COMPATIBILITY (EMC) Immunity: Complies with EN : 1992 Emissions: Complies with EN : 1992 FLOW RATE TO SENSOR 3 to 20 litres/minute CONNECTIONS Swagelok stainless steel coupling, 0.25 overall diameter compression tube fitting. WEIGHTS & DIMENSIONS 8.5 Kgs Net / 12.0 Kgs Gross 409 x 425 x 198 mm WARRANTY One year from date of delivery against faulty materials or workmanship. TYPICAL RESPONSE TIME 95% of reading within 20 seconds in normal operation.
2 Model DS2000 Single Channel Hygrometer A complete on-line solution for continuous measurement of dewpoint in process gas and compressed air, offering powerful functionality via a simple interface. Advanced functionality with user-friendly design Accuracy ±2 C Dewpoint Automatic Calibration RS485 serial communications 2 Independent, fully programmable alarm relays Linear retransmission signal of current or voltage Cable Length of up to 1km to sensor Calibration Traceable To National & International Standards The dewtec Model DS2000 from Alpha Moisture Systems is a single channel, online hygrometer designed to accurately measure dewpoint in process gases and dry compressed air. This panel mounted DIN style dewpoint meter combines advanced digital technology with the industry leading ultra high capacitance aluminium oxide sensors - the acknowledged industry benchmark for trace moisture measurement in process gases. The combination of advanced display electronics and full specification sensors makes this the unit of choice for a wide range applications due to it's flexibility and high configurability and ease of use through its menu driven operation.the instrument can be factory configured to order to read in C dewpoint, F dewpoint, ppm(v), g/m³ or lb/mmscf. The measured dewpoint is displayed by means of a large 5½ digit, seven segment LED display. Model DS2000 also offers two independent, fully user programmable alarm contacts and open collector outputs to provide all the versatility required for any application. The factory set, analogue, current or voltage output, that is linear with dewpoint, is available as well as RS485 serial connection with 1-99 address settings. The dewtec DS2000 is a fully self-contained display system in 1/8 DIN panel mounted case providing front panel protection to IP66 (NEMA 4). The dewtec DS2000 can be supplied together with any one of a range of dedicated, ultra high capacitance, sensors incorporating automatic calibration and covering an overall range from -110ºC to +20ºC dewpoint. The sensor has been conditioned for long term stability so ensuring a long and trouble free service life. Additionally the high sensitivity and linear output of the sensor allow up to 1km cable length between sensor and dewtec DS2000 instrument. The sensor is ATEX certified for intrinsic safety so can be used in hazardous environments with a suitable Zener barrier in the safe area. The long lasting sensing element offers excellent sensitivity, repeatability and response speed - especially on dry down. Each unit is calibrated to ±2 ºC accuracy and is supplied with a calibration certificate traceable to International Humidity Standards. The dewtec DS2000 incorporates two fully programmable alarm relays to control external devices, including process high, process low, direct acting, reverse and logical OR. The alarms can be independently set across the full operating range of the sensor with variable hysteresis for the chosen engineering unit. The dewtec Model DS2000 Dewpoint Hygrometer is supplied fully configured according to the customer's specifications and ready to use, with calibration certification traceable to national and international standards, instruction manual and 2 meters of connecting cable as standard. An optional sensor holder can be ordered with the instrument. Additional sensor cable upto 1000m can be ordered separately. Alpha Moisture Systems Alpha House, 96 City Road Bradford, BD8 8ES England BSI UKAS QUALITY MANAGEMENT OO3 Tel Fax Website +44 (0) (0) info@amsystems.co.uk DS2000/08/03
3 Model DS2000 DEWPOINT METER Specifications Model DS Dewpoint Hygrometer DISPLAY Type: Red/green, 7-segment, 5-digit primary display, 1-digit secondary display. Height: 18mm (0.71inches) primary display, 7mm (0.3 inches) secondary display Annunciators: Alarm 1 and Alarm 2 status SENSOR Dewpoint Range: From 110 C to +20 C Accuracy: ± 2 C over full range Operating Temperature: 10 C to +50 C Storage Temperature: 50 C to +70 C Operating Humidity: 10-90% RH non-condensing Mechanical Connections: M14x1.25mm pitch Electrical Connection: Male coaxial plug Approvals: CE, Atex - HYGROMETER INPUT Sampling Rate: 10Hz Resolution: 14-bits Sensor short circuit detection: Detected within two seconds. All alarms become active TRANSISTOR OUTPUTS Type: NPN open collector. Output 1 tied to Alarm 1. Output 2 tied to Alarm 2. RELAY OUTPUT 1 Contact Type: Single pole, double throw Rating: 2A resistive, 120/240V, 50/60Hz Lifetime: >500,000 operations at rated voltage/current Isolation: Inherent RELAY OUTPUT 2 Contact Type: Single pole, double throw Rating: 2A resistive, 120/240V, 50/60Hz Lifetime: >500,000 operations at rated voltage/current Isolation: Inherent Dimensions LINEAR (RE-TRANSMITTED) OUTPUT (VOLTAGE OR CURRENT IS A FACTORY SET OPTION) Ranges Available: 0-5V, 1-5V, 0-10V, 2-10V, 0-20mA and 4-20mA Accuracy: ±0.5% max of display value Load Impedance: 500. max for ma ranges, 500. min for voltage ranges SERIAL COMMUNICATIONS Type: Serial asynchronous, UART to UART Data Format: Open ASCII; one start bit, even parity, seven data bits and one stop bit. Physical Layer: RS485 Maximum Number of zones: 99 Baud Rate: Selectable 1200, 2400, 4800 or 9600 PERFORMANCE Reference Conditions Supply Voltage: V, 50Hz Power Consumption: 7.5VA max Operating Conditions Ambient Temperature: 0 C to +55 C Relative Humidity: 10 95% non-condensing ENVIRONMENTAL Approvals: CE EMC: Certified to EN61326 Safety: Complies with EN Front Panel Seal: To IP66 PHYSICAL Dimensions: Height 48mm Width 95mm Depth 110mm without sensor connector. Weight 0.21 kg max. Mounting: Panel-mount; press-fit fixing. Panel cut out 45mm x 92mm. Terminals: Screw type Ordering Information Process Indication Alarm LED s (3.94 ) 10.1 (0.4 ) 6.0 (0.25 ) Maximum Model DS2000- [XX] - [X] - [X] - [XX] - [XX] - [X] dewtec Function Keys P Function Identification Sensor Cable Connection Socket Autocal Potentiometer Input / Output Terminals R ALM 1 ALM (1.77 ) 91.6 (3.60 ) RS232 Port Clamp Locking Ratchets Sealing Gasket Panel Clamp 48.0 (1.89 ) Mounting Panel 96.0 (3.78 ) RANGE = XX SR: -110 C to -20 C (-166 F to -4 F) dewpoint PL: -100 C to 0 C (-148 F to +32 F) dewpoint RD: -80 C to -20 C (-112 F to +32 F) dewpoint GY: -80 C to 0 C (-112 F to +32 F) dewpoint BL: -80 C to +20 C (-112 F to +68 F) dewpoint DISPLAY UNITS = X C: C Dewpoint F: F Dewpoint V: ppm(v) L: lb/mmscf G: g/m³ RETRANSMISSION OUTPUT = X V: Voltage Output A: Current Output ALARM 1 CONFIGURATION = XX HD: High Alarm, Direct action (normally de-energised relay) HR: High Alarm, Reverse action (normally energised relay) LD: Low Alarm, Direct action (normally de-energised relay) LR: Low Alarm, Reverse action (normally energised relay) Your Local Distributor ALARM 2 CONFIGURATION = XX HD: High Alarm, Direct action (normally de-energised relay) HR: High Alarm, Reverse action (normally energised relay) LD: Low Alarm, Direct action (normally de-energised relay) LR: Low Alarm, Reverse action (normally energised relay) SENSOR HOLDER N: None 4: 1/4 Swagelok Stainless Steel Compression Fitting 8: 1/8 Swagelok Stainless Steel Compression Fitting 6: 6mm Swagelok Stainless Steel Compression Fitting Example: To order device with range -80 C to -20 C dewpoint, with retransmission set to current output, both alarms to trip on rising dewpoint with normally de-energised relays and sensor holder with 1/4 Swagelok fittings, order as : Model DS RD - C - A - HD - HD - 4
4 a alphamoisture Model SADPminiEx S Y S T E M S Intrinsically Safe Hand-Held Dewpoint Hygrometer Certified Intrinsically Safe for use in explosive atmospheres: II 1G EEx ia IIC T4 (Ta -20ºC to +50ºC) User Selectable Units - ºC, ºF, ppm(v), ppb(v), ppm(w), g/m3, lbs/mmscf Desiccant Dry Down Assembly for Quick Measurements Desiccant Dry Down Technology sample gas inlet Dry-Down Position sensor Sensor surrounded is by desiccant, ensuring that it is dry, when not in use. sample exhaust Unique design ensures purging of both void space and all chamber wall surfaces by sample gas prior to taking a reading Read Position Real Time Graphic Logging to PC Advanced DATA Logging & PC Download RS485 Serial Communication 4-20mA Analogue Output Electronic Automatic Calibration Pressure Correction Computation Various Ranges between -110ºC to + 20ºC dewpoint True Hand Held Portable Device Fully Self-Contained User Friendly Multi-Level Security to Prevent Unauthorised sensor exposed to gas sample The desiccant is always isolated from atmospheric air, and the sensor to ensure an accurate reading Desiccant Head assembly Keeping the sensor dry between tests ensures that the SADPmini is always ready to carry out rapid spot checks. The unique design of the Desiccant Head achieves this by surrounding the sensor with desiccant when taking a measurement. At no time is the sensor allowed to come into contact with ambient air. The chamber is also designed so that the void space and chamber wall surfaces are purged with sample gas, before exposure of the sensor, so giving faster, more accurate and reliable results. Alpha Moisture Systems Alpha House, 96 City Road Bradford, BD8 8ES England Tamper Robust Ergonomically Designed Custom Housing Tel BSI Fax UKAS QUALITY MANAGEMENT OO3 Website +44 (0) (0) info@amsystems.co.uk SADPminiEX/07/05
5 115 MODEL SADPminiEx - Intrinsically Safe Hand-Held Dewpoint Hygrometer Specifications TYPE B SENSING ELEMENT: Ultra Hi-Capacitance Aluminium Oxide Type INTRINSIC SAFETY: Standard : EN 50014:1997+A1+A2 / EN 50020:2002 / EN 50284:1999 Code : II 1G EEx ia IIC T4 (Ta -20ºC to +50ºC) RANGE IN DEW POINT: SR: -110ºC to -20ºC (-166ºF to -4ºF ) dew point PL: -100ºC to 0ºC (-148ºF to +32ºF ) dew point RD: -80ºC to -20ºC (-112ºF to -4ºF) dew point GY: -80ºC to 0ºC (-112ºF to +32ºF ) dew point BL: -80ºC to +20ºC (-112ºF to +68ºF ) dew point DISPLAY UNITS: ºC - Degrees Centigrade Dew / Frost Point ºF - Degrees Fahrenheit Dew / Frost Point ppm(v) - Parts per million (volume) ppb(v) - Parts per billion (volume) ppm(w) - Parts per million (weight) g/m 3 - Grams per cubic metre lbs/mmscf - Pounds per million standard cubic feet DISPLAY: Blue on Green, 128 x 64 pixel, Graphical LCD with LED backlight. SENSOR CALIBRATION ACCURACY: ± 2 C Dew point. Each unit supplied with a Certificate of Calibration, traceable to National & International Standards - National Physical Laboratory (UK) / NIST (USA). AUTOMATIC CALIBRATION: Electronic "Span Check". Performed by user following simple menu driven instructions. Can be password protected to avoid unauthorised tamper. REPEATABILITY: Better than ±0.2ºC dew point POWER SUPPLY: 3.6v, 2.3Ah Rechargeable Li-Ion Battery. Battery charger included. BATTERY LIFE: In excess of 80 hours of continuous use on full charge at 20ºC/68ºF. TIME TO FULL CHARGE: 14 hours KEYBOARD: 5 Membrane covered, metallic dome tactile keys. PRESSURE CORRECTION: Integral calculator to display pressure dew points. Gauge pressure can be entered in kpa, kg/cm2, bar or psi. TEMPERATURE COEFFICIENT: Temperature compensated for operating range. GAS SAMPLE CONNECTIONS: Ports accept Swagelok VCO type coupling (9/16" x 18 UNF). Supplied with either 6mm or 1/4" Swagelok SS compression fitting on one side. The other side is fitted with a stainless steel push-on, "fir tree" type, hose connector for 6mm ID tube.each unit is supplied with 2m length, 6mm ID, PTFE tube. ELECTRICAL CONNECTIONS: 9 Pin "D" type for 4-20mA analogue output, RS485 Serial Communications and PC interface. Separate socket for battery charger. OPERATING TEMPERATURE: -10 C to +50 C STORAGE TEMPERATURE & HUMIDITY: -40 C to +80 C / 95% RH Noncondensing OPERATING PRESSURE: Maximum 30kPag ( 0.3barg / 4 psig). OPERATING HUMIDITY (External): 95% RH Non-condensing TYPICAL RESPONSE TIMES: Wet to Dry: -10 C to -60 C - less than 120 seconds Dry to Wet: -110 C to -20 C - less than 20 seconds SAMPLE FLOW RATE: Flow independent, but ideally 2 to 5 litres per minute. Max: 10 litres/min. REPLACEMENT DESICCANT: Field Interchangeable. SENSOR LIFE: Between 5 & 10 years - depending on application. REPLACEMENT SENSOR: Field Interchangeable. ELECTROMAGNETIC COMPATIBILITY (EMC): Product complies with the objectives and requirements of EMC Directive 89/336/EEC, amended 92/31/EEC. SECURITY: Multi level password protection. WARM UP TIME: 10 seconds WEATHERPROOF CLASSIFICATION: IP54 / NEMA12 WARRANTY: 2 years for faulty workmanship and defective parts. WEIGHT: 1.2kgs (2.6lbs) DIMENSIONS: 192 x 97 x 115 mms (7.5 x 3.8 x 4.5 inches) MATERIALS OF CONSTRUCTION: Sensor in metal housing. Outer case custom manufactured in,stainless steel impregnated, high impact Polybutylene Terephthalate (PBT). ACCESSORIES INCLUDED: 2m PTFE Sampling pipe, Universal Battery Charger, User Manual. TYPE F As TYPE B plus following additional specifications & features: ANALOGUE OUTPUT: Externally powered 4-20mA loop. Linear output with unit selected. Span easily configured by user. Max. load = 50 x (VEXT - 0.6) VEXT = Supply voltage. (Eg. For 24V supply, Max. load = 1065Ω ACCESSORIES INCLUDED: 4-20mA Connector for analogue output. TYPE L As TYPE B plus following additional specifications and features: DATA LOGGING: 16,000 samples. Date and time stamped data, stored in chosen units of measurement, for download to PC. DATA LOCATION: 20 separate locations (Tags) can be entered (alpha-numerical) by user for data collection at pre-programmed locations. SAMPLING RATES: User selectable (in intervals of 6 seconds) from once every 6 seconds to once a day. DATA DISPLAY: Numerical and Graphical display of data on SADPmini screen. REAL TIME RECORDING: Device can be programmed to monitor, record and graphically present data in real time directly to PC. SERIAL COMMUNICATIONS: RS485, baud rate half duplex. ANALOGUE OUTPUT: Externally powered 4-20mA loop. Linear output with unit selected. Span easily configured by user. Max. load = 50 x (VEXT - 0.6) VEXT = Supply voltage. (Eg. For 24V supply, Max. load = 1065Ω) ACCESSORIES INCLUDED: 4-20mA / RS485 Connector for analogue output and serial communications. Serial or USB (Isolated / self powered) interface, with cables, for "Real Time Logging" and data download to PC. Software supplied on CD. SYSTEM REQUIREMENT: Windows 95 or later for Serial connection & Windows 2000 or later for USB connection to PC. OPTIONAL ACCESSORIES Hard wearing, padded nylon carrying case, custom manufactured with carry handle, belt loop and adjustable shoulder strap. DIMENSIONS Model SADPminiEx AUTOMATIC DEWPOINT HYGROMETER I O Product specification maybe subject to change without prior notice as a part of our ongoing product development programme. SADPminiEx/07/05
6 MODEL SADPminiEx Intrinsically Safe Hand-Held Dewpoint Hygrometer Ordering Information Model SADPminiEx - [XX] - [X] - [X] - [XX] - [X] RANGE = XX SR: -110 C to -20 C (-166 F to -4 F) dewpoint PL: -100 C to 0 C (-148 F to +32 F) dewpoint RD: -80 C to -20 C (-112 F to -4 F) dewpoint GY: -80 C to 0 C (-112 F to +32 F) dewpoint BL: -80 C to +20 C (-112 F to +68 F) dewpoint PC CONNECTION = XX S: Serial Port (Windows 95) U: USB (Windows 2000 or later) (Choose only for Type L. Leave blank for Type B or Type F) TYPE = X B: Standard F: With 4-20mA Analogue Output L: With logging, RS485 Comms. & 4-20mA Output MAINS PLUG TYPE- For Battery Charger = XX UK: Standard UK, 3 Square pin US: 2 Vertical Flat Blade EU: 2 Round Pins AU: 2 Oblique FlatBlades with Ground (see diagram below) SAMPLE CONNECTION = X 4: 0.25 Swagelok SS Compression Type 6: 0.6 mm Swagelok SS Compression Type UK Type Example: To order device with range -80 C to 0 C dewpoint, with logging capabilities, 6mm pipe connection, for use in UK and to be used with a USB type PC port, order as: US Type Model SADPminiEx - GY-L-6-UK-U EU Type OPTIONAL ACCESSORIES Carrying Case for SADPminiEx Order as: Model SADPminiEx - CC Australian Type Alpha Moisture Systems Alpha House, 96 City Road Bradford, BD8 8ES England BSI UKAS QUALITY MANAGEMENT OO3 Tel Fax Website +44 (0) (0) info@amsystems.co.uk SADPminiEX/07/05
7 FEATURES 3 Wire, 4-20mA Dewpoint Transmitter ATEX Certified Intrinsically Safe: II 1 GD For Gas - Ex ia IIC T4 (IP66) For Dust - Ex iad IIC T4 20 T135ºC (North American Equivalent: Class 1 Division 1 (Zone 0), Groups A&B T4 NEMA4X) Overall Range: -120ºC to +20ºC Dewpoint Can be configured for Linear 4-20mA signal in: ºC & ºF Dewpoint, ppm(v), g/m 3, lb/mmscf Accuracy ± 2ºC Dewpoint AUTOMATIC CALIBRATION Temperature Compensation Long Term Stability Fast Response Model AMT-Ex Intrinsically Safe Dewpoint Hygrometer is a three wire 4-20mA transmitter, certified safe for use in potentially explosive environments, for the continuous measurement of humidity in a process gas or compressed air. It is certified intrinsically safe to ATEX Ex ia IIC T4 for gas zones and ATEX Ex iad IIC T4 20 for dust zones. Powered from an approved galvanically isolated barrier, providing a safe 12v to 28v dc source, the unit is designed to output a safe and linear 4-20mA signal corresponding to the dewpoint temperature. The Model AMT-Ex transmitter can also be factory configured to output a 4-20mA linear signal for any other moisture unit e.g. ppm(v), ppm(w), g/m 3, lb/mmscf etc. Model AMT-Ex is certified intrinsically safe to ATEX Ex ia IIC T4 for gas zones and ATEX Ex iad IIC T4 20 for dust zones. The long lasting sensing element offers excellent sensitivity, repeatability and response speed. Each unit is supplied with a Certificate of Calibration, traceable to International Humidity Standards, validating accuracy to ±2ºC dewpoint. The transmitter also incorporates an Automatic Calibration (AUTOCAL) feature, which allows the user to carry out a field calibration / span check. AUTOCAL feature is operated by means of two small button switches built within the transmitter enclosure. To avoid tamper, the switches are locked in normal use and can only be activated by following a special routine. The integral RISC microprocessor circuitry of the Model AMT-Ex transmitter allows high resolution with advanced selfdiagnostics for fault conditions. It also enables periodic re-calibration of the moisture sensor, storing calibration data within the fully self contained unit. The Model AMT-Ex transmitter can also be factory set to provide a linear 4-20mA signal over a chosen range. The mechanics of the Model AMT-Ex transmitter have been designed to cope with extreme environmental conditions. The rugged stainless steel construction offers protection to IP66 (NEMA4X), with transmitter electrical connections made via secure industrial connector. Model AMT-Ex transmitter can withstand 35,000 kpa (350bar) maximum pressure and by employing a low resistance cable, the transmitter can be located significant distances, in excess of 1000 metres, from the control room. Designed with the operator in mind, for reliable and accurate measurements, Model AMT-Ex is extremely easy to install and operate, requiring little or no maintenance. Alpha Moisture Systems Alpha House, 96 City Road Bradford BD8 8ES England CERTIFICATE No.FM35600 BS EN ISO 9001:2000 SS-AMT-Ex1.0 05/08
8 SPECIFICATIONS MODEL AMT-Ex Dewpoint Hygrometer 4-20mA Loop Powered Transmitter TRANSMITTER TYPE: 3 Wire, 4-20mA Transmitter OUTPUT SIGNAL: 4 to 20mA Linear SENSING ELEMENT: Ultra High Capacitance - Aluminium Oxide Type RANGES: SD: -100 C to +20 C (-148ºF to +68ºF) dewpoint LD: -120 C to +20 C (-184ºF to +68ºF) dewpoint HD: -65 C to +20 C (-85ºF to +68ºF) dewpoint SP: 1 to 1000 ppm(v) MP: 0.1 to ppm(v) LP: 0.01 to ppm(v) PB: 1 to 1000 ppb(v) HG: 0.01 to g/m 3 LG: to g/m 3 LB: 0.1 to lbs/mmscf 1ppm resolution 0.1ppm resolution 0.01ppm resolution 1ppb resolution 0.01g/m 3 resolution 0.001g/m 3 resolution 0.1 lbs/mmscf resolution NOTE: ppm, ppb, g/m 3 & lbs/mmscf ranges are stated at atmospheric pressure. Model AMT Transmitter can be configured to provide corresponding ranges for operation at other constant pressure. This must be stated at time of order. AUTOCAL: Field calibration / Span check facility. Activated and operated via two button switches on the Transmitter body. Locked in normal operating mode to avoid tamper. FACTORY CALIBRATION: Supplied with Certificate of Calibration traceable to NPL / NIST ACCURACY: ± 2 C dewpoint (NPL / NIST traceable for range -90 C to +20 C) TEMPERATURE COEFFICIENT: Temperature compensated for operating range. RESOLUTION: 6 micro-amps REPEATABILITY: Better than ±0.3ºC dewpoint OPERATING TEMPERATURE: -20 C to +60 C STORAGE TEMPERATURE: -20 C to +70 C OPERATING PRESSURE: From 1kPa (0.01 bara) to Maximum 35,000kPa ( 350 bara) OPERATING HUMIDITY (External): Maximum - 95% RH Non-condensing SAMPLE FLOW RATE: Flow independent but ideally 2 to 5 litres per minute. Max: 25 litres/min. CABLE TERMINATIONS: IP66 (NEMA4X) rated, size C, DIN EN connector at the Transmitter and other end terminated with bootlace ferrules. CABLE: Supplied with 2m standard cable. Nominal diameter 3.4mm, 92ohms/km at 20ºC ELECTROMAGNETIC COMPATIBILITY (EMC): Immunity: Complies with EN :2001 Emissions: Complies with EN :2000 WARM UP TIME: 10 seconds FAULT CONDITIONS: Open Circuit : Output drives to 20.50mA Short Circuit : Output drives to 20.75mA System Error: Output drives to 21.00mA ISOLATION: Sensing Element connected to the 4-20mA loop but isolated from Transmitter body. TRANSMITTER ENCLOSURE: 316 Stainless steel body with size C, DIN EN connector. SENSOR PROTECTION: 316 Sintered stainless steel filter - 50 micron PROBE MATERIAL (WETTED PARTS): 316 Stainless Steel WEATHERPROOF CLASSIFICATION: IP66 / NEMA4X when Connector mated to Transmitter. MECHANICAL CONNECTION: 3 /4 UNF (16tpi) with integral Viton O ring seal. MECHANICAL WARRANTY: 24 months in case of faulty workmanship and defective parts. CALIBRATION WARRANTY: 12 months. 0.1 C / F resolution ORDERING INFORMATION Dewpoint Hygrometer Model AMT-Ex - [XX] Supplied with connector and 2m cable XX = Choose required range: SD: -100ºC to +20ºC (-148ºF to +68ºF) dewpoint LD: -120ºC to +20ºC (-184ºF to +68ºF) dewpoint HD: -65ºC to +20ºC (-85ºF to +68ºF) dewpoint SP: 1 to 1000 ppm(v) MP: 0.1 to ppm(v) LP: 0.01 to ppm(v) PB: 1 to 1000ppb(v) HG: 0.01 to g/m 3 LG: to g/m 3 LB: 0.1 to lbs/mmscf E.g. Transmitter with range 100 to +20 ºC dewpoint, order: Dewpoint Hygrometer Model AMT-SD Connection Cable Model CAB-AMT- [XXX] XXX =Choose required length of Cable Leave Blank: 2m (Included with Model AMT) XXX: Insert cable length in metres (Available in 10m, 15m, 25m, 50m, 75m & 100m lengths) E.g. For 60 metres of cable, order: Model CAB-AMT-75 Model SH-AMT Sensor Holder (Optional) (Flow Through Sample Cell) MATERIAL OF CONSTRUCTION: 316 grade stainless steel. PORTS: Inlet & oulet both 1 /8 NPT F PIPE FITTINGS: 2 x Swagelok male pipe fittings. ( ¼, 1 /8 or 6mm OD - inlet / outlet fitting size to be stated at time of order) MOUNTING PLATE: 68 x 54 x 3mm Passivated mild steelplate with 2 fixing holes - 5.5mm diameter. WEATHERPROOF CLASSIFICATION: IP66 / NEMA4X with Model AMT transmitter installed. WEIGHT: 500grams (includes 2 fittings & mounting plate) ORDERING INFORMATION Sensor Holder Model SH-AMT - [X]-[X] Supplied with Mounting Plate X = Choose Inlet Pipe Fitting 0: None 4: ¼ 8: 1 / 8 6: 6mm X = Choose Outlet Pipe Fitting 0: None 4: ¼ 8: 1 / 8 6: 6mm E.g. For Sensor Holder with two ¼ fittings, order: Model SH-AMT-4-4 Product specification may be subject to change, without prior notice, as part of our ongoing product development programme. SS-AMT1.2-09/06
Model DS1200-AMT-Ex Intrinsically Safe 4-20mA ATEX Certified Dewpoint Hygrometer Ranges available between -120 C to +20 C (-166 F to +68 F) dewpoint
Model DS1200-AMT-Ex Intrinsically Safe 4-20mA ATEX Certified Dewpoint Hygrometer Ranges available between -120 C to +20 C (-166 F to +68 F) dewpoint Model DS1200-AMT-Ex Dewpoint Hygrometer has been designed
More informationDM70 Hand-Held Dewpoint Meter for Spot-Checking Applications
DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications The Vaisala DRYCAP Hand-Held Dewpoint Meter DM70 offers accurate and fast measurement for industrial dewpoint applications, such as compressed
More informationED 701 General Industry Pressure Transmitter
ED 701 General Industry Pressure Transmitter Standard industrial process connections Complete range of electrical connections 4... 20 ma and Voltage outputs Accuracy: 0.1%, 0.2% and 0.4% FS Quick response
More informationDM70 Hand-Held Dewpoint Meter for Spot-Checking Applications
DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications The Vaisala DRYCAP Hand-Held Dewpoint Meter DM70 offers accurate and fast measurement for industrial dewpoint applications, such as compressed
More informationVaisala DRYCAP Hand-Held Dewpoint Meter DM70. Fast Response - Low Maintenance
Vaisala DRYCAP Hand-Held Dewpoint Meter DM70 Fast Response - Low Maintenance DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications The Vaisala DRYCAP Hand- Held Dewpoint Meter DM70 offers accurate
More informationCALYS 80P. Field pressure calibrator with high accuracy 0.01%
Field pressure calibrator with high accuracy 0.01% CALYS 80P is a field pressure calibrator within CALYS range with high accuracy of 0.01%. Description CALYS 80P is a field pressure calibrator within CALYS
More informationDM70. Hand-held Dewpoint Meter. Truly Fast Response - Superior Long-term Stability
DM70 Hand-held Dewpoint Meter Truly Fast Response - Superior Long-term Stability DM70 - When Dry Air Is Important The Vaisala DM70 Hand-held Dewpoint Meter offers stable and fast measurement for industrial
More informationCALYS 120 IS. ATEX documenting multifunction calibrator with high accuracy 0.006%
ATEX documenting multifunction calibrator with high accuracy 0.006% is the top instrument of our range of ATEX documenting multifunction calibrators. Description is the top instrument of our range of ATEX
More informationField IT 50T Series Transmitters
Data Sheet SS/G/A_ Field IT 0T Series Transmitters gauge/absolute pressure transmitter Ranges: -00 to 0000kPa - to 00bar -. to 6000psi Base accuracy : 0.% (BFSL) Chemical Vapour Deposition (CVD) technology
More informationTurbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas
Turbine Meter TRZ 0 PRODUCT INFORMATION Reliable Measurement of Gas TURBINE METER TRZ 0 Method of operation, Construction Method of operation The TRZ 0 turbine meter is a flow meter suitable for gas measurement
More information473-SHX Dew Point Mirror
Humidity and Temperature Reference Hygrometer For Temperature up to 125 C Precise and stable chilled mirror dew point mirror technology High temperature optical components High temperature sample fan Cable
More informationISEmax CAM40/CAS40. Technical Information
Technical Information Online measurement of nutrient parameters Ion-selective electrode system for the continuous measurement of ammonium and nitrate Application The ion-selective electrode system works
More informationDMP 331P. Industrial Pressure Transmitter. Process Connections With Flush Welded Stainless Steel Diaphragm
DMP P Industrial Pressure Transmitter Process Connections With Flush Welded Stainless Steel accuracy according to IEC 60770: standard: 0.5 % FSO option: 0.5 % FSO Nominal pressure from 0... 00 mbar up
More informationTHM80X Series Feature Application. Temperature & Humidity / Flow Measuring Specialist
THM80X Series Feature Application IP67 protection degree, rugged aluminum case, fit in variety harsh environment Capable of temperature compensation Linear adjustment temperature & humidity by computer,
More informationPressure Measurement. Transmitters for basic requirements SITRANS P MPS (submersible sensor) Transmitter for hydrostatic level 2/27
Overview Siemens AG 011 Function s are for measuring the liquid levels in wells, tanks, channels and dams. Diaphragm p Sensor U const. U + (brown) I EM I -1 (blue) 0, U B s are submersible sensors for
More informationOverview Brochure Comprehensive range of instrumentation for the measurement of moisture and oxygen
Overview Brochure 2016 Comprehensive range of instrumentation for the measurement of moisture and oxygen Overview Brochure Relative Humidity Probes and Transmitters Repeatable RH and temperature measurement
More informationDMP 331. Industrial Pressure Transmitter for Low Pressure. Stainless Steel Sensor
DMP Industrial Pressure Transmitter for Low Pressure Stainless Steel Sensor accuracy according to IEC 60770: : 0.5 % FSO option: 0.5 / 0. % FSO Nominal pressure from 0... 00 mbar up to 0... 60 bar Output
More informationPressure Transmitter ED 701
Pressure Transmitter ED 701 Hygienic and industrial process connections Zero setting function Accuracy < 0.1% FS Quick response time < 25 ms High long-term stability Excellent repeatability Active compensation
More informationPressure transmitters for industrial applications MBS 32 and MBS 33
Data sheet Pressure transmitters for industrial applications MBS 32 and MBS 33 The standard pressure transmitters MBS 32 and MBS 33 are designed for use in almost all industrial applications, and offer
More informationISOCON-6. 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER
ISOCON-6 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER Whilst every effort has been taken to ensure the accuracy of this document, we accept no responsibility for damage, injury, loss or expense resulting
More informationFLT5A FLT5A. abc. abc. Thermometer User Guide. Thermometer User Guide. Internet:
abc instruments international FLT5A Thermometer Land Instruments International Dronfield, S18 1DJ England Telephone: (1246) 417691 Facsimile: (1246) 41585 Email: infrared.sales@landinst.com Internet: www.landinst.com
More informationDMP 331Pi. Precision Pressure Transmitter. pressure ports and process connections with flush welded stainless steel diaphragm
DMP Pi Precision Pressure Transmitter pressure ports and process connections with flush welded stainless steel diaphragm accuracy according to IEC 600: 0, % FSO Nominal pressure from 0... 00 mbar up to
More informationOperating Instructions for Digital Manometer. Model: MAN-SD
Operating Instructions for Digital Manometer Model: MAN-SD 1. Contents 1. Contents...2 2. Note...3 3. Instrument Inspection...3 4. Regulation Use...4 5. Operating Principle...4 6. Mechanical Connection...5
More informationLatest generation testo air velocity & IAQ probes with cable handle. Instruction manual
Latest generation testo air velocity & IAQ probes with cable handle Instruction manual Contents Contents 1 About this document... 5 2 Safety and disposal... 5 3 System description... 6 4 Description of
More informationDS 300. Electronic Pressure Switch. with IO-Link interface. Stainless Steel Sensor. accuracy according to IEC 60770: 0.35 % FSO.
DS 00 Electronic Pressure Switch with IO-Link interface Stainless Steel Sensor accuracy according to IEC 60770: 0.5 % FSO Nominal pressure from 0... 00 mbar up to 0... 600 bar Digital output signal IO-Link
More information- compressed air plants (refrigerating/adsorption dryers) - granulate dryers - medical gases - non-corrosive gases, e. g. nitrogen
INSTRUCTION MANUAL FA 510 The CS dew point sensor FA 510 with 3-wire technology 4 20 ma and RS 485 output enables a reliable and long-term stable monitoring of the dew point in industrial applications
More information473 Dew Point Hygrometer
gh Performance Chilled Hygrometer With Cable Mounted Measuring Heads ghly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct insertion
More informationTF 212 / TF 212-Ex. Field mounted temperature transmitters, PROFIBUS PA, Pt 100 (RTD), thermocouples, 1 or 2 independent channels 10/11-8.
TF / TF -Ex Field mounted temperature transmitters, PROFIBUS PA, Pt 00 (RTD), thermocouples, or independent channels 0/-8.70 EN P R O F I PROCESS FIELD BUS B U S Input Resistance thermometer (,, -wire
More informationPressure Display Controller
MODEL DW Display Controller Low The Model DW pressure controller accurately monitors and controls very low pressures. It combines in one enclosure a low pressure transducer, a digital display and two alarms
More informationDMP 331Pi. Precision Pressure Transmitter. pressure ports and process connections with flush welded stainless steel diaphragm
DMP Pi Precision Pressure Transmitter pressure ports and process connections with flush welded stainless steel diaphragm accuracy according to IEC 600: 0, % FSO Nominal pressure from 0... 00 mbar up to
More informationPascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure
DATASHEET Pascal ET Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure - electrical signals (ma, mv, V, ) - temperature (TC and
More information473 Dew Point Hygrometer
High Performance Chilled Mirror Hygrometer With Cable Mounted Measuring Heads Highly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct
More informationTransmitters for basic requirements SITRANS P MPS (submersible sensor) Transmitter for hydrostatic level
Overview Function s are for measuring the liquid levels in wells, tanks, channels and dams. Sensor Diaphragm p U const. U +2 (brown) I EM I - (blue) 0, U B s are submersible sensors for hydrostatic level
More informationTF 202 / TF 202-Ex. Field mounted temperature transmitters, FOUNDATION Fieldbus (H1), Pt 100 (RTD), thermocouples, 1 or 2 independent channels
TF 0 / TF 0-Ex Field mounted temperature transmitters, FOUNDATION Fieldbus (H), Pt 00 (RTD), thermocouples, or independent channels 0/-8.69 EN USA Input Resistance thermometer (,, wire circuit) Thermocouples
More informationHumidity Calibration Solutions
Humidity Calibration Solutions www.michell.com Humidity Calibration Solutions The Importance of Regular Calibration The reliable operation of a hygrometer and indeed any measuring instrument, can only
More informationChilled Mirror Hygrometry
Chilled Mirror Hygrometry Precision instruments for critical measurement and control applications Chilled Mirror Hygrometry Wherever precise, fundamental measurements are required in your laboratory or
More informationData Sheet / Instructions. Dewpoint MeterO
Data Sheet / Instructions Dewpoint MeterO Dewpoint Meter ISO 8502-4: Preparation of steel substrates before application of paints and related products. Tests for the assessment of surface cleanliness.
More informationInstallation guide 862 MIT / MIR
Installation guide 862 MIT / MIR in combination with: 864 MTT or 863 MRT or (dual) spot element November 2005 Part no. 4416.232_Rev3 Enraf BV PO Box 812 2600 AV Delft Netherlands Tel. : +31 15 2701 100
More informationMinimum (-31) every 10 s -35 (-31)
Humidity, and Dew Point Data Logger with LCD ORDERING INFORMATION Standard Data Logger (Data Logger, Software on CD, Battery) Replacement Battery EL-USB-2-LCD BAT 3V6 1/2AA FEATURES 0 to 100 measurement
More informationThe Ordering Code for various standard model Recorders with an AC supply and without any additional options are as follows:
The basic PC software is supplied free with the recorder. There is an additional charge for the extensive Data Acquisition Software supplied with communication of RS-232/422/485 or Ethernet. The Ordering
More informationPressure Measurement Single-range transmitters for general applications
Siemens AG 07 Overview Function SITRANS LH00 Transmitter for hydrostatic level U const. p U I EM Sensor Connection for auxiliary power supply Vent pipe with humidity filter Protective conductor connection/
More informationISOCON-3 MAINS POWERED ISOLATING SIGNAL CONVERTER
ISOCON-3 MAINS POWERED ISOLATING SIGNAL CONVERTER CAUTION: This equipment is designed for connection to mains voltages and must be used in accordance with this guide. If it is not, the safety protection
More informationOverload Relays. SIRIUS 3RU1 Thermal Overload Relays. 3RU11 for standard applications. 5/46 Siemens LV 1 AO 2011
SIRIUS 3RU1 Thermal Overview 1 2 7 3 4 6 "Increased safety" type of EEx e according to ATEX directive 94/9/EC The 3RU11 thermal overload relays are suitable for the overload of explosion-proof motors with
More informationDS 200P. Electronic Pressure Switch. Pressure Ports And Process Connections With Flush Welded Stainless Steel Diaphragm
Pressure Ports And Process Connections With Flush Welded Stainless Steel Diaphragm accuracy according to IEC 600: standard: 0. % FSO option: 0. % FSO Nominal pressure from 0... 00 mbar up to 0... 0 bar
More informationDewpoint Meter Data Sheet / Instructions
Paint Test Equipment Dewpoint Meter Data Sheet / Instructions Dewpoint Meter Information ISO 8502-4: Preparation of steel substrates before application of paints and related products. Tests for the assessment
More informationHumidity Calibration Solutions
Instruments Humidity Solutions www.michell.com Instruments Humidity Solutions The Importance of Regular The reliable operation of a hygrometer and indeed any measuring instrument, can only be verified
More informationHumidity, and Dew Point Data Logger with LCD ORDERING INFORMATION Standard Data Logger (Data Logger, Software on CD, Battery) Replacement Battery EL-USB-2-LCD BAT 3V6 1/2AA FEATURES 0 to 100 measurement
More informationSECTION 1 - WHAT IS A BTU METER? BTU's = Flow x ΔT Any ISTEC BTU Meter System consists of the following main components:
SECTION 1 - WHAT IS A BTU METER? ISTEC BTU Meters measure energy usage by multiplying flow rate and temperature difference. As the water (or other liquid) passes through these lines, the multi-wing turbine
More informationEdgetech Instruments Inc.
2015_07_ETI OVERVIEW Rev 3 MADE IN THE USA Edgetech Instruments (ETI) is a USA company based in Massachusetts. We have over 50 years of expertise in the design and manufacture of a comprehensive range
More informationComplete Moisture Analysis
Complete Moisture Analysis Water Activity Instrumentation Dew Point Water Activity & Moisture Analysis Moisture Sorption Isotherms Capacitance Water Activity Accessories PRODUCT CATALOG 2013-2014 DEW POINT
More informationSTATIC GAS MONITOR Type SGM/DEW Revision A of 20 aprile 2016
APPLICATIONS Moisture monitoring of air or gas (SF6) Suitable for indoor or outdoor Industrial, medical or aerospace fields HIGHLIGHTS High voltage circuit breakers commonly used for distribution and transmission
More informationIDENTIFICATION STS400
IDENTIFICATION APPLICATIONS Tracking and Traceability is a primary need of tire manufactures. In logistics, WIP, quality control, and compliance, tires get tracked at every step of production process.
More informationJ R. TransTech Systems T S. The World s Number 1 Supplier Of Non Nuclear Density Meters for Asphalt Pavements and Engineered Soils
TransTech Systems The World s Number 1 Supplier Of Non Nuclear Density Meters for Asphalt Pavements and Engineered Soils Distributed in the UK and Ireland by JR Technical Services UK Ltd J R T S 1-800-724-6306
More informationISEmax CAS40D. Technical Information
Technical Information ISEmax CAS40D Online measurement of nutrient parameters Ion-selective electrodes for the continuous measurement of ammonium, nitrate and other ions Application The ion-selective sensor
More informationTechnical Data Sheet DPT & DPIT-4005 SERIES DIFFERENTIAL PRESSURE TRANSDUCERS. Features. Description
Technical Data Sheet DPT & DPIT-4005 SERIES DIFFERENTIAL PRESSURE TRANSDUCERS Description The 4005 series of differential pressure transducers include Models DPT-4005 and DPIT-4005-M. The 4005 series differential
More informationE THERMADATA LOGGER Humidity & temperature recording meter
E 07 00 THERMADATA LOGGER Humidity & temperature recording meter THERMADATA LOGGER Data-loggers are electronic devices that record temperature over a period of time for analysis at a later date. The user
More informationSINEAX VK 636 Programmable Temperature Transmitter with PROFIBUS-PA protocol
with PROFIBUS-PA procol for installation in the terminal head of a temperature sensor DIN 79, shape B Application 00 II () G FISCO The SINEAX VK 66 is a temperature transducer for use in aumation systems
More informationOPERATION MANUAL. 3 in 1, IR, Type K/J/R/E/T, Pt 100 ohm THERMOMETER. Model : TM-2000
3 in 1, IR, Type K/J/R/E/T, Pt 100 ohm THERMOMETER Model : TM-2000 Your purchase of this THERMOMETER marks a step forward for you into the field of precision measurement. Although this THERMOMETER is a
More informationAcceleration/Velocity/Displacement VIBRATION METER
Acceleration/Velocity/Displacement VIBRATION METER Model : VB-8220 Your purchase of this VIBRATION METER marks a step forward for you into the field of precision measurement. Although this METER is a complex
More informationDHS1900 DHS3000 High Temperature Hand Held Infrareds User Manual
DHS1900 DHS3000 High Temperature Hand Held Infrareds User Manual WD1063 Rev B Revised 02/23/15 Wahl Instruments Inc. 234 Old Weaverville Road Asheville, NC 28804 Toll Free: 800-421-2853 Phone: 828-658-3131
More informationS95.3U01 / S95.3U11. Head mounted temperature transmitters, programmable, Pt 100 (RTD), thermocouples, electrical isolation
S95.U0 / S95.U Head mounted temperature transmitters, programmable, Pt 00 (RTD), thermocouples, electrical isolation Resistance thermometer (,, wire circuit) Thermocouples Resistance remote signalling
More informationABB MEASUREMENT & ANALYTICS DATA SHEET. PGC5000 Temperature programmed gas chromatograph
ABB MEASUREMENT & ANALYTICS DATA SHEET PGC5000 Temperature programmed gas chromatograph 2 P G C 5 0 0 0 T E M P E R AT U R E P R O G R A M M E D D S/ P G C 5 0 0 0 T P - E N Measurement & Analytics The
More informationHeat stress index. Dewpoint. pressure. altitude
Kestrel 4200 Measures Volume air flow in cfm, L/s, m³/s, m³/m, m³/h Air velocity (current, average, and maximum) Heat stress index Windchill Relative humidity Dewpoint in F or C Humidity ratio Barometric
More informationSERIES 4000 PRECISION DEWPOINT HYGROMETER Integrale INSTALLATION, OPERATION AND MAINTENANCE MANUAL
SERIES 4000 PRECISION DEWPOINT HYGROMETER Integrale INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue July 2001 Page 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 4 1.1 General 4 1.2 Operating Principle
More informationDesigned to go where you work
Designed to go where you work Finally, a digital substitute for your mercury-in-glass thermometers! Accurate and repeatable to ± 0.05 C over its full range, the 55A/55A Stik Thermometer is the new gold
More informationSINEAX VK 626 Programmable Temperature Transmitter for RTD and TC inputs, with HART protocol
SINEAX VK 66 for RTD and TC inputs, with HART procol for installation in the terminal head of a temperature sensor DIN 79, shape B Application II () G SINEAX VK 66 is a two-wire head-mounted transmitter.
More informationHYGRASGARD KFF KFTF HYGRASGARD AFF AFTF HYGRASGARD AFF - LC AFTF - LC
HYGRASGARD AFF - LC AFTF - LC sensors, calibrateable, with and active passive output Quality product for HVAC sector, accuracy % r.h. The calibrateable room humidity and temperature sensor HYGRASGARD KFF
More informationMODEL: R7K4FML3. Remote I/O R7K4F Series. POWER INPUT DC power R: 24 V DC (Operational voltage range: ±10 %; ripple 10 %p-p max.)
Remote I/O R7KF Series I/O MODULE (-III) POWER INPUT DC power R: V DC (Operational voltage range: ±0 %; ripple 0 %p-p max.) 0 (8.7). (.) 7(.8) mm (inch) [] OPTIONS (multiple selections) Output data read
More informationThe Fundamentals of Moisture Calibration
The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity
More informationPHOENIX CONTACT - 04/2016. Features
Signal conditioner Data sheet 100238_de_06 1 Description PHOENIX CONTACT - 04/2016 Features The MCR-C-UI-UI(-450)-DCI(-NC) 3-way isolation amplifier is used to electrically isolate and convert analog signals.
More informationph/redox Measurement CPM 280
Technical Information TI 346C/07/en/07.03 51511752 ph/redox Measurement CPM 280 Battery-driven pocket meter for ph and redox measurement Application CPM 280 is a compact pocket meter for measuring ph value
More informationADI 2040 Process Analyzer
ADI 2040 Process Analyzer Multi-purpose wet chemical analysis Multi-purpose Wet Chemical Analysis ADI 2040 The Metrohm Applikon ADI 2040 Process Analyzer is a multipurpose wet chemical analyzer, that has
More informationSemi-automatic titrators PH-Burette. A great ph-meter with integrated burette. Incredible price! Infinite possibilities
Semi-automatic titrators PH-Burette. A great ph-meter with integrated burette. Incredible price! Infinite possibilities instruments 5 YEARS G U A R A N T E E Quality at a reasonable price The optimum combination
More informationEVEN THE FINEST CANNOT ESCAPE. Wöhler SM 500 Suspended Particulate Analyser
EVEN THE FINEST CANNOT ESCAPE. Wöhler SM 500 Suspended Particulate Analyser The latest innovation in total Suspended Particulate Matter (SPM) measurement The Measure of Technology Made in Germany TÜV CERTIFIED
More informationThermal Mass Flow Meter for Gases
Thermal Mass Flow Meter for Gases measuring monitoring analysing Direct mass flow rate measurement of gases Measuring accuracy: ±1.0% f.s. +0.5% m.v. pmax PN 16, tmax 175 C Fast response time No moving
More informationImpedance relay and protection assemblies
RXZK 21H, 22H, 23H 509 006-BEN Page 1 Issued June 1999 Changed since July 1998 Data subject to change without notice (SE970175) (SE970184) Features Micro-processor based impedance relay with R and X settings
More informationMetso Concentration Measurement 3300
Metso Concentration Measurement 3300 Metso Concentration 3300 Accurate direct in-line measurement The Concentration Measurement is provided with intelligent technology calculating the concentration of
More informationPAA-ANALYZERS PRINCIPLE OF OPERATION
PAA-ANALYZERS PRINCIPLE OF OPERATION The PAA process analyzer comprises a control unit and an analyzing section which includes the reaction chamber and measuring cell, reagent pumps and associated valves
More informationIon-Chamber Survey Meter OD-02
Ion-Chamber Survey Meter OD-02 Dose and dose rate meter for measuring the ambient dose equivalent H*(10) and dose rate equivalent dh*(10)/dt as well as the directional dose equivalent H (0,07) and dose
More informationAnalog - Digital multimeters
CAT-IV -3, MI, Satpur, Nashik-4 7,India. Tel.: +9 53 6, ax : +9 53 3564 Page of / 5 / I I / 7 / B Automatic Terminal Blocking System (ABS) The automatic Terminal blocking system prevents incorrect connection
More informationS95.3U02 / S95.3U12. Head mounted temperature transmitters, HART programmable, Pt 100 (RTD), thermocouples, electrical isolation
S95.U0 / S95.U Head mounted temperature transmitters, HART programmable, Pt 00 (RTD), thermocouples, electrical isolation Resistance thermometer (,, wire circuit) Thermocouples Resistance remote signalling
More informationVIBRATION METER. Instruction Manual.
VIBRATION METER 840063 Instruction Manual CONTENTS Title Page I. Introduction... 2 II. Panel Description... 3 III. Operating Instructions A. General Measurement Procedures... 3~4 B. Peak... 5 C. Data Hold...
More informationTHERMYS 150. Advanced field reference thermometer / temperature calibrator
Advanced field reference thermometer / temperature calibrator THERMYS 150 is an all-in-one high accurate documenting field reference thermometer for thermocouples (14 types), resistive probes (12 types)
More information1551A Ex/1552A Ex Intrinsically Safe Stik Thermometer. The new gold standard of industrial temperature calibration
1551A Ex/155A Ex Intrinsically Safe Stik Thermometer The new gold standard of industrial temperature calibration Designed to go where you work Finally, a digital substitute for your mercury-in-glass thermometers!
More informationWATER ACTIVITY MEASUREMENT WATER ACTIVITY
MEASUREMENT WATER ACTIVITY FAST AND PRECISE WATER ACTIVITY MEASUREMENTS. INTRODUCING AIRCHIP DIGITAL TECHNOLOGY. Determine product quality with ROTRONIC Aw products ROTRONIC Aw devices for laboratory and
More informationAnalogue Meters CQ2x07 series
Analogue Meters CQ2x07 series Analogue Power Meters Active, reactive, apparent power meters and Power factor meters with exchangeable scales EQ0207/2207/0107/2107 & YQ0207/2207/0107/2107 o Measurement
More informationFVBSS fixed volume bottle
N FVBSS fixed volume bottle Sampling System The SENSOR Fixed Volume Bottle Sampling System (FVBSS) provides a repeatable fixed volume of sample during each sampling process without stopping flow of the
More informationAir velocity & IAQ measuring instruments
Data sheet dp Air velocity & IAQ measuring instruments - air velocity & IAQ measuring instrument - dp air velocity & IAQ measuring instrument incl. differential pressure C CO Intuitive: clearly structured
More informationAir velocity & IAQ measuring instruments
Data sheet dp Air velocity & IAQ measuring instruments - air velocity & IAQ measuring instrument - dp air velocity & IAQ measuring instrument incl. differential pressure C CO Intuitive: clearly structured
More informationPRODUCT CATALOGUE PREssURE switch
PRODUCT CATALOGUE PREssURE switch w w w. p a a b. c o m Pressure AT THe HIGHesT LeVeL Successful mediumsized companies are not successful because they are active in many areas, but rather because they
More informationBOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER. Technical Description Version 2 BGRIMM AUTOMATION
BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER Technical Description Version 2 BGRIMM AUTOMATION, South 4th Ring Road West, Beijing, China TEL: +86(10)59069762 Fax: +86(10)68360101 Contents BGRIMM AUTOMATION
More informationv = m / s Thermal Flow Sensors TA with thin-film sensor element Probes Probe Guides Measuring Tubes
Thermal Flow Sensors TA with thin-film sensor element Probes Probe Guides Measuring Tubes Examples of Application Flow measurements also in hazardous locations X zone 0 or zone 1 in clean rooms in air
More informationAIR PREPARATION. ELECTRONICALLY CONTROLLED PRESSURE REGULATING VALVE (PROPORTIONAL PRESSURE REGULATING VALVE) SERIES control
AIR PREPARATION ELECTRONICALLY CONTROLLED PRESSRE REGLATING ALE (PROPORTIONAL PRESSRE REGLATING ALE) SERIES control Description Abb. Port size Recommended Type Data Sheet No. flow (l/min)* airfit tecno
More informationAcceleration, Velocity, Separate probe VIBRATION METER Model : VB-8202
Acceleration, Velocity, Separate probe VIBRATION METER Model : VB-8202 Your purchase of this VIBRATION METER marks a step forward for you into the field of precision measurement. Although this METER is
More informationLANCOM200 PORTABLE ACID DEWPOINT TEMPERATURE MONITORING. Acid Dewpoint SO3 H2 SO4 QUALITY CUSTOMER SOLUTIONS
PORTABLE ACID DEWPOINT TEMPERATURE MONITORING i Acid Dewpoint SO3 H2 SO4 QUALITY CUSTOMER SOLUTIONS AMETEK LAND HAS BEEN MANUFACTURING PRECISION MEASURING EQUIPMENT SINCE 1947. WE ARE SPECIALISTS IN NON-CONTACT
More informationAutomatic titrators, TitroMatic
112 TitroMatic These are the fourth generation of CRISON titrators. CRISON has over 25 years experience in this field and knows that, for certain applications, the performance and cost of large titrator
More informationField mounted Temperature Transmitter TF212/TF212-Ex
Data Sheet Field mounted Temperature Transmitter TF/TF-Ex PROFIBUS PA, Pt 00 (RTD), Thermocouples, or independent channels Input Resistance thermometer (,, -wire circuit) Thermocouples Resistance remote
More informationPro-V Multivariable Flowmeter Model M23 Insertion Vortex
Pro-V Multivariable Flowmeter Model M23 Insertion Vortex Pro-V TM Advantage: Volumetric or mass flow monitoring of most liquids, gases, and steam Multivariable meter delivers mass flow, temperature, pressure,
More informationHeat Meter WR 200-D. 2 temperature inputs for Pt100 in four-wire circuit. Exact calculation of the heat parameters according IAPWS-IF97
Heat Meter WR 200-D electronic computing device for heat quantity measuring Heat transfer medium water steam Any flow sensor with analog signal connectable 2 temperature inputs for Pt100 in four-wire circuit
More informationGC005 thru GC600. Wireless Load Cell. Features. Applications. General Description. Load Cell
Wireless Load Cell Features Standard single part line pull sizes ranging from 5,000 to 600,000 lb. All stainless frame and hardware Heavy duty construction from 17-4PH1150 stainless steel (17-4PH1025 on
More information