Response Characteristics of a Large-Area Ion Chamber with Various Radiotherapy Beams Makan Farrokhkish, B.Sc.
|
|
- Clementine Welch
- 5 years ago
- Views:
Transcription
1 Response Characteristics of a Large-Area Ion Chamber with Various Radiotherapy Beams Makan Farrokhkish, B.Sc. Department of Radiation Physics, Radiation Medicine Program Princess Margaret Cancer Centre, Toronto
2
3 S M Dose Area Photon Source Collimating elements Large Ion Chamber
4 Factors that influence IQM signals Linac Head Structure Beam Energy Beam Calibration
5 Varian Millennium Collimator Elekta MLCi2 Elekta Agility
6 IQM Response Characteristics With Different Radiotherapy Beam IQM Intrinsic and effective spatial sensitivity Response as a function of field size Response as a function of beam energy IQM response as a function of dose rates
7 Studied Linac Models TrueBeam (Varian Medical Systems, Palo Alto, CA). Millennium MLC. 6 MV, 10 MV, 6 FFF, 10 FFF Clinac ix (Varian Medical Systems, Palo Alto, CA). Millennium MLC. 6 MV, 18 MV Infinity ( Elekta AB, Stockholm, Sweden). Agility Collimator. 6MV, 18 MV Infinity ( Elekta AB, Stockholm, Sweden). MLCi2 Collimator. 6 MV, 18 MV
8 Objectives IQM Intrinsic and effective spatial sensitivity
9 -500 V 0 V
10 Normalized Signal Central Gradient Profile ( Normalized To Centre) Loss of lateral 0.8 Electron 0.6 Equilibrium Off Axis Distance (cm)
11 Normalized Signal Canteral None-Gradient Profile ( Normalized To Centre) Off Axis Distance (cm)
12 Normalized IQM Signal IQM Intrinsic Response (Gradient Direction) 1.60 TB 6X TB 6FFF 10 FFF 6X Agility Off Axis Distance In Transverse Plane (cm)
13 Normalized IQM Signal IQM Intrinsic Response (None Gradient Direction) 1.05 TB 6X TB 6 FFF 10 FFF 6X Agility Off Axis Distance In Radial Plane (cm)
14 IQM Signal Normalized to Centre Effective IQM Spatial Sensitivity for a 3X3 cm2 field True Beam 6 MV True Beam 10 MV Int Gradient Slope R Value Int Gradient True Beam 6 MV True Beam 10 MV Off axis field shift (cm)
15 IQM Signal Normalized to Centre Effective IQM Spatial Sensitivity for a 3X3 cm2 field MLCi2 6 MV MLCi2 18 MV Int Gradient Slope R Value Int Gradient MLCi2 6 MV MLCi2 18 MV Off axis field shift (cm)
16 IQM Signal Normalized to Centre Effective IQM Spatial Sensitivity for a 3X3 cm2 field Agility 6 MV Agility 18 MV Int Gradient Slope R Value Int Gradient Agility 6 MV Agility 18 MV Off axis field shift (cm)
17 IQM Signal Normalized to Centre 1.20 Effective IQM Spatial Sensitivity for a 3X3 cm2 field True Beam 6FFF True Beam 10FFF Int Gradient Off axis field shift (cm)
18 IQM Response Characteristics Response as a function of field size
19 Normalized IQM REF Signal REF IQM REF Signal = C x Side of square (cm)
20 Normalized IQM REF Signal REF Side of square (cm) Deviations from REF caused by: Difference in LINAC output with field size Difference in beam flatness and symmetry Difference in beam transmission through collimating elements
21 IQM Signal Normalized to 10X Signal % Diff With respect to REF Side of Square (cm) True Beam 6 MV True Beam 6 MV ix 6 MV REF ix 6 MV Side of square (cm)
22 IQM Signal Normalized to 10X Signal % Diff With respect to REF Side of Square (cm) True Beam 6 MV True Beam 6 MV ix 6 MV REF ix 6 MV Side of square (cm)
23 IQM Response Characteristics Response as a function of beam energy
24 IQM signal normalized to 6 MV 1.4 True Beam 10 MV Agility 18 MV MLCi2 18 MV ix 18MV Side of square (cm)
25 IQM signal normalized to 6 MV FFF 1.2 True Beam 10FFF 6 FFF 10 FFF Side of square (cm)
26 IQM Response Characteristics IQM response as a function of dose rates
27
28 IQM Dose Rate Dependency (Varian TB) Signal Normalized To 600 MU / Min For 6 MV,10 MV, 6 FFF And Normalized to 800 MU / Min For 10 FFF Dose Rate 6 MV IQM 10 MV IQM 6 FFF IQM 10 FFF IQM
29 Summary Only minor deviations in IQM intrinsic spatial sensitivity was observed across different energies and platforms The effective spatial sensitivity is affected by beam profile parameters and beam attenuation The signal for square field size depends on the design of the LINAC head components The IQM energy discrimination varies with field size, design of the LINAC head components, and beam profile IQM system exhibits negligible dose rate dependency
30 Acknowledgments Dr. Mohammad Islam Dr. Robert Heaton Bern Norrlinger Dr. David Jaffray Yinkun Wang Andrew Jung Graham Wilson Dr. Ivan Yeung Marlyn McIntosh Renata Chmielewski Thank You
A comparison of methods for monitoring photon beam energy constancy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 A comparison of methods for monitoring photon beam energy constancy Song Gao, 1a Peter A. Balter, 1 Mark Rose, 2 and William E. Simon
More informationNeutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 3, SUMMER 2003 Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators David S. Followill,*
More information8/4/2016. Determination of small field output factors, advantages and limitations of Monte Carlo simulations
etermination of small field output factors, advantages and limitations of onte Carlo simulations AAP-APR-SEF Joint Scientific Symposium Josep Puxeu Vaqué Institut Català d Oncologia ICO Edinburgh Cancer
More informationfactors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons
Calorimetric determination of k Q factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons PTB s water calorimeter in front of Elekta Precise medical linac Calorimetric determination
More information1 Introduction. A Monte Carlo study
Current Directions in Biomedical Engineering 2017; 3(2): 281 285 Sebastian Richter*, Stefan Pojtinger, David Mönnich, Oliver S. Dohm, and Daniela Thorwarth Influence of a transverse magnetic field on the
More information) for Varian TrueBeam high-dose-rate therapy beams
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 Ion recombination correction factors (P ion ) for Varian TrueBeam high-dose-rate therapy beams Stephen F. Kry, 1a Richard Popple,
More informationCommissioning measurements for photon beam data on three TrueBeam linear accelerators, and comparison with Trilogy and Clinac 2100 linear accelerators
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 1, 2013 Commissioning measurements for photon beam data on three TrueBeam linear accelerators, and comparison with Trilogy and Clinac 2100
More informationImplementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy
Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy M. Saiful Huq, PhD, FAAPM, FInstP Dept. of Radiation Oncology, University of
More informationMonitor backscatter factors for the Varian 21EX and TrueBeam linear accelerators: measurements and Monte Carlo modelling.
Monitor backscatter factors for the Varian 21EX and TrueBeam linear accelerators: measurements and Monte Carlo modelling. 5 Sergei Zavgorodni 1,2, Eyad Alhakeem 2,1 and Reid Townson 2,1 1 Department of
More informationThe Radiological Physics Center s standard dataset for small field size output factors
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 The Radiological Physics Center s standard dataset for small field size output factors David S. Followill, 1a Stephen F. Kry, 1 Lihong
More informationComparison of Primary Doses Obtained in Three 6 MV Photon Beams Using a Small Attenuator
Comparison of Primary Doses Obtained in Three 6 MV Photon Beams Using a Small Attenuator Christoph Trauernicht Groote Schuur Hospital & University of Cape Town Method is based on: Background A method of
More informationEstimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom
Progress in NUCLEAR SCIENCE and TECHNOLOGY, Vol. 2, pp.83-87 (2) ARTICLE Estimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom Toshioh FUJIBUCHI,2,*, Satoshi
More informationBattling Maxwell s Equations Physics Challenges and Solutions for Hybrid MRI Systems
Battling Maxwell s Equations Physics Challenges and Solutions for Hybrid MRI Systems SYDNEY MEDICAL SCHOOL PAUL KEALL, BRENDAN WHELAN and BRAD OBORN Representing The Australian MRI-Linac Program Team Patents:
More informationTime-resolved beam symmetry measurement for VMAT commissioning and quality assurance
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Time-resolved beam symmetry measurement for VMAT commissioning and quality assurance Michael P. Barnes 1,2 a and Peter B. Greer 1,3
More informationVisual Analysis of the Daily QA Results of Photon and Electron Beams of a Trilogy Linac over a Five-Year Period
International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2015, 4, 290-299 Published Online November 2015 in SciRes. http://www.scirp.org/journal/ijmpcero http://dx.doi.org/10.4236/ijmpcero.2015.44035
More informationCalculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials*
SLAC-PUB-70 Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials* J. C. Liu, K. R. Kase, X. S. Mao, W. R. Nelson, J. H. Kleck, and S. Johnson ) Stanford Linear
More informationLinac activation of radioisotopes and underground gammaspectrometric analyses
Linac activation of radioisotopes and underground gammaspectrometric analyses Patrick WEBER 1*, Jean-Luc VUILLEUMIER 2, Geoffroy GUIBERT 1, Claire TAMBURELLA 1 1 Hôpital Neuchâtelois, Service de Radio-oncologie,
More informationPhoton-beams monitor-unit calculations
Photon-beams monitor-unit calculations Narayan Sahoo March 24, 2011 The materials included in this lecture notes are from previous lecture notes for this course by Karl Prado, Ph.D. Introduction Standard
More informationACCELERATORS AND MEDICAL PHYSICS
ACCELERATORS AND MEDICAL PHYSICS 1 Ugo Amaldi University of Milano Bicocca and TERA Foundation EPFL 1-28.10.10 - U. Amaldi 1 Short history of Medical Physics with radiations (*) In physics radiation is
More informationMeasurements and comparisons for data of small beams of linear accelerators
[Chinese Journal of Cancer 28:3, 272-276; March 2009]; 2009 Landes Bioscience Methods and Technology Measurements and comparisons for data of small beams of linear accelerators Li Chen, 1 Li-Xin Chen,
More informationInternational Journal of Scientific & Engineering Research, Volume 7, Issue 2, February-2016 ISSN
1465 International Journal of Scientific & Engineering Research, Volume 7, Issue, February-016 ISSN 9-5518 Investigation of Medical Neutron Radiation Doses Using Two types of bubble Detectors Hsin-Ell
More informationObjectives. Tolerance Limits and Action Levels for IMRT QA. The Overall Process of IMRT Planning and Delivery. Chain of IMRT Process
Tolerance Limits and Action Levels for IMRT QA Jatinder R Palta, Ph.D Siyong Kim, Ph.D. Department of Radiation Oncology University of Florida Gainesville, Florida Objectives Describe the uncertainties
More informationSimulation Modeling in Dosimetry
Simulation Modeling in Dosimetry Aleksei Zhdanov Ural Federal University named after the first President of Russia B. N. Yeltsin, Yekaterinburg, Russian Federation jjj1994@yandex.ru Leonid Dorosinskiy
More informationIRPA 2 nd European Congress, Paris, May Session: RP in Medicine - Workers. Activation Products In Medical Linear Accelerators
IRPA 2 nd European Congress, Paris, May 15-19 2006 Session: RP in Medicine - Workers Activation Products In Medical Linear Accelerators FISCHER Helmut W.*, TABOT Ben*,+, POPPE, Björn + * University of
More informationRadiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp Advance Access publication 11 May 2007
Radiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp. 229 233 Advance Access publication 11 May 2007 doi:10.1093/rpd/ncm047 CHARACTERIZATION AND UTILIZATION OF A BONNER SPHERE SET BASED ON GOLD
More informationEvaluation of different methods for determining the magnitude of initial recombination in ionization chambers
Evaluation of different methods for determining the magnitude of initial recombination in ionization chambers Klaus Derikum Physikalisch-Technische undesanstalt undesallee 100, 38116 raunschweig, Germany
More informationQuality-Assurance Check of Collimator and Phantom- Scatter Factors
Quality-Assurance Check of Collimator and Phantom- Scatter Factors Ramesh C. Tailor, David S. Followill, Nadia Hernandez, Timothy S. Zhu, and Geoffrey S. Ibbott. UT MD Anderson Cancer Center, Houston TX.
More informationA Thesis. entitled. Micro Ionization Chamber. Vidheesha Arora. Master of Science Degree in. Biomedical Sciences: Medical Physics
A Thesis entitled An Investigation of the Polarity Effects in Small Field Based on the Orientation of the Micro Ionization Chamber by Vidheesha Arora Submitted to the Graduate Faculty as partial fulfillment
More informationVariations in daily quality assurance dosimetry from device levelling, feet position and backscatter material
Australas Phys Eng Sci Med (2012) 35:485 489 DOI 10.1007/s13246-012-0169-6 SCIENTIFIC NOTE Variations in daily quality assurance dosimetry from device levelling, feet position and backscatter material
More informationThe Influence of Brass Compensator Thickness and Field Size on Neutron Contamination Spectrum in 18MV Elekta SL 75/25 Medical Linear Accelerator
The Influence of Brass Compensator Thickness and Field Size on Neutron Contamination Spectrum in 18MV Elekta SL 75/25 Medical Linear Accelerator with and without Flattening Filter: A Monte Carlo Study
More informationDOI /
Provided by the author(s) and NUI Galway in accordance with publisher policies. Please cite the published version when available. Title Accounting for the fringe magnetic field from the bending magnet
More informationDiamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Diamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams Ramamoorthy Ravichandran,
More informationMichael Dunn Nuclear Data Group Leader Nuclear Science & Technology Division Medical Physics Working Group Meeting October 26, 2005
Nuclear Data Michael Dunn Nuclear Data Group Leader Nuclear Science & Technology Division Medical Physics Working Group Meeting October 26, 2005 ORELA LANSCE 0.1 00 Data Analyses ORELA data 0.0 75 Basic
More informationElectron therapy Class 2: Review questions
Electron therapy Class 2: Review questions 1 Raphex Question: T63, 2002 In what situation is electron backscatter likely to be a problem? A. Using 1cm of tissue equivalent bolus on the skin. B. Using a
More informationComparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification.
Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification. Antonio Lopez Medina, Antonio Teijeiro, Daniela Medal, Francisco Salvador, Julio Vazquez, Manuel
More informationDosimetric Properties of the Field Sizes of 12MV Photon Beams: A Monte Carlo Study
Journal of Nuclear and Particle Physics 2015, 5(3): 52-57 DOI: 10.5923/j.jnpp.20150503.02 Dosimetric Properties of the Field Sizes of 12MV Photon Beams: A Monte Carlo Study Maged Mohammed 1,2,*, E. Chakir
More informationMonte Carlo simulation of the photoneutron field in linac radiotherapy treatments with different collimation systems
INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 49 (2004) 571 582 PHYSICS IN MEDICINE AND BIOLOGY PII: S0031-9155(04)69196-2 Monte Carlo simulation of the photoneutron field in linac radiotherapy treatments
More informationEstimation of neutron and gamma radiation doses inside the concrete shield wall for 10 and 15 MV medical linear accelerators
DOI: 10.15669/pnst.4.280 Progress in Nuclear Science and Technology Volume 4 (2014) pp. 280-284 ARTICLE Estimation of neutron and gamma radiation doses inside the concrete shield wall for 10 and medical
More informationNEUTOR: Neutrons Monitor for. Project financed by: Radiotherapy
NEUTOR: Neutrons Monitor for Project financed by: Radiotherapy PURPOSES OF THE PROJECT To develop a Neutrons monitor for its application in radiotherapy, to control, in real time, the dose of neutrons
More informationBEAMnrc: a code to simulate radiotherapy external beam sources
BEAMnrc: a code to simulate radiotherapy external beam sources D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University Ottawa, Canada http://www.physics.carleton.ca/~drogers
More information1.1.4 What are the differences between the Varian 600C and the Siemens KD2?
1 Radiotherapy 1.1 Linear Accelerators 1.1.1 Sketch the main components of a linear accelerator. 1.1.2 Describe the theory of beam production and the part each component plays in the production of a clincally
More informationNeutron Fluence and Energy Spectra Around the Varian Clinac 2lOOC/23OOC Medical Accelerator
SLAC-PUB-7 190 June 1996 Neutron Fluence and Energy Spectra Around the Varian Clinac 2lOOC/23OOC Medical Accelerator K. R. Kase, X. S. Mao, W. R. Nelson, J. C. Liu Stanford Linear Accelerator Center, Stanford
More informationEvaluation of an electron Monte Carlo dose calculation algorithm for electron beams
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 3, SUMMER 2008 Evaluation of an electron Monte Carlo dose calculation algorithm for electron beams Ye Angela Hu, 1,a Haijun Song, 2 Zhe Chen,
More informationA Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators
A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators N. Golnik 1, P. Kamiński 1, M. Zielczyński 2 1 Institute of Precision and Biomedical Engineering,
More informationPhysics. Sunday, March 4, :30 a.m. 10:00 a.m.
Physics Sunday, March 4, 2018 9:30 a.m. 10:00 a.m. Social Q&A Use your phone, tablet, or laptop to Submit questions to speakers and moderators Answer interactive questions / audience response polls astro.org/refreshersocialqa
More information(INCLUDING THIS FRONT PAGE)
I'IFIITIIBIFI UNIVERSITY OF SCIEI'ICE RITD TECHNOLOGY FACULTY OF HEALTH AND APPLIED SCIENCES DEPARTMENT OF NATURAL AND APPLIED SCIENCES QUALIFICATION: BACHELOR OF SCIENCE (MAJOR AND MINOR) QUALIFICATION
More informationSecondary neutron spectra from modern Varian, Siemens, and Elekta linacs with multileaf collimators
Secondary neutron spectra from modern Varian, Siemens, and Elekta linacs with multileaf collimators Rebecca M. Howell a and Stephen F. Kry Department of Radiation Physics, The University of Texas M. D.
More informationProgress in Nuclear Science and Technology, Volume 6,
DOI: 1.15669/pnst.6 Progress in Nuclear Science and Technology Volume 6 (19) pp. 1-16 ARTICLE A study on calculation method of duct streaming from medical linac rooms Takuma Noto * Kazuaki Kosako and Takashi
More informationGeneral characteristics of radiation dosimeters
General characteristics of radiation dosimeters and a terminology to describe them D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers
More informationMedical Linac. Block diagram. Electron source. Bending magnet. Accelerating structure. Klystron or magnetron. Pulse modulator.
Block diagram Medical Linac Electron source Bending magnet Accelerating structure Pulse modulator Klystron or magnetron Treatment head 1 Medical Linac 2 Treatment Head 3 Important Accessories Wedges Dynamic
More informationClinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry
Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry TJ JORDAN Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE
More informationMagnetic removal of electron contamination in radiotherapy x-ray beams
University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2006 Magnetic removal of electron contamination in radiotherapy
More informationChapter 4. QUANTIFYING THE HAZARD II: DATA & ANALYSIS. The dose equivalents for spheres in air with 10 cm radius centred at a point in the
Chapter 4. QUANTIFYING THE HAZARD II: DATA & ANALYSIS Neutron Dose The dose equivalents for spheres in air with 10 cm radius centred at a point in the treatment room and at 9 points along the passage of
More informationCity University of Hong Kong
City University of Hong Kong Information on a Course offered by the Department of Physics and Materials Science with effect from Semester A in 2013 / 2014 Part I Course Title: Radiological Physics and
More informationVolume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved
Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah
More informationThe effect of dose calculation uncertainty on the evaluation of radiotherapy plans
The effect of dose calculation uncertainty on the evaluation of radiotherapy plans P. J. Keall a) and J. V. Siebers Department of Radiation Oncology, Medical College of Virginia Hospitals, Virginia Commonwealth
More informationMonte Carlo simulation with Geant4 for verification of rotational total skin electron therapy (TSET)
Monte Carlo simulation with Geant4 for verification of rotational total skin electron therapy (TSET) Christina Jarlskog Department of Radiation Physics Lund University, Malmö University Hospital (UMAS)
More informationEvaluation of the truebeam machine performance check (MPC): mechanical and collimation checks
Received: 5 December 2016 Revised: 30 January 2017 Accepted: 14 February 2017 DOI: 10.1002/acm2.12072 RADIATION ONCOLOGY PHYSICS Evaluation of the truebeam machine performance check (MPC): mechanical and
More informationRadiation protection issues in proton therapy
Protons IMRT Tony Lomax, Centre for Proton Radiotherapy, Paul Scherrer Institute, Switzerland Overview of presentation 1. Proton therapy: An overview 2. Radiation protection issues: Staff 3. Radiation
More informationLow energy linear acceleratordriven subcritical assembly
UNIVERSIDAD SIMÓN BOLÍVAR LABORATORIO DE FÍSICA NUCLEAR Low energy linear acceleratordriven subcritical assembly Eduardo D. Greaves 1 and Laszlo Sajo-Bohus 2 Universidad Simón Bolívar. Apartado 89000,
More informationThe neutron dose equivalent evaluation and shielding at the maze entrance of a Varian Clinac 23EX treatment room
This partial RIAS has been downloaded on 15 Apr 2019 The neutron dose equivalent evaluation and shielding at the maze entrance of a Varian Clinac 23EX treatment room Xudong Wang, Carlos Esquivel, Elena
More informationPoint dose measurements in VMAT: an investigation of detector choice and plan complexity.
Point dose measurements in VMAT: an investigation of detector choice and plan complexity. Daniel Phillip Goodwin Department of Physics and Astronomy University of Canterbury A thesis submitted in partial
More informationSmall Field Dosimetry and IAEA/AAPM Protocol
Small Field Dosimetry and IAEA/AAPM Protocol Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Emeritus Professor, Department of Medical Physics University of Wisconsin
More informationPHYS 5020 Computation and Image Processing
PHYS 5020 and Image Processing : Monte Carlo Thursday 2 August 2012 Monte Carlo (MC) is a numerical method that uses random sampling of probability distributions to simulate stochastic processes in nature,
More informationCHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR
ANALELE STIINTIFICE ALE UNIVERSITATII AL. I. CUZA IASI Tomul II, s. Biofizică, Fizică medicală şi Fizica mediului 2006 CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR Dan
More informationExperimental validation of the Eclipse AAA algorithm
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 8, NUMBER 2, SPRING 2007 Experimental validation of the Eclipse AAA algorithm Karen Breitman 1 3, Satyapal Rathee 4,5,a, Chris Newcomb 1,2, Brad Murray
More informationA Study on Effective Source-Skin Distance using Phantom in Electron Beam Therapy
Journal of Magnetics 19(1), 15-19 (2014) ISSN (Print) 1226-1750 ISSN (Online) 2233-6656 http://dx.doi.org/10.4283/jmag.2014.19.1.015 A Study on Effective Source-Skin Distance using Phantom in Electron
More informationScience, Ami, Ami-machi, Inashiki-gun, Ibaraki , Japan. Ibaraki , Japan. Japan. *Corresponding author:
Radiation Protection Dosimetry (2012), Vol. 148, No. 2, pp. 168 173 Advance Access publication 10 February 2011 doi:10.1093/rpd/ncr005 INDUCED RADIOACTIVE NUCLIDES OF 10-MeV RADIOTHERAPY ACCELERATORS DETECTED
More informationA Brief Introduction to Medical Imaging. Outline
A Brief Introduction to Medical Imaging Outline General Goals Linear Imaging Systems An Example, The Pin Hole Camera Radiations and Their Interactions with Matter Coherent vs. Incoherent Imaging Length
More informationAcquisition and analysis of megavoltage linac beam transmission data for direct verification of photon spectra models in a treatment planning system
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2014 Acquisition and analysis of megavoltage linac beam transmission data for direct verification of photon
More informationAnalysis of Image Noise in 3D Cone-Beam CT: Spatial and Fourier Domain Approaches under Conditions of Varying Stationarity
Analysis of Image Noise in 3D Cone-Beam CT: Spatial and Fourier Domain Approaches under Conditions of Varying Stationarity Angel R. Pineda *a, Jeffrey H. Siewerdsen b,c, Daniel J. Tward b a Department
More informationSurface Analysis - The Principal Techniques
Surface Analysis - The Principal Techniques Edited by John C. Vickerman Surface Analysis Research Centre, Department of Chemistry UMIST, Manchester, UK JOHN WILEY & SONS Chichester New York Weinheim Brisbane
More informationSensitivity of megavoltage photon beam Monte Carlo simulations to electron beam and other parameters
Sensitivity of megavoltage photon beam Monte Carlo simulations to electron beam and other parameters Daryoush Sheikh-Bagheri a) and D. W. O. Rogers b) Ionizing Radiation Standards, National Research Council
More informationComparison of two commercial detector arrays for IMRT quality assurance
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 2, SPRING 2009 Comparison of two commercial detector arrays for IMRT quality assurance Jonathan G. Li, 1,a Guanghua Yan, 1,2 and Chihray Liu
More informationComposite field dosimetry
Composite field dosimetry Hugo Bouchard, PhD, MCCPM Senior Research Scientist Radiation dosimetry group National Physical Laboratory May 2014 Overview 1. Introduction Dosimetry protocols IAEA formalism
More informationMonitor Unit Calculations for Photon and Electrons. AAMD Meeting Raleigh, NC October 3, John P. Gibbons Chief of Clinical Physics
Monitor Unit Calculations for Photon and Electrons AAMD Meeting Raleigh, NC October 3, 2014 John P. Gibbons Chief of Clinical Physics Outline I. TG71 Formation and Charge II. Photon Calculations III. Electron
More informationThe impact of detector volume on penumbral dose prediction in a 6 MV radiation therapy X-ray beam
University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2009 The impact of detector volume on penumbral dose prediction
More informationThe EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1
The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1 1 Medical Radiation Physics Group, University of Oldenburg and Pius
More informationMeasurement of Neutron Activation from a High Energy Varian Linear Accelerator
UNLV Theses, Dissertations, Professional Papers, and Capstones August 2016 Measurement of Neutron Activation from a High Energy Varian Linear Accelerator Timothy Thatcher University of Nevada, Las Vegas,
More informationABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *
Romanian Reports in Physics, Vol. 67, No. 3, P. 1152 1158, 2015 ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * ELENA STANCU 1,2, CATALIN VANCEA
More informationUpdating reference dosimetry a decade after TG-51
Updating reference dosimetry a decade after TG-51 Malcolm McEwen Ionizing Radiation Standards Institute for National Measurement Standards National Research Council, Canada CE Presentation at AAPM Annual
More informationI. INTRODUCTION EXPERIMENTAL
International Journal of Computational Engineering Research Vol, 04 Issue, 4 Simulation of Photon and Electron dose distributions by 5 code for the treatment area using the linear electron accelerator
More informationOverview and Status of the Austrian Particle Therapy Facility MedAustron. Peter Urschütz
Overview and Status of the Austrian Particle Therapy Facility MedAustron Peter Urschütz MedAustron Centre for ion beam therapy and non-clinical research Treatment of 1200 patients/year in full operation
More informationTowards efficient and accurate particle transport simulation in medical applications
Towards efficient and accurate particle transport simulation in medical applications L. Grzanka1,2, M. Kłodowska1, N. Mojżeszek1, N. Bassler3 1 Cyclotron Centre Bronowice, Institute of Nuclear Physics
More informationSmall Field Dosimetric Measurements with TLD-100, Alanine, and Ionization Chambers
Small Field Dosimetric Measurements with TLD-1, Alanine, and Ionization Chambers S. Junell a, L. DeWerd a, M. Saiful Huq b, J. Novotny Jr. b, M. uader b, M.F. Desrosiers c, G. Bednarz b a Department of
More informationDetermination of absorbed dose to water for 60Co by the scaling theorem. M. Boutillon and A.-M. Perroche (1)
Rapport BIPM92/1 Determination of absorbed dose to water for 60Co by the scaling theorem M. Boutillon and A.M. Perroche (1) Bureau International des Poids et Mesures, F92312 Sevres Cedex Abstract A simple
More informationMonte Carlo Analyses of X-Ray Absorption, Noise, and Detective Quantum Efficiency Considering Therapeutic X-Ray Spectrum in Portal Imaging Detector
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 48, NO. 4, AUGUST 2001 1423 Monte Carlo Analyses of X-Ray Absorption, Noise, and Detective Quantum Efficiency Considering Therapeutic X-Ray Spectrum in Portal
More informationInterpretation of in-air output ratio and its impact on Dose Calculation
Interpretation of in-air output ratio and its impact on Dose Calculation Timothy C Zhu 1 and Anders Ahnesjö 2,3 1 Dept. of Radiation Oncolog University of Pennsylvania, Philadelphia, PA, 2 Nucletron Scandinavia
More informationSimulations in Radiation Therapy
Simulations in Radiation Therapy D. Sarrut Directeur de recherche CNRS Université de Lyon, France CREATIS-CNRS ; IPNL-CNRS ; Centre Léon Bérard Numerical simulations and random processes treat cancer 2
More informationarxiv: v2 [physics.med-ph] 29 May 2015
The Proton Therapy Nozzles at Samsung Medical Center: A Monte Carlo Simulation Study using TOPAS Kwangzoo Chung, Jinsung Kim, Dae-Hyun Kim, Sunghwan Ahn, and Youngyih Han Department of Radiation Oncology,
More informationKeywords: mechanical property, polymer optical fibre, radiation, transmission.
5 th Australasian Congress on Applied Mechanics, ACAM 27 1-12 December 27, Brisbane, Australia Radiation damage to polymer optical fibres C. Yan 1, S.H. Law 2, N. Suchowerska 3,4 and S.H. Hong 5 1 School
More informationM d e i di l ca A pplilli t ca i ttions o f P arti ttic ti l P e h Physics Saverio Braccini INSEL
Medical la Applications of Particle Physics Saverio Braccini INSELSPITALSPITAL Department of Medical Radiation Physics University Hospital, Berne, Switzerland Rome - 14-15.06.07 - SB - 1/5 Saverio.Braccini@cern.ch
More informationNeutron Sources in the Varian Clinac 2100~2300C Medical Accelerator Calculated by the EGS4 Code
SLAC-PUB-7077 June 1996 Neutron Sources in the Varian Clinac 2100~2300C Medical Accelerator Calculated by the EGS4 Code X. S. Mao, K.R.Kase,J. C. Liu and W. R. Nelson Stanford Linear Accelerator Center
More informationRadiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe
Radiation Dosimetry Alun Beddoe Medical Physics University Hospital Birmingham NHS Trust Contents ABSOLUTE DOSIMETRY (CALIBRATION) Photon interactions (recap) Energy transfer and absorption Electron range
More informationThe volume effect of detectors in the dosimetry of small fields used in IMRT
The volume effect of detectors in the dosimetry of small fields used in IMRT Wolfram U. Laub a) Department of Medical Physics, University of Tübingen, Germany Tony Wong b) Department of Radiation Oncology,
More informationCurrent and Recent ICRU Activities in Radiation Protection Dosimetry and Measurements
Current and Recent ICRU Activities in Radiation Protection Dosimetry and Measurements Hans-Georg Menzel International Commission on Radiation Units and Measurements (ICRU) The principal objective of ICRU
More informationSPARCLAB. Source For Plasma Accelerators and Radiation Compton with Laser And Beam
SPARCLAB Source For Plasma Accelerators and Radiation Compton with Laser And Beam EMITTANCE X X X X X X X X Introduction to SPARC_LAB 2 BRIGHTNESS (electrons) B n 2I nx ny A m 2 rad 2 The current can be
More informationM [scale units/s] of the system
APPENDIX TO IAEA CALIBRATION CERTIFICATE RADIATION PROTECTION IONIZATION CHAMBER CALIBRATION PROCEDURES AT THE IAEA DOSIMETRY LABORATORY 1. INTRODUCTION 1.1 General Ionization chambers and electrometers
More information8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required
UPDATING TG-51 When will it end? Malcolm McEwen Ionizing Radiation Standards National Research Council, Canada AAPM Annual Meeting, Charlotte, 2012 What are these updates? Working Group review recommends:
More informationTERA CONTRIBUTIONS TO PARTNER
TERA CONTRIBUTIONS TO PARTNER Ugo Amaldi University of Milano Bicocca and TERA Foundation 1 CNAO status 2 The CNAO Foundation builds with INFN in Pavia the Centre designed by TERA on the basis of PIMMS.
More information