1 Instrumentation for Verification of Dose S. Vatnitsky MedAustron GmbH, Wiener Neustadt, Austria Presented to: Educational Workshop PTCOG 53 Shanghai, China, June 9-11, 2014
2 Consistent and harmonized dosimetry guidelines Accurate beam calibration Goals Ensure exact delivery of prescribed dose Perform planning of high-precision conformal therapy Provide interchange of clinical experience between facilities Dosimetry tasks Acceptance testing & commissioning BDS Reference calibration of clinical beams Acceptance testing & commissioning TPS Periodic QA checks Provide standardization of dosimetry in radiobiology experiments Verification of dose delivery 2
3 Absorbed dose determination in reference conditions for light ion beams Faraday Cup Calorimeter A tool for beam monitor calibration Lack of national and international dosimetry standards Thimble air-filled ionization chamber 3
4 Protocols/COP for proton and heavier ion beam dosimetry Only a Protocol based on standards of absorbed dose to water is being recommended by ICRU/IAEA Reports for protons and heavier ions 4
5 Absorbed dose determination in reference conditions for light ion beams: Step one At the standards laboratory ionization chamber is provided with a calibration factor in terms of the radiation quantity S in a beam of quality Q o : N S,Qo Q o (Co-60) S No national and international dosimetry standards Q o (Co-60) M Qo Q o (proton beam) calorimeter N S,Q M S o Q o reference conditions: N D, w, p D ( ) wp, cal M corr N D,w,Qo P, T, h, S, etc 5
6 Absorbed dose determination in reference conditions for light ion beams: Step two The ionization chamber is then subsequently placed at a reference depth in water in the user s beam, of quality Q. Q D D S w Q M Q N S Q f,,, Q, Q o o M Q f D, S Q, Q o is the instrument reading in the user s beam, suitably corrected to the reference conditions for which N S,Q is valid o is any overall factor necessary to convert both from the calibration quantity S to dose D and from the calibration quality Q o to the user s quality Q 6
7 N D,w - based formalism: IAEA TRS-398/ICRU 78 D w (z ref ) at any user quality Q (photons, electrons, protons, heavier ions) D w,q = M Q N D,w,Q o k Q,Qo corrected instrument reading at Q calibration coefficient at Q o beam quality correction factor Q o (proton beam) calorimeter D M N k w, Q Q D, w, p Q, p 7
8 N D,w - based formalism: IAEA TRS-398/ICRU 78 k Q => Q o = 60 Co s s w, air W p w, air 60 Q air Q Q Co W air Co p => Q o = proton beam Co k Qp, s s w, air w, air Q p 8
9 Stopping powers Proton beams water/air stopping-power ratio Heavier ion beams alpha particles carbon ions protons C (Salamon) Ne (Salamon) Ar (Salamon) He (Salamon) p (ICRU-49) He (ICRU-49) C (H and B) 1.13 Basic proton stopping powers from ICRU residual range (g cm -2 ) A constant value of s w,air = 1.13 adopted in TRS 398 (ignores fragments) Calculation using MC code PETRA following Spencer- Attix cavity theory Transport included secondary electrons and nuclear inelastic process Geitner et al
10 Calorimetry-based determination of W-values D ( Q, cal) cv (1 D ) w T D w ( Q, cal ) M cor N D, w, 60 Co ( S ( S w, air w, air ) ) 60 Q Co ( W ( W air air 60 Co ) ) Q ( p) ( p) Q Co
11 Values of W/e for protons and carbon ions deduced from comparison of IC and calorimeter measurements (w air/e )p / (J C -1 ) Proton beam 34.2 J C -1 Delacroix et al., 1997 Palmans et al., 2004 Palmans et al., 1996 Hashemian et al., 2003 Siebers et al., 1995 Schulz et al., 1992 Brede et al., 2006 Medin et al., J C J C -1 TRS Proton energy / MeV ICRU 78 Sakama et al
12 Transportable graphite calorimeters NPL protons at CCO NIRS - carbon ions Sakama et al 2008 Palmans et al
13 Transportable water calorimeters McGill PTB Protons Calorimetry Gy/MU Ionometry Gy/MU Difference % Calorimetry Gy/MU Ionometry Gy/MU Difference % Scattering 9.087* *10 3 Scanning 1.198* *10 3 Sarfehnia et al., Protons 182 MeV C MeV/u 2.95± ± ± ± 0.08 Brede et al.,
14 k Q s 60 air W 60 p60 sw, air W Q Q pq w, air Co air Co Nylon66-Al Co PMMA-Al &PTW30001 ExrT2 D w,ne2571 / D w,ch C-C &PTW30002 A150-Al &NE2581 IC Chamber # Data from Palmans et al 2001, and Palmans and Verhaegen, Montreal workshop for protons and ions 1 for 60 Co 14
15 Compilation of published data I water I water = 78 ev 400 I water (ev) mean excitation energy of liquid water ICRU 73 (2005) superseded ICRU (2009) tentative ICRU 37 (1984) ICRU 49 (1993) I-water DRF I-water EXP I-water EST Energy deposition (MeV/cm) C 400 MeV/u on water SHIELD-HIT variation with I water is ~1 mm for 2 ev 5 mm value # Depth (cm) 15
16 Influence of a change in the I water and I graphite values on basic dosimetry data and k Q values (results presented by P. Andreo at ESTRO 2013) Decrease of 0.6% in S w,air for Co-60 Decrease of 0.4% in S w,air for protons and heavier ions Net change in W air,protons - increase of 0.6% i.e. W air,protons = ev (current value ev) CONCLUSION: the net effect of all the changes leaves current calculated kq values unaltered 16
17 Recombination corrections for protons and heavier ions Protons and heavier Ion beams: Pulsed (passive) or pulsed scanned (active) beams, no continuous beams! k s a o M a a M M 2 M 2 Not all beams are pulsed for determination of recombination 2 Two-voltage method k s a o M a a M M 2 M 2 2 1/ M 1/ M b / V General recombination 17
18 Recombination corrections for proton beams ICRU 78 Synchrotrons (Repetition < 0.5 Hz, Acceleration 0.5 1s) Cyclotrons (small pulses, high repetition, high dose per pulse) dose per pulse (0.2 Gy) pulse length 400μs maximum transit time for the ionization chamber 152 μs (300 V) and 76 μs (600 V) Lorin et al, 2008 Effective pulse duration is long compared to ion collection time of ion chamber continuous beam k s 2 V N / VL 1 2 V / V M / M N L 18 N L Ion collection time of ion chamber shorter compared to pulse duration Scanned continuous beam The user should verify recombination corrections against independent method
19 Reference calibration: reference conditions Passive Scattering protons, carbon ions Calibration at SOBP ( Pedroni et al) Reference conditions are facility specific Protons spot scanning Carbon ions spot scanning Calibration at SOBP Calibration at plateau Plateau versus SOBP: superposition of beams with different intensities not continuous and reproducible mix of particles with high and low LET fluence corrections are small 19
20 Proton spot scanning calibration PSI (Pedroni et al, 2005) D w = (N/A) (S/) w * x Number of protons/mu IC 10x10x10 MD ACC (Gillin et al, 2010) Dose/MU 20
21 Standard uncertainties in determination of D w (TRS 398, ICRU 78) u(n D,w SSDL ) = 0.6 k Q calc k Q,p Co-60 gamma-rays 0.9 High-energy photons 1.5 High-energy electrons Proton beams Heavier ions
22 Dosimetry in non-reference conditions Relative dose measurements require no detector calibration other than verification of linearity of response within assumed dynamic range of measurement conditions Dosimetry tasks Routine daily clinical physics activity Beam line commissioning Collecting data for TPS Periodic QA Beam characteristics Depth dose Lateral profiles Output factors 22
23 Detectors for measurements in non-reference conditions Active detectors: Ion chambers, diodes, diamond detector, scintillators (single and multiple) Direct display of the current dose rate or the accumulating dose Passive detectors: Destructive TLD Non-destructive Films, alanine Probe accumulates the dose during irradiation. The value of dose is obtained after irradiation with read-out device 23
24 Characterization of small proton beams MeV: no modulation, aperture 30 mm 126 MeV protons, collimator 30 mm, no modulation Markus chamber Diode DEB 50 Diamond detector PinPoint chamber Relative dose Depth in water, mm The user should carefully select detectors depending on beam size 24
25 Characterization of scanned proton pencil beams Depth dose distribution Dose profiles Gillin et al, 2010 Kimstrand et al
26 Multi-detector systems for characterization and QA of proton and carbon beams courtesy by PTW courtesy by IBA Nichiporov et al courtesy by IBA 26
27 2-D dosimetry: fluorescent screen and CCD camera Boon et al 2000 Courtesy CMS Lin et al, 2013 E. Pedroni et al Rosenthal et al
28 2-D dosimetry: TLD Radiochromic film 1.0 MD- 55 film Kodak XV-2 film 0.8 A. Entrance B. SOBP region relative dose Olko et al off-axis data, mm 28
29 Radiochromic films and gels for characterization of clinical proton beams 155 MeV, modulation 3 cm, normalized at 11 cm parallel plate ionization chamber MD-55 film, 70 Gy relative dose Heufelder et al, water equivalent depth, cm 29
30 Additional reading 30
31 Conclusion Implementation of ICRU Report 78 IAEA TRS 398 harmonize clinical dosimetry at proton and heavier ion beam facilities provide a level of accuracy comparable to that in calibration of photon and electron beams 31
32 Acknowledgements Members of ICRU/IAEA Committee Report 78 O. Jäkel and colleagues at HIT T. Lomax and colleagues at PSI Thank you for your attention! 32
Instrumentation for Verification of Dose
Instrumentation for Verification of Dose MedAustron GmbH, Wiener Neustadt, Austria Presented to: Educational Workshop PTCOG 52 Essen, Germany, May 3 5, 2013 Consistent and harmonized dosimetry guidelines
More informationComment on Proton beam monitor chamber calibration
Physics in Medicine & Biology COMMENT on Proton beam monitor chamber calibration To cite this article: Hugo Palmans and Stanislav M Vatnitsky 216 Phys. Med. Biol. 61 6585 View the article online for updates
More informationDosimetry and beam calibration
Dosimetry and beam calibration Hugo Palmans 1,2 and Stanislav Vatnitksy 1 1 EBG MedAustron GmbH, Wiener Neustadt, Austria 2 National Physical Laboratory, Teddington, UK 1 Overview - Learning objectives
More informationNPL s progress towards absorbed dose standards for proton beams
NPL s rogress towards absorbed dose standards for roton beams H. Palmans 1 R. Thomas 1 D. Shiley 1 A. Kacerek 2 1 National Physical Laboratory Teddington United Kingdom 2 Clatterbridge Centre of Oncology
More informationABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *
Romanian Reports in Physics, Vol. 67, No. 3, P. 1152 1158, 2015 ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS * ELENA STANCU 1,2, CATALIN VANCEA
More informationDose detectors, sensors, and their applications
Dose detectors, sensors, and their applications Simona Giordanengo a) Istituto Nazionale di Fisica Nucleare, Section of Torino, Via Giuria 1, 10125 Torino, Italy Hugo Palmans National Physical Laboratory,
More informationDosimetry: Electron Beams
ICTP SChool On MEdical PHysics For RAdiation THerapy: DOsimetry And TReatment PLanning For BAsic And ADvanced APplications 13-24 April 2015 Miramare, Trieste, Italy Dosimetry: Electron Beams G. Hartmann
More informationDevelopment of beam delivery systems for proton (ion) therapy
7th 28th of July 213, JINR Dubna, Russia Development of beam delivery systems for proton (ion) therapy S t u d e n t : J o z e f B o k o r S u p e r v i s o r : D r. A l e x a n d e r M o l o k a n o v
More informationReferensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden
Referensdosimetri Crister Ceberg Medical Radiation Physics Lund University Sweden Reference dosimetry Determination of absorbed dose to water under reference conditions Not accounting for uncertainties
More informationERRATA LIST AND UPDATES TO IAEA TRS-398 (2000) 1
ERRATA LIST AND UPDATES TO IAEA TRS-398 (2000) 1 1. page 034 Table 3 (cont) Chambers SNC 100730 and 100740 are replaced, respectively, by model numbers SNC 100700-0 and 100700-1. Their radii are changed
More informationClinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry
Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry TJ JORDAN Royal Surrey County Hospital IPEM Electron Dosimetry Working Party: + DI Thwaites, AR DuSautoy, MR McEwen, AE
More informationChapter 9: Calibration of Photon and Electron Beams
Chapter 9: Calibration of Photon and Electron Beams Set of 189 slides based on the chapter authored by P. Andreo, J.P. Seuntjens, and E.B. Podgorsak of the IAEA publication (ISBN 92-0-107304-6): Radiation
More information7. a XV-2 high spatial resolution lm detector (Kodak). Important parameters of these detectors are given in Table1. The ionization chambers and the di
Proceedings of the Second International Workshop on EGS, 8.-12. August 2000, Tsukuba, Japan KEK Proceedings 200-20, pp.264-271 Variation of Dose Distribution by Detectors for Narrow Beam T. Fujisaki, H.
More informationOverview and Status of the Austrian Particle Therapy Facility MedAustron. Peter Urschütz
Overview and Status of the Austrian Particle Therapy Facility MedAustron Peter Urschütz MedAustron Centre for ion beam therapy and non-clinical research Treatment of 1200 patients/year in full operation
More informationCharacterization of heavy charged particle fields using fluorescent nuclear track detectors
PTCOG 53, Shanghai June 12, 2014 Characterization of heavy charged particle fields using fluorescent nuclear track detectors Grischa M. Klimpki 1, P. Incardona 2, H. Mescher 1, T. Pfeiler 1, M.S. Akselrod
More informationA Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators
A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators N. Golnik 1, P. Kamiński 1, M. Zielczyński 2 1 Institute of Precision and Biomedical Engineering,
More informationAn introduction to IAEA TRS-483
An introduction to IAEA TRS-483 P Andreo, Professor of Medical Radiation Physics Karolinska University Hospital and Karolinska Institute, Stockholm, Sweden Journées Scientifiques de la SFPM Toulouse 2018
More informationIAEA TRS May 2001 (V.10A) PUBLISHED BY THE IAEA ON BEHALF OF IAEA, WHO, PAHO, AND ESTRO INTERNATIONAL ATOMIC ENERGY AGENCY IAEA
IAEA TRS-398 Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water Pedro Andreo, Dosimetry and Medical
More informationBeam diagnostics: Alignment of the beam to prevent for activation. Accelerator physics: using these sensitive particle detectors.
Beam Loss Monitors When energetic beam particles penetrates matter, secondary particles are emitted: this can be e, γ, protons, neutrons, excited nuclei, fragmented nuclei... Spontaneous radiation and
More informationUpdating reference dosimetry a decade after TG-51
Updating reference dosimetry a decade after TG-51 Malcolm McEwen Ionizing Radiation Standards Institute for National Measurement Standards National Research Council, Canada CE Presentation at AAPM Annual
More informationAcademy of Sciences, Moscow , Russia
Analytical expressions for water-to-air stopping-power ratios relevant for accurate dosimetry in particle therapy arxiv:1010.5356v3 [physics.med-ph] 22 Dec 2010 Abstract Armin Lühr 1,2, David C. Hansen
More informationMonte Carlo Simulation concerning Particle Therapy
Monte Carlo Simulation concerning Particle Therapy Masaaki Takashina Graduate School of Medicine, Osaka University, Osaka 565-0871, Japan INTRODUCTION It is well known that the particle therapy has some
More informationAPPLIED RADIATION PHYSICS
A PRIMER IN APPLIED RADIATION PHYSICS F A SMITH Queen Mary & Westfield College, London fe World Scientific m Singapore * New Jersey London Hong Kong CONTENTS CHAPTER 1 : SOURCES of RADIATION 1.1 Introduction
More informationfactors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons
Calorimetric determination of k Q factors for NE2561 ionization chambers in 3 cm x 3 cm beams of 6 MV and 10 MV photons PTB s water calorimeter in front of Elekta Precise medical linac Calorimetric determination
More information8/3/2016. Chia-Ho, Hua St. Jude Children's Research Hospital. Kevin Teo The Hospital of the University of Pennsylvania
Bijan Arjomandy, Ph.D. Mclaren Proton Therapy Center Mark Pankuch, Ph.D. Cadence Health Proton Center Chia-Ho, Hua St. Jude Children's Research Hospital Kevin Teo The Hospital of the University of Pennsylvania
More informationPhysics of Novel Radiation Modalities Particles and Isotopes. Todd Pawlicki, Ph.D. UC San Diego
Physics of Novel Radiation Modalities Particles and Isotopes Todd Pawlicki, Ph.D. UC San Diego Disclosure I have no conflicts of interest to disclose. Learning Objectives Understand the physics of proton
More informationComparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2
Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST Ronaldo Minniti 1 and Ladislav Czap 2 1 National Institute of Standards and Technology (NIST), Gaithersburg,
More informationVolume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved
Assessment Of The Effectiveness Of Collimation Of Cs 137 Panoramic Beam On Tld Calibration Using A Constructed Lead Block Collimator And An ICRU Slab Phantom At SSDL In Ghana. C.C. Arwui 1, P. Deatanyah
More informationSimulation of light ion transport in a water phantom using Geant4.
Simulation of light ion transport in a water phantom using Geant4. I.Gudowska 1, A.Bagulya 2, V.Ivanchenko 3 and N.Starkov 2 1 Karolinska Institutet and Stockholm University, Stockholm, Sweden 2 Lebedev
More informationTHE mono-energetic hadron beam such as heavy-ions or
Verification of the Dose Distributions with GEANT4 Simulation for Proton Therapy T.Aso, A.Kimura, S.Tanaka, H.Yoshida, N.Kanematsu, T.Sasaki, T.Akagi Abstract The GEANT4 based simulation of an irradiation
More informationOutline. Physics of Charge Particle Motion. Physics of Charge Particle Motion 7/31/2014. Proton Therapy I: Basic Proton Therapy
Outline Proton Therapy I: Basic Proton Therapy Bijan Arjomandy, Ph.D. Narayan Sahoo, Ph.D. Mark Pankuch, Ph.D. Physics of charge particle motion Particle accelerators Proton interaction with matter Delivery
More informationThe heavy ion irradiation facility at KVI-CART
The heavy ion irradiation facility at KVI-CART Brian N. Jones 1, Marc-Jan van Goethem 1,2, Rob Kremers 1, Harry Kiewiet 1, Emiel van der Graaf 1, Sytze Brandenburg 1 1 University of Groningen, KVI-Center
More informationRadiation protection issues in proton therapy
Protons IMRT Tony Lomax, Centre for Proton Radiotherapy, Paul Scherrer Institute, Switzerland Overview of presentation 1. Proton therapy: An overview 2. Radiation protection issues: Staff 3. Radiation
More informationPTRAN. McPTRAN.MEDIA, McPTRAN.CAVITY & McPTRAN.RZ. Hugo Palmans
PTRAN McPTRAN.MEDIA, McPTRAN.CAVITY & McPTRAN.RZ Hugo Palmans Centre for Acoustics & Ionising Radiation, National Physical Laboratory, Teddington, Middlesex, UK Louvain-la-Neuve 1994 Why PTRAN? 1.07 1.06
More informationThe electron accelerator of the ISOF-CNR Institute: its characteristics and use
The electron accelerator of the ISOF-CNR Institute: its characteristics and use P. Fuochi, U. Corda, and M. Lavalle ISOF-CNR Institute, Via P. Gobetti 101, I-40129 Bologna, Italy LINAC Laboratory: ISOF-CNR,
More informationNACP-02 perturbation correction factors for the NPL primary standard of absorbed dose to water in high energy electron beams
NACP-02 perturbation correction factors for the NPL primary standard of absorbed dose to water in high energy electron beams E. Chin 1, J. Seuntjens 1, H. Palmans 2, A. DuSautoy 2, D. Shipley 2, M. Bailey
More informationCode of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams
Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams NEDERLANDSE COMMISSIE VOOR STRALINGSDOSIMETRIE Report 18 of the Netherlands Commission on Radiation Dosimetry
More informationStudy of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 1, WINTER 2008 Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration Pablo
More informationHigh-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers
Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society of Radiobiology
More informationStudy of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water
More informationPrompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research
Prompt gamma measurements for the verification of dose deposition in proton therapy Two Proton Beam Facilities for Therapy and Research Ion Beam Facilities in Korea 1. Proton therapy facility at National
More informationEfficiencies of Some Spherical Ion Chambers in Continuous and Pulsed Radiation: A Numerical Evaluation
Signature: Pol J Radiol, 05; 80: 55-5 DOI: 0.659/PJR.89450 ORIGINAL ARTICLE Received: 05.03.7 Accepted: 05.06.9 Published: 05..5 Authors Contribution: A Study Design B Data Collection C Statistical Analysis
More informationCHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR
ANALELE STIINTIFICE ALE UNIVERSITATII AL. I. CUZA IASI Tomul II, s. Biofizică, Fizică medicală şi Fizica mediului 2006 CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR Dan
More informationFLUKA simulations of selected topics regarding proton pencil beam scanning
FLUKA simulations of selected topics regarding proton pencil beam scanning C. Bäumer, J. Farr, J. Lambert and B. Mukherjee Westdeutsches Protonentherapiezentrum Essen, Germany T. Mertens, and B. Marchand
More informationPHITS calculation of the radiation field in HIMAC BIO
PHITS calculation of the radiation field in HIMAC BIO Ondřej Ploc, Yukio Uchihori, Hisashi Kitamura, Lembit Sihver National Institute of Radiological Sciences, Chiba, Japan Nuclear Physics Institute, Prague,
More informationImplementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy
Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy M. Saiful Huq, PhD, FAAPM, FInstP Dept. of Radiation Oncology, University of
More informationABSORBED DOSE IN ION BEAMS: COMPARISON OF IONISATION- AND FLUENCE-BASED MEASUREMENTS
ABSORBED DOSE IN ION BEAMS: COMPARISON OF IONISATION- AND FLUENCE-BASED MEASUREMENTS Julia-Maria Osinga 1,2,*, Stephan Brons 3, James A. Bartz 4,5, Mark S. Akselrod 5, Oliver Jäkel 1,2,3, Steffen Greilich
More informationPHYSICAL AND DOSIMETRIC BEAM CHARACTERISATION PROTOCOL AT CENTRO NAZIONALE DI ADROTERAPIA ONCOLOGICA (CNAO)
Project co-funded by the European Commission within the FP7 (2007 2013) ULICE Union of Light Centres in Europe Page 1 of 27 PHYSICAL AND DOSIMETRIC BEAM CHARACTERISATION PROTOCOL AT CENTRO NAZIONALE DI
More informationTITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137
TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 AUTHORS: J. Cardoso, L. Santos, C. Oliveira ADRESS: Instituto Tecnológico e Nuclear Estrada Nacional 10; 2686-953 Sacavém;
More informationABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS
ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS C. CAPORALI, AS. GUERRA, R.F. LAITANO, M. PIMPINELLA ENEA-Casaccia, Departimento
More informationGerhard Stucki, Sandor VörösV
idgenössisches Justiz- und Polizeidepartement JPD Bundesamt für Metrologie MTAS xperimental k Q,Q0 lectron Beam Quality Correction Factors for the Types NACP02 and PTW34001 Plane-parallel Chambers Gerhard
More informationMetrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT)
1- Ionizing radiation metrology for radiation protection 2- Metrological requirements for ionizing radiation measurement in radiotherapy and radiodiagnostics R. F. Laitano Part 2 Metrological traceability
More informationPhysics of Particle Beams. Hsiao-Ming Lu, Ph.D., Jay Flanz, Ph.D., Harald Paganetti, Ph.D. Massachusetts General Hospital Harvard Medical School
Physics of Particle Beams Hsiao-Ming Lu, Ph.D., Jay Flanz, Ph.D., Harald Paganetti, Ph.D. Massachusetts General Hospital Harvard Medical School PTCOG 53 Education Session, Shanghai, 2014 Dose External
More informationPRIMARY STANDARDS of AIR KERMA for 60 CO and X-RAYS & ABSORBED DOSE in PHOTON and ELECTRON BEAMS. Malcolm McEwen
PRIMARY STANDARDS of AIR KERMA for 60 CO and X-RAYS & ABSORBED DOSE in PHOTON and ELECTRON BEAMS Malcolm McEwen Definitions: Standard instrument/measurement/artifact intended to define, realize, conserve
More informationBeam Diagnostics and Instrumentation JUAS, Archamps Peter Forck Gesellschaft für Schwerionenforschnung (GSI)
Beam Diagnostics and Instrumentation JUAS, Archamps Peter Forck Gesellschaft für Schwerionenforschnung (GSI), 2003, A dedicated proton accelerator for 1p-physics at the future GSI Demands facilities for
More informationRadiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe
Radiation Dosimetry Alun Beddoe Medical Physics University Hospital Birmingham NHS Trust Contents ABSOLUTE DOSIMETRY (CALIBRATION) Photon interactions (recap) Energy transfer and absorption Electron range
More informationResponse characteristics of an imaging plate to clinical proton beams
Nuclear Instruments and Methods in Physics Research A 424 (1999) 569 574 Response characteristics of an imaging plate to clinical proton beams A. Nohtomi *, T. Terunuma, R. Kohno, Y. Takada, Y. Hayakawa,
More informationDosimetry. Introduction. General. CHAPTER 1
CHAPTER 1 Dosimetry Introduction. a. The detection and repeated measurement of radioactive fallout fields produced by nuclear explosions will give important information affecting the operation of field
More informationChapter V: Cavity theories
Chapter V: Cavity theories 1 Introduction Goal of radiation dosimetry: measure of the dose absorbed inside a medium (often assimilated to water in calculations) A detector (dosimeter) never measures directly
More informationHigh dose-per-pulse electron beam dosimetry A model to correct for the ion recombination in the Advanced Markus ionization chamber
High dose-per-pulse electron beam dosimetry A model to correct for the ion recombination in the Advanced Markus ionization chamber Kristoffer Petersson, a) Maud Jaccard, Jean-Francßois Germond, Thierry
More informationPrinciples of applied dosimetry - illustrated by ionometry. Lesson FYSKJM4710 Eirik Malinen
Principles of applied dosimetry - illustrated by ionometry Lesson FYSKJM4710 Eirik Malinen Ionometry Ionometry: the art of measuring ionizations Number of ionizations proportional to dose Air filled ionization
More informationA Monte Carlo Study of the Relationship between the Time. Structures of Prompt Gammas and in vivo Radiation Dose in.
A Monte Carlo Study of the Relationship between the Time Structures of Prompt Gammas and in vivo Radiation Dose in Proton Therapy Wook-Geun Shin and Chul Hee Min* Department of Radiation Convergence Engineering,
More informationCHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS
CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS Leonardo de Holanda Mencarini 1,2, Claudio A. Federico 1,2 and Linda V. E. Caldas 1 1 Instituto de Pesquisas Energéticas e Nucleares IPEN,
More informationSmall Field Dosimetric Measurements with TLD-100, Alanine, and Ionization Chambers
Small Field Dosimetric Measurements with TLD-1, Alanine, and Ionization Chambers S. Junell a, L. DeWerd a, M. Saiful Huq b, J. Novotny Jr. b, M. uader b, M.F. Desrosiers c, G. Bednarz b a Department of
More informationAir Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira
Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137 J. Cardoso, L. Santos, C. Oliveira SUMMARY: i. The primary standard ii. The irradiation room iii. Experimental results iv. Simulation
More informationEEE4106Z Radiation Interactions & Detection
EEE4106Z Radiation Interactions & Detection 2. Radiation Detection Dr. Steve Peterson 5.14 RW James Department of Physics University of Cape Town steve.peterson@uct.ac.za May 06, 2015 EEE4106Z :: Radiation
More informationUpdate on Calibration Studies of the Canadian High-Energy Neutron Spectrometry System (CHENSS)
Update on Calibration Studies of the Canadian High-Energy Neutron Spectrometry System (CHENSS) K. Garrow 1, B.J. Lewis 2, L.G.I. Bennett 2, M.B. Smith, 1 H. Ing, 1 R. Nolte, 3 S. Röttger, R 3 R. Smit 4
More informationSecondary Radiation and Shielding Design for Particle Therapy Facilities
Secondary Radiation and Shielding Design for Particle Therapy Facilities π± A p, n, π± A p, n A Nisy Elizabeth Ipe, Ph.D., C.H.P. Consultant, Shielding Design, Dosimetry & Radiation Protection San Carlos,
More informationMonte Carlo study of the potential reduction in out-of-field dose using a patient-specific aperture in pencil beam scanning proton therapy
University of Wollongong Research Online Faculty of Engineering - Papers (Archive) Faculty of Engineering and Information Sciences 2012 Monte Carlo study of the potential reduction in out-of-field dose
More informationSmall field dosimetry
Small ield dosimetry Hugo Palmans edaustron Wiener Neustadt Austria and National Physical Laboratory Teddington UK 1 Acknowledgements IAEA/AAP small and composite ield working group (Rodolo Alonso Pedro
More informationSTANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS
STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS M.A. HASSAN*, M.H. GABER**, E. ESMAT*, H.I. FARAG***, H.M. EISSA* *National Institute for Standards (NIS), Giza, Egypt **Biophysics Department,
More information8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required
UPDATING TG-51 When will it end? Malcolm McEwen Ionizing Radiation Standards National Research Council, Canada AAPM Annual Meeting, Charlotte, 2012 What are these updates? Working Group review recommends:
More informationEnergy device for monitoring 4-10 MeV industrial electron accelerators
Published in Nuclear Instruments and Methods in Physics Research A 546, 3,(2005), 385 390 2005 Elsevier B.V. All rights reserved. doi:10.1016/j.nima.2005.02.038 Energy device for monitoring 4-10 MeV industrial
More informationIntroduction. Neutron Effects NSEU. Neutron Testing Basics User Requirements Conclusions
Introduction Neutron Effects Displacement Damage NSEU Total Ionizing Dose Neutron Testing Basics User Requirements Conclusions 1 Neutron Effects: Displacement Damage Neutrons lose their energy in semiconducting
More informationIonizing Radiation Dosimetry and Medical Physics
Ionizing Radiation Dosimetry and Medical Physics D.W.O. Rogers Ionizing Radiation Standards, Institute for National Measurement Standards NRC, Ottawa, K1A OR6 Present e-mail: drogers at physics.carleton.ca
More informationRadiation Dose, Biology & Risk
ENGG 167 MEDICAL IMAGING Lecture 2: Sept. 27 Radiation Dosimetry & Risk References: The Essential Physics of Medical Imaging, Bushberg et al, 2 nd ed. Radiation Detection and Measurement, Knoll, 2 nd Ed.
More informationSpectral Correction Factors for Conventional Neutron Dose Meters Used in High-Energy Neutron Environments
Spectral Correction Factors for Conventional Neutron Dose Meters Used in High-Energy Neutron Environments A complete survey of all neutron spectra in IAEA-TRS-403 RadSynch2015, DESY Hamburg, Germany (June
More informationSTUDY ON THE ENERGY RESPONSE OF PLASTIC SCINTILLATION DETECTOR TO MEV NEUTRONS ABSTRACT
STUDY ON THE ENERGY RESPONSE OF PLASTIC SCINTILLATION DETECTOR TO 0.75-14.75 MEV NEUTRONS Jianfu Zhang 1, 2, Xiaoping Ouyang 1, 2, Suizheng Qiu 1, Xichao Ruan 3, Jinlu Ruan 2 1 School of Nuclear Science
More informationReference Dosimetry for Megavoltage Therapy Beams: Electrons
Reference Dosimetry for Megavoltage Therapy Beams: Electrons David Followill Ph.D Radiological Physics Center UT M.D.Anderson Cancer Center Houston TX Protocol for Clinical Reference Dosimetry of High-Energy
More informationPart E. Radiation monitors. Radiation Safety - JUAS 2014, X. Queralt. Part E. Radiation monitors
Part E. Radiation monitors 1 / 29 Part E. Radiation monitors dose rate ( Sv.h -1 ) ma 0 10 20 30 40 50 10.0 1.0 0.1 250 0 10 20 30 40 50 days since 18/01/02 200 150 100 50 0 Example: photon dose rate measurement
More informationComparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification.
Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification. Antonio Lopez Medina, Antonio Teijeiro, Daniela Medal, Francisco Salvador, Julio Vazquez, Manuel
More informationTowards efficient and accurate particle transport simulation in medical applications
Towards efficient and accurate particle transport simulation in medical applications L. Grzanka1,2, M. Kłodowska1, N. Mojżeszek1, N. Bassler3 1 Cyclotron Centre Bronowice, Institute of Nuclear Physics
More informationOptimization of hadron therapy proton beam using Monte Carlo code on GPU
Dottorato in Fisica degli Acceleratori, XXIX ciclo Optimization of hadron therapy proton beam using Monte Carlo code on GPU Candidata: Martina Senzacqua N matricola: 1163436 Supervisor: Prof. Vincenzo
More informationScreens for low current beams
Screens for low current beams P.Finocchiaro, L.Cosentino, A.Pappalardo INFN - Laboratori Nazionali del Sud Catania, Italy Status Workshop on 29-Apr-2010 on Scintillating - Preliminary Screen Applications
More informationDiamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Diamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams Ramamoorthy Ravichandran,
More informationFOREWORD In 1987 the IAEA published a Code of Practice entitled Absorbed Dose Determination in Photon and Electron Beams: An International Code of
IAEA-TECDOC-1173 Review of data and methods recommended in the international code of practice for dosimetry IAEA Technical Reports Series No. 381, The Use of Plane Parallel Ionization Chambers in High
More informationTERA CONTRIBUTIONS TO PARTNER
TERA CONTRIBUTIONS TO PARTNER Ugo Amaldi University of Milano Bicocca and TERA Foundation 1 CNAO status 2 The CNAO Foundation builds with INFN in Pavia the Centre designed by TERA on the basis of PIMMS.
More informationNeutron Spectroscopy in Proton Therapy
Neutron Spectroscopy in Proton Therapy Khalid Aloufi Thesis submitted for the Degree of Doctor of Philosophy Department of Medical Physics and Biomedical Engineering Faculty of Engineering Sciences University
More informationAccelerators for Hadrontherapy -- Present & Future --
IVICFA Institut Valencià d Investigació Cooperativa en Física Avançada Miniworkshop on Medical Physics Accelerators for Hadrontherapy -- Present & Future -- Silvia Verdú-Andrés TERA / IFIC (CSIC-UV) Valencia,
More informationRecent Activities on Neutron Standardization at the Electrotechnical Laboratory
Recent Activities on Neutron Standardization at the Electrotechnical Laboratory K. Kudo, N. Takeda, S. Koshikawa and A. Uritani Quantum Radiation Division, National Metrology Institute of Japan (NMIJ)
More informationSimulation Modeling in Dosimetry
Simulation Modeling in Dosimetry Aleksei Zhdanov Ural Federal University named after the first President of Russia B. N. Yeltsin, Yekaterinburg, Russian Federation jjj1994@yandex.ru Leonid Dorosinskiy
More informationKey comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation
Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation C. Kessler 1, D.T. Burns 1, S. Vörös 2 and B. Hofstetter-Boillat 2 1
More informationProton and neutron radiation facilities in the PS East hall at CERN
Proton and neutron radiation facilities in the PS East hall at CERN http://www.cern.ch/irradiation M. Glaser, CERN Division EP-TA1-SD Introduction CERN Accelerators CERN-PS East Hall Proton irradiation
More informationMeasurement ofdepth-dose distribution ofprotons by an imaging plate
Nuclear Instruments and Methods in Physics Research A 511 (2003) 382 387 Measurement ofdepth-dose distribution ofprotons by an imaging plate A. Nohtomi a, *, T. Sakae a, T. Terunuma a, Y. Tsunashima a,
More informationCity University of Hong Kong
City University of Hong Kong Information on a Course offered by the Department of Physics and Materials Science with effect from Semester A in 2013 / 2014 Part I Course Title: Radiological Physics and
More informationIon- and proton-beams: Experience with Monte Carlo Simulation
Ion- and proton-beams: Experience with Monte Carlo Simulation Katia Parodi, Ph.D. Heidelberg Ion Therapy Centre, Heidelberg, Germany (Previously: Massachusetts General Hospital, Boston, USA) Workshop on
More informationChapiter VII: Ionization chamber
Chapiter VII: Ionization chamber 1 Types of ionization chambers Sensitive volume: gas (most often air direct measurement of exposure) ionization chamber Sensitive volume: semiconductor (silicon, germanium,
More informationPublished text: Institute of Cancer Research Repository Please direct all s to:
This is an author produced version of an article that appears in: MEDICAL PHYSICS The internet address for this paper is: https://publications.icr.ac.uk/375/ Copyright information: http://www.aip.org/pubservs/web_posting_guidelines.html
More informationThe accelerators at LNS - INFN and diagnostics aspects of the radioactive beams
The accelerators at LNS - INFN and diagnostics aspects of the radioactive beams L. Cosentino, P. Finocchiaro, A. Pappalardo LNS INFN Catania J. Harasimowicz Univ. of Liverpool Cockroft Institute LNS INFN
More informationBeam Optics for a Scanned Proton Beam at Loma Linda University Medical Center
Beam Optics for a Scanned Proton Beam at Loma Linda University Medical Center George Coutrakon, Jeff Hubbard, Peter Koss, Ed Sanders, Mona Panchal Loma Linda University Medical Center 11234 Anderson Street
More information