Application Note LCMS-110 Development of a Targeted Quantitative LC-MS/MS Method for 431 Positive and Negative Ion Pesticides in a Single Analysis

Size: px
Start display at page:

Download "Application Note LCMS-110 Development of a Targeted Quantitative LC-MS/MS Method for 431 Positive and Negative Ion Pesticides in a Single Analysis"

Transcription

1 Application Note LCMS-110 Development of a Targeted Quantitative LC-MS/MS Method for 431 Positive and Negative Ion Pesticides in a Single Analysis Abstract A rapid, targeted quantitative UHPLC-triple quadrupole mass spectrometry method was developed for the simultaneous analysis of 431 pesticides in both positive and negative ion mode. Authors Zicheng Yang, Louis Maljers Bruker Daltonics Inc., Fremont, California, USA A Bruker Advance UHPLC system coupled to the Bruker EVOQ Elite LC-TQ demonstrated parts per trillion level sensitivity and excellent linearity when applying the method to the analysis of pesticides in a variety of food matrices. Introduction Pesticides are chemicals used to eliminate or control a variety of agricultural pests that can damage crops and livestock and reduce farm productivity. The most commonly applied pesticides are insecticides, herbicides, rodenticides, and fungicides. The US Environmental Protection Agency (EPA) information on US pesticide usage from 2007 estimated that over 2.4 billion kilogram of pesticides were used worldwide. Agricultural use is by far the biggest pesticide consumer. The increasing use of pesticides, illegal use and poor agricultural practices can create significant human, cattle, and pet health concerns through the presence of pesticides and metabolite residues in food. Keywords Pesticides EVOQ LC-TQ LC-MS/MS Compound Based Scanning Positive / negative switching Instrumentation and Software EVOQ LC-TQ

2 The number and frequency of pesticides used for horticultural purposes has increased considerably over the last 50 years. With over 1,100 pesticides documented globally, there has been a commensurate drive to shorten analysis times despite their widely varying physiochemical properties. Consequently, multi-residue pesticide analysis has become an increasingly complex and challenging task. Due to its wide scope of application, speed, sensitivity and selectivity, LC-MS/MS has now become the technique of choice for analyzing these complex samples. Experimental Sample Preparation Celery, bean, tomato and garlic matrix samples were prepared for LC-MS/MS analysis using a standard QuEChERS extraction procedure. Chromatography Parameters Instrument: Analytical column: Column Temperature: Injection Volume: Flow rate: Mobile Phase A: Bruker Advance UHPLC system YMC-Pack ODS-AQ 3 µm, 150mm x 3mm (I.D.) 40 C 5 µl (100 µl Loop) 400 µl/min. 5 mm ammonium fluoride in water Mobile Phase B: Gradient conditions: MS Parameters Methanol 0.0 min, 10% B 0.2 min, 10% B 2.0 min, 70% B 8.0 min, 80% B 10.0 min, 100% B 15.0 min, 100% B 15.1 min. 10% B 18.1 min. 10% B Instrument: Bruker EVOQ Elite with Heated Electrospray Ionization Spray Voltage (Positive): 4000 V Spray Voltage (Negative): 4000 V Cone Gas Flow: 20 units Cone Temperature: 250 C Heated Probe Gas Flow: 40 units Heated Probe Temperature: 400 C Nebulizer Gas Flow: 60 units Exhaust Gas: on q2 pressure: 2.0 mtorr (Argon) Table 1: Subset of the 431 pesticide list derived from the full mass spec method (refer to the Appendix for the full list). Compound Name Retention Time (min) Q1 First Q3 First Collision Energy (ev) Q1 First Q3 First Collision Energy (ev) Polarity Acephate Positive 23 Mesotrione Negative 17 Monocrotophos Positive 17 Bentazone Negative 17 Clethodim Isomer Positive 14 1-Naphthol Negative 14 Chlorotoluron Positive 14 Fludioxinil Negative 14 Ametryn Positive 14 Fipronil Negative 16 Prometryne Positive 16 Hexaflumuron Negative 31 Novaluron Isomer Positive 29 Fluazinam Negative 23 Benfuracarb Positive 23 Temephos Positive 23 Lufenuron Negative 23 Teflubenzuron Negative 23 Scan Time (ms)

3 Compound Based Scanning method Figure 1: Graphic representation of the Compound Based Scanning method for the 431 pesticides. Results The 431 pesticides cover a wide range of different chemical properties which effect the sensitivity and optimum polarity for their detection. The 5 mm ammonium fluoride in the UHPLC mobile phase assisted negative ion pesticide sensitivity without significantly affecting positive ion pesticide performance. The optimized method generated lower limits of detection (LLOD) of 100 ppt for most of the compounds with over 3.5 orders of dynamic range and good linearity (Figure 2). Continuous polarity switching was used in a single analysis without any negative performance effects. Sensitivity and linearity for co-eluting compounds (where one is negative and the other is positive) was still very good. For example, a concentration of 50 ppt for the positive compound Fenbuconazole and 100 ppt for the negative compound Fipronil with a r 2 >0.99 (Figure 3) were quantified easily by the described pesticide method.

4 Pesticides at LLOD level Figure 2: Example chromatogram of selected pesticide MRMs at the LLOD level. Linearity Figure 3: Linearity for positive and negative overlapping peaks.

5 Wide calibration ranges ppb Figure 4: Example calibration curves for a few selected compounds. Working with different matrix samples Figure 5: Example chromatograms of the matrix samples Celery, Bean, Tomato, and Garlic.

6 Bruker Daltonics is continually improving its products and reserves the right to change specifications without notice. Bruker Daltonics , LCMS-110, Conclusion The Bruker EVOQ Elite LC-TQ mass spectrometry system enabled the targeted quantification of 431 positive and negatively charged pesticides with a single method in 18 minutes. Good linearity, sensitivity and RSD s were achieved for all the components, proving the robustness and sensitivity of the EVOQ Elite for rapid multi-residue pesticide quantification. For research use only. Not for use in diagnostic procedures. Bruker Daltonik GmbH Bruker Daltonics Inc. Bremen Germany Phone +49 (0) Fax +49 (0) Billerica, MA USA Phone +1 (978) Fax +1 (978) ms.sales.bdal@bruker.com -

Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS

Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS Abstract A study was carried out using the Bruker EVOQ GC-TQ to screen a

More information

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS PesticideScreener Comprehensive Pesticide Screening and Quantitation Innovation with Integrity UHR-TOF MS The Challenge of Comprehensive Pesticide Residue Analysis The use of pesticides to reduce crop

More information

Multi-residue analysis of pesticides in crude food extracts using a simple extraction technique and LC/MS/MS

Multi-residue analysis of pesticides in crude food extracts using a simple extraction technique and LC/MS/MS PO-CON1529E Multi-residue analysis of pesticides in crude food extracts using a simple extraction technique and LC/MS/MS ASMS 215 WP 77 Miho Kawashima 1, Yuka Fujito 2, Yayoi Ichiki 3, Miho Sakai 4, Takashi

More information

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery Application Note: 3 Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery Yoko Yamagishi, Thermo Fisher Scientific, C-2F 3-9

More information

A Comprehensive, Fast, and Sensitive Method for the Quantitation of Synthetic Cannabinoids using the Elute UHPLC Coupled to an EVOQ LC-TQ MS

A Comprehensive, Fast, and Sensitive Method for the Quantitation of Synthetic Cannabinoids using the Elute UHPLC Coupled to an EVOQ LC-TQ MS A Comprehensive, Fast, and Sensitive Method for the Quantitation of Synthetic Cannabinoids using the Elute UHPLC Coupled to an EVOQ LC-TQ MS Rapid and Sensitive Quantitation of Synthetic Cannabinoids in

More information

Application Note LCMS-112 A Fully Automated Two-Step Procedure for Quality Control of Synthetic Peptides

Application Note LCMS-112 A Fully Automated Two-Step Procedure for Quality Control of Synthetic Peptides Application Note LCMS-112 A Fully Automated Two-Step Procedure for Quality Control of Synthetic Peptides Abstract Here we describe a two-step QC procedure for synthetic peptides. In the first step, the

More information

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Application Note Clinical Research Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Authors Yanan Yang 1, Victor Mandragon 2, and Peter Stone 1 1

More information

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS Application Note EPA Method 535 EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS API 3200 LC/MS/MS System Overview Described here is the

More information

TargetScreener. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS

TargetScreener. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS TargetScreener A Comprehensive Screening Solution for Forensic Toxicology Innovation with Integrity UHR-TOF MS TargetScreener Get the Complete Picture Forensic laboratories are frequently required to perform

More information

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Application Note Authors Kevin McCann, Sameer Nene, Doug McIntyre, Edmond Neo, Dennis Nagtalon, and

More information

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401 Gas Chromatograph Mass Spectrometer No. GCMS-1401 Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS Introduction Development of methods for analysis of drugs of abuse has become a high

More information

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539) Determination of ormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell P Column (EPA 539) Application Note Environmental Authors Rong-jie Fu and Chen-ao (Andy) Zhai Agilent Technologies

More information

A Method for Simultaneous Analysis of 174 Pesticides in Grape Using GC-MS/MS

A Method for Simultaneous Analysis of 174 Pesticides in Grape Using GC-MS/MS PO-CON1518E A Method for Simultaneous Analysis of 174 Pesticides Pittcon 2015 2220-12P Feifei Tian, Xizhi Wang, Shiheng Luo, Guixiang Yang, Taohong Huang, Shin-ichi Kawano, Yuki Hashi, Shimadzu Global

More information

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Application Note Food Authors Jianqiu Mi, Zhengxiang Zhang, Zhixu Zhang,

More information

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis PO-CON1768E Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide ASMS 2017 TP-194 David Baker 1, Christopher Titman 1, Neil Loftus 1, Jonathan Horner

More information

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS Christopher Borton AB SCIEX Golden, Colorado verview Described here is the analysis of

More information

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS Liquid Chromatography Mass Spectrometry SSI-LCMS-096 Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS LCMS-8045 Summary: A short, robust quantitative method for the analysis

More information

Determination of 198 pesticide residues in eggplant using Gas Chromatography tandem Mass Spectrometry/Mass Spectrometry

Determination of 198 pesticide residues in eggplant using Gas Chromatography tandem Mass Spectrometry/Mass Spectrometry PO-CON1619E Determination of 198 pesticide residues in eggplant using Gas Chromatography tandem Mass Spectrometry/Mass Spectrometry Pittcon 2016 480-25 Yan Wang, Feifei Tian, Yueqi Li, Guixiang Yang, Taohong

More information

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Heather A Brown, 1 Claudia Rossi, 2 and Lisa J Calton 1 1 Waters Corporation, Wilmslow, UK 2

More information

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers Application Note Forensic and Toxicology - Criminalistics Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers Using the Agilent 126 Infinity II SFC and Detection by Coupled

More information

MRMS axelerate rapidly detected micropollutants and plant response metabolites in poplar leaves

MRMS axelerate rapidly detected micropollutants and plant response metabolites in poplar leaves MRMS axelerate rapidly detected micropollutants and plant response metabolites in poplar leaves MRMS axelerate is demonstrated to be a new and powerful workflow to rapidly profile plant extracts in a context

More information

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS Application Note: 56 Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS Yang Shi, Catherine Lafontaine, Matthew Berube, John Fink, François Espourteille Thermo

More information

The Analysis of Organophosphate Pesticides by LC/MS Application

The Analysis of Organophosphate Pesticides by LC/MS Application The Analysis of Organophosphate Pesticides by LC/MS Application LC-MS Authors Paul Zavitsanos Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA Paul Yang MOE Ontario Canada

More information

At-a-Glance. Verapamil C 27 H 38 N 2 O 4 M+H + =

At-a-Glance. Verapamil C 27 H 38 N 2 O 4 M+H + = Application ote #LC-MS 80 Simultaneous Quantitative and Qualitative Measurements in a Single Workflow to Increase Productivity in Primary Drug Metabolism Investigations At-a-Glance n Increased productivity

More information

DETERMINATION OF PESTICIDES IN FOOD USING UPLC WITH POLARITY SWITCHING TANDEM QUADRUPOLE LC/MS/MS

DETERMINATION OF PESTICIDES IN FOOD USING UPLC WITH POLARITY SWITCHING TANDEM QUADRUPOLE LC/MS/MS DETERMINATION OF PESTICIDES IN FOOD USING UPLC WITH POLARITY SWITCHING TANDEM QUADRUPOLE LC/MS/MS Peter Hancock 1, Cristiana C. Leandro 2,3, Brendan J. Keely 2 and Richard J. Fussell 3 1 Waters Corporation,

More information

Theory and Practical Consequences.

Theory and Practical Consequences. Theory and Practical Consequences. Number of hits depending on reachable mass accuracy 250 hits @ ±10 ppm 137 hits @ ±5 ppm 78 hits @ ±3 ppm 29 hits @ ±1 ppm 3 hits @ ±0.1 ppm 1 hit @ ± 0.05 ppm Intens.

More information

Designed for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS

Designed for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS Designed for Accuracy High resolution quantitative proteomics Innovation with Integrity LC-MS Setting New Standards in Accuracy The development of mass spectrometry based proteomics approaches has dramatically

More information

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish. 1 Supporting Information 2 3 Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish. 4 5 John Moreman, Okhyun

More information

Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis

Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis Application Note AN772 Introduction This application note describes the extraction of cocaine and a full range

More information

Screening of pesticides in water using SPE on line

Screening of pesticides in water using SPE on line PO-CON E Screening of pesticides in water using SPE on line HPLC 2013 SAMP-38 Stefano Zaza 1, Stefano Maria Lucini 1 1 Shimadzu Corporation, Milano, Italy 1. Introduction Pesticides are widely used in

More information

Application Note GCMS-03 Reliable high-throughput quantification of melamine and related analogs below regulatory limits in food using the EVOQ GC-TQ

Application Note GCMS-03 Reliable high-throughput quantification of melamine and related analogs below regulatory limits in food using the EVOQ GC-TQ Application Note GCMS03 Reliable highthroughput quantification of melamine and related analogs below regulatory limits in food using the EVOQ GCTQ Introduction is a molecule that presents a high content

More information

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 646 Triple Quadrupole LC/MS system with Triggered MRM Application Note Authors Thomas Glauner Agilent Technologies,

More information

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer Improved Screening for 2 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer Mary lackburn, Jia Wang, Jonathan eck, Charles Yang, Dipankar Ghosh, Thermo Fisher Scientific, San Jose, C,

More information

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses Technical Overview Authors Abstract Peter Stone, Thomas Glauner, Frank Kuhlmann, Tim Schlabach and Ken Miller

More information

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes Application Note Food Safety Author Sunil Kulkarni Agilent Technologies

More information

Highly Sensitive and Rugged GC/MS/MS Tool

Highly Sensitive and Rugged GC/MS/MS Tool Highly Sensitive and Rugged GC/MS/MS Tool For Pesticide Multiresidue Analysis in Food Samples Agilent 7 Series Triple Quadrupole GC/MS. The world s first MS/MS designed specifically for GC Analysis Introduction

More information

Proven robustness for large-scale metabolomics studies using the Phenomics Workhorse

Proven robustness for large-scale metabolomics studies using the Phenomics Workhorse Proven robustness for large-scale metabolomics studies using the Phenomics Workhorse HRAM LC-MS profiling of > 1 urine samples demonstrates the outstanding analytical skills of the impact II Introduction

More information

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract. Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS Application Food Analysis Authors Sheher Mohsin Agilent Technologies, Inc. Schaumburg, Il USA Michael Zumwalt

More information

Routine Multiresidue Pesticide Analysis using the Agilent 6470 Triple Quadrupole Mass Spectrometer

Routine Multiresidue Pesticide Analysis using the Agilent 6470 Triple Quadrupole Mass Spectrometer Routine Multiresidue Pesticide Analysis using the Agilent 6470 Triple Quadrupole Mass Spectrometer Application Note Food safety Authors Mark Sartain and Anabel Fandino Agilent Technologies, Inc. Santa

More information

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120 LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120 Application Note Food Testing & Agriculture Author Irina Dioumaeva Agilent Technologies, Inc. Abstract

More information

Simultaneous analysis for forensic drugs in human blood and urine using ultra-high speed LC-MS/MS

Simultaneous analysis for forensic drugs in human blood and urine using ultra-high speed LC-MS/MS Simultaneous analysis for forensic drugs in human blood and urine using ultra-high speed LC-MS/MS TIAFT 204 Toshikazu Minohata, Keiko Kudo 2, Kiyotaka Usui 3, Noriaki Shima 4, Munehiro Katagi 4, Hitoshi

More information

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Adulteration in Dietary Supplements and Botanicals Application Note Food Testing & Agriculture Authors James Neal-Kababick

More information

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Application Note Food Testing and Agriculture Author Chen-Hao (Andy) Zhai and Rong-jie Fu Agilent Technologies (Shanghai)

More information

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS Application Note Food Authors Gerd Vanhoenacker, Frank David,

More information

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany Analysis of Pesticides by Supercritical Fluid Chromatography/ Mass Spectrometry Optimizing the Agilent 126 Infi nity Analytical SFC System in Combination with the Agilent 646 Triple Quadrupole LC/MS Application

More information

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120 LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 10 Application Note Food Testing & Agriculture Author Irina Dioumaeva Agilent Technologies, Inc. Abstract

More information

Application Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry

Application Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry Application Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry Introduction The oil industry requires detailed information on the composition of

More information

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis Application Note AN89.V.1 Extraction of Methylmalonic Acid from Serum Using ISOLUTE SAX Page 1 Extraction of Methylmalonic Acid from Serum Using ISOLUTE SAX Prior to LC-MS/MS Analysis Sample Preparation

More information

Application Note. Abstract. Authors. Environmental

Application Note. Abstract. Authors. Environmental Determination of Acidic Herbicides using an Agilent 6460 Triple Quadrupole LC/MS Equipped with Agilent Jet Stream Technology and Direct Aqueous Injection, for Potable and Environmental Samples Application

More information

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies APPLICATION BENEFITS Regulatory guidelines and recommendations place a greater emphasis on the detection

More information

Chemical Analysis. Low Level Oxygenates Analyzer. Trace Analysis of Oxygenates in Hydrocarbon Matrices. Gas Chromatography.

Chemical Analysis. Low Level Oxygenates Analyzer. Trace Analysis of Oxygenates in Hydrocarbon Matrices. Gas Chromatography. Chemical Analysis Low Level Oxygenates Analyzer Trace Analysis of Oxygenates in Hydrocarbon Matrices think forward Gas Chromatography The determination of sub to high ppm levels of ethers, alcohols, aldehydes

More information

Tyler Trent, Applications Sales Specialist, Teledyne Tekmar P a g e 1

Tyler Trent, Applications Sales Specialist, Teledyne Tekmar P a g e 1 Application Note AutoMate-Q40 Automated QuEChERS Extraction for Pesticide Residues in Botanicals Tyler Trent, Applications Sales Specialist, Teledyne Tekmar P a g e 1 Abstract QuEChERS is a Quick-Easy-Cheap-

More information

Overview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada

Overview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada LC-MS/MS Based Strategy for the Non-Targeted Screening of an Unlimited Number of Contaminants in Food Using the AB SCIEX TripleTOF 5600 System and Advanced Software Tools André Schreiber 1 and Yun Yun

More information

Highly sensitive quantitative analysis of Felodipine and Hydrochlorothiazide from plasma using LC/MS/MS

Highly sensitive quantitative analysis of Felodipine and Hydrochlorothiazide from plasma using LC/MS/MS PO-CON18E Highly sensitive quantitative analysis of Felodipine and Hydrochlorothiazide from plasma using LC/MS/MS ASMS 201 TP97 Shailendra Rane, Rashi Kochhar, Deepti Bhandarkar, Shruti Raju, Shailesh

More information

Agilent 6495 Triple Quadrupole LC/MS System EXPERIENCE A NEW LEVEL OF CONFIDENCE

Agilent 6495 Triple Quadrupole LC/MS System EXPERIENCE A NEW LEVEL OF CONFIDENCE Agilent 49 Triple Quadrupole LC/MS System EXPERIENCE A NEW LEVEL OF CONFIDENCE AGILENT 49 TRIPLE QUADRUPOLE LC/MS SYSTEM WORK SMARTER, DO MORE ifunnel new Q1 Optics new Ion Detector IMPROVED SENSITIVITY

More information

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma LC-MS/MS Method for the Determination of Diclofenac in Human Plasma J. Jones, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20569 Key Words SPE, SOLA, Accucore RP-MS, diclofenac, Core

More information

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories APPLICATION NOTE 65200 Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories Authors Neloni Wijeratne, Claudia Martins, Mary

More information

Highly sensitive and rapid analysis of synthetic dyes in sea food by LC/MS/MS

Highly sensitive and rapid analysis of synthetic dyes in sea food by LC/MS/MS PO-CON1745E Highly sensitive and rapid analysis of synthetic dyes in sea food by LC/MS/MS ASMS 2017 MP 189 Shailendra Rane 1, Ashutosh Shelar 1, Shailesh Damale 1, Rashi Kochhar 1, Purshottam Sutar 1,

More information

Charles Yang, Dipankar Ghosh, Jonathan Beck Thermo Fisher Scientific, San Jose, CA, USA

Charles Yang, Dipankar Ghosh, Jonathan Beck Thermo Fisher Scientific, San Jose, CA, USA Utilization of High Resolution LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using a Q Exactive Benchtop Orbitrap Platform Charles Yang, Dipankar Ghosh, Jonathan Beck Thermo

More information

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 66 Triple Quadrupole LC/MS System Application Note Environmental Authors Don Noot, Matthew Noestheden, and Ralph Hindle Vogon

More information

Application Note LCMS-116 What are we eating? MetaboScape Software; Enabling the De-replication and Identification of Unknowns in Food Metabolomics

Application Note LCMS-116 What are we eating? MetaboScape Software; Enabling the De-replication and Identification of Unknowns in Food Metabolomics Application Note LCMS-116 What are we eating? MetaboScape Software; Enabling the De-replication and Identification of Unknowns in Food Metabolomics Introduction Determining the structure of secondary metabolites

More information

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS Richard Jack, Xiaodong Liu, Leo Wang, and Chris Pohl OT70806_E 08/13S 1 The world leader in serving science

More information

Multi-residue analysis of pesticides by GC-HRMS

Multi-residue analysis of pesticides by GC-HRMS An Executive Summary Multi-residue analysis of pesticides by GC-HRMS Dr. Hans Mol is senior scientist at RIKILT- Wageningen UR Introduction Regulatory authorities throughout the world set and enforce strict

More information

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos

More information

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION Quantitation by High Resolution Mass Spectrometry: Using Target Enhancement and Tof-MRM to Achieve Femtogram-level On-column Sensitivity for Quantitation of Drugs in Human Plasma Yun W. Alelyunas, Mark

More information

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer Olaf Scheibner and Maciej Bromirski Thermo Fisher Scientific, Bremen, Germany Overview Purpose: Improve the performance

More information

Overview. Introduction

Overview. Introduction Comprehensive Quantitation and Identification of Pesticides in Food Samples Using the SCIEX UltraLC 100 and the SCIEX QTRAP 4500 System André Schreiber SCIEX, Concord, Ontario, Canada Overview Liquid Chromatography

More information

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis Application Note AN785 Extraction of Aflatoxins and chratoxin From Dried Chili Using ISLUTE Myco prior to LC-MS/MS Analysis Page Extraction of Aflatoxins and chratoxin from Dried Chili Using ISLUTE Myco

More information

microtof-q III Innovation with Integrity The bench-mark in accurate mass LC-MS/MS ESI-Qq-TOF

microtof-q III Innovation with Integrity The bench-mark in accurate mass LC-MS/MS ESI-Qq-TOF microtof-q III The bench-mark in accurate mass LC-MS/MS Innovation with Integrity ESI-Qq-TOF The Advantage of Confidence in Routine The microtof-q range is widely acknowledged in setting standards in performance

More information

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System Bertram ieland 1 and Stephen Lock 2 1 AB SCIEX ieuwerkerk aan den Ijssel, The etherlands;

More information

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry PO-CON111E Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry Pittcon 201 1120-2 Brahm Prakash, William Lipps, Di Wang, Shilpi Chopra, Nicole

More information

Determination of Pharmaceuticals in Environmental Samples

Determination of Pharmaceuticals in Environmental Samples Determination of Pharmaceuticals in Environmental Samples BACKGROUND The full effects of pharmaceutical substances in the environment are largely unknown however the risk is significant enough that many

More information

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry Application Note Food safety Authors Ismael Flores and Carlos Sepulveda Agrolab México Km 7 Carretera Pachuca-Actopan

More information

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd.

More information

Liquid Chromatography Mass Spectrometry. Shimadzu Corporation, UK; 2 Phytocontrol, France; 3 Shimadzu France, France. Trimesium S + CH 3 ETU

Liquid Chromatography Mass Spectrometry. Shimadzu Corporation, UK; 2 Phytocontrol, France; 3 Shimadzu France, France. Trimesium S + CH 3 ETU LAA-A-LM-E89 Liquid Chromatography Mass Spectrometry LCMS C8 Highly Polar Pesticide Multi-Residue Analysis in Food Safety by LC-MS/MS David R. Baker, Eric Capodanno, Mikaël Levi Shimadzu Corporation, UK;

More information

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward Bruker Daltonics EASY-nLC Tailored HPLC for nano-lc-ms Proteomics think forward Nano-HPLC World-Class Performance with a Small Footprint Bruker Daltonics presents a nano-lc system, perfectly integrated

More information

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies Jonathan Danaceau, Erin Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA, USA

More information

This document is a preview generated by EVS

This document is a preview generated by EVS TECHNICAL REPORT RAPPORT TECHNIQUE TECHNISCHER BERICHT CEN/TR 15641 August 2007 ICS 67.050 English Version Food analysis - Determination of pesticide residues by LC- MS/MS - Tandem mass spectrometric parameters

More information

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for Clinical Toxicology Page of 0 Materials and Methods Biomass Component Extraction: The uneaten cooked plant specimen was prepared for chemical analysis as follows. The sample was frozen, diced, pulverized

More information

Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode

Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode PO-CON1610E Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode Pittcon 2016 1110-2 Brahm Prakash, William Lipps, Di Wang, Shilpi Chopra, Nicole

More information

László Tölgyesi LC/MS scientist EFSC Demo Group. International Symposium on Food Safety and Quality: Applications of Nuclear and Related Techniques

László Tölgyesi LC/MS scientist EFSC Demo Group. International Symposium on Food Safety and Quality: Applications of Nuclear and Related Techniques Minimizing matrix effects in multiresidue pesticide analysis using a new sensitivity-enhanced triple quadrupole instrument in combination with extensive sample dilution László Tölgyesi LC/MS scientist

More information

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use Shih-Tse Jason Lai, Jia Wang, Brad Hart, Kristine Van Natta, Marta Kozak, Jorge Valdivia, Haibo Wang,

More information

LC-MS compatible Separation of the Fungicide Spiroxamine

LC-MS compatible Separation of the Fungicide Spiroxamine Spiroxamine is a systemic fungicide, which was brought to the market by Bayer CropScience. The substance is a mixture of diastereomers A and B again consisting of 4 enantiomers A1, A2, B1 and B2 (fig.

More information

LC Technical Information

LC Technical Information LC Technical Information Method Transfer to Accucore.6 μm Columns Containing solid core particles, which are engineered to a diameter of.6μm and a very narrow particle size distribution; Accucore HPLC

More information

Application News. No. C154. Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Application News. No. C154. Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis LAAN-A-LM-E124 Application Liquid Chromatography Mass Spectrometry Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis David R. Baker

More information

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS Application Note Drug Discovery Authors Srividya Kailasam Agilent

More information

Validation Report 18

Validation Report 18 EURL for Cereals and Feeding stuff National Food Institute Technical University of Denmark Validation Report 18 Determination of pesticide residues in maize for livestock feed by GC-MS/MS and LC-MS/MS

More information

Quantitative Analysis of Opioids Using a Triple-Quadrupole GC/MS/MS

Quantitative Analysis of Opioids Using a Triple-Quadrupole GC/MS/MS PO-CON1408E Quantitative Analysis of Opioids Using a Pittcon 2014 1090-8P Laura Chambers, Richard Whitney, Ph.D., Nicole M. Lock, Zhuangzhi Max Wang, Ph.D., Clifford M. Taylor; Shimadzu Scientific Instruments,

More information

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma Jon Bardsley 1, James Goldberg 2 1 Thermo Fisher Scientific, Runcorn, UK; 2 Thermo Fisher Scientific, West Palm Beach,

More information

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring High-Throughput Protein Quantitation Using Multiple Reaction Monitoring Application Note Authors Ning Tang, Christine Miller, Joe Roark, Norton Kitagawa and Keith Waddell Agilent Technologies, Inc. Santa

More information

Characterization of petroleum samples via thermal analysis coupled to APCI FTMS

Characterization of petroleum samples via thermal analysis coupled to APCI FTMS Characterization of petroleum samples via thermal analysis coupled to APCI FTMS Abstract Thermal analysis, by means of a thermo balance (TG), was coupled to an atmospheric pressure chemical ionization

More information

Analysis of Pesticides in Baby Foods Using a Triple-Quadrupole GC/MS/MS

Analysis of Pesticides in Baby Foods Using a Triple-Quadrupole GC/MS/MS PO-CONE Analysis of Pesticides in Baby Foods Using a Pittcon 01 -P Laura Chambers, Richard Whitney, Ph.D., Nicole M. Lock, Zhuangzhi Max Wang, Ph.D., Clifford M. Taylor; Shimadzu Scientific Instruments,

More information

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Application Note Doping Control Authors Prof. Maria Kristina Parr Freie Universität Berlin Institute

More information

Highly sensitive quantitative estimation of genotoxic impurities from API and drug formulation using LC/MS/MS

Highly sensitive quantitative estimation of genotoxic impurities from API and drug formulation using LC/MS/MS P-CN467E Highly sensitive quantitative estimation of genotoxic impurities from API and drug formulation using LC/MS/MS ASMS 24 TP496 Shruti Raju, Deepti Bhandarkar, Rashi Kochhar, Shailesh Damale, Shailendra

More information

Ultra-fast screening of pesticides in foods and agricultural products with desorption corona beam ionization (DCBI) tandem mass spectrometry

Ultra-fast screening of pesticides in foods and agricultural products with desorption corona beam ionization (DCBI) tandem mass spectrometry PO-CON18E Ultra-fast screening of pesticides in foods and agricultural products with desorption corona beam ionization (DCBI) ASMS 21 TP 3 Jing Dong, Satoshi Yamaki, Yuki Hashi, Naoki Hamada Shimadzu China,

More information

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS Dominic Roberts, 1 Sara Stead, 1 and Jennifer A. Burgess 2 1 Waters Corporation, Manchester, UK 2 Waters Corporation,

More information

Food & Environmental. Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou SCIEX, Concord, Ontario, Canada

Food & Environmental. Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou SCIEX, Concord, Ontario, Canada Food & Environmental Comprehensive Quantitation and Identification of Pesticides in Food Samples using LC-MS/MS with Scheduled MRM, Fast Polarity Switching, and MS/MS Library Searching André Schreiber

More information

Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou AB SCIEX, Concord, Ontario, Canada

Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou AB SCIEX, Concord, Ontario, Canada Comprehensive Quantitation and Identification of Pesticides in Food Samples using LC-MS/MS with Scheduled MRM, Fast Polarity Switching, and MS/MS Library Searching André Schreiber and Yun Yun Zou AB SCIEX,

More information

Analysis of Pesticides in Citrus Oil Using a Shimadzu GCMS-TQ8030. No. GCMS No. SSI-GCMS-1404

Analysis of Pesticides in Citrus Oil Using a Shimadzu GCMS-TQ8030. No. GCMS No. SSI-GCMS-1404 Gas Chromatograph Mass Spectrometer No. GCMS-1404 Analysis of Pesticides in Citrus Oil Using a Shimadzu GCMS-TQ8030 Introduction Contamination of consumer products with pesticides is a growing concern

More information

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada) Quantitation and Identification of Pharmaceuticals and Personal Care Products (PPCP) in Environmental Samples using Advanced TripleTOF MS/MS Technology André Schreiber AB SCIEX Concord, Ontario (Canada)

More information