Positron emission tomography with additional c-ray detectors for multipletracer

Size: px
Start display at page:

Download "Positron emission tomography with additional c-ray detectors for multipletracer"

Transcription

1 Positron emission tomography with additional c-ray detectors for multipletracer imaging Tomonori Fukuchi, a) Takashi Okauchi, and Mika Shigeta RIKEN Center for Life Science Technologies, Kobe , Japan Seiichi Yamamoto Department of Radiological and Medical Laboratory Sciences, Nagoya University Graduate School of Medicine, Nagoya , Japan Yasuyoshi Watanabe and Shuichi Enomoto RIKEN Center for Life Science Technologies, Kobe , Japan (Received 21 September 2016; revised 27 January 2017; accepted for publication 29 January 2017; published 4 May 2017) Purpose: Positron emission tomography (PET) is a useful imaging modality that quantifies the physiological distributions of radiolabeled tracers in vivo in humans and animals. However, this technique is unsuitable for multiple-tracer imaging because the annihilation photons used for PET imaging have a fixed energy regardless of the selection of the radionuclide tracer. This study developed a multiisotope PET (MI-PET) system and evaluated its imaging performance. Methods: Our MI-PET system is composed of a PET system and additional c-ray detectors. The PET system consists of pixelized gadolinium orthosilicate (GSO) scintillation detectors and has a ring geometry that is 95 mm in diameter with an axial field of view of 37.5 mm. The additional detectors are eight bismuth germanium oxide (BGO) scintillation detectors, each of which is mm 3, arranged into two rings mounted on each side of the PET ring with a 92-mminner diameter. This system can distinguish between different tracers using the additional c-ray detectors to observe prompt c-rays, which are emitted after positron emission and have an energy intrinsic to each radionuclide. Our system can simultaneously acquire double- (two annihilation photons) and triple- (two annihilation photons and a prompt c-ray) coincidence events. The system s efficiency for detecting prompt de-excitation c-rays was measured using a positron-c emitter, 22 Na. Dual-radionuclide ( 18 F and 22 Na) imaging of a rod phantom and a mouse was performed to demonstrate the performance of the developed system. Our system s basic performance was evaluated by reconstructing two images, one containing both tracers and the other containing just the second tracer, from list-mode data sets that were categorized by the presence or absence of the prompt c-ray. Results: The maximum detection efficiency for 1275 kev c-rays emitted from 22 Na was approximately 7% at the scanner s center, and the minimum detection efficiency was 5.1% at the edge of the field of view. Dual-radionuclide imaging of the point sources and rod phantom revealed that our system maintained PET s intrinsic spatial resolution and quantitative nature for the second tracer. We also successfully acquired simultaneous double- and triple-coincidence events from a mouse containing 18 F-fluoro-deoxyglucose and 22 Na dissolved in water. The dual-tracer distributions in the mouse obtained by our MI-PET were reasonable from the viewpoints of physiology and pharmacokinetics. Conclusions: This study demonstrates the feasibility of multiple-tracer imaging using PET with additional c-ray detectors. This method holds promise for enabling the reconstruction of quantitative multiple-tracer images and could be very useful for analyzing multiple-molecular dynamics American Association of Physicists in Medicine [ Key words: dual-radionuclide imaging, positron emission tomography, positron gamma emitter, small animal imaging 1. INTRODUCTION Positron emission tomography (PET) is a powerful tool for noninvasive imaging of the distribution of a tracer in living organisms. However, conventional PET is restricted to singletracer imaging because it is based on the detection of positron annihilation photons, which always have an energy of 511 kev regardless of the variety of the positron emitter. If PET could image multiple tracers, it would be a much more powerful tool. In previous work, several methods have been proposed for multiple-tracer imaging using PET, such as using the difference between the half-lives of the tracers radionuclide decays or the detection of the prompt c-rays emitted after positron emission. A method using the difference in tracer decay times is not actually simultaneous multiple-tracer imaging because all such methods either assume a static tracer distribution or estimate the tracer dynamics based on pharmacokinetics. 1,2 On the other hand, a method that detect prompt c-rays, a multiisotope PET (MI-PET), is genuine simultaneous multiple Med. Phys. 44 (6), June /2017/44(6)/2257/ American Association of Physicists in Medicine 2257

2 2258 Fukuchi et al.: PET for multiple tracer imaging 2258 tracer imaging and has recently been actively studied. 3 6 A method using MI-PET combined with the administration of two tracers at different times to increase the imaging performance has also been proposed by Andreyev et al. 7 In addition to the development of a MI-PET scanner, the production of specific radionuclides for MI-PET has also been investigated. 8 SPECT also works for multiple-tracer imaging based on c-ray energy separation. However, it is still desirable to accomplish multiple-tracer imaging using PET, because it has higher sensitivity and quantitativity than other modalities including SPECT, and also because PET has its own unique tracers, such as 18 F-fluorodeoxyglucose (FDG). Additionally, although the sensitivity for the second tracer in MI-PET is lower than that of the conventional PET, MI-PET still possesses high sensitivity for both tracers. In the MI-PET method, the PET detector itself works as the prompt c-ray detector; however, employing additional detectors enhances the detection efficiency. 4 6 To demonstrate MI-PET s feasibility, we developed a prototype MI- PET system that is composed of a conventional circular ring PET and additional detectors to detect prompt c-rays. The PET system and the additional detectors consist of gadolinium orthosilicate (GSO) pixel detectors and eight largevolume ( mm 3 ) bismuth germanium oxide (BGO) detectors, respectively. There is another method for only detecting prompt c-rays using PET detectors. 3 However, the detection efficiency of PET detectors is limited, especially for high-energy c-rays, because PET detectors are designed for identifying 511 kev annihilation c-rays. Therefore, to maintain the sensitivity for high-energy c-rays, our system uses a ring configuration for the additional c-ray detectors, but other detector configurations are also possible, such as the end-cap configuration proposed by Levin et al. 5 Some of the main considerations of our configuration are the accessibility and ease of animal handling on a moving bed. The uniformity of the sensitivity for prompt c-rays in a field of view (FOV) is also considered important for normalizing the second tracer. However, further study is required to optimize the configuration of additional detectors in MI-PET. In addition to conventional double coincidence between the PET detectors, triple coincidence among the additional detectors and the PET system can be performed. Using a positron-c emitter as the second tracer, tracer identification can be made using additional detectors that locate prompt c- rays whose energy is characteristic of the radionuclide. We expect that MI-PET will be used in various new applications that are impracticable with single-tracer imaging, such as simultaneously performing multiple diagnostic scans, analyzing time sequences from multiple tracers and the effect of additional dosing with the same drug, and validating drug delivery systems by dual labeling. In this study, we investigated dual-radionuclide imaging using our newly developed MI-PET system to evaluate its imaging performance. To evaluate its basic performance, images of both tracers and just the second tracer were reconstructed from list-mode data sets that categorized data based on whether prompt c-rays were detected. 2. MATERIALS AND METHODS 2.A. Principles The positron emitters employed in PET imaging can be categorized into two types: pure positron emitters and positron-c emitters. Figure 1 shows two typical b + -decay schemes. Decay type (A) is a pure positron emitter. In such of decay, the nucleus undergoes a transition to the ground state of a daughter nucleus through b + -decay. Then, only a positron (with a neutrino) is emitted in the b + -decay. In contrast, decay type (B) is a positron-c emitter. In this type of decay, the nucleus undergoes a transition to an excited state of the daughter nucleus through b + -decay. Then, a c-ray is emitted after positron emission to transfer the energy from the excited state to the ground state. This c-ray is called a deexcitation c-ray, which has characteristic energy for each radionuclide. In some nuclides, the de-excitation c-ray is emitted with some delay from the positron emission because the nuclide goes through the isomeric state. Therefore, we use the term prompt c-ray to indicate a promptly emitted de-excitation c-ray to distinguish them from the annihilation c-rays that are produced by positron-electron pair-annihilation. MI-PET distinguishes between different tracers based on the detection of prompt c-rays. In dual-radionuclide imaging, it is possible to use a pure positron emitter and a positron-c emitter and distinguish some decay between these radioisotopes by the presence or absence of prompt c-rays. By using more than two positron-c emitters and detecting the energy of the prompt c-rays, more than three radioisotopes can be imaged simultaneously. To be used in a MI-PET, a positron-c emitter is required to have such particular characteristics, as an adequate lifetime and a large enough positron and prompt c-ray emissions per decay. Considering all of the required characteristics, the suitable radionuclides, which can be used as a positron-c emitter for MI-PET, are shown in Table I. Here, the e + ratio is the probability of the positron emission per decay, and the prompt c-ray ratio is the probability of prompt c-ray emission after positron emission 9. This table contains 22 Na whose lifetime is too long for clinical use, but since it is useful for phantom or animal imaging, we used it FIG. 1. Decay schemes for two types of b + -decay: (a) pure positron emitter and (b) positron-c emitter. Z and Z-1 are the atomic numbers of the radionuclides.

3 2259 Fukuchi et al.: PET for multiple tracer imaging 2259 TABLE I. Candidate radionuclides for use as a positron-c emitter in MI-PET imaging. Nuclide Half-life e + Ratio (%) c Energy (kev) Prompt c ratio (%) 14 O s Na yr m Cl min K min Sc 3.97 h V days Mn days m Mn 21.1 min Fe h Cu 23.7 min As 26.0 h Br 16.2 h m Rb h Y h m Tc 52.0 min m In 69.2 min I days in this study. For multiple prompt c-ray emitters, only prompt c-ray energies with strong intensities are noted. 2.B. System description Figure 2 shows a photograph and a schematic illustration of the developed system. The prototype is composed of a PET system and additional c-ray detectors. The PET system has a ring geometry and consists of pixelized GSO scintillation detectors. The additional detectors are eight BGO scintillation detectors. The PET system is based on the open-close PET developed by Yamamoto et al. 10 which can change the scanner configuration between a closed mode (ring configuration) and an open mode (two separated semicircle configuration). In this system, the detectors are a combination of mm 3 and mm 3 GSO pixel scintillation detectors with fast (35 ns) and slow (60 ns) decay times, respectively. These two types of GSO detectors were optically coupled to determine the depth of interaction from the differences in decay times. These GSO pixels are arranged into a matrix with a 0.1-mm-thick BaSO 4 reflector at a pitch of 1.7 or 2.5 mm. To configure a ring geometry, two scintillator matrices were arranged on a flat panel photomultiplier tube (FPPMT; H8500, Hamamatsu) at an angle of The energy resolution of the PET GSO detector was 28.4% (FWHM) for a 511 kev c-ray. 11 The details of the structure and the performance of this angled geometry and contact with FPPMT were described by Okumura et al. 11 Eight FPPMTs with scintillator blocks were arranged in a hexagonal shape to form a ring-like PET system whose inner diameter was 95 mm and whose axial FOV was 37.5 mm. We used the closed mode in this study. For the detection of prompt c-rays, we added eight BGO scintillation detectors with dimensions of mm 3. All were optically coupled to a 2 in. square photomultiplier tube. The measured energy resolution of a BGO detector was 10.3% (FWHM) for 1275 kev c-rays. Two rings with a 92-mm inner diameter were formed using four BGO detectors per ring and mounted on each side of the PET ring with an axial distance of 46 mm between the centers of the PET and BGO detectors. For the signal processing, we used the same front-end electronics for the block detectors as those used for the position-sensitive photomultiplier tube in a brain PET FIG. 2. Photograph (left) and schematic illustration (right) of developed MI-PET system. [Color figure can be viewed at wileyonlinelibrary.com]

4 2260 Fukuchi et al.: PET for multiple tracer imaging 2260 system developed by Yamamoto et al. 12 The system was modified to measure the triple coincidence between the PET and the BGO detectors in addition to the PET double coincidence for the annihilation photons. This triple-coincidence measurement can be performed in parallel with the PET double-coincidence measurement with list-mode data acquisition. To distinguish between double- and triple-coincidence data, a tag was recorded that indicate triple coincidence. The width of the coincidence time window was set to 20 ns for both double and triple coincidence. Although this time window is wide for the intrinsic time resolution of the GSO and BGO detectors, it is set to correspond to the time jitter originating from the energy variation of the leading edge timing and the individuality of the timing performance in each detector. For the PET detectors, the lower energy threshold level was set to 350 kev, and the upper energy threshold level remained open. In contrast, the lower- and upper-level thresholds of the energy window for the additional BGO detectors were set to 1.0 and 1.5 MeV to focus on the 1275 kev c-rays originating from the 22 Na used in this study. The energy window of the BGO detectors was set to exclude the 511 kev annihilation c-rays. Consequently, in addition to the photo-peak events, some parts of the Compton scattering events of the 1275 kev c-rays, which deposit part of their initial energy, were also detected. Therefore, the efficiency of the BGO detectors includes these Compton scattering events. However, because the PET detectors have open thresholds for the upper energy limit, they also detect 1275 kev c-rays, if the energy upper level is set, although the Compton scattering events of the 1275 kev c-rays were still detected. These events are the same as the random coincidence events in a conventional PET and can be corrected using the delayed coincidence method. Thus, in our system, only energy discrimination is used to distinguish prompt and annihilation c-rays, although another method could distinguish between them based on the collinearity of annihilation photons. 8,13 2.C. Image reconstruction The list-mode coincidence data are classified into two groups based on the presence or absence of prompt c-rays. After the classification, a conventional image reconstruction algorithm is executed for each set of the data. In our system, first, the projection data are produced using single-slice rebinning with a maximum ring difference of eight. Image reconstruction from the projected sinogram was performed using inter-update Metz-filtered ordered subset expectation maximization (IMF-OSEM) 14 in a program called Software for Tomographic Image Reconstruction (STIR). 15 The resulting image had a size of voxels with a voxel size of mm 3. For the IMF-OSEM, the subset number was fixed to be four. The images reconstructed from the data without and with the prompt c-rays coincidence are called image-d (double coincidence) and image-t (triple coincidence). Based on the delayed coincidence, the data random coincidence background was subtracted. The random coincidence between the annihilation photons and the prompt c-ray, which causes biased noise reflecting the distribution of the pure positron emitter, was also subtracted using the delayed coincidence data. Perhaps triple coincidence could be generated by scattering photons between the two sets of detectors, 16,17 but having a lower level energy threshold of 1.0 MeV for the BGO detectors, which is approximately double the energy of the 511 kev annihilation photons, removes most of this scattering from the data. From the measurements, triple-coincidence events occur very rarely (below 0.1%) when using just a pure positron emitter 18 F with activity of ~1 MBq. However, in dual-tracer experiments, random coincidence events were detected between the annihilation c-rays from either the pure positron emitter or the positron-c emitter and the prompt c-rays. The random coincidence events between the annihilation c-rays from the pure positron emitter and the prompt c-rays from the positron-c emitter produced biased noise reflecting the distribution of the pure positron emitter in the image reconstructed from triple-coincidence events where only the positron-c emitter should appear without any correction. This biased noise can also be subtracted using a delayed coincidence method. In a mouse experiment using 18 F (810 kbq) and 22 Na (410 kbq) in Section 3.C, the delayed coincidence fraction of the true triple coincidence was 14.1%. 2.D. Image normalization In MI-PET, normalization of the positional dependence of the prompt c-ray detection efficiency is necessary in addition to the conventional image normalization of PET. Therefore, a long-period scan was performed using a cylindrical 22 Na phantom (diameter: 78 mm; length: 180 mm). Sodium-22 emits both a positron and a 1275 kev c-ray (Table I). The phantom had 1 MBq of total activity, which corresponds to 1.1 kbq/ml. A total of double- and triple-coincidence events were recorded over 87 h of measurement. Projection data were produced for each set of data and used for the normalization of the double- and triple-coincidence images. This normalization measurement was for the 1275 kev c-rays of 22 Na used in this study; if another radionuclide is used as a tracer, a specific measurement is necessary for normalization with it. If the measurement is difficult, some simulations or interpolations for different energies can be executed instead. In this study, c-ray absorption in the imaged objects was ignored, because the phantom and animal had small volumes. 3. PERFORMANCE EVALUATION 3.A. Point sources To evaluate the prompt c-ray detection efficiency and the spatial resolution for dual-radionuclide imaging, we made a 0.3-mm-diameter 22 Na point source with 42 kbq activity.

5 2261 Fukuchi et al.: PET for multiple tracer imaging A.1. Detection efficiency for prompt c-rays The positional dependence of the prompt c-ray detection efficiency was measured in the transverse (x-axis) and sagittal (z-axis) directions in 2.5-mm increments from the scanner s center. The detection efficiency of the triple coincidence (P t ) in each position is defined as P t ¼ P d P g ; where P d is the detection efficiency of the double coincidence for the annihilation c-rays and P g is the prompt c-ray detection efficiency. P g is the ratio of the double and triple coincidences: Pg ¼ P t =P d : We must consider the emission fraction of prompt c-rays, but 22 Na, which was used for efficiency measurements, emits a prompt c-ray after positron emission 99.9% of the time (Table I), and when PET detects annihilation photons generated by positron emission, a prompt c-ray is also emitted at the same time. Then, efficiency P g is the same as the absolute efficiency of the prompt c-ray detection in the first approximation. The acquisition time for each measurement was 5 min. We also measured the positional dependence along a line in the x-y plane at 45 to the x- and y-axes (the x-y direction for brevity). This radial direction differs from the x-direction, since all the BGO detectors are mounted, so they point in a direction 45 to the x- and y-axes. To obtain a global view, the prompt c-ray detection efficiency on the x-z plane was also measured on a 5-mm mesh. These measured points are depicted in Fig. 3. In addition to the measurement using 22 Na, to investigate the energy dependence of the prompt c-ray efficiency (P g ), we also measured the efficiency of the BGO detectors using single c-ray emitters 54 Mn (835 kev) and 137 Cs (662 kev). 3.A.2. Dual-radionuclide imaging To evaluate the spatial resolution of the second tracer ( 22 Na, positron fraction 90.4%, Table I) when used in combination with another positron emitter, we prepared an 18 F (positron fraction 96.7%) point source in addition to the 22 Na point source. The 18 F point source had a diameter of 0.8 mm with 47 kbq activity (comparable with the activity of the 22 Na point source) at the start of the measurement. The two point sources were arranged along the x-axis at 20-mm intervals in three arrangements: (a) 22 Na at the center, (b) the two point sources equidistant from the center, and (c) 18 F at the center. With respect to the z-axis, the point sources were arranged in the same way as on the x-axis with a 10-mm interval (arrangements (d)-(f)). These arrangements are shown in Fig 4. Measurements were performed for 5 min in each arrangement. After image reconstruction using three-dimensional filtered backprojection, the spatial resolutions for the second tracer ( 22 Na) were evaluated for each source arrangement. 3.B. Rod phantom To make a quantitative assessment of the imaging ability of the MI-PET system, we scanned a dual-radionuclide rod phantom that had three cylindrical rods with a diameter of 10 mm and a length of 76 mm. Fluorine-18 and 22 Na dissolved in water were poured into the rods. The first rod had 18 F (686 kbq at the start of the measurement) and 22 Na (585 kbq) activities. The second and third rods had 18 F (636 kbq at the start of the measurement) and 22 Na (585 kbq) activities, respectively. The three rods were arranged in an equilateral triangle at a distance of twice the rod diameter. The rods were positioned at the scanner s center parallel to the scanner s axial FIG. 3. Measurement positions for prompt c-ray detection efficiency using a 22 Na point source. Z-axis is axial direction, and x- and y-axes are transverse horizontal and vertical directions, respectively. The line labeled x-y lies in the x-y plane. The origin of coordinate axes is the scanner s center. FIG. 4. Arrangements of point sources for dual-radionuclide imaging. (a) (c) and (d) (f), respectively show x-axis (transverse) and z-axis (axial) arrangements.

6 2262 Fukuchi et al.: PET for multiple tracer imaging 2262 FIG. 5. Illustration of rod phantom configuration. Three rods with activity 18 F (636 kbq) + 22 Na (585 kbq), 18 F (636 kbq), and 22 Na (585 kbq) were called rods 1, 2, and 3, respectively. [Color figure can be viewed at wileyonlinelibrary.com] direction. The configuration and activities of the rod phantom are shown in Fig. 5. A 30-min scan was performed for the double and triple coincidences in parallel. For each of the three rods, the regions of interest (ROIs) were chosen to make a quantitative comparison between the original activities and those deduced from the PET images. FIG. 6. Axial (z-axis) positional dependence of efficiencies of additional detectors for 1275 kev c-rays of 22 Na. 3.C. Small animal study To test the practical performance of the developed MI- PET system, we carried out small animal imaging. In this experiment, 410 kbq 22 Na and 810 kbq 18 F-FDG were administered to an 8-week-old normal male mouse by oral administration and tail vein injection, respectively. Ten minutes after administration, a 30-min scan with bed motion was performed under anesthesia. Image reconstructions using the IMF-OSEM were carried out for data with and without the detection of the prompt c-rays. This animal experiment was performed in accordance with the Principles of Laboratory Animal Care (NIH Publication No , revised 1985) and approved by the Institutional Animal Care and Use Committee (IACUC) of RIKEN, Kobe Branch. FIG. 7. X- and x-y direction dependence of efficiency of additional detectors for 1275 kev c-rays of 22 Na. 4. RESULTS 4.A. Point sources 4.A.1. Prompt c-ray detection efficiency In the measurements of prompt c-ray detection efficiency, approximately 400,000 events were accumulated at the center position in 5-min measurements. Accumulated event numbers at the other positions were of the same order. The prompt c-ray detection efficiency in each position (P g ) contains accidental coincidence events between the annihilation photons and the prompt c-rays. However, the rate of accidental coincidence calculated using the coincidence window (20 ns) of the PET and the additional detectors count rate (PET: 450 cps; BGO: 3000 cps) in these measurements were less than 0.03 cps, which is negligible. The prompt c- FIG. 8. Measured efficiency of additional detectors for 1275 kev c-rays of 22 Na in x-z plane with a 5-mm mesh. ray detection efficiency for a 1275 kev c-ray at the center position was approximately 7%. Figures 6 and 7 show the axial (z-axis) and transverse (xaxis) dependence of the prompt c-ray detection efficiency. Figure 8 shows the prompt c-ray detection efficiency on the x-z plane with a 5-mm mesh. It monotonically decreased in both

7 2263 Fukuchi et al.: PET for multiple tracer imaging 2263 FIG. 9. Examples of dual-radionuclide images and profiles for point sources in arrangement 2 (a) and arrangement 4 (b). Images and profiles are center slices of image-d (left) and image-t (right). These images have a voxel size of mm 3. the x- and z-directions as the distance from the scanner s center increased, and so a position at the edge of the FOV should have the lowest prompt c-ray detection efficiency. The ratio of the maximum detection efficiency (7.0% at the scanner s center) and the minimum detection efficiency (5.1% at the edge of the FOV, x = 30 mm, z = 15 mm) was about From the measurements for energy dependence, the c-ray photo-peak efficiencies for 54 Mn (835 kev) and 137 Cs (662 kev) in the center position were 7.8% and 9.4%, respectively. 4.A.2. Dual-radionuclide imaging In dual-radionuclide imaging, the total accumulated numbers of double- and triple-coincidence events for the layout shown in Fig. 4 (a) were and , respectively. The double-coincidence number includes events originating from the annihilation c-rays from the pure positron emitter ( 18 F). For the other layouts in Figs. 4(b) (f), the total accumulated counts were approximately the same as those for the layout in Fig. 4(a). Figure 9 shows a slice image of the 18 F and 22 Na point sources. Image-D (left) and image-t (right) were reconstructed from data without and with the detection of the prompt c-rays, respectively. The upper and lower images are those of point sources placed along the x- and z-axes, respectively, equidistant from the scanner s center (Figs. 4(b) and 4(e)). In both layouts, we can clearly observe the 18 F and 22 Na point sources in the image-d and the 22 Na point source in image-t. Table II summarizes the spatial

8 2264 Fukuchi et al.: PET for multiple tracer imaging 2264 TABLE II. Spatial resolutions for a 22 Na point source in image-d and image-t. Layout Axis x x x z z z Image-D (mm) Image-T (mm) Values are full widths at half-maximum in millimeters. Values in layouts 1 3 are x-y plane spatial resolutions and those for layouts 4 6 are z-projection spatial resolutions. Layouts of 18 F and 22 Na point sources are shown in Fig. 4. resolutions of the 22 Na point source in image-d and image-t in the presence of another radionuclide ( 18 F). The spatial resolutions for the arrangements in Figs. 4(a) (c) in image-t are for the x-y plane, and those for the arrangements in Figs. 4(d) (f) are for the z projection. At the center, the x-y spatial resolution in image-t of the layout in Fig. 4(a) was 2.6 mm and the z-projection resolution was 2.4 mm. 4.B. Rod phantom In the rod phantom measurement, the count rate at the starting time was 15.5 kcps double- and triple-coincidence events were acquired. The double-coincidence number includes events originating from the annihilation c-rays from the pure positron emitter ( 18 F). Images were reconstructed using IMF-OSEM with six iterations and four subsets. In this image reconstruction, image normalization and delayed background subtraction were also executed. Figure 10 shows the x-y plane projection images of the rod phantom in image-d (left) and image-t (right). In both images, we can clearly observe the 18 F and 22 Na activities with reasonable intensities. Comparisons between ROIs and original activities are shown in Table III. The ROIs were set to the whole region of each rod in both images. The values in image-d and image-t were normalized by the measured total activity of rod 1 in image-d. From these values, the original activities of both radioisotopes were well reconstructed in both image-d and image-t. The ROI activity of rod 2 in image-t, which does not have 22 Na, is small. This nonzero activity is conceivably due to the accidental coincidence between the annihilation photons of 18 F and the 1275 kev c-rays of 22 Na that are TABLE III. Measured rod activities for image-d and image-t in dual-radionuclide rod phantom. Layout Rod 1 ( 18 F+22Na) Rod 2 ( 18 F) Rod 3 ( 22 Na) Image-D Image-T Activities were normalized by total activity of rod 1 in image-d. not fully removed by the delayed background correction. However, the amount of remaining activity is comparable to the error of the activities in other ROIs. The activities in every slice of the ROIs were also evaluated to check the uniformity along the z-axis. Figures 11 and 12 show the activities in each z slice for image-d and image-t, respectively. The nonuniformity of image-d was in the range of a few percent and this value is sufficiently small. The nonuniformity of image-t is rather large compared with that in image-d, but it is also acceptable. 4.C. Small animal study In the mouse experiment, the initial double- and triplecoincidence count rates were 9.7 kcps and 340 cps, leading to a total of accumulated events in a 30-min scan of FIG. 11. Uniformity of measured rod activities in z-axis slices of image-d. FIG. 10. X-y plane projection of rod phantom in image-d (left) and image-t (right).

9 2265 Fukuchi et al.: PET for multiple tracer imaging 2265 FIG. 12. Uniformity of measured rod activities in z-axis slices of image-t and counts, respectively. The doublecoincidence number includes events originating from the annihilation c-rays from the pure positron emitter ( 18 F). Images were reconstructed using IMF-OSEM with six iterations and four subsets. Figure 13 shows the results of dualradionuclide small animal imaging. We can clearly see 22 Na in the esophagus and stomach, whereas 18 F-FDG is distributed in the brain, heart, kidneys, and urinary bladder. 5. DISCUSSION For prompt c-ray detection efficiency, the sensitivity of the open-close PET to annihilation photons decreases proportionally with the distance from the scanner s center 10. In contrast, the sensitivity of the additional detectors for the prompt c-rays also decreases proportionally with the distance from the scanner s center but more drastically. This tendency may reflect the fact that the additional detectors on the far side are shadowed by the PET detectors. The transverse prompt c-ray detection efficiency in Fig. 7 also decreases proportionally with the distance from the scanner s center, but its variation is slower than the variation in the axial position. This difference is presumably due to the fact that the prompt c-ray detectors on both sides are not shadowed in this case. The detection efficiency along the slanting line in the x-y plane in Fig. 7 decreases more slowly than it does along the x-axis. This can be explained by the fact that the x-y direction is parallel to the axis of one of the additional BGO detectors. The minimum prompt c-ray detection efficiency of approximately 5% indicates that it takes at least 20 times longer to obtain a second tracer image with comparable quality to conventional PET imaging. However, Cal-Gonzalez et al. showed that the triple-coincidence fraction in a small animal PET (Argus small animal PET/CT scanner) for 22 Na (1275 kev) is about 2.5%. 17 Because the PET detectors were designed for 511 kev annihilation c-rays, the detection efficiency for highenergy c-rays is lower. Furthermore, the prompt c-ray efficiency of the Argus scanner for 76 Br (559 kev) is smaller than that of our system s 9.4% for 137 Cs (662 kev). The spatial resolution of the second tracer is comparable to the intrinsic open-close PET scanner performance of 2.4- mm full width at half-maximum of the z-projection image in the center position 10. The spatial resolutions away from the center also behave in the same manner as in single-tracer imaging with an open-close PET. As a result, we found that the spatial resolutions for the second tracer in dual-radionuclide imaging are comparable with the scanner s intrinsic spatial resolutions. This means that the first tracer does not affect the spatial resolution of the second positron-c emitter in a reconstructed image with prompt c-ray detection (image-t). The spatial resolution of image-t probably only depends on the accumulated number of events. In the rod phantom measurement, we reconstructed the original activity of the dual radioisotopes. The values in image-t for rod 2, which has no activity in the original phantom, are not zero. As discussed above, this might be because of incomplete delayed background subtraction, and the FIG. 13. Results of dual-radionuclide imaging of a mouse. 18 F-FDG and 22 Na were administered by tail vein injection and oral administration, respectively. Images are maximum intensity projections.

10 2266 Fukuchi et al.: PET for multiple tracer imaging 2266 magnitude of this remaining activity is comparable with the error in the other rods activities. From this result, we concluded that our normalization and delayed coincidence methods, which are the same as those used in conventional PET image reconstructions, are applicable for MI-PET. In this study, 22 Na was employed as the second tracer, and the data for normalization were also measured using a 1275 kev c-ray from 22 Na. However, if another radionuclide is used, such as one of those in Table I, measurements or simulations at specific energy for these radionuclides are necessary for image reconstruction. A dual-radionuclide experiment on a small animal shows that our developed system and the image reconstruction method work for distributed tracers in a living organism. The measured distributions of both the first and second tracers are reasonable from a physiological viewpoint. The additional detectors are sensitive to the prompt c-rays from out of the FOV. These c-rays increase the random coincidence noise, which, however, was subtracted using a delayed coincidence method, at least, at the activity levels in our measurements. To handle higher activities, further investigation and the development of equipment appear necessary, such as the addition of a shield for the c-rays from out of the FOV. In both the phantom and animal studies, we reconstructed images for the presence and absence of prompt c-ray detection. Therefore, a superposed image of the first ( 18 F: pure positron emitter) and the second ( 22 Na: positron-c emitter) tracers and an image of only the second tracer were reconstructed. To reconstruct an independent image of a pure positron emitter, we must differentiate the first tracer image from the images with and without prompt c-ray detection and perhaps develop another iterative method. 7 We intend to develop optimal image reconstruction methods for our system in future work. 6. CONCLUSION We developed an MI-PET system and successfully performed dual-radionuclide imaging for a phantom and a small animal. In this study, we reconstructed the superposed distribution of the first and second tracers and the distribution of the second tracer. Even though there are applications for which superposed tracer distributions may be useful, we need to develop a specific image reconstruction method for MI- PET to obtain independent images of the tracers. We will also use our developed MI-PET system for various applications of multiple-tracer imaging. ACKNOWLEDGMENTS The authors thank the members of the RIKEN Center for Life Science Technologies for their support. This work was supported by JSPS KAKENHI Grant Numbers JP15H04770 and JP CONFLICT OF INTEREST The authors have no relevant conflicts of interest to disclose. a) Author to whom correspondence should be addressed. Electronic mail: tfukuchi@riken.jp. REFERENCES 1. Kadrmas DJ. Feasibility of rapid multitracer PET tumor imaging. IEEE Trans Nucl Sci. 2005;52: Ben-Haim S, Kacperski K, Hain S, et al. Simultaneous dual-radionuclide myocardial perfusion imaging with a solid-state dedicated cardiac camera. Eur J Nucl Med Mol Imaging. 2010;37: Andreyev A, Celler A. Dual-isotope PET using positron-gamma emitters. Phys Med Biol. 2011;56: Miyaoka RS, Hunter WCJ, Andreyev A, et al. Dual-radioisotope PET data acquisition and analysis. In: Proceedings of the IEEE Nuclear Science Symposium and Medical Imaging Conference Valencia, Spain: IEEE; 2011: (2012). 5. Gonzalez E, Olcott PD, Bieniosek M, Levin CS. Methods for increasing the sensitivity of simultaneous multi-isotope positron emission tomography. In: Proceedings of the IEEE Nuclear Science Symposium and Medical Imaging Conference Valencia, Spain: IEEE; 2011: (2012). 6. Olcott PD, Levin CS. Methods and systems for increasing the sensitivity of simultaneous multi-isotope positron emission tomography. Patent US B2; Andreyev A, Sitek A, Celler A. EM reconstruction of dual isotope PET using staggered injections and prompt gamma positron emitters. Med Phys. 2014;41: Herraiz JL, Lage E, Parot V, et al. Production of positron-gamma emitters for multiplexed PET (mpet) imaging. Proc IEEE NNS/MIC. 2014;2013: National Nuclear Data Center. Brookhaven National Laboratory Yamamoto S, Okumura S, Watabe T, et al. Development of a prototype Open-close positron emission tomography system. Rev Sci Instrum. 2015;86: Okamura S, Yamamoto S, Watabe H, Kato N, Hamamura H. Development of dual-layer GSO depth-of-interaction block detector using angled optical fiber. Nucl Inst Meth Phys Res Sect A. 2015;781: Yamamoto S, Honda M, Oohashi T, Shimizu K, Senda M. Development of a Brain PET System, PET-Hat: a wearable PET system for Brain Research. IEEE Trans Nucl Sci. 2011;58: Lage E, Parot V, Moore SC, et al. Recovery and normalization of triple coincidences in PET. Med Phys. 2015;42: Jacobson M, Levkopvitz R, Ben-Tal A, et al. Enhanced 3D PET OSEM reconstruction using inter-update Metz filtering. Phys Med Biol. 2000;45: Thielemans K, Tsoumpas C, Mustafovic S, et al. STIR: software for tomographic image reconstruction release 2. Phys Med Biol. 2012;57: Gillam JE, Solevi P, Oliver JF, et al. Sensitivity recovery for the AX- PET prototype using inter-crystal scattering events. Phys Med Biol. 2014;59: Cal-Gonzalez J, Lage E, Herranz E, et al. Simulation of triple coincidences in PET. Phys Med Biol. 2015;60:

Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017

Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017 Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017 Marcus Ressner, PhD, Medical Radiation Physicist, Linköping University Hospital Content What is Nuclear medicine? Basic principles of Functional

More information

Radionuclide Imaging MII Positron Emission Tomography (PET)

Radionuclide Imaging MII Positron Emission Tomography (PET) Radionuclide Imaging MII 3073 Positron Emission Tomography (PET) Positron (β + ) emission Positron is an electron with positive charge. Positron-emitting radionuclides are most commonly produced in cyclotron

More information

Simulation of triple coincidences in PET

Simulation of triple coincidences in PET 136 Institute of Physics and Engineering in Medicine Physics in Medicine & Biology doi:10.1088/0031-9155/60/1/117 Simulation of triple coincidences in PET J Cal-González 1,4, E Lage 2, E Herranz 1, E Vicente

More information

1st Faculty of Medicine, Charles University in Prague Center for Advanced Preclinical Imaging (CAPI)

1st Faculty of Medicine, Charles University in Prague Center for Advanced Preclinical Imaging (CAPI) Radioation Resolution and Sensitivity Nuclear Imaging PET + SPECT Radioactive Decay (EC,Ɣ), (β -,Ɣ), (I.T.,Ɣ) β + Projection imaging collimator needed one angular view Projection imaging coincidence imaging,

More information

6: Positron Emission Tomography

6: Positron Emission Tomography 6: Positron Emission Tomography. What is the principle of PET imaging? Positron annihilation Electronic collimation coincidence detection. What is really measured by the PET camera? True, scatter and random

More information

www.aask24.com www.aask24.com www.aask24.com P=Positron E= Emission T=Tomography Positron emission or beta plus decay (+ ) is a particular type of radioactive decay, in which a proton inside a radionuclide

More information

Tomography is imaging by sections. 1

Tomography is imaging by sections. 1 Tomography is imaging by sections. 1 It is a technique used in clinical medicine and biomedical research to create images that show how certain tissues are performing their physiological functions. 1 Conversely,

More information

PET scan simulation. Meysam Dadgar. UMSU, Iran. IFMP, Elbasan, Fig 1: PET camera simulation in gate by cylindrical phantom

PET scan simulation. Meysam Dadgar. UMSU, Iran. IFMP, Elbasan, Fig 1: PET camera simulation in gate by cylindrical phantom PET scan simulation Meysam Dadgar UMSU, Iran IFMP, Elbasan, 2016 Meysamdadgar10@gmail.com 1 Fig 1: PET camera simulation in gate by cylindrical phantom 2 What is PET? Positron emission tomography (PET),

More information

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: NUCLEAR MEDICINE Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Introduction to Medical Imaging: Physics, Engineering and Clinical Applications, by Nadine Barrie Smith

More information

Positron Emission Tomography

Positron Emission Tomography Positron Emission Tomography Presenter: Difei Wang June,2018 Universität Bonn Contents 2 / 24 1 2 3 4 Positron emission Detected events Detectors and configuration Data acquisition Positron emission Positron

More information

CT-PET calibration : physical principles and operating procedures F.Bonutti. Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital.

CT-PET calibration : physical principles and operating procedures F.Bonutti. Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital. CT-PET calibration : physical principles and operating procedures Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital Topics Introduction to PET physics F-18 production β + decay and annichilation

More information

Nuclear Medicine Intro & Physics from Medical Imaging Signals and Systems, Chapter 7, by Prince and Links

Nuclear Medicine Intro & Physics from Medical Imaging Signals and Systems, Chapter 7, by Prince and Links Nuclear Medicine Intro & Physics from Medical Imaging Signals and Systems, Chapter 7, by Prince and Links NM - introduction Relies on EMISSION of photons from body (versus transmission of photons through

More information

CLINICALLY USEFUL RADIONUCLIDES:

CLINICALLY USEFUL RADIONUCLIDES: INTRODUCTION It is important that Nuclear Medicine Technologists be familiar with the imaging properties of all commonly used radionuclides to insure correct choice of isotope for a particular study as

More information

A. I, II, and III B. I C. I and II D. II and III E. I and III

A. I, II, and III B. I C. I and II D. II and III E. I and III BioE 1330 - Review Chapters 7, 8, and 9 (Nuclear Medicine) 9/27/2018 Instructions: On the Answer Sheet, enter your 2-digit ID number (with a leading 0 if needed) in the boxes of the ID section. Fill in

More information

Chapter 2 PET Imaging Basics

Chapter 2 PET Imaging Basics Chapter 2 PET Imaging Basics Timothy G. Turkington PET Radiotracers Positron emission tomography (PET) imaging is the injection (or inhalation) of a substance containing a positron emitter, the subsequent

More information

Radioisotopes and PET

Radioisotopes and PET Radioisotopes and PET 1 Radioisotopes Elements are defined by their number of protons, but there is some variation in the number of neutrons. Atoms resulting from this variation are called isotopes. Consider

More information

PET. Technical aspects

PET. Technical aspects PET Technical aspects 15 N 15 O Detector 1 β+ Detector 2 e- Evolution of PET Detectors CTI/Siemens 15 N 15 O Detector block 1 β+ Detector block 2 x e- x y y location line of response Constant fraction

More information

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques Radioisotopes in action Diagnostic application of radioisotopes Steps of diagnostic procedure - Radioactive material introduced into the patient - Distribution and alteration of activity is detected -

More information

Outline Chapter 14 Nuclear Medicine

Outline Chapter 14 Nuclear Medicine Outline Chapter 14 uclear Medicine Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther Introduction Detectors for nuclear

More information

The Physics of PET/CT scanners

The Physics of PET/CT scanners The Physics of PET/CT scanners Ruth E. Schmitz, Adam M. Alessio, and Paul E. Kinahan Imaging Research Laboratory Department of Radiology University of Washington What Makes PET Useful? Positron emission

More information

Study of the feasibility of a compact gamma camera for real-time cancer assessment

Study of the feasibility of a compact gamma camera for real-time cancer assessment Study of the feasibility of a compact gamma camera for real-time cancer assessment L. Caballero Instituto de Física Corpuscular - CSIC - University of Valencia; C/Catedrático José Beltrán, 2; E-46980;

More information

Compton Camera. Compton Camera

Compton Camera. Compton Camera Diagnostic Imaging II Student Project Compton Camera Ting-Tung Chang Introduction The Compton camera operates by exploiting the Compton Effect. It uses the kinematics of Compton scattering to contract

More information

What is scintigraphy? The process of obtaining an image or series of sequential images of the distribution of a radionuclide in tissues, organs, or

What is scintigraphy? The process of obtaining an image or series of sequential images of the distribution of a radionuclide in tissues, organs, or Let's remind... What is nuclear medicine? Nuclear medicine can be broadly divided into two branches "in vitro" and "in vivo" procedures. There are numerous radioisotopic "in vitro" procedures for genotyping

More information

Bioimage Informatics. Lecture 23, Spring Emerging Applications: Molecular Imaging

Bioimage Informatics. Lecture 23, Spring Emerging Applications: Molecular Imaging Bioimage Informatics Lecture 23, Spring 2012 Emerging Applications: Molecular Imaging Lecture 23 April 25, 2012 1 Outline Overview of molecular imaging Molecular imaging modalities Molecular imaging applications

More information

MINERVA MEDICA COPYRIGHT

MINERVA MEDICA COPYRIGHT Q J NUCL MED MOL IMAGING 2008;52:151-8 Image quality with non-standard nuclides in PET Non-standard positron emission tomography (PET) nuclides bring with them the prospect of new chemistry leading the

More information

Mayneord-Phillips Summer School St Edmund Hall, University of Oxford July Proton decays to n, e +, ν

Mayneord-Phillips Summer School St Edmund Hall, University of Oxford July Proton decays to n, e +, ν Positron Emission Tomography Physics & Instrumentation Dimitra G. Darambara, Ph.D Multimodality Molecular Imaging Joint Department of Physics RMH/ICR Outline Introduction PET Physics overview Types of

More information

Mitigation of External Radiation Exposures

Mitigation of External Radiation Exposures Mitigation of External Radiation Exposures The three (3) major principles to assist with maintaining doses ALARA are :- 1) Time Minimizing the time of exposure directly reduces radiation dose. 2) Distance

More information

Development of a High Precision Axial 3-D PET for Brain Imaging

Development of a High Precision Axial 3-D PET for Brain Imaging Development of a High Precision Axial 3-D PET for Brain Imaging On behalf of the AX-PET Collaboration SIENA - IPRD08 October 1st 4th, 2008 1 Outline Basics of Positron Emission Tomography (PET); Principle

More information

A liquid xenon PET camera - Simulation and position sensitive PMT tests

A liquid xenon PET camera - Simulation and position sensitive PMT tests A liquid xenon PET camera - Simulation and position sensitive PMT tests S. Jan, J. Collot, E. Tournefier To cite this version: S. Jan, J. Collot, E. Tournefier. A liquid xenon PET camera - Simulation and

More information

Gamma ray coincidence and angular correlation

Gamma ray coincidence and angular correlation University of Cape Town Department of Physics Course III laboratory Gamma ray coincidence and angular correlation Introduction Medical imaging based on positron emission tomography (PET) continues to have

More information

Nuclear Physics and Astrophysics

Nuclear Physics and Astrophysics Nuclear Physics and Astrophysics PHY-302 Dr. E. Rizvi Lecture 24 Medical Imaging Effects of Radiation We now know what radiation is But what does it mean for our bodies? Radioactivity is quantified in

More information

Design of a virtual model of a hand-held Germanium detector and a voxelized ICRP whole body phantom: A Monte Carlo study

Design of a virtual model of a hand-held Germanium detector and a voxelized ICRP whole body phantom: A Monte Carlo study Design of a virtual model of a hand-held Germanium detector and a voxelized ICRP whole body phantom: A Monte Carlo study ASM SABBIR AHMED 1, Gary H Kramer 2, Kurt Ungar 2 1 University of Saskatchewan,

More information

Physics in Nuclear Medicine

Physics in Nuclear Medicine SIMON R. CHERRY, PH.D. Professor Department of Biomedical Engineering University of California-Davis Davis, California JAMES A. SORENSON, PH.D. Emeritus Professor of Medical Physics University of Wisconsin-Madison

More information

Bases of radioisotope diagnostic methods

Bases of radioisotope diagnostic methods Medical, pharmaceutical applications of radioisotopes Bases of radioisotope diagnostic methods Dr. István Voszka Basis of application: radioisotopes have identical behavior in the organism to corresponding

More information

QUIZ: Physics of Nuclear Medicine Atomic Structure, Radioactive Decay, Interaction of Ionizing Radiation with Matter

QUIZ: Physics of Nuclear Medicine Atomic Structure, Radioactive Decay, Interaction of Ionizing Radiation with Matter QUIZ: Physics of Nuclear Medicine Atomic Structure, Radioactive Decay, Interaction of Ionizing Radiation with Matter 1. An atomic nucleus contains 39 protons and 50 neutrons. Its mass number (A) is a)

More information

A dual scintillator - dual silicon photodiode detector module for intraoperative gamma\beta probe and portable anti-compton spectrometer

A dual scintillator - dual silicon photodiode detector module for intraoperative gamma\beta probe and portable anti-compton spectrometer University of Wollongong Research Online Faculty of Engineering - Papers (Archive) Faculty of Engineering and Information Sciences 2008 A dual scintillator - dual silicon photodiode detector module for

More information

Lecture 5: Tomographic nuclear systems: SPECT

Lecture 5: Tomographic nuclear systems: SPECT Lecture 5: Tomographic nuclear systems: SPECT Field trip this saturday at 11 AM at UWMC meet in main hospital lobby at 11 AM if you miss the 'boat', page me at 540-4950 should take ~1 to 1.5 hours, depending

More information

Dana-Farber Cancer Institute, 44 Binney Street, Boston, MA 02115, USA ramsey

Dana-Farber Cancer Institute, 44 Binney Street, Boston, MA 02115, USA   ramsey SPECIAL FEATURE: MEDICAL PHYSICS www.iop.org/journals/physed Nuclear medicine Ramsey D Badawi Dana-Farber Cancer Institute, 44 Binney Street, Boston, MA 02115, USA E-mail: ramsey badawi@dfci.harvard.edu

More information

What is Albira and How Does the System Work and Why is it Differentiated Technology? Page 1

What is Albira and How Does the System Work and Why is it Differentiated Technology? Page 1 What is Albira and How Does the System Work and Why is it Differentiated Technology? Page 1 Why Albira? Tri-modality: PET, SPECT, CT Modular and state-of-the-art electronics 6 configurations Novel, Proprietary

More information

The 46g BGO bolometer

The 46g BGO bolometer Nature, 3 The g BGO bolometer 1 Photograph of the heat [g BGO] and light [Ge; =5 mm] bolometers: see Fig. 1c for description Current events: Amplification gains: 8, (heat channel) &, (light channel). The

More information

Dosimetry of patients injected with tracers Ga-68, Zr-89 and Lu-177. Bruno Vanderlinden

Dosimetry of patients injected with tracers Ga-68, Zr-89 and Lu-177. Bruno Vanderlinden Dosimetry of patients injected with tracers Ga-68, Zr-89 and Lu-177 Bruno Vanderlinden What is NM speciality? Imaging radiology Physics Diagnostic Treatment assessment Clinical pathology Biological marker

More information

anti-compton BGO detector

anti-compton BGO detector 1 2 3 Q β - measurements with a total absorption detector composed of through-hole HPGe detector and anti-compton BGO detector 4 5 Hiroaki Hayashi a,1, Michihiro Shibata b, Osamu Suematsu a, Yasuaki Kojima

More information

Travels with a Cyclotron. David Parker University of Birmingham

Travels with a Cyclotron. David Parker University of Birmingham Travels with a Cyclotron David Parker University of Birmingham Quick history Current uses of the cyclotron Transfer from Minneapolis 2 History of accelerators at Birmingham 60 Nuffield cyclotron (1948-1999)

More information

22.56J Noninvasive Imaging in Biology and Medicine Instructor: Prof. Alan Jasanoff Fall 2005, TTh 1-2:30

22.56J Noninvasive Imaging in Biology and Medicine Instructor: Prof. Alan Jasanoff Fall 2005, TTh 1-2:30 22.56J Noninvasive Imaging in Biology and Medicine Instructor: Prof. Alan Jasanoff Fall 2005, TTh 1-2:30 Sample problems HW1 1. Look up (e.g. in the CRC Manual of Chemistry and Physics www.hbcpnetbase.com)

More information

Nuclear Reactions A Z. Radioactivity, Spontaneous Decay: Nuclear Reaction, Induced Process: x + X Y + y + Q Q > 0. Exothermic Endothermic

Nuclear Reactions A Z. Radioactivity, Spontaneous Decay: Nuclear Reaction, Induced Process: x + X Y + y + Q Q > 0. Exothermic Endothermic Radioactivity, Spontaneous Decay: Nuclear Reactions A Z 4 P D+ He + Q A 4 Z 2 Q > 0 Nuclear Reaction, Induced Process: x + X Y + y + Q Q = ( m + m m m ) c 2 x X Y y Q > 0 Q < 0 Exothermic Endothermic 2

More information

A novel method to calibrate DOI function of a PET detector with a dual-ended-scintillator readout

A novel method to calibrate DOI function of a PET detector with a dual-ended-scintillator readout A novel method to calibrate DOI function of a PET detector with a dual-ended-scintillator readout Yiping Shao a Department of Imaging Physics, University of Texas M.D. Anderson Cancer Center, 55 Holcombe

More information

CHIPP Plenary Meeting University of Geneva, June 12, 2008 W. Lustermann on behalf of the AX PET Collaboration

CHIPP Plenary Meeting University of Geneva, June 12, 2008 W. Lustermann on behalf of the AX PET Collaboration CHIPP Plenary Meeting University of Geneva, June 12, 2008 W. Lustermann on behalf of the AX PET Collaboration INFN Bari, Ohio State University, CERN, University of Michigan, University of Oslo, INFN Roma,

More information

Positron-Electron Annihilation

Positron-Electron Annihilation Positron-Electron Annihilation Carl Akerlof September 13, 008 1. Introduction This experiment attempts to explore several features of positron-electron annihilation. One of the attractive aspects of e

More information

Recent advances and future perspectives of gamma imagers for scintimammography

Recent advances and future perspectives of gamma imagers for scintimammography 3rd International Conference on Imaging Technologies in Biomedical Sciences: ITBS2005 Innovation in Nuclear and Radiological Imaging: From Basic Research to Clinical Application Milos Conference Center,

More information

Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions

Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 2, APRIL 2012 469 Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions Weiyi Wang, Member,

More information

1. Motivation & Detector concept 2. Performance 3. Applications 4. Summary

1. Motivation & Detector concept 2. Performance 3. Applications 4. Summary A. Takada, T. Tanimori, H. Kubo, K. Miuchi, J. D. Parker, T. Mizumoto, Y. Mizumura, T. Sawano, Y. Matsuoka, S. Komura, S. Nakamura, M. Oda, S. Iwaki, K. Nakamura, S. Sonoda, D. Tomono (Kyoto Univ.) 1.

More information

Some nuclei are unstable Become stable by ejecting excess energy and often a particle in the process Types of radiation particle - particle

Some nuclei are unstable Become stable by ejecting excess energy and often a particle in the process Types of radiation particle - particle Radioactivity George Starkschall, Ph.D. Lecture Objectives Identify methods for making radioactive isotopes Recognize the various types of radioactive decay Interpret an energy level diagram for radioactive

More information

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN 308 Angular dependence of 662 kev multiple backscattered gamma photons in Aluminium Ravindraswami K a, Kiran K U b, Eshwarappa K M b and Somashekarappa H M c* a St Aloysius College (Autonomous), Mangalore

More information

Advances in PET technology

Advances in PET technology Advances in PET technology Jostein Sæterstøl September 15th 2009 Outline Introduction Detectors Time-of-flight PET PET / MRI Conclusions Introduction Fluor-18; FDG β + decay p n + e + + ν Annihilation

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? 2. Briefly discuss dead time in a detector. What factors are important

More information

CALCULATION OF SHIELDING AND RADIATION DOSES FOR PET/CT NUCLEAR MEDICINE FACILITY

CALCULATION OF SHIELDING AND RADIATION DOSES FOR PET/CT NUCLEAR MEDICINE FACILITY International Conference on Mathematics and Computational Methods Applied to Nuclear Science and Engineering (M&C 2011) Rio de Janeiro, RJ, Brazil, May 8-12, 2011, on CD-ROM, Latin American Section (LAS)

More information

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors Radiation Detectors 1 How do we detect ionizing radiation? Indirectly, by its effects as it traverses matter? What are these effects? Ionization and excitation of the atoms and molecules Heat 2 Types of

More information

Investigation of the Minimum Detectable Activity Level of a Preclinical LSO PET Scanner

Investigation of the Minimum Detectable Activity Level of a Preclinical LSO PET Scanner 27 IEEE Nuclear Science Symposium Conference Record M13-337 Investigation of the Minimum Detectable ctivity Level of a Preclinical LSO PET Scanner Nicolas Karakatsanis, Student Member, IEEE, Qinan Bao,

More information

There are three mechanisms by which gamma rays interact with absorber atoms from which two are important for nuclear medicine.

There are three mechanisms by which gamma rays interact with absorber atoms from which two are important for nuclear medicine. Measurement of radioactivity. Radioactive decay is a random process and therefore fluctuations are expected in the radioactivity measurement. That is why measurement of radioactivity must be treated by

More information

Nuclear Radiation. Natural Radioactivity. A person working with radioisotopes wears protective clothing and gloves and stands behind a shield.

Nuclear Radiation. Natural Radioactivity. A person working with radioisotopes wears protective clothing and gloves and stands behind a shield. Nuclear Radiation Natural Radioactivity A person working with radioisotopes wears protective clothing and gloves and stands behind a shield. 1 Radioactive Isotopes A radioactive isotope has an unstable

More information

Charge collection in PET detectors

Charge collection in PET detectors University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2007 Charge collection in PET detectors Tony Young University of

More information

Time-of-Flight PET using Cherenkov Photons Produced in PbF 2

Time-of-Flight PET using Cherenkov Photons Produced in PbF 2 Photons Produced in PbF 2 R. Dolenec a, S. Korpar b,a, P. Križan c,a, R. Pestotnik a, A. Stanovnik d,a a, Ljubljana, Slovenia b Faculty of Chemistry and Chemical Engineering, University of Maribor, Slovenia

More information

Differentiating Chemical Reactions from Nuclear Reactions

Differentiating Chemical Reactions from Nuclear Reactions Differentiating Chemical Reactions from Nuclear Reactions 1 CHEMICAL Occurs when bonds are broken or formed. Atoms remained unchanged, though may be rearranged. Involves valence electrons Small energy

More information

Design of a Lanthanum Bromide Detector for TOF PET

Design of a Lanthanum Bromide Detector for TOF PET Design of a Lanthanum Bromide Detector for TOF PET A. Kuhn, S. Surti, Member, IEEE, J. S. Karp, Senior Member, IEEE, P. S. Raby, K. S. Shah, A. E. Perkins, Member, IEEE, G. Muehllehner, Fellow Member,

More information

Positron Emission Tomography (PET)

Positron Emission Tomography (PET) Positron Emission Tomography (PET) A radiological technique for functional imaging Please note that this exercise takes place at the Stockholm Centre for Physics, Astronomy and Biotechniques (Alba Nova).

More information

Medical Neutron Science

Medical Neutron Science Medical Neutron Science 03 Neutron Activation Analysis The use of NAA techniques for medical applications was first reported in 1964 for measurement of sodium in the body J. Anderson, SB S.B. Ob Osborn,

More information

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques Radioisotopes in action Diagnostic application of radioisotopes Steps of diagnostic procedure - Radioactive material introduced into the patient - Distribution and alteration of activity is detected -Monitoring

More information

Year 12 Notes Radioactivity 1/5

Year 12 Notes Radioactivity 1/5 Year Notes Radioactivity /5 Radioactivity Stable and Unstable Nuclei Radioactivity is the spontaneous disintegration of certain nuclei, a random process in which particles and/or high-energy photons are

More information

Overview of Nuclear Medical Imaging Instrumentation and Techniques*

Overview of Nuclear Medical Imaging Instrumentation and Techniques* Overview of Nuclear Medical Imaging Instrumentation and Techniques* William W. Moses Lawrence Berkeley National Laboratory, University of California, Berkeley, CA 94720 USA Abstract. Nuclear medical imaging

More information

Positron Annihilation in Material Research

Positron Annihilation in Material Research Positron Annihilation in Material Research Introduction Positron sources, positron beams Interaction of positrons with matter Annihilation channels: Emission of 1, 2 or 3 γ-quanta Annihilation spectroscopies:

More information

Simultaneous PET/MRI imaging using nanoparticles

Simultaneous PET/MRI imaging using nanoparticles Simultaneous PET/MRI imaging using nanoparticles Ramesh Sharma 4, Yowen Xu 1, S. David Smith 1, Valerie Carroll 2, Dmitri Medvedev 1, Sri- Harsha Maramraju 3, David. Alexoff 1, Daniela Schulz 1, Paul Vaska

More information

Digital Gamma-ray Spectroscopy & Imaging with Semiconductor detectors Frontiers of gamma-ray spectroscopy

Digital Gamma-ray Spectroscopy & Imaging with Semiconductor detectors Frontiers of gamma-ray spectroscopy Digital Gamma-ray Spectroscopy & Imaging with Semiconductor detectors Frontiers of gamma-ray spectroscopy AGATA GRETA Dr Andy Boston ajboston@liv.ac.uk And its applications The AGATA Concept Without Compton

More information

A new timing model for calculating the intrinsic timing resolution of a scintillator detector

A new timing model for calculating the intrinsic timing resolution of a scintillator detector INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 5 (7) 3 7 PHYSICS IN MEDICINE AND BIOLOGY doi:.88/3-955/5/4/6 A new timing model for calculating the intrinsic timing resolution of a scintillator detector

More information

Energy resolution and absolute detection efficiency for LSO crystals: a comparison between Monte Carlo simulation and experimental data

Energy resolution and absolute detection efficiency for LSO crystals: a comparison between Monte Carlo simulation and experimental data Energy resolution and absolute detection efficiency for LSO crystals: a comparison between Monte Carlo simulation and experimental data Harold Rothfuss a,b, Larry Byars c, Michael E. Casey a, Maurizio

More information

Isotope Production for Nuclear Medicine

Isotope Production for Nuclear Medicine Isotope Production for Nuclear Medicine Eva Birnbaum Isotope Program Manager February 26 th, 2016 LA-UR-16-21119 Isotopes for Nuclear Medicine More than 20 million nuclear medicine procedures are performed

More information

Application of Nuclear Physics

Application of Nuclear Physics Application of Nuclear Physics Frontier of gamma-ray spectroscopy 0.1 IR visible light UV soft X-ray X-ray hard X-ray gamma-ray 1 10 100 1e3 1e4 1e5 1e6 energy [ev] Photoelectric effect e - Compton scattering

More information

Technical University of Denmark

Technical University of Denmark Technical University of Denmark Page 1 of 11 pages Written test, 9 December 2010 Course name: Introduction to medical imaging Course no. 31540 Aids allowed: none. "Weighting": All problems weight equally.

More information

2007 Fall Nuc Med Physics Lectures

2007 Fall Nuc Med Physics Lectures 2007 Fall Nuc Med Physics Lectures Tuesdays, 9:30am, NN203 Date Title Lecturer 9/4/07 Introduction to Nuclear Physics RS 9/11/07 Decay of radioactivity RS 9/18/07 Interactions with matter RM 9/25/07 Radiation

More information

Chapter 18: Radioactivity And Nuclear Transformation. Presented by Mingxiong Huang, Ph.D.,

Chapter 18: Radioactivity And Nuclear Transformation. Presented by Mingxiong Huang, Ph.D., Chapter 18: Radioactivity And Nuclear Transformation Presented by Mingxiong Huang, Ph.D., mxhuang@ucsd.edu 18.1 Radionuclide Decay Terms and Relationships Activity Decay Constant Physical Half-Life Fundamental

More information

Physics of Radioactive Decay. Purpose. Return to our patient

Physics of Radioactive Decay. Purpose. Return to our patient Physics of Radioactive Decay George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Purpose To demonstrate qualitatively the various processes by which unstable nuclides

More information

arxiv: v1 [physics.ins-det] 3 Feb 2011

arxiv: v1 [physics.ins-det] 3 Feb 2011 Nuclear Instruments and Methods in Physics Research A 00 (2018) 1 5 Alogo.pdf Nuclear Instruments and Methods in Physics Research A Scintillation decay time and pulse shape discrimination in oxygenated

More information

Copyright 2008, University of Chicago, Department of Physics. Gamma Cross-sections. NaI crystal (~2" dia) mounted on photo-multiplier tube

Copyright 2008, University of Chicago, Department of Physics. Gamma Cross-sections. NaI crystal (~2 dia) mounted on photo-multiplier tube Gamma Cross-sections 1. Goal We wish to measure absorption cross-sections for γ-rays for a range of gamma energies and absorber atomic number. 2. Equipment Pulse height analyzer Oscilloscope NaI crystal

More information

Figure 1. Decay Scheme for 60Co

Figure 1. Decay Scheme for 60Co Department of Physics The University of Hong Kong PHYS3851 Atomic and Nuclear Physics PHYS3851- Laboratory Manual A. AIMS 1. To learn the coincidence technique to study the gamma decay of 60 Co by using

More information

Figure1. Aerial view of the Thomas Jefferson National Accelerator Facility electron accelerator and experimental halls.

Figure1. Aerial view of the Thomas Jefferson National Accelerator Facility electron accelerator and experimental halls. Applying Nuclear Physics Research Detector Technology to Biomedical Applications Drew Weisenberger, Brian Kross, James Proffitt, Stan Majewski, John McKisson, Alexander Stolin and Carl Zorn Thomas Jefferson

More information

Neutron and Gamma Ray Imaging for Nuclear Materials Identification

Neutron and Gamma Ray Imaging for Nuclear Materials Identification Neutron and Gamma Ray Imaging for Nuclear Materials Identification James A. Mullens John Mihalczo Philip Bingham Oak Ridge National Laboratory Oak Ridge, Tennessee 37831-6010 865-574-5564 Abstract This

More information

Procesamiento de Imágenes y Bioseñales

Procesamiento de Imágenes y Bioseñales Procesamiento de Imágenes y Bioseñales Dr. Víctor Castañeda Agenda Physical basis of X-ray- CT, NMR, Ultrasound, Nuclear Medicine Sensors (cameras, gamma probes, microphone) Computational Tomography (CT)

More information

11/10/2014. Chapter 1: Introduction to Medical Imaging. Projection (Transmission) vs. Emission Imaging. Emission Imaging

11/10/2014. Chapter 1: Introduction to Medical Imaging. Projection (Transmission) vs. Emission Imaging. Emission Imaging Chapter 1: Introduction to Medical Imaging Overview of Modalities Properties of an Image: Limitations on Information Content Contrast (both object & image): Brightness difference Sharpness (blur): Smallest

More information

Quantitative Imaging with Y-90 Bremsstrahlung SPECT (Single Positron Emission Computed Tomography)

Quantitative Imaging with Y-90 Bremsstrahlung SPECT (Single Positron Emission Computed Tomography) Quantitative Imaging with Y-90 Bremsstrahlung SPECT (Single Positron Emission Computed Tomography) S. Beykan, M. Lassmann, S. Schlögl Klinik und Poliklinik für Nuklearmedizin Direktor: Prof. Dr. A. Buck

More information

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values.

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values. EXPERIMENT 2.1 GAMMA ENERGY CALIBRATION 1- Turn the power supply on to 900 V. Turn the NIM crate on to power the amplifiers. Turn the Oscilloscope on to check the gamma pulses. The main amplifier should

More information

THREE-QUANTA POSITRON ANNIHILATION IN BLOOD SAMPLES OF DIFFERENT OXYGENATION LEVELS

THREE-QUANTA POSITRON ANNIHILATION IN BLOOD SAMPLES OF DIFFERENT OXYGENATION LEVELS THREE-QUANTA POSITRON ANNIHILATION IN BLOOD SAMPLES OF DIFFERENT OXYGENATION LEVELS MARY CHIN & NICHOLAS SPYROU Department of Physics, University of Surrey, Guildford GU2 7XH, UK THREE-QUANTA POSITRON

More information

Synthesis and stability of [ 77 Br]-m-Bromobenzylguanidine ( 77 Br-MBBG)

Synthesis and stability of [ 77 Br]-m-Bromobenzylguanidine ( 77 Br-MBBG) Synthesis and stability of [ 77 Br]-m-Bromobenzylguanidine ( 77 Br-MBBG) Shigeki Watanabe 1 [ watanabe.shigeki@jaea.go.jp ] Noriko S. Ishioka 1, Ji Xin Liang 1, Hirofumi Hanaoka 2, Yasuhiko iida 2,3, Keigo

More information

Comparison of image quality of different iodine isotopes (I-123, I-124 and I-131)

Comparison of image quality of different iodine isotopes (I-123, I-124 and I-131) Comparison of image quality of different iodine isotopes (I-123, I-124 and I-131) Erwann Rault, Stefaan Vandenberghe, Roel Van Holen, Jan De Beenhouwer, Steven Staelens, and Ignace Lemahieu MEDISIP, ELIS

More information

Recent Advances of Planar Silicon APD Technology

Recent Advances of Planar Silicon APD Technology Recent Advances of Planar Silicon APD Technology M. McClish 1, R. Farrell 1, R. Myers 1, F. Olschner 2, G. Entine 1, K.S. Shah 1 1 Radiation Monitoring Devices Inc., Watertown, MA 2 Cremat Inc., Newton,

More information

Medical Physics. Nuclear Medicine Principles and Applications

Medical Physics. Nuclear Medicine Principles and Applications Medical Physics Nuclear Medicine Principles and Applications Dr Roger Fulton Department of PET & Nuclear Medicine Royal Prince Alfred Hospital Sydney Email: rfulton@mail.usyd.edu.au Lectures: http://www-personal.usyd.edu.au/~rfulton/medical_physics

More information

3. Which of the following statements is (are) TRUE about detector crystals in Anger cameras?

3. Which of the following statements is (are) TRUE about detector crystals in Anger cameras? BioE 1330 - Exam 2 11/13/2018 Answer Sheet - Correct answer is A for all questions 1. Unlike CT, in nuclear medicine A. Bremsstrahlung is not used to produce high-energy photons. B. signal can be increased

More information

Detector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples

Detector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples Aim of this tal Detector technology WMIC Educational Program Nuclear Imaging World Molecular Imaging Congress, Dublin, Ireland, Sep 5-8, 202 You can now the name of a bird in all the languages of the world,

More information

Structure of Biological Materials

Structure of Biological Materials ELEC ENG 3BA3: Structure of Biological Materials Notes for Lecture #19 Monday, November 22, 2010 6.5 Nuclear medicine imaging Nuclear imaging produces images of the distribution of radiopharmaceuticals

More information

Radiochemistry and Radiopharmacy III

Radiochemistry and Radiopharmacy III Radiochemistry and Radiopharmacy III Compact course held at UFSCAR, September 20123 Ulrich Abram Freie Universität Berlin Institute of Chemistry and Biochemistry Radiochemistry and Radiopharmacy 1. Fundamentals

More information

Objective Functions for Tomographic Reconstruction from. Randoms-Precorrected PET Scans. gram separately, this process doubles the storage space for

Objective Functions for Tomographic Reconstruction from. Randoms-Precorrected PET Scans. gram separately, this process doubles the storage space for Objective Functions for Tomographic Reconstruction from Randoms-Precorrected PET Scans Mehmet Yavuz and Jerey A. Fessler Dept. of EECS, University of Michigan Abstract In PET, usually the data are precorrected

More information