1 Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical Physics University of Wisconsin NCCAAPM meeting April 17, 2015

2 Larry DeWerd has partial interest in Standard Imaging Inc.

3 Determination of your uncertainty with TG 51 addendum Presented the TG 51 addendum last October. Short review of TG 51 addendum Uncertainty determination

4 What are reference class ionization chambers %dd(10) is used for k Q No lead foil to be used anymore caused errors. Use eq. 15 in TG 51 Use of small volume chambers in relative dosimetry

5 Majority are 0.6 cm 3 Farmer-type chambers A-150 chambers explicitly excluded 5 scanning chambers, NO microchambers (Exception A26 from some preliminary measurements. Long term to come) No parallel plate chambers are included

6 A. k Q factors for new chambers B. Comparison of measured and calculated k Q factors: Measured k Q data is available for some types of chambers C. Non-water phantoms only water for dosimetry D. Uncertainty analysis for implementation of TG-51

7 1. TG-51 made the deliberate decision not to include uncertainties. 2. Other protocols have included uncertainty budgets and/or detailed reviews of uncertainty components. 3. TG 51 addendum gives some guidance on: i. How to develop an uncertainty budget ii. Typical values for individual components. 4. The ISO GUM is the starting point or NIST publication

8 Uncertainty tables and determinations are important for the goodness of your measurements. Uncertainty gives an indication of where you would expect your measurements to fall each time you measure it. Uncertainty is larger than precision of measurement

9 Systematic errors: Results from a bias on part of observer or a faulty calibration of equipment. Random errors: Fluctuations in observations. Generally a standard deviation can account for these. Reduce all sources of error. Now use Type A (random) and Type B Errors are truly errors something done wrong

10

11 Accuracy and Precision Accurate but not precise Precise but not accurate Precise and accurate

12 1 0.9 primary standard measurement frequency of result clinical measurement conventional true value a.u.

13 Today possible variation in readings are given by uncertainties. The difference between the measured value - measurand - and the conventional true value is generally never zero. There are uncertainties involved in the measurement that can be expressed. Uncertainty is a methodology to express the goodness of the value as to its accuracy.

14 The uncertainty associated with a measurement characterizes the dispersion of values that can be reasonably associated with the measurand. What is the difference between uncertainty, precision and accuracy or the error in measurement?

15 Type A and Type B uncertainties are used Type A uncertainty is estimated by the standard deviation of the mean value. These are measured results. Any valid statistical method for treating data can be used for Type A uncertainties

16 Type B is based on scientific judgment using all relevant information available such as: previous measurement data experience or general knowledge of behavoir of instruments used manufacturers specifications data of calibration or other reports uncertainties assigned to reference data taken from handbooks.

17 A limit(confidence interval) is generally used for Type B, designated by k. For Guassian distributions, If we are certain that the value lies between + L then 99% lie here and the confidence limit is designated by k. For 99% k=3. Express L in %, call it L % The uncertainty is then u=l % /k and expressed in a %. For 95%, k=2 and for 66% k=1

18

19 Also can assume some type of probability distribution, e.g. rectangular or triangular Rectangular - all values fall within these maximum limits, +M (express in %) then u=m/ 3. For example, this is used when a manufacturer gives maximum limits for a parameter, such as the range of the calibration extends from 4.7 to 5.3. (3.5%)

20 Triangular is when all values fall with limits of + M, but the values are more weighted toward the central value. All values do not have equal probability. Then the uncertainty is estimated by u=m/ 6 This isn t used as much. Manufacturer may say: the general value is centered at 5, with a range from 4.7 to 5.3 Triangular would be value is 5.0, with some outliers at 4.7 to 5.3 Then 2.4%

21 Procedure is outlined in NIST Technical Note 1297 (1994) Each uncertainty component is propagated sequentially throughout the measurement pathway by quadrature summation: Square root of sum of squares for k=1

22 Generally the symbol u is used Say you measure a quantity a number of times and the standard deviation of the measurement is + 0.5% Therefore your type A uncertainty for this measurement is u A =+ 0.5% (precision of measurement)

23 Uncertainties in % are combined as u c =(u A2 + u B2 ) 1/2 Therefore in a simple example, we have u c =( ) 1/2 = (8.66) 1/2 ~ 2.92% Note that the greatest uncertainty governs the overall uncertainty So at k=2, total uncertainty is 5.84%

24 Standard Uncertainty: Expected range of values that are attributed to a measured quantity (source strength) Expanded Uncertainty: Defined here as the Standard Uncertainty multiplied by a coverage factor of 2. This corresponds to a coverage probability or confidence level of approximately 95%

25 The uncertainty in the laboratory from all aspects of your measurement are determined. - all possible contributions Make your best estimate. You can list your best (lowest) uncertainty and your worst (highest) uncertainty.

26 The calibration of the chamber progresses from NIST through the ADCL to the clinic. The traceability to NIST is given by the ADCL as 0.7% at k=1 u total u 2 clinic u 2 ADCL ADCL uncertainty tables available if necessary

27 Further treatment of Uncertainty AAPM Summer School 2009

28 NIST has produced an explanatory document (a guide to the Guide) - NIST Technical Note 1297 Uncertainty budget broken down into: Measurement Calibration data Influence quantities Typical values discussed but emphasis on individual users constructing site-specific uncertainty budgets for their calibration situations

29 Want to take into account your measurement process Values and quantities are representative. You need to determine your clinic values Start with dose measurement.

30 TG 51 gives a measure of the absorbed dose to water using an ADCL calibrated ion chamber. D N 60 Co k M w,q D,w Q ion k Q is the factor that converts from the calibration beam ( 60 Co) to the user linac beam, defined by beam quality Q

31 Quantity Type A Type B Comment Calibration N Dw ADCL 0.7% ADCL Traceability to NIST uncertainty k Q 0.4% Picking Maximum P elec 0.1% Correction for charge reading for M corr

32 M corr,w M raw P TP P ion P pol P elec Each of these parameters have an uncertainty associated with them. There is also an uncertainty involved in the setup, which we will deal with as yet.

33 Many times the manufacturer of the barometer (thermometer) will give a response range, e.g. +0.5%. Assume a rectangular distribution % Keep in mind the equation for TP correction. Generally most clinics have a 0.1% k=1 for P TP

34 Uncertainty components examples Temperature Pressure correction P TP T w P air For T in C, P in kpa Need water temperature at ion chamber: Water should be in equilibrium with room Ion chamber should be in equilibrium with water P TP does not take account of thermal expansion of thimble Need air pressure in air volume: Only realistic to measure room air pressure Need to confirm air communication of ion chamber (ADCL)

35 Thermal equilibration Chamber at 22 C placed in water at 10 C Equilibration is pretty quick, but not instantaneous Reference: Das and Zhu (Med. Phys., 2004)

36 Quantity Type A Type B Comment Influence Quantities P TP 0.10 Clinic Measurement of P TP P Pol 0.05 Measurement P ion 0.10 Measurement P leak 0.05 If present Pre irrad leak 0.10 If present

37 M raw M(x,y,z,SSD,FS) Wherever the chamber actually is positioned and whatever the actual geometry, M will be assigned to: x = y = 0 cm (on axis) z = 10 cm (d ref ) SSD = 100 cm (assuming SSD setup) Field size = 10 cm x 10 cm By writing the equation out this way we identify the influence quantities, and therefore uncertainty components Uncertainty analysis is also a procedural review

38 Quantity Type A Setup Parameters SSD Setting 0.10 Depth Setting Linac Stability Type B Comment If not measured, then type B estimated 0.25 If not measured, then type B estimated 0.10 If not measured, then type B estimated

39 Quantity Type A Type B Calibration 0.81 Influence 0.19 Quantities Setup 0.29 Parameters Combined uncertainty Comment 0.88% combined uor Relative Expanded u= 1.76%

40 The majority of the uncertainty comes from the calibration. What does this say to you You can measure with better precision than this. Your overall accuracy to the conventional true value is confident to 1.76% - or 1.8%

41 Thanks are due to Students and staff of the UW ADCL All those who send us calibration instruments that support the research program of the UW Medical Radiation Research Center.

Updating reference dosimetry a decade after TG-51

Updating reference dosimetry a decade after TG-51 Updating reference dosimetry a decade after TG-51 Malcolm McEwen Ionizing Radiation Standards Institute for National Measurement Standards National Research Council, Canada CE Presentation at AAPM Annual

More information

Reference Dosimetry for Megavoltage Therapy Beams: Electrons

Reference Dosimetry for Megavoltage Therapy Beams: Electrons Reference Dosimetry for Megavoltage Therapy Beams: Electrons David Followill Ph.D Radiological Physics Center UT M.D.Anderson Cancer Center Houston TX Protocol for Clinical Reference Dosimetry of High-Energy

More information

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required UPDATING TG-51 When will it end? Malcolm McEwen Ionizing Radiation Standards National Research Council, Canada AAPM Annual Meeting, Charlotte, 2012 What are these updates? Working Group review recommends:

More information

Small Field Dosimetric Measurements with TLD-100, Alanine, and Ionization Chambers

Small Field Dosimetric Measurements with TLD-100, Alanine, and Ionization Chambers Small Field Dosimetric Measurements with TLD-1, Alanine, and Ionization Chambers S. Junell a, L. DeWerd a, M. Saiful Huq b, J. Novotny Jr. b, M. uader b, M.F. Desrosiers c, G. Bednarz b a Department of

More information

Optimization of microionization chambers for small-field reference dosimetry

Optimization of microionization chambers for small-field reference dosimetry Optimization of microionization chambers for small-field reference dosimetry Jessica R. Snow University of Wisconsin-Madison Department of Medical Physics Medical Radiation Research Center NCCAAPM Fall

More information

8/2/2012. Larry DeWerd has a partial interest in Standard Imaging

8/2/2012. Larry DeWerd has a partial interest in Standard Imaging Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin AAPM July 2012 Larry DeWerd has a partial interest in Standard Imaging Traceability to NIST via calibration points

More information

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

More information

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden Referensdosimetri Crister Ceberg Medical Radiation Physics Lund University Sweden Reference dosimetry Determination of absorbed dose to water under reference conditions Not accounting for uncertainties

More information

Dosimetry: Electron Beams

Dosimetry: Electron Beams ICTP SChool On MEdical PHysics For RAdiation THerapy: DOsimetry And TReatment PLanning For BAsic And ADvanced APplications 13-24 April 2015 Miramare, Trieste, Italy Dosimetry: Electron Beams G. Hartmann

More information

Measurement And Uncertainty

Measurement And Uncertainty Measurement And Uncertainty Based on Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results, NIST Technical Note 1297, 1994 Edition PHYS 407 1 Measurement approximates or

More information

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * Romanian Reports in Physics, Vol. 67, No. 3, P. 1152 1158, 2015 ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * ELENA STANCU 1,2, CATALIN VANCEA

More information

Introduction to the evaluation of uncertainty

Introduction to the evaluation of uncertainty Manual of Codes of Practice for the Determination of Uncertainties in Mechanical Tests on Metallic Materials SECTION 1 Introduction to the evaluation of uncertainty F A Kandil National Physical Laboratory

More information

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC 17025 Table of contents 1 GENERAL... 2 2 THE STANDARD SIST EN ISO/IEC 17025... 2 3 SA'S POLICY IN IMPLEMENTING

More information

ABSTRACT. Keywords: Megavoltage, dosimetry, TG51 protocol, TG21 protocol, parallel-plate chambers, crosscomparison. INTRODUCTION

ABSTRACT. Keywords: Megavoltage, dosimetry, TG51 protocol, TG21 protocol, parallel-plate chambers, crosscomparison. INTRODUCTION CALCULATED TG51 TO TG21 ABSORBED-DOSE RATIOS FOR MOST WIDELY USED CYLINDRICAL AND PARALLEL-PLATE ION CHAMBERS OVER A RANGE OF PHOTON AND ELECTRON ENERGIES R.C. Tailor, and W.F. Hanson ABSTRACT The change

More information

Essentials of expressing measurement uncertainty

Essentials of expressing measurement uncertainty Essentials of expressing measurement uncertainty This is a brief summary of the method of evaluating and expressing uncertainty in measurement adopted widely by U.S. industry, companies in other countries,

More information

MEASUREMENT UNCERTAINTY PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER Prepared by MJ Mc Nerney for ENAO Assessor Calibration

MEASUREMENT UNCERTAINTY PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER Prepared by MJ Mc Nerney for ENAO Assessor Calibration MEASUREMENT PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE UNCERTAINTY OCTOBER/NOVEMBER 2012 Prepared by for ENAO Assessor Calibration B SCOPE Introduction House Rules Overview 17025 & 15189 MU Approaches

More information

Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification.

Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification. Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification. Antonio Lopez Medina, Antonio Teijeiro, Daniela Medal, Francisco Salvador, Julio Vazquez, Manuel

More information

Progress in calculations of k Q for TG-51

Progress in calculations of k Q for TG-51 1 Progress in calculations of k Q for TG-51 D. W. O. Rogers Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University Ottawa http://www.physics.carleton.ca/~drogers AAPM Charlotte,NC,

More information

Air kerma rate measurements from a miniature x-ray source using free-air ionization chambers

Air kerma rate measurements from a miniature x-ray source using free-air ionization chambers Air kerma rate measurements from a miniature x-ray source using free- ionization chambers Stephen D. Davis 1, John A. Micka 1, Larry A. DeWerd 1, and Thomas W. Rusch 2 1 University of Wisconsin, Madison,

More information

Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy

Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy M. Saiful Huq, PhD, FAAPM, FInstP Dept. of Radiation Oncology, University of

More information

Electron beam water calorimetry measurements to obtain beam quality conversion factors

Electron beam water calorimetry measurements to obtain beam quality conversion factors Electron beam water calorimetry measurements to obtain beam quality conversion factors Bryan R. Muir, a) Claudiu D. Cojocaru, Malcolm R. McEwen, and Carl K. Ross Measurement Science and Standards, National

More information

Doubt-Free Uncertainty In Measurement

Doubt-Free Uncertainty In Measurement Doubt-Free Uncertainty In Measurement Colin Ratcliffe Bridget Ratcliffe Doubt-Free Uncertainty In Measurement An Introduction for Engineers and Students Colin Ratcliffe United States Naval Academy Annapolis

More information

NACP-02 perturbation correction factors for the NPL primary standard of absorbed dose to water in high energy electron beams

NACP-02 perturbation correction factors for the NPL primary standard of absorbed dose to water in high energy electron beams NACP-02 perturbation correction factors for the NPL primary standard of absorbed dose to water in high energy electron beams E. Chin 1, J. Seuntjens 1, H. Palmans 2, A. DuSautoy 2, D. Shipley 2, M. Bailey

More information

CALCULATION OF UNCERTAINTY IN CHEMICAL ANALYSIS. A.Gnanavelu

CALCULATION OF UNCERTAINTY IN CHEMICAL ANALYSIS. A.Gnanavelu CALCULATION OF UNCERTAINTY IN CHEMICAL ANALYSIS A.Gnanavelu UNCERTAINTY (U) Every measurement has error which is unknown and unknowable. This unknown error is called measurement uncertainty. Types of Error

More information

Composite field dosimetry

Composite field dosimetry Composite field dosimetry Hugo Bouchard, PhD, MCCPM Senior Research Scientist Radiation dosimetry group National Physical Laboratory May 2014 Overview 1. Introduction Dosimetry protocols IAEA formalism

More information

Monitor Unit Calculations for Photon and Electrons. AAMD Meeting Raleigh, NC October 3, John P. Gibbons Chief of Clinical Physics

Monitor Unit Calculations for Photon and Electrons. AAMD Meeting Raleigh, NC October 3, John P. Gibbons Chief of Clinical Physics Monitor Unit Calculations for Photon and Electrons AAMD Meeting Raleigh, NC October 3, 2014 John P. Gibbons Chief of Clinical Physics Outline I. TG71 Formation and Charge II. Photon Calculations III. Electron

More information

ISO 376 Calibration Uncertainty C. Ferrero

ISO 376 Calibration Uncertainty C. Ferrero ISO 376 Calibration Uncertainty C. Ferrero For instruments classified for interpolation, the calibration uncertainty is the uncertainty associated with using the interpolation equation to calculate a single

More information

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society of Radiobiology

More information

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS M.A. HASSAN*, M.H. GABER**, E. ESMAT*, H.I. FARAG***, H.M. EISSA* *National Institute for Standards (NIS), Giza, Egypt **Biophysics Department,

More information

M [scale units/s] of the system

M [scale units/s] of the system APPENDIX TO IAEA CALIBRATION CERTIFICATE RADIATION PROTECTION IONIZATION CHAMBER CALIBRATION PROCEDURES AT THE IAEA DOSIMETRY LABORATORY 1. INTRODUCTION 1.1 General Ionization chambers and electrometers

More information

Uncertainty of Measurement

Uncertainty of Measurement Uncertainty of Measurement Contents Uncertainty of Measurement.... 3 Uncertainty of Measurement and Measurement Error.... 4 Measuring Uncertainty.... 5 Coverage factor k...6 Factors Affecting Uncertainty....

More information

Unit 4. Statistics, Detection Limits and Uncertainty. Experts Teaching from Practical Experience

Unit 4. Statistics, Detection Limits and Uncertainty. Experts Teaching from Practical Experience Unit 4 Statistics, Detection Limits and Uncertainty Experts Teaching from Practical Experience Unit 4 Topics Statistical Analysis Detection Limits Decision thresholds & detection levels Instrument Detection

More information

R. C. Tailor and W. F. Hanson. Department of Radiation Physics. The University of Texas, M.D. Anderson Cancer Center. Houston, Texas 77030

R. C. Tailor and W. F. Hanson. Department of Radiation Physics. The University of Texas, M.D. Anderson Cancer Center. Houston, Texas 77030 Page 1 of 231 Calculated absorbed-dose ratios, TG51/TG21, for most idely used cylindrical and parallel-plate ion chambers over a range of photon and electron energies R. C. Tailor and W. F. Hanson Department

More information

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah

More information

Uncertainty of Measurement A Concept

Uncertainty of Measurement A Concept Uncertainty of Measurement A Concept Shweta V. Matey 1, Dr. Nitin K. Mandavgade 2, and Dr. R.R. Lakhe 3 1 (Asst. Professor, Dept. of Mechanical Engg, Lokmanya Tilak College of Engineering, Mumbai University,

More information

) for Varian TrueBeam high-dose-rate therapy beams

) for Varian TrueBeam high-dose-rate therapy beams JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 Ion recombination correction factors (P ion ) for Varian TrueBeam high-dose-rate therapy beams Stephen F. Kry, 1a Richard Popple,

More information

Uncertainty sources of reference measurement procedures for enzymes

Uncertainty sources of reference measurement procedures for enzymes Medical School Hannover Institute for Clinical Chemistry Director: Prof. Dr. K. Brand Uncertainty sources of reference measurement procedures for enzymes R. Klauke and G. Schumann Renate Strache Rainer

More information

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014) SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February

More information

What is measurement uncertainty?

What is measurement uncertainty? What is measurement uncertainty? What is measurement uncertainty? Introduction Whenever a measurement is made, the result obtained is only an estimate of the true value of the property being measured.

More information

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry TJ JORDAN Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE

More information

POWER UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES

POWER UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES Proceedings of the ASME 2014 Power Conference POWER2014 July 28-31, 2014, Baltimore, Maryland, USA POWER2014-32205 UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES Michael S. Hering Rosemount DP

More information

Monte Carlo modeling of an electronic brachytherapy source using MCNP5 and EGSnrc

Monte Carlo modeling of an electronic brachytherapy source using MCNP5 and EGSnrc Monte Carlo modeling of an electronic brachytherapy source using MCNP5 and EGSnrc Stephen D. Davis and Larry A. DeWerd University of Wisconsin, Madison, WI 13 th UK Monte Carlo User Group meeting March

More information

A MEASUREMENT ASSURANCE PROGRAM FOR LOW FLOW CALIBRATION BY THE TRANSFER METHOD CS20

A MEASUREMENT ASSURANCE PROGRAM FOR LOW FLOW CALIBRATION BY THE TRANSFER METHOD CS20 A MEASUREMENT ASSURANCE PROGRAM FOR LOW FLOW CALIBRATION BY THE TRANSFER METHOD CS0 JEFF PEERY MAY. 18, 010 This document outlines the Measurement Assurance Program (MAP for Seametrics CS0 flo measurement

More information

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2 Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST Ronaldo Minniti 1 and Ladislav Czap 2 1 National Institute of Standards and Technology (NIST), Gaithersburg,

More information

General characteristics of radiation dosimeters

General characteristics of radiation dosimeters General characteristics of radiation dosimeters and a terminology to describe them D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers

More information

HOW TO ASSESS THE MEASUREMENT UNCERTAINTY

HOW TO ASSESS THE MEASUREMENT UNCERTAINTY 15th EFLM Continuing Postgraduate Course in Clinical Chemistry and Laboratory Medicine How to assess the quality of your method? Zagreb, 24-25 October 2015 HOW TO ASSESS THE MEASUREMENT UNCERTAINTY Centre

More information

Practical Statistics for the Analytical Scientist Table of Contents

Practical Statistics for the Analytical Scientist Table of Contents Practical Statistics for the Analytical Scientist Table of Contents Chapter 1 Introduction - Choosing the Correct Statistics 1.1 Introduction 1.2 Choosing the Right Statistical Procedures 1.2.1 Planning

More information

A Unified Approach to Uncertainty for Quality Improvement

A Unified Approach to Uncertainty for Quality Improvement A Unified Approach to Uncertainty for Quality Improvement J E Muelaner 1, M Chappell 2, P S Keogh 1 1 Department of Mechanical Engineering, University of Bath, UK 2 MCS, Cam, Gloucester, UK Abstract To

More information

Measurement Uncertainty, March 2009, F. Cordeiro 1

Measurement Uncertainty, March 2009, F. Cordeiro 1 Measurement Uncertainty, March 2009, F. Cordeiro 1 Uncertainty Estimation on the Quantification of Major Proteins in Milk by Liquid Chromatography Contribution to a Chemical Reference Measurement System

More information

Chapter 9: Calibration of Photon and Electron Beams

Chapter 9: Calibration of Photon and Electron Beams Chapter 9: Calibration of Photon and Electron Beams Set of 189 slides based on the chapter authored by P. Andreo, J.P. Seuntjens, and E.B. Podgorsak of the IAEA publication (ISBN 92-0-107304-6): Radiation

More information

A comparison of methods for monitoring photon beam energy constancy

A comparison of methods for monitoring photon beam energy constancy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 A comparison of methods for monitoring photon beam energy constancy Song Gao, 1a Peter A. Balter, 1 Mark Rose, 2 and William E. Simon

More information

1.11 Measurement Uncertainty

1.11 Measurement Uncertainty Document #: 4968 Page 1 of 10 1.11 Measurement Uncertainty In an effort to comply with accreditation requirements, and because scientific measurements in general are subject to variability, a budget estimating

More information

Know Your Uncertainty

Know Your Uncertainty April 2000 METROLOGY Know Your Uncertainty Understanding and documenting measurement uncertainty is key to gage calibration. By Henrik S. Nielsen, Ph.D. The process of developing uncertainty budgets requires

More information

The Determination of Uncertainties in Bend Tests on Metallic Materials

The Determination of Uncertainties in Bend Tests on Metallic Materials Manual of Codes of Practice for the Determination of Uncertainties in Mechanical Tests on Metallic Materials Code of Practice No. 09 The Determination of Uncertainties in end Tests on Metallic Materials

More information

Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration

Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 1, WINTER 2008 Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration Pablo

More information

Use of a radioactive check device for redundancy check of ionization chambers

Use of a radioactive check device for redundancy check of ionization chambers JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Use of a radioactive check device for redundancy check of ionization chambers N. P. S. Sidhu,* Alkis Rouvas, and Patrick Cadman

More information

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL A. Specification Gamma-spectrometry method is used to identify and determine the activity concentration

More information

Vocabulary of Metrology

Vocabulary of Metrology Vocabulary of Metrology For Understating Uncertainty and Traceability Prepared by: Kim, Sang Ho Engineering Team Leader Modal Shop, Inc A PCB Group Company SI National Metrology Institute Secondary Calibration

More information

Uncertainties associated with the use of a sound level meter

Uncertainties associated with the use of a sound level meter NPL REPORT DQL-AC 002 Uncertainties associated with the use of a sound level meter Richard Payne April 2004 April 2004 ABSTRACT Uncertainties associated with the use of a sound level meter Richard Payne

More information

The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1

The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1 The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1 1 Medical Radiation Physics Group, University of Oldenburg and Pius

More information

Efficiencies of Some Spherical Ion Chambers in Continuous and Pulsed Radiation: A Numerical Evaluation

Efficiencies of Some Spherical Ion Chambers in Continuous and Pulsed Radiation: A Numerical Evaluation Signature: Pol J Radiol, 05; 80: 55-5 DOI: 0.659/PJR.89450 ORIGINAL ARTICLE Received: 05.03.7 Accepted: 05.06.9 Published: 05..5 Authors Contribution: A Study Design B Data Collection C Statistical Analysis

More information

Small Field Dosimetry and IAEA/AAPM Protocol

Small Field Dosimetry and IAEA/AAPM Protocol Small Field Dosimetry and IAEA/AAPM Protocol Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Emeritus Professor, Department of Medical Physics University of Wisconsin

More information

Commissioning of the Beta Secondary Standard (BSS2)

Commissioning of the Beta Secondary Standard (BSS2) Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,

More information

Guide to the Expression of Uncertainty in Measurement (GUM)- An Overview

Guide to the Expression of Uncertainty in Measurement (GUM)- An Overview Estimation of Uncertainties in Chemical Measurement Guide to the Expression of Uncertainty in Measurement (GUM)- An Overview Angelique Botha Method of evaluation: Analytical measurement Step 1: Specification

More information

Measurement Uncertainty Knowing the Unknown

Measurement Uncertainty Knowing the Unknown Measurement Uncertainty Knowing the Unknown Anish Shah FOUNDER and Chief Metrology Officer at Metrologized, LLC, a company founded in 2014 to further the metrology knowledge-base within the manufacturing

More information

Investigation of the standard temperature- pressure correction factor at low x-ray energies

Investigation of the standard temperature- pressure correction factor at low x-ray energies Investigation of the standard temperaturepressure correction factor at low x-ray energies D. J. La Russa, M. R. McEwen and D. W. O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton

More information

Measurement Uncertainty - How to Calculate It In The Medical Laboratory

Measurement Uncertainty - How to Calculate It In The Medical Laboratory Measurement Uncertainty - How to Calculate It In The Medical Laboratory Godfrey C. Moses, PhD, FCACB National Director, Gamma-Dynacare Medical Laboratories 11/18/2009 Moses, GC; GDML 1 Introduction Disclaimers

More information

Good Practice Guide No. 130

Good Practice Guide No. 130 Good Practice Guide No. 130 Co-ordinate measuring machine task-specific measurement uncertainties David Flack Measurement Good Practice Guide No. 130 Co-ordinate measuring machine task-specific measurement

More information

Determination of contributions of scatter and distance error to the source strength of 192 Ir HDR brachytherapy source

Determination of contributions of scatter and distance error to the source strength of 192 Ir HDR brachytherapy source Polish Journal of Medical Physics and Engineering 2016;22(3):55-59 September 2016 doi: 10.1515/pjmpe-2016-0010 ISSN 1898-0309 Scientific Paper Determination of contributions of scatter and distance error

More information

Simulation Modeling in Dosimetry

Simulation Modeling in Dosimetry Simulation Modeling in Dosimetry Aleksei Zhdanov Ural Federal University named after the first President of Russia B. N. Yeltsin, Yekaterinburg, Russian Federation jjj1994@yandex.ru Leonid Dorosinskiy

More information

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition INTERNATIONAL STANDARD ISO 6974-1 Second edition 2012-05-15 Natural gas Determination of composition and associated uncertainty by gas chromatography Part 1: General guidelines and calculation of composition

More information

Guidelines on the Calibration of Static Torque Measuring Devices

Guidelines on the Calibration of Static Torque Measuring Devices European Association of National Metrology Institutes Guidelines on the Calibration of Static Torque Measuring Devices EURAMET/cg-14/v.01 Previously EA-10/14 July 2007 Calibration Guide EURAMET/cg-14/v.01

More information

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong Original Article PMP Progress in Medical Physics 28(2), June 2017 https://doi.org/10.14316/pmp.2017.28.2.49 pissn 2508-4445, eissn 2508-4453 Electron Energy Distribution for a Research Electron LINAC Heuijin

More information

Chapter 10 Acceptance Tests and Commissioning Measurements

Chapter 10 Acceptance Tests and Commissioning Measurements Chapter 10 Acceptance Tests and Commissioning Measurements This set of 189 slides is based on Chapter 10 authored by J. L. Horton of the IAEA publication (ISBN 92-0-107304-6): Radiation Oncology Physics:

More information

TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1

TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1 TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1 (1) Instituto de Pesquisas Energéticas e Nucelares ABSTRACT In order

More information

IAEA-TECDOC-1585 Measurement Uncertainty

IAEA-TECDOC-1585 Measurement Uncertainty IAEA-TECDOC-1585 Measurement Uncertainty A Practical Guide for Secondary Standards Dosimetry Laboratories May 008 IAEA-TECDOC-1585 Measurement Uncertainty A Practical Guide for Secondary Standards Dosimetry

More information

Accuracy, Calibration, Type Testing and Traceability General. Peter Ambrosi

Accuracy, Calibration, Type Testing and Traceability General. Peter Ambrosi Accuracy, Calibration, Type Testing and Traceability General [Chapter 6 & 7 of RP 160] Peter Ambrosi EURADOS Training Course Lisbon, Portugal, 18-22 Mai Content Introduction - Aim of Measurement & type

More information

Circuits for Analog System Design Prof. Gunashekaran M K Center for Electronics Design and Technology Indian Institute of Science, Bangalore

Circuits for Analog System Design Prof. Gunashekaran M K Center for Electronics Design and Technology Indian Institute of Science, Bangalore Circuits for Analog System Design Prof. Gunashekaran M K Center for Electronics Design and Technology Indian Institute of Science, Bangalore Lecture No. # 08 Temperature Indicator Design Using Op-amp Today,

More information

Quality-Assurance Check of Collimator and Phantom- Scatter Factors

Quality-Assurance Check of Collimator and Phantom- Scatter Factors Quality-Assurance Check of Collimator and Phantom- Scatter Factors Ramesh C. Tailor, David S. Followill, Nadia Hernandez, Timothy S. Zhu, and Geoffrey S. Ibbott. UT MD Anderson Cancer Center, Houston TX.

More information

Errors and Uncertainties in Chemistry Internal Assessment

Errors and Uncertainties in Chemistry Internal Assessment Errors and Uncertainties in Chemistry Internal Assessment The treatment of errors and uncertainties is relevant in the internal assessment criteria of: data collection, aspect 1 (collecting and recording

More information

ROLE AND SIGNIFICANCE OF UNCERTAINTY IN HV MEASUREMENT OF PORCELAIN INSULATORS A CASE STUDY

ROLE AND SIGNIFICANCE OF UNCERTAINTY IN HV MEASUREMENT OF PORCELAIN INSULATORS A CASE STUDY International Conference on Ceramics, Bikaner, India International Journal of Modern Physics: Conference Series Vol. (01) 8 5 World Scientific Publishing Company DOI: 10.11/S010195101019 ROLE AND SIGNIFICANCE

More information

VAM Project Development and Harmonisation of Measurement Uncertainty Principles

VAM Project Development and Harmonisation of Measurement Uncertainty Principles VAM Project 3.2.1 Development and Harmonisation of Measurement Uncertainty Principles Part (d): Protocol for uncertainty evaluation from validation data V J Barwick and S L R Ellison January 2000 LGC/VAM/1998/088

More information

CMM Uncertainty Budget

CMM Uncertainty Budget Table of Contents Purpose...3 Scope...3 Measurement Example...3 Uncertainty Groups...4 Probe Group...5 Calibration Artifact Group...6 Environment Group...7 CMM Group...8 Part Group...10 Conversion of Potential

More information

PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS

PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS ABSTRACT LEDs are used in many applications, ranging from signalling to ambient lighting to displays. Specifications and tolerances are used

More information

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

2.1. Accuracy, n- how close the indication of the thermometer is to the true value. AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria

More information

Temperatures: Measure Thrice

Temperatures: Measure Thrice Temperatures: Measure Thrice At Least Before Reporting Once G. Raymond Peacock Temperatures.com, Inc. Southampton, PA 18966 Ph: 215-436-9730 Web: www.temperatures.com Email: grp@mrpyro.com Abstract A measurement

More information

factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons

factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons Calorimetric determination of k Q factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons PTB s water calorimeter in front of Elekta Precise medical linac Calorimetric determination

More information

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE Publication Reference EA-10/13 EA Guidelines on the Calibration of Temperature Block Calibrators PURPOSE This document been produced by EA to improve the harmonisation in the calibration of temperature

More information

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE Publication Reference EA-10/14 EA Guidelines on the Calibration of Static Torque Measuring Devices PURPOSE This document has been produced by EA to improve harmonisation in determining the calibration

More information

Uncertainty of Measurement (Analytical) Maré Linsky 14 October 2015

Uncertainty of Measurement (Analytical) Maré Linsky 14 October 2015 Uncertainty of Measurement (Analytical) Maré Linsky 14 October 015 Introduction Uncertainty of Measurement GUM (Guide to the Expression of Uncertainty in Measurement ) (Analytical) Method Validation and

More information

City University of Hong Kong

City University of Hong Kong City University of Hong Kong Information on a Course offered by the Department of Physics and Materials Science with effect from Semester A in 2013 / 2014 Part I Course Title: Radiological Physics and

More information

Calibration of Temperature Block Calibrators

Calibration of Temperature Block Calibrators European Association of National Metrology Institutes Calibration of Temperature Block Calibrators EURAMET cg-13 Version 2.0 (03/2011) Previously EA-10/13 Calibration Guide EURAMET cg-13 Version 2.0 (03/2011)

More information

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR ANALELE STIINTIFICE ALE UNIVERSITATII AL. I. CUZA IASI Tomul II, s. Biofizică, Fizică medicală şi Fizica mediului 2006 CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR Dan

More information

Objectives. Tolerance Limits and Action Levels for IMRT QA. The Overall Process of IMRT Planning and Delivery. Chain of IMRT Process

Objectives. Tolerance Limits and Action Levels for IMRT QA. The Overall Process of IMRT Planning and Delivery. Chain of IMRT Process Tolerance Limits and Action Levels for IMRT QA Jatinder R Palta, Ph.D Siyong Kim, Ph.D. Department of Radiation Oncology University of Florida Gainesville, Florida Objectives Describe the uncertainties

More information

P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05

P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05 P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05 This document provides guidance on the estimation of uncertainty

More information

57:020 Mechanics of Fluids and Transfer Processes OVERVIEW OF UNCERTAINTY ANALYSIS

57:020 Mechanics of Fluids and Transfer Processes OVERVIEW OF UNCERTAINTY ANALYSIS 57:020 Mechanics of Fluids and Transfer Processes OVERVIEW OF UNCERTAINTY ANALYSIS INTRODUCTION Measurement is the act of assigning a value to some physical variable. In the ideal measurement, the value

More information

A hybrid Measurement Systems Analysis and Uncertainty of Measurement Approach for Industrial Measurement in the Light Controlled Factory

A hybrid Measurement Systems Analysis and Uncertainty of Measurement Approach for Industrial Measurement in the Light Controlled Factory A hybrid Measurement Systems Analysis and Uncertainty of Measurement Approach for Industrial Measurement in the Light Controlled Factory J E Muelaner, A Francis, M Chappell, P G Maropoulos Laboratory for

More information

Reproducibility within the Laboratory R w Control Sample Covering the Whole Analytical Process

Reproducibility within the Laboratory R w Control Sample Covering the Whole Analytical Process Flowchart for Nordtest Method (a) Specify measurand Quantify components for within lab reproducibility A control samples B possible steps, not covered by the control sample Quantify bias components Convert

More information

y, x k estimates of Y, X k.

y, x k estimates of Y, X k. The uncertainty of a measurement is a parameter associated with the result of the measurement, that characterises the dispersion of the values that could reasonably be attributed to the quantity being

More information

Uncertainty in Humidity Measurement. Helping you make a better measurement.

Uncertainty in Humidity Measurement. Helping you make a better measurement. Uncertainty in Humidity Measurement Helping you make a better measurement. Webinar Presenters Bruce McDuffee Michael Boetzkes Agenda Terms & Definitions Understand the Uncertainty of Your Instrument Understanding

More information