Column Selection for Size Exclusion Chromatography (SEC)
|
|
- Bernice Jordan
- 5 years ago
- Views:
Transcription
1 Column Selection for Size Exclusion Chromatography (SEC) Application Solvent Column Page Biological macromolecules (proteins, peptides, nucleic acids, etc.) Buffer etc. KW-800 series KW400 series High performance (solvent-saving) Aqueous SEC (GFC) Biological macromolecules (high MW range) Water-soluble polymers (polyacrylamide, polyethylenie, etc.) Polysaccharides Buffer etc. H O, Buffer Aqueous solution etc. SB-800 HQ series LB-800 series SB-800 HQ series LB-800 series Oligosaccharide, polysaccharides H O, Aqueous solution etc. KS-800 series GS-HQ series 4 4 KF-800 series 4 General polymers THF KF-00 series KF-400HQ series Rapid analysis (solvent-saving) High performance (solvent-saving) LF series High linearity of calibration curve Chloroform K-800 series 48 Organic SEC (GPC) Polar polymers (polyimides, polyvinylpyrrolidones etc.) DMF KD-800 series 0 SB-800 HQ series 8 LB-800 series 8 Analysis at high temperature (polyethylene, polypropylene etc.) ODCB etc. HT-800 series UT-800 series AT-80MS Engineering resin analysis at room temperature [polyamide (Nylon), polyethylene terephthalate (PET) etc.] HFIP HFIP-800 series HFIP-00 series Rapid analysis (solvent-saving) High linearity of calibration curve 0 0 LF series Aqueous/Organic SEC GF-HQ series 44 Guideline for column solvent usability High Polarity of solvent H O Methanol DMSO Acetonitrile DMF THF Chloroform Toluene Low KW-800 series, KW400 series SB-800 HQ series, LB-800 series GS-HQ series GF-HQ series KD series 4 KF, K, LF series * See page for the solvent replaceability of organic solvent SEC (GPC) packed columns.
2 Precautions for Polar Polymer Analysis Size exclusion chromatography analysis of polar polymers can be influenced by unexpected interactions in the column. These interactions may change elution patterns and results in an invalid molecular weight calculation. It is important to reduce them in order to obtain the accurate molecular weight distribution. Interactions between the analyte and the packing materials Hydrophobic interaction The analyte is adsorbed into the packing material. This delays the analyte elution, and thus results in under estimation of its molecular weight (Figure B, D). Ionic interaction () Ion Exclusion The analyte is repelled from the packing material. This accelerates the analyte elution, and thus results in over estimation of its molecular weight (Figure A, C). () Ion Exchange The analyte is adsorbed onto the packing material. This delays the analyte elution, and thus results in under estimation of its molecular weight (Figure B, D). Interaction within and between the analyte Ionic repulsion effects observed within the multivalent macromolecules causes structure expansion This accelerates the analyte elution, and thus results in over estimation of its molecular weight (Figure A). Association between the molecules Associated molecule detected as a larger molecule (Figure A). Interactions between the analyte and the solvent The multivalent ion of the solvent works as a bridge to bind ionic molecules (analyte). Aqueous SEC (GFC) Interfering interactions likely to be observed Exclusion limit molecular weight log MW (C) Methods to reduce interactions Elution will be accelerated Exclusion limit volume (A) Elution will be retarded Organic solvent SEC (GPC) (B) Permeation limit volume (D) Column Selection for Size Exclusion Chromatography (SEC) Precautions for Polar Polymer Analysis Ionic Interaction Add salt Hydrophobic interaction Increase dissociation of the analyte Cationic polymer Lower the ph Anionic polymer Higher the ph Lower the eluent polarity (Example) Add acetonitrile or methanol Ionic Interaction Add salt (Example) Add LiBr to DMF Add CF COONa to HFIP Hydrophobic interaction Lower the polarity of the eluent (Example) Change the eluent from DMF to THF Hydrophillic interaction Increase the polarity of the eluent (Example) Change the eluent from THF to DMF
3
4
5
6
7 Calibration curves for SB-800 HQ series using pullulan (eluent : H O) Calibration curves for SB-800 HQ series using PEG/PEO (eluent : DMF) Carboxymethylcellulose 7 7 Sample : Carboxymethylcellulose 0.% each, 0µL,00-,000cP Molecular weight (Pullulan) 4 SB-804 HQ SB-80 HQ SB-80. HQ SB-80 HQ SB-80 HQ SB-80 HQ SB-80M HQ Molecular weight (PEG/PEO) 4 SB-804 HQ SB-80 HQ SB-80. HQ SB-80 HQ SB-80 HQ SB-80M HQ Mn : 0,00 Mw :,940,00 Mw/Mn : cP Mn :,400 Mw :,,000 Mw/Mn : cP Mn : 4,00 Mw : 40,000 Mw/Mn : * Molecular weight was detered from the calibration curve of pullulan. Column : Shodex OHpak SB-800 HQ series Eluent : H O Column temp. : 0 C Column : Shodex OHpak SB-800 HQ series Eluent : DMF Column : Shodex OHpak SB-80M HQ x Eluent : 0.M NaCl aq. Effects of sodium nitrate in eluent on the analysis of polyallylae Gelatin For the analysis of cationic polymers, such as pollyallylae, undesired adsorption of the polymer is observed when low (0.0M) sodium nitrate eluent was used. By using higher concentration (> 0.M) salt, it suppresses the sample adsorption and enables to obtain accurate chromatograms. Sample : Polyallylae 0.%, 0µL 0. M 0.0 M Sample : 0.% each, 0μL Gelatin from bovine skin (Acid treatment, Gel strength : g) Gelatin from porcine skin (Alkali treatment, Gel strength : 90-0g) 0. M From bovine skin Mn :,400 Mw : 8,0 Mw/Mn : M From porcine skin Mn :,00 Mw : 8,700 Mw/Mn : Volume (ml) 0 0 * Molecular weight was detered from the calibration curve of pullulan. Column : Shodex OHpak SB-80M HQ x Eluent : 0.M Acetic acid + NaNO aq. Column : Shodex OHpak SB-80M HQ x Eluent : 0.M KH PO 4 aq./ 0.M Na HPO 4 aq.=0/0 Polyacrylamide Cellulose acetate SB-80 HQ Sample : Polyacrylamide, 0µL Sample : Cellulose acetate 0.% each, 0μL.0x 8 SB-807 HQ Mn~0,000 Molar Mass (g/mol).0x 7.0x Mw:,00,000 Mn~0, x Volume (ml) 40 Column : Shodex OHpak SB-807 HQ, SB-80 HQ Eluent : 0.M NaCl aq. Flow rate : 0.mL/ MALS (Multi angle light scattering) Column temp. : 0 C Column : Shodex OHpak SB-80M HQ x Eluent : 0mM LiBr in DMF
8 Copovidones Sample : 0μL Poly(-vinylpyrrolidone-co-vinyl acetate) 0.% each Copolymer 7: Mn :,000 Mw : 4,400 Mw/Mn : 7.40 Copolymer :7 Mn :,400 Mw : 8,900 Mw/Mn : * Molecular weight was detered from the calibration curve of PEG/PEO. Column : Shodex OHpak SB-80M HQ x Eluent : 0mM LiBr in DMF Calibration curves for LB-800 series using pullulan (eluent : H O) Molecular weight (Pullulan) 7 4 LB-80 LB-80M Column : Shodex OHpak LB-800 series Eluent : H O Column temp. : 0 C Calibration curves for LB-800 series using PEG/PEO (eluent : DMF) Molecular weight (PEG/PEO) LB-80 LB-80M Column : Shodex OHpak LB-800 series Eluent : DMF Aqueous SEC (GFC) Columns : Polymer-based Comparison of the detection of pullulan using multiangle light scattering detector The multiangle light scattering (MALS) signal of low molecular weight substances is low with competitors columns due to the baseline noise. New product OHpak LB-800 series enables the detection by improving the baseline noise level. LB-80M (8.0mm I.D. x 00mm) Sample : Pullulan (MW :,000) 0.%, µl 0 00 LB-80M (8.0mm I.D. x 00mm) SB-80M HQ (8.0mm I.D. x 00mm) SEC column from other manufacturer (7.8mm I.D. x 00mm) SB-80M HQ (8.0mm I.D. x 00mm) Signal to noise ratio (S/N) SEC column from other manufacturer (7.8mm I.D. x 00mm) 0,40,00,000 0,800 Molecular weight (Pullulan) 0 Column : Shodex OHpak LB-80M Shodex OHpak SB-80M HQ SEC column from other manufacturer Eluent : 0.M NaNO aq. Detector : MALS (Multi angle light scattering) (90 ) Column temp. : 0 C Sodium alginate Sodium heparin Sample : Sodium alginate 0.%, 0µL Sample : Sodium heparin 0.%, 0µL.0x 8.0x 8.0x 7.0x 7 Molar Mass (g/mol).0x.0x.0x Molar Mass (g/mol).0x.0x.0x Mn : 8,790 Mw :,00 Mw/Mn : Mn :,070 Mw :,0 Mw/Mn :, Time () Time () Column : Shodex OHpak LB-80M x Eluent : 0.M NaNO aq. MALS (Multi angle light scattering) Column temp. : 0 C Column : Shodex OHpak LB-80M x Eluent : 0.M NaNO aq. MALS (Multi angle light scattering) Column temp. : 0 C 4
9 Multimode Columns Features GS-HQ SEC is the main separation mode Depends on the eluent selected, the column adds multimode features of reversed phase, HILIC, and ion exchange modes to SEC Suitable for the separation of peptides or nucleic acids with similar molecular weights Suitable for desalting samples or substituting buffer in protein analysis Standard columns Pore Size (Å) Shipping Solvent F70000 Asahipak GS-0 HQ 9, H O/CH OH=70/0 F70000 Asahipak GS-0 HQ 9, H O/CH OH=70/0 F Asahipak GS-0 HQ 7, H O/CH OH=70/0 F Asahipak GS-0 HQ 7 7, H O/CH OH=70/0 F709 Asahipak GS-G 7B H O/CH OH=70/0 Semi-micro columns * The following semi-micro columns are made to order. Base Material : Polyvinyl alcohol Usable ph range : ph- (GS-0 HQ : ph-9) Usable concentration of methanol is up to 0% (GS-0 HQ : up to 0%) Usable concentration of acetonitrile is up to 0% Pore Size (Å) F770 GS0A-4D F770 GS0A-4E F770 GS0A-4B 7 7, Preparative columns * Preparative columns are made to order. Standard Column F807 Asahipak GS-0 0F 8, GS-0 HQ F808 Asahipak GS-0 0F 8, GS-0 HQ F809 Asahipak GS-0 0F 8, GS-0 HQ F800 Asahipak GS-0 0F 8, GS-0 HQ F804 Asahipak GS-0 0G 4, GS-0 HQ F80 Asahipak GS-0 0G 4, GS-0 HQ F80 Asahipak GS-0 0G 4, GS-0 HQ F807 Asahipak GS-0 0G 4, GS-0 HQ F70 Asahipak GS-0G 7B Target molecular weight range and Exclusion limit With Pullulan (eluent : ultrapure water) Target Molecular Weight Range Exclusion Limit GS-0 GS-0 00, ,000 7,000 40,000 4 GS-0 GS-0,000 00,000, ,000 00,000,000,000 * Please use the above table as a rough indication for the column selection.
10 Calibration curves for GS-HQ series using pullulan (eluent : H O) Molecular weight (Pullulan) 7 4 GS-0 HQ GS-0 HQ GS-0 HQ GS-0 HQ Peptides GS-HQ columns can be used not only for SEC (GFC) in an aqueous system, but also for multimodal analysis where hydrophobic interaction and ionic interaction are used together as separation criteria, under certain conditions of the eluent. This results in unprecedented separation analysis. GS-0 HQ columns are excellent in the performance for separating hydrophilic peptides, in particular, acidic or basic peptides, from each other. Glu-Ala-Glu Arg-Asp Gly-His-Lys Arg-Pro-Lys-Pro MW pl Σf Sample : 0μL. Glu-Ala-Glu 0.0%. Arg-Asp 0.0%. Gly-His-Lys 0.0% 4. Arg-Pro-Lys-Pro 0.0% Multimode Columns Σf : Hydrophobic parameter pl : Isoelectric point Column : Shodex Asahipak GS-HQ series Eluent : H O Flow rate : 0.mL/ Column temp. : 0 C Column : Shodex Asahipak GS-0 HQ Eluent : 0mM Ammonium acetate buffer (ph.7) Flow rate : 0.mL/ Detector : UV (0nm) Column temp. : 0 C Analysis of purine bases in beer Purine in foods is analyzed as purine base after a step of sample preparation; homogenization, freeze drying, hydolyzation with 70% perchloric acid, and neutralization. Example below shows the analysis of purin in regular beer and beer treated with guanase (an enzyme that degrades guanine to xanthine). The following data indicate that guanine was decreased and xanthine was increased by guanase. Purine bases in beer Normal beer Guanase treated beer adenine guanine guanine hypoxanthine xanthine xanthine uric acid xanthine Column : Shodex Asahipak GS-0 HQ Eluent : 0mM Sodium phosphate buffer (ph.) Flow rate : 0.mL/ Detector : UV (0nm) Column temp. : C Data provided by Kiyoko Kaneko Ph.D., Faculty of Pharmaceutical Sciences, Teikyo University Low molecular weight water-soluble dietary fiber Umami Lignosulfonic acid By using GS-0 HQ, monosaccharides, disaccharides, and sugar alcohols elute after indigestible component faction (indicated by an arrow on the chromatogram). This separation makes the method preferable for the quantification of low molecular weight water-soluble dietary fiber. Sample : 0μg/mL each, 0μL. IMP. GMP. AMP 4. Inosine. Hypoxanthine. Guanosine 7. Guanine 8. Adenosine 9. Adenine Sample : 0μL Lignosulfonic acid sodium salt 0.% Column : Shodex Asahipak GS-0 HQ x Eluent : H O Flow rate : 0.mL/ Column temp. : 0 C Column : Shodex Asahipak GS-0 HQ Eluent : mm NaH PO 4 aq./mm Na HPO 4 aq. =00/ Detector : UV (0nm) Column : Shodex Asahipak GS-0 HQ x Eluent : 0mM Na HPO 4 aq. Flow rate : 0.mL/ Detector : UV (4nm) 4
11 Aqueous/Organic SEC Columns Features GF-HQ Polymer-based SEC columns exhibit high solvent durability Supports both aqueous and organic solvents Standard columns Pore Size (Å) Shipping Solvent F Asahipak GF- HQ 9, H O F70000 Asahipak GF- HQ 9, H O/CH OH=70/0 F70000 Asahipak GF- HQ 9,000, H O/CH OH=70/0 F70000 Asahipak GF-7 HQ,000 9, H O/CH OH=70/0 F Asahipak GF-7M HQ,000 9, H O/CH OH=70/0 F708 Asahipak GF-G 7B H O/CH OH=70/0 F7000 MSpak GF- 4B, H O F700 MSpak GF- 4D, H O F70004 MSpak GF- 4E, H O F7000 MSpak GF- D, H O GF-7M HQ is a mixed-gel column capable of analyzing samples over a wide range of molecular weight. Base Material : Polyvinyl alcohol Usable ph range : ph-9 Semi-micro columns * The following semi-micro columns are made to order. Pore Size (Å) F70000 Asahipak GF- 4D F7000 Asahipak GF- 4E F770 GFA-4D, F770 GFA-4E, F7707 GF7A-4D 9, Preparative columns * Preparative columns are made to order. Standard Columns F800 Asahipak GS- 0F 8, GF- HQ F80 Asahipak GS- 0F 8, GF- HQ F80 Asahipak GS-7 0F 8, GF-7 HQ F80 Asahipak GSM-700 0F 8, GF-7M HQ F808 Asahipak GS- 0G 4, GF- HQ F809 Asahipak GS- 0G 4, GF- HQ F8040 Asahipak GS-7 0G 4, GF-7 HQ F804 Asahipak GSM-700 0G 4, GF-7M HQ F700 Asahipak GS-G 7B Target molecular weight range and Exclusion limit With Pullulan (eluent : ultrapure water) With *PEG/PEO (eluent : DMF) 44 GF- GF- GF- GF-7 GF-7M * Please use the above table as a rough indication for the column selection. Target Molecular Weight Range 00 4, ,000,000 00,000 0,000 *(,000,000) 00 *(,000,000) Exclusion Limit 9,000 40,000 00,000 *(,000,000) *(,000,000) GF- GF- GF- GF-7 GF-7M Target Molecular Weight Range 0, ,000,000 00,000 0,000 **(,000,000) 00 **(,000,000) *( ) Estimated value * Please use the above *PEG : polyethylene glycol table as a rough indication *PEO : polyethylene oxide for the column selection. **( ) Estimated value
12 Usable solvents Solvent Water 0-0.M sodium concentration Methanol Ethanol Acetonitrile THF DMF GF- GF- GF- GF-7 GF-7M * When replacing acetonitrile, ethyl acetate or chloroform with water, replace with methanol first and then replace with water. * When replacing water with ethyl acetate or chloroform, replace with methanol first and then replace with the required eluent condition. (Note) The usable solvents for preparative columns of GF-7 HQ, GS-7 0F and 0G, are water and methanol. GSM-700 0F or 0G is recommended when other solvents are required. Sodium polystyrene sulfonates MW:4, MW:8, Polymers having both hydrophobic and hydrophilic functional groups may exhibit hydrophobic interactions with packing materials. When analyzing such polymers, the hydrophobic interaction can be suppressed by adding organic solvents to the eluent. Sample : mg/ml each, 0µL Sodiumn polystyrene sulfonate CH CH SONa n Aqueous/Organic SEC Columns Acetone Chloroform Ethylacetate DMSO 0~0% 0 MW:8, Column : Shodex Asahipak GF- HQ Eluent : 0mM LiCl aq./ch CN =0/40 Flow rate : 0.mL/ Detector : UV (4nm) Column temp. : 0 C Calibration curves for GF-HQ series using pullulan (eluent : H O) Calibration curves for GF-HQ series using PEG/PEO (eluent : DMF) Polyacrylonitrile 7 7 Sample : Polyacrylonitrile 0.%, 0µL Molecular weight (Pullulan) 4 GF- HQ GF- HQ GF- HQ GF-7 HQ GF-7M HQ Molecular weight (PEG/PEO) 4 GF- HQ GF- HQ GF- HQ GF-7M HQ GF-7 HQ Column : Shodex Asahipak GF-HQ series Eluent : H O Flow rate : 0.mL/ Column temp. : 0 C Column : Shodex Asahipak GF-HQ series Eluent : DMF Flow rate : 0.mL/ Column : Shodex Asahipak GF-7 HQ x Eluent : 0mM LiBr in DMF Flow rate : 0.mL/ Comparison of two GF columns for the separation performance of maltooligosaccharides Comparison of polystyrenes separation using various solvents GF- HQ demonstrates an improved separation of low molecular substances. The sample below shows that the peaks obtained by GF- HQ are separated with deeper notches compared to peaks obtained by GF- HQ. GF- HQ is capable of separating oligosaccharides (trisaccharides to hexasaccharides) while those oligosaccharides were eluted all together when analyzed by GF- HQ. GF- HQ x GF- HQ x ~ 4 4 Sample : 0.% each, 0μL. Maltohexaose. Maltopentaose. Maltotetraose 4. Maltotriose. Maltose. Glucose (A) THF 4 Sample : Polystyrene mg/ml each, 0µL. Mw :,090,000. Mw : 90,000. Mw : 7, Mw : 9,0. Mw :,980. Mw : 00 (B) Chloroform 4 (C) DMF Column : Shodex Asahipak GF- HQ x Shodex Asahipak GF- HQ x Eluent : H O Flow rate : 0.mL/ Column temp. : 0 C Column : Shodex Asahipak GF- HQ + GF- HQ Eluent : (A); THF, (B); Chloroform, (C); DMF Flow rate : 0.mL/ Detector : (A),(B) UV (4nm), (C) UV (70nm) Column temp. : 0 C 4
13 Organic SEC (GPC) Columns (General Analysis) : THF Features KF-800 Standard organic solvent SEC (GPC) column Supports a wide range of applications from low to high molecular weight compounds Corresponds to USP L Standard columns KF-800 series : Shipping solvent Tetrahydrofuran (THF) Pore Size (Å) F080 GPC KF-80 0 F0800 GPC KF-80 0 F080 GPC KF F0800 GPC KF F0700 GPC KF-80L 00 F08040 GPC KF-804 7,00 F07040 GPC KF-804L 7,00 F0800 GPC KF-80,000,000 F0700 GPC KF-80L,000,000 F0800 GPC KF-80,000,000 F08090 GPC KF-80M,000,000 F0700 GPC KF-80L,000,000 F08070 GPC KF-807, ,000 F07070 GPC KF-807L, ,000 F70000 GPC KF-G 4A (GPC KF-G) 8 4. F7090 GPC KF-800D solvent-peak separation column * The columns with L or M at the end of column names are mixed-gel column capable of analyzing samples over a wide range of molecular weight distribution. * See page 0 for details of the solvent-peak separation columns. * See pages 4 and for details preparative columns. * See page for applicability of SEC (GPC) columns to solvent replacement. Target molecular weight range and Exclusion limit With Polystyrene (eluent : THF) Target Molecular Weight Range Exclusion Limit Target Molecular Weight Range Exclusion Limit KF ,00 KF-80 0,000,000,000 4,000,000 KF-80 00,000,000 KF-80L 00,000,000 4,000,000 KF ,000 0,000 KF-80 0,000 *(0,000,000) *(0,000,000) KF-80,000 0,000 70,000 KF-80M,000 *(0,000,000) *(0,000,000) KF-80L 0 0,000 70,000 KF-80L 00 *(0,000,000) *(0,000,000) KF-804 7,000 00, ,000 KF ,000 *(00,000,000) *(00,000,000) KF-804L 0 00, ,000 KF-807L 00 *(00,000,000) *(00,000,000) * Please use the above table as a rough indication for the column selection. *( ) Estimated value 4
14 Calibration curves for KF-800 series using polystyrene Molecular weight (Polystyrene) 7 4 KF-804 KF-80 KF-80 KF-80 KF-80 KF-80. KF-80 KF-807 KF-80M Column : Shodex GPC KF-800 series Calibration curves for KF-800L (linear type) series using polystyrene Molecular weight (Polystyrene) 7 4 KF-80L KF-80L KF-804L KF-80L KF-807L Column : Shodex GPC KF-800L series Fat-soluble vitas Sample : 0μL. Vita A palmitate 00μg/mL. Vita E 0μg/mL. Vita D 0μg/mL 4. Vita K 0μg/mL 4 0 Column : Shodex GPC KF-80 x Detector : UV (80nm) Organic SEC (GPC) Columns (General Analysis) : THF Standard polystyrenes 4 7 Sample : 0µL. Polystyrene (Mw :,000) 0.%. Polystyrene (Mw :,0) 0.%. Polystyrene (Mw : 9,00) 0.% 4. Polystyrene (Mw : 7,) 0.%. Polystyrene (Mw :,070) 0.%. Polystyrene (Mw : 80) 0.% 7. Toluene (Mw: 9) 0.% 4 Sample : 0μL. Polystyrene (Mw : 9,8,000) 0.0%. Polystyrene (Mw :,0,000) 0.%. Polystyrene (Mw :,000) 0.% 4. Polystyrene (Mw : 7,) 0.%. Polystyrene (Mw : 80) 0.%. Toluene (Mw : 9) 0.% Column : Shodex GPC KF-80L x Detector : UV (4nm) Column : Shodex GPC KF-807L x Detector : UV (4nm) Styrene isoprene ABA block copolymer Polycarbonate resin Raw rubber Sample : Styrene isoprene ABA block copolymer Sample : Polycarbonate resin 0.%, 0µL Sample : Rubber 0.%, 00μL CH O C O C CH O n Column : Shodex GPC KF-80M x Column temp. : 0 C Column : Shodex GPC KF-80L x Column : Shodex GPC KF-80M x + KF-80 Eluent : Toluene Column temp. : Room temp. 47
15 Organic SEC (GPC) Columns (General Analysis) : Chloroform Features K-800 Standard organic solvent SEC (GPC) column Supports a wide range of applications from low to high molecular weight compounds Corresponds to USP L Standard columns K-800 series : Shipping solvent Chloroform Pore Size (Å) F08 GPC K F080 GPC K-80 0 F08 GPC K F080 GPC K F0894 GPC K-80L 00 F0840 GPC K-804 7,00 F089 GPC K-804L 7,00 F080 GPC K-80,000,000 F089 GPC K-80L,000,000 F08 GPC K-80,000,000 F0890 GPC K-80M,000,000 F0897 GPC K-80L,000,000 F0870 GPC K-807, ,000 F0898 GPC K-807L, ,000 F70040 GPC K-G 4A (GPC K-G) 8 4. F70940 GPC K-800D solvent-peak separation column * The columns with L or M at the end of column names are mixed-gel column capable of analyzing samples over a wide range of molecular weight distribution. * See page 0 for details of the solvent-peak separation columns. * See pages 4 and for details preparative columns. * See page for applicability of SEC (GPC) columns to solvent replacement. Target molecular weight range and Exclusion limit With Polystyrene (eluent : Chloroform) Target Molecular Weight Range Exclusion Limit Target Molecular Weight Range Exclusion Limit K ,00 K-80 0,000,000,000 4,000,000 K-80 00,000,000 K-80L 00,000,000 4,000,000 K ,000 0,000 K-80 0,000 *(0,000,000) *(0,000,000) K-80,000 0,000 70,000 K-80M,000 *(0,000,000) *(0,000,000) K-80L 0 0,000 70,000 K-80L 00 *(0,000,000) *(0,000,000) K-804 7,000 00, ,000 K ,000 *(00,000,000) *(00,000,000) K-804L 0 00, ,000 K-807L 00 *(00,000,000) *(00,000,000) * Please use the above table as a rough indication for the column selection. *( ) Estimated value 48
16 Epoxy resin Sample : Epikote TM 0 0.%, 0μL 0 Column : Shodex GPC K-80L x Eluent : Chloroform Detector : UV (4nm) Column temp. : Room temp. Silicon oil Low molecular polydimethylsiloxanes Sample : 0.% each, 0μL. Polydimethylsiloxane (MW:770). Polydimethylsiloxane (MW:7). Hexamethyldisiloxane (MW:) 0 Column : Shodex GPC K-80 x Eluent : Chloroform (polarity : -) 0 Bionolle TM (Polybutylene succinate/adipate) Sample : Silicone oil 0.%, 00µL Sample : Polybutylene succinate/adipate 0.%, 0µL Organic SEC (GPC) Columns (General Analysis) : Chloroform Silicone oil Column : Shodex GPC K-80M x Eluent : Toluene Column temp. : 4 C Column : Shodex GPC K-80M x + K-80 Eluent : Chloroform Recycling fractionation of lauryl stearyl thiodipropionate Sample : %, 00μL. Distearyl stearyl thiodipropionate. Lauryl stearyl thiodipropionate. Dilauryl thiodipropionate CYCLE CYCLE CYCLE CYCLE Column : Shodex GPC K-LG + K-00 Eluent : Chloroform Flow rate :.0mL/ (preparative type) Column temp. : 0 C 0 0 * See page 4 for K-00 49
17 Organic SEC (GPC) Columns (General Analysis) : DMF Features KD-800 Standard organic solvent SEC (GPC) column Supports a wide range of applications from low to high molecular weight compounds Corresponds to USP L Standard columns KD-800 series : Shipping solvent Dimethylformamide (DMF) Pore Size (Å) F08 GPC KD-80 7,000 0 F080 GPC KD-80 7,000 0 F08 GPC KD-80. 7, F080 GPC KD-80 7, F0840 GPC KD-804 7,000 7,00 F080 GPC KD-80,000,000 F080 GPC KD-80,000,000 F0890 GPC KD-80M,000,000 F0870 GPC KD-807, ,000 F7004 GPC KD-G 4A (GPC KD-G) 8 4. * The columns with L or M at the end of column names are mixed-gel column capable of analyzing samples over a wide range of molecular weight distribution. * See page for applicability of SEC (GPC) columns to solvent replacement. Target molecular weight range and Exclusion limit With *PEG/PEO (eluent : DMF) Target Molecular Weight Range Exclusion Limit Target Molecular Weight Range Exclusion Limit KD-80 0,00,00 KD-80 0,000 **(4,000,000) **(4,000,000) KD ,000 7,000 KD-80 0,000 **(40,000,000) **(40,000,000) KD-80. KD-80 KD-804 Features KF-800D K-800D 400,000,000 0,000 4,000 00,000 0,000 70,000 00,000 KD-80M KD-807,000 **(40,000,000) 0,000 **(00,000,000) * Please use the above table as a rough indication for the column selection. Solvent-peak Separation Columns for Organic SEC (GPC) **(40,000,000) **(00,000,000) *PEG : polyethylene glycol *PEO : polyethylene oxide **( ) Estimated value Use this column in combination with a linear column By shifting the elution of monomers, polymer additives, and the solvent-peak in low molecular region, it reduces interferences in the calculation of the molecular weight distribution of polymers and oligomers Solvent-peak separation columns Column Combination Shipping Solvent F7090 GPC KF-800D KF-80L, 80L, 80M, 807L THF F70940 GPC K-800D K-80L, 80L, 80M, 807L Chloroform 0
18 Calibration curves for KD-800 series using PEG/PEO Molecular weight (PEG/PEO) Molecular mass Celluloses KD-80 KD Column : Shodex GPC KD-80M Eluent : % LiCl in *DMI Flow rate : 0.mL/, MALS (Multi angle light scattering) Column temp. : 0 C * DMI,-dimethyl--imidazolidinone Potato starch KD-80 KD-80 Sample : 0μL Potato starch in DMSO 0.% * solved at 80 C KD-80 KD-80 KD-80M KD-80. KD-807 Column : Shodex GPC KD-800 series Eluent : DMF Polyvinylpyrrolidones Sample : Polyvinylpyrrolidone 0.% each (A) K-0 400µL (B) K-0 00µL (C) K-90 00µL (A) (B) (C) Column : Shodex GPC KD-80M x Eluent : mm LiBr in DMF Column temp. : 0 C Sample : 0μL Cellulose ca. 0.0% each Cellulose is known to be difficult to dissolve. A cellulose solution is prepared by repeating solvent replacement. It is reported that it requires a long time to dissolve ( to 0 days), depending on the solvent type, crystallinity and molecular weight. Data provided by Dr. Masahiko Yanagisawa, Isogai group, Graduate School of Agricultural and Life Sciences, The University of Tokyo Effects of solvent-peak separation column Sample : Poly(vinyl chloride) DOP (Dioctyl phthalate) Melae formaldehyde resin Phenoxy resin Sample : 0µL Melae-formaldehyde resin % 0 Column : Shodex GPC KD-80 x Eluent : mm LiBr in DMF Column temp. : 0 C Sample : 0µL Phenoxy resin 0.% 0 Column : Shodex GPC KD-80M x Eluent : mm LiBr in DMF Column temp. : 0 C Organic SEC (GPC) Columns (General Analysis) : DMF Solvent-peak Separation Columns for Organic SEC (GPC) KF-80L 0 Molecular weight (Polystyrene) 4 KF-80L KF-80L+KF-800D KF-80L + KF-800D Column : Shodex GPC KD-80M Eluent : mm LiBr in DMSO/DMF=7/ Column temp. : 0 C 0 Column : Shodex GPC KF-80L Shodex GPC KF-80L + KF-800D
19 Organic SEC (GPC) Columns : Rapid Analysis, High Performance Analysis Features KF-00 KF-400HQ Approximately half of the analysis time compared with standard columns The amount of solvent used is reduced to about a third Improved applicability of solvent replacement Corresponds to USP L About. times better separation performance than standard columns, obtains higher resolution About 4 times better sensitivity than that of standard columns, supports high sensitivity analysis The amount of solvent used is reduced to about a third Improved applicability of solvent replacement Corresponds to USP L Rapid analysis downsized columns KF-00 series F0809 F0809 F0809 F08094 F0809 F0809 F08097 F08098 F08099 F70000 GPC KF-0 GPC KF-0 GPC KF-0. GPC KF-0 GPC KF-04 GPC KF-0 GPC KF-0 GPC KF-0M GPC KF-07 GPC KF-G 4A (GPC KF-G) 7,000 7,000 7,000 7,000,000 7,000 7,000 8,000,000 Use of the KF-00 series with semi-micro type devices is recommended. 8 8 Pore Size (Å) ,00,000,000,000 0, High performance semi-micro columns KF-400HQ series F08 F08 F084 F08 F088 F089 F08 F70000 GPC KF-40HQ GPC KF-40HQ GPC KF-40.HQ GPC KF-40HQ GPC KF-404HQ GPC KF-40LHQ GPC KF-40LHQ GPC KF-G 4A (GPC KF-G),000,000,000,000,000,000,000 Use of the KF-400HQ series with semi-micro type devices is recommended. 8 Pore Size (Å) ,00,000, [KF-00 series and KF-400HQ series] * The columns with L or M at the end of column names are mixed-gel column capable of analyzing samples over a wide range of molecular weight distribution. * See page for applicability of SEC (GPC) columns to solvent replacement. [KF-00 series and KF-400HQ series] Shipping Solvent : Tetrahydrofuran (THF) Target molecular weight range and Exclusion limit With Polystyrene (eluent : THF) Target Molecular Weight Range Exclusion Limit Target Molecular Weight Range Exclusion Limit KF ,00 KF-40HQ 0 700,00 KF-0 00,00 4,000 KF-40HQ 00,00 4,000 KF-0. 00,000 0,000 KF-40.HQ 00,000 0,000 KF ,000 70,000 KF-40HQ 00 0,000 70,000 KF-04 7,000 00,000,000,000 KF-404HQ 7,000 00,000,000,000 KF-0 0,000,000,000 4,000,000 KF-40LHQ 00,000,000 4,000,000 KF-0 0,000 *(0,000,000) *(0,000,000) KF-40LHQ 00 *(0,000,000) *(0,000,000) KF-0M KF-07,000 *(0,000,000) 00,000 *(00,000,000) *(0,000,000) *(00,000,000) * Please use the above table as a rough indication for the column selection. *( ) Estimated value
20 Calibration curves for KF-00 series using polystyrene Molecular weight (Polystyrene) Column : Shodex GPC KF-00 series Flow rate : 0.mL/ (small cell volume) Comparison of standard, rapid analysis, and high performance type columns Standard type KF-80. x 0µL injection KF-0 KF-04 KF-0. KF-0 KF-0 KF-0 KF-07 KF-0 KF-0M ml Calibration curves for KF-400HQ series using polystyrene Molecular weight (Polystyrene) 7 4 KF-404HQ KF-40HQ KF-40LHQ KF-40HQ KF-40.HQ KF-40HQ KF-40LHQ Column : Shodex GPC KF-400HQ series Flow rate : 0.mL/ (small cell volume) Sample : EPON0 0.% By using KF-0., the analysis time is reduced to less than half of that using KF-80.. Thus KF-00 series enables rapid analysis. On the other hand, KF-40.HQ has a theoretical plate number. times more than that of the standard column, thereby improving resolution especially in the analysis of molecules that have a small to medium molecular weight. Rapid analysis and high performance type columns use less than one third of solvent per analysis compared to standard type columns do. Organic SEC (GPC) Columns : Rapid Analysis, High Performance Analysis High performance type.7ml KF-40.HQ x µl injection Rapid analysis type KF-0. x µl injection.0ml Column : Shodex GPC KF-80. x Shodex GPC KF-40.HQ x Shodex GPC KF-0. x (KF-80.) 0.mL/ (KF-40.HQ) 0.mL/ (KF-0.) (conventional type) (KF-80.) RI (small cell volume) (KF-40.HQ, KF-0.) 0 0 Styrene acrylonitrile copolymer Liquid paraffin Phthalates Sample : Styrene-Acrylonitrile (0:70) copolymer Sample : Liquid paraffin %, μl Sample : µl. Dioctyl phthalate 0.%. Dibutyl phthalate 0.%. Dipropyl phthalate 0.% 4. Diethyl phthalate 0.%. Dimethyl phthalate 0.%. Propylbenzene 0.% 7. Ethylbenzene 0.% 8. Toluene 0.% 9. Benzene 0.% Column : Shodex GPC KF-0M x Eluent : mm LiBr in DMF Flow rate : 0.mL/ (small cell volume) Column : Shodex GPC KF-40HQ Eluent : Chloroform Flow rate : 0.mL/ (small cell volume) Column : Shodex GPC KF-40HQ x Flow rate : 0.mL/ Detector : UV (4nm) (small cell volume)
21 Organic SEC (GPC) Columns : For Preparative Preparative columns * Preparative columns are made to order. KF-000 series : Shipping solvent Tetrahydrofuran (THF) Standard Columns F40 GPC KF KF-80 F40 GPC KF KF-80 F4 GPC KF KF-80. F40 GPC KF KF-80 F404 GPC KF-004 4, KF-804 F40 GPC KF-00, KF-80 F40 GPC KF-00, KF-80 F409 GPC KF-00M, KF-80M F70040 GPC KF-G 8B (GPC KF-LG) * See page 4 for GPC KF-800 series. K-000 series : Shipping solvent Chloroform Standard Columns F0 GPC K K-80 F GPC K K-80 F GPC K K-80. F0 GPC K K-80 F04 GPC K-004 4, K-804 F0 GPC K-00, K-80 F0 GPC K-00, K-80 F09 GPC K-00M, K-80M F GPC K-G 8B (GPC K-LG) * See page 48 for GPC K-800 series. 4
22 Preparative columns * Preparative columns are made to order. H-000 series : Shipping solvent Chloroform F00 F00 F0 F00 F004 F00 GPC H-00 GPC H-00 GPC H-00. GPC H-00 GPC H-004 GPC H-00,000,000,000,000,000, Standard Columns K-80 K-80 K-80. K-80 K-804 K-80 Organic SEC (GPC) Columns : For Preparative F00 GPC H-00, K-80 F009 GPC H-00M, K-80M F700 GPC H-G 8B (GPC H-G) * See page 48 for GPC K-800 series. KF-000 series : Shipping solvent Tetrahydrofuran (THF) [Customized columns] Standard Columns F80 GPC KF KF-80 F800 GPC KF KF-80 F80 GPC KF KF-80. F800 GPC KF KF-80 F8040 GPC KF KF-804 F GPC KF-G 0B (GPC KF-LLG) * See page 4 for GPC KF-800 series. K-000 series : Shipping solvent Chloroform [Customized columns] Standard Columns F90 GPC K K-80 F900 GPC K K-80 F90 GPC K K-80. F900 GPC K K-80 F9040 GPC K K-804 F GPC K-G 0B (GPC K-LLG) * See page 48 for GPC K-800 series.
23 Organic SEC (GPC) Columns : Linear Calibration Type Features LF Employs a special packing material with a wide pore size distribution (multi pore type) Highly linear calibration curve without inflection points Molecular weight distribution can be detered with high precision Enables analysis over a broad range of molecular weights Column for rapid analysis (LF-04) and column for high performance analysis (LF-404) enabling reduction in solvent use are also available Corresponds to USP L Standard columns Pore Size (Å) F04 GPC LF-804 7,000,000 F709 GPC LF-G 4. * See page for applicability of SEC (GPC) columns to solvent replacement. Rapid analysis downsized column Shipping Solvent : Tetrahydrofuran (THF) Use of the LF-04 with semi-micro type devices is recommended. Pore Size (Å) F04 GPC LF-04 9,000, F709 GPC LF-G 4. * See page for applicability of SEC (GPC) columns to solvent replacement. High performance semi-micro column Shipping Solvent : Tetrahydrofuran (THF) Use of the LF-404 with semi-micro type devices is recommended. Pore Size (Å) F04 GPC LF-404 4,000, F709 GPC LF-G 4. * See page for applicability of SEC (GPC) columns to solvent replacement. Target molecular weight range and Exclusion limit With Polystyrene (eluent : THF) Shipping Solvent : Tetrahydrofuran (THF) LF-804 LF-04 Target Molecular Weight Range 00,000,000 00,000,000 Exclusion Limit,000,000,000,000 LF ,000,000,000,000 * Please use the above table as a rough indication for the column selection. Schematic diagram of linear calibration type packing Connecting linear calibration type columns (LF series) The linear calibration type column covers a broad range of molecular weights with only one kind of packing (column). Connecting mixed-gel columns (KF-804L, etc.) Connecting different grades columns (KF KF-80 + KF-80, etc.)
24 Calibration curve for LF-804 using polystyrene Molecular weight (Polystyrene) 7 4 Xylan Column : Shodex GPC LF-804 Calibration curves for LF-04 and LF-404 using polystyrene Molecular weight (Polystyrene) 7 4 LF-404 LF Column : Shodex GPC LF-04, LF-404 Flow rate : 0.mL/ (LF-04) 0.mL/ (LF-404) (small cell volume) Polyamide (Nylon/) Polyurethane Sample : Polyurethane 0.%, 0μL 0 0 Column : Shodex GPC LF-404 x Flow rate : 0.mL/ (small cell volume) Polymethyl methacrylate Organic SEC (GPC) Columns : Linear Calibration Type Sample : Xylan 0.%, 0μL Sample : Nylon TM / 0.%, 0µL Sample : 0μL Polymethyl methacrylate Column : Shodex GPC LF-804 Eluent : 0mM H PO 4 + 0mM LiBr in (DMSO/DMF=80/0) Flow rate : 0.mL/ Column temp. : 0 C Column : Shodex GPC LF-404 Eluent : mm CF COONa in HFIP Flow rate : 0.mL/ (small cell volume) Column : Shodex GPC LF-804 x Eluent : Methyl ethyl ketone Polyamic acid Sample : Poly(pyromellitic dianhydride-co-4,4'-oxydianiline), 0μL Comparison of polystyrenes separation with different numbers of LF-404 LF-404 x 4 7 Sample : µl. Polystyrene (Mw :,00,000). Polystyrene (Mw :,000). Polystyrene (Mw :,000) 4. Polystyrene (Mw :,000). Polystyrene (Mw :,00). Polystyrene (Mw : 80) 7. Ethylbenzene LF-404 x 4 7 Column : Shodex GPC LF-404 x n Flow rate : 0.mL/ (small cell volume) Column : Shodex GPC LF-804 x Eluent : 0mM LiBr + 0mM H PO 4 in NMP Flow rate : 0.7mL/ Column temp. : 0 C LF-404 x
25 Organic SEC (GPC) Columns : High Temperature/Ultra High Temperature Analysis Features HT-800 Varied product lineup to support a wide range of molecular weights Corresponds to USP L UT-800 Dedicated to SEC analysis at high/ultra high temperatures with a maximum usable temperature of C Suitable for the analysis of ultra high molecular weight polymer containing samples Corresponds to USP L Standard columns Usable Temperature ( C) Pore Size (Å) F08700 GPC HT-80 7,000 0 ~ 0 00 F087 GPC HT-804 7,000 0 ~ 0,00 F0870 GPC HT-80 7,000 0 ~ 0,000 F0870 GPC HT-80 7,000 0 ~ 0,000 F08740 GPC HT-80M 7,000 0 ~ 0,000 F08770 GPC HT-807 4,000 0 ~ 0 8 0,000 F7094 GPC HT-G 0 ~ F0800 GPC UT-80. 4,400 0 ~ 0 00 F08 GPC UT-80M 4,400 0 ~ 0,000 F080 GPC UT-807,00 0 ~ 0 0,000 F GPC UT-G 0 ~ F0890 GPC AT-80MS,000 *Ta ~ 0, F70080 GPC AT-G *Ta ~ * The columns with M at the end of column names are mixed-gel column capable of analyzing samples over a wide range of molecular weight distribution. Shipping Solvent : Toluene *Ta : Ambient temperature Target molecular weight range and Exclusion limit With Polystyrene (eluent : THF) HT-80 HT-804 HT-80 HT-80 HT-80M HT-807 UT-80. UT-80M UT-807 Target Molecular Weight Range,000 0,000 7,000 00,000 0,000,000,000 0,000 *(0,000,000),000 *(0,000,000) 00,000 *(00,000,000) 00,000,000 *(0,000,000) 00,000 *(00,000,000) Exclusion Limit 70, ,000 4,000,000 *(0,000,000) *(0,000,000) *(00,000,000) 0,000 *(0,000,000) *(00,000,000) AT-80MS,000 *(0,000,000) *(0,000,000) * Please use the above table as a rough indication for the column selection. *( ) Estimated value 8
26 Calibration curves for HT-800 series using polystyrene Molecular weight (Polystyrene) HT-80 HT-80 HT-804 HT-80 Molecular weight (Polystyrene) Effects of gel particle size in high temperature GPC columns High temperature SEC columns are suitable for the analysis of high molecular weight polymers that are difficult to be dissolved in ambient temperature solvents; examples of such polymers are polyethylene and polypropylene. The GPC UT-800 series packed with large particle size (0 um) are recommended for the analysis of macromolecules. The large particle size prevents potential molecular shear degradation of the sample. Column : UT-80M x (Particle size : 0µm) HT-807 HT-80M Column : Shodex GPC HT-800 series Calibration curves for UT-800 series using polystyrene UT-80. UT-80M Column : AT-80MS x (Particle size : µm) UT-807 Column : Shodex GPC UT-800 series Calibration curve for AT-80MS using polystyrene Molecular weight (Polystyrene) 7 4 Column : HT-80M x (Particle size : µm) Column : Shodex GPC AT-80MS Sample :. Polystyrene (MW : 0,000,000). High density polyethylene (HDPE-A) Organic SEC (GPC) Columns : High Temperature/Ultra High Temperature Analysis Vo Shear degradation Vo 0 Vo Shear degradation 0 Column : Shodex GPC UT-80M Shodex GPC HT-80M Shodex GPC AT-80MS Eluent : 0.% BHT in o-dichlorobenzene Column temp. : 40 C High density polyethylene Low density polyethylene Sample : High density polyethylene (HDPE-B) 8 Sample : Low density polyethylene (LDPE-C) 7 Molecular weight RI Column : Shodex GPC UT-80M x Eluent : 0.% BHT in o-dichlorobenzene, MALS (Multi angle light scattering) Column temp. : 4 C Column : Shodex GPC HT-80M x Eluent : 0.% BHT in o-dichlorobenzene Column temp. : 40 C 9
27 Organic SEC (GPC) Columns : HFIP Features HFIP-800 HFIP-00 Columns exclusively for use with hexafluoroisopropanol (HFIP) Corresponds to USP L Rapid analysis, solvent saving type Corresponds to USP L Standard columns HFIP-800 series Pore Size (Å) F080 GPC HFIP-80, F0840 GPC HFIP-804,000 7,00 F080 GPC HFIP-80,000,000 F080 GPC HFIP-80,000,000 F0890 GPC HFIP-80M,000,000 F0870 GPC HFIP-807 4, ,000 F70000 GPC HFIP-G 8B (GPC HFIP-LG) * The columns with M at the end of column names are mixed-gel column capable of analyzing samples over a wide range of molecular weight distribution. Shipping Solvent : Hexafluoroisopropanol (HFIP) Rapid analysis downsized columns HFIP-00 series Use of the HFIP-00 series with semi-micro type devices is recommended. Pore Size (Å) F00 GPC HFIP-0, F040 GPC HFIP-04,000, F00 GPC HFIP-0,000, F00 GPC HFIP-0,000, F080 GPC HFIP-0M,000, F070 GPC HFIP-07, , F700 GPC HFIP-G 4A (GPC HFIP-G) 8 4. * The columns with M at the end of column names are mixed-gel column capable of analyzing samples over a wide range of molecular weight distribution. Shipping Solvent : Hexafluoroisopropanol (HFIP) Target molecular weight range and Exclusion limit With *PMMA (eluent : HFIP) Target Molecular Weight Range Exclusion Limit Target Molecular Weight Range Exclusion Limit HFIP-80,000 0,000 0,000 HFIP-0,000 0,000 0,000 HFIP-804 0,000 00,000 00,000 HFIP-04 0,000 00,000 00,000 HFIP-80 0,000 00,000,000,000 HFIP-0 0,000 00,000,000,000 HFIP-80 70,000 **(8,000,000) **(8,000,000) HFIP-0 70,000 **(8,000,000) **(8,000,000) HFIP-80M,000 **(8,000,000) **(8,000,000) HFIP-0M,000 **(8,000,000) **(8,000,000) 0 HFIP ,000 **(0,000,000) **(0,000,000) HFIP-07 70,000 **(0,000,000) * Please use the above table as a rough indication for the column selection. **(0,000,000) *PMMA : Polymethylmethacrylate **( ) Estimated value
28 Calibration curves for HFIP-800 series using PMMA Molecular weight (PMMA) 7 4 HFIP-80 HFIP-80 HFIP-80 HFIP-804 HFIP-807 HFIP-80M Calibration curves for HFIP-00 series using PMMA Molecular weight (PMMA) 7 4 HFIP-04 HFIP-0 HFIP-0 HFIP-07 HFIP-0 HFIP-0M Polyethylene terephthalate (PET) Sample : Polyethylene terephthalate 0.%, 0µL Mw=9,00 0 Organic SEC (GPC) Columns : HFIP Column : Shodex GPC HFIP-800 series Eluent : HFIP Column : Shodex GPC HFIP-00 series Eluent : HFIP Flow rate : 0.mL/ (HFIP-0, 04) 0.mL/ (HFIP-0, 0, 0M, 07) Detctor : RI (small cell volume) Column : Shodex GPC HFIP-0M x Eluent : mm CF COONa in HFIP Flow rate : 0.mL/ (small cell volume) Polyamide (effects of salt addition) Polybutylene terephthalate (PBT) Sample : Polycaprolactum (Nylon ) Sample : Polybutylene terephthalate 0.0%, 00µL HFIP mm CFCOONa in HFIP In SEC analysis using HFIP, some samples may yield abnormal peaks as a result of ionic interaction. In this case, ionic interaction can be suppressed by adding sodium trifluoroacetate to HFIP Column : Shodex GPC HFIP-80M x Eluent : HFIP (solid line), mm CF COONa in HFIP (broken line) Column : Shodex GPC HFIP-80 + HFIP-80 Eluent : mm CF COONa in HFIP Polyamides (Nylon and Nylon ) Polyacetal Sample : Nylon 0.%, 0µL Sample : Nylon 0.%, 0µL Sample : Polyacetal 0.%, 0µL Mw=9,00 Mw=8,0 Mw=7, Column : Shodex GPC HFIP-0M x Eluent : mm CF COONa in HFIP Flow rate : 0.mL/ (small cell volume) Column : Shodex GPC HFIP-0M x Eluent : mm CF COONa in HFIP Flow rate : 0.mL/ (small cell volume)
29 Applicability of SEC (GPC) Columns to Solvent Replacement Shipping Solvent : THF Shipping Solvent : DMF Solvent KF-80 KF-80 KF-80. KF-80L KF-804L KF-80 KF-804 KF-80 KF-80 KF-807 KF-80M KF-80L KF-80L KF-807L KF-0 KF-0 KF-0. KF-0 KF-04 KF-0 KF-0 KF-07 KF-0M LF-804 LF-04 LF-404 KD-80 KD-80 KD-80. KD-80 KD-804 KD-80 KD-80 KD-807 KD-80M Shipping Solvent : Chloroform Shipping Solvent : THF K-80 K-80 K-80 K-80. K-80L K-804L K-804 K-80 K-80 K-807 K-80M K-80L K-80L K-807L KF-40HQ KF-40HQ KF-40.HQ KF-40HQ KF-404HQ KF-40LHQ KF-40LHQ Tetrahydrofuran (THF) Chloroform Carbon tetrachloride Benzene Toluene p-xylene o-dichlorobenzene (ODCB) Trichlorobenzene (TCB) Dioxane Diethyl ether Ethyl acetate Acetone Methyl ethyl ketone Dimethylformamide (DMF) Dimethylacetamide (DMAc) Hexafluoroisopropanol (HFIP) m-cresol o-chlorophenol Quinolin N-Methylpyrrolidone (NMP) Dimethylsulfoxide (DMSO) 0% m-cresol/chloroform 0% o-chlorophenol/chloroform 0% HFIP/Chloroform Hexane Acetonitrile Methanol Water : Solvent replacement possible : Solvent replacement possible, but this may cause column performance to deteriorate slightly : Usable at 40 C or higher : Solvent replacement not possible
30 Calibration Standards for SEC [Polystyrene (PS)] Features For organic solvent SEC (GPC) Less branched polystyrene with anionic polymerization Easily soluble in tetrahydrofuran (THF), chloroform, toluene, and o-dichlorobenzene (ODCB) Kit type F80 F80 F8007 SL- Std. No. Mp Mw/Mn S-9 S- S-9.9 S-.9 S-4.8 S-.7 S-.8 S-. S-0.8 S-0. 9,00,00 9,970,940 4,80,780,80, STANDARD SL- STANDARD SM- STANDARD SH S-700 S-90 S- S- S-4 S-47 S-8 S-.9 S-.0 S-.,700,000,90,000,000,000 4,000 47,00 8,00,940,980,0 Contents 0.g x kinds 0.g x kinds 0.g x 7 kinds Std. No. Mp Mw/Mn [ Polymethylmethacrylate (PMMA)] Features Contents F80407 STANDARD M-7 0.g x 7 kinds Features S-870 S-90 S-990 S-0 S-80 S-99 S-9 MW Range 80 9,00,0,700,000 9,000,870,000 Std. No. Mp Mw/Mn,870,000,90,000,990,000,0,000,80,000 99,000 9,000 For organic solvent SEC (GPC) Narrow molecular weight distribution range Easily soluble in hexafluoroisopropanol (HFIP) and dimethylformamide (DMF) Kit type [Pullulan] SM- SH-7 MW Range,880,00,000 (Note) Molecular weights (Mp, Mw/Mn) of each kit may vary depending on production lots Structural formula of S series M-0 M-9 M- M-9 M-8 M-.8 M-.9 C 4H 9 CH CH (Note) Molecular weights (Mp, Mw/Mn) of each kit may vary depending on production lots. Std. No. Mp Mw/Mn,00,000 9,000,000 8,800 7,800,80,880 n H Applicability of SEC (GPC) Columns to Solvent Replacement Calibration Standards for SEC For aqueous SEC (GFC) Unbranched pullulan standard High solubility in water eliates the possibility of recrystallization Kit type Contents MW Range Structural formula of P series F STANDARD P-8 Single type 0.g x 8 kinds,0 4,000 O CH H H H O OH OH H H OH O H CH OH H H CH OH H H O H O OH H O OH H H OH H OH n F F F F84000 F F F84000 F STD P-800 STD P-400 STD P-00 STD P-0 STD P-0 STD P-0 STD P- STD P- Contents 0.g 0.g 0.g 0.g 0.g 0.g 0.g 0.g Mp 4,000 7,000 94,000 7,000 47,0,0 9,00,0 Mw/Mn Std. No. Mp Mw/Mn STD P-800 STD P-400 STD P-00 STD P-0 STD P-0 STD P-0 STD P- STD P- 4,000 7,000 94,000 7,000 47,0,0 9,00, (Note) Molecular weights (Mp, Mw/Mn) of each kit may vary depending on production lots.
Solvent-Compatible High Performance GPC Columns: TSKgel HHR Series
ANALYSIS S e p a r a t i o n R e p o r t N o. 8 3 Solvent-Compatible High Performance GPC Columns: TSKgel HHR Series Table of Contents 1. Introduction 1 2. Features 1 3. Basic characteristics 5 3-1 Resolution
More informationPhenogel. GPC/SEC Columns. Sample Elution. Technical Specifications 10 3 Å 10 6 Å
phenogel Gpc/sec columns HPLC 5 and 10 μm particle sizes Narrow bore (4.6 mm ID) solvent-saver to preparative columns available Very good alternative to Polymer Labs PLgel and Waters Styragel, Ultrastyragel,
More informationTSKgel Size Exclusion Chromatography Columns
Silica-based for protein analysis: TSKgel SW mab TSKgel SW TSKgel SWXL TSKgel SuperSW Polymer-based for desalting: TSKgel BioAssist DS Columns Polymethacrylate-based for water-soluble polymers analysis:
More informationPHENOGEL GPC/SEC COLUMNS
5, 10, and 20 μm particle sizes Narrow bore (4.6 mm D) solvent-saver to preparative columns available Very good alternative to Polymer Labs PLgel and Waters Styragel, Ultrastyragel, Styragel HT, and Styragel
More informationOptimizing GPC Separations
Optimizing GPC Separations Criteria for Solvent Selection True sample solubility (Polarity and Time dependant) Compatibility with columns Avoid non-size exclusion effects (eg adsorption by reverse phase
More informationTECHNICAL NOTEBOOK No.1. GPC LF series columns. Linear calibration curves over a wide molecular weight range. TECHNICAL NOTEBOOK. No.
TECNICAL NOTEBOOK No.1 GPC LF series columns Linear calibration curves over a wide molecular weight range. TECNICAL NOTEBOOK No.1 Contents 1. Introduction 1 2. Shodex GPC LF series 2.1. Specifications
More informationGFC Analysis of Water-Soluble Polymers with TSKgel PW-type Columns. Contents
No.035 GFC Analysis of Water-Soluble Polymers with TSKgel PW-type Columns Contents Page 1. Introduction 2. Separation ranges for each grade of TSKgel PW-type column 3. GFC apparatus for water-soluble systems
More informationShodex TM ODP2 HP series columns
HPLC Columns Shodex TM ODP2 HP series columns Better retention of highly polar substances Technical notebook No. 6 Contents 1. Introduction 1-1. Specifications 1-2. Eluent Compatibility of ODP2 HP Series
More informationGel Permeation Chromatography Basics and Beyond eseminar March 13, Jean Lane Technical and Applications Support LSCA, Columns and Supplies
Gel Permeation Chromatography Basics and Beyond eseminar March 13, 2013 Jean Lane Technical and Applications Support LSCA, Columns and Supplies 1 Content Overview of GPC/SEC What is it? Why do we use it?
More informationQuick guide to selecting columns and standards for Gel Permeation Chromatography and Size Exclusion Chromatography SELECTION GUIDE
Quick guide to selecting columns and standards for Gel Permeation Chromatography and Size Exclusion Chromatography SELECTION GUIDE Introduction Gel permeation chromatography (GPC) and size exclusion chromatography
More informationGFC separation of water-soluble polymers with TSKgel PW-type columns (2) using the PWXL series. Contents
No.038 GFC separation of water-soluble polymers with TSKgel PW-type columns (2) using the PWXL series Contents Page 1. Introduction 2. Calibration curves, theoretical plate numbers and separation range
More informationGPC/SEC standards. Product guide
GPC/SEC standards Product guide Contents Polymer standards for GPC/SEC 3 Agilent EasiVial 5 Agilent EasiCal 8 Polystyrene 9 Polymethylmethacrylate 11 Polyethylene glycol/oxide 12 Other polymer standards
More informationNo.106. Aqueous SEC Columns for Analysis of Cationic Polymers: TSKgel PWXL-CP Series. Contents. Page
No.106 Aqueous SEC Columns for Analysis of Cationic Polymers: TSKgel PWXL-CP Series Contents Page 1. Introduction 2. Features 3. Basic characteristics 3-1. Pore characteristics 3-2. Sample injection volume
More informationGPC/SEC standards. Product guide
GPC/SEC standards Product guide Contents Polymer standards for GPC/SEC 3 Agilent EasiVial 5 Agilent EasiCal 8 Polystyrene 9 Polymethylmethacrylate 11 Polyethylene glycol/oxide 12 Other polymer standards
More informationAdvantages of polymerbased. an alternative for ODS
Advantages of polymerbased HPLC columns an alternative for ODS Yukiko Higai Shodex, 1 Content (1) Comparison between Silica and Polymer-based RP columns (2) Characteristics of Polymer RP columns - an example:
More informationSaccharides and Organic Acids
Saccharides and Organic Acids Contents Introduction... 2 Reducing Sugars... 2 HILIC The HILICpak and Ashipak Series... 3 Shodex HILIC Columns for Sugar Analysis... 7 Ligand Exchange Shodex SUGAR series...
More informationGPC/SEC Practical Tips and Tricks. Thomas Dent Applications Scientist Agilent Technologies. October, 2011 Gulf Coast Conference
GPC/SEC Practical Tips and Tricks Thomas Dent Applications Scientist Agilent Technologies October, 2011 Gulf Coast Conference 1 Section 1: Introduction Goals Brief introduction to GPC/SEC Highlight considerations
More informationAqueous and polar GPC/SEC columns. Product guide
Aqueous and polar GPC/SEC columns Product guide Contents Agilent PL aquagel-oh SEC columns 3 PL aquagel-oh preparative SEC columns 8 Agilent PolarGel GPC columns 9 PolarGel-L 10 PolarGel-M 11 Agilent PL
More informationHICHROM. Chromatography Columns and Supplies. LC COLUMNS Shodex. Catalogue 9. Hichrom Limited
HICHROM Chromatography Columns and Supplies LC COLUMNS Catalogue 9 Hichrom Limited The Markham Centre, Station Road Theale, Reading, Berks, RG7 PE, UK Tel: + (0)8 90 660 Fax: + (0)8 9 8 Email: sales@hichrom.co.uk
More informationGPC/SEC Column Selection & Method Development
GPC/SEC Column Selection & Method Development LIVE WEBCAST: Tuesday April 27, 2010 Topics Introduction to GPC/SEC Column Characteristics & Column Selection Additional Method Development Troubleshooting
More informationPractical Steps in GPC Method Development
Practical Steps in GPC Method Development Jean Lane Applications Engineer LC Columns & Consumables Technical Support March 23,2016 1 Overview Solvent Criteria for Solvent Selection Polymer Standards -
More informationLecture 4 : Gel Permeation or Size Exclusion Chromatography
Lecture 4 : Gel Permeation or Size Exclusion Chromatography Polymer Fractionation Sedimentation Centrifugation Evaporation of the solvent Gel permeation chromatography Gel Permeation Chromatography (GPC)
More informationGPC/SEC An essential tool for polymer analysis
GPC/SEC An essential tool for polymer analysis Ben MacCreath, PhD Product Manager GPC/SEC Instrumentation 26 th March 2013 Introduction to Polymers Where are they found? Polyolefins Engineering Polymers
More informationHarris: Quantitative Chemical Analysis, Eight Edition CHAPTER 25: CHROMATOGRAPHIC METHODS AND CAPILLARY ELECTROPHORESIS
Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 25: CHROMATOGRAPHIC METHODS AND CAPILLARY ELECTROPHORESIS CHAPTER 25: Opener Aa CHAPTER 25: Opener Ab CHAPTER 25: Opener B 25-1 Ion-Exchange
More informationClearing the Confusion: GPC, SEC, GFC What, When, Why, and How?
Clearing the Confusion: GPC, SEC, GFC What, When, Why, and How? Jean Lane Applications Engineer LC Columns & Consumables Technical Support January 19, 2017 What we will cover What - Clarification of the
More informationGuaranteed * Enhance Resolution FOR POLYMER ANALYSIS, Remarkable Reproducibility Exceptional Efficiency Long Lifetimes Economically Priced
Enhance Resolution FOR POLYMER ANALYSIS, Guaranteed * Remarkable Reproducibility Exceptional Efficiency Long Lifetimes Economically Priced www.phenomenex.com/phenogel Phenogel Non-Aqueous GPC/SEC Columns
More informationPolymer analysis by GPC-SEC. Technical Note. Introduction
Polymer analysis by GPC-SEC Technical Note Introduction Gel Permeation Chromatography (GPC), also referred to as Size Exclusion Chromatography (SEC) is a mode of liquid chromatography in which the components
More informationCation Exchange HPLC Columns
Cation Exchange HPLC Columns Hamilton offers seven polymeric packing materials for cation exchange separations. Type Recommended Application(s) PRP-X00 PRP-X00 PRP-X800 HC-0 HC-7 Ca + HC-7 H + HC-7 Pb
More informationTSKgel Size Exclusion Chromatography Columns
Silica-based for protein analysis: TSKgel SW mab TSKgel SW TSKgel SWXL TSKgel SuperSW Polymer-based for desalting: TSKgel BioAssist DS Columns Polymethacrylate-based for water-soluble polymers analysis:
More informationProducts for GPC. Products for GPC. Size exclusion chromatography. Basic principles
Products for GPC Products for GPC Size exclusion chromatography Basic principles injection A B C retention time mobile phase B Size exclusion chromatography (SEC) is a liquid chromatographic separation
More informationSHODEX. by Showa Denko K.K.
selectosil For Availability and Ordering nformation please contact your Phenomenex Technical Consultant. SHODEX Selectosil/ High efficiency polymer columns Excellent mechanical and chemical stability Wide
More informationOrganic GPC/SEC columns. Product guide
Organic GPC/SEC columns Product guide Contents Agilent PLgel GPC Columns 3 PLgel MIXED 5 PLgel 20 µm MIXED-A 6 PLgel 10 µm MIXED-B 7 PLgel 5 µm MIXED-C 8 PLgel 5 µm MIXED-D 9 PLgel 3 µm MIXED-E 10 PLgel
More informationAGILENT AQUEOUS AND POLAR GPC/SEC COLUMNS
AGILENT AQUEOUS AND POLAR GPC/SEC COLUMNS Contents Agilent GPC/SEC columns...3 InfinityLab PL Multisolvent...4 PL aquagel-oh SEC columns...7 PL Rapide Aqua columns...10 PL aquagel-oh preparative SEC columns...11
More informationComparison of Polymer Separation by Size Exclusion Chromatography and Asymmetric Flow Field Flow Fractionation
Comparison of Polymer Separation by Size Exclusion Chromatography and Asymmetric Flow Field Flow Fractionation Stepan Podzimek, 1 Christoph Johann 2 1 SYNPO / University of Pardubice, Czech Republic, stepan.podzimek@synpo.cz
More informationSize Exclusion Chromatography: Method Development
Size Exclusion Chromatography: Method Development To develop a successful Size Exclusion Chromatography (SEC) method it is desired to find a column/solvent combination under which the follow conditions
More informationCation Exchange HPLC Columns
Cation Exchange PRP-X00 Polymeric cation exchange packing for separation of inorganic cations and organic cations. Easily separate mono or divalent cations. ph stable from to. Use with organic solvent
More informationOptical Isomer Separation Columns and Packing Materials
02 Optical Isomer Separation s and Packing Materials CHIRAL ART----------------------------------- 26~29 YMC CHIRAL NEA (R), (S)-----------------------30 YMC CHIRAL CD BR------------------------------31
More informationChapter 2 The Chemistry of Biology. Dr. Ramos BIO 370
Chapter 2 The Chemistry of Biology Dr. Ramos BIO 370 2 Atoms, Bonds, and Molecules Matter - all materials that occupy space and have mass Matter is composed of atoms. Atom simplest form of matter not divisible
More informationGPC-/SEC-Columns. MZ-Gel SD. MZ-Gel Super FG SEC with fluorinated Eluents
GPC-/SEC-Columns & 2 accessories2 0 18 direct from Producer packed in new generation column hardware! MZ-Gel SD MZ-Gel SD LS MZ-Super FG Development Manufacturing Distribution Consulting HPLC-Columns New
More informationTSKgel SW TSKgel SW XL TSKgel SuperSW. TSKgel PW TSKgel PW XL -CP. TSKgel Alpha TSKgel SuperAW. TSKgel H XL TSKgel H HR TSKgel Super H TSKgel Super HZ
8 www.tosohbioscience.com size exclusion CHROMATOGRAPHY Tosoh bioscience ANALYSIS sec size exclusion chromatography 9 size exclusion CHROMATOGRAPHY PRODUCTS TSK-GEL SW-type TSKgel SW TSKgel SW XL TSKgel
More informationMolecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. H i
Gel Permeation Chromatography (GPC) : Size Exclusion Chromatography GPC : 1. Chromatogram (V R vs H) H i Detector response Baseline N i M i 130 135 140 145 150 155 160 165 Elution volume (V R ) (counts)
More informationLC and LC/MS Column Selection Flow Chart
LC and LC/MS Column Selection Flow Chart To use the column selection diagram below, simply follow the path for your analyte and mobile phase. At the far right, follow your final column selection to the
More informationGel Permeation Chromatography - GPC
Isolation and Separation Methods J. Poustka, VŠCHT Praha, ÚAPV 2014, http://web.vscht.cz/poustkaj Gel Permeation Chromatography - GPC Separation and clean-up method Group separation of compounds with similar
More informationEnhance Resolution FOR POLYMER ANALYSIS, Guaranteed *
Enhance Resolution FOR POLYMER ANALYSIS, Guaranteed * Remarkable Reproducibility Exceptional Efficiency Long Lifetimes Economically Priced www.phenomenex.com/phenogel Phenogel Non-Aqueous GPC/SEC Columns
More informationPacked Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents
No. 089 SEPARATION REPORT Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS Table of Contents 1. Introduction 1 2. Column Specification 1 3. Features of Packing Materials
More informationPackings for HPLC. Packings for HPLC
Summary of packings for HPLC In analytical HPLC, packings with particle sizes of 3 to 10 µm are preferred. For preparative separation tasks, also particles with diameters larger than 10 µm are applied.
More informationPhenogel GPC Materials
Phenogel GPC Materials Highly cross-linked for mechanical and chemical stability Narrow bore (4.6 mm ID) solvent-saver to preparative columns available Temperature stable to 140 C / 205 C* Phenogel packing
More informationA User s Guide to Gel Permeation. Chromatography
A User s Guide to Gel Permeation Chromatography Technical Notes 3 July 2000 U.S.A. PUERTO RICO UNITED KINGDOM CANADA AUSTRALIA NEW ZEALAND GERMANY PHONE: (310) 212-0555 (800) 541-HPLC 01625-501367 (800)
More informationSolvent Flexibility for Size-Based Polymer Analysis Using the Advanced Polymer Chromatography (APC) System
Solvent Flexibility for Size-Based Polymer Analysis Using the Advanced Polymer Chromatography (APC) System Mia Summers, Jeremy Shia, Kenneth J. Fountain Waters Corporation, Milford, MA, USA A P P L I C
More informationN6,6 Posidyne Membranes
Rev. Date: 20-Jun-2010 Ultipor Nylon 6,6 and N6,6 Posidyne Membrane Description Pall provides a complete range of Nylon 6,6 membranes for fluid filtration in the pharmaceutical, biotechnology, electronics,
More informationGPC - Gel Permeation Chromatography. aka Size Exclusion Chromatography- SEC
GPC - Gel Permeation Chromatography aka Size Exclusion Chromatography- SEC Wendy Gavin Biomolecular Characterization Laboratory Version 1 May 2016 1 Table of Contents 1. GPC Introduction. Page 3 2. How
More informationLiquid Chromatography
Liquid Chromatography 1. Introduction and Column Packing Material 2. Retention Mechanisms in Liquid Chromatography 3. Method Development 4. Column Preparation 5. General Instrumental aspects 6. Detectors
More informationGPC/SEC Start-up Kit
GPC/SEC Start-up Kit GPC/SEC Start-up Kit Congratulations for choosing GPC/SEC solutions from Agilent. This kit contains everything you need to get you up and running quickly and easily. The information
More informationMaximizing Performance Through GPC Column Selection
Maximizing Performance Through GPC Column Selection What Are Polymers? Polymers are long chain molecules produced by linking small repeat units (monomers) together There are many ways to link different
More informationPhenogelTM Non-Aqueous GPC/SEC Columns
PhenogelTM Non-Aqueous GPC/SEC Columns Improve Resolution Wide solvent compatibility Temperature stability up to 140 C Narrow bore columns available for reduced solvent consumption Phenogel GPC Materials
More informationHeadspace Technology for GC and GC/MS: Features, benefits & applications
Headspace Technology for GC and GC/MS: Features, benefits & applications Karima Baudin Oct 2015 Why use Headspace? Very Simple no to minimum sample prep Robust enhance uptime Non-detectable carry-over
More informationhttps://www.chemicool.com/definition/chromatography.html
CHROMATOGRAPHY 1 Chromatography - a physical method of mixture separation in which the components to be separated are distributed between two phases, one of which is stationary (stationary phase) while
More informationPhysical Polymer Science Lecture notes by Prof. E. M. Woo adapting from: Textbook of Physical Polymer Science (Ed: L. H.
Caution: Contents of the lecture notes are copyrighted ( 有版權 Textbook of Physical Polymer Science). (Do not use outside the class room purposes without permission from the publisher.) Physical Polymer
More informationMethod Development with ZirChrom's Ion Exchange Phases
For Peak Performance ZIRCHROM ION EXCHANGERS Phases for Sugars and Proteins Wide Range of Ion Exchange Selectivity No Shrinking or Swelling Use Any Organic Solvent Significantly Higher Efficiency than
More informationOpen Column Chromatography, GC, TLC, and HPLC
Open Column Chromatography, GC, TLC, and HPLC Murphy, B. (2017). Introduction to Chromatography: Lecture 1. Lecture presented at PHAR 423 Lecture in UIC College of Pharmacy, Chicago. USES OF CHROMATOGRAPHY
More informationLecture No. (1) Introduction of Polymers
Lecture No. (1) Introduction of Polymers Polymer Structure Polymers are found in nature as proteins, cellulose, silk or synthesized like polyethylene, polystyrene and nylon. Some natural polymers can also
More informationSPE. Simply Efficient! Watchers bond Solid Phase Extraction by ISU. Ver
SPE Simply Efficient! Watchers bond Solid Phase Extraction by ISU Ver. 05-11 Solid Phase Extraction Basic principles of Solid Phase Extraction Solid phase extraction (SPE) is a powerful method for sample
More informationWelcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster
Welcome to our E-Sear: Choosing HPLC Columns for Faster Analysis Smaller and Faster High Throughput/Fast LC Requires. Short columns 0 mm or shorter Small particle sizes. µm Rapid Resolution or new.8 µm
More informationHints for Strong Ion Exchange Resins
Hints for Strong Ion Exchange Resins Chromatography Application Note AN98 Abstract Ion exchange columns are a powerful means of isolating and purifying compounds, but their use is limited due to lack of
More informationLatest Developments in GPC Analysis of Adhesive and Sealant Polymers Mark Pothecary PhD Americas Product Manager Malvern Instruments
Latest Developments in GPC Analysis of Adhesive and Sealant Polymers Mark Pothecary PhD Americas Product Manager Malvern Instruments Molecular weight The most fundamental molecular property that controls
More information1. Ion exchange chromatography
1. Ion exchange chromatography Ion Exchange Chromatography Most popular method for the separation or purification of charged molecules. In cation exchange chromatography, positively charged molecules are
More informationExperiment 5. Synthetic Polymers.
Experiment 5. Synthetic Polymers. References: Brown & Foote, Chapters 24 INTRODUCTION: A polymer (Greek: polys + meros = many parts) is a giant or macromolecule made up of repeating structural units. The
More informationAnalysis - HPLC A.136. Primesep 5 µm columns B.136
Primesep 5 µm columns Primesep columns feature double functionality of the bonding i.e : alkyl chain with anionic or cationic group, chelating group. This feature creates unique selectivities when using
More informationSepax SRT -C, Zenix -C SEC Phases
Sepax SRT -C, Zenix -C SEC Phases Complimentary Phases to SRT for Derivatized Monoclonal Antibodies General Description Stationary Phase Structure SRT, Zenix, SRT-C and Zenix-C SEC phases developed based
More informationSPE Introduction. general chromatography. solid phase extraction. How to Choose an SPE Product. Solid-Phase Processing Methods
How to Choose an SPE Product 1. Characterize the Sample Factors such as the analyte s polarity relative to the matrix, the presence of charged functional groups, solubility, molecular weight, etc., determine
More informationPolymer Analysis by Gel Permeation Chromatography
Polymer Analysis by Gel Permeation Chromatography A Historical Perspective The development of polymer has had a profound effect on the modern world. These versatile materials are used in an extensive array
More informationProtein separation and characterization
Address:800 S Wineville Avenue, Ontario, CA 91761,USA Website:www.aladdin-e.com Email USA: tech@aladdin-e.com Email EU: eutech@aladdin-e.com Email Asia Pacific: cntech@aladdin-e.com Protein separation
More informationMolecular Weight Distribution of Living Chains in Polystyrene Pre-pared by Atom Transfer Radical Polymerization
Molecular Weight Distribution of Living Chains in Polystyrene Pre-pared by Atom Transfer Radical Polymerization Joongsuk Oh, a Jiae Kuk, a Taeheon Lee, b Jihwa Ye, b Huyn-jong Paik, b* Hyo Won Lee, c*
More informationHigh Performance Liquid Chromatography
Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN
More informationHigh Performance Liquid Chromatography
Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 1 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN
More informationChapter 13 - Polymers Introduction
Chapter 13 - Polymers Introduction I. Nomenclature A. Polymer/Macromolecule polymer - nonmetallic material consisting of large molecules composed of many repeating units - from Greek: poly (many) and meros
More informationHow to use GPC/SEC for compositional analysis
How to use GPC/SEC for compositional analysis Determining the relative concentration of two components in a polymer sample MOLECULAR SIZE MOLECULAR STRUCTURE MOLECULAR WEIGHT Introduction Over the last
More informationHICHROM. Chromatography Columns and Supplies. LC COLUMNS SIELC Primesep. Catalogue 9. Hichrom Limited
HICHROM Chromatography Columns and Supplies LC COLUMNS SIELC Primesep Catalogue 9 The Markham Centre, Station Road Theale, Reading, Berks, RG7 4PE, UK Tel: +44 (0)8 90 660 Fax: +44 (0)8 9 484 Email: sales@hichrom.co.uk
More informationComparison of different aqueous mobile phase HPLC techniques
June 009 ewsletter Pharmaceutical Analysis: What is Your Problem? SIELC Technologies, Inc., Prospect Heights, IL 0070 Pharmaceutical analysis involves liquid chromatography of various compounds, from active
More informationAqueous GPC Eluents and Calibrations
Eluent Name: Sodium 4-Styrenesulfonate Eluent Polymer Types: NaStS Eluent Composition: 70 % 0.2M NaNO 3, 0.01M NaH 2 PO 4, at ph 7.0; 30 % methanol Column(s): PL Aquagel-OH 40 and PL Aquagel-OH 30 Flow
More informationSRT -10/10C (10 µm ) Fast Protein Purification For FPLC, HPLC and Prep LC. Sepax Technologies, Inc. Preparative Size Exclusion Chromatography
Sepax Technologies, Inc. Preparative Size Exclusion Chromatography SRT -0/0C (0 µm ) Fast Protein Purification For FPLC, HPLC and Prep LC Better Surface Chemistry for Better Separation SRT -0/0C Size Exclusion
More informationCategory Major Groups Products
Polyester Filament Yarn 1562.48 1803.873 1874.268 1877.529 1811.283 2178.749 2179 2200.91 1127.08 Nylon Filament Yarn (NFY) 30.306 33.429 30.388 21.943 24.324 32.449 37.251 40.906 19.373 Nylon Industrial
More informationToo Polar for Reversed Phase What Do You Do?
Too Polar for Reversed Phase What Do You Do? June 20, 2013 Mark Powell Columns and Consumables Technical Support Page 1 C8 or C18 Doesn t Always Do the Job Typical reversed phase conditions involve water/buffer
More informationState-of-the-art C18 HPLC Columns
An HPLC GL Sciences Newest and Most Advanced ODS Phase-New For 00 State-of-the-art C HPLC s Improved Peak Shapes and Heights Enhancing Sensitivity High Resolution Fast Equilibration Compatible with 00%
More information2/25/2013. Electronic Configurations
1 2 3 4 5 Chapter 2 Chemical Principles The Structure of Atoms Chemistry is the study of interactions between atoms and molecules The atom is the smallest unit of matter that enters into chemical reactions
More informationChapter 2. The Structure of Atoms. The Structure of Atoms. The Structure of Atoms
1 The Structure of Atoms 2 Chapter 2 Chemical Principles Chemistry is the study of interactions between atoms and molecules The atom is the smallest unit of matter that enters into chemical reactions Atoms
More informationFoundations in Microbiology Seventh Edition
Lecture PowerPoint to accompany Foundations in Microbiology Seventh Edition Talaro Chapter 2 The Chemistry of Biology Copyright The McGraw-Hill Companies, Inc. Permission required for reproduction or display.
More informationFAQ's. 1 - Which column would be the most appropriate for my application?
1 - Which column would be the most appropriate for my application? Due to the complexity of the chiral recognition mechanism, it is not possible yet to establish rules for the selection of the best chiral
More informationGuide to Choosing and Using Polymer Reversed Phase Columns!
Guide to Choosing and Using Polymer Reversed Phase Columns! Choosing the best RP Column to use: Silica is normally used as the packing material for HPLC columns for a number of reasons. It is very strong,
More informationChemical Principles. PowerPoint Lecture Presentations prepared by Bradley W. Christian, McLennan Community College C H A P T E R
PowerPoint Lecture Presentations prepared by Bradley W. Christian, McLennan Community College C H A P T E R 2 Chemical Principles The Structure of Atoms Learning Objective 2-1 Describe the structure of
More informationCHEMICAL BONDS. Attraction that holds molecules together Involves valence electrons. Ionic Bonds Covalent Bonds. Involves sharing of.
CHEMICAL BONDS DEFINITION/DESCRIPTION: Attraction that holds molecules together Involves valence electrons TYPES: Ionic Bonds Covalent Bonds Involves sharing of electrons Electronegativities O = 3.5 N
More informationT.A Nouf Alshareef KAU-Faculty of Science- Biochemistry department Analytical biochemistry lab (Bioc 343) 2012
T.A Nouf Alshareef KAU-Faculty of Science- Biochemistry department Analytical biochemistry lab (Bioc 343) 2012 nf.shareef@hotmil.com False colour scanning electron micrograph of Sephadex beads. Background
More informationpenta-hilic UHPLC COLUMNS
penta-hilic UHPLC COLUMNS penta-hilic Highly retentive, proprietary penta-hydroxy-ligand Excellent peak shape for polar compounds with a variety of functional groups: acids, bases, zwitterions strong and
More informationRing-Opening Polymerization of N-Carboxyanhydrides Initiated by a Hydroxyl Group
SUPPRTING INFRMATIN Ring-pening Polymerization of N-Carboxyanhydrides Initiated by a Hydroxyl Group Špela Gradišar, Ema Žagar, and David Pahovnik* National Institute of Chemistry, Department of Polymer
More informationSupplementary Materials: SRG Inscription in Supramolecular Liquid Crystalline Polymer Film: Replacement of Mesogens
Supplementary Materials: SRG Inscription in Supramolecular Liquid Crystalline Polymer Film: Replacement of Mesogens Shun Mitsui, Mitsuo Hara, Shusaku Nagano, and Takahiro Seki S. Synthesis Materials Sodium
More informationAnion Exchange HPLC Columns
Anion Exchange PRP-X00 Polymeric anion exchange packings for separation of inorganic and organic anions. Easily separate the eight common anions (fluoride through sulfate). Good separation of fluoride
More informationGPC/SEC. VARIAN, INC. Engineering Polymers ANALYSIS OF ENGINEERING POLYMERS BY GPC/SEC.
GPC/SEC VARIAN, INC. Engineering Polymers ANALYSIS OF ENGINEERING POLYMERS BY GPC/SEC www.varianinc.com GPC/SEC of Engineering Polymers Contents Page Aggressive Solvents GPC Analysis of Polyether Ether
More informationDr. Christoph Johann Wyatt Technology Europe GmbH Copyright Wyatt Technology Europe GmbH All Rights reserved 1
Dr. Christoph Johann Wyatt Technology Europe GmbH 2010 Copyright Wyatt Technology Europe GmbH All Rights reserved 1 Introduction Overview The Nature of Scattered Light: Intensity of scattered light Angular
More informationChapter 5. Ionic Polymerization. Anionic.
Chapter 5. Ionic Polymerization. Anionic. Anionic Polymerization Dr. Houston S. Brown Lecturer of Chemistry UH-Downtown brownhs@uhd.edu What you should know: What is anionic polymerization? What is MWD,
More informationBio10 Cell and Molecular Lecture Notes SRJC
Basic Chemistry Atoms Smallest particles that retain properties of an element Made up of subatomic particles: Protons (+) Electrons (-) Neutrons (no charge) Isotopes Atoms of an element with different
More information