1 Calibration General Principles & Theory, Equipment Considerations

2 Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link to Atomic Clock) True Value: The best estimate of the true value. To the Metrologist the true value is never known Accuracy: The deviation from the estimated true value. Ex: the true value was 0.01 C and the instrument read 0.05 C (error of 0.04 C) Precision: How close measurements can be made Resolution: The smallest increment of the measurement Repeatability: The ability to reproduce the same reading Uncertainty: A statistical probability of a measurement being within a specified tolerance: Ex. ±0.5% with 95% confidence Calibration: The process of recording an instrument s reading against a reference standard Correction: The process of adjusting an instrument s reading based on the measurements of a reference standard.

3 Statement of Accuracy True value = The best estimate of the actual value Confidence Level: The probability that the true value lies in the given range Uncertainty is is defined as the tolerance band in which (at k=2) a measured value will be in agreement with the true value for 95% of the measurements made.

4 Calibration Traceability Hierarchy MICHELL TECHNOLOGIES FOR MOISTURE

5 UKAS Calibration Hierarchy MICHELL TECHNOLOGIES FOR MOISTURE

6 Factors Limiting Performance of a Calibration Methodology Physical facilities of the laboratory used temperature control air quality stability of electrical power supply Calibration equipment Accuracy Repeatability Traceability Quality of staff training

7 Calibration schedule MICHELL TECHNOLOGIES FOR MOISTURE

8 Calibration All Michell hygrometers go through a calibration regime in a purpose built laboratories in the UK, Holland and Japan. MICHELL TECHNOLOGIES FOR MOISTURE Traceable to NPL or NIST UKAS accredited We use the systems to calibrate: Capacitive dewpoint sensors Cooled mirror dewpoint meters Relative humidity probes Electrolytic hygrometers

9 Considerations What devices do you want to calibrate What parameters are the devices measuring? What are the %RH & T, or DP measurement ranges of the devices? What are the stated accuracies and resolutions of the devices? Does the calibration need to be performed over the entire operating range of the device? Do you want to calibrate at set points, or do they want to be able to make fine adjustments? Do you want a system which has manual, semi, or fully automated adjustment of set points? Do you want a system which has manual, semi, or fully automated logging of devices under test and the reference? Consider the dryer and generator Consider the reference instrument Are these devices sensors or instruments? What size and shape are they? Do the devices require a flow past/through, or are they making an environmental measurement? Consider an appropriate calibration manifold or chamber

10 Calibration Systems: general overview Traceability to National Standards Air Source Dryer Reference Dewpointmeter Generator Test Chamber Unit under test

11 Pressure Swing Dryers Provides dried air to moisture & dew point generators in absence of suitable site air Two column timed re-generation Oil-free compressed air feed Self-contained unit Two ranges available C dewpoint -100 C dewpoint

12 Moisture Generation to -75 C Enclosure is T controlled at 35 C Two-stage mixing of full dry and saturated gas flows allows high precision adjustment of generated dew-points Generation of dew-points from - 75 Cdp up to ambient T Designed to be fed by a pressure swing dryer or dry instrument air

13 Manual Dewpoint Generator Model DG-2 Range -75 to +20 o C Dewpoint Two Stage Flow Mixing Type MICHELL TECHNOLOGIES FOR MOISTURE

14 Automatic Dewpoint Generator Model DG-4 Range -75 to +20 o C Dewpoint Two Stage Flow Mixing up to 11 Points MICHELL TECHNOLOGIES FOR MOISTURE

15 Moisture generation to -100 C Temperature controlled enclosure Three-stage mass flow controlled flow mixing Generation of dew-points from -90 Cdp up to ambient T ±0.25 Cdp output stability

16 Chilled mirrors come in different types with different ranges.

17 Cooled Mirror Advantages fundamental measurement principle high accuracy drift free operation traceability to national standards

18 Basic Cooled mirror technology The cooled mirror sensor principle : cooling of a metal mirror by a thermoelectric cooler. At dew point, the mirror temperature is measured by a precision thermometer. Fundamental, high accuracy and long-term stability. Accuracy is limited by the PRT Range is limited by cooling methods and number of peltier used

19 Frost Formation on a cooled Mirror and ensuring frost is formed is key to accuracy Clean Mirror Mirror with even formation of frost Super-cooled water can form between 0 and -40 C leading to a 1 C error for every 10 C depression

20 Condensation Hygrometer Stabilization A cooled mirror modulates the power applied to a thermoelectric cooling module based on optical feedback. Stabilization typically requires oscillations around the actual dew point. PID control algorithms are employed to provide this function

21 FAST ensure ice not water is formed Super-cooled water causes 1 C /10 C depression error

22 NPL Calibration Certificate MICHELL TECHNOLOGIES FOR MOISTURE

23 Dewpoint Calibration Systems Various designs from -100 C MICHELL TECHNOLOGIES FOR MOISTURE

24 Combinations in Dew-Point Calibration Systems -60Cdp Capability DG2 S8000 Integrale PSD2-80Cdp Capability DG4 S8000 RS Pre-Dryer -100Cdp Capability PSD4 VDS3 S8000 TRS

25 Rh Calibration Still requires the basics - Air source - Controlled moisture generation Can use chambers - Bigger the chamber the more potential errors Reference can be CM or a suitable probe - What traceability - Accuracy Trained staff and methodology

26 Dew point v %Rh accuracies MICHELL TECHNOLOGIES FOR MOISTURE

27 Full chamber system. This complete, automatic calibration system is computer controlled to provide a calibration environment for various types of humidity sensor from -20 to +50 C and 1 to 95% relative humidity (-50 to +50 C) Systems can provide profiling of both parameters as well as a comprehensive data logging capability.

28 Full chamber schematic MICHELL TECHNOLOGIES FOR MOISTURE

29 HG-1 The HG-1 is an easy to use, bench-top generator designed to give a complete solution in a single box. Comprises: pump, dryer, divided flow generator, calibration test chamber and reference dewpointmeter, with direct display of relative humidity and chamber temperature.

30 Basic Rh generator using mixed flow can work with or without a chilled mirror reference

31 Simple Rh calibration at ambient conditions

32 Principle S503 MICHELL TECHNOLOGIES FOR MOISTURE

33 Inside the S503 MICHELL TECHNOLOGIES FOR MOISTURE

34 Rh calibration S904: Allows Rh and temperature control

35 Schematic MICHELL TECHNOLOGIES FOR MOISTURE

36 Inside the chamber: Mixing fan and control sensor

37 Outside of the chamber; 4 peltier ensure temperature uniformity

38 Chamber with chilled mirror reference MICHELL TECHNOLOGIES FOR MOISTURE

39 Questions? MICHELL TECHNOLOGIES FOR MOISTURE

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014 Calibration General Principles & Theory, Equipment Considerations Metrological Principles Accurate but not Precise Calibration A A+ Correction = Precise but not Accurate Correction B Ref Std (Radio Link

More information

Humidity Calibration Solutions

Humidity Calibration Solutions Instruments Humidity Solutions www.michell.com Instruments Humidity Solutions The Importance of Regular The reliable operation of a hygrometer and indeed any measuring instrument, can only be verified

More information

Humidity Calibration Solutions

Humidity Calibration Solutions Humidity Calibration Solutions www.michell.com Humidity Calibration Solutions The Importance of Regular Calibration The reliable operation of a hygrometer and indeed any measuring instrument, can only

More information

Chilled Mirror Hygrometry

Chilled Mirror Hygrometry Chilled Mirror Hygrometry Precision instruments for critical measurement and control applications Chilled Mirror Hygrometry Wherever precise, fundamental measurements are required in your laboratory or

More information

The Fundamentals of Moisture Calibration

The Fundamentals of Moisture Calibration The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer High Performance Chilled Mirror Hygrometer With Cable Mounted Measuring Heads Highly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct

More information

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length Yun-Kyung Bae and In-Jik Jeong Industrial & Physical Instrument Center, Korea

More information

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Introduction Humidity generators are mechanical devices that provide stable temperature and humidity

More information

REALIZATION OF HUMIDITY STANDARD FACILITY USING TWO-PRESSURE HUMIDITY GENERATOR AND HUMIDITY CHAMBER

REALIZATION OF HUMIDITY STANDARD FACILITY USING TWO-PRESSURE HUMIDITY GENERATOR AND HUMIDITY CHAMBER METROLOGY AND MEASUREMENT SYSTEMS Index 3393, ISSN 86-8229 www.metrology.pg.gda.pl REALIZATION OF HUMIDITY STANDARD FACILITY USING TWO-PRESSURE HUMIDITY GENERATOR AND HUMIDITY CHAMBER Essam El-Din, Mostafa

More information

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Technical Note The Fluke Calibration 5128A RHapid-Cal Humidity Generator provides a portable, stable test environment

More information

473-SHX Dew Point Mirror

473-SHX Dew Point Mirror Humidity and Temperature Reference Hygrometer For Temperature up to 125 C Precise and stable chilled mirror dew point mirror technology High temperature optical components High temperature sample fan Cable

More information

473 Dew Point Hygrometer

473 Dew Point Hygrometer gh Performance Chilled Hygrometer With Cable Mounted Measuring Heads ghly precise chilled mirror dew point technology Cable mounted dew point and temperature measurement Aspirated and direct insertion

More information

Overview Brochure Comprehensive range of instrumentation for the measurement of moisture and oxygen

Overview Brochure Comprehensive range of instrumentation for the measurement of moisture and oxygen Overview Brochure 2016 Comprehensive range of instrumentation for the measurement of moisture and oxygen Overview Brochure Relative Humidity Probes and Transmitters Repeatable RH and temperature measurement

More information

The calibration of Hygrometer (LectureandTraining )

The calibration of Hygrometer (LectureandTraining ) The calibration of Hygrometer (LectureandTraining ) HAGIYA Satoshi 8Nov.2016 Outline 1.Measu rementmethods of midity(theory) Hu 2.Traceab ili ty in JMA 3.Cali brationmethodsof Hyg rometer 4.Practice (Ca

More information

Edgetech Instruments Inc.

Edgetech Instruments Inc. 2015_07_ETI OVERVIEW Rev 3 MADE IN THE USA Edgetech Instruments (ETI) is a USA company based in Massachusetts. We have over 50 years of expertise in the design and manufacture of a comprehensive range

More information

Fundamentals of Moisture Measurement in Industrial Applications

Fundamentals of Moisture Measurement in Industrial Applications 2016 Fundamentals of Moisture Measurement in Industrial Applications Moisture is not only a process variable to be measured in industrial applications, but also a common contaminant which must be controlled.

More information

Science of Stability 2017

Science of Stability 2017 Science of Stability 2017 Humidity Measurement in Stability Testing Applications Defining Uncertainty and the Challenges at Higher Temperatures Humidity Measurement In Stability Testing Applications Some

More information

D501 and D501-RS. Microscanner D-Series IR Thermometers. no effect. no effect. no effect. no effect no effect

D501 and D501-RS. Microscanner D-Series IR Thermometers. no effect. no effect. no effect. no effect no effect Microscanner D-Series The Only Certified Accurate Surface Instruments in the World A MUST FOR ISO 9001 ISO 9002 ISO 9003 TRACEABILITY PROGRAMS Common Surface Measurement Errors D501 and D501-RS Microscanner

More information

Techniques for Fast Measurements of Low Dewpoint in Portable Hygrometers

Techniques for Fast Measurements of Low Dewpoint in Portable Hygrometers Techniques for Fast Measurements of Low Dewpoint in Portable Hygrometers Speaker / Author: Nick Malby, Michell Instruments Ltd, 48 Lancaster Way Business Park, Ely, Cambridgeshire. CB6 3NW. UK. +44 1353

More information

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Quality assurance for sensors at the Deutscher Wetterdienst (DWD) Paper submitted to ICAWS 2017: Topic 3 Sustainability of the measurements: Calibration, intercomparisons, laboratory and field performance tests, quality assurance and control assessment for traceable

More information

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Quality assurance for sensors at the Deutscher Wetterdienst (DWD) Quality assurance for sensors at the Deutscher Wetterdienst (DWD) Quality assurance / maintenance / calibration Holger Dörschel, Dr Tilman Holfelder WMO International Conference on Automatic Weather Stations

More information

STATUS OF THE WIGOS DEMONSTRATION PROJECTS

STATUS OF THE WIGOS DEMONSTRATION PROJECTS STATUS OF THE WIGOS DEMONSTRATION PROJECTS Demonstration Project Morocco Strengthening Moroccan RIC Capacities (Submitted by Rabia Merrouchi, National Meteorological Service of Morocco (DMN)) February

More information

Infrared Thermometer Calibration A Complete Guide

Infrared Thermometer Calibration A Complete Guide Infrared Thermometer Calibration A Complete Guide Application note With proper setup and planning, infrared calibrations can be accurate. The steps outlined below should be followed to perform accurate

More information

Many short reports are inadequate and need more work

Many short reports are inadequate and need more work University of Maryland Baltimore County - UMBC Phys650 - Special Topics in Experimental Atmospheric Physics (Spring 2009) J. V. Martins and M. H. Tabacniks Grading feedback: http://userpages.umbc.edu/~martins/phys650/

More information

CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR

CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR Davor Zvizdic, Tomislav Stasic, Lovorka

More information

Uncertainty in Humidity Measurement. Helping you make a better measurement.

Uncertainty in Humidity Measurement. Helping you make a better measurement. Uncertainty in Humidity Measurement Helping you make a better measurement. Webinar Presenters Bruce McDuffee Michael Boetzkes Agenda Terms & Definitions Understand the Uncertainty of Your Instrument Understanding

More information

Measuring Humidity in the Charters of Freedom Encasements Using a Moisture Condensation Method

Measuring Humidity in the Charters of Freedom Encasements Using a Moisture Condensation Method Measuring Humidity in the Charters of Freedom Encasements Using a Moisture Condensation Method Speaker: Cecil G. Burkett NASA Langley Research Center Mail Stop 236 Hampton VA 23681 PH: 757-864-4720 FAX:

More information

Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer

Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer Temperature Control without Compromise: rapid changes, precise temperatures, powerful complex functionality.

More information

Introduction to Blackbody Sources

Introduction to Blackbody Sources Introduction to s This section contains dedicated blackbody sources for low uncertainty calibration of infrared thermometers. A range of portable primary blackbody sources combine high emissivity with

More information

SERIES 4000 PRECISION DEWPOINT HYGROMETER Integrale INSTALLATION, OPERATION AND MAINTENANCE MANUAL

SERIES 4000 PRECISION DEWPOINT HYGROMETER Integrale INSTALLATION, OPERATION AND MAINTENANCE MANUAL SERIES 4000 PRECISION DEWPOINT HYGROMETER Integrale INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue July 2001 Page 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 4 1.1 General 4 1.2 Operating Principle

More information

Evaluation of Vaisala HMP45D Humidity Probes

Evaluation of Vaisala HMP45D Humidity Probes Instrument Test Report: 661 Evaluation of Vaisala HMP45D Humidity Probes J.Gorman Regional Instrument Centre 2 nd Jan 2002 Authorisation Bruce Forgan SRLR Regional Instrument Centre Distribution STNM STIE

More information

Dewpoint Meter Data Sheet / Instructions

Dewpoint Meter Data Sheet / Instructions Paint Test Equipment Dewpoint Meter Data Sheet / Instructions Dewpoint Meter Information ISO 8502-4: Preparation of steel substrates before application of paints and related products. Tests for the assessment

More information

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years Int J Thermophys (2017) 38:161 DOI 10.1007/s10765-017-2291-x TEMPMEKO 2016 Development of the High-Temperature Dew-Point Generator Over the Past 15 Years R. Bosma 1 J. Nielsen 2 A. Peruzzi 1 Received:

More information

Data Sheet / Instructions. Dewpoint MeterO

Data Sheet / Instructions. Dewpoint MeterO Data Sheet / Instructions Dewpoint MeterO Dewpoint Meter ISO 8502-4: Preparation of steel substrates before application of paints and related products. Tests for the assessment of surface cleanliness.

More information

SSi Super Systems Inc.

SSi Super Systems Inc. Model DPC2530 Continuous Digital Dew Point Analyzer For measurement of: Endothermic Atmosphere Exothermic Atmosphere Nitrogen / Hydrogen Atmosphere Plant Air Systems SPECIFICATIONS Measurement Range: -50

More information

ALMA Weather Instrumentation Specification

ALMA Weather Instrumentation Specification ALMA-90.03.00.00-00x-A-SPE 2006 12 01 Specification Document Jeff Mangum Page 2 Change Record Revision Date Author Section/ Remarks Page affected 1 2005-08-12 Jeff Mangum All Initial Draft 2 2006-08-30

More information

Measuring State Parameters of the Atmosphere

Measuring State Parameters of the Atmosphere Measuring State Parameters of the Atmosphere Some Applications of Atmospheric Thermodynamics Earth Observing Laboratory, NCAR IDEAS-4 Tutorial Introduction Goals of This Presentation Present two complementary

More information

DESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR

DESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR Copyright 2013 by ABCM DESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR Amaury Roteger de Oliveira Renato Nunes Teixeira Instituto Nacional de Metrologia, Qualidade e Tecnologia - Inmetro Diretoria

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 CONTROL COMPANY 12554 Galveston Road Suite B230 Webster, TX 77598 Aaron Judice Phone: 281 482 1714 CALIBRATION Valid To: March 31, 2018

More information

Module 2. Measurement Systems. Version 2 EE IIT, Kharagpur 1

Module 2. Measurement Systems. Version 2 EE IIT, Kharagpur 1 Module 2 Measurement Systems Version 2 EE IIT, Kharagpur 1 Lesson 8 Measurement of Level, Humidity and ph Version 2 EE IIT, Kharagpur 2 Instructional Objectives At the end of this lesson, the student will

More information

CALIBRATION CERTIFICATE # 503

CALIBRATION CERTIFICATE # 503 CALIBRATION CERTIFICATE # 503 Calibration date : 2011-02-23 Certificate issued : 2012-04-18 Company name Company address City, Province, Canada CLAS 2009-02 Calibration of Mass flow meter Micro Motion

More information

New Peltier-based chilled-mirror hygrometer SKYDEW

New Peltier-based chilled-mirror hygrometer SKYDEW GRUAN ICM-10, Potsdam, April 25, 2018 New Peltier-based chilled-mirror hygrometer SKYDEW Takuji Sugidachi, K. Shimizu (Meisei Electric Co. LTD., Japan) M. Fujiwara (Hokkaido University, Japan) Motivation

More information

Calibration of temperature sensors within Length Standard Section of NMIJ

Calibration of temperature sensors within Length Standard Section of NMIJ Calibration of temperature sensors within Length Standard Section of NMIJ Dr. Akiko Hirai Length Standard Section National Metrology Institute of Japan (NMIJ) 1 Contents Temperature measurement system

More information

Water Vapour Effects in Mass Measurement

Water Vapour Effects in Mass Measurement Water Vapour Effects in Mass Measurement N.-E. Khélifa To cite this version: N.-E. Khélifa. Water Vapour Effects in Mass Measurement. Measurement. Water Vapour Effects in Mass Measurement, May 2007, Smolenice,

More information

Comparison of HCD5000 DewPoint Duo to Condumax II

Comparison of HCD5000 DewPoint Duo to Condumax II EXECUTIVE SUMMARY This document compares the technology and products of ZEGAZ Instruments (US) to that of Michell Instruments (UK) in the field of hydrocarbon and water dewpoint measurement. INTRODUCTION

More information

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements Experimental Design 8.2.2013 Doc. Martti Heinonen etunimi.sukunimi@mikes.fi Outline 1.

More information

CONTENT 2. ORGANIZATION 3. SERVICES. instruments. (3)Activities of RIC Tsukuba

CONTENT 2. ORGANIZATION 3. SERVICES. instruments. (3)Activities of RIC Tsukuba 1 CONTENT 1. INTRODUCTION 2. ORGANIZATION 3. SERVICES (1)Quality assurance of meteorological instruments (2)Research and development (3)Activities of RIC Tsukuba 2 Meteorological Instruments Center (MIC)

More information

Thermodynamics Introduction and Basic Concepts

Thermodynamics Introduction and Basic Concepts Thermodynamics Introduction and Basic Concepts by Asst. Prof. Channarong Asavatesanupap Mechanical Engineering Department Faculty of Engineering Thammasat University 2 What is Thermodynamics? Thermodynamics

More information

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE Michael Bair and Pierre Delajoud 2002 DEC 10 Revised 2004 MAR 26 FORWARD FPG8601 is a pressure standard designed

More information

3URGXFWLQIRUPDWLRQ 5HVHUYDWLRQRISURSHUW\ULJKWV

3URGXFWLQIRUPDWLRQ 5HVHUYDWLRQRISURSHUW\ULJKWV 3URGXFWLQIRUPDWLRQ +\URSKLO Š '7W\SH 5HVHUYDWLRQRISURSHUW\ULKWV 7KHFRQWHQWRIWKLVGRFXPHQWLVWKHLQWHOOHFWXDOSURSHUW\RI%$57(&DQGPXVWQHLWKHUEHFRSLHGQRUPDGHDYDLO DEOHWRWKLUGSDUWLHVQRWHYHQLQH[WUDFWVZLWKRXWRXUH[SUHVVDSSURYDOLQZULWLQ$Q\LQIULQHPHQWZLOOUHVXOWLQD

More information

COMPARISON OF HUMIDITY MEASUREMENTS USING A DEW POINT METER AS A TRANSFER STANDARD APMP-IC-1-97 REPORT

COMPARISON OF HUMIDITY MEASUREMENTS USING A DEW POINT METER AS A TRANSFER STANDARD APMP-IC-1-97 REPORT COMPARISON OF HUMIDITY MEASUREMENTS USING A DEW POINT METER AS A TRANSFER STANDARD APMP-IC-1-97 REPORT BY NATIONAL METROLOGY CENTRE (A DIVISION OF SPRING SINGAPORE) COORDINATOR OF THE COMPARISON September

More information

Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C

Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C DOI 10.1007/s10765-014-1687-0 Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C Domen Hudoklin Regina Mnguni Hans Liedberg Igor Pušnik Jovan Bojkovski

More information

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

2.1. Accuracy, n- how close the indication of the thermometer is to the true value. AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria

More information

Process Analyzer and Sampling Systems

Process Analyzer and Sampling Systems Process Analyzer and Sampling Systems Course Price 3050 Course Description Analytical instruments used for online chemical analysis of process streams or plant environments are generally called Process

More information

LAB007 Industrial Analytical Chemistry & Process Analyzer System

LAB007 Industrial Analytical Chemistry & Process Analyzer System LAB007 Industrial Analytical Chemistry & Process Analyzer System H.H. Sheikh Sultan Tower (0) Floor Corniche Street Abu Dhabi U.A.E www.ictd.ae ictd@ictd.ae Course Introduction: Analytical instruments

More information

Measuring State Parameters of the Atmosphere

Measuring State Parameters of the Atmosphere Measuring State Parameters of the Atmosphere Some Applications of Atmospheric Thermodynamics Earth Observing Laboratory, NCAR IDEAS-4 Tutorial Introduction Goals of This Presentation Present two complementary

More information

AL108 Process Analyzers & Sampling Systems

AL108 Process Analyzers & Sampling Systems AL108 Process Analyzers & Sampling Systems AL108 Rev.002 CMCT COURSE OUTLINE Page 1 of 8 Training Description: Analytical instruments used for online chemical analysis of process streams or plant environments

More information

Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000

Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000 Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000 Welcome to the Isotech Journal of Thermometry Fundamentals of Thermometry Part 1 Temperature Scales Practical Calibration of Thermometers

More information

CERTIFICATE OF CALIBRATION

CERTIFICATE OF CALIBRATION Page 1 of 5 CERTIFICATE OF CALIBRATION 1K Thermistor CERTIFICATE NO.: MANUFACTURER: Measurement Specialties DATE OF CALIBRATION: 26-Apr-17 TC15249 CUSTOMER: AMBIENT TEMPERATURE: (21 ± 3) C JOB NUMBER:

More information

Heat stress index. Dewpoint. pressure. altitude

Heat stress index. Dewpoint. pressure. altitude Kestrel 4200 Measures Volume air flow in cfm, L/s, m³/s, m³/m, m³/h Air velocity (current, average, and maximum) Heat stress index Windchill Relative humidity Dewpoint in F or C Humidity ratio Barometric

More information

Towards SI traceable humidity measurements with radiosondes

Towards SI traceable humidity measurements with radiosondes VTT TECHNICAL RESEARCH CENTRE OF FINLAND LTD Centre for Metrology MIKES MIKES METROLOGY Towards SI traceable humidity measurements with radiosondes GRUAN ICM-9, MeteoMet Training Course FMI, Helsinki,

More information

e-newsletter n 6 August 31 st, 2018 New calibration facility developed at PTB for relative humidity, associated to dtdlas transfer standard

e-newsletter n 6 August 31 st, 2018 New calibration facility developed at PTB for relative humidity, associated to dtdlas transfer standard e-newsletter n 6 August 31 st, 2018 New calibration facility developed at PTB for relative humidity, associated to dtdlas transfer standard PTB has developed a test facility to characterize humidity sensors

More information

Measurement Uncertainty of Dew-Point Temperature in a Two-Pressure Humidity Generator

Measurement Uncertainty of Dew-Point Temperature in a Two-Pressure Humidity Generator Int J Thermophys (2012) 33:1568 1582 DOI 10.1007/s10765-011-1005-z Measurement Uncertainty of Dew-Point Temperature in a Two-Pressure Humidity Generator L. Lages Martins A. Silva Ribeiro J. Alves e Sousa

More information

DeFelsko PosiTector Dew Point Meter with Anemometer (DPMA)

DeFelsko PosiTector Dew Point Meter with Anemometer (DPMA) Management Procedure 2562 Revision: A Date Issued: March 24, 2017 Date Revised: Calibration Procedure DeFelsko PosiTector Dew Point Meter with Anemometer (DPMA) Table of Contents 1. Introduction and UUC

More information

Performance of the Meteolabor Snow White chilled-mirror hygrometer. Masatomo Fujiwara Hokkaido University, Japan

Performance of the Meteolabor Snow White chilled-mirror hygrometer. Masatomo Fujiwara Hokkaido University, Japan Performance of the Meteolabor Snow White chilled-mirror hygrometer Masatomo Fujiwara Hokkaido University, Japan Coauthors and Collaborators F. Hasebe, M. Shiotani, H. Voemel, S. J. Oltmans, H. Takashima,

More information

Atmosphere Properties and Instruments. Outline. AT351 Lab 2 January 30th, 2008

Atmosphere Properties and Instruments. Outline. AT351 Lab 2 January 30th, 2008 Atmosphere Properties and Instruments AT351 Lab 2 January 30th, 2008 Outline 1. Atmospheric Variables and How We Measure Them 2. Composition of the Atmosphere 3. How to Represent Weather Data Visually

More information

COMPARISON OF INMETRO AND INTI HUMIDITY STANDARDS. Final Report

COMPARISON OF INMETRO AND INTI HUMIDITY STANDARDS. Final Report SIM.T-K6.4 INMETRO / INTI Final Report COMPARISON OF INMETRO AND INTI HUMIDITY STANDARDS Final Report Instituto Nacional de Metrologia, Qualidade e Tecnologia (INMETRO) Instituto Nacional de Tecnología

More information

Metrology and Environment

Metrology and Environment Bridging the Gap Responding to Environmental Change: from Words to Deeds 14-16 May 2008, Portorož Metrology and Environment Prof Janko Drnovšek and Laboratory group Janko.Drnovsek@fe.uni-lj.si Basic dilemma

More information

Certificate of Accreditation to ISO/IEC 17025:2005

Certificate of Accreditation to ISO/IEC 17025:2005 United States Department of Commerce National Institute of Standards and Technology Certificate of Accreditation to ISO/IEC 17025:2005 NVLAP LAB CODE: 200508-0 Insco Metrology, Inc. Miami, FL is accredited

More information

CREATION AND DEVELOPMENT OF THE NATIONAL HUMIDITY STANDARD OF GASES

CREATION AND DEVELOPMENT OF THE NATIONAL HUMIDITY STANDARD OF GASES CREATION AND DEVELOPMENT OF THE NATIONAL HUMIDITY STANDARD OF GASES Efimia LUCHIAN, Constantin BORDIANU, Grigore BUZUC, Victoria CEBAN National Institute of Metrology, Chisinau, Republic of Molova tel.:

More information

The Vaisala Reference Radiosonde Program: First Results and Future Plans

The Vaisala Reference Radiosonde Program: First Results and Future Plans The Vaisala Reference Radiosonde Program: First Results and Future Plans Presentation in the NMM 2010, Helsinki, Finland, 9 June2010 Heikki Turtiainen, Hans von Bagh, Hannu Jauhiainen, Terhi Lehtola, Johanna

More information

Psychrometer construction for performance testing in. temperature and humidity chambers, and the precision

Psychrometer construction for performance testing in. temperature and humidity chambers, and the precision Technology report Psychrometer construction for performance testing in temperature and humidity chambers, and the precision of humidity measurements Meeting the challenge of quality engineering Hirokazu

More information

MEASUREMENT OF WBGT INDEX IN AXIAL FLOW AS THE MAIN THERMODYNAMIC PARAMETERS VARY G.

MEASUREMENT OF WBGT INDEX IN AXIAL FLOW AS THE MAIN THERMODYNAMIC PARAMETERS VARY G. MEASUREMENT OF WBGT INDEX IN AXIAL FLOW AS THE MAIN THERMODYNAMIC PARAMETERS VARY G. Buonanno, M. Dell'Isola, A. Frattolillo and L. Vanoli Di.M.S.A.T. - Università di Cassino via Di Biasio 43-03043 Cassino

More information

Kestrel 4500 Pocket Weather Tracker

Kestrel 4500 Pocket Weather Tracker Kestrel 4500 Pocket Weather Tracker For years our customers have been asking for wind direction along with wind speed. New for 2007, the Kestrel 4500 does just that with its built in digital compass. But

More information

Typical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge

Typical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge Typical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge 2090TN05 Revised 2004 March 26 Technical Note FPG8601 is a pressure standard designed to cover the range from

More information

A Case for Periodic Calibration or Verification of RTDs

A Case for Periodic Calibration or Verification of RTDs A Case for Periodic Calibration or Verification of RTDs If you are experiencing audio problems please call the teleconference number below Phone 650-479-3208 Access code 668 676 434 Your Host and Presenter

More information

A calibration facility for automatic weather stations

A calibration facility for automatic weather stations METEOROLOGICAL APPLICATIONS Meteorol. Appl. 22: 842 846 (2015) Published online 16 November 2015 in Wiley Online Library (wileyonlinelibrary.com) DOI: 10.1002/met.1514 A calibration facility for automatic

More information

Evaluating New Generation Vibrating Tube Sensor for Density Measurement under Process Conditions

Evaluating New Generation Vibrating Tube Sensor for Density Measurement under Process Conditions Products Solutions Services Evaluating New Generation Vibrating Tube Sensor for Density Measurement under Process Conditions Dr. Alfred Rieder, Slide 1 Overview Introduction Process meter Features Working

More information

Welcome. Vaisala Industrial Measure Webinar Series - Humidity Theory, Terms & Definitions. Yumi Alanoly Vaisala Application Engineer

Welcome. Vaisala Industrial Measure Webinar Series - Humidity Theory, Terms & Definitions. Yumi Alanoly Vaisala Application Engineer Welcome Vaisala Industrial Measure Webinar Series - Humidity Theory, Terms & Definitions Yumi Alanoly Vaisala Application Engineer Agenda 1. Why do we measure humidity? 2. Dalton s Law 3. Relative humidity

More information

Elcometer 116C. Sling Hygrometer. Operating Instructions

Elcometer 116C. Sling Hygrometer. Operating Instructions Elcometer 116C English Sling Hygrometer Operating Instructions English is a registered trademark of Elcometer Limited. All other trademarks acknowledged. Copyright Elcometer Limited. 2008. All rights reserved.

More information

Laird Thermal Systems Application Note. Precise Temperature Control for Medical Analytical, Diagnostic and Instrumentation Equipment

Laird Thermal Systems Application Note. Precise Temperature Control for Medical Analytical, Diagnostic and Instrumentation Equipment Laird Thermal Systems Application Note Precise Temperature Control for Medical Analytical, Diagnostic and Instrumentation Equipment Contents Overview...3 Thermal Management Techniques...3 Application Examples...3

More information

Vaisala DRYCAP Hand-Held Dewpoint Meter DM70. Fast Response - Low Maintenance

Vaisala DRYCAP Hand-Held Dewpoint Meter DM70. Fast Response - Low Maintenance Vaisala DRYCAP Hand-Held Dewpoint Meter DM70 Fast Response - Low Maintenance DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications The Vaisala DRYCAP Hand- Held Dewpoint Meter DM70 offers accurate

More information

OverviewofRIC Tsukuba (Japan,RAII )

OverviewofRIC Tsukuba (Japan,RAII ) OverviewofRIC Tsukuba (Japan,RAII ) 7Nov.2016 Hiroshi Yoshimura Scientific Officer Regional Instrument Centre Tsukuba Observing Division, Observing Department Outline 1. WMO Regional Instrument Center

More information

Document No: TR 12 Issue No: 1

Document No: TR 12 Issue No: 1 ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Prepared by: SADCAS Technical Manage Approved

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ARCONIC POWER AND PROPULSION RESEARCH CENTER 1500 S. Warner St. Whitehall, MI 49461-1895 Michelle McDonald Phone: 231 894 7586 CALIBRATION

More information

DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications

DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications The Vaisala DRYCAP Hand-Held Dewpoint Meter DM70 offers accurate and fast measurement for industrial dewpoint applications, such as compressed

More information

D501 and D501-RS. Figure 1. Unique Automatic Emissivity Compensation System (AECS) produces accurate

D501 and D501-RS. Figure 1. Unique Automatic Emissivity Compensation System (AECS) produces accurate Microscanner D-Series The Only Certified Accurate Surface Temperature Instrument in the World A MUST FOR ISO 9001 ISO 9002 ISO 9003 TRACEABILITY PROGRAMS D501 and D501-RS Figure 1. Unique Automatic Emissivity

More information

Introduction of RIC Tsukuba (Japan, RAII)

Introduction of RIC Tsukuba (Japan, RAII) Introduction of RIC Tsukuba (Japan, RAII) 20 March, 2018 Kouichi NAKASHIMA Scientific Officer Regional Instrument Centre Tsukuba Observing Division, Observing Department Outline 1. Overview of Organization

More information

Loadcell Calibration - Evaluation of Uncertainties

Loadcell Calibration - Evaluation of Uncertainties Loadcell Calibration - Evaluation of Uncertainties This document is a revision of the evaluation issued in June 2015. Novatech s calibration laboratory is not UKAS accredited due to the high number of

More information

Can a -20 C reference point be established in a laboratory?

Can a -20 C reference point be established in a laboratory? Can a -20 C reference point be established in a laboratory? Speaker / Author: Yvette Volschenk Company: AFRIQLABS Centurion South Africa Cell: 083 658 8373 E-mail: yvette7137@iburst.co.za Abstract: When

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 FLORIDA METROLOGY, LLC 645 NW Enterprise Dr. Port St. Lucie FL 34986 Ron Kupper Phone: (772)212-7158 CALIBRATION Valid To: October 31,

More information

SHEBA GLASS SOUNDING DATA

SHEBA GLASS SOUNDING DATA SHEBA GLASS SOUNDING DATA Erik R. Miller and Kathryn Beierle - 12 June 2000 NCAR ATD Considerable speculation has been brought to bear as to whether there is a dry bias in the SHEBA radiosonde data. Two

More information

LABORATORY EXPERIMENTS OF MIXED- PHASE CLOUD FORMATION

LABORATORY EXPERIMENTS OF MIXED- PHASE CLOUD FORMATION LABORATORY EXPERIMENTS OF MIXED- PHASE CLOUD FORMATION Takuya Tajiri 1*, Katsuya Yamashita 1, Masataka Murakami 1, Narihiro Orikasa 1, Atsushi Saito 1, Tomohiro Nagai 1, Tetsu Sakai 1, and Hiroshi Ishimoto

More information

SIMPLE DEW DIGITAL DEW POINT ANALYZER

SIMPLE DEW DIGITAL DEW POINT ANALYZER SIMPLE DEW DIGITAL DEW POINT ANALYZER OPERATIONS MANUAL Super Systems Inc. 7205 Edington Drive Cincinnati, OH 45249 513-772-0060 Fax: 513-772-9466 www.supersystems.com Super Systems Inc. USA Office Corporate

More information

COPLEY S C I E N T I F I C. A multi-function aerosol system with aerosol generation, classification and monitoring capabilities for:

COPLEY S C I E N T I F I C. A multi-function aerosol system with aerosol generation, classification and monitoring capabilities for: A multi-function aerosol system with aerosol generation, classification and monitoring capabilities for: generating monodisperse aerosol by mobility classification with automatic concentration detection

More information

: INSTRUMENTATION AND PROCESS CONTROL COURSE CODE : 6071 COURSE CATEGORY : A PERIODS/ WEEK : 5 PERIODS/ SEMESTER : 75 CREDIT : 5 TIME SCHEDULE

: INSTRUMENTATION AND PROCESS CONTROL COURSE CODE : 6071 COURSE CATEGORY : A PERIODS/ WEEK : 5 PERIODS/ SEMESTER : 75 CREDIT : 5 TIME SCHEDULE COURSE TITLE : INSTRUMENTATION AND PROCESS CONTROL COURSE CODE : 6071 COURSE CATEGORY : A PERIODS/ WEEK : 5 PERIODS/ SEMESTER : 75 CREDIT : 5 TIME SCHEDULE MODULE TOPIC PERIODS 1 Measuring Instruments

More information

DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications

DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications The Vaisala DRYCAP Hand-Held Dewpoint Meter DM70 offers accurate and fast measurement for industrial dewpoint applications, such as compressed

More information

Hygrostat based on adsorption processes controlled by a high precision chilled dew point mirror

Hygrostat based on adsorption processes controlled by a high precision chilled dew point mirror URN (Paper): urn:nbn:de:gbv:ilm1-2014iwk-122:7 58 th ILMENAU SCIENTIFIC COLLOQUIUM Technische Universität Ilmenau, 08 12 September 2014 URN: urn:nbn:de:gbv:ilm1-2014iwk:3 Hygrostat based on adsorption

More information

THM80X Series Feature Application. Temperature & Humidity / Flow Measuring Specialist

THM80X Series  Feature Application. Temperature & Humidity / Flow Measuring Specialist THM80X Series Feature Application IP67 protection degree, rugged aluminum case, fit in variety harsh environment Capable of temperature compensation Linear adjustment temperature & humidity by computer,

More information

Specific Accreditation Guidance. Infrastructure and Asset Integrity. Measurement Uncertainty in Geotechnical Testing

Specific Accreditation Guidance. Infrastructure and Asset Integrity. Measurement Uncertainty in Geotechnical Testing Specific Accreditation Guidance Infrastructure and Asset Integrity Measurement Uncertainty in Geotechnical Testing January 2018 Copyright National Association of Testing Authorities, Australia 2018 This

More information