Technical Note # TN-24 Extending the Capacity in Modern Spherical High Capacity Ion Traps (HCT)
|
|
- Henry Young
- 6 years ago
- Views:
Transcription
1 Bruker Daltonics Technical Note # TN-24 Extending the Capacity in Modern Spherical High Capacity Ion Traps (HCT) During the last decade, RF ion traps have become a workhorse of the analytical laboratory [1]. The ion capacity is a key factor affecting ion trap performance. Recently, substantial increases to the ion capacity have been achieved through the introduction of two novel instruments: linear ion traps [2] and high capacity spherical traps (HCT) [3]. The HCT uses phase-coupled excitation at a non-linear resonance [4] resulting from the optimized high-order fields (HOF) introduced by the new HCT geometry. The rationale behind the increased ion capacity of the linear trap seems to be well understood; however, the mechanisms for the increased ion capacity of the HCT still requires a more detailed explanation. This technical note will analyze and describe the mechanisms for increasing the ion capacity in a spherical high capacity trap. Background Before understanding the mechanisms for increasing ion capacity it is important to understand the major factor that influences ion capacity and what is meant by ion capacity. In addition to the trapping and excitation fields in the ion trap influencing ion motion, the electric field produced by the ions themselves has a significant impact on their motion. The interaction of ions with other ions as they are being excited and or ejected is one of the key factors that influence the ion capacity of the ion trap. There are three main different ion capacities to be defined: spectrum limit, isolation limit, and the storage limit [5]. The storage limit is perhaps the simplest to understand; this limit is the maximum number of ions that can be put into the ion trap. Any further ions are either pushed out of the ion trap by Columbic forces or cause already trapped ions to be pushed out. This limit has no qualifications on the spectra quality (e.g. the resolution of peaks). The next isolation and spectrum limits depend on the definition of spectral quality. The isolation limit is the number of ions that can be trapped and still allow a given isolation efficiency. For instance, the isolation limit might be defined as the number of ions that can be trapped and still isolate within a 1 Da window. The isolation limit is affected by ion-ion coupling which prevents resonant selection and thereby achieving the desired isolation width. Typically the isolation limit is roughly ten fold lower than storage limit. Finally the spectrum limit is the number of ions that can be trapped while still achieving the desired resolution, speed and mass accuracy. This limit is the strictest in that as ions are being ejected, the ion-ion interaction can retard the resonant ejection, cause ion coupling, and cause mass shifts. This limit can be fifty to a few hundred times smaller than the isolation limit. In mass spectrometry we are mainly interested in the final mass spectrum and thus the spectrum limit is the most important limit on the experiment. For the remainder of this note ion capacity will be referring to the spectrum limit.
2 Recent developments in linear ion traps yielded a higher storage limit because of the geometrically larger volume. If the number of ions is kept constant, as shown in Figure 1, their density is decreased, or vice versa, if the density is kept constant, the total number of ions in the trap can be increased. However, from the definitions of spectrum and storage limit it can be seen that if all we did was to increase the storage limit it does not necessarily lead to an increase in the spectrum limit. As discussed above the storage limit is a result of trapping and storage processes where as the spectrum limit occurs during the ejection process. To increase the capacity without principally altering the volume of the system, the ejection process must be changed to limit ion-ion interaction during ion ejection. This is achieved in the HCT through the judicious use of non-linear fields and proper phase correlation of the resonant excitation process. Spherical and linear ion trap designs Spherical Ion Trap Linear Ion Trap Fig. 1: Spherical and linear ion traps have been developed in recent years. Ion trap theory To make a document like this understandable we will cover some ion trap fundaments that may be useful as a refresher; however, so as to keep the document as short as possible these descriptions will not go into details. If these topics seem to be too short or more information is required it may be advantageous to first read a more detailed description of the conventional ion trap operation, some helpful references follow. Ions in the quadrupole ion trap are trapped and stored via the quadrupole field that is formed by the voltage applied to the ring electrode. The Bruker ion traps also have a unique design that gives distortions to this field that are advantageous to the performance of the ion trap. These field distortions are most commonly termed higher order fields and the Bruker ion traps specifically exploit hexapole and octapole fields. The voltage that is applied to the ring electrode is in the radio frequency region and is therefore referred to as the main rf voltage. The ion trap can trap ions of a given m/z range. Theoretically the lowest m/z that can be trapped is defined by the amplitude of the main rf voltage and increasing the rf voltage will result in ions below an increasing m/z to be ejected from the ion trap. In general for this technical note the ions have already been trapped and kinetically cooled by collisions with the background gas (He). Cooled ions are moving near the center of the ion trap. When cooled in this manner the ions are referred to as the ion cloud. These ions can interact with each other, because like charges repel (columbic interaction) each other, ions will spread out against the focusing forces by the main rf to avoid each other. Ions of a given m/z value that are trapped in the ion trap have a unique oscillatory motion that is a function of the ion trap dimensions, the voltage that is applied to the ring electrode, and the ion s m/z. Every ion of a different m/z has a different oscillation frequency (secular frequency), thus if an ac voltage with the same frequency is applied across the endcaps the ion with that frequency is excited. This process is quite similar to pushing a child on a swing or increasing the motion of an already moving pendulum and is referred to as resonance excitation. If the amplitude of the applied voltage is large enough ions can be ejected out of the ion trap, termed resonance ejection. Keeping the frequency of the excitation ac voltage constant while increasing the amplitude of the main rf will drive ions of increasing m/z into resonance with this excitation ac voltage, causing the ejection of these ions. A mass spectrum is created by detecting the amount of resonantly ejected ions as a function of the currently applied main rf amplitude. For Bruker ion traps, the frequency of the exciting ac voltage applied to the endcaps is chosen to be exactly 1/3 of the frequency of the main rf applied to the ring electrode. This additional ac voltage is at a frequency which is also in resonance with the hexapole higher order fields. Because ions can interact with each other the ejection process is influenced by the number of the ions in the ion trap. This technical note will focus on how to exploit the ejection process of the ion trap so as to allow more ions to be in the into trap during the ejection process.
3 Model An improvement of the ion capacity can be achieved during resonance ejection at a specific non-linear resonance. [5] A model for this is illustrated in Figure 2 as a sequence of events during the acquisition of a mass spectrum. For this improvement to occur, the ions of interest are first excited (t 1 ) away from the other ions into a higher orbit by the dipolar resonance excitation. As shown in Figure 2a and Figure 2b, the ions of red m/z are resonantly excited to a higher orbit away from the other ions in the ion trap. Because these ions are at a higher orbit they now have a lower density than before resonant excitation. As a result of this separation both spatially and by their velocities these ions have a significantly reduced interaction with the ions that are not resonantly excited in the center of the ion trap. As the ions orbit slowly continues to increase (t 1 + t d ) they are precisely ejected utilizing the higher order fields that are greater in strength near the end caps, Figure 2c. This process would then begin again for the green m/z ions and so on, Figure 2d. As can be seen the elongated ion cloud that would be created in this process is analogous to the elongation of the ion cloud achieved in the linear ion trap. And perhaps even more important: this larger volume is populated by ions of one m/z only, thus further decreasing the effect of ion-ion coupling. To achieve this process it is necessary to have two distinct resonant activation events. In this case the first excitation process is achieved with a simple dipolar resonance excitation and the second is achieved utilizing excitation at the non-linear resonances that are inherent to the Bruker Daltonik s novel spherical ion trap geometry, the HCT. Experimental These experiments will explore the exact timing of the ejection of ions from the ion trap. To experimentally demonstrate the enhancements achieved with the new geometry of the HCT over our previous generations, experiments were performed with electrode geometries of a Bruker esquire 3000 and Bruker HCTultra. These experiments were conducted in the same instrument with only changes to the electrode configuration between sets of experiments. The supplemental AC voltage applied during activation and ejection of ions is derived from the fundamental RF voltage for all experiments. The scan rate of 26,000 Th/sec was used for both electrode configurations. ESI Tuning Mix (Agilent; Palo Alto, CA), which has ions covering the mass range from m/z , was used in these experiments. The exact timing of the ejection process was analyzed by varying the phase between the fundamental RF voltage and the resonance excitation voltage. The relative mass position, mass resolving power, and ion abundance were monitored, at each phase. Simulations were performed using SIMION 7.0 and custom calculation programs were written to understand excitation with dipolar fields coupled in the presence Non-linear resonance ejection improves ion capacity Fig. 2: Non-linear ion ejection is a two-step process: First (see 2a), ions are cooled and located in the trap center. Second (see 2b), during the scan, the RF is ramped and ions of a first mass (red) come into resonance with the AC. They are driven to a higher orbit, while the ions of higher mass (green, blue) remain cooled at the center. Third, the excited ions (red, see 2c) meet the non-linear resonance and are ejected instantly. The cooled ions in the center are unaffected by this. Then, the process repeats itself for higher masses (see 2d, green). The ion capacity is improved because right before the non-linear ejection the ions fill up a large volume (see 2b) which reduced space charge effects for these ions.
4 Non-linear resonance ejection at different phases alters ion ejection Ejection by non-linear resonance Ions excited by AC over entire time delay Ion motion (a) (b) (c) RF or m/z Fig.3: The process of Fig. 2 is explained on a time (=RF-amplitude, amu) axis. (a) corresponds to Fig.2 (a), (b) to Fig. 2 (b) and (c) to Fig. 2 (c). Ions are resonantly excited (red line) and ions motion increases. For a given phase correlation between the resonance ac and the main RF ions would be ejected quickly (blue line). At a different more optimal phase correlation between the resonance ac and the main RF ion ejection is delayed allowing ions to spend more time in the RF at higher orbit and a lower space charge conditions. of non-linear excitation fields. These data and simulations/ calculations provide insight into how the ion capacity can be influenced as a result of the ejection process, and also how it is improved with proper experimental design. Results and discussion An improvement of the ion capacity can be achieved during resonance ejection at a specific non-linear resonance. For this to occur, the ions of interest are first excited away from the other ions into a higher orbit by the dipolar resonance excitation. After the ions are excited to this higher orbit, they are later precisely ejected from the ion trap by the non-linear resonance. Importantly the non-linear resonance ejection occurs with high temporal precision. The decoupling of the initial excitation and ejection can be observed by altering the relative phase between the supplemental AC and the main RF. As shown in Figure 3, during the resonance ejection process ions experience a delay from the initial excitation (red line) process to the ejection of the ions. In conventional experiments found in the literature ions are ejected at higher amplitudes to minimize ion-ion interaction. In the experiments described here, as the relative phase is changed, a delay in the ejection process occurs from the normal delay (blue line) to the optimal delay (vertical green line). In the experiments this would be manifested as a shift in the entire mass spectrum; however, because the scan rate is known the relative ejection delay can be calculated. Simulations Simulations of the dipolar excitation process coupled with a non-linear resonance give an insight into what range ejection delay can be expected. Shown in Figure 4 are the results of dipolar excitation coupled with hexapole fields performed at two different phases. It is important to note that higher order fields are strongest near the endcaps of the ion trap and these simulations are predominately impacted by the dipolar fields until the ions are closer to the endcaps and higher order fields take effect. The plots show the arbitrary amplitude of the ions motion, where the amplitude of ~30 would result in ejection of the ions from the ion trap. At the displacement where higher order fields take effect and when phases of the dipolar excitation and hexapole fields are matching at this point in time (plot on the left) ions can be quickly excited by both the dipolar resonance as well as the non-linear resonance, thus ions are rapidly ejected from the ion trap (150μsec). When the relative phases at the same excitation level are appropriate (plot on the right) the ions are first quickly excited to an intermediate level. The amplitude of their motion is then retarded until their phase matches the phase of the higher order fields allowing for rapid excitation and ejection by the higher order fields. In the second simulation because the non-linear resonance and the dipole field are not in phase the two fields deconstructively interact. After the ions are rapidly excited by the dipolar field and when the ions approach a region where
5 Simulation of AC excitation and non-linear ejection process 30 (a) Ejection 30 (b) Ion motion (arb. units) 10 0 Amplitude Amplitude Time (μsec) Time (μsec) Fig. 4: Simulations of the AC ion excitation and non-linear ejection process, for two different settings of the relative phase between AC and RF. In these simulations ions are ejected with their amplitude of their motion reaches a value of 30 arb units (noted by the green line). In (a), the AC excitation and the non-linear ejection are in phase, ion are rapidly ejected and no distinction between two excitation processes is possible. In (b), they are out of phase, and the non-linear ejection is delayed by several hundred microseconds (ejection delay). This gives the ions the time needed to fill a large volume before being ejected by the non-linear resonance. Only macromotion is shown. the non-linear resonance is stronger the ions are slowed or de-excited by the non-linear resonance. Then once the ions become in phase with non-linear resonance they are quickly excited out of the ion trap. Even though the ions are de-excited by the non-linear resonance they do not cool back down to the center of the trap because the dipolar excitation is still present during this time. These simulations agree with the model of two distinct and coupled excitation events. Ejection experiments Figure 5 shows the ejection delays as a function of phase observed for the m/z 1522 ion ejected from the HCTultra and the esquire For both instruments the fastest ejection (smallest delay) occurs at the same phase ~225. The maximum delay for both instruments was roughly 170 μsec. This corresponds to roughly 44 oscillation cycles, so this delay is not a result of slight damping or columbic interaction. This plot illustrates the need for strong phase control during the excitation and ejection process. Without it mass spectra could be acquired at random phases which would result in a variety of ejection delays (wider mass peaks). In these instruments, the excitation RF is electronically derived from the main RF in these instruments, which results in an excellent phase control. The real question that arises is which phase is a good phase to utilize? From a reliability standpoint, operating in a region which has minimal change in the ejection delay at a given change in phase is ideal (horizontal tangent). This is desirable because if the region is flat, then for a given fluctuation in the ion motion there will be minimal impact on the MS peaks shape and position. For the esquire 3000 there is only one option for operation, which is at the earliest ejection point (~225, Figure 5). Whereas for the HCTultra there are two options, one at a delayed ejection (~120, Figure 5) and one at the earliest ejection point (~225, Figure 5). Figure 6 and Figure 7 show how the other figures of merit, abundance and resolving power, change as a function of relative phase for the esquire 3000 and for the HCTultra respectively. From a phase of 50 to 140 the resolving power of the esquire 3000 (Figure 6) is increasing reaching a maximum around 125 ; however, neither the signal intensity nor the ejection delay are reaching steady plateaus in this phase region. For the esquire 3000 the stable and robust operation would be at about 240. It is important to note that the absolute performance of the esquire 3000 in these experiments is poorer than standard instruments because it is operating at the faster scan rate of 26,000 Th/sec (The scan rate for the HCT). The general trends do not change at the standard scan rates for this instrument (13,000 Th/sec) but the absolute resolving power and intensities become greater. In the HCTultra both signal intensity and resolving power reach a maximum at 120. Both of these parameters also reach a plateau for about 15 showing further robustness at this operating point. The resolving power at the operating points for the HCT and the esquire 3000 were
6 Ejection delay as a function of phase Ejection Delay (µsec) μ sec) Time Delay ( HCT Esquire Phase (Degrees) between main RF and AC Fig. 5: Ejection delay as function of phase between AC and RF. Instruments are best operated at a point of zero derivative. For the esquire3000, one such point exists (at 225 ). For the high capacity ion trap (HCT), two points exist (at 120, 225 ). Operation at the additional point at 120 results in a large ejection delay and hence improved ion capacity and 1600 respectively. This resolving power occurs at an ICC value that is five fold greater for the HCT than the esquire This result shows that this ejection delay means a higher capacity with improved resolving power. In this delayed ejection process, ions are initially excited to a higher orbit separating them from other ions in the trap. The excited ions still pass through the remaining ion cloud at the lower orbit but at a higher speed. Thus, the excited ions cannot couple with other ions and the final excitation is performed in a lower charge density environment, which is the key for achieving the higher capacity ion trap. Conclusions A model and experimental evidence for achieving a higher capacity ion trap have been presented. This method involves decreasing the ion density of the ions being excited before ejecting them from the ion trap. For this to occur, the ions of interest are first excited away from the other ions into a higher orbit by the dipolar resonance excitation. After the ions are excited to this higher orbit, they are later precisely in-time ejected from the ion trap by the non-linear resonance. Importantly this ejection occurs with great temporal precision, which leads to rapid ejection with high resolving power at a high capacity. Simulations and calculations were in good agreement with this model. Experimentally the ejection timing can be controlled by setting the relative phase of dipolar excitation voltage and the main RF voltage. The ejection delay is advantageous because ions spend longer in the excitation field resulting in better resolving power. For the HCT there is an optimal phase and ejection delay where both the resolving power and ion abundance are at a maximum. This maximum has a plateau, which makes for stable robust operation. In these experiments, under the optimal operating conditions, the HCT showed a greater than ten fold increase in ion capacity over an esquire These results show that a precise design of the ion trap geometry together with the appropriate ejection methods lead to a robust high capacity ion trap that also has improvements in speed, resolving power, and mass accuracy.
7 Large ejection delay results in higher capacity, improved resolving power, and higher ion abundance esquire 3000 HCT 1.2 Operating point 1.2 Operating point Normalized Value Normalized Value Ph as e (Deg rees) Ph ase (Deg rees) Resolution (NL 3.5x10 3 ) Ejection Delay (NL170 µs) Signal Intensity (NL 4.5x10 7 ) Resolution (NL 3.5x10 3 ) Ejection Delay (NL170 µs) Signal Intensity (NL 4.5x10 7 ) Fig. 6: Ejection delay, mass resolution and ion abundance as function of the phase between AC and RF, for esquire3000 (operated at 26,000 Th/s) and high capacity trap (HCT). For HCT, resolution, delay (correlated to the ion capacity) and ion abundance all maximize at the same operating point (120 ), resulting in highest scan speeds, highest ion capacity and highest sensitivty all at the same time (NL: Normalization Level). References [1] March, R. E. and Todd, J. F. J., A historical Review of the Devlopment of the Quadrupole Ion Trap, in Quadrupole Ion Trap Mass Spectrometry, 2 nd ed., John Wiley & Sons, Inc., Hoboken, NJ, 2005, Chapter 1 [2] March, R. E. and Todd, J. F. J., Linear Quadrupole Ion Trap Mass Spectrometer, in Quadrupole Ion Trap Mass Spectrometry, 2 nd ed., John Wiley & Sons, Inc., Hoboken, NJ, 2005, Chapter 5 [3] C. Baessmann, A. Brekenfeld, G. Zurek, U. Schweiger-Hufnagel, M. Lubek, T. Ledertheil, R. Hartmer, M. Schubert, Proceedings of the 51 st American Society of Mass Spectrometry and Allied Topics, Montreal, PQ, June [4] March, R. E. and Todd, J. F. J., Recent Applications of ESI combined with QIT, in Quadrupole Ion Trap Mass Spectrometry, 2 nd ed., John Wiley & Sons, Inc., Hoboken, NJ, 2005, Chapter 8 [5] March, R. E. and Todd, J. F. J., Dynamics of Ion Trapping, in Quadrupole Ion Trap Mass Spectrometry, 2 nd ed., John Wiley & Sons, Inc., Hoboken, NJ, 2005, Chapter 3 Authors Desmond Kaplan, Andreas Brekenfeld, Ralf Hartmer, Thorsten Ledertheil, Christoph Gebhardt, Michael Schubert Bruker Daltonics Keywords Ion Capacity Ion Density Higher Order Fields Quadrupole Ion Trap Spherical Ion Trap Instrumentation & Software HCTultra esquire 3000
8 Bruker Daltonics is continually improving its products and reserves the right to change specifications without notice. Bruker Daltonics , TN-24 For research use only. Not for use in diagnostic procedures. Bruker Daltonik GmbH Bremen Germany Phone +49 (421) Fax +49 (421) Bruker Daltonics Inc. Billerica, MA USA Phone +1 (978) Fax +1 (978)
Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward
Bruker Daltonics EASY-nLC Tailored HPLC for nano-lc-ms Proteomics think forward Nano-HPLC World-Class Performance with a Small Footprint Bruker Daltonics presents a nano-lc system, perfectly integrated
More informationMass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.
Mass Analyzers After the production of ions in ion sources, the next critical step in mass spectrometry is to separate these gas phase ions according to their mass-to-charge ratio (m/z). Ions are extracted
More informationIon trap. 3D Quadrupole Trap. -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April
Ion trap 3D Quadrupole Trap -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April 011 1 The trajectories of the ions can be given: They form Mathieu differential equations: with r + Ion
More informationApplication Note LCMS-112 A Fully Automated Two-Step Procedure for Quality Control of Synthetic Peptides
Application Note LCMS-112 A Fully Automated Two-Step Procedure for Quality Control of Synthetic Peptides Abstract Here we describe a two-step QC procedure for synthetic peptides. In the first step, the
More informationThe Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results
The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results A. Kürten 1, J. Curtius 1 and S. Borrmann 1,2 1 Johannes Gutenberg-University Mainz, Germany 2 Max Planck
More informationTechnical Note # TN-35 amazon Ion Trap: New Performance Levels Meeting Demanding Analytical Requirements
Bruker Daltonics Technical Note # TN-35 amazon Ion Trap: New Performance Levels Meeting Demanding Analytical Requirements The amazon mass spectrometer platform The amazon TM series represents a new generation
More informationIon traps. Quadrupole (3D) traps. Linear traps
Ion traps Quadrupole (3D) traps Linear traps 3D-Ion traps A 3D-ion trap can be considered the tridimensional analogue of the linear quadrupole mass analyzer: B A C D The central ring electrode of the 3D-ion
More informationMass Spectrometry. What is Mass Spectrometry?
Mass Spectrometry What is Mass Spectrometry? Mass Spectrometry (MS): The generation of gaseous ions from a sample, separation of these ions by mass-to-charge ratio, and measurement of relative abundance
More informationCharacterization of petroleum samples via thermal analysis coupled to APCI FTMS
Characterization of petroleum samples via thermal analysis coupled to APCI FTMS Abstract Thermal analysis, by means of a thermo balance (TG), was coupled to an atmospheric pressure chemical ionization
More informationApplication Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry
Application Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry Introduction The oil industry requires detailed information on the composition of
More informationAt-a-Glance. Verapamil C 27 H 38 N 2 O 4 M+H + =
Application ote #LC-MS 80 Simultaneous Quantitative and Qualitative Measurements in a Single Workflow to Increase Productivity in Primary Drug Metabolism Investigations At-a-Glance n Increased productivity
More informationIntroduction to the Q Trap LC/MS/MS System
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap LC/MS/MS Introduction to the Q Trap LC/MS/MS System The Q Trap
More informationQuadrupole Storage Mass Spectrometry
Quadrupole Storage Mass Spectrometry RAYMOND E. MARCH AND RICHARD J. HUGHES Trent University Peterborough, Ontario, Canada with a historical review by John F. J. Tbdd University of Kent Canterbury, Kent,
More informationChemical Analysis. PIONA+ Analyzer. Characterization of Engine Fuels by Hydrocarbon Group Type. Gas Chromatography. think forward
Chemical Analysis PIONA+ Analyzer Characterization of Engine Fuels by Hydrocarbon Group Type think forward Gas Chromatography Bruker 450 PIONA+ Analyzer Bruker s innovative PIONA+ TM Analyzer provides
More informationThe development of algebraic methods to compute
Ion Energy in Quadrupole Mass Spectrometry Vladimir Baranov MDS SCIEX, Concord, Ontario, Canada Application of an analytical solution of the Mathieu equation in conjunction with algebraic presentation
More informationMass Selective Ejection by Axial Resonant Excitation from a Linear Ion Trap
Mass Selective Ejection by Axial Resonant Excitation from a Linear Ion Trap Yuichiro Hashimoto, Hideki Hasegawa, Takashi Baba, and Izumi Waki Central Research Laboratory, Hitachi, Ld., Tokyo, Japan We
More informationMass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez
Mass Analyzers Ion Trap, FTICR, Orbitrap CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography Prof. Jose-Luis Jimenez Last Update: Oct. 014 Some slides from Dr. Joel Kimmel (007) MS Interpretation
More informationChemistry Instrumental Analysis Lecture 35. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 35 Principle components: Inlet Ion source Mass analyzer Ion transducer Pumps Signal processor Mass analyzers Quadrupole Time of Flight Double Focusing Ion
More informationThermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3
Second-Generation Electron Transfer Dissociation (ETD) on the Thermo Scientific Orbitrap Fusion Mass Spectrometer with Improved Functionality, Increased Speed, Christopher Mullen, 1 Lee Earley, 1 Jean-Jacques
More informationTranslational Biomarker Core
Translational Biomarker Core Instrumentation Thermo Scientific TSQ Quantum Triple Quadrupole Mass Spectrometers. There are two TSQ Quantum Ultra AM instruments available in the TBC. The TSQ Quantum Ultra
More information(Refer Slide Time 00:09) (Refer Slide Time 00:13)
(Refer Slide Time 00:09) Mass Spectrometry Based Proteomics Professor Sanjeeva Srivastava Department of Biosciences and Bioengineering Indian Institute of Technology, Bombay Mod 02 Lecture Number 09 (Refer
More informationTypes of Analyzers: Quadrupole: mass filter -part1
16 Types of Analyzers: Sector or double focusing: magnetic and electric Time-of-flight (TOF) Quadrupole (mass filter) Linear ion trap Quadrupole Ion Trap (3D trap) FTICR fourier transform ion cyclotron
More informationDesigned for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS
Designed for Accuracy High resolution quantitative proteomics Innovation with Integrity LC-MS Setting New Standards in Accuracy The development of mass spectrometry based proteomics approaches has dramatically
More informationChemical Analysis. Low Level Oxygenates Analyzer. Trace Analysis of Oxygenates in Hydrocarbon Matrices. Gas Chromatography.
Chemical Analysis Low Level Oxygenates Analyzer Trace Analysis of Oxygenates in Hydrocarbon Matrices think forward Gas Chromatography The determination of sub to high ppm levels of ethers, alcohols, aldehydes
More informationprofileanalysis Innovation with Integrity Quickly pinpointing and identifying potential biomarkers in Proteomics and Metabolomics research
profileanalysis Quickly pinpointing and identifying potential biomarkers in Proteomics and Metabolomics research Innovation with Integrity Omics Research Biomarker Discovery Made Easy by ProfileAnalysis
More informationThree dimensional Paul traps may benefit from. Linear Quadrupoles with Added Hexapole Fields
Linear Quadrupoles with Added Hexapole Fields Nikolai Konenkov Department of General Physics, Ryazan State Pedagogical University, Ryazan, Russia Frank Londry Pan Galactic Scientific, Omemee, Ontario,
More informationGRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2
DEPARTMENTS OF CHEMISTRY AND BIOCHEMISTRY GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 Mass Analysers Shabaz Mohammed October 20, 2015 High Vacuum System Turbo pumps Diffusion pumps Rough pumps Rotary
More informationProven robustness for large-scale metabolomics studies using the Phenomics Workhorse
Proven robustness for large-scale metabolomics studies using the Phenomics Workhorse HRAM LC-MS profiling of > 1 urine samples demonstrates the outstanding analytical skills of the impact II Introduction
More informationCHAPTER D3 TOF ION OPTICS
Back to Basics Section D: Ion Optics CHAPTER D3 TOF ION OPTICS TABLE OF CONTENTS QuickGuide...399 Summary...401 Background...403 EquationsofMotionofIons...403 Resolution...405 Reflectron...407 Comparison
More information11/17/10. Chapter 14. Oscillations. Chapter 14. Oscillations Topics: Simple Harmonic Motion. Simple Harmonic Motion
11/17/10 Chapter 14. Oscillations This striking computergenerated image demonstrates an important type of motion: oscillatory motion. Examples of oscillatory motion include a car bouncing up and down,
More informationPesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS
PesticideScreener Comprehensive Pesticide Screening and Quantitation Innovation with Integrity UHR-TOF MS The Challenge of Comprehensive Pesticide Residue Analysis The use of pesticides to reduce crop
More information1 Pushing your Friend on a Swing
Massachusetts Institute of Technology MITES 017 Physics III Lecture 05: Driven Oscillations In these notes, we derive the properties of both an undamped and damped harmonic oscillator under the influence
More informationPractical Quadrupole Theory: Peak Shapes at Various Ion Energies
Practical Quadrupole Theory: Peak Shapes at Various Ion Energies Randall E. Pedder ABB Inc., Analytical-QMS Extrel Quadrupole Mass Spectrometry, 575 Epsilon Drive, Pittsburgh, PA 15238 (Poster presented
More informationMM 2. Innovation with Integrity. The new Generation of Mobile Mass Spectrometers. Defence CBRNE
MM 2 The new Generation of Mobile Mass Spectrometers Innovation with Integrity Defence CBRNE Picture by: MOWAG GmbH Mobile CWA Detection and Identification Low weight and high performance detector Introducing
More informationMASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps
ION ANALYZERS MASS ANALYSER sample Vacuum pumps Mass analysers - separate the ions according to their mass-to-charge ratio MASS ANALYSER Separate the ions according to their mass-to-charge ratio in space
More informationPhysics Mechanics. Lecture 32 Oscillations II
Physics 170 - Mechanics Lecture 32 Oscillations II Gravitational Potential Energy A plot of the gravitational potential energy U g looks like this: Energy Conservation Total mechanical energy of an object
More informationSSSC Discovery Series NMR2 Multidimensional NMR Spectroscopy
SSSC Discovery Series NMR2 Multidimensional NMR Spectroscopy Topics: 1. Some Common Experiments 2. Anatomy of a 2D experiment 3. 3D NMR spectroscopy no quantum mechanics! Some Common 2D Experiments Very
More informationThis is the total charge on an ion divided by the elementary charge (e).
12.2 Fundamentals and general terms Accelerator mass spectrometry An ultra-sensitive technique using tandem accelerators employed mainly for the study of long-lived radioisotopes, and stable isotopes at
More informationmicrotof-q III Innovation with Integrity The bench-mark in accurate mass LC-MS/MS ESI-Qq-TOF
microtof-q III The bench-mark in accurate mass LC-MS/MS Innovation with Integrity ESI-Qq-TOF The Advantage of Confidence in Routine The microtof-q range is widely acknowledged in setting standards in performance
More informationOSCILLATIONS ABOUT EQUILIBRIUM
OSCILLATIONS ABOUT EQUILIBRIUM Chapter 13 Units of Chapter 13 Periodic Motion Simple Harmonic Motion Connections between Uniform Circular Motion and Simple Harmonic Motion The Period of a Mass on a Spring
More informationNIGAS. i o n. N B C D e t e c t. The Neutron Induced Gamma Spectrometer
NIGAS The Neutron Induced Gamma Spectrometer N B C D e t e c t i o n Po Non-invasive Identification of Explosives and Chemical Warfare Agents Positioning System The NIGAS utilizes Neutron Induced Gamma
More informationLUMINOSITY LEVELLING TECHNIQUES FOR THE LHC
Published by CERN in the Proceedings of the ICFA Mini-Workshop on Beam Beam Effects in Hadron Colliders, CERN, Geneva, Switzerland, 18 22 March 2013, edited by W. Herr and G. Papotti, CERN 2014 004 (CERN,
More informationIntroduction to GC/MS
Why Mass Spectrometry? Introduction to GC/MS A powerful analytical technique used to: 1.Identify unknown compounds 2. Quantify known materials down to trace levels 3. Elucidate the structure of molecules
More informationEnd correction = 0.2cm Actual length of object X = = 1.3cm. Constant gradient (straight line) = uniform acceleration. = 98 cm/s 100 cm/s
# Ans Workings/Remarks C End correction = 0.2cm Actual length of object X =. + 0.2 =.3cm 2 D Force, Acceleration, Velocity and Displacement are vectors. Work, Time and Mass are scalars. 3 D Gradient of
More informationSimulation of External Ion Injection, Cooling and Extraction Processes with SIMION 6.0 for the Ion Trap/Reflectron Time-of-flight Mass Spectrometer
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, VOL. 11, 1467 1477 (1997) Simulation of External Ion Injection, Cooling and Extraction Processes with SIMION 6.0 for the Ion Trap/Reflectron Time-of-flight Mass
More informationCerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs
Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs Application Note Author Harry Prest, Ph.D. Senior Chemist in GC-MS Agilent Technologies, Inc. 5301 Stevens Creek Blvd Santa Clara,
More informationDesign considerations for linear Paul trap mass spectrometer under development
Design considerations for linear Paul trap mass spectrometer under development S.Sevugarajan and A.G.Menon Department of Instrumentation, Indian Institute of Science, Bangalore 561, India. Abstract This
More informationCEE 772: Instrumental Methods in Environmental Analysis
Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511-524) (Harris, Chapt.
More informationMass Spectrometry in MCAL
Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring
More informationPhysics of and in Ion Traps
Physics of and in Ion Traps Proposed Topics: TRIUMF, Vancouver June 01 Basics of Paul- and Penning-traps (equ. of motion, trap geometries, influence of trap imperfections,) Ion detection and cooling (Buffer
More informationREDUCING PROCESS VARIABLITY BY USING FASTER RESPONDING FLOWMETERS IN FLOW CONTROL
REDUCING PROCESS VARIABLITY BY USING FASTER RESPONDING FLOWMETERS IN FLOW CONTROL David Wiklund Marcos Peluso Sr. Principal Engineer Director of Temperature and Plantweb Development Rosemount, Inc. Rosemount,
More informationFourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS)
Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS) The first steps along the way to FT- ICR-MS date back to the 1930s, when Ernest Lawrence, at the University of California - Berkeley,
More informationRequirements for scaleable QIP
p. 1/25 Requirements for scaleable QIP These requirements were presented in a very influential paper by David Divincenzo, and are widely used to determine if a particular physical system could potentially
More informationTargetScreener. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS
TargetScreener A Comprehensive Screening Solution for Forensic Toxicology Innovation with Integrity UHR-TOF MS TargetScreener Get the Complete Picture Forensic laboratories are frequently required to perform
More informationThe phenomenon: complex motion, unusual geometry
Part I The phenomenon: complex motion, unusual geometry Chapter 1 Chaotic motion 1.1 What is chaos? Certain long-lasting, sustained motion repeats itself exactly, periodically. Examples from everyday life
More informationLecture 8: Mass Spectrometry
intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?
More informationRED. BLUE Light. Light-Matter
1 Light-Matter This experiment demonstrated that light behaves as a wave. Essentially Thomas Young passed a light of a single frequency ( colour) through a pair of closely spaced narrow slits and on the
More informationA new seismic testing method E. Kausel Professor of Civil and Environmental Engineering, Massachusetts 7-277, OamWd^e, ^ 027 JP,
A new seismic testing method E. Kausel Professor of Civil and Environmental Engineering, Massachusetts 7-277, OamWd^e, ^ 027 JP, Introduction The bulleted enumeration that follows shows five experimental
More informationCEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis
Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511 524) (Harris, Chapt.
More informationPlasma Diagnostics Introduction to Langmuir Probes
Plasma Diagnostics Technical Information Sheet 531 Plasma Diagnostics Introduction to Langmuir Probes Introduction A Langmuir Probe is a powerful plasma diagnostic tool which capable of determining the
More informationFundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics
Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass
More informationMass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS) Mass Analyzers
More informationAtomic masses. Atomic masses of elements. Atomic masses of isotopes. Nominal and exact atomic masses. Example: CO, N 2 ja C 2 H 4
High-Resolution Mass spectrometry (HR-MS, HRAM-MS) (FT mass spectrometry) MS that enables identifying elemental compositions (empirical formulas) from accurate m/z data 9.05.2017 1 Atomic masses (atomic
More informationRaising intensity of the LHC beam in the SPS - longitudinal plane
SL-Note-- MD Raising intensity of the LHC beam in the SPS - longitudinal plane Ph. Baudrenghien, T. Bohl, T. Linnecar, E. Shaposhnikova Abstract Different aspects of the LHC type beam capture and acceleration
More informationAN-1301 APPLICATION NOTE
AN-131 APPLICATION NOTE One Technology Way P.O. Box 916 Norwood, MA 262-916, U.S.A. Tel: 781.329.47 Fax: 781.461.3113 www.analog.com Using CDCs to Control Motion for Sample Aspiration by Jim Scarlett INTRODUCTION
More informationApplication Note LCMS-110 Development of a Targeted Quantitative LC-MS/MS Method for 431 Positive and Negative Ion Pesticides in a Single Analysis
Application Note LCMS-110 Development of a Targeted Quantitative LC-MS/MS Method for 431 Positive and Negative Ion Pesticides in a Single Analysis Abstract A rapid, targeted quantitative UHPLC-triple quadrupole
More informationImproved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer
Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer Reiko Kiyonami, Mary Blackburn, Andreas FR Hühme: Thermo Fisher
More informationPreClass Notes: Chapter 13, Sections
PreClass Notes: Chapter 13, Sections 13.3-13.7 From Essential University Physics 3 rd Edition by Richard Wolfson, Middlebury College 2016 by Pearson Education, Inc. Narration and extra little notes by
More informationNew Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses
New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses Technical Overview Authors Abstract Peter Stone, Thomas Glauner, Frank Kuhlmann, Tim Schlabach and Ken Miller
More informationAn Introduction to Hyperfine Structure and Its G-factor
An Introduction to Hyperfine Structure and Its G-factor Xiqiao Wang East Tennessee State University April 25, 2012 1 1. Introduction In a book chapter entitled Model Calculations of Radiation Induced Damage
More informationCHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS
Back to Basics Section D: Ion Optics CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS TABLE OF CONTENTS QuickGuide...413 Summary...415 Introduction...417 The physical basis of orthogonal TOF....... 419 Pulsedmainbeamsofions...421
More informationLecture 8: Mass Spectrometry
intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?
More informationTendenze nell innovazione della strumentazione in spettrometria di massa:
Tendenze nell innovazione della strumentazione in spettrometria di massa: Trappola lineare, orbitrap, ion mobility e nuova strumentazione René Magritte La Condition Humaine 1 Ion Trap Mass Spectrometry
More informationChem 3502/4502 Physical Chemistry II (Quantum Mechanics) 3 Credits Spring Semester 2006 Christopher J. Cramer. Lecture 10, February 10, / 4
Chem 350/450 Physical Chemistry II (Quantum Mechanics 3 Credits Spring Semester 006 Christopher J. Cramer Lecture 10, February 10, 006 Solved Homework We are asked to find and for the first two
More informationThe ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode
The ultratrace determination of iodine in aqueous samples using the 7700x ICP-MS with oxygen reaction mode Application note Nuclear Authors Kazumi Nakano, Yasuyuki Shikamori, Naoki Sugiyama and Shinichiro
More informationLaser Dissociation of Protonated PAHs
100 Chapter 5 Laser Dissociation of Protonated PAHs 5.1 Experiments The photodissociation experiments were performed with protonated PAHs using different laser sources. The calculations from Chapter 3
More informationApplication Note LCMS-116 What are we eating? MetaboScape Software; Enabling the De-replication and Identification of Unknowns in Food Metabolomics
Application Note LCMS-116 What are we eating? MetaboScape Software; Enabling the De-replication and Identification of Unknowns in Food Metabolomics Introduction Determining the structure of secondary metabolites
More informationSEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW
SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW Paula Hong, John Van Antwerp, and Patricia McConville Waters Corporation, Milford, MA, USA Historically UV detection has been
More informationHR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput
HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput Yi Zhang 1, Zhiqi Hao 1, Markus Kellmann 2 and Andreas FR. Huhmer
More informationodes I. HARMONIC OSCILLATOR
odes I. HARMONIC OSCILLATOR This is the name that physicists give to almost any thing which vibrates. The vibration is always a competition between two aspects there is some mass which moves and some source
More informationMass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University
Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk Previously: Resolution and other basics MALDI Electrospray 40 Lecture 2 Mass analysers Detectors
More informationCourtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.
ESS The RGA freenotes Theory page 1 of 14 RGA Theory Notes Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.
More informationExemplar for Internal Achievement Standard. Mathematics and Statistics Level 3
Exemplar for internal assessment resource Mathematics and Statistics for Achievement Standard 91580 Exemplar for Internal Achievement Standard Mathematics and Statistics Level 3 This exemplar supports
More informationFourier Transform Ion Cyclotron Resonance Mass Spectrometry
Fourier Transform Ion Cyclotron Resonance Mass Spectrometry What is Mass Spectrometry? Gravity F= G m M/r 2 r M m Newton Newton s second law F = m a Mass-Energy equivalence E = m c 2 1 ev = ~10-9 amu Einstein
More informationInformation Dependent Acquisition (IDA) 1
Information Dependent Acquisition (IDA) Information Dependent Acquisition (IDA) enables on the fly acquisition of MS/MS spectra during a chromatographic run. Analyst Software IDA is optimized to generate
More informationExtrel Application Note
Extrel Application Note Real-Time Plasma Monitoring and Detection of Trace H 2 O and HF Species in an Argon Based Plasma Jian Wei, 575 Epsilon Drive, Pittsburgh, PA 15238. (Presented at the 191st Electrochemical
More informationProgress with the. MPIK / UW - PTMS in Heidelberg. Max Planck Institute for Nuclear Physics / University of Washington Penning Trap Mass Spectrometer
Progress with the MPIK / UW - PTMS in Heidelberg Max Planck Institute for Nuclear Physics / University of Washington Penning Trap Mass Spectrometer TCP 010, Saariselkä, April 1, 010 David Pinegar, MPI-K
More information10. SPONTANEOUS EMISSION & MULTIPOLE INTERACTIONS
P4533 Fall 1995 (Zorn) Atoms in the Radiation Field, 2 page 10.1 10. SPONTANEOUS EMISSION & MULTIPOLE INTERACTIONS In this chapter we address two of the questions that remain as we build a theoretical
More informationStatus of the ESR And Future Options
Status of the ESR And Future Options M. Steck for the Storage Ring Division (C. Dimopoulou, A. Dolinskii, S. Litvinov, F. Nolden, P. Petri, U. Popp, I. Schurig) Outline 1) New Old ESR 2) Slow (Resonant)
More informationAutoresonant Ion Trap Mass Spectrometer The RGA Alternative
Autoresonant Ion Trap Mass Spectrometer The GA Alternative 1 What is a Mass Spectrometer? Ionizer Mass Spectrometer Mass Separator Magnetic Sector, Quadrupole or Ion Trap Detector Partial Pressure Measurement
More informationSupporting Information for: A. Portz 1, C. R. Gebhardt 2, and M. Dürr 1, Giessen, D Giessen, Germany
Supporting Information for: Real-time Investigation of the H/D Exchange Kinetics of Porphyrins and Oligopeptides by means of Neutral Cluster Induced Desorption/Ionization Mass Spectrometry A. Portz 1,
More informationPHY217: Vibrations and Waves
Assessed Problem set 1 Issued: 5 November 01 PHY17: Vibrations and Waves Deadline for submission: 5 pm Thursday 15th November, to the V&W pigeon hole in the Physics reception on the 1st floor of the GO
More informationAB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences
AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching s by Removal of Isobaric Interferences Differential Mobility Used as a Tool to Address Selectivity Challenges Adrian M. Taylor
More informationSimplified Approaches to Impurity Identification using Accurate Mass UPLC/MS
Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS Marian Twohig, Michael D. Jones, Dominic Moore, Peter Lee, and Robert Plumb Waters Corporation, Milford, MA, USA APPLICATION
More informationThermoacoustic Instabilities Research
Chapter 3 Thermoacoustic Instabilities Research In this chapter, relevant literature survey of thermoacoustic instabilities research is included. An introduction to the phenomena of thermoacoustic instability
More informationMRMS axelerate rapidly detected micropollutants and plant response metabolites in poplar leaves
MRMS axelerate rapidly detected micropollutants and plant response metabolites in poplar leaves MRMS axelerate is demonstrated to be a new and powerful workflow to rapidly profile plant extracts in a context
More informationDecoupling Theory and Practice
Decoupling Theory and Practice General Concepts We have spent the last two lectures discussing the structural implications of -coupling analysis and overall have discovered that the interaction may readily
More informationInstrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai
303351 Instrumental Analysis Mass Spectrometry Lecturer:! Somsak Sirichai Mass Spectrometry What is Mass spectrometry (MS)? An analytic method that employs ionization and mass analysis of compounds in
More informationELECTRON DYNAMICS WITH SYNCHROTRON RADIATION
ELECTRON DYNAMICS WITH SYNCHROTRON RADIATION Lenny Rivkin Ecole Polythechnique Federale de Lausanne (EPFL) and Paul Scherrer Institute (PSI), Switzerland CERN Accelerator School: Introduction to Accelerator
More informationAnalysis of Vibrating Plates with Acoustic Holography and Eddy Currents
University of Illinois at Urbana-Champaign Analysis of Vibrating Plates with Acoustic Holography and Eddy Currents Tyler Lindemann and Andrea Vella PHYS 406 Acoustical Physics of Music Professor Steven
More information