PET. Technical aspects
|
|
- Bethany Walters
- 5 years ago
- Views:
Transcription
1 PET Technical aspects
2 15 N 15 O Detector 1 β+ Detector 2 e-
3 Evolution of PET Detectors CTI/Siemens
4 15 N 15 O Detector block 1 β+ Detector block 2 x e- x y y location line of response Constant fraction discriminator Constant fraction discriminator
5 PET Detectors Most modern PET system use a different detector technology where a large number of scintillation crystals are coupled to a smaller number of PMTs. For a given crystal, the lower the cross section is the shorter the length of the crystal has to be due to the light transport.
6 Scintillator Materials NaI (Tl) BGO GSO LSO LYSO LaBr 3 Density [g/ml] /µ [cm] ~2 Index of Refraction Hygroscopic Yes No No No No Yes Rugged No Yes No Yes Yes Yes Peak Emission [nm] Decay Constant [ns] Light Output >100 Energy Resolution <
7 PET Detectors In the block detector, a matrix of cuts are made into a solid block of scintillator material to define the detector elements. The depth of the cuts are adjusted to direct the light to the PMTs. The light produced in each crystal, will produce a unique combination of signals in the PMTs, which will allow the detector to be identified. The more light output the more crystals can be identified.
8 PET Detectors To identify the detector elements, the two following ratios are calculated (similar to the logic used in scintillation cameras). A+ B C D X = A + B + C + D A B+ C D Y = A + B + C + D Below is the flood response (i.e., X and Y density distribution when exposed to a flood source of 511 kev photons) for a block detector from the ECAT EXACT HR. Element (4,4) Element (5,3)
9 The Technology : HiRez Standard Detector 6.4 mm x 6.4 mm 64 crystals/block 144 blocks/scanner 9216 crystals/scanner 3.4 mm slice width 47 slices HI-REZ Detector 4.0 mm x 4.0 mm 169 crystals/block 144 blocks/scanner crystals/scanner 2 mm slice width 81 slices
10 Detector block 1 Detector block 2 x x y y location 12 ns AND line of response coincidence? Constant fraction discriminator Constant fraction discriminator
11 Scintillator Materials NaI (Tl) BGO GSO LSO LYSO LaBr 3 Density [g/ml] /µ [cm] ~2 Index of Refraction Hygroscopic Yes No No No No Yes Rugged No Yes No Yes Yes Yes Peak Emission [nm] Decay Constant [ns] Light Output >100 Energy Resolution <
12 Detector block 1 Detector block 2 x x y y location 12 ns AND MCA line of response coincidence? Energy spectrum Constant fraction discriminator Constant fraction discriminator
13 Scintillator Materials NaI (Tl) BGO GSO LSO LYSO LaBr 3 Density [g/ml] /µ [cm] ~2 Index of Refraction Hygroscopic Yes No No No No Yes Rugged No Yes No Yes Yes Yes Peak Emission [nm] Decay Constant [ns] Light Output >100 Energy Resolution <
14 Event Types True Event Scattered Event Random Event Multiple Event
15 Random Coincidences Because of the finite width of the logic pulses that are fed into the coincidence circuit, there is a probability for random or accidental coincidences between unrelated events. True Coinc. Single Event Random Coinc. Detector 1 Detector 2 Time τ True Coinc. Single Event
16 Random Coincidences If N 1 and N 2 are the individual average count rates of detector 1 and 2, respectively, then it can be shown that the random coincidence rate for the pair of detectors is: N R = 2τ N 1 N 2 where 2τ is the coincidence window (or τ is the width of the singles pulses)
17 Random Coincidences Randoms can be estimated by adding a second coincidence circuit. The logic pulse from one detector is delayed beyond the time resolution of the detector pairs. This does not require any knowledge of the coincidence time window, but is not as statistically accurate as the singles method Amp./ PHA Amp./ PHA Coinc. Delay Prompts Coinc. Randoms
18 Estimation of Random Coincidences Singles Method Monitor singles for each detector Statistically accurate Requires accurate knowledge of τ Dead-time different between prompt coinc. & singles Delayed Coincidence Method Separate coincidence circuitry/processor Poor statistical accuracy (improvement with variance reduction)
19 Scatter Scatter can be reduced by better energy resolution more collimation (2D)
20 Signal-to-Noise True Coincidences ~ Activity Good events! / ~ T S N T
21 Signal-to-Noise Random Coincidences ~ Activity 2 Can be accurately corrected for Correction increases image noise Detector material dependent S/ N ~ T T + 2R
22 Signal-to-Noise Scattered Coincidences ~ Activity Reduces Image Contrast Requires correction Analytical estimation Correction increases image noise S/ N ~ T T + S+ 2R
23 Signal-to-Noise Multiple Coincidences: ~ Activity 3 Never saved Source of Dead time
24 Noise Equivalent Count Rate - NEC The correction for random and scattered events increases the noise in the net true events The correction therefore reduce the effective true count rate The NEC describes the equivalent true count rate in the absence of random and scattered events: NEC = T T + S + 2kR ~ S / N
25 Count Rate [cps] Trues Prompts Randoms NEC Specific Activity [MBq/ml]
26 Spatial Resolution The spatial resolution in PET is primarily determined by: Detector size Physics of positron decay System geometry Detector material
27 Spatial Resolution For a source placed at the midpoint between two scintillation detectors with a width w d, the geometric line spread function has a triangular shape with a FWHM of w d /2. w d FWHM = w d 2
28 Spatial Resolution -Tangential For sources located between the midpoint and the detector surface the LSF will have a trapezoidal shape with width varying from w d /2 (at the center) and w d at the detector surface.
29 Spatial Resolution - Radial Transaxial Resolution or ECAT EXACT HR Axial section Radial FWHM (mm) 8 Tangential FWHM tang FWHM rad R= R (cm)
30 Depth-of-Interaction cm etector eparate
31 Depth of Interaction Decoding Techniques Multiple layers of scintillators with different decay constants. Wong, 1985; Carrier et. al., 1987; Casey et. al Additional light readouts on the scintillator. McIntyre et. al., 1980; Moses et. al., 1991 Modulation of light output Rogers et. al., 1995
32 30 mm 2x15 mm 3x10 mm 4x7.5 mm 30 cm 40 cm UCLA School of Medicine
33 DOI Detectors Siemens Molecular Imaging
34 SSPM/SiPM UCLA School of Medicine
35 SSPM/SiPM Counts ns FWHM Channel UCLA School of Medicine
36 Avalanche Photo Diodes (APD) vs. Photo Multiplier Tubes (PMT) PMTs Size 10-52mm dia. APDs 5x5mm Gain Up to 10 6 Up to 200 Rise time ~1 ns ~5 ns QE 20% 70% Magnetically sensitive insensitive
37 Spatial Resolution Although the most energetic positrons can travel several mm before annihilating, only a few of these are emitted. The average positron energy emitted is approximately 1/3-1/2 of the maximum energy. The total path length the positrons travel is not along a straight path. Through inelastic interactions with electrons in the positrons path is deflected. The distance from the mother nucleus is therefore much shorter.
38 Positron Range From Levin & Hoffman PMB 44, 1999
39 Resolution loss due to the positron range Isotope β + Energy FWHM FWTM [MeV] [mm] [mm] 18 F C N O m Rb
40 Positron Range ~65% ~50% 18 F 635 kev 124 I 1.53 & 2.14 MeV
41 Non-colinearity ~0.3 FWHM 100 cm Ø ~ 2.5 mm FWHM 15 cm Ø ~ 0.3 mm FWHM
42 Spatial Resolution The measured resolution (intrinsic resolution) of the system is a convolution of the various resolution components. If the different resolution components are assumed to be Gaussian in shape and are described by a FWHM then the combined resolution is the squared sum of the individual resolution components: FWHM = FWHM + FWHM + FWHM total det ector positron angulation FWHM = FWHM + FW M + FWHM total det ector 20% positron angulation Sánchez Crespo et. al. EJNM, 2004
43 Improved NEC Performance
44 Time-of-flight PET R R Det 2 Det 1 x R - x R + x s= vt R + x = vt R x = vt x= vt ( t) x= 2 1 c t 2
45 Time-of-flight PET For ideal detectors, TOF would eliminate the need for image reconstruction, since the measurement would allow each event to be accurately positioned in space. All detectors have a finite time resolution, or uncertainty in timing. This translates to an uncertainty in positioning. BGO ~ 5 ns NaI ~ 1.5 ns CsF, LaBr 3 ~ 0.45 ns BaF 2, LSO, LYSO ~ 0.3 ns 75 cm 22.5 cm 6.7 cm 4.5 cm
46 Time-of-flight PET Figure 1. Image elements contributing to a LOR, for conventional PET (left) and TOF PET (right).
47 Time-of-flight PET Even with a finite time resolution, using the TOF information an improvement in signal-to-noise ratio (S/N) can be achieved: 2 D D SNR SNR = SNR x c t TOF conv. conv.
48 Can we see TOF improvement? non TOF TOF 5 min 3 min 1 min 6-to-1 contrast; 35-cm phantom Noticeable improvement with TOF with large size phantom J. Karp, U of Penn
49 QUALITY CONTROL (QC)
50 Quality Control Ensure operational integrity of the system Maintain consistent and high image quality Minimize chance for artifacts Catch potential problems early Maintain quantitative accuracy
51 PET System, storage of data Each detector is in coincidence with a sector of opposite detectors. The angle of the section defines the Field of View (FOV)
52 PET System, storage of data All coincidence lines that are parallel at a given angle form a projection which is the accumulated radioactivity in that angle (1D) of the object (2D). In the computer we store the projection in a matrix x-axis radioactivity profile, y-axis the different angles
53 Projection & Sinogram Projection: All ray-sums in a direction P(θ,t) y t Sinogram: All projections θ π θ x f(x,y) X-rays Sinogram t
54 PET System θ r In most PET scanners, a large number of scintillation detectors are arranged in a circle. Each detector is in coincidence with a number of opposite detectors. The field-of-view (FOV) of the scanner is defined by the width or the angle of the fan. All coincidence lines that are parallel at a given angle form a projection in the sinogram.
55 PET System θ r All coincidence lines (or lines of response) for a given detector form a diagonal trace in the sinogram.
56
57 Daily QC / Blank Scan
58 The sinogram as a help to identify errors Normal QC Detector Failure Detector Controller Failure Coincidence Processor Failure
59 Buchert et. al. JNM 40, 1999
60 Daily QC / Cylinder Scan Before Normalization After Normalization
61 Daily QC Procedure Daily Detector Check Sources & Phantoms needed Transmission/Rotating Rod Sources Uniform 68 Ge cylinder phantom
62 Daily / Weekly Quantitation Scan Scan uniform 20 cm Ø 68 Ge or 18 F Cylinder Reconstruct: All corrections applied Standard reconstruction parameters Visual inspection Compare image ROI activity to calibrated activity Always perform after: Service Re-tuning Re-normalization
63
64 Quantitation 2D Cylinder ( µci/ml) FBP OSEM Calculated Atten. Corr % % Meas. Atten. Corr. CT % % Meas. Atten. Corr. Rods % %
65 Quarterly QC Procedures Detector setup (if needed) PMT tuning Detector setup Coincidence timing Normalization Gantry alignment (for PET/CT) Always after Service Software upgrades Other calibrations (well counter, etc)
66 Jaszczak Phantom
67
68 PET/CT Gantry Alignment CT PET
69 Annual QC Procedures Perform a sub-set of the acceptance test: Uniformity Resolution Count Rate Test Dead Time correction
70 Test Phantoms
Radionuclide Imaging MII Positron Emission Tomography (PET)
Radionuclide Imaging MII 3073 Positron Emission Tomography (PET) Positron (β + ) emission Positron is an electron with positive charge. Positron-emitting radionuclides are most commonly produced in cyclotron
More informationMayneord-Phillips Summer School St Edmund Hall, University of Oxford July Proton decays to n, e +, ν
Positron Emission Tomography Physics & Instrumentation Dimitra G. Darambara, Ph.D Multimodality Molecular Imaging Joint Department of Physics RMH/ICR Outline Introduction PET Physics overview Types of
More informationPositron Emission Tomography
Positron Emission Tomography Presenter: Difei Wang June,2018 Universität Bonn Contents 2 / 24 1 2 3 4 Positron emission Detected events Detectors and configuration Data acquisition Positron emission Positron
More informationProspects for achieving < 100 ps FWHM coincidence resolving time in time-of-flight PET
Prospects for achieving < 100 ps FWHM coincidence resolving time in time-of-flight PET, 28-Feb-2012, ICTR-PHE, Geneva, Switzerland 1 Time-of-flight PET Colon cancer, left upper quadrant peritoneal node
More informationDetector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples
Aim of this tal Detector technology WMIC Educational Program Nuclear Imaging World Molecular Imaging Congress, Dublin, Ireland, Sep 5-8, 202 You can now the name of a bird in all the languages of the world,
More information6: Positron Emission Tomography
6: Positron Emission Tomography. What is the principle of PET imaging? Positron annihilation Electronic collimation coincidence detection. What is really measured by the PET camera? True, scatter and random
More informationTiming and Energy Response of Six Prototype Scintillators
Timing and Energy Response of Six Prototype Scintillators CCM Kyba 1, J Glodo 2, EVD van Loef 2, JS Karp 1, KS Shah 2 1 University of Pennsylvania 2 Radiation Monitoring Devices SCINT 2007 June 7, 2007
More informationCT-PET calibration : physical principles and operating procedures F.Bonutti. Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital.
CT-PET calibration : physical principles and operating procedures Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital Topics Introduction to PET physics F-18 production β + decay and annichilation
More informationDevelopment of a High Precision Axial 3-D PET for Brain Imaging
Development of a High Precision Axial 3-D PET for Brain Imaging On behalf of the AX-PET Collaboration SIENA - IPRD08 October 1st 4th, 2008 1 Outline Basics of Positron Emission Tomography (PET); Principle
More informationThe Physics of PET/CT scanners
The Physics of PET/CT scanners Ruth E. Schmitz, Adam M. Alessio, and Paul E. Kinahan Imaging Research Laboratory Department of Radiology University of Washington What Makes PET Useful? Positron emission
More informationFigure 1. Decay Scheme for 60Co
Department of Physics The University of Hong Kong PHYS3851 Atomic and Nuclear Physics PHYS3851- Laboratory Manual A. AIMS 1. To learn the coincidence technique to study the gamma decay of 60 Co by using
More informationMEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah
MEDICAL EQUIPMENT: NUCLEAR MEDICINE Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Introduction to Medical Imaging: Physics, Engineering and Clinical Applications, by Nadine Barrie Smith
More informationDesign of a Lanthanum Bromide Detector for TOF PET
Design of a Lanthanum Bromide Detector for TOF PET A. Kuhn, S. Surti, Member, IEEE, J. S. Karp, Senior Member, IEEE, P. S. Raby, K. S. Shah, A. E. Perkins, Member, IEEE, G. Muehllehner, Fellow Member,
More informationTime-of-Flight PET using Cherenkov Photons Produced in PbF 2
Photons Produced in PbF 2 R. Dolenec a, S. Korpar b,a, P. Križan c,a, R. Pestotnik a, A. Stanovnik d,a a, Ljubljana, Slovenia b Faculty of Chemistry and Chemical Engineering, University of Maribor, Slovenia
More informationRadiation Detection and Measurement
Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron
More informationIntroduction to SPECT & PET TBMI02 - Medical Image Analysis 2017
Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017 Marcus Ressner, PhD, Medical Radiation Physicist, Linköping University Hospital Content What is Nuclear medicine? Basic principles of Functional
More informationLecture 5: Tomographic nuclear systems: SPECT
Lecture 5: Tomographic nuclear systems: SPECT Field trip this saturday at 11 AM at UWMC meet in main hospital lobby at 11 AM if you miss the 'boat', page me at 540-4950 should take ~1 to 1.5 hours, depending
More informationhν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?
Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? 2. Briefly discuss dead time in a detector. What factors are important
More informationCHIPP Plenary Meeting University of Geneva, June 12, 2008 W. Lustermann on behalf of the AX PET Collaboration
CHIPP Plenary Meeting University of Geneva, June 12, 2008 W. Lustermann on behalf of the AX PET Collaboration INFN Bari, Ohio State University, CERN, University of Michigan, University of Oslo, INFN Roma,
More informationScintillation Detectors
Scintillation Detectors J.L. Tain Jose.Luis.Tain@ific.uv.es http://ific.uv.es/gamma/ Instituto de Física Corpuscular C.S.I.C - Univ. Valencia Scintillation detector: SCINTILLATION MATERIAL LIGHT-GUIDE
More informationDual Isotope Imaging with LaBr3:Ce Crystal and H8500 PSPMT
Dual Isotope Imaging with LaBr3:Ce Crystal and H8500 PSPMT Dr. Andrea Fabbri, University of Rome Roma Tre I.N.F.N. (National Institue of Nuclear Physics) γ-ray imaging with scintillator and PSPMT γ-ray
More informationChapter 2 PET Imaging Basics
Chapter 2 PET Imaging Basics Timothy G. Turkington PET Radiotracers Positron emission tomography (PET) imaging is the injection (or inhalation) of a substance containing a positron emitter, the subsequent
More informationGamma ray coincidence and angular correlation
University of Cape Town Department of Physics Course III laboratory Gamma ray coincidence and angular correlation Introduction Medical imaging based on positron emission tomography (PET) continues to have
More informationSample Spectroscopy System Hardware
Semiconductor Detectors vs. Scintillator+PMT Detectors Semiconductors are emerging technology - Scint.PMT systems relatively unchanged in 50 years. NaI(Tl) excellent for single-photon, new scintillation
More informationSimulation of triple coincidences in PET
136 Institute of Physics and Engineering in Medicine Physics in Medicine & Biology doi:10.1088/0031-9155/60/1/117 Simulation of triple coincidences in PET J Cal-González 1,4, E Lage 2, E Herranz 1, E Vicente
More informationSub-Nanosecond Timing for In-Beam PET in hadrontherapy
Sub-Nanosecond Timing for In-Beam PET in hadrontherapy Baptiste Joly Clermont Université, Université Blaise Pascal, CNRS/IN2P3, Laboratoire de Physique Corpusculaire, BP 10448, F-63000 CLERMONT-FERRAND,
More informationRecent advances and future perspectives of gamma imagers for scintimammography
3rd International Conference on Imaging Technologies in Biomedical Sciences: ITBS2005 Innovation in Nuclear and Radiological Imaging: From Basic Research to Clinical Application Milos Conference Center,
More informationAdvances in PET technology
Advances in PET technology Jostein Sæterstøl September 15th 2009 Outline Introduction Detectors Time-of-flight PET PET / MRI Conclusions Introduction Fluor-18; FDG β + decay p n + e + + ν Annihilation
More informationList of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide
Nuclear Medicine Imaging Systems: The Scintillation Camera List of Nuclear Medicine Radionuclides Tc99m 140.5 kev 6.03 hours I-131 364, 637 kev 8.06 days I-123 159 kev 13.0 hours I-125 35 kev 60.2 days
More informationPositron Emission Tomography
Positron Emission Tomography CERN Accelerator School Small Accelerators Zeegse, the Netherlands A.M.J. Paans Nuclear Medicine & Molecular Imaging UMC Groningen Elements of Life PET-nuclide Hydrogen Carbon
More informationWhat is scintigraphy? The process of obtaining an image or series of sequential images of the distribution of a radionuclide in tissues, organs, or
Let's remind... What is nuclear medicine? Nuclear medicine can be broadly divided into two branches "in vitro" and "in vivo" procedures. There are numerous radioisotopic "in vitro" procedures for genotyping
More information1st Faculty of Medicine, Charles University in Prague Center for Advanced Preclinical Imaging (CAPI)
Radioation Resolution and Sensitivity Nuclear Imaging PET + SPECT Radioactive Decay (EC,Ɣ), (β -,Ɣ), (I.T.,Ɣ) β + Projection imaging collimator needed one angular view Projection imaging coincidence imaging,
More informationMitigation of External Radiation Exposures
Mitigation of External Radiation Exposures The three (3) major principles to assist with maintaining doses ALARA are :- 1) Time Minimizing the time of exposure directly reduces radiation dose. 2) Distance
More informationGAMMA RAY SPECTROSCOPY
GAMMA RAY SPECTROSCOPY Gamma Ray Spectroscopy 1 In this experiment you will use a sodium iodide (NaI) detector along with a multichannel analyzer (MCA) to measure gamma ray energies from energy level transitions
More informationA. I, II, and III B. I C. I and II D. II and III E. I and III
BioE 1330 - Review Chapters 7, 8, and 9 (Nuclear Medicine) 9/27/2018 Instructions: On the Answer Sheet, enter your 2-digit ID number (with a leading 0 if needed) in the boxes of the ID section. Fill in
More informationDELAYED COINCIDENCE METHOD FOR PICOSECOND LIFETIME MEASUREMENTS
306 DELAYED COINCIDENCE METHOD FOR PICOSECOND LIFETIME MEASUREMENTS ZHANG WEIJIE China Institute of Atomic Energy E-mail: zhangreatest@163.com The advanced time delay (ATD) technique, based by delayed
More informationLAB 4: Gamma-ray coincidence spectrometry (2018)
LAB 4: Gamma-ray coincidence spectrometry (2018) As you have seen, in several of the radioactive sources we encountered so far, they typically emit more than one gamma photon per decay or even more than
More informationA POSITION SENSITIVE ALPHA PARTICLE DETECTOR BASED ON A LYSO CRYSTAL AND A MICRO-PIXEL AVALANCHE PHOTODIODE
A POSITION SENSITIVE ALPHA PARTICLE DETECTOR BASED ON A LYSO CRYSTAL AND A MICRO-PIXEL AVALANCHE PHOTODIODE Ahmadov G.S. a,b,c, Ahmadov F.I. b,c, Kopatch Yu.N. a, Telezhnikov S.A. a, Nuriyev S.M. a,b,
More information4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values.
EXPERIMENT 2.1 GAMMA ENERGY CALIBRATION 1- Turn the power supply on to 900 V. Turn the NIM crate on to power the amplifiers. Turn the Oscilloscope on to check the gamma pulses. The main amplifier should
More informationSolid State LightBurst New PET Technology GE PET/CT and PET/MR
Solid State LightBurst New PET Technology GE PET/CT and PET/MR Osama Mawlawi PhD. Dept. of Imaging Physics MD Anderson Cancer Center Disclosures SIEMENS Research grant GE research grant Discovery MI LYSO
More informationA new timing model for calculating the intrinsic timing resolution of a scintillator detector
INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 5 (7) 3 7 PHYSICS IN MEDICINE AND BIOLOGY doi:.88/3-955/5/4/6 A new timing model for calculating the intrinsic timing resolution of a scintillator detector
More informationAcknowledgements. PET Fundamentals: Ideal Case. Why Are We Excited About PET/CT? PET Fundamentals: Real Case
PET/CT and Fusion Issues Jon A. Anderson Department of Radiology The University of Texas Southwestern Medical Center at Dallas American Associate of Physicists in Medicine 2003 Annual Meeting Acknowledgements
More informationInorganic Scintillators
Inorganic Scintillators Inorganic scintillators are inorganic materials (usually crystals) that emit light in response to ionizing radiation NaI is the protypical example Scintillation mechanism is different
More informationQuality Assurance. Purity control. Polycrystalline Ingots
Quality Assurance Purity control Polycrystalline Ingots 1 Gamma Spectrometry Nuclide Identification Detection of Impurity Traces 1.1 Nuclides Notation: Atomic Mass Atomic Number Element Neutron Atomic
More informationarxiv: v1 [physics.ins-det] 22 Dec 2013
arxiv:1312.6334v1 [physics.ins-det] 22 Dec 213 Test of the prototype of electron detector for LHAASO project using cosmic rays * WANG Xu 1 XU Tongye 1 DU Yanyan 1 SHAO Ruobin 1 ZHU Chengguang 1;1) for
More informationScintillating Crystals and their Applications in Particle and Nuclear Physics November 17-18, 2003
Scintillating Crystals and their Applications in Particle and Nuclear Physics November 17-18, 2003 Takayuki Yanagida, Hiromitsu Takahashi, Daisuke Kasama, Takeshi Ito, Hisako Niko, Motohide Kokubun, and
More informationRadiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors
Radiation Detectors 1 How do we detect ionizing radiation? Indirectly, by its effects as it traverses matter? What are these effects? Ionization and excitation of the atoms and molecules Heat 2 Types of
More informationScintillation Detector
Scintillation Detector Introduction The detection of ionizing radiation by the scintillation light produced in certain materials is one of the oldest techniques on record. In Geiger and Marsden s famous
More informationNuclear Reactions A Z. Radioactivity, Spontaneous Decay: Nuclear Reaction, Induced Process: x + X Y + y + Q Q > 0. Exothermic Endothermic
Radioactivity, Spontaneous Decay: Nuclear Reactions A Z 4 P D+ He + Q A 4 Z 2 Q > 0 Nuclear Reaction, Induced Process: x + X Y + y + Q Q = ( m + m m m ) c 2 x X Y y Q > 0 Q < 0 Exothermic Endothermic 2
More informationPET scan simulation. Meysam Dadgar. UMSU, Iran. IFMP, Elbasan, Fig 1: PET camera simulation in gate by cylindrical phantom
PET scan simulation Meysam Dadgar UMSU, Iran IFMP, Elbasan, 2016 Meysamdadgar10@gmail.com 1 Fig 1: PET camera simulation in gate by cylindrical phantom 2 What is PET? Positron emission tomography (PET),
More informationDevelopment of a new PET detector module with improved Depth of Interaction and Time of Flight capabilities
Development of a new PET detector module with improved Depth of Interaction and Time of Flight capabilities Ana Rita Borrego Instituto Superior Técnico IST, Universidade de Lisboa UL, Lisbon, Portugal
More informationChapter 4 Scintillation Detectors
Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 4-1 4.1. Basic principle of the scintillator Chapter 4 Scintillation Detectors Scintillator Light sensor Ionizing radiation Light (visible,
More informationMulti Channel Analyzer (MCA) Analyzing a Gamma spectrum
Multi Channel Analyzer (MCA) Analyzing a Gamma spectrum Objective: Using the MCA to acquire spectrums for different gamma sources and to identify an unknown source from its spectrum, furthermore to investigate
More informationPositron-Electron Annihilation
Positron-Electron Annihilation Carl Akerlof September 13, 008 1. Introduction This experiment attempts to explore several features of positron-electron annihilation. One of the attractive aspects of e
More informationInternational Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN
308 Angular dependence of 662 kev multiple backscattered gamma photons in Aluminium Ravindraswami K a, Kiran K U b, Eshwarappa K M b and Somashekarappa H M c* a St Aloysius College (Autonomous), Mangalore
More informationUsing new digital SiPM from Philips with AX-PET a new geometrical concept for PET
Using new digital SiPM from Philips with AX-PET a new geometrical concept for PET Matthieu Heller CERN - PH/DT Marie Curie network MC-PAD Matthieu.heller@cern.ch On behalf of the AX-PET collaboration https://twiki.cern.ch/twiki/bin/view/axialpet
More informationPropagation Speed of γ-radiation (Rγ) in Air (a)
1 Propagation Speed of γ-radiation (Rγ) in Air (a) Osmar F.S.L. Neto, (b)marcelo A.V. Macedo Jr., (c)josé T.P.D. Cavalcante, (c)henrique Saitovitch (a)fundação Técnico-Educacional Souza Marques Av. Ernani
More informationA Brief Introduction to Medical Imaging. Outline
A Brief Introduction to Medical Imaging Outline General Goals Linear Imaging Systems An Example, The Pin Hole Camera Radiations and Their Interactions with Matter Coherent vs. Incoherent Imaging Length
More informationAbsolute activity measurement
Absolute activity measurement Gábor Veres, Sándor Lökös Eötvös University, Department of Atomic Physics January 12, 2016 Financed from the financial support ELTE won from the Higher Education Restructuring
More informationGamma Spectroscopy. References: Objectives:
Gamma Spectroscopy References: G.F. Knoll, Radiation Detection and Measurement (John Wiley & Sons, New York, 2000) W. R. Leo, Techniques for Nuclear and Particle Physics Experiments: A How-to Approach,
More informationNuclear Lifetimes. = (Eq. 1) (Eq. 2)
Nuclear Lifetimes Theory The measurement of the lifetimes of excited nuclear states constitutes an important experimental technique in nuclear physics. The lifetime of a nuclear state is related to its
More informationPositron Emission Tomography (PET)
Positron Emission Tomography (PET) A radiological technique for functional imaging Please note that this exercise takes place at the Stockholm Centre for Physics, Astronomy and Biotechniques (Alba Nova).
More informationPositron emission tomography with additional c-ray detectors for multipletracer
Positron emission tomography with additional c-ray detectors for multipletracer imaging Tomonori Fukuchi, a) Takashi Okauchi, and Mika Shigeta RIKEN Center for Life Science Technologies, Kobe 650-0047,
More informationSCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator
88 The 1 st NPRU Academic Conference SCI-O11 Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator P. Limkitjaroenporn and W.Chewpraditkul Radiation Physics Laboratory,
More informationDetecting high energy photons. Interactions of photons with matter Properties of detectors (with examples)
Detecting high energy photons Interactions of photons with matter Properties of detectors (with examples) Interactions of high energy photons with matter Cross section/attenution length/optical depth Photoelectric
More informationEvaluation of the Nonlinear Response Function and Efficiency of a Scintillation Detector Using Monte Carlo and Analytical Methods
Asian J. Exp. Sci., Vol. 28, No. 2, 2014; 23-31 Evaluation of the Nonlinear Response Function and Efficiency of a Scintillation Detector Using Monte Carlo and Analytical Methods Rahim Khabaz, Farhad Yaghobi
More informationCharge collection in PET detectors
University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2007 Charge collection in PET detectors Tony Young University of
More informationCompton Camera. Compton Camera
Diagnostic Imaging II Student Project Compton Camera Ting-Tung Chang Introduction The Compton camera operates by exploiting the Compton Effect. It uses the kinematics of Compton scattering to contract
More informationDETECTORS. I. Charged Particle Detectors
DETECTORS I. Charged Particle Detectors A. Scintillators B. Gas Detectors 1. Ionization Chambers 2. Proportional Counters 3. Avalanche detectors 4. Geiger-Muller counters 5. Spark detectors C. Solid State
More informationAlpha particle scintillation detector based on micro pixel avalanche photodiode and LYSO crystal
Alpha particle scintillation detector based on micro pixel avalanche photodiode and LYSO crystal G.S. Ahmadov, F.I. Ahmadov Institute of Radiation Problems of ANAS, Baku, Azerbaijan C. Granja, S. Pospíšil
More informationRice University Physics 332 LIFETIME OF THE MUON I. INTRODUCTION...2! II. MEASUREMENT PROCEDURES...3! III. ANALYSIS PROCEDURES...7!
Rice University Physics 332 LIFETIME OF THE MUON I. INTRODUCTION...2! II. MEAUREMENT PROCEDURE...3! III. ANALYI PROCEDURE...7! Revised July 2011 I. Introduction In this experiment you will measure the
More informationOslo Cyclotron Laboratory
Oslo Cyclotron Laboratory Laboratory Excercise in FYS3180 - Experimental Methods in Physics Morten A. Salvesen University of Oslo, Oslo, Norway 2007-09-24 to 2007-11-15 Contents 1 Abstract 1 2 Oslo Cyclotrone
More informationRadioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques
Radioisotopes in action Diagnostic application of radioisotopes Steps of diagnostic procedure - Radioactive material introduced into the patient - Distribution and alteration of activity is detected -
More informationScintillation Detectors
Scintillation Detectors Introduction Components Scintillator Light Guides Photomultiplier Tubes Formalism/Electronics Timing Resolution Elton Smith JLab 2006 Detector/Computer Summer Lecture Series Experiment
More informationA Measurement of Monoenergetic Neutrons from 9 Be(p,n) 9 B
Journal of the Korean Physical Society, Vol. 32, No. 4, April 1998, pp. 462 467 A Measurement of Monoenergetic Neutrons from 9 Be(p,n) 9 B J. H. Kim, H. Bhang, J. H. Ha, J. C. Kim, M. J. Kim, Y. D. Kim
More informationCompton Imaging with Coincidence Measurements for Treaty Verification Purposes
1 Compton Imaging with Coincidence Measurements for Treaty Verification Purposes Scott Wandel, The Pennsylvania State University Abstract Nuclear arms control is a cooperative effort among many government
More informationOPPORTUNITY TO JOIN IEEE AND NPSS
OPPORTUNITY TO JOIN IEEE AND NPSS If you are NOT an IEEE Member, IEEE & NPSS offers you a FREE: Half-year membership in IEEE (value= ~$80)* Half-year membership in NPSS (value= ~$13)* Half-year subscription
More informationTHE COMPTON EFFECT Last Revised: January 5, 2007
B2-1 THE COMPTON EFFECT Last Revised: January 5, 2007 QUESTION TO BE INVESTIGATED: How does the energy of a scattered photon change after an interaction with an electron? INTRODUCTION: When a photon is
More informationApplication of positrons in materials research
Application of positrons in materials research Trapping of positrons at vacancy defects Using positrons, one can get defect information. R. Krause-Rehberg and H. S. Leipner, Positron annihilation in Semiconductors,
More informationWhat is Albira and How Does the System Work and Why is it Differentiated Technology? Page 1
What is Albira and How Does the System Work and Why is it Differentiated Technology? Page 1 Why Albira? Tri-modality: PET, SPECT, CT Modular and state-of-the-art electronics 6 configurations Novel, Proprietary
More information1 Introduction. KOPIO charged-particle vetos. K - RARE Meeting (Frascati) May Purpose of CPV: veto Kl
Introduction - Purpose of CPV: veto Kl decay modes with a real or apparent π and a pair of charged particles - Examples of background modes: (i) K l π π + π (ii) K l π π ± eν there are always (iii) K l
More informationThe 46g BGO bolometer
Nature, 3 The g BGO bolometer 1 Photograph of the heat [g BGO] and light [Ge; =5 mm] bolometers: see Fig. 1c for description Current events: Amplification gains: 8, (heat channel) &, (light channel). The
More informationProgress in using prompt gammas for ion range monitoring during hadrontherapy
Progress in using prompt gammas for ion range monitoring during hadrontherapy J. Krimmer 1, M. Chevallier 1, J. Constanzo 1, M. Dahoumane 1, D. Dauvergne 1, M. De Rydt 1,2, G. Dedes 1, N. Freud 3, J.M.
More informationEEE4106Z Radiation Interactions & Detection
EEE4106Z Radiation Interactions & Detection 2. Radiation Detection Dr. Steve Peterson 5.14 RW James Department of Physics University of Cape Town steve.peterson@uct.ac.za May 06, 2015 EEE4106Z :: Radiation
More informationCopyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy
Experiment VI Gamma Ray Spectroscopy 1. GAMMA RAY INTERACTIONS WITH MATTER In order for gammas to be detected, they must lose energy in the detector. Since gammas are electromagnetic radiation, we must
More informationThe Compton Effect. Martha Buckley MIT Department of Physics, Cambridge, MA (Dated: November 26, 2002)
The Compton Effect Martha Buckley MIT Department of Physics, Cambridge, MA 02139 marthab@mit.edu (Dated: November 26, 2002) We measured the angular dependence of the energies of 661.6 kev photons scattered
More informationAmorphous selenium (a-se) avalanche photodetector for applications in Positron Emission Tomography (PET)
Amorphous selenium (a-se) avalanche photodetector for applications in Positron Emission Tomography (PET) By OLEKSANDR BUBON Master of Science in Physics Lakehead University Thunder Bay, Ontario 2011 Amorphous
More informationChapter 6: Basic radiation detectors
Chapter 6: Basic radiation detectors Set of 60 slides based on the chapter authored by C.W.E. VAN EIJK Faculty of Applied Sciences, Delft University of Technology, Delft, Netherlands of the publication
More informationCharacterization and Monte Carlo simulations for a CLYC detector
Characterization and Monte Carlo simulations for a CLYC detector A. Borella 1, E. Boogers 1, R.Rossa 1, P. Schillebeeckx 1 aborella@sckcen.be 1 SCK CEN, Belgian Nuclear Research Centre JRC-Geel, Joint
More informationNew Photonis XP20D0 photomultiplier for fast timing in nuclear medicine
New Photonis XP20D0 photomultiplier for fast timing in nuclear medicine M. Moszyński, M. Gierlik, M. Kapusta, A. Nassalski, T. Szczęśniak, Soltan Institute for Nuclear Studies, PL 05-400 Świerk-Otwock,
More informationMEASURING THE LIFETIME OF THE MUON
B6-1 MEASURING THE LIFETIME OF THE MUON Last Revised September 19, 2006 QUESTION TO BE INVESTIGATED What is the lifetime τ of a muon? INTRODUCTION AND THEORY Muons are a member of a group of particles
More information3. Perturbed Angular Correlation Spectroscopy
3. Perturbed Angular Correlation Spectroscopy Dileep Mampallil Augustine K.U.Leuven, Belgium Perturbed Angular Correlation Spectroscopy (PAC) is a gamma ray spectroscopy and can be used to investigate
More informationMeasurements of liquid xenon s response to low-energy particle interactions
Measurements of liquid xenon s response to low-energy particle interactions Payam Pakarha Supervised by: Prof. L. Baudis May 5, 2013 1 / 37 Outline introduction Direct Dark Matter searches XENON experiment
More informationGamma Ray Spectroscopy
Gamma Ray Spectroscopy Uzair Latif, Imran Younus Department of Physics Lahore University of Management Sciences November 4, 2014 1 Objectives 1. To acquaint the students with some of the basic techniques
More informationScintillation Detectors
Radiation Measurement Systems Scintillation Detectors Ho Kyung Kim Pusan National University Scintillation detector = scintillator + light sensor Scintillators Inorganic alkali halide crystals Best light
More informationDetection of X-Rays. Solid state detectors Proportional counters Microcalorimeters Detector characteristics
Detection of X-Rays Solid state detectors Proportional counters Microcalorimeters Detector characteristics Solid State X-ray Detectors X-ray interacts in material to produce photoelectrons which are collected
More informationStudy of the feasibility of a compact gamma camera for real-time cancer assessment
Study of the feasibility of a compact gamma camera for real-time cancer assessment L. Caballero Instituto de Física Corpuscular - CSIC - University of Valencia; C/Catedrático José Beltrán, 2; E-46980;
More informationScintillation Detectors
Scintillation Detectors Introduction Components Scintillator Light Guides Photomultiplier Tubes Formalism/Electronics Timing Resolution Elton Smith JLab 2009 Detecto Summer Lecture Series Experiment basics
More informationarxiv: v2 [physics.ins-det] 8 Feb 2013
Preprint typeset in JINST style - HYPER VERSION arxiv:1302.0278v2 [physics.ins-det] 8 Feb 2013 Investigation of gamma ray detection performance of thin LFS scintillator with MAPD readout E.Guliyev a, F.Ahmadov
More informationKey words: avalanche photodiode, soft X-ray detector, scintillation γ-ray detector, imaging device PACS: 07.85;95.55.A;85.60.D
We have studied the performance of large area avalanche photodiodes (APDs) recently developed by Hamamatsu Photonics K.K, in high-resolution X-rays and γ- rays detections. We show that reach-through APD
More information