1 Radiation Protection Dosimetry (2006), Vol. 118, No. 3, pp Advance Access publication 6 October 2005 doi: /rpd/nci353 DOSE BUILD UP CORRECTION FOR RADIATION MONITORS IN HIGH-ENERGY BREMSSTRAHLUNG PHOTON RADIATION FIELDS Haridas G. Nair 1,, M. K. Nayak 1, Vipin Dev 1, K. K. Thakkar 1, P. K. Sarkar 2 and D. N. Sharma 2 1 Health Physics Unit, Centre for Advanced Technology, Indore , India 2 Radiation Safety Systems Division, Bhabha Atomic Research Centre, Mumbai , India Received June , amended August , accepted September Conventional radiation monitors have been found to underestimate the personal dose equivalent in the high-energy bremsstrahlung photon radiation fields encountered near electron storage rings. Depth-dose measurements in a water phantom were carried out with a radiation survey meter in the bremsstrahlung photon radiation fields from a 450 MeV electron storage ring to find out the magnitude of the underestimation. Dose equivalent indicated by the survey meter was found to build up with increase in thickness of water placed in front of the meter up to certain depth and then reduce with further increase in thickness. A dose equivalent build up factor was estimated from the measurements. An absorbed dose build up factor in a water phantom was also estimated from calculations performed using the Monte Carlo codes, EGS-4 and EGSnrc. The calculations are found to be in very good agreement with the measurements. The studies indicate inadequacy of commercially available radiation monitors for radiation monitoring within shielded enclosures and in streaming high-energy photon radiation fields from electron storage rings, and the need for proper correction for use in such radiation fields. INTRODUCTION Bremsstrahlung X-ray photons form the major radiation hazard at electron storage rings. These photons have a continuous energy spectrum, whose maximum energy extends up to the energy of electrons in the ring. Bremsstrahlung X rays in electron storage rings are produced as a result of electron beam losses and their interaction with accelerator structures, mainly vacuum envelopes and residual gas molecules (1,2). Within shielded enclosures of the storage ring, the radiation field is mainly composed of these high-energy photons, whereas outside the radiation shield, photons are of relatively low energy owing to spectral degradation, due to the shield. There are locations like shield joints, holes in the shield structure for accommodating synchrotron radiation beam lines, front ends, etc. through which the high-energy bremsstrahlung photons can stream out to accessible areas. Also, there are occasions during which the accelerator operation crew has to go near the storage ring, within the shielded enclosure, at low electron beam currents for minor adjustments. In all these circumstances, workers are exposed to high-energy photon radiation. For monitoring of these photons no proper radiation survey instruments are commercially available since nearly all conventional radiation monitors have a useful energy response only up to a few MeV. When a high-energy photon is incident on a medium, an energetic electron and positron pair is Corresponding author: haridas@cat.ernet.in formed, as pair production is the dominant interaction mechanism at high photon energies. After the pair formation, the energetic electron and positron radiate out photons by radiative loss and these photons in turn produce further pairs, giving rise to a cascade of electrons, positrons and photons (3) in the medium. As a result, the particle fluence increases with depth of the medium and their mean energy decreases. Therefore, energy deposition in the medium increases (4) with depth giving rise to build up in absorbed dose. Thus, absorbed dose and, hence, the dose equivalent would be higher, deep within the body than the dose equivalent indicated by a conventional radiation survey meter. Conventional radiation monitors, usually calibrated in terms of personal dose equivalent, H p (10), when used to survey such high-energy photon radiation field do not indicate the true dose received by the worker as they show only the dose equivalent at a depth of 10 mm in tissue. Though the dose build up effect is taken into account in high-energy electron accelerator shielding applications and in electron beam shutter designs (5,6) it is usually not accounted for in monitors used for radiation protection purposes. Therefore, a study was carried out in the high-energy photon radiation field at Indus-1 (a 450 MeV electron storage ring (7), operational at the Centre for Advanced Technology, Indore, India) to find out the dose equivalent build up factor using a conventional radiation survey meter and a water phantom. In the study, a Victoreen survey meter, model 450 P (which is a pressurised ion chamber based survey meter, used routinely for Ó The Author Published by Oxford University Press. All rights reserved. For Permissions, please journals.permissions@oxfordjournals.org

2 H. G. NAIR ET AL. Figure 1. Schematic diagram of the experimental set-up used for the measurement of photon depth-dose profiles in water. radiation surveying in the facility) was used to measure the depth-dose profile in a water phantom. From the measurements, the dose equivalent build up factor was estimated. The build up factor was also calculated using the Monte Carlo codes, EGS-4 (8) and EGSnrc (9). From the dose build up studies, a conservative high-energy correction factor for the survey meter was deduced. DEPTH-DOSE MEASUREMENTS During the operation of the Indus-1 storage ring, depth-dose measurements in water phantom were carried out at (i) the bremsstrahlung photon radiation field near a bending dipole magnet within the shielded enclosure and (ii) the streaming radiation field at a front end, in an accessible area. For the measurements, the Victoreen survey meter was positioned at an appropriate distance from the bending dipole magnet at the median plane (1.25 m from floor level). The schematic diagram of the experimental set-up used for the measurements is shown in Figure 1. Initially the reading of the meter was noted without any water phantom. Thereafter, cylindrical water slabs made out of plastic containers of radius 150 mm and width of 50 mm were added in front of the survey meter and readings were taken at each depth. A similar experiment was carried out at the front end location. The measurement were carried out at a low stored electron beam current of 10 ma. It was essential to carry out the measurement at low beam current as it gave a steady radiation field due to higher beam lifetime in the ring. The depth-dose profile obtained is presented in Dose equivalent rate (msv/h) At front end At bending magnet Depthinwater(cm) Figure 2. Measured photon depth-dose profile at the front end and near the bending magnet of the 450 MeV storage ring, Indus-1. Figure 2. In both cases the dose equivalent rate was found to increase until a certain depth in water and then decrease. The dose equivalent build up factor was calculated as the ratio of the maximum dose equivalent rate with water phantom, recorded by the survey meter to that recorded without water phantom. The maximum dose was found to occur at different depths at each location and the build up factor was also found to be different. 234

3 MONTE CARLO CALCULATIONS Monte Carlo calculations were performed to estimate the dose build up factors. The different build up factors observed during measurement indicates difference in bremsstrahlung spectra at the measurement locations. Therefore to simulate different geometries, the calculation was done in two stages. In stage I, a set of bremsstrahlung spectra was generated by impinging a pencil beam of 450 MeV electrons on different thicknesses of semi-infinite copper Photon fluence (photon / cm 2 /e) DOSE BUILD UP CORRECTION FOR RADIATION MONITORS 3mm 50 mm 100 mm Photon energy (MeV) Figure 3. Calculated bremsstrahlung photon spectra from semi-infinite copper slabs on 450 MeV electron bombardment using EGS-4. plates and scoring the emergent bremsstrahlung spectra in the forward direction with a central axis scoring radius of 5 mm, using the EGS-4 code. Spectra from 3, 6, 15, 33, 50 and 100 mm thick copper targets were calculated. For radiation transport an electron cut-off (ECUT) of 521 kev and a photon cut-off (PCUT) of 10 kev was used in the calculations. Ten million histories were used for the depthdose calculation to have better statistical accuracy. The statistical accuracy obtained in the calculation was within 1%. Figure 3 shows the spectra obtained from 3, 50 and 100 mm thick copper targets. Copper was selected as the target material because materials in and around the storage ring are mainly of medium atomic number. In stage II, the photon spectra calculated in stage I were allowed to be incident as a broad parallel beam on cylindrical water slabs of radius 0.15 m, and the depth dose was scored using the DOSRZ user code from EGSnrc. A schematic diagram of the geometry used for the calculation is shown in Figure 4. Depth-dose curves obtained are presented in Figure 5. The maximum absorbed dose (D max ) and the depth where it occurs, hereafter called the shower maximum (X max ), was observed for each incident spectrum. From the depth-dose curves, dose build up factor was calculated as the ratio of the maximum absorbed dose (D max ) to the absorbed dose at a depth of 1 cm in water. The use of absorbed dose at a depth of 1 cm in water for the calculation of build up factor is justified since, for routine radiation monitoring, dosemeters are usually calibrated in terms of personal dose equivalent or ambient dose equivalent, at a depth of 1 cm in tissue. Therefore, absorbed dose at a depth of 1 cm in water Figure 4. Schematic diagram of the geometry used for Monte Carlo calculations of the depth-dose curves in water. 235

4 H. G. NAIR ET AL. Absorbed dose per incident photon fluence (Gy.cm 2 ) 4.50E E E E E E E E E-12 3mm 6mm 15 mm 33 mm 50 mm 100 mm 0.00E Depth in water (cm) Figure 5. Calculated photon depth-dose curves in water due to various incident photon spectra. corresponds to the dose rate the survey meter indicates without the water phantom. The calculated build up factor plotted as a function of the shower maximum is presented in Figure 6. RESULTS AND ANALYSIS The depth-dose profiles measured in the bremsstrahlung photon radiation field near the bending dipole magnet of the storage ring and in the streaming photon radiation field through the front end (Figure 2) indicate that the dose equivalent rate increases up to a certain depth in water and declines thereafter. The profile near the bending magnet showed that the maximum dose equivalent rate occurred at a depth of 15 cm in water and the build up factor obtained was 2.5. The respective values obtained at the front end were 20 cm and 3.3. The shower maximum and the build up factor obtained at the front end were found to be higher than those at the bending magnet. The different shower maximum and build up factor at each place may be attributed to the difference in bremsstrahlung spectra at these locations. Photon spectra calculated for different thickness of copper target (Figure 3) indicate that the photon spectra emerging from thin targets are more penetrating (harder) than those from thick targets. Since the structural materials in the storage ring vary in thickness from a few millimetres to tens of millimetres, the spectra generated from different thickness of copper targets were used to generate depth-dose curves in water phantom (Figure 5). Dose equivalent build up factor Calculated build up factor Fit Shower maximum in water, X max (cm) Figure 6. Dose equivalent build up factor for various incident bremsstrahlung photon spectra as a function of shower maximum. Table 1. Comparison of dose build up factor obtained from measurements and Monte Carlo calculations. Location Shower maximum, X max (cm) (measured) Build up factor Measured Simulated a Bending magnet Front end a From Equation 1 in the text. The build up factor, B, obtained from the calculated depth-dose curves in water was found to increase with the shower maximum (Figure 6). An exponential fit to the build up factor as a function of the shower maximum, X max assumes the following form B ¼ 1:08 exp½x max =16:9Š ð1þ where X max is in centimetres. The dose build up factors obtained from the exponential fit as a function of the shower maximum were found to be in very good agreement with measured values (Table 1). As the vacuum chamber is generally of only few millimetres thick, harder bremsstrahlung spectra are possible around the ring, and hence the shower maximum shifts to higher values. Therefore, evaluating the fit function for X max ¼ 30 cm, the size of human body, the build up factor works out to be 6.4. This build up factor can be used as a 236

5 DOSE BUILD UP CORRECTION FOR RADIATION MONITORS conservative correction factor for the survey meter reading for radiation protection in the high-energy photon radiation field emerging from the storage ring. Since the corrected dose equivalent is nearly equal to the maximum dose equivalent in a 30 cm diameter sphere consisting of material equivalent to soft tissue with unit density the corrected result equals the dose equivalent index, which is suggested for radiation protection at high-energy accelerator facilities (10). storage ring to carry out experiments. We are highly indebted to Dr R. V. Nandedkar, Head Synchrotron Utilisation Division for technical discussions on the experimental results. We also thank S. K. Gupta for helping us in the use of EGS-4 code for the work. The authors are indebted to Dr M. V. Dingankar and D. S. Joshi, Accelerator Radiation Safety Section, for reviewing the manuscript and giving useful suggestions. We would also like to thank S. Jafar Ali for preparing the drawings. CONCLUSION Experimentally measured dose build up factors around the storage ring, Indus-1, were found to be in very good agreement with those obtained from Monte Carlo simulations. It is concluded from the studies that the radiation field around the storage ring and at streaming locations is comprised of high-energy photons and with spectra varying at different locations. Conventional radiation survey meters when used for monitoring in such radiation fields show only the dose equivalent at 1 cm, which a worker receives, whereas deep within the body the dose received might be much higher than that indicated by the radiation survey meter. Therefore, survey meters, area monitors, personnel dosemeters, etc. when used for measurements in high-energy photon radiation fields have to be properly corrected for high-energy response by applying correction factors deduced from dose equivalent measurements using a water phantom. It is planned to study the dose build up effects in radiation monitors for still higher electron energy electron storage rings. ACKNOWLEDGEMENTS We are highly thankful to Dr H. S. Kushwaha, Director, Health, Safety & Environment Group, Bhabha Atomic Research Centre, for providing all the necessary support and motivation in carrying out the present investigations. The authors acknowledge with gratitude Gurnam Singh, Head, Beam Dynamics Section and his entire team including the acceleration operation crew for providing the REFERENCES 1. Rindi, A. Gas bremsstrahlung from electron storage rings. Health Phys. 42(2), (1982). 2. Ipe, N. E. and Fasso, A. Gas bremsstrahlung considerations in the shielding design of the Advanced Photon Source synchrotron radiation beam lines. Nucl. Instrum. Meth. Phys. Res. A 351, (1994). 3. International Commission on Radiation Units and Measurements. Basic aspects of high energy particle interactions and radiation dosimetry. ICRU Report 28 (Bethesda, MD: ICRU) (1978). 4. Hubbel, J. H. Photon cross-sections, attenuation coefficients and energy absorption coefficients from 10 kev to 100 GeV. NSRDS-NBS Report No. 29 US Department of Commerce, US Government Printing Office (1969). 5. Dinter, H. and Tesch, K. Dose and shielding parameters of electron photon stray radiation from a high-energy electron Beam. Nucl. Instrum. Meth. Phys. Res. 143, (1977). 6. Dinter, H., Pang, J. and Tesch, K. Calculation of electron and photon doses behind beam absorbers at high-energy electron accelerators. Radiat. Prot. Dosim. 28(3), (1989). 7. Angal-Kalinin, D. et al. Synchrotron Radiation source Indus-1. Curr. Sci. 82(3), (2002). 8. Nelson, W. R., Hirayama, H. and Rogers, D. W. O. The EGS-4 code system. Report SLAC-265 (Stanford Linear Accelerator Center, Stanford, CA) (1985). 9. Rogers, D. W. O., Kawrakow, I., Seuntjens, J. P., Walters, B. R. B. and Mainegra Hing, E. NRC user codes for EGSnrc. NRCC Report PIRS-702 National Research Council of Canada. (Rev. B) (2003). Available on EGSnrc.html 10. International Commission on Radiation Units and Measurements. Radiation quantities and units. ICRU Report 19 (Bethesda, MD: ICRU) (1971). 237

Radiation shielding for undulator beamline in Indus-2 synchrotron radiation source

Radiation shielding for undulator beamline in Indus-2 synchrotron radiation source Radiation shielding for undulator beamline in Indus-2 synchrotron radiation source P. K. Sahani 1,5, A. K. Das 2, Haridas G. 3, A. K. Sinha 4,5, B. N. Rajasekhar 2,5, T. A. Puntambekar 1 and N K Sahoo

More information

SLAC-PUB Submitted to Radiation Protection and Dosimetry. Work supported by Department of Energy contract DE-AC02-76SF00515

SLAC-PUB Submitted to Radiation Protection and Dosimetry. Work supported by Department of Energy contract DE-AC02-76SF00515 SLAC-PUB-11088 CALCULATIONS OF NEUTRON AND PHOTON SOURCE TERMS AND ATTENUATION PROFILES FOR THE GENERIC DESIGN OF THE SPEAR3 STORAGE RING SHIELD S. H. Rokni, H. Khater, J. C. Liu, S. Mao and H. Vincke

More information

Occupational Radiation Protection at Accelerator Facilities: Challenges

Occupational Radiation Protection at Accelerator Facilities: Challenges Occupational Radiation Protection at Accelerator Facilities: Challenges Haridas.G Health Physics Division Bhabha Atomic Research Centre INDIA Int. Conf. on Occupational Radiation Protection: Enhancing

More information

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR ANALELE STIINTIFICE ALE UNIVERSITATII AL. I. CUZA IASI Tomul II, s. Biofizică, Fizică medicală şi Fizica mediului 2006 CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR Dan

More information

PHOTON-NEUTRON SHIELDING CALCULATIONS FOR ELECTRON STORAGE RINGS. J. C. Liu, W. R. Nelson and K. R. Kase ~~~_

PHOTON-NEUTRON SHIELDING CALCULATIONS FOR ELECTRON STORAGE RINGS. J. C. Liu, W. R. Nelson and K. R. Kase ~~~_ ~~~. ~-_ SLAC-PUB-6532 June 1994 (A) GAS BREMStiRAHLUNG AND ASSOCIATED PHOTON-NEUTRON SHIELDING CALCULATIONS FOR ELECTRON STORAGE RINGS by J. C. Liu, W. R. Nelson and K. R. Kase ~~~_ Stanford Linear Accelerator

More information

Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials*

Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials* SLAC-PUB-70 Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials* J. C. Liu, K. R. Kase, X. S. Mao, W. R. Nelson, J. H. Kleck, and S. Johnson ) Stanford Linear

More information

Dosimetric Quantities and Neutron Spectra Outside the Shielding of Electron Accelerators

Dosimetric Quantities and Neutron Spectra Outside the Shielding of Electron Accelerators SLAC-PUB-15257 Dosimetric Quantities and Neutron Spectra Outside the Shielding of Electron Accelerators Alberto Fassò a,b, James C. Liu a and Sayed H. Rokni a* a SLAC National Accelerator Laboratory, 2575

More information

Radiation Protection At Synchrotron Radiation Facilities

Radiation Protection At Synchrotron Radiation Facilities 3 rd ILSF Advanced School on Synchrotron Radiation and Its Applications September 14-16, 2013 Radiation Protection At Synchrotron Radiation Facilities Ehsan Salimi Shielding and Radiation Safety Group

More information

Shielding calculations for the design of new Beamlines at ALBA Synchrotron

Shielding calculations for the design of new Beamlines at ALBA Synchrotron Shielding calculations for the design of new Beamlines at ALBA Synchrotron A. Devienne 1, M.J. García-Fusté 1 1 Health & Safety Department, ALBA Synchrotron, Carrer de la Llum -6, 0890 Cerdanyola del Vallès,

More information

7. a XV-2 high spatial resolution lm detector (Kodak). Important parameters of these detectors are given in Table1. The ionization chambers and the di

7. a XV-2 high spatial resolution lm detector (Kodak). Important parameters of these detectors are given in Table1. The ionization chambers and the di Proceedings of the Second International Workshop on EGS, 8.-12. August 2000, Tsukuba, Japan KEK Proceedings 200-20, pp.264-271 Variation of Dose Distribution by Detectors for Narrow Beam T. Fujisaki, H.

More information

Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry

Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry F. Gutermuth *, S. Beceiro, H. Emling, G. Fehrenbacher, E. Kozlova, T. Radon, T. Aumann, T. Le Bleis, K. Boretzky, H. Johansson,

More information

Radiation Safety Assessment of the CLS Beamlines Using FLUKA Monte-Carlo Code

Radiation Safety Assessment of the CLS Beamlines Using FLUKA Monte-Carlo Code Radiation Safety Assessment of the CLS Beamlines Using FLUKA Monte-Carlo Code Mo Benmerrouche Fluka Advanced Workshop - Oct 07, 2010 Ericeira, Portugal M. Benmerrouche, HSE Manager http://www.lightsource.ca

More information

Characterization of Low-Energy (6-30 kev) Response of Polish TLDs (MTS-N, MCP-N) with Synchrotron Radiation and Determination of Some TLD Quantities.

Characterization of Low-Energy (6-30 kev) Response of Polish TLDs (MTS-N, MCP-N) with Synchrotron Radiation and Determination of Some TLD Quantities. SLAC-PUB-7906 November, 1998 Characterization of Low-Energy (6-30 kev) Response of Polish TLDs (MTS-N, MCP-N) with Synchrotron Radiation and Determination of Some TLD Quantities. N. E. Ipe, A. Fassò, K.

More information

Comparison of FLUKA and STAC8 for shielding calculations of the hard X-ray line of the LCLS

Comparison of FLUKA and STAC8 for shielding calculations of the hard X-ray line of the LCLS SLAC RADIATION PHYSICS NOTE RP-08-11 September 23, 2008 Comparison of FLUKA and STAC8 for shielding calculations of the hard X-ray line of the LCLS J. Vollaire, A. Prinz Radiation Protection Department,

More information

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT Proceedings of the Eleventh EGS4 Users' Meeting in Japan, KEK Proceedings 2003-15, p.107-113 ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT K. Noto and

More information

Since the beam from the JNC linac is a very high current, low energy beam, energy loss induced in the material irradiated by the beam becomes very lar

Since the beam from the JNC linac is a very high current, low energy beam, energy loss induced in the material irradiated by the beam becomes very lar Proceedings of the Second International Workshop on EGS, 8.-12. August 2000, Tsukuba, Japan KEK Proceedings 200-20, pp.255-263 Beam Dump for High Current Electron Beam at JNC H. Takei and Y. Takeda 1 Japan

More information

Energy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration Fields of KAERI

Energy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration Fields of KAERI Energy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration s of KAERI B.H. Kim 1, J.L. Kim 1, S.Y. Chang 1, J.K. Chang 1, G. Cho 2 1 Korea

More information

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong Original Article PMP Progress in Medical Physics 28(2), June 2017 https://doi.org/10.14316/pmp.2017.28.2.49 pissn 2508-4445, eissn 2508-4453 Electron Energy Distribution for a Research Electron LINAC Heuijin

More information

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 AUTHORS: J. Cardoso, L. Santos, C. Oliveira ADRESS: Instituto Tecnológico e Nuclear Estrada Nacional 10; 2686-953 Sacavém;

More information

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator Determination of Ambient Dose quivalent at INFLPR 7 MeV Linear Accelerator F. Scarlat, A. Scarisoreanu, M. Oane,. Badita,. Mitru National Institute for Laser, Plasma and Radiation Physics - INFLPR, Bucharest-Magurele,

More information

Transport under magnetic fields with the EGSnrc simulation toolkit

Transport under magnetic fields with the EGSnrc simulation toolkit Transport under magnetic fields with the EGSnrc simulation toolkit Ernesto Mainegra-Hing, Frédéric Tessier, Blake Walters Measurement Science and Standards, National Research Council Canada Hugo Bouchard

More information

Shielding Design for the Imaging and Medical Beamline at the Australian Synchrotron

Shielding Design for the Imaging and Medical Beamline at the Australian Synchrotron Shielding Design for the Imaging and Medical Beamline at the Australian Synchrotron P. Berkvens and D. Häusermann European Synchrotron Radiation Facility BP 0, Grenoble Cedex 0, France Australian Synchrotron

More information

PERSPECTIVES OF PERSONNEL EXTERNAL DOSIMETRY AT STANFORD LINEAR ACCELERATOR CENTER

PERSPECTIVES OF PERSONNEL EXTERNAL DOSIMETRY AT STANFORD LINEAR ACCELERATOR CENTER SLAC-PUB-95-6749 (March 1995) PERSPECTIVES OF PERSONNEL EXTERNAL DOSIMETRY AT STANFORD LINEAR ACCELERATOR CENTER J. C. Liu, D. Busick 1, K. R. Kase, R. C. McCall 2, R. Sit and H. Tran 3 Stanford Linear

More information

SHIELDING CALCULATIONS FOR THE HARD X-RAY GENERATED BY LCLS MEC LASER SYSTEM R. QIU, J. C. LIU, S. H. ROKNI AND A. A. PRINZ

SHIELDING CALCULATIONS FOR THE HARD X-RAY GENERATED BY LCLS MEC LASER SYSTEM R. QIU, J. C. LIU, S. H. ROKNI AND A. A. PRINZ SLAC-PUB-14159 SHIELDING CALCULATIONS FOR THE HARD X-RAY GENERATED BY LCLS MEC LASER SYSTEM R. QIU, J. C. LIU, S. H. ROKNI AND A. A. PRINZ SLAC National Accelerator Laboratory: 2575 Sand Hill Road, Menlo

More information

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators N. Golnik 1, P. Kamiński 1, M. Zielczyński 2 1 Institute of Precision and Biomedical Engineering,

More information

Radiation Shielding of Extraction Absorbers for a Fermilab Photoinjector

Radiation Shielding of Extraction Absorbers for a Fermilab Photoinjector Fermilab FERMILAB-TM-2220 August 2003 Radiation Shielding of Extraction Absorbers for a Fermilab Photoinjector I.L. Rakhno Fermilab, P.O. Box 500, Batavia, IL 60510, USA August 12, 2003 Abstract Results

More information

Monte Carlo Calculations Using MCNP4B for an Optimal Shielding Design. of a 14-MeV Neutron Source * James C. Liu and Tony T. Ng

Monte Carlo Calculations Using MCNP4B for an Optimal Shielding Design. of a 14-MeV Neutron Source * James C. Liu and Tony T. Ng SLAC-PUB-7785 November, 1998 Monte Carlo Calculations Using MCNP4B for an Optimal Shielding Design of a 14-MeV Neutron Source * James C. Liu and Tony T. Ng Stanford Linear Accelerator Center MS 48, P.O.

More information

Gas Bremsstrahlung at the Diamond Light Source. P. Bonner, R. Ryder

Gas Bremsstrahlung at the Diamond Light Source. P. Bonner, R. Ryder Gas Bremsstrahlung at the Diamond Light Source. P. Bonner, R. Ryder Diamond Light Source Limited, Rutherford Appleton Laboratory, Chilton, Didcot, OX11 0QX, United Kingdom. E-mail: paul.bonner@diamond.ac.uk

More information

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER M. Bovi (1), R.F. Laitano (1), M. Pimpinella (1), M. P. Toni (1), K. Casarin(2), E. Quai(2), G. Tromba(2), A. Vascotto(2),

More information

Journal of Radiation Protection and Research

Journal of Radiation Protection and Research 1) JONG WOON KIM AND YOUNG-OUK LEE: DETAILED ANALYSIS OF THE KAERI ntof FACILITY Journal of Radiation Protection and Research pissn 2508-1888 eissn 2466-2461 http://dx.doi.org/10.14407/jrpr.2016.41.2.141

More information

THE ACTIVE PERSONNEL DOSIMETER - APFEL ENTERPRISES SUPERHEATED DROP DETECTOR*

THE ACTIVE PERSONNEL DOSIMETER - APFEL ENTERPRISES SUPERHEATED DROP DETECTOR* SLAC-PUB-5122 Rev March 1991 w THE ACTIVE PERSONNEL DOSIMETER - APFEL ENTERPRISES SUPERHEATED DROP DETECTOR* N. E. Ipe, R. J. Donahue, and D. D. Busick Stanford Linear Accelerator Center Stanford University,

More information

1. RADIOACTIVITY AND RADIATION PROTECTION

1. RADIOACTIVITY AND RADIATION PROTECTION 1. Radioactivity and radiation protection 1 1. RADIOACTIVITY AND RADIATION PROTECTION Revised August 2011 by S. Roesler and M. Silari (CERN). 1.1. Definitions [1,2] 1.1.1. Physical quantities: Fluence,

More information

Radiological safety studies for the TeraFERMI beamline at

Radiological safety studies for the TeraFERMI beamline at Radiological safety studies for the TeraFERMI beamline at FERMI@elettra K.Casarin 1, L. Fröhlich 2, G.Tromba 1, A.Vascotto 1 1 Elettra - Sincrotrone Trieste S.C.p.A., Trieste, Italy 2 Deutsches Elektronen-Synchrotron

More information

1 Introduction. A Monte Carlo study

1 Introduction. A Monte Carlo study Current Directions in Biomedical Engineering 2017; 3(2): 281 285 Sebastian Richter*, Stefan Pojtinger, David Mönnich, Oliver S. Dohm, and Daniela Thorwarth Influence of a transverse magnetic field on the

More information

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors 6 th International Congress of Metrology, 05004 (203) DOI: 0.05/ metrology/20305004 C Owned by the authors, published by EDP Sciences, 203 Design, construction and characterization of a portable irradiator

More information

Analyzing the Sensitivity of a Hard X Ray Detector Using Monte Carlo Methods

Analyzing the Sensitivity of a Hard X Ray Detector Using Monte Carlo Methods 1 Analyzing the Sensitivity of a Hard X Ray Detector Using Monte Carlo Methods Junhong Sam Zhou Victor Senior High School LLE Advisor: Christian Stoeckl Laboratory for Laser Energetics University of Rochester

More information

Geant4 Monte Carlo code application in photon interaction parameter of composite materials and comparison with XCOM and experimental data

Geant4 Monte Carlo code application in photon interaction parameter of composite materials and comparison with XCOM and experimental data Indian Journal of Pure & Applied Physics Vol. 54, Februray 2016, pp. 137-143 Geant4 Monte Carlo code application in photon interaction parameter of composite materials and comparison with XCOM and experimental

More information

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS Leonardo de Holanda Mencarini 1,2, Claudio A. Federico 1,2 and Linda V. E. Caldas 1 1 Instituto de Pesquisas Energéticas e Nucleares IPEN,

More information

Comparison of some human tissues and some commonly used thermoluminescent dosimeters for photon energy absorption

Comparison of some human tissues and some commonly used thermoluminescent dosimeters for photon energy absorption American Journal of Physics and Applications 2014; 2(6): 145-149 Published online December 16, 2014 (http://www.sciencepublishinggroup.com/j/ajpa) doi: 10.11648/j.ajpa.20140206.15 ISSN: 2330-4286 (Print);

More information

Calculation of photon energy deposition kernels and electron dose point kernels in water

Calculation of photon energy deposition kernels and electron dose point kernels in water Calculation of photon energy deposition kernels and electron dose point kernels in water Ernesto Mainegra-Hing a Ionizing Radiation Standards, National Research Council of Canada, Ottawa K1A OR6, Canada

More information

Energy response for high-energy neutrons of multi-functional electronic personal dosemeter

Energy response for high-energy neutrons of multi-functional electronic personal dosemeter Energy response for high-energy neutrons of multi-functional electronic personal dosemeter T. Nunomiya 1, T. Ishikura 1, O. Ueda 1, N. Tsujimura 2,, M. Sasaki 2,, T. Nakamura 1,2 1 Fuji Electric Systems

More information

Calculation of Effective Dose Conversion Coefficients for Electrons

Calculation of Effective Dose Conversion Coefficients for Electrons Calculation of Effective Dose Conversion Coefficients for Electrons S. Tsuda 1, A. Endo 1, Y. Yamaguchi 1 and O. Sato 2 1 Department of Health Physics, Japan Atomic Energy Research Institute (JAERI) Tokai-mura,

More information

Chapter 30 X Rays GOALS. When you have mastered the material in this chapter, you will be able to:

Chapter 30 X Rays GOALS. When you have mastered the material in this chapter, you will be able to: Chapter 30 X Rays GOALS When you have mastered the material in this chapter, you will be able to: Definitions Define each of the following terms, and use it in an operational definition: hard and soft

More information

Estimation of Radioactivity and Residual Gamma-ray Dose around a Collimator at 3-GeV Proton Synchrotron Ring of J-PARC Facility

Estimation of Radioactivity and Residual Gamma-ray Dose around a Collimator at 3-GeV Proton Synchrotron Ring of J-PARC Facility Estimation of Radioactivity and Residual Gamma-ray Dose around a Collimator at 3-GeV Proton Synchrotron Ring of J-PARC Facility Y. Nakane 1, H. Nakano 1, T. Abe 2, H. Nakashima 1 1 Center for Proton Accelerator

More information

Radiation Protection Considerations for the Cryogenic In-Vacuum Undulator of the EMIL Project at BESSY

Radiation Protection Considerations for the Cryogenic In-Vacuum Undulator of the EMIL Project at BESSY Radiation Protection Considerations for the Cryogenic In-Vacuum Undulator of the EMIL Project at BESSY Yvonne Bergmann, Klaus Ott Helmholtz- Zentrum Berlin BESSY II Radiation Protection Department yvonne.bergmann@helmholtz-berlin.de

More information

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements Specific Accreditation Criteria Calibration ISO/IEC 17025 Annex Ionising radiation measurements January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected

More information

In vacuum ID beam line shielding commissioning and direct gasbremsstrahlung measurements at Synchrotron SOLEIL

In vacuum ID beam line shielding commissioning and direct gasbremsstrahlung measurements at Synchrotron SOLEIL In vacuum ID beam line shielding commissioning and direct gasbremsstrahlung measurements at Synchrotron SOLEIL J-B. Pruvost 1, P. Berkvens 2, F. Justine 1, C. Mage 1 1: Synchrotron SOLEIL BP.48, 91192

More information

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic More Info at Open Access Database www.ndt.net/?id=16530 Quantitative Assessment of Scattering

More information

Comparative Analysis of Nuclear Cross Sections in Monte Carlo Methods for Medical Physics Applications

Comparative Analysis of Nuclear Cross Sections in Monte Carlo Methods for Medical Physics Applications Comparative Analysis of Nuclear Cross Sections in Monte Carlo Methods for Medical Physics Applications Christopher T. Myers 1 Georgia Institute of Technology Bernadette L. Kirk 2 Luiz C. Leal 2 Oak Ridge

More information

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Recent Activities on Neutron Calibration Fields at FRS of JAERI Recent Activities on Neutron Calibration Fields at FRS of JAERI Michio Yoshizawa, Yoshihiko Tanimura, Jun Saegusa and Makoto Yoshida Department of Health Physics, Japan Atomic Energy Research Institute

More information

Monte Carlo-based investigation of absorbed-dose energy dependence of radiochromic films in high energy brachytherapy dosimetry

Monte Carlo-based investigation of absorbed-dose energy dependence of radiochromic films in high energy brachytherapy dosimetry JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Monte Carlo-based investigation of absorbed-dose energy dependence of radiochromic films in high energy brachytherapy dosimetry Mishra

More information

The photoneutron yield predictions by PICA and comparison with the measurements

The photoneutron yield predictions by PICA and comparison with the measurements The photoneutron yield predictions by PICA and comparison with the measurements P. K. Job Advanced Photon Source Argonne National Laboratory Argonne, IL 60349 T. G Gabriel OakRidge Detector Center OakRidge

More information

A comprehensive study on energy absorption and exposure buildup factors for some Vitamins and Tissue Equivalent Materials

A comprehensive study on energy absorption and exposure buildup factors for some Vitamins and Tissue Equivalent Materials International Journal of Scientific and Research Publications, Volume 2, Issue 9, September 2012 1 A comprehensive study on energy absorption and exposure buildup factors for some Vitamins and Tissue Equivalent

More information

BEAMnrc: a code to simulate radiotherapy external beam sources

BEAMnrc: a code to simulate radiotherapy external beam sources BEAMnrc: a code to simulate radiotherapy external beam sources D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University Ottawa, Canada http://www.physics.carleton.ca/~drogers

More information

Laboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3)

Laboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3) PROPOSAL FOR EYE-LENS DOSEMETER CALIBRATION AND TYPE TESTING ORAMED WP2 J.-M. Bordy (1), J. Daures (1), M. Denozière (1), G. Gualdrini (2), M. Guijaume (3), E. Carinou (4),F. Vanhavere (5) (1) Laboratoire

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 6980-2 Première edition 2004-10-15 Nuclear energy Reference beta-particle radiation Part 2: Calibration fundamentals related to basic quantities characterizing the radiation

More information

PHYS 5020 Computation and Image Processing

PHYS 5020 Computation and Image Processing PHYS 5020 and Image Processing : Monte Carlo Thursday 2 August 2012 Monte Carlo (MC) is a numerical method that uses random sampling of probability distributions to simulate stochastic processes in nature,

More information

Update on Calibration Studies of the Canadian High-Energy Neutron Spectrometry System (CHENSS)

Update on Calibration Studies of the Canadian High-Energy Neutron Spectrometry System (CHENSS) Update on Calibration Studies of the Canadian High-Energy Neutron Spectrometry System (CHENSS) K. Garrow 1, B.J. Lewis 2, L.G.I. Bennett 2, M.B. Smith, 1 H. Ing, 1 R. Nolte, 3 S. Röttger, R 3 R. Smit 4

More information

Radiation Safety Considerations for the TPS Accelerators

Radiation Safety Considerations for the TPS Accelerators Radiation Safety Considerations for the TPS Accelerators R.J. Sheu, J. Liu, and J.P. Wang National Synchrotron Radiation Research Center, 101 Hsin-Ann Road, Hsinchu Science Park, Hsinchu 30076, TAIWAN

More information

Secondary Particles Produced by Hadron Therapy

Secondary Particles Produced by Hadron Therapy Iranian Journal of Medical Physics Vol. 12, No. 2, Spring 2015, 1-8 Received: March 10, 2015; Accepted: July 07, 2015 Original Article Secondary Particles Produced by Hadron Therapy Abdolkazem Ansarinejad

More information

Development And Testing of a Thermoluminescent Dosemeter for Mixed Neutron-Photon-Beta Radiation Fields

Development And Testing of a Thermoluminescent Dosemeter for Mixed Neutron-Photon-Beta Radiation Fields SLAC-PUB-7901 August 1998 Development And Testing of a Thermoluminescent Dosemeter for Mixed Neutron-Photon-Beta Radiation Fields John J. Jummo and James C. Liu Presented at 12th International Conference

More information

Neutron Spectrum Measurement in a Medical Cyclotron

Neutron Spectrum Measurement in a Medical Cyclotron Neutron Spectrum Measurement in a Medical Cyclotron V. Sathian 1, Deepa. S. 2, U.V. Phadnis 1, P.S. Soni 3 and D.N. Sharma 1 1 Radiation Safety Systems Division, 2 Radiological Physics and Advisory Division

More information

Calibration of the GNU and HSREM neutron survey instruments

Calibration of the GNU and HSREM neutron survey instruments Calibration of the GNU and HSREM neutron survey instruments Neutron Users Club Meeting National Physical Laboratory 20 th October 2015 J. S. Eakins 1, L. G. Hager 1, J. W. Leake 2, R. S. Mason 2 and R.

More information

Gamma Radiation Absorption Characteristics of Concrete with Components of Different Type Materials

Gamma Radiation Absorption Characteristics of Concrete with Components of Different Type Materials Vol. 117 (2010) ACTA PHYSICA POLONICA A No. 5 Selected papers presented at the Eleventh Annual Conference of the Materials Research Society of Serbia, YUCOMAT 2009 Gamma Radiation Absorption Characteristics

More information

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah

More information

Measurements of Radiation Doses Induced by High Intensity Laser between and W/cm 2 onto Solid Targets at LCLS MEC Instrument

Measurements of Radiation Doses Induced by High Intensity Laser between and W/cm 2 onto Solid Targets at LCLS MEC Instrument Measurements of Radiation Doses Induced by High Intensity Laser between 10 16 and 10 21 W/cm 2 onto Solid Targets at LCLS MEC Instrument T. Liang 1,2, J. Bauer 1, M. Cimeno 1, A. Ferrari 3, E. Galtier

More information

RESPONSE CALCULATIONS OF THE CdZnTe DETECTOR USING EGS4. James C. Liu, W. R. Nelson and R. Seefred

RESPONSE CALCULATIONS OF THE CdZnTe DETECTOR USING EGS4. James C. Liu, W. R. Nelson and R. Seefred SLAC-PUB-8606 (September 5, 2000) RESPONSE CALCULATIONS OF THE CdZnTe DETECTOR USING EGS4 James C. Liu, W. R. Nelson and R. Seefred Stanford Linear Accelerator Center, MS 48, P. O. Box 4349, Stanford,

More information

Vacuum System of Synchrotron radiation sources

Vacuum System of Synchrotron radiation sources 3 rd ILSF Advanced School on Synchrotron Radiation and Its Applications September 14-16, 2013 Vacuum System of Synchrotron radiation sources Prepared by: Omid Seify, Vacuum group, ILSF project Institute

More information

Radiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp Advance Access publication 11 May 2007

Radiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp Advance Access publication 11 May 2007 Radiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp. 229 233 Advance Access publication 11 May 2007 doi:10.1093/rpd/ncm047 CHARACTERIZATION AND UTILIZATION OF A BONNER SPHERE SET BASED ON GOLD

More information

Advanced Storage Photon Ring Source Upgrade Project:

Advanced Storage Photon Ring Source Upgrade Project: Advanced Storage Photon Ring Source Upgrade Project: The Shielding World s for Leading the Hard X-ray Light Source Advanced Photon Source - Upgrade Bradley J. Micklich Radiation Physicist Argonne National

More information

RADIATION PROTECTION AT SYNCHROTRON RADIATION FACILITIES* James C. Liu 1 and Vaclav Vylet 2. Abstract

RADIATION PROTECTION AT SYNCHROTRON RADIATION FACILITIES* James C. Liu 1 and Vaclav Vylet 2. Abstract SLAC-PUB-9006 September 27, 2001 RADIATION PROTECTION AT SYNCHROTRON RADIATION FACILITIES* James C. Liu 1 and Vaclav Vylet 2 1. Stanford Linear Accelerator Center, MS48, P. O. Box 4349, Stanford, CA 94309,

More information

Characterizations and Diagnostics of Compton Light Source

Characterizations and Diagnostics of Compton Light Source Characterizations and Diagnostics of Compton Light Source Advance Light Source (ALS) (LBNL) Ying K. Wu Duke Free Electron Laser Laboratory (DFELL) Acknowledgments: DFELL: B. Jia, G. Swift, H. Hao, J. Li,

More information

Electron Impact Ionization in EGSnrc

Electron Impact Ionization in EGSnrc Electron Impact Ionization in EGSnrc by Leila Lukhumaidze A thesis submitted to the Faculty of Graduate and Postdoctoral Affairs in partial fulfillment of the requirements for the degree of Master of Science

More information

Ranges of Electrons for Human Body Substances

Ranges of Electrons for Human Body Substances Abstract Research Journal of Chemical Sciences ISSN 2231-606X Ranges of Electrons for Human Body Substances Singh Hemlata 1, Rathi S.K. 1,2 and Verma A.S. 3 1 Department of physics, B. S. A. College, Mathura

More information

Neutron dose assessments for MATROSHKA using the HPA PADC dosemeter

Neutron dose assessments for MATROSHKA using the HPA PADC dosemeter HAMLET FP7 GA 218817 Neutron dose assessments for MATROSHKA using the HPA PADC dosemeter Jon Eakins,Luke Hager and Rick Tanner for the HAMLET consortium Health Protection Agency, Centre for Radiation,

More information

Neutron Fluence and Energy Spectra Around the Varian Clinac 2lOOC/23OOC Medical Accelerator

Neutron Fluence and Energy Spectra Around the Varian Clinac 2lOOC/23OOC Medical Accelerator SLAC-PUB-7 190 June 1996 Neutron Fluence and Energy Spectra Around the Varian Clinac 2lOOC/23OOC Medical Accelerator K. R. Kase, X. S. Mao, W. R. Nelson, J. C. Liu Stanford Linear Accelerator Center, Stanford

More information

Monte Carlo commissioning of photon beams in medical LINACS using wide-field profiles in a water phantom

Monte Carlo commissioning of photon beams in medical LINACS using wide-field profiles in a water phantom Monte Carlo commissioning of photon beams in medical LINACS using wide-field profiles in a water phantom F.Gómez 1, L. Franco 1, A. Iglesias 1, J. Pardo 1, J. Pena 1, A. Rodríguez 1, R.Lobato 2, J. Mosquera

More information

FLUKA calculations for the beam dump system of the LHC : Energy deposition in the dump core and particle spectra in the beam loss monitors

FLUKA calculations for the beam dump system of the LHC : Energy deposition in the dump core and particle spectra in the beam loss monitors EDMS Document Number: 880178 ORGANISATION EUROPENNE POUR LA RECHERCHE NUCLEAIRE EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH Laboratoire Européen pour la Physique des Particules European Laboratory for Particle

More information

Radiation safety of the Danish Center for Proton Therapy (DCPT) Lars Hjorth Præstegaard Dept. of Medical Physics, Aarhus University Hospital

Radiation safety of the Danish Center for Proton Therapy (DCPT) Lars Hjorth Præstegaard Dept. of Medical Physics, Aarhus University Hospital Radiation safety of the Danish Center for Proton Therapy (DCPT) Lars Hjorth Præstegaard Dept. of Medical Physics, Aarhus University Hospital Rationale of proton therapy Dose deposition versus depth in

More information

Geant4 Based Space Radiation Application for Planar and Spherical Geometries

Geant4 Based Space Radiation Application for Planar and Spherical Geometries Advances in Applied Sciences 2017; 2(6): 110-114 http://www.sciencepublishinggroup.com/j/aas doi: 10.11648/j.aas.20170206.13 ISSN: 2575-2065 (Print); ISSN: 2575-1514 (Online) Geant4 Based Space Radiation

More information

Secondary Radiation and Shielding Design for Particle Therapy Facilities

Secondary Radiation and Shielding Design for Particle Therapy Facilities Secondary Radiation and Shielding Design for Particle Therapy Facilities π± A p, n, π± A p, n A Nisy Elizabeth Ipe, Ph.D., C.H.P. Consultant, Shielding Design, Dosimetry & Radiation Protection San Carlos,

More information

Theoretical Assessment of Aircrew Exposure to Galactic Cosmic Radiation Using the FLUKA Monte Carlo Code

Theoretical Assessment of Aircrew Exposure to Galactic Cosmic Radiation Using the FLUKA Monte Carlo Code Theoretical Assessment of Aircrew Exposure to Galactic Cosmic Radiation Using the FLUKA Monte Carlo Code R. Ashkenazi 1, 2, J. Koch 1 and I. Orion 2 1 Radiation Safety Division, Soreq Nuclear Research

More information

Monte Carlo modeling of an electronic brachytherapy source using MCNP5 and EGSnrc

Monte Carlo modeling of an electronic brachytherapy source using MCNP5 and EGSnrc Monte Carlo modeling of an electronic brachytherapy source using MCNP5 and EGSnrc Stephen D. Davis and Larry A. DeWerd University of Wisconsin, Madison, WI 13 th UK Monte Carlo User Group meeting March

More information

Chapter V: Cavity theories

Chapter V: Cavity theories Chapter V: Cavity theories 1 Introduction Goal of radiation dosimetry: measure of the dose absorbed inside a medium (often assimilated to water in calculations) A detector (dosimeter) never measures directly

More information

8 th International Workshop on Radiation Safety at Synchrotron Radiation Sources

8 th International Workshop on Radiation Safety at Synchrotron Radiation Sources 8 th International Workshop on Radiation Safety at Synchrotron Radiation Sources DESY Hamburg, 3 5 June 2015 Proposed material release plan for The decommissioning of the ESRF storage ring Paul Berkvens

More information

Progress in Nuclear Science and Technology, Volume 6,

Progress in Nuclear Science and Technology, Volume 6, DOI: 1.15669/pnst.6 Progress in Nuclear Science and Technology Volume 6 (19) pp. 1-16 ARTICLE A study on calculation method of duct streaming from medical linac rooms Takuma Noto * Kazuaki Kosako and Takashi

More information

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory Recent Activities on Neutron Standardization at the Electrotechnical Laboratory K. Kudo, N. Takeda, S. Koshikawa and A. Uritani Quantum Radiation Division, National Metrology Institute of Japan (NMIJ)

More information

NEUTRON H*(10) INSIDE A PROTON THERAPY FACILITY: COMPARISON BETWEEN MONTE CARLO SIMULATIONS AND WENDI-2 MEASUREMENTS

NEUTRON H*(10) INSIDE A PROTON THERAPY FACILITY: COMPARISON BETWEEN MONTE CARLO SIMULATIONS AND WENDI-2 MEASUREMENTS Radiation Protection Dosimetry (year), Vol. 0, No. 0, pp. 0 0 DOI: 10.1093/rpd/nc0000 NEUTRONS AND IONS IN MEDICINE NEUTRON H*(10) INSIDE A PROTON THERAPY FACILITY: COMPARISON BETWEEN MONTE CARLO SIMULATIONS

More information

Neutron Metrology Activities at CIAE (2009~2010)

Neutron Metrology Activities at CIAE (2009~2010) Neutron Metrology Activities at CIAE (2009~2010) Ionizing Radiation Metrology Division China Institute of Atomic Energy (P.O.Box 275(20), Beijing 102413, China) 1. Neutron calibration fields So far the

More information

(a) Mono-absorber. (b) 4-segmented absorbers. (c) 64-segmented absorbers

(a) Mono-absorber. (b) 4-segmented absorbers. (c) 64-segmented absorbers Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 2001-22, p.37-42 EVALUATION OF ABSORPTION EFFICIENCY FOR NIS TUNNEL JUNCTION DETECTOR WITH SEGMENTED ABSORBERS R. Nouchi, I. Yamada,

More information

Georgia Institute of Technology. Radiation Detection & Protection (Day 3)

Georgia Institute of Technology. Radiation Detection & Protection (Day 3) Georgia Institute of Technology The George W. Woodruff School of Mechanical Engineering Nuclear & Radiological Engineering/Medical Physics Program Ph.D. Qualifier Exam Spring Semester 2009 Your ID Code

More information

Limitations and benchmarks of EGSnrc

Limitations and benchmarks of EGSnrc Limitations and benchmarks of EGSnrc D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers AIFM Workshop, Rome,

More information

Radiation Shielding of a 230 MeV Proton Cyclotron For Cancer Therapy

Radiation Shielding of a 230 MeV Proton Cyclotron For Cancer Therapy Radiation Shielding of a 230 MeV Proton Cyclotron For Cancer Therapy BHASKAR MUKHERJEE Joint DESY and University of Hamburg Accelerator Physics Seminar 27 August 2009 WPE is located within the Campus of

More information

Measurements with the new PHE Neutron Survey Instrument

Measurements with the new PHE Neutron Survey Instrument Measurements with the new PHE Neutron Survey Instrument Neutron Users Club Meeting National Physical Laboratory 16 th October 2013 Jon Eakins, Rick Tanner and Luke Hager Centre for Radiation, Chemicals

More information

a). Air measurements : In-air measurements were carried out at a distance of 82cm from the surface of the light beam

a). Air measurements : In-air measurements were carried out at a distance of 82cm from the surface of the light beam Determination of Backscatter Factors For Diagnostic X-Ray Beams By Experimental And Monte Carlo Methods And Determination Of Air Kerma To Dose Equivalent Conversion Factors For The Calibration Of Personal

More information

Beam diagnostics: Alignment of the beam to prevent for activation. Accelerator physics: using these sensitive particle detectors.

Beam diagnostics: Alignment of the beam to prevent for activation. Accelerator physics: using these sensitive particle detectors. Beam Loss Monitors When energetic beam particles penetrates matter, secondary particles are emitted: this can be e, γ, protons, neutrons, excited nuclei, fragmented nuclei... Spontaneous radiation and

More information

A spectrometer for pulsed and continuous photon radiation

A spectrometer for pulsed and continuous photon radiation Journal of Instrumentation OPEN ACCESS A spectrometer for pulsed and continuous photon radiation To cite this article: R Behrens View the article online for updates and enhancements. Related content -

More information

Commissioning of the Beta Secondary Standard (BSS2)

Commissioning of the Beta Secondary Standard (BSS2) Commissioning of the Beta Secondary Standard (BSS2) Speaker / Author: R.W. Thoka* Co-author: S. Jozela* * National Metrology Institute of South Africa (NMISA), Private Bag X 34, Lynnwood Ridge, Pretoria,

More information

Geant4 simulation of SOI microdosimetry for radiation protection in space and aviation environments

Geant4 simulation of SOI microdosimetry for radiation protection in space and aviation environments Geant4 simulation of SOI microdosimetry for radiation protection in space and aviation environments Dale A. Prokopovich,2, Mark I. Reinhard, Iwan M. Cornelius 3 and Anatoly B. Rosenfeld 2 Australian Nuclear

More information

Optimization studies of photo-neutron production in high-z metallic targets using high energy electron beam for ADS and transmutation

Optimization studies of photo-neutron production in high-z metallic targets using high energy electron beam for ADS and transmutation PRAMANA c Indian Academy of Sciences Vol. 68, No. 2 journal of February 2007 physics pp. 235 241 Optimization studies of photo-neutron production in high-z metallic targets using high energy electron beam

More information