Harris: Quantitative Chemical Analysis, Eight Edition

Size: px
Start display at page:

Download "Harris: Quantitative Chemical Analysis, Eight Edition"

Transcription

1 Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 21: MASS SPECTROMETRY

2 CHAPTER 21: Opener

3 21.0 Mass Spectrometry Mass Spectrometry provides information about 1) The elemental composition of samples (atomic mass spectrometry) 2) The structures of inorganic, organic, and biological molecules (molecular mass spectrometry) 3) Qualitative and quantitative composition of complex mixtures 4) The structure and composition of solid surfaces 5) Isotopic ratios of atoms in samples

4 21-1 What is Mass Spectrometry? The mass spectrum shows detector current as a function of m/z selected by the magnetic field Atomic Mass Spectrum

5 21-1 What is Mass Spectrometry? Base Peak : the most intense peak in a mass spectrum Molecular Mass Spectrum

6 Box 21-2 How ions of different masses are separated by a magnetic field - When an ion (m, +ze ) is accelerated through a potential difference, it aquires a kinetic energy equal to the electric potential difference : Kinetic energy (v = velocity) Potential energy - An ion with charge (+ze ) and velocity (v) traveling perpendicular to a magnetic field B experiences a force zevb that is perpendicular to both the velocity vector and the magnetic field vector. - This force deflects the ion through a circular path of radius r. - The centripetal force (mv 2 /r ) required to deflect the particle is provided by the magnetic field.

7 Box 21-2 How ions of different masses are separated by a magnetic field Equating velocities from two equations above gives - This equation tells us the radius of curvature of the path traveled by an ion with mass m and charge z. - The radius of curvature is fixed by the geometry of the hard ware (Fig. 21-2). Ions can be selected to reach the detector by adjusting the magnetic field B, or the accelerating voltage V. - Normally B is varied to select ions and V is fixed near 3,000 V.

8 21-1 What is Mass Spectrometry?

9 21-1 components of a mass spectrometer 1)Inlet: introduction of a micro amount of sample 2) Ion source: components are converted into gaseous ions 3) Mass analyzer: dispersion of components depending on the mass-to-charge ratio of analyte ions 4) Ion transducer: a transducer converts the beams of ions into an electrical signal. 5) Signal processor the signal is processed, stored in the memory of a computer and displayed. A mass spectrometer requires a vacuum system to maintain a low pressure (10-5 to 10-8 torr )

10 21-1 Resolving Power - The actual detector signals are not computer-generated bar graphs, but are like Fig If peaks are too close, they appear to be a single peak. - The higher the resolving power of a mass spectrometer, the better it is able to separate two peaks with similar mass.

11 21-1 *Resolving Power - Resolution is the smallest difference in m/z values that can be detected as separate peaks and should be reported with the m/z value where it is measured.

12 21-1 components of a mass spectrometer 1)Inlet: introduction of a micro amount of sample 2) Ion source: components are converted into gaseous ions 3) Mass analyzer: dispersion of components depending on the mass-to-charge ratio of analyte ions 4) Ion transducer: a transducer converts the beams of ions into an electrical signal. 5) Signal processor the signal is processed, stored in the memory of a computer and displayed. A mass spectrometer requires a vacuum system to maintain a low pressure (10-5 to 10-8 torr )

13 21-1 Ion Sources - The starting point for a mass spectrometeric analysis is the formation of gaseous analyte ions. - The scope and the utility of a mass spectrometric method is dictated by the ionization process. - The appearance of mass spectra for a given molecular species strongly depends on the method used for ion formation. - Compare the differences between ion sources for atomic mass and molecular mass spectrometers.

14 21-1 Ion Sources (Atomic Mass vs. Molecular Mass)

15 21-2 Ion Source (Gas Phase)-Electron Impact Ionization - Electrons are emitted from a heated tungsten or rhenium filament and accelerated by applying approximately 70 V between the filament and the anode. - The paths of the electrons and molecules are at right angles and intersect near the center of the source, where collision and ionization occur.

16 21-2 Ion Source (Gas Phase)-Electron Impact Ionization M + e - M +. + e - + e - 70 ev Molecular ion ~ 55eV 0.1 ev e - from filament e - from M - The primary product is singly charged positive ions (M +. ) formed when the energetic electrons approach molecules closely enough to cause them to lose electrons by electrostatic repulsion. - Electron-impact ionization is not very efficient, and only about one molecule in a million undergoes the primary reaction. In other words, some (~ 0.01 %) molecules (M) absorb as much as ev, which is enough for ionization.

17 21-2 Ion Source (Gas Phase)-Electron Impact Ionization M + e - M +. + e - + e - 70 ev Molecular ion ~ 55eV 0.1 ev - After ionization, M +. Usually has enough internal energy (~ 1eV) to break into fragments. There might be so little M +. that its peak is small in the mass spectrum.

18 21-2 Ion Source (Gas Phase)-Electron Impact Ionization - Base Peak: the most intense peak in a mass spectrum (M/Z=141 or M/Z = 157 in Fig ) - A computer search is commonly used to match the spectrum of an unknown to similar spectra in a library

19 21-2 Ion Source (Gas Phase)-Electron Impact Ionization M + e - M +. + e - + e - 70 ev Molecular ion ~ 55eV 0.1 ev - If you lower the kinetic energy of electrons in the ionization source, to say 20eV, there will be lower yield of ions and less fragmentation. - We customarily use 70eV because it gives reproducible fragmentation patterns that can be compared with library spectra.

20 21-2 Ion Source (Gas Phase)- Electron Impact Ionization - Electrons with an energy near 70eV almost exclusively create cationic molecular products. However, if the electron energy is lower, it is possible to form negative ions from molecules with a sufficient great electron affinity. Resonance capture Dissociative capture M + e - M -. < 0.1 ev XY + e - X + Y ev

21 21-2 Ion Source (Gas Phase)-Chemical Ionization - In chemical ionization, gaseous molecules of the sample are ionized by collision with ions (CH 5 + ) produced by electron bombardment of an excess of a reagent gas (CH 4 ). - One of the most common reagents is methane, which reacts with high energy electrons to give several ions such as CH 4 +, CH 3 + CH 5 +, C 2 H 5 + CH CH CH CH 4 + e CH + CH CH 4 4 CH CH C 2 + 2e H H CH3 + H 2 CH C C 2 2 CH + 5 H H M + M + M + M CH C C CH H H 4 + MH MH + (M -1) + M + + ( protonated molecule, most abundant in chem. ioniz.) + + ( proton transfer) ( hydride ( molecular ion) transfer)

22 21-2 Ion Source (Gas Phase)-Chemical Ionization - It is necessary to modify the electron beam ionization area used for electron ionization : a reagent gas pressure of about 1 Torr (1 mbar) while maintaining the pressure of the analyzer below 10-5 Torr. - Because of this large concentration difference, the high energy electron beam ( eV) reacts nearly exclusively with reagent molecules. -Ammonia or isobutane are used in place of CH 4 to reduce the fragmentation of MH +. These reagents bind H + more strongly than CH 4 does and impart less energy to MH + when the proton is transferred to M.

23 Comparison of Mass spectra for Pentobarbital from electron ionization vs. chemical ionization - Peaks in Fig provide clues about the structure of molecule. A computer search is commonly used to match the spectrum of unknown to similar spectra in a library. - A reasonable match of the experimental spectrum to one in the computer library is not proof of molecular structure it is just a clue

24 Copmarison of mass spectra of 1-decanol from electron ionization vs. chemical ionization Figure. Mass spectrum of 1-decanol a) electron ionization b) chemical ionization

25 21-2 Ion Source (Desorption Sources) - The gas phase ionizations are not applicable to nonvolatile or thermally unstable samples - A number of desorption ionization methods have been developed for dealing with this type of sample. - These methods have enabled mass spectra to be obtained for thermally delicate biochemical species and species having molecular masses of greater than 100,000 Da. - In desorption methods, energy in various forms is introduced into the solid or liquid sample in such a way to cause direct formation of gaseous ions. - Consequently, spectra are greatly simplified and often consist of only the molecular ion or the protonated molecular ion. - In most cases the exact mechanism of how ions are formed without fragmentation is not well understood.

26 21-2 Ion Source (Desorption Sources)- Matrix-Assisted Laser Desorption-Ionization (MALDI)

27 21-2 Ion Source (Desorption Sources)- Matrix-Assisted Laser Desorption-Ionization (MALDI) - MALDI is an ionization method that can be used to obtain accurate molecular mass about polar biopolymers (Mass = a few thousands ~ several hundred thousands Da.). - MALDI was first described nearly simultaneously in 1988 by two research groups, one German and the other Japan. - Analyte is mixed with an UV absorbing compound (matrix) and placed on a metal sample plate.. - A brief IR or UV pulse (600 ps) from a laser is directed onto the sample. - The matrix vaporizes and expands into the gas phase, carrying analyte along with it. - It is thought to involve absorption of the laser beam by the matrix, followed by transfer of energy from the matrix to the analyte. (The mechanism is not yet clear) - The analyte can be protonated, be deprotonated, or form adducts (i.e., cation in Fig.) before entering the TOF (Time Of Flight ) analyzer.

28 Figure. MALDI TOF spectrum from nicotinic acid matrix irradiated with a 266-nm laser beam - Analyte : a monoclonal antibody from a mouse (M.W.= ca.150,000 da). Note peaks for multiply charged ions, dimer and trimer species.

29 21-2 *Ion Sources : Hard vs. Soft - Ion sources are also classified as being hard sources or soft sources. 1) Hard ionization sources - They impart enough energy to analyte molecules to leave them in a highly excited energy state. - Relaxation then involves rupture of bonds, producing fragment ions that have m/z ratios less than that of molecular ion (M +. ). 2) Soft ionization sources - They cause little fragmentation and thus the mass spectrum often consists of the molecular ion (M +. ) peak and a only a few, if any, other peaks.

30 21-2 *Ion Sources : Hard vs. Soft Figure. Mass spectrum of 1-decanol a) a hard ionization source (electron impact) b) a soft ionization source (chemical ionization)

31 21-2 *Ion Sources : Hard vs. Soft

32 21-3 Reactions in an electron source - The low mass and high kinetic energy of the resulting electrons cause little increase in the translation energy of impacted molecules. - Instead, the molecules are left in a highly excited vibrational and rotational states. Subsequent relaxation usually takes place by extensive fragmentation, giving a large number of positive ions of various masses. - They are higher or lower mass than the molecular ion. The lower mass ions are called daughter ions.

33 - More often than not, the base peaks in electron impact spectra correspond to fragments such as daughter ions having a mass significantly less than the molecular mass of the original compound. Figure. Electron impact mass spectra. a) methylene chloride (MW=84) b) 1-pentanol (MW=88)

34 21-3 Reactions in an electron source Sometimes peaks occur at mass -to -charge ratios greater than that of the molecular ion. - Analyte : a monoclonal antibody from a mouse (M.W.= ca.150,000 da). - Note peaks for multiply charged ions, dimer and trimer species.

35 21-4. Types of Mass Spectrometers (Analyzers) 1) Magnetic Sector Analyzer (or Spectrometer) 2) Double-Focusing Spectrometer 3) Time-of-Fleight (TOF) Mass Spectrometer 4) Quadrupole Mass Spectrometer

36 ) Magnetic Sector Spectrometer Figure. Schematic of a magnetic sector spectrometer

37 ) Double-Focusing Spectrometer M + e - M +. + e - + e - 70 ev Molecular ion ~ 55eV 0.1 ev - Ions are created from the neutral molecule with a small spread of kinetic energies, depending on where in the ion source they were formed. So they are accelerated to different extents. - The spread of kinetic energies of ions leaving ion source causes a broadening of the beam reaching the transducer and thus a loss of resolution. - The resolving power of a mass spetrometer is limited by the variation in kinetic energy of ions emerging from the source, which is typically ~ 0.1 %. It limits the resolving power to ~ 1,000 (or a resolution of 0.1 at m/z 100)

38 ) Double-Focusing Spectrometer -The magnetic sector analyzer is called a single-focusing spectrometer. - In a double-focusing mass spectrometer, ions ejected from the source pass through an electric sector as well as a magnetic sector. - In an electric sector, ions with energies greater than or lower than averages are removed. - With both sectors in series, it is possible to achieve a resolving power of ~ 10 5.

39 ) Time-of-Fleight (TOF) Mass Spectrometer. - The ions produced are accelerated by an electric field pulse of 10 3 to 10 4 V. - The accelerated particles pass into a field-free drift tube (~ 1m). - Because all the ions entering the tube have the same kinetic energy, their velocity in the tube must vary inversely with their masses, with the lighter particles arriving at the detector earlier than the heavier ones.

40 ) Time-of-Fleight (TOF) Mass Spectrometer. - The ions produced are accelerated by an electric field pulse of 10 3 to 10 4 V. - The accelerated particles pass into a field-free drift tube (~ 1m). - Because all the ions entering the tube have the same kinetic energy, their velocity in the tube must vary inversely with their masses, with the lighter particles arriving at the detector earlier than the heavier ones.

41 ) Quadrupole Mass Spectrometer - Quadrupole mass spectrometers are usually less expensive more rugged than their magnetic sector counterparts. They are considerably more compact than magnetic sector analyzers (benchtop). They are most common mass analyzers in use today. - They offer the advantage of low scan times ( < 100 ms), which is particularly useful for real-time scanning of chromatographic peaks.

42 - The heart is the rour parralel cylindrical rods that serve as electrodes. - Ions are accelerated into the space between the rods by a potential difference of 5 to 10 V. Meanwhile, the ac and dc voltages on the rods are increased simultaneously while maintaining their ratio constant. - At any given moment, all of the ions except those having a certain m/z values strike the rods and converted to neutral molecules. Thus, only ions having a limited range of m/z values reach the transducer. - Typically quadrupole analyzers easily resolve ions that differ in mass by one unit. - They are sometimes referred to as a mass filter.

43 21-5. Ion transducers for Mass Spectrometry 1) Electron Multipliers - Several types of transducers are commercially available for mass spectrometers. - The electron multiplier is the transducer of choice for most routine experiments. - It is designed for collecting and converting positive ions into an electrical signal. Dynode: a very sensitive device in which electrons are emmitted from the first surface strikes a second surface, and so on. Figure. a) Discrete-dynode electron multiplier. b) Continuous-dynode electron multiplier.

44 21-5. Transducers for Mass Spectrometry 1) Electron Multipliers Figure. a) Discrete-dynode electron multiplier. - Each dynode is held at a successively higher voltage. - When energetic electrons or ions strike the Cu-Be surfaces of the cathode and the dynodes, burst of electrons are emitted. - The electrons are then attracted to the next dynode down the chain until, at the last dynode, a huge number of electrons appear for every ion that strikes the cathode. - Electron multipliers with up to twenty dynodes are available that typically provide a current gain of 10 7.

45 - Ions that strike the surface near the entrance eject electrons that are then attracted to higher-voltage points farther along the tube. These secondary electrons skip along the surface, ejecting more electrons with each impact. - This type typically have gains of 10 5, but in certain applications gains as high as 10 8 can be achieved Transducers for Mass Spectrometry b) Continuous-dynode electron multiplier - The device is shaped like a cornucopia ( 염소뿔 ) 뭉 is made of glass heavily doped with lead to give the material a small conductivity. - A voltage of 1.8 to 2 kv applied across the length of the transducer produces a voltage - gradient from one end to the other.

46 21-5. Ion transducers for Mass Spectrometry 1) Electron Multipliers - Electron multipliers are rugged and reliable and are capable of providing high current gains and nanosecond response times. - They can be placed directly behind the exit slit of a magnetic sector mass spectrometer ( next slide), because the ions reaching the transducer usually have enough kinetic energy to eject electrons from the first stage of the device. - If the ions have low-kinetic energy, then the ion beam exiting the mass analyzer is - accelerated to several thousand electron volts prior to striking the first stage.

47 21-4 Types of Mass Spectrometers (Analyzers) Discrete-dynode electron multiplier

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai 303351 Instrumental Analysis Mass Spectrometry Lecturer:! Somsak Sirichai Mass Spectrometry What is Mass spectrometry (MS)? An analytic method that employs ionization and mass analysis of compounds in

More information

Chemistry Instrumental Analysis Lecture 35. Chem 4631

Chemistry Instrumental Analysis Lecture 35. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 35 Principle components: Inlet Ion source Mass analyzer Ion transducer Pumps Signal processor Mass analyzers Quadrupole Time of Flight Double Focusing Ion

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?

More information

Chemistry Instrumental Analysis Lecture 34. Chem 4631

Chemistry Instrumental Analysis Lecture 34. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 34 From molecular to elemental analysis there are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry

More information

Chemistry 311: Topic 3 - Mass Spectrometry

Chemistry 311: Topic 3 - Mass Spectrometry Mass Spectroscopy: A technique used to measure the mass-to-charge ratio of molecules and atoms. Often characteristic ions produced by an induced unimolecular dissociation of a molecule are measured. These

More information

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n 20.2 Ion Sources electron ionization produces an M + ion and extensive fragmentation chemical ionization produces an M +, MH +, M +, or M - ion with minimal fragmentation MALDI uses laser ablation to transfer

More information

Molecular Mass Spectrometry

Molecular Mass Spectrometry Molecular Mass Spectrometry Mass Spectrometry: capable of providing information about (1) Elemental composition of samples of matter: atomic mass (2) Structures of inorganic, organic, and biological molecules

More information

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass

More information

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511 524) (Harris, Chapt.

More information

Mass Spectrometry. What is Mass Spectrometry?

Mass Spectrometry. What is Mass Spectrometry? Mass Spectrometry What is Mass Spectrometry? Mass Spectrometry (MS): The generation of gaseous ions from a sample, separation of these ions by mass-to-charge ratio, and measurement of relative abundance

More information

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511-524) (Harris, Chapt.

More information

MS Goals and Applications. MS Goals and Applications

MS Goals and Applications. MS Goals and Applications MS Goals and Applications 3 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis

More information

History of Mass spectroscopy. Mass Spectroscopy. Introduction... Uses of Mass Spec. Where are Mass Spectrometers Used? Mass Spectroscopy Units

History of Mass spectroscopy. Mass Spectroscopy. Introduction... Uses of Mass Spec. Where are Mass Spectrometers Used? Mass Spectroscopy Units History of Mass spectroscopy Mass Spectroscopy CHEM 466 Upali Siriwardane Marilyn Cox Jim Plamer http://www.chemistry.ohiostate.edu/~allen/587%20w04/587%20w0 4%20130-136%20std.pdf http://www.cem.msu.edu/~cem333/week1

More information

Mass Spectrometry. Electron Ionization and Chemical Ionization

Mass Spectrometry. Electron Ionization and Chemical Ionization Mass Spectrometry Electron Ionization and Chemical Ionization Mass Spectrometer All Instruments Have: 1. Sample Inlet 2. Ion Source 3. Mass Analyzer 4. Detector 5. Data System http://www.asms.org Ionization

More information

Lecture 15: Introduction to mass spectrometry-i

Lecture 15: Introduction to mass spectrometry-i Lecture 15: Introduction to mass spectrometry-i Mass spectrometry (MS) is an analytical technique that measures the mass/charge ratio of charged particles in vacuum. Mass spectrometry can determine masse/charge

More information

Molecular Mass Spectrometry

Molecular Mass Spectrometry Molecular Mass Spectrometry Mass Spectrometry: capable of providing information about (1) Elemental composition of samples of matter: atomic mass (2) Structures of inorganic, organic, and biological molecules

More information

ICPMS Doherty Lecture 1

ICPMS Doherty Lecture 1 ICPMS Doherty Lecture 1 Mass Spectrometry This material provides some background on how to measure isotope abundances by means of mass spectrometry. Mass spectrometers create and separate ionized atoms

More information

MS Goals and Applications. MS Goals and Applications

MS Goals and Applications. MS Goals and Applications MS Goals and Applications 1 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis

More information

Ionization Techniques Part IV

Ionization Techniques Part IV Ionization Techniques Part IV CU- Boulder CHEM 5181 Mass Spectrometry & Chromatography Presented by Prof. Jose L. Jimenez High Vacuum MS Interpretation Lectures Sample Inlet Ion Source Mass Analyzer Detector

More information

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein. Mass Analyzers After the production of ions in ion sources, the next critical step in mass spectrometry is to separate these gas phase ions according to their mass-to-charge ratio (m/z). Ions are extracted

More information

TANDEM MASS SPECTROSCOPY

TANDEM MASS SPECTROSCOPY TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER

More information

M M e M M H M M H. Ion Sources

M M e M M H M M H. Ion Sources Ion Sources Overview of Various Ion Sources After introducing samples into a mass spectrometer, the next important step is the conversion of neutral molecules or compounds to gas phase ions. The ions could

More information

Secondary Ion Mass Spectroscopy (SIMS)

Secondary Ion Mass Spectroscopy (SIMS) Secondary Ion Mass Spectroscopy (SIMS) Analyzing Inorganic Solids * = under special conditions ** = semiconductors only + = limited number of elements or groups Analyzing Organic Solids * = under special

More information

Mass Spectrometry in MCAL

Mass Spectrometry in MCAL Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring

More information

Mass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+.

Mass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+. Mass spectrometry is the study of systems causing the formation of gaseous ions, with or without fragmentation, which are then characteried by their mass to charge ratios (m/) and relative abundances.

More information

MASS SPECTROMETRY. Topics

MASS SPECTROMETRY. Topics MASS SPECTROMETRY MALDI-TOF AND ESI-MS Topics Principle of Mass Spectrometry MALDI-TOF Determination of Mw of Proteins Structural Information by MS: Primary Sequence of a Protein 1 A. Principles Ionization:

More information

Mass spectrometry.

Mass spectrometry. Mass spectrometry Mass spectrometry provides qualitative and quantitative information about the atomic and molecular composition of inorganic and organic materials. The mass spectrometer produces charged

More information

MS/MS .LQGVRI0606([SHULPHQWV

MS/MS .LQGVRI0606([SHULPHQWV 0DVV6SHFWURPHWHUV Tandem Mass Spectrometry (MS/MS) :KDWLV0606" Mass spectrometers are commonly combined with separation devices such as gas chromatographs (GC) and liquid chromatographs (LC). The GC or

More information

3 Use of Mass Spectra to Obtain Structural Information

3 Use of Mass Spectra to Obtain Structural Information 3 Use of Mass Spectra to Obtain Structural Information 1 Mass Spectrometry One of the most sensitive and versatile analytical tools More sensitive than other spectroscopic methods (e.g. IR spectroscopy)

More information

Introduction to GC/MS

Introduction to GC/MS Why Mass Spectrometry? Introduction to GC/MS A powerful analytical technique used to: 1.Identify unknown compounds 2. Quantify known materials down to trace levels 3. Elucidate the structure of molecules

More information

RECOMMENDATIONS FOR NOMENCLATURE OF MASS SPECTROMETRY

RECOMMENDATIONS FOR NOMENCLATURE OF MASS SPECTROMETRY international UNION OF PURE AND APPLIED CHEMISTRY ANALYTICAL CHEMISTRY DIVISION COMMISSION ON ANALYTICAL NOMENCLATURE RECOMMENDATIONS FOR NOMENCLATURE OF MASS SPECTROMETRY RULES APPROVED 1973 LONDON BUTTER

More information

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps ION ANALYZERS MASS ANALYSER sample Vacuum pumps Mass analysers - separate the ions according to their mass-to-charge ratio MASS ANALYSER Separate the ions according to their mass-to-charge ratio in space

More information

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Module - 6 Spectroscopic Techniques Lecture - 6 Atomic Spectroscopy

More information

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

(Refer Slide Time 00:09) (Refer Slide Time 00:13) (Refer Slide Time 00:09) Mass Spectrometry Based Proteomics Professor Sanjeeva Srivastava Department of Biosciences and Bioengineering Indian Institute of Technology, Bombay Mod 02 Lecture Number 09 (Refer

More information

1. The range of frequencies that a measurement is sensitive to is called the frequency

1. The range of frequencies that a measurement is sensitive to is called the frequency CHEM 3 Name Exam 1 Fall 014 Complete these problems on separate paper and staple it to this sheet when you are finished. Please initial each sheet as well. Clearly mark your answers. YOU MUST SHOW YOUR

More information

Mass Spectrometry. General Principles

Mass Spectrometry. General Principles General Principles Mass Spectrometer: Converts molecules to ions Separates ions (usually positively charged) on the basis of their mass/charge (m/z) ratio Quantifies how many units of each ion are formed

More information

Mass spectrometry and elemental analysis

Mass spectrometry and elemental analysis Mass spectrometry and elemental analysis A schematic representation of a single-focusing mass spectrometer with an electron-impact (EI) ionization source. M: + e _ M +. + 2e _ Ionization and fragmentation

More information

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS Back to Basics Section D: Ion Optics CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS TABLE OF CONTENTS QuickGuide...413 Summary...415 Introduction...417 The physical basis of orthogonal TOF....... 419 Pulsedmainbeamsofions...421

More information

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS) Mass Analyzers

More information

Welcome to Organic Chemistry II

Welcome to Organic Chemistry II Welcome to Organic Chemistry II Erika Bryant, Ph.D. erika.bryant@hccs.edu Class Syllabus 3 CHAPTER 12: STRUCTURE DETERMINATION 4 What is this solution Soda Tea Coffee??? 5 What is this solution Soda Tea

More information

CHAPTER D3 TOF ION OPTICS

CHAPTER D3 TOF ION OPTICS Back to Basics Section D: Ion Optics CHAPTER D3 TOF ION OPTICS TABLE OF CONTENTS QuickGuide...399 Summary...401 Background...403 EquationsofMotionofIons...403 Resolution...405 Reflectron...407 Comparison

More information

2. Separate the ions based on their mass to charge (m/e) ratio. 3. Measure the relative abundance of the ions that are produced

2. Separate the ions based on their mass to charge (m/e) ratio. 3. Measure the relative abundance of the ions that are produced I. Mass spectrometry: capable of providing both quantitative and qualitative information about samples as small as 100 pg (!) and with molar masses in the 10 4-10 5 kdalton range A. The mass spectrometer

More information

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency. Light We can use different terms to describe light: Color Wavelength Frequency Light is composed of electromagnetic waves that travel through some medium. The properties of the medium determine how light

More information

CHAPTER A2 LASER DESORPTION IONIZATION AND MALDI

CHAPTER A2 LASER DESORPTION IONIZATION AND MALDI Back to Basics Section A: Ionization Processes CHAPTER A2 LASER DESORPTION IONIZATION AND MALDI TABLE OF CONTENTS Quick Guide...27 Summary...29 The Ionization Process...31 Other Considerations on Laser

More information

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Chemistry Instrumental Analysis Lecture 37. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 37 Most analytes separated by HPLC are thermally stable and non-volatile (liquids) (unlike in GC) so not ionized easily by EI or CI techniques. MS must be at

More information

This is the total charge on an ion divided by the elementary charge (e).

This is the total charge on an ion divided by the elementary charge (e). 12.2 Fundamentals and general terms Accelerator mass spectrometry An ultra-sensitive technique using tandem accelerators employed mainly for the study of long-lived radioisotopes, and stable isotopes at

More information

Introduction to the Q Trap LC/MS/MS System

Introduction to the Q Trap LC/MS/MS System www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap LC/MS/MS Introduction to the Q Trap LC/MS/MS System The Q Trap

More information

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #22 Mass Spectrometry: Chemical Ionization (Skoog,) (Harris, Chapt.) Mercer/Goodwill CEE 772 #22

More information

AQA Chemistry A-Level : Atomic Structure

AQA Chemistry A-Level : Atomic Structure AQA Chemistry A-Level 3.1.1: Atomic Structure Detailed Notes 3.1.1.1 - Fundamental Particles The model for atomic structure has evolved over time as knowledge and scientific understanding changes. Plum

More information

Analysis of Polar Metabolites using Mass Spectrometry

Analysis of Polar Metabolites using Mass Spectrometry Analysis of Polar Metabolites using Mass Spectrometry TransMed Course: Basics in Clinical Proteomics and Metabolomics. Oct 10-19, 2012 dd.mm.yyyy Vidya Velagapudi, Ph.D, Adjunct Professor Head of the Metabolomics

More information

Chapter 4 Scintillation Detectors

Chapter 4 Scintillation Detectors Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 4-1 4.1. Basic principle of the scintillator Chapter 4 Scintillation Detectors Scintillator Light sensor Ionizing radiation Light (visible,

More information

Mass Spectrometry. Introduction EI-MS and CI-MS Molecular mass & formulas Principles of fragmentation Fragmentation patterns Isotopic effects

Mass Spectrometry. Introduction EI-MS and CI-MS Molecular mass & formulas Principles of fragmentation Fragmentation patterns Isotopic effects Mass Spectrometry Introduction EI-MS and CI-MS Molecular mass & formulas Principles of fragmentation Fragmentation patterns Isotopic effects 1 Introduction to MS Mass spectrometry is the method of analysis

More information

Propose a structure for an alcohol, C4H10O, that has the following

Propose a structure for an alcohol, C4H10O, that has the following Propose a structure for an alcohol, C4H10O, that has the following 13CNMR spectral data: Broadband _ decoupled 13CNMR: 19.0, 31.7, 69.5 б DEPT _90: 31.7 б DEPT _ 135: positive peak at 19.0 & 31.7 б, negative

More information

12. Structure Determination: Mass Spectrometry and Infrared Spectroscopy

12. Structure Determination: Mass Spectrometry and Infrared Spectroscopy 12. Structure Determination: Mass Spectrometry and Infrared Spectroscopy Determining the Structure of an Organic Compound The analysis of the outcome of a reaction requires that we know the full structure

More information

15.04.jpg. Mass spectrometry. Electron impact Mass spectrometry

15.04.jpg. Mass spectrometry. Electron impact Mass spectrometry Mass spectrometry Electron impact Mass spectrometry 70 ev = 1614 kcal/mol - contrast with energy from IR (1-10 kcal/mol) or NMR (0.2 cal/mol) - typical C-C bond = 100 kcal/mol Point: lots of energy in

More information

MASS SPECTRA measure a compound s Mol. Wt. This ionization type is called: electron impact MS

MASS SPECTRA measure a compound s Mol. Wt. This ionization type is called: electron impact MS MASS SPECTRA measure a compound s Mol. Wt. p. 213 M + Molecule e - Molecule + 2 e - + + Mole cule + + Mol ecule IONIZATION CHAMBER repellor plate accelerating plates variable field magnet + Mo + lecule

More information

Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. H i

Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. H i Gel Permeation Chromatography (GPC) : Size Exclusion Chromatography GPC : 1. Chromatogram (V R vs H) H i Detector response Baseline N i M i 130 135 140 145 150 155 160 165 Elution volume (V R ) (counts)

More information

Secondary Ion Mass Spectrometry (SIMS)

Secondary Ion Mass Spectrometry (SIMS) CHEM53200: Lecture 10 Secondary Ion Mass Spectrometry (SIMS) Major reference: Surface Analysis Edited by J. C. Vickerman (1997). 1 Primary particles may be: Secondary particles can be e s, neutral species

More information

( 1+ A) 2 cos2 θ Incident Ion Techniques for Surface Composition Analysis Ion Scattering Spectroscopy (ISS)

( 1+ A) 2 cos2 θ Incident Ion Techniques for Surface Composition Analysis Ion Scattering Spectroscopy (ISS) 5.16 Incident Ion Techniques for Surface Composition Analysis 5.16.1 Ion Scattering Spectroscopy (ISS) At moderate kinetic energies (few hundred ev to few kev) ion scattered from a surface in simple kinematic

More information

Types of Analyzers: Quadrupole: mass filter -part1

Types of Analyzers: Quadrupole: mass filter -part1 16 Types of Analyzers: Sector or double focusing: magnetic and electric Time-of-flight (TOF) Quadrupole (mass filter) Linear ion trap Quadrupole Ion Trap (3D trap) FTICR fourier transform ion cyclotron

More information

LC-MS Based Metabolomics

LC-MS Based Metabolomics LC-MS Based Metabolomics Analysing the METABOLOME 1. Metabolite Extraction 2. Metabolite detection (with or without separation) 3. Data analysis Metabolite Detection GC-MS: Naturally volatile or made volatile

More information

Chemistry Instrumental Analysis Lecture 17. Chem 4631

Chemistry Instrumental Analysis Lecture 17. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 17 Introduction to Optical Atomic Spectrometry From molecular to elemental analysis there are three major techniques used for elemental analysis: Optical spectrometry

More information

EXPERIMENT 2-6. e/m OF THE ELECTRON GENERAL DISCUSSION

EXPERIMENT 2-6. e/m OF THE ELECTRON GENERAL DISCUSSION Columbia Physics: Lab -6 (ver. 10) 1 EXPERMENT -6 e/m OF THE ELECTRON GENERAL DSCUSSON The "discovery" of the electron by J. J. Thomson in 1897 refers to the experiment in which it was shown that "cathode

More information

EEE4106Z Radiation Interactions & Detection

EEE4106Z Radiation Interactions & Detection EEE4106Z Radiation Interactions & Detection 2. Radiation Detection Dr. Steve Peterson 5.14 RW James Department of Physics University of Cape Town steve.peterson@uct.ac.za May 06, 2015 EEE4106Z :: Radiation

More information

An ion source performs the following two functions:

An ion source performs the following two functions: Ionization The Ion Source An ion source performs the following two functions: 1) converts sample atoms or molecules to ionized particles (ions) in the gas phase (sometimes the task of introducing the atoms

More information

Worksheet 2.1. Chapter 2: Atomic structure glossary

Worksheet 2.1. Chapter 2: Atomic structure glossary Worksheet 2.1 Chapter 2: Atomic structure glossary Acceleration (in a mass spectrometer) The stage where the positive ions are attracted to negatively charged plates. Alpha decay The emission of an alpha

More information

Auxiliary Techniques Soft ionization methods

Auxiliary Techniques Soft ionization methods Auxiliary Techniques The limitations of the structural information in the normal mass spectrum can be partly offset by special mass-spectral techniques. Although a complete description of these is beyond

More information

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS Mass Spectrometry Hyphenated Techniques GC-MS LC-MS and MS-MS Reasons for Using Chromatography with MS Mixture analysis by MS alone is difficult Fragmentation from ionization (EI or CI) Fragments from

More information

Laser Dissociation of Protonated PAHs

Laser Dissociation of Protonated PAHs 100 Chapter 5 Laser Dissociation of Protonated PAHs 5.1 Experiments The photodissociation experiments were performed with protonated PAHs using different laser sources. The calculations from Chapter 3

More information

Mass Spectrometry (MS)

Mass Spectrometry (MS) Mass Spectrometry (MS) Alternative names: Mass spectrometric (selective) detector (MSD) Spectrometry - methods based on interaction of matter and radiation Mass spectrometry - method based on formation

More information

Introduction to Elemental Analysis

Introduction to Elemental Analysis Introduction to Elemental Analysis A dust storm charges across the Sahara desert, whipping up sand and blowing it west, towards the ocean. Meanwhile, in the middle of the Atlantic Ocean, a team of scientist

More information

4. How can fragmentation be useful in identifying compounds? Permits identification of branching not observed in soft ionization.

4. How can fragmentation be useful in identifying compounds? Permits identification of branching not observed in soft ionization. Homework 9: Chapters 20-21 Assigned 12 April; Due 17 April 2006; Quiz on 19 April 2006 Chap. 20 (Molecular Mass Spectroscopy) Chap. 21 (Surface Analysis) 1. What are the types of ion sources in molecular

More information

Mass Spectrometry: Introduction

Mass Spectrometry: Introduction Mass Spectrometry: Introduction Chem 8361/4361: Interpretation of Organic Spectra 2009 Andrew Harned & Regents of the University of Minnesota Varying More Mass Spectrometry NOT part of electromagnetic

More information

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 DEPARTMENTS OF CHEMISTRY AND BIOCHEMISTRY GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 Mass Analysers Shabaz Mohammed October 20, 2015 High Vacuum System Turbo pumps Diffusion pumps Rough pumps Rotary

More information

Questions on Electric Fields

Questions on Electric Fields Questions on Electric Fields 1. The diagram shows a positively charged oil drop held at rest between two parallel conducting plates A and B. Oil drop A B 2.50 cm The oil drop has a mass 9.79 x 10 15 kg.

More information

Mass spectrometry gas phase transfer and instrumentation

Mass spectrometry gas phase transfer and instrumentation Objectives of the Lecture spectrometry gas phase transfer and instrumentation Matt Renfrow January 15, 2014 1. Make ions 2. Separate/Analyze 3. Detect ions 4. What is mass resolution and mass accuracy?

More information

CHM 424 EXAM 4 CRIB - COVER PAGE FALL

CHM 424 EXAM 4 CRIB - COVER PAGE FALL CHM 44 EXAM 4 CRIB - COVER PAGE FALL 007 There are six numbered pages with five questions. Answer the questions on the exam. Exams done in ink are eligible for regrade, those done in pencil will not be

More information

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk Previously: Resolution and other basics MALDI Electrospray 40 Lecture 2 Mass analysers Detectors

More information

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY Chapter 23. Gas Chromatography What did they eat in the year 1,000? GC of Cholesterol and other lipids extracted from

More information

Chapter 7. Mass Spectrometry

Chapter 7. Mass Spectrometry Chapter 7 Mass Spectrometry 1 Advantages of Atomic MS over Atomic Optical Spectroscopy Detection limits up to three times better Simple spectra, unique and easily interpretable Ability to measure atomic

More information

ATOMIC STRUCTURE, ELECTRONS, AND PERIODICITY

ATOMIC STRUCTURE, ELECTRONS, AND PERIODICITY ATOMIC STRUCTURE, ELECTRONS, AND PERIODICITY All matter is made of atoms. There are a limited number of types of atoms; these are the elements. (EU 1.A) Development of Atomic Theory Atoms are so small

More information

Extrel Application Note

Extrel Application Note Extrel Application Note Real-Time Plasma Monitoring and Detection of Trace H 2 O and HF Species in an Argon Based Plasma Jian Wei, 575 Epsilon Drive, Pittsburgh, PA 15238. (Presented at the 191st Electrochemical

More information

Secondary ion mass spectrometry (SIMS)

Secondary ion mass spectrometry (SIMS) Secondary ion mass spectrometry (SIMS) ELEC-L3211 Postgraduate Course in Micro and Nanosciences Department of Micro and Nanosciences Personal motivation and experience on SIMS Offers the possibility to

More information

Monday (March 28)- Mass spectrometry Tuesday (March 29)- Experiment 6: Separa>ons ICP-MS calcula>ons

Monday (March 28)- Mass spectrometry Tuesday (March 29)- Experiment 6: Separa>ons ICP-MS calcula>ons LOOKING AHEAD Monday (March 28)- Mass spectrometry Tuesday (March 29)- Experiment 6: Separa>ons ICP-MS calcula>ons Thursday (March 31)- Poster crea>on? Hayes on travel Monday (April 4)- Chromatographic

More information

Skoog/Holler/Crouch Chapter 19 Principles of Instrumental Analysis, 6th ed. CHAPTER 19

Skoog/Holler/Crouch Chapter 19 Principles of Instrumental Analysis, 6th ed. CHAPTER 19 Skoog/Holler/Crouch Chapter 19 Principles of Instrumental Analysis, 6th ed. Instructor s Manual CHAPTER 19 19-1. In a continuous wave NMR experiment, the intensity of the absorption signal is monitored

More information

Quadrupole Mass Spectrometry Concepts. Mass spectrometers for residual gas analysis: Intermediate Level Users Guide

Quadrupole Mass Spectrometry Concepts. Mass spectrometers for residual gas analysis: Intermediate Level Users Guide Quadrupole Mass Spectrometry Concepts Mass spectrometers for residual gas analysis: Intermediate Level Users Guide What does Residual Gas Analysis allow us to do? RGA is the examination of the molecular

More information

sample was a solution that was evaporated in the spectrometer (such as with ESI-MS) ions such as H +, Na +, K +, or NH 4

sample was a solution that was evaporated in the spectrometer (such as with ESI-MS) ions such as H +, Na +, K +, or NH 4 Introduction to Spectroscopy V: Mass Spectrometry Basic Theory: Unlike other forms of spectroscopy used in structure elucidation of organic molecules mass spectrometry does not involve absorption/emission

More information

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez Mass Analyzers Ion Trap, FTICR, Orbitrap CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography Prof. Jose-Luis Jimenez Last Update: Oct. 014 Some slides from Dr. Joel Kimmel (007) MS Interpretation

More information

Lecture 22 Ion Beam Techniques

Lecture 22 Ion Beam Techniques Lecture 22 Ion Beam Techniques Schroder: Chapter 11.3 1/44 Announcements Homework 6/6: Will be online on later today. Due Wednesday June 6th at 10:00am. I will return it at the final exam (14 th June).

More information

Photoelectron spectroscopy Instrumentation. Nanomaterials characterization 2

Photoelectron spectroscopy Instrumentation. Nanomaterials characterization 2 Photoelectron spectroscopy Instrumentation Nanomaterials characterization 2 RNDr. Věra V Vodičkov ková,, PhD. Photoelectron Spectroscopy general scheme Impact of X-ray emitted from source to the sample

More information

Ion sources. Ionization and desorption methods

Ion sources. Ionization and desorption methods Ion sources Ionization and desorption methods 1 2 Processes in ion sources 3 Ionization/ desorption Ionization Desorption methods Electron impact ionization Chemical ionization Electro-spray ionisation

More information

ELECTROMAGNETIC WAVES

ELECTROMAGNETIC WAVES VISUAL PHYSICS ONLINE MODULE 7 NATURE OF LIGHT ELECTROMAGNETIC WAVES SPECTRA PRODUCED BY DISCHARGE TUBES CATHODE RAYS (electron beams) Streams of electrons (negatively charged particles) observed in vacuum

More information

A process whereby an electron is either removed from or added to the atom or molecule producing an ion in its ground state.

A process whereby an electron is either removed from or added to the atom or molecule producing an ion in its ground state. 12.3 Processes and techniques 12.3.1 Ionization nomenclature Adiabatic ionization A process whereby an electron is either removed from or added to the atom or molecule producing an ion in its ground state.

More information

Time-of-Flight Mass Analyzers

Time-of-Flight Mass Analyzers Time-of-Flight Mass Analyzers Jonathan Karty C613 lecture 1 March 6, 8 (Section 4. in Gross, pages 115-18) TOF Overview Time-of-flight (TOF) is the least complex mass analyzer in terms of its theory Ions

More information

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy Topic 2b: X-ray Fluorescence Spectrometry Text: Chapter 12 Rouessac (1 week) 4.0 X-ray Fluorescence Download, read and understand EPA method 6010C ICP-OES Winter 2009 Page 1 Atomic X-ray Spectrometry Fundamental

More information

Mass Spectrometry (MS)

Mass Spectrometry (MS) Mass Spectrometry (MS) MW Molecular formula Structural information GC-MS LC-MS To Do s Read Chapter 7, and complete the endof-chapter problem 7-4. Answer Keys are available in CHB204H MS Principles Molecule

More information

v = E B FXA 2008 UNIT G485 Module Magnetic Fields BQv = EQ THE MASS SPECTROMETER

v = E B FXA 2008 UNIT G485 Module Magnetic Fields BQv = EQ THE MASS SPECTROMETER UNIT G485 Module 1 5.1.2 Magnetic Fields 11 Thus, in order for the particle to suffer NO DEFLECTION and so exit the device at Y : From which : MAGNETIC FORCE UP = ELECTRIC FORCE DOWN BQv = EQ THE MASS

More information

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC INTRODUCTION Principle: LC/MS is a technique that combines physical separation capabilities of liquid

More information

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy Experiment VI Gamma Ray Spectroscopy 1. GAMMA RAY INTERACTIONS WITH MATTER In order for gammas to be detected, they must lose energy in the detector. Since gammas are electromagnetic radiation, we must

More information