Time-of-Flight Mass Analyzers
|
|
- Lizbeth Lynch
- 6 years ago
- Views:
Transcription
1 Time-of-Flight Mass Analyzers Jonathan Karty C613 lecture 1 March 6, 8 (Section 4. in Gross, pages )
2 TOF Overview Time-of-flight (TOF) is the least complex mass analyzer in terms of its theory Ions are given a defined kinetic energy and allowed to drift through a field-free region (.5 to several meters) The time ions arrive at the detector is measured and related to the m/z ratio
3 TOF Concept A packet of stationary ions is accelerated to a defined kinetic energy and the time required to move through a fixed distance is measured First TOF design published in 1946 by W.E. Stephens Detector
4 TOF advantages Theoretically unlimited mass range Ions are not trapped (quad, IT, FTICR) nor are their flight paths curved (BE sectors) Detection efficiencies induce practical limits of a few hundred kda (M+H)+ Instrument is not scanning (it is dispersive) Analysis is very rapid (4+ khz acquisition possible) Wide range of m/z s can be measured with good sensitivity Moderate to high resolving powers (5,-,+) Moderate cost ($1k to $5k) Relatively high duty cycle Couples extremely well with pulsed ion sources (e.g. MALDI)
5 TOF Disadvantages Requires high vacuum (<1-6 torr) Coupling to continuous ion sources (e.g. ESI or EI) not straight forward Requires complex and high speed electronics High acceleration voltages (5-3 kv) Fast detectors (ns or faster) GHz sampling digital conversion Large volumes of data can be generated quickly Limited dynamic range Often 1 or 1 3 at most High resolution instruments can get rather large
6 From Physics 1: (1) KE = ½mv From Physics : () KE = z*u = ½mv All ions accelerated by the same voltage, U From Physics 1: (3) ΔX= v TOF + ½aTOF * U * z ( 4) v = m 1, Th 19 kv, v 6 km/sec X m ( 6) TOFflight _ tube = * * U z (5) TOF = ΔX/v Time-of-Flight Theory ΔX same for all ions = D (flight tube length) No acceleration in flight tube TOF α U -1 α (m/z) ½
7 Mass Scale Calibration TOF α (m/z) 1/ or m/z α TOF Mass scale is calibrated measuring flight times known m/z ions and fitting them to a polynomial equation (7) TOF = a*(m/z) 1/ + b also Higher order calibrations are often used 5 th order on some commercial instruments Form can be (7a) m/z = A*TOF + B*TOF + C
8 Resolution in TOF MS ( 8) z * U = * m* v * z * U ( 1) v t D (9) m = * z * U v m * z * U ( t D = D / t m = * 11) = *t * z * U 1) m = *t * t D ( ( * z * U ) * t m t (13) = D = m (* z * U ) *t * dt t D Easy way to improve resolution is to increase flight tube length (assuming excellent vacuum)
9 Real-world TOF MS Previous examples all assumed ions formed at rest, at the same time, and all at the same position in the source In reality, ions are formed throughout the source at various times, in various locations, with a range of initial kinetic energies Practical TOF instrument design relies on minimizing the contributions of each of these realities
10 Influence of Initial Position In EI, CI, and ESI sources, ions are NOT formed all in the exact same position These differences in initial position have profound effects on the mass spectrum Ions spend some time in the source prior to crusing through the flight tube (14) TOF obs = TOF source + TOF flight_tube A more complete treatment of the TOF requires that we consider how the ions get accelerated
11 A brief discussion of acceleration Ions are usually accelerated by using parallel plates at different potentials to create electric fields Ions in an electric field gain energy according to the equation: (15) KE = z*(e*s) or U = z*(e*s) z is charge, E is electric field strength in V/m, and s is distance the ion travels in the field (Lorentz force equation) Time of flight in the source must be computed with a differential equation
12 Initial Position Math * U * z ( 4) v = m ( 15)U = E * s (17) v s = * z * E * s m = s t (16) s = * z * E * s m t (19) s * z * E * s = m t (18) t m = * z * E 1 * 1/ s s (1) t s t = s * m z * E s 1/ s +1 V V () t s = m * z * E *s m m () t s = * s (3) t d = * D * z * E * s * z * E * s (14) TOF m = t + td = (4) * * z * E * s ( s D) s +
13 Starting Position Example Both ions are 1 m/z (1.364*1-6 kg/c) Red ion is 6 mm from V plate, blue ion is 5 mm from V plate s for red is.6 m; s for blue is.5 m Distance between plates is 1 cm Electric field is 1, V/m Detector is 1 m from nd grid (4) TOF m = t + td = * * z * E * s TOF red = 94 usec TOF blue = 11 usec E red = 6 ev, E blue = 5 ev ( s D) s + Detector +1 V V This effect tends to cause peaks to tail to longer TOF
14 Initial Kinetic Energy Spread If the ion has a non-zero kinetic energy along the axis of the flight tube prior to acceleration: (5) KE total = z*u+u (U is initial KE) At 5 K, kt =.43 ev Kinetic energies perpendicular to this axis can cause the ions to miss the detector entirely The initial kinetic energy spread and initial position can be accounted for mathematically m* v 6) ( = U + z * E * s ( 7) v = *( U ) + z * E * s m
15 How long do ions spend in the source? 7) v s = *( U + z * E * s) m ( = s t 8) s = *( U + z * E * s) * t m ( (9) t s t = s m * ( U + s z * E * s ) ( m) [( ) ] U * z * E * s ± U ( 3) t s = * + t z * E (initial KE and position) (initial direction) (uncertainty of ion formation)
16 Putting this all together ( m) [( ) ] U * z * E * s ± U ( 3) t s = * + t z * E (31) t D * + ( m) * D ( U z * E * s) = D ( m) (14) TOF = t s + t [( ) ] ( ) m * D U * z * E * s ± U + * ( U + z * E * s) ( 3) TOF = * + t z * E Source design is used to minimize the variables that contribute to peak broadening High U to minimize U Narrow ion formation regions to minimize s Pulsing ions out of source to minimize t
17 TOF designs to maximize resolving power High acceleration voltages can reduce impact of initial KE distribution Initial position effects are harder to minimize Additional ion optical elements can be introduced to compensate for these two effects Set flight tube axis perpendicular to major vector of KE distribution orthogonal extraction
18 Time-lag focusing There is a point in the flight tube where ions starting at different positions arrive simultaneously m m ( 33) ts = t D (34) * s = * D (35) D = s Adding a third grid allows the analyst to move this focus plane to the detector +1 V +9 V ( * z * E * s) ( * z * E * s) 1955, Wiley and McLaren 1994, Colby, King, and IU (special case for MALDI) s is related to initial KE since all ions formed simultaneously Detector V +1 V
19 Reflectrons? In 1966, B. Mamyrin patented an ion mirror device for energy focusing and resolution improvement A reflectron is a long series of electrodes that create an electric field to reverse the direction the ions travel Higher energy ions spend more time in reflectron than lower energy ions By adjusting parameters, ions with same m/z but slightly different KE s can be made to arrive at a detector simultaneously
20 Reflectron drawing +11 V V V +1 V Detector m 36) TOF = * 1 4 * z * E ( ( D + D + ) D ref m 37) TOF = * 1 4 *( z * E + U ) ( z * E + U (38) D = * z * E ref D ref ( D + D + ) D ref
21 Reflectron Pros and Cons Advantages Focuses kinetic energies Better resolution Allows one to use same flight tube twice Remember, res power increases with D Cons Only a narrow range of KE can be focused Usually 85%-15% of E Metastable ions are not focused by reflectron If ion fragments in flight tube, products have same velocity as precursors (but KE is proportional to mass of product) Reflectron focuses by energy
22 Measuring TOF Signals Oscilloscope (aka multi-channel scaler) Analog recorder (ion current vs. time) Great for intense signals Generates large data files (all times recorded) Noise recorded as well as signal Time to Digital Converter Records only when an event occurred Threshold is set so an event is arrival of 1 ion Excellent for low signals (single ion counting) Does NOT take into account intensity ions arriving simultaneously counts as 1 event Limited dynamic range
23 Factors that Influence Resolution Laser pulse width in MALDI Detector response profile Digitization rate and amplifier bandwidths Kinetic energy distribution of ions Initial position of ions in source Power supply stability Response profiles of pulsing electronics
24 Waters LCT Drawing Mass range: 6-, m/z Resolving power: ~5, 54 psec resolution TDC ESI or APCI source MS recorded at ~ khz Table-top MS (1.5x1x1 m) Cost ~$,
Chemistry Instrumental Analysis Lecture 35. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 35 Principle components: Inlet Ion source Mass analyzer Ion transducer Pumps Signal processor Mass analyzers Quadrupole Time of Flight Double Focusing Ion
More informationCEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis
Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511 524) (Harris, Chapt.
More informationCEE 772: Instrumental Methods in Environmental Analysis
Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511-524) (Harris, Chapt.
More informationMass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University
Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk Previously: Resolution and other basics MALDI Electrospray 40 Lecture 2 Mass analysers Detectors
More information1. The range of frequencies that a measurement is sensitive to is called the frequency
CHEM 3 Name Exam 1 Fall 014 Complete these problems on separate paper and staple it to this sheet when you are finished. Please initial each sheet as well. Clearly mark your answers. YOU MUST SHOW YOUR
More informationMASS SPECTROMETRY. Topics
MASS SPECTROMETRY MALDI-TOF AND ESI-MS Topics Principle of Mass Spectrometry MALDI-TOF Determination of Mw of Proteins Structural Information by MS: Primary Sequence of a Protein 1 A. Principles Ionization:
More information(Refer Slide Time 00:09) (Refer Slide Time 00:13)
(Refer Slide Time 00:09) Mass Spectrometry Based Proteomics Professor Sanjeeva Srivastava Department of Biosciences and Bioengineering Indian Institute of Technology, Bombay Mod 02 Lecture Number 09 (Refer
More informationCHAPTER D3 TOF ION OPTICS
Back to Basics Section D: Ion Optics CHAPTER D3 TOF ION OPTICS TABLE OF CONTENTS QuickGuide...399 Summary...401 Background...403 EquationsofMotionofIons...403 Resolution...405 Reflectron...407 Comparison
More informationMS Goals and Applications. MS Goals and Applications
MS Goals and Applications 3 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis
More informationCHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS
Back to Basics Section D: Ion Optics CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS TABLE OF CONTENTS QuickGuide...413 Summary...415 Introduction...417 The physical basis of orthogonal TOF....... 419 Pulsedmainbeamsofions...421
More informationMS Goals and Applications. MS Goals and Applications
MS Goals and Applications 1 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis
More informationMass Spectrometry. What is Mass Spectrometry?
Mass Spectrometry What is Mass Spectrometry? Mass Spectrometry (MS): The generation of gaseous ions from a sample, separation of these ions by mass-to-charge ratio, and measurement of relative abundance
More informationGRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2
DEPARTMENTS OF CHEMISTRY AND BIOCHEMISTRY GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 Mass Analysers Shabaz Mohammed October 20, 2015 High Vacuum System Turbo pumps Diffusion pumps Rough pumps Rotary
More informationMass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS) Mass Analyzers
More informationHarris: Quantitative Chemical Analysis, Eight Edition
Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 21: MASS SPECTROMETRY CHAPTER 21: Opener 21.0 Mass Spectrometry Mass Spectrometry provides information about 1) The elemental composition of
More informationMASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps
ION ANALYZERS MASS ANALYSER sample Vacuum pumps Mass analysers - separate the ions according to their mass-to-charge ratio MASS ANALYSER Separate the ions according to their mass-to-charge ratio in space
More informationLecture 8: Mass Spectrometry
intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?
More informationLecture 8: Mass Spectrometry
intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?
More informationMass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.
Mass Analyzers After the production of ions in ion sources, the next critical step in mass spectrometry is to separate these gas phase ions according to their mass-to-charge ratio (m/z). Ions are extracted
More informationChemistry 311: Topic 3 - Mass Spectrometry
Mass Spectroscopy: A technique used to measure the mass-to-charge ratio of molecules and atoms. Often characteristic ions produced by an induced unimolecular dissociation of a molecule are measured. These
More informationTypes of Analyzers: Quadrupole: mass filter -part1
16 Types of Analyzers: Sector or double focusing: magnetic and electric Time-of-flight (TOF) Quadrupole (mass filter) Linear ion trap Quadrupole Ion Trap (3D trap) FTICR fourier transform ion cyclotron
More informationInstrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai
303351 Instrumental Analysis Mass Spectrometry Lecturer:! Somsak Sirichai Mass Spectrometry What is Mass spectrometry (MS)? An analytic method that employs ionization and mass analysis of compounds in
More informationFundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics
Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass
More informationMass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez
Mass Analyzers Ion Trap, FTICR, Orbitrap CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography Prof. Jose-Luis Jimenez Last Update: Oct. 014 Some slides from Dr. Joel Kimmel (007) MS Interpretation
More informationMass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+.
Mass spectrometry is the study of systems causing the formation of gaseous ions, with or without fragmentation, which are then characteried by their mass to charge ratios (m/) and relative abundances.
More informationScreening of pesticides residues by the time of flight analyzer (ToF) : Myth or reality?
Screening of pesticides residues by the time of flight analyzer (ToF) : Myth or reality? Laure Joly & Vincent Hanot WIV-ISP, Juliette Wytsmanstraat 14, 15 Brussel The passengers pesticides are requested......
More informationMolecular Mass Spectrometry
Molecular Mass Spectrometry Mass Spectrometry: capable of providing information about (1) Elemental composition of samples of matter: atomic mass (2) Structures of inorganic, organic, and biological molecules
More informationMolecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. H i
Gel Permeation Chromatography (GPC) : Size Exclusion Chromatography GPC : 1. Chromatogram (V R vs H) H i Detector response Baseline N i M i 130 135 140 145 150 155 160 165 Elution volume (V R ) (counts)
More informationMS/MS .LQGVRI0606([SHULPHQWV
0DVV6SHFWURPHWHUV Tandem Mass Spectrometry (MS/MS) :KDWLV0606" Mass spectrometers are commonly combined with separation devices such as gas chromatographs (GC) and liquid chromatographs (LC). The GC or
More informationAnalysis of Polar Metabolites using Mass Spectrometry
Analysis of Polar Metabolites using Mass Spectrometry TransMed Course: Basics in Clinical Proteomics and Metabolomics. Oct 10-19, 2012 dd.mm.yyyy Vidya Velagapudi, Ph.D, Adjunct Professor Head of the Metabolomics
More informationICPMS Doherty Lecture 1
ICPMS Doherty Lecture 1 Mass Spectrometry This material provides some background on how to measure isotope abundances by means of mass spectrometry. Mass spectrometers create and separate ionized atoms
More informationLECTURE-13. Peptide Mass Fingerprinting HANDOUT. Mass spectrometry is an indispensable tool for qualitative and quantitative analysis of
LECTURE-13 Peptide Mass Fingerprinting HANDOUT PREAMBLE Mass spectrometry is an indispensable tool for qualitative and quantitative analysis of proteins, drugs and many biological moieties to elucidate
More informationTime of Flight Mass Spectroscopy and Velocity Map Imaging
Time of Flight Mass Spectroscopy and Velocity Map Imaging Geet Ghanshyam January 26, 2013 Velocity map imaging (VMI) is used to study the dynamics of various dissociative electron attachment (DEA) processes
More informationMass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS
Mass Spectrometry Hyphenated Techniques GC-MS LC-MS and MS-MS Reasons for Using Chromatography with MS Mixture analysis by MS alone is difficult Fragmentation from ionization (EI or CI) Fragments from
More informationInterface (backside) & Extraction Lens
Plasma Interface Interface (backside) & Extraction Lens Extraction Lens (-2000 volts) ION OPTICS Tip of the sampler cone is positioned to be in the region of maximum ionization Ions no longer under control
More informationQuadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365
Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365 Effortless Performance The LCMS-9030 quadrupole time-of-flight (Q-TOF) mass spectrometer integrates the world s fastest
More informationCollision Cross Section: Ideal elastic hard sphere collision:
Collision Cross Section: Ideal elastic hard sphere collision: ( r r 1 ) Where is the collision cross-section r 1 r ) ( 1 Where is the collision distance r 1 r These equations negate potential interactions
More informationA Simple Multi-Turn Time of Flight Mass Spectrometer MULTUM II
J. Mass Spectrom. Soc. Jpn. Vol. 51, No. 2, 23 REGULAR PAPER A Simple Multi-Turn Time of Flight Mass Spectrometer MULTUM II Daisuke O@JBJG6, a) Michisato TDND96, a) Morio IH=>=6G6, a) and Itsuo K6I6@JH:
More informationMass spectrometry.
Mass spectrometry Mass spectrometry provides qualitative and quantitative information about the atomic and molecular composition of inorganic and organic materials. The mass spectrometer produces charged
More information1. Prepare the MALDI sample plate by spotting an angiotensin standard and the test sample(s).
Analysis of a Peptide Sequence from a Proteolytic Digest by MALDI-TOF Post-Source Decay (PSD) and Collision-Induced Dissociation (CID) Standard Operating Procedure Purpose: The following procedure may
More informationTANDEM MASS SPECTROSCOPY
TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER
More informationLECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile
LECTURE-11 Hybrid MS Configurations HANDOUT PREAMBLE As discussed in our previous lecture, mass spectrometry is by far the most versatile technique used in proteomics. We had also discussed some of the
More informationChemistry Instrumental Analysis Lecture 34. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 34 From molecular to elemental analysis there are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry
More information422 Brazilian Journal of Physics, vol. 29, no. 3, September, An Introduction to the Time-of-Flight Technique. Per Hakansson
422 Brazilian Journal of Physics, vol. 29, no. 3, September, 1999 An Introduction to the Time-of-Flight Technique Per Hakansson The Angstrom Laboratory, Div. of Ion Physics, Box 534, S-751 21 Uppsala,
More informationIntroduction to LC-MS
Wednesday April 5, 2017 10am Introduction to LC-MS Amy Patton, MS Laboratory Manager, Pinpoint Testing, LLC Little Rock, AR DESCRIPTION: Amy Patton, laboratory manager for Pinpoint Testing, will begin
More information1) Introduction. 2) Types of Mass Analyzers - Overview. Assigned Reading: Nier mass spectrometry paper Extra Reading:
Assigned Reading: Nier mass spectrometry paper Extra Reading: 1) Introduction a) We are now on to the heart of the mass spectrometer. How do you separate ions of different mass, so that you can detect/quantify
More informationTheory English (Official)
Q3-1 Large Hadron Collider (10 points) Please read the general instructions in the separate envelope before you start this problem. In this task, the physics of the particle accelerator LHC (Large Hadron
More informationPhysicsAndMathsTutor.com 1
PhysicsAndMathsTutor.com 1 1. Millikan determined the charge on individual oil droplets using an arrangement as represented in the diagram. The plate voltage necessary to hold a charged droplet stationary
More informationIntroduction to GC/MS
Why Mass Spectrometry? Introduction to GC/MS A powerful analytical technique used to: 1.Identify unknown compounds 2. Quantify known materials down to trace levels 3. Elucidate the structure of molecules
More informationLaser Dissociation of Protonated PAHs
100 Chapter 5 Laser Dissociation of Protonated PAHs 5.1 Experiments The photodissociation experiments were performed with protonated PAHs using different laser sources. The calculations from Chapter 3
More informationMacromolecular Chemistry
Macromolecular Chemistry Lecture 8 Measuring Molecular Weight Membrane Osmometry Alfredo Vapor Phase Osmometry Linda Viscometry GW Gel Permeation Chromatography Size exclusion Chromatography Light Scattering
More informationTranslational Biomarker Core
Translational Biomarker Core Instrumentation Thermo Scientific TSQ Quantum Triple Quadrupole Mass Spectrometers. There are two TSQ Quantum Ultra AM instruments available in the TBC. The TSQ Quantum Ultra
More informationWeek 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap
Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap 1 Last Time Mass Analyzers; CAD and TOF mass analyzers: 2 Fourier Transforms A transform is when you change your analytical space without
More informationMass Spectrometry in MCAL
Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring
More information[ instrument specifications ]
Designed for leading researchers working at the limits of conventional mass spectrometry capabilities who need to further characterize and define their samples the Waters SYNAPT High Definition MS (HDMS
More informationMass spectrometry gas phase transfer and instrumentation
Objectives of the Lecture spectrometry gas phase transfer and instrumentation Matt Renfrow January 15, 2014 1. Make ions 2. Separate/Analyze 3. Detect ions 4. What is mass resolution and mass accuracy?
More informationLecture 9. PMTs and Laser Noise. Lecture 9. Photon Counting. Photomultiplier Tubes (PMTs) Laser Phase Noise. Relative Intensity
s and Laser Phase Phase Density ECE 185 Lasers and Modulators Lab - Spring 2018 1 Detectors Continuous Output Internal Photoelectron Flux Thermal Filtered External Current w(t) Sensor i(t) External System
More informationChem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY:
Chem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY: I. BASIC ELEMENTS OF A MASS SPECTROMETER Inlet System or Chromatograph Ion Source Mass Analyzer Detector Computer II. ION SOURCES A. Electron
More informationTime of Flight Mass Spectrometry of Ions Generated by Molecules in Intense Laser Fields
Time of Flight Mass Spectrometry of Ions Generated by Molecules in Intense Laser Fields Mingtong Han 5 th August, 013 Abstract Photoionization of acetylene, dimethylacetylene and methylacetylene in an
More informationFactors Affecting Mass Measurement Accuracy in a TOF Mass Spectrometer. Curtis D. Cleven, James Little, Eastman Chemical Company, Kingsport, TN
Factors Affecting Mass Measurement Accuracy in a TOF Mass Spectrometer Curtis D. Cleven, James Little, Eastman Chemical Company, Kingsport, TN INTRODUCTION 1 Mass measurement accuracy and precision are
More informationFourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS)
Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS) The first steps along the way to FT- ICR-MS date back to the 1930s, when Ernest Lawrence, at the University of California - Berkeley,
More informationLecture 9: Introduction to QM: Review and Examples
Lecture 9: Introduction to QM: Review and Examples S 1 S 2 Lecture 9, p 1 Photoelectric Effect V stop (v) KE e V hf F max stop Binding energy F The work function: F is the minimum energy needed to strip
More informationLC-MS Based Metabolomics
LC-MS Based Metabolomics Analysing the METABOLOME 1. Metabolite Extraction 2. Metabolite detection (with or without separation) 3. Data analysis Metabolite Detection GC-MS: Naturally volatile or made volatile
More informationSolutions for Exercise session I
Solutions for Exercise session I 1. The maximally polarisation-entangled photon state can be written as Ψ = 1 ( H 1 V V 1 H ). Show that the state is invariant (i.e. still maximally entangled) after a
More informationLarge Plasma Device (LAPD)
Large Plasma Device (LAPD) Over 450 Access ports Computer Controlled Data Acquisition Microwave Interferometers Laser Induced Fluorescence DC Magnetic Field: 0.05-4 kg, variable on axis Highly Ionized
More informationLecture 22 Ion Beam Techniques
Lecture 22 Ion Beam Techniques Schroder: Chapter 11.3 1/44 Announcements Homework 6/6: Will be online on later today. Due Wednesday June 6th at 10:00am. I will return it at the final exam (14 th June).
More informationMass Analysers for LC MS
Mass Analysers for LC MS Filip Lemière, Dept of Chemistry, University of Antwerp, Belgium. Image courtesy of Bruker Daltonik GmbH Introduction The major technical challenge in LC MS is interfacing the
More informationMass Spectrometry. Fundamental LC-MS. Mass Analysers
Mass Spectrometry Fundamental LC-MS Mass Analysers i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference
More informationParticle physics experiments
Particle physics experiments Particle physics experiments: collide particles to produce new particles reveal their internal structure and laws of their interactions by observing regularities, measuring
More informationMass Spectrometry (MS)
Mass Spectrometry (MS) Alternative names: Mass spectrometric (selective) detector (MSD) Spectrometry - methods based on interaction of matter and radiation Mass spectrometry - method based on formation
More informationFall Quarter 2010 UCSB Physics 225A & UCSD Physics 214 Homework 1
Fall Quarter 2010 UCSB Physics 225A & UCSD Physics 214 Homework 1 Problem 2 has nothing to do with what we have done in class. It introduces somewhat strange coordinates called rapidity and pseudorapidity
More informationJoint Institute for Nuclear Research. Flerov Laboratory of Nuclear Reactions. University Center LABORATORY WORK
Joint Institute for Nuclear Research Flerov Laboratory of Nuclear Reactions University Center LABORATORY WORK TOF measurement of the charged particles using the MCP detector Supervisor: Dauren Aznabayev
More informationAbstract... I. Acknowledgements... III. Table of Content... V. List of Tables... VIII. List of Figures... IX
Abstract... I Acknowledgements... III Table of Content... V List of Tables... VIII List of Figures... IX Chapter One IR-VUV Photoionization Spectroscopy 1.1 Introduction... 1 1.2 Vacuum-Ultraviolet-Ionization
More informationExperiment 2 Deflection of Electrons
Name Partner(s): Experiment 2 Deflection of Electrons Objectives Equipment Preparation Pre-Lab To study the effects of electric fields on beams of fast moving electrons. Cathode-ray tube (CRT), voltage
More information(Total for Question = 5 marks) PhysicsAndMathsTutor.com
1 Rutherford designed an experiment to see what happened when alpha particles were directed at a piece of gold foil. Summarise the observations and state the conclusions Rutherford reached about the structure
More informationChapter-4 Stimulated emission devices LASERS
Semiconductor Laser Diodes Chapter-4 Stimulated emission devices LASERS The Road Ahead Lasers Basic Principles Applications Gas Lasers Semiconductor Lasers Semiconductor Lasers in Optical Networks Improvement
More informationin Electronic Devices and Circuits
in Electronic Devices and Circuits Noise is any unwanted excitation of a circuit, any input that is not an information-bearing signal. Noise comes from External sources: Unintended coupling with other
More informationMass Spectrometry. Electron Ionization and Chemical Ionization
Mass Spectrometry Electron Ionization and Chemical Ionization Mass Spectrometer All Instruments Have: 1. Sample Inlet 2. Ion Source 3. Mass Analyzer 4. Detector 5. Data System http://www.asms.org Ionization
More informationMALDI-TOF: a powerful tool for the analysis of biomolecules and large organic molecules
III Workshop 2013-2014 Corso del Dottorato in Ingegneria dei Materiali del Politecnico di Milano Methods of Characterization of Materials MALDI-TOF: a powerful tool for the analysis of biomolecules and
More informationCh. 22 Mass Spectrometry (MS)
Ch. Mass Spectroetry (MS).1 MS easures ass of atos, olecules, or fragents of olecules -1. What is MS? Gaseous ato fro condensed phase ionized Accelerated & separated By ass to charge ratio /z M=00: 1 ion
More informationHomework 2: Forces on Charged Particles
Homework 2: Forces on Charged Particles 1. In the arrangement shown below, 2 C of positive charge is moved from plate S, which is at a potential of 250 V, to plate T, which is at a potential of 750 V.
More informationRadioactivity APPARATUS INTRODUCTION PROCEDURE
Radioactivity APPARATUS. Geiger Counter / Scaler. Cesium-7 sealed radioactive source. 0 pieces of paper. 8 aluminum plates. 0 lead plates 6. Graph paper - log-log and semi-log 7. Survey Meter ( unit for
More informationKE = 1 2 mv2 = ev. (1)
The e/m ratio Objective To measure the electronic charge-to-mass ratio e/m, by injecting electrons into a magnetic field and examining their trajectories. We also estimate the magnitude of the earth s
More informationLecture 11: Doppler wind lidar
Lecture 11: Doppler wind lidar Why do we study winds? v Winds are the most important variable studying dynamics and transport in the atmosphere. v Wind measurements are critical to improvement of numerical
More informationMechanism of Insulin Emission in Matrix Assisted Laser Ionization
170 Brazilian Journal of Physics, vol. 35, no. 1, March, 2005 Mechanism of Insulin Emission in Matrix Assisted Laser Ionization F. Fernández-Lima 1, V. M. Collado 2, C. R. Ponciano 2, L. S. Farenzena 2,
More informationpickup from external sources unwanted feedback RF interference from system or elsewhere, power supply fluctuations ground currents
Noise What is NOISE? A definition: Any unwanted signal obscuring signal to be observed two main origins EXTRINSIC NOISE examples... pickup from external sources unwanted feedback RF interference from system
More informationv = E B FXA 2008 UNIT G485 Module Magnetic Fields BQv = EQ THE MASS SPECTROMETER
UNIT G485 Module 1 5.1.2 Magnetic Fields 11 Thus, in order for the particle to suffer NO DEFLECTION and so exit the device at Y : From which : MAGNETIC FORCE UP = ELECTRIC FORCE DOWN BQv = EQ THE MASS
More informationSecondary Ion Mass Spectroscopy (SIMS)
Secondary Ion Mass Spectroscopy (SIMS) Analyzing Inorganic Solids * = under special conditions ** = semiconductors only + = limited number of elements or groups Analyzing Organic Solids * = under special
More informationMagnetic fields applied to laser-generated plasma to enhance the ion yield acceleration
Magnetic fields applied to laser-generated plasma to enhance the ion yield acceleration L. Torrisi, G. Costa, and G. Ceccio Dipartimento di Scienze Fisiche MIFT, Università di Messina, V.le F.S. D Alcontres
More informationMass Spectrometry. A truly interdisciplinary and versatile analytical method
Mass Spectrometry A truly interdisciplinary and versatile analytical method MS is used for the characterization of molecules ranging from small inorganic and organic molecules to polymers and proteins.
More informationDifferential Cross Section Measurements in Ion-molecule Collisions
Differential Cross Section Measurements in Ion-molecule Collisions Song Cheng Department of Physics and Astronomy, University of Toledo, Toledo, Ohio 43606 A 14 m long beam line dedicated to study very
More informationThermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer
IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific LTQ Orbitrap Velos
More informationChapter 4 Scintillation Detectors
Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 4-1 4.1. Basic principle of the scintillator Chapter 4 Scintillation Detectors Scintillator Light sensor Ionizing radiation Light (visible,
More informationMass spectrometry of proteins, peptides and other analytes: principles and principal methods. Matt Renfrow January 11, 2008
Mass spectrometry of proteins, peptides and other analytes: principles and principal methods Matt Renfrow January 11, 2008 Objectives of the Lecture 1. Make ions 2. Separate/Analyze/Detect ions 3. What
More informationRANGE OF ALPHAS. Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706
(revised 10/20/10) RANGE OF ALPHAS Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706 Abstract A silicon solid state detector is used to measure the energy of alphas which
More informationM M e M M H M M H. Ion Sources
Ion Sources Overview of Various Ion Sources After introducing samples into a mass spectrometer, the next important step is the conversion of neutral molecules or compounds to gas phase ions. The ions could
More informationAutomatic Control Systems. -Lecture Note 15-
-Lecture Note 15- Modeling of Physical Systems 5 1/52 AC Motors AC Motors Classification i) Induction Motor (Asynchronous Motor) ii) Synchronous Motor 2/52 Advantages of AC Motors i) Cost-effective ii)
More informationVIII. QUADRUPOLE ION TRAPS: AN ION TRAPPING INST. A. BASIC DESIGN (REF: JMS, 1997, 32, 351)
VIII. QUADRUPOLE ION TRAPS: AN ION TRAPPING INST. A. BASIC DESIGN (REF: JMS, 1997, 32, 351) The trap consists of a ring electrode (internal radius is r 0 ) and two end caps separated by z 0.; all internal
More informationExtrel Application Note
Extrel Application Note Real-Time Plasma Monitoring and Detection of Trace H 2 O and HF Species in an Argon Based Plasma Jian Wei, 575 Epsilon Drive, Pittsburgh, PA 15238. (Presented at the 191st Electrochemical
More informationChapter 10: Wave Properties of Particles
Chapter 10: Wave Properties of Particles Particles such as electrons may demonstrate wave properties under certain conditions. The electron microscope uses these properties to produce magnified images
More information