Analysis of Image Noise in 3D Cone-Beam CT: Spatial and Fourier Domain Approaches under Conditions of Varying Stationarity

Size: px
Start display at page:

Download "Analysis of Image Noise in 3D Cone-Beam CT: Spatial and Fourier Domain Approaches under Conditions of Varying Stationarity"

Transcription

1 Analysis of Image Noise in 3D Cone-Beam CT: Spatial and Fourier Domain Approaches under Conditions of Varying Stationarity Angel R. Pineda *a, Jeffrey H. Siewerdsen b,c, Daniel J. Tward b a Department of Mathematics, California State University, Fullerton CA, USA 9834 b Ontario Cancer Institute, Princess Margaret Hospital, Toronto ON, Canada M5G M9 c Department of Medical Biophysics, University of Toronto, Toronto ON, Canada M5G M9 ABSTRACT The statistical properties of medical images are central in characterizing the performance of imaging systems. The noise in cone-beam CT (CBCT) is often characterized using Fourier-based metrics, such as the 3D noise-power spectrum (NPS). Under a stationarity assumption, the NPS provides a complete representation of the covariance of the images, since the covariance matrix of the Fourier transform of the image is diagonal. In practice, such assumptions are obeyed to varying degrees. The objective of this work is to investigate the degree to which such assumptions apply in CBCT and to experimentally characterize the NPS and off-diagonal elements under a range of experimental conditions. A benchtop CBCT system was used to acquire 3D image reconstructions of various objects (air and a water cylinder) across a range of experimental conditions that could affect stationarity (bowtie filter and dose). We test the stationarity assumption under such varying experimental conditions using both spatial and frequency domain measures of stationarity. The results indicate that experimental conditions affect the degree of stationarity and that under some imaging conditions, local descriptions of the noise need to be developed to appropriately describe CBCT images. The off-diagonal elements of the DFT covariance matrix may not always be ignored. Keywords: noise-power spectrum, image quality, covariance matrix, stationarity, image noise, cone-beam CT, 3D imaging, flat-panel detector 1. INTRODUCTION Evaluating the performance of imaging systems should take into account three elements: the task, the statistics, and the observer [1-4]. The task refers to the intended use of the images, such as the detection of a lesion or the estimation of the size of a vessel. The statistics characterize the sources of variation in the images. The observer deals with how to extract information from the data and may be a human or a computer. Each of these components is critical in evaluating the usefulness of a given imaging system. Noise in computed tomography (CT) is often characterized using measurements of the noise-power spectrum (NPS) by implicitly assuming that the noise is stationary or at least wide-sense stationary [4-7]. Since stationarity implies that the correlations of the image are the same for all the locations in the 3D volume, both the experimental procedure and the analysis of the noise are greatly simplified, because the NPS gives a complete description of the covariance of the noise i.e., the covariance matrix is diagonal, and the diagonal is the NPS. The stationarity assumption arises naturally from an ideal and continuous description of CT but may not be rigorously justified for real experimental data. In this paper, we experimentally investigate the validity of the stationarity assumption in characterizing the statistics of images produced in cone-beam computed tomography (CBCT). The purpose is to understand to what degree the NPS fully characterizes the magnitude and correlation of fluctuations in the image data as we change the experimental conditions. There has been an active discussion in the research community of the non-stationarity nature of medical images [8-9]. For this study, we test the extent to which we can control the stationarity in the images by changing the experimental imaging conditions. We consider both spatial and frequency domain characterizations of the noise to better understand the degree to which the stationarity assumption holds in real CBCT images. The degree of applicability of the stationarity assumption is an initial step in understanding the validity of summarizing image quality and object detectability using the NPS and related Fourier-based metrics, such as the noise-equivalent quanta (NEQ). * apineda@fullerton.edu; phone Medical Imaging 8: Physics of Medical Imaging, edited by Jiang Hsieh, Ehsan Samei, Proc. of SPIE Vol. 6913, 69131Q, (8) /8/$18 doi: / Proc. of SPIE Vol Q-1 Downloaded from SPIE Digital Library on 13 Nov 9 to Terms of Use:

2 . THEORY The evaluation of noise in CBCT images based on the NPS assumes that the images are stationary. While a finite and discrete image cannot be rigorously stationary, this approximation can be useful to different degrees depending on how well the NPS characterizes the covariance of the images in the Fourier domain. For a wide-sense stationary continuous random process, the Fourier transform diagonalizes the autocovariance function [4]. The discrete version of this property is that the image noise in the DFT space would have a diagonal covariance matrix with the NPS along the diagonal. There are different methods to estimate digital image noise-power spectrum. The technique used below is the directdigital method, which considers the ensemble average of the Fourier transform of zero-mean ( noise-only ) data which is equivalent to averaging periodograms: dxdydz NPS( fn) = DFT ( I) NPS() f = FLLL {( a x )} x y z n subvolumes where DFT(I) is the Discrete Fourier Transform of zero-mean (de-trended) images (I), f n is a 3-D spatial frequency, dx, dy, and dz are the dimensions of the reconstructed voxels, and L x, L y, and L z are the dimensions of the subvolumes used to estimate the NPS. (1) The elements of the full covariance DFT covariance matrix can be computed similarly as follows: dxdydz K = DFT I DFT I * ( ())( () ) nm n m LLL x y z subvolumes () Note that the diagonal of this matrix (for which n = m) is simply the NPS, illustrating the interpretation of the NPS as the variance of the Fourier components of the noise. In a stationary ensemble of images, the off-diagonal elements of this matrix are zero. In experiments described below, we test how well the DFT diagonalizes the covariance matrix of the images by computing the off-diagonal elements and comparing them to the NPS. Task-based evaluation of imaging systems using the Hotelling observer either in the frequency or spatial domain requires a description of the covariance of the data. The covariance matrix enters the computation of the detectability of a signal through the inverse of the covariance matrix. In the DFT space and if the DFT covariance matrix is diagonal then this would be simply related to 1/NPS. In the general case where the DFT covariance is not diagonal, the offdiagonal elements enter the detectability estimate in a complicated manner depending on the object being imaged. Future work in this project is to map the effect of the off-diagonal elements in computations of detectability. 3. EXPERIMENTAL METHODS 3.1 Experimental Setup The imaging bench of Fig. 1 was used in all experiments below. The bench incorporates an x-ray tube, a flat-panel detector (FPD), and motion control system (translation and rotation axes) for CBCT imaging under precise, reproducible conditions. The imaging components, system geometry, acquisition parameters, and reconstruction parameters are summarized in Table I. The FPD incorporated a scintillator (5 mg/cm CsI:Tl x-ray converter) and a (41 41 cm ) active matrix of a-si:h photodiodes and thin-film transistors at 4 µm pixel pitch (8% fill factor). The geometry approximated that common in CBCT-guided radiation therapy viz., a geometric magnification of ~1.54, giving a ~(5.6 x 5.6 x 5.6) cm 3 volumetric field of view. For each CBCT scan, 3 projections were acquired from a circular orbit over 36 o. The Feldkamp algorithm for 3D filtered backprojection was used to reconstruct volumetric images at (.5 x.5 x.5) mm 3 voxel size, equal to the detector pixel pitch divided by the magnification. The reconstruction filter consisted of a ramp and apodization window (equivalent to a Hann filter). Proc. of SPIE Vol Q- Downloaded from SPIE Digital Library on 13 Nov 9 to Terms of Use:

3 Figure 1. Experimental setup for cone-beam CT imaging. The x-ray tube, object rotation, and detector are synchronized within a precisely controlled geometry, with images reconstructed using a modified Feldkamp algorithm. A variety of objects (illustrated in blue) were considered, including air and various water cylinders. Images were acquired with and without a bowtie filter and as a function of dose to examine factors that affect the degree of noise stationarity. Experimental Methods Imaging Components Image Acquisition Parameters X-Ray Tube W target /.4 mm FS kvp 1 kvp Generator 8 kw const. Added Filtration Variable (below) potential Flat-Panel Detector RID-164 mas / Projection Variable (below) X-ray Converter 5 mg/cm CsI:Tl Frame Rate 1 fps Pixel Pitch.4 mm Number of Projections 3 Format 14 x 14 Orbital Extent (Angular Increment) 36 o (1.15 o ) Projection FOV 41 x 41 cm Antiscatter Grid None Bowtie Filter Optional Imaging Geometry Reconstruction Parameters Source-to-Axis Distance (SAD) 93.5 cm Filter Hann Source-to-Detector Distance 144. cm Voxel Size.5 mm isotropic (SDD) Geometric Magnification 1.54 Volume Image Format (14 x 14) axial < 14 longitudinal Volumetric FOV up to 5.6 cm Imaging Phantom Configurations Phantom Added Filtration Radiation Dose Level (mas / Bowtie Filter projection) Air 5.1 mm Cu (.4, 1.,., and 4.) mas / projection No cm Diameter.1 mm Cu (.4,.8, 1.6, and 3.) mas / projection No Water Cylinder cm Diameter Water Cylinder.1 mm Cu (.4,.8, 1.6, and 3.) mas / projection Yes Table I. Summary of experimental parameters and imaging phantom configurations. A variety of experimental conditions were tested to investigate the influence on the NPS and DFT covariance matrix. The scope of experimental parameters that could potentially affect the degree of stationarity in CBCT is vast, including: geometry, scatter conditions (grid, air gap, and longitudinal FOV), detector type, pixel binning, reconstruction filters, voxel size, radiation dose level, use of a bowtie filter, and phantom configuration. In the present study, we investigated the last 3, giving the phantom configurations described in Table I: Air (no bowtie); Water Cylinder (with bowtie); and Water Cylinder (without bowtie), each imaged at 4 dose levels roughly spanning the sensitive range of the detector (bare-beam signal maintained below sensor saturation). The 1 resulting data sets formed the basis for experimental analysis of 3D NPS and DFT covariance matrix, described below. Proc. of SPIE Vol Q-3 Downloaded from SPIE Digital Library on 13 Nov 9 to Terms of Use:

4 The bowtie filter was an Elekta bowtie filter cassette (F , Elekta Oncology Systems, Atlanta GA), taken from an Elekta Synergy TM system for CBCT-guided radiation therapy and adapted to the imaging bench. The bowtie was precisely centered at the exit face of the collimator and was in place during flood-field calibration. Ideally, the bowtie would provide full lateral modulation of the x-ray beam projected through the water cylinder, such that the resulting mean detector signal (in pre-flood-field-corrected projections) was constant (1:1 modulation between the center and edge of the image). Without the bowtie filter, the signal modulation from the center to the edge of the image was 1:1. With the bowtie, the modulation improved to 5:1, and while this was less than ideal, it was believed sufficient to give a measurable effect on noise and stationarity. Future work will incorporate a stronger (thicker / higher density) bowtie. (b) Water + Bowtie Figure. Example axial images of the three phantom configurations, giving qualitative illustration of possible variation in noise stationarity. The hypothesis underlying these experiments was that various experimental configurations would result in varying degrees of noise stationarity, measurable in terms of: 1.) locally varying NPS; and.) relative magnitude of off-diagonal elements of the DFT covariance matrix. Specifically, we hypothesize that Air images would exhibit the highest degree of stationarity overall and that images of the Water Cylinder with a bowtie filter would exhibit higher stationarity than without a bowtie filter. The basis for this hypothesis is the degree of lateral uniformity in detector SNR highest for Air and worst for Water without a bowtie. Example axial image reconstructions for the three phantom configurations are shown in Fig., in qualitative agreement with these assertions. The stationarity of first-order statistics (i.e., the mean) appears highest for the Air phantom and worst for the Water (No Bowtie) case (highest level of x-ray scatter shading artifact). The stationarity of the second-order statistics (i.e., the correlations) deserves closer inspection. For the Air scan, there appears to be a rich variation of spatial-frequency dependence in the noise from the center (ring correlation) to intermediate radii (uniform mid-pass noise) to the periphery (rings and view aliasing). Similarly for the Water phantom configurations: one would expect radial dependence of the variance according to the transmittance of the cylinder (lower quantum noise at the periphery); however, the images suggest a tradeoff in noise texture among ring correlation, midpass quantum noise, and view aliasing. 3. Measurement of the 3D Noise-Power Spectrum and DFT Covariance Matrix The resulting 1 volume reconstructions (3 phantoms x 4 exposure levels) provided the basis for analysis of the NPS, local NPS, and DFT covariance matrix. Each volume was divided into non-overlapping sub-volumes ROIs of (64 x 64 x 64) voxels, 6 of which were taken from completely within the cylindrical volume. Prior to spectral estimation, each ROI was detrended by subtraction of a 3D planar fit, giving zero-mean data free from first-order signal non-uniformity (e.g., x-ray scatter shading artifact). The 3D NPS was estimated from the 3D FFT of the zero-mean data using equation (1). Any element of the full covariance matrix can be estimated using equation (). Because CT reconstruction applies a high-pass filter in the in-plane (x,y) direction but not in the longitudinal (z) direction, we focus on the frequencies in the x direction. We compute the NPS by averaging the periodograms (square of the magnitude of the normalized DFT) over the 6 zero-mean image blocks. The covariance matrix in the DFT domain has 64 3 x 64 3 complex elements. There are methods for dealing with such large covariance matrices [1] but for the purposes of this work we limit ourselves to estimating diagonals. We explore experimentally the components of the DFT Proc. of SPIE Vol Q-4 Downloaded from SPIE Digital Library on 13 Nov 9 to Terms of Use:

5 covariance matrix that are typically ignored in NPS analysis. For example, the 1 st off-diagonal in the x-direction describes the correlations with the nearest frequency in the x-direction which likely represents the greatest component of off-diagonal (non-stationary) noise. 4. RESULTS 4.1 The 3D Noise-Power Spectrum Figure 3 shows slices of the 3D NPS for the Air scan reconstructions. Note the asymmetric characteristic of the NPS between axial and sagittal planes, as previously reported and in agreement with theoretical expectations. The axial domain exhibits a mid-pass spectrum that is characteristic of filtered backprojection (ramp + apodization filters), whereas the NPS in the longitudinal direction is band-limited according to the low-pass characteristics of the detector (indirect detection FPD), interpolation filter, etc. Such analysis is the basis for assessment of noise-equivalent quanta (NEQ) and is becoming an increasingly important characterization of the noise in CT systems evaluation, quality assurance, and optimization. NPSInAXLPInsforAir Spatial Frequency, f (mmd) xlo' (a) (b) NPS In S.glttai Pluis for Air Spatial Frequency, f1 (mmd) I io' Figure 3. Orthogonal slices of the 3D NPS estimated from Air scan reconstructions: (a) axial NPS and (b) sagittal NPS. The asymmetry in the noise between axial and saggital planes is a characterization of asymmetric filtration in 3D filtered backprojection. Data shown here are from reconstructions of Air at mas per projection. The extent to which the 3D NPS is invariant with location throughout the reconstruction FOV was investigated in terms of the local NPS estimated from ROIs at a specific radius (assuming NPS constant at a given radius). The resulting local NPS were examined as a function of distance from the reconstruction center as illustrated in Figs. 4 and 5. A number of features come to light from such analysis. For the Water phantom, the reduction in NPS at larger radius is attributable to increased fluence (reduced attenuation) nearer the edges of the detector (i.e., reduced fluence higher attenuation at the center of the detector in the shadow of the thickest part of the phantom, resulting in higher NPS). A competing effect is view sampling, which becomes more significant at larger radius and tends to increase the NPS. Second, the extent to which the NPS varies with position for a given phantom configuration is evident in the results. Each phantom exhibits significant variation of NPS with radius, but for the Air data, the NPS varies less (i.e., more stationary) than for the water data consistent with hypothesis. Finally, the axial NPS is seen to exhibit a marked change not only in magnitude, but also in spatial-frequency content, changing from a radially symmetric donut at smaller radii to a more lobular characteristic at greater radii. Moreover, the change in such frequency characteristics is slightly different among the three phantom configurations a somewhat more pronounced, less radially symmetric shape observed for the water phantom data. Proc. of SPIE Vol Q-5 Downloaded from SPIE Digital Library on 13 Nov 9 to Terms of Use:

6 (a) Air (b) Water (with Bowtie Filter) (c) Water (No Bowtie)..5 (x1o)i R-.5 (x15) 4cm R 4 cm.. U- a). z (a)air i :: d.5 1:i.5 :Lo 6.5. Spatial Frequency (mm-') Spatial Frequency (mm-') Figure 4. The local 3D NPS measured for various phantom configurations. (a-c) Axial NPS estimated as a function of distance from reconstruction center (R ~4-8 cm). The Air phantom shows a near independence in NPS with varying radius, while the water phantoms show a distinct reduction in NPS with radius. (b) Water (with Bowtie Filter) (C) Water (No Bowtie) Figure 5. The local 3D NPS measured for various phantom configurations as a function of distance from reconstruction center. (a-c) Axial NPS are illustrated as a function of distance from center, demonstrating the increasingly asymmetric, lobular structure of the NPS at extended radii a clear demonstration of non-stationarity. 4. Off-Diagonal Elements of the DFT Covariance Matrix Figure 6 shows the 1 st off-diagonal element of the DFT covariance matrix, which quantifies the correlation of two neighboring frequencies in the x-direction. Such likely represents the greatest component of off-diagonal components in the DFT covariance matrix. Proc. of SPIE Vol Q-6 Downloaded from SPIE Digital Library on 13 Nov 9 to Terms of Use:

7 First Of.dI.gonL In x In AXL Pluis for Air xl o' First Of.dI.gonL In x In S.glttai Pluis for Air xl o' Spatial Frequency, f (mmd) (a) (b) Spatial Frequency, f1 (mmd) Figure 6. First off diagonal of the DFT covariance matrix in the x-frequency direction, i.e. the correlation of the given frequency with the next closest frequency. These components are commonly ignored in typical NPS analysis. By comparing the axial NPS [Fig. 3(a)] with the axial representation of the 1 st off-diagonal element [Fig. 6(a)], we see that the off-diagonal elements are small but not negligible with respect to the NPS. In this case, the two exhibit similar shape in the frequency domain and differ in magnitude by approximately a factor of 1. Subsequent analysis (not shown here) demonstrates that the nd, 3 rd, etc. off-diagonal elements are reduced still further, but persist in a manner that may not be entirely negligible in system noise characterization and evaluation of detectability. i O Radial Average of NPS and ODE in Axial Plane (Air) 6 NPS ' FirstODEinx I Frequency mm Figure 7. Axial NPS in comparison to off-diagonal elements (ODE) of the DFT covariance matrix. The plots represent radial averages of the axial plots in Figs. 3(a) and 6(a). Data are for Air scan reconstructions at mas per projection. 4.3 Simple Measures of Stationarity We are interested in quantifying the degree of stationarity of the images. A simple spatial domain measure of stationarity is the variability of the standard deviation along the radius in the axial plane of the volume. For a stationary random process, the standard deviation would be constant. Figure 8 demonstrates how the standard deviation of the noise changes with radius for the three phantom configurations described in Table I. The Air scan reconstructions are seen to exhibit the greatest degree of stationarity, with image noise increasing with radius by a few percent. The increase in noise at larger radius is attributed to view sampling (finite number of projections). The Water+Bowtie and Water (no bowtie) configurations are considerably less stationary, each demonstrating a reduction in noise at larger radius of Proc. of SPIE Vol Q-7 Downloaded from SPIE Digital Library on 13 Nov 9 to Terms of Use:

8 several percent. The reduction is attributed to higher detector signal (less attenuation and, therefore, higher incident fluence) at greater distances from center. Of the three phantom configurations, Water (no bowtie) is the least stationary (largest variation in noise) consistent with hypothesis. I.5 Normalized SD of noise at.4 mas = a (.95.9 water -water + bowtie air Radius (cm) Figure 8. Standard deviation of the noise as a function of radius (distance from reconstruction center). The standard deviation is normalized to 1 at a radius of 4 cm for all three imaging conditions. The variability with radius is consistent with a rank order that Air is more stationary than Water+Bowtie, which in turn is more stationary than Water (without bowtie). We can define simple measures of relative stationarity by computing the standard deviation of the noise as a function of radius and dividing by the average standard deviation: Κ σ stdev = ( σ ) σ where σ is the sample standard deviation in voxel values for a given region of interest (ROI). The term stdev(σ) is the standard deviation in the noise e.g., exemplified in the variation with radius shown in Fig. 8. For a constant standard deviation across the image, therefore, the quantity K σ should be zero. Further, we hypothesize smaller values of K σ for images that are more stationary specifically, K air σ < K water+bowtie σ < K water σ. Similarly, a measure of stationarity based on the off-diagonal elements of the Fourier covariance matrix may be defined as: Κ 1 NPS = 1 ODE x NPS were the superscript 1 refers to the 1 st off-diagonal element. The numerator is based on the correlations with the nearest frequency in the x-direction, and the denominator to the variance as defined by the integral over the NPS. We compare the degree of stationarity of the three experimental configurations based on these simple measures of relative stationarity. As shown in Table II, the results are consistent with hypothesis specifically, the K metrics are smallest for the Air scan reconstructions (most stationary) and largest for the Water (without bowtie) reconstructions (least stationary). (3) (4) Proc. of SPIE Vol Q-8 Downloaded from SPIE Digital Library on 13 Nov 9 to Terms of Use:

9 (a) Spatial Domain: K σ (Noise Non-Stationarity) Exposure Level (mas / projection) Phantom Configuration.4 mas.8-1 mas 1.6- mas 3.-4 mas Water Water + Bowtie Air (b) Fourier Domain: K 1 NPS (NPS Non-Stationarity) Exposure Level (mas / projection) Phantom Configuration.4 mas.8-1 mas 1.6- mas 3.-4 mas Water Water + Bowtie Air Table II. Summary of (a) spatial and (b) Fourier domain (non-) stationarity metrics as defined in Eqs (3) and (4), respectively. Both spatial and Fourier metrics should be zero for stationary images. We observe overall consistency with hypothesis regarding the extent to which experimental conditions affect stationarity viz., air data are more stationary than water data, and the bowtie filter helps somewhat to restore the degree of stationarity. 5. DISCUSSION AND CONCLUSIONS These results explore for the first time the complex correlations in the Fourier components of CBCT data. The analysis suggests that as the non-stationarity increases so do the off-diagonal elements of the DFT covariance matrix; hence, ignoring such effects by characterizing noise solely in terms of the NPS would lead to significant errors or misinterpretation. The spatial domain analog of the current investigation is moving from looking solely at the variance across the image to looking at the correlations between pixels. Since the NPS is the variance of the Fourier coefficients, we are looking to do the same in the Fourier domain by understanding the complex correlations between different frequencies. In light of the non-negligible magnitude of off-diagonal terms under certain experimental conditions, we need to develop further a proper interpretation and intuition as to their meaning and impact on the practical evaluation of noise properties of CT reconstructions. In particular, two questions beckon: (i) what level of non-stationarity is tolerable such that NPS is a reliable, meaningful metric of image noise?; and (ii) to what extent (and in what manner) do the observed non-stationarities affect detectability? Future work also involves refinement of the experimental and analysis techniques in estimating elements of the DFT covariance matrix. The variability in the correlations in the data associated with multiple realizations of the noise volumes needs to be better understood to avoid the effects of structured artifacts (e.g., rings and streaks). Furthermore, possible effects of arising from the size of the subvolumes and associated spectral leakage need to be more fully investigated and minimized in order to obtain accurate estimates of the DFT covariance matrix. The measures of relative stationarity defined in this paper were for the most part heuristic. While the results (Table II) agreed overall with the hypothesis regarding experimental conditions and degree of stationarity, we still need to understand their variability and proper interpretation. The single data point in Table II that did not agree with hypothesis (Air data at.4 mas) is believed to be an outlier. Future work will include more fundamental metrics of stationarity in terms of the difference in detectability estimates arising from the NPS alone versus the full DFT covariance matrix as well as statistics related to testing stationarity as a hypothesis test. The hypothesis underlying these experiments was that implementing a bowtie filter would improve stationarity for the water reconstructions, which was indeed observed, but to a somewhat smaller extent than ideal. There are several Proc. of SPIE Vol Q-9 Downloaded from SPIE Digital Library on 13 Nov 9 to Terms of Use:

10 reasons why the improvement in stationarity was not as much as might be expected. (i.) An ideal bowtie filter would modulate the beam such that the (raw, uncorrected) projection image is completely flat. Without the bowtie, we measured approximately 1:1 modulation between the center of the beam and the edge (bare-beam). With the bowtie, the modulation dropped to 5:1, which is less than the 1:1 ideal. One might expect only a slight improvement in stationarity given this (fairly weak) bowtie. Future work will include a more aggressive bowtie that gives stronger (nearly 1:1) modulation. (ii.) It is worth noting that even for a perfect (1:1 modulation) bowtie filter, there is still an important potential source of non-stationarity: hardening of the beam at the periphery (more than at the center). While such a bowtie may perfectly modulate the mean signal, one must consider how the filter has modulated the energy of the beam - - and therefore, the DQE of the detector. Even with the mean signal perfectly modulated, if the beam incident on the detector is 'harder' (higher mean energy) at the periphery than at the center, the DQE() is likely lower at the periphery than at the center, and one would expect higher quantum noise at the periphery. The converse is also true. Therefore, there are underlying factors in modulating the mean signal that affect the noise and, hence, the stationarity. Such should be investigated in future work, including modeling of the x-ray beam and use of bowtie filters of various material types. It is also worth noting that the metrics of relative stationarity were designed to characterize the degree of stationarity but do not necessarily tell if the noise is higher or lower in one area of the image than another. The local NPS does provide such information and conveys further how the noise correlation differs. Future work will study different metrics related to the DFT covariance matrix to better understand how to characterize both the degree of non-stationarity as well as the variations in noise correlation. In summary, the degree of stationarity (as characterized by the off-diagonal elements of the DFT covariance matrix with respect to the NPS) depends on the experimental imaging conditions, and conditions can be selected that improve the stationarity assumption. Our work shows that the relative magnitude of off-diagonal elements conventionally ignored in NPS analysis warrants investigation of their effect on object detectability and that ignoring such terms may lead to an inaccurate characterization of the image noise. ACKNOWLEDGEMENTS This work was performed in a research consortium between the University Health Network (Toronto ON), California State University (Fullerton CA), and Stanford University (Palo Alto CA). The work was supported by the National Institutes of Health, R1-CA REFERENCES 1. Barrett, H.H., Objective assessment of image quality: Effects of quantum noise and object variability, J. Opt. Soc. Am. A 7, (199).. Hanson, K.M., "The characteristics of computed-tomographic reconstruction noise and their effect on detectability," IEEE Trans. Nucl. Sci. NS-5, (1978). 3. Siewerdsen, J.H. and Jaffray, D.A., Optimization of x-ray imaging geometry (with specific application to flat-panel cone-beam computed tomography, Med. Phys. 7(8), (). 4. Barrett, H.H. and Myers K.J., Foundations of image science (Wiley-Interscience Hoboken, N.J. 4) 5. Riederer, S.J., Pelc, N.J., and Chesler, D.A., The Noise Power Spectrum in Computed X-ray Tomography, Phys. Med. Biol., 3 (3), (1978). 6. Siewerdsen, J.H., Cunningham, I.A. and Jaffray, D.A., A framework for noise-power spectrum analysis of multidimensional images, Med. Phys. 9(11), (). 7. Kay, S.M., Fundamentals of Statistical Signal Processing (Prentice-Hall, Upper Saddel River, NJ. 1993) 8. Gagne, R.M., Myers, K.J., and Quin P.W., Effect of shift invariance and stationary assumptions on simple detection tasks: Spatial and spatial frequency domains, Proc. SPIE 43, (1). 9. Pineda, A.R. and Barrett, H.H., Figures of merit for detectors in digital radiography. I. Flat background and deterministic blurring, Med. Phys. 31() (4) 1. Barrett H.H., Myers K.J., Gallas B., Clarkson E. and Zhang H., Megalopinakophobia: Its symptoms and cures, Proc. SPIE 43, (1). Proc. of SPIE Vol Q-1 Downloaded from SPIE Digital Library on 13 Nov 9 to Terms of Use:

Generalized DQE analysis of radiographic and dual-energy imaging using flat-panel detectors

Generalized DQE analysis of radiographic and dual-energy imaging using flat-panel detectors Generalized DQE analysis of radiographic and dual-energy imaging using flat-panel detectors S. Richard Department of Medical Biophysics, University of Toronto, Ontario, Canada M5G M9 J. H. Siewerdsen a

More information

The Generalized NEQ and Detectability Index for Tomosynthesis and Cone-Beam CT

The Generalized NEQ and Detectability Index for Tomosynthesis and Cone-Beam CT The Generalized NEQ and Detectability Inde for Tomosynthesis and Cone-Beam CT Grace J. Gang 1, Junghoon Lee 2, J. Webster Stayman 3, Daniel Tward 3,W. Zbijewski 3, Jerry L. Prince 2, and Jeffrey H. Siewerdsen

More information

Ke Li and Guang-Hong Chen

Ke Li and Guang-Hong Chen Ke Li and Guang-Hong Chen Brief introduction of x-ray differential phase contrast (DPC) imaging Intrinsic noise relationship between DPC imaging and absorption imaging Task-based model observer studies

More information

Steven Tilley Fully 3D Recon 2015 May 31 -June 4

Steven Tilley Fully 3D Recon 2015 May 31 -June 4 Fully3D Advanced System Models for Reconstruction in Flat-Panel Detector Cone-Beam CT Steven Tilley, Jeffrey Siewerdsen, Web Stayman Johns Hopkins University Schools of Medicine and Engineering Acknowledgements

More information

Midterm Review. Yao Wang Polytechnic University, Brooklyn, NY 11201

Midterm Review. Yao Wang Polytechnic University, Brooklyn, NY 11201 Midterm Review Yao Wang Polytechnic University, Brooklyn, NY 11201 Based on J. L. Prince and J. M. Links, Medical maging Signals and Systems, and lecture notes by Prince. Figures are from the textbook.

More information

Neutron and Gamma Ray Imaging for Nuclear Materials Identification

Neutron and Gamma Ray Imaging for Nuclear Materials Identification Neutron and Gamma Ray Imaging for Nuclear Materials Identification James A. Mullens John Mihalczo Philip Bingham Oak Ridge National Laboratory Oak Ridge, Tennessee 37831-6010 865-574-5564 Abstract This

More information

Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions

Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 2, APRIL 2012 469 Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions Weiyi Wang, Member,

More information

Initial Studies in Proton Computed Tomography

Initial Studies in Proton Computed Tomography SCIPP Initial Studies in Proton Computed Tomography L. R. Johnson, B. Keeney, G. Ross, H. F.-W. Sadrozinski, A. Seiden, D.C. Williams, L. Zhang Santa Cruz Institute for Particle Physics, UC Santa Cruz,

More information

The relationship between image noise and spatial resolution of CT scanners

The relationship between image noise and spatial resolution of CT scanners The relationship between image noise and spatial resolution of CT scanners Sue Edyvean, Nicholas Keat, Maria Lewis, Julia Barrett, Salem Sassi, David Platten ImPACT*, St George s Hospital, London *An MDA

More information

Towards Proton Computed Tomography

Towards Proton Computed Tomography SCIPP Towards Proton Computed Tomography L. R. Johnson, B. Keeney, G. Ross, H. F.-W. Sadrozinski, A. Seiden, D.C. Williams, L. Zhang Santa Cruz Institute for Particle Physics, UC Santa Cruz, CA 95064 V.

More information

Reconstruction for Proton Computed Tomography: A Monte Carlo Study

Reconstruction for Proton Computed Tomography: A Monte Carlo Study Reconstruction for Proton Computed Tomography: A Monte Carlo Study T Li, Z. Liang, K. Mueller, J. Heimann, L. Johnson, H. Sadrozinski, A. Seiden, D. Williams, L. Zhang, S. Peggs, T. Satogata, V. Bashkirov,

More information

Error Reporting Recommendations: A Report of the Standards and Criteria Committee

Error Reporting Recommendations: A Report of the Standards and Criteria Committee Error Reporting Recommendations: A Report of the Standards and Criteria Committee Adopted by the IXS Standards and Criteria Committee July 26, 2000 1. Introduction The development of the field of x-ray

More information

Factors affecting Probability of Detection with Computed Radiography

Factors affecting Probability of Detection with Computed Radiography 17th World Conference on Nondestructive Testing, 5-8 Oct 008, Shanghai, China Factors affecting Probability of Detection with Computed Radiography Gregory A. MOHR 1 and Peter WILLEMS 1 GE Inspection Technologies

More information

AUTOMATED TEMPLATE MATCHING METHOD FOR NMIS AT THE Y-12 NATIONAL SECURITY COMPLEX

AUTOMATED TEMPLATE MATCHING METHOD FOR NMIS AT THE Y-12 NATIONAL SECURITY COMPLEX AUTOMATED TEMPLATE MATCHING METHOD FOR NMIS AT THE Y-1 NATIONAL SECURITY COMPLEX J. A. Mullens, J. K. Mattingly, L. G. Chiang, R. B. Oberer, J. T. Mihalczo ABSTRACT This paper describes a template matching

More information

Health Physics Services Ltd

Health Physics Services Ltd 1 Title: Adept Medical STARTable: Scatter Radiation Shielding Grid Format Methodology 2 Background: Adept Medical STARTable Shield is embedded with 0.5mm Lead (Pb), offering protection from scatter radiation

More information

Proton radiography with a range telescope and its use in proton therapy

Proton radiography with a range telescope and its use in proton therapy Proton radiography with a range telescope and its use in proton therapy Juan L. Romero Department of Physics, University of California, Davis, CA Invited talk at Massachusetts General Hospital, Boston,

More information

Neural Network Approach for Photon-counting Detection The First Step: PPE Correction

Neural Network Approach for Photon-counting Detection The First Step: PPE Correction Neural Network Approach for Photon-counting Detection The First Step: PPE Correction Ruibin Feng, Ph.D. Biomedical Imaging Center, CBIS/BME, RPI fengr@rpi.edu David Rundle JairiNovus Technologies Ltd.

More information

Two-Material Decomposition From a Single CT Scan Using Statistical Image Reconstruction

Two-Material Decomposition From a Single CT Scan Using Statistical Image Reconstruction / 5 Two-Material Decomposition From a Single CT Scan Using Statistical Image Reconstruction Yong Long and Jeffrey A. Fessler EECS Department James M. Balter Radiation Oncology Department The University

More information

Image Assessment San Diego, November 2005

Image Assessment San Diego, November 2005 Image Assessment San Diego, November 005 Pawel A. Penczek The University of Texas Houston Medical School Department of Biochemistry and Molecular Biology 6431 Fannin, MSB6.18, Houston, TX 77030, USA phone:

More information

1. Which of the following statements is true about Bremsstrahlung and Characteristic Radiation?

1. Which of the following statements is true about Bremsstrahlung and Characteristic Radiation? BioE 1330 - Review Chapters 4, 5, and 6 (X-ray and CT) 9/27/2018 Instructions: On the Answer Sheet, enter your 2-digit ID number (with a leading 0 if needed) in the boxes of the ID section. Fill in the

More information

Comparison of Polychromatic and Monochromatic X-rays for Imaging

Comparison of Polychromatic and Monochromatic X-rays for Imaging Comparison of Polychromatic and Monochromatic X-rays for Imaging M. Hoheisel 1, P. Bernhardt 1, R. Lawaczeck 2, and H. Pietsch 2 1 Siemens AG Medical Solutions, Forchheim, Germany 2 Schering AG Imaging

More information

Foundations of Image Science

Foundations of Image Science Foundations of Image Science Harrison H. Barrett Kyle J. Myers 2004 by John Wiley & Sons,, Hoboken, 0-471-15300-1 1 VECTORS AND OPERATORS 1 1.1 LINEAR VECTOR SPACES 2 1.1.1 Vector addition and scalar multiplication

More information

Response of DIMM turbulence sensor

Response of DIMM turbulence sensor Response of DIMM turbulence sensor A. Tokovinin Version 1. December 20, 2006 [tdimm/doc/dimmsensor.tex] 1 Introduction Differential Image Motion Monitor (DIMM) is an instrument destined to measure optical

More information

Electron therapy Class 2: Review questions

Electron therapy Class 2: Review questions Electron therapy Class 2: Review questions 1 Raphex Question: T63, 2002 In what situation is electron backscatter likely to be a problem? A. Using 1cm of tissue equivalent bolus on the skin. B. Using a

More information

Noise Analysis of Regularized EM for SPECT Reconstruction 1

Noise Analysis of Regularized EM for SPECT Reconstruction 1 Noise Analysis of Regularized EM for SPECT Reconstruction Wenli Wang and Gene Gindi Departments of Electrical Engineering and Radiology SUNY at Stony Brook, Stony Brook, NY 794-846 Abstract The ability

More information

Introduction to the Mathematics of Medical Imaging

Introduction to the Mathematics of Medical Imaging Introduction to the Mathematics of Medical Imaging Second Edition Charles L. Epstein University of Pennsylvania Philadelphia, Pennsylvania EiaJTL Society for Industrial and Applied Mathematics Philadelphia

More information

Electron density and effective atomic number images generated by dual energy imaging with a 320-detector CT system: A feasibility study

Electron density and effective atomic number images generated by dual energy imaging with a 320-detector CT system: A feasibility study Electron density and effective atomic number images generated by dual energy imaging with a 320-detector CT system: A feasibility study Poster No.: C-0403 Congress: ECR 2014 Type: Scientific Exhibit Authors:

More information

Objective Functions for Tomographic Reconstruction from. Randoms-Precorrected PET Scans. gram separately, this process doubles the storage space for

Objective Functions for Tomographic Reconstruction from. Randoms-Precorrected PET Scans. gram separately, this process doubles the storage space for Objective Functions for Tomographic Reconstruction from Randoms-Precorrected PET Scans Mehmet Yavuz and Jerey A. Fessler Dept. of EECS, University of Michigan Abstract In PET, usually the data are precorrected

More information

Index. p, lip, 78 8 function, 107 v, 7-8 w, 7-8 i,7-8 sine, 43 Bo,94-96

Index. p, lip, 78 8 function, 107 v, 7-8 w, 7-8 i,7-8 sine, 43 Bo,94-96 p, lip, 78 8 function, 107 v, 7-8 w, 7-8 i,7-8 sine, 43 Bo,94-96 B 1,94-96 M,94-96 B oro!' 94-96 BIro!' 94-96 I/r, 79 2D linear system, 56 2D FFT, 119 2D Fourier transform, 1, 12, 18,91 2D sinc, 107, 112

More information

In today s lecture, we want to see what happens when we hit the target.

In today s lecture, we want to see what happens when we hit the target. In the previous lecture, we identified three requirements for the production of x- rays. We need a source of electrons, we need something to accelerate electrons, and we need something to slow the electrons

More information

Monte Carlo Characterization of a Pulsed Laser-Wakefield Driven Monochromatic X-Ray Source

Monte Carlo Characterization of a Pulsed Laser-Wakefield Driven Monochromatic X-Ray Source 2009 IEEE Nuclear Science Symposium Conference Record N30-3 Monte Carlo Characterization of a Pulsed Laser-Wakefield Driven Monochromatic X-Ray Source S. D. Clarke, S. A. Pozzi, IEEE Member, N. Cunningham,

More information

Monte Carlo Analyses of X-Ray Absorption, Noise, and Detective Quantum Efficiency Considering Therapeutic X-Ray Spectrum in Portal Imaging Detector

Monte Carlo Analyses of X-Ray Absorption, Noise, and Detective Quantum Efficiency Considering Therapeutic X-Ray Spectrum in Portal Imaging Detector IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 48, NO. 4, AUGUST 2001 1423 Monte Carlo Analyses of X-Ray Absorption, Noise, and Detective Quantum Efficiency Considering Therapeutic X-Ray Spectrum in Portal

More information

Maximal Entropy for Reconstruction of Back Projection Images

Maximal Entropy for Reconstruction of Back Projection Images Maximal Entropy for Reconstruction of Back Projection Images Tryphon Georgiou Department of Electrical and Computer Engineering University of Minnesota Minneapolis, MN 5545 Peter J Olver Department of

More information

Simulation Modeling in Dosimetry

Simulation Modeling in Dosimetry Simulation Modeling in Dosimetry Aleksei Zhdanov Ural Federal University named after the first President of Russia B. N. Yeltsin, Yekaterinburg, Russian Federation jjj1994@yandex.ru Leonid Dorosinskiy

More information

Henry Ford NERS/BIOE 481. Lecture 07 Metrics of Image Quality

Henry Ford NERS/BIOE 481. Lecture 07 Metrics of Image Quality NERS/BIOE 481 Lecture 07 Metrics of Image Quality Michael Flynn, Adjunct Prof Nuclear Engr & Rad. Science mikef@umich.edu mikef@rad.hfh.edu Henry Ford Health System RADIOLOGY RESEARCH I.A.1 - General Model

More information

Combined neutron and X-ray imaging on different length scales

Combined neutron and X-ray imaging on different length scales More Info at Open Access Database www.ndt.net/?id=18740 Combined neutron and X-ray imaging on different length scales Anders Kaestner, David Mannes, Jan Hovind, Pierre Boillat, and Eberhard Lehmann Neutron

More information

arxiv:hep-ex/ v1 5 Apr 2000

arxiv:hep-ex/ v1 5 Apr 2000 Track Fit Hypothesis Testing and Kink Selection using Sequential Correlations Robert V. Kowalewski and Paul D. Jackson arxiv:hep-ex/48v 5 Apr 2 Abstract Department of Physics and Astronomy, University

More information

A quantum accounting and detective quantum efficiency analysis for video-based portal imaging

A quantum accounting and detective quantum efficiency analysis for video-based portal imaging A quantum accounting and detective quantum efficiency analysis for video-based portal imaging Jean-Pierre Bissonnette a),b) London Regional Cancer Centre, 790 Commissioners Rd. E., London, Ontario, N6A

More information

A Monte Carlo Study of the Relationship between the Time. Structures of Prompt Gammas and in vivo Radiation Dose in.

A Monte Carlo Study of the Relationship between the Time. Structures of Prompt Gammas and in vivo Radiation Dose in. A Monte Carlo Study of the Relationship between the Time Structures of Prompt Gammas and in vivo Radiation Dose in Proton Therapy Wook-Geun Shin and Chul Hee Min* Department of Radiation Convergence Engineering,

More information

FLUKA simulations of selected topics regarding proton pencil beam scanning

FLUKA simulations of selected topics regarding proton pencil beam scanning FLUKA simulations of selected topics regarding proton pencil beam scanning C. Bäumer, J. Farr, J. Lambert and B. Mukherjee Westdeutsches Protonentherapiezentrum Essen, Germany T. Mertens, and B. Marchand

More information

Practical Statistics

Practical Statistics Practical Statistics Lecture 1 (Nov. 9): - Correlation - Hypothesis Testing Lecture 2 (Nov. 16): - Error Estimation - Bayesian Analysis - Rejecting Outliers Lecture 3 (Nov. 18) - Monte Carlo Modeling -

More information

Imploded Shell Parameter Estimation Based on Radiograph Analysis. George Liu. Pittsford Sutherland High School. LLE Advisor: Reuben Epstein

Imploded Shell Parameter Estimation Based on Radiograph Analysis. George Liu. Pittsford Sutherland High School. LLE Advisor: Reuben Epstein Imploded Shell Parameter Estimation Based on Radiograph Analysis George Liu Pittsford Sutherland High School LLE Advisor: Reuben Epstein Laboratory for Laser Energetics University of Rochester Summer High

More information

Data Acquisition and Image Formation Methods for Multi Energy CT

Data Acquisition and Image Formation Methods for Multi Energy CT Data Acquisition and Image Formation Methods for Multi Energy CT Cynthia H. McCollough, PhD, DABR, FAIMBE, FAAPM, FACR Professor of Medical Physics and Biomedical Engineering Director, CT Clinical Innovation

More information

THE OBSERVATION AND ANALYSIS OF STELLAR PHOTOSPHERES

THE OBSERVATION AND ANALYSIS OF STELLAR PHOTOSPHERES THE OBSERVATION AND ANALYSIS OF STELLAR PHOTOSPHERES DAVID F. GRAY University of Western Ontario, London, Ontario, Canada CAMBRIDGE UNIVERSITY PRESS Contents Preface to the first edition Preface to the

More information

Nuclear Instruments and Methods in Physics Research A 417 (1998) 86 94

Nuclear Instruments and Methods in Physics Research A 417 (1998) 86 94 Nuclear Instruments and Methods in Physics Research A 417 (1998) 86 94 Experimental determination of detector gain, zero frequency detective quantum efficiency, and spectral compatibility of phosphor screens:

More information

The Thermal Sieve: a diffractive baffle that provides thermal isolation of a cryogenic optical system from an ambient temperature collimator

The Thermal Sieve: a diffractive baffle that provides thermal isolation of a cryogenic optical system from an ambient temperature collimator The Thermal Sieve: a diffractive baffle that provides thermal isolation of a cryogenic optical system from an ambient temperature collimator James H. Burge * and Dae Wook Kim College of Optical Sciences

More information

Development of a Monte Carlo Simulation Code and Its Test Results for Optimal Design of a Digital Radiographic System Based on a CMOS Image Sensor

Development of a Monte Carlo Simulation Code and Its Test Results for Optimal Design of a Digital Radiographic System Based on a CMOS Image Sensor Journal of the Korean Physical Society, Vol. 46, No. 2, February 2005, pp. 418 424 Development of a Monte Carlo Simulation Code and Its Test Results for Optimal Design of a Digital Radiographic System

More information

X-RAY MICRO-TOMOGRAPHY OF PORE-SCALE FLOW AND TRANSPORT. University of California Davis. Dorthe Wildenschild & Annette Mortensen

X-RAY MICRO-TOMOGRAPHY OF PORE-SCALE FLOW AND TRANSPORT. University of California Davis. Dorthe Wildenschild & Annette Mortensen X-RAY MICRO-TOMOGRAPHY OF PORE-SCALE FLOW AND TRANSPORT Jan W. Hopmans University of California Davis Volker Clausnitzer Dorthe Wildenschild & Annette Mortensen ISSUES: Measurements and modeling of water

More information

Lecture 5: Tomographic nuclear systems: SPECT

Lecture 5: Tomographic nuclear systems: SPECT Lecture 5: Tomographic nuclear systems: SPECT Field trip this saturday at 11 AM at UWMC meet in main hospital lobby at 11 AM if you miss the 'boat', page me at 540-4950 should take ~1 to 1.5 hours, depending

More information

Medical Physics. Image Quality 1) Ho Kyung Kim. Pusan National University

Medical Physics. Image Quality 1) Ho Kyung Kim. Pusan National University Medical Physics Prince & Links 3 Image Quality 1) Ho Kyung Kim hokyung@pusan.ac.kr Pusan National University 1) The degree to which an image allows medical professionals to accomplish their goals (e.g.,

More information

Evaluation of influencing factors in Dual Energy X-ray imaging

Evaluation of influencing factors in Dual Energy X-ray imaging 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Evaluation of influencing factors in Dual Energy X-ray imaging More Info at Open Access Database

More information

Simulations of an etched spiral axial attenuation scheme for an on-axis reflecting telescope

Simulations of an etched spiral axial attenuation scheme for an on-axis reflecting telescope Journal of Physics: Conference Series PAPER OPEN ACCESS Simulations of an etched spiral axial attenuation scheme for an on-axis reflecting telescope To cite this article: Aaron Spector and Guido Mueller

More information

Notes on the point spread function and resolution for projection lens/corf. 22 April 2009 Dr. Raymond Browning

Notes on the point spread function and resolution for projection lens/corf. 22 April 2009 Dr. Raymond Browning Notes on the point spread function and resolution for projection lens/corf Abstract April 009 Dr. Raymond Browning R. Browning Consultants, 4 John Street, Shoreham, NY 786 Tel: (63) 8 348 This is a calculation

More information

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT Proceedings of the Eleventh EGS4 Users' Meeting in Japan, KEK Proceedings 2003-15, p.107-113 ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT K. Noto and

More information

A Brief Introduction to Medical Imaging. Outline

A Brief Introduction to Medical Imaging. Outline A Brief Introduction to Medical Imaging Outline General Goals Linear Imaging Systems An Example, The Pin Hole Camera Radiations and Their Interactions with Matter Coherent vs. Incoherent Imaging Length

More information

1-D Fourier Transform Pairs

1-D Fourier Transform Pairs 1-D Fourier Transform Pairs The concept of the PSF is most easily explained by considering a very small point source being placed in the imaging field-of-view The relationship between the image, I, and

More information

5.1 2D example 59 Figure 5.1: Parabolic velocity field in a straight two-dimensional pipe. Figure 5.2: Concentration on the input boundary of the pipe. The vertical axis corresponds to r 2 -coordinate,

More information

COMPUTATIONAL IMAGING. Berthold K.P. Horn

COMPUTATIONAL IMAGING. Berthold K.P. Horn COMPUTATIONAL IMAGING Berthold K.P. Horn What is Computational Imaging? Computation inherent in image formation What is Computational Imaging? Computation inherent in image formation (1) Computing is getting

More information

A comparison of methods for monitoring photon beam energy constancy

A comparison of methods for monitoring photon beam energy constancy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 A comparison of methods for monitoring photon beam energy constancy Song Gao, 1a Peter A. Balter, 1 Mark Rose, 2 and William E. Simon

More information

The effect of dose calculation uncertainty on the evaluation of radiotherapy plans

The effect of dose calculation uncertainty on the evaluation of radiotherapy plans The effect of dose calculation uncertainty on the evaluation of radiotherapy plans P. J. Keall a) and J. V. Siebers Department of Radiation Oncology, Medical College of Virginia Hospitals, Virginia Commonwealth

More information

ELEG 479 Lecture #6. Mark Mirotznik, Ph.D. Associate Professor The University of Delaware

ELEG 479 Lecture #6. Mark Mirotznik, Ph.D. Associate Professor The University of Delaware ELEG 479 Lecture #6 Mark Mirotznik, Ph.D. Associate Professor The University of Delaware Summary of Last Lecture X-ray Physics What are X-rays and when are they useful for medical imaging? How are X-rays

More information

Study of the phase contrast for the characterization of the surface of microshell

Study of the phase contrast for the characterization of the surface of microshell 19 th World Conference on Non-Destructive Testing 2016 Study of the phase contrast for the characterization of the surface of microshell Alexandre CHOUX 1,*, Vincent DUTTO 1, Eric BUSVELLE 2, Jean-Paul

More information

HYDROBETA: A NEW INSTRUMENT FOR MEASURING IN-SITU PROFILES OF THE VOLUME SCATTERING FUNCTION FROM 10 TO 170 DEGREES

HYDROBETA: A NEW INSTRUMENT FOR MEASURING IN-SITU PROFILES OF THE VOLUME SCATTERING FUNCTION FROM 10 TO 170 DEGREES HYDROBETA: A NEW INSTRUMENT FOR MEASURING IN-SITU PROFILES OF THE VOLUME SCATTERING FUNCTION FROM 10 TO 170 DEGREES David R. Dana, Robert A. Maffione Hydro-Optics, Biology and Instrumentation Laboratories,

More information

The statistics of ocean-acoustic ambient noise

The statistics of ocean-acoustic ambient noise The statistics of ocean-acoustic ambient noise Nicholas C. Makris Naval Research Laboratory, Washington, D.C. 0375, USA Abstract With the assumption that the ocean-acoustic ambient noise field is a random

More information

Influence of Collimation and Detector Length on CT Exposures Measured in a 60 cm Long Body Phantom. Victor J Weir ABSTRACT

Influence of Collimation and Detector Length on CT Exposures Measured in a 60 cm Long Body Phantom. Victor J Weir ABSTRACT Influence of Collimation and Detector Length on CT Exposures Measured in a 60 cm Long Body Phantom Victor J Weir ABSTRACT Background: Approach to equilibrium functions are typically presented as growth

More information

Uncertainty due to Finite Resolution Measurements

Uncertainty due to Finite Resolution Measurements Uncertainty due to Finite Resolution Measurements S.D. Phillips, B. Tolman, T.W. Estler National Institute of Standards and Technology Gaithersburg, MD 899 Steven.Phillips@NIST.gov Abstract We investigate

More information

CT-PET calibration : physical principles and operating procedures F.Bonutti. Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital.

CT-PET calibration : physical principles and operating procedures F.Bonutti. Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital. CT-PET calibration : physical principles and operating procedures Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital Topics Introduction to PET physics F-18 production β + decay and annichilation

More information

An experimental study of dual-energy CT imaging using synchrotron radiation

An experimental study of dual-energy CT imaging using synchrotron radiation Nuclear Science and Techniques 24 (2013) 020102 An experimental study of dual-energy CT imaging using synchrotron radiation HAO Jia 1,2 ZHANG Li 1,2 XING Yuxiang 1,2 KANG Kejun 1,2 1 Department of Engineering

More information

ADAPTIVE ALGEBRAIC RECONSTRUCTION OF CT IMAGES IN COMPARISON TO FREQUENCY DOMAIN METHODS. Bernhard Bundschuh Universität-GH-Siegen, Germany

ADAPTIVE ALGEBRAIC RECONSTRUCTION OF CT IMAGES IN COMPARISON TO FREQUENCY DOMAIN METHODS. Bernhard Bundschuh Universität-GH-Siegen, Germany ADAPIVE ALGEBRAIC RECONSRUCION OF C IMAGES IN COMPARISON O FREQUENCY DOMAIN MEHODS Bernhard Bundschuh Universität-GH-Siegen German Abstract his paper presents a new algebraic algorithm for the reconstruction

More information

NIH Public Access Author Manuscript Proc Soc Photo Opt Instrum Eng. Author manuscript; available in PMC 2013 December 17.

NIH Public Access Author Manuscript Proc Soc Photo Opt Instrum Eng. Author manuscript; available in PMC 2013 December 17. NIH Public Access Author Manuscript Published in final edited form as: Proc Soc Photo Opt Instrum Eng. 2008 March 18; 6913:. doi:10.1117/12.769604. Tomographic Reconstruction of Band-limited Hermite Expansions

More information

MAP Reconstruction From Spatially Correlated PET Data

MAP Reconstruction From Spatially Correlated PET Data IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL 50, NO 5, OCTOBER 2003 1445 MAP Reconstruction From Spatially Correlated PET Data Adam Alessio, Student Member, IEEE, Ken Sauer, Member, IEEE, and Charles A Bouman,

More information

Minimum detection window and inter-helix PI-line with triple-source helical cone-beam scanning

Minimum detection window and inter-helix PI-line with triple-source helical cone-beam scanning Journal of X-Ray Science and Technology 14 2006) 95 107 95 IOS Press Minimum detection window and inter-helix PI-line with triple-source helical cone-beam scanning Jun Zhao a,, Ming Jiang b, Tiange Zhuang

More information

Imaging of built-in electric field at a p-n junction by scanning transmission electron microscopy

Imaging of built-in electric field at a p-n junction by scanning transmission electron microscopy Imaging of built-in electric field at a p-n junction by scanning transmission electron microscopy N. Shibata, S.D. Findlay, H. Sasaki, T. Matsumoto, H. Sawada, Y. Kohno, S. Otomo, R. Minato and Y. Ikuhara

More information

A. I, II, and III B. I C. I and II D. II and III E. I and III

A. I, II, and III B. I C. I and II D. II and III E. I and III BioE 1330 - Review Chapters 7, 8, and 9 (Nuclear Medicine) 9/27/2018 Instructions: On the Answer Sheet, enter your 2-digit ID number (with a leading 0 if needed) in the boxes of the ID section. Fill in

More information

PRISMATIC COVERS FOR BOOSTING THE EFFICIENCY OF HIGH-CONCENTRATION PV SYSTEMS

PRISMATIC COVERS FOR BOOSTING THE EFFICIENCY OF HIGH-CONCENTRATION PV SYSTEMS PRISMATIC COVERS FOR BOOSTING THE EFFICIENCY OF HIGH-CONCENTRATION PV SYSTEMS Andreea Boca, Kenneth M. Edmondson, and Richard R. King Spectrolab, Inc., 12500 Gladstone Ave., Sylmar, CA 91342 U.S.A. ABSTRACT

More information

Scan Time Optimization for Post-injection PET Scans

Scan Time Optimization for Post-injection PET Scans Presented at 998 IEEE Nuc. Sci. Symp. Med. Im. Conf. Scan Time Optimization for Post-injection PET Scans Hakan Erdoğan Jeffrey A. Fessler 445 EECS Bldg., University of Michigan, Ann Arbor, MI 4809-222

More information

Let us consider a typical Michelson interferometer, where a broadband source is used for illumination (Fig. 1a).

Let us consider a typical Michelson interferometer, where a broadband source is used for illumination (Fig. 1a). 7.1. Low-Coherence Interferometry (LCI) Let us consider a typical Michelson interferometer, where a broadband source is used for illumination (Fig. 1a). The light is split by the beam splitter (BS) and

More information

ULTRAFAST LASER PULSE TRAIN RADIATION TRANSFER IN A SCATTERING-ABSORBING 3D MEDIUM WITH AN INHOMOGENEITY

ULTRAFAST LASER PULSE TRAIN RADIATION TRANSFER IN A SCATTERING-ABSORBING 3D MEDIUM WITH AN INHOMOGENEITY Heat Transfer Research 46(9), 861 879 (2015) ULTRAFAST LASER PULSE TRAIN RADIATION TRANSFER IN A SCATTERING-ABSORBING 3D MEDIUM WITH AN INHOMOGENEITY Masato Akamatsu 1,* & Zhixiong Guo 2 1 Graduate School

More information

Persistence in the WFC3 IR Detector: Spatial Variations

Persistence in the WFC3 IR Detector: Spatial Variations SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA WFC3 Instrument Science Report 2015-16 Persistence in the WFC3 IR Detector: Spatial Variations Knox S. Long, Sylvia M. Baggett, & John W. MacKenty

More information

This is a repository copy of TACTIC : The TRIUMF Annular Chamber for Tracking and Identification of Charged particles.

This is a repository copy of TACTIC : The TRIUMF Annular Chamber for Tracking and Identification of Charged particles. This is a repository copy of TACTIC : The TRIUMF Annular Chamber for Tracking and Identification of Charged particles. White Rose Research Online URL for this paper: http://eprints.whiterose.ac.uk/128325/

More information

Optimal resolutions for optical and NIR spectroscopy S. Villanueva Jr.* a, D.L. DePoy a, J. L. Marshall a

Optimal resolutions for optical and NIR spectroscopy S. Villanueva Jr.* a, D.L. DePoy a, J. L. Marshall a Optimal resolutions for optical and NIR spectroscopy S. Villanueva Jr.* a, D.L. DePoy a, J. L. Marshall a a Department of Physics and Astronomy, Texas A&M University, 4242 TAMU, College Station, TX, USA

More information

Medical Visualization - Tensor Visualization. J.-Prof. Dr. Kai Lawonn

Medical Visualization - Tensor Visualization. J.-Prof. Dr. Kai Lawonn Medical Visualization - Tensor Visualization J.-Prof. Dr. Kai Lawonn Lecture is partially based on the lecture by Prof. Thomas Schultz 2 What is a Tensor? A tensor is a multilinear transformation that

More information

Response Characteristics of a Large-Area Ion Chamber with Various Radiotherapy Beams Makan Farrokhkish, B.Sc.

Response Characteristics of a Large-Area Ion Chamber with Various Radiotherapy Beams Makan Farrokhkish, B.Sc. Response Characteristics of a Large-Area Ion Chamber with Various Radiotherapy Beams Makan Farrokhkish, B.Sc. Department of Radiation Physics, Radiation Medicine Program Princess Margaret Cancer Centre,

More information

Principal Components Analysis. Sargur Srihari University at Buffalo

Principal Components Analysis. Sargur Srihari University at Buffalo Principal Components Analysis Sargur Srihari University at Buffalo 1 Topics Projection Pursuit Methods Principal Components Examples of using PCA Graphical use of PCA Multidimensional Scaling Srihari 2

More information

3.5 Cavities Cavity modes and ABCD-matrix analysis 206 CHAPTER 3. ULTRASHORT SOURCES I - FUNDAMENTALS

3.5 Cavities Cavity modes and ABCD-matrix analysis 206 CHAPTER 3. ULTRASHORT SOURCES I - FUNDAMENTALS 206 CHAPTER 3. ULTRASHORT SOURCES I - FUNDAMENTALS which is a special case of Eq. (3.30. Note that this treatment of dispersion is equivalent to solving the differential equation (1.94 for an incremental

More information

The Use of Attenuation Standards for CT Scanning

The Use of Attenuation Standards for CT Scanning The Use of Attenuation Standards for CT Scanning by M. E. Coles and E. L. Muegge Mobil Exploration and Producing Technical Center Dallas, Texas and F. M. Auzerais and P. Frulla Schlumberger-Doll Research

More information

PET. Technical aspects

PET. Technical aspects PET Technical aspects 15 N 15 O Detector 1 β+ Detector 2 e- Evolution of PET Detectors CTI/Siemens 15 N 15 O Detector block 1 β+ Detector block 2 x e- x y y location line of response Constant fraction

More information

2. SPECTRAL ANALYSIS APPLIED TO STOCHASTIC PROCESSES

2. SPECTRAL ANALYSIS APPLIED TO STOCHASTIC PROCESSES 2. SPECTRAL ANALYSIS APPLIED TO STOCHASTIC PROCESSES 2.0 THEOREM OF WIENER- KHINTCHINE An important technique in the study of deterministic signals consists in using harmonic functions to gain the spectral

More information

Objectives. Tolerance Limits and Action Levels for IMRT QA. The Overall Process of IMRT Planning and Delivery. Chain of IMRT Process

Objectives. Tolerance Limits and Action Levels for IMRT QA. The Overall Process of IMRT Planning and Delivery. Chain of IMRT Process Tolerance Limits and Action Levels for IMRT QA Jatinder R Palta, Ph.D Siyong Kim, Ph.D. Department of Radiation Oncology University of Florida Gainesville, Florida Objectives Describe the uncertainties

More information

Part III Minor Option in Medical Physics 2018 Examples Sheet

Part III Minor Option in Medical Physics 2018 Examples Sheet Part III Minor Option in Medical Physics 2018 Examples Sheet Any errors or comments should be addressed sent to: seb53@cam.ac.uk URLs that may be useful: Stanford Event Generation Simulator: http://tinyurl.com/pkg476r

More information

What is Image Deblurring?

What is Image Deblurring? What is Image Deblurring? When we use a camera, we want the recorded image to be a faithful representation of the scene that we see but every image is more or less blurry, depending on the circumstances.

More information

Optimized Gradient Waveforms for Spiral Scanning

Optimized Gradient Waveforms for Spiral Scanning Optimized Gradient Waveforms for Spiral Scanning Kevin F. King, Thomas K. F. Foo, Carl R. Crawford Spiral scanning gradient waveforms can be optimized with respect to blurring from off-resonance effects

More information

Field Tests of elongated Sodium LGS wave-front sensing for the E-ELT

Field Tests of elongated Sodium LGS wave-front sensing for the E-ELT Florence, Italy. May 2013 ISBN: 978-88-908876-0-4 DOI: 10.12839/AO4ELT3.13437 Field Tests of elongated Sodium LGS wave-front sensing for the E-ELT Gérard Rousset 1a, Damien Gratadour 1, TIm J. Morris 2,

More information

CHAPTER 2 RANDOM PROCESSES IN DISCRETE TIME

CHAPTER 2 RANDOM PROCESSES IN DISCRETE TIME CHAPTER 2 RANDOM PROCESSES IN DISCRETE TIME Shri Mata Vaishno Devi University, (SMVDU), 2013 Page 13 CHAPTER 2 RANDOM PROCESSES IN DISCRETE TIME When characterizing or modeling a random variable, estimates

More information

Biomedical Imaging. X ray imaging. Patrícia Figueiredo IST

Biomedical Imaging. X ray imaging. Patrícia Figueiredo IST Biomedical Imaging X ray imaging Patrícia Figueiredo IST 2013-2014 Overview Production of X rays Interaction of electrons with matter X ray spectrum X ray tube Interaction of X rays with matter Photoelectric

More information

arxiv: v1 [physics.class-ph] 10 Feb 2009

arxiv: v1 [physics.class-ph] 10 Feb 2009 Ground-Plane Quasi-Cloaking for Free Space Efthymios Kallos, Christos Argyropoulos, and Yang Hao School of Electronic Engineering and Computer Science, Queen Mary University of London, Mile End Road, London,

More information

gives rise to multitude of four-wave-mixing phenomena which are of great

gives rise to multitude of four-wave-mixing phenomena which are of great Module 4 : Third order nonlinear optical processes Lecture 26 : Third-order nonlinearity measurement techniques: Z-Scan Objectives In this lecture you will learn the following Theory of Z-scan technique

More information

429 LIGHT DIFFRACTION MEASUREMENT OF PARTICLE SIZE

429 LIGHT DIFFRACTION MEASUREMENT OF PARTICLE SIZE Search USP29 429 LIGHT DIFFRACTION MEASUREMENT OF PARTICLE SIZE Light diffraction is one of the most widely used techniques for measuring the size of a wide range of particles from very fine to very coarse.

More information

Analysis of Communication Systems Using Iterative Methods Based on Banach s Contraction Principle

Analysis of Communication Systems Using Iterative Methods Based on Banach s Contraction Principle Analysis of Communication Systems Using Iterative Methods Based on Banach s Contraction Principle H. Azari Soufiani, M. J. Saberian, M. A. Akhaee, R. Nasiri Mahallati, F. Marvasti Multimedia Signal, Sound

More information

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research Prompt gamma measurements for the verification of dose deposition in proton therapy Two Proton Beam Facilities for Therapy and Research Ion Beam Facilities in Korea 1. Proton therapy facility at National

More information