PROCESS AND ENVIRONMENTAL APPLICATIONS OF GAS CHROMATOGRAPHY WITH THE REDUCTION GAS DETECTOR
|
|
- Adela Anis Nicholson
- 5 years ago
- Views:
Transcription
1 JPACSM 18 PROCESS AND ENVIRONMENTAL APPLICATIONS OF GAS CHROMATOGRAPHY WITH THE REDUCTION GAS DETECTOR Marco Succi SAES Getters S.p.A., Milano, Italy James G. Moncur Independent Consultant KEYWORDS: REDUCTION GAS DETECTOR, PPB, PPT, TRACE GAS ANALYSIS, H 2, CO, HYDRO- CARBONS, ETHYLENE OXIDE, ETHYLENE, PROPYLENE, ARSINE, PHOSPHINE, CARBONYL SULFIDE ABSTRACT The Reduction Gas Detector (RGD) is routinely used for gas chromatographic analysis of volatile compounds that are readily oxidized. The RGD has a tiny amount of HgO in a heated assembly that reacts with volatile compounds as they pass over this bed of material. This causes traces of Hg to be released into a low volume flow cell with photometer for quantitative determination of the analyte. The traces of Hg are then scrubbed with activated charcoal. The HgO bed lasts for six months to two years and is easily changed. Impurities in nitrogen, oxygen, argon, helium, ambient air, carbon dioxide, and ammonia are readily analyzed by GC as the RGD does not react with these relatively stable gases. Hydrogen, carbon monoxide, arsine, phosphine, carbonyl sulfide are measured in ethylene and propylene for catalyst protection purposes. Ethylene, ethylene oxide, arsine, phosphine, nitric oxide, ketones, aldehydes, isoprene, 1,3-butadiene, and benzene are analyzed in ambient air for industrial hygiene and environmental reasons. This paper discusses examples of GC/RGD analyses of ambient air and in chemical process monitoring and the comparison between the flame ionization detector (FID) and the RGD for organic compounds detection. INTRODUCTION The Reduction Gas Detector (RGD) is a unique, ultrasensitive detector developed for the analysis of trace hydrogen, carbon monoxide and reducing species. In the last ten years it has been widely used in several applications where its specificity, reliability, and sensitivity has been a major advantage over more traditional detectors, such as thermal conductivity detectors (TCD) or discharge ionization detectors (DID). This paper discusses some of the various ambient air and process monitoring 1-7 applications of the Reduction Gas Detector for gas chromatography. The difference between the RGD and the flame ionization detection (FID) of organic compounds is also discussed. OPERATING PRINCIPLE OF THE RGD The basic operating principle of the detector is based upon the strong absorption of UV light by mercury vapor. Compounds eluting from the column pass through a mercuric oxide bed that is heated. Reducing components, i.e. compounds that can combine with oxygen, react with the mercuric oxide and release mercury vapor. The free vapor evolved in the reaction is then detected by measuring the absorption of UV light in a photometric cell. The absorption is directly proportional to the concentration of the mercury vapor which is proportional to the concentration of the reducing analyte from the column. The following general reaction takes place: X + HgO (solid) XO + Hg (vapor) For H 2 and CO the reactions are: H 2 + HgO H 2 O + Hg CO + HgO CO 2 + Hg This technique of sensing mercury vapor lends itself to part-per-trillion (ppt) and part-per-billion (ppb) detection limits of reducing compounds. Since the detector reacts only to reducing compounds, there is no interference from the matrix such as O 2, Air, He, N 2, Ar, CO 2, etc. The actual detection limits are related to the reaction efficiency of different compounds in the mercury oxide bed as well as the sample volume injected into the column, the path length of the photometer cell, and the signal to noise level achieved through the system.
2 JPACSM 19 A cross sectional drawing of the detector is shown in Figure 1. The UV source is a mercury lamp with an optical filter and a reference photo diode detector to monitor lamp intensity fluctuations at 254 nm. Figure 1. Reduction Gas Detector schematic. The rest of the light travels via the light guide through the cell and out the other light guide through the 254 nm optical filter to the signal photo diode detector. The column effluent enters the HgO reaction bed which is heated to about 265 C, where the reducing compounds react and release mercury vapor. The light intensity will be different between the two photo diodes, depending upon the amount of mercury vapor in the cell. Signal processing electronics measure the difference between the two photodiodes, automatically canceling lamp flicker noise, and displaying the mercury vapor concentration. This is directly proportional to the concentration of a reducing compound eluting from the column system. The low level of mercury vapor in the carrier gas flowing from the photo cell is trapped with a scrubber before entering the ambient atmosphere. For the ultimate detection limit, it is important to have a carrier gas with a reliable negligible concentration of reducing species to optimize the detector signal to noise ratio. ENVIRONMENTAL APPLICATIONS H 2 and CO Analysis in the Atmosphere The RGD is ideally suited for the determination of ppb to part-per-million (ppm) levels of H 2 and CO in atmospheric studies. CO is involved in ozone depletion, even at very low concentrations. The RGD was mounted in an aircraft and was used by Dr. Paul Crutznen for his analyses of CO in the upper stratosphere. In these analyses, the Reduction Gas Detector is used to provide rapid and reproducible quantitative measurements 8,9. This system was designed for trace levels of H 2 and CO in the atmosphere, where there is a possibility of other high molecular weight pollutants that could interfere with the determination. Air carrier gas is preferred because it gives a minimum of upset in the baseline from the injection, is inexpensive, and readily available. A 10-port valve is used to inject the sample from the sample loop to the stripper column. Lighter compounds, such as H 2 and CO, elute quickly from the stripper column into the analytical column, while higher molecular weight components such as hydrocarbons and moisture elute very slowly from the stripper column. After the light components have had sufficient time to elute from the stripper column into the analytical column, the 10-port valve returns to the load position. In addition to preparing the analyzer for the next sample injection, this valve position is used to backflush the higher molecular weight components from the stripper column to vent. Backflushing of the stripper column accomplishes two objectives. It shortens the analysis time and it vents contaminants which can produce an unstable baseline away from the detector. Oxygen and nitrogen from the air will elute through the analytical column. However these components do not interfere with the determination of H 2 and CO because of the chromatography and the selectivity of the RGD. Ethylene Oxide in Ambient Air The presence of ethylene oxide in ambient and indoor air should be monitored at less than the OSHA regulatory limit of one ppm to reduce the possibility of cancers in workers exposed to it. A Reduction Gas Detector, with appropriate injection system and columns, responds to ethylene oxide at ppb in ambient air (Table I and Figure 1). In the RGD determination of ethylene oxide, a 10-port valve and a 1 ml sample loop is in the flow path of continuously flowing ambient air. The valve rotates to inject the sample to a stripper column, (inorganic phase). Light components, such as oxygen, are diverted from the stripper column to vent using a 4-port valve located downstream from the 10-port valve. After the oxygen elutes from the stripper column, this 4-port valve is programmed to return to its original position to transfer ethylene oxide to the analytical column (HayeSep). In this way, the 10-port valve and the 4-port valve divert the oxygen, carbon monoxide, and hydrogen to vent and away from the analytical column and RGD detector. The 10- port valve is programmed to return to its original sampling position once the ethylene oxide has been transferred to the analytical column, and at the same time, puts the stripper column in backflush mode to prevent the build up of moisture. The total cycle time between sequential injections in this work is 500 seconds (8.3 minutes). Table I summarizes the results obtained generating different concentrations.
3 JPACSM 20 Table I EtO in Air Analytical Data EtO (ppb) Peak Area Average %SD , , , , , , , , , ,864 12, ,856 Figure 2 shows a typical spectrum with less than 1 ppm EtO. A 10-port valve is used to inject sample from the sample loop to the stripper column. Lighter compounds such as H 2 and CO, elute quickly from the stripper column into the analytical column while higher molecular weight components, such as hydrocarbons (principally ethylene and propylene) and moisture remain in the stripper column. After the light components have had sufficient time to elute from the stripper columns into the analytical column, the 10-port valve returns to the load position. In addition to preparing the analyzer for the next sample injection, the valve in this position is used to backflush the higher molecular weight components from the stripper columns to vent. As mentioned previously, backflushing of the stripper column to vent shortens the analysis time and vents contaminants which can produce an unstable baseline. Figure 3 shows a typical sample of propylene when the backflush is set after the ethylene peak; in the case of high ethylene concentrations, the backflush should be set before the ethylene peak to avoid excessive consumption of the mercury oxide bed. Figure 2. Typical chromatogram of about 0.8 ppm EtO in air. PROCESS APPLICATIONS H 2 and CO in Ethylene or Propylene Making polyethylene or polypropylene from ethylene or propylene requires a catalyst, such as the Ziegler- Natta polymerization catalyst. The performance of this catalyst is adversely affected by CO, and it is of interest to determine this component at the ppb levels. In this analysis, the RGD is used to provide rapid and reproducible CO measurements 10. Unsaturated hydrocarbons give strong response in the RGD. When ethylene or propylene is a major component in the sample, it can overload the RGD as well as the analytical column, producing a long tailing peak that requires a long time to return to baseline. This application package is designed to overcome interference from the major unsaturated component by using special plumbing and column configuration. Nitrogen carrier gas is used because it gives a minimum of upset in the baseline from the injection, is inexpensive, and readily available. Figure 3. Propylene sample with trace level of H 2, CO, and ethylene. The RGD is also sensitive to phosphine, arsine, and carbonyl sulfide, additional impurities that can adversely affect catalyst performances, as shown in Figure 4.
4 JPACSM 21 RGD millivolt response CO PH3 18 ppb AsH3 27 ppb COS retention time (sec) Figure 4. CO, PH 3, AsH 3, and COS in propylene H 2 and CO in Ammonia Ammonia purity used in the production of light emitting diodes or diode lasers is becoming more and more important to a successful manufacturing operation. Setting the injection system in a way similar to the one mentioned previously, letting H 2 and CO pass over a stripper column and trapping NH 3, allows the analysis of H 2 and CO at trace level to be easily performed. As shown in Figures 5 and 6, the calibration performed generating various concentrations in the ppb range provided good linear results and allowed detection in the range of 10 ppb. peak area (counts) 1.E+05 8.E+04 6.E+04 4.E+04 2.E+04 y = x R 2 = E ,000 1,200 inlet concentrations (ppb) Even better results, in terms of sensitivity, can be obtained for less reactive matrix gases such as Ar, N 2, H 2, O 2, and CO 2. Figure 7 shows a chromatogram of about 1 ppb H 2 and CO in a sample of ultra pure nitrogen RGD Signal (counts) 50,000 49,000 48,000 47,000 46,000 45,000 44,000 Detector Auto Zero Normal Valve Pressure Switch H 2 43, Time (seconds) 1 ppb Figure 7. One ppb H 2 and CO in ultra pure N 2. CONCLUSIONS Some applications of a Reduction Gas Detector for gas chromatography have been briefly discussed. Even if the RGD is not commonly used for the detection of organic compounds, it offers several advantages over the more traditional FID, such as a sensitivity about times better and the lack of a requirement for service gases such as air or hydrogen. The data presented in this paper were obtained with packed column; however, the low carrier gas flow rate required for its operation, 10 ml/minute, makes it suitable for use in combination with capillary columns, therefore allowing use for other applications. CO Figure 5. H 2 calibration in an ammonia sample. 3.E+06 peak area (counts) 3.E+06 2.E+06 2.E+06 1.E+06 y = x R 2 = E+05 0.E concentration (ppb) Figure 6. CO calibration in an ammonia sample.
5 JPACSM 22 REFERENCES 1. R.F. Hickey and M.S. Switzenbaum, Environmental Science and Technology, 24, 11, , (1990). 2. D.W. Smith, A.O. Hopper, et al, J. Of Pediatric Gastroenterology and Nutrition, 3, 77-80, (1984). 3. C.P.R. Ostrander, D.K. Stevenson, et al, Analytical Biochemistry, 119, , (1982). 4. H.J. Vreman, L.M. Baxter, R.T. Stone, and D.K. Stevenson, Chinical Chemistry, 42, 1, 50-56, (1996). 5. M. Khalil and R.A. Rasmussen, JAPCA, 38, , (1988). 6. Xu-Liang Cao, C.N. Hewitt, and K.S. Waterhouse, J. of Chromatography, A, 679, , (1994). 7. Xu-Liang Cao, C.N. Hewitt, and K.S. Waterhouse, Analytica Chimica Acta, 300, , (1995). 8. H. B. Singh, M. Kanakidou, P.J. Crutzen, and D.J. Jacob, Nature, 378, 2, November, (1995). 9. D. O'Hara, and H.B. Singh, Atmospheric Environment, 22, 11, , (1988). 10. G. Frising, F. Testoni, V. Furlan, and C. Ostrander, LC GC Int., 8, 2, February, (1995). 11. R.G. Ridgeway, S.N. Ketkar, J.V. Martinez, and C.R. Ostrander, Microcontamination Conference Proceedings, November, (1991). 12. R.G. Ridgeway, S. Ketkar, and J.V. Martinez, Technology Circuit, 1, 3, (1992). 13. S. Guadagnuolo and L. Pusterla, PITTCON 2000, paper J.G. Moncur, PITTCON 2000, paper 917.
AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES
JPACSM 127 AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES Trace Analytical Inc. Menlo Park, CA ABSTRACT GC based gas analyzers with Reduction Gas Detector (RGD) and Flame Ionization
More informationApplication Note S/SL
Analysis of Ultra low Sulfur Compounds in Natural Gas and Gaseous Fuels by Gas Chromatography and Chemiluminescence according to ASTM D Ultra low detection limits Excellent Sensitivity, Repeatability &
More informationNON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25)
Gas Chromatography NON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25) The Non-Methane Organic Compounds (NMOC) Analyzer is a gas chromatograph configured for analyzing gaseous samples for total organic
More information2B Technologies, Inc. An InDevR Company
2B Technologies, Inc. An InDevR Company Technical Note No. 40 UV-Absorbing Interferences in Ozone Monitors Date: 22 April 2015 Author: John Birks Background Ozone measurements by absorbance of the 253.7-nm
More informationFast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary.
Application Note 36720111 Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary PLOT Column Keywords Pulsed Flame Photometric
More informationFar UV Absorbance Detector
Far UV Absorbance Detector Theory Most organic and inorganic species absorb strongly in the far UV. Notable exceptions are the inert gases, helium and nitrogen which absorb very weakly in this region.
More informationGas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5
Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH 5730-6730 Analysis Topic 5 Chromatography! Group of separation techniques based on partitioning (mobile phase/stationary phase). Two immiscible
More informationThe Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer
The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS Agilent Technologies/Wasson ECE Monomer Analyzer Application Authors Fred Feyerherm 119 Forest Cove Dr. Kingwood, TX
More informationChemistry Instrumental Analysis Lecture 27. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 27 Gas Chromatography Introduction GC covers all chromatographic methods in which the mobile phase is gas. It may involve either a solid stationary phase (GSC)
More informationMonomer Analysis. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.
Monomer Analysis Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Polymer Grade Monomer Analysis Monomer Analysis Impurities in feedstocks can adversely
More informationHarris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY
Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY Chapter 23. Gas Chromatography What did they eat in the year 1,000? GC of Cholesterol and other lipids extracted from
More informationPetrochemical. Transformer Oil Gas Analysis - TOGA
Petrochemical Transformer Oil Gas Analysis - TOGA www.dps-instruments.com The DPS TOGA GC Systems are designed to analyze oil from electrical insulation materials that may have decomposed under thermal,
More informationAnalyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column
Analyze Hydrocarbon Impurities in,3-butadiene with an Agilent J&W GS-Alua PT Column Application Note Energy and Chemicals Authors Yun Zou and Chunxiao Wang Agilent Technologies Shanghai Ltd Abstract The
More information2401 Gas (liquid) Chromatography
2401 Gas (liquid) Chromatography Chromatography Scheme Gas chromatography - specifically gas-liquid chromatography - involves a sample being vaporized and injected onto the head of the chromatographic
More informationGas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC
Gas Chromatography Gas Chromatography Presented By Mr. Venkateswarlu Mpharm KTPC What is Gas Chromatography? It is also known as Gas-Liquid Chromatography (GLC) GAS CHROMATOGRAPHY Separation of gaseous
More informationHigh-Speed Gas and Headspace Analysis for the Process-Line and Laboratory: SIFT- MS IFPAC 2017
High-Speed Gas and Headspace Analysis for the Process-Line and Laboratory: SIFT- MS IFPAC 2017 Y.J. Mange D.B. Milligan V.S. Langford B.J. Prince M. Perkins C. Anderson T. Wilks Who is using Syft Technologies
More informationTHE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS
International Gas Union Research Conference 14 THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS Main author Hironori IMANISHI Tokyo Gas Co., Ltd. JAPAN himanishi@tokyo-.co.jp
More informationGas Chromatography. Rosa Yu, David Reckhow CEE772 Instrumental Methods in Environmental Analysis CEE 772 #16 2
Print version Gas Chromatography Rosa Yu, David Reckhow CEE772 Instrumental Methods in Environmental Analysis CEE 772 #16 1 Contents The primary components to a GC system 1. Carrier Gas System (including
More informationGas Chromatography. Introduction
Gas Chromatography Introduction 1.) Gas Chromatography Mobile phase (carrier gas) is a gas - Usually N 2, He, Ar and maybe H 2 - Mobile phase in liquid chromatography is a liquid Requires analyte to be
More informationAPPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline
A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline Fast : Cycles in
More informationUV Hound Series. Dependable High Quality Air Monitoring. Accurate Readings Within Seconds. Portable UVDOAS Multi-gas Analyzers. Multi-Gas Capability
UV Hound Series Portable UVDOAS Multi-gas Analyzers Multi-Gas Capability Dependable High Quality Air Monitoring Individual VOCs PPB Sensitivity Non-Contact Optical Measurement Automated Reporting Configurable
More informationCH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian
CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography M. Subramanian Assistant Professor Department of Chemical Engineering Sri Sivasubramaniya Nadar College of Engineering Kalavakkam 603
More informationAnalysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application
Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies
More informationGC Instruments. GC Instruments - Sample Introduction
GC Instruments 1 Fairly simple instrumentation Maintaining constant average pressure is important! Pressure controls flow rate T influences retention (k ) Flow rate monitoring Changing flow rate changes
More informationRS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers
RS DYNAMICS ECOPROBE 5 Portable IR/PID Gas Analyzer PID + IR PID and IR Analyzers General ECOPROBE 5 has two autonomous analyzers in one case. The combination of analyzers provides a set of data designed
More informationAnalytical techniques: Environmental samples. Lecture 2 Universidade do Algarve
Analytical techniques: Environmental samples Lecture 2 Universidade do Algarve Terms, definitions & applications Difference between technique and method: Analytical technique: Fundamental scientific application
More informationGAS CHROMATOGRAPHY (GC)
GAS CHROMATOGRAPHY (GC) Pre-Lab Questions Questions are to be answered before the beginning of the laboratory. The answers are due at the beginning of each experiment (the questions are for credit and
More informationAtmospheric Analysis Gases. Sampling and analysis of gaseous compounds
Atmospheric Analysis Gases Sampling and analysis of gaseous compounds Introduction - External environment (ambient air) ; global warming, acid rain, introduction of pollutants, etc - Internal environment
More informationIntroduction to Gas Chromatography
Introduction to Gas Chromatography 31-1 Objectives To know what is chromatography To understand the mechanism of compound separation To know the basic of gas chromatography system 31-2 Chromatography Definition
More informationThe Claviature of Gas Analysis
The Claviature of Gas Analysis Mass spectrometric gas analysis for quality assurance of gases and gas mixtures High purity and special gases are becoming increasingly important in the development of modern
More informationChapter 31 Gas Chromatography. Carrier Gas System
Chapter 31 Gas Chromatography GAS-LIQUID CHROMATOGRAPHY In gas chromatography, the components of a vaporized sample are fractionated as a consequence of being partitioned between a mobile gaseous phase
More informationPRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC
PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC CHROMATOGRAPHY Laboratory technique for the Separation of mixtures Chroma -"color" and graphein
More informationSub-ppt Atmospheric Measurements Using PTV-GC-FID and Real-Time Digital Signal Processing
Sub-ppt Atmospheric Measurements Using PTV-GC-FID and Real-Time Digital Signal Processing Application Note No. 036 J.B McQuaid 1, A.C Lewis 1, K.D Bartle 1 and S.J Walton 2 1 School of Chemistry, University
More informationCHAPTER 6 GAS CHROMATOGRAPHY
CHAPTER 6 GAS CHROMATOGRAPHY Expected Outcomes Explain the principles of gas chromatography Able to state the function of each components of GC instrumentation Able to state the applications of GC 6.1
More informationExperiment 6 Simple and Fractional Distillation
Experiment 6 Simple and Fractional Distillation Vapor Pressure vs Temperature of Water Vapor Pressure vs Temperature of Water 25 Vapor Pressure vs Temperature of Water 25 Vapor Pressure (kpa) (kpa) 2 2
More informationChapter 27: Gas Chromatography
Chapter 27: Gas Chromatography Gas Chromatography Mobile phase (carrier gas): gas (He, N 2, H 2 ) - do not interact with analytes - only transport the analyte through the column Analyte: volatile liquid
More informationGas Chromatography (Chapter 2 and 3 in The essence of chromatography)
Gas Chromatography 1. Introduction. Stationary phases 3. Retention in Gas-Liquid Chromatography 4. Capillary gas-chromatography 5. Sample preparation and injection 6. Detectors (Chapter and 3 in The essence
More informationGas Chromatography (GC)
Gas Chromatography (GC) Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11541 Saudi Arabia Office: AA53
More informationVirtually Particle-Free Rt -Silica BOND Columns
Petrochemical Applications Virtually Particle-Free Rt -Silica BOND s Provide Reliable PLOT Performance With Less Time Lost for Maintenance By Amanda Rigdon, Bill Bromps, Tom Vezza, and Jaap de Zeeuw Optimized
More information2Fe 2 O 3 +3H 2 S FeS+FeS x +S+3H 2 O
Elemental analysis of hydrocarbon streams using Dry colorimetry analyzers, a catalyst saviour Bushra Dawood, Application Coordinator C.I. Analytics www.cianalytics.com The Petrochemical industry has refined
More informationDEPARTMENT OF ELECTRONICS AND INSTRUMENTATION ENGINEERING SRM NAGAR, KATTANKULATHUR-603203 EI 2302 ANALYTICAL INSTRUMENTS QUESTION BANK UNIT I COLORIMETRY AND SPECTROPHOTOMETRY Part A 1. State Lambert
More informationGas chromatography. Advantages of GC. Disadvantages of GC
Advantages of GC Gas chromatography Fast analysis, typically minutes Effi cient, providing high resolution Sensitive, easily detecting ppm and often ppb Nondestructive, making possible on - line coupling;
More informationAnalysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software
Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software NEMC, Washington D.C. August 6 th 2012 Nicola Watson Environmental Specialist Markes International
More informationThermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS
Thermo Scientific ConFlo IV Universal Interface Continuous Flow Interface Isotope Ratio MS 3 ConFlo IV Universal Interface for Continuous Flow Isotope Ratio MS The development of Continuous Flow carrier
More informationSTANDARD OPERATING PROCEDURES
PAGE: 1 of 17 CONTENTS 1.0 SCOPE AND APPLICATION* 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
More information1.0 Reactive Gas Chromatography Design and Setup Procedure for Catalyst Pelletization & Inlet Packing Software and Method.
Electronic Supplementary Material (ESI) for Catalysis Science & Technology. This journal is The Royal Society of Chemistry 207 Supplementary Information Simple Quantification of Zeolite Acid Site Density
More informationIntroduction. David B. Wurm, Kefu Sun, and William L. Winniford. Abstract
Analysis of Low Levels of Oxygen, Carbon Monoxide, and Carbon Dioxide in Polyolefin Feed Streams Using a Pulsed Discharge Detector and Two PLOT Columns David B. Wurm, Kefu Sun, and William L. Winniford
More informationProduct Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas
Agilent Porous Polymer PLOT Columns: New Products, Expanded Uses, Prices Cut in Half! Product Brief Need improved resolution of small volatile compounds? Didn't try a PLOT column due to high price, short
More informationGUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY
Page 1 of 13 CIPAC/4105/R GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Prepared for CIPAC by Dr M J Tandy*, P M Clarke and B White (UK) The rapid
More informationTOUR OF SOIL CHEMISTRY LABS (09/22/07)
Department of Crop and Soil Sciences Commemorating Our Centennial, 1907 2007 TOUR OF SOIL CHEMISTRY LABS (09/22/07) WELCOME TO THE TOUR OF SOIL CHEMISTRY LABS Instrument: Lachat QuikChem FIA + 8000 Analyzer
More informationCantilever enhanced tunable diode laser photoacoustic spectroscopy in gas purity measurement case study: acetylene in ethylene
Cantilever enhanced tunable diode laser photoacoustic spectroscopy in gas purity measurement case study: acetylene in ethylene Juho Uotila, Jussi Raittila, Ismo Kauppinen ¹Gasera Ltd., Tykistökatu 4, 20520
More informationChemical Analysis. Low Level Oxygenates Analyzer. Trace Analysis of Oxygenates in Hydrocarbon Matrices. Gas Chromatography.
Chemical Analysis Low Level Oxygenates Analyzer Trace Analysis of Oxygenates in Hydrocarbon Matrices think forward Gas Chromatography The determination of sub to high ppm levels of ethers, alcohols, aldehydes
More informationLuminescence transitions. Fluorescence spectroscopy
Luminescence transitions Fluorescence spectroscopy Advantages: High sensitivity (single molecule detection!) Measuring increment in signal against a dark (zero) background Emission is proportional to excitation
More informationThe Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity
The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity Janet Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS High
More informationMuliple Gas#3 plus Sulfur GC Configuration
The SRI Model 8610C Gas Chromatograph ( GC ) configured as a MultipleGas#3 plus Sulfur is designed to measure H2, O2, N2, CO, CO2, H2O, C1 through C5 hydrocarbons and also H2S, COS/SO2, and other sulfur
More informationPermanent Gas Analysis by Gas Chromatography Vacuum Ultraviolet Spectroscopy
Application Note Permanent Gas Analysis by Gas Chromatography Vacuum Ultraviolet Spectroscopy Permanent Gas Analysis by Gas Chromatography Vacuum Ultraviolet Spectroscopy Introduction Gas Chromatography
More informationMeasurement Technique and its Application to Trace Components of Atmospheric Gas
F e a t u r e A r t i c l e Feature Article Measurement Technique and its Application to Trace Components of Atmospheric Gas Junji Kato From early on, HORIBA has been developing and adopting various measurement
More informationGas Chromatography in the Low Concentration Determination of Nitrous Oxide and Volatile Anaesthetics
Turk J Chem (998), 67 7. c TÜBİTAK Gas Chromatography in the Low Concentration Determination of Nitrous Oxide and Volatile Anaesthetics Ahmet UYANIK Ondokuz Mayıs University, Faculty of Arts and Science
More informationTHE APPLICATION OF PROCESS MASS SPECTROMETRY TO FUMED SILICA PRODUCTION
JPACSM 5 Process Analytical Chemistry THE APPLICATION OF PROCESS MASS SPECTROMETRY TO FUMED SILICA PRODUCTION Peter Traynor and Steven Trimuar Thermo Electron Corporation Houston, TX Robert G. Wright Thermo
More informationPrinciples of Gas- Chromatography (GC)
Principles of Gas- Chromatography (GC) Mohammed N. Sabir January 2017 10-Jan-17 1 GC is a chromatographic technique utilizes gas as the mobile phase which is usually an inert gas (Hydrogen, Helium, Nitrogen
More informationChromatography. Gas Chromatography
Chromatography Chromatography is essentially the separation of a mixture into its component parts for qualitative and quantitative analysis. The basis of separation is the partitioning of the analyte mixture
More informationMETHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION
METHOD 80 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION 1.0 SCOPE AND APPLICATION 1.1 Method 80 may be used to determine the concentration of acetonitrile (CAS No. 75-05-8) in aqueous
More informationMultidimensional GC. Multidimensional GC. Multidimensional GC. Why bother? Examples of complex samples. Multidimensional Gas Chromatography
Multidimensional Gas Chromatography Examples of complex samples The use of two or more columns to resolve a sample (or using a single column in more than one direction. Current column technology is very
More informationPrima PRO Process Mass Spectrometer
APPLICATION NOTE Prima PRO Process Mass Spectrometer No. xxxx Controlling emissions by multi-component analysis of Volatile Organic Compounds (VOC) Authors: Peter Traynor, Thermo Fisher Scientific, Franklin,
More informationGas Chromatography CHEM Dr. Reem M. Alghanmi st term
Gas Chromatography CHEM 313-5 Dr. Reem M. Alghanmi 2017 1 st term 17.7 Gas Chromatography Introduction There are two types of gas chromatography: Gas-solid (adsorption) chromatography. Gas-liquid (partition)
More informationChromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed
Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed Gas Chromatography (GC) In gas chromatography, the sample is vaporized and injected onto the head of a chromatographic
More informationPolymer plants continue to seek ways to increase production and efficiency without compromising safety.
Polyethylene Polypropylene APPLICATION NOTE NOTE Polymer plants continue to seek ways to increase production and efficiency without compromising safety. Process gas analysis is integral to the control
More informationUnderstanding Gas Chromatography
Understanding Gas Chromatography What is Really Going on Inside the Box? Simon Jones GC Applications Engineer Page 1 Group/Presentation Title Month ##, 200X ?? K? Page 2 Typical GC System Gas supply Injector
More informationVUV ANALYTICS VGA-100 GC DETECTOR Gas Chromatograph Agilent 6890 equipped with a 7683 model autosampler Restek 30m x 0.25mm x 0.
System Description The VUV PIONA + Analyzer is a VGA-100 Gas Chromatography (GC) detector configured for PIONA compound analysis by the VUV PIONA + method. The VUV PIONA + method uses relatively simple
More informationResidual Solvents in Pharmaceuticals by USP Chapter <467> Methodology
APPLICATION NOTE Gas Chromatography Author: David Scott PerkinElmer, Inc. Shelton, CT Residual Solvents in Pharmaceuticals by USP Chapter Methodology Introduction The synthesis of active pharmaceutical
More informationcostech instruments ECS 4010 Elemental Combustion System CHNS-O
instruments ECS 4010 Elemental Combustion System CHNS-O costech Costech Analytical Tecnologies Inc. 26074 Avenue Hall, Suite 14 Valencia, CA 91355 - USA Phone: (661) 297-2395 Fax: (661) 297-5492 e-mail:
More informationFast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography
Fast Analysis of Aromatic Solvent with.8 mm ID GC column Application Gas Chromatography Authors Yun Zou Agilent Technologies (Shanghai) Co. Ltd. Ying Lun Road Waigaoqiao Free Trade Zone Shanghai 3 P.R.
More informationChapter 11 Conventional Gas Chromatography
Chapter 11 Conventional Gas Chromatography Gas Chromatography GC is the first instrumental chromatographic method developed commercially It is relatively easy to introduce a stable flow and pressure for
More informationEarth s Atmosphere. Composition
Earth s Atmosphere Earth s atmosphere is a layer of gases surrounding the planet that is held in place by gravity. The atmosphere protects life on Earth by absorbing ultraviolet radiation, warming the
More information10/8/2013. Alternatives to Mass Spectrometry and Some Real-time Monitoring Measurements. Project Driven Tasks -- Process Gas Measurements
Alternatives to Mass Spectrometry and Some Real-time Monitoring Measurements Co-worker Acknowledgement Laura Tovo, Ronald Hooper, Louis Boone, Jack Zamecnik William A. Spencer SRNL Science and Technology
More informationATOC 3500/CHEM 3151 Air Pollution Chemistry Lecture 1
ATOC 3500/CHEM 3151 Air Pollution Chemistry Lecture 1 Note Page numbers refer to Daniel Jacob s online textbook: http://acmg.seas.harvard.edu/publications/ jacobbook/index.html Atmos = vapor + sphaira
More informationReal Time On-Site Odor and VOC Emission Measurements Using a znose
Real Time On-Site Odor and VOC Emission Measurements Using a znose Edward J. Staples, Electronic Sensor Technology, EST@ESTCAL.COM Remediation Site Description Remediation of contaminated soil from where
More informationIntroduction to Capillary GC
?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention
More informationAnalysis for Natural Gas and Similar Gaseous Mixtures by Gas Chromatography
GPA Standard 2261-13 Analysis for Natural Gas and Similar Gaseous Mixtures by Gas Chromatography Adopted as Tentative Standard, 1961 Revised and Adopted as a Standard, 1964 Revised 1972, 1986, 1989, 1990,
More informationTrace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection
Trade Science Inc. September 2009 Volume 8 Issue 3 ACAIJ, 8(3) 2009 [346-349] Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization
More informationANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis
DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes
More informationHighly Sensitive and Rugged GC/MS/MS Tool
Highly Sensitive and Rugged GC/MS/MS Tool For Pesticide Multiresidue Analysis in Food Samples Agilent 7 Series Triple Quadrupole GC/MS. The world s first MS/MS designed specifically for GC Analysis Introduction
More informationIntroduction to IH Analytical Chemistry
Introduction to IH Analytical Chemistry Sampling and Analysis Manning, Eide, Van Etten (Apr 2013) Foreword Thank you for purchasing the Introduction to IH Analytical Chemistry self-study workbook. Course
More informationUV3000. Accurate, precise, and portable ambient gas point analyzer
UV3000 Accurate, precise, and portable ambient gas point analyzer The Cerex UV3000 is a multifunction analyzer designed to detect part per billion (ppb) to percent level concentrations of multiple gases
More informationVALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur
VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur 603 203 DEPARTMENT OF ELECTRONICS AND INSTRUMENTATION ENGINEERING QUESTION BANK V SEMESTER EI6501 Analytical Instruments Regulation 2013 Academic
More informationOdor Detection and Analysis using GC/SAW znose
Complimentary Copy Odor Detection and Analysis using GC/SAW znose Edward J. Staples Electronic Sensor Technology 1077 Business Center Circle Newbury Park, California 91320 Ph. 1-805-480-1994 FAX 1-805-480-1984
More informationIntroduction. Chapter 1. Learning Objectives
Chapter 1 Introduction Learning Objectives To understand the need to interface liquid chromatography and mass spectrometry. To understand the requirements of an interface between liquid chromatography
More informationWhite Paper. Overview: NDIR Definition:
Title: NDIR Technology Overview, Compliance, and Comparison to Other Generally Available Gas Measurement Technologies TSN Number: 06 File:\\MII- SRV1\Metron\Bridge_Analyzers\Customer_Service_Documentation\White_Papers\06
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Applied Biosystems QStar Pulsar i Features of the API QSTAR Pulsar i The
More informationCHAPTER 1. MEASURES OF ATMOSPHERIC COMPOSITION
1 CHAPTER 1. MEASURES OF ATMOSPHERIC COMPOSITION The objective of atmospheric chemistry is to understand the factors that control the concentrations of chemical species in the atmosphere. In this book
More informationThe end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry
The end of mass-speculation Pre-cleaned and certified vials for mass spectrometry Bonded, ultra-pure closures LC/MS and GC/MS Manufacturing the Ultimate Vial for Mass Spectrometry More than a quarter century
More informationUltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation
UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation Technical Overview Introduction Inert flow path technology Modern GC and GC/MS instrumentation is an important analytical tool
More informationA STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC
A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC Aysel T. ATIMTAY Middle East Technical University, Environmental Engineering Department, 06531 Ankara Turkey (aatimtay@metu.edu.tr)
More informationTHEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES
THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES J Kouzaki 1*, S Sato 1, S Nakai 1, Y Shirasuna 2, K Hirano 2 1 Graduate School of Environmental and Information
More informationSERVOTOUGH SpectraScan
SERVOTOUGH SpectraScan Continuous, real-time analysis of light hydrocarbons Jan Hordijk - Servomex Hydrocarbon stream analysis 5 major technologies used in HydroCarbon land for gas: IR (Infra Red) Laser
More informationDetermination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC
Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 77A mini TD and 70A GC Application Note Environmental Authors Tingting Bu, Xiaohua Li Agilent Technologies (Shanghai) Co.,
More informationGAS CHROMATOGRAPHY. Mobile phase is a gas! Stationary phase could be anything but a gas
GAS CHROMATOGRAPHY Mobile phase is a gas! Stationary phase could be anything but a gas Gas Chromatography (GC) GC is currently one of the most popular methods for separating and analyzing compounds. This
More informationSTANDARD OPERATING PROCEDURES
PAGE: 1 of 12 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
More informationMethod for the determination of 1,3-butadiene
Federation of the Employment Accidents Insurance Institutions of Germany (Hauptverband der Berufsgenossenschaften) Centre for Accident Prevention and Occupational Medicine Alte Heerstraße 111, 53757 Sankt
More information