Theoretical analysis of comparative patient skin dose and exposure technique approaches in planar radiography

Size: px
Start display at page:

Download "Theoretical analysis of comparative patient skin dose and exposure technique approaches in planar radiography"

Transcription

1 Exposure technique Australian Institute of Radiography The Radiographer 2009; 56 (1): Theoretical analysis of comparative patient skin dose exposure technique approaches in planar radiography Faculty of Science, Charles Sturt University, Wagga Wagga, New South Wales 2678, Australia. Correspondence Abstract A theoretical analysis of patient skin dose changes with manipulation of technique factors (mas kv) was carried out in order to obtain a description of the relationship between patient dose the exposure technique approach employed. The stard exposure technique approaches in variable part thickness radiography of fixed kvp variable kvp the 15% rule in constant part thickness radiography were analysed in terms of the changes in patient skin dose. Analytical expressions describing the variation of dose change ratios have been developed in respect of each approach also in comparison between the two stard approaches. The resultant model is limited to essentially qualitative predictions due to its inherent limitation of neglecting secondary radiation at the entrance site of the beam, but nevertheless indicates some non intuitive benefits in utilisation of mas (only) changes in compensating for the thinner part for lower image densities, as well as indicating that most gain is to be had (in terms of lower skin dose), where only modest part thickness compensation is required. The model also describes the influence of the chosen voltage in a reference technique (upon which further compensatory changes in exposure factors are based) on the patient skin dose, when the variable kv approach is utilised. Keywords: exposure techniques, patient dose, planar radiography Introduction Elementary textbooks (for example 1,2,3 ) in radiography when discussing exposure technique in two-dimensional planar or common projection modes, whether screen-film or computed radiography versions commonly cite the 15% exposure rule as applicable to a constant part thickness. They also discuss stard approaches of variable kv fixed kv in order to compensate for changes in part thickness. In addition, it is generally pointed out in such texts that less absorption of x-ray energy occurs at higher kilovoltages (kv) this implies less patient dose occurs by using a higher kv rather than using a higher tube current (mas) where an exposure increase is required. Most radiography texts or works, old or new, do not actually explore to any significant degree the relationships between patient dose technique settings. One notable exception being that of van der Plaats 4 who examines the relationship between patient skin dose, tube loading, focus size in the context of higher kilovoltages exposure techniques. Automatic exposure control in planar radiography is being more widely used. As manual techniques in radiography have become less of a topic, naturally fundamental work in the area of radiographic exposure technique has also lessened. While much experimental work in computed radiography digital radiography has been published in regards to image quality patient dose, there still appears to be essentially no cohesive nor general theoretical/ analytical model that explicitly describes the relationship between the chosen employed exposure technique, patient dose. The basis of exposure technique as developed initially for screen film receptors hence consequently the ramifications in terms of patient dose have direct relevance to the use also of digital receptors, as indicated by Fauber. 3 This current study investigates this relationship. The aim of analysis was to develop a simple analytical model of dose change in terms of the operational quantities involved, that of tube voltage, tube charge, patient part thickness increase. The analysis used restricts itself to purely the consideration of patient skin absorbed dose, does not attempt to address the other consideration of image quality. Image quality is a consideration that is important in any actual imaging practice. However, if any understing of the balance between image quality patient dose is to be finally achieved, then it is necessary that each of these aspects be separately understood in terms of the relationships at play. Analysis The main points of the analysis are discussed in this section, however further mathematical details of the analysis are included in the appendices. Assumptions relevant quantities of consideration This analysis ignores, as is generally traditional to do so, the use of distance, focus to film distance (FFD) or source to image distance (SID), as a technique factor employed to manipulate the exposure value upon the image receptor. For simplicity, the patient skin absorbed dose, D, proximal to the x-ray tube is considered here as the measure of patient dose. The image density signal, S, as measured by a scale such as the optical density, or its equivalent on a digital receptor (e.g. the voltage output upon digital readout); is considered here (in this simple modelling exercise) in terms of its average value across the image receptor. The technique exposure quantities involved (that are under operator control) are the tube charge, q (or mas ), the tube voltage, V (or kv ). The part thickness, x, is also an

2 22 The Radiographer important quantity ( outside operator control). Last, there is an involvement of the beam area, A (as measured at some convenient location, such as the level of the image receptor or the patient entrance beam area), however for the most part the case of small area values will be considered, restricting analysis for simplicity to consideration of patient skin dose essentially to that component due to the primary beam. Some consideration is then later given to modification of the analytical model to embrace other than small beam areas evaluation of the patient skin dose considering backscatter radiation contribution. In a given case (whether constant part thickness technique or variable part thickness technique is being considered) these quantities exist as either variables or constant values. Throughout, a simple assumption will be made in the case of absorbed dose. The general relationship relating exposure, X, to absorbed dose, D, is given by the classic equation: The f factor, f, is actually a function of the photon energy, or more generally in the case of a multi-energetic x-ray beam, it will be a function of anything affecting the beam spectrum. In particular, for a given x-ray tube, this means that the f factor will strictly be a function of the tube voltage, V, used. In addition, as has already been shown 5, the image density can be usefully described for a given typical part thickness by the model: which simplifies to case of constant FFD, L, constant part thickness x: (3)...an expression which has been either implicitly or explicitly related by several authors. 4,6 (The exponent of 5 above agrees perfectly with the exposure rules 5 though some works cite the exponent of 4. In fact the exponent varies truly with part thickness at least, as indicated by van der Plaats 4 ) On the other h, the actual exposure value at some distance d from the focus is usefully given by: this reduces for the case of constant distance d to: (note each k in the above three expressions is a constant value) As can be seen from the following analysis, it is this difference in the power of the voltage (2 versus 5) describing the (skin) exposure the photographic effect respectively, that leads to notable differences in skin absorbed dose, according as to whether kv or mas is chosen to be varied. Now, in the case of soft tissue across the diagnostic energy range, there is some very slight decrease in the f factor, but this small change can be neglected for the purpose of the elementary treatment here. For the purposes that follow it is assumed that the f factor is essentially constant in value. Thus: where c is a constant for all systems (patients apparatus) (1) (2) (4) (5) (6) considered. For ease of presentation, further analysis is broken down under two main items of consideration, (a) constant part thickness technique (b) variable part thickness techniques. (a) Constant part thickness radiography Here, consider changing the average image density signal, S, by changing both q V together. This means that all of S, q, V are variables, whilst the part thickness, x, is a constant value. The so called 15% rule of radiography is generally related as applicable in cases where the part thickness is constant the image density is held constant. However, it is easily shown that this rule is mathematically equivalent to the statement:...where k is a constant whose value is determined by the system, that is the part radiographed, image receptor x-ray apparatus used. This realisation allows cases constant part thickness to be considered as radiography of the same part, but where now the image density is altered, utilising either q (mas) or V (kv) only. In doing so (see Appendix A) then in consideration of the assumptions relations already stated, the ratio change in skin absorbed dose ( new divided by old value) when using only a change in mas to change the average image density signal by a multiplicative factor of F, is given by: Likewise a similar skin dose change ratio where only kv is used to effect an increase in image density by a factor of F is given by: (9) These are further usefully compared by the further ratio this gives: (7) (8) (10) Equation (10) expresses effectively the ratio of the resultant skin absorbed dose to obtain an image density signal altered by a factor F using a change only in tube kv, to the skin absorbed dose value that results from using only mas to effect the same image density signal change. (b) Variable part thickness radiography. The stard approaches used here are embraced in the fixed kv variable kv approach rules. The fixed kv rule, typically expressed as a 25% increase in mas per cm, is easily shown to be equivalent to the mathematical statement: (11) where x r is some reference thickness where known settings of V (kv) q (mas) produce an acceptable average image density signal, m is the number of cm increments in part thickness. The variable kv rule is typically expressed as add 2 kv per cm increase in part thickness. This rule can be easily shown as equivalent to the expression: where V is measured in units of kv. (12)

3 Theoretical analysis of comparative patient skin dose exposure technique approaches in planar radiography The Radiographer 23 Both approaches confer that the average image density signal is held constant, which is the aim of these approaches. This is despite changes in part thickness, for example as in a thickness increase of m cm from some reference thickness x = x r to a thickness x = x r + m. Thus the part thickness x is a variable in both approaches. The only variable will in fact be either q(mas) in the fixed kv approach, or V(kV) in the variable kv approach. Again, conveniently for simplicity at this point tube to image distance beam area effects are ignored. Analogously to the previous case of constant part thickness, dose change ratios can be constructed (see Appendix B) related to the increase in part thickness, m. This gives dose change factors, old skin absorbed dose divided by new skin absorbed dose, of: (13) (14) for the cases of variable kv fixed kv approaches respectively. If t v is divided by t q then this gives the ratio of the skin absorbed dose as results from utilising increase in V (kv) only, using the variable kv approach, to the skin absorbed dose as results from changing only tube charge q (mas), using the fixed kv approach. This ratio, denoted T is shown (Appendix B) to be given by: (15) Discussion Equations (10) (15) above constitute a model of dose variation with exposure technique selection. In order to interpret the predictions of the first of these, it is convenient to consider graphical representations as shown in Figure 1 Figure 2 below: Figure 1 shows the variation of R with increase image density change factor F. This means that F naturally starts at the value of unity, signifying no alteration in image density, if just increases in image density values is consider. Since R <1 for all F> 1, then this means that there is always a lower skin absorbed dose utilising voltage increase to effect the increase in image density, than would result from using an increase in tube charge (mas) to effect the same increase in image density. The benefit, purely in terms of skin absorbed dose, is quantified by the value of R, which shows that since R is less for larger F, then the greater the increase required in average image density, the greater the benefit in terms of reduced skin dose in using tube voltage increases rather than tube charge increases. The rate of change of benefit, however clearly is greatest where the slope of the plot is descending most rapidly, in fact the rate of change of R with F is easily shown to be given by: (16) Using this gives that the value of R falls with F at a value of about 0.6 (at F = 1) the steepness of this fall decreases to half of this, 0.3, at about F = 1.5. Thus the rate of change of benefit is greatest for small changes in the image density increase factor, but decreases significantly for larger increases in F. Indeed between F = 5 F = 6 there is virtually no change in the value of R (~ 1% change only). This means that the greatest gains in terms of reduction of patient skin dose using tube voltage (kv) rather than tube charge (mas), will be had over quite small increases in image density. Nevertheless, there is always some benefit to be had in lesser comparative skin dose as a result of using kv increases in preference to mas increases. If values of F are considered where 0 <F <1, corresponding to a required decrease in the image density, then Figure 2 indicates the results. In this case R> 1 increases as the value of F is decreased further towards zero. Since R> 1, then there is now a relative increase in skin dose in the case of decreasing tube voltage, in comparison to the skin dose that would ensue using decreases in tube charge to obtain the required decreased image density. Of course the skin dose will be lessened if either V or q is decreased, but it will be a lesser value in the case of changing the tube charge, q(mas). Considering the cases of variable part thickness approaches, again pertinent results can be more easily seen by consulting a graph of T with thickness increase, m. [Note that the increment m assumes both negative positive values. The interpretation of this is that a negative increment is a decrement in thickness. Thus a value of m = -2 signifies a decrease in part thickness (from some reference technique thickness) by 2 cm. Also the plot shows the particular case whereby the reference technique voltage (at m = 0) is 50 kv]. Because of the definition of T, then if T <1, this means there is an advantage in terms of comparative skin dose in utilising a Figure 1. Figure 2.

4 24 The Radiographer Figure 3. Figure 4. change in voltage only through using the variable kv approach rule rather than utilising the fixed kv approach whereby only the mas is used to compensate for change in part thickness. Figure 3 indicates generally that there is always a skin dose benefit to be had in utilising the variable kv approach when compensation is required for the larger anatomical part or patient. Conversely, for m <0, there is benefit in choosing the fixed kv approach to compensate for the thinner part, lower skin doses will result in this case. The variation is now much less though than the former case (variation of R with F), thus the rate of change of the benefit here (dt/dm), whilst decreasing with increasing m, is not dramatic. T is a function not only of the change in thickness (m), but also the reference thickness voltage,, used (this corresponds to the value of V at m = 0). In fact simple consideration of the expression for T (Appendix C) shows that where a larger voltage is used for the stard thickness technique, then the values of T are correspondingly less, for positive increments in thickness. To illustrate this, Figure 4 shows T curves for reference voltages of 50 kv also 100 kv. Figure 4 also shows that the skin dose benefit to be had in using the fixed kv approach, in compensating for the thinner patient, is larger when larger tube voltages are used in the reference technique. To further illustrate the change in T with various choice of then it is possible to plot T against for some particular m value/s, as in Figure 5 for example. Figure 5 shows the variation of T with. In the case of compensating for a decrement in thickness, then decreasing the mas only, via the fixed kv rule, will always be of benefit (T> 1). However, more benefit is to be had where the reference technique voltage is larger, as shown by the gentle increase of the curve for m = -2 cm, from about 1.26 to 1.46 over the voltage interval from 40 to 120 kv. When the part to be imaged is thicker than the stard technique, then the solid curve of Figure 5 shows that an increased benefit occurs in terms of skin dose, where higher values of reference voltage are used within technique. In the case of thicker parts, the change of benefit, as evaluated using the ratio T, is though about half of that for the case of thinner parts, since there is now an average fall of about 0.1 in T over the same voltage interval. Limitations of the analysis: secondary radiation The simple model so far has not considered the effect of secondary, essentially scatter, radiation. This will alter the values of both R T as evaluated in the model here, though naturally, where beam area is smaller, then the simple model developed here is more applicable. The complication occurs in skin-absorbed dose by virtue of the Figure 5. necessity to consider the backscatter factor, B, that describes the variation of scatter radiation at the skin surface at beam entry. Since in general, B can be regarded as a function of tube voltage V, at least for a given system, then it should possible to incorporate this into a more general model using the same general approach. The inherent benefit in using changes of voltage, to accommodate for the larger part or to generate higher image density values, is essentially caused by the difference in the V 2 the V 5 terms within the analysis. V 2 V 5 refer respectively to the variation of absorbed dose the photographic effect (image density) with tube voltage V (kv). When the backscatter factor is considered, then the term for the absorbed dose will alter effectively increase the index or power above the value of 2. However, cursory consideration of the energy dependency of B (B values as listed in ICRU Report 74) 7 indicates this is unlikely to push the absorbed dose voltage dependency term much higher than the power of 2. This means there will still be a significant difference between any exponent chosen for the absorbed dose term the image density term. Thus the qualitative changes, if not the numerical values of R T, should be essentially unchanged. Consequently, the qualitative changes inferred by this model should also apply to the more general case where scatter radiation is significant. Conclusion A simple theoretical model has been constructed that describes the variation in proximal skin absorbed dose with technique factor changes, in both variable part radiography constant part thickness radiography. Qualitatively, accepting the limitations of the model, the model predicts importantly that patient dose benefits are gained by utilisation of increases in voltage where either larger image density

5 Theoretical analysis of comparative patient skin dose exposure technique approaches in planar radiography The Radiographer 25 values are sought, with a given radiographic part, also where compensation is required over a larger thickness. These predictions agree with the common sense intuitive notions of course using the basic reasoning of higher kv therefore higher transmission, less absorption. However, less intuitively, the predictions imply that major benefits are gained in compensating for modest part thickness increases, also that, where compensation is sought for the thinner part, then benefit is gained by utilising changes in tube charge (mas) rather than using the tube voltage. Improvement of this model by incorporation of secondary radiation effects should enable more general applicability more accurate predictions. This might be expected to reasonably extend the capabilities of the model from essentially purely qualitative predictions, into one that has a reasonable quantitative predictability, that can further be compared effectively with experimental or other empirically obtained data. The construction of such a model would be useful in further building a complete logistic approach to effecting the ALARA (As Low As Reasonably Achievable) dose principle of modern radiological protection by further consideration of the independent requirement of a necessary level of image quality. References 1 Fossbinder RA, Kelsey CA, Essentials Of Radiologic Science. New York: McGraw Hill; Bushong SC, Radiologic Science for Technologists: Physics, Biology, Protection. St Louis: Mosby; Fauber TL, Radiographic Imaging Exposure, 3 rd ed., St Louis, Mosby, van der Plaats GJ, Medical X ray techniques in diagnostic radiology. London & Basingstoke: Macmillan Press; pp & pp Garbett IT, Derivation of Exposure Rules In Conventional Radiographic Imaging, The Radiographer 1999; (46) 3: Johns HE, Cunningham JR. The physics of radiology, 4th ed. Springfield Illinois: Charles C Thomas; pp Journal of the ICRU Appendix A: Backscatter Factors, Table A1. Oxford University Press, 2005: 5 No 2 Report 74, pp 66. Appendix A Using the earlier related equations, then we consider the skin dose in two cases old new ( denoted here by subscripts o n ), so as to consider the change effected. Writing: Then for the case where q is used to change S (so V n = V o ): is similarly obtained as: Suppose an increase factor in image density, given by: Then, in general (A2) Substituting for the cases of changing q (mas) also changing V (kv) only leads to: For the case of changing only q (to effect the image density change): (A3) for the instance of increasing V only (to effect the change in image density): (A4) This allows the dose change ratio to be expressed in terms of the image density change factor in each case: (A5) (A6) The ratio R can now be constructed in order to compare these two change factors, where: This leads to the general result that: (A7) Then the ratio of change (from old to new) in skin absorbed dose, (denote as r q ) is: (A1)...a result that may be intuitively expected since the skin dose is the directly proportional to the value of mas used. For the case where the image density signal is changed using only tube voltage change then the equivalent dose change ratio r V Appendix B Again (as in Appendix A) we consider the skin dose in two cases old new (as denoted here by subscripts o n ), so as to consider the change effected. Using: Now generally: Then for the fixed kv approach (V n = ): X n = k 4 q n 2

6 26 The Radiographer Between old new cases we suppose a change in thickness by a positive amount m cm. Then in the fixed kv approach using equation (A7) we have: In the variable case we have: (B8) (B9) in the variable kv case a similar ratio change, denoted as t V : Defining now a new ratio T: (B13) Then for the skin absorbed dose in the fixed kv case: (B10) This gives the general result: (B14) For the variable kv case: (B11) The old skin absorbed dose values are c 3 q 0 c 4 2 for fixed variable cases respectively (these are necessarily the same value because in fact c 3 = c 2 2 c 4 = c 2 q 0 ) Then there is a skin dose change ratio (denote this as t q ) in the fixed kv case of: (B 12) Appendix C The voltage term in equation (B13) can be expressed as: (C15) thus as then this term reduces approaches unity. The smaller the value of, then the larger is this ratio (denoted as t V above). Thus for any chosen value of m, then the value of T will be smaller in the case of a larger choice for this reference technique voltage,.

X-ray Interaction with Matter

X-ray Interaction with Matter X-ray Interaction with Matter 10-526-197 Rhodes Module 2 Interaction with Matter kv & mas Peak kilovoltage (kvp) controls Quality, or penetrating power, Limited effects on quantity or number of photons

More information

Rad T 290 Worksheet 2

Rad T 290 Worksheet 2 Class: Date: Rad T 290 Worksheet 2 1. Projectile electrons travel from a. anode to cathode. c. target to patient. b. cathode to anode. d. inner shell to outer shell. 2. At the target, the projectile electrons

More information

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT Proceedings of the Eleventh EGS4 Users' Meeting in Japan, KEK Proceedings 2003-15, p.107-113 ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT K. Noto and

More information

For the next several lectures, we will be looking at specific photon interactions with matter. In today s lecture, we begin with the photoelectric

For the next several lectures, we will be looking at specific photon interactions with matter. In today s lecture, we begin with the photoelectric For the next several lectures, we will be looking at specific photon interactions with matter. In today s lecture, we begin with the photoelectric effect. 1 The objectives of today s lecture are to identify

More information

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9 Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation

More information

Health Physics Services Ltd

Health Physics Services Ltd 1 Title: Adept Medical STARTable: Scatter Radiation Shielding Grid Format Methodology 2 Background: Adept Medical STARTable Shield is embedded with 0.5mm Lead (Pb), offering protection from scatter radiation

More information

We have seen how the Brems and Characteristic interactions work when electrons are accelerated by kilovolts and the electrons impact on the target

We have seen how the Brems and Characteristic interactions work when electrons are accelerated by kilovolts and the electrons impact on the target We have seen how the Brems and Characteristic interactions work when electrons are accelerated by kilovolts and the electrons impact on the target focal spot. This discussion will center over how x-ray

More information

LECTURE 4 PRINCIPLE OF IMAGE FORMATION KAMARUL AMIN BIN ABDULLAH

LECTURE 4 PRINCIPLE OF IMAGE FORMATION KAMARUL AMIN BIN ABDULLAH LECTURE 4 PRINCIPLE OF IMAGE FORMATION KAMARUL AMIN BIN ABDULLAH Lesson Objectives At the end of the lesson, student should able to: Define attenuation Explain interactions between x-rays and matter in

More information

Electron density and effective atomic number images generated by dual energy imaging with a 320-detector CT system: A feasibility study

Electron density and effective atomic number images generated by dual energy imaging with a 320-detector CT system: A feasibility study Electron density and effective atomic number images generated by dual energy imaging with a 320-detector CT system: A feasibility study Poster No.: C-0403 Congress: ECR 2014 Type: Scientific Exhibit Authors:

More information

Chapter 30 X Rays GOALS. When you have mastered the material in this chapter, you will be able to:

Chapter 30 X Rays GOALS. When you have mastered the material in this chapter, you will be able to: Chapter 30 X Rays GOALS When you have mastered the material in this chapter, you will be able to: Definitions Define each of the following terms, and use it in an operational definition: hard and soft

More information

Basic physics Questions

Basic physics Questions Chapter1 Basic physics Questions S. Ilyas 1. Which of the following statements regarding protons are correct? a. They have a negative charge b. They are equal to the number of electrons in a non-ionized

More information

Radhakrishnan B*, Kurup P G G**, Ramakrishnan G***, Chandralingam S****

Radhakrishnan B*, Kurup P G G**, Ramakrishnan G***, Chandralingam S**** Photon Interaction Cross Section of Materials in Heterogeneous Energy Spectrum of Medical Diagnostic X-Ray Beam Radhakrishnan B*, Kurup P G G**, Ramakrishnan G***, Chandralingam S**** *Senior Medical Physicist,

More information

1Meherun Nahar, Sazzad, 3Abdus Sattar Mollah and 4Mir Mohammad Akramuzzaman Correspondence Address INTRODUCTION MATERIALS AND METHODS Mathematical

1Meherun Nahar, Sazzad, 3Abdus Sattar Mollah and 4Mir Mohammad Akramuzzaman Correspondence Address INTRODUCTION MATERIALS AND METHODS Mathematical ORIGINAL ARTICLE Development of MGDA Software for Calculation of Patient Specific Mean Glandular Dose during Mammography 1Meherun Nahar, 2 M Sazzad, 3 Abdus Sattar Mollah and 4 Mir Mohammad Akramuzzaman

More information

City University of Hong Kong

City University of Hong Kong City University of Hong Kong Information on a Course offered by the Department of Physics and Materials Science with effect from Semester A in 2013 / 2014 Part I Course Title: Radiological Physics and

More information

DICOM Correction Item

DICOM Correction Item DICOM Correction Item Correction Number CP-687 Log Summary: of Modification Addition Name of Standard DICOM PS 3.16 2007 Rationale for Correction: When the Diagnostic X-Ray Radiation Dose Reporting SOP

More information

H e a l t h P h y s i c s S e r v i c e s L t d

H e a l t h P h y s i c s S e r v i c e s L t d 1 Title: Adept Medical X-Ray Shield: Scatter Radiation Shielding Grid Format Methodology 2 Background: Adept Medical X-Ray Shield (X-Ray Shield) is embedded with 0.5mm Lead (pb), offering protection from

More information

CHAPTER 2 RADIATION INTERACTIONS WITH MATTER HDR 112 RADIATION BIOLOGY AND RADIATION PROTECTION MR KAMARUL AMIN BIN ABDULLAH

CHAPTER 2 RADIATION INTERACTIONS WITH MATTER HDR 112 RADIATION BIOLOGY AND RADIATION PROTECTION MR KAMARUL AMIN BIN ABDULLAH HDR 112 RADIATION BIOLOGY AND RADIATION PROTECTION CHAPTER 2 RADIATION INTERACTIONS WITH MATTER PREPARED BY: MR KAMARUL AMIN BIN ABDULLAH SCHOOL OF MEDICAL IMAGING FACULTY OF HEALTH SCIENCE Interactions

More information

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe Radiation Dosimetry Alun Beddoe Medical Physics University Hospital Birmingham NHS Trust Contents ABSOLUTE DOSIMETRY (CALIBRATION) Photon interactions (recap) Energy transfer and absorption Electron range

More information

Study of Monte Carlo Simulator for Estimation of Anti-Scatter Grid Physical Characteristics on IEC 60627:2013-Based

Study of Monte Carlo Simulator for Estimation of Anti-Scatter Grid Physical Characteristics on IEC 60627:2013-Based American Journal of Physics and Applications 2018; 6(2): 35-42 http://www.sciencepublishinggroup.com/j/ajpa doi: 10.11648/j.ajpa.20180602.12 ISSN: 2330-4286 (Print); ISSN: 2330-4308 (Online) Study of Monte

More information

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements Specific Accreditation Criteria Calibration ISO/IEC 17025 Annex Ionising radiation measurements January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected

More information

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS M.A. HASSAN*, M.H. GABER**, E. ESMAT*, H.I. FARAG***, H.M. EISSA* *National Institute for Standards (NIS), Giza, Egypt **Biophysics Department,

More information

CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE

CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE Page no. 5.1 Introduction 132 5.2 Methods and measurements 132 5.3 Saturation thickness of elements and polymers

More information

Matter, energy, power and heat Units and prefixes used in radiography Radiological units Useful mathematics Proportions and the inverse square law

Matter, energy, power and heat Units and prefixes used in radiography Radiological units Useful mathematics Proportions and the inverse square law Chapter 1 Essential Mathematics and Physics Introduction Chapter contents Matter, energy, power and heat Units and prefixes used in radiography Radiological units Useful mathematics Proportions and the

More information

Basic Radiation Physics

Basic Radiation Physics Continuing Education Brought to you by Basic Radiation Physics Course Author(s): Hassem Geha, DDS, MDS CE Credits: 2 hours Intended Audience: Dentists, Dental Hygienists, Dental Assistants, Dental Students,

More information

Current and Recent ICRU Activities in Radiation Protection Dosimetry and Measurements

Current and Recent ICRU Activities in Radiation Protection Dosimetry and Measurements Current and Recent ICRU Activities in Radiation Protection Dosimetry and Measurements Hans-Georg Menzel International Commission on Radiation Units and Measurements (ICRU) The principal objective of ICRU

More information

Midterm Radiation Physics

Midterm Radiation Physics Name: Student ID: CHIR 204 Radiographic Physics and Protection Midterm Radiation Physics Friday 12 September, 2003 DO NOT TURN OVER THIS PAGE UNTIL INSTRUCTED TO DO SO Meanwhile, fill in your name and

More information

Basic principles of x-ray production

Basic principles of x-ray production Production of X-Rays part 1 George Starkschall, Ph.D. Lecture Objectives Identify what is needed to produce x-rays Describe how a diagnostic x-ray tube produces x-rays Describe the types of interactions

More information

INTERACTIONS OF RADIATION WITH MATTER

INTERACTIONS OF RADIATION WITH MATTER INTERACTIONS OF RADIATION WITH MATTER Renée Dickinson, MS, DABR Medical Physicist University of Washington Medical Center Department of Radiology Diagnostic Physics Section Outline Describe the various

More information

1 Random and systematic errors

1 Random and systematic errors 1 ESTIMATION OF RELIABILITY OF RESULTS Such a thing as an exact measurement has never been made. Every value read from the scale of an instrument has a possible error; the best that can be done is to say

More information

Some Thoughts on the Notion of Kinetic Energy.

Some Thoughts on the Notion of Kinetic Energy. Some Thoughts on the Notion of Kinetic Energy. Jeremy Dunning-Davies, Institute for Basic Research, Palm Harbor, Florida, U.S.A. and Institute for Theoretical Physics and Advanced Mathematics (IFM) Einstein-Galilei,

More information

CURRICULUM COURSE OUTLINE

CURRICULUM COURSE OUTLINE CURRICULUM COURSE OUTLINE Course Name(s): Grade(s): Department: Course Length: Pre-requisite: Introduction to Physics 9 th grade Science 1 semester Textbook/Key Resource: Conceptual Physical Science Explorations

More information

1-D Fourier Transform Pairs

1-D Fourier Transform Pairs 1-D Fourier Transform Pairs The concept of the PSF is most easily explained by considering a very small point source being placed in the imaging field-of-view The relationship between the image, I, and

More information

Absorption spectra variations of EBT radiochromic film from radiation exposure

Absorption spectra variations of EBT radiochromic film from radiation exposure INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 5 (25) N35 N4 PHYSICS IN MEDICINE AND BIOLOGY doi:.88/3-955/5/3/n2 NOTE Absorption spectra variations of EBT radiochromic film from radiation exposure M

More information

Ba (Z = 56) W (Z = 74) preferred target Mo (Z = 42) Pb (Z = 82) Pd (Z = 64)

Ba (Z = 56) W (Z = 74) preferred target Mo (Z = 42) Pb (Z = 82) Pd (Z = 64) Produced by accelerating electrons with high voltage and allowing them to collide with metal target (anode), e.g, Tungsten. Three Events (Two types of x-ray) a) Heat X-Ray Tube b) bremsstrahlung (braking

More information

FXA UNIT G485 Module X-Rays. Candidates should be able to : I = I 0 e -μx

FXA UNIT G485 Module X-Rays. Candidates should be able to : I = I 0 e -μx 1 Candidates should be able to : HISTORY Describe the nature of X-rays. Describe in simple terms how X-rays are produced. X-rays were discovered by Wilhelm Röntgen in 1865, when he found that a fluorescent

More information

Initial Certification

Initial Certification Initial Certification Medical Physics Part 1 Content Guide Part 1 Content Guides and Sample Questions PLEASE NOTE: List of Constants and Physical Values for Use on the Part 1 Physics Exam The ABR provides

More information

Chemical Engineering 412

Chemical Engineering 412 Chemical Engineering 412 Introductory Nuclear Engineering Lecture 26 Radiation Detection & Measurement II Spiritual Thought 2 I would not hold the position in the Church I hold today had I not followed

More information

CHAPTER 4 RADIATION ATTENUATION

CHAPTER 4 RADIATION ATTENUATION HDR202 PHYSICS FOR RADIOGRAPHERS 2 CHAPTER 4 RADIATION ATTENUATION PREPARED BY: MR KAMARUL AMIN BIN ABDULLAH SCHOOL OF MEDICAL IMAGING FACULTY OF HEALTH SCIENCES Learning Objectives At the end of the lesson,

More information

All instruction should be three-dimensional. Page 1 of 11

All instruction should be three-dimensional. Page 1 of 11 High School Modified Domains Model Course II - Physics Bundle 4: How Do We Use Energy to Communicate with Each Other? This is the fourth bundle of the High School Domains Model Course II - Physics. Each

More information

Development of new educational apparatus to visualize scattered X-rays

Development of new educational apparatus to visualize scattered X-rays Development of new educational apparatus to visualize scattered X-rays Poster No.: C-0073 Congress: ECR 2015 Type: Scientific Exhibit Authors: H. Hayashi, K. Takegami, H. Okino, K. Nakagawa, Y. 1 2 1 1

More information

Thermal Characteristics of Rotating Anode X-ray Tube with Emissivity in Aging Process for Digital Radiography

Thermal Characteristics of Rotating Anode X-ray Tube with Emissivity in Aging Process for Digital Radiography Research Paper Applied Science and Convergence Technology Vol.24 No.5, September 2015, pp.125 131 http://dx.doi.org/10.5757/asct.2015.24.5.125 Thermal Characteristics of Rotating Anode X-ray Tube with

More information

Chapter 16 Basic Precautions

Chapter 16 Basic Precautions Chapter 16 Basic Precautions 16.1 Basic Principles of Radiation Protection The four basic methods used to control radiation exposure are time, distance, shielding, and contamination control. The first

More information

Radiation Quantities and Units

Radiation Quantities and Units Radiation Quantities and Units George Starkschall, Ph.D. Lecture Objectives Define and identify units for the following: Exposure Kerma Absorbed dose Dose equivalent Relative biological effectiveness Activity

More information

ROC7080: RADIATION THERAPY PHYSICS LABORATORY LAB A : MECHANICAL MEASUREMENTS LINAC QA

ROC7080: RADIATION THERAPY PHYSICS LABORATORY LAB A : MECHANICAL MEASUREMENTS LINAC QA ROC7080: RADIATION THERAPY PHYSICS LABORATORY LAB A : MECHANICAL MEASUREMENTS LINAC QA GROUP I SPRING 2014 KEVIN JORDAN GRADUATE STUDENT, RADIOLOGICAL PHYSICS KARMANOS CANCER CENTER WAYNE STATE UNIVERSITY

More information

Georgia Institute of Technology. Radiation Detection & Protection (Day 3)

Georgia Institute of Technology. Radiation Detection & Protection (Day 3) Georgia Institute of Technology The George W. Woodruff School of Mechanical Engineering Nuclear & Radiological Engineering/Medical Physics Program Ph.D. Qualifier Exam Spring Semester 2009 Your ID Code

More information

A small object is placed a distance 2.0 cm from a thin convex lens. The focal length of the lens is 5.0 cm.

A small object is placed a distance 2.0 cm from a thin convex lens. The focal length of the lens is 5.0 cm. TC [66 marks] This question is about a converging (convex) lens. A small object is placed a distance 2.0 cm from a thin convex lens. The focal length of the lens is 5.0 cm. (i) Deduce the magnification

More information

INDUSTRIAL RADIOGRAPHY

INDUSTRIAL RADIOGRAPHY 1 BASICS INDUSTRIAL RADIOGRAPHY -Dr.O.Prabhakar, OP-TECH INTRODUCTION From the time Roentgen discovered X-rays and used it to radiograph his rifle, X-rays are being used in the industry to reveal internal

More information

CHARGED PARTICLE INTERACTIONS

CHARGED PARTICLE INTERACTIONS CHARGED PARTICLE INTERACTIONS Background Charged Particles Heavy charged particles Charged particles with Mass > m e α, proton, deuteron, heavy ion (e.g., C +, Fe + ), fission fragment, muon, etc. α is

More information

October 4th, Bioengineering 508: Physical Aspects of Medical Imaging Some Elementary Particles

October 4th, Bioengineering 508: Physical Aspects of Medical Imaging   Some Elementary Particles Bioengineering 508: Physical Aspects of Medical Imaging http://courses.washington.edu/bioen508/ For questions, remarks, discussions, errors in the book: Class Discussion Board (link from class website)

More information

X-ray imaging with amorphous selenium: Theoretical feasibility of the liquid crystal light valve for radiography

X-ray imaging with amorphous selenium: Theoretical feasibility of the liquid crystal light valve for radiography X-ray imaging with amorphous selenium: Theoretical feasibility of the liquid crystal light valve for radiography Pia-Krista Rieppo and J. A. Rowlands a) Departments of Medical Biophysics and Medical Imaging,

More information

Regulatory Considerations in the Licensing of a Mobile Backscatter X-ray Device used in Security Screening

Regulatory Considerations in the Licensing of a Mobile Backscatter X-ray Device used in Security Screening Regulatory Considerations in the Licensing of a Mobile Backscatter X-ray Device used in Security Screening Jim Scott a* and Leon Railey a a Australian Radiation Protection and Nuclear Safety Agency, Regulatory

More information

Training program for x-ray inspection film (RT)

Training program for x-ray inspection film (RT) of MTU Aero Engines AG Training program for x-ray inspection film (RT) In the NDT training course the following knowledge will be educated. approved by NANDTB Germany Preface + + + Introduction + + + Properties

More information

The LM741C Integrated Circuit As A Differential Amplifier Building The Electronic Thermometer

The LM741C Integrated Circuit As A Differential Amplifier Building The Electronic Thermometer BE 209 Group BEW6 Jocelyn Poruthur, Justin Tannir Alice Wu, & Jeffrey Wu November 19, 1999 The LM741C Integrated Circuit As A Differential Amplifier Building The Electronic Thermometer INTRODUCTION: A

More information

PHYS 3650L - Modern Physics Laboratory

PHYS 3650L - Modern Physics Laboratory PHYS 3650L - Modern Physics Laboratory Laboratory Advanced Sheet Photon Attenuation 1. Objectives. The objectives of this laboratory exercise are: a. To measure the mass attenuation coefficient at a gamma

More information

Photoelectric Effect

Photoelectric Effect PC1144 Physics IV Photoelectric Effect 1 Purpose Demonstrate the different predictions of the classical wave and quantum model of light with respect to the photoelectric effect. Determine an experimental

More information

Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays

Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays D T Burns Bureau International des Poids et Mesures, Pavillon de Breteuil, F-92312

More information

Nuclear Instruments and Methods in Physics Research A 417 (1998) 86 94

Nuclear Instruments and Methods in Physics Research A 417 (1998) 86 94 Nuclear Instruments and Methods in Physics Research A 417 (1998) 86 94 Experimental determination of detector gain, zero frequency detective quantum efficiency, and spectral compatibility of phosphor screens:

More information

Radiation Protection & Radiation Therapy

Radiation Protection & Radiation Therapy Radiation Protection & Radiation Therapy For Medical Students Professor of Medical Physics Radiation Units Activity Number disintegrations per second (Curie, Becquerel) Exposure (Roentgen, C/kg) Absorbed

More information

ADVANCED GCE 2825/02 PHYSICS A Health Physics FRIDAY 25 JANUARY Time: 1 hour 30 minutes

ADVANCED GCE 2825/02 PHYSICS A Health Physics FRIDAY 25 JANUARY Time: 1 hour 30 minutes *CUP/T36925* ADVANCED GCE 2825/02 PHYSICS A Health Physics FRIDAY 25 JANUARY 2008 Candidates answer on the question paper. Additional materials: Electronic calculator Morning Time: 1 hour 30 minutes INSTRUCTIONS

More information

Absorption of X-rays

Absorption of X-rays Absorption of X-rays TEP Related topics Bremsstrahlung, characteristic X-radiation, Bragg scattering, law of absorption, mass absorption coefficient, absorption edges, half-value thickness, photoelectric

More information

In today s lecture, we want to see what happens when we hit the target.

In today s lecture, we want to see what happens when we hit the target. In the previous lecture, we identified three requirements for the production of x- rays. We need a source of electrons, we need something to accelerate electrons, and we need something to slow the electrons

More information

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah

More information

Mitigation of External Radiation Exposures

Mitigation of External Radiation Exposures Mitigation of External Radiation Exposures The three (3) major principles to assist with maintaining doses ALARA are :- 1) Time Minimizing the time of exposure directly reduces radiation dose. 2) Distance

More information

Open Access. Suman Chakraborty* Q T + S gen = 1 S 1 S 2. Department of Mechanical Engineering, Indian Institute of Technology, Kharagpur , India

Open Access. Suman Chakraborty* Q T + S gen = 1 S 1 S 2. Department of Mechanical Engineering, Indian Institute of Technology, Kharagpur , India he Open hermodynamics Journal, 8,, 6-65 6 Open Access On the Role of External and Internal Irreversibilities towards Classical Entropy Generation Predictions in Equilibrium hermodynamics and their Relationship

More information

Chapter V: Cavity theories

Chapter V: Cavity theories Chapter V: Cavity theories 1 Introduction Goal of radiation dosimetry: measure of the dose absorbed inside a medium (often assimilated to water in calculations) A detector (dosimeter) never measures directly

More information

Modern physics ideas are strange! L 36 Modern Physics [2] The Photon Concept. How are x-rays produced? The uncertainty principle

Modern physics ideas are strange! L 36 Modern Physics [2] The Photon Concept. How are x-rays produced? The uncertainty principle L 36 Modern Physics [2] X-rays & gamma rays How lasers work Medical applications of lasers Applications of high power lasers Medical imaging techniques CAT scans MRI s Modern physics ideas are strange!

More information

Name: COMBINED SCIENCE Topics 4, 5 & 6 LEARNING OUTCOMES. Maintain a record of your progress Use the booklet to guide revision

Name: COMBINED SCIENCE Topics 4, 5 & 6 LEARNING OUTCOMES. Maintain a record of your progress Use the booklet to guide revision Name: COMBINED SCIENCE Topics 4, 5 & 6 LEARNING OUTCOMES Maintain a record of your progress Use the booklet to guide revision Close the Gap Contemporary record of the Topics / Learning outcomes that I

More information

Magnetic removal of electron contamination in radiotherapy x-ray beams

Magnetic removal of electron contamination in radiotherapy x-ray beams University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2006 Magnetic removal of electron contamination in radiotherapy

More information

Physics of Radiography

Physics of Radiography Physics of Radiography Yao Wang Polytechnic Institute of NYU Brooklyn, NY 11201 Based on J L Prince and J M Links Medical Imaging Signals and Based on J. L. Prince and J. M. Links, Medical Imaging Signals

More information

ATTENUATION STUDIES ON DRY AND HYDRATED CROSS-LINKED HYDROPHILIC COPOLYMER MATERIALS AT 8.02 TO kev USING X-RAY FLUORESCENT SOURCES

ATTENUATION STUDIES ON DRY AND HYDRATED CROSS-LINKED HYDROPHILIC COPOLYMER MATERIALS AT 8.02 TO kev USING X-RAY FLUORESCENT SOURCES Journal of Physical Science, Vol. 18(1), 23 32, 27 23 ATTENUATION STUDIES ON DRY AND HYDRATED CROSS-LINKED HYDROPHILIC COPOLYMER MATERIALS AT 8.2 TO 28.43 kev USING X-RAY FLUORESCENT SOURCES Sabar Bauk

More information

LAB 01 X-RAY EMISSION & ABSORPTION

LAB 01 X-RAY EMISSION & ABSORPTION LAB 0 X-RAY EMISSION & ABSORPTION REPORT BY: TEAM MEMBER NAME: Ashley Tsai LAB SECTION No. 05 GROUP 2 EXPERIMENT DATE: Feb., 204 SUBMISSION DATE: Feb. 8, 204 Page of 3 ABSTRACT The goal of this experiment

More information

Schriften der Naturforschenden Gesellschaft in Danzig Kikuchi Bands with X-Rays (preliminary communication)

Schriften der Naturforschenden Gesellschaft in Danzig Kikuchi Bands with X-Rays (preliminary communication) Schriften der Naturforschenden Gesellschaft in Danzig 1936 Kikuchi Bands with X-Rays (preliminary communication) by Hermann Determann By transmission of X-rays through a copper crystal we have observed

More information

Lab #13: Polarization

Lab #13: Polarization Lab #13: Polarization Introduction In this experiment we will investigate various properties associated with polarized light. We will study both its generation and application. Real world applications

More information

A) m B) m C) m D) m E) m. 5. Which one of the following circuits has the largest resistance?

A) m B) m C) m D) m E) m. 5. Which one of the following circuits has the largest resistance? Use the following to answer question 1. Two point charges, A and B, lie along a line separated by a distance L. The point x is the midpoint of their separation. 1. Which combination of charges would yield

More information

Possible Interactions. Possible Interactions. X-ray Interaction (Part I) Possible Interactions. Possible Interactions. section

Possible Interactions. Possible Interactions. X-ray Interaction (Part I) Possible Interactions. Possible Interactions. section Possible Interactions X-ray Interaction (Part I) Three types of interaction 1. Scattering Interaction with an atom Deflected May or may not loss of energy 1 Possible Interactions Three types of interaction

More information

CHAPTER 2 INTERACTION OF RADIATION WITH MATTER

CHAPTER 2 INTERACTION OF RADIATION WITH MATTER CHAPTER 2 INTERACTION OF RADIATION WITH MATTER 2.1 Introduction When gamma radiation interacts with material, some of the radiation will be absorbed by the material. There are five mechanisms involve in

More information

An industrial radiography exposure device based on measurement of transmitted gamma-ray intensity

An industrial radiography exposure device based on measurement of transmitted gamma-ray intensity Journal of Physics: Conference Series PAPER OPEN ACCESS An industrial radiography exposure device based on measurement of transmitted gamma-ray intensity To cite this article: C Polee et al 2015 J. Phys.:

More information

Two-Port Networks Admittance Parameters CHAPTER16 THE LEARNING GOALS FOR THIS CHAPTER ARE THAT STUDENTS SHOULD BE ABLE TO:

Two-Port Networks Admittance Parameters CHAPTER16 THE LEARNING GOALS FOR THIS CHAPTER ARE THAT STUDENTS SHOULD BE ABLE TO: CHAPTER16 Two-Port Networks THE LEARNING GOALS FOR THIS CHAPTER ARE THAT STUDENTS SHOULD BE ABLE TO: Calculate the admittance, impedance, hybrid, and transmission parameter for two-port networks. Convert

More information

A. I, II, and III B. I C. I and II D. II and III E. I and III

A. I, II, and III B. I C. I and II D. II and III E. I and III BioE 1330 - Review Chapters 7, 8, and 9 (Nuclear Medicine) 9/27/2018 Instructions: On the Answer Sheet, enter your 2-digit ID number (with a leading 0 if needed) in the boxes of the ID section. Fill in

More information

(INCLUDING THIS FRONT PAGE)

(INCLUDING THIS FRONT PAGE) I'IFIITIIBIFI UNIVERSITY OF SCIEI'ICE RITD TECHNOLOGY FACULTY OF HEALTH AND APPLIED SCIENCES DEPARTMENT OF NATURAL AND APPLIED SCIENCES QUALIFICATION: BACHELOR OF SCIENCE (MAJOR AND MINOR) QUALIFICATION

More information

PHYSICS A 2825/02. Health Physics. OXFORD CAMBRIDGE AND RSA EXAMINATIONS Advanced GCE. 1 hour 30 minutes

PHYSICS A 2825/02. Health Physics. OXFORD CAMBRIDGE AND RSA EXAMINATIONS Advanced GCE. 1 hour 30 minutes OXFORD CAMBRIDGE AND RSA EXAMINATIONS Advanced GCE PHYSICS A 2825/02 Health Physics Monday 27 JUNE 2005 Afternoon 1 hour 30 minutes Candidates answer on the question paper. Additional materials: Electronic

More information

X-RAY SPECTRA. Theory:

X-RAY SPECTRA. Theory: 12 Oct 18 X-ray.1 X-RAY SPECTRA In this experiment, a number of measurements involving x-rays will be made. The spectrum of x-rays emitted from a molybdenum target will be measured, and the experimental

More information

TEACHERS OF SCIENCE: Physics 9-12 FORM I-D GRID

TEACHERS OF SCIENCE: Physics 9-12 FORM I-D GRID 8710.4750 TEACHERS OF SCIENCE: sics 9-12 FORM I-D GRID Professional Education Program Evaluation Report (PEPER II) 8710.4750 Teachers of Science: sics 9-12 Subp. 7. Subject matter standards for teachers

More information

Basic structure of SEM

Basic structure of SEM Table of contents Basis structure of SEM SEM imaging modes Comparison of ordinary SEM and FESEM Electron behavior Electron matter interaction o Elastic interaction o Inelastic interaction o Interaction

More information

BI-LAYER PROTECTION AGAINST SCATTER RADIATION

BI-LAYER PROTECTION AGAINST SCATTER RADIATION BI-LAYER PROTECTION AGAINST SCATTER RADIATION An all new innovation in radiation protection core material THE STORY Kiarmor is an all new innovation in radiation protection core material, the combination

More information

11/10/2014. Chapter 1: Introduction to Medical Imaging. Projection (Transmission) vs. Emission Imaging. Emission Imaging

11/10/2014. Chapter 1: Introduction to Medical Imaging. Projection (Transmission) vs. Emission Imaging. Emission Imaging Chapter 1: Introduction to Medical Imaging Overview of Modalities Properties of an Image: Limitations on Information Content Contrast (both object & image): Brightness difference Sharpness (blur): Smallest

More information

Supplementary Information Supplementary Figures

Supplementary Information Supplementary Figures Supplementary Information Supplementary Figures Supplementary Figure S1. Schematic defining the relevant physical parameters in our grating coupler model. The schematic shows a normally-incident light

More information

h p λ = mν Back to de Broglie and the electron as a wave you will learn more about this Equation in CHEM* 2060

h p λ = mν Back to de Broglie and the electron as a wave you will learn more about this Equation in CHEM* 2060 Back to de Broglie and the electron as a wave λ = mν h = h p you will learn more about this Equation in CHEM* 2060 We will soon see that the energies (speed for now if you like) of the electrons in the

More information

MAT 17A - UHP - DISCUSSION #10 December 1, 2015

MAT 17A - UHP - DISCUSSION #10 December 1, 2015 MAT 17A - UHP - DISCUSSION #10 December 1, 2015 PROBLEM 1. Linear recursion equations Consider the linear recursion (finite difference) equation x j+1 = L x j = a x j + b, x 0 (1), where a and b are: (i)

More information

1. Module Details RADIOGRAPHY AND RADIATION SAFETY. 2. Module Purpose To provide learners with knowledge of the principles of industrial

1. Module Details RADIOGRAPHY AND RADIATION SAFETY. 2. Module Purpose To provide learners with knowledge of the principles of industrial 1. Module Details Module name Nominal duration RADIOGRAPHY AND RADIATION SAFETY 1 modules It is anticipated that a learner holding the prescribed entry level skills will achieve the module purpose in 35

More information

11/19/2014. Chapter 3: Interaction of Radiation with Matter in Radiology and Nuclear Medicine. Nuclide Families. Family Nuclides with Same: Example

11/19/2014. Chapter 3: Interaction of Radiation with Matter in Radiology and Nuclear Medicine. Nuclide Families. Family Nuclides with Same: Example 2014-2015 Residents' Core Physics Lectures Mondays 7:00-8:00 am in VA Radiology and UCSDMC Lasser Conference Rooms Topic Chapters Date Faculty 1 Introduction and Basic Physics 1, 2 M 11/17 Andre 2 Interaction

More information

HYDROBETA: A NEW INSTRUMENT FOR MEASURING IN-SITU PROFILES OF THE VOLUME SCATTERING FUNCTION FROM 10 TO 170 DEGREES

HYDROBETA: A NEW INSTRUMENT FOR MEASURING IN-SITU PROFILES OF THE VOLUME SCATTERING FUNCTION FROM 10 TO 170 DEGREES HYDROBETA: A NEW INSTRUMENT FOR MEASURING IN-SITU PROFILES OF THE VOLUME SCATTERING FUNCTION FROM 10 TO 170 DEGREES David R. Dana, Robert A. Maffione Hydro-Optics, Biology and Instrumentation Laboratories,

More information

Comparison of Polychromatic and Monochromatic X-rays for Imaging

Comparison of Polychromatic and Monochromatic X-rays for Imaging Comparison of Polychromatic and Monochromatic X-rays for Imaging M. Hoheisel 1, P. Bernhardt 1, R. Lawaczeck 2, and H. Pietsch 2 1 Siemens AG Medical Solutions, Forchheim, Germany 2 Schering AG Imaging

More information

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications Bedanta Kr. Deka, D. Thakuria, H. Bora and S. Banerjee # Department of Physicis, B. Borooah College, Ulubari,

More information

ASTM E545 TUTORIAL ON PROPER IMAGE QUALITY INDICATOR USAGE

ASTM E545 TUTORIAL ON PROPER IMAGE QUALITY INDICATOR USAGE ASTM E545 IQI USAGE TUTORIAL Page 1 of 9 ASTM E545 TUTORIAL ON PROPER IMAGE QUALITY INDICATOR USAGE Neutron radiography is a non-destructive testing method similar to the more familiar x- ray, but which

More information

GA A25658 PRECISION X-RAY OPTICAL DEPTH MEASUREMENTS IN ICF SHELLS

GA A25658 PRECISION X-RAY OPTICAL DEPTH MEASUREMENTS IN ICF SHELLS GA A25658 PRECISION X-RAY OPTICAL DEPTH MEASUREMENTS IN ICF SHELLS by S.A. EDDINGER, R.B. STEPHENS, H. HUANG, T.J. DRAKE, A. NIKROO, G. FLINT, and C.R. BYSTEDT JANUARY 2007 DISCLAIMER This report was prepared

More information

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields Journal of NUCLEAR SCIENCE and TECHNOLOGY, Vol. 44, No. 2, p. 109 113 (2007) ARTICLE A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard

More information

X-ray Absorption Spectroscopy

X-ray Absorption Spectroscopy X-ray Absorption Spectroscopy Nikki Truss November 26, 2012 Abstract In these experiments, some aspects of x-ray absorption spectroscopy were investigated. The x-ray spectrum of molybdenum was recorded

More information

TOPIC 3: READING AND REPORTING NUMERICAL DATA

TOPIC 3: READING AND REPORTING NUMERICAL DATA Page 1 TOPIC 3: READING AND REPORTING NUMERICAL DATA NUMERICAL DATA 3.1: Significant Digits; Honest Reporting of Measured Values Why report uncertainty? That is how you tell the reader how confident to

More information

APPLIED RADIATION PHYSICS

APPLIED RADIATION PHYSICS A PRIMER IN APPLIED RADIATION PHYSICS F A SMITH Queen Mary & Westfield College, London fe World Scientific m Singapore * New Jersey London Hong Kong CONTENTS CHAPTER 1 : SOURCES of RADIATION 1.1 Introduction

More information