Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation

Size: px
Start display at page:

Download "Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation"

Transcription

1 Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation C. Kessler 1, D.T. Burns 1, S. Vörös 2 and B. Hofstetter-Boillat 2 1 Bureau International des Poids et Mesures, F Sèvres Cedex 2 Federal Institute of Metrology (METAS), Bern, Switzerland Abstract An indirect comparison has been made of the standards for absorbed dose to water in 60 Co radiation of the Federal Institute of Metrology (METAS), Switzerland and of the Bureau International des Poids et Mesures (BIPM). The measurements at the BIPM were carried out in June The comparison result, based on the calibration coefficients for two transfer standards and evaluated as a ratio of the METAS and the BIPM standards for absorbed dose to water, is with a combined standard uncertainty of The results are analysed and presented in terms of degrees of equivalence for entry in the BIPM key comparison database. 1. Introduction An indirect comparison has been made of the standards for absorbed dose to water in 60 Co radiation of the Federal Institute of Metrology (METAS), Switzerland and of the Bureau International des Poids et Mesures (BIPM) to update the previous comparison result of 2000 (Allisy Roberts et al 2003) published in the BIPM.RI(I)-K4 key comparison database (KCDB 2014). The comparison was undertaken using two ionization chambers of the METAS as transfer standards. The measurements at the BIPM were carried out in June 2013 and final results were supplied by the METAS in July Details of the standards The METAS determination of absorbed dose to water is based on a water calorimeter (Domen 1994). The BIPM primary standard is a parallel-plate graphite cavity ionization chamber positioned at the reference depth in a water phantom (Boutillon et al 1993). 3. Determination of the absorbed dose to water At the BIPM the absorbed dose to water rate is determined using a cavity ionization chamber, with measuring volume V, by the relation I W D a w, BIPM ( µ en ) w,c sc, aπk i, (1) V e where a I W a a is the density of air under reference conditions, is the ionization current measured by the standard, is the mean energy expended in dry air per ion pair formed, 1/10

2 e is the electronic charge, (µ en / ) w,c is the ratio of the mean mass energy-absorption coefficients for water and graphite s c,a k i is the ratio of the mean mass stopping powers for graphite and air, and is the product of the correction factors to be applied to the standard. The values for the physical constants, the correction factors entering in equation (1), the volume of the BIPM primary standard and the associated uncertainties for the standard are given in Table 1 (Allisy-Roberts et al 2011). Table 1. Physical constants, correction factors and relative standard uncertainties for the BIPM ionometric standard for absorbed dose to water Symbol Parameter / unit Value Physical constants Relative standard uncertainty (1) 100 s i 100 u i a dry air density (0 C, kpa) / kg m ( en / ) w,c ratio of mass energy-absorption coefficients (2) 0.01 (2) 0.14 (2) s c,a ratio of mass stopping powers (3) W a /e mean energy per charge / J C Correction factors k p fluence perturbation (includes ( en / ) w,c ) (2) 0.05 (2) 0.17 (2) k ps polythene envelope of the chamber k pf front face of the phantom k rn radial non-uniformity k s saturation k h humidity Measurement of I / effective volume of CH4-1/ cm I ionization current (T, P, air compressibility) 0.02 short-term reproducibility (including positioning and current measurement) 0.02 Combined uncertainty of the BIPM determination of absorbed dose to water rate quadratic summation combined relative standard uncertainty 0.29 (1) expressed as one standard deviation. s i represents the relative uncertainty estimated by statistical methods, type A u i represents the relative uncertainty estimated by other methods, type B. (2) the ratio (µ en / ) w,c is included in k p and likewise its uncertainty is included in the uncertainty for k p. (3) uncertainty value for the product s c,a W/e. At the METAS, absorbed dose to water is based on a water calorimeter, described by Medin et al (1999). The detector is a glass vessel filled with ultra-pure water, in which 2 NTC probes measure the temperature increase due to the absorbed radiation energy, positioned in a thermally isolated cubic water phantom of 30 cm side length. The absorbed dose to water is determined using the equation D w, METAS C p T k i (2) 2/10

3 where C p ΔT k i is the specific heat capacity at constant pressure, is the measured temperature increase, and is the product of the correction factors to be applied to the standard. The relative standard uncertainty of the METAS absorbed dose to water determination is given in Table 2. Table 2. Relative standard uncertainties for the METAS standard for absorbed dose to water Symbol Parameter / unit Value Relative standard uncertainty (1) 100 s i 100 u i Water calorimetry Thermistor calibration Thermistor self-heating / K W Calibration of bridge sensitivity - <0.01 <0.01 Absolute temperature from PT100 - < Irradiation duration ( 60 Co) - <0.01 <0.01 k p Field perturbation ~ k rho Density k c Conductive heat flow k v Convection 1 k t Transient thermistor response <0.01 Depth in water Lateral position Field size k HD Heat defect C w Specific heat capacity of water / J kg 1 K Reproducibility 0.02 Combined uncertainty of the METAS determination of absorbed dose to water rate quadratic summation combined relative standard uncertainty 0.41 (1) expressed as one standard deviation. s i represents the relative uncertainty estimated by statistical methods, type A u i represents the relative uncertainty estimated by other methods, type B. The correction factors, k i, applied in (2) are the following: k p k c is the effect of the perturbation of the radiation field due to the presence of the glass vessel and NTC probes; it was measured for each different vessel using a diode located at the reference depth, corrects for conductive heat flow due to the presence of glass, which has a different heat capacity from that of water, and to the presence of temperature gradients due to non-uniformities in the absorbed dose distribution; it was calculated using finite elements method software, 3/10

4 k HD is the heat defect due to endothermic or exothermic chemical reactions in the water, which has been calculated using various models to be zero in pure water saturated with H 2, N 2 or Ar (Klassen and Ross 1991; Klassen and Ross 1997). The measurement conditions are chosen in such a way that the remaining factors are taken to be strictly unity, or with the uncertainty stated in Table 2: k rho is the effect of the different water densities in calorimetric measurements at 4 C and in secondary standard measurements at 20 C (the depth of measurement being determined in the units g cm 2 at both temperatures and hence no additional correction is needed), k v corrects for convective heat flow due to the temperature gradients in the water (convection does not take place when measuring at 4 C, the temperature of maximum water density), k t is the effect of transient behaviour of the thermistor just after the end of irradiation, which might affect the post-irradiation drift curve (data during the first 20 s after irradiation are discarded from the fit to avoid this effect). Reference conditions Absorbed dose is determined at the BIPM under reference conditions defined by the Consultative Committee for Ionizing Radiation (CCEMRI 1985): the distance from the source to the reference plane (centre of the detector) is 1 m; the beam size in air at the reference plane is 10 cm 10 cm, the photon fluence rate at the centre of each side of the square being 50 % of the photon fluence rate at the centre of the square; and the reference depth in the water phantom is 5 g cm 2. The reference conditions at the METAS are the same as those at the BIPM. The transfer standards used in this comparison were measured in the same water phantom as that used for water calorimetry, a cubic tank of 30 cm side length. The source to reference distance is 100 cm with the depth to the reference point in the tank set to 5 g cm 2, including the Polystyrene window of 2.6 mm thickness as a water-equivalent thickness in g cm 2. The field size at this reference point is 10 cm x 10 cm. There is 10 cm of Styrofoam insulation in the beam path outside the tank that is not included in the 5 g cm 2, but is the same for reference and transfer measurements. Reference values The BIPM reference absorbed dose to water rate D w, BIPM is taken as the mean of the four measurements made around the period of the comparison, corrected to the reference date of , 0 h UTC, as is the ionization current of the transfer chambers. The half-life of 60 Co is taken as days (u = 0.29 days) (Bé et al 2006). The value of D w, METAS used for the comparison was obtained as the mean of calorimetric measurements made between December 2012 and August 2013 with the water calorimeter, all calorimetric and ionometric measurements corrected to the reference date The half-life is also taken as days (u = 0.29 days). 4. Experimental method The comparison of the METAS and BIPM standards was made indirectly using the calibration coefficients N for two transfer chambers given by Dw 4/10

5 , (3) ND w, lab Dw,lab Ilab where D w, lab is the water absorbed dose rate at the METAS or the BIPM and I lab is the corresponding ionization current for a transfer chamber measured under reference conditions at that laboratory. The experimental method for measurements at the BIPM is described by Allisy-Roberts et al (2011); the essential details for the determination of the calibration coefficients N Dw for the transfer chambers are reproduced here. The ionization chambers NE 2571 and NE 2611, serial numbers 2807 and 147, respectively, both belonging to the METAS, were used as transfer chambers for this comparison. Their main characteristics are listed in Table 3. These chambers were calibrated twice (in May 2013 and January 2014) at the METAS. Each chamber was inserted in its own PMMA waterproof sleeve 1 mm thick, provided by the METAS. Table 3. Characteristics of the METAS transfer chambers Characteristic/Nominal values NE 2571 NE 2611 Dimensions Inner diameter / mm Wall mass thickness / g cm Cavity length / mm Tip to reference point / mm Electrode Length / mm Diameter / mm Air cavity Volume / cm Wall Material graphite graphite Voltage applied / V Density / g cm Negative polarity to outer electrode at the BIPM Positive polarity to collector at the METAS Positioning At each laboratory the chambers were positioned with the stem perpendicular to the beam direction and with the same orientation (line and text on the stem of the chambers and line on the waterproof sleeves facing the source). Applied voltage and polarity A collecting voltage as indicated in Table 3 was applied to each chamber at least 30 min before any measurements were made. No polarity correction was made as both laboratories apply the polarity in an equivalent way. Recombination loss Volume recombination is negligible at a dose rate of less than 20 mgy s 1 for these chamber types at this polarizing voltage, and the initial recombination loss will be the same in the two laboratories. Consequently, no correction for recombination was applied. Radial non-uniformity correction At the METAS, the correction for the radial non-uniformity of the beam over the section of the transfer chambers, relative to the section of the water calorimeter, is taken as and for the NE 2571 and NE 2611, respectively, with an uncertainty of At the 5/10

6 BIPM, the corresponding corrections applied to the ionization current for the transfer chambers NE 2571 and NE 2611 are and , respectively, with an uncertainty of Phantom window and waterproof sleeve Both laboratories use a horizontal radiation beam and the thickness of the PMMA (BIPM) and Polystyrene (METAS) front window of the water phantom is included as a water-equivalent thickness in g cm 2 when positioning the chamber. In addition, the BIPM applies a correction factor k pf (0.9996) that accounts for the non-equivalence to water of the PMMA in terms of interaction coefficients. Two waterproof sleeves of PMMA were supplied by the METAS for use with the chambers. The same sleeves were used at both laboratories and, consequently, no correction for the influence of the sleeve was necessary at either laboratory. Charge and leakage measurements The charge collected for each transfer chamber was measured using a Keithley electrometer, model 642 at the BIPM and model 6517 at the METAS. The chambers were pre-irradiated for at least 30 min ( 10 Gy) at the BIPM before any measurements were made. At the METAS, the pre-irradiation lasted only 10 min to achieve the same 10 Gy. The ionization current measured for each transfer chamber was corrected for the leakage current; at both laboratories, this correction was less than in relative value. Ambient conditions During a series of measurements, the water temperature is measured for each current measurement and was stable to better than 0.01 C at the BIPM and to better than 0.05 C at the METAS. The ionization current is corrected to the reference conditions of K and kpa at both laboratories. Relative humidity is controlled at (50 5) % both at the BIPM and at the METAS. Consequently, no correction for humidity is applied to the transfer chamber ionization current measured at either laboratory. 5. Results of the comparison The NE transfer chambers were set-up and measured in the BIPM 60 Co beam on three separate occasions. The reproducibility of the results was around At the METAS, the transfer chambers were measured in May 2013, just before sending them to the BIPM and in August On the first occasion, 5 series of measurements were performed, and 3 series on the second occasion, each series consisting of 10 leakage current measurements followed by 25 measurements during irradiation and finally 10 more leakage current measurements, each measurement lasting for 60 s. The reproducibility of the results over the 8 months separating the METAS measurements was around The result of the comparison, R Dw Dw, is expressed in the form R N (4) D METAS, BIPM, w, N Dw in which the average value of measurements made at the METAS before and after those made at the BIPM is compared with the mean of the measurements made at the BIPM. The results for each chamber are presented in Table 4. Contributions to the relative standard uncertainty of combined standard uncertainty u c for the comparison result N D w, lab are listed in Table 5 and the R D is presented in Table 6. w 6/10

7 Table 4. Final result of the METAS/BIPM comparison of standards for 60 Co absorbed dose to water Transfer chamber N D w, METAS / Gy µc 1 N D w, BIPM R / Gy µc 1 Dw u c NE NE Mean values Table 5. Uncertainties associated with the calibration of the transfer chambers BIPM METAS Relative standard uncertainty 100 s i 100 u i 100 s i 100 u i Absorbed dose to water rate Ionization current for the transfer chambers (1) < Temperature < Pressure < Humidity 0.03 Distance 0.02 Lateral position 0.03 Depth in water Field size 0.03 Short-term stability 0.01 <0.01 Recombination loss k s,tr 0.09 Radial non-uniformity Relative standard uncertainties of N D w, lab Quadratic summation Combined uncertainty (1) includes the uncertainty of the BIPM temperature and pressure normalization. Table 6. Uncertainties associated with the comparison result Relative standard uncertainty 100 s i 100 u i N N (1) Dw, METAS Dw, BIPM Relative standard uncertainties of (1) R D, w Takes account of correlation in type B uncertainties (k s and k h ) u c = /10

8 The comparison result is taken as the unweighted mean of the results for the two transfer chambers, R D, w = , with a combined standard uncertainty for the comparison of demonstrating the agreement between the two standards for absorbed dose to water. 6. Degrees of equivalence Following a decision of the CCRI, the BIPM determination of the dosimetric quantity, here D w,bipm, is taken as the key comparison reference value (KCRV) (Allisy-Roberts et al 2009). It follows that for each NMI i having a BIPM comparison result x i with combined standard uncertainty u i, the degree of equivalence with respect to the reference value is the relative difference D i = (D wi D w,bipmi )/ D w,bipmi = x i 1 and its expanded uncertainty U i = 2 u i. The results for D i and U i are usually expressed in mgy/gy. Table 7 gives the values for D i and U i for each NMI, i, taken from the KCDB of the CIPM MRA (1999) and this report. These data are presented graphically in Figure 1. Table 7. Degrees of equivalence For each laboratory i, the degree of equivalence with respect to the key comparison reference value is the difference D i and its expanded uncertainty U i. Tables formatted as they appear in the BIPM key comparison database BIPM.RI(I)-K4 - SIM.RI(I)-K4 (2002) - EUROMET.RI(I)-K4 (2005 to 2008) D i U i D i U i Lab i / (mgy/gy) Lab i / (mgy/gy) MKEH CIEMAT PTB CMI VSL RMTC ENEA SSM NPL STUK BEV NRPA VNIIFTRI SMU NRC IAEA NMIJ HIRCL ARPANSA ITN LNE-LNHB NIST METAS LNMRI CNEA ININ When required, the degree of equivalence between two laboratories i and j can be evaluated as the difference D ij = D i D j = x i x j and its expanded uncertainty U ij = 2 u ij, both expressed in mgy/gy. In evaluating u ij, account should be taken of correlation between u i and u j Following the advice of the CCRI(I) in 2011, results for D ij and U ij are no longer published in the KCDB. 8/10

9 MKEH PTB VSL ENEA NPL BEV VNIIFTRI NRC NMIJ ARPANSA LNE-LNHB METAS ININ CIEMAT CMI RMTC SSM STUK NRPA SMU IAEA HIRCL ITN NIST LNMRI CNEA Metrologia 52 (2015) Tech. Suppl Figure Graph of the degrees of equivalence with the KCRV BIPM.RI(I)-K4, 2002 SIM.RI(I)-K4 and 2005 to 2008 EUROMET.RI(I)-K4 Degrees of equivalence for absorbed dose to water Di / (mgy / Gy) N.B. Black squares indicate results that are more than 10 years old. 7. Conclusions A key comparison has been carried out between the METAS and the BIPM standards for absorbed dose to water in 60 Co gamma rays, using two ionization chambers as transfer instruments. The comparison result is evaluated as the ratio of the calibration coefficients measured by the METAS and the BIPM. The comparison result is (52) and so the METAS standard is in agreement with the KCRV within the standard uncertainty for the comparison. The result of the comparison made in 2000 was (54), in good agreement with the present result. When compared with the results for the other laboratories that have carried out comparisons in terms of absorbed dose to water at the BIPM, the METAS standard for absorbed dose to water is in good agreement. Note that the data presented in the tables, while correct at the time of publication of the present report, become out of date as laboratories make new comparisons with the BIPM. The formal results under the CIPM MRA are those available in the BIPM key comparison database. 9/10

10 References Allisy-Roberts P J, Burns D T and Stucki G 2003 Comparison of the standards for absorbed dose to water of the METAS and the BIPM for 60 Co gamma radiation Rapport BIPM- 03/02 Allisy-Roberts P J, Burns D T and Kessler C 2011 Measuring conditions and uncertainties for the comparison and calibration of national dosimetric standards at the BIPM Rapport BIPM-11/04 Allisy P J, Burns D T and Andreo P 2009 International framework of traceability for radiation dosimetry quantities Metrologia, 2009, 46(2), S1-S8. Bé M M, Chisté V, Dulieu C, Browne E, Baglin C, Chechev V, Kuzmenco N, Helmer R, Kondev F, MacMahon D and Lee K B Table of Radionuclides (Vol. 3 A = 3 to 244) Monographie BIPM-5 Boutillon M and Perroche A-M 1993 Ionometric determination of absorbed dose to water for cobalt-60 gamma rays Phys. Med. Biol CCEMRI 1985 Constantes physiques pour les étalons de mesure de rayonnement BIPM Com. Cons. Etalons Mes. Ray. Ionisants (Section I) 11 R45 (Paris : Offilib) CIPM MRA: Mutual recognition of national measurement standards and of calibration and measurement certificates issued by national metrology institutes, International Committee for Weights and Measures, 1999, 45 pp Domen S R 1994 A sealed water calorimeter for measuring absorbed dose J. Res. Nat. Bur. Stand KCDB 2014 BIPM key comparison database 60 Co absorbed dose to water comparisons BIPM.RI(I)-K4 Klassen N V and Ross C K 1991 Absorbed dose calorimetry using various aqueous solutions Radiat. Phys. Chem Klassen N V and Ross C K 1997 Water calorimetry: The heat defect J. Res. Natl. Inst. Stand. Technol Medin J, Seuntjens J, Klassen N, Ross C K and Stucki G 1999 The OFMET sealed water calorimeter Proceedings of NPL Workshop on Recent advances in Calorimetric Absorbed Dose Standards NPL Report CIRM 42 10/10

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the LNE-LNHB, France and the BIPM in 60 Co gamma radiation

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the LNE-LNHB, France and the BIPM in 60 Co gamma radiation Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the LNE-LNHB, France and the BIPM in 60 Co gamma radiation C. Kessler 1, D.T. Burns 1, F. Delaunay 2, M. Donois 2 1 Bureau International des Poids

More information

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the NIM, China and the BIPM in 60 Co gamma radiation

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the NIM, China and the BIPM in 60 Co gamma radiation ey comparison BIPM.RI(I)-1 of the air-kerma standards of the NIM, China and the BIPM in 60 Co gamma radiation C. essler 1, D. Burns 1,. Wang 2,Y. Fan 2,S. Jin 2, X. Yang 2 1 Bureau International des Poids

More information

Comparison of the standards for absorbed dose to water of the ENEA-INMRI (Italy) and the BIPM for 60 Co γ rays

Comparison of the standards for absorbed dose to water of the ENEA-INMRI (Italy) and the BIPM for 60 Co γ rays Final Report 2010-01-12 Comparison of the standards for absorbed dose to water of the ENEA-INMRI (Italy) and the BIPM for 60 Co γ rays C. Kessler 1, P.J. Allisy-Roberts 1, D.T. Burns 1, A.S. Guerra 2,

More information

Comparison of the standards for absorbed dose to water of the VSL and the BIPM for 60 Co gamma rays

Comparison of the standards for absorbed dose to water of the VSL and the BIPM for 60 Co gamma rays Comparison of the standards for absorbed dose to water of the VSL and the BIPM for 60 Co gamma rays C. Kessler 1, P.J. Allisy-Roberts 1 D.T. Burns 1, P. Roger 1, L.A. de Prez 2, J.A. de Pooter 2 P.M.G.

More information

Comparisons of the standards for air kerma of the GUM and the BIPM for 60 Co and 137 Cs gamma radiation

Comparisons of the standards for air kerma of the GUM and the BIPM for 60 Co and 137 Cs gamma radiation Comparisons of the standards for air kerma of the GUM and the BIPM for 60 Co and 137 Cs gamma radiation P. J. Allisy-Roberts, C. Kessler, D.T. Burns Bureau International des Poids et Mesures, F-92312 Sèvres

More information

Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays

Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays C Kessler 1, D T Burns 1, P Roger 1, V Sochor 2 1 Bureau International des Poids et

More information

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Rapport BIPM 02/01 Bureau International des Poids et Mesures Comparison of the standards of air kerma of the Yugoslavia and the BIPM for 60 Co γ rays by P.J. Allisy-Roberts, D.T. Burns, C Kessler BIPM

More information

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NRC, Canada and the BIPM in medium-energy x-rays

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NRC, Canada and the BIPM in medium-energy x-rays Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NRC, Canada and the BIPM in medium-energy x-rays D T Burns 1, C Kessler 1, E Mainegra-Hing 2, H Shen 2 and M R McEwen 2 1 Bureau International

More information

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays D.T. Burns, W. Tiefenböck* and J. Witzani* Bureau International des Poids et Mesures, Pavillon

More information

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Bureau International des Poids et Mesures Comparison of the air-kerma standards of the VNIIM and the BIPM in the medium-energy x-ray range D.T. Burns, N.D. Villevalde, A.V. Oborin and E.N. Yurjatin September

More information

Comparison of the standards of air kerma of the GUM and the BIPM for 60Co 'Y rays

Comparison of the standards of air kerma of the GUM and the BIPM for 60Co 'Y rays Rapport BIPM-9712 Comparison of the standards of air kerma of the GUM and the BIPM for 60Co 'Y rays by P.J. Allisy-Roberts and M. Boutillon Bureau International des Poids et Mesures, F-92312 Sevres Cedex,

More information

Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays

Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays D T Burns Bureau International des Poids et Mesures, Pavillon de Breteuil, F-92312

More information

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden Referensdosimetri Crister Ceberg Medical Radiation Physics Lund University Sweden Reference dosimetry Determination of absorbed dose to water under reference conditions Not accounting for uncertainties

More information

Abstract An indirect comparison of the standards for reference air kerma rate for

Abstract An indirect comparison of the standards for reference air kerma rate for Comparison BIPM.RI(I)-K8 of high dose-rate Ir-19 brachyerapy standards for reference air kerma rate of e VSL and e BIPM J.T. Alvarez 1, J. A. de Pooter, C. Andersen 3, A.H.L. Aalbers, P.J. Allisy-Roberts

More information

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2 Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST Ronaldo Minniti 1 and Ladislav Czap 2 1 National Institute of Standards and Technology (NIST), Gaithersburg,

More information

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Bureau International des Poids et Mesures Comparison of the air-kerma standards of the ENEA-INMRI and the BIPM in the low-energy x-ray range D.T. Burns, M.P. Toni and M. Bovi November 1999 Pavillon de

More information

BIPM Key Comparison BIPM.RI(I)-K6 PROTOCOL 3.0 CCRI(I) 6 November 2017

BIPM Key Comparison BIPM.RI(I)-K6 PROTOCOL 3.0 CCRI(I) 6 November 2017 COMPARISON OF ABSORBED DOSE TO WATER AT HIGH PHOTON ENERGIES AT A REFERENCE FACILITY BIPM Key Comparison BIPM.RI(I)-K6 PROTOCOL 3.0 CCRI(I) 6 November 2017 Bureau International des Poids et Mesures Pavillon

More information

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck Bilateral Comparisons of Air Kerma and Absorbed Dose to Water standards in 60 Co radiation beams for radiotherapy and Air Kerma standards in 137 Cs radiation beams for radiation protection between BEV

More information

Comparison of the standards of air kerma of the BNM-LPRI and the BIPM for 137CS y rays

Comparison of the standards of air kerma of the BNM-LPRI and the BIPM for 137CS y rays Rapport BIPM-96/6 Comparison of the standards of air kerma of the BNM-LPRI and the BIPM for 137CS y rays by P.J. Allisy-Roberts and M. Boutillon Bureau International des Poids et Mesures, F-92312 Sevres

More information

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 AUTHORS: J. Cardoso, L. Santos, C. Oliveira ADRESS: Instituto Tecnológico e Nuclear Estrada Nacional 10; 2686-953 Sacavém;

More information

Determination of absorbed dose to water for 60Co by the scaling theorem. M. Boutillon and A.-M. Perroche (1)

Determination of absorbed dose to water for 60Co by the scaling theorem. M. Boutillon and A.-M. Perroche (1) Rapport BIPM92/1 Determination of absorbed dose to water for 60Co by the scaling theorem M. Boutillon and A.M. Perroche (1) Bureau International des Poids et Mesures, F92312 Sevres Cedex Abstract A simple

More information

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

More information

PRIMARY STANDARDS of AIR KERMA for 60 CO and X-RAYS & ABSORBED DOSE in PHOTON and ELECTRON BEAMS. Malcolm McEwen

PRIMARY STANDARDS of AIR KERMA for 60 CO and X-RAYS & ABSORBED DOSE in PHOTON and ELECTRON BEAMS. Malcolm McEwen PRIMARY STANDARDS of AIR KERMA for 60 CO and X-RAYS & ABSORBED DOSE in PHOTON and ELECTRON BEAMS Malcolm McEwen Definitions: Standard instrument/measurement/artifact intended to define, realize, conserve

More information

Activity measurements of the radionuclide 111 In for the LNE-LNHB, France in the ongoing comparison BIPM.RI(II)-K1.In-111

Activity measurements of the radionuclide 111 In for the LNE-LNHB, France in the ongoing comparison BIPM.RI(II)-K1.In-111 Activity measurements of the radionuclide 111 In for the LNE-LNHB, France in the ongoing comparison BIPM.RI(II)-K1.In-111 C. Michotte*, G. Ratel*, S. Courte*, E. Verdeau, M.-N. Amiot *BIPM, LNE-LNHB, France

More information

First international comparison of primary absorbed dose to water standards in the medium-energy X-ray range

First international comparison of primary absorbed dose to water standards in the medium-energy X-ray range First international comparison of primary absorbed dose to water standards in the medium-energy X-ray range Ludwig Büermann (a), Antonio Stefano Guerra (b), Maria Pimpinella (b), Massimo Pinto (b), Jacco

More information

C. Michotte*, G. Ratel* and S. Courte*, T. Dziel #, A. Listkowska #, I. Csete, K. Rózsa, L. Szücs, A. Zsinka, C. Fréchou**, C. Bobin**, S.

C. Michotte*, G. Ratel* and S. Courte*, T. Dziel #, A. Listkowska #, I. Csete, K. Rózsa, L. Szücs, A. Zsinka, C. Fréchou**, C. Bobin**, S. Update of the BIPM comparison BIPM.RI(II)-K1.Am-241 of activity measurements of the radionuclide 241 Am to include the 2009 results of the POLATOM (Poland), MKEH (Hungary) and the 2011 results of the LNE-LNHB

More information

Final report EURAMET project 813 1/46

Final report EURAMET project 813 1/46 Comparison of air kerma and absorbed dose to water measurements of 60 Co radiation beams for radiotherapy Report on EUROMET project no. 813, identifiers in the BIPM key comparison database (KCDB) are EUROMET.RI(I)-K1

More information

Activity measurements of the radionuclide 153 Sm for the ANSTO, Australia in the ongoing comparison BIPM.RI(II)-K1.Sm-153

Activity measurements of the radionuclide 153 Sm for the ANSTO, Australia in the ongoing comparison BIPM.RI(II)-K1.Sm-153 Activity measurements of the radionuclide 153 Sm for the ANSTO, Australia in the ongoing comparison BIPM.RI(II)-K1.Sm-153 G. Ratel*, C. Michotte*, M. Reinhard, D. Alexiev, L. Mo *BIPM, ANSTO, Australia

More information

APMP key comparison for the measurement of air kerma for 60 Co (APMP.RI(I)-K1.1)

APMP key comparison for the measurement of air kerma for 60 Co (APMP.RI(I)-K1.1) APMP key comparison for the measurement of air kerma for 60 Co (APMP.RI(I)-K1.1) D V Webb a, J-H Lee b, C T Budiantari c, J Laban d, N Saito e, S Srimanoroth f, N E Khaled g a Australian Radiation Protection

More information

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 J. Cardoso, L. Santos, C. Oliveira SUMMARY: i. The primary standard ii. The irradiation room iii. Experimental results iv. Simulation

More information

PRELIMINARY CHARACTERIZATION OF THE NIS FREE-AIR CHAMBER STANDARD AT THE BIPM

PRELIMINARY CHARACTERIZATION OF THE NIS FREE-AIR CHAMBER STANDARD AT THE BIPM Rapport BIPM-09/02 PRELIMINARY CHARACTERIZATION OF THE NIS FREE-AIR CHAMBER STANDARD AT THE BIPM D T Burns, C Kessler, P Roger BIPM A R El-Sersy NIS (Egypt) July 2009 BUREAU INTERNATIONAL DES POIDS ET

More information

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS C. CAPORALI, AS. GUERRA, R.F. LAITANO, M. PIMPINELLA ENEA-Casaccia, Departimento

More information

BIPM.RI(II)-K1.Am-241, CCRI(II)-K2.Am-241, and COOMET.RI(II)-K2.Am.241

BIPM.RI(II)-K1.Am-241, CCRI(II)-K2.Am-241, and COOMET.RI(II)-K2.Am.241 MEASURAND : Key comparison BIPM.RI(II)-K1.Am-241 x i : u i : Equivalent activity of 241 Am result of measurement carried out in the SIR for the sample submitted by laboratory i combined standard uncertainty

More information

H Bjerke 1, J Plagnard 2, J-M Bordy 2, A Kosunen 3, J Huikari 3, L Persson 4, and P O Hetland 1

H Bjerke 1, J Plagnard 2, J-M Bordy 2, A Kosunen 3, J Huikari 3, L Persson 4, and P O Hetland 1 Comparison of the air-kerma x-ray standards of the NRPA, the STUK, the SSM and the LNE-LNHB in the ISO 4037 narrow spectrum series in the range 40 kv to 300 kv H Bjerke 1, J Plagnard 2, J-M Bordy 2, A

More information

C. Michotte 1, G. Ratel 1, S. Courte 1, Y. Nedjadi 2, C. Bailat 2, L. Johansson 3, Y. Hino 4. Abstract

C. Michotte 1, G. Ratel 1, S. Courte 1, Y. Nedjadi 2, C. Bailat 2, L. Johansson 3, Y. Hino 4. Abstract Update of the BIPM comparison BIPM.RI(II)-K1.Ho-166m activity measurements to include the IRA and the NPL and a re-evaluation of the degrees of equivalence for the APMP.RI(II)-K2.Ho-166m comparison C.

More information

Electron beam water calorimetry measurements to obtain beam quality conversion factors

Electron beam water calorimetry measurements to obtain beam quality conversion factors Electron beam water calorimetry measurements to obtain beam quality conversion factors Bryan R. Muir, a) Claudiu D. Cojocaru, Malcolm R. McEwen, and Carl K. Ross Measurement Science and Standards, National

More information

M [scale units/s] of the system

M [scale units/s] of the system APPENDIX TO IAEA CALIBRATION CERTIFICATE RADIATION PROTECTION IONIZATION CHAMBER CALIBRATION PROCEDURES AT THE IAEA DOSIMETRY LABORATORY 1. INTRODUCTION 1.1 General Ionization chambers and electrometers

More information

Activity measurements of the radionuclide 99m Tc for the NIST, USA in the ongoing comparison BIPM.RI(II)-K4.Tc-99m. C. Michotte 1, R.

Activity measurements of the radionuclide 99m Tc for the NIST, USA in the ongoing comparison BIPM.RI(II)-K4.Tc-99m. C. Michotte 1, R. Activity measurements of the radionuclide 99m Tc for the NIST, USA in the ongoing comparison BIPM.RI(II)-K4.Tc-99m C. Michotte 1, R. Fitzgerald 2 1 Bureau International des Poids et Mesures (BIPM), 2 National

More information

Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams

Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams NEDERLANDSE COMMISSIE VOOR STRALINGSDOSIMETRIE Report 18 of the Netherlands Commission on Radiation Dosimetry

More information

Comments on ICRU Report 64: Dosimetry of High-Energy Photon Beams based on Standards of Absorbed Dose to Water

Comments on ICRU Report 64: Dosimetry of High-Energy Photon Beams based on Standards of Absorbed Dose to Water Comments on ICRU Report 64: Dosimetry of High-Energy Photon Beams based on Standards of Absorbed Dose to Water D.W.O. Rogers Ionizing Radiation Standards National Research Council of Canada Ottawa, K1A

More information

Updating reference dosimetry a decade after TG-51

Updating reference dosimetry a decade after TG-51 Updating reference dosimetry a decade after TG-51 Malcolm McEwen Ionizing Radiation Standards Institute for National Measurement Standards National Research Council, Canada CE Presentation at AAPM Annual

More information

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society of Radiobiology

More information

Bureau International des Poids et Mesures (BIPM),

Bureau International des Poids et Mesures (BIPM), Activity measurements of the radionuclides 18 F and 99m Tc for the NMISA, South Africa in the ongoing comparisons BIPM.RI(II)-K4.F-18 and BIPM.RI(II)-K4.Tc-99m (with erratum) C. Michotte 1, M. Nonis 1,

More information

G. Ratel 1, C. Michotte 1, D. Kryeziu 1, M. Moune 2, A. Iwahara 3. Abstract

G. Ratel 1, C. Michotte 1, D. Kryeziu 1, M. Moune 2, A. Iwahara 3. Abstract Update of the BIPM comparison BIPM.RI(II)-K1.Cr-51 to include new activity measurements for the LNE-LNHB (France) and the pilot study result of the LNMRI (Brazil) G. Ratel 1, C. Michotte 1, D. Kryeziu

More information

factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons

factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons Calorimetric determination of k Q factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons PTB s water calorimeter in front of Elekta Precise medical linac Calorimetric determination

More information

Comparisons between IRD and BIPM exposure and air-kerma standards. for cobalt-60 gamma rays. by C.E. de Almeida* and M.-T. Niatel

Comparisons between IRD and BIPM exposure and air-kerma standards. for cobalt-60 gamma rays. by C.E. de Almeida* and M.-T. Niatel Rapport BIPM-86/12 Comparisons between IRD and BIPM exposure and air-kerma standards for cobalt-60 gamma rays by C.E. de Almeida* and M.-T. Niatel Bureau International des Poids et Mesures, F-92310 Sevres

More information

Ionizing Radiation Dosimetry and Medical Physics

Ionizing Radiation Dosimetry and Medical Physics Ionizing Radiation Dosimetry and Medical Physics D.W.O. Rogers Ionizing Radiation Standards, Institute for National Measurement Standards NRC, Ottawa, K1A OR6 Present e-mail: drogers at physics.carleton.ca

More information

Development of a water calorimeter for medium energy x-rays

Development of a water calorimeter for medium energy x-rays Development of a water calorimeter for medium energy x-rays Leon de Prez Eduard van Dijk Patricia Damen Nederlands Meetinstituut Introduction Introduction Design and construction Measurement assembly Correction

More information

Work Programme. of the International Bureau of Weights and Measures. for the four years Comité international des poids et mesures

Work Programme. of the International Bureau of Weights and Measures. for the four years Comité international des poids et mesures Work Programme of the International Bureau of Weights and Measures for the four years 2016-2019 Comité international des poids et mesures Section II : BIPM Work Programme for 2016-2019 35 Priority activities

More information

National Physical Laboratory (NPL), UK

National Physical Laboratory (NPL), UK BIPM comparison BIPM.RI(II)-K1.Lu-177 of activity measurements of the radionuclide 177 Lu for the NPL (UK) and the IRMM (EU), with linked results for the CCRI(II)-K2.Lu-177 comparison C. Michotte, G. Ratel,

More information

Comparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kv to 300 kv

Comparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kv to 300 kv EUROMET 545: Final report 25/9/28 page 1 of 48 Comparison of national air kerma standards for ISO 437 narrow spectrum series in the range 3 kv to 3 kv L. Büermann (a), M. O Brien (b), D. Butler (c), I.

More information

Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI

Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI Pierino De Felice and Marco Capogni - ENEA C.R. Casaccia Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti (ENEA-INMRI) Roma ITALY Yoshio

More information

Calorimetry for Absorbed-dose Measurements at BNM-LNHB

Calorimetry for Absorbed-dose Measurements at BNM-LNHB Calorimetry for Absorbed-dose Measurements at BNM-LNHB J. Daures, A. Ostrowsky, P. Gross J.P. Jeannot and J.Gouriou BNM-LNHB, CEA/Saclay, Bat 534, F-91191 Gif-sur-Yvette Cedex Abstract At BNM-LNHB, the

More information

Chapter 9: Calibration of Photon and Electron Beams

Chapter 9: Calibration of Photon and Electron Beams Chapter 9: Calibration of Photon and Electron Beams Set of 189 slides based on the chapter authored by P. Andreo, J.P. Seuntjens, and E.B. Podgorsak of the IAEA publication (ISBN 92-0-107304-6): Radiation

More information

Towards the general application of water calorimetry as absorbed dose standard for radiotherapy dosimetry

Towards the general application of water calorimetry as absorbed dose standard for radiotherapy dosimetry Towards the general application of water calorimetry as absorbed dose standard for radiotherapy dosimetry A. Krauss, M. Bambynek, H.-J. Selbach Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116

More information

Final Report August 2010

Final Report August 2010 Bilateral Comparison of 100 pf Capacitance Standards (ongoing BIPM key comparison BIPM.EM-K14.b) between the CMI, Czech Republic and the BIPM, January-July 2009 J. Streit** and N. Fletcher* *Bureau International

More information

G. Ratel 1, C. Michotte 1, M. Sahagia 2, A. Yunoki 3. Abstract

G. Ratel 1, C. Michotte 1, M. Sahagia 2, A. Yunoki 3. Abstract Update of the BIPM comparison BIPM.RI(II)-K1.Ba-133 of activity measurements of the radionuclide 133 Ba to include the results of the IFIN-HH (Romania) and the NMIJ (Japan) G. Ratel 1, C. Michotte 1, M.

More information

Comparison of V and 10 V DC Voltage References

Comparison of V and 10 V DC Voltage References Comparison of 1.018 V and 10 V DC Voltage References Technical protocol for BIPM.EM-K11.a and.b comparisons 1. Introduction The Mutual Recognition Arrangement among National Metrology Institutes (NMIs)

More information

BIPM.RI(II)-K1.F-18, BIPM.RI(II)-K4.F18, CCRI(II)-K3.F-18 and APMP.RI(II)-K3.F-18

BIPM.RI(II)-K1.F-18, BIPM.RI(II)-K4.F18, CCRI(II)-K3.F-18 and APMP.RI(II)-K3.F-18 BIPM.RI(II)-K1.F-18, BIPM.RI(II)-K4.F18, CCRI(II)-K3.F-18 and APMP.RI(II)-K3.F-18 MEASURAND : Equivalent activity of 18 F Key comparison BIPM.RI(II)-K1.F-18 result of measurement carried out in the SIR

More information

Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the PTB, Germany and the BIPM in low-energy x-rays

Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the PTB, Germany and the BIPM in low-energy x-rays Key comprison BIPM.RI(I)-K2 of the ir-kerm stndrds of the PTB, Germny nd the BIPM in low-energy x-rys D.T. Burns 1, C Kessler 1 nd L Büermnn 2 1 Bureu Interntionl des Poids et Mesures, Pvillon de Breteuil,

More information

Activity measurements of the radionuclide 99m Tc for the CNEA, Argentina and the LNMRI/IRD, Brazil in the ongoing comparison BIPM.RI(II)-K4.

Activity measurements of the radionuclide 99m Tc for the CNEA, Argentina and the LNMRI/IRD, Brazil in the ongoing comparison BIPM.RI(II)-K4. Activity measurements of the radionuclide 99m Tc for the CNEA, Argentina and the LNMRI/IRD, Brazil in the ongoing comparison BIPM.RI(II)-K4.Tc-99m C. Michotte 1, M. Nonis 1, P. Arenillas 2, G. Cerutti

More information

4.1b - Cavity Theory Lecture 2 Peter R Al mond 2011 Overview of Lecture Exposure (W/e)air Exposure Exposure and and and Air Air Kerma

4.1b - Cavity Theory Lecture 2 Peter R Al mond 2011 Overview of Lecture Exposure (W/e)air Exposure Exposure and and and Air Air Kerma 4.1b - Cavity Theory Lecture 2 Peter R Almond 2011 Overview of Lecture Exposure (W/e) air Exposure and Air Kerma Exposure Exposure is symbolized as X and defined by the ICRU as the quotient of dq by dm,

More information

Chapiter VII: Ionization chamber

Chapiter VII: Ionization chamber Chapiter VII: Ionization chamber 1 Types of ionization chambers Sensitive volume: gas (most often air direct measurement of exposure) ionization chamber Sensitive volume: semiconductor (silicon, germanium,

More information

Determination of the Specific Heat Capacity of Graphite Using Absolute and Differential Methods Susanne Picard David Burns Philippe Roger BIPM

Determination of the Specific Heat Capacity of Graphite Using Absolute and Differential Methods Susanne Picard David Burns Philippe Roger BIPM Determination of the Specific Heat Capacity of Graphite Using Absolute and Differential Methods Susanne Picard David Burns Philippe Roger BIPM Absorbed Dose and Air Kerma Dosimetry Workshop - Paris 9-11

More information

G. Ratel*, C. Michotte*, Y. Hino** *BIPM and **NMIJ, Japan

G. Ratel*, C. Michotte*, Y. Hino** *BIPM and **NMIJ, Japan Update of the BIPM comparison BIPM.RI(II)-K1.Ce-139 of activity measurements of the radionuclide 139 Ce to include the 2004 NMIJ result and links for the 2004 regional comparison APMP.RI(II)-K2.Ce-139

More information

RADIOTHERAPY: GRAPHITE CALORIMETRY, A BASIS OF THE TRACEABILITY CHAIN

RADIOTHERAPY: GRAPHITE CALORIMETRY, A BASIS OF THE TRACEABILITY CHAIN RADIOTHERAPY: GRAPHITE CALORIMETRY, A BASIS OF THE TRACEABILITY CHAIN Since medical exposure is by far the most common source of artificial irradiation, there is a clear need to provide specific metrological

More information

Reference Dosimetry for Megavoltage Therapy Beams: Electrons

Reference Dosimetry for Megavoltage Therapy Beams: Electrons Reference Dosimetry for Megavoltage Therapy Beams: Electrons David Followill Ph.D Radiological Physics Center UT M.D.Anderson Cancer Center Houston TX Protocol for Clinical Reference Dosimetry of High-Energy

More information

Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration

Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 1, WINTER 2008 Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration Pablo

More information

APMP supplementary comparison of absorbed dose rate in tissue for beta radiation

APMP supplementary comparison of absorbed dose rate in tissue for beta radiation APMP supplementary comparison of absorbed dose rate in tissue for beta radiation BIPM KCDB: APMP.R(I)-S2 Technical Protocol N. Saito and M. Kato National Metrology Institute of Japan (NMIJ), Tsukuba, Ibaraki

More information

Metrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT)

Metrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT) 1- Ionizing radiation metrology for radiation protection 2- Metrological requirements for ionizing radiation measurement in radiotherapy and radiodiagnostics R. F. Laitano Part 2 Metrological traceability

More information

CCM short note on the dissemination process after the proposed redefinition of the kilogram

CCM short note on the dissemination process after the proposed redefinition of the kilogram CCM short note on the dissemination process after the proposed redefinition of the kilogram Consultative Committee for Mass and Related Quantities 1. Introduction This note proposes how the mise en pratique

More information

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe Radiation Dosimetry Alun Beddoe Medical Physics University Hospital Birmingham NHS Trust Contents ABSOLUTE DOSIMETRY (CALIBRATION) Photon interactions (recap) Energy transfer and absorption Electron range

More information

Progress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima

Progress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima Progress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima Norio Saito, Tadahiro Kurosawa, Masahiro Kato, Yuichiro Morishita, Takahiro Tanaka, and Morihito

More information

Improvements to the UK Primary Standard Therapy Level Electron Beam Calorimeter

Improvements to the UK Primary Standard Therapy Level Electron Beam Calorimeter Radiotherapy Standards Users Group meeting NPL, 5 th June 2007 Improvements to the UK Primary Standard Therapy Level Electron Beam Calorimeter Mark Bailey, G A Bass, N D Lee, G A Mapp Contents 1. History

More information

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators N. Golnik 1, P. Kamiński 1, M. Zielczyński 2 1 Institute of Precision and Biomedical Engineering,

More information

Dosimetry: Electron Beams

Dosimetry: Electron Beams ICTP SChool On MEdical PHysics For RAdiation THerapy: DOsimetry And TReatment PLanning For BAsic And ADvanced APplications 13-24 April 2015 Miramare, Trieste, Italy Dosimetry: Electron Beams G. Hartmann

More information

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR Jin Wan Chung 1, Shih Mean Lee, Sungjun Lee 1 1 Korea Research Institute of Standards

More information

Principles of applied dosimetry - illustrated by ionometry. Lesson FYSKJM4710 Eirik Malinen

Principles of applied dosimetry - illustrated by ionometry. Lesson FYSKJM4710 Eirik Malinen Principles of applied dosimetry - illustrated by ionometry Lesson FYSKJM4710 Eirik Malinen Ionometry Ionometry: the art of measuring ionizations Number of ionizations proportional to dose Air filled ionization

More information

Comparison of air kerma measurements for tungsten anode based mammography x-ray beam qualities (EURAMET.RI(I)-S4.1)

Comparison of air kerma measurements for tungsten anode based mammography x-ray beam qualities (EURAMET.RI(I)-S4.1) Comparison of air kerma measurements for tungsten anode based mammography x-ray beam qualities (EURAMET.RI(I)-S4.1) István Csete 1, Ludwig Büermann 2, Igor Gomola 1 1 International Atomic Energy Agency

More information

Commissioning of the Beta Secondary Standard (BSS2)

Commissioning of the Beta Secondary Standard (BSS2) Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,

More information

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * Romanian Reports in Physics, Vol. 67, No. 3, P. 1152 1158, 2015 ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * ELENA STANCU 1,2, CATALIN VANCEA

More information

General characteristics of radiation dosimeters

General characteristics of radiation dosimeters General characteristics of radiation dosimeters and a terminology to describe them D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers

More information

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry TJ JORDAN Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE

More information

Bureau International des Poids et Mesures

Bureau International des Poids et Mesures Bureau International des Poids et Mesures Comparison of the standards for absorbed dose to water of the ARPANSA and the BIPM for 60 Co gamma radiation P J Allisy-Roberts and D T Burns BIPM J F Boas, R

More information

N. E. Ipe*, K. E. Rosser, C. J. Moretti, J. W. Manning and M. J. Palmer

N. E. Ipe*, K. E. Rosser, C. J. Moretti, J. W. Manning and M. J. Palmer SLAC-PUB-8099 July 1999 Air Kerma Calibration Factors and k ch Values for PTW Soft X-ray, NACP and Roos Ionization Chambers at Very Low X-ray Energies (0.035 mm - 1.0 mm Al HVL) N. E. Ipe*, K. E. Rosser,

More information

International comparison of extrapolation chamber measurements of the absorbed dose rate in tissue for beta radiation

International comparison of extrapolation chamber measurements of the absorbed dose rate in tissue for beta radiation International comparison of extrapolation chamber measurements of the absorbed dose rate in tissue for beta radiation EUROMET project No. 739 BIPM KCDB: EUROMET.RI(I)-S2 Final Report R. Behrens, K. Helmstädter,

More information

BUREAU INTERNATIONAL DES POIDS ET MESURES

BUREAU INTERNATIONAL DES POIDS ET MESURES 1 BUREAU INTERNATIONAL DES POIDS ET MESURES International comparison of activity measurements of a solution of 3 H (January 2009) Participating laboratory: T ½ = (4 496.862 d; u = 9.131 d)* Ampoule number

More information

Bureau International des Poids et Mesures (BIPM), Sèvres. National Institute of Standards and Technology (NIST), USA. Becquerel (LNE-LNHB), France

Bureau International des Poids et Mesures (BIPM), Sèvres. National Institute of Standards and Technology (NIST), USA. Becquerel (LNE-LNHB), France BIPM comparison BIPM.RI(II)-K1.I-131 of activity measurements of the radionuclide 131 I for the NMIJ (Japan), the NIST (USA) and the LNE-LNHB (France), with linked results for the APMP.RI(II)-K2.I-131

More information

Investigation of the standard temperature- pressure correction factor at low x-ray energies

Investigation of the standard temperature- pressure correction factor at low x-ray energies Investigation of the standard temperaturepressure correction factor at low x-ray energies D. J. La Russa, M. R. McEwen and D. W. O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton

More information

NPL s progress towards absorbed dose standards for proton beams

NPL s progress towards absorbed dose standards for proton beams NPL s rogress towards absorbed dose standards for roton beams H. Palmans 1 R. Thomas 1 D. Shiley 1 A. Kacerek 2 1 National Physical Laboratory Teddington United Kingdom 2 Clatterbridge Centre of Oncology

More information

Final Report of the SIM 60 Co Absorbed- Dose-to-Water Comparison

Final Report of the SIM 60 Co Absorbed- Dose-to-Water Comparison Fnal Report of the SIM 60 Co Absorbed- Dose-to-Water Comparson (KCDB Entry SIM.RI(I)-K4) May 2008 C. K. Ross and K. R. Shortt (NRC, Canada; plot laboratory) M. Sarav (CNEA, Argentna) A. Meghzfene (IAEA)

More information

Evaluation of Correction Factors Applied in Photon Calibration of NIS TE Neutron Ionization Chambers

Evaluation of Correction Factors Applied in Photon Calibration of NIS TE Neutron Ionization Chambers Evaluation of orrection Factors Applied in Photon alibration of NIS TE Neutron Ionization hambers A. I. Abd El-Hafez * and M. Ezzat Radiation Metrology Dept., National Institute for Standards (NIS), Giza,

More information

*BIPM, # CNEA, BARC, $ NRC, **NIM. Abstract

*BIPM, # CNEA, BARC, $ NRC, **NIM. Abstract Update of the BIPM comparison BIPM.RI(II)-K1.Co-60 of activity measurements of the radionuclide 60 Co to include the 2011 result of the CNEA (Argentina), the 2012 results of the BARC (India) and the NRC

More information

Instrumentation for Verification of Dose

Instrumentation for Verification of Dose Instrumentation for Verification of Dose MedAustron GmbH, Wiener Neustadt, Austria Presented to: Educational Workshop PTCOG 52 Essen, Germany, May 3 5, 2013 Consistent and harmonized dosimetry guidelines

More information

The metrology of radioactivity. C. Michotte, BIPM

The metrology of radioactivity. C. Michotte, BIPM The metrology of radioactivity C. Michotte, BIPM SUMMARY Introduction Nuclear data of interest in activity measurements Activity measurement methods (primary and secondary) Monte Carlo simulations The

More information

Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network /2017

Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network /2017 Journal of Physics: Conference Series PAPER OPEN ACCESS Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network - 2016/2017 To cite this article: TS Cabral et al 2018

More information

RECOMBINATION PARAMETERS OF SOME FABRICATED IONIZATION CHAMBERS

RECOMBINATION PARAMETERS OF SOME FABRICATED IONIZATION CHAMBERS Bangladesh Journal of Medical Physics ol. 4, No.1, 011 RECOMBNATON PARAMETERS OF SOME FABRCATED ONZATON CHAMBERS Afia Begum 1 and Nobuhisa Takata 1 Department of Physics, Bangladeah University of Engineering

More information

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements Specific Accreditation Criteria Calibration ISO/IEC 17025 Annex Ionising radiation measurements January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected

More information

Draft protocol of APMP.RI(I)-K8 Reference air kerma rate for HDR Ir-192 brachytherapy sources

Draft protocol of APMP.RI(I)-K8 Reference air kerma rate for HDR Ir-192 brachytherapy sources Draft protocol of APMP.RI(I)-K8 Reference air kerma rate for HDR Ir-19 brachytherapy sources T. Kurosawa and N. Saito National Metrology Institute of Japan (NMIJ), Tsukuba, Ibaraki 305-8568 JAPAN Revised

More information

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3.

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3. AFRIMETS Secretariat Private Bag X34 Lynnwood Ridge 0040 AFRIMETS Supplementary Comparison Programme Calibration of Gauge Blocks by Mechanical Comparison Method AFRIMETS.L S3 Final Report Giza, Egypt,

More information