Mayneord-Phillips Summer School St Edmund Hall, University of Oxford July Proton decays to n, e +, ν

Size: px
Start display at page:

Download "Mayneord-Phillips Summer School St Edmund Hall, University of Oxford July Proton decays to n, e +, ν"

Transcription

1 Positron Emission Tomography Physics & Instrumentation Dimitra G. Darambara, Ph.D Multimodality Molecular Imaging Joint Department of Physics RMH/ICR Outline Introduction PET Physics overview Types of events Acquisition modes Detectors Performance parameters Data reconstruction and corrections Detector developments New development in PET Medical Imaging Techniques Anatomical CT MRI US Functional or metabolic SPECT PET (f)mri(s) (probe) optical (SAI) High resolution morphological capabilities with physiological info Real-time physiological info Biological processes at molecular level PET in Medical Imaging Powerful and sensitive means to non-invasively investigate biological processes in-vivo Radiotracer-imagingimaging technique Inject tracer compounds labelled with positron-emitting radionuclides Images of the static or dynamic distribution of tracer Info on blood flow, metabolism, proliferation, angiogenesis, hypoxia, gene expression In clinic: Diagnosis, treatment follow-up, therapy assessment Oncology,cardiology,neurology, psychiatry,treatment planning Drug development PET: unique molecular imaging modality 3 main factors: No need for heavy absorptive collimators higher sensitivity can detect and image nano- and pico- molar levels of tracers in-vivo Absolute quantification capability through accurate attenuation correction Availability of β -emitting radionuclides that are organic atoms can be substituted in molecules without modifying their biological activity PET radiotracers to investigate any biological process without interfering with normal biochemistry Positron Emission and Annihilation Proton decays to n, e, ν e e combines with e - and annihilates ν 511 kev γ e e - γ 511 kev 1

2 Annihilation Coincidence Detection A molecular probe labelled with e -emitting radionuclides annihilation -- simultaneous 2 x 511 kev gamma-rays Line of Response (LOR) -- electronic collimation Isotope Positron-emitting Radionuclides Half-life life Max E Range (MeV) (mm) Production C mins Cyclotron N mins Cyclotron Scanner: rings of position-sensitive photodetectors -- millions of coincident pairs detected -- coincidence time window (6-12 ns) -- coincidence logic electronics a time stamp to the record of detected events -- scan time: 5-60 mins sensitivity, acquisition mode, size of ROI, amount of injected activity Reconstructed images of radiotracer distribution e e - O secs Cyclotron F mins Cyclotron Ga mins Generator Rb secs Generator Cu mins Generator Cu hrs Cyclotron e -emitting radionuclide Positron Range --- Non-Colinearity Effective e range e range 511 kev e e- Non-colinearity Error due to noncolinearity 511 kev Positron Range Non-Colinearity 2 fundamental limits of spatial resolution: 2-2.5mm for clinical whole-body scanners Positron Range: --- its distributions have exponential shape with long tails not well described by Gaussian function rms effective range better indicator than FWHM --- radionuclide-specific, depends on positron emission E e 511 kev 511 kev Non-Colinearity: --- can be described as an approximate Gaussian angular distribution around 180 with FWHM~ linearly dependent on the distance D between coincident detectors: R = xD Detected Events in PET A detected event in PET is valid if: 2 photons are detected within a predefined electronic time window coincidence window Types of Coincidence Events The LOR between the 2 photons is within a valid acceptance angle of the scanner The E deposited in the detector by both photons is within the selected E window E-gating technique around the photopeak in the E spectrum need for detectors with good E resolution v narrow E gate Such coincidence events prompts TRUE RANDOM SCATTERED 2

3 Types of Coincidence Events Single: a single photon recorded by a detector True coincidence : detection of 2 singles from the same positron annihilation within the coincidence window True Count Rate: T = Aε 2 Ω coin Where A = activity concentration of the source ε = detector detection efficiency for 511 kev Ω coin = solid angle for detection of coincidence annihilation photons Types of Coincidence Events Random or accidental coincidence: detection of 2 singles from 2 different annihilations within the coincidence window Count Rate: R = 2τN x N z where 2τ the width of time window N single event rate incident upon detector x and z High Random count rate impact on noise, dead time and pulse pile-up limit sensitivity Types of Coincidence Events Scattered Events: one or both of the photons from the same annihilation have undergone a Compton interaction (body, detector, scanner gantry) Count Rate: S = Aε 2 Ω d where Ω d the solid angle for the FOV for coincidence scatter events Scatter and Random events undesirable source of background counts loss of contrast resolution and quantitative accuracy in final reconstructed image Can be reduced by narrowing E and coincidence window and limiting FOV activity Prompt Count Rate: Sum {TSR} Noise Equivalent Count Rate The performance of a PET scanner characterised by a figure of merit a trade-off between undesired contributions and scanner sensitivity NECR, defined as: T 2 NECR = T S kr measured as a function of activity meaningful way to compare performance of different scanners Factor k : 2 or 1 determined by whether Randoms measured in a delayed time window and subtracted or estimated from the single count rate 2D PET Acquisition Modes 3D PET Acquisition Modes 2D Mode: Interplane septa orthogonal to system axis prevent photons entering at oblique angles efficient rejection of photons scattered in the body reduce single-channel counting rate lower R rate min dead time losses Interplane septa Centre of FOV: sensitivity in 3D significantly greater 3D Mode 3D Mode: septa removed data for all possible LORs significant increase (4-8 fold) in photon sensitivity but considerable increase in S and R rates and system dead time sensitivity: a triangular function peaked at the centre of FOV 3

4 PET Detectors Properties of ideal PET detector: High stopping power for 511 kev photons High spatial resolution V good energy resolution to reject scatter events V high timing resolution Be inexpensive to produce Physical properties of commonly used scintillators Reproduced from D. Bailey, J.S. Karp, S. Surti, Positron Emission Tomography, Springer-Verlag, London, 2005, pp 31 Possible detectors: Proportional gas chambers Semiconductors Scintillators photo-detectors (PMTs, semiconductor-based photodiodes) Inorganic crystals emit visible light photons 4 main properties: stopping power of 511keV, signal decay time, light output, intrinsic E resolution Single Crystal mounted on individual PMT one-to-one coupling Scintillation Crystal PMT Front-end Electronics 511 kev γ-ray BGO or LSO 8x8 or 12x12 matrix of 6x6x30 mm 3 or 4x4x30mm 3 Block Detector A B γ photons convert to light photons, proportional to γ energy Light converted to electrical signal and amplified Crystal dimensions determine the spatial resolution 30mm depth, 10-30mm high (axial), 3-10mm width (in-plane) Registration and further processing of the signal y x 30 mm C y = x = 4 PMTs D (AB) (CD) ABCD (BD) (AC) ABCD Important Performance Parameters in PET Photon Sensitivity: Ability to detect coincident photons emitted from inside the FOV Determined by: scanner geometry and stopping efficiency of detectors for 511 kev Denser, higher Z, thicker (longer) detector elements to improve the 511 kev stopping power (BGO) Scanner geometry defines solid angle covering object Small diameter & large axial FOV high sensitivity scanner The higher the sensitivity the better the SNR in the reconstructed image High stopping power reduction of parallax error Important Performance Parameters in PET Spatial Resolution: Positron range effect its extent depends on the range of Es of the emitted positrons and the medium Photon Non-Colinearity worse for larger systems Size of the photon detector element pixel size (4-6 mm in typical clinical systems) Directly affects the spatial resolution in reconstructed image R sys = ( R det2 R range2 R nonc 2 ) 1/2 4

5 Important Performance Parameters in PET Energy and Coincidence Time Resolution: Improved by: Using crystals with brighter & faster light pulses Using low-noise photo-detectors Collecting a higher fraction of the light into the photodetector for larger electronic pulses Good E resolution narrow E window reduce S & R events without compromising sensitivity Good T resolution narrow time window reduce R events Typical values: 25% FWHM at 511 kev, 3ns FWHM Important Performance Parameters in PET Count Rate: For each signal registered a finite processing time If too many photons hit detector saturation of electronics due to pile-up of more than one pulses Pile-up depends on: scintillator decay time; effective integration time of electronics; photon event rate seen by the detector For best count rate performance: Crystals with fast decay time Detector with excellent T resolution Fast processing electronics Limited activity within the sensitive FOV Depth of Interaction (DOI) Effect A photon hits the detector travels a short distance within the detector material deposits its E in the detector the detectors used in PET do not measure this point the exact depth at which photons interact is unknown DOI within the crystal The measured position of E deposition projected to the entrance surface of the detector parallax error Parallax Error due to DOI Effect For photons at oblique angles parallax error produces significant deviations from real position blurring of reconstructed image degradation of spatial resolution DOI depends on: dimensions of the detector element, source location, scanner diameter, crystal length Annihilation photon path Assigned LOR DOI Effect A thin crystal with high stopping power reduces distance travelled by photons in the detector reduces parallax error But thin crystal reduces sensitivity An accurate measurement of DOI within crystal required Need for detectors with DOI measurement capabilities Phoswich detectors: stacking thin layers of different scintillators with different decay time on top of each other and implementing pulse shape discrimination Use photo-detectors at both ends of a thick (long) scintillator Depth sensitive detectors (e.g. CdZnTe) Ongoing research PET Data Reconstruction mathematical algorithms to calculate 3D probe distribution volume from the 2D projection data 2 basic reconstruction schemes: 1. Analytic Methods: acquisition process, measurements, reconstructed image as continuous functions (e.g. FBP) Directly compute an inverse transform formula to convert the recorded hits into an image Require spatial frequency filtering to reduce statistical noise resulting in a loss of spatial resolution Linear, more computational efficient, fast and simple to implement 5

6 PET Data Reconstruction 2. Iterative Methods: acquisition process, measurements, reconstructed image as discrete quantities start with an initial estimate of the 3D distribution and go through iterative modifications of that estimate until a solution is reached (e.g. MLEM) may incorporate statistical methods Appropriate for photon count limited data & PET systems with non-standard geometry Allow an improved trade-off between spatial resolution and More computationally intensive Corrections for PET Data There are several undesired physical effects inherent in the detection process of annihilation photons in PET measured PET data must be corrected for these physical factors either before or during image reconstruction Photon Attenuation within the tissue Most important correction For any given LOR, the attenuation depends on the total path travelled by the 2 annihilation photons (the total thickness of the body along that line) and is the same whether a point source (origin of the 2γ emission) is outside or inside the object Attenuation may be corrected for every LOR by measuring the total attenuation factor of an external radiation source that transmits activity through any LOR In clinical PET/CT use of X-ray CT the 511 kev attenuation coefficients determined from an appropriate scaling of attenuation coefficients measured at X-ray Es Detector Response Non-uniformity Variations of the coincidence detection efficiency between LORs affect the image uniformity in PET Variations result from imperfections related to physical, geometric, mechanical and electronic properties of individual detector elements This artefact is normalised by measuring the non-uniform response for every LOR with an external radiation source and applying this normalisation to every measured data set for correcting them Detector Dead Time or Saturation Dead time: the finite time required to process and record an event while no other events can be recorded Also includes pulse pile-up loss of counts Causes loss in spatial and contrast resolutions Saturation when the incoming photon flux is higher than the system processing bandwidth allows Dead time losses significant in high count rates with continuous or block detectors An analytic model of the dead time/saturation to calculate the correction factors that applied to every LOR Random Coincidence Events Cause loss in quantitative accuracy and contrast resolution R effects worse for high count rates Estimates of the R coincidence rate for every LOR are obtained from measurements or calculations and are subtracted or estimated from the single count rate Single count rate as a function of time is required Common technique is to measure the count rate in a delayed time window where true coincidences are impossible Then real-time subtraction for each LOR implemented in hardware 6

7 Scatter Coincidence Events Cause degradation of quantitative accuracy and contrast resolution Worse for larger objects, higher γ rates, poor E resolution Using a narrow E window rejection of large angle S events but small angle S events still present Small angle S events are calculated for each LOR and subtracted Most promising scatter compensation use simplified Monte Carlo simulations to estimate the scatter distribution Time consuming but already practical implementation in commercial PET scanners Isotope Decay Compensates for changes in tracer activity over time Knowledge of the half-time of the isotope as well as the time record of when each data acquired is required Partial Volume Effect Causes spatial resolution blurring Due to the finite spatial resolution of the system and the inherent sampling of discrete pixel image representation reduces intensity for structures that are on the order of the system resolution or smaller The activity concentrations of such structures either overestimated or under-estimated depending on the regional distribution of the radioactivity Correction factor by measuring the intensity reduction effects vs structure size in a phantom with known activity concentration in spheres of various known diameters Detector Developments New scintillators with fast decay time, high light output, high stopping power PS-PMT and MC-PMT Si PIN Diodes APDs SiPMs CdZnTe HpGe Si microstrip Avalanche Photo Diodes (APD) In a single package or arrays Compact size Position sensitive Large excess noise Internal signal amplification, so improved SNR High quantum efficiency Need for low-noise, fast frontend electronics Gain sensitive to small temperature variations and changes in applied bias voltage MR compatible Reproduced from T.K. Lewellen, Phys Med Biol 53 (2008) R287-R317 Silicon Photomultipliers (SiPM) Geiger-mode APD High intrinsic gain Low operating voltage Excellent timing resolution Single photoelectron sensitivity Excellent SNR Robustness V low excess noise MR compatible Low cost, wide range of pixel sizes Dead space between cells reduces overall QE Lower detection efficiency than APD at short wavelengths High dark current (cooling) 7

8 CdZnTe Direct conversion --- true semiconductor High density and effective Z high absorption efficiency High intrinsic spatial resolution obtained by the electrode pixellation on the front and in the depth rather than cutting crystals High energy resolution (better scatter & random rejection, multiple-isotope imaging, energyresolved CT) CdZnTe Operate in room temperature High resistivity due to wide band gap lowleakage current low-noise characteristics Small system FOV versatile, flexible, light weight and configurable detectors Compatible with MRI: operate normally with MRI acquisitions up to 7 and 9.4 T Multimodal detector ({MRI/}SPECT/PET{/CT}) Time of Flight (TOF) PET In TOF PET for each annihilation event the difference in arrival times between the 2 coincident photons is also measured Images have higher SNR than images without TOF info Scintillators with v fast timing decay, high light output and high stopping power (LSO, LaBr 3 ) V fast electronics Potential to improve the image quality in heavy patients (more attenuation and scatter) CT Non-TOF MLEM TOF MLEM Patient 1: colon cancer, 119Kg, BMI=46.5 Patient 2: abdominal cancer, 115Kg, BMI=38 Reproduced from J.S. Karp et al, J Nucl Med 2008; 49: Multimodality Imaging PET/CT PET/MR TOF PET New Developments Specific applications dedicated systems breast, brain, prostate V high resolution pre-clinical imaging systems small animal imaging 8

Radionuclide Imaging MII Positron Emission Tomography (PET)

Radionuclide Imaging MII Positron Emission Tomography (PET) Radionuclide Imaging MII 3073 Positron Emission Tomography (PET) Positron (β + ) emission Positron is an electron with positive charge. Positron-emitting radionuclides are most commonly produced in cyclotron

More information

PET. Technical aspects

PET. Technical aspects PET Technical aspects 15 N 15 O Detector 1 β+ Detector 2 e- Evolution of PET Detectors CTI/Siemens 15 N 15 O Detector block 1 β+ Detector block 2 x e- x y y location line of response Constant fraction

More information

1st Faculty of Medicine, Charles University in Prague Center for Advanced Preclinical Imaging (CAPI)

1st Faculty of Medicine, Charles University in Prague Center for Advanced Preclinical Imaging (CAPI) Radioation Resolution and Sensitivity Nuclear Imaging PET + SPECT Radioactive Decay (EC,Ɣ), (β -,Ɣ), (I.T.,Ɣ) β + Projection imaging collimator needed one angular view Projection imaging coincidence imaging,

More information

Positron Emission Tomography

Positron Emission Tomography Positron Emission Tomography Presenter: Difei Wang June,2018 Universität Bonn Contents 2 / 24 1 2 3 4 Positron emission Detected events Detectors and configuration Data acquisition Positron emission Positron

More information

Time-of-Flight PET using Cherenkov Photons Produced in PbF 2

Time-of-Flight PET using Cherenkov Photons Produced in PbF 2 Photons Produced in PbF 2 R. Dolenec a, S. Korpar b,a, P. Križan c,a, R. Pestotnik a, A. Stanovnik d,a a, Ljubljana, Slovenia b Faculty of Chemistry and Chemical Engineering, University of Maribor, Slovenia

More information

Detector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples

Detector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples Aim of this tal Detector technology WMIC Educational Program Nuclear Imaging World Molecular Imaging Congress, Dublin, Ireland, Sep 5-8, 202 You can now the name of a bird in all the languages of the world,

More information

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: NUCLEAR MEDICINE Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Introduction to Medical Imaging: Physics, Engineering and Clinical Applications, by Nadine Barrie Smith

More information

Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017

Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017 Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017 Marcus Ressner, PhD, Medical Radiation Physicist, Linköping University Hospital Content What is Nuclear medicine? Basic principles of Functional

More information

Current Challenges and Opportunities in Positron Emission Tomography. Paul Marsden King s College London

Current Challenges and Opportunities in Positron Emission Tomography. Paul Marsden King s College London Current Challenges and Opportunities in Positron Emission Tomography Paul Marsden King s College London STFC Particle Physics Department Rutherford Appleton Laboratory - 3 Sept 2014 Current Challenges

More information

6: Positron Emission Tomography

6: Positron Emission Tomography 6: Positron Emission Tomography. What is the principle of PET imaging? Positron annihilation Electronic collimation coincidence detection. What is really measured by the PET camera? True, scatter and random

More information

Development of a High Precision Axial 3-D PET for Brain Imaging

Development of a High Precision Axial 3-D PET for Brain Imaging Development of a High Precision Axial 3-D PET for Brain Imaging On behalf of the AX-PET Collaboration SIENA - IPRD08 October 1st 4th, 2008 1 Outline Basics of Positron Emission Tomography (PET); Principle

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information

Advances in PET technology

Advances in PET technology Advances in PET technology Jostein Sæterstøl September 15th 2009 Outline Introduction Detectors Time-of-flight PET PET / MRI Conclusions Introduction Fluor-18; FDG β + decay p n + e + + ν Annihilation

More information

Prospects for achieving < 100 ps FWHM coincidence resolving time in time-of-flight PET

Prospects for achieving < 100 ps FWHM coincidence resolving time in time-of-flight PET Prospects for achieving < 100 ps FWHM coincidence resolving time in time-of-flight PET, 28-Feb-2012, ICTR-PHE, Geneva, Switzerland 1 Time-of-flight PET Colon cancer, left upper quadrant peritoneal node

More information

CHIPP Plenary Meeting University of Geneva, June 12, 2008 W. Lustermann on behalf of the AX PET Collaboration

CHIPP Plenary Meeting University of Geneva, June 12, 2008 W. Lustermann on behalf of the AX PET Collaboration CHIPP Plenary Meeting University of Geneva, June 12, 2008 W. Lustermann on behalf of the AX PET Collaboration INFN Bari, Ohio State University, CERN, University of Michigan, University of Oslo, INFN Roma,

More information

A. I, II, and III B. I C. I and II D. II and III E. I and III

A. I, II, and III B. I C. I and II D. II and III E. I and III BioE 1330 - Review Chapters 7, 8, and 9 (Nuclear Medicine) 9/27/2018 Instructions: On the Answer Sheet, enter your 2-digit ID number (with a leading 0 if needed) in the boxes of the ID section. Fill in

More information

Timing and Energy Response of Six Prototype Scintillators

Timing and Energy Response of Six Prototype Scintillators Timing and Energy Response of Six Prototype Scintillators CCM Kyba 1, J Glodo 2, EVD van Loef 2, JS Karp 1, KS Shah 2 1 University of Pennsylvania 2 Radiation Monitoring Devices SCINT 2007 June 7, 2007

More information

Chapter 2 PET Imaging Basics

Chapter 2 PET Imaging Basics Chapter 2 PET Imaging Basics Timothy G. Turkington PET Radiotracers Positron emission tomography (PET) imaging is the injection (or inhalation) of a substance containing a positron emitter, the subsequent

More information

EEE4106Z Radiation Interactions & Detection

EEE4106Z Radiation Interactions & Detection EEE4106Z Radiation Interactions & Detection 2. Radiation Detection Dr. Steve Peterson 5.14 RW James Department of Physics University of Cape Town steve.peterson@uct.ac.za May 06, 2015 EEE4106Z :: Radiation

More information

www.aask24.com www.aask24.com www.aask24.com P=Positron E= Emission T=Tomography Positron emission or beta plus decay (+ ) is a particular type of radioactive decay, in which a proton inside a radionuclide

More information

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors Radiation Detectors 1 How do we detect ionizing radiation? Indirectly, by its effects as it traverses matter? What are these effects? Ionization and excitation of the atoms and molecules Heat 2 Types of

More information

Sub-Nanosecond Timing for In-Beam PET in hadrontherapy

Sub-Nanosecond Timing for In-Beam PET in hadrontherapy Sub-Nanosecond Timing for In-Beam PET in hadrontherapy Baptiste Joly Clermont Université, Université Blaise Pascal, CNRS/IN2P3, Laboratoire de Physique Corpusculaire, BP 10448, F-63000 CLERMONT-FERRAND,

More information

PoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo

PoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo ICEPP, the University of Tokyo E-mail: sawada@icepp.s.u-tokyo.ac.jp The MEG experiment yielded the most stringent upper limit on the branching ratio of the flavorviolating muon decay µ + e + γ. A major

More information

The Physics of PET/CT scanners

The Physics of PET/CT scanners The Physics of PET/CT scanners Ruth E. Schmitz, Adam M. Alessio, and Paul E. Kinahan Imaging Research Laboratory Department of Radiology University of Washington What Makes PET Useful? Positron emission

More information

Compton Camera. Compton Camera

Compton Camera. Compton Camera Diagnostic Imaging II Student Project Compton Camera Ting-Tung Chang Introduction The Compton camera operates by exploiting the Compton Effect. It uses the kinematics of Compton scattering to contract

More information

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques Radioisotopes in action Diagnostic application of radioisotopes Steps of diagnostic procedure - Radioactive material introduced into the patient - Distribution and alteration of activity is detected -

More information

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide Nuclear Medicine Imaging Systems: The Scintillation Camera List of Nuclear Medicine Radionuclides Tc99m 140.5 kev 6.03 hours I-131 364, 637 kev 8.06 days I-123 159 kev 13.0 hours I-125 35 kev 60.2 days

More information

3. Which of the following statements is (are) TRUE about detector crystals in Anger cameras?

3. Which of the following statements is (are) TRUE about detector crystals in Anger cameras? BioE 1330 - Exam 2 11/13/2018 Answer Sheet - Correct answer is A for all questions 1. Unlike CT, in nuclear medicine A. Bremsstrahlung is not used to produce high-energy photons. B. signal can be increased

More information

A dual scintillator - dual silicon photodiode detector module for intraoperative gamma\beta probe and portable anti-compton spectrometer

A dual scintillator - dual silicon photodiode detector module for intraoperative gamma\beta probe and portable anti-compton spectrometer University of Wollongong Research Online Faculty of Engineering - Papers (Archive) Faculty of Engineering and Information Sciences 2008 A dual scintillator - dual silicon photodiode detector module for

More information

Gamma ray coincidence and angular correlation

Gamma ray coincidence and angular correlation University of Cape Town Department of Physics Course III laboratory Gamma ray coincidence and angular correlation Introduction Medical imaging based on positron emission tomography (PET) continues to have

More information

CT-PET calibration : physical principles and operating procedures F.Bonutti. Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital.

CT-PET calibration : physical principles and operating procedures F.Bonutti. Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital. CT-PET calibration : physical principles and operating procedures Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital Topics Introduction to PET physics F-18 production β + decay and annichilation

More information

Study of the feasibility of a compact gamma camera for real-time cancer assessment

Study of the feasibility of a compact gamma camera for real-time cancer assessment Study of the feasibility of a compact gamma camera for real-time cancer assessment L. Caballero Instituto de Física Corpuscular - CSIC - University of Valencia; C/Catedrático José Beltrán, 2; E-46980;

More information

Positron Emission Tomography (PET)

Positron Emission Tomography (PET) Positron Emission Tomography (PET) A radiological technique for functional imaging Please note that this exercise takes place at the Stockholm Centre for Physics, Astronomy and Biotechniques (Alba Nova).

More information

arxiv: v1 [physics.ins-det] 29 Jun 2011

arxiv: v1 [physics.ins-det] 29 Jun 2011 Performance simulation of a MRPC-based PET Imaging System arxiv:1106.5877v1 [physics.ins-det] 29 Jun 2011 A. Banerjee, S. Chattopadhyay April 16, 2018 Abstract The low cost and high resolution gas-based

More information

Tomography is imaging by sections. 1

Tomography is imaging by sections. 1 Tomography is imaging by sections. 1 It is a technique used in clinical medicine and biomedical research to create images that show how certain tissues are performing their physiological functions. 1 Conversely,

More information

Design of a Lanthanum Bromide Detector for TOF PET

Design of a Lanthanum Bromide Detector for TOF PET Design of a Lanthanum Bromide Detector for TOF PET A. Kuhn, S. Surti, Member, IEEE, J. S. Karp, Senior Member, IEEE, P. S. Raby, K. S. Shah, A. E. Perkins, Member, IEEE, G. Muehllehner, Fellow Member,

More information

Chapter 4 Scintillation Detectors

Chapter 4 Scintillation Detectors Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 4-1 4.1. Basic principle of the scintillator Chapter 4 Scintillation Detectors Scintillator Light sensor Ionizing radiation Light (visible,

More information

Radionuclide Imaging MII Detection of Nuclear Emission

Radionuclide Imaging MII Detection of Nuclear Emission Radionuclide Imaging MII 3073 Detection of Nuclear Emission Nuclear radiation detectors Detectors that are commonly used in nuclear medicine: 1. Gas-filled detectors 2. Scintillation detectors 3. Semiconductor

More information

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam (closed book, 1 sheets of notes double sided allowed, no calculator or other electronic devices allowed) 1. Ultrasound Physics (15 pt) A) (9

More information

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS TSOKOS OPTION I-2 MEDICAL IMAGING Reading Activity Answers IB Assessment Statements Option I-2, Medical Imaging: X-Rays I.2.1. I.2.2. I.2.3. Define

More information

Technical University of Denmark

Technical University of Denmark Technical University of Denmark Page 1 of 11 pages Written test, 9 December 2010 Course name: Introduction to medical imaging Course no. 31540 Aids allowed: none. "Weighting": All problems weight equally.

More information

DETECTORS. I. Charged Particle Detectors

DETECTORS. I. Charged Particle Detectors DETECTORS I. Charged Particle Detectors A. Scintillators B. Gas Detectors 1. Ionization Chambers 2. Proportional Counters 3. Avalanche detectors 4. Geiger-Muller counters 5. Spark detectors C. Solid State

More information

Outline Chapter 14 Nuclear Medicine

Outline Chapter 14 Nuclear Medicine Outline Chapter 14 uclear Medicine Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther Introduction Detectors for nuclear

More information

Introduction to Radiation Monitoring

Introduction to Radiation Monitoring Introduction to Radiation Monitoring Iain Darby Honorary Research Fellow, University of Glasgow iain.darby@glasgow.ac.uk https://at.linkedin.com/in/idarby https://www.facebook.com/iain.darby.662 Outline

More information

Application of Nuclear Physics

Application of Nuclear Physics Application of Nuclear Physics Frontier of gamma-ray spectroscopy 0.1 IR visible light UV soft X-ray X-ray hard X-ray gamma-ray 1 10 100 1e3 1e4 1e5 1e6 energy [ev] Photoelectric effect e - Compton scattering

More information

Charge collection in PET detectors

Charge collection in PET detectors University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2007 Charge collection in PET detectors Tony Young University of

More information

Scintillation Detectors

Scintillation Detectors Scintillation Detectors J.L. Tain Jose.Luis.Tain@ific.uv.es http://ific.uv.es/gamma/ Instituto de Física Corpuscular C.S.I.C - Univ. Valencia Scintillation detector: SCINTILLATION MATERIAL LIGHT-GUIDE

More information

Physics in Nuclear Medicine

Physics in Nuclear Medicine SIMON R. CHERRY, PH.D. Professor Department of Biomedical Engineering University of California-Davis Davis, California JAMES A. SORENSON, PH.D. Emeritus Professor of Medical Physics University of Wisconsin-Madison

More information

Sample Spectroscopy System Hardware

Sample Spectroscopy System Hardware Semiconductor Detectors vs. Scintillator+PMT Detectors Semiconductors are emerging technology - Scint.PMT systems relatively unchanged in 50 years. NaI(Tl) excellent for single-photon, new scintillation

More information

Nuclear Medicine Intro & Physics from Medical Imaging Signals and Systems, Chapter 7, by Prince and Links

Nuclear Medicine Intro & Physics from Medical Imaging Signals and Systems, Chapter 7, by Prince and Links Nuclear Medicine Intro & Physics from Medical Imaging Signals and Systems, Chapter 7, by Prince and Links NM - introduction Relies on EMISSION of photons from body (versus transmission of photons through

More information

DEVELOPMENT OF A NEW SILICON BASED DETECTOR MODULE FOR PET

DEVELOPMENT OF A NEW SILICON BASED DETECTOR MODULE FOR PET DEVELOPMENT OF A NEW SILICON BASED DETECTOR MODULE FOR PET A thesis submitted in fulfilment of the requirements for the award of the degree Master of Science (Honours) from UNIVERSITY OF WOLLONGONG By

More information

Structure of Biological Materials

Structure of Biological Materials ELEC ENG 3BA3: Structure of Biological Materials Notes for Lecture #19 Monday, November 22, 2010 6.5 Nuclear medicine imaging Nuclear imaging produces images of the distribution of radiopharmaceuticals

More information

Bioimage Informatics. Lecture 23, Spring Emerging Applications: Molecular Imaging

Bioimage Informatics. Lecture 23, Spring Emerging Applications: Molecular Imaging Bioimage Informatics Lecture 23, Spring 2012 Emerging Applications: Molecular Imaging Lecture 23 April 25, 2012 1 Outline Overview of molecular imaging Molecular imaging modalities Molecular imaging applications

More information

Compton Camera with PositionSensitive Silicon Detectors

Compton Camera with PositionSensitive Silicon Detectors University of Ljubljana Faculty of mathematics and physics Andrej Studen Compton Camera with PositionSensitive Silicon Detectors Doctoral thesis Supervisor: Professor Marko Mikuž Outline: Motivation Basic

More information

Recent advances and future perspectives of gamma imagers for scintimammography

Recent advances and future perspectives of gamma imagers for scintimammography 3rd International Conference on Imaging Technologies in Biomedical Sciences: ITBS2005 Innovation in Nuclear and Radiological Imaging: From Basic Research to Clinical Application Milos Conference Center,

More information

Comparison of image quality of different iodine isotopes (I-123, I-124 and I-131)

Comparison of image quality of different iodine isotopes (I-123, I-124 and I-131) Comparison of image quality of different iodine isotopes (I-123, I-124 and I-131) Erwann Rault, Stefaan Vandenberghe, Roel Van Holen, Jan De Beenhouwer, Steven Staelens, and Ignace Lemahieu MEDISIP, ELIS

More information

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research Prompt gamma measurements for the verification of dose deposition in proton therapy Two Proton Beam Facilities for Therapy and Research Ion Beam Facilities in Korea 1. Proton therapy facility at National

More information

What is scintigraphy? The process of obtaining an image or series of sequential images of the distribution of a radionuclide in tissues, organs, or

What is scintigraphy? The process of obtaining an image or series of sequential images of the distribution of a radionuclide in tissues, organs, or Let's remind... What is nuclear medicine? Nuclear medicine can be broadly divided into two branches "in vitro" and "in vivo" procedures. There are numerous radioisotopic "in vitro" procedures for genotyping

More information

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques Radioisotopes in action Diagnostic application of radioisotopes Steps of diagnostic procedure - Radioactive material introduced into the patient - Distribution and alteration of activity is detected -Monitoring

More information

Single Photon detectors

Single Photon detectors Single Photon detectors Outline Motivation for single photon detection Semiconductor; general knowledge and important background Photon detectors: internal and external photoeffect Properties of semiconductor

More information

PET scan simulation. Meysam Dadgar. UMSU, Iran. IFMP, Elbasan, Fig 1: PET camera simulation in gate by cylindrical phantom

PET scan simulation. Meysam Dadgar. UMSU, Iran. IFMP, Elbasan, Fig 1: PET camera simulation in gate by cylindrical phantom PET scan simulation Meysam Dadgar UMSU, Iran IFMP, Elbasan, 2016 Meysamdadgar10@gmail.com 1 Fig 1: PET camera simulation in gate by cylindrical phantom 2 What is PET? Positron emission tomography (PET),

More information

Positron Annihilation in Material Research

Positron Annihilation in Material Research Positron Annihilation in Material Research Introduction Positron sources, positron beams Interaction of positrons with matter Annihilation channels: Emission of 1, 2 or 3 γ-quanta Annihilation spectroscopies:

More information

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber CYRIC Annual Report 2001 V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber Kawata N. Baba M. Kato M.*, Miura T.**, and Yamadera A.***, Cyclotron and Radioisotope Center, Tohoku

More information

Positron-Electron Annihilation

Positron-Electron Annihilation Positron-Electron Annihilation Carl Akerlof September 13, 008 1. Introduction This experiment attempts to explore several features of positron-electron annihilation. One of the attractive aspects of e

More information

Radioisotopes and PET

Radioisotopes and PET Radioisotopes and PET 1 Radioisotopes Elements are defined by their number of protons, but there is some variation in the number of neutrons. Atoms resulting from this variation are called isotopes. Consider

More information

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors 5. Radiation Microsensors Radiation µ-sensors convert incident radiant signals into standard electrical out put signals. Radiant Signals Classification

More information

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? 2. Briefly discuss dead time in a detector. What factors are important

More information

Radiation Dose, Biology & Risk

Radiation Dose, Biology & Risk ENGG 167 MEDICAL IMAGING Lecture 2: Sept. 27 Radiation Dosimetry & Risk References: The Essential Physics of Medical Imaging, Bushberg et al, 2 nd ed. Radiation Detection and Measurement, Knoll, 2 nd Ed.

More information

Nuclear Instruments and Methods in Physics Research A

Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A () Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima The

More information

Readout of LYSO using a new silicon photodetector for positron emission tomography

Readout of LYSO using a new silicon photodetector for positron emission tomography University of Wollongong Research Online Faculty of Engineering - Papers (Archive) Faculty of Engineering and Information Sciences 23 Readout of LYSO using a new silicon photodetector for positron emission

More information

Dual Isotope Imaging with LaBr3:Ce Crystal and H8500 PSPMT

Dual Isotope Imaging with LaBr3:Ce Crystal and H8500 PSPMT Dual Isotope Imaging with LaBr3:Ce Crystal and H8500 PSPMT Dr. Andrea Fabbri, University of Rome Roma Tre I.N.F.N. (National Institue of Nuclear Physics) γ-ray imaging with scintillator and PSPMT γ-ray

More information

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY ABSORBED DOSE: The amount of energy absorbed, as a result of radiation passing through a material, per unit mass of material. Measured in rads (1 rad

More information

Nuclear Medicine RADIOPHARMACEUTICAL CHEMISTRY

Nuclear Medicine RADIOPHARMACEUTICAL CHEMISTRY Nuclear Medicine RADIOPHARMACEUTICAL CHEMISTRY An alpha particle consists of two protons and two neutrons Common alpha-particle emitters Radon-222 gas in the environment Uranium-234 and -238) in the environment

More information

A Measurement of Monoenergetic Neutrons from 9 Be(p,n) 9 B

A Measurement of Monoenergetic Neutrons from 9 Be(p,n) 9 B Journal of the Korean Physical Society, Vol. 32, No. 4, April 1998, pp. 462 467 A Measurement of Monoenergetic Neutrons from 9 Be(p,n) 9 B J. H. Kim, H. Bhang, J. H. Ha, J. C. Kim, M. J. Kim, Y. D. Kim

More information

11/10/2014. Chapter 1: Introduction to Medical Imaging. Projection (Transmission) vs. Emission Imaging. Emission Imaging

11/10/2014. Chapter 1: Introduction to Medical Imaging. Projection (Transmission) vs. Emission Imaging. Emission Imaging Chapter 1: Introduction to Medical Imaging Overview of Modalities Properties of an Image: Limitations on Information Content Contrast (both object & image): Brightness difference Sharpness (blur): Smallest

More information

Overview of Nuclear Medical Imaging Instrumentation and Techniques*

Overview of Nuclear Medical Imaging Instrumentation and Techniques* Overview of Nuclear Medical Imaging Instrumentation and Techniques* William W. Moses Lawrence Berkeley National Laboratory, University of California, Berkeley, CA 94720 USA Abstract. Nuclear medical imaging

More information

What is Albira and How Does the System Work and Why is it Differentiated Technology? Page 1

What is Albira and How Does the System Work and Why is it Differentiated Technology? Page 1 What is Albira and How Does the System Work and Why is it Differentiated Technology? Page 1 Why Albira? Tri-modality: PET, SPECT, CT Modular and state-of-the-art electronics 6 configurations Novel, Proprietary

More information

Technical University of Denmark

Technical University of Denmark Technical University of Denmark Page 1 of 10 pages Written test, 12 December 2012 Course name: Introduction to medical imaging Course no. 31540 Aids allowed: None. Pocket calculator not allowed "Weighting":

More information

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN 308 Angular dependence of 662 kev multiple backscattered gamma photons in Aluminium Ravindraswami K a, Kiran K U b, Eshwarappa K M b and Somashekarappa H M c* a St Aloysius College (Autonomous), Mangalore

More information

Optimization of the parameters in the electronics of a PET detector.

Optimization of the parameters in the electronics of a PET detector. Optimization of the parameters in the electronics of a PET detector. (Optimización de los parámetros de la electrónica de un detector para PET) Ana María Barragán Montero July 4, 2013 Academic project

More information

Development of a new PET detector module with improved Depth of Interaction and Time of Flight capabilities

Development of a new PET detector module with improved Depth of Interaction and Time of Flight capabilities Development of a new PET detector module with improved Depth of Interaction and Time of Flight capabilities Ana Rita Borrego Instituto Superior Técnico IST, Universidade de Lisboa UL, Lisbon, Portugal

More information

Characterisation of Silicon Photomultipliers for the T2K Experiment

Characterisation of Silicon Photomultipliers for the T2K Experiment Characterisation of Silicon Photomultipliers for the T2K Experiment, 18th May 2010 Martin Haigh, University of Oxford Outline Brief introduction to the T2K experiment. Overall configuration and goals.

More information

Simulation of triple coincidences in PET

Simulation of triple coincidences in PET 136 Institute of Physics and Engineering in Medicine Physics in Medicine & Biology doi:10.1088/0031-9155/60/1/117 Simulation of triple coincidences in PET J Cal-González 1,4, E Lage 2, E Herranz 1, E Vicente

More information

arxiv: v1 [physics.ins-det] 3 Feb 2011

arxiv: v1 [physics.ins-det] 3 Feb 2011 Nuclear Instruments and Methods in Physics Research A 00 (2018) 1 5 Alogo.pdf Nuclear Instruments and Methods in Physics Research A Scintillation decay time and pulse shape discrimination in oxygenated

More information

Radioactivity. Lecture 6 Detectors and Instrumentation

Radioactivity. Lecture 6 Detectors and Instrumentation Radioactivity Lecture 6 Detectors and Instrumentation The human organs Neither humans nor animals have an organ for detecting radiation from radioactive decay! We can not hear it, smell it, feel it or

More information

Lecture 5: Tomographic nuclear systems: SPECT

Lecture 5: Tomographic nuclear systems: SPECT Lecture 5: Tomographic nuclear systems: SPECT Field trip this saturday at 11 AM at UWMC meet in main hospital lobby at 11 AM if you miss the 'boat', page me at 540-4950 should take ~1 to 1.5 hours, depending

More information

Amorphous selenium (a-se) avalanche photodetector for applications in Positron Emission Tomography (PET)

Amorphous selenium (a-se) avalanche photodetector for applications in Positron Emission Tomography (PET) Amorphous selenium (a-se) avalanche photodetector for applications in Positron Emission Tomography (PET) By OLEKSANDR BUBON Master of Science in Physics Lakehead University Thunder Bay, Ontario 2011 Amorphous

More information

Slides by: Prof. Abeer Alharbi

Slides by: Prof. Abeer Alharbi Slides by: Prof. Abeer Alharbi electromagnetic radiation of high energy. They are produced by sub-atomic particle interactions, such as electron-positron annihilation, neutral pion decay, radioactive decay,

More information

Towards Proton Computed Tomography

Towards Proton Computed Tomography SCIPP Towards Proton Computed Tomography L. R. Johnson, B. Keeney, G. Ross, H. F.-W. Sadrozinski, A. Seiden, D.C. Williams, L. Zhang Santa Cruz Institute for Particle Physics, UC Santa Cruz, CA 95064 V.

More information

Silicon Drift Detectors for gamma-ray detection: 15 years of research (and collaboration between Politecnico and INAF-Milano)

Silicon Drift Detectors for gamma-ray detection: 15 years of research (and collaboration between Politecnico and INAF-Milano) Silicon Drift Detectors for gamma-ray detection: 15 years of research (and collaboration between Politecnico and INAF-Milano) Outline The Silicon Drift Detector (SDD) Gamma-ray detectors based on scintillators

More information

Properties of the nucleus. 8.2 Nuclear Physics. Isotopes. Stable Nuclei. Size of the nucleus. Size of the nucleus

Properties of the nucleus. 8.2 Nuclear Physics. Isotopes. Stable Nuclei. Size of the nucleus. Size of the nucleus Properties of the nucleus 8. Nuclear Physics Properties of nuclei Binding Energy Radioactive decay Natural radioactivity Consists of protons and neutrons Z = no. of protons (Atomic number) N = no. of neutrons

More information

Scintillators 1. YEAR, I. CYCLE. AUTHOR: BPhys. Manja Ščetinec ADVISOR/MENTOR: Prof. Boštjan Golob

Scintillators 1. YEAR, I. CYCLE. AUTHOR: BPhys. Manja Ščetinec ADVISOR/MENTOR: Prof. Boštjan Golob Scintillators SEMINAR 1. YEAR, I. CYCLE AUTHOR: BPhys. Manja Ščetinec ADVISOR/MENTOR: Prof. Boštjan Golob Faculty of Mathematics and Physics, University in Ljubljana Ljubljana, October 2017 Abstract In

More information

Using new digital SiPM from Philips with AX-PET a new geometrical concept for PET

Using new digital SiPM from Philips with AX-PET a new geometrical concept for PET Using new digital SiPM from Philips with AX-PET a new geometrical concept for PET Matthieu Heller CERN - PH/DT Marie Curie network MC-PAD Matthieu.heller@cern.ch On behalf of the AX-PET collaboration https://twiki.cern.ch/twiki/bin/view/axialpet

More information

There are three mechanisms by which gamma rays interact with absorber atoms from which two are important for nuclear medicine.

There are three mechanisms by which gamma rays interact with absorber atoms from which two are important for nuclear medicine. Measurement of radioactivity. Radioactive decay is a random process and therefore fluctuations are expected in the radioactivity measurement. That is why measurement of radioactivity must be treated by

More information

Alpha-Gamma discrimination by Pulse Shape in LaBr 3 :Ce and LaCl 3 :Ce

Alpha-Gamma discrimination by Pulse Shape in LaBr 3 :Ce and LaCl 3 :Ce Alpha-Gamma discrimination by Pulse Shape in LaBr 3 :Ce and LaCl 3 :Ce F.C.L. Crespi 1,2, F.Camera 1,2, N. Blasi 2, A.Bracco 1,2, S. Brambilla 2, B. Million 2, R. Nicolini 1,2, L.Pellegri 1, S. Riboldi

More information

David B. Cassidy. Department of Physics and Astronomy, University of California, Riverside, USA. Varenna, July 09

David B. Cassidy. Department of Physics and Astronomy, University of California, Riverside, USA. Varenna, July 09 Experimental production of many- positron systems: L2, techniques David B. Cassidy Department of Physics and Astronomy, University of California, Riverside, USA cassidy@physics.ucr.edu Varenna, July 09

More information

Chicago PET Development and

Chicago PET Development and Chicago PET Development and Recent Progress in Digital i TOFPET 1. Heejong Kim, Chien-Min Kao, Qingguo Xie, Yun Dong, Ming-Chi Shih, Antonio Machado, and Chin-Tu Chen 2. Octavia Biris, Jialin Lin, Fukun

More information

Modular Survey Spectrometer and Compton Imager

Modular Survey Spectrometer and Compton Imager 1 Modular Survey Spectrometer and Compton Imager Audrey MacLeod Ionizing Radiation Standards Measurement Science and Standards National Research Council Canada On behalf of the SCoTSS collaboration (Silicon

More information

PHOTODETECTORS AND SILICON PHOTO MULTIPLIER

PHOTODETECTORS AND SILICON PHOTO MULTIPLIER ESE seminar Photodetectors - Sipm, P. Jarron - F. Powolny 1 PHOTODETECTORS AND SILICON PHOTO MULTIPLIER ESE seminar Pierre Jarron, Francois Powolny OUTLINE 2 Brief history and overview of photodetectors

More information

MEASUREMENT AND DETECTION OF RADIATION

MEASUREMENT AND DETECTION OF RADIATION MEASUREMENT AND DETECTION OF RADIATION Second Edition Nicholas Tsoulfanidis University of Missouri-Rolla Ж Taylor &Francis * Publishers since I79H CONTENTS Preface to the First Edition Preface to the Second

More information