MB Scientific AB Angular-Resolved Photoemission Spectroscopy System MBS A1SYS_V
|
|
- Amie Chase
- 5 years ago
- Views:
Transcription
1 Products catalogue MB Scientific AB Angular-Resolved Photoemission Spectroscopy System MBS A1SYS_V 1/15
2 MBS A1SYS_V; Angular-Resolved Photoemission Spectroscopy System MBS A1SYS_V is MB Scientific AB s complete system for Angular-Resolved Photoemission Spectroscopy at Ultra High Resolution. This system has basic but state of the art components for the UPS, XPS* including ARPES measurements. (*X-ray system is optional) Here is a description of our system; Speciality 1. Smart, easy handling UHV system. 2. Extremely high energy resolution photoelectron analyser combined with ultra narrow energy line multi-gas VUV photon source provides absolute front line science at your own lab. 3. Compactly designed, but with full capability for expansion of the system with various options. Basic content of MBS A1SYS_V system; 1. UHV chambers and Aluminum Frame 2. UHV components including Sample cleaver SKTH Photoelectron energy analyser: MBS A-1 with lens4 4. VUV photon source: MBS L-1 (He gas) MBS T-1 (multi gas including Xe) 5. UHV dual grating Monochromator MBS TM-1 for VUV photon sources 6. UHV sample manipulator with low temperature cryostat 7. Water cooling unit 8. Air cooling unit 9. Baking system Options; 1. Different working energy range of electronics 2. Extra pumping system 3. Additional component i.e. X-ray source, Sputtering Ion source, Electron source, LEED- Auger, evaporator. 4. Additional detection system such as Spin polarimeter, Gated MCP detector and so on. There is a flexible choice of the system configuration. For your specific requirements please contact info@mbscientific.se 2/15
3 MBS A1SYS_V; Angular-Resolved Photoemission Spectroscopy System Detailed contents and technical specifications of the system components; 1. UHV chambers with Aluminum frame 1-1 Sample Chamber piece 1-2 Vacuum manifold set 1-3 Sample preparation chamber set 1-4 Al frame set Image photo* *Actual design might deviate from the picture. 2-1 Sample chamber: a. Mu-metal/ Permalloy UHV chamber with 3mm thickness as body material. (All flanges are SUS316L) b. Guaranteed residual magnetic field at sample point less than 500nT c. Guaranteed sample chamber reachable vacuum level less than 2 x 10-8 Pa d. Multiple ports for viewing and optional equipment e. Baking temperature up to 200 degree Celsius d. Every system which contain the Mu-metal/Permalloy chamber supplied with heat treatment certificate 2-2 Vacuum manifold: a. UHV chamber/connection tube made by SUS316L b. Guaranteed reachable vacuum level less than 2 x 10-8 Pa c. Baking temperature up to 200 degree Celsius 2-3 Sample preparation chamber: a. UHV chamber made by SUS316L b. Multiple ports for the sample treatments such as cleaving, evaporation*, annealing* also LEED-AES* optics with convenient opening/pumping. It has also ports for viewing and optional equipment. c. Guaranteed reachable vacuum level less than 2 x 10-8 Pa d. Baking temperature up to 200 degree Celsius *: optional component 3/15
4 2-4 Aluminum frame: a. Flexible standard extruded aluminum frame b. Stable but right weight with high extendibility c. Easy to mount/dismount with stably fastened feet and wheels for transport A-1_#11 UPS system with Aluminium Frame installed at Konan University in Japan 4/15
5 MBS A1SYS; Angular-Resolved Photoemission Spectroscopy System 2. UHV components 1-1 Turbo molecular pumps pcs 1-2 Ion pump 300l pce 1-3 Scroll dry pumps pcs 1-4 UHV gate valves pcs 1-5 MFC:s pcs 1-6 Ion gauges pcs 1-7 Q-mass pce 1-8 Sample cleaver pce 1-9 Magnet coupled rotary feedthrough with Ti Allen key for fastening sample --- 1pcs 1-10 Sampletransfer feedthroughs with magnet coupling pcs 1-11 Vacuum connection tubes and parts --- q.s. Fig. Sample preparation side photo of the A-1SYS #0001 at Shiga University 5/15
6 MBS A1SYS; Angular-Resolved Photoemission Spectroscopy System 2-1 Above components make to reach/allow following condition: a. Guaranteed reachable vacuum level less than 2 x 10-8 Pa in the sample chamber b. Baking temperature up to 200 degree Celsius c. In case of power cut the sample chamber with photoelectron energy analyzer can be kept in a good vacuum by closing 5 gate valves. Fig. UHV system schematic Fig. A-1SYS_#0001 photographed from the light source side 6/15
7 MBS A1SYS_V; Angular-Resolved Photoemission Spectroscopy System 3. Photoelectron energy analyzer_mbs A-1 analyser 1-1 UHV chambers set 1-2 Electron static lens Lans set 1-3 Hemispherical electron energy analyser set 1-4 2D MCP-CCD detection system set 2-1 UHV chambers: a. A1 analyser dome is made of Mu-metal/Permalloy with 3mm thickness as body material. (All flanges, base plate and lens chamber are SUS316L) b. SUS316L parts are electro polished as standard c. Double Mu-metal/Permalloy shield d. Guaranteed reachable vacuum level less than 2 x 10-8 Pa e. CF100/ICF150 pumping port for hemisphere part f. All parts used inside is nonmagnetic material g. 72 hour/ 300 C vacuum fired insulators f. CF40/ICF70 port for mounting ion gauge 2-2 Electrostatic lens: MBS A-1 Lens4 a. Multiple angular modes: 4, 7 and 36 degree total acceptance standard set b. Applicable kinetic energy range UPS ev, XPS ev c. In vacuum adjustable lens axis d. Very good focusing and flexibility by using 7 elements and electrostatic x-y deflector e. All inner surfaces is coated by proprietary 'micro graphite' f. Wide input acceptance angle (36degree) with 35 mm working distance. g. Double Mu-metal/Permalloy shielded h. Mounting flange size is CF200/ICF253, port length 300 mm. i. Angular Resolution (along the slit) within an angular interval of ±3.5 degree Less than 0.1 degree* (Ang3d5) *in the case of 1mmΦ spot size, if spot size is 2mmΦ resolution be less than 0.3 degree 7/15
8 j. Angular Mapping Mode 2-3 Hemispherical electron energy analyser : a. Double Mu-metal/Permalloy shielded b. Baking temperature up to 150 degree Celsius c. Guaranteed reachable vacuum level less than 2 x 10-8 Pa d. Applicable kinetic energy range UPS ev, XPS ev, HE ev e. 200mm mean radius f. Standard Slit Set 9 pairs of 25 mm long BeCu slits Straight: 0.05, 0.1, 0.1(wide aperture), 0.2, 0.4, 0.8, 1.6, 3.2 Curved: 0.2 mm with 1 mm Aperture for adjustments g. Guaranteed energy resolution less than 1meV h. Minimum pre defined pass energy 1eV i. All inner surfaces is coated by micro graphite j. Guaranteed residual magnetic field less than 10nT 2-4 2D MCP-CCD detection system: a. Chevron mount Φ50mm MCP plus fiber optic phosphor screen. b. 2D detection with high sensitivity CCD camera which has 525x910 pixels c. Real time view can be obtained with separate LCD-TV monitor 2-5 Other: a. Easy to be up graded to attach different optional detector such as Spin detector, Gate detector, or other. b. Interface program LabView.vi Energy Resolution -----Better than 1meV The following data was obtained by Dr. Seigo Souma at Tohoku University, Japan. The system consists of MBS A1 with spin option, MBS T1 multi-gas light source, in this case using Xenon gas excitation. 8/15
9 A-1SYS Outlook photo* *System might deviate from the picture depending on the options. 9/15
10 MBS A1SYS_V; Angular-Resolved Photoemission Spectroscopy System 4. VUV photon source_mbs L-1/T-1 lamp 1-1 MBS L-1 He Discharge head set 1-2 MBS T-1 Multi Gas Discharge head set 1-3 Microwave generator set 1-4 Power supply set 1-5 control unit set 2-1 Discharge head: a. Magnetron Microwave ECR (Electron Cyclotron Resonance) plasma. Can use all rare gases from He up to Xe b. Narrow peak wdith less than 0.15meV FWHM at Xe 8.436eV c. Extremely High Brightness More than photon/sr,s(xe) d. Cooling type: Model T-1 air and water, Model L-1 water e. CF40/ICF70 pumping port(s) 2-2 Microwave generator: a. Air cooling with fan T-1 Lamp Head Energies and approximate strengths of He lines HeI: ev 78% ev 8% ev 2% HeII: ev 10% ev 2% Illumination from Xe gas Microwave Generator 10/15
11 MBS A1SYS_V; Angular-Resolved Photoemission Spectroscopy System 5. UHV Monochromator for VUV photon source_mbs TM UHV Twin Monochromator set 2-1 UHV Twin Monochromator: a. Working wavelength between 0nm and 250nm b. Baking temperature up to 150 degree Celsius c. Blaze Wavelength 1350Å d. CF100/ICF150 pumping port e. Accept double gratings to be mounted f. Two ports for mounting L-1 and T-1 lamp head Monochromatised Xe I ( 8.5eV ) Test set up of the Xe VUV system with TM-1 mono 11/15
12 MBS A1SYS_V; Angular-Resolved Photoemission Spectroscopy System 12/15
13 MBS A1SYS_V; Angular-Resolved Photoemission Spectroscopy System 6. UHV sample manipulator with low temperature cryostat 1-1 UHV sample manipulator with Goniometer piece 2-1 UHV sample manipulator with Goniometer: a. Temperature range: 10K to 300K b. Minimum achievable temperature: refrigerator: 4.0K (As system: 10K guarantee) c. Thermal stability: ±0.2K d. Cryostat 1. Cryo capability: exceed 3/5W@60K at 1st Stage(50/60Hz), 0.1W@4.2K at 2nd Stage. 2. He hose:5m 3. Closed liquid He cycle flow e. Interface 1. Sample holder: Gold coated OFHC copper 2. Thermal shield: Body/Head: gold coated OFHC copper 3. Sample rod, Heater mount: Gold coated OFHC copper 4. Skirt: SUSφCF62/ICF114, 6PIN-2pieces f. Manipulation 1. XYZ: manual, Moving distance: XY-axis±10mm Z-axis more than 150mm 2. Rotation: manual 3. Mounting flange size: CF100/ICF Working distance: 250mm, +50mm, -100mm g. Temperature control 1. Temperature control: PID controlled 2. Thermo Heater: Si sensor 2pieces 3. Heater: 50w cartridge 1piece 4. Connection: GP-IB, RS232C h. Utility 1. Power supply for Cryostat AC100/240V, 1φ, 50/60Hz 2. Power supply for temperature control AC100/240V, 1φ, 50/60Hz,200VA 3. Working environment: less than 28 degree Celsius 4. Required vacuum level: less than Pa 13/15
14 MBS A1SYS_V ; Angular-Resolved Photoemission Spectroscopy System 7. Water cooling unit 1-1 Water cooling unit set 2-1 Water cooling unit: It is for cooling the L-1 VUV lamp and thickness monitor, and also for TMP cooling. 8. Air cooling unit 1-1 Air cooling unit set 2-1 Air cooling unit: a. It is for cooling the T-1 VUV lamp. b. Compressor power consumption maximum 2.2kW c. Maximum pressure up to 0.8MPa, controllable pressure between 0.6 and 0.8MPa d. Output air up to 250 l/min e. Power consumption less than 210W for the air dryer f. Relative humidity of output air 15 % within operational temperature 5 to 40 degree Celsius g. Up to 18 liter air tank h. Noise level less than 45dB 14/15
15 MBS A1SYS_V ; Angular-Resolved Photoemission Spectroscopy System 9. Baking system 1-1 Controller unit with thermocouple sets 1-2 Heating tapes 1-3 Fuse box set Heating tapes Controller units Fuse box 15/15
ARPES (Angular Resolved Photoemission Spectroscopy) with a Gas Discharge Lamp
ARPES (Angular Resolved Photoemission Spectroscopy) with a Gas Discharge Lamp Outline - Experimental Setup - Individual Components - How to conduct an experiment - VUV lamps - Electron analyzer essentials
More informationTechnical description of photoelectron spectrometer Escalab 250Xi
Technical description of photoelectron spectrometer Escalab 250Xi Resource center Physical Methods of Surface Investigations 2014 Table of contents Common description 3 Analytical chamber 8 Preparation
More informationMicroscopy Cryostat System
OF AMERICA, INC. Microscopy Cryostat System RC102-CFM Microscopy Cryostat offers fast cooldown, high efficiency, lowest thermal drift, excellent temperature stability and ultra low vibration Optical cryostat
More informationUV Source Series. UVS 10/35 gas discharge, UVS 300 duoplasmatron. Key Features
UV Source Series UVS 10/35 gas discharge, UVS 300 duoplasmatron and UVLS microwave source for UPS and arpes Key Features Ultimate intensity and stability UV sources Ease of operation Variable excitation
More informationThe Use of Synchrotron Radiation in Modern Research
The Use of Synchrotron Radiation in Modern Research Physics Chemistry Structural Biology Materials Science Geochemical and Environmental Science Atoms, molecules, liquids, solids. Electronic and geometric
More informationPHI 5000 Versaprobe-II Focus X-ray Photo-electron Spectroscopy
PHI 5000 Versaprobe-II Focus X-ray Photo-electron Spectroscopy The very basic theory of XPS XPS theroy Surface Analysis Ultra High Vacuum (UHV) XPS Theory XPS = X-ray Photo-electron Spectroscopy X-ray
More informationX-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu
X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu X-ray Photoelectron Spectroscopy Introduction Qualitative analysis Quantitative analysis Charging compensation Small area analysis and XPS imaging
More informationPhotoemission Spectroscopy
FY13 Experimental Physics - Auger Electron Spectroscopy Photoemission Spectroscopy Supervisor: Per Morgen SDU, Institute of Physics Campusvej 55 DK - 5250 Odense S Ulrik Robenhagen,
More informationExperimental 2.1 Introduction. Ultra high vacuum.
2 Experimental 2.1 Introduction. Ultra high vacuum. The interest in solid-gas (or solid-vacuum) interfaces has been increasing rapidly during the last three decades. On one hand this is due to the practical
More informationLecture 5. X-ray Photoemission Spectroscopy (XPS)
Lecture 5 X-ray Photoemission Spectroscopy (XPS) 5. Photoemission Spectroscopy (XPS) 5. Principles 5.2 Interpretation 5.3 Instrumentation 5.4 XPS vs UV Photoelectron Spectroscopy (UPS) 5.5 Auger Electron
More informationLecture 5-8 Instrumentation
Lecture 5-8 Instrumentation Requirements 1. Vacuum Mean Free Path Contamination Sticking probability UHV Materials Strength Stability Permeation Design considerations Pumping speed Virtual leaks Leaking
More informationAdvanced Lab Course. X-Ray Photoelectron Spectroscopy 1 INTRODUCTION 1 2 BASICS 1 3 EXPERIMENT Qualitative analysis Chemical Shifts 7
Advanced Lab Course X-Ray Photoelectron Spectroscopy M210 As of: 2015-04-01 Aim: Chemical analysis of surfaces. Content 1 INTRODUCTION 1 2 BASICS 1 3 EXPERIMENT 3 3.1 Qualitative analysis 6 3.2 Chemical
More informationAtomic Absorption Spectrometer ZEEnit P series
Atomic Absorption Spectrometer ZEEnit P series Technical Data ZEEnit series Update 07/2014 OBue 1/ 5 ZEEnit P series Variable high-end AA Spectrometer with Deuterium and Zeeman Background Correction with
More information3. Experimental Methods and Instrumentation.
3. Experimental Methods and Instrumentation. 3.1 UHV Setup An ultrahigh vacuum chamber pumped by a turbo molecular pump with a base pressure in the low 10-10 mbar range was used (figure 3.1). This chamber
More informationA photoelectron spectrometer for k-space mapping above the Fermi level
A photoelectron spectrometer for k-space mapping above the Fermi level T. Greber a) Physik Institut der Universität Zürich, Winterthurerstr. 190, CH-8057 Zürich, Switzerland O. Raetzo Institut de Physique,
More informationX-ray Photoelectron Spectroscopy (XPS)
X-ray Photoelectron Spectroscopy (XPS) As part of the course Characterization of Catalysts and Surfaces Prof. Dr. Markus Ammann Paul Scherrer Institut markus.ammann@psi.ch Resource for further reading:
More informationStudying Metal to Insulator Transitions in Solids using Synchrotron Radiation-based Spectroscopies.
PY482 Lecture. February 28 th, 2013 Studying Metal to Insulator Transitions in Solids using Synchrotron Radiation-based Spectroscopies. Kevin E. Smith Department of Physics Department of Chemistry Division
More informationAngle-resolved photoemission spectroscopy (ARPES) Overview-Physics 250, UC Davis Inna Vishik
Angle-resolved photoemission spectroscopy (ARPES) Overview-Physics 250, UC Davis Inna Vishik Outline Review: momentum space and why we want to go there Looking at data: simple metal Formalism: 3 step model
More informationVacuum Pumps. Two general classes exist: Gas transfer physical removal of matter. Mechanical, diffusion, turbomolecular
Vacuum Technology Vacuum Pumps Two general classes exist: Gas transfer physical removal of matter Mechanical, diffusion, turbomolecular Adsorption entrapment of matter Cryo, sublimation, ion Mechanical
More informationnovaa 800 D Atomic Absorption Spectrometer
Technical Data Atomic Absorption Spectrometer Cpt : +27 (0) 21 905 0476 Jhb : +27 (0) 11 794 Dbn : +27 (0) 31 266 2454 1/7 General The is a compact atomic absorption spectrometer with deuterium background
More informationM2 TP. Low-Energy Electron Diffraction (LEED)
M2 TP Low-Energy Electron Diffraction (LEED) Guide for report preparation I. Introduction: Elastic scattering or diffraction of electrons is the standard technique in surface science for obtaining structural
More informationChapter 3. Experimental Techniques. 3.1 Optical bending beam method
Chapter 3 Experimental Techniques Stress measurements using optical bending beam method can be applied for UHV system as well as in air. However, the magnetic properties of ultra thin films are better
More informationMSE 321 Structural Characterization
Auger Spectroscopy Auger Electron Spectroscopy (AES) Scanning Auger Microscopy (SAM) Incident Electron Ejected Electron Auger Electron Initial State Intermediate State Final State Physical Electronics
More information6.5 Optical-Coating-Deposition Technologies
92 Chapter 6 6.5 Optical-Coating-Deposition Technologies The coating process takes place in an evaporation chamber with a fully controlled system for the specified requirements. Typical systems are depicted
More informationThe Standard Auger Electron Spectra in The AIST DIO-DB(111); Ge(111)
Paper The Standard Auger Electron Spectra in The AIST DIO-DB(111); Ge(111) K. Goto, a,* Adel Alkafri, b Y. Ichikawa, b A.Kurokawa, c and Y. Yamauchi a a AIST-Chubu Center, Anagahora 2266, Moriyama-ku,
More informationExtrel is widely respected for the quality of mass spectrometer systems that are
Extrel is widely respected for the quality of mass spectrometer systems that are available to the world's top research scientists. In response to increasing requests for complete turn-key systems built
More informationPhotoelectron spectroscopy Instrumentation. Nanomaterials characterization 2
Photoelectron spectroscopy Instrumentation Nanomaterials characterization 2 RNDr. Věra V Vodičkov ková,, PhD. Photoelectron Spectroscopy general scheme Impact of X-ray emitted from source to the sample
More informationCRYOGENIC CONDUCTION COOLING TEST OF REMOVABLE PANEL MOCK-UP FOR ITER CRYOSTAT THERMAL SHIELD
CRYOGENIC CONDUCTION COOLING TEST OF REMOVABLE PANEL MOCK-UP FOR ITER CRYOSTAT THERMAL SHIELD K. Nam, a D. K. Kang, a W. Chung, a C. H. Noh, a J. Yu, b N. I. Her, b C. Hamlyn-Harris, b Y. Utin, b and K.
More informationVibration-Free Pulse Tube Cryocooler System for Gravitational Wave Detectors I
1 Vibration-Free Pulse Tube Cryocooler System for Gravitational Wave Detectors I - Vibration-Reduction Method and Measurement - T. Tomaru A, T. Suzuki A, T. Haruyama A, T. Shintomi A, N. Sato A, A. Yamamoto
More informationMa5: Auger- and Electron Energy Loss Spectroscopy
Ma5: Auger- and Electron Energy Loss Spectroscopy 1 Introduction Electron spectroscopies, namely Auger electron- and electron energy loss spectroscopy are utilized to determine the KLL spectrum and the
More informationIon Implanter Cyclotron Apparatus System
Ion Implanter Cyclotron Apparatus System A. Latuszyñski, K. Pyszniak, A. DroŸdziel, D. M¹czka Institute of Physics, Maria Curie-Sk³odowska University, Lublin, Poland Abstract In this paper the authors
More informationAtomic Spectra HISTORY AND THEORY
Atomic Spectra HISTORY AND THEORY When atoms of a gas are excited (by high voltage, for instance) they will give off light. Each element (in fact, each isotope) gives off a characteristic atomic spectrum,
More informationParts Manual R410A Models
Parts Manual R410A Models PTAC - PD Series (Electronic Controls) Packaged Terminal Air Conditioners Packaged Terminal Heat Pumps PTAC PARTS E (1-2010) Table of Contents Unit Identification... 2 PTAC System
More informationCIG B5. Capitolato Tecnico Technical requirements
Gara a procedura aperta per l'affidamento della fornitura di uno strumento per X-ray Photoelectron Spetroscopy (XPS) destinato alla Nanochemistry Facility dell Istituto Italiano di Tecnologia CIG 05092316B5
More informationApplication of the GD-Profiler 2 to the PV domain
Application of the GD-Profiler 2 to the PV domain GD Profiler 2 RF GDOES permits to follow the distribution of the elements as function of depth. This is an ultra fast characterisation technique capable
More informationA DIVISION OF ULVAC-PHI
A DIVISION OF ULVAC-PHI X-ray photoelectron spectroscopy (XPS/ESCA) is the most widely used surface analysis technique and has many well established industrial and research applications. XPS provides
More informationXPS/UPS and EFM. Brent Gila. XPS/UPS Ryan Davies EFM Andy Gerger
XPS/UPS and EFM Brent Gila XPS/UPS Ryan Davies EFM Andy Gerger XPS/ESCA X-ray photoelectron spectroscopy (XPS) also called Electron Spectroscopy for Chemical Analysis (ESCA) is a chemical surface analysis
More informationJARA FIT Ferienprakticum Nanoelektronik Experiment: Resonant tunneling in quantum structures
JARA FIT Ferienprakticum Nanoelektronik 2013 Experiment: Resonant tunneling in quantum structures Dr. Mihail Ion Lepsa, Peter Grünberg Institut (PGI 9), Forschungszentrum Jülich GmbH 1. Introduction The
More informationThe ISIS Penning H - SPS and Diagnostic Developments at RAL
The ISIS Penning H - SPS and Diagnostic Developments at RAL D.C. Faircloth 1, A.P. Letchford 1, J. Pozimski 1+2, M.O. Whitehead 1, T. Wood 1, S. Jolly 2, P. Savage 2, M. Haigh 3, J. Morrison 3, I. Yew
More information5) Surface photoelectron spectroscopy. For MChem, Spring, Dr. Qiao Chen (room 3R506) University of Sussex.
For MChem, Spring, 2009 5) Surface photoelectron spectroscopy Dr. Qiao Chen (room 3R506) http://www.sussex.ac.uk/users/qc25/ University of Sussex Today s topics 1. Element analysis with XPS Binding energy,
More informationNova 600 NanoLab Dual beam Focused Ion Beam IITKanpur
Nova 600 NanoLab Dual beam Focused Ion Beam system @ IITKanpur Dual Beam Nova 600 Nano Lab From FEI company (Dual Beam = SEM + FIB) SEM: The Electron Beam for SEM Field Emission Electron Gun Energy : 500
More informationGeneral Physics II Summer Session 2013 Review Ch - 16, 17, 18
95.104 General Physics II Summer Session 2013 Review Ch - 16, 17, 18 A metal ball hangs from the ceiling by an insulating thread. The ball is attracted to a positivecharged rod held near the ball. The
More informationReplication Of MHI Transmutation Experiment By D 2 Gas Permeation Through Pd Complex
Higashiyama, T., et al. Replication Of MHI Transmutation Experiment By D2 Gas Permeation Through Pd Complex. in Tenth International Conference on Cold Fusion. 2003. Cambridge, MA: LENR-CANR.org. This paper
More informationResults on a-c tubes subjected to synchrotron irradiation
Results on a-c tubes subjected to synchrotron irradiation V. Baglin, P. Chiggiato, P. Costa-Pinto, B. Henrist (CERN, Geneva) V. Anashin, D. Dorokhov. A. Semenov, A. Krasnov, D. Shwartz, A. Senchenko (,
More informationVacuum. Kai Schwarzwälder, Institut für Physik Universität Basel October 6 th 2006
Physics,, Technology and Techniques of the Vacuum Kai Schwarzwälder, Institut für Physik Universität Basel October 6 th 2006 Outline Introduction and basics Defintion of Vacuum Vacuum A vacuum is a volume
More information1) Introduction 2) Photo electric effect 3) Dual nature of matter 4) Bohr s atom model 5) LASERS
1) Introduction 2) Photo electric effect 3) Dual nature of matter 4) Bohr s atom model 5) LASERS 1. Introduction Types of electron emission, Dunnington s method, different types of spectra, Fraunhoffer
More informationMass Spectrometers. for Surface Analysis SURFACE ANALYSIS
Mass Spectrometers for Surface Analysis SURFACE ANALYSIS MASS SPECTROMETERS for Surface Analysis Hiden Analytical have been designing and developing the highest quality quadrupole mass spectrometer based
More informationChemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES
Chemistry 524--Final Exam--Keiderling May 4, 2011 3:30 -?? pm -- 4286 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils are permitted. No open books or
More informationPARTS CATALOG MUZ-FE09NA MUZ-FE09NA - 1 MUZ-FE09NAH MUZ-FE12NA MUZ-FE12NA1 MUZ-FE12NAH MUZ-FE18NA OUTDOOR UNIT. No. OBB543 REVISED EDITION-C.
SPLIT-TYPE AIR CONDITIONERS Revision C: H and FE12NAH have been added. Please void OBB543 REVISED EDITION-B. OUTDOOR UNIT PARTS CATALOG No. OBB543 REVISED EDITION-C Models - 1 H FE12NA FE12NA1 FE12NAH
More informationMulti-technique photoelectron spectrometer for micro-area spectroscopy and imaging
Multi-technique photoelectron spectrometer for micro-area spectroscopy and imaging U. Manju, M. Sreemony and A. K. Chakraborty In this note we present the new multipurpose photoelectron spectroscopy facility
More informationobject objective lens eyepiece lens
Advancing Physics G495 June 2015 SET #1 ANSWERS Field and Particle Pictures Seeing with electrons The compound optical microscope Q1. Before attempting this question it may be helpful to review ray diagram
More informationDefining quality standards for the analysis of solid samples
Defining quality standards for the analysis of solid samples Thermo Scientific Element GD Plus Glow Discharge Mass Spectrometer Redefine your quality standards for the elemental analysis of solid samples
More informationARGUS VI. Static Vacuum Mass Spectrometer. Static Vacuum ARGUS VI. Multicollection Low Volume Precision
ARGUS VI Static Vacuum Mass Spectrometer Static Vacuum ARGUS VI Multicollection Low Volume Precision Based on more than 20 years of experience in noble gas mass spectrometry instrumentation, we have developed
More informationPhysics 24, Spring 2007 Lab 2 - Complex Spectra
Physics 24, Spring 2007 Lab 2 - Complex Spectra Theory The optical spectra of isolated atoms consist of discrete, unequally spaced lines. This fact could not be understood on the basis of classical atomic
More informationThe photoelectric effect
The photoelectric effect E K hν-e B E F hν E B A photoemission experiment Lifetime broadening ΔE.Δτ~ħ ΔE~ħ/Δτ + Experimental resolution Hüfner, Photoelectron Spectroscopy (Springer) A photoemission experiment
More informationX-Rays From Laser Plasmas
X-Rays From Laser Plasmas Generation and Applications I. C. E. TURCU CLRC Rutherford Appleton Laboratory, UK and J. B. DANCE JOHN WILEY & SONS Chichester New York Weinheim Brisbane Singapore Toronto Contents
More informationBeamlines on Indian synchrotron radiation source Indus-1
Beamlines on Indian synchrotron radiation source Indus-1 R. V. Nandedkar Centre for Advanced Technology, Indore 452 013, India In India, at the Center for Advanced Technology (CAT), Indore a 450 MeV electron
More informationMethods of surface analysis
Methods of surface analysis Nanomaterials characterisation I RNDr. Věra Vodičková, PhD. Surface of solid matter: last monoatomic layer + absorbed monolayer physical properties are effected (crystal lattice
More informationThermo Scientific K-Alpha + XPS Spectrometer. Fast, powerful and accessible chemical analysis for surface and thin film characterization
Thermo Scientific K-Alpha + XPS Spectrometer Fast, powerful and accessible chemical analysis for surface and thin film characterization X-ray Photoelectron Spectroscopy Quantitative, chemical identification
More informationAtomic and Nuclear Physics
Atomic and Nuclear Physics Nuclear physics Rutherford scattering LD Physics Leaflets P6.5.2.1 Rutherford scattering: measuring the scattering rate as a function of the scattering angle and the atomic number
More informationFlight Demonstration of Electrostatic Thruster Under Micro-Gravity
Flight Demonstration of Electrostatic Thruster Under Micro-Gravity Shin SATORI*, Hiroyuki MAE**, Hiroyuki OKAMOTO**, Ted Mitsuteru SUGIKI**, Yoshinori AOKI # and Atsushi NAGATA # * Hokkaido Institute of
More informationSpin-resolved photoelectron spectroscopy
Spin-resolved photoelectron spectroscopy Application Notes Spin-resolved photoelectron spectroscopy experiments were performed in an experimental station consisting of an analysis and a preparation chamber.
More informationUPGRADE AND COMMISSIONING OF THE PIAVE-ALPI ECR INJECTOR AT LNL
UPGRADE AND COMMISSIONING OF THE PIAVE-ALPI ECR INJECTOR AT LNL A. Galatà, M. Sattin, L. Bertazzo, L. Boscagli, S. Contran, A. Dainese, A. Lombardi, D. Maniero, M. Poggi, F. Scarpa, A. Facco, INFN-LNL,
More informationPrevious Lecture. Electron beam lithoghraphy e - Electrons are generated in vacuum. Electron beams propagate in vacuum
Previous Lecture Electron beam lithoghraphy e - Electrons are generated in vacuum Electron beams propagate in vacuum Lecture 6: Vacuum & plasmas Objectives From this vacuum lecture you will learn: What
More informationPHOTOELECTRON SPECTROSCOPY (PES)
PHOTOELECTRON SPECTROSCOPY (PES) NTRODUCTON Law of Photoelectric effect Albert Einstein, Nobel Prize 1921 Kaiser-Wilhelm-nstitut (now Max-Planck- nstitut) für Physik Berlin, Germany High-resolution electron
More informationKP Technology System Catalogue
Sample: Al Tip: 2mm S/S Sig Ave: 60 WF Ave: 2 T: 18.9 C Gain: 4 KP Technology System Catalogue Winter 2013-2014 Al Relative Humidity Ramp Test (10% steps) @ 2 Hr/step 09-10/Nov/2013 RH (%) 92 90 88 86
More informationMSE 321 Structural Characterization
Auger Spectroscopy Auger Electron Spectroscopy (AES) Scanning Auger Microscopy (SAM) Incident Electron Ejected Electron Auger Electron Initial State Intermediate State Final State Physical Electronics
More informationPractice Final Name. m 3 /s b) 8.75 x 10 4 c) 8.21 x 10 4 d) 7.45 x 10 4 e) 7.21 x 10 4 Ans: a
I included more than 35 problems only for practice purposes. In the final you will have 35 problems, as I stated during the last class meeting on Thursday, December 7, 2006. Practice Final Name 1) In a
More informationA beam of coherent monochromatic light from a distant galaxy is used in an optics experiment on Earth.
Waves_P2 [152 marks] A beam of coherent monochromatic light from a distant galaxy is used in an optics experiment on Earth. The beam is incident normally on a double slit. The distance between the slits
More informationHigh Resolution Photoemission Study of the Spin-Dependent Band Structure of Permalloy and Ni
High Resolution Photoemission Study of the Spin-Dependent Band Structure of Permalloy and Ni K. N. Altmann, D. Y. Petrovykh, and F. J. Himpsel Department of Physics, University of Wisconsin, Madison, 1150
More informationH. R. Hirsch R. J. Hull R. H. Kohler Mrs. Ilana Levitan K. K. Y. Li J. H. Loehlin
IIIL - -~ V. NUCLEAR MAGNETIC RESONANCE AND HYPERFINE STRUCTURE Prof. F. Bitter Prof. L. C. Bradley III Prof. J. S. Waugh Dr. H. H. Stroke Dr. J. F. Waymouth B. B. Aubrey M. Ciftan H. R. Hirsch R. J. Hull
More informationInstructions for the advanced lab course Angle-resolved photoelectron spectroscopy
Instructions for the advanced lab course Angle-resolved photoelectron spectroscopy Contents 1 Introduction 2 2 References 2 3 Preparation 3 3.1 Basics of Condensed Matter Physics............... 3 3.2 Photoelectron
More informationHow to distinguish EUV photons from out-of-band photons
How to distinguish EUV photons from out-of-band photons Thesis submitted in partial fulfillment of the requirements for the degree of Master of Science in Physics Author: Student ID: Supervisors: 2 nd
More informationHELIX SFT. Static Vacuum ARGUS VI HELIX SFT. Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer
ARGUS VI HELIX SFT Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer Static Vacuum HELIX SFT Low Volume Multicollection High Mass Resolution 3 Based on more than 20 years of experience in
More informationX-ray Photoelectron Spectroscopy/ Electron spectroscopy for chemical analysis (ESCA), By Francis Chindeka
X-ray Photoelectron Spectroscopy/ Electron spectroscopy for chemical analysis (ESCA), By Francis Chindeka X-ray photoelectron spectroscopy (XPS) or Electron spectroscopy for chemical analysis (ESCA), Surface
More informationVisualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source
3rd International EUVL Symposium NOVEMBER 1-4, 2004 Miyazaki, Japan Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source H. Tanaka, A. Matsumoto, K. Akinaga, A. Takahashi
More informationHarmonic Generation for Photoionization Experiments Christian J. Kornelis Physics REU Kansas State University
Harmonic Generation for Photoionization Experiments Christian J. Kornelis Physics REU Kansas State University The Basic Setup for the KLS Photoionization Experiment V. Kumarappan Femtosecond Pump-Probe
More informationLIBSlab ANALYZERS ANALYZERS
ANALYZERS ANALYZERS Chemical multi-elemental analysis with LIBS in modular benchtop design LIBSlab LIBSpector compact sample chamber for the LIBS analysis of solid, liquid and gaseous samples. Sample chamber
More informationWhich Optistat is right for you?
OPTISTAT Optistat liquid cryogen optical cryostats for spectroscopy Which Optistat is right for you? Our easy to use, compact and efficient Optistat cryostats offer a wide range of options to suit most
More informationHELIX MC Plus. Static Vacuum. ARGUS VI Thermo Scientific HELIX MC Plus Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer
ARGUS VI Thermo Scientific HELIX MC Plus Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer Static Vacuum HELIX MC Plus Dynamic range Multi collection High mass resolution Based on more than
More informationDirect Neutrino Mass Measurement with KATRIN. Sanshiro Enomoto (University of Washington) for the KATRIN Collaboration
Direct Neutrino Mass Measurement with KATRIN Sanshiro Enomoto (University of Washington) for the KATRIN Collaboration DBD16, Osaka, Japan, 8 Nov 2016 Neutrino Mass Measurement with Single Beta Decay 2
More informationBirck Nanotechnology Center XPS: X-ray Photoelectron Spectroscopy ESCA: Electron Spectrometer for Chemical Analysis
Birck Nanotechnology Center XPS: X-ray Photoelectron Spectroscopy ESCA: Electron Spectrometer for Chemical Analysis Dmitry Zemlyanov Birck Nanotechnology Center, Purdue University Outline Introduction
More informationPHYSICS QUESTION PAPER CLASS-XII
PHYSICS QUESTION PAPER CLASS-XII Time : 3.00 Hours] [Maximum Marks : 100 Instructions : There are four sections and total 60 questions in this question paper. ( 1) (2) All symbols used in this question
More informationBeetle UHV VT AFM / STM
UHV VT AFM / STM RHK Technology Imaging the Future of Nanoscience UHV VT AFM / STM Engineered Excellence Everyday, in university and government labs around the globe, RHK research platforms lead to new
More informationBecause light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.
Light We can use different terms to describe light: Color Wavelength Frequency Light is composed of electromagnetic waves that travel through some medium. The properties of the medium determine how light
More informationDevelopment of a High-Speed VUV Camera System for 2-Dimensional Imaging of Edge Turbulent Structure in the LHD
Development of a High-Speed VUV Camera System for 2-Dimensional Imaging of Edge Turbulent Structure in the LHD Masaki TAKEUCHI, Satoshi OHDACHI and LHD experimental group National Institute for Fusion
More informationPuebla, C.P , Mexico; ABSTRACT 1. INTRODUCTION 2. CRYOSTAT INTEGRATION
Integration and characterization of the cryogenic system of MEGARA Ferrusca R.,D.* a, Cisneros G., M. E. a, Velázquez, M. a, Zenteno H., J. A. a, Gil de Paz, A. b, Gallego, J. b, Carrasco, E. a, Sánchez-Moreno
More informationPlasma Spectroscopy in ISTTOK
Plasma Spectroscopy in ISTTOK J. Figueiredo 1, R. B. Gomes 1, T. Pereira 1, H. Fernandes 1, A. Sharakovski 2 1 Associação EURATOM/IST, Centro de Fusão Nuclear, IST, 1049-001 Lisboa, Portugal 2 Association
More informationEmission Spectrum of Atomic Gases. Prelab Questions
Emission Spectrum of Atomic Gases Prelab Questions Before this coming to this lab, please review your text for the physics of the spectrum of visible light and of diffraction grating spectrometer.. Which
More informationORMS HARS. SpectrA LAB EQUIPMENTS Optical Rotation Resolution nm Wavelength Range 0.1nm Wavelength Resolution -70 C Cooling 10 Vacuum
All Products are manufactured and marketed by Opto-Mechatronics Pvt. Ltd HARS ADVANCED RESEARCH SYSTEMS & SOLUTIONS FARADAY & KERR OPTO-MECHATRONICS PVT.LTD OPTICAL ROTATION MEASUREMENT SYSTEM 0-3 0.005
More informationCover Page. The handle holds various files of this Leiden University dissertation.
Cover Page The handle http://hdl.handle.net/1887/44295 holds various files of this Leiden University dissertation. Author: Badan, C. Title: Surface-structure dependence of water-related adsorbates on platinum
More informationAES - Auger Electron Spectrosopy
Advanced Materials - Lab Intermediate Physics Ulm University Institute of Solid State Physics AES - Auger Electron Spectrosopy Sebastian Schnurr March 13, 2013 Safety Precautions MAKE SURE THAT YOU UNDERSTAND
More informationSpectroscopic studies of impurities in the LHD plasmas
Spectroscopic studies of impurities in the LHD plasmas Visitor: Zhenwei Wu (the institute of plasma physics, CAS -ASIPP) Host: Shigeru Morita (the national institute for fusion science -NIFS) Content 1.
More informationThis experiment is included in the XRP 4.0 X-ray solid state, XRS 4.0 X-ray structural analysis, and XRC 4.0 X-ray characteristics upgrade sets.
The intensity of characteristic X-rays as a TEP Related topics Characteristic X-radiation, energy levels, Bragg s law, and intensity of characteristic X-rays Principle The X-ray spectrum of an X-ray tube
More informationIn the investigation, a pressure on the order of 10-6
Production of a 131m Xe calibration source for Cyclotron Radiation Emission Spectroscopy and use with Project 8 setup Annie Ramey Physics Department and CENPA, University of Washington (Dated: August 23,
More information= 6 (1/ nm) So what is probability of finding electron tunneled into a barrier 3 ev high?
STM STM With a scanning tunneling microscope, images of surfaces with atomic resolution can be readily obtained. An STM uses quantum tunneling of electrons to map the density of electrons on the surface
More informationThermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples
Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer Defining quality standards for the analysis of solid samples Redefine your quality standards for the elemental analysis of solid samples
More informationLecture 3 Vacuum Science and Technology
Lecture 3 Vacuum Science and Technology Chapter 3 - Wolf and Tauber 1/56 Announcements Homework will be online from noon today. This is homework 1 of 4. 25 available marks (distributed as shown). This
More informationIntroduction to Blackbody Sources
Introduction to s This section contains dedicated blackbody sources for low uncertainty calibration of infrared thermometers. A range of portable primary blackbody sources combine high emissivity with
More informationA fast, parallel acquisition, electron energy analyzer: The hyperbolic field analyzer
REVIEW OF SCIENTIFIC INSTRUMENTS VOLUME 70, NUMBER 5 MAY 1999 A fast, parallel acquisition, electron energy analyzer: The hyperbolic field analyzer M. Jacka a) and M. Kirk Department of Physics, University
More information