CIG B5. Capitolato Tecnico Technical requirements

Size: px
Start display at page:

Download "CIG B5. Capitolato Tecnico Technical requirements"

Transcription

1 Gara a procedura aperta per l'affidamento della fornitura di uno strumento per X-ray Photoelectron Spetroscopy (XPS) destinato alla Nanochemistry Facility dell Istituto Italiano di Tecnologia CIG B5 Capitolato Tecnico Technical requirements

2 The Nanochemistry facility at IIT needs an Instrument to perform X-ray photoelectron spectroscopy that aims at advancing the characterization of nanostructures, fabricated by chemical approaches, as building blocks for engineered self assembly architectures across multiple length scales, from the molecular level up to the macroscopic world. The steps of synthesis and functionalization of nanomaterials require chemical characterization of material surfaces that could not be easily obtained from other bulk techniques. The use of the XPS, whose sampling depth is less than 10nm (i.e. surface analysis) could drastically contribute to develop new strategies of synthesis and to gain important insight both into nanomaterials and bulk materials. XPS characterizations could be used to support the research activities of all IIT departments that work on materials for different kind of applications. For these reasons the instrument proposed must be scalable with time, due to the continuous evolution of the research topics of IIT. A high versatility in terms of the possibility to integrate the system with additional components and optional is a fundamental requirement. The XPS instrument must be able to obtain qualitative (low spectral resolution) chemical analysis, elemental chemical composition through acquired high resolution spectra of all materials, from conducting to insulating ones, like as an example: metals, semiconductors, polymers and topographically difficult materials like 3D porous polymeric scaffolds or powders. A high spatial resolution (less than 15µm) spectroscopy is highly desired to have the possibility to gain the most important information from the nanomaterial point of view. A good characterization with a small spot (high counts per seconds specification) will be one of the most important criteria that will be evaluated. The instrument of IIT interest must be able to perform, in UHV environment, surface analysis using, as excitation source, both non monochromatized twin anode (Mg Kα and Al Kα ) and Al-Kα monochromatized X-ray beam, and an hemispherical electron analyzer. Moreover the instrument must be a fully multi-technique tool allowing the installation and use, at their top performances, of UPS source and twin anode. The possibility in future to perform Auger Electron Spectroscopy is desirable including REELS The instrument must be able to perform all the analyses on samples that are thicker than 5 mm whatever their nature is, without any problem of charge compensation. A stage having 5 degree of freedom (i.e. x, y, z, angle with respect to the analyzer and compucentric rotation) in chamber of analysis is required. The ability to image and chemically mapping the area of analysis with a high spatial resolution is one of the most important features due to the latest research developments at IIT concerning synthesis of nanoparticles. The system must be able to perform angle resolved XPS (ARXPS) analyses in automated way by collecting all the information in a fast way and with high energy and spatial resolution. It is tightly requested that the instrument maintains the exact position of the sample at all angles by applying an autofocus and contemporary automatic charge 2/9

3 compensation during the ARXPS analysis. The instrument must be able to perform depth profile with Ar ion source and it must be arranged to be integrated with a polyatomic source if future research lines at IIT requires the use of this option (eg. sputtering of soft homogeneous materials for cleaning and depth profile). During the sputtering with Ar ion source and in the future with polyatomic ion source the equipment must have the ability to maintain a constant charge neutralization and constant binding energy peak position whatever the nature of the resulting interface is. In the future if both polyatomic and Ar ion guns are present on the instrument, the ability of equipment to perform sputtering with the two guns alternatively or simultaneously is required without compromising the possibility to work at a low Ar ion beam voltage (below 200eV). A hot/cold stage (-120 C to 500 C) with a thermocouple monitor on the sample holder is highly desired in the chamber of analysis with the possibility to maintain all 5 degree of freedom over the entire temperature range. A similar option (cool/hot sample holder) is required for the intro-chamber with the same performances in terms of temperature. If required in the future, upgrading the instrument with additional extensions has to be readily carried out, e.g. pre-treatment chamber for heterogeneous catalysis, an UHV chamber for thermal desorption spectroscopy and etc., all of which need to be connected through a radial distribution chamber. 1. DETAILED INSTRUMENT DESCRIPTION The system should meet or exceed the following specifications. All the specifications indicated below must be obtained on the system before shipment. They must also be obtained using X-ray power recommended for routine analysis without scaling of the count rate. Where sensitivity and lateral resolution are specified these must be demonstrated simultaneously with the sample in the same position for each measurement. The system must be assembled and tested before shipment. It must obtain all the values indicated in the following specifications tables if the relevant source is supplied. The calibration of binding energy must be performed by choosing Au, Cu and Ag as reference materials. Silver (Ag3d spectrum) must be used as a material of reference to assess sensitivity of detection. The performances of ion sputtering, in terms of ion beam energy and incidence angle, must be defined by means of the analysis of SiO 2 /Si, or Ta 2 O 5 /Ta. Polyethylene therephtalate and polytetrafluoro ethylene will be considered as standard material to assess the performances of neutralization of insulating polymers and performances of sputtering by means of polyatomic source. The possibility to have reliable results on stoichiometry, a complete library of sensitivity factors, of well know inorganic materials whatever their nature is one of the most important feature that will be evaluated. 1.1 System 3/9

4 The system must be equipped with an intro chamber and a chamber of analysis and some other components connected to it (see below). The chamber of analysis must be of Mumetal to perform, in future, Auger spectroscopy. The Intro chamber, the analysis chamber and all the components should allow UHV conditions in the analysis chamber, the minimum pressure obtainable should be stated below; the chamber should shield the sample from external magnetic fields and allow the installation, later, of additional «optional analysis tools»like as an example the equipment needed for Auger electron spectroscopy (AES) a polyatomic ion gun for polymer depth profiling or a glove box. Particular care must be dedicated to the alignement and dimensions of the ports and to the materials used, so as to reduce evacuation time and pressure recovery. The vacuum in the analysis chamber have to be better than 5E10-10 Torr after baking of the system with no samples inside the chamber of analysis; in the intro chamber the pressure must be less than 10-8 Torr. Fast Entry Air Lock should allow fast sample insertion in the analysis chamber. The pumping system should allow fast pump down time and fast recovery. During ion gun sample cleaning at typical operating pressure the pressure must be not higher than two orders of magnitude with respect the base pressure with samples inside the chamber, while, in case of the presence of an electron flood gun as charge compensation system, the pressure must be not higher than one order of magnitude with respect the base pressure with samples inside the chamber. Vacuum recovery time after ion gun sample cleaning at typical operating pressure and after flood gun use should be indicated. An integral baking system built into the structure of the spectrometer is required. This integral baking must be complete with automatic timer and all the components necessary to minimize cold spots and provide an average baking temperature up to150 C. A fitted flexible baking cover must be included with the equipment. Different programs of baking must be available for the operator to be changed depending on the status of the machine. Concerning the pumping system, the supplier should be able to provide Ion pumps, turbomolecular pumps (rough pumping) with rotary backing pumps on different chambers (i.e. analysis chamber and intro chamber ) and a combination of differential ion pump and Titanium sublimation pump (TSP). A combination of Ion / Pirani gauge for pressure measurement of the analysis chamber is required. Continuous monitoring and logging of vacuum system parameters and an automatic control and interlock of all valves is also required for the introduction of the sample to the analysis chamber. The pumping arrangement for the introduction and analysis chamber must be designed specifically for automated vacuum control and integrated with the software and hardware functions of the system. 4/9

5 1.2 Sample handling Sample holder The sample holder minimum 50 mm in diameter must be able to fit a number of samples no less than 10 of 5mm in diameter). Three sample holders with these characteristics must be purchased to speed up the time of analysis. A sample holder for heating/cooling and one for compucentric rotation must be included in case the same sample holder cannot be used for the three operations. All the accessories to opportunely fix the samples on the sample holder must be included. In case of thick samples, when the instrument is not able to perform the analyses with standard sample holder, a specially designed one for this particular use must be included in the instrument equipment list. Sample holders with a mounted sample must be transferred into and out of the analysis chamber using a rack-and-pinion or magnetically coupled transfer rod. A Hot/Cool (H/C) system is required for the intro chamber. In the chamber of analysis and intro chamber with a nominal temperature range of -120 C to +500 C In case of standard sample holders which are used with the ambient stage cannot be used with the H/C stage, sample holders designed for the H/C stage must be supplied with this option Stage The maximum sample thickness to be inserted in the system must be no less than 10mm. The sample stage in the analysis chamber must be controlled by the computer system managing all other functions of the XPS system. A remote internet control is desirable for the instrument. A laboratory license for data processing software is a valuable option for the instrument. The system must have the ability to define and store different analysis positions and have the ability to return to these positions with high precision for automated analysis. The stage must have X, Y, Z, Rotation and Tilt capability to enable precise positioning of the sample. Moving along Z must enable the analysis of samples up to 10 mm thick. The movement along the axis Tilt must tilt the sample between 0-90 degrees with respect to the analyzer in order to automatically carry out analytical ARXPS mode in case the instrument cannot collect the information of ARXPS without tilting the samples. The stage must have the ability to heat and cool the samples between -120 C and +500 C while maintaining the ability to control all 5 axes of the sample stage. Bakeable motors on all axes are required Specimen viewing X-ray shielded viewports on all the chamber of the system. Microscope specimen alignment facility with high-resolution, color CCD TV video camera and digital zoom facility 5/9

6 Microscope is required for easy sample navigation. LED light source for illumination is preferable. Other forms of imaging such as x-ray induced secondary electron imaging are desirable Analyzer and detector Electron Energy Analyzer should have 180 deflection and a mean radius of at least 150mm or higher. It should be fitted with a multi-channel detector, for high sensitivity XPS spectroscopy, and 2-dimensional XPS imaging. The multi-channel detector must have a large dynamic range of > 1 Mcps. It must be possible for the user to limit the area of analysis to 15 µm or better for small area applications. The system must be optimized for angle resolved XPS studies and solid angle of collection must be +/- 4 degrees. The energy resolution the full width at half maximum (FWHM) will be better than 0,5eV determined on Ag3d 3/2 peak. At 0,6eV resolution the count rate must be higher than 100,000CPS. The instrument should have at least the following performance for large area and small area monochromated XPS: the count rates warranted on Ag3d 3/2 peak working at high energy resolution must be at least higher than 10 6 cts*s -1 for large area (energy resolution better than 1.0eV) and better than 3000 cts*s -1 for 20µm area of analysis (energy resolution better than 0,6eV). The highest performance at the smallest area for spectroscopy for nanotechnology applications is the highest desired capability for the new instrument. The performance specifications at additional smaller areas is requested Multipoint spectroscopy must be performed with a spatial resolution of 10µm or better. A snapshot mode must be available with a high number of channels. The snapshot spectra must be effectively used to identify chemical shifts and fingerprint different chemical species for post analysis processing of maps and sputter depth profiles based on snapshot spectra. Magnetic and/or multi-element electrostatic input lenses, computer programmed are requested to generate a good XPS imaging. It must be possible to collect XPS chemical maps rapidly. For reasons of speed and avoiding sample damage parallel acquisition of images must be possible. The ultimate resolution within the image of <3 µm is desirable. The supplier should be able to demonstrate that images and spectroscopy data are collected within the same point of the sample up to the ultimate spatial resolution 1.3 Sources and Charge neutralizer system Achromatic and X-ray monochromatized source The X-ray Photoelectron system must be equipped with a monochromatized AlKα source which allows spectral information from areas less than 20µm dia. to 300µm dia. The 6/9

7 minimum x-ray spot size on the sample shall be specified. The cooling system (chiller) for this source must be provided with the XPS system. The maximum anode power should be specified together with anode lifetime. A multi-position x-ray anode is desirable. Twin anode source, if not bought initially must be straight forward upgrade. Mg/Al anode achromatic source mounted on linear retractor with a mechanism for optimal positioning of each source face. The dual anode must operate at a nominal operating voltage of 15kV and a maximum power of 400W Ion gun The X-ray Photoelectron system must be equipped with an argon gun to clean the samples, to perform depth profiles and to be used for charge neutralization for insulating samples, if the system provides for this. This gun must be able to operate at 100V - 4kV. To keep a high current at low energy, the gun must be designed with a floating ion column. The gun must be differentially pumped and the gas introduction must be easily controlled to maintain constant the ion density during long sputtering. The maximum current of sputtering must be at least 5µA at 5KV. All electronic parameters for the ion gun must be under full digital control of the data system. The spot size must be better than 200µm and the possibility to raster this beam is necessary. If the system permit to use the ion gun as an analytical tool this must be stated (ISS, scattered electrons imaging) Polyatomic ion gun The system must be designed to allow a future addition of a polyatomic source that could be installed in the future. This addition must be realized without compromising the correct function of the instrument UPS source A windowless discharge source of fully bakeable metal and ceramic construction, which is readily optimized for maximum He (I) or He (II) (or Ne (l)/ne (ll)) output. The lamp must be differentially pumped. The pumping system should also be supplied and integrated with the one of the main system. Sensitivity higher than 1,000,000 cps at 120 mev resolution is required Charge compensation: The equipment must be fitted with a charge compensation system to minimize the the effect of induced charge during the analysis of insulating materials. Flood gun should allow efficient charge compensation on insulating samples even if patterned, porous and 7/9

8 micro and nano-structured. Electron energy must be very low. The neutralization technology must be specified. The resolution in terms of FWHM of the ester component ( C-C(=O)-O-) at ca.289 ev of the C1s peak of PET, must be less than 0.85 ev with cps higher than 2000 at 100um of analyzed area. The flood gun must be controlled by computer. The charge neutralization system shall be capable of self regulating and must work automatically for most types of samples without any intervention during multi-sample analysis. In case of very complicated samples (i.e. porous scaffolds) an alternative adjustment of the charge compensation (i.e. different programs of work) could be required. 1.4 Software and Data Processing All the devices controlled by data system should be listed. It should allow an easy configuration of all experiment parameters. Data system software should allow control, acquisition, processing and printout for provided techniques (XPS, SAXPS, ISS, UPS, AES). Data system should also perform line-scans, imaging, multi-point analysis, multispecimen analysis and depth profiling. Data processing with all the most frequent tools like qualitative and quantitative analysis, best fitting procedures, normalization, background subtracting etc... is required. A windows cascade showing all the spectra during the acquisition is required. The computer operating system must be indicated together with the computer specifications. The interface between the computer and the instrument must be versatile. Remote internet control of the system is desired. If any third party software is required that should be stated. Up to 5 licenses or a laboratory license is required. 1.5 Installation Installation must be performed by the vendors engineers or approved agent of the company that will provide the instrument. Installation will follow successful completion of the site survey confirming that the requirements of the Site Preparation Guide are met. Installation will include the following: -Assembly of the instrument to the layout detailed in the current specification documents -System start-up and basic functionality testing -Operation and performance testing -Performance and documentation of agreed specifications -Three/four days of hands-on analytical training by a laboratory scientist of the company at the user s site three to six months after system installation. An advanced training must be scheduled within one year of installation. 8/9

A DIVISION OF ULVAC-PHI

A DIVISION OF ULVAC-PHI A DIVISION OF ULVAC-PHI X-ray photoelectron spectroscopy (XPS/ESCA) is the most widely used surface analysis technique and has many well established industrial and research applications. XPS provides

More information

A DIVISION OF ULVAC-PHI. Quantera II. Scanning XPS Microprobe

A DIVISION OF ULVAC-PHI. Quantera II. Scanning XPS Microprobe A DIVISION OF ULVAC-PHI Quantera II Scanning XPS Microprobe X-ray Photoelectron Spectroscopy (XPS/ESCA) is the most widely used surface analysis technique and has many well established industrial and

More information

Thermo Scientific K-Alpha + XPS Spectrometer. Fast, powerful and accessible chemical analysis for surface and thin film characterization

Thermo Scientific K-Alpha + XPS Spectrometer. Fast, powerful and accessible chemical analysis for surface and thin film characterization Thermo Scientific K-Alpha + XPS Spectrometer Fast, powerful and accessible chemical analysis for surface and thin film characterization X-ray Photoelectron Spectroscopy Quantitative, chemical identification

More information

Technical description of photoelectron spectrometer Escalab 250Xi

Technical description of photoelectron spectrometer Escalab 250Xi Technical description of photoelectron spectrometer Escalab 250Xi Resource center Physical Methods of Surface Investigations 2014 Table of contents Common description 3 Analytical chamber 8 Preparation

More information

Multi-technique photoelectron spectrometer for micro-area spectroscopy and imaging

Multi-technique photoelectron spectrometer for micro-area spectroscopy and imaging Multi-technique photoelectron spectrometer for micro-area spectroscopy and imaging U. Manju, M. Sreemony and A. K. Chakraborty In this note we present the new multipurpose photoelectron spectroscopy facility

More information

PHI 5000 Versaprobe-II Focus X-ray Photo-electron Spectroscopy

PHI 5000 Versaprobe-II Focus X-ray Photo-electron Spectroscopy PHI 5000 Versaprobe-II Focus X-ray Photo-electron Spectroscopy The very basic theory of XPS XPS theroy Surface Analysis Ultra High Vacuum (UHV) XPS Theory XPS = X-ray Photo-electron Spectroscopy X-ray

More information

A DIVISION OF ULVAC-PHI. Time-of-Flight Secondary Ion Mass Spectrometer with Parallel Imaging MS/MS for Confident Molecular Identification

A DIVISION OF ULVAC-PHI. Time-of-Flight Secondary Ion Mass Spectrometer with Parallel Imaging MS/MS for Confident Molecular Identification A DIVISION OF ULVAC-PHI Time-of-Flight Secondary Ion Mass Spectrometer with Parallel Imaging MS/MS for Confident Molecular Identification Designed for Confident Molecular Identification and Superior Imaging

More information

Auger Electron Spectroscopy Overview

Auger Electron Spectroscopy Overview Auger Electron Spectroscopy Overview Also known as: AES, Auger, SAM 1 Auger Electron Spectroscopy E KLL = E K - E L - E L AES Spectra of Cu EdN(E)/dE Auger Electron E N(E) x 5 E KLL Cu MNN Cu LMM E f E

More information

PHI. Scanning XPS Microprobe

PHI. Scanning XPS Microprobe PHI Scanning XPS Microprobe Unique Scanning XPS Microprobe X-ray photoelectron spectroscopy (XPS/ESA) is the most widely used surface analysis technique and has many well established industrial and research

More information

Photoemission Spectroscopy

Photoemission Spectroscopy FY13 Experimental Physics - Auger Electron Spectroscopy Photoemission Spectroscopy Supervisor: Per Morgen SDU, Institute of Physics Campusvej 55 DK - 5250 Odense S Ulrik Robenhagen,

More information

ToF-SIMS or XPS? Xinqi Chen Keck-II

ToF-SIMS or XPS? Xinqi Chen Keck-II ToF-SIMS or XPS? Xinqi Chen Keck-II 1 Time of Flight Secondary Ion Mass Spectrometry (ToF-SIMS) Not ToF MS (laser, solution) X-ray Photoelectron Spectroscopy (XPS) 2 3 Modes of SIMS 4 Secondary Ion Sputtering

More information

The design of an integrated XPS/Raman spectroscopy instrument for co-incident analysis

The design of an integrated XPS/Raman spectroscopy instrument for co-incident analysis The design of an integrated XPS/Raman spectroscopy instrument for co-incident analysis Tim Nunney The world leader in serving science 2 XPS Surface Analysis XPS +... UV Photoelectron Spectroscopy UPS He(I)

More information

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu X-ray Photoelectron Spectroscopy Introduction Qualitative analysis Quantitative analysis Charging compensation Small area analysis and XPS imaging

More information

Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer 1

Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer 1 Designation: E 2108 00 Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer 1 This standard is issued under the fixed designation E 2108; the number

More information

Secondary Ion Mass Spectrometry (SIMS)

Secondary Ion Mass Spectrometry (SIMS) CHEM53200: Lecture 10 Secondary Ion Mass Spectrometry (SIMS) Major reference: Surface Analysis Edited by J. C. Vickerman (1997). 1 Primary particles may be: Secondary particles can be e s, neutral species

More information

Hiden SIMS Secondary Ion Mass Spectrometers. Analysers for surface, elemental and molecular analysis

Hiden SIMS Secondary Ion Mass Spectrometers. Analysers for surface, elemental and molecular analysis Hiden SIMS Secondary Ion Mass Spectrometers Analysers for surface, elemental and molecular analysis vacuum analysis surface science plasma diagnostics gas analysis SIMS Versatility SIMS is a high sensitivity

More information

Advanced Lab Course. X-Ray Photoelectron Spectroscopy 1 INTRODUCTION 1 2 BASICS 1 3 EXPERIMENT Qualitative analysis Chemical Shifts 7

Advanced Lab Course. X-Ray Photoelectron Spectroscopy 1 INTRODUCTION 1 2 BASICS 1 3 EXPERIMENT Qualitative analysis Chemical Shifts 7 Advanced Lab Course X-Ray Photoelectron Spectroscopy M210 As of: 2015-04-01 Aim: Chemical analysis of surfaces. Content 1 INTRODUCTION 1 2 BASICS 1 3 EXPERIMENT 3 3.1 Qualitative analysis 6 3.2 Chemical

More information

Assessment of the Azimuthal Homogeneity of the Neutral Gas in a Hall Effect Thruster using Electron Beam Fluorescence

Assessment of the Azimuthal Homogeneity of the Neutral Gas in a Hall Effect Thruster using Electron Beam Fluorescence Assessment of the Azimuthal Homogeneity of the Neutral Gas in a Hall Effect Thruster using Electron Beam Fluorescence IEPC-2015-91059 / ISTS-2015-b-91059 Presented at Joint Conference of 30th International

More information

Experimental 2.1 Introduction. Ultra high vacuum.

Experimental 2.1 Introduction. Ultra high vacuum. 2 Experimental 2.1 Introduction. Ultra high vacuum. The interest in solid-gas (or solid-vacuum) interfaces has been increasing rapidly during the last three decades. On one hand this is due to the practical

More information

Photoelectron spectroscopy Instrumentation. Nanomaterials characterization 2

Photoelectron spectroscopy Instrumentation. Nanomaterials characterization 2 Photoelectron spectroscopy Instrumentation Nanomaterials characterization 2 RNDr. Věra V Vodičkov ková,, PhD. Photoelectron Spectroscopy general scheme Impact of X-ray emitted from source to the sample

More information

SNMS. SNMS Applications. Combined SIMS and SNMS

SNMS. SNMS Applications. Combined SIMS and SNMS Hiden SIMS SNMS Sputtered Neutral Mass Spectrometry is a quantitative technique using essentially the same instrumentation as SIMS. However, instead of detecting the secondary ions which are formed at

More information

Mass Spectrometers. for Surface Analysis SURFACE ANALYSIS

Mass Spectrometers. for Surface Analysis SURFACE ANALYSIS Mass Spectrometers for Surface Analysis SURFACE ANALYSIS MASS SPECTROMETERS for Surface Analysis Hiden Analytical have been designing and developing the highest quality quadrupole mass spectrometer based

More information

Introduction to X-ray Photoelectron Spectroscopy (XPS) XPS which makes use of the photoelectric effect, was developed in the mid-1960

Introduction to X-ray Photoelectron Spectroscopy (XPS) XPS which makes use of the photoelectric effect, was developed in the mid-1960 Introduction to X-ray Photoelectron Spectroscopy (XPS) X-ray Photoelectron Spectroscopy (XPS), also known as Electron Spectroscopy for Chemical Analysis (ESCA) is a widely used technique to investigate

More information

X-ray Photoelectron Spectroscopy/ Electron spectroscopy for chemical analysis (ESCA), By Francis Chindeka

X-ray Photoelectron Spectroscopy/ Electron spectroscopy for chemical analysis (ESCA), By Francis Chindeka X-ray Photoelectron Spectroscopy/ Electron spectroscopy for chemical analysis (ESCA), By Francis Chindeka X-ray photoelectron spectroscopy (XPS) or Electron spectroscopy for chemical analysis (ESCA), Surface

More information

Auger Electron Spectroscopy

Auger Electron Spectroscopy Auger Electron Spectroscopy Auger Electron Spectroscopy is an analytical technique that provides compositional information on the top few monolayers of material. Detect all elements above He Detection

More information

Chemical Analysis in TEM: XEDS, EELS and EFTEM. HRTEM PhD course Lecture 5

Chemical Analysis in TEM: XEDS, EELS and EFTEM. HRTEM PhD course Lecture 5 Chemical Analysis in TEM: XEDS, EELS and EFTEM HRTEM PhD course Lecture 5 1 Part IV Subject Chapter Prio x-ray spectrometry 32 1 Spectra and mapping 33 2 Qualitative XEDS 34 1 Quantitative XEDS 35.1-35.4

More information

MB Scientific AB Angular-Resolved Photoemission Spectroscopy System MBS A1SYS_V

MB Scientific AB Angular-Resolved Photoemission Spectroscopy System MBS A1SYS_V Products catalogue MB Scientific AB Angular-Resolved Photoemission Spectroscopy System MBS A1SYS_V 1/15 MBS A1SYS_V; Angular-Resolved Photoemission Spectroscopy System MBS A1SYS_V is MB Scientific AB s

More information

Methods of surface analysis

Methods of surface analysis Methods of surface analysis Nanomaterials characterisation I RNDr. Věra Vodičková, PhD. Surface of solid matter: last monoatomic layer + absorbed monolayer physical properties are effected (crystal lattice

More information

The Benefit of Wide Energy Range Spectrum Acquisition During Sputter Depth Profile Measurements

The Benefit of Wide Energy Range Spectrum Acquisition During Sputter Depth Profile Measurements The Benefit of Wide Energy Range Spectrum Acquisition During Sputter Depth Profile Measurements Uwe Scheithauer, 82008 Unterhaching, Germany E-Mail: scht.uhg@googlemail.com Internet: orcid.org/0000-0002-4776-0678;

More information

Application of the GD-Profiler 2 to the PV domain

Application of the GD-Profiler 2 to the PV domain Application of the GD-Profiler 2 to the PV domain GD Profiler 2 RF GDOES permits to follow the distribution of the elements as function of depth. This is an ultra fast characterisation technique capable

More information

Ion Pump Applications

Ion Pump Applications Gamma Vacuum Gamma Vacuum designs, manufactures, and services ion pumps, titanium sublimation pumps, and their controls. We have grown to be an industry leading ion pump supplier through superior quality,

More information

Secondaryionmassspectrometry

Secondaryionmassspectrometry Secondaryionmassspectrometry (SIMS) 1 Incident Ion Techniques for Surface Composition Analysis Mass spectrometric technique 1. Ionization -Electron ionization (EI) -Chemical ionization (CI) -Field ionization

More information

UV Source Series. UVS 10/35 gas discharge, UVS 300 duoplasmatron. Key Features

UV Source Series. UVS 10/35 gas discharge, UVS 300 duoplasmatron. Key Features UV Source Series UVS 10/35 gas discharge, UVS 300 duoplasmatron and UVLS microwave source for UPS and arpes Key Features Ultimate intensity and stability UV sources Ease of operation Variable excitation

More information

Reduced preferential sputtering of TiO 2 (and Ta 2 O 5 ) thin films through argon cluster ion bombardment.

Reduced preferential sputtering of TiO 2 (and Ta 2 O 5 ) thin films through argon cluster ion bombardment. NATIOMEM Reduced preferential sputtering of TiO 2 (and Ta 2 O 5 ) thin films through argon cluster ion bombardment. R. Grilli *, P. Mack, M.A. Baker * * University of Surrey, UK ThermoFisher Scientific

More information

X-Ray Photoelectron Spectroscopy (XPS) Auger Electron Spectroscopy (AES)

X-Ray Photoelectron Spectroscopy (XPS) Auger Electron Spectroscopy (AES) X-Ray Photoelectron Spectroscopy (XPS) Auger Electron Spectroscopy (AES) XPS X-ray photoelectron spectroscopy (XPS) is one of the most used techniques to chemically characterize the surface. Also known

More information

Supporting Information s for

Supporting Information s for Supporting Information s for # Self-assembling of DNA-templated Au Nanoparticles into Nanowires and their enhanced SERS and Catalytic Applications Subrata Kundu* and M. Jayachandran Electrochemical Materials

More information

SurPASS. Electrokinetic Analyzer for Solid Samples. ::: Innovation in Materials Science

SurPASS. Electrokinetic Analyzer for Solid Samples. ::: Innovation in Materials Science SurPASS Electrokinetic Analyzer for Solid Samples ::: Innovation in Materials Science SurPASS For Solid Surface Analysis The SurPASS electrokinetic analyzer helps material scientists to improve surface

More information

Developments & Limitations in GSR Analysis

Developments & Limitations in GSR Analysis Developments & Limitations in GSR Analysis ENFSI Working Group Meeting June 2006 Jenny Goulden Oxford Instruments NanoAnalysis Overview Introduction Developments in GSR Software Importance of EDS Hardware

More information

XPS/UPS and EFM. Brent Gila. XPS/UPS Ryan Davies EFM Andy Gerger

XPS/UPS and EFM. Brent Gila. XPS/UPS Ryan Davies EFM Andy Gerger XPS/UPS and EFM Brent Gila XPS/UPS Ryan Davies EFM Andy Gerger XPS/ESCA X-ray photoelectron spectroscopy (XPS) also called Electron Spectroscopy for Chemical Analysis (ESCA) is a chemical surface analysis

More information

Jlab FEL Photoemission DC Guns

Jlab FEL Photoemission DC Guns Jlab FEL Photoemission DC Guns Fay Hannon On behalf of the FEL team FLS 2010, 2 nd March 2 Operational Guns 1. FEL Gun 60m Gun Test Stand (GTS) 2. Backup gun, test stand with beam characterization beamline

More information

Lecture 5. X-ray Photoemission Spectroscopy (XPS)

Lecture 5. X-ray Photoemission Spectroscopy (XPS) Lecture 5 X-ray Photoemission Spectroscopy (XPS) 5. Photoemission Spectroscopy (XPS) 5. Principles 5.2 Interpretation 5.3 Instrumentation 5.4 XPS vs UV Photoelectron Spectroscopy (UPS) 5.5 Auger Electron

More information

Applications of XPS, AES, and TOF-SIMS

Applications of XPS, AES, and TOF-SIMS Applications of XPS, AES, and TOF-SIMS Scott R. Bryan Physical Electronics 1 Materials Characterization Techniques Microscopy Optical Microscope SEM TEM STM SPM AFM Spectroscopy Energy Dispersive X-ray

More information

Birck Nanotechnology Center XPS: X-ray Photoelectron Spectroscopy ESCA: Electron Spectrometer for Chemical Analysis

Birck Nanotechnology Center XPS: X-ray Photoelectron Spectroscopy ESCA: Electron Spectrometer for Chemical Analysis Birck Nanotechnology Center XPS: X-ray Photoelectron Spectroscopy ESCA: Electron Spectrometer for Chemical Analysis Dmitry Zemlyanov Birck Nanotechnology Center, Purdue University Outline Introduction

More information

for XPS surface analysis

for XPS surface analysis Thermo Scientific Avantage XPS Software Powerful instrument operation and data processing for XPS surface analysis Avantage Software Atomic Concentration (%) 100 The premier software for surface analysis

More information

MSE 321 Structural Characterization

MSE 321 Structural Characterization Auger Spectroscopy Auger Electron Spectroscopy (AES) Scanning Auger Microscopy (SAM) Incident Electron Ejected Electron Auger Electron Initial State Intermediate State Final State Physical Electronics

More information

AZtec LayerProbe TM. Layer thickness and composition characterisation in the SEM. ...the ultimate non-destructive, high-resolution solution.

AZtec LayerProbe TM. Layer thickness and composition characterisation in the SEM. ...the ultimate non-destructive, high-resolution solution. AZtec LayerProbe TM EDS Layer thickness and composition characterisation in the SEM...the ultimate non-destructive, high-resolution solution 300 nm 253 nm 198 nm ds AZtec LayerProbe Overview Measure layer

More information

Characterization of Secondary Emission Materials for Micro-Channel Plates. S. Jokela, I. Veryovkin, A. Zinovev

Characterization of Secondary Emission Materials for Micro-Channel Plates. S. Jokela, I. Veryovkin, A. Zinovev Characterization of Secondary Emission Materials for Micro-Channel Plates S. Jokela, I. Veryovkin, A. Zinovev Secondary Electron Yield Testing Technique We have incorporated XPS, UPS, Ar-ion sputtering,

More information

Dual Beam Helios Nanolab 600 and 650

Dual Beam Helios Nanolab 600 and 650 Dual Beam Helios Nanolab 600 and 650 In the Clean Room facilities of the INA LMA, several lithography facilities permit to pattern structures at the micro and nano meter scale and to create devices. In

More information

Catalysis CAPABILITIES

Catalysis   CAPABILITIES Catalysis www.extrel.com CAPABILITIES Contents Extrel instruments have been recognized for their exceptional performance by the world s leading researchers for more than 50 years. Reliability and flexibility

More information

Check the LCLS Project website to verify 2 of 6 that this is the correct version prior to use.

Check the LCLS Project website to verify 2 of 6 that this is the correct version prior to use. 1. Introduction The XTOD Offset Systems are designed to spatially separate the useful FEL radiation from high-energy spontaneous radiation and Bremsstrahlung γ-rays. These unwanted radiations are generated

More information

Lecture 5-8 Instrumentation

Lecture 5-8 Instrumentation Lecture 5-8 Instrumentation Requirements 1. Vacuum Mean Free Path Contamination Sticking probability UHV Materials Strength Stability Permeation Design considerations Pumping speed Virtual leaks Leaking

More information

An Introduction to Auger Electron Spectroscopy

An Introduction to Auger Electron Spectroscopy An Introduction to Auger Electron Spectroscopy Spyros Diplas MENA3100 SINTEF Materials & Chemistry, Department of Materials Physics & Centre of Materials Science and Nanotechnology, Department of Chemistry,

More information

Nova 600 NanoLab Dual beam Focused Ion Beam IITKanpur

Nova 600 NanoLab Dual beam Focused Ion Beam IITKanpur Nova 600 NanoLab Dual beam Focused Ion Beam system @ IITKanpur Dual Beam Nova 600 Nano Lab From FEI company (Dual Beam = SEM + FIB) SEM: The Electron Beam for SEM Field Emission Electron Gun Energy : 500

More information

Information on the test material EDS-TM002 and the BAM software package EDX Spectrometer Test for determination of the spectrometer performance

Information on the test material EDS-TM002 and the BAM software package EDX Spectrometer Test for determination of the spectrometer performance BAM 6.8 8.5.213 Information on the test material EDS-TM2 and the BAM software package EDX Spectrometer Test for determination of the spectrometer performance 1. Introduction Energy dispersive spectrometers

More information

Keywords: Abstract. Uwe Scheithauer. Unterhaching

Keywords: Abstract. Uwe Scheithauer. Unterhaching Sputter Induced Cross Contaminations in Analytical AES and XPS Instrumentation: Utilization of the effect for the In situ Deposition of Ultrathin Functional Layers Uwe Scheithauer Unterhaching Phone: +49

More information

Gaetano L Episcopo. Scanning Electron Microscopy Focus Ion Beam and. Pulsed Plasma Deposition

Gaetano L Episcopo. Scanning Electron Microscopy Focus Ion Beam and. Pulsed Plasma Deposition Gaetano L Episcopo Scanning Electron Microscopy Focus Ion Beam and Pulsed Plasma Deposition Hystorical background Scientific discoveries 1897: J. Thomson discovers the electron. 1924: L. de Broglie propose

More information

Electronics Supplementary Information for. Manab Kundu, Cheuk Chi Albert Ng, Dmitri Y. Petrovykh and Lifeng Liu*

Electronics Supplementary Information for. Manab Kundu, Cheuk Chi Albert Ng, Dmitri Y. Petrovykh and Lifeng Liu* Electronics Supplementary Information for Nickel foam supported mesoporous MnO 2 nanosheet arrays with superior lithium storage performance Manab Kundu, Cheuk Chi Albert Ng, Dmitri Y. Petrovykh and Lifeng

More information

Auger Analyses Using Low Angle Incident Electrons

Auger Analyses Using Low Angle Incident Electrons Auger Analyses Using Low Angle Incident Electrons Kenichi Tsutsumi, Yuji agasawa and Toyohiko Tazawa Electron ptics Division, JEL Ltd. Introduction Auger Electron Spectroscopy (AES) is widely used, as

More information

Surface Science Spectra

Surface Science Spectra Surface Science Spectra WCF Submission 226 Proof - SSS Submission # 14-015 (20141102)V22 Analysis of Silicon Germanium Standards for the Quantification of SiGe Microelectronic Devices using AES SECTION

More information

X-ray fluorescence analysis - calibration of the X-ray energy detector

X-ray fluorescence analysis - calibration of the X-ray energy detector X-ray fluorescence analysis - TEP Related topics Bremsstrahlung, characteristic X-radiation, energy levels, fluorescence radiation, conduction processes in semiconductors, doping of semiconductors, pin-diodes,

More information

AMOR the time-of-flight neutron reflectometer at SINQ/PSI

AMOR the time-of-flight neutron reflectometer at SINQ/PSI PRAMANA c Indian Academy of Sciences Vol. 63, No. 1 journal of July 2004 physics pp. 57 63 AMOR the time-of-flight neutron reflectometer at SINQ/PSI MUKUL GUPTA 1, T GUTBERLET 1, J STAHN 1, P KELLER 1

More information

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy Hiroshi Onodera Application & Research Center, JEOL Ltd. Introduction um, PBB and PBDE) are subject to usage restrictions in Europe.

More information

Large Area TOF-SIMS Imaging of the Antibacterial Distribution in Frozen-Hydrated Contact Lenses

Large Area TOF-SIMS Imaging of the Antibacterial Distribution in Frozen-Hydrated Contact Lenses Large Area TOF-SIMS Imaging of the Antibacterial Distribution in Frozen-Hydrated Contact Lenses Overview: Imaging by time-of-flight secondary ion mass spectrometry (TOF-SIMS) is accomplished in a vacuum

More information

Fig 1: Auger Electron Generation (a) Step 1 and (b) Step 2

Fig 1: Auger Electron Generation (a) Step 1 and (b) Step 2 Auger Electron Spectroscopy (AES) Physics of AES: Auger Electrons were discovered in 1925 but were used in surface analysis technique in 1968. Auger Electron Spectroscopy (AES) is a very effective method

More information

Dr. Tim Nunney Thermo Fisher Scientific, East Grinstead, UK Dr. Nick Bulloss Thermo Fisher Scientific, Madison, WI, USA Dr. Harry Meyer III Oak Ridge

Dr. Tim Nunney Thermo Fisher Scientific, East Grinstead, UK Dr. Nick Bulloss Thermo Fisher Scientific, Madison, WI, USA Dr. Harry Meyer III Oak Ridge Dr. Tim Nunney Thermo Fisher Scientific, East Grinstead, UK Dr. Nick Bulloss Thermo Fisher Scientific, Madison, WI, USA Dr. Harry Meyer III Oak Ridge National Laboratory, TN, USA Introduction New materials

More information

X- ray Photoelectron Spectroscopy and its application in phase- switching device study

X- ray Photoelectron Spectroscopy and its application in phase- switching device study X- ray Photoelectron Spectroscopy and its application in phase- switching device study Xinyuan Wang A53073806 I. Background X- ray photoelectron spectroscopy is of great importance in modern chemical and

More information

Chapter 10. Nanometrology. Oxford University Press All rights reserved.

Chapter 10. Nanometrology. Oxford University Press All rights reserved. Chapter 10 Nanometrology Oxford University Press 2013. All rights reserved. 1 Introduction Nanometrology is the science of measurement at the nanoscale level. Figure illustrates where nanoscale stands

More information

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) The Agilent

More information

MSE 321 Structural Characterization

MSE 321 Structural Characterization Auger Spectroscopy Auger Electron Spectroscopy (AES) Scanning Auger Microscopy (SAM) Incident Electron Ejected Electron Auger Electron Initial State Intermediate State Final State Physical Electronics

More information

electron spectroscopy for chemical analysis under environmental conditions Key Features

electron spectroscopy for chemical analysis under environmental conditions Key Features electron spectroscopy for chemical analysis under environmental conditions Key Features Fast Quality Control High Throughput Analysis Controllable Atmosphere Revolutionary Technology Ergonomic all-in-one

More information

Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer

Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer Temperature Control without Compromise: rapid changes, precise temperatures, powerful complex functionality.

More information

Defining quality standards for the analysis of solid samples

Defining quality standards for the analysis of solid samples Defining quality standards for the analysis of solid samples Thermo Scientific Element GD Plus Glow Discharge Mass Spectrometer Redefine your quality standards for the elemental analysis of solid samples

More information

Practical Surface Analysis

Practical Surface Analysis Practical Surface Analysis SECOND EDITION Volume 1 Auger and X-ray Photoelectron Spectroscopy Edited by D. BRIGGS ICI PLC, Wilton Materials Research Centre, Wilton, Middlesbrough, Cleveland, UK and M.

More information

vacuum analysis plasma diagnostics surface science gas analysis

vacuum analysis plasma diagnostics surface science gas analysis Hiden EQP Systems High Sensitivity Mass and Energy Analysers for Monitoring, Control and Characterisation of Ions, Neutrals and Radicals in Plasma. vacuum analysis surface science gas analysis plasma diagnostics

More information

M2 TP. Low-Energy Electron Diffraction (LEED)

M2 TP. Low-Energy Electron Diffraction (LEED) M2 TP Low-Energy Electron Diffraction (LEED) Guide for report preparation I. Introduction: Elastic scattering or diffraction of electrons is the standard technique in surface science for obtaining structural

More information

CYCLOTRON DEVELOPMENT PROGRAM AT JYVÄSKYLÄ

CYCLOTRON DEVELOPMENT PROGRAM AT JYVÄSKYLÄ CYCLOTRON DEVELOPMENT PROGRAM AT JYVÄSKYLÄ P. Heikkinen, E. Liukkonen, Department of Physics, University of Jyväskylä, Finland Abstract The Jyväskylä K130 cyclotron has been modified to allow also negative

More information

X-Ray Photoelectron Spectroscopy (XPS)

X-Ray Photoelectron Spectroscopy (XPS) X-Ray Photoelectron Spectroscopy (XPS) Louis Scudiero http://www.wsu.edu/~scudiero; 5-2669 Electron Spectroscopy for Chemical Analysis (ESCA) The basic principle of the photoelectric effect was enunciated

More information

Surface and Interface Characterization of Polymer Films

Surface and Interface Characterization of Polymer Films Surface and Interface Characterization of Polymer Films Jeff Shallenberger, Evans Analytical Group 104 Windsor Center Dr., East Windsor NJ Copyright 2013 Evans Analytical Group Outline Introduction to

More information

Strain Measurement Techniques for Composite Coupon Testing

Strain Measurement Techniques for Composite Coupon Testing Strain Measurement Techniques for Composite Coupon Testing Introduction Characterization of the properties of anisotropic and inhomogeneous composite materials for use in demanding structural applications

More information

FRAM V5.2. Plutonium and Uranium Isotopic Analysis Software

FRAM V5.2. Plutonium and Uranium Isotopic Analysis Software V5.2 Plutonium and Uranium Isotopic Analysis Software Advanced Isotopic Ratio Analysis Software for HPGe Gamma-Ray Spectra Analyzes Pu, and a wide variety of heterogeneous samples containing Pu, Am, U,

More information

RamanStation 400: a Versatile Platform for SERS Analysis

RamanStation 400: a Versatile Platform for SERS Analysis FIELD APPLICATION REPORT Raman Spectroscopy Author: Dean H. Brown PerkinElmer, Inc. Shelton, CT USA RamanStation 400 RamanStation 400: a Versatile Platform for SERS Analysis Introduction Surface Enhanced

More information

Study of Distributed Ion-Pumps in CESR 1

Study of Distributed Ion-Pumps in CESR 1 Study of Distributed Ion-Pumps in CESR 1 Yulin Li, Roberto Kersevan, Nariman Mistry Laboratory of Nuclear Studies, Cornell University Ithaca, NY 153-001 Abstract It is desirable to reduce anode voltage

More information

Determination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P )

Determination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P ) Determination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P2541001) Curricular Relevance Area of Expertise: ILIAS Education Level: Physik Topic: Hochschule Subtopic: Moderne

More information

Vacuum Pumps. Two general classes exist: Gas transfer physical removal of matter. Mechanical, diffusion, turbomolecular

Vacuum Pumps. Two general classes exist: Gas transfer physical removal of matter. Mechanical, diffusion, turbomolecular Vacuum Technology Vacuum Pumps Two general classes exist: Gas transfer physical removal of matter Mechanical, diffusion, turbomolecular Adsorption entrapment of matter Cryo, sublimation, ion Mechanical

More information

Introduction to X-ray Photoelectron Spectroscopy (XPS) Introduction to X-ray Photoelectron Spectroscopy (XPS) Comparison of Sensitivities

Introduction to X-ray Photoelectron Spectroscopy (XPS) Introduction to X-ray Photoelectron Spectroscopy (XPS) Comparison of Sensitivities Introduction to X-ray Photoelectron Spectroscopy (XPS) Sources of Information Principles of XPS and Auger How to prepare samples for XPS Instrumentation, X rays, Photoelectron detection Data acquisition

More information

X-Ray Photoelectron Spectroscopy (XPS)-2

X-Ray Photoelectron Spectroscopy (XPS)-2 X-Ray Photoelectron Spectroscopy (XPS)-2 Louis Scudiero http://www.wsu.edu/~scudiero; 5-2669 Fulmer 261A Electron Spectroscopy for Chemical Analysis (ESCA) The 3 step model: 1.Optical excitation 2.Transport

More information

Lecture 23 X-Ray & UV Techniques

Lecture 23 X-Ray & UV Techniques Lecture 23 X-Ray & UV Techniques Schroder: Chapter 11.3 1/50 Announcements Homework 6/6: Will be online on later today. Due Wednesday June 6th at 10:00am. I will return it at the final exam (14 th June).

More information

CASSY Lab. Manual ( )

CASSY Lab. Manual ( ) CASSY Lab Manual (524 202) Moseley's law (K-line x-ray fluorescence) CASSY Lab 271 can also be carried out with Pocket-CASSY Load example Safety notes The X-ray apparatus fulfils all regulations on the

More information

Supporting Information. Re-Investigation of the Alleged Formation of CoSi Nanoparticles on Silica. Van An Du, Silvia Gross and Ulrich Schubert

Supporting Information. Re-Investigation of the Alleged Formation of CoSi Nanoparticles on Silica. Van An Du, Silvia Gross and Ulrich Schubert Supporting Information Re-Investigation of the Alleged Formation of CoSi Nanoparticles on Silica Van An Du, Silvia Gross and Ulrich Schubert Experimental All manipulations were carried out under an atmosphere

More information

Nanoscale IR spectroscopy of organic contaminants

Nanoscale IR spectroscopy of organic contaminants The nanoscale spectroscopy company The world leader in nanoscale IR spectroscopy Nanoscale IR spectroscopy of organic contaminants Application note nanoir uniquely and unambiguously identifies organic

More information

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific LTQ Orbitrap Velos

More information

X-Ray Photoelectron Spectroscopy (XPS)

X-Ray Photoelectron Spectroscopy (XPS) X-Ray Photoelectron Spectroscopy (XPS) Louis Scudiero http://www.wsu.edu/~scudiero; 5-2669 Fulmer 261A Electron Spectroscopy for Chemical Analysis (ESCA) The basic principle of the photoelectric effect

More information

ICP-MS. High Resolution ICP-MS.

ICP-MS. High Resolution ICP-MS. ICP-MS attom ES High Resolution ICP-MS www.nu-ins.com Attom ES Enhanced Sensitivity Enhanced Speed Enhanced Selectivity Enhanced Software The Attom ES from Nu Instruments is a double focussing inductively

More information

REX Evaluation Guide. American Micro Detection Systems Inc March Lane, Suite 200 Stockton, CA 95219

REX Evaluation Guide. American Micro Detection Systems Inc March Lane, Suite 200 Stockton, CA 95219 REX Evaluation Guide American Micro Detection Systems Inc. 2800 March Lane, Suite 200 Stockton, CA 95219 I. INTRODUCTION REX (Real-time Elemental X-ray Fluorescence System) is the only instrument capable

More information

5) Surface photoelectron spectroscopy. For MChem, Spring, Dr. Qiao Chen (room 3R506) University of Sussex.

5) Surface photoelectron spectroscopy. For MChem, Spring, Dr. Qiao Chen (room 3R506) University of Sussex. For MChem, Spring, 2009 5) Surface photoelectron spectroscopy Dr. Qiao Chen (room 3R506) http://www.sussex.ac.uk/users/qc25/ University of Sussex Today s topics 1. Element analysis with XPS Binding energy,

More information

Low Temperature (LT), Ultra High Vacuum (UHV LT) Scanning Probe Microscopy (SPM) Laboratory

Low Temperature (LT), Ultra High Vacuum (UHV LT) Scanning Probe Microscopy (SPM) Laboratory Low Temperature (LT), Ultra High Vacuum (UHV LT) Scanning Probe Microscopy (SPM) Laboratory The laboratory of Low Temperature, Ultra High Vacuum (UHV LT) is specifically designed for surface science microscopy

More information

TECHNIC A L WORK ING GROUP ITWG GUIDELINE ON SECONDARY ION MASS SPECTROMETRY (SIMS)

TECHNIC A L WORK ING GROUP ITWG GUIDELINE ON SECONDARY ION MASS SPECTROMETRY (SIMS) NUCLE A R FORENSIC S INTERN ATION A L TECHNIC A L WORK ING GROUP ITWG GUIDELINE ON SECONDARY ION MASS SPECTROMETRY (SIMS) EXECUTIVE SUMMARY Secondary Ion Mass Spectrometry (SIMS) is used for elemental

More information

Extrel Application Note

Extrel Application Note Extrel Application Note Real-Time Plasma Monitoring and Detection of Trace H 2 O and HF Species in an Argon Based Plasma Jian Wei, 575 Epsilon Drive, Pittsburgh, PA 15238. (Presented at the 191st Electrochemical

More information

Extrel is widely respected for the quality of mass spectrometer systems that are

Extrel is widely respected for the quality of mass spectrometer systems that are Extrel is widely respected for the quality of mass spectrometer systems that are available to the world's top research scientists. In response to increasing requests for complete turn-key systems built

More information