Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C

Size: px
Start display at page:

Download "Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C"

Transcription

1 DOI /s Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C Domen Hudoklin Regina Mnguni Hans Liedberg Igor Pušnik Jovan Bojkovski Received: 30 July 2013 / Accepted: 4 July 2014 Springer Science+Business Media New York 2014 Abstract Traceability in the field of relative humidity (RH) measurements is typically assured indirectly through dew point and temperature scales. Conducting an interlaboratory comparison at the national metrology institute (NMI) level, using a direct approach with a precision RH hygrometer as a transfer standard would, therefore, be of a particular interest, especially if the measurement setups were of a different type. This paper presents an RH comparison at the NMI level between the National Metrology Institute of South Africa (NMISA) and University of Ljubljana, Faculty of Electrical Engineering, Laboratory of Metrology and Quality (MIRS/UL-FE/LMK). In scope of this inter-comparison, calibration of an impedance-type hygrometer in the range from 10 %rh to 95 %rh at air temperatures of 5 C, 25 C, and 55 C, respectively, was performed. It was recommended that the participants use their standard procedure for the calibration of RH sensors and, at the same time, follow the specific criteria of the review protocol for uncertainty estimation accepted by Bureau International des Poids et Mesures (BIPM), marked as BIPM CCT-WG8/CMC-10. An interesting part of the comparison was the two different calibration methods which were used by the two partners and which also have different traceability routes. MIRS/UL-FE/LMK calibrated the sensor in the humidity generator by comparison against the reference chilled mirror hygrometer, which is traceable to the MIRS/UL-FE/LMK primary dewpoint generator. NMISA calibrated the transfer standard against certified salt solutions, which were kept in a temperature-controlled chamber. Results showed acceptable agreement at all 15 calibration points. D. Hudoklin (B) I. Pušnik J. Bojkovski Metrology Institute of the Republic of Slovenia/University of Ljubljana-Faculty of Electrical Engineering/Laboratory of Metrology and Quality (MIRS/UL-FE/LMK), Ljubljana, Slovenia domen.hudoklin@fe.uni-lj.si R. Mnguni H. Liedberg National Metrology Institute of South Africa (NMISA), Pretoria, South Africa

2 Keywords Dew-point measurement Inter-laboratory comparison Relative humidity Salt solutions 1 Introduction Relative humidity (RH), as the most frequently measured humidity parameter, is predominantly measured directly by impedance-based hygrometers. Due to the limited accuracy, these types of hygrometers typically provide traceability indirectly through the dew point and temperature. This is also one of the reasons, why at the level of national metrology institutes (NMI), calibration and measurement capabilities (CMC) are inter-compared for dew-point realizations using dew-point hygrometers as transfer standards. On the other hand, RH calibration introduces additional intrinsic uncertainty components which result from both humidity and temperature measurements. Moreover, the uncertainty contribution of the RH hygrometer as a device under calibration (DUC) adds additional uncertainties, such as reproducibility, long-term instability, etc. For this reason, inter-laboratory comparison in the field of RH using an RH hygrometer as a transfer standard is of great interest, because it gives needed supporting evidence to CMC claims [1 3] of the laboratory and to hygrometry in general (also in terms of [4,5]), particularly because such intercomparisons at the level of NMIs are scarcely performed and reported [6 8]. The aim of this bilateral inter-laboratory comparison was to evaluate the metrological equivalence of NMISA s RH calibration capabilities, over a range of air temperatures from 5 Cto55 C, with those of LMK (short for MIRS/UL-FE/LMK), the NMI of Slovenia. The institutes are accredited according to ISO 17025:2005 and ISO 17025:2010, for humidity calibrations, respectively. NMISA acted as the pilot (coordinating) laboratory, while the LMK provided the comparison reference value. The participating laboratories used their standard procedure for the calibration of RH sensors and at the same time followed the specific criteria of the BIPM CCT-WG8/12-10 CMC review protocol for relative humidity. An interesting part of this comparison was the fact that the two RH realizations were based on two different methods; salt solutions [9] and primary dew-point generation [10 12], while for the transfer standard [13], a precision RH hygrometer with a capacitive sensor was used. 2 Technical Objectives of the Intercomparison The objective of this comparison was to assess the procedures and standards used in calibrating RH hygrometers, particularly at temperatures other than ambient. The transfer standard, Vaisala MI70/HMP77 (Serial Nos. D and D ) was provided by NMISA. It is a precision capacitive RH hygrometer with a known and stable calibration history. During the transport, the transfer standard artifact was packed in the accompanying protective case and shipped via courier. Both labs, first the NMISA followed by the LMK, calibrated the artifact at different RH set points of 10 %rh, 35 %rh, 50 %rh, 75 %rh, and 95 %rh, all at air temperatures other than ambient: 5 C, 25 C, and 55 C.

3 2.1 Calibration Facilities NMISA used unsaturated salt-solution ampoules, traceable to an accredited commercial calibration laboratory, as reference standards for RH measurements. The unsaturated salt solutions were supplied in sets of 5 or 50 in a box. Two ampoules of the same salt solution were poured onto clean fiber pads in a 100 ml stainless-steel chamber, placed in a large temperature-controlled chamber. The transfer standard was passed through the O-ring seals into a 100 ml stainless chamber. The measurements were done at three temperature points, i.e., 5 C, 25 C, and 55 C, however, not in a particular order for a set of ampoules. The ampoules used were for 10 %rh, 35 %rh, 50 %rh, 75 %rh, and 95 %rh salt solutions. The chamber was set to the three temperatures points consecutively, with typically a 5 h stabilization time for each set of measurements, and a total time for all three sets of 15 h. The traceability of the unsaturated salt solutions at different temperatures was provided through the supplier of the ampoules. At LMK, the transfer standards were calibrated in a humidity generator. The reference RH was calculated according to the Sonntag formulae [14] from the measured dew-point temperature and the temperature of the air. Both parameters were directly measured by a chilled mirror hygrometer MBW 373H (Serial No ). Traceability of the dew-point temperature is assured by a LMK primary dew-point generator [10 12]. Air-temperature measurements are traceable to an LMK temperature primary standard, realized by temperature fixed points. At each temperature point, 5 C, 25 C, and 55 C, RH points were set in rising and falling order to check for hysteresis of the instruments. At each set point, stability criteria were set according to the standard deviation of the calculated RH. 3 Method for Comparison Results Analysis Each participating lab determined corrections D lab,η,t of the artifact indication at each nominal measurement point of RH η and temperature t according to D lab,η,t = η ref,η,t η ind,η,t, (1) where η ref,η,t denotes the lab s reference RH at a nominal measurement point and η ind,η,t a RH indicated by the artifact. LMK s corrections served as the intercomparison reference value D RV,η,t for each nominal point D RV,η,t = D LMK,η,t, (2) where the index lab is substituted by LMK. For each correction, both labs provided also the expanded uncertainty (with coverage factor k = 2) of the correction, U(D lab,η,t ). The uncertainties were estimated according to the lab s calibration procedure taking into account also specific acceptance criteria for uncertainty budget (inter-laboratory-evaluation-based criteria) of the BIPM CCT-WG8/12-10 CMC review protocol [15]. The uncertainty calculation included the contribution due to hysteresis of ±0.3%rh[8] for the transfer standard.

4 As the LMK provided the reference value of the comparison, the expanded uncertainty of the comparison reference value U(D RV,η,t ), is taken as a combination of the expanded uncertainty U(D LMK,η,t ), where the index lab is substituted by LMK, and the expanded uncertainty due to the long-term instability of the artifact U drift : U ( D RV,η,t ) = U 2 ( D LMK,η,t ) + U 2 drift, (3) where U drift is estimated from a history of calibration by NMISA to 0.23 %rh by conservatively taking into account the worst case maximum value of differences between two corrections at the set points of 5 C and 75 % [8]. To compare the results of the two NMIs, individual differences D lab,η,t to the comparison reference value were calculated for each nominal point and for each lab D lab,η,t = D lab,η,t D RV,η,t, (4) To show bilateral equivalence of the comparison results, the differences from Eq. 4 were normalized to the expanded (k = 2) uncertainty of the bilateral difference that was calculated to obtain the value E n,η,t. E n,η,t = D NMISA,η,t U 2 ( ) D RV,η,t + U 2 ( ) D NMISA,η,t, (5) where the index lab is substituted only by NMISA, because the LMK provided the reference value and therefore LMK s D n,η,t is meaningful (zero values). 4 Results and Discussion Table 1 shows a summary of comparison results between the LMK and NMISA at different nominal points of RH and temperature. Results are graphically shown also in Figs. 1, 2, and 3. From Table 1, it can be concluded that results at all nominal points of RH and temperature meet the condition E n,η,t 1, which means that the differences to the reference value agree to within the expanded uncertainties with a coverage factor k = 2. From Figs. 1, 2, and 3, it can be seen that the agreement of the results is better at the extremes of the RH range, 10 %rh and 95 %rh at temperatures above ambient, 25 C and 55 C, respectively. At 5 C, however, the agreement is better in the middle of the range at 50 %rh with an E n,η,t value of Except for one point at 5 C and 50 %rh, the deviations do, however, show small systematic behavior. If the reason for that would be a deviation in either the dew-point or temperature measurement of the LMK setup, then the deviation would be higher at a higher relative humidity because of the higher sensitivity coefficient. Additionally, the deviations of the dew-point reference from the reference value obtained in the inter-laboratory comparison [12] are smaller or equal to C, which in the worst case results in a RH deviation of <0.1%rh. This is significantly smaller than the observed deviation in Figs. 1, 2, and 3.

5 Table 1 Summary of comparison results between LMK and NMISA at different nominal points of RH and temperature Nominal Nominal relative temperature, humidity, η (%rh) t ( C) LMK exp. unc., U(D LMK,η,t ) (%rh) NIMSA exp. unc., U(D NIMSA,η,t ) (%rh) NMISA difference, D NMISA,η,t (%rh) Normalized diff., E n,η,t D lab,,t - D RV,,t, %rh lab=nmisa lab=lmk Relative humidity, %rh Fig. 1 Comparison results at nominal temperature t = 5 C; bars on the graph show the expanded uncertainty (k = 2) of bilateral difference between the individual correction and reference value 5 Conclusion This paper presents an inter-laboratory comparison in the field of RH at the level of national metrology institutes. Set-ups of the two participating laboratories, NMISA and MIRS/UL-FE/LMK, used a different type of humidity generation and different traceability routes. The use of a single impedance-based RH hygrometer as a transfer standard showed to be sufficient for this bilateral inter-comparison. It is to be noted,

6 D lab,,t - D RV,,t, %rh lab=nmisa lab=lmk Relative humidity, %rh Fig. 2 Comparison results at nominal temperature t = 25 C; bars on the graph show the expanded uncertainty (k = 2) ofbilateral difference between theindividual correctionandreference value D lab,,t - D RV,,t, %rh lab=nmisa lab=lmk Relative humidity, %rh Fig. 3 Comparison results at nominal temperature t = 55 C; bars on the graph show the expanded uncertainty (k = 2) ofbilateral difference between theindividual correctionandreference value however, that for longer intercomparisons with more participants, it would be wise to use more transfer standards in order to avoid potential long-term instability. The results of this comparison show agreement with all E n,η,t values being <1, which is in accordance also with one of the specific criteria of the BIPM CMC review protocol CCT-WG8/ The participants agreed also to align their uncertainty estimations with other specific criteria of the mentioned protocol. References 1. D. Zvizdic, M. Heinonen, D. Sestan, Int. J. Thermophys. 33, 1536 (2012) 2. V. Fernicola, M. Banfo, L. Rosso, D. Smorgon, Int. J. Thermophys. 29, 1668 (2008) 3. M. Heinonen, D. Zvizdic, D. Sestan, Int. J. Thermophys. 33, 1451 (2012) 4. G. Begeš, J. Drnovšek, L.R. Pendrill, Accredit. Qual. Assur. 15, 147 (2010) 5. G. Begeš, H.J. Dalsgaadr, J. Drnovšek, J. Test. Eval. 36, 345 (2008) 6. M. Stevens, R. Benyon, S.A. Bell, T. Vicente, Int. J. Thermophys. 29, 1685 (2008)

7 7. R.M. Gee, R.S. Farley, P. Sax, R. Benyon, M. Stevens, N. Böse, Int. J. Thermophys. 29, 1696 (2008) 8. D. Jonker, M. Heinonen, H.G. Liedberg, M.R. Mnguni, Int. J. Thermophys. 33, 1458 (2012) 9. D. Jonker, M.R. Mnguni, H.G. Liedberg, Int. J. Thermophys. 29, 1644 (2008) 10. D. Hudoklin, J. Bojkovski, J. Nielsen, J. Drnovšek, Measurement 41, 950 (2008) 11. D. Hudoklin, J. Drnovšek, Int. J. Thermophys. 29, 1652 (2008) 12. M. Heinonen, M. Anagnostou, S. Bell, M. Stevens, R. Benyon, R. Anita Bergerud, J. Bojkovski, R. Bosma, J. Nielsen, N. Böse, P. Cromwell, A. Kartal Dogan, S. Aytekin, A. Uytun, V. Fernicola, K. Flakiewicz, B. Blanquart, D. Hudoklin, P. Jacobson, A. Kentved, I. Lóio, G. Mamontov, A. Masarykova, H. Mitter, R. Mnguni, J. Otych, A. Steiner, N. Szilágyi Zsófia, D. Zvizdic, Int. J. Thermophys. 33, 1422 (2012) 13. A.B. Kentved, M. Heinonen, D. Hudoklin, Int. J. Thermophys. 33, 1408 (2012) 14. D. Sonntag, Meteorol. Z. 3, 51 (1994) 15. BIPM web page.

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years Int J Thermophys (2017) 38:161 DOI 10.1007/s10765-017-2291-x TEMPMEKO 2016 Development of the High-Temperature Dew-Point Generator Over the Past 15 Years R. Bosma 1 J. Nielsen 2 A. Peruzzi 1 Received:

More information

Uncertainty Evaluation of the New Setup for Measurement of Water-Vapor Permeation Rate by a Dew-Point Sensor

Uncertainty Evaluation of the New Setup for Measurement of Water-Vapor Permeation Rate by a Dew-Point Sensor Int J Thermophys (2012) 33:1595 1605 DOI 10.1007/s10765-012-1213-1 Uncertainty Evaluation of the New Setup for Measurement of Water-Vapor Permeation Rate by a Dew-Point Sensor D. Hudoklin J. Šetina J.

More information

CCT Working Group for Humidity

CCT Working Group for Humidity CCT/17-41 CCT Working Group for Humidity Report to the CCT 2017 Stephanie Bell CCT Working Group for Humidity Terms of reference: to advise the CCT on matters relating to humidity; to pursue harmonization

More information

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations Calibration General Principles & Theory, Equipment Considerations Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link

More information

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length Yun-Kyung Bae and In-Jik Jeong Industrial & Physical Instrument Center, Korea

More information

EURAMET P1061 COMPARISON OF AIR TEMPERATURE CALIBRATIONS

EURAMET P1061 COMPARISON OF AIR TEMPERATURE CALIBRATIONS EURAMET P1061 COMPARISON OF AIR TEMPERATURE CALIBRATIONS Final Report version 4 M. Heinonen, M. Anagnostou, J. Bartolo, S. Bell, R. Benyon, R. A. Bergerud, J. Bojkovski, N. Böse, V. Fernicola, K. Flakiewicz,

More information

CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR

CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR Davor Zvizdic, Tomislav Stasic, Lovorka

More information

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR Jin Wan Chung 1, Shih Mean Lee, Sungjun Lee 1 1 Korea Research Institute of Standards

More information

Final Report August 2010

Final Report August 2010 Bilateral Comparison of 100 pf Capacitance Standards (ongoing BIPM key comparison BIPM.EM-K14.b) between the CMI, Czech Republic and the BIPM, January-July 2009 J. Streit** and N. Fletcher* *Bureau International

More information

COMPARISON OF INMETRO AND INTI HUMIDITY STANDARDS. Final Report

COMPARISON OF INMETRO AND INTI HUMIDITY STANDARDS. Final Report SIM.T-K6.4 INMETRO / INTI Final Report COMPARISON OF INMETRO AND INTI HUMIDITY STANDARDS Final Report Instituto Nacional de Metrologia, Qualidade e Tecnologia (INMETRO) Instituto Nacional de Tecnología

More information

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014 Calibration General Principles & Theory, Equipment Considerations Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link

More information

A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C

A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C Int J Thermophys (2017) 38:37 DOI 10.1007/s10765-016-2171-9 TEMPMEKO 2016 A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C R. Bosma 1 A. Peruzzi 1 R. Van Breugel 1 C. Bruin-Barendregt 1

More information

A bilateral comparison of a 1 kg platinum standard

A bilateral comparison of a 1 kg platinum standard DANSK FUNDAMENTAL METROLOGI DANISH FUNDAMENTAL METROLOGY Report on A bilateral comparison of a 1 kg platinum standard by Lars Nielsen Danish Fundamental Metrology B 307, Matematiktorvet, DK-800 Kgs. Lyngby

More information

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Introduction Humidity generators are mechanical devices that provide stable temperature and humidity

More information

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3.

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3. AFRIMETS Secretariat Private Bag X34 Lynnwood Ridge 0040 AFRIMETS Supplementary Comparison Programme Calibration of Gauge Blocks by Mechanical Comparison Method AFRIMETS.L S3 Final Report Giza, Egypt,

More information

DESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR

DESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR Copyright 2013 by ABCM DESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR Amaury Roteger de Oliveira Renato Nunes Teixeira Instituto Nacional de Metrologia, Qualidade e Tecnologia - Inmetro Diretoria

More information

REALIZATION OF HUMIDITY STANDARD FACILITY USING TWO-PRESSURE HUMIDITY GENERATOR AND HUMIDITY CHAMBER

REALIZATION OF HUMIDITY STANDARD FACILITY USING TWO-PRESSURE HUMIDITY GENERATOR AND HUMIDITY CHAMBER METROLOGY AND MEASUREMENT SYSTEMS Index 3393, ISSN 86-8229 www.metrology.pg.gda.pl REALIZATION OF HUMIDITY STANDARD FACILITY USING TWO-PRESSURE HUMIDITY GENERATOR AND HUMIDITY CHAMBER Essam El-Din, Mostafa

More information

Comparison of V and 10 V DC Voltage References

Comparison of V and 10 V DC Voltage References Comparison of 1.018 V and 10 V DC Voltage References Technical protocol for BIPM.EM-K11.a and.b comparisons 1. Introduction The Mutual Recognition Arrangement among National Metrology Institutes (NMIs)

More information

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 2017-03-06 revised 2018-02-13 (changed schedule) Task and Purpose of the Comparison According to the rules set up by the CIPM MRA the consultative

More information

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Technical Note The Fluke Calibration 5128A RHapid-Cal Humidity Generator provides a portable, stable test environment

More information

The Fundamentals of Moisture Calibration

The Fundamentals of Moisture Calibration The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity

More information

Metrology and Environment

Metrology and Environment Bridging the Gap Responding to Environmental Change: from Words to Deeds 14-16 May 2008, Portorož Metrology and Environment Prof Janko Drnovšek and Laboratory group Janko.Drnovsek@fe.uni-lj.si Basic dilemma

More information

Science of Stability 2017

Science of Stability 2017 Science of Stability 2017 Humidity Measurement in Stability Testing Applications Defining Uncertainty and the Challenges at Higher Temperatures Humidity Measurement In Stability Testing Applications Some

More information

Comparison protocol EURAMET project

Comparison protocol EURAMET project Comparison protocol EURAMET project Bilateral three-phase AC and energy comparison at frequency 1. Introduction This bilateral comparison was triggered by MIKES need to get support for their planned extension

More information

STATISTIQUE DE PERFORMANCE

STATISTIQUE DE PERFORMANCE INTERCOMPARISON OF STANDARD CALIBRATION, METHODS AND MEASUREMENTS FOR PRESSURE, TEMPERATURE AND HUMIDITY BETWEEN RIC CASABLANCA, MOROCCO AND RIC Ljubljana SLOVENIA 3(10) azizmounir@gmail.com The operators

More information

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2 Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST Ronaldo Minniti 1 and Ladislav Czap 2 1 National Institute of Standards and Technology (NIST), Gaithersburg,

More information

H1: Metrological support. Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič

H1: Metrological support. Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič H1: Metrological support Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič Basic requirement How to ensure appropriate measurements and monitor changes in time and space? Analytical

More information

Euramet project 1075 Bilateral comparison

Euramet project 1075 Bilateral comparison Bilateral comparison Between the gas flow standards of the national Metrology Institutes of Bulgaria (BIM-NCM) and The Netherlands (VSL) Copyright VSL B.V. Dutch Metrology Institute Page 2 of 19 VSL Thijsseweg

More information

BUREAU INTERNATIONAL DES POIDS ET MESURES

BUREAU INTERNATIONAL DES POIDS ET MESURES BUREAU INTERNATIONAL DES POIDS ET MESURES Bilateral comparison of 1 Ω and 10 kω standards (ongoing BIPM key comparisons BIPM.EM-K13.a and 13.b) between the NIMT (Thailand) and the BIPM December 016 Final

More information

Final report to the CCT on key comparison CCT-K6 Comparison of local realisations of dew-point temperature scales in the range -50 C to +20 C

Final report to the CCT on key comparison CCT-K6 Comparison of local realisations of dew-point temperature scales in the range -50 C to +20 C NPL REPORT ENG 57 Final report to the CCT on key comparison CCT-K6 Comparison of local realisations of dew-point temperature scales in the range -50 to +20 S Bell, M Stevens, H Abe, R Benyon, R Bosma,

More information

ISO 376 Calibration Uncertainty C. Ferrero

ISO 376 Calibration Uncertainty C. Ferrero ISO 376 Calibration Uncertainty C. Ferrero For instruments classified for interpolation, the calibration uncertainty is the uncertainty associated with using the interpolation equation to calculate a single

More information

Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities

Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities T. J. Wiandt 1,2 1 Fluke Corporation, Hart Scientific Division, American Fork, Utah United States. 2 E-mail: tom.wiandt@hartscientific.com

More information

Supplementary comparison SIM.M.FF-S12. Final Report for Volume of Liquids at 20 L

Supplementary comparison SIM.M.FF-S12. Final Report for Volume of Liquids at 20 L Supplementary comparison SIM.M.FF-S12 Final Report for Volume of Liquids at 20 L Pilot: M. Maldonado (CENAM, Mexico) mmaldona@cenam.mx km 4.5 Carretera a los Cués, El Marqués, Qro 76241, Mexico Participants:

More information

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands Calibration, traceability, uncertainty and comparing measurement results Workshop 16 March 2015 VSL, Delft The Netherlands 1 Calibration Calibrating an instrument involves comparing the measurements of

More information

Chilled Mirror Hygrometry

Chilled Mirror Hygrometry Chilled Mirror Hygrometry Precision instruments for critical measurement and control applications Chilled Mirror Hygrometry Wherever precise, fundamental measurements are required in your laboratory or

More information

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web:  sales: Morehouse 1 Morehouse Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA 17403 PH: 717-843-0081 web: www.mhforce.com sales: edlane@mhforce.com 2 This presentation will cover the calibration

More information

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1 Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1 Updated at 12-December-2012 Task and Purpose of the Comparison According to the rules set up by the CIPM

More information

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR M. Araya 1, D. Almeida 2 1 National Laboratory of Temperature of Chile, LCPNT, Santiago, Chile 2 Ecuadorian Standardization

More information

Measurement Uncertainty of Dew-Point Temperature in a Two-Pressure Humidity Generator

Measurement Uncertainty of Dew-Point Temperature in a Two-Pressure Humidity Generator Int J Thermophys (2012) 33:1568 1582 DOI 10.1007/s10765-011-1005-z Measurement Uncertainty of Dew-Point Temperature in a Two-Pressure Humidity Generator L. Lages Martins A. Silva Ribeiro J. Alves e Sousa

More information

Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu

Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu Bilateral Comparison of between NMIA (Australia) and SPRING (Singapore) (KCDB reference No. CCPR-K1.a.1) Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu Contents 1. Introduction..2 2.

More information

Humidity Calibration Solutions

Humidity Calibration Solutions Humidity Calibration Solutions www.michell.com Humidity Calibration Solutions The Importance of Regular Calibration The reliable operation of a hygrometer and indeed any measuring instrument, can only

More information

Correlation Effects in the Uncertainty Estimation of Two-Pressure Humidity Generators

Correlation Effects in the Uncertainty Estimation of Two-Pressure Humidity Generators Journal of Physics: Conference Series PAPER OPEN ACCESS Correlation Effects in the Uncertainty Estimation of Two-Pressure Humidity Generators To cite this article: S A Carvajal and C A Sánchez 08 J. Phys.:

More information

Available online at ScienceDirect. Andrej Godina*, Bojan Acko

Available online at  ScienceDirect. Andrej Godina*, Bojan Acko Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 69 ( 014 ) 191 198 4th DAAAM International Symposium on Intelligent Manufacturing and Automation, 013 Measurement Uncertainty

More information

METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES AFTER REINSTALLATION

METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES AFTER REINSTALLATION IMEKO 22 nd TC3, 12 th TC5 and 3 rd TC22 International Conferences 3 to 5 February, 2014, Cape Town, Republic of South Africa METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES

More information

e-newsletter n 6 August 31 st, 2018 New calibration facility developed at PTB for relative humidity, associated to dtdlas transfer standard

e-newsletter n 6 August 31 st, 2018 New calibration facility developed at PTB for relative humidity, associated to dtdlas transfer standard e-newsletter n 6 August 31 st, 2018 New calibration facility developed at PTB for relative humidity, associated to dtdlas transfer standard PTB has developed a test facility to characterize humidity sensors

More information

2 Calibration and test facilities

2 Calibration and test facilities Towards improved humidity measurements at high temperatures and transient conditions Martti Heinonen 1, Fausto Arpino 2, Rien Bosma 3, Laura Cavallarin 4, Gino Cortellessa 2, Marco Dell Isola 2, Volker

More information

RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS

RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS The Fifth International Proficiency Testing Conference Timisoara, Romania () th 8 th September, 0 RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS Adriana Vâlcu, Sterică Baicu National Institute of Metrology,

More information

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS Jan P.M. de Vreede Department of Electricity, Radiation and Length NMi Van Swinden Laboratorium Thijsseweg 11, 2629 JA Delft, the

More information

BUREAU INTERNATIONAL DES POIDS ET MESURES

BUREAU INTERNATIONAL DES POIDS ET MESURES BUREAU INTERNATIONAL DES POIDS ET MESURES Bilateral comparison of 1 Ω and 10 kω standards (ongoing BIPM key comparisons BIPM.EM-K13.a and 13.b) between the SMD (Belgium) and the BIPM October 2017 Final

More information

Interamerican Metrology System (SIM) Regional Metrology Organization (RMO) Capacitance Comparison, Final Report

Interamerican Metrology System (SIM) Regional Metrology Organization (RMO) Capacitance Comparison, Final Report Interamerican Metrology System (SIM) Regional Metrology Organization (RMO) Capacitance Comparison, Final Report SIM.EM-K4., 0 pf fused-silica standard capacitor at 000 Hz SIM.EM-S4., 00 pf fused-silica

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer High Performance Chilled Mirror Hygrometer With Cable Mounted Measuring Heads Highly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct

More information

Final Report on Comparison of Hydraulic Pressure Balance Effective Area Determination in the Range up to 80 MPa

Final Report on Comparison of Hydraulic Pressure Balance Effective Area Determination in the Range up to 80 MPa EURAMET - THE EUROPEAN ASSOCIATION OF NATIONAL METROLOGY INSTITUTES Comparison Identifier: EURAMET 1414 (Registered in KCDB as EURAMET.M.P-S18) Final Report on Comparison of Hydraulic Pressure Balance

More information

Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz

Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range to 10 k June 016 Author: Co-author: Danuta Dobrowolska (GUM, pilot) Alexander Kosterov

More information

A calibration facility for automatic weather stations

A calibration facility for automatic weather stations METEOROLOGICAL APPLICATIONS Meteorol. Appl. 22: 842 846 (2015) Published online 16 November 2015 in Wiley Online Library (wileyonlinelibrary.com) DOI: 10.1002/met.1514 A calibration facility for automatic

More information

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688 Supplementary Comparison EURAMET.EMS19 EURAMET Project No. 688 Bilateral Comparison of Measurements of Current Transformers (CTs) Between and Final Report Hüseyin Çaycı (, Pilot Laboratory) January 211

More information

TEMPERATURE AND HUMIDITY DEPENDENCE ON STABILITY OF TORQUE MEASURING DEVICES Tassanai Sanponpute and Nittaya Arksonnarong

TEMPERATURE AND HUMIDITY DEPENDENCE ON STABILITY OF TORQUE MEASURING DEVICES Tassanai Sanponpute and Nittaya Arksonnarong IMEKO 22 nd TC3, 15 th TC5 and 3 rd TC22 International Conferences 3 to 5 February, 2014, Cape Town, Republic of South Africa TEMPERATURE AND HUMIDITY DEPENDENCE ON STABILITY OF TORQUE MEASURING DEVICES

More information

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs IMEKO 20 th TC3, 3 rd TC16 and 1 st TC22 International Conference Cultivating metrological knowledge 27 th to 30 th November, 2007. Merida, Mexico. A comparison of primary platinum-iridium kilogram mass

More information

Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar

Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) Dr. Steven D. Phillips Precision Engineering Division National Institute

More information

Towards SI traceable humidity measurements with radiosondes

Towards SI traceable humidity measurements with radiosondes VTT TECHNICAL RESEARCH CENTRE OF FINLAND LTD Centre for Metrology MIKES MIKES METROLOGY Towards SI traceable humidity measurements with radiosondes GRUAN ICM-9, MeteoMet Training Course FMI, Helsinki,

More information

Humidity Calibration Solutions

Humidity Calibration Solutions Instruments Humidity Solutions www.michell.com Instruments Humidity Solutions The Importance of Regular The reliable operation of a hygrometer and indeed any measuring instrument, can only be verified

More information

From MeteoMet to MeteoMet2. and more. Andrea Merlone

From MeteoMet to MeteoMet2. and more. Andrea Merlone From MeteoMet to MeteoMet2 and more Presented by Tom Gardiner WP1 WP2 Upper air measurements: sensors and techniques Novel methods, instruments and measurements WP3 Traceable measurement and protocols

More information

Development of a Heat-Pipe-Based Hot Plate for Surface-Temperature Measurements

Development of a Heat-Pipe-Based Hot Plate for Surface-Temperature Measurements Int J Thermophys (2009) 30:257 264 DOI 10.1007/s10765-008-0495-9 Development of a Heat-Pipe-Based Hot Plate for Surface-Temperature Measurements L. Rosso N. Koneva V. Fernicola Published online: 12 August

More information

WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty

WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty October 21, 2013 Summary The Working Group for Fluid Flow (WGFF) defines Calibration and Measurement Capability (CMC) uncertainty

More information

792A AC/DC Transfer Standard

792A AC/DC Transfer Standard 92A AC/DC Transfer Standard Technical Data Support for your most demanding ac measurement requirements ppm total uncertainty Traceable to national standards range: 2 mv to 00 V Frequency range: Hz to 1

More information

792A AC/DC Transfer Standard

792A AC/DC Transfer Standard 92A AC/DC Transfer Standard Technical Data Support for your most demanding ac measurement requirements ppm total uncertainty Traceable to national standards range: 2 mv to 00 V Frequency range: Hz to 1

More information

CCQM-K27.2 Second Subsequent Study: Determination of Ethanol in Aqueous Media Final Report

CCQM-K27.2 Second Subsequent Study: Determination of Ethanol in Aqueous Media Final Report CCQM-K27.2 Second Subsequent Study: Determination of Ethanol in Aqueous Media Final Report Michele M. Schantz, Reenie M. Parris, and Willie E. May National Institute of Standards and Technology (NIST)

More information

by A.Tonina*, R.Iuzzolino*, M.Bierzychudek* and M.Real* S. Solve + R. Chayramy + and M. Stock +

by A.Tonina*, R.Iuzzolino*, M.Bierzychudek* and M.Real* S. Solve + R. Chayramy + and M. Stock + Bilateral Comparison of 1.018 V and 10 V Standards between the INTI (Argentina) and the BIPM, August to October 2009 (part of the ongoing BIPM key comparison BIPM.EM-K11.a and b) by A.Tonina*, R.Iuzzolino*,

More information

Final Report. CCEM Comparison of 10 pf Capacitance Standards. Anne-Marie Jeffery Electricity Division NIST May 2000 Revised March 2002

Final Report. CCEM Comparison of 10 pf Capacitance Standards. Anne-Marie Jeffery Electricity Division NIST May 2000 Revised March 2002 Final Report CCEM Comparison of 10 pf Capacitance Standards Anne-Marie Jeffery Electricity Division NIST May 2000 Revised March 2002 Abstract Comparison of electrical standards must be carried out periodically

More information

Commissioning of the Beta Secondary Standard (BSS2)

Commissioning of the Beta Secondary Standard (BSS2) Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,

More information

473-SHX Dew Point Mirror

473-SHX Dew Point Mirror Humidity and Temperature Reference Hygrometer For Temperature up to 125 C Precise and stable chilled mirror dew point mirror technology High temperature optical components High temperature sample fan Cable

More information

MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT

MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT Michael Borys, Michael Gläser, Michael Mecke Physikalisch-Technische Bundesanstalt (PTB), Germany ABSTRACT Spheres made of a silicon

More information

A SPRT Intercomparison at Hg, TPW, Ga, Sn and Zn ITS-90 Fixed Points between PTB and LATU with PTB as Pilot Laboratory

A SPRT Intercomparison at Hg, TPW, Ga, Sn and Zn ITS-90 Fixed Points between PTB and LATU with PTB as Pilot Laboratory A SPRT Intercomparison at Hg, TPW, Ga, Sn and Zn ITS-90 Points between PTB and LATU with PTB as Pilot Laboratory O. Robatto, Departamento de Metrología Científica, Laboratorio Tecnológico del Uruguay(LATU),

More information

Document No: TR 12 Issue No: 1

Document No: TR 12 Issue No: 1 ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Prepared by: SADCAS Technical Manage Approved

More information

Calibration Traceability Guidelines

Calibration Traceability Guidelines Calibration Traceability Guidelines Check the website of accrediting bodies for the accredited laboratories individual scope of accreditation www.a2la.org ts.nist.gov/standards/accreditation/index.cfm

More information

Uncertainty of Calibration. Results in Force Measurements

Uncertainty of Calibration. Results in Force Measurements Publication Reference EAL-G Uncertainty of Calibration Results in Force Measurements PURPOSE This document has been produced to improve the harmonization in determination of uncertainties in force measurements.

More information

Interlaboratory Comparison in the field of Temperature, Humidity and Pressure, in the WMO Regional Association VI (MM-ILC-2015-THP)

Interlaboratory Comparison in the field of Temperature, Humidity and Pressure, in the WMO Regional Association VI (MM-ILC-2015-THP) Instruments and Observing Methods Report No. 128 WEATHER CLIMATE WATER Interlaboratory Comparison in the field of Temperature, Humidity and Pressure, in the WMO Regional Association VI (MM-ILC-2015-THP)

More information

Book of Abstracts Volume B

Book of Abstracts Volume B Joint International Symposium on Temperature, Humidity, Moisture and Thermal Measurements in Industry and Science Book of Abstracts Volume B TEMPMEKO & ISHM 2010 31 May - 4 June 2010, Portorož, Slovenia

More information

Abstract. 1. Introduction

Abstract. 1. Introduction The importance of establishing a very-high temperature radiation thermometry measurement capability at the National Metrology Institute of South Africa (NMISA) Speaker: Dr Efrem K Ejigu National Metrology

More information

Nanometrology and its role in the development of nanotechnology

Nanometrology and its role in the development of nanotechnology Nanometrology and its role in the development of nanotechnology Rob Bergmans Nederlands Meetinstituut Van Swinden Laboratorium 1 NMi Van Swinden Laboratorium The Art of Measurement Dutch national metrology

More information

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT Scott Crone North American Sales Manager AMETEK Calibration Instruments 8600 Somerset Drive Largo, Florida 33773 USA INTRODUCTION Like any other piece of equipment,

More information

Low gauge pressure comparisons between MIKES, Metrosert and FORCE Technology

Low gauge pressure comparisons between MIKES, Metrosert and FORCE Technology J2/2007 Low gauge pressure comparisons between MIKES, Metrosert and FORCE Technology Range -2000 to +2000, gauge Report on EUROMET Project 921 Markku Rantanen, Sari Semenoja, Guliko Peterson and Jesper

More information

CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon

CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon Tohru Nakano and Osamu Tamura National Metrology Institute of Japan (NMIJ), AIST Tsukuba, Japan

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer gh Performance Chilled Hygrometer With Cable Mounted Measuring Heads ghly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct insertion

More information

SIM FORCE STANDARDS COMPARISON UP TO 10 kn

SIM FORCE STANDARDS COMPARISON UP TO 10 kn SIM FORCE STANDARDS COMPARISON UP TO 10 kn Jorge C. Torres-Guzmán, Daniel A. Ramírez-Ahedo Centro Nacional de Metrología, CENAM.. ABSTRACT A comparison in the quantity of force was carried out among the

More information

SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Force Proving Instruments

SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Force Proving Instruments NABL 122-06 NATIONAL ACCREDITATION BOARD FOR TESTING AND CALIBRATION LABORATORIES SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Force Proving Instruments Reviewed by MASTER

More information

Uncertainty Propagation for Force Calibration Systems

Uncertainty Propagation for Force Calibration Systems Uncertainty Propagation for Force Calibration Systems Morehouse Instruments Figure 1. Measurement Uncertainty Pyramid Introduction There are several labs operating throughout the world, whom do not follow

More information

MASS AND VOLUME COMPARISONS AT MIKES

MASS AND VOLUME COMPARISONS AT MIKES MITTATEKNIIKAN KESKUS CENTRE FOR METROLOGY AND ACCREDITATION Julkaisu J4/2000 MASS AND VOLUME COMPARISONS AT MIKES Additional results to the EA intercomparison of weights 1 mg 100 g (Ma1) and to the EUROMET

More information

STATUS OF THE WIGOS DEMONSTRATION PROJECTS

STATUS OF THE WIGOS DEMONSTRATION PROJECTS STATUS OF THE WIGOS DEMONSTRATION PROJECTS Demonstration Project Morocco Strengthening Moroccan RIC Capacities (Submitted by Rabia Merrouchi, National Meteorological Service of Morocco (DMN)) February

More information

Calibration of Temperature Block Calibrators

Calibration of Temperature Block Calibrators European Association of National Metrology Institutes Calibration of Temperature Block Calibrators EURAMET cg-13 Version 2.0 (03/2011) Previously EA-10/13 Calibration Guide EURAMET cg-13 Version 2.0 (03/2011)

More information

UNCERTAINTY SCOPE OF THE FORCE CALIBRATION MACHINES. A. Sawla Physikalisch-Technische Bundesanstalt Bundesallee 100, D Braunschweig, Germany

UNCERTAINTY SCOPE OF THE FORCE CALIBRATION MACHINES. A. Sawla Physikalisch-Technische Bundesanstalt Bundesallee 100, D Braunschweig, Germany Measurement - Supports Science - Improves Technology - Protects Environment... and Provides Employment - No and in the Future Vienna, AUSTRIA, 000, September 5-8 UNCERTAINTY SCOPE OF THE FORCE CALIBRATION

More information

ASIA PACIFIC METROLOGY PROGRAME

ASIA PACIFIC METROLOGY PROGRAME APMP.M.M-K6 (SUB) multiples mass key comparison ASIA PACIFIC METROLOGY PROGRAME Final Report of APMP Comparison of Mass standards APMP.M.M-K6 (SUB) multiples mass key comparison Coordinated by: Vietnam

More information

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM 83.8058 K TO 1234.93 K: METHODS AND UNCERTAINTIES Gregory F. Strouse National Institute of Standards and Technology, Gaithersburg, Maryland, USA ABSTRACT

More information

Evaluation of measurement uncertainty in testing laboratories

Evaluation of measurement uncertainty in testing laboratories Elektrotehniški vestnik 68(5): 286 293, 2001 Electrotechnical Review, Ljubljana, Slovenija Evaluation of measurement uncertainty in testing laboratories Gaber Begeš, Janko Drnovšek, Igor Pušnik, Jovan

More information

ANNEXE 8. Technical protocols for the interlaboratory comparisons

ANNEXE 8. Technical protocols for the interlaboratory comparisons ANNEXE 8 Technical protocols for the interlaboratory comparisons TECHNICAL PROTOCOL FOR BILATERAL COMPARISON (Magnitude of the complex charge sensitivity of accelerometers) EURAMET.AUV.V-S1 1. INTRODUCTION

More information

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET CERTIFIED

More information

Final Report On COOMET Vickers PTB/VNIIFTRI Key Comparison (COOMET.M.H- K1.b and COOMET.M.H- K1.c)

Final Report On COOMET Vickers PTB/VNIIFTRI Key Comparison (COOMET.M.H- K1.b and COOMET.M.H- K1.c) Version 7 of 01-19-07 Final Report On COOMET Vickers PTB/VNIIFTRI Key Comparison (COOMET.M.H- K1.b and COOMET.M.H- K1.c) E. Aslanyan¹, K. Herrmann² ¹ All Russian Research Institute for Physical -Technical

More information

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE Publication Reference EA-10/13 EA Guidelines on the Calibration of Temperature Block Calibrators PURPOSE This document been produced by EA to improve the harmonisation in the calibration of temperature

More information

FORCE STANDARDS COMPARISON BETWEEN PTB (GERMANY) AND CENAM (MEXICO).

FORCE STANDARDS COMPARISON BETWEEN PTB (GERMANY) AND CENAM (MEXICO). FORCE STANDARDS COMPARISON BETWEEN PTB (GERMANY) AND CENAM (MEXICO). Jorge C. Torres-Guzmán*, Amritlal Sawla**, Daniel Ramírez-Ahedo* * Centro Nacional de Metrología, México. **Physikalisch-Technische

More information

CERTIFICATE OF ACCREDITATION

CERTIFICATE OF ACCREDITATION CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Hards Laboratories cc, t/a Technology Solutions

More information

CCM short note on the dissemination process after the proposed redefinition of the kilogram

CCM short note on the dissemination process after the proposed redefinition of the kilogram CCM short note on the dissemination process after the proposed redefinition of the kilogram Consultative Committee for Mass and Related Quantities 1. Introduction This note proposes how the mise en pratique

More information