REPORT issued by an Accredited Testing Laboratory
|
|
- Martha Webb
- 5 years ago
- Views:
Transcription
1 issued by an Accredited Testing Laboratory Contact person RISE Tobias Eriksson P rev1 1 (3) Energy and circular economy tobias.eriksson@ri.se Camfil AB Industrigatan TROSA Testing Testing of Air Filter according to ISO 16890:2016 (6 appendices) Rev 1 1) The energy calculation according to Eurovent 4/21 has been removed from the report. 2) Testing organisation has been changed from SP to RISE. This report replace 6P dated A test according to ISO 16890:2016 was carried out by request from Camfil AB. Tested item Camfil AB, Hi Flo II, XLT 7/ (HFGX-F7-592/592/ ), a 592 mm x 592 mm x 640 mm, 10 Pocket bag filter. Article number: Pictures can be found in appendix 5. The item was sent to SP by Camfil AB and was received by SP on February 14, The item was without visible defects. Test method The test was carried out according to standard ISO 16890:2016 Air filters for general ventilation. The standard consists of four parts: - ISO : Technical specifications, requirements and classification system based upon particulate matter efficiency (epm) - ISO : Measurement of fractional efficiency and air flow resistance Measurements were performed with dual particle counters according to section Testing sequence for dual OPC testing. - ISO : Determination of the gravimetric efficiency and the airflow resistance versus the mass of test dust captured Postal address Office location Phone / Fax / This report may not be reproduced other than in full, except Box 857 SE BORÅS Sweden Brinellgatan 4 SE BORÅS info@ri.se with the prior written approval of the issuing laboratory. Accred. No. 2 Testing ISO/IEC 17025
2 P rev1 2 (3) - ISO : Conditioning method to determine the minimum fractional test efficiency Eight cabinets with a total surface area of 1.82 m 2 were placed in the chamber according to ISO section 7. The purity of the 2-propanol was 99.5 %. The test item was conditioned for /- 0.5 hours. Efficiency at 50% nominal air flow was measured with DEHS in the range µm. Deviation from the standard: Section and 9.2.8, the evaporated amount of 2-propanol was not determined. Section 7.3, the atmospheric pressure during the conditioning procedure was not measured. Date and Place The test was carried out at SP s laboratory of Energy and circular economy in Borås, Sweden on February 16-23, Tests according to ISO were carried out on February, 16, Tests according to ISO were carried out on February, 23, Tests according to ISO were carried out on February, 23, 2017, conditioning procedure according to ISO was carried out on February, 20-21, Results The results are presented in appendix 1-4 and are valid only for the item tested. In appendix 1 a summary of the results are reported according ISO It also includes the fractional efficiencies and the calculation of PM-efficiencies. In appendix 2 fractional efficiency and air flow resistance are reported according to ISO In appendix 3 determination of the gravimetric efficiency (arrestance) and the air flow resistance versus the mass of test dust capture (test dust capacity) are reported according to ISO In appendix 4 the minimum fractional efficiency is reported according to ISO The measured particle concentrations are reported in appendix 2 and appendix 4. Table A6 (upstream count data), A7 (downstream count data) and A9 Uncertainty in ISO Annex A are reported.
3 P rev1 3 (3) Measurement equipment Pressure gauge Furness model 318, SP's inventory no (static P Filter) Pressure gauge Furness model 318, SP's inventory no (static P Flow) Pressure gauge Furness FC012, SP's inventory no ( P Filter) Pressure gauge Furness FC012, SP's inventory no ( P Flow) Particle counter TSI, OPS 3330, SP's inventory no Particle counter TSI, OPS 3330, SP's inventory no Barometer, Testo 511, SP's inventory no Temperature and RH, Testo 635, SP's inventory no Weighing scale, Mettler PC16, SP's inventory no Flow meter, MFS-C-250, SP's inventory no Temperature and RH, Tinytag, DIV 94 S Barometer, Druck PACE 1, SP's inventory no Uncertainty of measurement The uncertainty of the Air flow is better than ± 5 % The uncertainty of the Pressure Drop is better than ± 3 % The uncertainty of the Temperature is better than ± 0.5 ºC The uncertainty of the Relative Humidity is better than ± 2 % RH The uncertainty of the Atmospheric Pressure is better than ± 1 mbar The uncertainty of the Measured mass is better than ± 0.5 g The uncertainty has been calculated according to EA-4/16 with a coverage factor k=2. The uncertainty of the filtration efficiency according to ISO 16890:2016 is presented in appendices 2 and 4. Energy and circular economy - Building Services Engineering Performed by Examined by Signature_1 Tobias Eriksson Signature_2 Christian Mossberg Appendices 1. Summary test report according to ISO : Test report according to ISO : Test report according to ISO : Test report according to ISO : Additional pictures of the test item. 6. The interpretation of test reports
4 Pressure differential, 1.2 kg/m 3 (Pa) Arrestance (% ) Fractional efficiency (% ) REPORT P rev1 1 (3) Appendix 1 ISO : Air Filter Test Results Testing Organization: RISE Research Institute of Sweden Brinellgatan 4, Borås, Swedan GENERAL Report no.: 6P rev1 Date of tests: Date of report: Supervisor: CM Device obtained (when and how obtained): Test(s) requested by: Camfil AB The device was sent and obtained on DEVICE TESTED Model: Hi Flo II Manufacturer: Construction: XLT 7/ (HFGX-F7-592/592/ ) Camfil AB Pocket filter, 10 Pockets Article number: Type of medium: Net effective filtering area: Filter dimensions (width x height x depth) Glass 7.3 m 2 592x592x640 mm TEST DATA AND ATTACHED TEST REPORTS Test air flow rate: Test aersole: Test report to ISO Report no. 6P rev1 Appendix m 3 /s KCl (1-10 µm) Test report to ISO (optional) Report no. 6P rev1 Appendix 3 DEHS (0.3-1 µm) Test report to ISO Report no. 6P rev1 Appendix 4 RESULTS Initial pressure differential: Initial grav. arrestance: epm 1, min epm 2.5, min ISO rating 72 Pa 97 % 63 % 73 % Final test pressure differential: Test dust capacity: epm 1 epm 2.5 epm 10 ISO epm 1 60 % 300 Pa 1160 g 64 % 73 % 91 % Remarks: Particle size (µm) Initial fractional efficiency Ei (ISO ) Conditioned fractional efficiency ED,i (ISO ) Average fractional efficiency EA, i (ISO ) Test dust capture (g) Air flow rate (m 3 /s) Pressure differential as a function of the air flow rate (clean filter) (ISO ) Pressure differential as a function of the test dust captured (ISO ) Grav. arrestance as a function of the test dust captured (ISO ) NOTE: The results of this test relate only to the test device in the condition stated herein. The performance results cannot by themselves be quantitatively applied to predict filtration performance in all "real life" environments.
5 P rev1 2 (3) Appendix 1 ISO : Fractional efficiency values Testing organisation: RISE Research Institute of Sweden Report no: 6P rev1 Model: Hi Flo II XLT 7/ (HFGX-F7-592/592/ ) Manufacturer: Camfil AB Test air flow rate: m 3 /s Date of report: i d i d i+1 d a, i Δln d i E i E D, i E A, i µm µm µm µm % % % d i : Lower limit particle diameter in a size range i, µm d i+1 : Upper limit particle diameter in a size range i, µm d a, i : Geometric mean diamter of a size range i, µm Δln d i : Logarithmic width of a particle diameter size in range i; ln is the natural logaritm to the base of e, where e is an irrational and transcendetal constant approximately equal to , dimensionless ln d i = ln (d i+1 /d i ) E i : Initial fractional efficiency of partcile size range i of the untreated and unloaded filter element, % E D, i : Fractional efficiency of particle size range i of the filter element after an artifical conditioning step, % E A, i : Average fractional efficeincy (E i + E D, i )/2 of particle size range i, %
6 P rev1 3 (3) Appendix 1 ISO : Calculation of PM-efficiencies Testing organisation: Model: RISE Research Institute of Sweden Hi Flo II XLT 7/ (HFGX-F7-592/592/ ) Report no.: Manufacturer: 6P rev1 Camfil AB Test air flow rate: m 3 /s Date of report: i urban E Δln d i q 3u (d a, i ) D, i E A, i d a, i epm x, min epm x distribution q ln d 3u (d a, i ) q 3u (d a, i ) µm µm q 3u (d a, i ) i ln d i ln d i % % epm 1, min epm 1 line epm 2.5, min epm 2.5 line i rural E d a, i Δln d i q 3u (d a, i ) A, i epm x distribution q ln d 3u (d a, i ) µm µm q 3u (d a, i ) i ln d i % epm 10 line
7 P rev1 1 (3) Appendix 2 ISO : AIR FILTER TEST RESULTS SUMMARY GENERAL Test ID: SP Date of test: Operator: CM Particle counter information Air flow measurement: Device obtained (when and how Manufacturer: Model: Coincidence value Annubar, Micatrone obtained): TSI Gmbh OPS 3330 (p/cm 3 ): Air flow sensor MFS-SS The device was sent and obtained on DEVICE TESTED Model: Hi Flo II Manufacturer: Construction: XLT 7/ (HFGX-F7-592/592/ ) Camfil AB Pocket filter, 10 Pockets Article number: Type of media: Glass Filter/media electrostatic charge: No Device Condition: Clean / Initial Net effective media area (m 2 ) 7.3 m 2 Media colour: green Testing Organization: RISE Research Institute of Sweden Brinellgatan 4, Borås, Swedan Filter dimensions (width x height x depth) 592x592x640 mm Media adhesive: N/A Other descriptive information: TEST DATA SUMMARY Test air flow rate: m 3 /s RESULTS Measured: Remarks: Test air temperature: C Resistance to airflow (Pa) Rated initial: - 72 Pa Rated Final: - Test Device Photo Test air RH: % Test aersol: DEHS (0.3-1 µm) KCl (1-10 mm) Fractional Efficiency (%) Range (µm) Measured Rated Upstream concentration Efficiency Efficiency (particles / dm 3 ) NOTE: The results of this test relate only to the test device in the condition stated herein. The performance results cannot by themselves be quantitatively applied to predict filtration performance in all "real life" environments.
8 Fractional efficiency (%) Pressure differential, 1.2 kg/m 3 (Pa) REPORT P rev1 2 (3) Appendix 2 ISO : AIR FILTER TEST RESULTS DETAILS Testing Organization: RISE Research Institute of Sweden Brinellgatan 4, Borås, Swedan Test ID: SP Date of test: Operator: CM TEST DATA DETAILS Resistance to Airflow 1.2 kg/m 3 % of rated airflow Airflow Resistance to (m 3 /s) Airflow (Pa) % % % % % Airflow (m 3 /s) Fractional Efficiency by Particle Size Particle size (µm) NOTE: The results of this test relate only to the test device in the condition stated herein. The performance results cannot by themselves be quantitatively applied to predict filtration performance in all "real life" environments.
9 P rev1 3 (3) Appendix 2 Efficiency measurement Upstream count data OPC bin d a, i Upstream efficiency count data µm U e,tot Note: All data shown is the number of particle counts for 60 s Efficiency measurement Downstream count data OPC bin d a, i Downstream efficiency count data µm D e,tot Note: All data shown is the number of particle counts for 60 s Efficiency measurement Final results and uncertainty OPC bin d a, i Penetration data reduction Uncertainty limits Uncertainty Efficiency µm P a δ e Static Dynamic Pass/Fail % Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass 99.9 d a, i : Geometric mean diameter of a size range i, µm P a δ e the final penetration for a given particle size the standard deviation of the penetration for a given particle size the uncertainty of the penetration for a given particle size
10 P rev1 1 (4) Appendix 3 ISO : AIR FILTER TEST RESULTS SUMMARY GENERAL Test ID: SP Date of test: Operator: CM DEVICE TESTED Model: Hi Flo II XLT 7/ (HFGX-F7-592/592/ ) Camfil AB Article number: Type of media: Glass Manufacturer: Air flow measurement: Annubar, Micatrone Air flow sensor MFS-SS Net effective media area (m 2 ) 7.3 m 2 Filter/media electrostatic charge: Media colour: No green Device Condition: Conditioned per ISO Testing Organization RISE Research Institute of Sweden Brinellgatan 4, Borås, Swedan Test sample obtained: The device was sent and obtained on Construction: Pocket filter, 10 Pockets Filter dimension (W x H x D) (mm) 592x592x640 mm Media adhesive: N/A Other descriptive information: TEST DATA SUMMARY Test air flow rate: Test air temperature: Test air RH: Loading dust: m 3 /s C % Particle Technology, ISO A2-fine RESULTS Resistance to airflow Dust loading results Measured: 72 Pa Final test pressure: 300 Pa Rated initial: Rated Final: - Pa Test Device Photo Initial arrestance (%) Average arrestance(%) Test dust capacity (g) - Pa 97 % >99 % 1160 g Remarks: NOTE: The results of this test relate only to the test device in the condition stated herein. The performance results cannot by themselves be quantitatively applied to predict filtration performance in all "real life" environments.
11 Arrestance (% ) Resistance to Airflow, 1.2 kg/m 3 (Pa) REPORT P rev1 2 (4) Appendix 3 ISO : AIR FILTER TEST RESULTS DETAILS Testing Organization: Test ID: SP Date of test: Operator: CM TEST DATA DETAILS Resistance to Airflow 1.2 kg/m 3 % of rated Airflow Resistance to airflow (m 3 /s) Airflow (Pa) % % % % % Resistance to Airflow, 1.2 kg/m 3 (Pa) RISE Research Institute of Sweden Brinellgatan 4, Borås, Swedan Airflow (m 3 /s) Dust fed (g) Dust fed (g) NOTE: The results of this test relate only to the test device in the condition stated herein. The performance results cannot by themselves be quantitatively applied to predict filtration performance in all "real life" environments.
12 P rev1 3 (4) Appendix 3 ISO : Air flow rate and resistance to air flow after different dust loading phases Test device: Test no.: Test dust: Hi Flo II XLT 7/ (HFGX-F7-592/592/ ) SP Particle Technology, ISO A2-fine, Batch nr: 9173 Air flow rate: m 3 /s Date Loaded Air flow meter dust Filter m tot t f p sf p f q m t ϕ p a ρ q v p p 1.20 g C Pa Pa kg/s C % kpa kg/m 3 m 3 /s Pa Pa Clean filter Clean filter pressure drop is proportional to (q v ) n, where n = 1.28 Dust loading phase Symbols and units p f air flow meter differential pressure, Pa q m mass flow rate, kg/s m tot cumulative mass of dust fed to filter, g q v air flow rate filter, m 3 /s p measured filter pressure drop, Pa t f temperature at air flow meter, C p 1.20 resistance to air flow at air density 1.20 kg/m 3, Pa t temperature upstream of filter, C p a absolute air pressure upstream of filter, kpa ϕ relative humidity upstream of the filter, % p sf air flow meter static pressure, kpa ρ air density upstream of filter, kg/m 3
13 P rev1 4 (4) Appendix 3 ISO : Resistance to air flow and arrestance after different dust loading phases Test device: Test no.: Test dust: Hi Flo II XLT 7/ (HFGX-F7-592/592/ ) SP Particle Technology, ISO A2-fine, Batch nr: 9173 Air flow rate: m 3 /s Date p 1 m m tot Δp 2 m 1 m 2 Δm ff m d A j A m Pa g g Pa g g g g % % Mass of tested device Initial mass of tested device: Final mass of tested device: g g Test dust: Particle Technology, ISO A2-fine, Batch nr: 9173 Symbols and units A j arrestance, % A m average arrestance, % m dust increment, g p 1 resistance to air flow before dust increment (air density 1.20 kg/m³), Pa p 2 resistance to air flow after dust increment (air density 1.20 kg/m³), Pa m d m 1 m 2 m tot m ff dust in duct after device, g mass of final filter before dust increment, g mass of final filter after dust increment, g cumulative mass of dust fed to filter, g mass gain of final filter, g
14 P rev1 1 (4) Appendix 4 ISO : AIR FILTER TEST RESULTS SUMMARY GENERAL Test ID: SP Date of test: Operator: CM Particle counter information Air flow measurement: Device obtained (when and how Manufacturer: Model: Coincidence value Annubar, Micatrone obtained): TSI Gmbh OPS 3330 (p/cm 3 ): Air flow sensor MFS-SS The device was sent and obtained on DEVICE TESTED Model: Hi Flo II Manufacturer: Construction: XLT 7/ (HFGX-F7-592/592/ ) Camfil AB Pocket filter, 10 Pockets Article number: Type of media: Net effective media area (m 2 ) Filter dimensions (width x height x depth) Glass 7.3 m 2 592x592x640 mm Filter/media electrostatic charge: Media colour: No green Device Condition: Conditioned per ISO Testing Organization: RISE Research Institute of Sweden Brinellgatan 4, Borås, Swedan Media adhesive: N/A Other descriptive information: TEST DATA SUMMARY Test air flow rate: Test air temperature: Test air RH: Test aersole: m 3 /s C % DEHS (0.3-1 µm) KCl (1-10 mm) RESULTS Resistance to airflow (Pa) Fractional Efficiency (%) Measured: 72 Pa Rated initial: - Measured Rated Upstream concentration Range (µm) Rated Final: - Efficiency Efficiency (particles / dm 3 ) Test item photo Remarks: NOTE: The results of this test relate only to the test device in the condition stated herein. The performance results cannot by themselves be quantitatively applied to predict filtration performance in all "real life" environments.
15 Pressure differential, 1.2 kg/m 3 (Pa) Fractional efficiency (% ) REPORT P rev1 2 (4) Appendix 4 ISO : AIR FILTER TEST RESULTS DETAILS Airflow (m 3 /s) Resistance to Airflow (Pa) Conditioned Airflow (m 3 /s) Resistance to Airflow (Pa) Testing Organization: RISE Research Institute of Sweden Brinellgatan 4, Borås, Swedan Test ID: SP Date of test: Operator: CM TEST DATA DETAILS Resistance to Airflow, 1.2 kg/m 3 Fractional efficiency Initial E i, % nominal E d, % nominal Range (µm) air flow air flow Particle size (µm) E d, 50% nominal air flow Ed, % nominal air flow Ei, % nominal air flow Ed, 50% nominal air flow Airflow (m 3 /s) Pressure differential as a function of the air flow rate, Initial Pressure differential as a function of the air flow rate, Conditioned CONDITIONING PROCEDURE Date: Temperature in the chamber: C Relative humidity in the chamber: % Atmospheric pressure: - NOTE: The results of this test relate only to the test device in the condition stated herein. The performance results cannot by themselves be quantitatively applied to predict filtration performance in all "real life" environments.
16 P rev1 3 (4) Appendix 4 Efficiency measurement Upstream count data OPC bin d a, i Upstream efficiency count data µm U e,tot Note: All data shown is the number of particle counts for 60 s Efficiency measurement Downstream count data OPC bin d a, i Downstream efficiency count data µm D e,tot Note: All data shown is the number of particle counts for 60 s Efficiency measurement Final results and uncertainty OPC bin d a, i Penetration data reduction Uncertainty limits Uncertainty Efficiency µm P δ e Static Dynamic Pass/Fail % Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass Pass.0 d a, i : Geometric mean diameter of a size range i, µm P a δ e the final penetration for a given particle size the standard deviation of the penetration for a given particle size the uncertainty of the penetration for a given particle size
17 P rev1 4 (4) Appendix 4 Efficiency measurement, 50% nominal air flow Upstream count data OPC bin d a, i Upstream efficiency count data µm U e,tot Note: All data shown is the number of particle counts for 60 s Efficiency measurement, 50% nominal air flow Downstream count data OPC bin d i Downstream efficiency count data µm D e,tot Note: All data shown is the number of particle counts for 60 s Efficiency measurement, 50% nominal air flow Final results and uncertainty OPC bin d a, i Penetration data reduction Uncertainty limits Uncertainty Efficiency µm P δ e Static Dynamic Pass/Fail % Pass Pass Pass Pass 76.3 d a, i : Geometric mean diameter of a size range i, µm P a δ e the final penetration for a given particle size the standard deviation of the penetration for a given particle size the uncertainty of the penetration for a given particle size
18 P rev1 1 (1) Appendix 5 Fig 1. Overview of the test item. Fig 2. Label on the test item. Fig 3. Media in the test item.
19 P rev1 1 (1) Appendix 6 The interpretation of test reports according to ISO 16890:2016 This brief review of the test procedures, including those for addressing the testing of electrostatic charged filters, is provided for those unfamiliar with the procedures of this series of ISO standards. It is intended to assist in understanding and interpreting the results in the test report/summary. (For further details of procedures the full ISO document series shall be consulted). Air filters may rely on the effects of passive static electric charges on the fibres to achieve high efficiencies, particularly in the initial stages of their working life. Environmental factors encountered in service may affect the action of these electric charges so that the initial efficiency may drop substantially after an initial period of service. This could be offset or countered by an increase in efficiency ( mechanical efficiency ) as dust deposits build up. The reported, untreated and conditioned (discharged) efficiency shows the extent of the electrical charge effect on initial performance and indicates the potential loss of particle removal efficiency when the charge effect is completely removed and when at the same time there is no compensating increase of the mechanical efficiency. These test results should not be assumed to represent the filter performance in all possible environmental conditions or to represent all possible real life behavior.
REPORT issued by an Accredited Testing Laboratory
issued by an Accredited Testing Laboratory Contact person RISE Maria Nilsson Tengelin 2018-01-09 MTt7P09138 1 (2) Measurement Science and Technology +46 516 54 51 maria.nilssontengelin@ri.se Universelt
More informationDetermination of sound absorption coefficient in impedance tubes according to ISO (2 appendices)
issued by an Accredited Testing Laboratory Date Reference Page 2011-05-16 PX03759-02-5A 1 (3) Contact person Xuetao Zhang Energy Technology +46 10 516 50 21, xuetao.zhang@sp.se SP Wood Technology Att:
More informationCALIBRATION CERTIFICATE # 503
CALIBRATION CERTIFICATE # 503 Calibration date : 2011-02-23 Certificate issued : 2012-04-18 Company name Company address City, Province, Canada CLAS 2009-02 Calibration of Mass flow meter Micro Motion
More informationREPORT. Client. Test object. Date of testt. Results. Rönnåsgatan 5B Ulricehamn. (7 appendices) content of the
Riss sued by an Accreditedd Testing Laboratory Contact person Malin Lindgren Sustainable Builtt Environment +46 10 516 50 722 malin.lindgren@ @sp.se 2016-01-11 5P08641-A/Engg 1 (8)) Stellan Bygård Rönnåsgatan
More informationEmission measurements according to, M1 (3 appendices)
issued by an Accredited Testing Laboratory Contact person Ulrika Johansson 2017-01-30 6F024225-1 1 (6) Chemistry, Materials and Surfaces +46 10 516 53 22 ulrika.johansson@sp.se Accred. No. 1002 Testing
More informationOmega Chemicals Emissions Testing Report
Omega Chemicals 18-Feb-2015 Doc No. 60336610_1.1_RPT Omega Chemicals Emissions Testing Report February 2015 NATA ACCREDITATION No. 2778 (14391) Accredited for compliance with ISO/IEC 17025 This document
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 8426 Second edition 2008-02-01 Hydraulic fluid power Positive displacement pumps and motors Determination of derived capacity Transmissions hydrauliques Pompes et moteurs volumétriques
More informationMason Technology Ltd
Mason Technology Ltd 228 South Circular Road, Dublin 8 Calibration Laboratory Registration number: 043C is accredited by the Irish National Board (INAB) to undertake calibration as detailed in the Schedule
More informationGuidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018
Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018 Guidelines on the Calibration of Automatic Instruments for
More informationData Sheet CAPACITIVE ABSOLUTE 1.2 BAR SCB10H-B012FB PRESSURE SENSOR ELEMENT. Features. Applications. General Description
Data Sheet CAPACITIVE ABSOLUTE 1.2 BAR SCB10H-B012FB PRESSURE SENSOR ELEMENT Features Small size 1.4 x 1.4 x 0.85 mm (w x l x h) High isolation resistance and low passive capacitance enables very low power
More informationCommenting template CEN/TC 102 Date : Document: pren 13060
Commenting template CEN/TC 0 Date : 006-09-3 Document: pren 3060 MB Clause / 3.) ) FR general Ge There have been so many discussions during the CEN/TC 0/WG5 meetings these years. The committee has recognised
More informationCERTIFICATE OF ACCREDITATION
CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Hards Laboratories cc, t/a Technology Solutions
More informationWindow air permeability measurement, watertightness and resistance to wind load (designation of the test)
KTU INSTITUTE OF ARCHITECTURE AND CONSTRUCTION BUILDING THERMAL PHYSICS Notified Body number: 2018 BANDYMAI ISO/IEC 17025 Nr. LA. 01.031 Test performed: Product: Page (pages) 05 of August 2013 1 (8) Window
More informationResearch and Development on European Cleaning Solutions for Radioactive Applications. Clean air solutions
Research and Development on European Cleaning Solutions for Radioactive Applications Dipl. Ing. Axel Mahler, R&D Director Europe, Camfil AB, Sweden 19.03.2014 Your speaker: Axel Mahler, R&D Director Europe
More informationTest Report EN (2008)
Test Report EN 62233 (2008) Product Name and address of the applicant Name and address of the manufacturer Model Rating Brand name Heating elements for building into floors Advanced Heating Technologies
More informationMason Technology Ltd
Mason Technology Ltd 228 South Circular Road, Dublin 8 Calibration Laboratory Registration number: 043C is accredited by the Irish National Board (INAB) to undertake calibration as detailed in the Schedule
More informationSAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0
076-0-067-ACC-USL-S Mettler Toledo, LLC 900 Polaris Parkway Columbus, OH 0.800.METTLER Customer Accredited by the American Association for Laboratory Accreditation (ALA) CALIBRATION CERT #788.0 SAMPLE
More informationSAMPLE. Accuracy Calibration Certificate. Mettler Toledo 1900 Polaris Parkway Contact: Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC WI v.1.
076-00-067-ACC-S Mettler Toledo, LLC 900 Polaris Parkway Columbus, OH 0.800.METTLER Customer SAMPLE Accuracy Calibration Certificate Company: Address: City: Zip / Postal: State / Province: Mettler Toledo
More informationOEM Silicon Pressure Die
OEM Silicon Pressure Die SM9520 Series FEATURES High volume, cost effective Gauge configuration Constant current or constant voltage drive Millivolt output Available in 0.15, 0.60 & 1.50 PSIG full-scale
More informationDetermination of particle size distribution Single particle light interaction methods. Part 4:
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 21501-4 Second edition 2018-05 Determination of particle size distribution Single particle light interaction methods Part 4: Light scattering airborne
More informationInterlaboratory Comparison & Proficiency Testing Program
Extended scope for the organization of proficiency tests in accordance with the requirements of ISO/IEC 17043 VSL, the Dutch Metrology Institute, has significantly extended the scope for their accreditation
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: Palenque 1 A, Bello Horizonte,. (Hereinafter called
More informationAckred. nr 1678 Kalibrering ISO/IEC 17025
Calibration Tvinnargatan 25 SE-507 30 Brämhult SWEDEN Phone: +4633 22 56 30 info@kalibrera.se www.kalibrera.se Ackred. nr 1678 Kalibrering ISO/IEC 17025 Our calibration lab is accredited by Swedac for:
More informationMedium Pressure Sensor Analog Output
Medium Pressure Sensor Analog Output SM6844-015-A-B-5-S FEATURES Analog pressure calibrated and temperature compensated output Amplified analog output Compensated temperature range: 0 to 85oC Absolute
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
Laboratory locations: Schedule of Accreditation 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Unit 1G Kilroot Business Park Carrickfergus Co Antrim BT38 7PR Contact: Mr R E Hughes Tel:
More informationUSING THE ELECTRET FILTER TO REMOVE THE SUBMICRON AEROSOLS
USING THE ELECTRET FILTER TO REMOVE THE SUBMICRON AEROSOLS SH Yang 1,*, GWM Lee 1, CH Luo 2, CC Wu 1, KP Yu 1, CL Lou 1 1 Graduate Institute of Environmental Engineering, National Taiwan University, 71,
More informationVOC TEST REPORT CDPH
Adolf Würth GmbH und Co. KG Reinhold-Würth Str. 12-17 74653 Künzelsau GERMANY Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing VOC TEST REPORT
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. ALLOMETRICS, INC Bayport Blvd. Seabrook, TX Terry Baldwin Phone: CALIBRATION
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ALLOMETRICS, INC. 2500 Bayport Blvd. Seabrook, TX 77586 Terry Baldwin Phone: 281 474 3329 CALIBRATION Valid To: April 30, 2019 Certificate Number: 2039.01 In
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares
More informationHigh Pressure fi lters. FHB series. Maximum working pressure up to 32 MPa (320 bar) - Flow rate up to 485 l/min
High Pressure fi lters FHB series Maximum working pressure up to MPa (0 bar) - Flow rate up to 8 l/min High Pressure fi lters Introduction FILTER SIZING INDEX CALCULATION CORRECTIVE FACTOR Page Introduction
More informationChapter 7 Separation of Particles from a Gas
Chapter 7 Separation of Particles from a Gas For either gas cleaning (removal of dusts) or recovery of particulate products Separation Mechanisms Sedimentation : Settling chamber, centrifuge Migration
More informationSMD Switch SI-C3436A General specification. Soldering conditions Reflow Soldering condition. Standard Reel Dimensions (mm) Packing specification
SMD Switch SI - C3436A - 7017 Features External Dimensions : 3.4mm x 3.6mm, Height 0.70mm Suitable for thinner and lighter portable devices Suitable for water proof devices Reflow soldering Package With
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: AMC LABS S.A de C.V (Hereinafter called the Organization)
More informationData Sheet CAPACITIVE ABSOLUTE 1.2 BAR SCB10H-B012FB PRESSURE SENSOR ELEMENT. Features. Applications. General Description
Data Sheet CAPACITIVE ABSOLUTE.2 BAR SCB0H-B02FB PRESSURE SENSOR ELEMENT Features Small size.4 x.4 x 0.85 mm (w x l x h) High isolation resistance and low passive capacitance enables very low power consumption,
More informationTest report CVN Project number Page 2 of 2
Project number 604545 Page 2 of 2 Tests : Results : The diaphragm gas meters as specified in annex 2 were tested for compliance with the European Standard BS EN 1359 : 1999 "Gas meters - Diaphragm gas
More informationVK2 volume flow controller Overview
New option with sound attenuator Maintenance-free VK2 volume flow controller that operates without an auxiliary power supply, for ventilation and air conditioning systems. Adjustable on site. Outstanding
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of:. Av. Rio Consulado # 2614, Col. San Juan de Aragón
More informationAccuracy Calibration Certificate
gl06-0-067-acc Mettler Toledo Service Business Unit Industrial 900 Polaris Parkway Columbus, OH 0-800-- Services Provided By: Customer Accuracy Calibration Certificate Company: Sample Company Test Address:
More informationTurbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas
Turbine Meter TRZ 0 PRODUCT INFORMATION Reliable Measurement of Gas TURBINE METER TRZ 0 Method of operation, Construction Method of operation The TRZ 0 turbine meter is a flow meter suitable for gas measurement
More informationProblem 1 in the section "self-thinking" Team «MG 12»
Problem 1 in the section "self-thinking" Team «MG 12» 1 Problem 1 in the section "self-thinking" Calculate the error in the measurement of mass of object by balance, taking into account external factors.
More informationTraceability research activities in the field of airborne particle number/ mass concentration measurement Liu Junjie, Zhang Wenge, Song Xiaoping
Traceability research activities in the field of airborne particle number/ mass concentration measurement Liu Junjie, Zhang Wenge, Song Xiaoping Division of Nano metrology and Materials Measurement National
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 CONTROL COMPANY 12554 Galveston Road Suite B230 Webster, TX 77598 Aaron Judice Phone: 281 482 1714 CALIBRATION Valid To: March 31, 2018
More informationProvläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition
INTERNATIONAL STANDARD ISO 17123-4 Second edition 2012-06-01 Optics and optical instruments Field procedures for testing geodetic and surveying instruments Part 4: Electro-optical distance meters (EDM
More informationTYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE
TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE Michael Bair and Pierre Delajoud 2002 DEC 10 Revised 2004 MAR 26 FORWARD FPG8601 is a pressure standard designed
More informationDirect reading aerosol monitors
Health and Safety Executive Direct reading aerosol monitors 27 th June 2018 Andrew Thorpe Introduction Direct reading Aerosol Monitors (DRAMs) are often used as a complementary approach to conventional
More informationStandard Practices for Air Speed Calibration Testing
Standard Practices for Air Speed Calibration Testing Rachael V. Coquilla Bryza Wind Lab, Fairfield, California Air speed calibration is a test process where the output from a wind measuring instrument
More information473 Dew Point Hygrometer
High Performance Chilled Mirror Hygrometer With Cable Mounted Measuring Heads Highly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct
More informationAddendum to ARI Standard , Air Terminals September 2002
Addendum to ARI Standard 880-98, Air Terminals September 2002 The September 2002 Addendum to ARI Standard 880-98, Air Terminals, is comprised of only the shaded portions shown, which has been incorporated
More informationInterlaboratory Comparison & Proficiency Testing Program
Extended scope for the organization of proficiency tests in accordance with the requirements of ISO/IEC 17043 VSL, the Dutch Metrology Institute, has significantly extended the scope for their accreditation
More informationTEST REPORT RP
Page 1 of 36 Environmental and material tests (ENV) SECTOR Product description: Tested Models: Manufacturer: Led lights Sample 01 -> Led Puma -> 28234/LED/AS Sample 02 -> Extreme Led -> 556600_04 Sample
More informationPMLCAP. Polymer Multi Layer Capacitor Specification Sheet RUBYCON CORPORATION PML DIVISION. PMLCAP ST series RPR Rubycon PART No. Drawing No.
To Polymer Multi Layer Capacitor Specification Sheet Rubycon PART No. ST series Drawing No. RPR-0024 Issued Date October 15, 2008 PML DIVISION 2932, MOTOOSHIMA, MATSUKAWA-MACHI, SHIMOINA-GUN, NAGANO-KEN,
More informationCERTIFICATE OF CALIBRATION
Page 1 of 10 CALIBRATION OF Sound Level Meter: Brüel & Kjær Type 3050-A-040 No: 3050-100751 Id: - Microphone: Brüel & Kjær Type 4189 No: 2621142 Preamplifier: Brüel & Kjær Type 2669 No: 2679073 Supplied
More informationCERTIFICATE OF ACCREDITATION
CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Compania Nacional de Metrologia S.A.S - Conamet
More informationHigh Pressure fi lters. FHA 051 series. Maximum working pressure up to 56 MPa (560 bar) - Flow rate up to 140 l/min. High Pressure fi lters
High Pressure fi lters FHA 0 series Maximum working pressure up to 6 MPa (60 bar) Flow rate up to 0 l/min 99 High Pressure fi lters Introduction FILTER SIZG DEX CALCULATION CORRECTIVE FACTOR Page Introduction
More informationComparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576
NPL REPORT CMAM 73 Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576 Peter Hanes October 2001 The National Physical Laboratory is operated on behalf
More informationEUROVENT 8/12 SOUND TEST METHOD FOR DUCTED FAN COIL UNITS. FCP 2008 testing procedures FCP acoustical test method pag. 1
EUROVENT 8/12 SOUND TEST METHOD FOR DUCTED FAN COIL UNITS FCP 2008 testing procedures FCP acoustical test method pag. 1 CONTENTS 1 - PURPOSE... 3 2 - NORMATIVE REFERENCES... 3 3 - DEFINITIONS... 3 3.1
More informationTechnical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017
MICROMATTER TECHNOLOGIES INC. #1-8333 130 th Street, Surrey, B.C. V3W 7X4, Canada Phone: +1.604.594.9720 or 1.866.966.2286 Fax: +1.604.594.9726 info@micromatter.com www.micromatter.com Technical Note 2015-01A
More informationCertificate of Accreditation to ISO/IEC 17025:2005
United States Department of Commerce National Institute of Standards and Technology Certificate of Accreditation to ISO/IEC 17025:2005 NVLAP LAB CODE: 200508-0 Insco Metrology, Inc. Miami, FL is accredited
More information473 Dew Point Hygrometer
gh Performance Chilled Hygrometer With Cable Mounted Measuring Heads ghly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct insertion
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: Ciudad Madero, Tamaulipas, México. C.P. 89520 (Hereinafter
More informationInterlaboratory Comparison & Proficiency Testing Program
Extended scope for the organization of proficiency tests in accordance with the requirements of ISO/IEC 17043 VSL, the Dutch Metrology Institute, has significantly extended the scope for their accreditation
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares
More informationv = m / s Thermal Flow Sensors TA with thin-film sensor element Probes Probe Guides Measuring Tubes
Thermal Flow Sensors TA with thin-film sensor element Probes Probe Guides Measuring Tubes Examples of Application Flow measurements also in hazardous locations X zone 0 or zone 1 in clean rooms in air
More informationApproval Block. Prepared by: Signature Date Evan Parnell 08 NOV Reviewed by: Signature Date. Approved by: Signature Date
ATS-SOI-3660 Page: 1 of 6 Approval Block Prepared by: Signature Date Evan Parnell 08 NOV 2013 Reviewed by: Signature Date Brian Flynn 08 NOV 2013 Approved by: Signature Date Kristal Jewell 08 NOV 2013
More informationTÜV RHEINLAND ENERGY GMBH ADDENDUM
TÜV RHEINLAND ENERGY GMBH ADDENDUM Addendum to TÜV test report no. 936/21207522/A dated 23 March 2009 on performance testing of the SWAM 5a Dual Channel Monitor, SWAM 5a Monitor and SWAM 5a Dual Channel
More informationThe energy performance of an airflow window
The energy performance of an airflow window B.(Bram) Kersten / id.nr. 0667606 University of Technology Eindhoven, department of Architecture Building and Planning, unit Building Physics and Systems. 10-08-2011
More informationTEST REPORT Formaldehyde Reduction
Sigma Coatings Amsterdamseweg 14 1422 AA Uithoorn The Netherlands Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing TEST REPORT Formaldehyde
More informationSAIOH Tutorial Ventilation 1 pressures and basic air flow calculations
SAIOH Tutorial Ventilation 1 pressures and basic air flow calculations Acknowledgement This tutorial was provided by SAIOH as an assessment support aid for prospective candidates. The tutorial is free
More informationMulti Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE
1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for
More informationCalculation of Pressure Drop Across Sunray Louvre. Prepared for: David Evans Sunray Engineering Ltd. November Client report number
Calculation of Pressure Drop Across Sunray Louvre Prepared for: David Evans Sunray Engineering Ltd November 2009 Client report number 258-256 1 Sunray Louvre Calculations Prepared by Name Gordon Breeze
More informationISO 5136 INTERNATIONAL STANDARD. Acoustics Determination of sound power radiated into a duct by fans and other air-moving devices In-duct method
INTERNATIONAL STANDARD ISO 5136 Second edition 2003-04-01 Acoustics Determination of sound power radiated into a duct by fans and other air-moving devices In-duct method Acoustique Détermination de la
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 21501-2 First edition 2007-05-15 Determination of particle size distribution Single particle light interaction methods Part 2: Light scattering liquid-borne particle counter
More informationNew Instruments from GRIMM
New Instruments from GRIMM Markus Pesch Symposium at Stockholm, 01.11.2016 1 Outline Motivation for new Aerosol measurement devices Objectives for measurements of particle number and sizes GRIMM EDM 665
More informationOIML R 50-3 RECOMMENDATION. Edition 2014 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION
INTERNATIONAL RECOMMENDATION OIML R 50-3 Edition 2014 (E) Continuous totalizing automatic weighing instruments (belt weighers). Part 3: Test report format Instruments de pesage totalisateurs continus à
More informationBrian Sargent, Assistant Director, Designee 3/21/18. See Thread for Approval
Cali Jones Cali Jones, Coordinator 3/21/18 Brian Sargent Brian Sargent, Assistant Director, Designee 3/21/18 See Email Thread for Approval 3/21/18 @ 3:30 p.m. 3/26/18 @ 3:30 p.m. 1818 Cali Jones Cali Jones
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: AMC LABS S.A de C.V Av. México # 329 Col. Valle De
More informationSonic Anemometer Classification
Summary report VT121243 - Version 1.1 Sonic Anemometer Classification According to (2017-03) Classification Scheme Description of Anemometer Manufacturer: Adolf Thies GmbH & Co. KG Hauptstrasse 76 37083
More informationNOVALYNX CORPORATION MODEL 110-WS-16BP BAROMETRIC PRESSURE SENSOR INSTRUCTION MANUAL
NOVALYNX CORPORATION MODEL 110-WS-16BP BAROMETRIC PRESSURE SENSOR INSTRUCTION MANUAL REVISION DATE: OCT 2005 Receiving and Unpacking Carefully unpack all components and compare to the packing list. Notify
More informationModel Selection 1. Process Valves. Air. Vacuum
Process Valves Model Selection For product specifications such as maximum operating pressure differentials and operating temperature ranges, refer to the relevent pages of each product. Air One-touch fittings
More informationv = 00,20 m/s Thermal Flow Sensors TA with thin-film sensor element Probes Probe guides Measuring tubes
with thin-film sensor element Probes Probe guides Measuring tubes Examples of Application Flow measurements also in hazardous locations Ex zone 0 or zone 1 in clean rooms in air conditioning and ventilating
More informationTEST METHOD FOR STILL- AND FORCED-AIR JUNCTION-TO- AMBIENT THERMAL RESISTANCE MEASUREMENTS OF INTEGRATED CIRCUIT PACKAGES
SEMI G38-0996 N/A SEMI 1987, 1996 TEST METHOD FOR STILL- AND FORCED-AIR JUNCTION-TO- AMBIENT THERMAL RESISTANCE MEASUREMENTS OF INTEGRATED CIRCUIT PACKAGES 1 Purpose The purpose of this test is to determine
More informationALUMINIUM CENTRIFUGAL FANS. 4 x M6 x OUTLET x M6 x 20
DESCRIPTION Dutair LK-series industrial ventilators consist of a robust body made from die-cast aluminium alloy. Impellers are made of extruded aluminium. The asynchronous electric motors are maintenance
More informationALUMINIUM CENTRIFUGAL FANS x M6 x x M6 x 16
LK4 DESCRIPTION Dutair LK-series industrial ventilators consist of a robust body made from die-cast aluminium alloy. Impellers are made of extruded aluminium. The asynchronous electric motors are maintenance
More informationETS-LINDGREN ACOUSTIC RESEARCH LABORATORY OFFICIAL LABORATORY REPORT ARL-SA6956 Revision 1 ARL
ETS-LINDGREN ACOUSTIC RESEARCH LABORATORY OFFICIAL LABORATORY REPORT -SA6956 Revision 1 Subject: Sound Absorption Test Date: 04 January 2017 Contents: on Sound Absorption Data, One-third Octave bands Sound
More informationINSTRUMENT TEST REPORT 679
INSTRUMENT TEST REPORT 679 The Uncertainty of the Hydrological Services Field Calibration Device Authorisation J.D. Gorman Regional Instrument Centre 11 October 2004 Bruce W Forgan Superintendent Atmosphere
More informationCERTIFICATE OF. Single Analyte Mass Spec Solution ± 0.1 µg/ml g/ml (measured at 20 ± 4 ºC)
CERTIFICATE OF ANALYSIS 300 Technology Drive Christiansburg, VA 24073. USA inorganicventures.com tel: 800.669.6799. 540.585.3030 fax: 540.585.3012 info@inorganicventures.com 1.0 ACCREDITATION / REGISTRATION
More informationSwemaMan 7 Operating instructions vers 1.13 MB_CW
SwemaMan 7 Operating instructions vers 1.13 MB_CW20180425 WARNING! Before start measuring please read chapter 6. Settings (Set). At delivery k2-factor is turned on. SWEMA AB Pepparv. 27 123 56 FARSTA Tel:
More informationTypical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge
Typical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge 2090TN05 Revised 2004 March 26 Technical Note FPG8601 is a pressure standard designed to cover the range from
More informationLow Pressure Sensor Amplified Analog Output SM6295-BCM-S
Low Pressure Sensor Amplified Analog Output SM6295-BCM-S-040-000 FEATURES Pressure range from 0 to 40 cmh 2 O 5.0 V operation Amplified analog output (10 to 90%Vdd) Compensated temperature range: 0 to
More informationclose to performances of the turbine gas meters designed for custody transfer measurements.
QUANTOMETER DESIGN AND FUNCTION CPT Quantometers have been designed in order to provide our customers with reliable and inexpensive measuring instruments for secondary flow measurements. Our great experience
More informationParticle number concentration standards at NMIJ, including a new calibration service with an inkjet aerosol generator.
Particle number concentration standards at NMIJ, including a new calibration service with an inkjet aerosol generator Hiromu Sakurai hiromu.sakurai@aist.go.jp Particle Measurement Research Group National
More informationISO INTERNATIONAL STANDARD. Determination of particle size distribution Differential electrical mobility analysis for aerosol particles
INTERNATIONAL STANDARD ISO 15900 First edition 2009-05-15 Determination of particle size distribution Differential electrical mobility analysis for aerosol particles Détermination de la distribution granulométrique
More informationFabric RUNNER Manufacturer Gabriel A/S
Müller-BBM GmbH Robert-Koch-Str. 11 82152 Planegg bei München Telephone +49(89)85602 0 Telefax +49(89)85602 111 www.muellerbbm.com Telephone +49(89)85602 3206 Eva.Mueller@mbbm.com 2015-03-03 Fabric RUNNER
More informationExperimental measurement of parameters governing flow rates and partial saturation in paper-based microfluidic devices
Experimental measurement of parameters governing flow rates and partial saturation in paper-based microfluidic devices Dharitri Rath 1, Sathishkumar N 1, Bhushan J. Toley 1* 1 Department of Chemical Engineering
More information405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook
405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Book and Flow Handbook www.rosemount.com 405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Book NOTICE Read
More informationDS 300. Electronic Pressure Switch. with IO-Link interface. Stainless Steel Sensor. accuracy according to IEC 60770: 0.35 % FSO.
DS 00 Electronic Pressure Switch with IO-Link interface Stainless Steel Sensor accuracy according to IEC 60770: 0.5 % FSO Nominal pressure from 0... 00 mbar up to 0... 600 bar Digital output signal IO-Link
More informationComparison protocol EURAMET project
Comparison protocol EURAMET project Bilateral three-phase AC and energy comparison at frequency 1. Introduction This bilateral comparison was triggered by MIKES need to get support for their planned extension
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: Asesoría Industrial en Mantenimiento y Calibración
More informationLoadcell Calibration - Evaluation of Uncertainties
Loadcell Calibration - Evaluation of Uncertainties This document is a revision of the evaluation issued in June 2015. Novatech s calibration laboratory is not UKAS accredited due to the high number of
More informationANSI/AHRI Standard (Formerly ARI Standard ) 2006 Standard for Performance Rating of Desuperheater/Water Heaters
ANSI/AHRI Standard 470-2006 (Formerly ARI Standard 470-2006) 2006 Standard for Performance Rating of Desuperheater/Water Heaters IMPORTANT SAFETY DISCLAIMER AHRI does not set safety standards and does
More information