1 DOI: /pnst.6 Progress in Nuclear Science and Technology Volume 6 (19) pp ARTICLE A study on calculation method of duct streaming from medical linac rooms Takuma Noto * Kazuaki Kosako and Takashi Nakamura Shimizu Corporation Etchujima Koto-ku Tokyo Japan The leakage radiation dose from medical linac systems must be suppressed to prevent public exposure to radiation. Therefore leakage dose through ducts or sleeves installed in shielding walls need to be evaluated because radiation can pass through these without undergoing scattering or attenuation. However conventional methods for calculating radiation streaming have not been sufficiently evaluated for the leakage dose from the ducts in mazes. In this study the method provided by the Japan Radiological Society for neutrons and X-rays Nakamura and Uwamino s formula for neutrons the method described by the IAEA Safety Reports Series No.47 and the method given by McGinley for X-rays were evaluated by comparing the results with a Monte Carlo calculation using MCNP5 in calculation models. The rectangular ducts in these models were located in the maze near the entrance of linac room. Compared with MCNP5 most results for duct streaming obtained using conventional methods for neutrons were within a factor of whereas the results of methods using X-rays were more than a factor of. The results for maze streaming using Nakamura and Uwamino s formula were in good agreement and differed by not more than 16%. Furthermore the ratios of the X-ray dose at the duct entrance to that at the exit were calculated to investigate X-ray duct streaming with the results of MCNP5 and showed the relationships against the ratio of the opening area of the duct to square length. Keywords: duct streaming; maze streaming; X-ray; neutron; Monte Carlo; radiotherapy; shielding design 1. Introduction 1 The number of cancer patients around the world is increasing and medical linac systems are effective treatment tools for radiotherapy. When a medical linac is introduced at a facility the leakage radiation from the linac room must be suppressed to prevent public exposure to radiation. For example in Japan the leakage radiation dose must be less than 1.3 msv over a 3-month period at the boundary of the radiation controlled area. To suppress leakage of radiation of X-rays and neutrons to the outside of the room thick radiation shields such as concrete walls and iron plates are used and a maze is placed between the irradiation room and the entrance. Therefore methods for evaluating the leakage radiation dose are required to design appropriate shields. The leakage radiation dose through the shielding wall and maze streaming can be calculated numerically using the method described in IAEA Safety Reports Series No.47 [1]. The method considers operating conditions and room layout in the calculations. Conventional methods such as those provided by the Japan Radiological Society (JRS) [] McGinley [3] and Nakamura and Uwamino [4] can also be used to evaluate the leakage radiation dose. Heating ventilation and air conditioning ducts and *Corresponding author. t.noto@shimz.co.jp sleeves for cables installed through the shielding walls are also important structures to consider for the calculation of leakage radiation dose from a linac room. The leakage radiation doses through them must be evaluated because radiation can pass through these without undergoing scattering or attenuation which may result in relatively high doses. The Duct-III code [5] which is based on Shin s semi-empirical formula [6] is a calculation tool for radiation streaming and is often used to estimate duct streaming. However the methods can overestimate or underestimate the leakage radiation dose when the streaming path is complicated. In this case a Monte Carlo calculation such as MCNP5 [7] may provide calculations with better accuracy; however such methods require a much longer calculation time and skills. Although conventional streaming calculation methods are advantageous because their operation is easy and fast they have not been sufficiently evaluated for the leakage radiation dose from the duct in the maze. In this study conventional numerical calculation methods such as the method provided by JRS for calculating leakage dose of neutrons and X-rays Nakamura and Uwamino s formula for calculating leakage dose of neutrons the method described in IAEA Safety Reports Series No.47 and the McGinley method for calculating leakage dose of X-rays are evaluated by comparing the obtained results with MCNP5. 19 Atomic Energy Society of Japan. All rights reserved.

2 Progress in Nuclear Science and Technology Volume Calculation methods.1. Linac system and room layout A linac system with an energy of 1 MeV is assumed for the calculation model since this is the most common system in Japan. The irradiation conditions of the model are an accelerator workload of 6 Gy over 3 months at an isocenter that is 1 m away from the target in the linac head an absorbed dose to water of 6 Gy/min at the isocenter a field size of primary beam of.4.4 m at the isocenter and the ratio of the dose sourced from head leakage relative to the primary beam of.1 that is conservative upper limit of regulated value in ordinance for enforcement of the medical care act in Japan. The linac head is located parallel to the maze (Figure 1). The room height is 3.5 m and the beam is at a height of 1.95 m from the floor. A rectangular duct with opening areas of or.3.3 m and lengths of.7 1. or 1.3 m is installed on the maze wall near the entrance at the same height as the beam. The dose rates at the maze corner (A) the center of the maze on the duct axis (B) the duct entrance (C) for maze streaming and the duct exit (D) for duct streaming were calculated. Scattering radiation from the patient was ignored in this calculation. Figure 1. Study room layout and calculation points A B C and D... Neutron streaming..1 JRS method The equation for calculating the neutron dose rate E at a scattered point is given as E α1a1 (1) E 1.43 ( d d1 ) where E is neutron dose rate at a point 1. m away from the target and is calculated by multiplying the workload of the accelerator by the ratio of Sv/Gy for a 1 MeV linac with a copper target [3]. A 1 is scattering area d and d 1 are the distances from the target to scattering area and from the scattering area to calculation point respectively and 1.43 is a factor that accounts for the contribution of thermal and fast neutrons. The reflection coefficient α 1 can be calculated by the equation α.11( cosθ ) / 3 1 cosθ () where θ is the incident angle and θ is the reflection angle. For a multi-scattered path Eq. (1) may be modified by multiplying with a factor of α i A i /d i for leg i... Nakamura and Uwamino s formula This method evaluates the dose at the first leg in the maze (point A in Figure 1) with 1 α A E E + r r + 4s rs A L 1 L s (3) where E is the neutron dose rate at a point 1. m away from the target r is distance between the target and the evaluation point A is the total surface area of the irradiation room A is the surface area of the irradiation room seen directly from the evaluation point L 1 and L are the width and length of the irradiation room respectively and α is an adjustable parameter. E was calculated by MCNP5 at the isocenter. The parameter α was set to a value of.5 in this study by comparing with the result of MCNP5 whereas study [4] used a value of 4 by comparing with the results of DOT 3.5. The dose rate E i at the evaluation point for ith leg was calculated by the equation ai (4) Ei Ei 1 ri where r i is the distance of ith leg and a i is the half width of the maze or the duct. The evaluation points on the streaming path were located at a distance of a i away from the entrance of the maze or the duct..3. X-ray streaming.3.1 IAEA safety reports series No.47 and McGinley methods The X-ray streaming paths from the target are categorized as the primary beam scattered by the wall into the maze (E W ) the head leakage radiation scattered by the wall into the maze (E L ) and the head leakage radiation transmitted through the maze (E T ). The primary beam transmitted through the maze and photons generated by the interaction of neutrons are negligible for a linac system with an energy of 1 MeV. E W E L and E T are calculated using the equations WαW 1AW 1 (5) EW ( d ) W dw 1 WLα L1 AL 1 (6) E L ( d ) Ld L1 WLB E T (7) dt where W is the workload of the accelerator L is the ratio of the dose generated by head leakage at 1. m from the target to the dose at the isocenter α W1 and α L1 are reflection coefficients described in references [1 3] and d W d W1 d L d L1 and d T are the distances of the legs from the target or scattering points. The equation for calculating the barrier transmission factor B is

3 14 T. Noto et al. t TVL B 1 (8) where t is the wall thickness and TVL is the tenth value layer of the wall material. For multi-scattered paths Eqs.(5) and (6) may be modified by multiplying by a factor of (α i A i /d i ) for leg i which is similar to Eq.(1)..3. JRS method The streaming paths considered for E W E L and E T are similar to those described previously. E W and E L are calculated using Eqs. (5) and (6) by multiplying with a safety factor of and 1.43 Sv/Gy respectively with a constant reflection coefficient of.1. E T is calculated using Eq. (7). The equation for calculating B is t TVL B F 1 (9) where F is the correction factor described in reference []..4. Monte Carlo calculation MCNP5 was used to calculate the dose rate. The calculations were separated into two steps to distinguish radiation type and paths. Firstly spectra of the neutron primary X-ray and head leakage X-ray spectra shown in Figure were calculated using a simplified linac head model. A copper target with 15-mm thickness was irradiated with monoenergetic electrons with an energy of 1 MeV to generate X-rays and neutrons which were tallied at the isocenter. A tungsten collimator was located around the target to reduce the dose rate of head leakage to 1/1 of the primary X-ray dose rate. The head leakage X-ray was tallied at a point 1. m away from the target at 9 to the beam axis. In the second step transport calculations of neutrons primary X-rays head leakage X-rays and the head leakage X-rays transmitted through the maze wall without entering aperture of the maze were performed separately. The dose rates were normalized to the absorbed dose in water at the isocenter which is 6 Gy/min for primary X-rays. Figure 3 shows the dose map of the primary X-rays. The dose rates of the head leakage X-rays scattered on the wall were calculated by Figure. Spectra of primary X-rays head leakage X-rays and neutrons. Figure 3. Dose map of primary X-ray irradiation. subtracting the dose rate of the head leakage X-ray transmitted through the maze wall from that of the whole head leakage X-rays. 3. Results and discussion 3.1. Evaluation of conventional methods The results of the neutrons streaming through the maze are shown in Figure 4 (a) and the results of the neutrons streaming through the duct are shown in Figures 4 (b)-(d). Statistical errors of dose rates in the MCNP calculations were less than %. The duct streaming results were within a factor of from the MCNP5 results except at low dose rates or for an opening area of.1 m with duct lengths of 1. or 1.3 m. Compared with maze streaming the results of Nakamura and Uwamino s formula differed by not more than 16% whereas the results of the JRS method were overestimated by at least a factor of 6. The results of X-rays streaming through the maze are shown in Figures 5 (a)-(c) and duct streaming of X-rays is shown in Figures 5 (d)-(l). Statistical errors of dose rates in MCNP calculations were less than 5%. The E L values of the McGinley method were within a factor of whereas other streaming paths were dissimilar by at least a factor of from the MCNP results. The JRS method and the method from the IAEA Safety Reports Series No.47 also contained results that were overestimated by a factor of compared with MCNP5. Therefore these methods can overestimate and underestimate the shielding design. 3.. Duct streaming ratio The ratios of the X-ray dose at the duct entrance to that at the exit were also calculated to investigate X-ray duct streaming. As the opening area of the duct (A D ) was enlarged or the length was (L D ) shortened the X-ray dose at the duct exit increased. The streaming ratio of the primary X-ray (R W ) head leakage X-ray scattered by the wall (R L ) and head leakage X-ray transmitted through the maze (R T ) was plotted against A D /L D in Figure 6. Opening areas of ducts of.4.5 and.6 m were also calculated. Statistical errors in these additional calculations were less than 3%. R W and R L exhibited a linear relationship with A D /L D whereas R T

4 Progress in Nuclear Science and Technology Volume Figure 4. Results of (a) maze streaming and (b)-(d) duct streaming of neutrons. N&U stands for Nakamura & Uwamino s formula. Figure 5. Results of maze streaming (a)-(c) and duct streaming (d)-(l) of X-ray. Streaming paths of (a) (d) (g) and (j) are EW (b) (e) (h) and (k) are EL and (c) (f) (i) and (l) are ET. 1-3 R W R L 1 R W R L R T A D /L D.6.4. R T Figure 6. Plots of streaming ratios RW RL and RT against AD/LD and fitting lines. neutrons primary X-rays head leakage X-rays scattered by the wall and head leakage X-rays transmitted through the maze were evaluated by comparing them with the results from MCNP5 for the duct in the maze in a 1 MeV linac room. The results from Nakamura and Uwamino s formula for neutron streaming were within a factor of at the evaluation points in the maze and the duct. The results from the X-ray streaming methods were overestimated by at least a factor of. The MCNP5 results also yielded the relation of streaming ratios which is the ratio of the dose at the duct entrance to that at the exit against the ratio of the duct opening area to square length. The dose rates at the duct exit may be calculated by multiplying the streaming ratios to the doses at the duct entrance. was found to be related to the saturation curve. Based on these results the dose rates at the duct exit may be calculated by multiplying the streaming ratios R W R L and R T to the doses at the duct entrance which is calculated using conventional streaming methods. These calculations may use improved maze streaming method in the future. However these ratios should be analyzed for the effects of changing the room layout such as changing the irradiation room size maze geometry and duct position and then used to calculate duct streaming. 4. Conclusions Conventional streaming calculation methods for References [1] International Atomic Energy Agency (IAEA) Safety Report Series No.47 Radiation Protection in the Design of Radiotherapy Facilities IAEA Austria (6) pp ISBN [] Nuclear Safety Technology Center (NUSTEC) Manual for Shield Calculation of Radiation Facilities NUSTEC Japan (15). [in Japanese] [3] P.H. McGinley Shielding Techniques for Radiation Oncology Facilities Medical Physics Wisconsin () pp ISBN [4] Y. Uwamino T. Nakamura T. Ohkubo and A. Hara Measurement and calculation of neutron leakage from a medical electron accelerator Med.

5 16 T. Noto et al. Phys. 13 (1986) pp [5] R. Tayama H. Nakano H. Handa K. Hayashi H. Hirayama K. Shin F. Masukawa H. Nakashima and N. Sasamoto DUCT-III: A Simple Design Code for Duct-Streaming Radiations KEK Internal 1-8 High Energy Accelerator Research Organization (1). [6] K. Shin Evaluation formula for radiation duct streaming J. Nucl. Sci. Technol. 6 (1989) pp [7] X-5 Monte Carlo Team MCNP Version 5 Vol. I: Overview and Theory LA-UR Los Alamos National Laboratory (3).

Estimation of Radioactivity and Residual Gamma-ray Dose around a Collimator at 3-GeV Proton Synchrotron Ring of J-PARC Facility

Estimation of Radioactivity and Residual Gamma-ray Dose around a Collimator at 3-GeV Proton Synchrotron Ring of J-PARC Facility Estimation of Radioactivity and Residual Gamma-ray Dose around a Collimator at 3-GeV Proton Synchrotron Ring of J-PARC Facility Y. Nakane 1, H. Nakano 1, T. Abe 2, H. Nakashima 1 1 Center for Proton Accelerator

More information

Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials*

Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials* SLAC-PUB-70 Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials* J. C. Liu, K. R. Kase, X. S. Mao, W. R. Nelson, J. H. Kleck, and S. Johnson ) Stanford Linear

More information

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT Proceedings of the Eleventh EGS4 Users' Meeting in Japan, KEK Proceedings 2003-15, p.107-113 ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT K. Noto and

More information

Estimation of neutron and gamma radiation doses inside the concrete shield wall for 10 and 15 MV medical linear accelerators

Estimation of neutron and gamma radiation doses inside the concrete shield wall for 10 and 15 MV medical linear accelerators DOI: 10.15669/pnst.4.280 Progress in Nuclear Science and Technology Volume 4 (2014) pp. 280-284 ARTICLE Estimation of neutron and gamma radiation doses inside the concrete shield wall for 10 and medical

More information

Neutron Fluence and Energy Spectra Around the Varian Clinac 2lOOC/23OOC Medical Accelerator

Neutron Fluence and Energy Spectra Around the Varian Clinac 2lOOC/23OOC Medical Accelerator SLAC-PUB-7 190 June 1996 Neutron Fluence and Energy Spectra Around the Varian Clinac 2lOOC/23OOC Medical Accelerator K. R. Kase, X. S. Mao, W. R. Nelson, J. C. Liu Stanford Linear Accelerator Center, Stanford

More information

Chapter 4. QUANTIFYING THE HAZARD II: DATA & ANALYSIS. The dose equivalents for spheres in air with 10 cm radius centred at a point in the

Chapter 4. QUANTIFYING THE HAZARD II: DATA & ANALYSIS. The dose equivalents for spheres in air with 10 cm radius centred at a point in the Chapter 4. QUANTIFYING THE HAZARD II: DATA & ANALYSIS Neutron Dose The dose equivalents for spheres in air with 10 cm radius centred at a point in the treatment room and at 9 points along the passage of

More information

Units S H I E L D I N H = D Q. H: Dose equivalent (Sv) D: Dose (Gy) Q: Quality Factor. 1Sv = 1 J/Kg. 1Gy = 1 J/Kg

Units S H I E L D I N H = D Q. H: Dose equivalent (Sv) D: Dose (Gy) Q: Quality Factor. 1Sv = 1 J/Kg. 1Gy = 1 J/Kg S H I E L D I N G H = D Q Units H: Dose equivalent (Sv) D: Dose (Gy) Q: Quality Factor 1Sv = 1 J/Kg 1Gy = 1 J/Kg if dose is expressed in units of cgy (rad) then dose equivalent is expressed in units of

More information

Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators

Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 3, SUMMER 2003 Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators David S. Followill,*

More information

DESIGN OF A TREATMENT ROOM FOR AN 18-MV LINAC

DESIGN OF A TREATMENT ROOM FOR AN 18-MV LINAC DESIGN OF A TREATMENT ROOM FOR AN 18-MV LINAC LUIS HERNANDEZ-ADAME, a * HECTOR CONTRERAS-SANDOVAL, a HECTOR RENE VEGA-CARRILLO, a and LEONEL HUMBERTO PEREZ LANDEROS b ACCELERATORS KEYWORDS: Monte Carlo

More information

Benchmark Test of JENDL High Energy File with MCNP

Benchmark Test of JENDL High Energy File with MCNP Benchmark Test of JENDL High Energy File with MCNP Masayuki WADA, Fujio MAEKAWA, Chikara KONNO Intense Neutron Source Laboratory, Department of Materials Science Japan Atomic Energy Research Institute,

More information

Measurement of induced radioactivity in air and water for medical accelerators

Measurement of induced radioactivity in air and water for medical accelerators Measurement of induced radioactivity in air and water for medical accelerators K. Masumoto 1, K. Takahashi 1, H. Nakamura 1, A. Toyoda 1, K. Iijima 1, K. Kosako 2, K. Oishi 2, F. Nobuhara 1 High Energy

More information

Peter J. Biggs Ph.D. Massachusetts General Hospital Harvard Medical School Boston, MA 02114

Peter J. Biggs Ph.D. Massachusetts General Hospital Harvard Medical School Boston, MA 02114 Basic Design Principles and Latest Recommended Data, Neutrons and Structural Concerns Winter on Cape Cod Peter J. Biggs Ph.D. Massachusetts General Hospital Harvard Medical School Boston, MA 02114 Outline

More information

Radiation protection issues in proton therapy

Radiation protection issues in proton therapy Protons IMRT Tony Lomax, Centre for Proton Radiotherapy, Paul Scherrer Institute, Switzerland Overview of presentation 1. Proton therapy: An overview 2. Radiation protection issues: Staff 3. Radiation

More information

Estimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom

Estimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom Progress in NUCLEAR SCIENCE and TECHNOLOGY, Vol. 2, pp.83-87 (2) ARTICLE Estimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom Toshioh FUJIBUCHI,2,*, Satoshi

More information

Study on Radiation Shielding Performance of Reinforced Concrete Wall (2): Shielding Analysis

Study on Radiation Shielding Performance of Reinforced Concrete Wall (2): Shielding Analysis 20th International Conference on Structural Mechanics in Reactor Technology (SMiRT 20) Espoo, Finland, August 9-14, 2009 SMiRT 20-Division V, Paper 1865 Study on Radiation Shielding Performance of Reinforced

More information

Shielding Design Considerations for Proton Therapy Facilities

Shielding Design Considerations for Proton Therapy Facilities Shielding Design Considerations for Proton Therapy Facilities p p n π ± INC π 0 Nisy Elizabeth Ipe, Ph.D., C.H.P. Consultant, Shielding Design, Dosimetry & Radiation Protection San Carlos, CA, U.S.A. Email:

More information

Radiation Shielding of a 230 MeV Proton Cyclotron For Cancer Therapy

Radiation Shielding of a 230 MeV Proton Cyclotron For Cancer Therapy Radiation Shielding of a 230 MeV Proton Cyclotron For Cancer Therapy BHASKAR MUKHERJEE Joint DESY and University of Hamburg Accelerator Physics Seminar 27 August 2009 WPE is located within the Campus of

More information

1.E Neutron Energy (MeV)

1.E Neutron Energy (MeV) Proceedings of the Second International Workshop on EGS, 8.-12. August 2000, Tsukuba, Japan KEK Proceedings 200-20, pp.130-134 Measurements of Photoneutron Spectra from Thick Pb Target Bombarded by 1.2

More information

BULK-I. Radiation Shielding Tool for Proton Accelerator Facilities

BULK-I. Radiation Shielding Tool for Proton Accelerator Facilities KEK Internal 2004-1 June 2004 A/R BULK-I Radiation Shielding Tool for Proton Accelerator Facilities English Part R. Tayama, K. Hayashi and H. Hirayama High Energy Accelerator Research Organization Legal

More information

I. INTRODUCTION EXPERIMENTAL

I. INTRODUCTION EXPERIMENTAL International Journal of Computational Engineering Research Vol, 04 Issue, 4 Simulation of Photon and Electron dose distributions by 5 code for the treatment area using the linear electron accelerator

More information

Neutronics Experiments for ITER at JAERI/FNS

Neutronics Experiments for ITER at JAERI/FNS Neutronics Experiments for ITER at JAERI/FNS C. Konno 1), F. Maekawa 1), Y. Kasugai 1), Y. Uno 1), J. Kaneko 1), T. Nishitani 1), M. Wada 2), Y. Ikeda 1), H. Takeuchi 1) 1) Japan Atomic Energy Research

More information

COMPARISON OF COMPUTER CODES APPLICABILITY IN SHIELDING DESIGN FOR HADRON THERAPY FACILITIES *

COMPARISON OF COMPUTER CODES APPLICABILITY IN SHIELDING DESIGN FOR HADRON THERAPY FACILITIES * Romanian Reports in Physics, Vol. 66, No. 1, P. 142 147, 2014 COMPARISON OF COMPUTER CODES APPLICABILITY IN SHIELDING DESIGN FOR HADRON THERAPY FACILITIES * D. SARDARI, M. HAMEDINEJAD Islamic Azad University,

More information

Secondary Radiation and Shielding Design for Particle Therapy Facilities

Secondary Radiation and Shielding Design for Particle Therapy Facilities Secondary Radiation and Shielding Design for Particle Therapy Facilities π± A p, n, π± A p, n A Nisy Elizabeth Ipe, Ph.D., C.H.P. Consultant, Shielding Design, Dosimetry & Radiation Protection San Carlos,

More information

Radiation Shielding for Megavoltage Photon Therapy Machines

Radiation Shielding for Megavoltage Photon Therapy Machines Radiation Shielding for Megavoltage Photon Therapy Machines Peter J. Biggs Ph.D. Department of Radiation Oncology, Massachusetts General Hospital, Harvard Medical School Boston, MA 02114 56 th Annual Meeting,

More information

Journal of Radiation Protection and Research

Journal of Radiation Protection and Research 1) JONG WOON KIM AND YOUNG-OUK LEE: DETAILED ANALYSIS OF THE KAERI ntof FACILITY Journal of Radiation Protection and Research pissn 2508-1888 eissn 2466-2461 http://dx.doi.org/10.14407/jrpr.2016.41.2.141

More information

Peter J. Biggs Ph.D., Massachusetts General lhospital, Harvard Medical School, Boston, MA 02114

Peter J. Biggs Ph.D., Massachusetts General lhospital, Harvard Medical School, Boston, MA 02114 National Council on Radiation Protection Report #151 Structural Shielding Design and Evaluation for Megavoltage X- and Gamma-Ray Radiotherapy Facilities Peter J. Biggs Ph.D., Massachusetts General lhospital,

More information

Energy response for high-energy neutrons of multi-functional electronic personal dosemeter

Energy response for high-energy neutrons of multi-functional electronic personal dosemeter Energy response for high-energy neutrons of multi-functional electronic personal dosemeter T. Nunomiya 1, T. Ishikura 1, O. Ueda 1, N. Tsujimura 2,, M. Sasaki 2,, T. Nakamura 1,2 1 Fuji Electric Systems

More information

Evaluate Shielding Design of the Brachytherapy Unit by Using Monte Carlo Simulation Code

Evaluate Shielding Design of the Brachytherapy Unit by Using Monte Carlo Simulation Code Journal of Modern Physics, 15, 6, 9-911 Published Online June 15 in SciRes. http://www.scirp.org/journal/jmp http://dx.doi.org/1.436/jmp.15.6794 Evaluate Shielding Design of the Brachytherapy Unit by Using

More information

A study on the cost of concrete shielding in a standard radiotherapy facility room

A study on the cost of concrete shielding in a standard radiotherapy facility room BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 06-0 (018) 01-18 A study on the cost of concrete shielding in a standard radiotherapy facility room Eduardo de Paiva a a Instituto de Radioproteção e Dosimetria/Divisão

More information

Radiation Shielding. PTCOG 57 Cincinnati, USA (2018)

Radiation Shielding. PTCOG 57 Cincinnati, USA (2018) Radiation Shielding PTCOG 57 Cincinnati, USA (2018) Meissner Consulting GmbH Prof.-Messerschmitt-Str. 3 D-85579 Neubiberg (München) phone +49 89 30765220 email meissner@meissner-consulting.com PTCOG 57-2018

More information

Radiation safety of the Danish Center for Proton Therapy (DCPT) Lars Hjorth Præstegaard Dept. of Medical Physics, Aarhus University Hospital

Radiation safety of the Danish Center for Proton Therapy (DCPT) Lars Hjorth Præstegaard Dept. of Medical Physics, Aarhus University Hospital Radiation safety of the Danish Center for Proton Therapy (DCPT) Lars Hjorth Præstegaard Dept. of Medical Physics, Aarhus University Hospital Rationale of proton therapy Dose deposition versus depth in

More information

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah

More information

Monte Carlo Simulation concerning Particle Therapy

Monte Carlo Simulation concerning Particle Therapy Monte Carlo Simulation concerning Particle Therapy Masaaki Takashina Graduate School of Medicine, Osaka University, Osaka 565-0871, Japan INTRODUCTION It is well known that the particle therapy has some

More information

Since the beam from the JNC linac is a very high current, low energy beam, energy loss induced in the material irradiated by the beam becomes very lar

Since the beam from the JNC linac is a very high current, low energy beam, energy loss induced in the material irradiated by the beam becomes very lar Proceedings of the Second International Workshop on EGS, 8.-12. August 2000, Tsukuba, Japan KEK Proceedings 200-20, pp.255-263 Beam Dump for High Current Electron Beam at JNC H. Takei and Y. Takeda 1 Japan

More information

Comparison with simulations to experimental data for photoneutron reactions using SPring-8 Injector

Comparison with simulations to experimental data for photoneutron reactions using SPring-8 Injector Comparison with simulations to experimental data for photoneutron reactions using SPring-8 Injector Yoshihiro Asano 1,* 1 XFEL/SPring-8 Center, RIKEN 1-1 Koto Sayo Hyogo 679-5148, Japan Abstract. Simulations

More information

Accelerator Facility Shielding Design

Accelerator Facility Shielding Design University of Tennessee, Knoxville Trace: Tennessee Research and Creative Exchange University of Tennessee Honors Thesis Projects University of Tennessee Honors Program 5-2015 Accelerator Facility Shielding

More information

Secondary Particles Produced by Hadron Therapy

Secondary Particles Produced by Hadron Therapy Iranian Journal of Medical Physics Vol. 12, No. 2, Spring 2015, 1-8 Received: March 10, 2015; Accepted: July 07, 2015 Original Article Secondary Particles Produced by Hadron Therapy Abdolkazem Ansarinejad

More information

Proposed Room Requirements for CT System

Proposed Room Requirements for CT System Siemens Proposed Room Requirements for CT System Semarang, 4-5 May 2017 Restricted Siemens Healthcare GmbH, 2016 Page 1 Roles of Medical Physicist CT Image Quality Radiation Protection Optimization Medical

More information

Work supported by Department of Energy contract DE-AC03-76SF00515

Work supported by Department of Energy contract DE-AC03-76SF00515 SLAC-PUB-7742 February 1998 SHIELDING OF RADIATION FIELDS GENERATED BY 252 Cf IN A CONCRETE MAZE PART II SIMULATION A. Fassò et al. Stanford Linear Accelerator Center Stanford University Stanford. CA 94309

More information

Characterization of the 3 MeV Neutron Field for the Monoenergetic Fast Neutron Fluence Standard at the National Metrology Institute of Japan

Characterization of the 3 MeV Neutron Field for the Monoenergetic Fast Neutron Fluence Standard at the National Metrology Institute of Japan Characterization of the 3 MeV Neutron Field for the Monoenergetic Fast Neutron Fluence Standard at the National Metrology Institute of Japan Hideki Harano * National Metrology Institute of Japan, National

More information

Radiation Safety Considerations for the TPS Accelerators

Radiation Safety Considerations for the TPS Accelerators Radiation Safety Considerations for the TPS Accelerators R.J. Sheu, J. Liu, and J.P. Wang National Synchrotron Radiation Research Center, 101 Hsin-Ann Road, Hsinchu Science Park, Hsinchu 30076, TAIWAN

More information

Neutron Shielding Properties Of Concrete With Boron And Boron Containing Mineral

Neutron Shielding Properties Of Concrete With Boron And Boron Containing Mineral 15 APJES I-I (2013) 15-19 Neutron Shielding Properties Of Concrete With Boron And Boron Containing Mineral *1 Salim Orak, 2 Derya Yilmaz Baysoy 1 Istanbul Commerce University, Faculty of Arts and Science,

More information

neutron building Project Title: Moderator design of RANS2 and investigating of radiation equivalent dose for Name: Sheng Wang

neutron building Project Title: Moderator design of RANS2 and investigating of radiation equivalent dose for Name: Sheng Wang Project Title: Moderator design of RANS2 and investigating of radiation equivalent dose for Name: Sheng Wang neutron building Laboratory at RIKEN: Neutron Beam Technology Team Description of the project

More information

Radiation shielding for gamma stereotactic radiosurgery units

Radiation shielding for gamma stereotactic radiosurgery units JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 8, NUMBER 3, SUMMER 2007 Radiation shielding for gamma stereotactic radiosurgery units Patrick N. McDermott a Radiation Oncology Center, William Beaumont

More information

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators N. Golnik 1, P. Kamiński 1, M. Zielczyński 2 1 Institute of Precision and Biomedical Engineering,

More information

The neutron dose equivalent evaluation and shielding at the maze entrance of a Varian Clinac 23EX treatment room

The neutron dose equivalent evaluation and shielding at the maze entrance of a Varian Clinac 23EX treatment room This partial RIAS has been downloaded on 15 Apr 2019 The neutron dose equivalent evaluation and shielding at the maze entrance of a Varian Clinac 23EX treatment room Xudong Wang, Carlos Esquivel, Elena

More information

Radiation Shielding of Extraction Absorbers for a Fermilab Photoinjector

Radiation Shielding of Extraction Absorbers for a Fermilab Photoinjector Fermilab FERMILAB-TM-2220 August 2003 Radiation Shielding of Extraction Absorbers for a Fermilab Photoinjector I.L. Rakhno Fermilab, P.O. Box 500, Batavia, IL 60510, USA August 12, 2003 Abstract Results

More information

TRAINING IN EXTERNAL DOSIMETRY CALCULATIONS WITH COMPUTATIONAL CODES

TRAINING IN EXTERNAL DOSIMETRY CALCULATIONS WITH COMPUTATIONAL CODES TRAINING IN EXTERNAL DOSIMETRY CALCULATIONS WITH COMPUTATIONAL CODES S. MORATÓ, A.BERNAL, A. QUEROL, A. ABARCA, C. GÓMEZ-ZARZUELA, R.MIRÓ, G.VERDÚ Institute for Industrial, Radiophysical and Environmental

More information

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR ANALELE STIINTIFICE ALE UNIVERSITATII AL. I. CUZA IASI Tomul II, s. Biofizică, Fizică medicală şi Fizica mediului 2006 CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR Dan

More information

Radiation Protection At Synchrotron Radiation Facilities

Radiation Protection At Synchrotron Radiation Facilities 3 rd ILSF Advanced School on Synchrotron Radiation and Its Applications September 14-16, 2013 Radiation Protection At Synchrotron Radiation Facilities Ehsan Salimi Shielding and Radiation Safety Group

More information

Neutron and Gamma-ray Emission Double Dierential Cross Sections. *5 Energy Conversion Engineering, Kyushu University, Kasuga-koen, Kasuga-shi 816.

Neutron and Gamma-ray Emission Double Dierential Cross Sections. *5 Energy Conversion Engineering, Kyushu University, Kasuga-koen, Kasuga-shi 816. Neutron and Gamma-ray Emission Double Dierential Cross Sections for the Nuclear Reaction by 1.5 GeV + Incidence Kiminori IGA 1, Kenji ISHIBASHI 1, Nobuhiro SHIGYO 1, Naruhiro MATSUFUJI 1;+1, Tatsushi NAKAMOTO

More information

ARTICLE. Progress in Nuclear Science and Technology Volume 4 (2014) pp Yoshiko Harima a*, Naohiro Kurosawa b and Yukio Sakamoto c

ARTICLE. Progress in Nuclear Science and Technology Volume 4 (2014) pp Yoshiko Harima a*, Naohiro Kurosawa b and Yukio Sakamoto c DOI: 10.15669/pnst.4.548 Progress in Nuclear Science and Technology Volume 4 (2014) pp. 548-552 ARTICLE Parameter search of geometric-progression formula for gamma-ray isotropic point source buildup factors

More information

Monte Carlo simulation for the estimation of iron in human whole blood and comparison with experimental data

Monte Carlo simulation for the estimation of iron in human whole blood and comparison with experimental data Pramana J. Phys. (2017) 88: 49 DOI 10.1007/s12043-016-1344-1 c Indian Academy of Sciences Monte Carlo simulation for the estimation of iron in human whole blood and comparison with experimental data M

More information

Neutronic Design on a Small Accelerator based 7 Li (p, n) Neutron Source for Neutron Scattering Experiments

Neutronic Design on a Small Accelerator based 7 Li (p, n) Neutron Source for Neutron Scattering Experiments 2010-08-16 Neutronic Design on a Small Accelerator based 7 Li (p, n) Neutron Source for Neutron Scattering Experiments Fujio Hiraga, Takanori Okazaki and Yoshiaki Kiyanagi Hokkaido University 1 Technical

More information

Shutdown dose calculations for the IFMIF test facility and the high flux test module

Shutdown dose calculations for the IFMIF test facility and the high flux test module DOI: 10.15669/pnst.4.233 Progress in Nuclear Science and Technology Volume 4 (2014) pp. 233-237 ARTICLE Shutdown dose calculations for the IFMIF test facility and the high flux test module Keitaro Kondo

More information

Measurement of 40 MeV Deuteron Induced Reaction on Fe and Ta for Neutron Emission Spectrum and Activation Cross Section

Measurement of 40 MeV Deuteron Induced Reaction on Fe and Ta for Neutron Emission Spectrum and Activation Cross Section Measurement of 40 MeV Deuteron Induced Reaction on Fe and Ta for Neutron Emission Spectrum and Activation Cross Section Toshiro Itoga, Masayuki Hagiwara, Takuji Oishi, So Kamada, Mamoru Baba Cyclotron

More information

QUESTION 8 GIVEN. Electron beam kinetic energy = 20 MeV Peak current = 1 A Beam pulse length = 1 microsecond Beam pulse frequency = 10 Hz

QUESTION 8 GIVEN. Electron beam kinetic energy = 20 MeV Peak current = 1 A Beam pulse length = 1 microsecond Beam pulse frequency = 10 Hz Page 1 of 2 QUESTION 8 You are asked to design the shielding for an electron accelerator facility with the facility information given below. Use the figures copied from the NCRP Report No. 51 (1977), Radiation

More information

USE OF DLP FOR ESTABLISHING THE SHIELDING OF MULTI- DETECTOR COMPUTED TOMOGRAPHY ROOMS

USE OF DLP FOR ESTABLISHING THE SHIELDING OF MULTI- DETECTOR COMPUTED TOMOGRAPHY ROOMS USE OF DLP FOR ESTABLISHING THE SHIELDING OF MULTI- DETECTOR COMPUTED TOMOGRAPHY ROOMS F.R. Verdun 1, A. Aroua 1, P.R. Trueb 2, F.O. Bochud 1* 1 University Institute for Radiation Physics, Switzerland

More information

Secondary Neutron Dose Measurement for Proton Line Scanning Therapy

Secondary Neutron Dose Measurement for Proton Line Scanning Therapy Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.162 pissn 2508-4445, eissn 2508-4453 Secondary Neutron Dose Measurement for Proton Line Scanning

More information

PHYS 5020 Computation and Image Processing

PHYS 5020 Computation and Image Processing PHYS 5020 and Image Processing : Monte Carlo Thursday 2 August 2012 Monte Carlo (MC) is a numerical method that uses random sampling of probability distributions to simulate stochastic processes in nature,

More information

Radiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp Advance Access publication 11 May 2007

Radiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp Advance Access publication 11 May 2007 Radiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp. 229 233 Advance Access publication 11 May 2007 doi:10.1093/rpd/ncm047 CHARACTERIZATION AND UTILIZATION OF A BONNER SPHERE SET BASED ON GOLD

More information

Analyses of shielding benchmark experiments using FENDL-3 cross-section data starter library for ITER and IFMIF applications

Analyses of shielding benchmark experiments using FENDL-3 cross-section data starter library for ITER and IFMIF applications DOI: 10.15669/pnst.4.322 Progress in Nuclear Science and Technology Volume 4 (2014) pp. 322-326 ARTICLE Analyses of shielding benchmark experiments using FENDL-3 cross-section data starter library for

More information

Accelerator Facility Accident Report

Accelerator Facility Accident Report Accelerator Facility Accident Report 31 May 2013 Incorporated Administrative Agency - Japan Atomic Energy Agency Inter-University Research Institute - High Energy Accelerator Research Organization Subject:

More information

Characteristics of Filtered Neutron Beam Energy Spectra at Dalat Reactor

Characteristics of Filtered Neutron Beam Energy Spectra at Dalat Reactor World Journal of Nuclear Science and Technology, 2014, 4, 96-102 Published Online April 2014 in SciRes. http://www.scirp.org/journal/wjnst http://dx.doi.org/10.4236/wjnst.2014.42015 Characteristics of

More information

Convertible source system of thermal neutron and X-ray at Hokkaido University electron linac facility

Convertible source system of thermal neutron and X-ray at Hokkaido University electron linac facility IL NUOVO CIMENTO 38 C (2015) 187 DOI 10.1393/ncc/i2015-15187-9 Colloquia: UCANS-V Convertible source system of thermal neutron and X-ray at Hokkaido University electron linac facility T. Kamiyama( ), K.

More information

Monte Carlo Calculations Using MCNP4B for an Optimal Shielding Design. of a 14-MeV Neutron Source * James C. Liu and Tony T. Ng

Monte Carlo Calculations Using MCNP4B for an Optimal Shielding Design. of a 14-MeV Neutron Source * James C. Liu and Tony T. Ng SLAC-PUB-7785 November, 1998 Monte Carlo Calculations Using MCNP4B for an Optimal Shielding Design of a 14-MeV Neutron Source * James C. Liu and Tony T. Ng Stanford Linear Accelerator Center MS 48, P.O.

More information

arxiv: v1 [physics.ins-det] 9 Apr 2018

arxiv: v1 [physics.ins-det] 9 Apr 2018 arxiv:1804.02889v1 [physics.ins-det] 9 Apr 2018 Study of neutron shielding collimators for curved beamlines at the European Spallation Source 1. Introduction V. Santoro 1,2, D. D. DiJulio 1,2, S. Ansell

More information

Integral Benchmark Experiments of the Japanese Evaluated Nuclear Data Library (JENDL)-3.3 for the Fusion Reactor Design

Integral Benchmark Experiments of the Japanese Evaluated Nuclear Data Library (JENDL)-3.3 for the Fusion Reactor Design 1 Integral Benchmark Experiments of the Japanese Evaluated Nuclear Data Library (JENDL)-3.3 for the Fusion Reactor Design T. Nishitani 1), K. Ochiai 1), F. Maekawa 1), K. Shibata 1), M. Wada 2), I. Murata

More information

Induced photonuclear interaction by Rhodotron-TT MeV electron beam

Induced photonuclear interaction by Rhodotron-TT MeV electron beam PRAMANA c Indian Academy of Sciences Vol. 78, No. 2 journal of February 2012 physics pp. 257 264 Induced photonuclear interaction by Rhodotron-TT200 10 MeV electron beam FARSHID TABBAKH 1,, MOJTABA MOSTAJAB

More information

Comparison of FLUKA and STAC8 for shielding calculations of the hard X-ray line of the LCLS

Comparison of FLUKA and STAC8 for shielding calculations of the hard X-ray line of the LCLS SLAC RADIATION PHYSICS NOTE RP-08-11 September 23, 2008 Comparison of FLUKA and STAC8 for shielding calculations of the hard X-ray line of the LCLS J. Vollaire, A. Prinz Radiation Protection Department,

More information

Determination of Photon Ambient Dose Buildup Factors for Radiological Applications for Points and Plaque Source Configurations Using MCNP5

Determination of Photon Ambient Dose Buildup Factors for Radiological Applications for Points and Plaque Source Configurations Using MCNP5 Determination of Photon Ambient Dose Buildup Factors for Radiological Applications for Points and Plaque Source Configurations Using MCNP5 P. Deatanyah 1, C.C. Arwui 1, S. Wotorchi- Gordon 1, H. Lawluvi

More information

Monte Carlo simulation for bremsstrahlung and photoneutron yields in high-energy x-ray radiography

Monte Carlo simulation for bremsstrahlung and photoneutron yields in high-energy x-ray radiography Monte Carlo simulation for bremsstrahlung and photoneutron yields in high-energy x-ray radiography Xu Hai-Bo( 许海波 ), Peng Xian-Ke( 彭现科 ), and Chen Chao-Bin( 陈朝斌 ) Institute of Applied Physics and Computational

More information

Compton suppression spectrometry

Compton suppression spectrometry Compton suppression spectrometry In gamma ray spectrometry performed with High-purity Germanium detectors (HpGe), the detection of low intensity gamma ray lines is complicated by the presence of Compton

More information

The scanning microbeam PIXE analysis facility at NIRS

The scanning microbeam PIXE analysis facility at NIRS Nuclear Instruments and Methods in Physics Research B 210 (2003) 42 47 www.elsevier.com/locate/nimb The scanning microbeam PIXE analysis facility at NIRS Hitoshi Imaseki a, *, Masae Yukawa a, Frank Watt

More information

Shielding Design for the Imaging and Medical Beamline at the Australian Synchrotron

Shielding Design for the Imaging and Medical Beamline at the Australian Synchrotron Shielding Design for the Imaging and Medical Beamline at the Australian Synchrotron P. Berkvens and D. Häusermann European Synchrotron Radiation Facility BP 0, Grenoble Cedex 0, France Australian Synchrotron

More information

X-ray Interaction with Matter

X-ray Interaction with Matter X-ray Interaction with Matter 10-526-197 Rhodes Module 2 Interaction with Matter kv & mas Peak kilovoltage (kvp) controls Quality, or penetrating power, Limited effects on quantity or number of photons

More information

In our recent paper [1], applicability of the self-activation of an NaI scintillator has been studied for the photo-neutron monitoring around

In our recent paper [1], applicability of the self-activation of an NaI scintillator has been studied for the photo-neutron monitoring around Proc. Int. Symp. on Radiation Detectors and Their Uses (ISRD2016) https://doi.org/10.7566/jpscp.11.050002 High Sensitive Neutron-detection by using a Self-activation of Iodine-containing Scintillators

More information

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong Original Article PMP Progress in Medical Physics 28(2), June 2017 https://doi.org/10.14316/pmp.2017.28.2.49 pissn 2508-4445, eissn 2508-4453 Electron Energy Distribution for a Research Electron LINAC Heuijin

More information

CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE

CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE Page no. 5.1 Introduction 132 5.2 Methods and measurements 132 5.3 Saturation thickness of elements and polymers

More information

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 AUTHORS: J. Cardoso, L. Santos, C. Oliveira ADRESS: Instituto Tecnológico e Nuclear Estrada Nacional 10; 2686-953 Sacavém;

More information

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS Leonardo de Holanda Mencarini 1,2, Claudio A. Federico 1,2 and Linda V. E. Caldas 1 1 Instituto de Pesquisas Energéticas e Nucleares IPEN,

More information

Simulation Modeling in Dosimetry

Simulation Modeling in Dosimetry Simulation Modeling in Dosimetry Aleksei Zhdanov Ural Federal University named after the first President of Russia B. N. Yeltsin, Yekaterinburg, Russian Federation jjj1994@yandex.ru Leonid Dorosinskiy

More information

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9 Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation

More information

Shielding verification and neutron dose evaluation of the Mevion S250 proton therapy unit

Shielding verification and neutron dose evaluation of the Mevion S250 proton therapy unit Received: 12 September 2017 Revised: 14 November 2017 Accepted: 22 November 2017 DOI: 10.1002/acm2.12256 RADIATION PROTECTION & REGULATIONS Shielding verification and neutron dose evaluation of the Mevion

More information

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN 308 Angular dependence of 662 kev multiple backscattered gamma photons in Aluminium Ravindraswami K a, Kiran K U b, Eshwarappa K M b and Somashekarappa H M c* a St Aloysius College (Autonomous), Mangalore

More information

Calculations of Neutron Yield and Gamma Rays Intensity by GEANT4

Calculations of Neutron Yield and Gamma Rays Intensity by GEANT4 Armenian Journal of Physics, 2016, vol. 9, issue 4, pp. 315-323 Calculations of Neutron Yield and Gamma Rays Intensity by GEANT4 R. Avagyan, R. Avetisyan, V. Ivanyan*, I. Kerobyan A.I. Alikhanyan National

More information

Proposed Room Requirements for CT System

Proposed Room Requirements for CT System Siemens Proposed Room Requirements for CT System Semarang, 4-5 May 2017 Restricted Siemens Healthcare GmbH, 2016 Page 1 Roles of Medical Physicist CT Image Quality Radiation Protection Optimization Medical

More information

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Annibal T. Baptista Neto, Carlos M. A. Soares, Teógenes A. da Silva Nuclear Technology Development Center, Radiation Dosimetry

More information

Neutron collimator design of neutron radiography based on the BNCT facility arxiv: v1 [physics.ins-det] 3 May 2013

Neutron collimator design of neutron radiography based on the BNCT facility arxiv: v1 [physics.ins-det] 3 May 2013 CPC(HEP & NP), 2009, 33(X): 1 5 Chinese Physics C Vol. 33, No. X, Xxx, 2009 Neutron collimator design of neutron radiography based on the BNCT facility arxiv:1305.0672v1 [physics.ins-det] 3 May 2013 YANG

More information

WM 07 Conference, February 25 March 01, 2007, Tucson, AZ

WM 07 Conference, February 25 March 01, 2007, Tucson, AZ Design and Construction of a High Energy X-Ray R&D Facility, and the Development and Optimization of Real Time Radioisotopic Characterization of Remote Handled Waste at MeV Energies. S. Halliwell, V.J.Technologies

More information

SLAC-PUB Submitted to Radiation Protection and Dosimetry. Work supported by Department of Energy contract DE-AC02-76SF00515

SLAC-PUB Submitted to Radiation Protection and Dosimetry. Work supported by Department of Energy contract DE-AC02-76SF00515 SLAC-PUB-11088 CALCULATIONS OF NEUTRON AND PHOTON SOURCE TERMS AND ATTENUATION PROFILES FOR THE GENERIC DESIGN OF THE SPEAR3 STORAGE RING SHIELD S. H. Rokni, H. Khater, J. C. Liu, S. Mao and H. Vincke

More information

Accelerator Upgrade Problem

Accelerator Upgrade Problem Accelerator Upgrade Problem You are a health physicist at a university. Physics faculty approached you with a proposal to upgrade the exis=ng 6 MeV electron accelerator to 30 MeV energy. The upgraded facility

More information

Internal Report DESY D3-86 January Production of radioactive nuclides in soil and groundwater near the beam dump of a Linear Collider. K.

Internal Report DESY D3-86 January Production of radioactive nuclides in soil and groundwater near the beam dump of a Linear Collider. K. Internal Report DESY D3-86 January 1997 Production of radioactive nuclides in soil and groundwater near the beam dump of a Linear Collider K. Tesch Internal Report DESY D3-86 January 1997 Production of

More information

Current issues of radiation safety regulation for accelerator facilities in Japan

Current issues of radiation safety regulation for accelerator facilities in Japan Current issues of radiation safety regulation for accelerator facilities in Japan K. MASUMOTO Radiation Science Center, High Energy Accelerator Research Organization, Japan Introduction In Japan, the clearance

More information

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory Recent Activities on Neutron Standardization at the Electrotechnical Laboratory K. Kudo, N. Takeda, S. Koshikawa and A. Uritani Quantum Radiation Division, National Metrology Institute of Japan (NMIJ)

More information

ACTIVATION ANALYSIS OF DECOMISSIONING OPERATIONS FOR RESEARCH REACTORS

ACTIVATION ANALYSIS OF DECOMISSIONING OPERATIONS FOR RESEARCH REACTORS ACTIVATION ANALYSIS OF DECOMISSIONING OPERATIONS FOR RESEARCH REACTORS Hernán G. Meier, Martín Brizuela, Alexis R. A. Maître and Felipe Albornoz INVAP S.E. Comandante Luis Piedra Buena 4950, 8400 San Carlos

More information

Medical Neutron Science

Medical Neutron Science Medical Neutron Science 03 Neutron Activation Analysis The use of NAA techniques for medical applications was first reported in 1964 for measurement of sodium in the body J. Anderson, SB S.B. Ob Osborn,

More information

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation S. Moriuchi, M.Tsutsumi2 and K. Saito2 Nuclear safety technology Center,

More information

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Recent Activities on Neutron Calibration Fields at FRS of JAERI Recent Activities on Neutron Calibration Fields at FRS of JAERI Michio Yoshizawa, Yoshihiko Tanimura, Jun Saegusa and Makoto Yoshida Department of Health Physics, Japan Atomic Energy Research Institute

More information

(This is a sample cover image for this issue. The actual cover is not yet available at this time.)

(This is a sample cover image for this issue. The actual cover is not yet available at this time.) (This is a sample cover image for this issue. The actual cover is not yet available at this time.) This article appeared in a journal published by Elsevier. The attached copy is furnished to the author

More information