Test report CVN Project number Page 2 of 2
|
|
- Ferdinand Caldwell
- 5 years ago
- Views:
Transcription
1
2 Project number Page 2 of 2 Tests : Results : The diaphragm gas meters as specified in annex 2 were tested for compliance with the European Standard BS EN 1359 : 1999 "Gas meters - Diaphragm gas meters", the EEC regulations on gas meters. The performed tests are stated in annex 1 of this test report. If applicable specific test conditions are stated at each test. See annex 1 of this test report. For all performed tests the gas meter fulfils the requirements of the European Standard BS EN 1359 (1999). The investigation has resulted in an EEC type approval no. E276 rev. 01. Uncertainty : The reported uncertainty of measurement is based on the standard uncertainty multiplied by a coverage factor k=2, which for a normal distribution corresponds to a coverage probability of approximately 95%. The total uncertainty of the measurements of the error of indication is 0,3% for flow rates greater than 400 dm³/h and 0,6% for flow rates smaller than or equal to 400 dm³/h. Annexes : The complete test report consists of the following annexes: annex 1 annex 2 annex 3 : performed tests : characteristics of the tested meters : test data Remark : The test data as presented in the annex 4 of this test report is performed under RvA accreditation with reference number L029, in which conformity to ISO/IEC has been demonstrated. The data as presented in the annexes 1, 2 and 3 gives extra information and is not included in the scope of this accreditation.
3 Test report no. CVN Annex 1 Page 1 of 2 Performed tests The diaphragm gas meter, type GNM, manufactured by LLC AEM, was tested for compliance with: - the European Standard BS EN 1359 : 1999 "Gas meters - Diaphragm gas meters". - the EEC regulations on gas meters; Particular requirements of the EN1359 article 5.1 or B tests: passed not not annex 4 applicable performed page Error of indication Pressure absorption Starting flow rate 5.4 Metrological stability External leak tightness Resistance to internal pressure Metercase sealing Connections Resistance to vibration Resistance to impact Resistance to mishandeling 6.4 Resistance to storage temperature range Pressure measuring point Insulating feet Magnetic index drive Reverse flow devices Resistance to high ambient temperature or B.2.2 Durability Meter error of indication at declared ambient and gas temperature limits 7.2 Index Toluene/iso-octane vapour test Water vapour test Ageing test 8.3 Durability and legibility of marking Ultraviolet exposure test
4 Test report no. CVN Annex 1 Page 2 of 2 Performed tests Particular requirements of EEC regulations on gasmeters. article EEC regulations 44, 45, 50 3 / appendix, II, 6 *) 3 / appendix, II, 4.3 *) 52, 55 tests: passed not not annex 3 applicable performed page accuracy tests under reference conditions 1 pressure absorption 2 repeatability test 30 x 0,1 Qmax endurance test Remark: The measurements are performed at a reference temperature of +20 C, unless an other temperature is stated. *) See chapter II of the appendix of the EEC regulations on gas meters.
5 Test report no. CVN Annex 2 Page 1 of 1 Annex 2: Characteristics of the tested meters The tested diaphragm gas meters have the following characteristics: Diaphragm gas meter: - Manufacturer: LLC AEM - Model: GNM - Q min [m 3 /h]: 0,04 - Q max [m 3 /h]: 6 - V [dm 3 ]: 1,2 - p max [bar]: 0,5 - Upper ambient temperature [ C]: Lower ambient temperature [ C]: 50 - Connections: ISO Markings on the individual meters: Sample Model Serial Year of number number manufacture 1 GNM G GNM G GNM G GNM G Remark: The results as mentioned in this test report relate only to the meters which are tested.
6 Annex 3 Page 1 of 4 Test: Errors of indication The errors of indication of the diaphragm gas meters are measured under reference conditions at several flow rates, by performing 6 accuracy curves in an increasing and decreasing manner. Before performing the accuracy tests the meters have been running in at a flow rate of 6 m³/h and passing a volume of 18 m³. Results: Running in: 18 m 3 at 6 m 3 /h Sample 3 Flow Test Errors [%] maximum rate volume differrence [m 3 /h] [dm 3 ] [%] Q max > ,2 + 0,4 + 0,5 + 0,4 + 0,3 + 0,3 0,3 0,7 Q max > ,0 + 1,1 + 1,0 + 1,2 + 1,2 + 1,3 0,3 0,4 Q max > ,1 + 1,1 + 1,2 + 1,3 + 1,2 + 1,4 0,3 0,2 Q max > ,2 + 1,3 + 1,4 + 1,5 + 1,4 + 1,5 0,4 0,1 Q max > ,3 + 1,2 + 1,2 + 1,4 + 1,2 + 1,2 0,3 3 Q min > ,6 + 0,9 + 0,5 + 0,8 + 0,9 + 0,9 Q min > 50 0,0 + 0,1 0,2 0,2 0,2 0,6 Sample 4 Flow Test Errors [%] maximum rate volume differrence [m 3 /h] [dm 3 ] [%] Q max > 500 1,4 1,1 0,8 1,2 1,2 1,3 0,6 0,7 Q max > 500 0,2 0,0 + 0,2 0,4 0,1 0,3 0,5 0,4 Q max > 500 0,1 0,1 + 0,3 + 0,2 + 0,1 + 0,0 0,4 0,2 Q max > ,6 + 1,0 + 1,0 + 1,0 + 0,8 + 1,1 0,4 0,1 Q max > ,2 + 1,5 + 1,4 + 1,5 + 1,2 + 1,2 0,3 3 Q min > ,4 + 1,6 + 1,6 + 1,4 + 1,4 + 1,7 Q min > ,5 + 1,7 + 1,2 + 0,8 + 1,0 + 0,9 Flow Mean error for mpe result each meter [%] rate Sample no. [m 3 /h] 3 4 [%] +/- Q max + 0,3 1,1 ± 1,5 + 0,7 Q max + 1,1 0,1 ± 1,5 + 0,4 Q max + 1,2 + 0,1 ± 1,5 + 0,2 Q max + 1,4 + 0,9 ± 1,5 + 0,1 Q max + 1,2 + 1,3 ± 1,5 + 3 Q min + 0,8 + 1,5 ± 3,0 + Q min 0,2 + 1,2 ± 3,0 +
7 Annex 3 Page 2 of 4 Test: Pressure absorption The pressure absorption of the meters is determined at Q max and at Q min. Results: Pressure absorption: Average value at Q max : 180 Pa Average value at Q min : 20 Pa
8 Annex 3 Page 3 of 4 Test: Resistance to high ambient temperatures The meter under test is installed inside a furnace with its connenctions positioned identical as used in practice. The meter is pressurized with nitrogen at a pressure of 100 mbar and maintained constant during the whole test. The test is performed with special prepared sample *). Before performing the high temperatures test the meter under test is checked on leak tighness and was found leak tight. Next the furnace is set to a temperature of 650 ºC, the rise curve of the increasing temperature is according to ISO 834. Results: After regarding a period of 30 minutes at 650 C the leakage rate is registered. The registered leakage rate did not exceed 150 dm³/h. *) Manufacturer has prepared a special sample meter for performing the high temperature test. In this prepared meter the internal metering part was replaced with a metal weight piece equivalent to the mass of the metering apparatus.
9 Annex 3 Page 4 of 4 Test: Meter error of indication at declared ambient and gas temperature limits The constant temperature test is performed with 2 meters at +20, +50 and at -20 C. For each meter the measured errors are stated at mentioned temperatures. Sample 3 Flow Test Measured errors [%] rate volume Error at Error at Error at [m 3 /h] [dm 3 ] -20 C +20 C +50 C Q max > ,2 0,2 + 2,2 0,4 Q max > 200 0,2 + 0,3 + 1,5 0,1 Q max > 100 1,4 0,1 + 1,6 Sample 4 Flow Test Measured errors [%] rate volume Error at Error at Error at [m 3 /h] [dm 3 ] -20 C +20 C +50 C Q max > ,4 + 0,6 + 0,3 0,4 Q max > ,6 + 0,5 + 1,0 0,1 Q max > 100 0,3 1,2 + 2,4 Remark: This test is performed on completion of the endurance test
CALIBRATION CERTIFICATE # 503
CALIBRATION CERTIFICATE # 503 Calibration date : 2011-02-23 Certificate issued : 2012-04-18 Company name Company address City, Province, Canada CLAS 2009-02 Calibration of Mass flow meter Micro Motion
More informationWIESON TECHNOLOGIES CO., LTD.
TABLE OF CONTENT 1. Scope.... 2 2. Reference Documents. 2 3. Material and Components.... 2 4. Design and Construction.... 2 5. Rating.. 2 6. Performance and Test Descriptions. 2 7. Test Requirements and
More informationSolderSleeve Devices Specification For Commercial Wire Splices. Table of Contents
Page 1 of 15 SolderSleeve Devices Specification For Commercial Wire Splices Table of Contents Chapter Pages 1.0 General Requirements 1.1 Design of the Electrical Connection-Reminder Concerning The Approval
More informationACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES
ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES (RPB-01) Approved by: Date: SECTION PAGE # 1.0 Scope 2.0 Definitions 3.0 References 4.0 Inspection
More informationWithdrawn at December 31, Guideline DKD-R 5-7. Calibration of Climatic Chambers DEUTSCHER KALIBRIERDIENST
DEUTSCHER KALIBRIERDIENST Guideline DKD-R 5-7 Calibration of Climatic Chambers Edition 07/2004 English translation 02/2009 Page 1 of 31 Published by the Accreditation Body of the Deutscher Kalibrierdienst
More informationGuidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018
Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018 Guidelines on the Calibration of Automatic Instruments for
More informationE C t y p e - e x a m i n a t i o n C e r t i f i c a t e
E C t y p e - e x a m i n a t i o n C e r t i f i c a t e Page 1 of 1 Issued by Manufacturer NMi Certin B.V., designated and notified by the Netherlands to perform tasks with respect to conformity modules
More informationEngineering Specification
Drawing No: 23 338 00001 Page 2 of 5 1. Purpose: To specify performance requirements for Pressure Sensitive Altimeters for use in B.I.C. manufactured ground support test equipment. 2. Scope: This PRODUCT
More informationTest Report EN (2008)
Test Report EN 62233 (2008) Product Name and address of the applicant Name and address of the manufacturer Model Rating Brand name Heating elements for building into floors Advanced Heating Technologies
More informationEA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE
Publication Reference EA-10/13 EA Guidelines on the Calibration of Temperature Block Calibrators PURPOSE This document been produced by EA to improve the harmonisation in the calibration of temperature
More informationCNG measuring systems for vehicles
CNG measuring systems for vehicles Sistemas de medición de GNC de uso vehicular Part 3 Review (1) METROLOGY - Accuracy - Trueness - Precision - Error - Uncertainty, Type A and Type B - Standard deviation
More informationMeasuring Instruments Directive (MID) MID/EN14154 Short Overview
Measuring Instruments Directive (MID) MID/EN14154 Short Overview STARTING POSITION Approval vs. Type examination In the past, country specific approvals were needed to sell measuring instruments in EU
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 8426 Second edition 2008-02-01 Hydraulic fluid power Positive displacement pumps and motors Determination of derived capacity Transmissions hydrauliques Pompes et moteurs volumétriques
More informationOA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents
UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC 17025 Table of contents 1 GENERAL... 2 2 THE STANDARD SIST EN ISO/IEC 17025... 2 3 SA'S POLICY IN IMPLEMENTING
More informationFLT5A FLT5A. abc. abc. Thermometer User Guide. Thermometer User Guide. Internet:
abc instruments international FLT5A Thermometer Land Instruments International Dronfield, S18 1DJ England Telephone: (1246) 417691 Facsimile: (1246) 41585 Email: infrared.sales@landinst.com Internet: www.landinst.com
More informationCalibration of Temperature Block Calibrators
European Association of National Metrology Institutes Calibration of Temperature Block Calibrators EURAMET cg-13 Version 2.0 (03/2011) Previously EA-10/13 Calibration Guide EURAMET cg-13 Version 2.0 (03/2011)
More informationUncertainties associated with the use of a sound level meter
NPL REPORT DQL-AC 002 Uncertainties associated with the use of a sound level meter Richard Payne April 2004 April 2004 ABSTRACT Uncertainties associated with the use of a sound level meter Richard Payne
More informationMOY/SCMI/36 SPECIFICATION OF ACCURACY FOR A PRECISION CLINOMETER
Centre for Basic, Thermal and Length Metrology National Physical Laboratory MOY/SCMI/36 SPECIFICATION OF ACCURACY FOR A PRECISION CLINOMETER A Watts Precision Clinometer fitted with a circular glass scale
More informationDocument No: TR 12 Issue No: 1
ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Prepared by: SADCAS Technical Manage Approved
More informationDetermination of particle size distribution Single particle light interaction methods. Part 4:
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 21501-4 Second edition 2018-05 Determination of particle size distribution Single particle light interaction methods Part 4: Light scattering airborne
More informationCAPACITORS, FIXED, TUBULAR, POROUS TANTALUM CATHODE AND ANODE, GELLED ELECTROLYTE, HERMETICALLY SEALED BASED ON TYPE ST79
Pages 1 to 15 CAPACITORS, FIXED, TUBULAR, POROUS TANTALUM CATHODE AND ANODE, GELLED ELECTROLYTE, HERMETICALLY SEALED BASED ON TYPE ST79 Issue 1 November 2007 Document Custodian: European Space Agency -
More informationCERTIFICATE OF CALIBRATION
Page 1 of 10 CALIBRATION OF Sound Level Meter: Brüel & Kjær Type 3050-A-040 No: 3050-100751 Id: - Microphone: Brüel & Kjær Type 4189 No: 2621142 Preamplifier: Brüel & Kjær Type 2669 No: 2679073 Supplied
More informationCalibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands
Calibration, traceability, uncertainty and comparing measurement results Workshop 16 March 2015 VSL, Delft The Netherlands 1 Calibration Calibrating an instrument involves comparing the measurements of
More informationAPPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT
APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. ESTIMATION OF UNCERTAINTY OF MEASUREMENT POLICY...
More informationOIML R 141 RECOMMENDATION. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION
INTERNATIONAL RECOMMENDATION OIML R 141 Edition 2008 (E) Procedure for calibration and verification of the main characteristics of thermographic instruments Procédure pour l'étalonnage et la vérification
More informationPage 2 of 14 Report Ref. No
age 2 of 14 Summary of testing: Tests performed (name of test and test clause): Clause 6.1, Initial measurements: - MST 01: Visual inspection - 10.2: Maximum power determination - MST 16: Dielectric withstand
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 EQUIPMENT SERVICES GROUP, LLC 3269 Hilton Rd. Ferndale, MI 48220 Michael Macauley Phone: 248 227 5310 CALIBRATION Valid To: December
More informationREPORT RESULTS OF RADON DIFFUSION COEFFICIENT (SAMPLES SYNTHESIA O-14015)
REPORT RESULTS OF RADON DIFFUSION COEFFICIENT (SAMPLES SYNTHESIA O-14015) Prof. Dr. Carlos Sainz Fernandez Technical Director REPORT RESULTS RADON DIFFUSION COEFFICIENT - SYNTHESIA CONTENTS 1. INTRODUCTION
More informationCertificate of Accreditation to ISO/IEC 17025:2005
United States Department of Commerce National Institute of Standards and Technology Certificate of Accreditation to ISO/IEC 17025:2005 NVLAP LAB CODE: 200508-0 Insco Metrology, Inc. Miami, FL is accredited
More informationAPPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY
APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. EVALUATION OF MEASUREMENT UNCERTAINTY POLICY... 5
More informationTITLE STATUS SPEC. NO. DATE PAGE NO. 3.5Φ EARPHONE JACK 4 POLEs / 2 SWITCH GENERAL SMT TYPE BS S-01
Specification TITLE STATUS SPEC. NO. DATE PAGE NO. 3.5Φ EARPHONE JACK 4 POLEs / 2 SWITCH GENERAL SMT TYPE BS123-000S-01 1 OF 4 1999/09/30 1. GENERAL SCOPE THIS SPECIFICATION COVERs THE GENERAL REQUIREMENTs
More information9.1 Compatibility of modules for nonautomatic weighing instruments (NAWIs) Common requirements
Übersetzung aus: Verwaltungsvorschriften gesetzliches Messwesen (administrative regulations legal metrology) 9.1 Compatibility of modules for nonautomatic weighing instruments (NAWIs) 9.1.1 Common requirements
More informationMorehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:
Morehouse 1 Morehouse Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA 17403 PH: 717-843-0081 web: www.mhforce.com sales: edlane@mhforce.com 2 This presentation will cover the calibration
More informationCERTIFICATE OF CALIBRATION
Page 1 of 10 CALIBRATION OF Sound Level Meter: Brüel & Kjær Type 3050-A-040 No: 3050-100751 Id: - Microphone: Brüel & Kjær Type 4189 No: 2621142 Preamplifier: Brüel & Kjær Type 2669 No: 2803547 Supplied
More informationISO 6395 INTERNATIONAL STANDARD. Earth-moving machinery Determination of sound power level Dynamic test conditions
INTERNATIONAL STANDARD ISO 6395 Second edition 2008-03-15 Earth-moving machinery Determination of sound power level Dynamic test conditions Engins de terrassement Détermination du niveau de puissance acoustique
More informationPRODUCT SPECIFICATION FOR Micro SIM +Nano SIM/Micro SD (T/F)Card Connector SERIES
Page 1 of 5 1.0 SCOPE This Specification covers Micro SIM +Nano SIM/Micro SD Card Connector SERIES 2.0 APPLICABLE DOCUMENTS AND SPECIFICATIONS See sales drawings and other sections of this specification
More informationCommenting template CEN/TC 102 Date : Document: pren 13060
Commenting template CEN/TC 0 Date : 006-09-3 Document: pren 3060 MB Clause / 3.) ) FR general Ge There have been so many discussions during the CEN/TC 0/WG5 meetings these years. The committee has recognised
More informationTÜV RHEINLAND ENERGY GMBH ADDENDUM
TÜV RHEINLAND ENERGY GMBH ADDENDUM Addendum to TÜV test report no. 936/21207522/A dated 23 March 2009 on performance testing of the SWAM 5a Dual Channel Monitor, SWAM 5a Monitor and SWAM 5a Dual Channel
More informationSAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0
076-0-067-ACC-USL-S Mettler Toledo, LLC 900 Polaris Parkway Columbus, OH 0.800.METTLER Customer Accredited by the American Association for Laboratory Accreditation (ALA) CALIBRATION CERT #788.0 SAMPLE
More informationSTATISTIQUE DE PERFORMANCE
INTERCOMPARISON OF STANDARD CALIBRATION, METHODS AND MEASUREMENTS FOR PRESSURE, TEMPERATURE AND HUMIDITY BETWEEN RIC CASABLANCA, MOROCCO AND RIC Ljubljana SLOVENIA 3(10) azizmounir@gmail.com The operators
More informationISO Radiological protection Sealed radioactive sources General requirements and classification
INTERNATIONAL STANDARD ISO 2919 Third edition 2012-02-15 Radiological protection Sealed radioactive sources General requirements and classification Radioprotection Sources radioactives scellées Exigences
More informationEA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement
Publication Reference EA-10/11 EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical PURPOSE This document has been produced by EA as a means of giving advice for calibrating
More informationTRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS
TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS W. Beyer and W. Pahl Physikalisch-Technische Bundesanstalt (PTB) 38116 Braunschweig, Germany Abstract: In accordance with ISO 17025,
More informationMetallic materials Brinell hardness test. Part 3: Calibration of reference blocks
INTERNATIONAL STANDARD ISO 6506-3 Third edition 2014-10-01 Metallic materials Brinell hardness test Part 3: Calibration of reference blocks Matériaux métalliques Essai de dureté Brinell Partie 3: Étalonnage
More informationR SANAS Page 1 of 7
ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Approved By: Chief Executive Officer:
More informationDetermining of Thermal Conductivity Coefficient of Pressed Straw (Name of test)
CONSTRUCTION HEAT PHYSICS LABORATORY INSTITUTE OF ARCHITECTURE AND CONSTRUCTION OF KAUNAS UNIVERSITY Test and calculation carried out according to: Product: 12 December 2012 LITHUANIAN NATIONAL ACCREDITATION
More informationVEIKI-VNL ELECTRIC LARGE LABORATORIES Ltd. TYPE TEST REPORT
Page 1 of 13 TYPE TEST REPORT Test object: Designation: All Aluminium Alloy Conductor (AAAC) 50 mm 2 All Aluminium Alloy Conductor (Code: 49-AL3) Manufacturer: BMET Energy Telecom Industry and Trade LLC.
More informationEstablishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements
Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements Experimental Design 8.2.2013 Doc. Martti Heinonen etunimi.sukunimi@mikes.fi Outline 1.
More informationSANMUR CONTROLS PVT. LTD.
Decide with confidence NMUR CONTROLS NMUR CONTROLS PVT. LTD. An ISO 900:008 Certified Company The name NMUR has been synonymous with reliability and quality in terms of products, services and systems integration.
More informationE v a l u a t i o n C e r t i f i c a t e
E v a l u a t i o n C e r t i f i c a t e Page 1 of 1 Issued by In accordance with Producer WELMEC guide 8.8 General and Administrative Aspects of the Voluntary System of Modular Evaluation of Measuring
More informationLaboratory Evaluation Report for: Dielectric Pipe Unions
American Society of Sanitary Engineering Seal (Certification) Program Laboratory Evaluation Report for: Dielectric Pipe Unions Tested under ASSE Standard 1079 Issued: September, 2005 Laboratory File Number
More informationDetermination of installed thermal resistance into a roof of TRISO-SUPER 12 BOOST R
INSTITUTE OF ARCHITECTURE AND CONSTRUCTION OF KAUNAS UNIVERSITY OF TECHNOLOGY LABORATORY OF BUILDING PHYSICS Notified Body number: 2018 TEST REPORT No. 037-10/11(C) SF/15 Date: 26 of November 2015 page
More informationCollector test according to EN ,2 : 2006
Test Report: KTB No. 2006-27 Collector test according to EN 12975-1,2 : 2006 for: Himin Solar Energy Group Co., Ltd., China Brand Name: HUJ 16/2.1, HUJ 12/2.1, HUJ 16/1.8, HUJ 12/1.8, HUJ 16/1.6 and HUJ
More informationThe Fundamentals of Moisture Calibration
The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity
More information1 General information about the gas meter
Page 1 f 8 1 General infrmatin abut the gas meter All prperties f the gas meter, whether mentined r nt, may nt be in cnflict with the legislatin. This revisin is issued with respect t the extensin f the
More informationMULTILAYER CERAMIC CAPACITORS Low Inductance Series 0612 Size, 50V X7R Dielectric Halogen Free & RoHS Compliance
MULTILAYER CERAMIC CAPACITORS Low Inductance Series 0612 Size, 50V X7R Dielectric Halogen Free & RoHS Compliance *Contents in this sheet are subject to change without prior notice. Page 1 of 7 ASC_Low
More informationCERTIFICATE OF ACCREDITATION
CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Hards Laboratories cc, t/a Technology Solutions
More informationUrban Legends in Gasmetering
Urban Legends in Gasmetering Jos van der Grinten Scientist High Pressure Gas Department Enquiry three statements A national standards institute can disseminate the m³ accurately without re-calibration.
More informationMulti Analyte Custom Grade Solution. Calcium, Iron, Potassium,
1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for
More informationMULTILAYER CERAMIC CAPACITORS Open-Mode Series (50V to 630V) 0805 to 1812 Sizes X7R Dielectric Halogen Free & RoHS Compliance
MULTILAYER CERAMIC CAPACITORS Open-Mode Series (50V to 630V) 0805 to 1812 Sizes X7R Dielectric Halogen Free & RoHS Compliance *Contents in this sheet are subject to change without prior notice. Page 1
More informationEUROVENT 8/12 SOUND TEST METHOD FOR DUCTED FAN COIL UNITS. FCP 2008 testing procedures FCP acoustical test method pag. 1
EUROVENT 8/12 SOUND TEST METHOD FOR DUCTED FAN COIL UNITS FCP 2008 testing procedures FCP acoustical test method pag. 1 CONTENTS 1 - PURPOSE... 3 2 - NORMATIVE REFERENCES... 3 3 - DEFINITIONS... 3 3.1
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 140-6 Second edition 1998-08-15 Acoustics Measurement of sound insulation in buildings and of building elements Part 6: Laboratory measurements of impact sound insulation of
More informationAIR TEST CALCULATION METHODOLOGY
AIR TEST CALCULATIO METHODOLOGY Produced by: Marc Cowlin Compliance Manager Version: 2.0 Issue Date: 0 th August 206 www.stromatech.com ITRODUCTIO This document has been compiled to explain how the figures
More informationA61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012
Revision 1.2 October 15, 2012 PAGE II TABLE OF CONTENTS TABLE OF CONTENTS... 1 GUIDANCE...2 1.0 Introduction... 2 2.0 Uncertainty and Traceability... 2 3.0 Need for Traceability... 3 3.1 What does traceability
More informationCERTIFICATE OF CALIBRATION
Issued By Transmille Ltd. Date of Issue 20 January 2016 0324 Approved Signatory Transmille Ltd. Unit 4, Select Business Centre Lodge Road Staplehurst, Kent. TN12 0QW. TEL 01580 890700 FAX 01580 890711
More informationWindow air permeability measurement, watertightness and resistance to wind load (designation of the test)
KTU INSTITUTE OF ARCHITECTURE AND CONSTRUCTION BUILDING THERMAL PHYSICS Notified Body number: 2018 BANDYMAI ISO/IEC 17025 Nr. LA. 01.031 Test performed: Product: Page (pages) 05 of August 2013 1 (8) Window
More informationReport on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz
Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range to 10 k June 016 Author: Co-author: Danuta Dobrowolska (GUM, pilot) Alexander Kosterov
More informationProduct Specification HYCAP HIGH TEMP, EXTREME SHOCK &VIBRATION. Family >>> HCD HC2D HC3D Max mass 18g 15g 15.5g
CPCITOR HYCP HIGH TEMP, &VIBRTION DTE 3/14/16 1.0 SCOPE This document contains specific electrical, mechanical, and environmental requirements and specifications for double- sealed, axial- leaded hybrid
More informationCALIBRATION REPORT FOR THERMOHYDROMETER
CALIBRATION REPORT FOR THERMOHYDROMETER SAMPLE CUSTOMER S/N: XXXXX Report Number: Y205188 Page 1 of 5 Report No. Y205188 ICL CALIBRATION LABORATORIES, INC. ISO/IEC 17025 and ANSI/NCSL Z540-1 accredited
More informationCollector test according to EN ,2:2002
Test Report: KTB Nr. 2003-18-b-en Collector test according to EN 12975-1,2:2002 for: RM Solar Ltd Brand Name: S-Class Responsible for Testing: Dipl.-Ing. (FH) A. Schäfer Date: 17th July 2006 Address: Fraunhofer-Institute
More informationISO INTERNATIONAL STANDARD. Thermal performance of windows, doors and shutters Calculation of thermal transmittance Part 1: Simplified method
INTERNATIONAL STANDARD ISO 10077-1 First edition 2000-07-15 Thermal performance of windows, doors and shutters Calculation of thermal transmittance Part 1: Simplified method Performance thermique des fenêtres,
More informationProduct Description. Table of Contents. CLD-DS30 Rev 11. Product family data sheet
Cree XLamp ML-E LEDs Product family data sheet CLD-DS3 Rev 11 Product Description The Cree XLamp ML-E LED brings lighting-class reliability and performance to ½-watt LEDs. The XLamp ML-E LED expands Cree
More informationS-S35B-F
Specification S-S35B-F2-285-01-2-110 RoHS SETi Customer Drawn Approval Approval 1. Description Table of Contents: 2. Mechanical Dimensions 3. Characteristics of S-S35B-F2-285-01-2 4. Characteristic Diagrams
More informationANSI/AHRI Standard (Formerly ARI Standard ) 2006 Standard for Performance Rating of Desuperheater/Water Heaters
ANSI/AHRI Standard 470-2006 (Formerly ARI Standard 470-2006) 2006 Standard for Performance Rating of Desuperheater/Water Heaters IMPORTANT SAFETY DISCLAIMER AHRI does not set safety standards and does
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 61788-6 First edition 2000-12 Superconductivity Part 6: Mechanical properties measurement Room temperature tensile test of Cu/Nb-Ti composite superconductors Supraconductivité
More informationMeasurement Uncertainty in Mechanical Testing
Intelligent testing Measurement Uncertainty in Mechanical Testing 27. testxpo 2018 ZwickRoell GmbH & Co. KG Ulm Dr. Eduard Schenuit Industry Manager Metal Agenda Why do we need measurement uncertainty?
More informationThis document is a preview generated by EVS
INTERNATIONAL STANDARD ISO 13168 First edition 2015-07-01 Water quality Simultaneous determination of tritium and carbon 14 activities Test method using liquid scintillation counting Qualité de l eau Détermination
More informationTurbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas
Turbine Meter TRZ 0 PRODUCT INFORMATION Reliable Measurement of Gas TURBINE METER TRZ 0 Method of operation, Construction Method of operation The TRZ 0 turbine meter is a flow meter suitable for gas measurement
More informationSpecific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements
Specific Accreditation Criteria Calibration ISO/IEC 17025 Annex Ionising radiation measurements January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected
More informationIMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT
IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT JO Favreau 1, E Georgin 1, B Savanier 1, A. Merlone 2 1 CETIAT, 96100 Villeurbanne, France 2 INRIM, Torino, Italy Abstract. CETIAT is a calibration laboratory
More informationFlow Management Devices, LLC An ISO 9000:2008 Certified Company
Flow Management Devices, LLC An ISO 9000:2008 Certified Company Liquid Flow Prover US Patent # 7650775 UNCERTAINTY ANALYSIS Model FMD-060 For the proving of volumetric pulse output flow meters Flow Management
More informationComparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576
NPL REPORT CMAM 73 Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576 Peter Hanes October 2001 The National Physical Laboratory is operated on behalf
More informationREPORT. Client. Test object. Date of testt. Results. Rönnåsgatan 5B Ulricehamn. (7 appendices) content of the
Riss sued by an Accreditedd Testing Laboratory Contact person Malin Lindgren Sustainable Builtt Environment +46 10 516 50 722 malin.lindgren@ @sp.se 2016-01-11 5P08641-A/Engg 1 (8)) Stellan Bygård Rönnåsgatan
More informationTechnical Report. Project. The Laboratory Determination of The Random Incidence sound Absorption Coefficient of VertiQ Panels.
Technical Report Ref Number C/21387/R01a Date 06 April 2011 Project The Laboratory Determination of The Random Incidence sound Absorption Coefficient of VertiQ Panels Prepared for Rockfon Rockwool A/S
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 17331 First edition 2004-05-15 Surface chemical analysis Chemical methods for the collection of elements from the surface of silicon-wafer working reference materials and their
More informationPrecision Thin Film Chip Resistor Array
ACAC 0612 (concave terminations) and ACAS 0612 (convex terminations) thin chip resistor arrays combine the proven reliability of precision thin film chip resistor products with the advantages of chip resistor
More informationMeasurement method for the proficiency testing program
APLAC T088 Appendix Measurement method for the proficiency testing program Introductions This measurement method is prepared for use by the APLAC Proficiency Testing Program Photometric measurement of
More informationREPORT issued by an Accredited Testing Laboratory
issued by an Accredited Testing Laboratory Contact person RISE Tobias Eriksson 2017-03-02 6P07577-25-rev1 1 (3) Energy and circular economy +46 10 516 57 07 tobias.eriksson@ri.se Camfil AB Industrigatan
More informationPart 5: Total stations
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 17123-5 Third edition 2018-02 Optics and optical instruments Field procedures for testing geodetic and surveying instruments Part 5: Total stations
More informationTYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE
TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE Michael Bair and Pierre Delajoud 2002 DEC 10 Revised 2004 MAR 26 FORWARD FPG8601 is a pressure standard designed
More informationCalibration Traceability Guidelines
Calibration Traceability Guidelines Check the website of accrediting bodies for the accredited laboratories individual scope of accreditation www.a2la.org ts.nist.gov/standards/accreditation/index.cfm
More informationDC/DC converter Input 9-36 and Vdc Output up to 0.5A/3W
PKV 3000 I PKV 5000 I DC/DC converter Input 9-36 and 18-72 Vdc up to 0.5A/3W Key Features Industry standard DIL24 Wide input voltage range, 9 36V, 18 72V High efficiency 74 83% typical Low idling power
More informationANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis
DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes
More informationTECHNICAL INFORMATION
TECHNICAL INFORMATION RATED PRIMARY OLTAGE () This is the supply assigned to the transformer by the manufacturer. RATED SECONDARY OLTAGE () This is the secondary output assigned to the transformer when
More informationProvläsningsexemplar / Preview. First edition
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 12764 First edition 2017-08 Measurement of fluid flow in closed conduits Flowrate measurement by means of vortex shedding flowmeters inserted in
More information@Copyright 2016 SKAPS Industries.
SKAPS INDUSTRIES 571 Industrial Pkwy, Commerce, GA 30529 Phone: (706) 336 7000 Fax: (706) 336 7007 E Mail: contact@skaps.com SKAPS GEOCOMPOSITE DROP IN SPECIFICATIONS @Copyright 2016 SKAPS Industries www.skaps.com
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 60076-10 First edition 2001-05 Power transformers Part 10: Determination of sound levels Transformateurs de puissance Partie 10: Détermination des niveaux de bruit IEC 2001 Copyright
More informationNon-destructive testing of steel forgings- Part 1: Magnetic particle inspection (BS EN :1999)
Non-destructive testing of steel forgings- Part 1: Magnetic particle inspection (BS EN 10228-1:1999) 1 Scope This part of EN 10228 describes the method and acceptance criteria to be used for the magnetic
More informationMULTILAYER CERAMIC CAPACITORS High Q / Low ESR Series (HH) 0201 to 0805 Sizes NP0 Dielectric Halogen Free & RoHS Compliance
MULTILAYER CERAMIC CAPACITORS High Q / Low ESR Series (HH) 0201 to 0805 Sizes NP0 Dielectric Halogen Free & RoHS Compliance *Contents in this sheet are subject to change without prior notice. Page 1 of
More information2.1. Accuracy, n- how close the indication of the thermometer is to the true value.
AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria
More information