Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical Physics University of Wisconsin
|
|
- Lauren Ross
- 5 years ago
- Views:
Transcription
1 Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical Physics University of Wisconsin NCCAAPM meeting April 17, 2015
2 Larry DeWerd has partial interest in Standard Imaging Inc.
3 Determination of your uncertainty with TG 51 addendum Presented the TG 51 addendum last October. Short review of TG 51 addendum Uncertainty determination
4 What are reference class ionization chambers %dd(10) is used for k Q No lead foil to be used anymore caused errors. Use eq. 15 in TG 51 Use of small volume chambers in relative dosimetry
5 Majority are 0.6 cm 3 Farmer-type chambers A-150 chambers explicitly excluded 5 scanning chambers, NO microchambers (Exception A26 from some preliminary measurements. Long term to come) No parallel plate chambers are included
6 A. k Q factors for new chambers B. Comparison of measured and calculated k Q factors: Measured k Q data is available for some types of chambers C. Non-water phantoms only water for dosimetry D. Uncertainty analysis for implementation of TG-51
7 1. TG-51 made the deliberate decision not to include uncertainties. 2. Other protocols have included uncertainty budgets and/or detailed reviews of uncertainty components. 3. TG 51 addendum gives some guidance on: i. How to develop an uncertainty budget ii. Typical values for individual components. 4. The ISO GUM is the starting point or NIST publication
8 Uncertainty tables and determinations are important for the goodness of your measurements. Uncertainty gives an indication of where you would expect your measurements to fall each time you measure it. Uncertainty is larger than precision of measurement
9 Systematic errors: Results from a bias on part of observer or a faulty calibration of equipment. Random errors: Fluctuations in observations. Generally a standard deviation can account for these. Reduce all sources of error. Now use Type A (random) and Type B Errors are truly errors something done wrong
10
11 Accuracy and Precision Accurate but not precise Precise but not accurate Precise and accurate
12 1 0.9 primary standard measurement frequency of result clinical measurement conventional true value a.u.
13 Today possible variation in readings are given by uncertainties. The difference between the measured value - measurand - and the conventional true value is generally never zero. There are uncertainties involved in the measurement that can be expressed. Uncertainty is a methodology to express the goodness of the value as to its accuracy.
14 The uncertainty associated with a measurement characterizes the dispersion of values that can be reasonably associated with the measurand. What is the difference between uncertainty, precision and accuracy or the error in measurement?
15 Type A and Type B uncertainties are used Type A uncertainty is estimated by the standard deviation of the mean value. These are measured results. Any valid statistical method for treating data can be used for Type A uncertainties
16 Type B is based on scientific judgment using all relevant information available such as: previous measurement data experience or general knowledge of behavoir of instruments used manufacturers specifications data of calibration or other reports uncertainties assigned to reference data taken from handbooks.
17 A limit(confidence interval) is generally used for Type B, designated by k. For Guassian distributions, If we are certain that the value lies between + L then 99% lie here and the confidence limit is designated by k. For 99% k=3. Express L in %, call it L % The uncertainty is then u=l % /k and expressed in a %. For 95%, k=2 and for 66% k=1
18
19 Also can assume some type of probability distribution, e.g. rectangular or triangular Rectangular - all values fall within these maximum limits, +M (express in %) then u=m/ 3. For example, this is used when a manufacturer gives maximum limits for a parameter, such as the range of the calibration extends from 4.7 to 5.3. (3.5%)
20 Triangular is when all values fall with limits of + M, but the values are more weighted toward the central value. All values do not have equal probability. Then the uncertainty is estimated by u=m/ 6 This isn t used as much. Manufacturer may say: the general value is centered at 5, with a range from 4.7 to 5.3 Triangular would be value is 5.0, with some outliers at 4.7 to 5.3 Then 2.4%
21 Procedure is outlined in NIST Technical Note 1297 (1994) Each uncertainty component is propagated sequentially throughout the measurement pathway by quadrature summation: Square root of sum of squares for k=1
22 Generally the symbol u is used Say you measure a quantity a number of times and the standard deviation of the measurement is + 0.5% Therefore your type A uncertainty for this measurement is u A =+ 0.5% (precision of measurement)
23 Uncertainties in % are combined as u c =(u A2 + u B2 ) 1/2 Therefore in a simple example, we have u c =( ) 1/2 = (8.66) 1/2 ~ 2.92% Note that the greatest uncertainty governs the overall uncertainty So at k=2, total uncertainty is 5.84%
24 Standard Uncertainty: Expected range of values that are attributed to a measured quantity (source strength) Expanded Uncertainty: Defined here as the Standard Uncertainty multiplied by a coverage factor of 2. This corresponds to a coverage probability or confidence level of approximately 95%
25 The uncertainty in the laboratory from all aspects of your measurement are determined. - all possible contributions Make your best estimate. You can list your best (lowest) uncertainty and your worst (highest) uncertainty.
26 The calibration of the chamber progresses from NIST through the ADCL to the clinic. The traceability to NIST is given by the ADCL as 0.7% at k=1 u total u 2 clinic u 2 ADCL ADCL uncertainty tables available if necessary
27 Further treatment of Uncertainty AAPM Summer School 2009
28 NIST has produced an explanatory document (a guide to the Guide) - NIST Technical Note 1297 Uncertainty budget broken down into: Measurement Calibration data Influence quantities Typical values discussed but emphasis on individual users constructing site-specific uncertainty budgets for their calibration situations
29 Want to take into account your measurement process Values and quantities are representative. You need to determine your clinic values Start with dose measurement.
30 TG 51 gives a measure of the absorbed dose to water using an ADCL calibrated ion chamber. D N 60 Co k M w,q D,w Q ion k Q is the factor that converts from the calibration beam ( 60 Co) to the user linac beam, defined by beam quality Q
31 Quantity Type A Type B Comment Calibration N Dw ADCL 0.7% ADCL Traceability to NIST uncertainty k Q 0.4% Picking Maximum P elec 0.1% Correction for charge reading for M corr
32 M corr,w M raw P TP P ion P pol P elec Each of these parameters have an uncertainty associated with them. There is also an uncertainty involved in the setup, which we will deal with as yet.
33 Many times the manufacturer of the barometer (thermometer) will give a response range, e.g. +0.5%. Assume a rectangular distribution % Keep in mind the equation for TP correction. Generally most clinics have a 0.1% k=1 for P TP
34 Uncertainty components examples Temperature Pressure correction P TP T w P air For T in C, P in kpa Need water temperature at ion chamber: Water should be in equilibrium with room Ion chamber should be in equilibrium with water P TP does not take account of thermal expansion of thimble Need air pressure in air volume: Only realistic to measure room air pressure Need to confirm air communication of ion chamber (ADCL)
35 Thermal equilibration Chamber at 22 C placed in water at 10 C Equilibration is pretty quick, but not instantaneous Reference: Das and Zhu (Med. Phys., 2004)
36 Quantity Type A Type B Comment Influence Quantities P TP 0.10 Clinic Measurement of P TP P Pol 0.05 Measurement P ion 0.10 Measurement P leak 0.05 If present Pre irrad leak 0.10 If present
37 M raw M(x,y,z,SSD,FS) Wherever the chamber actually is positioned and whatever the actual geometry, M will be assigned to: x = y = 0 cm (on axis) z = 10 cm (d ref ) SSD = 100 cm (assuming SSD setup) Field size = 10 cm x 10 cm By writing the equation out this way we identify the influence quantities, and therefore uncertainty components Uncertainty analysis is also a procedural review
38 Quantity Type A Setup Parameters SSD Setting 0.10 Depth Setting Linac Stability Type B Comment If not measured, then type B estimated 0.25 If not measured, then type B estimated 0.10 If not measured, then type B estimated
39 Quantity Type A Type B Calibration 0.81 Influence 0.19 Quantities Setup 0.29 Parameters Combined uncertainty Comment 0.88% combined uor Relative Expanded u= 1.76%
40 The majority of the uncertainty comes from the calibration. What does this say to you You can measure with better precision than this. Your overall accuracy to the conventional true value is confident to 1.76% - or 1.8%
41 Thanks are due to Students and staff of the UW ADCL All those who send us calibration instruments that support the research program of the UW Medical Radiation Research Center.
Updating reference dosimetry a decade after TG-51
Updating reference dosimetry a decade after TG-51 Malcolm McEwen Ionizing Radiation Standards Institute for National Measurement Standards National Research Council, Canada CE Presentation at AAPM Annual
More informationReference Dosimetry for Megavoltage Therapy Beams: Electrons
Reference Dosimetry for Megavoltage Therapy Beams: Electrons David Followill Ph.D Radiological Physics Center UT M.D.Anderson Cancer Center Houston TX Protocol for Clinical Reference Dosimetry of High-Energy
More information8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required
UPDATING TG-51 When will it end? Malcolm McEwen Ionizing Radiation Standards National Research Council, Canada AAPM Annual Meeting, Charlotte, 2012 What are these updates? Working Group review recommends:
More informationSmall Field Dosimetric Measurements with TLD-100, Alanine, and Ionization Chambers
Small Field Dosimetric Measurements with TLD-1, Alanine, and Ionization Chambers S. Junell a, L. DeWerd a, M. Saiful Huq b, J. Novotny Jr. b, M. uader b, M.F. Desrosiers c, G. Bednarz b a Department of
More informationOptimization of microionization chambers for small-field reference dosimetry
Optimization of microionization chambers for small-field reference dosimetry Jessica R. Snow University of Wisconsin-Madison Department of Medical Physics Medical Radiation Research Center NCCAAPM Fall
More information8/2/2012. Larry DeWerd has a partial interest in Standard Imaging
Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin AAPM July 2012 Larry DeWerd has a partial interest in Standard Imaging Traceability to NIST via calibration points
More informationStudy of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water
More informationReferensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden
Referensdosimetri Crister Ceberg Medical Radiation Physics Lund University Sweden Reference dosimetry Determination of absorbed dose to water under reference conditions Not accounting for uncertainties
More informationDosimetry: Electron Beams
ICTP SChool On MEdical PHysics For RAdiation THerapy: DOsimetry And TReatment PLanning For BAsic And ADvanced APplications 13-24 April 2015 Miramare, Trieste, Italy Dosimetry: Electron Beams G. Hartmann
More informationMeasurement And Uncertainty
Measurement And Uncertainty Based on Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results, NIST Technical Note 1297, 1994 Edition PHYS 407 1 Measurement approximates or
More informationABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *
Romanian Reports in Physics, Vol. 67, No. 3, P. 1152 1158, 2015 ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * ELENA STANCU 1,2, CATALIN VANCEA
More informationIntroduction to the evaluation of uncertainty
Manual of Codes of Practice for the Determination of Uncertainties in Mechanical Tests on Metallic Materials SECTION 1 Introduction to the evaluation of uncertainty F A Kandil National Physical Laboratory
More informationOA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents
UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC 17025 Table of contents 1 GENERAL... 2 2 THE STANDARD SIST EN ISO/IEC 17025... 2 3 SA'S POLICY IN IMPLEMENTING
More informationABSTRACT. Keywords: Megavoltage, dosimetry, TG51 protocol, TG21 protocol, parallel-plate chambers, crosscomparison. INTRODUCTION
CALCULATED TG51 TO TG21 ABSORBED-DOSE RATIOS FOR MOST WIDELY USED CYLINDRICAL AND PARALLEL-PLATE ION CHAMBERS OVER A RANGE OF PHOTON AND ELECTRON ENERGIES R.C. Tailor, and W.F. Hanson ABSTRACT The change
More informationEssentials of expressing measurement uncertainty
Essentials of expressing measurement uncertainty This is a brief summary of the method of evaluating and expressing uncertainty in measurement adopted widely by U.S. industry, companies in other countries,
More informationMEASUREMENT UNCERTAINTY PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER Prepared by MJ Mc Nerney for ENAO Assessor Calibration
MEASUREMENT PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE UNCERTAINTY OCTOBER/NOVEMBER 2012 Prepared by for ENAO Assessor Calibration B SCOPE Introduction House Rules Overview 17025 & 15189 MU Approaches
More informationComparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification.
Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification. Antonio Lopez Medina, Antonio Teijeiro, Daniela Medal, Francisco Salvador, Julio Vazquez, Manuel
More informationProgress in calculations of k Q for TG-51
1 Progress in calculations of k Q for TG-51 D. W. O. Rogers Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University Ottawa http://www.physics.carleton.ca/~drogers AAPM Charlotte,NC,
More informationAir kerma rate measurements from a miniature x-ray source using free-air ionization chambers
Air kerma rate measurements from a miniature x-ray source using free- ionization chambers Stephen D. Davis 1, John A. Micka 1, Larry A. DeWerd 1, and Thomas W. Rusch 2 1 University of Wisconsin, Madison,
More informationImplementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy
Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy M. Saiful Huq, PhD, FAAPM, FInstP Dept. of Radiation Oncology, University of
More informationElectron beam water calorimetry measurements to obtain beam quality conversion factors
Electron beam water calorimetry measurements to obtain beam quality conversion factors Bryan R. Muir, a) Claudiu D. Cojocaru, Malcolm R. McEwen, and Carl K. Ross Measurement Science and Standards, National
More informationDoubt-Free Uncertainty In Measurement
Doubt-Free Uncertainty In Measurement Colin Ratcliffe Bridget Ratcliffe Doubt-Free Uncertainty In Measurement An Introduction for Engineers and Students Colin Ratcliffe United States Naval Academy Annapolis
More informationNACP-02 perturbation correction factors for the NPL primary standard of absorbed dose to water in high energy electron beams
NACP-02 perturbation correction factors for the NPL primary standard of absorbed dose to water in high energy electron beams E. Chin 1, J. Seuntjens 1, H. Palmans 2, A. DuSautoy 2, D. Shipley 2, M. Bailey
More informationCALCULATION OF UNCERTAINTY IN CHEMICAL ANALYSIS. A.Gnanavelu
CALCULATION OF UNCERTAINTY IN CHEMICAL ANALYSIS A.Gnanavelu UNCERTAINTY (U) Every measurement has error which is unknown and unknowable. This unknown error is called measurement uncertainty. Types of Error
More informationComposite field dosimetry
Composite field dosimetry Hugo Bouchard, PhD, MCCPM Senior Research Scientist Radiation dosimetry group National Physical Laboratory May 2014 Overview 1. Introduction Dosimetry protocols IAEA formalism
More informationMonitor Unit Calculations for Photon and Electrons. AAMD Meeting Raleigh, NC October 3, John P. Gibbons Chief of Clinical Physics
Monitor Unit Calculations for Photon and Electrons AAMD Meeting Raleigh, NC October 3, 2014 John P. Gibbons Chief of Clinical Physics Outline I. TG71 Formation and Charge II. Photon Calculations III. Electron
More informationISO 376 Calibration Uncertainty C. Ferrero
ISO 376 Calibration Uncertainty C. Ferrero For instruments classified for interpolation, the calibration uncertainty is the uncertainty associated with using the interpolation equation to calculate a single
More informationHigh-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers
Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society of Radiobiology
More informationSTANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS
STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS M.A. HASSAN*, M.H. GABER**, E. ESMAT*, H.I. FARAG***, H.M. EISSA* *National Institute for Standards (NIS), Giza, Egypt **Biophysics Department,
More informationM [scale units/s] of the system
APPENDIX TO IAEA CALIBRATION CERTIFICATE RADIATION PROTECTION IONIZATION CHAMBER CALIBRATION PROCEDURES AT THE IAEA DOSIMETRY LABORATORY 1. INTRODUCTION 1.1 General Ionization chambers and electrometers
More informationUncertainty of Measurement
Uncertainty of Measurement Contents Uncertainty of Measurement.... 3 Uncertainty of Measurement and Measurement Error.... 4 Measuring Uncertainty.... 5 Coverage factor k...6 Factors Affecting Uncertainty....
More informationUnit 4. Statistics, Detection Limits and Uncertainty. Experts Teaching from Practical Experience
Unit 4 Statistics, Detection Limits and Uncertainty Experts Teaching from Practical Experience Unit 4 Topics Statistical Analysis Detection Limits Decision thresholds & detection levels Instrument Detection
More informationR. C. Tailor and W. F. Hanson. Department of Radiation Physics. The University of Texas, M.D. Anderson Cancer Center. Houston, Texas 77030
Page 1 of 231 Calculated absorbed-dose ratios, TG51/TG21, for most idely used cylindrical and parallel-plate ion chambers over a range of photon and electron energies R. C. Tailor and W. F. Hanson Department
More informationVolume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved
Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah
More informationUncertainty of Measurement A Concept
Uncertainty of Measurement A Concept Shweta V. Matey 1, Dr. Nitin K. Mandavgade 2, and Dr. R.R. Lakhe 3 1 (Asst. Professor, Dept. of Mechanical Engg, Lokmanya Tilak College of Engineering, Mumbai University,
More information) for Varian TrueBeam high-dose-rate therapy beams
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 Ion recombination correction factors (P ion ) for Varian TrueBeam high-dose-rate therapy beams Stephen F. Kry, 1a Richard Popple,
More informationUncertainty sources of reference measurement procedures for enzymes
Medical School Hannover Institute for Clinical Chemistry Director: Prof. Dr. K. Brand Uncertainty sources of reference measurement procedures for enzymes R. Klauke and G. Schumann Renate Strache Rainer
More informationSAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)
SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February
More informationWhat is measurement uncertainty?
What is measurement uncertainty? What is measurement uncertainty? Introduction Whenever a measurement is made, the result obtained is only an estimate of the true value of the property being measured.
More informationClinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry
Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry TJ JORDAN Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE
More informationPOWER UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES
Proceedings of the ASME 2014 Power Conference POWER2014 July 28-31, 2014, Baltimore, Maryland, USA POWER2014-32205 UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES Michael S. Hering Rosemount DP
More informationMonte Carlo modeling of an electronic brachytherapy source using MCNP5 and EGSnrc
Monte Carlo modeling of an electronic brachytherapy source using MCNP5 and EGSnrc Stephen D. Davis and Larry A. DeWerd University of Wisconsin, Madison, WI 13 th UK Monte Carlo User Group meeting March
More informationA MEASUREMENT ASSURANCE PROGRAM FOR LOW FLOW CALIBRATION BY THE TRANSFER METHOD CS20
A MEASUREMENT ASSURANCE PROGRAM FOR LOW FLOW CALIBRATION BY THE TRANSFER METHOD CS0 JEFF PEERY MAY. 18, 010 This document outlines the Measurement Assurance Program (MAP for Seametrics CS0 flo measurement
More informationComparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2
Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST Ronaldo Minniti 1 and Ladislav Czap 2 1 National Institute of Standards and Technology (NIST), Gaithersburg,
More informationGeneral characteristics of radiation dosimeters
General characteristics of radiation dosimeters and a terminology to describe them D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers
More informationHOW TO ASSESS THE MEASUREMENT UNCERTAINTY
15th EFLM Continuing Postgraduate Course in Clinical Chemistry and Laboratory Medicine How to assess the quality of your method? Zagreb, 24-25 October 2015 HOW TO ASSESS THE MEASUREMENT UNCERTAINTY Centre
More informationPractical Statistics for the Analytical Scientist Table of Contents
Practical Statistics for the Analytical Scientist Table of Contents Chapter 1 Introduction - Choosing the Correct Statistics 1.1 Introduction 1.2 Choosing the Right Statistical Procedures 1.2.1 Planning
More informationA Unified Approach to Uncertainty for Quality Improvement
A Unified Approach to Uncertainty for Quality Improvement J E Muelaner 1, M Chappell 2, P S Keogh 1 1 Department of Mechanical Engineering, University of Bath, UK 2 MCS, Cam, Gloucester, UK Abstract To
More informationMeasurement Uncertainty, March 2009, F. Cordeiro 1
Measurement Uncertainty, March 2009, F. Cordeiro 1 Uncertainty Estimation on the Quantification of Major Proteins in Milk by Liquid Chromatography Contribution to a Chemical Reference Measurement System
More informationChapter 9: Calibration of Photon and Electron Beams
Chapter 9: Calibration of Photon and Electron Beams Set of 189 slides based on the chapter authored by P. Andreo, J.P. Seuntjens, and E.B. Podgorsak of the IAEA publication (ISBN 92-0-107304-6): Radiation
More informationA comparison of methods for monitoring photon beam energy constancy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 A comparison of methods for monitoring photon beam energy constancy Song Gao, 1a Peter A. Balter, 1 Mark Rose, 2 and William E. Simon
More information1.11 Measurement Uncertainty
Document #: 4968 Page 1 of 10 1.11 Measurement Uncertainty In an effort to comply with accreditation requirements, and because scientific measurements in general are subject to variability, a budget estimating
More informationKnow Your Uncertainty
April 2000 METROLOGY Know Your Uncertainty Understanding and documenting measurement uncertainty is key to gage calibration. By Henrik S. Nielsen, Ph.D. The process of developing uncertainty budgets requires
More informationThe Determination of Uncertainties in Bend Tests on Metallic Materials
Manual of Codes of Practice for the Determination of Uncertainties in Mechanical Tests on Metallic Materials Code of Practice No. 09 The Determination of Uncertainties in end Tests on Metallic Materials
More informationStudy of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 1, WINTER 2008 Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration Pablo
More informationUse of a radioactive check device for redundancy check of ionization chambers
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Use of a radioactive check device for redundancy check of ionization chambers N. P. S. Sidhu,* Alkis Rouvas, and Patrick Cadman
More informationInvestigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL
Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL A. Specification Gamma-spectrometry method is used to identify and determine the activity concentration
More informationVocabulary of Metrology
Vocabulary of Metrology For Understating Uncertainty and Traceability Prepared by: Kim, Sang Ho Engineering Team Leader Modal Shop, Inc A PCB Group Company SI National Metrology Institute Secondary Calibration
More informationUncertainties associated with the use of a sound level meter
NPL REPORT DQL-AC 002 Uncertainties associated with the use of a sound level meter Richard Payne April 2004 April 2004 ABSTRACT Uncertainties associated with the use of a sound level meter Richard Payne
More informationThe EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1
The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1 1 Medical Radiation Physics Group, University of Oldenburg and Pius
More informationEfficiencies of Some Spherical Ion Chambers in Continuous and Pulsed Radiation: A Numerical Evaluation
Signature: Pol J Radiol, 05; 80: 55-5 DOI: 0.659/PJR.89450 ORIGINAL ARTICLE Received: 05.03.7 Accepted: 05.06.9 Published: 05..5 Authors Contribution: A Study Design B Data Collection C Statistical Analysis
More informationSmall Field Dosimetry and IAEA/AAPM Protocol
Small Field Dosimetry and IAEA/AAPM Protocol Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Emeritus Professor, Department of Medical Physics University of Wisconsin
More informationCommissioning of the Beta Secondary Standard (BSS2)
Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,
More informationGuide to the Expression of Uncertainty in Measurement (GUM)- An Overview
Estimation of Uncertainties in Chemical Measurement Guide to the Expression of Uncertainty in Measurement (GUM)- An Overview Angelique Botha Method of evaluation: Analytical measurement Step 1: Specification
More informationMeasurement Uncertainty Knowing the Unknown
Measurement Uncertainty Knowing the Unknown Anish Shah FOUNDER and Chief Metrology Officer at Metrologized, LLC, a company founded in 2014 to further the metrology knowledge-base within the manufacturing
More informationInvestigation of the standard temperature- pressure correction factor at low x-ray energies
Investigation of the standard temperaturepressure correction factor at low x-ray energies D. J. La Russa, M. R. McEwen and D. W. O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton
More informationMeasurement Uncertainty - How to Calculate It In The Medical Laboratory
Measurement Uncertainty - How to Calculate It In The Medical Laboratory Godfrey C. Moses, PhD, FCACB National Director, Gamma-Dynacare Medical Laboratories 11/18/2009 Moses, GC; GDML 1 Introduction Disclaimers
More informationGood Practice Guide No. 130
Good Practice Guide No. 130 Co-ordinate measuring machine task-specific measurement uncertainties David Flack Measurement Good Practice Guide No. 130 Co-ordinate measuring machine task-specific measurement
More informationDetermination of contributions of scatter and distance error to the source strength of 192 Ir HDR brachytherapy source
Polish Journal of Medical Physics and Engineering 2016;22(3):55-59 September 2016 doi: 10.1515/pjmpe-2016-0010 ISSN 1898-0309 Scientific Paper Determination of contributions of scatter and distance error
More informationSimulation Modeling in Dosimetry
Simulation Modeling in Dosimetry Aleksei Zhdanov Ural Federal University named after the first President of Russia B. N. Yeltsin, Yekaterinburg, Russian Federation jjj1994@yandex.ru Leonid Dorosinskiy
More informationProvläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition
INTERNATIONAL STANDARD ISO 6974-1 Second edition 2012-05-15 Natural gas Determination of composition and associated uncertainty by gas chromatography Part 1: General guidelines and calculation of composition
More informationGuidelines on the Calibration of Static Torque Measuring Devices
European Association of National Metrology Institutes Guidelines on the Calibration of Static Torque Measuring Devices EURAMET/cg-14/v.01 Previously EA-10/14 July 2007 Calibration Guide EURAMET/cg-14/v.01
More informationHeuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong
Original Article PMP Progress in Medical Physics 28(2), June 2017 https://doi.org/10.14316/pmp.2017.28.2.49 pissn 2508-4445, eissn 2508-4453 Electron Energy Distribution for a Research Electron LINAC Heuijin
More informationChapter 10 Acceptance Tests and Commissioning Measurements
Chapter 10 Acceptance Tests and Commissioning Measurements This set of 189 slides is based on Chapter 10 authored by J. L. Horton of the IAEA publication (ISBN 92-0-107304-6): Radiation Oncology Physics:
More informationTA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1
TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1 (1) Instituto de Pesquisas Energéticas e Nucelares ABSTRACT In order
More informationIAEA-TECDOC-1585 Measurement Uncertainty
IAEA-TECDOC-1585 Measurement Uncertainty A Practical Guide for Secondary Standards Dosimetry Laboratories May 008 IAEA-TECDOC-1585 Measurement Uncertainty A Practical Guide for Secondary Standards Dosimetry
More informationAccuracy, Calibration, Type Testing and Traceability General. Peter Ambrosi
Accuracy, Calibration, Type Testing and Traceability General [Chapter 6 & 7 of RP 160] Peter Ambrosi EURADOS Training Course Lisbon, Portugal, 18-22 Mai Content Introduction - Aim of Measurement & type
More informationCircuits for Analog System Design Prof. Gunashekaran M K Center for Electronics Design and Technology Indian Institute of Science, Bangalore
Circuits for Analog System Design Prof. Gunashekaran M K Center for Electronics Design and Technology Indian Institute of Science, Bangalore Lecture No. # 08 Temperature Indicator Design Using Op-amp Today,
More informationQuality-Assurance Check of Collimator and Phantom- Scatter Factors
Quality-Assurance Check of Collimator and Phantom- Scatter Factors Ramesh C. Tailor, David S. Followill, Nadia Hernandez, Timothy S. Zhu, and Geoffrey S. Ibbott. UT MD Anderson Cancer Center, Houston TX.
More informationErrors and Uncertainties in Chemistry Internal Assessment
Errors and Uncertainties in Chemistry Internal Assessment The treatment of errors and uncertainties is relevant in the internal assessment criteria of: data collection, aspect 1 (collecting and recording
More informationROLE AND SIGNIFICANCE OF UNCERTAINTY IN HV MEASUREMENT OF PORCELAIN INSULATORS A CASE STUDY
International Conference on Ceramics, Bikaner, India International Journal of Modern Physics: Conference Series Vol. (01) 8 5 World Scientific Publishing Company DOI: 10.11/S010195101019 ROLE AND SIGNIFICANCE
More informationVAM Project Development and Harmonisation of Measurement Uncertainty Principles
VAM Project 3.2.1 Development and Harmonisation of Measurement Uncertainty Principles Part (d): Protocol for uncertainty evaluation from validation data V J Barwick and S L R Ellison January 2000 LGC/VAM/1998/088
More informationCMM Uncertainty Budget
Table of Contents Purpose...3 Scope...3 Measurement Example...3 Uncertainty Groups...4 Probe Group...5 Calibration Artifact Group...6 Environment Group...7 CMM Group...8 Part Group...10 Conversion of Potential
More informationPRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS
PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS ABSTRACT LEDs are used in many applications, ranging from signalling to ambient lighting to displays. Specifications and tolerances are used
More information2.1. Accuracy, n- how close the indication of the thermometer is to the true value.
AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria
More informationTemperatures: Measure Thrice
Temperatures: Measure Thrice At Least Before Reporting Once G. Raymond Peacock Temperatures.com, Inc. Southampton, PA 18966 Ph: 215-436-9730 Web: www.temperatures.com Email: grp@mrpyro.com Abstract A measurement
More informationfactors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons
Calorimetric determination of k Q factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons PTB s water calorimeter in front of Elekta Precise medical linac Calorimetric determination
More informationEA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE
Publication Reference EA-10/13 EA Guidelines on the Calibration of Temperature Block Calibrators PURPOSE This document been produced by EA to improve the harmonisation in the calibration of temperature
More informationEA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE
Publication Reference EA-10/14 EA Guidelines on the Calibration of Static Torque Measuring Devices PURPOSE This document has been produced by EA to improve harmonisation in determining the calibration
More informationUncertainty of Measurement (Analytical) Maré Linsky 14 October 2015
Uncertainty of Measurement (Analytical) Maré Linsky 14 October 015 Introduction Uncertainty of Measurement GUM (Guide to the Expression of Uncertainty in Measurement ) (Analytical) Method Validation and
More informationCity University of Hong Kong
City University of Hong Kong Information on a Course offered by the Department of Physics and Materials Science with effect from Semester A in 2013 / 2014 Part I Course Title: Radiological Physics and
More informationCalibration of Temperature Block Calibrators
European Association of National Metrology Institutes Calibration of Temperature Block Calibrators EURAMET cg-13 Version 2.0 (03/2011) Previously EA-10/13 Calibration Guide EURAMET cg-13 Version 2.0 (03/2011)
More informationCHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR
ANALELE STIINTIFICE ALE UNIVERSITATII AL. I. CUZA IASI Tomul II, s. Biofizică, Fizică medicală şi Fizica mediului 2006 CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR Dan
More informationObjectives. Tolerance Limits and Action Levels for IMRT QA. The Overall Process of IMRT Planning and Delivery. Chain of IMRT Process
Tolerance Limits and Action Levels for IMRT QA Jatinder R Palta, Ph.D Siyong Kim, Ph.D. Department of Radiation Oncology University of Florida Gainesville, Florida Objectives Describe the uncertainties
More informationP103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05
P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05 This document provides guidance on the estimation of uncertainty
More information57:020 Mechanics of Fluids and Transfer Processes OVERVIEW OF UNCERTAINTY ANALYSIS
57:020 Mechanics of Fluids and Transfer Processes OVERVIEW OF UNCERTAINTY ANALYSIS INTRODUCTION Measurement is the act of assigning a value to some physical variable. In the ideal measurement, the value
More informationA hybrid Measurement Systems Analysis and Uncertainty of Measurement Approach for Industrial Measurement in the Light Controlled Factory
A hybrid Measurement Systems Analysis and Uncertainty of Measurement Approach for Industrial Measurement in the Light Controlled Factory J E Muelaner, A Francis, M Chappell, P G Maropoulos Laboratory for
More informationReproducibility within the Laboratory R w Control Sample Covering the Whole Analytical Process
Flowchart for Nordtest Method (a) Specify measurand Quantify components for within lab reproducibility A control samples B possible steps, not covered by the control sample Quantify bias components Convert
More informationy, x k estimates of Y, X k.
The uncertainty of a measurement is a parameter associated with the result of the measurement, that characterises the dispersion of the values that could reasonably be attributed to the quantity being
More informationUncertainty in Humidity Measurement. Helping you make a better measurement.
Uncertainty in Humidity Measurement Helping you make a better measurement. Webinar Presenters Bruce McDuffee Michael Boetzkes Agenda Terms & Definitions Understand the Uncertainty of Your Instrument Understanding
More information