CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014
|
|
- Blaze Roberts
- 5 years ago
- Views:
Transcription
1 Calibration General Principles & Theory, Equipment Considerations
2 Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link to Atomic Clock) True Value: The best estimate of the true value. To the Metrologist the true value is never known Accuracy: The deviation from the estimated true value. Ex: the true value was 0.01 C and the instrument read 0.05 C (error of 0.04 C) Precision: How close measurements can be made Resolution: The smallest increment of the measurement Repeatability: The ability to reproduce the same reading Uncertainty: A statistical probability of a measurement being within a specified tolerance: Ex. ±0.5% with 95% confidence Calibration: The process of recording an instrument s reading against a reference standard Correction: The process of adjusting an instrument s reading based on the measurements of a reference standard.
3 Statement of Accuracy True value = The best estimate of the actual value Confidence Level: The probability that the true value lies in the given range Uncertainty is is defined as the tolerance band in which (at k=2) a measured value will be in agreement with the true value for 95% of the measurements made.
4 Measurement Uncertainty CALIBRATION Normal Distribution True Value 1s = 68.27% 2s = 95.45% 3s = 99.73% 1s 2s 3s
5 Calibration Traceability Hierarchy
6 UKAS Calibration Hierarchy
7 Factors Limiting Performance of a Calibration Methodology Physical facilities of the laboratory used temperature control air quality stability of electrical power supply Calibration equipment Accuracy Repeatability Traceability Quality of staff training
8 Calibration schedule
9 Calibration All Michell hygrometers go through a calibration regime in a purpose built laboratories in the UK, Holland and Japan. DP sensors- Dry down for 5 days Calibration run using high accuracy dew point generators and chilled mirror references Traceable to NPL, Nmi, or NIST UKAS accredited
10 Calibration Systems: general overview Our systems can be configured to calibrate various types of hygrometer - Capacitive dewpoint sensors Cooled mirror dewpoint meters Relative humidity probes Electrolytic hygrometers
11 Considerations What devices do you want to calibrate What parameters are the devices measuring? What are the %RH & T, or DP measurement ranges of the devices? What are the stated accuracies and resolutions of the devices? Does the calibration need to be performed over the entire operating range of the device? Do you want to calibrate at set points, or do they want to be able to make fine adjustments? Do you want a system which has manual, semi, or fully automated adjustment of set points? Do you want a system which has manual, semi, or fully automated logging of devices under test and the reference? Consider the dryer and generator Consider the reference instrument Are these devices sensors or instruments? What size and shape are they? Do the devices require a flow past/through, or are they making an environmental measurement? Consider an appropriate calibration manifold or chamber
12 Calibration Systems: general overview Traceability to National Standards Air Source Dryer Reference Dewpointmeter Generator Test Chamber Unit under test
13 Pressure Swing Dryers Provides dried air to moisture & dew point generators in absence of suitable site air Two column timed re-generation Oil-free compressed air feed Self-contained unit Two ranges available C dewpoint -100 C dewpoint
14 Moisture Generation to -75 C Enclosure is T controlled at 35 C Two-stage mixing of full dry and saturated gas flows allows high precision adjustment of generated dew-points Generation of dew-points from - 75 Cdp up to ambient T Designed to be fed by a pressure swing dryer or dry instrument air
15 Moisture generation to -100 C Temperature controlled enclosure Three-stage mass flow controlled flow mixing Generation of dew-points from -90 Cdp up to ambient T ±0.25 Cdp output stability
16 Chilled mirrors come in different types with different range-one size does NOT fit all.
17 Cooled Mirror Advantages fundamental measurement principle high accuracy drift free operation traceability to national standards
18 Basic Cooled mirror technology The cooled mirror sensor principle : cooling of a metal mirror by a thermoelectric cooler. At dew point, the mirror temperature is measured by a precision thermometer. Fundamental, high accuracy and long-term stability. Accuracy is limited by the PRT Range is limited by cooling methods and number of peltier used
19 Dew-Point Depression of Cooled Mirror Sensors 1 Stage Minimum Measureable Dew point ( C) Sensor Temperature ( C) 2 Stage Cooled Mirror Hygrometers that use Peltier cooling have a finite ability to cool Multiple Peltier stages may be stacked to achieve low temperatures Additional cooling methods such as refrigeration, fluid jackets may be used to augment the thermoelectric cooling The amount of depression is not equal to the minimum dew point reading attainable. A cooled mirror typically has to cool 5-10 C below the dew/frost point to collect a dew/frost layer
20 Frost Formation on a cooled Mirror and ensuring frost is formed is key to accuracy Clean Mirror Mirror with even formation of frost Super-cooled water can form between 0 and -40 C leading to a 1 C error for every 10 C depression
21 Condensation Hygrometer Stabilization CALIBRATION A cooled mirror modulates the power applied to a thermoelectric cooling module based on optical feedback. Stabilization typically requires oscillations around the actual dew point. PID control algorithms are employed to provide this function
22 Cooled Mirror Problems Instruments have always perceived to be :- Expensive instrumentation Maintenance intensive Perception of being too good for application But there are lower cost solutions if restriction on range is accepted Today these are low maintenance, stable instruments
23 Contamination Correction CALIBRATION Any optical system can be affected by contamination of the light path. Cooled mirror dew-point hygrometers are no exception. Contamination of the mirror surface or the detection window will reduce the detection of the light reflected by the mirror. During normal operation mirror contamination will have no measurable effect on the performance of the instrument. However, gross contamination of the sensor can result in inaccurate measurements. The mirror condition is measured and a percentage value should be displayed on the instruments For this reason, all Michell cooled mirror instruments incorporate an automatic compensation system.
24 Contamination Correction (2) CALIBRATION This works by heating the mirror to a temperature above that of the dewpoint temperature, which burns off any condensation and contaminants. The mirror surface is now clean and the sensor optics are rebalanced to compensate for any reduction in light intensity caused by the contamination. With heavy contamination the automatic system may not be able to compensate and so manual cleaning or the mirror surface may be periodically required. This should be performed with a Q tip bud and distilled water. The sensor window and the mirror should be cleaned.
25 CALIBRATION
26 NPL Calibration Certificate
27 Dewpoint Calibration Systems Various designs from -100 C
28 Rh Calibration Still requires the basics - Air source - Controlled moisture generation Can use chambers - Bigger the chamber the more potential errors Reference can be CM or a suitable probe - What traceability - Accuracy Trained staff and methodology
29 Dew point v %Rh
30 Basic Rh generator using mixed flow can work with or without a chilled mirror reference
31 Full chamber system. This complete, automatic calibration system is computer controlled to provide a calibration environment for various types of humidity sensor from -20 to +50 C and 1 to 95% relative humidity (-50 to +50 C) Systems can provide profiling of both parameters as well as a comprehensive data logging capability.
32 Full chamber schematic CALIBRATION
33 Simple Rh calibration at ambient conditions
34 Principle S503
35 Inside the S503
36 Rh calibration S904: Allows Rh and temperature control
37 Schematic
38 Inside the chamber: Mixing fan and control sensor
39 Outside of the chamber; 4 peltier ensure temperature uniformity
40 Chamber with chilled mirror reference
41 Questions?
MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations
Calibration General Principles & Theory, Equipment Considerations Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link
More informationHumidity Calibration Solutions
Instruments Humidity Solutions www.michell.com Instruments Humidity Solutions The Importance of Regular The reliable operation of a hygrometer and indeed any measuring instrument, can only be verified
More informationHumidity Calibration Solutions
Humidity Calibration Solutions www.michell.com Humidity Calibration Solutions The Importance of Regular Calibration The reliable operation of a hygrometer and indeed any measuring instrument, can only
More informationThe Fundamentals of Moisture Calibration
The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity
More informationChilled Mirror Hygrometry
Chilled Mirror Hygrometry Precision instruments for critical measurement and control applications Chilled Mirror Hygrometry Wherever precise, fundamental measurements are required in your laboratory or
More information473 Dew Point Hygrometer
High Performance Chilled Mirror Hygrometer With Cable Mounted Measuring Heads Highly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct
More informationREALIZATION OF HUMIDITY STANDARD FACILITY USING TWO-PRESSURE HUMIDITY GENERATOR AND HUMIDITY CHAMBER
METROLOGY AND MEASUREMENT SYSTEMS Index 3393, ISSN 86-8229 www.metrology.pg.gda.pl REALIZATION OF HUMIDITY STANDARD FACILITY USING TWO-PRESSURE HUMIDITY GENERATOR AND HUMIDITY CHAMBER Essam El-Din, Mostafa
More informationUnderstanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator
Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Introduction Humidity generators are mechanical devices that provide stable temperature and humidity
More information473 Dew Point Hygrometer
gh Performance Chilled Hygrometer With Cable Mounted Measuring Heads ghly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct insertion
More information473-SHX Dew Point Mirror
Humidity and Temperature Reference Hygrometer For Temperature up to 125 C Precise and stable chilled mirror dew point mirror technology High temperature optical components High temperature sample fan Cable
More informationResponse Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length
Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length Yun-Kyung Bae and In-Jik Jeong Industrial & Physical Instrument Center, Korea
More informationUnderstanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator
Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Technical Note The Fluke Calibration 5128A RHapid-Cal Humidity Generator provides a portable, stable test environment
More informationScience of Stability 2017
Science of Stability 2017 Humidity Measurement in Stability Testing Applications Defining Uncertainty and the Challenges at Higher Temperatures Humidity Measurement In Stability Testing Applications Some
More informationFundamentals of Moisture Measurement in Industrial Applications
2016 Fundamentals of Moisture Measurement in Industrial Applications Moisture is not only a process variable to be measured in industrial applications, but also a common contaminant which must be controlled.
More informationOverview Brochure Comprehensive range of instrumentation for the measurement of moisture and oxygen
Overview Brochure 2016 Comprehensive range of instrumentation for the measurement of moisture and oxygen Overview Brochure Relative Humidity Probes and Transmitters Repeatable RH and temperature measurement
More informationThe calibration of Hygrometer (LectureandTraining )
The calibration of Hygrometer (LectureandTraining ) HAGIYA Satoshi 8Nov.2016 Outline 1.Measu rementmethods of midity(theory) Hu 2.Traceab ili ty in JMA 3.Cali brationmethodsof Hyg rometer 4.Practice (Ca
More informationEdgetech Instruments Inc.
2015_07_ETI OVERVIEW Rev 3 MADE IN THE USA Edgetech Instruments (ETI) is a USA company based in Massachusetts. We have over 50 years of expertise in the design and manufacture of a comprehensive range
More informationD501 and D501-RS. Microscanner D-Series IR Thermometers. no effect. no effect. no effect. no effect no effect
Microscanner D-Series The Only Certified Accurate Surface Instruments in the World A MUST FOR ISO 9001 ISO 9002 ISO 9003 TRACEABILITY PROGRAMS Common Surface Measurement Errors D501 and D501-RS Microscanner
More informationUncertainty in Humidity Measurement. Helping you make a better measurement.
Uncertainty in Humidity Measurement Helping you make a better measurement. Webinar Presenters Bruce McDuffee Michael Boetzkes Agenda Terms & Definitions Understand the Uncertainty of Your Instrument Understanding
More informationCHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR
XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR Davor Zvizdic, Tomislav Stasic, Lovorka
More informationQuality assurance for sensors at the Deutscher Wetterdienst (DWD)
Quality assurance for sensors at the Deutscher Wetterdienst (DWD) Quality assurance / maintenance / calibration Holger Dörschel, Dr Tilman Holfelder WMO International Conference on Automatic Weather Stations
More informationDevelopment of the High-Temperature Dew-Point Generator Over the Past 15 Years
Int J Thermophys (2017) 38:161 DOI 10.1007/s10765-017-2291-x TEMPMEKO 2016 Development of the High-Temperature Dew-Point Generator Over the Past 15 Years R. Bosma 1 J. Nielsen 2 A. Peruzzi 1 Received:
More informationTemperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer
Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer Temperature Control without Compromise: rapid changes, precise temperatures, powerful complex functionality.
More informationTechniques for Fast Measurements of Low Dewpoint in Portable Hygrometers
Techniques for Fast Measurements of Low Dewpoint in Portable Hygrometers Speaker / Author: Nick Malby, Michell Instruments Ltd, 48 Lancaster Way Business Park, Ely, Cambridgeshire. CB6 3NW. UK. +44 1353
More informationSTATUS OF THE WIGOS DEMONSTRATION PROJECTS
STATUS OF THE WIGOS DEMONSTRATION PROJECTS Demonstration Project Morocco Strengthening Moroccan RIC Capacities (Submitted by Rabia Merrouchi, National Meteorological Service of Morocco (DMN)) February
More informationQuality assurance for sensors at the Deutscher Wetterdienst (DWD)
Paper submitted to ICAWS 2017: Topic 3 Sustainability of the measurements: Calibration, intercomparisons, laboratory and field performance tests, quality assurance and control assessment for traceable
More informationInfrared Thermometer Calibration A Complete Guide
Infrared Thermometer Calibration A Complete Guide Application note With proper setup and planning, infrared calibrations can be accurate. The steps outlined below should be followed to perform accurate
More informationEvaluation of Vaisala HMP45D Humidity Probes
Instrument Test Report: 661 Evaluation of Vaisala HMP45D Humidity Probes J.Gorman Regional Instrument Centre 2 nd Jan 2002 Authorisation Bruce Forgan SRLR Regional Instrument Centre Distribution STNM STIE
More informationEstablishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements
Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements Experimental Design 8.2.2013 Doc. Martti Heinonen etunimi.sukunimi@mikes.fi Outline 1.
More informationSSi Super Systems Inc.
Model DPC2530 Continuous Digital Dew Point Analyzer For measurement of: Endothermic Atmosphere Exothermic Atmosphere Nitrogen / Hydrogen Atmosphere Plant Air Systems SPECIFICATIONS Measurement Range: -50
More informationALMA Weather Instrumentation Specification
ALMA-90.03.00.00-00x-A-SPE 2006 12 01 Specification Document Jeff Mangum Page 2 Change Record Revision Date Author Section/ Remarks Page affected 1 2005-08-12 Jeff Mangum All Initial Draft 2 2006-08-30
More informationDewpoint Meter Data Sheet / Instructions
Paint Test Equipment Dewpoint Meter Data Sheet / Instructions Dewpoint Meter Information ISO 8502-4: Preparation of steel substrates before application of paints and related products. Tests for the assessment
More informationIntroduction to Blackbody Sources
Introduction to s This section contains dedicated blackbody sources for low uncertainty calibration of infrared thermometers. A range of portable primary blackbody sources combine high emissivity with
More informationDESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR
Copyright 2013 by ABCM DESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR Amaury Roteger de Oliveira Renato Nunes Teixeira Instituto Nacional de Metrologia, Qualidade e Tecnologia - Inmetro Diretoria
More informationComparison of HCD5000 DewPoint Duo to Condumax II
EXECUTIVE SUMMARY This document compares the technology and products of ZEGAZ Instruments (US) to that of Michell Instruments (UK) in the field of hydrocarbon and water dewpoint measurement. INTRODUCTION
More informationData Sheet / Instructions. Dewpoint MeterO
Data Sheet / Instructions Dewpoint MeterO Dewpoint Meter ISO 8502-4: Preparation of steel substrates before application of paints and related products. Tests for the assessment of surface cleanliness.
More informationMeasurement Uncertainty of Dew-Point Temperature in a Two-Pressure Humidity Generator
Int J Thermophys (2012) 33:1568 1582 DOI 10.1007/s10765-011-1005-z Measurement Uncertainty of Dew-Point Temperature in a Two-Pressure Humidity Generator L. Lages Martins A. Silva Ribeiro J. Alves e Sousa
More informationNew Peltier-based chilled-mirror hygrometer SKYDEW
GRUAN ICM-10, Potsdam, April 25, 2018 New Peltier-based chilled-mirror hygrometer SKYDEW Takuji Sugidachi, K. Shimizu (Meisei Electric Co. LTD., Japan) M. Fujiwara (Hokkaido University, Japan) Motivation
More informationCALIBRATION CERTIFICATE # 503
CALIBRATION CERTIFICATE # 503 Calibration date : 2011-02-23 Certificate issued : 2012-04-18 Company name Company address City, Province, Canada CLAS 2009-02 Calibration of Mass flow meter Micro Motion
More informationModule 2. Measurement Systems. Version 2 EE IIT, Kharagpur 1
Module 2 Measurement Systems Version 2 EE IIT, Kharagpur 1 Lesson 8 Measurement of Level, Humidity and ph Version 2 EE IIT, Kharagpur 2 Instructional Objectives At the end of this lesson, the student will
More information3URGXFWLQIRUPDWLRQ 5HVHUYDWLRQRISURSHUW\ULJKWV
3URGXFWLQIRUPDWLRQ +\URSKLO Š '7W\SH 5HVHUYDWLRQRISURSHUW\ULKWV 7KHFRQWHQWRIWKLVGRFXPHQWLVWKHLQWHOOHFWXDOSURSHUW\RI%$57(&DQGPXVWQHLWKHUEHFRSLHGQRUPDGHDYDLO DEOHWRWKLUGSDUWLHVQRWHYHQLQH[WUDFWVZLWKRXWRXUH[SUHVVDSSURYDOLQZULWLQ$Q\LQIULQHPHQWZLOOUHVXOWLQD
More informationSERIES 4000 PRECISION DEWPOINT HYGROMETER Integrale INSTALLATION, OPERATION AND MAINTENANCE MANUAL
SERIES 4000 PRECISION DEWPOINT HYGROMETER Integrale INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue July 2001 Page 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 4 1.1 General 4 1.2 Operating Principle
More informationTYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE
TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE Michael Bair and Pierre Delajoud 2002 DEC 10 Revised 2004 MAR 26 FORWARD FPG8601 is a pressure standard designed
More informationMeasuring Humidity in the Charters of Freedom Encasements Using a Moisture Condensation Method
Measuring Humidity in the Charters of Freedom Encasements Using a Moisture Condensation Method Speaker: Cecil G. Burkett NASA Langley Research Center Mail Stop 236 Hampton VA 23681 PH: 757-864-4720 FAX:
More informationElcometer 116C. Sling Hygrometer. Operating Instructions
Elcometer 116C English Sling Hygrometer Operating Instructions English is a registered trademark of Elcometer Limited. All other trademarks acknowledged. Copyright Elcometer Limited. 2008. All rights reserved.
More informationMany short reports are inadequate and need more work
University of Maryland Baltimore County - UMBC Phys650 - Special Topics in Experimental Atmospheric Physics (Spring 2009) J. V. Martins and M. H. Tabacniks Grading feedback: http://userpages.umbc.edu/~martins/phys650/
More informationPerformance of the Meteolabor Snow White chilled-mirror hygrometer. Masatomo Fujiwara Hokkaido University, Japan
Performance of the Meteolabor Snow White chilled-mirror hygrometer Masatomo Fujiwara Hokkaido University, Japan Coauthors and Collaborators F. Hasebe, M. Shiotani, H. Voemel, S. J. Oltmans, H. Takashima,
More informationMeasuring State Parameters of the Atmosphere
Measuring State Parameters of the Atmosphere Some Applications of Atmospheric Thermodynamics Earth Observing Laboratory, NCAR IDEAS-4 Tutorial Introduction Goals of This Presentation Present two complementary
More informationA Case for Periodic Calibration or Verification of RTDs
A Case for Periodic Calibration or Verification of RTDs If you are experiencing audio problems please call the teleconference number below Phone 650-479-3208 Access code 668 676 434 Your Host and Presenter
More informationProcess Analyzer and Sampling Systems
Process Analyzer and Sampling Systems Course Price 3050 Course Description Analytical instruments used for online chemical analysis of process streams or plant environments are generally called Process
More informationCONTENT 2. ORGANIZATION 3. SERVICES. instruments. (3)Activities of RIC Tsukuba
1 CONTENT 1. INTRODUCTION 2. ORGANIZATION 3. SERVICES (1)Quality assurance of meteorological instruments (2)Research and development (3)Activities of RIC Tsukuba 2 Meteorological Instruments Center (MIC)
More informationLAB007 Industrial Analytical Chemistry & Process Analyzer System
LAB007 Industrial Analytical Chemistry & Process Analyzer System H.H. Sheikh Sultan Tower (0) Floor Corniche Street Abu Dhabi U.A.E www.ictd.ae ictd@ictd.ae Course Introduction: Analytical instruments
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 CONTROL COMPANY 12554 Galveston Road Suite B230 Webster, TX 77598 Aaron Judice Phone: 281 482 1714 CALIBRATION Valid To: March 31, 2018
More informationLaird Thermal Systems Application Note. Precise Temperature Control for Medical Analytical, Diagnostic and Instrumentation Equipment
Laird Thermal Systems Application Note Precise Temperature Control for Medical Analytical, Diagnostic and Instrumentation Equipment Contents Overview...3 Thermal Management Techniques...3 Application Examples...3
More informationAL108 Process Analyzers & Sampling Systems
AL108 Process Analyzers & Sampling Systems AL108 Rev.002 CMCT COURSE OUTLINE Page 1 of 8 Training Description: Analytical instruments used for online chemical analysis of process streams or plant environments
More informationCREATION AND DEVELOPMENT OF THE NATIONAL HUMIDITY STANDARD OF GASES
CREATION AND DEVELOPMENT OF THE NATIONAL HUMIDITY STANDARD OF GASES Efimia LUCHIAN, Constantin BORDIANU, Grigore BUZUC, Victoria CEBAN National Institute of Metrology, Chisinau, Republic of Molova tel.:
More informationCalibration of temperature sensors within Length Standard Section of NMIJ
Calibration of temperature sensors within Length Standard Section of NMIJ Dr. Akiko Hirai Length Standard Section National Metrology Institute of Japan (NMIJ) 1 Contents Temperature measurement system
More informationEvaluating New Generation Vibrating Tube Sensor for Density Measurement under Process Conditions
Products Solutions Services Evaluating New Generation Vibrating Tube Sensor for Density Measurement under Process Conditions Dr. Alfred Rieder, Slide 1 Overview Introduction Process meter Features Working
More informationCOMPARISON OF HUMIDITY MEASUREMENTS USING A DEW POINT METER AS A TRANSFER STANDARD APMP-IC-1-97 REPORT
COMPARISON OF HUMIDITY MEASUREMENTS USING A DEW POINT METER AS A TRANSFER STANDARD APMP-IC-1-97 REPORT BY NATIONAL METROLOGY CENTRE (A DIVISION OF SPRING SINGAPORE) COORDINATOR OF THE COMPARISON September
More informationCOMPARISON OF INMETRO AND INTI HUMIDITY STANDARDS. Final Report
SIM.T-K6.4 INMETRO / INTI Final Report COMPARISON OF INMETRO AND INTI HUMIDITY STANDARDS Final Report Instituto Nacional de Metrologia, Qualidade e Tecnologia (INMETRO) Instituto Nacional de Tecnología
More informationCan a -20 C reference point be established in a laboratory?
Can a -20 C reference point be established in a laboratory? Speaker / Author: Yvette Volschenk Company: AFRIQLABS Centurion South Africa Cell: 083 658 8373 E-mail: yvette7137@iburst.co.za Abstract: When
More informationTRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS
TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS W. Beyer and W. Pahl Physikalisch-Technische Bundesanstalt (PTB) 38116 Braunschweig, Germany Abstract: In accordance with ISO 17025,
More informatione-newsletter n 6 August 31 st, 2018 New calibration facility developed at PTB for relative humidity, associated to dtdlas transfer standard
e-newsletter n 6 August 31 st, 2018 New calibration facility developed at PTB for relative humidity, associated to dtdlas transfer standard PTB has developed a test facility to characterize humidity sensors
More informationTypical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge
Typical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge 2090TN05 Revised 2004 March 26 Technical Note FPG8601 is a pressure standard designed to cover the range from
More informationA calibration facility for automatic weather stations
METEOROLOGICAL APPLICATIONS Meteorol. Appl. 22: 842 846 (2015) Published online 16 November 2015 in Wiley Online Library (wileyonlinelibrary.com) DOI: 10.1002/met.1514 A calibration facility for automatic
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ARCONIC POWER AND PROPULSION RESEARCH CENTER 1500 S. Warner St. Whitehall, MI 49461-1895 Michelle McDonald Phone: 231 894 7586 CALIBRATION
More informationAN EXPERIMENTAL STUDY OF THE FROST FORMATION ON A COLD SURFACE IN FREE CONVECTIVE FLOW
AN EXPERIMENTAL STUDY OF THE FROST FORMATION ON A COLD SURFACE IN FREE CONVECTIVE FLOW Giovanni Tanda, Marco Fossa DITEC, Università degli Studi di Genova via all Opera Pia 15a, I-16145 Genova, ITALY E-mail:
More informationAtmosphere Properties and Instruments. Outline. AT351 Lab 2 January 30th, 2008
Atmosphere Properties and Instruments AT351 Lab 2 January 30th, 2008 Outline 1. Atmospheric Variables and How We Measure Them 2. Composition of the Atmosphere 3. How to Represent Weather Data Visually
More informationInter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C
DOI 10.1007/s10765-014-1687-0 Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C Domen Hudoklin Regina Mnguni Hans Liedberg Igor Pušnik Jovan Bojkovski
More informationTowards SI traceable humidity measurements with radiosondes
VTT TECHNICAL RESEARCH CENTRE OF FINLAND LTD Centre for Metrology MIKES MIKES METROLOGY Towards SI traceable humidity measurements with radiosondes GRUAN ICM-9, MeteoMet Training Course FMI, Helsinki,
More informationDM70. Hand-held Dewpoint Meter. Truly Fast Response - Superior Long-term Stability
DM70 Hand-held Dewpoint Meter Truly Fast Response - Superior Long-term Stability DM70 - When Dry Air Is Important The Vaisala DM70 Hand-held Dewpoint Meter offers stable and fast measurement for industrial
More informationIsotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000
Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000 Welcome to the Isotech Journal of Thermometry Fundamentals of Thermometry Part 1 Temperature Scales Practical Calibration of Thermometers
More informationPsychrometer construction for performance testing in. temperature and humidity chambers, and the precision
Technology report Psychrometer construction for performance testing in temperature and humidity chambers, and the precision of humidity measurements Meeting the challenge of quality engineering Hirokazu
More informationCERTIFICATE OF CALIBRATION
Page 1 of 5 CERTIFICATE OF CALIBRATION 1K Thermistor CERTIFICATE NO.: MANUFACTURER: Measurement Specialties DATE OF CALIBRATION: 26-Apr-17 TC15249 CUSTOMER: AMBIENT TEMPERATURE: (21 ± 3) C JOB NUMBER:
More informationHeat stress index. Dewpoint. pressure. altitude
Kestrel 4200 Measures Volume air flow in cfm, L/s, m³/s, m³/m, m³/h Air velocity (current, average, and maximum) Heat stress index Windchill Relative humidity Dewpoint in F or C Humidity ratio Barometric
More informationCalibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands
Calibration, traceability, uncertainty and comparing measurement results Workshop 16 March 2015 VSL, Delft The Netherlands 1 Calibration Calibrating an instrument involves comparing the measurements of
More informationSIMPLE DEW DIGITAL DEW POINT ANALYZER
SIMPLE DEW DIGITAL DEW POINT ANALYZER OPERATIONS MANUAL Super Systems Inc. 7205 Edington Drive Cincinnati, OH 45249 513-772-0060 Fax: 513-772-9466 www.supersystems.com Super Systems Inc. USA Office Corporate
More informationOverviewofRIC Tsukuba (Japan,RAII )
OverviewofRIC Tsukuba (Japan,RAII ) 7Nov.2016 Hiroshi Yoshimura Scientific Officer Regional Instrument Centre Tsukuba Observing Division, Observing Department Outline 1. WMO Regional Instrument Center
More informationVaisala DRYCAP Hand-Held Dewpoint Meter DM70. Fast Response - Low Maintenance
Vaisala DRYCAP Hand-Held Dewpoint Meter DM70 Fast Response - Low Maintenance DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications The Vaisala DRYCAP Hand- Held Dewpoint Meter DM70 offers accurate
More informationMEASUREMENT OF WBGT INDEX IN AXIAL FLOW AS THE MAIN THERMODYNAMIC PARAMETERS VARY G.
MEASUREMENT OF WBGT INDEX IN AXIAL FLOW AS THE MAIN THERMODYNAMIC PARAMETERS VARY G. Buonanno, M. Dell'Isola, A. Frattolillo and L. Vanoli Di.M.S.A.T. - Università di Cassino via Di Biasio 43-03043 Cassino
More informationIsothermal Technology Ltd
INDUSRIAL MEASUREMENS WIH VERY SHOR IMMERSION J. P. avener, D. Southworth, D. Ayres, N. Davies. Isothermal echnology Ltd lab@isotech.co.uk Abstract One major problem that keeps recurring is the request
More informationDM70 Hand-Held Dewpoint Meter for Spot-Checking Applications
DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications The Vaisala DRYCAP Hand-Held Dewpoint Meter DM70 offers accurate and fast measurement for industrial dewpoint applications, such as compressed
More informationIntroduction of RIC Tsukuba (Japan, RAII)
Introduction of RIC Tsukuba (Japan, RAII) 20 March, 2018 Kouichi NAKASHIMA Scientific Officer Regional Instrument Centre Tsukuba Observing Division, Observing Department Outline 1. Overview of Organization
More informationLoadcell Calibration - Evaluation of Uncertainties
Loadcell Calibration - Evaluation of Uncertainties This document is a revision of the evaluation issued in June 2015. Novatech s calibration laboratory is not UKAS accredited due to the high number of
More informationMeasuring State Parameters of the Atmosphere
Measuring State Parameters of the Atmosphere Some Applications of Atmospheric Thermodynamics Earth Observing Laboratory, NCAR IDEAS-4 Tutorial Introduction Goals of This Presentation Present two complementary
More informationSHEBA GLASS SOUNDING DATA
SHEBA GLASS SOUNDING DATA Erik R. Miller and Kathryn Beierle - 12 June 2000 NCAR ATD Considerable speculation has been brought to bear as to whether there is a dry bias in the SHEBA radiosonde data. Two
More informationModule 11: Meteorology Topic 3 Content: Weather Instruments Notes
Introduction In order for meteorologists to accurately predict the weather, they take thousands of different weather measurements each day. Meteorologists need to use many tools in order to draw an accurate
More information2.1. Accuracy, n- how close the indication of the thermometer is to the true value.
AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria
More informationDeFelsko PosiTector Dew Point Meter with Anemometer (DPMA)
Management Procedure 2562 Revision: A Date Issued: March 24, 2017 Date Revised: Calibration Procedure DeFelsko PosiTector Dew Point Meter with Anemometer (DPMA) Table of Contents 1. Introduction and UUC
More informationAncillary Calibration Instruments Specifications and Requirements
SCID-90.05.13.00-001-A-SPE Version: A 2004 June 15 Prepared By: Name(s) and Signature(s) Organization Date 2004-06-15 Approved By: Name and Signature Organization Date Released By: Name and Signature Organization
More information: INSTRUMENTATION AND PROCESS CONTROL COURSE CODE : 6071 COURSE CATEGORY : A PERIODS/ WEEK : 5 PERIODS/ SEMESTER : 75 CREDIT : 5 TIME SCHEDULE
COURSE TITLE : INSTRUMENTATION AND PROCESS CONTROL COURSE CODE : 6071 COURSE CATEGORY : A PERIODS/ WEEK : 5 PERIODS/ SEMESTER : 75 CREDIT : 5 TIME SCHEDULE MODULE TOPIC PERIODS 1 Measuring Instruments
More informationWater Vapour Effects in Mass Measurement
Water Vapour Effects in Mass Measurement N.-E. Khélifa To cite this version: N.-E. Khélifa. Water Vapour Effects in Mass Measurement. Measurement. Water Vapour Effects in Mass Measurement, May 2007, Smolenice,
More informationAutomated multi-vapor gravimetric sorption analyzer for advanced research applications
Automated multi-vapor gravimetric sorption analyzer for advanced research applications Automated multi-vapor gravimetric sorption analyzer for advanced research applications Key benefits of the DVS Advantage
More informationThe Vaisala Reference Radiosonde Program: First Results and Future Plans
The Vaisala Reference Radiosonde Program: First Results and Future Plans Presentation in the NMM 2010, Helsinki, Finland, 9 June2010 Heikki Turtiainen, Hans von Bagh, Hannu Jauhiainen, Terhi Lehtola, Johanna
More informationMeasurement method for the proficiency testing program
APLAC T088 Appendix Measurement method for the proficiency testing program Introductions This measurement method is prepared for use by the APLAC Proficiency Testing Program Photometric measurement of
More informationGood practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors
Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors 1. Scope... 2 2. Introduction... 2 3. Selection of thermocouples
More informationMETHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone
METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT Scott Crone North American Sales Manager AMETEK Calibration Instruments 8600 Somerset Drive Largo, Florida 33773 USA INTRODUCTION Like any other piece of equipment,
More informationMetrology and Environment
Bridging the Gap Responding to Environmental Change: from Words to Deeds 14-16 May 2008, Portorož Metrology and Environment Prof Janko Drnovšek and Laboratory group Janko.Drnovsek@fe.uni-lj.si Basic dilemma
More informationTECORA Isostack G4 and DECS Stack Emissions Sampling. Methods for sampling emissions of Dioxins, Mercury and Heavy Metals
Methods for sampling emissions of Dioxins, Mercury and Heavy Metals Outline of the session: Sampling using the Standard Reference Method Continuous sampling Dioxins Furans What are dioxins and dioxin-like
More informationDM70 Hand-Held Dewpoint Meter for Spot-Checking Applications
DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications The Vaisala DRYCAP Hand-Held Dewpoint Meter DM70 offers accurate and fast measurement for industrial dewpoint applications, such as compressed
More information