STATISTIQUE DE PERFORMANCE

Size: px
Start display at page:

Download "STATISTIQUE DE PERFORMANCE"

Transcription

1 INTERCOMPARISON OF STANDARD CALIBRATION, METHODS AND MEASUREMENTS FOR PRESSURE, TEMPERATURE AND HUMIDITY BETWEEN RIC CASABLANCA, MOROCCO AND RIC Ljubljana SLOVENIA 3(10) The operators who participated in the calibration relating to this comparison are: - Mr. Mounir AZIZ: RIC Responsible, Casablanca Morocco - Mr. Ales Razpotnic: Pression Field, Slovenian Environment Agency - -Mr. Drago Groselj, RIC Responsible, Slovenian Environment Agency, drago.groselj@gov.si ABSTRACT A CIL (Comparison interlaboratory) of standard calibration instruments and methods for pressure, humidity and temperature was organised in 2015 between the laboratories of metrology of Meteo Morocco and the Slovenian Environment Agency. The two RIC laboratories are equipped and work identically in terms of equipments and methods. Both laboratories use the ISO 17025:2005 standard and have an unbroken chain of traceability of their standards. The results show and prove an agreement between the measurements from the laboratories for both pressure, temperature and humidity calibration.

2 1. Introduction 3 2. REFERENCE DOCUMENTS 3 3. Instrument used 3 4. Program and pressure calibration procedures: 3 5. Diagram measurement benches mounting pressure: 3 6. RESULALTS: Uncertainties: Drift values obtained from testing: 5 7. STATISTIQUE DE PERFORMANCE 6 8. ERRORS DUE TO TRANSPORTATION : 7 9. Conclusion: 7

3 1. INTRODUCTION This paper presents the results of the comparison inter laboratory for calibration of barometers, between the laboratories of Moroccan Meteorology Department and Slovenian Environment Agency, performed on the occasion of Expert Team on Operational Metrology (ET OM) meeting in Ljubljana, Slovenia, from 1st 4th December 2015, hosted by Regional Instrument Center of the Slovenian Environment Agency. The first calibrations were made in Morocco on November 30, 2015, and return on March 1, 2016, the calibration by Slovenian Environment Agenc, took place on 3 and 4 December Interlaboratory comparison (ILC) serves as a tool for comparison of measurement results carried out by both calibration laboratories in the field of Pressure, temperature and humidity measurement. ILC represents the best means to demonstrate technical competence of the participant and also serves as a part of technical bases for accreditation, it is the most important element for monitoring of quality of measurement results as required by ISO/IEC 17025:2005 standard for laboratories in part REFERENCE DOCUMENTS The certificates produced or used in this comparison are: The calibration certificate No Labo. Slovenia and the certificate No. CEP15170 Labo. Morocco. The labo. Morocco joined his traceability certificate (Cal Cert ID: 121,595 in the accredited Mensor laboratory Cert ZL. 3. INSTRUMENT USED The instrument calibrated by both laboratories is: In Calibration pressure: Display PTB 330 barometer (VAISALA), Serial C PROGRAM AND PRESSURE CALIBRATION PROCEDURES: The laboratory of Slovenian Environment Agency calibrated on its own calibration bench with a standard traceable to the international system, the calibration was done in 11 points between 850 hpa and 1100 hpa. Three down climb cycles were applied for 2 Slovénen and Moroccan laboratories, with bearings 2500 Pa. The Morocco Laboratory calibrate the barometer compared to its generator and reference standard "MENSOR CPC 6000" in 11 points from Pascal to pascal. 5. DIAGRAM MEASUREMENT BENCHES MOUNTING PRESSURE: Installation diagrams of two laboratories are fairly similar, but with two different brands.

4 calibration bench assembly diagram of Moroccan laboratory, with integrated standard automatic generator CPC6000 POMPE PC Baromètre réf : pression PPC1 Générateur A Z O T E Liaison série Circuit pneumatique Alimentation secteur Distributeur de pression calibration bench assembly diagram of Slovenian laboratory Baromètre à étalonner 6. RESULALTS:

5 The results of calibrations used for intercomparison, are averages of three cycles of correction term measures, accompanied by the uncertainty calculated for each calibration laboratory and according to the means used. The ambient conditions in laboratory of Slovenia is 22 C to 25.5 C with a relative humidity between 25% and 32%; and in the laboratory of Morocco is 23 C to 23.6 C Uncertainties: Both laboratories express calibration uncertainty with a coverage factor k = 2 And take into account four important components: Uncertainty calibration means Uncertainty due to the repeatability of the sensor at 3 cycles, The maximum hysteresis observed for all of the bearings. The uncertainty due to the resolution of the instrument under test Laboratory of Slovenian Environment Agency declared uncertainties are set at the maximum pressure used = 1100 hpa. This method simplifie the use of uncertainty whereas the range of maximum pressure measurements (850 to 1100 hpa) does not justify the method of polynomial calculation and the discrepancies between the low pressure and considered high pressure, are small compared to the overall uncertainty. Details uncertainties determined by the two laboratories: Laboratory Slovenia ( Pa) Laboratory Morocco ( Pa) u -moyens / 2 u -moyens / 2 U= 1,5 Pa + 3,3 *10-5 * P u- répétabilité u- répétabilité 3 u_ hyster ptb u_ hyster ptb 0.005/ 3=0,289 u_ resolution ptb u_ resolution ptb 0.005/ 3=0,289 U (k=2) = 8 Pa U(k=2) = 6 Pa + 2*10-3 P soit au max. = 8,2 Pa Slovénie : 2* Root (u -moyens 2 +u -répétabilité2 +u _hyster_ptb 2 +u _resolution_ptb2 ) U = 8 Pa Maroc : 2*Root (u -moyens 2 +u -répétabilité 2 +u _hyster_ptb 2 +u _resolution_ptb 2 ) U = (6 Pa +2*10-3 P) = 8,2 Pa The error means the maximum difference observed between the standard and the test sensor. Tolerance or compliance of customer, declared by the Moroccan Laboratory is 0,5 mbar or 50 Pa The decision criterion is "the greater error (E Accuracy, Fidelity E, E hysteresis) increased the uncertainty (U) must be less than the specified tolerance. Ei + U <(T / 2), T = customer tolerance Similarly, the adjustment criterion in test barometers established by the formula Error + U <( C / 2), C = Customer Compliance = T 6.2. Drift values obtained from testing:

6 All the points of steps executed by Laboratory Slovenia and Laboratory Morocco is reported in this table, where measurements are made using the same bearings are analyzed. SETPOINTS (Pa) UNCERTAINTY U (Pa) Morocco Slovenia U Maroc U Slovénie Réf = Pe Pe-Pi Réf = Pe Pe-Pi , , , , , , , , , , , , , , , ,2 8 Comparative curves in Pascal Note that the sensor used, presents a gap between 32 and 33Pa for the values obtained by Laboratory Slovenia and 30.5 and 36 Pa to the values obtained by Laboratory Morocco. The uncertainty bars are representative of U declared 8 Pa Laboratory Slovenia and between 7.7 and 8.2 for Laboratory Morocco, these error bars show the overlay measurements of this sensor by 2 calibration methods. To this end, it seems necessary to calculate the standard deviation of all measurements, to check whether the inter comparison pressure is considered satisfactory. 7. STATISTIQUE DE PERFORMANCE Performance statistic used in the context of this intercomparison is the standard deviation "EN"; described according to DIN EN ISO / IEC 17043: It is obtained as follows: En M S / (U M U S )

7 X M and X S are the results, respectively, error indications or corrections of participating laboratory from morocco and slovenia. U Labi et U Labp are respectively the expanded uncertainties of participating laboratory from morocco and slovenia. ERRORS MOROCCO SLOVENIA Ecarts = M U Ecarts = S U EN Maroc Slovenie 34 7, , , ,09 33,5 7, , , ,09 31,5 7, ,04 30,5 7, , ,00 33, , , ,26 34,5 8, , , ,26 8. ERRORS DUE TO TRANSPORTATION : SETPOINTS DERIVE /REF DERIVE/REF BEFORE AFTER DRIFT UP DOWN UP DOWN UP DOWN For the travel of the sensor, calibration report shows excellent retention of the equipment after transportation, the maximum deviation is 5 Pascal between calibrations values before and after transport. 9. CONCLUSION:

8 RESULTATS DE L INTER COMPARAISON PRESSION EN-S Version: A Date: 14/08/2013 In this case comparing two laboratories and in particular for the pressure calibration method, it is found that the values obtained by the two laboratories have a superposition inducing a good agreement. The standard deviation is well below the limit value of 1. The comparison is satisfactory. Both laboratories have unbroken traceability chains for humidity and pressure and their calibration results are in good agreement. Full results will be available at TECO2016.

STATUS OF THE WIGOS DEMONSTRATION PROJECTS

STATUS OF THE WIGOS DEMONSTRATION PROJECTS STATUS OF THE WIGOS DEMONSTRATION PROJECTS Demonstration Project Morocco Strengthening Moroccan RIC Capacities (Submitted by Rabia Merrouchi, National Meteorological Service of Morocco (DMN)) February

More information

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands Calibration, traceability, uncertainty and comparing measurement results Workshop 16 March 2015 VSL, Delft The Netherlands 1 Calibration Calibrating an instrument involves comparing the measurements of

More information

World Meteorological Organization

World Meteorological Organization World Meteorological Organization CIMO/WIGOS EXPLORATORY WORKSHOP: IMPROVING SURFACE-BASED DATA QUALITY THOUGH IMPROVED STANDARDIZATION OF PRACTICES AND PROCEDURES Langen, Germany, 5-3 December 2014 Regional

More information

RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS

RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS The Fifth International Proficiency Testing Conference Timisoara, Romania () th 8 th September, 0 RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS Adriana Vâlcu, Sterică Baicu National Institute of Metrology,

More information

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Quality assurance for sensors at the Deutscher Wetterdienst (DWD) Paper submitted to ICAWS 2017: Topic 3 Sustainability of the measurements: Calibration, intercomparisons, laboratory and field performance tests, quality assurance and control assessment for traceable

More information

CALIBRATION CERTIFICATE # 503

CALIBRATION CERTIFICATE # 503 CALIBRATION CERTIFICATE # 503 Calibration date : 2011-02-23 Certificate issued : 2012-04-18 Company name Company address City, Province, Canada CLAS 2009-02 Calibration of Mass flow meter Micro Motion

More information

Interlaboratory Comparison in the field of Temperature, Humidity and Pressure, in the WMO Regional Association VI (MM-ILC-2015-THP)

Interlaboratory Comparison in the field of Temperature, Humidity and Pressure, in the WMO Regional Association VI (MM-ILC-2015-THP) Instruments and Observing Methods Report No. 128 WEATHER CLIMATE WATER Interlaboratory Comparison in the field of Temperature, Humidity and Pressure, in the WMO Regional Association VI (MM-ILC-2015-THP)

More information

Introduction of RIC Tsukuba (Japan, RAII)

Introduction of RIC Tsukuba (Japan, RAII) Introduction of RIC Tsukuba (Japan, RAII) 20 March, 2018 Kouichi NAKASHIMA Scientific Officer Regional Instrument Centre Tsukuba Observing Division, Observing Department Outline 1. Overview of Organization

More information

CALIBRATION REPORT FOR THERMOHYDROMETER

CALIBRATION REPORT FOR THERMOHYDROMETER CALIBRATION REPORT FOR THERMOHYDROMETER SAMPLE CUSTOMER S/N: XXXXX Report Number: Y205188 Page 1 of 5 Report No. Y205188 ICL CALIBRATION LABORATORIES, INC. ISO/IEC 17025 and ANSI/NCSL Z540-1 accredited

More information

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Quality assurance for sensors at the Deutscher Wetterdienst (DWD) Quality assurance for sensors at the Deutscher Wetterdienst (DWD) Quality assurance / maintenance / calibration Holger Dörschel, Dr Tilman Holfelder WMO International Conference on Automatic Weather Stations

More information

Metrology and Environment

Metrology and Environment Bridging the Gap Responding to Environmental Change: from Words to Deeds 14-16 May 2008, Portorož Metrology and Environment Prof Janko Drnovšek and Laboratory group Janko.Drnovsek@fe.uni-lj.si Basic dilemma

More information

CERTIFICATE OF CALIBRATION

CERTIFICATE OF CALIBRATION CERTIFICATE OF CALIBRATION 12100 WEST 6TH AVE LAKEWOOD, CO 80228 ANAB AC-2489.02 Manufacturer: Ruska Instruments Corp. Model Number: 7250i : Pressure Calibrator Serial Number: 69718 ID: TE10768 As-Found:

More information

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement Publication Reference EA-10/11 EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical PURPOSE This document has been produced by EA as a means of giving advice for calibrating

More information

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS W. Beyer and W. Pahl Physikalisch-Technische Bundesanstalt (PTB) 38116 Braunschweig, Germany Abstract: In accordance with ISO 17025,

More information

INTER-COMPARISONON TEMPERATURE MEASUREMENT USING FLUKE THERMOMETER MODEL 1502A

INTER-COMPARISONON TEMPERATURE MEASUREMENT USING FLUKE THERMOMETER MODEL 1502A INTER-COMPARISONON TEMPERATURE MEASUREMENT USING FLUKE THERMOMETER MODEL 1502A Drs. Untung Merdijanto M.Si, Drs. Damianus Tri Heryanto, Drs. Kanton Lumban Toruan, M.Si, and Budi Santoso Agency for Meteorology,

More information

OverviewofRIC Tsukuba (Japan,RAII )

OverviewofRIC Tsukuba (Japan,RAII ) OverviewofRIC Tsukuba (Japan,RAII ) 7Nov.2016 Hiroshi Yoshimura Scientific Officer Regional Instrument Centre Tsukuba Observing Division, Observing Department Outline 1. WMO Regional Instrument Center

More information

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure DATASHEET Pascal ET Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure - electrical signals (ma, mv, V, ) - temperature (TC and

More information

Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018

Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018 Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018 Guidelines on the Calibration of Automatic Instruments for

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

STATUS OF INSTRUMENTS TRACEABILITY IN NMHSs

STATUS OF INSTRUMENTS TRACEABILITY IN NMHSs WORLD METEOROLOGICAL ORGANIZATION COMMISSION FOR INSTRUMENTS AND METHODS OF OBSERVATION OPAG CAPACITY BUILDING CIMO EXPERT TEAM ON REGIONAL INSTRUMENT CENTRES, QUALITY MANAGEMENT SYSTEMS AND COMMERCIAL

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 CONTROL COMPANY 12554 Galveston Road Suite B230 Webster, TX 77598 Aaron Judice Phone: 281 482 1714 CALIBRATION Valid To: March 31, 2018

More information

Test report CVN Project number Page 2 of 2

Test report CVN Project number Page 2 of 2 Project number 604545 Page 2 of 2 Tests : Results : The diaphragm gas meters as specified in annex 2 were tested for compliance with the European Standard BS EN 1359 : 1999 "Gas meters - Diaphragm gas

More information

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC 17025 Table of contents 1 GENERAL... 2 2 THE STANDARD SIST EN ISO/IEC 17025... 2 3 SA'S POLICY IN IMPLEMENTING

More information

SAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0

SAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0 076-0-067-ACC-USL-S Mettler Toledo, LLC 900 Polaris Parkway Columbus, OH 0.800.METTLER Customer Accredited by the American Association for Laboratory Accreditation (ALA) CALIBRATION CERT #788.0 SAMPLE

More information

Measurement Uncertainty in Mechanical Testing

Measurement Uncertainty in Mechanical Testing Intelligent testing Measurement Uncertainty in Mechanical Testing 27. testxpo 2018 ZwickRoell GmbH & Co. KG Ulm Dr. Eduard Schenuit Industry Manager Metal Agenda Why do we need measurement uncertainty?

More information

Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C

Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C DOI 10.1007/s10765-014-1687-0 Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C Domen Hudoklin Regina Mnguni Hans Liedberg Igor Pušnik Jovan Bojkovski

More information

Vocabulary of Metrology

Vocabulary of Metrology Vocabulary of Metrology For Understating Uncertainty and Traceability Prepared by: Kim, Sang Ho Engineering Team Leader Modal Shop, Inc A PCB Group Company SI National Metrology Institute Secondary Calibration

More information

SIM AUV.A-K1.prev Microphone Intercomparison

SIM AUV.A-K1.prev Microphone Intercomparison Participants: Report on SIM AUV.A-K1.prev Microphone Intercomparison by Knud Rasmussen, DPLA CENAM, División de Vibraciones y Acústica km 4,5 Carretera a los Cués C.P. 76241 Querétaro, México DPLA, Danish

More information

Analysis of a national low dc resistance inter-laboratories comparison

Analysis of a national low dc resistance inter-laboratories comparison 16 th International Congress of Metrology, 04008 (2013) DOI: 10.1051/ metrology/201304008 C Owned by the authors, published by EDP Sciences, 2013 Analysis of a national low dc resistance inter-laboratories

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 PROTEMP MECHANICAL, INC. 3350 Scott Blvd, Bldg. #3 Santa Clara, CA 95054 Martin L. Reich Phone: 408 244 9821 CALIBRATION Valid To: August

More information

Meteorological Service

Meteorological Service Meteorological Service The Meteorological Service in Oman was established in 1973. Oman joined ICAO in 1973 Oman joined WMO in 1975. Meteorological Service is under the framework of the Directorate General

More information

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations Calibration General Principles & Theory, Equipment Considerations Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 FLORIDA METROLOGY, LLC 645 NW Enterprise Dr. Port St. Lucie FL 34986 Ron Kupper Phone: (772)212-7158 CALIBRATION Valid To: October 31,

More information

Metallic materials Brinell hardness test. Part 3: Calibration of reference blocks

Metallic materials Brinell hardness test. Part 3: Calibration of reference blocks INTERNATIONAL STANDARD ISO 6506-3 Third edition 2014-10-01 Metallic materials Brinell hardness test Part 3: Calibration of reference blocks Matériaux métalliques Essai de dureté Brinell Partie 3: Étalonnage

More information

Calibration and verification: Two procedures having comparable objectives and results

Calibration and verification: Two procedures having comparable objectives and results Calibration and verification: Two procedures having comparable objectives and results KLAUS-DIETER SOMMER, Landesamt für Messund Eichwesen Thüringen (LMET), Germany SAMUEL E. CHAPPELL, Consultant, Formerly

More information

Interlaboratory Comparison & Proficiency Testing Program

Interlaboratory Comparison & Proficiency Testing Program Extended scope for the organization of proficiency tests in accordance with the requirements of ISO/IEC 17043 VSL, the Dutch Metrology Institute, has significantly extended the scope for their accreditation

More information

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT Scott Crone North American Sales Manager AMETEK Calibration Instruments 8600 Somerset Drive Largo, Florida 33773 USA INTRODUCTION Like any other piece of equipment,

More information

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1 Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1 Updated at 12-December-2012 Task and Purpose of the Comparison According to the rules set up by the CIPM

More information

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE Publication Reference EA-10/14 EA Guidelines on the Calibration of Static Torque Measuring Devices PURPOSE This document has been produced by EA to improve harmonisation in determining the calibration

More information

The Fundamentals of Moisture Calibration

The Fundamentals of Moisture Calibration The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 CLEVELAND SPECIALTY INSPECTION SERVICES, INC. 8562 East Avenue Mentor, OH 44060 James Popovic Phone: 440 974 1818 CALIBRATION Valid

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 EQUIPMENT SERVICES GROUP, LLC 3269 Hilton Rd. Ferndale, MI 48220 Michael Macauley Phone: 248 227 5310 CALIBRATION Valid To: December

More information

A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY Dr Jérôme DUVERNOY Météo-France DSO/LM 7 Rue Teisserenc de Bort 78195 TRAPPES France Phone +33130136350 Fax +33130136020 www.meteo.fr

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 RS CALIBRATION SERVICES, INC. 1047 Serpentine Lane Pleasanton, CA 94566 Mr. Ralph Sabiel Phone: 925 462 4217 CALIBRATION Valid To: April

More information

EURAMET Project no Inter-comparison of a 1000 L proving tank

EURAMET Project no Inter-comparison of a 1000 L proving tank EURAMET Project no. 1157 Inter-comparison of a 1000 L proving tank Final Report IPQ (Elsa Batista) Coordinator of the comparison VSL (Erik Smits) Pilot and coordinator of the comparison September 2013

More information

CERTIFICATE OF CALIBRATION

CERTIFICATE OF CALIBRATION Page 1 of 5 CERTIFICATE OF CALIBRATION 1K Thermistor CERTIFICATE NO.: MANUFACTURER: Measurement Specialties DATE OF CALIBRATION: 26-Apr-17 TC15249 CUSTOMER: AMBIENT TEMPERATURE: (21 ± 3) C JOB NUMBER:

More information

CONTENT 2. ORGANIZATION 3. SERVICES. instruments. (3)Activities of RIC Tsukuba

CONTENT 2. ORGANIZATION 3. SERVICES. instruments. (3)Activities of RIC Tsukuba 1 CONTENT 1. INTRODUCTION 2. ORGANIZATION 3. SERVICES (1)Quality assurance of meteorological instruments (2)Research and development (3)Activities of RIC Tsukuba 2 Meteorological Instruments Center (MIC)

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

UNCERTAINTY ANALYSIS FOR LABORATORY ACCREDITATION. Peter B. Crisp. Fluke Precision Measurement Ltd, 52 Hurricane Way, Norwich, UK

UNCERTAINTY ANALYSIS FOR LABORATORY ACCREDITATION. Peter B. Crisp. Fluke Precision Measurement Ltd, 52 Hurricane Way, Norwich, UK UNCERTAINTY ANALYSIS FOR LABORATORY ACCREDITATION Peter B. Crisp Fluke Precision Measurement Ltd, 5 Hurricane Way, Norwich, UK Introduction NAMAS Accredited laboratories are required to follow recommended

More information

SUB-REGIONAL PYRANOMETER INTERCOMPARISON FOR THE SOUTH-EASTERN EUROPE ABSTRACT

SUB-REGIONAL PYRANOMETER INTERCOMPARISON FOR THE SOUTH-EASTERN EUROPE ABSTRACT SUB-REGIOAL PYRAOMETER ITERCOMPARISO FOR THE SOUTH-EASTER EUROPE Krunoslav Premec Meteorological and Hydrological Service of Croatia, HR-10000 Zagreb, Gric 3, CROATIA Tel: + 385 1 4565 607, Fax: +385 1

More information

High pressure comparison among seven European national laboratories

High pressure comparison among seven European national laboratories IMEKO 20 th TC3, 3 rd TC16 and 1 st TC22 International Conference Cultivating metrological knowledge 27 th to 30 th November, 2007. Merida, Mexico. High pressure comparison among seven European national

More information

Document No: TR 12 Issue No: 1

Document No: TR 12 Issue No: 1 ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Prepared by: SADCAS Technical Manage Approved

More information

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements Experimental Design 8.2.2013 Doc. Martti Heinonen etunimi.sukunimi@mikes.fi Outline 1.

More information

CERTIFICATE OF CALIBRATION

CERTIFICATE OF CALIBRATION Page 1 of 10 CALIBRATION OF Sound Level Meter: Brüel & Kjær Type 3050-A-040 No: 3050-100751 Id: - Microphone: Brüel & Kjær Type 4189 No: 2621142 Preamplifier: Brüel & Kjær Type 2669 No: 2679073 Supplied

More information

Survey on alternatives for dangerous and obsolete instruments WMO - IOM 117

Survey on alternatives for dangerous and obsolete instruments WMO - IOM 117 VAISALA AWARD 2016 FOR THE DEVELOPMENT AND IMPLEMENTATION OF INSTRUMENTATION AND METHODS OF OBSERVATION IN DEVELOPING COUNTRIES Survey on alternatives for dangerous and obsolete instruments WMO - IOM 117

More information

Final Report August 2010

Final Report August 2010 Bilateral Comparison of 100 pf Capacitance Standards (ongoing BIPM key comparison BIPM.EM-K14.b) between the CMI, Czech Republic and the BIPM, January-July 2009 J. Streit** and N. Fletcher* *Bureau International

More information

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014) SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February

More information

Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz

Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range to 10 k June 016 Author: Co-author: Danuta Dobrowolska (GUM, pilot) Alexander Kosterov

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 F. D. HURKA COMPANY 4731 Stockholm Court Charlotte, NC 28273 Charles Jenkins Phone: 704 552 0008 CALIBRATION Valid To: December 31,

More information

FINAL REPORT. European co-operation for Accreditation (EA) Interlaboratory Comparison IR3: Calibration of a Radiation Protection Dosimeter

FINAL REPORT. European co-operation for Accreditation (EA) Interlaboratory Comparison IR3: Calibration of a Radiation Protection Dosimeter - 1 (59) - EA IR3 Final Report ENEA-INMRI SIT Servizio di Taratura in Italia FINAL REPORT European co-operation for Accreditation (EA) Interlaboratory Comparison IR3: Calibration of a Radiation Protection

More information

Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar

Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) Dr. Steven D. Phillips Precision Engineering Division National Institute

More information

Withdrawn at December 31, Guideline DKD-R 5-7. Calibration of Climatic Chambers DEUTSCHER KALIBRIERDIENST

Withdrawn at December 31, Guideline DKD-R 5-7. Calibration of Climatic Chambers DEUTSCHER KALIBRIERDIENST DEUTSCHER KALIBRIERDIENST Guideline DKD-R 5-7 Calibration of Climatic Chambers Edition 07/2004 English translation 02/2009 Page 1 of 31 Published by the Accreditation Body of the Deutscher Kalibrierdienst

More information

Interlaboratory Comparison & Proficiency Testing Program

Interlaboratory Comparison & Proficiency Testing Program Extended scope for the organization of proficiency tests in accordance with the requirements of ISO/IEC 17043 VSL, the Dutch Metrology Institute, has significantly extended the scope for their accreditation

More information

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017 Metrology Principles for Earth Observation: the NMI view Emma Woolliams 17 th October 2017 Interoperability Decadal Stability Radiometric Accuracy Identical worldwide Century-long stability Absolute accuracy

More information

Science of Stability 2017

Science of Stability 2017 Science of Stability 2017 Humidity Measurement in Stability Testing Applications Defining Uncertainty and the Challenges at Higher Temperatures Humidity Measurement In Stability Testing Applications Some

More information

H1: Metrological support. Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič

H1: Metrological support. Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič H1: Metrological support Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič Basic requirement How to ensure appropriate measurements and monitor changes in time and space? Analytical

More information

8001 E. BLOOMINGTON FREEWAY BLOOMINGTON, MN Service:

8001 E. BLOOMINGTON FREEWAY BLOOMINGTON, MN Service: simco e l e c t r o n i c s 806 CITY CENTER BOULEVARD NEWPORT NEWS, VA 23606 CERTIFICATE OF CALIBRATION FOR GN RESOUND-FIELD SERVICE KITS 8001 E. BLOOMINGTON FREEWAY BLOOMINGTON, MN 55420 Certificate No.

More information

AN EXPLANATION OF THE DG-1000 ACCURACY SPECIFICATIONS

AN EXPLANATION OF THE DG-1000 ACCURACY SPECIFICATIONS WHITE PAPER AN EXPLANATION OF THE DG-1000 ACCURACY SPECIFICATIONS The Energy Conservatory Minneapolis, MN Introduction: The purpose of this document is to explain the details of the accuracy specifications

More information

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 2017-03-06 revised 2018-02-13 (changed schedule) Task and Purpose of the Comparison According to the rules set up by the CIPM MRA the consultative

More information

Accuracy Calibration Certificate

Accuracy Calibration Certificate gl06-0-067-acc Mettler Toledo Service Business Unit Industrial 900 Polaris Parkway Columbus, OH 0-800-- Services Provided By: Customer Accuracy Calibration Certificate Company: Sample Company Test Address:

More information

Calibration of Paroscientific Model 205 Pressure Sensor for use at Heard Island.

Calibration of Paroscientific Model 205 Pressure Sensor for use at Heard Island. INSTRUMENT TEST REPORT NUMBER 67 Calibration of Paroscientific Model 05 Pressure Sensor for use at Heard Island. Kent Gregory & Paul Huysing Physics Laboratory, OEB 15 July, 1994 Authorisation Jane Warne

More information

Technical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017

Technical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017 MICROMATTER TECHNOLOGIES INC. #1-8333 130 th Street, Surrey, B.C. V3W 7X4, Canada Phone: +1.604.594.9720 or 1.866.966.2286 Fax: +1.604.594.9726 info@micromatter.com www.micromatter.com Technical Note 2015-01A

More information

Method Validation and Accreditation

Method Validation and Accreditation SELAMAT Mycotoxins Workshop China, 11-15th December 2006 Method Validation and Accreditation Dr Hamide Z Şenyuva Senior Research Scientist TÜBİTAK-ATAL, TURKEY hamide.senyuva@tubitak.gov.tr SELAMAT Mycotoxins

More information

POWER QUALITY MEASUREMENT PROCEDURE. Version 4 October Power-Quality-Oct-2009-Version-4.doc Page 1 / 12

POWER QUALITY MEASUREMENT PROCEDURE. Version 4 October Power-Quality-Oct-2009-Version-4.doc Page 1 / 12 POWER QUALITY MEASUREMENT PROCEDURE Version 4 October 2009 Power-Quality-Oct-2009-Version-4.doc Page 1 / 12 MEASNET 2009 Copyright all rights reserved This publication may not be reproduced or utilized

More information

MASS AND VOLUME COMPARISONS AT MIKES

MASS AND VOLUME COMPARISONS AT MIKES MITTATEKNIIKAN KESKUS CENTRE FOR METROLOGY AND ACCREDITATION Julkaisu J4/2000 MASS AND VOLUME COMPARISONS AT MIKES Additional results to the EA intercomparison of weights 1 mg 100 g (Ma1) and to the EUROMET

More information

Test Report EN (2008)

Test Report EN (2008) Test Report EN 62233 (2008) Product Name and address of the applicant Name and address of the manufacturer Model Rating Brand name Heating elements for building into floors Advanced Heating Technologies

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ARCONIC POWER AND PROPULSION RESEARCH CENTER 1500 S. Warner St. Whitehall, MI 49461-1895 Michelle McDonald Phone: 231 894 7586 CALIBRATION

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 FRANK COX METROLOGY (Formerly CANADIAN CENTRAL GAUGE LABORATORY) 40 West Drive Brampton, Ontario, Canada L6T 3T6 Hilliard Cox Phone: 905 457 9190 CALIBRATION

More information

R SANAS Page 1 of 7

R SANAS Page 1 of 7 ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Approved By: Chief Executive Officer:

More information

IS INTERNATIONAL STANDARD. Natural gas - Guidelines to traceability in analysis

IS INTERNATIONAL STANDARD. Natural gas - Guidelines to traceability in analysis INTERNATIONAL STANDARD IS0 14111 First edition 1997-03-15 Natural gas - Guidelines to traceability in analysis Gaz na turel - Lignes directrices pour la tracabilit6 en analyse Reference number IS0 14111

More information

Calibration Traceability Guidelines

Calibration Traceability Guidelines Calibration Traceability Guidelines Check the website of accrediting bodies for the accredited laboratories individual scope of accreditation www.a2la.org ts.nist.gov/standards/accreditation/index.cfm

More information

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis TECHNICAL SPECIFICATION ISO/TS 13530 First edition 2009-03-15 Water quality Guidance on analytical quality control for chemical and physicochemical water analysis Qualité de l'eau Lignes directrices pour

More information

R & D AKUSTIKA * SIA. Test Report of laboratory Sound Insulation measurements Nr. 569 / 2011 AL 8.4. CUSTOMER: AEROC AS (Estonia)

R & D AKUSTIKA * SIA. Test Report of laboratory Sound Insulation measurements Nr. 569 / 2011 AL 8.4. CUSTOMER: AEROC AS (Estonia) SIA * R & D AKUSTIKA * ============================================================================================= LATVIJAS REPUBLIKA, LV - 1067, R GA, KURZEMES PROSPEKTS 3, Re istr. Nr. : LV 010310045

More information

Uncertainty of Calibration. Results in Force Measurements

Uncertainty of Calibration. Results in Force Measurements Publication Reference EAL-G Uncertainty of Calibration Results in Force Measurements PURPOSE This document has been produced to improve the harmonization in determination of uncertainties in force measurements.

More information

Physikalisch- Technische Bundesanstalt

Physikalisch- Technische Bundesanstalt Physikalisch- Technische Bundesanstalt Guideline Calibration of Multiple Delivery Dispensers Edition 01/2018 Page: 2 / 22 Deutscher Kalibrierdienst (DKD) German Calibration Service Since its foundation

More information

ERROR LIMITS AND MEASUREMENT UNCERTAINTY IN LEGAL METROLOGY

ERROR LIMITS AND MEASUREMENT UNCERTAINTY IN LEGAL METROLOGY ERROR LIMITS AND MEASUREMENT UNCERTAINTY IN LEGAL METROLOGY K.-D. Sommer 1, M. Kochsiek and W. Schulz 1 Landesamt fuer Mess- und Eichwesen Thueringen, German Phsikalisch-Technische Bundesanstalt, German

More information

Appendix B1. Reports of SMU

Appendix B1. Reports of SMU Appendix B1 Reports of SMU Euromet Project 600 Comparison of Surface Roughness Standards /13 EUROMET SUPPLEMENTARY COMPARISON SURFACE TEXTURE Project No. 600 Final Report Elaborated by: M. Szmicskova Bratislava,

More information

ANNEXE 8. Technical protocols for the interlaboratory comparisons

ANNEXE 8. Technical protocols for the interlaboratory comparisons ANNEXE 8 Technical protocols for the interlaboratory comparisons TECHNICAL PROTOCOL FOR BILATERAL COMPARISON (Magnitude of the complex charge sensitivity of accelerometers) EURAMET.AUV.V-S1 1. INTRODUCTION

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 LAKE SHORE INDUSTRIAL SERVICES, INC. 2230 Caughey Road Erie, PA 16506 Suzanne R. Zuba / James B. Cross Phone: 814 838 3539 CALIBRATION

More information

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576 NPL REPORT CMAM 73 Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576 Peter Hanes October 2001 The National Physical Laboratory is operated on behalf

More information

The AWS based operational urban network in Milano: achievements and open questions.

The AWS based operational urban network in Milano: achievements and open questions. The AWS based operational urban network in Milano: achievements and open questions. Frustaci Giuseppe, Curci Savino, Pilati Samantha, Lavecchia Cristina, Paganelli Chiara Fondazione Osservatorio Meteorologico

More information

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014 Calibration General Principles & Theory, Equipment Considerations Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 MACHINIST TOOL REPAIR INC. 6484 Grafton Road Valley City, OH 44280 Susan Novak Phone: 330 483 3202 CALIBRATION Valid To: December 31,

More information

Commissioning of the Beta Secondary Standard (BSS2)

Commissioning of the Beta Secondary Standard (BSS2) Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,

More information

Draft EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement. Second Edition. Draft: June 1999

Draft EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement. Second Edition. Draft: June 1999 Draft EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement Second Edition Draft: June 1999 Prepared by the EURACHEM Measurement Uncertainty Working Group in collaboration with members

More information

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. EVALUATION OF MEASUREMENT UNCERTAINTY POLICY... 5

More information

SAMPLE. Accuracy Calibration Certificate. Mettler Toledo 1900 Polaris Parkway Contact: Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC WI v.1.

SAMPLE. Accuracy Calibration Certificate. Mettler Toledo 1900 Polaris Parkway Contact: Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC WI v.1. 076-00-067-ACC-S Mettler Toledo, LLC 900 Polaris Parkway Columbus, OH 0.800.METTLER Customer SAMPLE Accuracy Calibration Certificate Company: Address: City: Zip / Postal: State / Province: Mettler Toledo

More information

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition INTERNATIONAL STANDARD ISO 17123-4 Second edition 2012-06-01 Optics and optical instruments Field procedures for testing geodetic and surveying instruments Part 4: Electro-optical distance meters (EDM

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: AMC LABS S.A de C.V (Hereinafter called the Organization)

More information