Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry
|
|
- Patrick Lawson
- 5 years ago
- Views:
Transcription
1 Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry TJ JORDAN Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE Nahum, A Nisbet & WG Pitchford Practical Course in Reference Dosimetry, NPL 2016
2 Basic Methodology Calorimeter based Absorbed Dose to Water calibration in electron beams Pictures courtesy of NPL
3 Basic Methodology Calorimeter based Absorbed Dose to Water calibration in electron beams Individualised parallel plate electron chamber calibration at 6 electron beam qualities based on R 50,D (depth of 50% dose) Original NPL accelerator, R 50,D : 1.97 to 6.60cm [6 to 19 MeV] LESS than the range of hospital clinical beams Elekta Clinical Linac, R 50,D : 1.6 up to 8.9*cm [4 to 22* MeV] *(not guaranteed)
4 Basic Methodology NPL Calibration certificate Calorimeter based Absorbed Dose to Water calibration in electron beams Individualised parallel plate electron chamber calibration at 6 electron beam qualities based on R 50,D Uses reference depth defined as Z ref = 0.6 R 50,D 0.1cm NPL determined Recombination f ion = m(dose per pulse) + c
5 Basic Methodology Calorimeter based Absorbed Dose to Water calibration in electron beams 100 %Depth Dose Individualised parallel plate electron chamber calibration at 7 electron beam qualities based on R 50,D Uses reference depth defined as Z ref = 0.6 R 50,D 0.1cm % Dose dmax Zref NPL determined Recombination f ion = m(dose per pulse) + c R50 Depth (mm)
6 Basic Methodology NPL Calibration certificate Calorimeter based Absorbed Dose to Water calibration in electron beams Individualised parallel plate electron chamber calibration at 7 electron beam qualities based on R 50,D Uses reference depth defined as Z ref = 0.6 R 50,D 0.1cm NPL determined Recombination f ion = m(dose per pulse) + c
7 Basic Methodology NPL Calibration certificate Calorimeter based Absorbed Dose to Water calibration in electron beams Individualised parallel plate electron chamber calibration at 6 electron beam qualities based on R 50,D Uses reference depth defined as Z ref = 0.6 R 50,D 0.1cm NPL determined Recombination f ion = m(dose per pulse) + c
8 Determination of R 50,D Measurement of Depth- Dose For photons, Dose w Air ionisation x W/e x [µ en / ] w,air [µ en / ] w,air the Mass Energy Absorption Coefficient for water to air is nearly constant with depth and relative Depth-Ionisation is assumed to Depth-Dose in water. For electrons Dose w Air ionisation x W/e x S w,air As the electron energy spectrum decreases rapidly with depth (from incident energy to 0), the Stopping Power ratio S w,air is NOT constant Depth-Ionisation curve needs renormalising at every point by S w,air ratio to be representative of a Depth-Dose curve
9 Sw,med Stopping Power Ratios (20MeV Electrons) Water/Air Depth (mm)
10 Sw,med Stopping Power Ratios (20MeV Electrons) Water/Si Water/Air Depth (mm)
11 Determination of R 50,D Protocol gives 3 options; Find R 50,D either from; R 50,D = x R 50,I cm eqn. 2.1 Use measured Depth-Ionisation curve (Need %DD curve clinically, see later) S w,air corrected Depth-Ionisation curve To produce Depth-Dose curve Directly from diode measured Depth-Ionisation curve as S w,silicon constant until deep Measure Depth-Dose directly
12 % DOSE Depth Dose Measurement Si diode NACP S w,air Corr DEPTH (mm)
13 (corrected) NACP vs Diode: Energy Dependence MeV NACP 9MeV NACP 12MeV NACP 16MeV NACP 20MeV NACP % Dose Depth (mm) Bremsstrahlung tail
14 (corrected) NACP vs Diode: Consistent Error % Dose MeV NACP 9MeV NACP 12MeV NACP 16MeV NACP 20MeV NACP Diode Depth (mm)
15 Effective Point of Measurement, P eff An air ionisation chamber introduces an air bubble into the water phantom The effective point of measurement is where the fluence is equivalent to the fluence in the undisturbed medium For a parallel plate chamber P eff is just inside the front window [For a cylindrical chamber it is around 0.6 x Internal Radius forward of the physical centre]
16 Effective Point of Measurement, P eff ROOS 1.18x 1mm NACP FARMER 0.125cc DIODE 1.7x 0.6mm 1.8mm 1.7mm 0.5mm Effective Depths
17 Polarity In an electron beam a polarity effect may arise as some of the primary beam collides with the collecting electrode F pol = ( M + + M - )/2M M is reading with normal polarity M + and M - readings with respective polarity In an electron beam polarity can change with depth (and energy) as electrons scatter obliquely
18 Polarity NPL polarity negative volts (to front window) Collecting electrode positive wrt front window Uses conventional connection of separate HV Connects to flying lead, or outer braid of cable
19 Polarity (conventional) NACP FLYING LEAD -100V EARTH NEGATIVE TYPE B
20 Polarity (electrometer floating) NACP +100V EARTH POSITIVE TYPE A
21 Reading Polarity: Measurement Accuracy NACP: 4MeV, R50-200V +200V USE A GOOD REFERENCE CHAMBER Switch off and disconnect Earth cable Reverse polarity Reconnect and switch on Take readings until stable Repeat Do +/- volts consecutively (not as part of recombination study) If in doubt use F pol = 1.0(!) -50V Time
22 1/Reading Ion Recombination Theory Plotting 1/Reading vs 1/Voltage is a straight line This can easily be extrapolated to volts, or 100% collection efficiency 0 ( volt) 1/Volts Linear dependence permits use of the 2 voltage method f ion - 1 = (M 1 /M 2-1)/(V 1 /V 2-1) In particular, if V 1 = 2x V 2 then f ion = M 1 / M 2 (M is reading)
23 Recombination Correction f ion Actual: Farmer Well behaved 1/Reading mev 20mev Parallel lines means curves overlay when normalised This means recombination is independent of energy Unless dose per pulse, Dp changes /Voltage Volts
24 1/Reading Recombination Correction f ion Actual: Parallel Plate chambers MeV, Markus 4 MeV NACP 10 MeV NACP 15 MeV NACP /VOLTS oo VOLTS
25 Recombination Correction Use of lower voltages (~100V) on parallel plate ensures linear region but may give large correction factor (2-3% on Varian) Use of higher voltages will give a smaller correction but loss of linearity means the error on the volt intercept is high It is recommended to stay close to NPL calibration voltage (100V) to reduce uncertainty MEASURE LOCALLY and use NPL as check only Due to uncertainty in Dose per Pulse
26 Recombination Correction f ion Roos chamber ion recombination /reading MeV 9MeV 12MeV 16MeV 20MeV /voltage
27 Recombination/Polarity NPL Certificate
28 Recombination Correction Recombination depends only on the Dose per Pulse (D p ) and chamber voltage Recombination is independent of energy But different energies might use a different Dose per Pulse!! Recombination is linearly related to D p F ion = 1 + (c-1)(v cal /V user ) + m(v cal /V user ) D p F ion = c + m x D p (at V user = 100Volt)
29 Linear Accelerator Calibration Roos or NACP with inside of front window at Z ref D w = R x N D,w x f t,p x f ion x f pol x %DD From %Depth Dose curve establish correction for dose at Z ref depth to D max depth Clinically quote Dose at d max and calibrate linac for 1cGy/mu at that point (NOT Z ref ) Negligible up to >12MeV Can be 4-5% at higher energy
30 Spreadsheet Calculates composite N D,W factors including polarity, recombination and Stopping power extrapolation User Electron Beam Definition 2 Type data into yellow cells only Accelerator Varian 2100cd Field Size (cmxcm) 10x10 at F.S.D (cm) 100 Energy 1 Energy 2 Energy 3 Energy 4 Energy 5 Energy 6 Nominal Energy (MeV) 6 MeV 9 MeV 12 MeV 16 MeV 20 MeV 20 MeV dmax (cm) in water R50,I (cm) in water R50,D (cm) in water zref (cm) in water Mean Surface Energy (MeV) Stopping Power Ratio (Eqn D.2) Reference Chamber Factor Summary: Calibration Factor x 10 7 (Gy/C) Polarity Corr Recombination Corr Electrometer Corr. (nc) % Depth Dose at Zref Composite Factor (to cgy)
31 Extrapolation to Higher (& Lower) Beam Qualities The previous NPL calibration covers a limited range of electron beam qualities The higher end was equivalent to a ~16MeV clinical beam To extrapolate to a higher energy use the ratio of stopping powers at the two (different) reference positions N w,user = N w,r50=6.6 x [S w,air, R50=User, dref ] / [S w,air, R50=6.6 ] N w,user = N w,r50=6.6 x [S w,air, R50=User, dref ] / Use SPR table in protocol or spreadsheet
32 Nw,u Spreadsheet: Extrapolation to Higher Beam Qualities PLOT Nw,u calibration and extrapolation N w,u x 10 7 (Gy/C) R 50,D (cm) Type extrapolation value S w,air R50,D
33 Nw,u Spreadsheet: Extrapolation to Higher Beam Qualities PLOT Nw,u calibration and extrapolation Type extrapolation value S w,air lin regrssn N w,u x 10 7 (Gy/C) R 50,D (cm) % R50,D
34 Nw,u Spreadsheet: Extrapolation to Higher Beam Qualities PLOT Nw,u calibration and extrapolation Type extrapolation value S w,air lin regrssn cubic N w,u x 10 7 (Gy/C) R 50,D (cm) % -0.47% R50,D
35 2,3,4 Rule of Thumb
36 Last Word Feedback on implementation and 2003/1996 differences Recommendations on 2 year calibration may be reexamined
ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *
Romanian Reports in Physics, Vol. 67, No. 3, P. 1152 1158, 2015 ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * ELENA STANCU 1,2, CATALIN VANCEA
More informationNACP-02 perturbation correction factors for the NPL primary standard of absorbed dose to water in high energy electron beams
NACP-02 perturbation correction factors for the NPL primary standard of absorbed dose to water in high energy electron beams E. Chin 1, J. Seuntjens 1, H. Palmans 2, A. DuSautoy 2, D. Shipley 2, M. Bailey
More informationDosimetry: Electron Beams
ICTP SChool On MEdical PHysics For RAdiation THerapy: DOsimetry And TReatment PLanning For BAsic And ADvanced APplications 13-24 April 2015 Miramare, Trieste, Italy Dosimetry: Electron Beams G. Hartmann
More informationReference Dosimetry for Megavoltage Therapy Beams: Electrons
Reference Dosimetry for Megavoltage Therapy Beams: Electrons David Followill Ph.D Radiological Physics Center UT M.D.Anderson Cancer Center Houston TX Protocol for Clinical Reference Dosimetry of High-Energy
More informationUpdating reference dosimetry a decade after TG-51
Updating reference dosimetry a decade after TG-51 Malcolm McEwen Ionizing Radiation Standards Institute for National Measurement Standards National Research Council, Canada CE Presentation at AAPM Annual
More information) for Varian TrueBeam high-dose-rate therapy beams
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 Ion recombination correction factors (P ion ) for Varian TrueBeam high-dose-rate therapy beams Stephen F. Kry, 1a Richard Popple,
More informationImprovements to the UK Primary Standard Therapy Level Electron Beam Calorimeter
Radiotherapy Standards Users Group meeting NPL, 5 th June 2007 Improvements to the UK Primary Standard Therapy Level Electron Beam Calorimeter Mark Bailey, G A Bass, N D Lee, G A Mapp Contents 1. History
More information8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required
UPDATING TG-51 When will it end? Malcolm McEwen Ionizing Radiation Standards National Research Council, Canada AAPM Annual Meeting, Charlotte, 2012 What are these updates? Working Group review recommends:
More informationN. E. Ipe*, K. E. Rosser, C. J. Moretti, J. W. Manning and M. J. Palmer
SLAC-PUB-8099 July 1999 Air Kerma Calibration Factors and k ch Values for PTW Soft X-ray, NACP and Roos Ionization Chambers at Very Low X-ray Energies (0.035 mm - 1.0 mm Al HVL) N. E. Ipe*, K. E. Rosser,
More informationRadiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe
Radiation Dosimetry Alun Beddoe Medical Physics University Hospital Birmingham NHS Trust Contents ABSOLUTE DOSIMETRY (CALIBRATION) Photon interactions (recap) Energy transfer and absorption Electron range
More informationReferensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden
Referensdosimetri Crister Ceberg Medical Radiation Physics Lund University Sweden Reference dosimetry Determination of absorbed dose to water under reference conditions Not accounting for uncertainties
More informationHigh-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers
Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society of Radiobiology
More informationInstrumentation for Verification of Dose
Instrumentation for Verification of Dose S. Vatnitsky MedAustron GmbH, Wiener Neustadt, Austria Presented to: Educational Workshop PTCOG 53 Shanghai, China, June 9-11, 2014 Consistent and harmonized dosimetry
More informationCode of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams
Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams NEDERLANDSE COMMISSIE VOOR STRALINGSDOSIMETRIE Report 18 of the Netherlands Commission on Radiation Dosimetry
More informationComparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification.
Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification. Antonio Lopez Medina, Antonio Teijeiro, Daniela Medal, Francisco Salvador, Julio Vazquez, Manuel
More informationPRIMARY STANDARDS of AIR KERMA for 60 CO and X-RAYS & ABSORBED DOSE in PHOTON and ELECTRON BEAMS. Malcolm McEwen
PRIMARY STANDARDS of AIR KERMA for 60 CO and X-RAYS & ABSORBED DOSE in PHOTON and ELECTRON BEAMS Malcolm McEwen Definitions: Standard instrument/measurement/artifact intended to define, realize, conserve
More informationImplementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy
Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy M. Saiful Huq, PhD, FAAPM, FInstP Dept. of Radiation Oncology, University of
More informationInstrumentation for Verification of Dose
Instrumentation for Verification of Dose MedAustron GmbH, Wiener Neustadt, Austria Presented to: Educational Workshop PTCOG 52 Essen, Germany, May 3 5, 2013 Consistent and harmonized dosimetry guidelines
More informationAir Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira
Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 J. Cardoso, L. Santos, C. Oliveira SUMMARY: i. The primary standard ii. The irradiation room iii. Experimental results iv. Simulation
More informationElectron beam water calorimetry measurements to obtain beam quality conversion factors
Electron beam water calorimetry measurements to obtain beam quality conversion factors Bryan R. Muir, a) Claudiu D. Cojocaru, Malcolm R. McEwen, and Carl K. Ross Measurement Science and Standards, National
More informationPrinciples of applied dosimetry - illustrated by ionometry. Lesson FYSKJM4710 Eirik Malinen
Principles of applied dosimetry - illustrated by ionometry Lesson FYSKJM4710 Eirik Malinen Ionometry Ionometry: the art of measuring ionizations Number of ionizations proportional to dose Air filled ionization
More informationChapter 9: Calibration of Photon and Electron Beams
Chapter 9: Calibration of Photon and Electron Beams Set of 189 slides based on the chapter authored by P. Andreo, J.P. Seuntjens, and E.B. Podgorsak of the IAEA publication (ISBN 92-0-107304-6): Radiation
More informationNPL s progress towards absorbed dose standards for proton beams
NPL s rogress towards absorbed dose standards for roton beams H. Palmans 1 R. Thomas 1 D. Shiley 1 A. Kacerek 2 1 National Physical Laboratory Teddington United Kingdom 2 Clatterbridge Centre of Oncology
More informationHigh dose-per-pulse electron beam dosimetry A model to correct for the ion recombination in the Advanced Markus ionization chamber
High dose-per-pulse electron beam dosimetry A model to correct for the ion recombination in the Advanced Markus ionization chamber Kristoffer Petersson, a) Maud Jaccard, Jean-Francßois Germond, Thierry
More informationComparison of Primary Doses Obtained in Three 6 MV Photon Beams Using a Small Attenuator
Comparison of Primary Doses Obtained in Three 6 MV Photon Beams Using a Small Attenuator Christoph Trauernicht Groote Schuur Hospital & University of Cape Town Method is based on: Background A method of
More informationTITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137
TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 AUTHORS: J. Cardoso, L. Santos, C. Oliveira ADRESS: Instituto Tecnológico e Nuclear Estrada Nacional 10; 2686-953 Sacavém;
More informationFOREWORD In 1987 the IAEA published a Code of Practice entitled Absorbed Dose Determination in Photon and Electron Beams: An International Code of
IAEA-TECDOC-1173 Review of data and methods recommended in the international code of practice for dosimetry IAEA Technical Reports Series No. 381, The Use of Plane Parallel Ionization Chambers in High
More informationA Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators
A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators N. Golnik 1, P. Kamiński 1, M. Zielczyński 2 1 Institute of Precision and Biomedical Engineering,
More informationDiamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Diamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams Ramamoorthy Ravichandran,
More informationCHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR
ANALELE STIINTIFICE ALE UNIVERSITATII AL. I. CUZA IASI Tomul II, s. Biofizică, Fizică medicală şi Fizica mediului 2006 CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR Dan
More informationStudy of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 1, WINTER 2008 Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration Pablo
More informationLarry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical Physics University of Wisconsin
Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical Physics University of Wisconsin NCCAAPM meeting April 17, 2015 Larry DeWerd has partial interest in Standard Imaging Inc. Determination of your uncertainty
More information7. a XV-2 high spatial resolution lm detector (Kodak). Important parameters of these detectors are given in Table1. The ionization chambers and the di
Proceedings of the Second International Workshop on EGS, 8.-12. August 2000, Tsukuba, Japan KEK Proceedings 200-20, pp.264-271 Variation of Dose Distribution by Detectors for Narrow Beam T. Fujisaki, H.
More informationEvaluation of different methods for determining the magnitude of initial recombination in ionization chambers
Evaluation of different methods for determining the magnitude of initial recombination in ionization chambers Klaus Derikum Physikalisch-Technische undesanstalt undesallee 100, 38116 raunschweig, Germany
More informationSTANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS
STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS M.A. HASSAN*, M.H. GABER**, E. ESMAT*, H.I. FARAG***, H.M. EISSA* *National Institute for Standards (NIS), Giza, Egypt **Biophysics Department,
More informationProgress in calculations of k Q for TG-51
1 Progress in calculations of k Q for TG-51 D. W. O. Rogers Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University Ottawa http://www.physics.carleton.ca/~drogers AAPM Charlotte,NC,
More informationStudy of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water
More informationThe EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1
The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1 1 Medical Radiation Physics Group, University of Oldenburg and Pius
More informationM [scale units/s] of the system
APPENDIX TO IAEA CALIBRATION CERTIFICATE RADIATION PROTECTION IONIZATION CHAMBER CALIBRATION PROCEDURES AT THE IAEA DOSIMETRY LABORATORY 1. INTRODUCTION 1.1 General Ionization chambers and electrometers
More informationCOMPARISON OF ABSORBED DOSE TO AIR CALIBRATION FACTORS FOR A PARALLEL PLATE IONIZATION CHAMBER*
Original Article COMPARISON OF ABSORBED DOSE TO AIR CALIBRATION FACTORS FOR A PARALLEL PLATE IONIZATION CHAMBER* Roseli T. Bulla 1, Linda V.E. Caldas 2 * Study developed at Instituto de Pesquisas Energéticas
More informationElectron therapy Class 2: Review questions
Electron therapy Class 2: Review questions 1 Raphex Question: T63, 2002 In what situation is electron backscatter likely to be a problem? A. Using 1cm of tissue equivalent bolus on the skin. B. Using a
More informationABSTRACT. Keywords: Megavoltage, dosimetry, TG51 protocol, TG21 protocol, parallel-plate chambers, crosscomparison. INTRODUCTION
CALCULATED TG51 TO TG21 ABSORBED-DOSE RATIOS FOR MOST WIDELY USED CYLINDRICAL AND PARALLEL-PLATE ION CHAMBERS OVER A RANGE OF PHOTON AND ELECTRON ENERGIES R.C. Tailor, and W.F. Hanson ABSTRACT The change
More informationELECTRON INTERACTIONS
Applications of Electron beam therapy: The principal applications are (a) the treatment of skin and lip cancers, (b) chest wall irradiation for breast cancer, (c) administering boost dose to nodes, and
More informationPhoton-beams monitor-unit calculations
Photon-beams monitor-unit calculations Narayan Sahoo March 24, 2011 The materials included in this lecture notes are from previous lecture notes for this course by Karl Prado, Ph.D. Introduction Standard
More informationOptimization of microionization chambers for small-field reference dosimetry
Optimization of microionization chambers for small-field reference dosimetry Jessica R. Snow University of Wisconsin-Madison Department of Medical Physics Medical Radiation Research Center NCCAAPM Fall
More informationComparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2
Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST Ronaldo Minniti 1 and Ladislav Czap 2 1 National Institute of Standards and Technology (NIST), Gaithersburg,
More informationAn introduction to IAEA TRS-483
An introduction to IAEA TRS-483 P Andreo, Professor of Medical Radiation Physics Karolinska University Hospital and Karolinska Institute, Stockholm, Sweden Journées Scientifiques de la SFPM Toulouse 2018
More informationGerhard Stucki, Sandor VörösV
idgenössisches Justiz- und Polizeidepartement JPD Bundesamt für Metrologie MTAS xperimental k Q,Q0 lectron Beam Quality Correction Factors for the Types NACP02 and PTW34001 Plane-parallel Chambers Gerhard
More informationGeneral Overview of Gas Filled Detectors
GAS-FILLED DETECTOR General Overview of Gas Filled Detectors Gas-Filled Detectors Ion chamber Proportional counter G-M (Geiger-Miller) counter Diagram of a Generic Gas-Filled Detector A Anode High-voltage
More information1.1.4 What are the differences between the Varian 600C and the Siemens KD2?
1 Radiotherapy 1.1 Linear Accelerators 1.1.1 Sketch the main components of a linear accelerator. 1.1.2 Describe the theory of beam production and the part each component plays in the production of a clincally
More informationKCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck
Bilateral Comparisons of Air Kerma and Absorbed Dose to Water standards in 60 Co radiation beams for radiotherapy and Air Kerma standards in 137 Cs radiation beams for radiation protection between BEV
More informationfactors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons
Calorimetric determination of k Q factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons PTB s water calorimeter in front of Elekta Precise medical linac Calorimetric determination
More informationGeneral characteristics of radiation dosimeters
General characteristics of radiation dosimeters and a terminology to describe them D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers
More informationDUAL NATURE OF RADIATION AND MATTER
Chapter Eleven DUAL NATURE OF RADIATION AND MATTER MCQ I 111 A particle is dropped from a height H The de Broglie wavelength of the particle as a function of height is proportional to (a) H (b) H 1/2 (c)
More informationERRATA LIST AND UPDATES TO IAEA TRS-398 (2000) 1
ERRATA LIST AND UPDATES TO IAEA TRS-398 (2000) 1 1. page 034 Table 3 (cont) Chambers SNC 100730 and 100740 are replaced, respectively, by model numbers SNC 100700-0 and 100700-1. Their radii are changed
More informationMonte Carlo commissioning of photon beams in medical LINACS using wide-field profiles in a water phantom
Monte Carlo commissioning of photon beams in medical LINACS using wide-field profiles in a water phantom F.Gómez 1, L. Franco 1, A. Iglesias 1, J. Pardo 1, J. Pena 1, A. Rodríguez 1, R.Lobato 2, J. Mosquera
More informationR. C. Tailor and W. F. Hanson. Department of Radiation Physics. The University of Texas, M.D. Anderson Cancer Center. Houston, Texas 77030
Page 1 of 231 Calculated absorbed-dose ratios, TG51/TG21, for most idely used cylindrical and parallel-plate ion chambers over a range of photon and electron energies R. C. Tailor and W. F. Hanson Department
More information4.1b - Cavity Theory Lecture 2 Peter R Al mond 2011 Overview of Lecture Exposure (W/e)air Exposure Exposure and and and Air Air Kerma
4.1b - Cavity Theory Lecture 2 Peter R Almond 2011 Overview of Lecture Exposure (W/e) air Exposure and Air Kerma Exposure Exposure is symbolized as X and defined by the ICRU as the quotient of dq by dm,
More informationComposite field dosimetry
Composite field dosimetry Hugo Bouchard, PhD, MCCPM Senior Research Scientist Radiation dosimetry group National Physical Laboratory May 2014 Overview 1. Introduction Dosimetry protocols IAEA formalism
More informationEfficiencies of Some Spherical Ion Chambers in Continuous and Pulsed Radiation: A Numerical Evaluation
Signature: Pol J Radiol, 05; 80: 55-5 DOI: 0.659/PJR.89450 ORIGINAL ARTICLE Received: 05.03.7 Accepted: 05.06.9 Published: 05..5 Authors Contribution: A Study Design B Data Collection C Statistical Analysis
More informationIonizing Radiation Dosimetry and Medical Physics
Ionizing Radiation Dosimetry and Medical Physics D.W.O. Rogers Ionizing Radiation Standards, Institute for National Measurement Standards NRC, Ottawa, K1A OR6 Present e-mail: drogers at physics.carleton.ca
More informationESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT
Proceedings of the Eleventh EGS4 Users' Meeting in Japan, KEK Proceedings 2003-15, p.107-113 ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT K. Noto and
More informationKey comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation
Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation C. Kessler 1, D.T. Burns 1, S. Vörös 2 and B. Hofstetter-Boillat 2 1
More informationABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS
ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS C. CAPORALI, AS. GUERRA, R.F. LAITANO, M. PIMPINELLA ENEA-Casaccia, Departimento
More informationMetrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT)
1- Ionizing radiation metrology for radiation protection 2- Metrological requirements for ionizing radiation measurement in radiotherapy and radiodiagnostics R. F. Laitano Part 2 Metrological traceability
More informationTutorial: simulating a rod pinch diode for pulsed radiography with Trak and GamBet
Tutorial: simulating a rod pinch diode for pulsed radiography with Trak and GamBet Stanley Humphries, Copyright 2012 Field Precision PO Box 13595, Albuquerque, NM 87192 U.S.A. Telephone: +1-505-220-3975
More information1 Introduction. A Monte Carlo study
Current Directions in Biomedical Engineering 2017; 3(2): 281 285 Sebastian Richter*, Stefan Pojtinger, David Mönnich, Oliver S. Dohm, and Daniela Thorwarth Influence of a transverse magnetic field on the
More informationBureau International des Poids et Mesures
Bureau International des Poids et Mesures Comparison of the air-kerma standards of the ENEA-INMRI and the BIPM in the low-energy x-ray range D.T. Burns, M.P. Toni and M. Bovi November 1999 Pavillon de
More informationDetermination of absorbed dose to water for 60Co by the scaling theorem. M. Boutillon and A.-M. Perroche (1)
Rapport BIPM92/1 Determination of absorbed dose to water for 60Co by the scaling theorem M. Boutillon and A.M. Perroche (1) Bureau International des Poids et Mesures, F92312 Sevres Cedex Abstract A simple
More informationChapiter VII: Ionization chamber
Chapiter VII: Ionization chamber 1 Types of ionization chambers Sensitive volume: gas (most often air direct measurement of exposure) ionization chamber Sensitive volume: semiconductor (silicon, germanium,
More informationShielding of Ionising Radiation with the Dosimetry & Shielding Module
Shielding of Ionising Radiation with the Dosimetry & Shielding Module J. Magill Overview Biological Effects of Ionising Radiation - Absorber dose, Quality or Weighting Factor, Equivalent Dose Attenuation
More informationContents. Charged Particles. Coulomb Interactions Elastic Scattering. Coulomb Interactions - Inelastic Scattering. Bremsstrahlung
Contents Marcel MiGLiERiNi Nuclear Medicine, Radiology and Their Metrological Aspects. Radiation in Medicine. Dosimetry 4. Diagnostics & Therapy 5. Accelerators in Medicine 6. Therapy Planning 7. Nuclear
More informationUse of a radioactive check device for redundancy check of ionization chambers
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Use of a radioactive check device for redundancy check of ionization chambers N. P. S. Sidhu,* Alkis Rouvas, and Patrick Cadman
More informationObjectives. Tolerance Limits and Action Levels for IMRT QA. The Overall Process of IMRT Planning and Delivery. Chain of IMRT Process
Tolerance Limits and Action Levels for IMRT QA Jatinder R Palta, Ph.D Siyong Kim, Ph.D. Department of Radiation Oncology University of Florida Gainesville, Florida Objectives Describe the uncertainties
More informationpp physics, RWTH, WS 2003/04, T.Hebbeker
3. PP TH 03/04 Accelerators and Detectors 1 pp physics, RWTH, WS 2003/04, T.Hebbeker 2003-12-16 1.2.4. (Inner) tracking and vertexing As we will see, mainly three types of tracking detectors are used:
More informationRadiation Dose, Biology & Risk
ENGG 167 MEDICAL IMAGING Lecture 2: Sept. 27 Radiation Dosimetry & Risk References: The Essential Physics of Medical Imaging, Bushberg et al, 2 nd ed. Radiation Detection and Measurement, Knoll, 2 nd Ed.
More informationMonitor Unit Calculations for Photon and Electrons. AAMD Meeting Raleigh, NC October 3, John P. Gibbons Chief of Clinical Physics
Monitor Unit Calculations for Photon and Electrons AAMD Meeting Raleigh, NC October 3, 2014 John P. Gibbons Chief of Clinical Physics Outline I. TG71 Formation and Charge II. Photon Calculations III. Electron
More informationDetermination of contributions of scatter and distance error to the source strength of 192 Ir HDR brachytherapy source
Polish Journal of Medical Physics and Engineering 2016;22(3):55-59 September 2016 doi: 10.1515/pjmpe-2016-0010 ISSN 1898-0309 Scientific Paper Determination of contributions of scatter and distance error
More informationRecombination factor dependancy of high and low dose pulsed accelerators for electron beam energies
Recombination factor dependancy of high and low dose plsed accelerators for electron beam energies Giseppe Felici Sordina IORT Technologies S.p.A. Talk Overview Dose per plse what is the so called high
More informationDosimetry and beam calibration
Dosimetry and beam calibration Hugo Palmans 1,2 and Stanislav Vatnitksy 1 1 EBG MedAustron GmbH, Wiener Neustadt, Austria 2 National Physical Laboratory, Teddington, UK 1 Overview - Learning objectives
More informationComparison of the standards of air kerma of the GUM and the BIPM for 60Co 'Y rays
Rapport BIPM-9712 Comparison of the standards of air kerma of the GUM and the BIPM for 60Co 'Y rays by P.J. Allisy-Roberts and M. Boutillon Bureau International des Poids et Mesures, F-92312 Sevres Cedex,
More informationHeuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong
Original Article PMP Progress in Medical Physics 28(2), June 2017 https://doi.org/10.14316/pmp.2017.28.2.49 pissn 2508-4445, eissn 2508-4453 Electron Energy Distribution for a Research Electron LINAC Heuijin
More informationManipulation on charged particle beam for RT benefit.
High Electron Beam Dose Modification using Transverse Magnetic Fields Ion Chamber Response Modification under Strong Magnetic Field Conditions Sion Koren, Radiation Oncology Preface Manipulation on charged
More informationA comparison of methods for monitoring photon beam energy constancy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 A comparison of methods for monitoring photon beam energy constancy Song Gao, 1a Peter A. Balter, 1 Mark Rose, 2 and William E. Simon
More informationChapter 8 Electron Beams: Physical and Clinical Aspects
1 Chapter 8 Electron Beams: Physical and Clinical Aspects This set of 91 slides is based on Chapter 8 authored by W. Strydom, W. Parker, and M. Olivares of the IAEA publication (ISBN 92-0-107304-6): Radiation
More informationChapter V: Cavity theories
Chapter V: Cavity theories 1 Introduction Goal of radiation dosimetry: measure of the dose absorbed inside a medium (often assimilated to water in calculations) A detector (dosimeter) never measures directly
More informationComparison of the standards of air kerma of the BNM-LPRI and the BIPM for 137CS y rays
Rapport BIPM-96/6 Comparison of the standards of air kerma of the BNM-LPRI and the BIPM for 137CS y rays by P.J. Allisy-Roberts and M. Boutillon Bureau International des Poids et Mesures, F-92312 Sevres
More informationVariations in daily quality assurance dosimetry from device levelling, feet position and backscatter material
Australas Phys Eng Sci Med (2012) 35:485 489 DOI 10.1007/s13246-012-0169-6 SCIENTIFIC NOTE Variations in daily quality assurance dosimetry from device levelling, feet position and backscatter material
More informationMeasurement of induced radioactivity in air and water for medical accelerators
Measurement of induced radioactivity in air and water for medical accelerators K. Masumoto 1, K. Takahashi 1, H. Nakamura 1, A. Toyoda 1, K. Iijima 1, K. Kosako 2, K. Oishi 2, F. Nobuhara 1 High Energy
More informationComment on Proton beam monitor chamber calibration
Physics in Medicine & Biology COMMENT on Proton beam monitor chamber calibration To cite this article: Hugo Palmans and Stanislav M Vatnitsky 216 Phys. Med. Biol. 61 6585 View the article online for updates
More informationCommissioning of the Beta Secondary Standard (BSS2)
Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,
More informationRadiation Protection Dosimetry (2006), Vol. 118, No. 3, pp Advance Access publication 6 October 2005
Radiation Protection Dosimetry (2006), Vol. 118, No. 3, pp. 233 237 Advance Access publication 6 October 2005 doi:10.1093/rpd/nci353 DOSE BUILD UP CORRECTION FOR RADIATION MONITORS IN HIGH-ENERGY BREMSSTRAHLUNG
More informationBEAMnrc: a code to simulate radiotherapy external beam sources
BEAMnrc: a code to simulate radiotherapy external beam sources D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University Ottawa, Canada http://www.physics.carleton.ca/~drogers
More informationPhotoelectric effect
Laboratory#3 Phys4480/5480 Dr. Cristian Bahrim Photoelectric effect In 1900, Planck postulated that light is emitted and absorbed in discrete but tiny bundles of energy, E = hν, called today photons. Here
More informationInstructions for Use for the transfer chamber and the electronic measuring device Revision 1
EURAMET supplementary comparison of the personal dose equivalent at 0.07 mm and 3 mm depth, H p (0.07) and H p (3), for beta radiation EURAMET Project No. 1398 BIPM KCDB: EURAMET.R(I)-S16 Instructions
More informationNon-radioactive radiation sources. Lesson FYSKJM4710 Eirik Malinen
Non-radioactive radiation sources Lesson FYSKJM4710 Eirik Malinen X-ray tube Electrons are released from the cathode (negative electrode) by thermionic emission accelerated in an evacuated tube hit the
More information3. Gas Detectors General introduction
3. Gas Detectors 3.1. General introduction principle ionizing particle creates primary and secondary charges via energy loss by ionization (Bethe Bloch, chapter 2) N0 electrons and ions charges drift in
More information8/4/2016. Determination of small field output factors, advantages and limitations of Monte Carlo simulations
etermination of small field output factors, advantages and limitations of onte Carlo simulations AAP-APR-SEF Joint Scientific Symposium Josep Puxeu Vaqué Institut Català d Oncologia ICO Edinburgh Cancer
More informationSimulation Modeling in Dosimetry
Simulation Modeling in Dosimetry Aleksei Zhdanov Ural Federal University named after the first President of Russia B. N. Yeltsin, Yekaterinburg, Russian Federation jjj1994@yandex.ru Leonid Dorosinskiy
More informationSpecific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements
Specific Accreditation Criteria Calibration ISO/IEC 17025 Annex Ionising radiation measurements January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected
More information