Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Size: px
Start display at page:

Download "Quality assurance for sensors at the Deutscher Wetterdienst (DWD)"

Transcription

1 Paper submitted to ICAWS 2017: Topic 3 Sustainability of the measurements: Calibration, intercomparisons, laboratory and field performance tests, quality assurance and control assessment for traceable and nontraceable quantities (e.g. temperature and lightning). Quality assurance for sensors at the Deutscher Wetterdienst (DWD) Quality assurance on site / maintenance Holger Dörschel, Dr.-Ing. Tilman Holfelder 1 Introduction At the Deutscher Wetterdienst (DWD), the national meteorological service of Germany, the data for many meteorological parameters are collected automatically with data acquisition systems for decades, even if an observer has been present at the stations. In the nineties of the last century DWD renewed the technical infrastructure and the data acquisition systems at the international airports. In the following years the primary and secondary observing stations were harmonized and equipped with automated weather stations (AWS) and sensors in a project. During this period the maintenance department has been restructured with four service and logistic divisions in Hamburg, Offenbach, Potsdam and Munich and three minor service offices in Essen, Leipzig and Stuttgart. Each division and service office is responsible for regular preventive maintenance and troubleshooting of different types of manned and automated weather stations and several remote sensing systems. Figure 1: Service locations of DWD 1

2 Within the next years, by 2021, the DWD measuring network will be fully automated. As all Stations are already equipped with AWS only minor changes have to be done for automation, but quality assurance of the measurements will remain a very important task. One part of this quality assurance is an adequate maintenance concept including regular field checks and calibration of the sensors in the laboratory. To achieve the requested high availability of meteorological data from the measurement systems and to assure the requested high quality of the meteorological data several requirements need to be considered. Specific technical know-how for the meteorological measurement systems and the process of data acquisition has to be obtained. A short response time in case of technical failures, especially for the weather radar network and AWOS at international airports is needed, And of course the specific know-how for procurement and operation of the meteorological measurement systems has to be preserved. To fulfil these requirements DWD has its own technical staff for the maintenance of the meteorological observing systems, located at the before mentioned locations within Germany, to reduce travel costs and achieve short reaction times. DWD has the necessary specific facilities to support the technical staff like stock at each division of the department, electronical and mechanical workshops and calibration laboratories for meteorological sensors in Hamburg and Munich. 2 Actual maintenance concept The first step for the assurance of data quality is a regular technical maintenance at the sites. Depending on the type of weather station, the interval range varies from several months at primary observing stations (automated and manned) to at least once a year at some automated secondary observing stations with only a few sensors. In Fehler! Verweisquelle konnte nicht gefunden werden. the number of stations maintained by each service and logistic division and the actual interval range for maintenance or inspection for each station type is given. Table 1: Maintenance per year depending on the type of station To assure that the technical staff in each division is using the same procedures for maintenance, responsible engineers for every type of weather station and sensor type are nominated. They issue the necessary papers for handling, installation and maintenance of the technical equipment. 2

3 To assure a high quality of the measurements, most of the sensors are exchanged regularly with calibrated sensors (pressure, temperature, humidity, wind, radiation and precipitation). After refurbishing at DWD workshops the dismantled sensors are recalibrated in a laboratory and marked with a sticker with the date of calibration on it and then stocked. Additional field checks are performed at each station visit, for example with a pressure transfer standard. If an obvious malfunction or inaccuracy has been determined during maintenance the sensor will be exchanged. Via remote maintenance access all weather stations are checked regularly for technical issues. Any malfunction will be investigated and if needed an additional maintenance onsite will be initiated. Additional service is required if the post-processing with the interactive data control and monitoring system QualiMET indicates malfunctions or if a meteorological observer reported a problem. The technical staff is using a computer based database including mainly the primary and secondary observing stations with their specific instrumentation. Every maintenance work, either remote or on site, is documented in this database too. For each sensor the next date for calibration is calculated automatically, depending on the date of the last calibration and the calibration interval of the sensor type. So the database is indicating the sensors which have to be exchanged at the next inspection. For long time planning of maintenance trips a calendar overview is included too. 3 First definition and further development of calibration intervals Even if the DWD has been operating calibration facilities for several decades a regular and systematic exchange, refurbishing and calibration of sensors has started in To archive the requirements of the ISO9001 certification regarding to the control of monitoring and measuring devices (old chapter 7.6 of ISO 9000) DWD introduced a news system. New relevant documents have been added: a document with rules for calibration and traceability a document with rules for the achievement of valid calibration a document with a table of the valid time after calibration Regular calibration intervals for a variety of sensors were defined. The decision for the individual calibration interval depends on the type of sensor, the experiences of its stability, manufacturer s instructions and the capacity of the calibration laboratories. Some examples are shown in Fehler! Verweisquelle konnte nicht gefunden werden.. Type calibration interval in month value in 2006 actual value Pt100 air thermometer Pt100 soil thermometer 120 not regularly Cup anemometer Ultrasonic anemometer Pressure gauge Humidity transmitter Weighing rain gauge Table 2: Defined calibration interval for different sensors in 2006 and

4 Comparing the two columns in Fehler! Verweisquelle konnte nicht gefunden werden. it can be seen, that only minor changes were made to these intervals. In all cases, the intervals have been slightly increased due to the analysis of the sensor stability or special requirements of service have to be respected. For example in 2013 the calibration interval for the soil thermometers was investigated, because an exchange of soil thermometers has an impact on the surrounding soil and in consequence on the homogeneity of the measurement data. This effect might be larger than the possible drift of the sensor itself after the originally defined calibration interval of 10 years. As high class Pt 100 sensors have in general a very good long time stability, the actual drift of some sensors has been determined. Seven soil thermometers were recalibrated after an operation period of ten years. Figure 2: Result of recalibration of soil thermometers after 10 years of operation After ten years all soil thermometers were within a tolerance of ±0.1 Kelvin. In addition to that 22 other soil thermometers, which have been stored in the warehouse, were recalibrated also. Only two sensors had a tolerance up to 0,2K, the other were within a tolerance of ±0,1 Kelvin too. In consequence it was decided that soil thermometers will not have to be recalibrated after installation. They will only be replaced, if they have a malfunction. Another example is the calibration interval of pressure gauges. DWDs first calibration interval was set to 18 month. In 2014 a bigger sample was investigated. With a few exceptions all pressure gauges were within the accepted tolerance of ±0.15 hpa in the relevant pressure range ( hpa) so DWD decided to increase the interval to 24 month. 4

5 Figure 3: Result of recalibration of pressure gauges after 24 month of operation After improving the equipment in the laboratories a review of the calibration interval was carried out in September All pressure gauges of Vaisala PTB220 type were within the accepted tolerance. Figure 4: Result of recalibration of pressure gauges after 24 month of operation 5

6 DWD is using weighing rain gauges for more than 10 years. In 2013 the calibration interval has been enhanced from 4 to 5 years, because of a good long term stability. In 2017 DWD will replace this rain gauge with a new sensor type. Based on the manufacturers proposal, the calibration interval has been defined to 48 month. This proposal will be investigated in the future, because DWD has only limited experiences with the stability of this rain gauge. The installation of the sensors started in September To prove this interval the calibration laboratories will calibrate a sample from different locations after an operation period of 12, 24, 36 and of course 48 month. Hopefully the calibration interval of 48 month can be confirmed. 4 Development of the quality system at the calibration laboratory To achieve best results and traceability of the measurements, the calibration laboratory has been subjected the accreditation process according to DIN EN ISO/IEC 17025:2005 with the status as a testing laboratory ( ). In 2015 DWD decided to become accredited as a calibration laboratory according to DIN EN ISO/IEC 17025:2005. This accreditation process has been finalized now, so the status has been changed to an accredited calibration laboratory in September 2017 (s. Fehler! Verweisquelle konnte nicht gefunden werden.). The accredited parameters are temperature, relative humidity, absolute pressure, wind speed and direction. 6

7 Figure 5: The accreditation certificate issued by the German accreditation body DAkkS. All instruments in the laboratory are regularly calibrated and the measurements are traceable to national standards provided by the National Metrological Institute of Germany (PTB). The following table gives an overview of the capabilities of the calibration laboratory with its two locations in Hamburg and Oberschleißheim (near Munich). Instrument Undergoing Calibration Calibration Range Reference wind speed sensors 0,5 50 ms -1 anemometer 2D laser doppler burst spectrum analyzer Calibration and Measurement Capability (CMC) 1,2 % of reference but not smaller than 0,1 ms -1 wind direction sensors precision turntable 0,9 Pt C special selected IPRT or SPRT 50 mk temperature transmitter C IPRT with electronic thermometer 60 mk 7

8 heated and unheated humidity transmitters and probes % of relative humidity chilled mirror dewpoint meter 0, x U in % of reference but not smaller than 0,4 % absolute pressure transmitters hpa piston gauge, vacuum gauge Table 3: Capabilities of the accredited DWD calibration laboratories 0,0025 hpa + 0,00002 x P + uncertainty of the reference vacuum measurement It is verified whether the calibration of rain gauges could be accredited in future, too. 8

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Quality assurance for sensors at the Deutscher Wetterdienst (DWD) Quality assurance for sensors at the Deutscher Wetterdienst (DWD) Quality assurance / maintenance / calibration Holger Dörschel, Dr Tilman Holfelder WMO International Conference on Automatic Weather Stations

More information

STATUS OF THE WIGOS DEMONSTRATION PROJECTS

STATUS OF THE WIGOS DEMONSTRATION PROJECTS STATUS OF THE WIGOS DEMONSTRATION PROJECTS Demonstration Project Morocco Strengthening Moroccan RIC Capacities (Submitted by Rabia Merrouchi, National Meteorological Service of Morocco (DMN)) February

More information

Meteorological Service

Meteorological Service Meteorological Service The Meteorological Service in Oman was established in 1973. Oman joined ICAO in 1973 Oman joined WMO in 1975. Meteorological Service is under the framework of the Directorate General

More information

for JMA/WMO Training Workshop on CALIBRATION AND MAINTENANCE OF METEOROLOGICAL INSTRUMENTS

for JMA/WMO Training Workshop on CALIBRATION AND MAINTENANCE OF METEOROLOGICAL INSTRUMENTS for JMA/WMO Training Workshop on CALIBRATION AND MAINTENANCE OF METEOROLOGICAL INSTRUMENTS in RA II (ASIA) 0. Overview of JMA 3 4 1. Surface observation station 5-8 2. Instruments in operational use 9-14

More information

CONTENT 2. ORGANIZATION 3. SERVICES. instruments. (3)Activities of RIC Tsukuba

CONTENT 2. ORGANIZATION 3. SERVICES. instruments. (3)Activities of RIC Tsukuba 1 CONTENT 1. INTRODUCTION 2. ORGANIZATION 3. SERVICES (1)Quality assurance of meteorological instruments (2)Research and development (3)Activities of RIC Tsukuba 2 Meteorological Instruments Center (MIC)

More information

OverviewofRIC Tsukuba (Japan,RAII )

OverviewofRIC Tsukuba (Japan,RAII ) OverviewofRIC Tsukuba (Japan,RAII ) 7Nov.2016 Hiroshi Yoshimura Scientific Officer Regional Instrument Centre Tsukuba Observing Division, Observing Department Outline 1. WMO Regional Instrument Center

More information

World Meteorological Organization

World Meteorological Organization World Meteorological Organization CIMO/WIGOS EXPLORATORY WORKSHOP: IMPROVING SURFACE-BASED DATA QUALITY THOUGH IMPROVED STANDARDIZATION OF PRACTICES AND PROCEDURES Langen, Germany, 5-3 December 2014 Regional

More information

Introductions to RIC-Beijing. NAN Xuejing, CUI Xiai Meteorological Observation Center China Meteorological Administration March,2018

Introductions to RIC-Beijing. NAN Xuejing, CUI Xiai Meteorological Observation Center China Meteorological Administration March,2018 Introductions to RIC-Beijing NAN Xuejing, CUI Xiai China Meteorological Administration March,2018 Contents 1. China Meteorological Administration (CMA) Organization 2. (MOC) Functional Structure Responsibilities

More information

QualiMET 2.0. The new Quality Control System of Deutscher Wetterdienst

QualiMET 2.0. The new Quality Control System of Deutscher Wetterdienst QualiMET 2.0 The new Quality Control System of Deutscher Wetterdienst Reinhard Spengler Deutscher Wetterdienst Department Observing Networks and Data Quality Assurance of Meteorological Data Michendorfer

More information

1 Introduction. Station Type No. Synoptic/GTS 17 Principal 172 Ordinary 546 Precipitation

1 Introduction. Station Type No. Synoptic/GTS 17 Principal 172 Ordinary 546 Precipitation Use of Automatic Weather Stations in Ethiopia Dula Shanko National Meteorological Agency(NMA), Addis Ababa, Ethiopia Phone: +251116639662, Mob +251911208024 Fax +251116625292, Email: Du_shanko@yahoo.com

More information

Introduction of RIC Tsukuba (Japan, RAII)

Introduction of RIC Tsukuba (Japan, RAII) Introduction of RIC Tsukuba (Japan, RAII) 20 March, 2018 Kouichi NAKASHIMA Scientific Officer Regional Instrument Centre Tsukuba Observing Division, Observing Department Outline 1. Overview of Organization

More information

AUTOMATION OF OBSERVING NETWORK IN BMKG

AUTOMATION OF OBSERVING NETWORK IN BMKG Agency for Meteorology, Climatology, and Geophysics of the Republic of Indonesia AUTOMATION OF OBSERVING NETWORK IN BMKG Agung Saifulloh Majid, G.S. Budhi Dharmawan, Damianus Tri Heryanto, Untung Merdijanto

More information

MeteoSwiss acceptance procedure for automatic weather stations

MeteoSwiss acceptance procedure for automatic weather stations MeteoSwiss acceptance procedure for automatic weather stations J. Fisler, M. Kube, E. Grueter and B. Calpini MeteoSwiss, Krähbühlstrasse 58, 8044 Zurich, Switzerland Phone:+41 44 256 9433, Email: joel.fisler@meteoswiss.ch

More information

REGIONAL INSTRUMENT CENTER (RIC) MANILA Philippines (Regional Association V)

REGIONAL INSTRUMENT CENTER (RIC) MANILA Philippines (Regional Association V) REGIONAL INSTRUMENT CENTER (RIC) MANILA Philippines (Regional Association V) Ferdinand Barcenas Contact Person, RIC Manila fybarce8@yahoo.com, ferdie@pagasa.dost.gov.ph PAGASA STRUCTURE PAGASA STRUCTURE

More information

Practicaluseof standardmeteorolo gicalinstruments indomesticmeteorolo gical observationnetworks

Practicaluseof standardmeteorolo gicalinstruments indomesticmeteorolo gical observationnetworks Practicaluseof standardmeteorolo gicalinstruments indomesticmeteorolo gical observationnetworks Satoshi HAGIYA Meteorolo gicalinst rumentcenter WMO/RegionalInst rumentcentre (RIC) Tsukuba JapanMeteorologicalAgency(JMA)

More information

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations Calibration General Principles & Theory, Equipment Considerations Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link

More information

Operational Applications of Awos Network in Turkey

Operational Applications of Awos Network in Turkey Operational Applications of Awos Network in Turkey by Soner Karatas Turkish State Meteorological Service Electronic Observing Systems Division Kütükcü Alibey Cad. No:4 06120 Kalaba-Ankara-TURKEY Tel:+90-312-302

More information

Wali Ullah Khan Pakistan Meteorological Department

Wali Ullah Khan Pakistan Meteorological Department An overview of Weather Observation practices over Pakistan By Wali Ullah Khan Pakistan Meteorological Department JMA/WMO TRAINING WORKSHOP ON CALIBRATION AND MAINTENANCE OF METEOROLOGICAL INSTRUMENTS IN

More information

Parallel measurements at German climate reference stations: DWD s approach to compare manual vs. automatic observations

Parallel measurements at German climate reference stations: DWD s approach to compare manual vs. automatic observations Parallel measurements at German climate reference stations: DWD s approach to compare manual vs. automatic observations Frank Kaspar and Lisa Hannak Deutscher Wetterdienst, National Climate Monitoring,

More information

THE CLIMATE INFORMATION MODULE

THE CLIMATE INFORMATION MODULE Climate Information and Prediction Services (CLIPS) -Curriculum- THE CLIMATE INFORMATION MODULE designed by Dipl.-Met. Peer Hechler Deutscher Wetterdienst P.O. Box 10 04 65 63004 Offenbach Germany THE

More information

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014 Calibration General Principles & Theory, Equipment Considerations Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link

More information

Deutscher Wetterdienst

Deutscher Wetterdienst WakeNet3-Greenwake Workshop Wake Vortex & Wind Monitoring Sensors in all weather conditions DWD s new Remote Wind Sensing Equipment for an Integrated Terminal Weather System (ITWS) Frank Lehrnickel Project

More information

TESTING METEOROLOGICAL SENSORS ON SHIPS SENSORS FOR SHIPBORNE AUTOMATIC WEATHER STATIONS. Simone Griesel

TESTING METEOROLOGICAL SENSORS ON SHIPS SENSORS FOR SHIPBORNE AUTOMATIC WEATHER STATIONS. Simone Griesel P (4) TESTING METEOROLOGICAL SENSORS ON SHIPS SENSORS FOR SHIPBORNE AUTOMATIC WEATHER STATIONS Simone Griesel Deutscher Wetterdienst, Frahmredder 95, 9 Hamburg, Germany simone.griesel@dwd.de ABSTRACT Beside

More information

WeatherHawk Weather Station Protocol

WeatherHawk Weather Station Protocol WeatherHawk Weather Station Protocol Purpose To log atmosphere data using a WeatherHawk TM weather station Overview A weather station is setup to measure and record atmospheric measurements at 15 minute

More information

Withdrawn at December 31, Guideline DKD-R 5-7. Calibration of Climatic Chambers DEUTSCHER KALIBRIERDIENST

Withdrawn at December 31, Guideline DKD-R 5-7. Calibration of Climatic Chambers DEUTSCHER KALIBRIERDIENST DEUTSCHER KALIBRIERDIENST Guideline DKD-R 5-7 Calibration of Climatic Chambers Edition 07/2004 English translation 02/2009 Page 1 of 31 Published by the Accreditation Body of the Deutscher Kalibrierdienst

More information

The Fundamentals of Moisture Calibration

The Fundamentals of Moisture Calibration The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity

More information

Interview with Dr. Jaya Singh. 1. How critical is weather monitoring in a solar power plant and what are all the parameters to be monitored?

Interview with Dr. Jaya Singh. 1. How critical is weather monitoring in a solar power plant and what are all the parameters to be monitored? Interview with Dr. Jaya Singh Dr. Jaya Singh is an entrepreneur and scientist with experience in a range of industries including biotech, environmental sciences, and agriculture. She currently drives strategy

More information

BIAN ZEQIANG Senior Engineer National Center for Meteorological Metrology Meteorological Observation Cent China Meteorological Administration

BIAN ZEQIANG Senior Engineer National Center for Meteorological Metrology Meteorological Observation Cent China Meteorological Administration Introduction of RIC-Beijing BIAN ZEQIANG Senior Engineer National Center for Meteorological Metrology Meteorological Observation Cent China Meteorological Administration Outline 1. China Meteorology Administration

More information

* * * Table (1) Table (2)

* * * Table (1) Table (2) A step Forward to Atomize the Sudan Meteorological Authority (SMA) Net work Y.S. Odan Surface Instruments Department Tel: 00249 912220246 E-mail yaseen@ersad.gov.sd Abstract AWS has been introduced to

More information

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

2.1. Accuracy, n- how close the indication of the thermometer is to the true value. AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria

More information

Country Report for Japan (Submitted by Kenji Akaeda, Japan Meteorological Agency)

Country Report for Japan (Submitted by Kenji Akaeda, Japan Meteorological Agency) Japan JMA/WMO Workshop on Quality Management in Surface, Climate and Upper-air Observations in RA II (Asia) Tokyo, Japan 27 30 July 2010 Doc. Japan (10.VII.2010) Country Report for Japan (Submitted by

More information

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR M. Araya 1, D. Almeida 2 1 National Laboratory of Temperature of Chile, LCPNT, Santiago, Chile 2 Ecuadorian Standardization

More information

Lisa Hannak Karsten Friedrich, Florian Imbery, Frank Kaspar Deutscher Wetterdienst National climate monitoring

Lisa Hannak Karsten Friedrich, Florian Imbery, Frank Kaspar Deutscher Wetterdienst National climate monitoring Changes in the measurement systems and the impact on the homogeneity of long time series A study based on parallel measurements from German reference stations Lisa Hannak (Lisa.Hannak@dwd.de) Karsten Friedrich,

More information

CAPACITY BUILDING AND DEVELOPMENT OF THE RIC IN MANILA (PHILIPPINES)

CAPACITY BUILDING AND DEVELOPMENT OF THE RIC IN MANILA (PHILIPPINES) CAPACITY BUILDING AND DEVELOPMENT OF THE RIC IN MANILA (PHILIPPINES) by Ferdinand Y. Barcenas 1 ABSTRACT The Regional Instrument Centre for the South-West Pacific in Manila, Philippines was recognized

More information

Quality Assurance and Quality Control

Quality Assurance and Quality Control Quality Assurance and Quality Control of Surface Observations in JMA Japan Meteorological Agency Hakaru MIZUNO "Guide to Meteorological Instruments and Methods of Observation", WMO-No.8, 7th ed., 2008.

More information

Interaction between wind turbines and the radar systems operated by the meteorological services

Interaction between wind turbines and the radar systems operated by the meteorological services Interaction between wind turbines and the radar systems operated by the meteorological services Challenges facing the DWD Carmen Diesner, Measurement Technology Division (TI 22), Deutscher Wetterdienst,

More information

Instruments and Methods of Observation Programme, the Report of the President of CIMO. Report to Cg-XV May 2007 Dr. J. Nash President of CIMO

Instruments and Methods of Observation Programme, the Report of the President of CIMO. Report to Cg-XV May 2007 Dr. J. Nash President of CIMO Instruments and Methods of Observation Programme, the Report of the President of CIMO Report to Cg-XV May 2007 Dr. J. Nash President of CIMO 1 CIMO Mission To promote and facilitate international standardisation

More information

Figure 1. Daily variation of air temperature

Figure 1. Daily variation of air temperature Comparative analysis of the meteorological data acquired on standard equipment and by automatic weather station of CAMPBELL SCIENTIFIC, INC Company Kudekov T.K. Director-General of the KAZHYDROMET 050022

More information

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS W. Beyer and W. Pahl Physikalisch-Technische Bundesanstalt (PTB) 38116 Braunschweig, Germany Abstract: In accordance with ISO 17025,

More information

g/cm³ DAkkS Calibration Laboratory for Temperature and Density High-precision calibration for temperature and density

g/cm³ DAkkS Calibration Laboratory for Temperature and Density High-precision calibration for temperature and density High-precision calibration for temperature and density g/cm³ F K C DAkkS Calibration Laboratory for Temperature and Density Accredited DAkkS laboratory (D-K-15223-01-00) for temperature and density measurements

More information

Development of Innovative Technology to Provide Low-Cost Surface Atmospheric Observations in Data-sparse Regions

Development of Innovative Technology to Provide Low-Cost Surface Atmospheric Observations in Data-sparse Regions Development of Innovative Technology to Provide Low-Cost Surface Atmospheric Observations in Data-sparse Regions Paul Kucera and Martin Steinson University Corporation for Atmospheric Research/COMET 3D-Printed

More information

Site assessment for a type certification icing class Fraunhofer IWES Kassel, Germany

Site assessment for a type certification icing class Fraunhofer IWES Kassel, Germany Site assessment for a type certification icing class Kai Freudenreich 1, Michael Steiniger 1, Zouhair Khadiri-Yazami 2, Ting Tang 2, Thomas Säger 2 1 DNV GL Renewables Certification, Hamburg, Germany kai.freudenreich@dnvgl.com,

More information

Commissioning of the Beta Secondary Standard (BSS2)

Commissioning of the Beta Secondary Standard (BSS2) Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,

More information

Specifications for a Reference Radiosonde for the GCOS Reference. Upper-Air Network (GRUAN)

Specifications for a Reference Radiosonde for the GCOS Reference. Upper-Air Network (GRUAN) Specifications for a Reference Radiosonde for the GCOS Reference Upper-Air Network (GRUAN) By the Working Group on Atmospheric Reference Observations (WG-ARO) Final Version, October 2008 1. Introduction

More information

3.2 Wind direction / wind velocity

3.2 Wind direction / wind velocity 3.2 Wind direction / wind velocity The direction from which air moves to is called the wind direction, and the distance air moves per unit time is the wind velocity. Wind has to be measured not only as

More information

CALIBRATION REPORT FOR THERMOHYDROMETER

CALIBRATION REPORT FOR THERMOHYDROMETER CALIBRATION REPORT FOR THERMOHYDROMETER SAMPLE CUSTOMER S/N: XXXXX Report Number: Y205188 Page 1 of 5 Report No. Y205188 ICL CALIBRATION LABORATORIES, INC. ISO/IEC 17025 and ANSI/NCSL Z540-1 accredited

More information

CERTIFICATE OF CALIBRATION

CERTIFICATE OF CALIBRATION Page 1 of 5 CERTIFICATE OF CALIBRATION 1K Thermistor CERTIFICATE NO.: MANUFACTURER: Measurement Specialties DATE OF CALIBRATION: 26-Apr-17 TC15249 CUSTOMER: AMBIENT TEMPERATURE: (21 ± 3) C JOB NUMBER:

More information

STATISTIQUE DE PERFORMANCE

STATISTIQUE DE PERFORMANCE INTERCOMPARISON OF STANDARD CALIBRATION, METHODS AND MEASUREMENTS FOR PRESSURE, TEMPERATURE AND HUMIDITY BETWEEN RIC CASABLANCA, MOROCCO AND RIC Ljubljana SLOVENIA 3(10) azizmounir@gmail.com The operators

More information

CALIBRATION CENTER. Republic of Turkey. Ministry of Forestry and. Water Affairs. Turkish State Meteorological Service.

CALIBRATION CENTER. Republic of Turkey. Ministry of Forestry and. Water Affairs. Turkish State Meteorological Service. Republic of Turkey Ministry of Forest and Water Affairs Republic of Turkey Ministry of Forestry and Water Affairs Turkish State Meteorological Service CALIBRATION CENTER Turkish State Meteorological Service

More information

STATUS OF INSTRUMENTS TRACEABILITY IN NMHSs

STATUS OF INSTRUMENTS TRACEABILITY IN NMHSs WORLD METEOROLOGICAL ORGANIZATION COMMISSION FOR INSTRUMENTS AND METHODS OF OBSERVATION OPAG CAPACITY BUILDING CIMO EXPERT TEAM ON REGIONAL INSTRUMENT CENTRES, QUALITY MANAGEMENT SYSTEMS AND COMMERCIAL

More information

IMPROVING SOIL MEASUREMENTS FOR BOUNDARY LAYER RESEARCH

IMPROVING SOIL MEASUREMENTS FOR BOUNDARY LAYER RESEARCH IMPROVING SOIL MEASUREMENTS FOR BOUNDARY LAYER RESEARCH Hannelore Bloemink, Jan Bijma Royal Netherlands Meteorological Institute (KNMI) Instrumentation Division P.O. Box 201, 3730 AE De Bilt, The Netherlands

More information

Meteorological QA/QC

Meteorological QA/QC Meteorological QA/QC Howard Schmidt, MS, MBA US EPA Region 3 Air Protection Division Air Monitoring Quality Assurance Workshop June 26, 2014 Overview Current state of R3 agency met monitoring Why? Where?

More information

Complex Terrain (EDUCT) experiment, conducted by the National Center for Atmospheric

Complex Terrain (EDUCT) experiment, conducted by the National Center for Atmospheric Alex Ameen Shenandoah Trip Paper I visited Shenandoah National Park on April 11, 2009 to investigate the Education in Complex Terrain (EDUCT) experiment, conducted by the National Center for Atmospheric

More information

Comparison of Vaisala Radiosondes RS41 and RS92 WHITE PAPER

Comparison of Vaisala Radiosondes RS41 and RS92 WHITE PAPER Comparison of Vaisala Radiosondes RS41 and RS92 WHITE PAPER Table of Contents CHAPTER 1 Introduction... 4 CHAPTER 2 Key Improvements in RS41... 5 CHAPTER 3 RS41 and RS92 Comparison Tables... 6 CHAPTER

More information

and soils characterizing would be defined.

and soils characterizing would be defined. Working out of the basic network for monitoring of natural and climatic processes in Siberia M.V. Kabanov, V.V. Zuev, V.А. Krutikov Institute of Monitoring of Climatic and Ecological Systems SB RAS, Tomsk,

More information

Expedited Filing Draft August 22, 2017

Expedited Filing Draft August 22, 2017 Applicability 1 Section 304.9 applies to: (a) (b) Requirements Expedited Filing Draft August 22, 2017 the legal owner of a wind or solar aggregated generating facility connected to the interconnected electric

More information

Advisory Circular. U.S. Department of Transportation Federal Aviation Administration

Advisory Circular. U.S. Department of Transportation Federal Aviation Administration U.S. Department of Transportation Federal Aviation Administration Subject: Use of Liquid Water Equivalent System to Determine Holdover Times or Check Times for Anti-Icing Fluids Advisory Circular Date:

More information

Interlaboratory Comparison & Proficiency Testing Program

Interlaboratory Comparison & Proficiency Testing Program Extended scope for the organization of proficiency tests in accordance with the requirements of ISO/IEC 17043 VSL, the Dutch Metrology Institute, has significantly extended the scope for their accreditation

More information

Windcube TM Pulsed lidar wind profiler Overview of more than 2 years of field experience J.P.Cariou, R. Parmentier, M. Boquet, L.

Windcube TM Pulsed lidar wind profiler Overview of more than 2 years of field experience J.P.Cariou, R. Parmentier, M. Boquet, L. Windcube TM Pulsed lidar wind profiler Overview of more than 2 years of field experience J.P.Cariou, R. Parmentier, M. Boquet, L.Sauvage 15 th Coherent Laser Radar Conference Toulouse, France 25/06/2009

More information

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 2017-03-06 revised 2018-02-13 (changed schedule) Task and Purpose of the Comparison According to the rules set up by the CIPM MRA the consultative

More information

Comparative analysis of data collected by installed automated meteorological stations and manual data in Central Asia.

Comparative analysis of data collected by installed automated meteorological stations and manual data in Central Asia. Comparative analysis of data collected by installed automated meteorological stations and manual data in Central Asia. Merkushkin Aleksandr Uzhydromet 71 K. Makhsumov str., 100052 Tashkent, Uzbekistan

More information

Optimizing the energy consumption and operational availability of point heaters using weather forecasts

Optimizing the energy consumption and operational availability of point heaters using weather forecasts Computers in Railways XIV 757 Optimizing the energy consumption and operational availability of point heaters using weather forecasts C. Weiß, T. Käsemann & K. A. Klinge i3mainz Institute for Spatial Information

More information

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years Int J Thermophys (2017) 38:161 DOI 10.1007/s10765-017-2291-x TEMPMEKO 2016 Development of the High-Temperature Dew-Point Generator Over the Past 15 Years R. Bosma 1 J. Nielsen 2 A. Peruzzi 1 Received:

More information

The AWS based operational urban network in Milano: achievements and open questions.

The AWS based operational urban network in Milano: achievements and open questions. The AWS based operational urban network in Milano: achievements and open questions. Frustaci Giuseppe, Curci Savino, Pilati Samantha, Lavecchia Cristina, Paganelli Chiara Fondazione Osservatorio Meteorologico

More information

Interlaboratory Comparison & Proficiency Testing Program

Interlaboratory Comparison & Proficiency Testing Program Extended scope for the organization of proficiency tests in accordance with the requirements of ISO/IEC 17043 VSL, the Dutch Metrology Institute, has significantly extended the scope for their accreditation

More information

Optimal combination of NWP Model Forecasts for AutoWARN

Optimal combination of NWP Model Forecasts for AutoWARN ModelMIX Optimal combination of NWP Model Forecasts for AutoWARN Tamas Hirsch, Reinhold Hess, Sebastian Trepte, Cristina Primo, Jenny Glashoff, Bernhard Reichert, Dirk Heizenreder Deutscher Wetterdienst

More information

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60 Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60 Bernd Meißner Physikalisch-Technische Bundesanstalt Weighing Instruments Laboratory Braunschweig Abstract

More information

Quality control methodology for temperature data of Automatic Weather Stations with non-wooden radiation shield

Quality control methodology for temperature data of Automatic Weather Stations with non-wooden radiation shield Quality control methodology for temperature data of Automatic Weather Stations with non-wooden radiation shield Eng. Nuwan Kumarasinghe Chief Electronics Engineer Department of Meteorology, Sri Lanka nuwan1960@gmail.com

More information

Certificate: / 29 April 2014

Certificate: / 29 April 2014 Test report: 936/21218896/A of 20 September 2013 Initial certification: 01 April 2014 Date of expiry: 31 March 2019 Publication: BAnz AT 01 April 2014 B12, chapter IV, No. 5.1 Approved application The

More information

Interlaboratory Comparison & Proficiency Testing Program

Interlaboratory Comparison & Proficiency Testing Program Extended scope for the organization of proficiency tests in accordance with the requirements of ISO/IEC 17043 VSL, the Dutch Metrology Institute, has significantly extended the scope for their accreditation

More information

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION IMEKO 22 nd TC3, 15 th TC5 and 3 rd TC22 International Conferences 3 to 5 February, 2014, Cape Town, Republic of South Africa PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION R. Kumme

More information

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012 Revision 1.2 October 15, 2012 PAGE II TABLE OF CONTENTS TABLE OF CONTENTS... 1 GUIDANCE...2 1.0 Introduction... 2 2.0 Uncertainty and Traceability... 2 3.0 Need for Traceability... 3 3.1 What does traceability

More information

Certificate: / 30. April 2015

Certificate: / 30. April 2015 Test report: 936/21226418/A of 29 September 2014 Initial certification: 2. April 2015 Date of expiry: 1. April 2020 Publication: BAnz AT 02 April 2015 B5, chapter III number 3.1 Approved application The

More information

Components of uncertainty in measurements by means of AWS

Components of uncertainty in measurements by means of AWS Components of uncertainty in measurements by means of AWS G. Coppa, F. Sanna, C. Musacchio, A. Merlone INRiM Istituto Nazionale di Ricerca Metrologica Outline Metrology and Meteorology Uncertainty in temperature

More information

Meteorological Instruments for Solar Energy

Meteorological Instruments for Solar Energy Meteorological Instruments for Solar Energy gillinstruments.com Gill Instruments has been designing and manufacturing professional use compact weather stations and ultrasonic wind speed and direction sensors

More information

CERTIFICATE of Product Conformity (QAL1 )

CERTIFICATE of Product Conformity (QAL1 ) ATÜVRheinland Precisely Right. CERTIFICATE of Product Conformity (QAL1 ) Certificate No.: 0000040212 03 Certified AMS: Manufacturer: Test lnstitute: Fidas 200 S respectively Fidas 200 for pariiculate matter

More information

A Case for Periodic Calibration or Verification of RTDs

A Case for Periodic Calibration or Verification of RTDs A Case for Periodic Calibration or Verification of RTDs If you are experiencing audio problems please call the teleconference number below Phone 650-479-3208 Access code 668 676 434 Your Host and Presenter

More information

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1 Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1 Updated at 12-December-2012 Task and Purpose of the Comparison According to the rules set up by the CIPM

More information

The new Closed Loop pigsar and its projected CMC. Jos van der Grinten, Detlef Vieth, Bodo Mickan

The new Closed Loop pigsar and its projected CMC. Jos van der Grinten, Detlef Vieth, Bodo Mickan The new Closed Loop pigsar and its projected CMC Jos van der Grinten, Detlef Vieth, Bodo Mickan Introduction pigsar Calibration facilities Closed Loop pigsar (CLP) Calibration model Uncertainty analysis

More information

A COMPARISON OF TWO WEATHER DATA ACQUISITION METHODS FOR THE CALIBRATION OF THE PTB GEODETIC BASELINE

A COMPARISON OF TWO WEATHER DATA ACQUISITION METHODS FOR THE CALIBRATION OF THE PTB GEODETIC BASELINE XX IMEKO World Congress Metrology for Green Growth 9-14 September 212, Busan, Republic of Korea A COMPARISON OF TWO WEATHER DATA ACQUISITION METHODS FOR THE CALIBRATION OF THE PTB GEODETIC BASELINE J.

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: AMC LABS S.A de C.V (Hereinafter called the Organization)

More information

METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES AFTER REINSTALLATION

METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES AFTER REINSTALLATION IMEKO 22 nd TC3, 12 th TC5 and 3 rd TC22 International Conferences 3 to 5 February, 2014, Cape Town, Republic of South Africa METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES

More information

Standard Practices for Air Speed Calibration Testing

Standard Practices for Air Speed Calibration Testing Standard Practices for Air Speed Calibration Testing Rachael V. Coquilla Bryza Wind Lab, Fairfield, California Air speed calibration is a test process where the output from a wind measuring instrument

More information

Guidelines on Quality Control Procedures for Data from Automatic Weather Stations

Guidelines on Quality Control Procedures for Data from Automatic Weather Stations Guidelines on Quality Control Procedures for Data from Automatic Weather Stations Igor Zahumenský Slovak Hydrometeorological Institute SHMI, Jeséniova 17, 833 15 Bratislava, Slovakia Tel./Fax. +421 46

More information

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT Scott Crone North American Sales Manager AMETEK Calibration Instruments 8600 Somerset Drive Largo, Florida 33773 USA INTRODUCTION Like any other piece of equipment,

More information

DOE S RADIOLOGICAL TRACEABILITY PROGRAM FOR RADIOANALYTICAL MEASUREMENTS: WORK IN PROGRESS

DOE S RADIOLOGICAL TRACEABILITY PROGRAM FOR RADIOANALYTICAL MEASUREMENTS: WORK IN PROGRESS DOE S RADIOLOGICAL TRACEABILITY PROGRAM FOR RADIOANALYTICAL MEASUREMENTS: WORK IN PROGRESS D. E. McCurdy, Duke Engineering & Services, Bolton, MA and J. S. Morton, Department of Energy, Office of Environmental

More information

Michael de Podesta. A Global Land-Surface Climate Reference Network: Technical Reflections. AOPC Exeter 30 th March 2017

Michael de Podesta. A Global Land-Surface Climate Reference Network: Technical Reflections. AOPC Exeter 30 th March 2017 A Global Land-Surface Climate Reference Network: Technical Reflections Thank you! To Michael Palleckiand Mark Hall from US CRN for extended conversations AOPC Exeter 30 th March 2017 Michael de Podesta

More information

CLIMATE CHANGE ADAPTATION BY MEANS OF PUBLIC PRIVATE PARTNERSHIP TO ESTABLISH EARLY WARNING SYSTEM

CLIMATE CHANGE ADAPTATION BY MEANS OF PUBLIC PRIVATE PARTNERSHIP TO ESTABLISH EARLY WARNING SYSTEM CLIMATE CHANGE ADAPTATION BY MEANS OF PUBLIC PRIVATE PARTNERSHIP TO ESTABLISH EARLY WARNING SYSTEM By: Dr Mamadou Lamine BAH, National Director Direction Nationale de la Meteorologie (DNM), Guinea President,

More information

Examination of different versions of bird repelling techniques for ultrasonic anemometers

Examination of different versions of bird repelling techniques for ultrasonic anemometers Examination of different versions of bird repelling techniques for ultrasonic anemometers K. Schubotz, H. Dörschel, Dr. E. Lanzinger, M. Theel Abstract: German Meteorological Service Section: TI 23 / TI

More information

Increased Requirements for Higher Nominal Forces, Necessities and Possibilities to Measure Them

Increased Requirements for Higher Nominal Forces, Necessities and Possibilities to Measure Them B 3.1 Increased Requirements for Higher Nominal Forces, Necessities and Possibilities to Measure Them Dr. Schäfer, André Hottinger Baldwin Messtechnik GmbH Im Tiefen See 45, 64293 Darmstadt andre.schaefer@hbm.com

More information

IMS4 ARWIS. Airport Runway Weather Information System. Real-time data, forecasts and early warnings

IMS4 ARWIS. Airport Runway Weather Information System. Real-time data, forecasts and early warnings Airport Runway Weather Information System Real-time data, forecasts and early warnings Airport Runway Weather Information System FEATURES: Detection and prediction of runway conditions Alarms on hazardous

More information

D501 and D501-RS. Microscanner D-Series IR Thermometers. no effect. no effect. no effect. no effect no effect

D501 and D501-RS. Microscanner D-Series IR Thermometers. no effect. no effect. no effect. no effect no effect Microscanner D-Series The Only Certified Accurate Surface Instruments in the World A MUST FOR ISO 9001 ISO 9002 ISO 9003 TRACEABILITY PROGRAMS Common Surface Measurement Errors D501 and D501-RS Microscanner

More information

MetConsole AWOS. (Automated Weather Observation System) Make the most of your energy SM

MetConsole AWOS. (Automated Weather Observation System) Make the most of your energy SM MetConsole AWOS (Automated Weather Observation System) Meets your aviation weather needs with inherent flexibility, proven reliability Make the most of your energy SM Automated Weather Observation System

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer High Performance Chilled Mirror Hygrometer With Cable Mounted Measuring Heads Highly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct

More information

Integrating third party data from partner networks: Quality assessment using MeteoSwiss Meteorological Certification procedure

Integrating third party data from partner networks: Quality assessment using MeteoSwiss Meteorological Certification procedure Federal Department of Home Affairs FDHA Federal Office of Meteorology and Climatology MeteoSwiss Integrating third party data from partner networks: Quality assessment using MeteoSwiss Meteorological Certification

More information

MSC Monitoring Renewal Project. CMOS 2012 Montreal, Quebec Thursday, May 31 Martin Elie on behalf of Dave Wartman

MSC Monitoring Renewal Project. CMOS 2012 Montreal, Quebec Thursday, May 31 Martin Elie on behalf of Dave Wartman MSC Monitoring Renewal Project CMOS 2012 Montreal, Quebec Thursday, May 31 Martin Elie on behalf of Dave Wartman Presentation Overview Context Monitoring Renewal Components Conclusions Q & A Page 2 Context

More information

Planning and Situation of the Meteorological Observation in The Republic of Korea

Planning and Situation of the Meteorological Observation in The Republic of Korea Republic of Korea JMA/WMO Workshop on Quality Management in Surface, Climate and Upper-air Observations in RA II (Asia) Tokyo, Japan 27-30 July 2010 Doc. Country (.VII.2010) Planning and Situation of the

More information

STATISTICAL PROCESS CONTROL - THE IMPORTANCE OF USING CALIBRATED MEASUREMENT EQUIPMENT. CASE STUDY

STATISTICAL PROCESS CONTROL - THE IMPORTANCE OF USING CALIBRATED MEASUREMENT EQUIPMENT. CASE STUDY Proceedings of the 6th International Conference on Mechanics and Materials in Design, Editors: J.F. Silva Gomes & S.A. Meguid, P.Delgada/Azores, 26-30 July 2015 PAPER REF: 5376 STATISTICAL PROCESS CONTROL

More information

Kestrel 4500 Pocket Weather Tracker

Kestrel 4500 Pocket Weather Tracker Kestrel 4500 Pocket Weather Tracker For years our customers have been asking for wind direction along with wind speed. New for 2007, the Kestrel 4500 does just that with its built in digital compass. But

More information

A calibration facility for automatic weather stations

A calibration facility for automatic weather stations METEOROLOGICAL APPLICATIONS Meteorol. Appl. 22: 842 846 (2015) Published online 16 November 2015 in Wiley Online Library (wileyonlinelibrary.com) DOI: 10.1002/met.1514 A calibration facility for automatic

More information