Chapter 11 Conventional Gas Chromatography

Size: px
Start display at page:

Download "Chapter 11 Conventional Gas Chromatography"

Transcription

1 Chapter 11 Conventional Gas Chromatography

2 Gas Chromatography GC is the first instrumental chromatographic method developed commercially It is relatively easy to introduce a stable flow and pressure for the mobile phase, that is the carrier gas The equipment is simple. All that is really needed is cylinder of compressed air, pressure regulator and a valve

3 Introduction Principles Partition of molecules between mobile phase and stationary phase Separation technique Gas is the mobile phase and liquid (GLC) or solid (GSC) is the stationary phase GLC: liquid is coated on an inert solid; separation is the result of variable solubilities in the liquid phase GSC: Particulate solid like molecular sieve is the stationary phase; separation is the result of variable adsorption on the solid surface

4 Abundanc e Gas Chromatography E A D Gas Chromatograph C Sample: mixture of volatile liquids (~1 L) B Gas Chromatogram A B C E D Time (minutes)

5 Uses of GC Separation and analysis of organic compounds Testing purity of compounds Determine relative amounts of components in mixtures Compound identification Isolation of pure compounds (micro scale work)

6 Advantages of the GC Speed: minutes or seconds Resolution: complex samples Sensitivity: 10-9 or even g/s Versatility: Gases, liquids or solids (qualitative & quantitative) GLC is more common than GSC in terms of flexibility and resolution Salts and other ionic compounds + high molecular weight may not be determined by the GC. They can be determined by liquid chromatography or pyrolysis chromatography

7 GC Process 1. Column is selected, packed with liquid phase, and installed 2. Sample injected with microliter syringe into the injection port where it is vaporized and mixed into the carrier gas stream (helium, nitrogen, argon). 3. Sample becomes partitioned between the moving gas phase and the stationary liquid phase. 4. The time spent by different compounds of the sample in vapor phase is a function of their vapor pressure 5. The more volatile compounds arrive at the end of the column first and pass into the detector

8 Factors Affecting Separation Boiling Points of Components in Sample Low boiling compounds have higher vapor pressures. Boiling point increases with increasing molecular weight Flow Rate of Carrier Gas Choice of Liquid Phase (Solubility in the liquid stationary phase determines the retention time in the stationary phase) Molecular weights, functional groups, and polarities of component molecules are factors in selecting liquid phase. Length of Column Similar compounds require longer columns than dissimilar compounds. Isomeric mixtures often require quite long columns

9 Mechanism of Separation of Components of a Mixture by Chromatography Stationary Phase Start Mobile Phase Affinity of molecule For the stationary Phase determines Residence time In this phase. high affinity low affinity

10 GC Instrumentation Carrier gas Sample Injector Column Detector

11 Schematic Diagram of a Gas Chromatograph

12 RESET Gas Chromatography System Filters/Traps Data system H Regulators Syringe/Sampler Inlets Air Hydrogen Gas Carrier Column Detectors gas system Inlet (Injector) column detector Readout system

13

14 Gas Flow

15 Carrier Gas Carrier gases, must be chemically inert, Include helium, argon, nitrogen, carbon dioxide, and hydrogen. The choice of gases is often dictated by the detector used. Associated with the gas supply are pressure regulators, gauges, and flow meters. The carrier gas system often contains a molecular sieve to remove water or other impurities.

16 Purity of Carrier Gases Impurities (particularly O 2 & H 2 O) can chemically change the liquid phase and thus the t R and reduce the lifetime of the column. False peaks may appear Liquid stationary phases (polyesters, polysiloxanes, polyamides) degrade by O 2 & H 2 O Contaminant from column may desorb in H 2 O causing a high detector background (baseline drift and noise) Traces of hydrocarbons cause a high background in the FID Purity should be >99.995%

17 Carrier gas: mainly He and N 2 Two stage pressure regulator 1. It indicates the pressure left in the cylinder (min. 40 psi) 2. It indicates the increasing pressure delivered to the GC (min. = 20 psi)

18 Function Introduce sample Vaporize sample Split sample Injection systems Main Components It is a metal block containing: Heaters Temperature sensors Septum holder on the front Connection for the column on the rear

19 Packed column injection Injector is a simple portion of the GC system when packed columns are used There are two basic approaches for these injectors Injection ports Sampling loops/ valves

20 Various Types of Injection ports Split - only a small portion (10:1 and 100:1) of the injected sample goes on column Splitless - all material injected goes on column Usually split/splitless modes are provided to the same instrument. Cool on-column injection (Direct on column cold injection. It is used for sensitive materials) Programmed temperature - sensitive materials (more durable method than OC)

21 Rubber septum serves for about 30 injections in ordinary care 5-10 injections in case of large syringes

22

23 Injection Port/ Temperature Effects The purpose of the port is to flash vaporize the sample and introduce it into the column T injector should be > 50 o C above T column Septum of the injector must be: Stable at the T inj Replaced regularly to maintain seal

24 Syringe Injection Methods Major source of precision error results from poor injection technique Both automatic and manual injection methods are available If automatic equipment is present, use it. If not, several approaches can be tried to help reduce injection errors

25 Injection of liquid sample 10 µl syringe is the most popular device Load the desired volume of liquid, then draw the plunger back to pull the liquid out of the needle Insert the needle quickly through the septum as far as it will go Depress the plunger and immediately remove the needle from the injection port

26 Gas Sampling Loops (Sampling Valves) Introducing a constant amount of a gas can be difficult with a syringe Gas sampling loops and valves offer a high precision (+/- 1%) means of introducing gases Equipment is relatively inexpensive and only requires a constant temperature for easy use

27 Gas Sampling Loops/Valves Valves give better reproducibility Require less skill Can be easily automated

28 Injection valve

29 Injection Port Temperature The temp. should be o C hotter than the boiling point of the least volatile component But low enough to prevent sample decomposition and septum bleed Temp. may be checked by raising it and watching : Position, or area, or shape of the peaks. Drastic changes mean the temp. setting is high It should be 10% above that of the column to ensure rapid volatilization of the sample. The efficiency of the column is almost constant under this condition Components may be vaporized at a temp. ~100 o C below its atmospheric boiling point

30 Very high boiling point or temp. sensitive material can be handled by dilution with volatile solvent that permits lowering the injection temp. This will lower the sensitivity! Try various temperatures until peak broadening becomes apparent With temp. programming techniques low injection temps become very practical. No rush to vaporize the high boiling components

31 Effect of injection port temperature on resolution a: Methanol; b: Ethanol; c: ispropanol Boling points: o C

32 Gas Chromatographic Columns 1. Packed Columns for GLC 2. Packed Columns for GSC 3. Capillary Columns

33 Types of columns Packed columns 1/8 1/4 OD, stainless steel or glass tube 6-20 feet in length Capillary (open tubular) columns mm ID meters in length

34

35

36 Capillary (Open tubular columns)

37

38

39 Packed Columns 1. Column material 2. packing material (solid or inert support) 3. Stationary phase

40 Column Material Stainless Steel : most common. adsorbs some compounds particularly polar ones & especially water. Copper tubing. It is good for trace water analysis. Plastics: are limited due to permeability & temperature limit (used for reactive or highly corrosive chemicals H 2 S, HF Teflon, polypropylene and nylon tubing are available. Glass : If glass were not difficult to form into columns & relatively fragile it would be the very best choice for tubing ( used for pesticides & steroid).

41 Solid Support (Stationary Phase Support) Thermally stable Diatomaceous earth is the most common (nearly, all silica) solid support. Diatomaceous earth: Skeletons of microscopic unicellular algae (diatoms) consisting of microamorphous hydrous silica marketed under Chromosorb Polar analyte species such as, Alcohols or Aromatic hydrocarbons are adsorbed physically on the silicate surfaces of diatomaceous earth. Adsorption results in distorted peaks broadened with tail This catalytic activity may lead to sample decomposition

42 Reasons for adsorption activity Silicates + Water Silanol groups on the silicate surface OH OH OH OH Si O Si O Si O Si Si-OH groups have strong affinity for polar organic molecules

43 Treatment of Solid Supports Non-acid washed (NAW) an untreated form Acid washed (AW) use HCl Removes metals, impurities, Reduces surface activity and absorption Acid washed Dimethyldichlorosilane treated (AW-DMCS) Cl OH Si + (CH 3 ) 2 SiCl 2 HCl CH 3 O Si CH 3 Si CH 3 CH 3 O Si CH 3 Si HCl CH 3 OH

44 Stationary Liquid Phase

45 Characteristics of Stationary Liquid Phase The stationary phase should provide separation of the sample with a reasonable column life Suitable phase is chosen on the basis of : Experience or Experiment. It is desirable to have maximum information about the sample composition : bp.range, components expected & their structure Stationary phases should have similar chemical structure to the sample components

46 ( is the separation factor)

47 Commonly Used Liquid Phases PHASE TEM. LIMITS Good for 1. SQUALANE 0/125oC Nonpolar 2. OV-1, SE /350 oc 3. DEXSIL /350 Oc (Most thermally stable) 4. OV-17; SP Moderately polar 5. QF-1; OV-210; SP CARBOWAX-20M 60/225 Strongle polar 7. DEGS 20/ OV /250

48

49 Squalane Saturated, highly branches, C-30 hydrocarbon Non-polar Limited temperature range: oc Separate hydrocarbons Standard reference for Rohrschneider and McReynolds constants

50 Non-polar Temperature range: oC Most widely used liquid phase Separate all sample types OV-1 ( SE-30 )AND SP-2100

51 OV-17 (SP-2250) 50% methyl, 50% phenylpolysiloxane Semi-polar Temperature range : 0-350oC Widely used to separate drugs, steroids, carbohydrates

52 Carbowax 20-M HO ( - CH 2 CH 2 O-) n H Polymeric polyethylene glycol Polar Temperature range : oC Widely used for polar samples

53 Column Temperature 1. As the Column temp. increases a sample component spends more time in the mobile phase. * This will cause a decrease in the t R (Faster separation) * t R doubles for every 30 o C decrease. 2. Increasing the temp. decreases the band broadening since it leads to a decrease in the available time for diffusion in the column. 3. The lower the temperature the better is the separation * The column temp. should not be less than about 10 o C below the bp of the highest-boiling sample component (other wise distorted peaks may result). * Roughly, a temp. equal or slightly above the average b.p of a sample results in a reasonable elution time (20 to 30 min.). 4. Optimum column temp. depends upon: b.p. of the sample Degree of separation required.

54 Isothermal and temperature programming analysis The column is placed in the oven If the temperature is held constant during the entire analysis then it is called analysis under an isothermal mode. If the temperature is varied during the analysis, then the process is called temperature programming

55 Column Temperature Effect (Isothermal Analysis) Isothermal chromatographic analysis is one which is performed at a constant column temperature.

56 Conclusion Higher temp. enables rapid analysis but loss in resolution. Lower temp. achieves better resolution but longer analysis time

57 Narrow Boiling Range Samples Isothermal column temperature should be used. Select temperature o C lower than boiling range of sample when thin films are handled. Use highest temperature that still allows adequate resolution and stability to shorten analysis time.

58

59 PERCENT LOADING OF STATIONARY PHASE How much liquid phase should be coated on the support? For analytical columns: 5 10% Higher loading ( up to ~ 30% ) can be applied for light gases (C 1 to C 4 hydrocarbons) Low Loading (1-5% ) with very high boiling compounds The liquid phase should cover the support almost completely. If the support is inert, this condition is not a must

60

61 LIQUID PHASE AND TEMPERATURE LIMIT To avoid decomposition, column bleeding and high retention Practical Temperature Limits depend upon the following: * Nature of liquid phase * Percent loading * Sensitivity of detector Temp. limits reported in most cases for TCD Limits; for FID may be about 100 o C less than that for the TCD.

62 Conditioning the column Why? Remove: impurities in liquid phase ; residual solvent from coating How? Install the column but do not connect to the detector Set the carrier flow 30 ml/min for 1/8 inch Heat for 1 hr at 100 o C Raise the temperature to slightly below the b.p. of liqid phase (about 30 o C higher than expected operating column operation temp.(. Continue heating over night Cap the column when it is removed from the instrument

63 Deatectors Features of Detectors A device that measures physical properties (preferred), not chemical properties The detector generates an electrical signal proportional to the sample concentration Detector and connections must be hot enough (20 to 30 o C above the column temp. or the boiling point of the highest boiling component) so that condensation of the sample or liquid phase does not occur

64 Peak broadening or disappearance is characteristic for condensation in the connections Ionization type detectors must be maintained at temp. high enough to avoid not only condensation of sample but also the water or by-products formed in thr ionization process

65 Most Common GC Detectors Most common detectors roughly in order from most common Thermal Conductivity Detector (TCD or hot wire detector), Flame Ionization Detector (FID), Electron Capture Detector (ECD), Photo Ionization Detector (PID), Flame Photometric Detector (FPD), Thermionic Detector VERY expensive choices: Atomic Emission Detector (AED) Ozone- or Fluorine-Induced Chemiluminescence Detectors. All of these (except the AED) produce an electrical signal that varies with the amount of analyte exiting the chromatographic column. Fourier Transform Infrared Detector (FTIR) Mass Spectrometer (MS) Other: UV, FT-NMR

66 Schematic of a thermal conductivity detector, TCD

67 Two pairs of TCDs are used in gas chromatographs. One pair is placed in the column effluent to detect the separated components as they leave the column. Another pair is placed before the injector or in a separate reference column. The resistances of the two sets of pairs are then arranged in a bridge circuit. The heated element may be a fine platinum, gold, or tungsten wire or, alternatively, a semi conducting thermistor. The resistance of the wire or thermistor gives a measure of the thermal conductivity of the gas. Elution heat loss increased resistance needed to balance bridge = recorded

68 Flame Ionization Detector Basic Principle The effluent from the column is mixed with hydrogen and air, and ignited. Organic compounds burning in the flame produce ions and electrons which can conduct electricity through the flame. A large electrical potential is applied at the burner tip, and a collector electrode is located above the flame. The current resulting from the pyrolysis of any organic compound is measured which is proportional to the carbon content of the molecule entering.

69 Ring electrode: stainless Steel gauze (+ve electrode) Flame jet serves As Ve Electrode (FID) sample burned in H 2 /air flame sample must be combustible must use electrometer ppm sensitivity destructive

70 Qualitative analysis by Chromatographic methods Qualitative analysis is based on retention data Retention time t R is characteristic of a substance, compared to a standard. Reproducibility of retention depends upon several experimental conditions: column length and diameter, stationary and mobile phases, column packing, column temperature, mobile phase flow rate and others

71 Retention time, t R t R : It is the time elapsed from the point of injection to the peak maximum Adjusted t R : It is the time from the maximum of unretained peak (the peak of the mobile phase or the air peak) to the peak maximum of a certain component t M (hold up time): is the time required for the mobile phase to be eluted completely from the column

72 Same column under Same conditions Has been used Unknown alcohol components Standard sample

73 Component 1 is used as the reference; it should be present or added to the sample and compatible with the sample Peak of component 1 must be close (but resolved) to the sample peak

74 When component 3 is suspected, add more of this component to the sample and watch any change in its peak

75 Basis for Quantitative Analysis The peaks in the chromatogram are the basis for quantitative analysis Peaks of interest should fulfill the following requirements: must be undistorted must be well separated Must have a large S/N ratio must have a flat baseline Peak shape: The ideal chromatographic peak is symmetric and narrow Peak integration The peak height or, better, the area must be determined and this is done by the computer Calculation External standard method Internal standard method Internal normalization

76 External standard method This method is common to most quantitative analysis techniques. It allows the measurement of the concentration of one or more components that elute in a chromatogram containing, perhaps, many peaks. This method, employing the absolute response factor, K, is used in the following way (Single point calibration method)

77 Multilevel calibration In a multilevel calibration several different amounts of the standard are prepared and analyzed. A regression method is used (e.g. linear least-square) and this leads to a more precise value for C unk. This quantitative method is the only one adapted to gaseous samples. This simple method is used in industry for repetitive analyses. For such analyses, chromatograph must be equipped with an autosampler,

78 Multilevel External Calibration of Fatty Acids Detector Response C 18 Peak Area (cm 2 ) 10 C C Retention Time The content % of C fatty acids = 14 C C + C + C Sample Concentration (mg/ml) = the content % of C fatty acids 14

Instrumentation. Components of a gas chromatograph

Instrumentation. Components of a gas chromatograph Gas chromatography Instrumentation Components of a gas chromatograph The components include the Mobile phase (Carrier gas) supply and pressure and flow rate regulators Injector the column the detector

More information

Gas Chromatography (GC)

Gas Chromatography (GC) Gas Chromatography (GC) Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11541 Saudi Arabia Office: AA53

More information

2401 Gas (liquid) Chromatography

2401 Gas (liquid) Chromatography 2401 Gas (liquid) Chromatography Chromatography Scheme Gas chromatography - specifically gas-liquid chromatography - involves a sample being vaporized and injected onto the head of the chromatographic

More information

GAS CHROMATOGRAPHY. Mobile phase is a gas! Stationary phase could be anything but a gas

GAS CHROMATOGRAPHY. Mobile phase is a gas! Stationary phase could be anything but a gas GAS CHROMATOGRAPHY Mobile phase is a gas! Stationary phase could be anything but a gas Gas Chromatography (GC) GC is currently one of the most popular methods for separating and analyzing compounds. This

More information

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC Gas Chromatography Gas Chromatography Presented By Mr. Venkateswarlu Mpharm KTPC What is Gas Chromatography? It is also known as Gas-Liquid Chromatography (GLC) GAS CHROMATOGRAPHY Separation of gaseous

More information

CHAPTER 6 GAS CHROMATOGRAPHY

CHAPTER 6 GAS CHROMATOGRAPHY CHAPTER 6 GAS CHROMATOGRAPHY Expected Outcomes Explain the principles of gas chromatography Able to state the function of each components of GC instrumentation Able to state the applications of GC 6.1

More information

Chapter 31 Gas Chromatography. Carrier Gas System

Chapter 31 Gas Chromatography. Carrier Gas System Chapter 31 Gas Chromatography GAS-LIQUID CHROMATOGRAPHY In gas chromatography, the components of a vaporized sample are fractionated as a consequence of being partitioned between a mobile gaseous phase

More information

Chapter 27: Gas Chromatography

Chapter 27: Gas Chromatography Chapter 27: Gas Chromatography Gas Chromatography Mobile phase (carrier gas): gas (He, N 2, H 2 ) - do not interact with analytes - only transport the analyte through the column Analyte: volatile liquid

More information

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed Gas Chromatography (GC) In gas chromatography, the sample is vaporized and injected onto the head of a chromatographic

More information

GC Instruments. GC Instruments - Sample Introduction

GC Instruments. GC Instruments - Sample Introduction GC Instruments 1 Fairly simple instrumentation Maintaining constant average pressure is important! Pressure controls flow rate T influences retention (k ) Flow rate monitoring Changing flow rate changes

More information

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5 Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH 5730-6730 Analysis Topic 5 Chromatography! Group of separation techniques based on partitioning (mobile phase/stationary phase). Two immiscible

More information

Chromatography. Gas Chromatography

Chromatography. Gas Chromatography Chromatography Chromatography is essentially the separation of a mixture into its component parts for qualitative and quantitative analysis. The basis of separation is the partitioning of the analyte mixture

More information

CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian

CH 2252 Instrumental Methods of Analysis Unit V  Gas Chromatography.  M. Subramanian CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography M. Subramanian Assistant Professor Department of Chemical Engineering Sri Sivasubramaniya Nadar College of Engineering Kalavakkam 603

More information

10/27/10. Chapter 27. Injector typically 50 C hotter than oven

10/27/10. Chapter 27. Injector typically 50 C hotter than oven Sample and solvent are vaporized onto the head of a column Vaporized solvent and solute are carried through the column by an inert gas (mobile phase) The mobile phase does not interact with compounds of

More information

Gas Chromatography. Introduction

Gas Chromatography. Introduction Gas Chromatography Introduction 1.) Gas Chromatography Mobile phase (carrier gas) is a gas - Usually N 2, He, Ar and maybe H 2 - Mobile phase in liquid chromatography is a liquid Requires analyte to be

More information

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY Chapter 23. Gas Chromatography What did they eat in the year 1,000? GC of Cholesterol and other lipids extracted from

More information

Introduction to Gas Chromatography

Introduction to Gas Chromatography Introduction to Gas Chromatography 31-1 Objectives To know what is chromatography To understand the mechanism of compound separation To know the basic of gas chromatography system 31-2 Chromatography Definition

More information

Gas chromatography. Advantages of GC. Disadvantages of GC

Gas chromatography. Advantages of GC. Disadvantages of GC Advantages of GC Gas chromatography Fast analysis, typically minutes Effi cient, providing high resolution Sensitive, easily detecting ppm and often ppb Nondestructive, making possible on - line coupling;

More information

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Chemistry Instrumental Analysis Lecture 27. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 27 Gas Chromatography Introduction GC covers all chromatographic methods in which the mobile phase is gas. It may involve either a solid stationary phase (GSC)

More information

Disadvantage: Destructive Technique once analyzed by GC, the sample is lost

Disadvantage: Destructive Technique once analyzed by GC, the sample is lost Gas Chromatography Like other methods of chromatography, a partitioning of molecules must occur between the stationary phase and the mobile phases in order to achieve separation. This is the same equilibrium

More information

Chemistry Gas Chromatography: Separation of Volatile Organics

Chemistry Gas Chromatography: Separation of Volatile Organics Chemistry 3200 Gas chromatography (GC) is an instrumental method for separating volatile compounds in a mixture. A small sample of the mixture is injected onto one end of a column housed in an oven. The

More information

Understanding Gas Chromatography

Understanding Gas Chromatography Understanding Gas Chromatography What is Really Going on Inside the Box? Simon Jones GC Applications Engineer Page 1 Group/Presentation Title Month ##, 200X ?? K? Page 2 Typical GC System Gas supply Injector

More information

Gas Chromatography. Chromatography Laboratory Course. Dr. Christian Jungnickel Chromatography Course GC September 2005

Gas Chromatography. Chromatography Laboratory Course. Dr. Christian Jungnickel Chromatography Course GC September 2005 Gas Chromatography Chromatography Laboratory Course The laboratory course experiments General Aim: Gain general experience using a GC Constant Injection technique Temperature variations Qualitative and

More information

Chapter 1. Chromatography. Abdul Muttaleb Jaber

Chapter 1. Chromatography. Abdul Muttaleb Jaber Chapter 1 Chromatography Abdul Muttaleb Jaber What is Chromatography? Chromatography is a physico-chemical process that belongs to fractionation methods same as distillation, crystallization or fractionated

More information

Chemistry Instrumental Analysis Lecture 28. Chem 4631

Chemistry Instrumental Analysis Lecture 28. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 28 Two types in general use: -packed (stationary phase) -open tubular or capillary determine selectivity and efficiency of the sample. Column Materials Column

More information

Ch24. Gas Chromatography (GC)

Ch24. Gas Chromatography (GC) Ch24. Gas Chromatography (GC) 24.1 What did they eat in the year 1000? From 13 C content of cholesterol in ancient bone 13 C : 1.1%, 12 C: 98.9% 13 C/ 12 C ratio types of plants Bones of 50 people in Barton-on-Humber

More information

GC Instruments. GC Instruments - Columns

GC Instruments. GC Instruments - Columns GC Instruments 1 Fairly simple instrumentation Maintaining constant average pressure is important! Pressure controls flow rate T influences retention (k ) Flow rate monitoring Changing flow rate changes

More information

Principles of Gas- Chromatography (GC)

Principles of Gas- Chromatography (GC) Principles of Gas- Chromatography (GC) Mohammed N. Sabir January 2017 10-Jan-17 1 GC is a chromatographic technique utilizes gas as the mobile phase which is usually an inert gas (Hydrogen, Helium, Nitrogen

More information

Gas Chromatography. Rosa Yu, David Reckhow CEE772 Instrumental Methods in Environmental Analysis CEE 772 #16 2

Gas Chromatography. Rosa Yu, David Reckhow CEE772 Instrumental Methods in Environmental Analysis CEE 772 #16 2 Print version Gas Chromatography Rosa Yu, David Reckhow CEE772 Instrumental Methods in Environmental Analysis CEE 772 #16 1 Contents The primary components to a GC system 1. Carrier Gas System (including

More information

Chapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications

Chapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications Chapter 27: Gas Chromatography Principles Instrumentation Detectors Columns and Stationary Phases Applications GC-MS Schematic Interface less critical for capillary columns Several types of Mass Specs

More information

Biochemistry. Biochemical Techniques. 12 Gas Liquid Chromatography

Biochemistry. Biochemical Techniques. 12 Gas Liquid Chromatography Description of Module Subject Name Paper Name 12 Module Name/Title 12 Gas - liquid Chromatography 1. Objectives 1.1 To understand principle of Gas Liquid Chromatography 1.2 To explain the different components

More information

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC CHROMATOGRAPHY Laboratory technique for the Separation of mixtures Chroma -"color" and graphein

More information

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011 ?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention

More information

Gas Chromatography CHEM Dr. Reem M. Alghanmi st term

Gas Chromatography CHEM Dr. Reem M. Alghanmi st term Gas Chromatography CHEM 313-5 Dr. Reem M. Alghanmi 2017 1 st term 17.7 Gas Chromatography Introduction There are two types of gas chromatography: Gas-solid (adsorption) chromatography. Gas-liquid (partition)

More information

7 INSTRUMENTAL CHROMATOGRAPHY

7 INSTRUMENTAL CHROMATOGRAPHY 7 INSTRUMENTAL CHROMATOGRAPHY 7.1 Introduction There are two forms of chromatography, very widely used in analytical laboratories, which rely on electronic control of the process and detection of the species.

More information

Introduction to Capillary GC

Introduction to Capillary GC ?? Kβ? Page 1 Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 2 CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention

More information

GAS CHROMATOGRAPHY (GC)

GAS CHROMATOGRAPHY (GC) GAS CHROMATOGRAPHY (GC) Pre-Lab Questions Questions are to be answered before the beginning of the laboratory. The answers are due at the beginning of each experiment (the questions are for credit and

More information

Luminescence transitions. Fluorescence spectroscopy

Luminescence transitions. Fluorescence spectroscopy Luminescence transitions Fluorescence spectroscopy Advantages: High sensitivity (single molecule detection!) Measuring increment in signal against a dark (zero) background Emission is proportional to excitation

More information

Principles of Instrumental Analysis

Principles of Instrumental Analysis Principles of Instrumental Analysis Chapter 27 Gas Chromatography Gas Chromatography (GC): vaporized analytes (solutes) are partitioned between a mobile gaseous phase and a liquid or a solid stationary

More information

Gas Chromatography. A schematic diagram of a gas chromatograph

Gas Chromatography. A schematic diagram of a gas chromatograph Gas Chromatography In gas liquid chromatography (GLC) partition of solutes occurs between a mobile gas phase (the "carrier gas") and a stationary liquid phase present in the column. The gas-phase concentration

More information

Volatile organic compounds (VOCs):

Volatile organic compounds (VOCs): Volatile organic compounds (VOCs): Organic chemicals with a high vapour pressure at room temperature. High vapour pressure results from a low boiling point. The World Health Organization (WHO) defined

More information

Chapter content. Reference

Chapter content. Reference Chapter 7 HPLC Instrumental Analysis Rezaul Karim Environmental Science and Technology Jessore University of Science and Technology Chapter content Liquid Chromatography (LC); Scope; Principles Instrumentation;

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

Skoog/Holler/Crouch Chapter 26 Principles of Instrumental Analysis, 6th ed. CHAPTER 26

Skoog/Holler/Crouch Chapter 26 Principles of Instrumental Analysis, 6th ed. CHAPTER 26 Skoog/Holler/Crouch Chapter 26 Principles of Instrumental Analysis, 6th ed. Instructor s Manual CHAPTE 26 26-1. (a) Elution is a process in which species are washed through a chromatographic column by

More information

High Pressure/Performance Liquid Chromatography (HPLC)

High Pressure/Performance Liquid Chromatography (HPLC) High Pressure/Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) is a form of column chromatography that pumps a sample mixture or analyte in a solvent (known as the

More information

Activity in the FID Detection Port: A Big Problem if Underestimated

Activity in the FID Detection Port: A Big Problem if Underestimated Activity in the FID Detection Port: A Big Problem if Underestimated Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands It is commonly known in gas chromatography, that many problems can be

More information

Experiment 6 Simple and Fractional Distillation

Experiment 6 Simple and Fractional Distillation Experiment 6 Simple and Fractional Distillation Vapor Pressure vs Temperature of Water Vapor Pressure vs Temperature of Water 25 Vapor Pressure vs Temperature of Water 25 Vapor Pressure (kpa) (kpa) 2 2

More information

Instrumental Chemical Analysis

Instrumental Chemical Analysis L2 Page1 Instrumental Chemical Analysis Chromatography (General aspects of chromatography) Dr. Ahmad Najjar Philadelphia University Faculty of Pharmacy Department of Pharmaceutical Sciences 2 nd semester,

More information

Lab 3 Guide: Gas Chromatography (GC) (Sept 8-14)

Lab 3 Guide: Gas Chromatography (GC) (Sept 8-14) Lab 3 Guide: Gas Chromatography (GC) (Sept 8-14) How GC works The Basic Idea Gas chromatography (GC) is mainly used for the qualitative analysis of samples: it answers the question What chemicals are present

More information

Determination of Volatile Substances Proof of Food Adulteration

Determination of Volatile Substances Proof of Food Adulteration ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of Volatile Substances Proof of Food Adulteration (method: gas chromatography with mass spectrometric detection) Exercise guarantor:

More information

Gas Chromatography notes

Gas Chromatography notes Gas Chromatography notes 1. Here is some terminology related to this experiment: Chromatography is a technique in which compounds to be separated are distributed between a mobile phase and a stationary

More information

Rapid Catalyst Screening Reactors

Rapid Catalyst Screening Reactors Rapid Catalyst Screening Reactors Rapid screening of catalysts Analysis of a variety of sample types Multi-modes of operation Rx-3050TR Single µ-reactor Rx-3050SR Overview Two types of rapid catalyst screening

More information

Gas Chromatography. Vaporization of sample Gas-solid Physical absorption Gas-liquid Liquid immobilized on inert solid

Gas Chromatography. Vaporization of sample Gas-solid Physical absorption Gas-liquid Liquid immobilized on inert solid Gas Chromatography Vaporization of sample Gas-solid Physical absorption Gas-liquid Liquid immobilized on inert solid Principles Instrumentation Applications 18-1 Retention Volumes Volumes rather than times

More information

Gas chromatography. Flow measurement. Flow control. Injection methods. Flow measurement

Gas chromatography. Flow measurement. Flow control. Injection methods. Flow measurement Gas chromatography Schematic of a packed column gas chromatograph First instrumental chromatographic method developed commercially. Reason - it is relatively easy to produce a stable flow and pressure

More information

https://www.chemicool.com/definition/chromatography.html

https://www.chemicool.com/definition/chromatography.html CHROMATOGRAPHY 1 Chromatography - a physical method of mixture separation in which the components to be separated are distributed between two phases, one of which is stationary (stationary phase) while

More information

Gas Chromatography (Chapter 2 and 3 in The essence of chromatography)

Gas Chromatography (Chapter 2 and 3 in The essence of chromatography) Gas Chromatography 1. Introduction. Stationary phases 3. Retention in Gas-Liquid Chromatography 4. Capillary gas-chromatography 5. Sample preparation and injection 6. Detectors (Chapter and 3 in The essence

More information

Chemistry Instrumental Analysis Lecture 31. Chem 4631

Chemistry Instrumental Analysis Lecture 31. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 31 High Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) High Performance Liquid Chromatography (HPLC) Solvent Delivery

More information

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS 1. List two advantages of temperature programming in GC. a) Allows separation of solutes with widely varying retention factors in a reasonable

More information

Trajan SGE GC Columns

Trajan SGE GC Columns Trajan Scientific and Medical Trajan SGE GC Columns Trajan Scientific and Medical Our focus is on developing and commercializing technologies that enable analytical systems to be more selective, sensitive

More information

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols. An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols. Jaap de Zeeuw, Ron Stricek and Gary Stidsen Restek Corp Bellefonte, USA To analyze basic compounds at

More information

Gas Chromatography (GC)

Gas Chromatography (GC) Gas Chromatography (GC) Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia Building:

More information

Partitioning. Separation is based on the analyte s relative solubility between two liquid phases or a liquid and solid.

Partitioning. Separation is based on the analyte s relative solubility between two liquid phases or a liquid and solid. Chromatography Various techniques for the separation of complex mixtures that rely on the differential affinities of substances for a gas or liquid mobile medium and for a stationary adsorbing medium through

More information

1. Carrier gas supply. - Flow rate measurement

1. Carrier gas supply. - Flow rate measurement Chapter 27 Gas chromatography Gas Chromatography - The components of a vaporized sample are separated as a consequence of being partitioned between a mobile gaseous phase and a liquid or a solid stationary

More information

/Chapter 27.ppt

/Chapter 27.ppt Information given in these slides are, either in part or all, recollection from the followings: http://bionmr.unl.edu/courses/chem421-821/lectures/chapter-2... http://faculty.atu.edu/abhuiyan/course/chem

More information

Selection of a Capillary

Selection of a Capillary Selection of a Capillary GC Column - Series 3 Mark Sinnott Application Engineer March 19, 2009 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector

More information

Fall 2012 Due In Class Friday, Oct. 19. Complete the following on separate paper. Show your work and clearly identify your answers.

Fall 2012 Due In Class Friday, Oct. 19. Complete the following on separate paper. Show your work and clearly identify your answers. CHEM 322 Name Fall 2012 Due In Class Friday, Oct. 19 Complete the following on separate paper. Show your work and clearly identify your answers. General Separations 1. Describe the relative contributions

More information

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note Abstract Area and retention time are the two primary measurements in gas chromatography. The precision with which

More information

Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary.

Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary. Application Note 36720111 Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary PLOT Column Keywords Pulsed Flame Photometric

More information

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis Updated: 3 November 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #14 Chromatography: Theory (Skoog, Chapt. 26, pp.674-693) (Harris, Chapt. 23) (641-664) David Reckhow

More information

High Performance Liquid Chromatography

High Performance Liquid Chromatography High Performance Liquid Chromatography What is HPLC? It is a separation technique that involves: Injection of small volume of liquid sample Into a tube packed with a tiny particles (stationary phase).

More information

NON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25)

NON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25) Gas Chromatography NON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25) The Non-Methane Organic Compounds (NMOC) Analyzer is a gas chromatograph configured for analyzing gaseous samples for total organic

More information

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds Atmospheric Analysis Gases Sampling and analysis of gaseous compounds Introduction - External environment (ambient air) ; global warming, acid rain, introduction of pollutants, etc - Internal environment

More information

Selection of a Capillary GC Column

Selection of a Capillary GC Column Selection of a Capillary GC Column Mark Sinnott Application Engineer March 13, 2008 Page 1 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector Electrometer

More information

Introduction and Principles of Gas Chromatography

Introduction and Principles of Gas Chromatography Introduction and Principles of Gas Chromatography Jaap de Zeeuw Restek, Middelburg, The Netherlands Jaap.dezeeuw@restek.com Definition and Uses of Gas Chromatography GC Components and Types of Columns

More information

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments. Chromatography Primer Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments. At its heart, chromatography is a technique

More information

Capillary GC Column Selection and Method Development A Primer on Column Parameters and Instrument Conditions

Capillary GC Column Selection and Method Development A Primer on Column Parameters and Instrument Conditions Capillary GC Column Selection and Method Development A Primer on Column Parameters and Instrument Conditions Michael D. Buchanan September 11, 2014 sigma-aldrich.com/analytical 2012 Sigma-Aldrich Co. All

More information

Experiment 3: Analysis of Unknown Hydrocarbons by GC, NMR, and Chemical Tests for Unsaturation and Aromaticity

Experiment 3: Analysis of Unknown Hydrocarbons by GC, NMR, and Chemical Tests for Unsaturation and Aromaticity Experiment 3: Analysis of Unknown Hydrocarbons by GC, NMR, and Chemical Tests for Unsaturation and Aromaticity In this experiment you will be determining the structures and checking the purity of the hydrocarbon

More information

G a s C h r o m a t o g r a p h y Determination of Ethanol in Wine by Head-Space Gas Chromatography

G a s C h r o m a t o g r a p h y Determination of Ethanol in Wine by Head-Space Gas Chromatography G a s C h r o m a t o g r a p h y Determination of Ethanol in Wine by Head-Space Gas Chromatography Pibulsongkram Rajabhat University Department of Agro-Industry Faculty of Food and Agricultural Technology

More information

Method for the determination of 1,3-butadiene

Method for the determination of 1,3-butadiene Federation of the Employment Accidents Insurance Institutions of Germany (Hauptverband der Berufsgenossenschaften) Centre for Accident Prevention and Occupational Medicine Alte Heerstraße 111, 53757 Sankt

More information

Thermal Soil Desorber Instructions September 2006

Thermal Soil Desorber Instructions September 2006 The SRI Thermal Soil Desorber is mounted in the valve oven section of the 8610C GC chassis. The 8610C GC shown in the photo is sitting next to the H2-40XR Hydrogen generator on the benchtop. With both

More information

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Page 1 of 13 CIPAC/4105/R GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Prepared for CIPAC by Dr M J Tandy*, P M Clarke and B White (UK) The rapid

More information

Course goals: Course goals: Lecture 1 A brief introduction to chromatography. AM Quality parameters and optimization in Chromatography

Course goals: Course goals: Lecture 1 A brief introduction to chromatography. AM Quality parameters and optimization in Chromatography Emqal module: M0925 - Quality parameters and optimization in is a separation technique used for quantification of mixtures of analytes Svein.mjos@kj.uib.no Exercises and lectures can be found at www.chrombox.org/emq

More information

Fundamentals of GC: Introduction: Sample introduction:

Fundamentals of GC: Introduction: Sample introduction: Fundamentals of GC: Introduction: If you are already familiar with gas chromatography, we would recommend you go visit the Links/References page for more in-depth discussion of chromatography. This page

More information

Introduction to Chromatographic Separations

Introduction to Chromatographic Separations Introduction to Chromatographic Separations Analysis of complex samples usually involves previous separation prior to compound determination. Two main separation methods instrumentation are available:

More information

Analytical Chemistry

Analytical Chemistry Analytical Chemistry Chromatographic Separations KAM021 2016 Dr. A. Jesorka, 6112, aldo@chalmers.se Introduction to Chromatographic Separations Theory of Separations -Chromatography Terms Summary: Chromatography

More information

Paints and varnishes Determination of volatile organic compound (VOC) content. Part 2: Gas-chromatographic method

Paints and varnishes Determination of volatile organic compound (VOC) content. Part 2: Gas-chromatographic method Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 11890-2 Third edition 2013-03-15 Paints and varnishes Determination of volatile organic compound (VOC) content Part 2: Gas-chromatographic method

More information

Chromatography and Functional Group Analysis

Chromatography and Functional Group Analysis Chromatography Chromatography separates individual substances from a mixture. - to find out how many components there are - to match the components with known reference materials - to use additional analytical

More information

Methods of pollution control and waste management - laboratory. Adsorptive removal of volatile organic compounds from gases streams

Methods of pollution control and waste management - laboratory. Adsorptive removal of volatile organic compounds from gases streams Methods of pollution control and waste management - laboratory Adsorptive removal of volatile organic compounds from gases streams Manual for experiment 17 dr Hanna Wilczura-Wachnik and dr inż. Jadwiga

More information

Introduction to Capillary GC

Introduction to Capillary GC Introduction to Capillary GC LC Columns and Consumables Simon Jones Chromatography Applications Engineer February 20, 2008 Page 1 Introduction to Capillary GC t r K c?? Kβ k = - tr t m? t m R s Page 2

More information

High Performance Liquid Chromatography

High Performance Liquid Chromatography Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN

More information

High Performance Liquid Chromatography

High Performance Liquid Chromatography Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 1 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN

More information

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS International Gas Union Research Conference 14 THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS Main author Hironori IMANISHI Tokyo Gas Co., Ltd. JAPAN himanishi@tokyo-.co.jp

More information

LEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT. Thomas Wenzel, Bates College. In-class Problem Set Extraction.

LEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT. Thomas Wenzel, Bates College. In-class Problem Set Extraction. LEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT Thomas Wenzel, Bates College In-class Problem Set Extraction Problem #1 1. Devise a scheme to be able to isolate organic acids, bases

More information

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #24 Special Applications: Chromatographic Retention Time and Environmental Properties (Skoog, nothing)

More information

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID Asian Journal of Chemistry Vol. 21, No. 3 (2009), 1739-1746 Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID PRAVEEN KUMAR BALIYAN*, R.P.

More information

HPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25

HPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25 High Performance Liquid Chromatography (HPLC) Harris Chapter 25 12/1/2005 Chem 253 - Chapter 25 1 HPLC Separation of nonvolatile or thermally unstable compounds. If the analyte/sample can be found to be

More information

Experiment 1: Thin Layer Chromatography

Experiment 1: Thin Layer Chromatography Experiment 1: Thin Layer Chromatography Part A: understanding R f values Part B: R f values & solvent polarity Part C: R f values & compound functionality Part D: identification of commercial food dye

More information

Muliple Gas#3 plus Sulfur GC Configuration

Muliple Gas#3 plus Sulfur GC Configuration The SRI Model 8610C Gas Chromatograph ( GC ) configured as a MultipleGas#3 plus Sulfur is designed to measure H2, O2, N2, CO, CO2, H2O, C1 through C5 hydrocarbons and also H2S, COS/SO2, and other sulfur

More information

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #24 Special Applications: Chromatographic Retention Time and Environmental Properties (Skoog, nothing)

More information

DET REPORT. 1.) CATALYTIC COMBUSTION IONIZATION DETECTION (CCID) - MORE CHARACTERISTICS OF THE SELECTIVITY FOR n PARAFFINS. NO.

DET REPORT. 1.) CATALYTIC COMBUSTION IONIZATION DETECTION (CCID) - MORE CHARACTERISTICS OF THE SELECTIVITY FOR n PARAFFINS. NO. 1.) CATALYTIC COMBUSTION IONIZATION DETECTION (CCID) - MORE CHARACTERISTICS OF THE SELECTIVITY FOR n-paraffins. 2.) FINGERPRINT CHROMATOGRAMS OF GASOLINES BY SELECTIVELY DETECTING ALKANE AND OXYGENATE

More information