BORON IN GLASS DETERMINATION USING WDXRF

Size: px
Start display at page:

Download "BORON IN GLASS DETERMINATION USING WDXRF"

Transcription

1 269 ABSTRACT BORON IN GLASS DETERMINATION USING WDXRF Alexander Seyfarth Bruker AXS Inc., 5465 East Cheryl Parkway, Madison, Wisconsin The application of wavelength-dispersive XRF to the analysis of boron in glass is reviewed with respect to physical, instrumental, calibration, and maintenance considerations. Advances in XRF instrumentation have enabled improvement in the analysis of boron. Reduced window thickness and high-current capabilities of X-ray tubes, as well as new analyzer crystals, produce a higher intensity boron signal. With correct determination of background and matrix interferences for successful quantification, only the sample itself limits the analysis. Observed boron migration in some glasses needs to be understood in order to establish stable, routine quantification in a process environment. INTRODUCTION Boron is an important component of glass. It exerts a strong influence on the viscosity of the melt, imparting a beneficial effect to the chemical resistance and lowering the expansion coefficient of the finished article. Boron is widely used in the production of E type glasses (7 wt%), neutral glass for pharmaceutical applications (7 to 11 wt%), and Pyrex -type borosilicate glasses (13 wt%). Because of the high volatility of boron in the melting furnace, a very accurate procedure is needed to meet the strict boron concentration specifications required for the finished product and to permit any necessary batch adjustment to be implemented almost in real time. XRF is well suited for glass analyses. Since the 1980s, XRF has been the method of choice because of its ease of sample preparation and rapid results. In the production control of soda-lime glass, the technique has already replaced wet chemical techniques because of its high precision and fast measurement time. Boron, being a very light element, poses challenges to the analyst as well as to the instrument. Demand for boron analysis in the glass industry prompted the development of new generation wavelength-dispersive X-ray fluorescence (WDXRF) spectrometers. ASPECTS OF THE ANALYSIS OF BORON To put the difficulty of the analysis of boron in glass into perspective, consider its physical and chemical properties [1 3]. Physical data for boron is given in Table I. Table I. Physical data for boron. Atomic number (Z) K shell absorption edge (ev) Kα radiation (ev) Mass absorption coefficient µλ (cm 2 /g) Fluorescence yield (Ω) , Boron s low atomic number and low K-shell energy cause various issues for XRF analysis which will be addressed further.

2 This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website ICDD Website -

3 270 EXCITATION AND FLUORESCENCE YIELD The emitted intensity, I B-Kα, of the characteristic fluorescence radiation is governed by the overlap of the excitation spectrum and the absorption coefficient µ λ of boron [4]. Absorption by the tube window cuts off the low energy (long wavelength) part of the excitation tube spectrum. Because of this, boron is excited with the primary X-ray continuum and the rhodium L lines. For optimum primary continuum, the tube window must be as thin as possible. Currently window thicknesses range from 150 to 30 µm; 75 µm has become the industry standard since it balances price, performance, and durability. Fluorescence yield is another important factor in the analysis of boron. Fluorescence yield is proportional to Z 4 and is therefore close to zero for low atomic number elements. Using the highest available tube current at the lowest possible voltage, the emission of boron photons from the sample can be optimized. Starting in 1988, the possible settings for tube current have increased from 100 ma to today s maximum of 170 ma [Siemens SRS 303 (1988) to Bruker S8 TIGER 4 kw (2006)]. Figure 1. Comparison of boron intensity in 3 kw and 4 kw WDXRF systems. As can be seen in Figure 1, 40% greater boron intensity was obtained from a boronphosphosilicate glass (BPSG) sample (4 wt% B) analyzed on a 4 kw instrument in 1999 than a 3 kw instrument in 1988 [SRS 3000 with 75 µm end window X-ray tube and 3 and 4 kw generator (1999)].

4 271 ANALYZED LAYER, SAMPLE PREPARATION, AND X-RAY BEAM PATH Because the energy of boron Kα radiation is so low (183 ev), only a thin 0.6 µm (600 nm) surface layer of the sample can be analyzed. The density and composition of the glass can, in some cases, reduce this down to 200 nm. The glass samples to be analyzed must therefore be prepared with a good surface finish. Mellott compiled and investigated the surface roughnesses of commercial glasses using Atomic Force Microscopy (AFM) and found that many are in the range from 2 to 0.2 nm [5]. Quantitative boron analysis on non-polished glasses was much less accurate and repeatable than on the same polished material (private communication in 2004 with Brian Simpkins, Alcan Packaging, formerly Wheaton Glass, Millville, New Jersey). Reducing the path length of the X-ray beam (tube-to-sample and sample-to-detector) and using thin-window flow proportional detectors, combined with a controlled vacuum of better than 0.1 mbar, enables higher yield of detected boron radiation. Comparing the published sensitivities from a 1998 dual-turret spectrometer with those from a direct-loading S4-type system using the same X-ray tube, the type of generator shows a substantial effect, attributable to the optimized beam path and window design [6 8]. Changes in the vacuum during the measurement affect the count rate of the detected boron and needs to be taken into consideration for those systems that have large sample compartments (e.g., dual-turret systems). In contrast, direct-loading systems can usually achieve a stable vacuum without changeover during the whole measurement period. These direct-loading systems feature pressure-controlled sample introduction. Measurement is started only when the pressure in the sample compartment reaches a certain level, resulting in constant pressure and higher, more stable count rates for boron. ANALYZER CRYSTALS Figure 2. Comparison of boron intensity collected with a traditional crystal versus multilayer crystal. Ricardo et al. showed in 2001 that with a new type of patented multilayer detector crystal, the boron signal can be raised by 28% [9]. The XS-B crystal depicted in Figure 2 is a LaB 4 C multilayer and is available from Incoatec GmbH (Gesthacht, Germany). Schuster et al. performed a feasibility study for the analysis of boron in a coating of BPSG, a common semiconductor material, in 1987 [10]. Using a WDXRF system equipped with a 125 µm endwindow X-ray tube, 80 ma tube current, and early multilayer crystal (OVO 160), he obtained a sensitivity of 2.7 counts/s per wt% B. The detection limit was reported to be 0.04 wt% for 600 s measurement time.

5 272 Using a 200 Å Mo-B 4 C multilayer crystal (OVO-B) and an S4 Series WDXRF system optimized for beam path, Mauser achieved a detection limit of 0.02 wt % B in 100 s and a sensitivity of 330 counts/s per wt% B on the same type of material [7]. Using the XS-B multilayer crystal on the same S4 instrument further improved the sensitivity to 420 counts/s per wt% B as was shown by Behrens in 2004 [11]. CHOICE OF COLLIMATORS AND BACKGROUND POSITIONS Multilayer crystals produce wide peak profiles and should be used with coarse collimators. The rule of thumb in WDXRF is to use the coarsest collimator for the lightest element detection. Since achieving a high count rate is paramount for the analysis, peak resolution and interference corrections are not an issue. This holds true for BPSG, but in order to create a meaningful widerrange calibration for commercial borosilicate and E glass as well, count rate needs to be balanced against resolution [12]. K Ll, Ca Ll Cl Ll, B KA1 O (3) Zr Mz Si LB/ S Ll Ln Ca Ca and K L l lines, as well as Ca Ln and Zr M lines, should be separated as shown in Figure 3. The effect of higher order reflections of O cannot be reduced much by using pulse height discrimination and needs to be minimized by collimation. With Ca and Zr present in the analyzed material, the selection of background positions becomes difficult. The approach from Feather and Willis is particularly useful in this regard, as it can be used to perform interference correction on the background [13]. CALIBRATION APPROACH < Y : 0 to KCps > Figure 3. Interference on and around the boron K signal. Obtaining a sufficient number of glass standards, as well as including enough samples to perform the interference corrections needed, pose the biggest challenge to the analyst. The quality of the boron certification by Manitol-based volumetric means is very much dependent on the experience of the analyst with that methodology. Few commercial labs now offer Manitol-based boron analysis. Inductively coupled plasma (ICP) methods are increasingly being offered instead, but these methods are time- and cost-prohibitive because of the requirement for sample digestion. Few glass samples of defined boron composition are commercially available. Many borosilicate glass samples, such as those available from Breitlaender GmbH (Hamm, Germany), are just premelt batch and not chemically certified. Borosilicate glass suppliers are the best source of sample material. They can produce a range of experimental matrix for the calibration based on past batch-melt samples. A comparison and discussion of the quality of certification techniques for boron in glass is unfortunately absent from the literature.

6 273 Using over 40 glass samples of standard 32 mm size and a small 28 mm aperture, a calibration for boron was established from 0.00 to wt%. The calibration has a standard error of calibration (SEC) of wt%. The sensitivity is 94 counts/s per wt% B 2 O 3. An empirical calculation using a concentration-based correction approach produced better results than any fundamental parameters (FP) model, even with variable influence coefficients. FP for boron, as taken from the NIST-compiled tables, are not well characterized and would need further improvements. Of greatest importance is interference correction of both the background and peak position to obtain a good net intensity for boron. MAINTENANCE OF CALIBRATION In order to maintain the calibration, drift correction is needed to account for changes in detector gas, foil, and tube signal that can influence the intensity and therefore the analysis. A very interesting study by Guadagnino, Sundberg, Michiels, and Brochot showed that some glass types exhibit a migration of boron when measured repeatedly [14]. Figure 4 shows the change in B 2 O 3 content in Pyrex-type and E-type glass as a function of the number of WDXRF measurements. Borosilicate glass should therefore be used as a drift correction and QC sample only after it has been conditioned. Figure 4. Long-term stability of boron in E-type and Pyrex-type glass

7 274 CONCLUSION The analysis of boron in glass using WDXRF has dramatically improved since its first application as a process control tool in the late 1980s. Nevertheless, it is still one of the most challenging applications, since the analyst must gather enough reference material and optimize conditions to fit the spectrometer capabilities. The choice of the right collimator, background and peak counting positions, correct pulse height analysis settings, and interference correction method greatly influences the range and durability of the resulting calibration. Maintenance of the calibration and a daily check of the system and calibration need to take the stability of the glass standards into account. ACKNOWLEDGMENTS The author thanks Brian Simpkins (Alcan Packaging, formerly Wheaton Glass, Millville, New Jersey) for his time performing the scans and tests on the borosilicate glass samples, as well as for evaluating the XS-B and OVO-B crystals side-by-side in a process environment. Many thanks as well to my colleagues at Bruker AXS Larry Arias, Kai Behrens, Karl Mauser, and Karen Roscoe for contributing to the presentation and article. REFERENCES [1] Bearden, J. A. Rev. Mod. Phys. 1967, 39, [2] Krause, M. O. J. Phys. Chem. Ref. Data 1979, 8, [3] White, E. W.; Johnson Jr., G. G. X-ray Emission and Absorption Wavelengths and Two- Theta Tables (ASTM Data Series DS 37A); American Society for Testing and Materials: Philadelphia, [4] Shiraiwa, T.; Fujino, N. Jpn. J. Appl. Phys. 1966, 5, [5] Mellott, N. P.; Brantley, S. L.; Hamilton, J. P.; Pantano, C. G. Surf. Interface Anal. 2001, 31, [6] Mauser, K. E.; Uhlig, S. SRS 3400 Boron in BPSG. Bruker AXS GmbH Application Note, [7] Mauser, K. E. S4 PIONEER Boron in BPSG. Bruker AXS GmbH Application Note, [8] Buehler, A.; Seyfarth, A.; Schlotz, R. Introduction to XRF. Bruker AXS GmbH Training Manual, [9] Ricardo, P.; Wiesmann, J.; Nowak, C.; Michaelsen, C.; Bormann, R. Appl. Opt. 2001, 40, [10] Schuster, M.; Müller, L.; Mauser, K. E.; Straub, R. Thin Solid Films 1988, 157, [11] Behrens, K. New Applications with the XS-B; S4 PIONEER. Bruker AXS GmbH Application Note, [12] Uhlig, S.; Müller, L. Boron Analysis in Ceramic and Glass Industries using Wavelength Dispersive X-ray Fluorescence Analysis. Siemens Analytical Application Notes 326, [13] Feather, C. E.; Willis, J. P. X-Ray Spectrom. 1976, 5, [14] Guadagnino, E.; Sundberg, P.; Michiels, D.; Brochot, D. A collaborative study into the determination of boron in glass using X-ray fluorescence (XRF) spectroscopy. ICG/TC2 report, 2006.

RADIOACTIVE SAMPLE EFFECTS ON EDXRF SPECTRA

RADIOACTIVE SAMPLE EFFECTS ON EDXRF SPECTRA 90 RADIOACTIVE SAMPLE EFFECTS ON EDXRF SPECTRA Christopher G. Worley Los Alamos National Laboratory, MS G740, Los Alamos, NM 87545 ABSTRACT Energy dispersive X-ray fluorescence (EDXRF) is a rapid, straightforward

More information

AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD

AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN 1097-0002 219 AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD Takao Moriyama 1), Atsushi Morikawa 1), Makoto

More information

MEASUREMENT CAPABILITIES OF X-RAY FLUORESCENCE FOR BPSG FILMS

MEASUREMENT CAPABILITIES OF X-RAY FLUORESCENCE FOR BPSG FILMS , MEASUREMENT CAPABILITIES OF X-RAY FLUORESCENCE FOR BPSG FILMS K.O. Goyal, J.W. Westphal Semiconductor Equipment Group Watkins-Johnson Company Scotts Valley, California 95066 Abstract Deposition of borophosphosilicate

More information

FUNDAMENTAL PARAMETER METHOD USING SCATTERING X-RAYS IN X-RAY FLUORESCENCE ANALYSIS

FUNDAMENTAL PARAMETER METHOD USING SCATTERING X-RAYS IN X-RAY FLUORESCENCE ANALYSIS FUNDAMENTAL PARAMETER METHOD USING SCATTERING X-RAYS IN X-RAY FLUORESCENCE ANALYSIS 255 Yoshiyuki Kataoka 1, Naoki Kawahara 1, Shinya Hara 1, Yasujiro Yamada 1, Takashi Matsuo 1, Michael Mantler 2 1 Rigaku

More information

S8 TIGER Series 2 for ASTM D 6443

S8 TIGER Series 2 for ASTM D 6443 Lab Report XRF 139 S8 TIGER Series 2 for ASTM D 6443 Standard Test Method for Determination of Ca, Cl, Cu, Mg, P, S and Zn in Unused Lubricating Oils and Additives Introduction Lubricating oils are generally

More information

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K. 783 SCOPE AND LIMITATIONS XRF ANALYSIS FOR SEMI-QUANTITATIVE Introduction Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K. Historically x-ray fluorescence spectrometry has

More information

THE IMPORTANCE OF THE SPECIMEN DISPLACEMENT CORRECTION IN RIETVELD PATTERN FITTING WITH SYMMETRIC REFLECTION-OPTICS DIFFRACTION DATA

THE IMPORTANCE OF THE SPECIMEN DISPLACEMENT CORRECTION IN RIETVELD PATTERN FITTING WITH SYMMETRIC REFLECTION-OPTICS DIFFRACTION DATA Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 96 THE IMPORTANCE OF THE SPECIMEN DISPLACEMENT CORRECTION IN RIETVELD PATTERN FITTING WITH SYMMETRIC REFLECTION-OPTICS

More information

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF 2016 Fall Semester MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF Byungha Shin Dept. of MSE, KAIST 1 Course Information Syllabus 1. Overview of various characterization techniques (1 lecture)

More information

A COMPACT X-RAY SPECTROMETER WITH MULTI-CAPILLARY X-RAY LENS AND FLAT CRYSTALS

A COMPACT X-RAY SPECTROMETER WITH MULTI-CAPILLARY X-RAY LENS AND FLAT CRYSTALS Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 320 A COMPACT X-RAY SPECTROMETER WITH MULTI-CAPILLARY X-RAY LENS AND FLAT CRYSTALS Hiroyoshi SOEJIMA and

More information

ULTRATHIN LAYER DEPOSITIONS A NEW TYPE OF REFERENCE SAMPLES FOR HIGH PERFORMANCE XRF ANALYSIS

ULTRATHIN LAYER DEPOSITIONS A NEW TYPE OF REFERENCE SAMPLES FOR HIGH PERFORMANCE XRF ANALYSIS 298 299 ULTRATHIN LAYER DEPOSITIONS A NEW TYPE OF REFERENCE SAMPLES FOR HIGH PERFORMANCE XRF ANALYSIS M. Krämer 1), R. Dietsch 1), Th. Holz 1), D. Weißbach 1), G. Falkenberg 2), R. Simon 3), U. Fittschen

More information

FUNDAMENTAL PARAMETERS ANALYSIS OF ROHS ELEMENTS IN PLASTICS

FUNDAMENTAL PARAMETERS ANALYSIS OF ROHS ELEMENTS IN PLASTICS 45 ABSTRACT FUNDAMENTAL PARAMETERS ANALYSIS OF ROHS ELEMENTS IN PLASTICS W. T. Elam, Robert B. Shen, Bruce Scruggs, and Joseph A. Nicolosi EDAX, Inc. Mahwah, NJ 70430 European Community Directive 2002/95/EC

More information

Elemental analysis by X-ray f luorescence. Sequential benchtop WDXRF spectrometer

Elemental analysis by X-ray f luorescence. Sequential benchtop WDXRF spectrometer Elemental analysis by X-ray f luorescence Sequential benchtop WDXRF spectrometer Elemental analysis is one of the most important fundamental measurements made for industrial quality control and research

More information

ADVANTAGES AND DISADVANTAGES OF BAYESIAN METHODS FOR OBTAINING XRF NET INTENSITIES

ADVANTAGES AND DISADVANTAGES OF BAYESIAN METHODS FOR OBTAINING XRF NET INTENSITIES 187 188 ADVANTAGES AND DISADVANTAGES OF BAYESIAN METHODS FOR OBTAINING XRF NET INTENSITIES ABSTRACT W. T. Elam, B. Scruggs, F. Eggert, and J. A. Nicolosi EDAX, a unit of Ametek Inc., 91 McKee Drive, Mahwah,

More information

ELECTRIC FIELD INFLUENCE ON EMISSION OF CHARACTERISTIC X-RAY FROM Al 2 O 3 TARGETS BOMBARDED BY SLOW Xe + IONS

ELECTRIC FIELD INFLUENCE ON EMISSION OF CHARACTERISTIC X-RAY FROM Al 2 O 3 TARGETS BOMBARDED BY SLOW Xe + IONS 390 ELECTRIC FIELD INFLUENCE ON EMISSION OF CHARACTERISTIC X-RAY FROM Al 2 O 3 TARGETS BOMBARDED BY SLOW Xe + IONS J. C. Rao 1, 2 *, M. Song 2, K. Mitsuishi 2, M. Takeguchi 2, K. Furuya 2 1 Department

More information

Time-Resolved μ-xrf and Elemental Mapping of Biological Materials

Time-Resolved μ-xrf and Elemental Mapping of Biological Materials 296 Time-Resolved μ-xrf and Elemental Mapping of Biological Materials K. Tsuji 1,2), K. Tsutsumimoto 1), K. Nakano 1,2), K. Tanaka 1), A. Okhrimovskyy 1), Y. Konishi 1), and X. Ding 3) 1) Department of

More information

IMPROVING THE ACCURACY OF RIETVELD-DERIVED LATTICE PARAMETERS BY AN ORDER OF MAGNITUDE

IMPROVING THE ACCURACY OF RIETVELD-DERIVED LATTICE PARAMETERS BY AN ORDER OF MAGNITUDE Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 158 IMPROVING THE ACCURACY OF RIETVELD-DERIVED LATTICE PARAMETERS BY AN ORDER OF MAGNITUDE B. H.

More information

TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER

TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 236 ABSTRACT TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER Hagen Stosnach Röntec GmbH,

More information

EFFECT OF CALIBRATION SPECIMEN PREPARATION TECHNIQUES ON NARROW RANGE X-RAY FLUORESCENCE CALIBRATION ACCURACY

EFFECT OF CALIBRATION SPECIMEN PREPARATION TECHNIQUES ON NARROW RANGE X-RAY FLUORESCENCE CALIBRATION ACCURACY Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 424 EFFECT OF CALIBRATION SPECIMEN PREPARATION TECHNIQUES ON NARROW RANGE X-RAY FLUORESCENCE CALIBRATION

More information

INFLUENCE OF GROWTH INTERRUPTION ON THE FORMATION OF SOLID-STATE INTERFACES

INFLUENCE OF GROWTH INTERRUPTION ON THE FORMATION OF SOLID-STATE INTERFACES 122 INFLUENCE OF GROWTH INTERRUPTION ON THE FORMATION OF SOLID-STATE INTERFACES I. Busch 1, M. Krumrey 2 and J. Stümpel 1 1 Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany

More information

NEW CORRECTION PROCEDURE FOR X-RAY SPECTROSCOPIC FLUORESCENCE DATA: SIMULATIONS AND EXPERIMENT

NEW CORRECTION PROCEDURE FOR X-RAY SPECTROSCOPIC FLUORESCENCE DATA: SIMULATIONS AND EXPERIMENT Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 266 NEW CORRECTION PROCEDURE FOR X-RAY SPECTROSCOPIC FLUORESCENCE DATA: SIMULATIONS AND EXPERIMENT

More information

ABNORMAL X-RAY EMISSION FROM INSULATORS BOMBARDED WITH LOW ENERGY IONS

ABNORMAL X-RAY EMISSION FROM INSULATORS BOMBARDED WITH LOW ENERGY IONS 302 ABNORMAL X-RAY EMISSION FROM INSULATORS BOMBARDED WITH LOW ENERGY IONS M. Song 1, K. Mitsuishi 1, M. Takeguchi 1, K. Furuya 1, R. C. Birtcher 2 1 High Voltage Electron Microscopy Station, National

More information

ACCURATE QUANTIFICATION OF RADIOACTIVE MATERIALS BY X-RAY FLUORESCENCE: GALLIUM IN PLUTONIUM METAL

ACCURATE QUANTIFICATION OF RADIOACTIVE MATERIALS BY X-RAY FLUORESCENCE: GALLIUM IN PLUTONIUM METAL Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 369 ACCURATE QUANTIFICATION OF RADIOACTIVE MATERIALS BY X-RAY FLUORESCENCE: GALLIUM IN PLUTONIUM

More information

PERFORMANCE OF A ROOM TEMPERATURE GAS PROPORTIONAL SCINTILLATION COUNTER IN X-RAY ANALYSIS OF METALLIC ALLOYS EXCITED WITH ALPHA PARTICLES

PERFORMANCE OF A ROOM TEMPERATURE GAS PROPORTIONAL SCINTILLATION COUNTER IN X-RAY ANALYSIS OF METALLIC ALLOYS EXCITED WITH ALPHA PARTICLES 249 PERFORMANCE OF A ROOM TEMPERATURE GAS PROPORTIONAL SCINTILLATION COUNTER IN X-RAY ANALYSIS OF METALLIC ALLOYS EXCITED WITH ALPHA PARTICLES F. I. G. M. Borges, S. J. C. do Carmo, T. H. V. T. Dias, F.

More information

USABILITY OF PORTABLE X-RAY SPECTROMETER FOR DISCRIMINATION OF VALENCE STATES

USABILITY OF PORTABLE X-RAY SPECTROMETER FOR DISCRIMINATION OF VALENCE STATES Copyright (c)jcpds-international Centre for Diffraction Data 00, Advances in X-ray Analysis, Volume 45. 409 ISSN 1097-000 USABIITY OF POTABE X-AY SPECTOMETE FO DISCIMINATION OF VAENCE STATES I.A.Brytov,.I.Plotnikov,B.D.Kalinin,

More information

BENEFITS OF IMPROVED RESOLUTION FOR EDXRF

BENEFITS OF IMPROVED RESOLUTION FOR EDXRF 135 Abstract BENEFITS OF IMPROVED RESOLUTION FOR EDXRF R. Redus 1, T. Pantazis 1, J. Pantazis 1, A. Huber 1, B. Cross 2 1 Amptek, Inc., 14 DeAngelo Dr, Bedford MA 01730, 781-275-2242, www.amptek.com 2

More information

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY Copyright -International Centre for Diffraction Data 2010 ISSN 1097-0002 CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY B. Chyba, M. Mantler, H. Ebel, R. Svagera Technische Universit Vienna,

More information

S2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF

S2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF S2 PICOFOX Spectrometry Solutions Innovation with Integrity TXRF S2 PICOFOX True Trace Analysis with XRF for the First Time! You need to know the concentration of trace elements in environmental samples?

More information

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 511 FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON

More information

Overview of X-Ray Fluorescence Analysis

Overview of X-Ray Fluorescence Analysis Overview of X-Ray Fluorescence Analysis AMPTEK, INC., Bedford, MA 01730 Ph: +1 781 275 2242 Fax: +1 781 275 3470 sales@amptek.com 1 What is X-Ray Fluorescence (XRF)? A physical process: Emission of characteristic

More information

ANALYSIS OF LOW MASS ABSORPTION MATERIALS USING GLANCING INCIDENCE X-RAY DIFFRACTION

ANALYSIS OF LOW MASS ABSORPTION MATERIALS USING GLANCING INCIDENCE X-RAY DIFFRACTION 173 ANALYSIS OF LOW MASS ABSORPTION MATERIALS USING GLANCING INCIDENCE X-RAY DIFFRACTION N. A. Raftery, L. K. Bekessy, and J. Bowpitt Faculty of Science, Queensland University of Technology, GPO Box 2434,

More information

SEM. Chemical Analysis in the. Elastic and Inelastic scattering. Chemical analysis in the SEM. Chemical analysis in the SEM

SEM. Chemical Analysis in the. Elastic and Inelastic scattering. Chemical analysis in the SEM. Chemical analysis in the SEM THE UNIVERSITY Chemical Analysis in the SEM Ian Jones Centre for Electron Microscopy OF BIRMINGHAM Elastic and Inelastic scattering Electron interacts with one of the orbital electrons Secondary electrons,

More information

Praktikum zur. Materialanalytik

Praktikum zur. Materialanalytik Praktikum zur Materialanalytik Energy Dispersive X-ray Spectroscopy B513 Stand: 19.10.2016 Contents 1 Introduction... 2 2. Fundamental Physics and Notation... 3 2.1. Alignments of the microscope... 3 2.2.

More information

LASER-COMPTON SCATTERING AS A POTENTIAL BRIGHT X-RAY SOURCE

LASER-COMPTON SCATTERING AS A POTENTIAL BRIGHT X-RAY SOURCE Copyright(C)JCPDS-International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Vol.46 74 ISSN 1097-0002 LASER-COMPTON SCATTERING AS A POTENTIAL BRIGHT X-RAY SOURCE K. Chouffani 1, D. Wells

More information

Issues With TXRF Angle Scans and Calibration

Issues With TXRF Angle Scans and Calibration Copyright (C) JCPDS-International Centre for Diffraction Data 1999 794 Issues With TXRF Angle Scans and Calibration Dennis Werho, Stephen N. Schauer, and George F. Carney, Motorola, Inc., AZ Abstract Previous

More information

Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source

Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source 3rd International EUVL Symposium NOVEMBER 1-4, 2004 Miyazaki, Japan Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source H. Tanaka, A. Matsumoto, K. Akinaga, A. Takahashi

More information

MATERIALS CHARACTERIZATION USING A NOVEL SIMULTANEOUS NEAR-INFRARED/X-RAY DIFFRACTION INSTRUMENT

MATERIALS CHARACTERIZATION USING A NOVEL SIMULTANEOUS NEAR-INFRARED/X-RAY DIFFRACTION INSTRUMENT Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 249 MATERIALS CHARACTERIZATION USING A NOVEL SIMULTANEOUS NEAR-INFRARED/X-RAY DIFFRACTION INSTRUMENT

More information

IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE

IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE Copyright JCPDS-International Centre for Diffraction Data 2012 ISSN 1097-0002 281 IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE Susumu Imashuku 1, Deh Ping Tee 1, Yasukazu

More information

XUV 773: X-Ray Fluorescence Analysis of Gemstones

XUV 773: X-Ray Fluorescence Analysis of Gemstones Fischer Application report vr118 HELM UT FISCHER GMBH + CO. KG Institut für Elektronik und Messtechnik Industriestrasse 21-7169 Sindelfingen, Germany Tel.: (+49) 731 33- - Fax: (+49) 731 33-79 E-Mail:

More information

Standardless Analysis by XRF but I don t know what s in my sample!! Dr Colin Slater Applications Scientist, XRF Bruker UK Limited

Standardless Analysis by XRF but I don t know what s in my sample!! Dr Colin Slater Applications Scientist, XRF Bruker UK Limited by XRF but I don t know what s in my sample!! Dr Colin Slater Applications Scientist, XRF Bruker UK Limited XRF Standardless Analysis In this talk What is meant by standardless analysis? Fundamental Parameters

More information

DEVELOPMENT OF A NEW POSITRON LIFETIME SPECTROSCOPY TECHNIQUE FOR DEFECT CHARACTERIZATION IN THICK MATERIALS

DEVELOPMENT OF A NEW POSITRON LIFETIME SPECTROSCOPY TECHNIQUE FOR DEFECT CHARACTERIZATION IN THICK MATERIALS Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 59 DEVELOPMENT OF A NEW POSITRON LIFETIME SPECTROSCOPY TECHNIQUE FOR DEFECT CHARACTERIZATION IN

More information

APPLICATION OF MICRO X-RAY FLUORESCENCE SPECTROMETRY FOR LOCALIZED AREA ANALYSIS OF BIOLOGICAL AND ENVIRONMENTAL MATERIALS

APPLICATION OF MICRO X-RAY FLUORESCENCE SPECTROMETRY FOR LOCALIZED AREA ANALYSIS OF BIOLOGICAL AND ENVIRONMENTAL MATERIALS Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 540 APPLICATION OF MICRO X-RAY FLUORESCENCE SPECTROMETRY FOR LOCALIZED AREA ANALYSIS OF BIOLOGICAL AND

More information

FACTORS AFFECTING IN-LINE PHASE CONTRAST IMAGING WITH A LABORATORY MICROFOCUS X-RAY SOURCE

FACTORS AFFECTING IN-LINE PHASE CONTRAST IMAGING WITH A LABORATORY MICROFOCUS X-RAY SOURCE Copyright JCPDS-International Centre for Diffraction Data 26 ISSN 197-2 FACTORS AFFECTING IN-LINE PHASE CONTRAST IMAGING WITH A LABORATORY MICROFOCUS X-RAY SOURCE 31 K. L. Kelly and B. K. Tanner Department

More information

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis MT-0.6026 Electron microscopy Scanning electron microscopy and electron probe microanalysis Eero Haimi Research Manager Outline 1. Introduction Basics of scanning electron microscopy (SEM) and electron

More information

GLANCING INCIDENCE XRF FOR THE ANALYSIS OF EARLY CHINESE BRONZE MIRRORS

GLANCING INCIDENCE XRF FOR THE ANALYSIS OF EARLY CHINESE BRONZE MIRRORS 176 177 GLANCING INCIDENCE XRF FOR THE ANALYSIS OF EARLY CHINESE BRONZE MIRRORS Robert W. Zuneska, Y. Rong, Isaac Vander, and F. J. Cadieu* Physics Dept., Queens College of CUNY, Flushing, NY 11367. ABSTRACT

More information

X-RAY MICRODIFFRACTION STUDY OF THE HALF-V SHAPED SWITCHING LIQUID CRYSTAL

X-RAY MICRODIFFRACTION STUDY OF THE HALF-V SHAPED SWITCHING LIQUID CRYSTAL Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 321 X-RAY MICRODIFFRACTION STUDY OF THE HALF-V SHAPED SWITCHING LIQUID CRYSTAL Kazuhiro Takada 1,

More information

Infrastructure of Thin Films Laboratory in Institute of Molecular Physics Polish Academy of Sciences

Infrastructure of Thin Films Laboratory in Institute of Molecular Physics Polish Academy of Sciences Infrastructure of Thin Films Laboratory in Institute of Molecular Physics Polish Academy of Sciences Outline Sample preparation Magnetron sputtering Ion-beam sputtering Pulsed laser deposition Electron-beam

More information

Silicon Drift Detectors: Understanding the Advantages for EDS Microanalysis. Patrick Camus, PhD Applications Scientist March 18, 2010

Silicon Drift Detectors: Understanding the Advantages for EDS Microanalysis. Patrick Camus, PhD Applications Scientist March 18, 2010 Silicon Drift Detectors: Understanding the Advantages for EDS Microanalysis Patrick Camus, PhD Applications Scientist March 18, 2010 EDS Detector Requirements Detect whole energy range of x-rays 50 ev

More information

ED(P)XRF: SCREENING ANALYSIS AND QUANTITATIVE ANALYSIS with POLARIZED

ED(P)XRF: SCREENING ANALYSIS AND QUANTITATIVE ANALYSIS with POLARIZED 384 ED(P)XRF: SCREENING ANALYSIS AND QUANTITATIVE ANALYSIS with POLARIZED X-RAYS R. Schramm, J. Heckel, K. Molt 2 Spectra Analytical Instruments, Kleve, Germany. 2Gerhard-Mercator- University Duisburg,

More information

Quantitative XRF Analysis. algorithms and their practical use

Quantitative XRF Analysis. algorithms and their practical use Joint ICTP-IAEA School on Novel Experimental Methodologies for Synchrotron Radiation Applications in Nano-science and Environmental Monitoring Quantitative XRF Analysis algorithms and their practical use

More information

EDS User School. Principles of Electron Beam Microanalysis

EDS User School. Principles of Electron Beam Microanalysis EDS User School Principles of Electron Beam Microanalysis Outline 1.) Beam-specimen interactions 2.) EDS spectra: Origin of Bremsstrahlung and characteristic peaks 3.) Moseley s law 4.) Characteristic

More information

In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer*

In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer* In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer* Anthony Thomas 1, Joachim Heckel 1, and Dirk Wissmann 1 Search and Discovery Article #41836 (2016)

More information

CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE

CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE 218 Chris M. Sparks 1, Elizabeth P. Hastings 2, George J. Havrilla 2, and Michael Beckstead 2 1. ATDF,

More information

OXEA - Online Elemental Analyzer

OXEA - Online Elemental Analyzer 02 25 08 OXEA - Online Elemental Analyzer OXEA (Online X-ray Elemental Analyzer) is based on the X-ray fluorescence technology (XRF) which is well known in the laboratory field. With the aid of a patented

More information

In Situ High-Temperature Study Of Silver Behenate Reduction To Silver Metal Using Synchrotron Radiation

In Situ High-Temperature Study Of Silver Behenate Reduction To Silver Metal Using Synchrotron Radiation Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 371 In Situ High-Temperature Study Of Silver Behenate Reduction To Silver Metal Using Synchrotron

More information

Electron-Induced X-Ray Intensity Ratios of Pb Lα/Lβ and As Kα/Kβ by kev Applied Voltages

Electron-Induced X-Ray Intensity Ratios of Pb Lα/Lβ and As Kα/Kβ by kev Applied Voltages Electron-Induced X-Ray Intensity Ratios of Pb Lα/Lβ and As Kα/Kβ by 18-30 kev Applied Voltages Bolortuya DAMDINSUREN and Jun KAWAI Department of Materials Science and Engineering, Kyoto University Sakyo-ku,

More information

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN 316 Effective atomic number of composite materials by Compton scattering - nondestructive evaluation method Kiran K U a, Ravindraswami K b, Eshwarappa K M a and Somashekarappa H M c* a Government Science

More information

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy Topic 2b: X-ray Fluorescence Spectrometry Text: Chapter 12 Rouessac (1 week) 4.0 X-ray Fluorescence Download, read and understand EPA method 6010C ICP-OES Winter 2009 Page 1 Atomic X-ray Spectrometry Fundamental

More information

LAB 01 X-RAY EMISSION & ABSORPTION

LAB 01 X-RAY EMISSION & ABSORPTION LAB 0 X-RAY EMISSION & ABSORPTION REPORT BY: TEAM MEMBER NAME: Ashley Tsai LAB SECTION No. 05 GROUP 2 EXPERIMENT DATE: Feb., 204 SUBMISSION DATE: Feb. 8, 204 Page of 3 ABSTRACT The goal of this experiment

More information

ION-EXCHANGE FILMS FOR ELEMENT CONCENTRATION IN X-RAY FLUORESCENCE ANALYSIS WITH TOTAL REFLECTION OF THE PRIMARY BEAM.

ION-EXCHANGE FILMS FOR ELEMENT CONCENTRATION IN X-RAY FLUORESCENCE ANALYSIS WITH TOTAL REFLECTION OF THE PRIMARY BEAM. 822 ION-EXCHANGE FILMS FOR ELEMENT CONCENTRATION IN X-RAY FLUORESCENCE ANALYSIS WITH TOTAL REFLECTION OF THE PRIMARY BEAM. Abstract A.P.Morovov, L.D.Danilin, V.V.Zhmailo, Yu.V.Ignatiev, A.E.Lakhtikov,

More information

Electron probe microanalysis - Electron microprobe analysis EPMA (EMPA) What s EPMA all about? What can you learn?

Electron probe microanalysis - Electron microprobe analysis EPMA (EMPA) What s EPMA all about? What can you learn? Electron probe microanalysis - Electron microprobe analysis EPMA (EMPA) What s EPMA all about? What can you learn? EPMA - what is it? Precise and accurate quantitative chemical analyses of micron-size

More information

MICRO-TOMOGRAPHY AND X-RAY ANALYSIS OF GEOLOGICAL SAMPLES

MICRO-TOMOGRAPHY AND X-RAY ANALYSIS OF GEOLOGICAL SAMPLES THE PUBLISHING HOUSE PROCEEDINGS OF THE ROMANIAN ACADEMY, Series A, OF THE ROMANIAN ACADEMY Volume 18, Number 1/2017, pp. 42 49 MICRO-TOMOGRAPHY AND X-RAY ANALYSIS OF GEOLOGICAL SAMPLES Ion GRUIA University

More information

STRESS ANALYSIS USING BREMSSTRAHLUNG RADIATION

STRESS ANALYSIS USING BREMSSTRAHLUNG RADIATION Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 106 STRESS ANALYSIS USING BREMSSTRAHLUNG RADIATION F. A. Selim 1, D.P. Wells 1, J. F. Harmon 1,

More information

X-Ray Fluorescence and Natural History

X-Ray Fluorescence and Natural History X-Ray Fluorescence and Natural History How XRF Helps XRF can be used both quantitatively (homogenous samples) and quantitatively (heterogenous samples).! Trace elements in a fossil can help identify source,

More information

AN EXAFS STUDY OF PHOTOGRAPHIC DEVELOPMENT IN THERMOGRAPHIC FILMS

AN EXAFS STUDY OF PHOTOGRAPHIC DEVELOPMENT IN THERMOGRAPHIC FILMS 96 AN EXAFS STUDY OF PHOTOGRAPHIC DEVELOPMENT IN THERMOGRAPHIC FILMS T. N. Blanton 1, D.R Whitcomb 2, and S.T. Misture 3 1 Eastman Kodak Company, Kodak Research Laboratories, Rochester, NY 14650-2106,

More information

Horst Ebel, Robert Svagera, Christian Hager, Maria F.Ebel, Christian Eisenmenger-Sittner, Johann Wernisch, and Michael Mantler

Horst Ebel, Robert Svagera, Christian Hager, Maria F.Ebel, Christian Eisenmenger-Sittner, Johann Wernisch, and Michael Mantler DETECTION OF SUBMONOLAYERS BY MEASUREMENT OF THE TOTAL ELECTRON YIELD (TEY) OF X-RAY EXCITED ELECTRON EMISSION Horst Ebel, Robert Svagera, Christian Hager, Maria F.Ebel, Christian Eisenmenger-Sittner,

More information

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis MT-0.6026 Electron microscopy Scanning electron microscopy and electron probe microanalysis Eero Haimi Research Manager Outline 1. Introduction Basics of scanning electron microscopy (SEM) and electron

More information

REALIZATION OF AN ASYMMETRIC MULTILAYER X-RAY MIRROR

REALIZATION OF AN ASYMMETRIC MULTILAYER X-RAY MIRROR Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 218 REALIZATION OF AN ASYMMETRIC MULTILAYER X-RAY MIRROR S. M. Owens Laboratory for High Energy Astrophysics,

More information

MSE 321 Structural Characterization

MSE 321 Structural Characterization Auger Spectroscopy Auger Electron Spectroscopy (AES) Scanning Auger Microscopy (SAM) Incident Electron Ejected Electron Auger Electron Initial State Intermediate State Final State Physical Electronics

More information

1 of 5 14/10/ :21

1 of 5 14/10/ :21 X-ray absorption s, characteristic X-ray lines... 4.2.1 Home About Table of Contents Advanced Search Copyright Feedback Privacy You are here: Chapter: 4 Atomic and nuclear physics Section: 4.2 Absorption

More information

Pros and Cons of Water Analysis Methods

Pros and Cons of Water Analysis Methods Water Lens, LLC 4265 San Felipe, Suite 1100 Houston, Texas 77027 Office: (844) 987-5367 www.waterlensusa.com Pros and Cons of Water Analysis Methods Prepared by: Adam Garland, CTO Water Lens, LLC ICP-MS/OES

More information

AN IMPROVED AEROSOL GENERATION SYSTEM FOR THE PREPARATION OF XRF CALIBRATION FILTERS

AN IMPROVED AEROSOL GENERATION SYSTEM FOR THE PREPARATION OF XRF CALIBRATION FILTERS Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 449 AN IMPROVED AEROSOL GENERATION SYSTEM FOR THE PREPARATION OF XRF CALIBRATION FILTERS C. Vanhoof, V.

More information

QUANTITATIVE ENERGY-DISPERSIVE ELECTRON PROBE X-RAY MICROANALYSIS OF INDIVIDUAL PARTICLES

QUANTITATIVE ENERGY-DISPERSIVE ELECTRON PROBE X-RAY MICROANALYSIS OF INDIVIDUAL PARTICLES QUANTITATIVE ENERGY-DISPERSIVE ELECTRON PROBE X-RAY MICROANALYSIS OF INDIVIDUAL PARTICLES 287 Chul-Un Ro Department of Chemistry, Inha University 253, Yonghyun-dong, Nam-gu, Incheon 402-751, Korea ABSTRACT

More information

An Analysis of Secondary Enhancement Effects in Quantitative XRFA

An Analysis of Secondary Enhancement Effects in Quantitative XRFA An Analysis of Secondary Enhancement Effects in Quantitative XRFA Michael Mantler Institut fur Angewandte und Technische Physik Vienna University of Technology, Vienna, Austria Secondary enhancement effects

More information

X-ray Absorption Spectroscopy

X-ray Absorption Spectroscopy X-ray Absorption Spectroscopy Nikki Truss November 26, 2012 Abstract In these experiments, some aspects of x-ray absorption spectroscopy were investigated. The x-ray spectrum of molybdenum was recorded

More information

PHYS General Physics II Lab The Balmer Series for Hydrogen Source. c = speed of light = 3 x 10 8 m/s

PHYS General Physics II Lab The Balmer Series for Hydrogen Source. c = speed of light = 3 x 10 8 m/s PHYS 1040 - General Physics II Lab The Balmer Series for Hydrogen Source Purpose: The purpose of this experiment is to analyze the emission of light from a hydrogen source and measure and the wavelengths

More information

DATA MINING WITH DIFFERENT TYPES OF X-RAY DATA

DATA MINING WITH DIFFERENT TYPES OF X-RAY DATA DATA MINING WITH DIFFERENT TYPES OF X-RAY DATA 315 C. K. Lowe-Ma, A. E. Chen, D. Scholl Physical & Environmental Sciences, Research and Advanced Engineering Ford Motor Company, Dearborn, Michigan, USA

More information

Portable type TXRF analyzer: Ourstex 200TX

Portable type TXRF analyzer: Ourstex 200TX Excerpted from Adv. X-Ray. Chem. Anal., Japan: 42, pp. 115-123 (2011) H. Nagai, Y. Nakajima, S. Kunimura, J. Kawai Improvement in Sensitivity and Quantification by Using a Portable Total Reflection X-Ray

More information

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy Hiroshi Onodera Application & Research Center, JEOL Ltd. Introduction um, PBB and PBDE) are subject to usage restrictions in Europe.

More information

Chemical State Analysis of SiO 2 /Si by Wavelength-Dispersive X-Ray Fluorescence

Chemical State Analysis of SiO 2 /Si by Wavelength-Dispersive X-Ray Fluorescence Chemical State Analysis of SiO 2 /Si by Wavelength-Dispersive X-Ray Fluorescence Shinji OZAKI, Matsushita Technoresearch Inc. ozaki.s@jp.panasonic.com The chemical states of a SiO 2 /Si govern the conductivity

More information

XRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures:

XRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures: 1 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures: XRF spectroscopy 1 exam question. Notes on: www.nuigalway.ie/nanoscale/3rdspectroscopy.html XRF books: Analytical Chemistry,

More information

FINDING DESCRIPTORS USEFUL FOR DATA MINING IN THE CHARACTERIZATION DATA OF CATALYSTS

FINDING DESCRIPTORS USEFUL FOR DATA MINING IN THE CHARACTERIZATION DATA OF CATALYSTS Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 338 FINDING DESCRIPTORS USEFUL FOR DATA MINING IN THE CHARACTERIZATION DATA OF CATALYSTS C. K. Lowe-Ma,

More information

EUV-Technology with Discharge EUV-Lamp"

EUV-Technology with Discharge EUV-Lamp EUV-Technology with Discharge EUV-Lamp" Rainer Lebert, Larissa Juschkin, Christian Wies, Bernhard Jägle, Manfred Meisen, Ulrich Bieberle, Willi Neff, Juri Barthel, Konstantin Walter, Klaus Bergmann, Fraunhofer

More information

X-rays. X-ray Radiography - absorption is a function of Z and density. X-ray crystallography. X-ray spectrometry

X-rays. X-ray Radiography - absorption is a function of Z and density. X-ray crystallography. X-ray spectrometry X-rays Wilhelm K. Roentgen (1845-1923) NP in Physics 1901 X-ray Radiography - absorption is a function of Z and density X-ray crystallography X-ray spectrometry X-rays Cu K α E = 8.05 kev λ = 1.541 Å Interaction

More information

Automated Determination of Crystal Reflectivity in the X-ray Laboratory

Automated Determination of Crystal Reflectivity in the X-ray Laboratory Automated Determination of Crystal Reflectivity in the X-ray Laboratory Bradley Wideman University of Rochester Laboratory for Laser Energetics 2008 High School Summer Research Program INTRODUCTION The

More information

Introduction to Fourier Transform Infrared Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy molecular spectroscopy Introduction to Fourier Transform Infrared Spectroscopy Part of Thermo Fisher Scientific Introduction What is FT-IR? FT-IR stands for Fourier Transform InfraRed, the preferred method

More information

2D XRD Imaging by Projection-Type X-Ray Microscope

2D XRD Imaging by Projection-Type X-Ray Microscope 0/25 National Institute for Materials Science,Tsukuba, Japan 2D XRD Imaging by Projection-Type X-Ray Microscope 1. Introduction - What s projection-type X-ray microscope? 2. Examples for inhomogeneous/patterned

More information

INTRODUCTION Atomic fluorescence spectroscopy ( AFS ) depends on the measurement of the emission ( fluorescence ) emitted from gasphase analyte atoms

INTRODUCTION Atomic fluorescence spectroscopy ( AFS ) depends on the measurement of the emission ( fluorescence ) emitted from gasphase analyte atoms INTRODUCTION Atomic fluorescence spectroscopy ( AFS ) depends on the measurement of the emission ( fluorescence ) emitted from gasphase analyte atoms that have been excited to higher energy levels by absorption

More information

Imaging Methods: Scanning Force Microscopy (SFM / AFM)

Imaging Methods: Scanning Force Microscopy (SFM / AFM) Imaging Methods: Scanning Force Microscopy (SFM / AFM) The atomic force microscope (AFM) probes the surface of a sample with a sharp tip, a couple of microns long and often less than 100 Å in diameter.

More information

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu X-ray Photoelectron Spectroscopy Introduction Qualitative analysis Quantitative analysis Charging compensation Small area analysis and XPS imaging

More information

Multi-element process analyzer

Multi-element process analyzer Multi-element process analyzer Elemental analysis by X-ray fluorescence Compact multi-element process analyzer for liquid streams or f ixed position web applications Featuring advanced third generation

More information

D y n a m a r. Polymer Processing Additives. Analysis of Dynamar Polymer Processing Additives in Polyolefins Using X-ray Fluorescence Spectrometry

D y n a m a r. Polymer Processing Additives. Analysis of Dynamar Polymer Processing Additives in Polyolefins Using X-ray Fluorescence Spectrometry D y n a m a r Polymer Processing Additives Analysis of Dynamar Polymer Processing Additives in Polyolefins Using X-ray Fluorescence Spectrometry Introduction Preparation of Calibration Standards and Samples

More information

Reversible uptake of HgCl 2 in a porous coordination polymer based on the dual functions of carboxylate and thioether

Reversible uptake of HgCl 2 in a porous coordination polymer based on the dual functions of carboxylate and thioether Supplementary Information Reversible uptake of HgCl 2 in a porous coordination polymer based on the dual functions of carboxylate and thioether Xiao-Ping Zhou, a Zhengtao Xu,*,a Matthias Zeller, b Allen

More information

X-Ray Emission and Absorption

X-Ray Emission and Absorption X-Ray Emission and Absorption Author: Mike Nill Alex Bryant February 6, 20 Abstract X-rays were produced by two bench-top diffractometers using a copper target. Various nickel filters were placed in front

More information

X-RAY EMISSION WAVELENGTHS AND KEV TABLES FOR NONDIFFRACTIVE ANALYSIS

X-RAY EMISSION WAVELENGTHS AND KEV TABLES FOR NONDIFFRACTIVE ANALYSIS X-RAY EMISSION WAVELENGTHS AND KEV TABLES FOR NONDIFFRACTIVE ANALYSIS Prepared by G. G. Johnson, Jr., and E. W. White ASTM Data Series DS 46 Price $5.00 AMERICAN SOCIETY FOR TESTING AND MATERIALS 1916

More information

MCSHAPE: A MONTE CARLO CODE FOR SIMULATION OF POLARIZED PHOTON TRANSPORT

MCSHAPE: A MONTE CARLO CODE FOR SIMULATION OF POLARIZED PHOTON TRANSPORT Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 363 MCSHAPE: A MONTE CARLO CODE FOR SIMULATION OF POLARIZED PHOTON TRANSPORT J.E. Fernández, V.

More information

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 19 Chapter 12 There are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry X-ray Techniques include:

More information

Ziessel a* Supporting Information (75 pages) Table of Contents. 1) General Methods S2

Ziessel a* Supporting Information (75 pages) Table of Contents. 1) General Methods S2 S1 Chemistry at Boron: Synthesis and Properties of Red to Near-IR Fluorescent Dyes based on Boron Substituted Diisoindolomethene Frameworks Gilles Ulrich, a, * Sebastien Goeb a, Antoinette De Nicola a,

More information

QUANTIFICATION OF MERCURY, LEAD, AND CADMIUM IN AQUEOUS SOLUTIONS BY ENERGY K X-RAY FLUORESCENCE SPECTROSCOPY

QUANTIFICATION OF MERCURY, LEAD, AND CADMIUM IN AQUEOUS SOLUTIONS BY ENERGY K X-RAY FLUORESCENCE SPECTROSCOPY Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 505 QUANTIFICATION OF MERCURY, LEAD, AND CADMIUM IN AQUEOUS SOLUTIONS BY ENERGY K X-RAY FLUORESCENCE

More information

Improvements for Absorption Spectroscopy at Beamlines A1, E4, X1

Improvements for Absorption Spectroscopy at Beamlines A1, E4, X1 Improvements for Absorption Spectroscopy at Beamlines A1, E4, X1 U. Brüggmann 1, N. Haack, M. Herrmann 2, S.K.J. Johnas 3, P. Kappen, K. Klementiev 4, E. Welter For an improvement of the conditions of

More information

CHARACTERIZATION OF Pu-CONTAINING PARTICLES BY X-RAY MICROFLUORESCENCE

CHARACTERIZATION OF Pu-CONTAINING PARTICLES BY X-RAY MICROFLUORESCENCE Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 534 CHARACTERIZATION OF Pu-CONTAINING PARTICLES BY X-RAY MICROFLUORESCENCE Marco Mattiuzzi, Andrzej Markowicz,

More information