VIII. QUADRUPOLE ION TRAPS: AN ION TRAPPING INST. A. BASIC DESIGN (REF: JMS, 1997, 32, 351)

Size: px
Start display at page:

Download "VIII. QUADRUPOLE ION TRAPS: AN ION TRAPPING INST. A. BASIC DESIGN (REF: JMS, 1997, 32, 351)"

Transcription

1 VIII. QUADRUPOLE ION TRAPS: AN ION TRAPPING INST. A. BASIC DESIGN (REF: JMS, 1997, 32, 351) The trap consists of a ring electrode (internal radius is r 0 ) and two end caps separated by z 0.; all internal surfaces are hyperbolic. RF (and sometimes DC) voltages are applied to the ring electrode, establishing a quadrupolar electric field. Force on the ion increases as displacement from trap center increases; i.e., there is a restoring force along r when there is a repelling force along the z axis. Buffer gas (~ 1 mtorr) is added to damp stable ion motions and bunch ions at the center of the trap, where ejection occurs by increasing the rf amplitude, pushing ions in increasing m/z through the end-cap holes. Typical operating conditions: r 0 ~ 1 cm; f ~ 1 MHz (applied to ring electrode) Amplitude of rf (V) in range of 1000 to 10,000 V f 0 = 0.5 to 1 MHz Scan rate 5000 u/sec (normal); 500 u/sec (zoom scan) He press: 1 mtorr CI gas (if used):.01 mtorr Ionization gate: I ms and adjustable Ion source: both internal (EI and CI) and external (ESI). 43

2 B. ELECTRIC FIELD P = P 0 2 (x r 2 + y 2 2z 2 ) 0 Since x = r cos, y = r sin, and z = z, then P = P 0 r 0 2 (r 2 cos 2 + r 2 sin 2 2z 2 ) Since cos 2 + sin 2 = 1, then P = P 0 2 (r r 2 2z 2 ) 0 Where r 0 is internal radius of the ring electrode. This equation gives the potential for any point within the quadrupolar field of the trap. In the ideal trap r 2 0 = 2z 2 0 C. EQUATIONS OF MOTION Since P 0 = U + V cos t as defined previously for the quadrupole, then Fz = m Ø2 z Øt 2 = q ØP Øz =+q 4z r 0 2 (U Vcos t) Fr = m Ø2 r Øt 2 = q ØP Ør 0 = q 2r r 0 2 (U V cos t) and m Ø 2 z, and Øt 4q 2 2 (U V cos t)z = 0 r 0 m Ø 2 r Øt 2 + 2q r 0 2 (U V cos t)r = 0 In a manner identical to that of the quadrupole mass analyzer, we can obtain solutions for the equations of motion and determine: az = 16qU and m(r z 02 ) = 2a r qz = 2 8qV m(r z 02 ) 2 = 2q r Usually, we can ignore az and ar because all commercial ion traps do not use any DC potential, which is proportional to U. The operation of the ion trap is along the qz axis. Note that commercial traps are stretched such that the value of z 0 is increased by ~10% or more so that there is no longer a purely quadrupolar geometry. The equations cannot be simplified further as r 2. 0! 2z

3 D. STABILITY DIAGRAMS As for the linear quadrupole, the solutions to the Mathieu equation are of two classes: (1) periodic and stable and (2) periodic and unstable. Note the strong similarity of this stability diagram with that of the linear quadrupole. Ions can be stored in the trap if their trajectories are stable in r and z simultaneously; that is where the r-stable and z-stable regions overlap, as seen in the next figure. Here one sees the two regions where ion trajectories are stable (labeled A and B). Region A is chosen because it allows for U to be zero or nearly so and for V to have reasonable values (not excessively high voltages). Blow up of region A is shown in the next figure. 45

4 This is the stability region in (a z, q z ) for region A in the previous figure. The iso z and r are shown and explained below. The q z axis intersects the z = 1 where q z = 0.908, which is q max in the mass-selective instability mode. This means that stored ions can be scanned out (from low to high m/z) by raising V (moving along the q z axis). Upper m/z for 1-cm trap at q z = is 650 and if r 0 is decreased to 7 mm, the value is 850. Improvements in upper m/z are described later. E. SECULAR FREQUENCIES The 3-D representation of an ion trajectory has the appearance of a Lissajous curve. There are not only fundamental frequencies but also higher order (n) frequencies. These frequencies are given by the eqns below: u,n =(n u) for pos values of n where a u + q u

5 E. OBTAINING SPECTRA: ION MOTION SIMULATION 1.To manipulate the ions, one must be able to store them in the trap. (Taken from web site of R.G. Cooks at Purdue) 2. The ions are then cooled by collisions with helium gas and focused to the center of the trap. This focusing is very important as when ions are ejected, they start from a relatively sharp point, where they are bunched. 47

6 3. A mass spectrum is then scanned: Ions of increasing mass are successively made unstable and ejected from the trap owing to an increase in V (the amplitude of the rf). With a standard sized trap, this gives an upper m/z limit of 650. Higher m/z limits can be achieved by using resonant excitation (see below). 4. MS/MS and extended mass use resonant excitation: Ions can be transl. excited by applying a suppl. oscillating potential (up to 7 V) across the end caps. This excites the axial motion, ejecting ions from the trap in precursor-ion selection for MS/MS, collisionally exciting (lower amplitude) for MS/MS, or exciting to promote endothermic ion-molecule rxs. MS/MS can be done to MS

7 Resonant excitation at lower frequencies also extends the m/z range of the ion trap (see below). F. CALCULATIONS 1. q z determines lower mass limit: For ions of m/z 200 in a trap where r 0 = 1.0 cm and z 0 = cm, U = 0, V = 1130 V (0-p), and f = 1.05 MHz, 8qV q z = m(r z 0 2 ) 2 Or q z = The low mass (LM) cut off is given by: m LM m = q z q z LM = since q z is proportional to m. m LM = 99. This means that when 1130 V are applied to the trap, only ions with m/z > 99 will be stored. This shows a limitation of the method: in MS/MS, not all ions will be stored in the trap, and the lower mass region of the product-ion spectrum will be missing. 2. β z and ω z : a u + q u 2 From 2, a q z of 0.45, β z = 0.32 (recall that a z is zero). For m/z 2000, q z =.045 and β z = Where are these on the stability diagram? Now we can calculate a fundamental( n= 0) axial secular frequency, ω z. Recall that u,n =(n u). Therefore, when β z =0.32, ω z = 1,050,000 rad/s or 167 khz. 3. Upper m/z: The upper m/z limit will be for an ion having a q z of 0.90 for the maximum rf amplitude (V = 7340 V), or m/z ~ Extended m/z range: Obviously, for LC/MS, we need an upper m/z > 650. An ion of m/z 1300 will have a q z of 0.45 and remain trapped. The ω z for this ion is 167 khz (see #2 above). If we excite with a low amplitude oscillating voltage of 167 khz applied to end caps, then this ion could be ejected if the maximum rf amplitude (V) is applied to the ring. If one used a q z of (ω z = 16.7 khz), the upper m/z limit would be 13,000. The ac applied to the end caps causes new regions of instability to be 8qV created, thus extending the mass range. Given that m max = can also extend the m/z range by: lowering, 49 q eject (r z 0 2 ) 2, one

8 increasing V, reducing r 0 (smaller trap). Recall the is fixed at q eject 5. Mass Resolving Power: The narrowest peak width in commercial instruments is ~0.2 Da, obtained by a slow ( zoom ) scan (rate can be 1000 times slower). G. EXTERNAL SOURCES The most successful instrument has been a trap with an external ESI source: The quadrupoles thermalize ions and focus them to a center trajectory so that they can be injected into the trap. This also has been done with MALDI, but only in a research setting. Note that high vacuum is not needed. H. AUTOMATIC GAIN CONTROL Space charge and unwanted ion-molecule reactions in trapping instruments can be a problem. To control number of ions for each scan (called a micro scan), two ionization stages are used. The first is of a fixed duration (e.g., 0.2 ms), and the ions detected. This ion signal is then used to calculate an optimization ionization (or admittance) time to fill the trap. The computer system keeps track of these times so the spectra can be normalized. I. SUMMARY Simple, robust, relatively inexpensive, easy to operate Fast scanning, detects sequentially all ions (except low m/z) upper m/z limit on commercial instruments Capable of Ms n but low m/z ions not trapped Commercially available as GC/MS and LC/MS (low voltage) Sensitivity comparable to that of most contemporary instruments Medium RP available but no exact mass Fourier transform possible. 50

9 IX. FOURIER TRANSFORM ION CYCLOTRON RESONANCE MS (FT-ICR-MS, FTMS) (MSR, 1998, 17, 1; JMS, 1996, 31, 1325; SCIENCE, 10/19/84) A. NATURAL MOTIONS OF IONS IN MAGNETIC TRAP Figure shows the motion of an ion in a B field in a plane perpendicular to the page. One motion is for + ions the other for - ions. Equating centrifugal force and that due v r = c = qb m where c = 2 f and f is in Hz. to the B field, we obtain: 2 mv r = qvb where v is the velocity (m/s) in xy plane. So we may write: 1. What is the radius of the circular Motion? mv 2 2 = kt r = qb 1 2mkT r Since r = mv and, then. At 3 tesla for m/z 100, qb 0.08 mm. Note that ions of greater m/z will have larger orbits at any T. 2. What is the KE of an ion for a given radius of trajectory? Since v = qbr and mv 2 m, 2 = KE KE = q 2 B 2 r 2 2m. If the radius is known, the KE of the ion can be calculated. Note that energy goes up with the square of radius and B field and decreases with increasing mass. For example, an ion of m/z 1000 orbiting at 1.0 cm in a B field of 7.00 Tesla has an energy of 240 ev (compare to ~50 ev for ions in quad.) 3. Ions stored and Detected in a Penning trap. Ions can be stored in a cubic trap, shown in the lower part of the figure on next page. Ions are constrained to the xy plane by the B field and are trapped in that plane. To contrain the ions along z, small trapping voltages (~ 1 V) or placed on the trapping plates. Ion excitation is done by applying to the excitation plates an alternating voltage with a frequency equal to the natural cyclotron frequency (qb/m) of the ion. This frequency is in the khz to low MHz range. The excitation imparts coherent motion to the ion packet. 51

10 Detection is done by using the receiver (detection) plates. As the ion packet nears the receiver plate, an image current is induced on the plates (electrons in the plate are attracted toward positive ions). That alternating image current across the shunt resistor give an image voltage (V = ir) for detection. 4. Three basic motions in the trap as result of E field: Because ions are held along the z axis by a small trapping field, they oscillate along z. The frequency in the trap field is that of a harmonic oscillator and given by: T = 2 fv T = Ez (Trapping) where a is the width along z axis, V T is the trapping voltage, and α is a geometric constant (1.3869), and Ez = 4qV T a. For an ion of m/z 100 in a m 3-T field and a 1-V, 1-cm cubic trap, T = 12 khz, whereas c = 460 khz. The three important forces on ions in a FT mass spectrometer cell are centrifugal (an outward force), magnetic (an inward force), and electric (due to the trap voltage--outward in the xy plane). We may balance these forces: 52

11 mv 2 r = qb 0 v qe 0 r The electric field complicates the cyclotron frequency: + = qb + q2 B 2 4mqE 0 2m (Cyclotron) Compute the value of + when the electric field goes to zero! When ions are off center, they are in a EXB field, and they will undergo a low frequency oscillation around the center of the trap: magnetron motion. = qb q2 B 2 4mqE 0 2m (Magnetron) Compute the value of when the electric field goes to zero. B. EVENT SEQUENCE AND FOURIER TRANSFORM 53 The quench is accomplished by raising the voltage on one trapping plate, causing ejection of any ions in the trap. Ionization could be an electron-beam pulse (EI), a laser beam pulse (MALDI, MPI), or injection of ions from external source. The time between ionization and

12 excite can be used by ion-molecule reactions, CID, ion spectroscopy, etc. Excitation is by a pulse at one frequency, a rapidly scanned frequency ( chirp ), or by selected waveform inverse FT. Figures a and b show that a rectangular waveform can be used by excitation. The narrower the waveform (pulse), the wider the range of excitation, as shown by the FT. Subfigure c shows a chirp, i.e., a rapid scan from low to high freqency, excited all ions have corresponding cyclotron frequencies. Figures d and e show a tailored excitation done by SWIFT. Figure e shows how ejection of some frequency (mass) range can be implemented. One could also use a single frequency to excite one ion only. The outcome of excitation can be a prelude to detection, MS/MS via collisional activation, or ejection (called double resonance). 54

13 C. MULTICHANNEL ADVANTAGE An important feature of FTMS is that the signal output (the sum of the cyclotron frequencies of all the excited ions) contains all the spectral information in the time domain. Transformation to the frequency domain gives a complete mass spectrum of all the excited ions (f = 2πqB/m). The signal decays (is transient) due to collisions of the excited ion packet with background gas. D. HIGH RESOLVING POWER AND RELATION TO S/N: Besides ion trapping, the most notable feature of FTMS is its high mass resolving power. Note that the longer the transient that is sampled, the higher the mass resolving power (the more certain is the frequency or the more narrow the peak width). Further, as resolving power increases, so does peak height, giving an interesting phenomenon that both resolving power and S/N are improved 55

14 simultaneously, but at the expense of time of observation. Besides long observation time, one needs low pressure. High resolving power detection of gramicidin S for ions produced by ESI. Usually resolving power is computed by m/ m where m is the width at half height (peak shapes tend to be Lorenzian, and broadened artificially at the base). Note the wide separation of isotopic peaks. E. FACTORS DETERMINING RESOLVING POWER The length of the transient is the principal factor that determines RP. One wishes to have a long transient--therefore, low pressure and long observation time are needed. 1. Pressure in the trap: R [ k qb mp( /n) where k is 8.6x10-10 u. cm 3 torrt -1 s -1 and /n is 1x10-9 cm 3 s -1 molecule -1, q is mrp expressed as units of charge (1, 2,...), and m as u. The term B 0.86 for +1 or -1 ions. For example, if a resolving power (R) of 50,000 is sought at 1.2 T for m/z 200, then P ~ 1x10-7 torr. 2. Observation time: R [ kqbt m Where k = 1.3x10 7 u.t -1 s -1 and charge is +1, 2,... Note that R is inverse to pressure and directly proportional to observation time. For example, if R = 1,000,000 is sought for an ion of m/z 100 at 3 T, 56

15 then t > 2.6 s, provided the pressure is sufficiently low to allow the transient decay to last for the required time. 3. Computer limitations: How many data points are needed if R = 1,000,000 is sought for an ion of m/z 100 (f = 460,000 Hz)? 920,000 points are needed to satisfy the Nyquist theorem: 920,000 pts/s x 2.5 s (see #2 above) = 2.3 million points or 2.3 Mbytes of fast memory. If we start the scan at m/z 50, then we need 4.6 million points (the number of points is determined by the lowest m/z, which has the highest frequency). If computer memory size becomes a problem, then turn to the heterodyne detection. Consider f 1 = 461,000 Hz and f 2 = 460,000 Hz. The difference frequency is 1000 Hz, and multiplied by 2 is 2000 Hz (to satisfy Nyquist) pts x 2.5 s = 5000 pts or 5K of memory. What is sacrificed is mass range. 57

16 F. UPPER MASS LIMIT (CRITICAL MASS) mc = qb 2 a 2 8V eff Where a is the width of a cubic trap, V eff is the effective trap voltage, and is a geometry constant depending on trap design (1.3869). For a 2.5-cm trap in a 13-T field, and a trapping voltage of 1 V, the critical m/z is ~920,000. Thermal motions of ions also determine the upper m/z (no consideration of trap field), and this is more limiting: mupper = q2 B 2 r 2 2kT G. DYNAMIC RANGE Dynamic range is always a problem with trapping instruments, owing to space charge. In FTMS, image current can be detected for ions. The trap saturates at ions (depending on trap volume and B). Therefore, the dynamic range is Improvements can be made for minor ions by ejecting major ones and repeating the experiment. H. ION STORAGE TIME t 1/2 = (ab) 2 P(V T ) Where is polarizability, is reduce mass (m 1 m 2 /(m 1 + m 2 ). We can detect 30% of the ions after 13.5 hr at P = 10-8 torr, 4.7 T, and ion m/z of 18. I. EXACT MASS MEASUREMENT One can measure frequency to 1 ppb. In principle, m/z measurements should be equally precise and accurate. Problem is the trap field, which affects ion frequency in ways not yet completely understood. The calibration is two-point, and state-of-the-art is ppm (at best 0.1 ppm) although a thorough evaluation has not been done (can be derived from the equation showing balance of electrical, magnetic, and centrifugal forces. 58

17 J. MS/MS IN FTMS ( TANDEM IN TIME ) Translational excitation is simply another event in the FTMS sequence. Instead of translational excitation, one can also add internal energy via an IR or UV laser or via collisions with surfaces. 1. Collisions via cyclotron excitation: Excitation makes use of ω = qb/m. Problem is that the laboratory energy varies with m/z of parent and decreases as m/z increases. 2. Sustained off resonance irradiation and related methods: SORI uses an excitation frequency that is slightly off resonance, allowing ions to be excited in deexcited--the ion orbit repeatedly increases then decreases In very-low-energy excitation, the excitation is on resonance but the phase alternates between 0 and 180 o. In multiple excitation for collisional activation, the parent ions are excited briefly and then allowed to relax by collisions. In each method, a collision gas (He, Ne, etc) is pulsed to the torr range and then turned off for mass analysis. SORI is most common. 3. IR Multiphoton Dissociation (IRMPD). An IR laser dumps in many photons as the internal energy is raised slowly by a ladder-climbing means. Available on commercial instruments. 59

18 4. Electron-Capture Dissociation (ECD) Low energy electrons for a filament (as in EI) are captured by multiply charged ions (e.g., from proteins) to form a hypervalent species. Fragmentation occurs at many sites of the multiply protonated species and is thought to be nonergodic (not governed by QET). K. IMPROVING PERFORMANCE Current research is directed at: making the electric fields in the cell more ideal for both excitation and detection, using external sources and intermediate storage devices to decouple ion source and mass analyzer, cooling ions for focusing to center of cell and for ion-molecule reactions and improved mass analysis, developing new means of MS/MS. 1. Lowering background pressure with dual cell: Ions are made in the high pressure region on left, where ion-molecule reactions or CAD are conducted. The low-pressure region on the right is for high RP mass analysis. Just prior to mass analysis, ions are allowed to partition on both sides of the trap (conductance limit at zero volts), and then the voltage on the conductance limit is increased, isolating half the ions in the low-pressure trap on the right, where mass analysis takes place. The ions left on the left are discarded. 2. External ion source. Here the source and mass analysis are nearly completely decoupled. Any ion source can be used, and the ions are transmitted via quadrupole or electrostatic lens into the ion trap. Ions are stopped in the cell by gated trapping. Once the ions are stored, mass analysis or other experiments 60

19 (MS/MS, ion-molecule reactions) can be conducted. Most commercial instrument now use this strategy--see figure below. Notice that the single magnet can be equipped with MALDI (on left) and ESI (on right) simultaneously, although the two ionization modes are used separately and independently. 3. Increasing inward force on ion (minimize trap field): The magnet force is inward but the trapping electric field provides an outward force, which works against B and leads to poor performance (note the radial force in the equipotential field lines). Ion a is subjected to only an inward trapping force whereas ion b experiences an upward and inward force. Problem with cubic and cylindrical traps. Efforts are underway to minimize upward or radial forces. Elongated cells (McIver) Screened cells (Marshall) 61

20 Hyperbolic cell (like ion trap) (Rempel and Gross) Example cells (traps) used in FTMS. (a) cubic, (b) cylindrical (can also be open with trapping plates replaced by trapping rings), (c) hyperbolic, (d) multisection (e.g., dual), (e) elongated and screened. E = excitation plate or electrode, D = detector plate, T = trap plate, S = screen. 62

21 Compensated cell (Rempel and Gross) Exploded view of compensated trap. 1.7 V is applied to outer trapping electrode whereas 1.0 V is applied to inner electrode. The field lines from 1.7 to 1.0 give a compensating inward force for ions such as b in the previous drawing. 4. Cooling ions translationally and internally: Especially needed for MALDI-produced ions formed at high translational energy (750 m/s, in a micro supersonic expansion), and for remeasurement strategies to relax excited ions to center of trap for re excitation. Gas pulsing (limited to 10-5 torr) Comatrix that decomposes to give high local P for MALDI Active cooling by quadrupole axialization: Use the four, nontrapping plates. Ions absorb at 2ω +, 2ω -, and (ω + + ω - ). At the latter frequency, ion motion is converted between pure cyclotron and pure magnetron. During the cyclotron mode, collisions with buffer gas cause motion to damp and ion cloud collapses. Mass range is a factor of 2 times. Active cooling by rf-only mode event: By applying 1 MHz to the trap plates, the Penning trap is converted into a Paul trap (ion trap). In the Paul mode, pressure can be increase to mtorr range, promoting ion collisions and collapse of ion cloud to center of trap. Useful for ion chemistry studies where intermediates can be stabilized by collisions. 63

22 More on Compensated traps Below is the cyclindrical cell used in the Ion Spec FT mass spectrometer. The end cylinders serve the same function as flat trapping plates in the cubic or rectangular traps. The center section is comprised of four segments: two excitation electrodes and two receiver electrodes. Note how laser and electron beams can be easily admitted to the trap from the right and intersect in the center the trapped ions. Below is the compensated cylindrical trap, designed by D. Rempel. The center section is as before except slightly shorter. Three addn ring electrodes are added, as per the design, and to these electrodes are applied voltages, also calculated according to the design, to shim the electric fields in the trap. The outcome is higher RP (3-5 X) and sensitivity (10 X)

23 L. TRAP/FTMS: Ions are stored in linear ion trap (similar to quadrupole ion trap but with 10X more capacity). There they can be thermally relaxed, focussed to trap center, and made ready for injection into FTMS trap for high resolving power and accurate mass. Following injection into FTMS, the linear trap is filled again, and MS/MS experiments can be done in a data-dependent manner, giving 5 product-ion spectra per second. FTMS is used for accurate mass measurements with a transient just under 1 sec, giving a RP of 100,000 at m/z 400. RP can be increased by increasing the observation time for transient. M. SUMMARY Works at low pressure (10-8 torr) Relatively simple mass analyzer but complicated by external sources and high-pumping demands Traps ions for minutes; good for ion chemistry, spectroscopy Nondestructive detection (remeasurement) Interfaces well with pulsed ionization (lasers)--multichannel Highest mass resolving power, but needs time for transient Requires expensive and maintenance-demanding magnet High mass limit (critical mass) Subject to space-charge effects, giving low dynamic range (improvable by ejecting unwanted ions) Highest accuracy m/z and can be improved (improvement not likely for sectors or TOFs) 64

24 X. ORBITRAP (A. Makarov, Anal Chem, 2006, 2113; rev JMS, 40, 430) A. DESIGN OF KINGDOM TRAP The ions are injected with velocity perpendicular to long axis of the Orbitrap (z-axis). Injection at a point displaced from z = 0 gives ions pot energy in z. Injection at this pt on the z-potential analogous to pulling back a pendulum bob and then releasing it to oscillate... Ion motion along z is an harmonic oscillator and is completely independent of radial (r) and angular (w) motion. Ion mass/charge ratio m/z is simply related to the frequency of ion oscillation along the z-axis. = [(z/m) k]

25 B. DESIGN OF INSTRUMENT The instrument is coupled with a linear ion trap to produce a high resolving power tandem instrument, interfaced to ESI. C. ACQUISITION OF MASS SPECTRUM Mass spectra are found in a time-domain, transient decay that contains all the ion frequencies. When transformed to the frequency domain, one has a mass spectrum. 66

26 Figure. Transient and mass spectrum of bovine insulin (MW ~5700). Mass calibrants, separated by 100 m/z, are also seen in the spectrum. D. MASS RESOLVING POWER RP determined by acquisition time (like FTMS). Commercial instrument has RP of 40,000 at m/z 400. Limited by pressure and imperfections in the machining. In MS/MS mode, can obtain high RP for precursor and product ions and accurate m/z in both modes! The spectrum is that of vancomycin. Note good agreement between calc and exp 67

27 isotope patterns. High RP and accurate m/z available for product ions in fast scanning. E. ACCURATE MASS Accurate m/z with a two point calibration is ~ 2 ppm. Long-term (days) is ~ 5 ppm, indicating that calibration is not needed for every spectrum. The accuracy remain high for 1,000,000 ions stored in trap (good dynamic range) F. S/N, TRANSMISSION, & DETECTION LIMIT Transmission from linear ion trap to orbitrap is 30-50%, lower for higher m/z ions. S/N = 4 for ~ 20 ions stored in the orbitrap. Detection limit for bradykinin (peptide) ~ 4 attomol--see figure; [Bradykinin] = 3 nm. G. APPLICATIONS LC/MS/MS for proteomics, metabolomics, std mass spectrometry. Various modes of ion activation--in ion trap or in the C trap. High RP for product ions (RP ~ 40,000 at m/z 400), but the RP is inverse to m/z and time of acquisition just like the FTICR. Question: is this the instrument to couple with high performance UPLC where, in peptide analysis, the chromatographic peaks are a few secs wide? 68

28 Hyperbolic quadrupole mass filter precise selection of precursor ions and outstanding ion transmission Q-Exactive Orbitrap Mass Spectrometer Orbitrap mass analyzer unrivaled mass resolution and unmatched spectrum quality S-lens ion source excellent ion transmission for enhanced sensitivity C-trap premier in-spectrum dynamic range HCD collision cell higher fragmentation efficiency for unsurpassed MS/MS spectra quality Carbofuran Formetanate 100 C12H15NO3 C11H15N3O2 M+H+ = M+H+ = R=17,500 R=35,000 R=70,000 R=140, Relative Abundance m/z m/z m/z m/z Nominally isobaric pesticides carbofuran and formetanate are unresolved or only poorly resolved at the highest resolutions available in Q-TOF instruments. The higher resolutions easily achieved by the Q Exactive instrument clearly resolve analytes from complex matrix interferences without sacrificing sensitivity. 3

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS)

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS) Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS) The first steps along the way to FT- ICR-MS date back to the 1930s, when Ernest Lawrence, at the University of California - Berkeley,

More information

Types of Analyzers: Quadrupole: mass filter -part1

Types of Analyzers: Quadrupole: mass filter -part1 16 Types of Analyzers: Sector or double focusing: magnetic and electric Time-of-flight (TOF) Quadrupole (mass filter) Linear ion trap Quadrupole Ion Trap (3D trap) FTICR fourier transform ion cyclotron

More information

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez Mass Analyzers Ion Trap, FTICR, Orbitrap CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography Prof. Jose-Luis Jimenez Last Update: Oct. 014 Some slides from Dr. Joel Kimmel (007) MS Interpretation

More information

Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap

Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap 1 Last Time Mass Analyzers; CAD and TOF mass analyzers: 2 Fourier Transforms A transform is when you change your analytical space without

More information

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps ION ANALYZERS MASS ANALYSER sample Vacuum pumps Mass analysers - separate the ions according to their mass-to-charge ratio MASS ANALYSER Separate the ions according to their mass-to-charge ratio in space

More information

Tendenze nell innovazione della strumentazione in spettrometria di massa:

Tendenze nell innovazione della strumentazione in spettrometria di massa: Tendenze nell innovazione della strumentazione in spettrometria di massa: Trappola lineare, orbitrap, ion mobility e nuova strumentazione René Magritte La Condition Humaine 1 Ion Trap Mass Spectrometry

More information

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein. Mass Analyzers After the production of ions in ion sources, the next critical step in mass spectrometry is to separate these gas phase ions according to their mass-to-charge ratio (m/z). Ions are extracted

More information

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511 524) (Harris, Chapt.

More information

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511-524) (Harris, Chapt.

More information

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS) Mass Analyzers

More information

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk Previously: Resolution and other basics MALDI Electrospray 40 Lecture 2 Mass analysers Detectors

More information

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass

More information

Ion trap. 3D Quadrupole Trap. -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April

Ion trap. 3D Quadrupole Trap. -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April Ion trap 3D Quadrupole Trap -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April 011 1 The trajectories of the ions can be given: They form Mathieu differential equations: with r + Ion

More information

MS/MS .LQGVRI0606([SHULPHQWV

MS/MS .LQGVRI0606([SHULPHQWV 0DVV6SHFWURPHWHUV Tandem Mass Spectrometry (MS/MS) :KDWLV0606" Mass spectrometers are commonly combined with separation devices such as gas chromatographs (GC) and liquid chromatographs (LC). The GC or

More information

Harris: Quantitative Chemical Analysis, Eight Edition

Harris: Quantitative Chemical Analysis, Eight Edition Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 21: MASS SPECTROMETRY CHAPTER 21: Opener 21.0 Mass Spectrometry Mass Spectrometry provides information about 1) The elemental composition of

More information

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 DEPARTMENTS OF CHEMISTRY AND BIOCHEMISTRY GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 Mass Analysers Shabaz Mohammed October 20, 2015 High Vacuum System Turbo pumps Diffusion pumps Rough pumps Rotary

More information

Chem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY:

Chem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY: Chem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY: I. BASIC ELEMENTS OF A MASS SPECTROMETER Inlet System or Chromatograph Ion Source Mass Analyzer Detector Computer II. ION SOURCES A. Electron

More information

Ion traps. Quadrupole (3D) traps. Linear traps

Ion traps. Quadrupole (3D) traps. Linear traps Ion traps Quadrupole (3D) traps Linear traps 3D-Ion traps A 3D-ion trap can be considered the tridimensional analogue of the linear quadrupole mass analyzer: B A C D The central ring electrode of the 3D-ion

More information

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

(Refer Slide Time 00:09) (Refer Slide Time 00:13) (Refer Slide Time 00:09) Mass Spectrometry Based Proteomics Professor Sanjeeva Srivastava Department of Biosciences and Bioengineering Indian Institute of Technology, Bombay Mod 02 Lecture Number 09 (Refer

More information

Mass Spectrometry in MCAL

Mass Spectrometry in MCAL Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring

More information

Mass Spectrometry. What is Mass Spectrometry?

Mass Spectrometry. What is Mass Spectrometry? Mass Spectrometry What is Mass Spectrometry? Mass Spectrometry (MS): The generation of gaseous ions from a sample, separation of these ions by mass-to-charge ratio, and measurement of relative abundance

More information

Chemistry 311: Topic 3 - Mass Spectrometry

Chemistry 311: Topic 3 - Mass Spectrometry Mass Spectroscopy: A technique used to measure the mass-to-charge ratio of molecules and atoms. Often characteristic ions produced by an induced unimolecular dissociation of a molecule are measured. These

More information

This is the total charge on an ion divided by the elementary charge (e).

This is the total charge on an ion divided by the elementary charge (e). 12.2 Fundamentals and general terms Accelerator mass spectrometry An ultra-sensitive technique using tandem accelerators employed mainly for the study of long-lived radioisotopes, and stable isotopes at

More information

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry Fourier Transform Ion Cyclotron Resonance Mass Spectrometry What is Mass Spectrometry? Gravity F= G m M/r 2 r M m Newton Newton s second law F = m a Mass-Energy equivalence E = m c 2 1 ev = ~10-9 amu Einstein

More information

Chemistry Instrumental Analysis Lecture 35. Chem 4631

Chemistry Instrumental Analysis Lecture 35. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 35 Principle components: Inlet Ion source Mass analyzer Ion transducer Pumps Signal processor Mass analyzers Quadrupole Time of Flight Double Focusing Ion

More information

Atomic masses. Atomic masses of elements. Atomic masses of isotopes. Nominal and exact atomic masses. Example: CO, N 2 ja C 2 H 4

Atomic masses. Atomic masses of elements. Atomic masses of isotopes. Nominal and exact atomic masses. Example: CO, N 2 ja C 2 H 4 High-Resolution Mass spectrometry (HR-MS, HRAM-MS) (FT mass spectrometry) MS that enables identifying elemental compositions (empirical formulas) from accurate m/z data 9.05.2017 1 Atomic masses (atomic

More information

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS Mass Spectrometry Hyphenated Techniques GC-MS LC-MS and MS-MS Reasons for Using Chromatography with MS Mixture analysis by MS alone is difficult Fragmentation from ionization (EI or CI) Fragments from

More information

Introduction to the Q Trap LC/MS/MS System

Introduction to the Q Trap LC/MS/MS System www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap LC/MS/MS Introduction to the Q Trap LC/MS/MS System The Q Trap

More information

1. The range of frequencies that a measurement is sensitive to is called the frequency

1. The range of frequencies that a measurement is sensitive to is called the frequency CHEM 3 Name Exam 1 Fall 014 Complete these problems on separate paper and staple it to this sheet when you are finished. Please initial each sheet as well. Clearly mark your answers. YOU MUST SHOW YOUR

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?

More information

What is Tandem Mass Spectrometry? (MS/MS)

What is Tandem Mass Spectrometry? (MS/MS) What is Tandem Mass Spectrometry? (MS/MS) Activate MS1 MS2 Gas Surface Photons Electrons http://www.iupac.org/goldbook/t06250.pdf Major difference between MS and MS/MS: two analysis steps Ionization Analysis

More information

LECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile

LECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile LECTURE-11 Hybrid MS Configurations HANDOUT PREAMBLE As discussed in our previous lecture, mass spectrometry is by far the most versatile technique used in proteomics. We had also discussed some of the

More information

Translational Biomarker Core

Translational Biomarker Core Translational Biomarker Core Instrumentation Thermo Scientific TSQ Quantum Triple Quadrupole Mass Spectrometers. There are two TSQ Quantum Ultra AM instruments available in the TBC. The TSQ Quantum Ultra

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?

More information

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. MAGLAB Summer School Ryan P. Rodgers

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. MAGLAB Summer School Ryan P. Rodgers Fourier Transform Ion Cyclotron Resonance Mass Spectrometry MAGLAB Summer School 2015 Ryan P. Rodgers Electrospray Ionization Source Drying Gas Capillary Evaporation Chamber High Voltage Supply Charged

More information

MS Goals and Applications. MS Goals and Applications

MS Goals and Applications. MS Goals and Applications MS Goals and Applications 3 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis

More information

Quadrupole Storage Mass Spectrometry

Quadrupole Storage Mass Spectrometry Quadrupole Storage Mass Spectrometry RAYMOND E. MARCH AND RICHARD J. HUGHES Trent University Peterborough, Ontario, Canada with a historical review by John F. J. Tbdd University of Kent Canterbury, Kent,

More information

TANDEM MASS SPECTROSCOPY

TANDEM MASS SPECTROSCOPY TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER

More information

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific LTQ Orbitrap Velos

More information

The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results

The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results A. Kürten 1, J. Curtius 1 and S. Borrmann 1,2 1 Johannes Gutenberg-University Mainz, Germany 2 Max Planck

More information

Autoresonant Ion Trap Mass Spectrometer The RGA Alternative

Autoresonant Ion Trap Mass Spectrometer The RGA Alternative Autoresonant Ion Trap Mass Spectrometer The GA Alternative 1 What is a Mass Spectrometer? Ionizer Mass Spectrometer Mass Separator Magnetic Sector, Quadrupole or Ion Trap Detector Partial Pressure Measurement

More information

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai 303351 Instrumental Analysis Mass Spectrometry Lecturer:! Somsak Sirichai Mass Spectrometry What is Mass spectrometry (MS)? An analytic method that employs ionization and mass analysis of compounds in

More information

Time-of-Flight Mass Analyzers

Time-of-Flight Mass Analyzers Time-of-Flight Mass Analyzers Jonathan Karty C613 lecture 1 March 6, 8 (Section 4. in Gross, pages 115-18) TOF Overview Time-of-flight (TOF) is the least complex mass analyzer in terms of its theory Ions

More information

Introduction to GC/MS

Introduction to GC/MS Why Mass Spectrometry? Introduction to GC/MS A powerful analytical technique used to: 1.Identify unknown compounds 2. Quantify known materials down to trace levels 3. Elucidate the structure of molecules

More information

Interazioni di ioni con elettroni (ECD, ETD) e fotoni (Ion spectroscopy) Gianluca Giorgi. via Aldo Moro Siena

Interazioni di ioni con elettroni (ECD, ETD) e fotoni (Ion spectroscopy) Gianluca Giorgi. via Aldo Moro Siena Interazioni di ioni con elettroni (ECD, ETD) e fotoni (Ion spectroscopy) Gianluca Giorgi Università degli Studi di Siena Dipartimento di Biotecnologie, Chimica e Farmacia via Aldo Moro 53100 Siena e-mail:

More information

MASS SPECTROMETRY. Topics

MASS SPECTROMETRY. Topics MASS SPECTROMETRY MALDI-TOF AND ESI-MS Topics Principle of Mass Spectrometry MALDI-TOF Determination of Mw of Proteins Structural Information by MS: Primary Sequence of a Protein 1 A. Principles Ionization:

More information

Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.

Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS. ESS The RGA freenotes Theory page 1 of 14 RGA Theory Notes Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.

More information

MS Goals and Applications. MS Goals and Applications

MS Goals and Applications. MS Goals and Applications MS Goals and Applications 1 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis

More information

LC-MS Based Metabolomics

LC-MS Based Metabolomics LC-MS Based Metabolomics Analysing the METABOLOME 1. Metabolite Extraction 2. Metabolite detection (with or without separation) 3. Data analysis Metabolite Detection GC-MS: Naturally volatile or made volatile

More information

Thermo Finnigan LTQ. Specifications

Thermo Finnigan LTQ. Specifications IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Finnigan LTQ Specifications

More information

Mass Spectrometry. Fundamental LC-MS. Mass Analysers

Mass Spectrometry. Fundamental LC-MS. Mass Analysers Mass Spectrometry Fundamental LC-MS Mass Analysers i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

Ionization Techniques Part IV

Ionization Techniques Part IV Ionization Techniques Part IV CU- Boulder CHEM 5181 Mass Spectrometry & Chromatography Presented by Prof. Jose L. Jimenez High Vacuum MS Interpretation Lectures Sample Inlet Ion Source Mass Analyzer Detector

More information

Mass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+.

Mass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+. Mass spectrometry is the study of systems causing the formation of gaseous ions, with or without fragmentation, which are then characteried by their mass to charge ratios (m/) and relative abundances.

More information

Physics of and in Ion Traps

Physics of and in Ion Traps Physics of and in Ion Traps Proposed Topics: TRIUMF, Vancouver June 01 Basics of Paul- and Penning-traps (equ. of motion, trap geometries, influence of trap imperfections,) Ion detection and cooling (Buffer

More information

Mass spectrometry.

Mass spectrometry. Mass spectrometry Mass spectrometry provides qualitative and quantitative information about the atomic and molecular composition of inorganic and organic materials. The mass spectrometer produces charged

More information

for the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York

for the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York Mass Spectrometry for the Novice John Greaves and John Roboz (^>, yc**' CRC Press J Taylor & Francis Group Boca Raton London New York CRC Press is an imprint of the Taylor & Francis Croup, an informa business

More information

Penning Traps. Contents. Plasma Physics Penning Traps AJW August 16, Introduction. Clasical picture. Radiation Damping.

Penning Traps. Contents. Plasma Physics Penning Traps AJW August 16, Introduction. Clasical picture. Radiation Damping. Penning Traps Contents Introduction Clasical picture Radiation Damping Number density B and E fields used to increase time that an electron remains within a discharge: Penning, 936. Can now trap a particle

More information

Precision Nuclear Mass Measurements Matthew Redshaw Exotic Beam Summer School, Florida State University Aug 7 th 2015

Precision Nuclear Mass Measurements Matthew Redshaw Exotic Beam Summer School, Florida State University Aug 7 th 2015 Precision Nuclear Mass Measurements Matthew Redshaw Exotic Beam Summer School, Florida State University Aug 7 th 2015 Outline WHAT are we measuring? - Nuclear/atomic masses WHY do we need/want to measure

More information

Quan/Qual Analyses. Unmatched Confidence for. Thermo Scientific Q Exactive Orbitrap LC-MS/MS System. Identify Quantify Confirm

Quan/Qual Analyses. Unmatched Confidence for. Thermo Scientific Q Exactive Orbitrap LC-MS/MS System. Identify Quantify Confirm 26.1645 165.91 33.267 Thermo Scientific Q Exactive rbitrap LC-MS/MS System Unmatched Confidence for Quan/Qual Analyses Identify Quantify Confirm Quanfirmation \ kwän-f r-'m -shän\ n 1 : the ability to

More information

Laser Dissociation of Protonated PAHs

Laser Dissociation of Protonated PAHs 100 Chapter 5 Laser Dissociation of Protonated PAHs 5.1 Experiments The photodissociation experiments were performed with protonated PAHs using different laser sources. The calculations from Chapter 3

More information

Analysis of Polar Metabolites using Mass Spectrometry

Analysis of Polar Metabolites using Mass Spectrometry Analysis of Polar Metabolites using Mass Spectrometry TransMed Course: Basics in Clinical Proteomics and Metabolomics. Oct 10-19, 2012 dd.mm.yyyy Vidya Velagapudi, Ph.D, Adjunct Professor Head of the Metabolomics

More information

Mass Selective Ejection by Axial Resonant Excitation from a Linear Ion Trap

Mass Selective Ejection by Axial Resonant Excitation from a Linear Ion Trap Mass Selective Ejection by Axial Resonant Excitation from a Linear Ion Trap Yuichiro Hashimoto, Hideki Hasegawa, Takashi Baba, and Izumi Waki Central Research Laboratory, Hitachi, Ld., Tokyo, Japan We

More information

ICPMS Doherty Lecture 1

ICPMS Doherty Lecture 1 ICPMS Doherty Lecture 1 Mass Spectrometry This material provides some background on how to measure isotope abundances by means of mass spectrometry. Mass spectrometers create and separate ionized atoms

More information

Space Charge Effects in Linear. Quadrupole Ion Traps

Space Charge Effects in Linear. Quadrupole Ion Traps Space Charge Effects in Linear Quadrupole Ion Traps by Cong Gao B.Sc., Peking University, 2010 A THESIS SUBMITTED IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREE OF DOCTOR OF PHILOSOPHY in The

More information

Technical Note # TN-24 Extending the Capacity in Modern Spherical High Capacity Ion Traps (HCT)

Technical Note # TN-24 Extending the Capacity in Modern Spherical High Capacity Ion Traps (HCT) Bruker Daltonics Technical Note # TN-24 Extending the Capacity in Modern Spherical High Capacity Ion Traps (HCT) During the last decade, RF ion traps have become a workhorse of the analytical laboratory

More information

Mass spectrometry gas phase transfer and instrumentation

Mass spectrometry gas phase transfer and instrumentation Objectives of the Lecture spectrometry gas phase transfer and instrumentation Matt Renfrow January 15, 2014 1. Make ions 2. Separate/Analyze 3. Detect ions 4. What is mass resolution and mass accuracy?

More information

ATHENA / AD-1. First production and detection of cold antihydrogen atoms. ATHENA Collaboration. Rolf Landua CERN

ATHENA / AD-1. First production and detection of cold antihydrogen atoms. ATHENA Collaboration. Rolf Landua CERN ATHENA / AD-1 First production and detection of cold antihydrogen atoms ATHENA Collaboration Rolf Landua CERN 1 LONG TERM PHYSICS GOALS Antihydrogen = Hydrogen? CPT Gravity But... 2 FIRST GOAL PRODUCTION

More information

Mass Spectrometry. Quantitative Mass Spectrometry Chiral Mass Spectrometry

Mass Spectrometry. Quantitative Mass Spectrometry Chiral Mass Spectrometry Mass Spectrometry Quantitative Mass Spectrometry Chiral Mass Spectrometry Quantitation by MS Goal is to develop methodology to sensitively, specifically, accurately and rapidly measure one or more compounds

More information

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n 20.2 Ion Sources electron ionization produces an M + ion and extensive fragmentation chemical ionization produces an M +, MH +, M +, or M - ion with minimal fragmentation MALDI uses laser ablation to transfer

More information

Chemistry Instrumental Analysis Lecture 34. Chem 4631

Chemistry Instrumental Analysis Lecture 34. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 34 From molecular to elemental analysis there are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry

More information

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching s by Removal of Isobaric Interferences Differential Mobility Used as a Tool to Address Selectivity Challenges Adrian M. Taylor

More information

Choosing the metabolomics platform

Choosing the metabolomics platform GBS 748 Choosing the metabolomics platform Stephen Barnes, PhD 4 7117; sbarnes@uab.edu So, I have my samples what s next? You ve collected your samples and you may have extracted them Protein precipitation

More information

Practical Quadrupole Theory: Peak Shapes at Various Ion Energies

Practical Quadrupole Theory: Peak Shapes at Various Ion Energies Practical Quadrupole Theory: Peak Shapes at Various Ion Energies Randall E. Pedder ABB Inc., Analytical-QMS Extrel Quadrupole Mass Spectrometry, 575 Epsilon Drive, Pittsburgh, PA 15238 (Poster presented

More information

arxiv:physics/ v1 [physics.plasm-ph] 5 Nov 2004

arxiv:physics/ v1 [physics.plasm-ph] 5 Nov 2004 Ion Resonance Instability in the ELTRAP electron plasma G. Bettega, 1 F. Cavaliere, 2 M. Cavenago, 3 A. Illiberi, 1 R. Pozzoli, 1 and M. Romé 1 1 INFM Milano Università, INFN Sezione di Milano, Dipartimento

More information

Collision Cross Section: Ideal elastic hard sphere collision:

Collision Cross Section: Ideal elastic hard sphere collision: Collision Cross Section: Ideal elastic hard sphere collision: ( r r 1 ) Where is the collision cross-section r 1 r ) ( 1 Where is the collision distance r 1 r These equations negate potential interactions

More information

Mass spectrometry of proteins, peptides and other analytes: principles and principal methods. Matt Renfrow January 11, 2008

Mass spectrometry of proteins, peptides and other analytes: principles and principal methods. Matt Renfrow January 11, 2008 Mass spectrometry of proteins, peptides and other analytes: principles and principal methods Matt Renfrow January 11, 2008 Objectives of the Lecture 1. Make ions 2. Separate/Analyze/Detect ions 3. What

More information

Ion traps. Trapping of charged particles in electromagnetic. Laser cooling, sympathetic cooling, optical clocks

Ion traps. Trapping of charged particles in electromagnetic. Laser cooling, sympathetic cooling, optical clocks Ion traps Trapping of charged particles in electromagnetic fields Dynamics of trapped ions Applications to nuclear physics and QED The Paul trap Laser cooling, sympathetic cooling, optical clocks Coulomb

More information

Dynamics of Ions in an Electrostatic Ion Beam Trap

Dynamics of Ions in an Electrostatic Ion Beam Trap Dynamics of Ions in an Electrostatic Ion Beam Trap Daniel Zajfman Dept. of Particle Physics Weizmann Institute of Science Israel and Max-Planck Institute for Nuclear Physics Heidelberg, Germany Charles

More information

1 Mass Analysis. Gene Hart-Smith and Stephen J. Blanksby

1 Mass Analysis. Gene Hart-Smith and Stephen J. Blanksby j5 1 Mass Analysis Gene Hart-Smith and Stephen J. Blanksby 1.1 Introduction Modern day mass analyzer technologies have, together with soft ionization techniques, opened powerful new avenues by which insights

More information

high temp ( K) Chapter 20: Atomic Spectroscopy

high temp ( K) Chapter 20: Atomic Spectroscopy high temp (2000-6000K) Chapter 20: Atomic Spectroscopy 20-1. An Overview Most compounds Atoms in gas phase high temp (2000-6000K) (AES) (AAS) (AFS) sample Mass-to-charge (ICP-MS) Atomic Absorption experiment

More information

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies Page 1 Agilent GC/MS Portfolio Smallest footprint, single GC column 5975VL with 6850GC Industry leading

More information

Welcome to Organic Chemistry II

Welcome to Organic Chemistry II Welcome to Organic Chemistry II Erika Bryant, Ph.D. erika.bryant@hccs.edu Class Syllabus 3 CHAPTER 12: STRUCTURE DETERMINATION 4 What is this solution Soda Tea Coffee??? 5 What is this solution Soda Tea

More information

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3 Second-Generation Electron Transfer Dissociation (ETD) on the Thermo Scientific Orbitrap Fusion Mass Spectrometer with Improved Functionality, Increased Speed, Christopher Mullen, 1 Lee Earley, 1 Jean-Jacques

More information

Overview of NETCHEM MSc & PhD courses: mass spectrometry in EFSC

Overview of NETCHEM MSc & PhD courses: mass spectrometry in EFSC AQualEER Chemistry faculty (Belgrade, Serbia, December 2017) Overview of NETCHEM MSc & PhD courses: mass spectrometry in EFSC Emeritus Prof. Dr Jean-Claude Tabet 1,2,3 1 Institut Parisien de Chimie Moléculaire

More information

RECOMMENDATIONS FOR NOMENCLATURE OF MASS SPECTROMETRY

RECOMMENDATIONS FOR NOMENCLATURE OF MASS SPECTROMETRY international UNION OF PURE AND APPLIED CHEMISTRY ANALYTICAL CHEMISTRY DIVISION COMMISSION ON ANALYTICAL NOMENCLATURE RECOMMENDATIONS FOR NOMENCLATURE OF MASS SPECTROMETRY RULES APPROVED 1973 LONDON BUTTER

More information

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy Section I Q1. Answer (i) (b) (ii) (d) (iii) (c) (iv) (c) (v) (a) (vi) (b) (vii) (b) (viii) (a) (ix)

More information

Design considerations for linear Paul trap mass spectrometer under development

Design considerations for linear Paul trap mass spectrometer under development Design considerations for linear Paul trap mass spectrometer under development S.Sevugarajan and A.G.Menon Department of Instrumentation, Indian Institute of Science, Bangalore 561, India. Abstract This

More information

CHM 424 EXAM 4 CRIB - COVER PAGE FALL

CHM 424 EXAM 4 CRIB - COVER PAGE FALL CHM 44 EXAM 4 CRIB - COVER PAGE FALL 006 There are seven numbered pages with five questions. Answer the questions on the exam. Exams done in ink are eligible for regrade, those done in pencil will not

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION Systematic shift caused by trap asymmetry The major systematic correction in the reported cyclotron frequency ratio comparison of an antiproton at ν c, p and a negatively charged hydrogen ion (H ) at ν

More information

Building Blocks for Quantum Computing Part IV. Design and Construction of the Trapped Ion Quantum Computer (TIQC)

Building Blocks for Quantum Computing Part IV. Design and Construction of the Trapped Ion Quantum Computer (TIQC) Building Blocks for Quantum Computing Part IV Design and Construction of the Trapped Ion Quantum Computer (TIQC) CSC801 Seminar on Quantum Computing Spring 2018 1 Goal Is To Understand The Principles And

More information

Lecture 15: Introduction to mass spectrometry-i

Lecture 15: Introduction to mass spectrometry-i Lecture 15: Introduction to mass spectrometry-i Mass spectrometry (MS) is an analytical technique that measures the mass/charge ratio of charged particles in vacuum. Mass spectrometry can determine masse/charge

More information

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS Back to Basics Section D: Ion Optics CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS TABLE OF CONTENTS QuickGuide...413 Summary...415 Introduction...417 The physical basis of orthogonal TOF....... 419 Pulsedmainbeamsofions...421

More information

Auxiliary Techniques Soft ionization methods

Auxiliary Techniques Soft ionization methods Auxiliary Techniques The limitations of the structural information in the normal mass spectrum can be partly offset by special mass-spectral techniques. Although a complete description of these is beyond

More information

Molecular Mass Spectrometry

Molecular Mass Spectrometry Molecular Mass Spectrometry Mass Spectrometry: capable of providing information about (1) Elemental composition of samples of matter: atomic mass (2) Structures of inorganic, organic, and biological molecules

More information

1) Introduction. 2) Types of Mass Analyzers - Overview. Assigned Reading: Nier mass spectrometry paper Extra Reading:

1) Introduction. 2) Types of Mass Analyzers - Overview. Assigned Reading: Nier mass spectrometry paper Extra Reading: Assigned Reading: Nier mass spectrometry paper Extra Reading: 1) Introduction a) We are now on to the heart of the mass spectrometer. How do you separate ions of different mass, so that you can detect/quantify

More information

Skoog/Holler/Crouch Chapter 19 Principles of Instrumental Analysis, 6th ed. CHAPTER 19

Skoog/Holler/Crouch Chapter 19 Principles of Instrumental Analysis, 6th ed. CHAPTER 19 Skoog/Holler/Crouch Chapter 19 Principles of Instrumental Analysis, 6th ed. Instructor s Manual CHAPTER 19 19-1. In a continuous wave NMR experiment, the intensity of the absorption signal is monitored

More information

Traps for Rare Isotopes

Traps for Rare Isotopes Traps for Rare Isotopes Georg Bollen National Superconducting Cyclotron Laboratory and Department for Physics and Astronomy, Michigan State University, East Lansing, MI, 48824, USA bollen@nscl.msu.edu

More information

Mass Analysers for LC MS

Mass Analysers for LC MS Mass Analysers for LC MS Filip Lemière, Dept of Chemistry, University of Antwerp, Belgium. Image courtesy of Bruker Daltonik GmbH Introduction The major technical challenge in LC MS is interfacing the

More information

11 SEPTEMBER This document consists of printed pages.

11 SEPTEMBER This document consists of printed pages. S 11 SEPTEMBER 2017 6 Write your name, centre number, index number and class in the spaces at the top of this page and on all work you hand in. Write in dark blue or black pen on both sides of the paper.

More information

Mini review for final

Mini review for final Mini review for final NB: This does not cover all material! (very) Rough outline Units, Conversions, Powers of 10, Prefixes Errors: Sig Figs, Error propagation Statistics: Normal distribution, t-test,

More information