[ instrument specifications ]

Size: px
Start display at page:

Download "[ instrument specifications ]"

Transcription

1 Designed for leading researchers working at the limits of conventional mass spectrometry capabilities who need to further characterize and define their samples the Waters SYNAPT High Definition MS (HDMS ) System offers unique, enabling functionality. Utilizing Waters patented Triwave technology, the SYNAPT HDMS System combines high-efficiency, ion-mobility based measurements and separations with high-performance quadrupole, time-of-flight (TOF) mass spectrometry. The SYNAPT HDMS System enables researchers to analyze samples that are differentiated by size, shape and charge as well as mass to deliver increased specificity and sample definition beyond that achievable by conventional mass spectrometers. PERFORMANCE SPECIFICATIONS LOCKMASS REFERENCE SPRAY T-WAVE ION GUIDE DRE LENS PUSHER DETECTOR QUADRUPOLE The SYNAPT HDMS System can operate in two modes: ANALYTE SPRAY TRAP ION MOBILITY SEPARATION TRANSFER (a) TOF mode (b) Mobility-TOF mode TOF Mass Resolution in Positive Ion 1 2 OIL-FREE AIR-COOLED TURBOMOLECULAR PUMPS SCROLL PUMP 5 10,000 measured on the (M + 6H) 6+ isotope cluster from bovine insulin (m/z 956) (b) W Mode 17,500 measured on the (M + 6H) 6+ isotope cluster from bovine insulin (m/z 956) TOF Mass Resolution in Negative Ion 10,000 measured on the (M - 4H) 4- isotope cluster from bovine insulin (m/z 1431) (b) W Mode 17,500 measured on the (M - 4H) 4- isotope cluster from bovine insulin (m/z 1431) Positive Ion MS Sensitivity The peak at m/z 556 from a solution of 50 pg/μl leucine enkephalin in 50/50 acetonitrile/water + 0.1% formic acid, infused at a flow rate of 5 μl/min, will have an intensity of greater than 1,700 counts per second. The instrument will be tuned to 10,000 resolution (as demonstrated on bovine insulin) and the mass range will be set to 1,000 Da. (b) EDC The peak at m/z 556 from a solution of 10 pg/μl leucine enkephalin in 50/50 acetonitrile/water + 0.1% formic acid, infused at a flow rate of 5 μl/min, will have an intensity of greater than 2,000 counts per second. The instrument will be tuned to 10,000 resolution (as demonstrated on bovine insulin), with sensitivity set to a maximum at 556 Da. Negative Ion MS Sensitivity The peak at m/z 503 from a solution of 500 pg/μl raffinose in 70/30 acetonitrile/water (no additives), infused at a flow rate of 5 μl/min, will have an intensity of greater than 1,800 counts per second. The instrument will be tuned to 10,000 resolution (as demonstrated on bovine insulin), and the mass range will be set to 1,000 Da.

2 Positive Ion MS/MS Sensitivity Using a [Glu] -Fibrinopeptide B solution of 100 fmol/μl, at a flow rate of 5μL/min and with the instrument tuned for 10,000 resolution (as demonstrated on bovine insulin), the intensity of the most intense y sequence ion from the MS/MS spectrum of the doubly charged precursor ion (785.8 Da) will be greater than 130 counts per second. This will correspond to a signal to noise ratio of greater than 50:1 (after a 3x9 smooth) on the most intense y sequence ion, for a 1 second scan. The instrument mass range will be set to 2,000 Da. Negative Ion MS/MS Sensitivity Using a solution of 500 pg/μl raffinose in 70/30 acetonitrile/water, at a flow rate of 5 μl/min and with the instrument tuned for 10,000 resolution (as demonstrated on bovine insulin), the intensity of the fragment ion at Da in the MS/MS spectrum of the precursor ion at Da will be greater than 130 counts per second. The instrument mass range will be set to 1,000 Da. Mass Scale Calibration Accuracy The mass measurement accuracy of the instrument in V-mode, using an internal lockmass, is such that the RMS error between the measured and the accepted masses of peaks which have sufficient intensity, and are free from interference from other masses, will be less than 2 ppm over the range Da. Mass Measurement Accuracy The mass measurement accuracy of the instrument, in W-mode, will be better than 2 ppm RMS, based on 10 consecutive repeat measurements of the [M + Na] + ion of raffinose (m/z ), using the [M + H] + ion of leucine enkephalin ( ) as the LockSpray lockmass. Analyte and lockmass peaks must have sufficient intensity and be free of interference from other masses. Mass Range The TOF mass range is ,000 Da in V-mode, and 20 26,500 Da in W-mode. The quadrupole mass range in non-resolving mode is 20 16,000 Da for a 4,000 Da quadrupole, and 20 32,000 Da for an 8,000 Da quadrupole.

3 Acquisition Rate Spectra can be acquired at a rate of 10 scans per second, except in pdre mode, when the maximum rate is 5 scans per second. Conditions: inter-scan time 20 ms, centroid or continuum, mass range 100 1,000 Da, single spray source. The sensitivity at 10 scans per second should be greater than 80% of the sensitivity at 1 scan per second. Dynamic Range The dynamic range, defined as the range of peak intensities that will give better than 3 ppm accurate mass (95% confidence) for 10 seconds of data, is at least 4 orders of magnitude, when measured on the m/z peak from leucine enkephalin. Isotope Ratios In the molecular ion isotope cluster of leucine enkephalin, the [M+H+1] + isotope will have an intensity of 33.8 (±1.0)%, the [M+H+2] + isotope will have an intensity of 6.9 (±1.0) %, and the [M+H+3] + isotope will have an intensity of 1.1 (±1.0) %, all measured with respect to the intensity of the [M+H] + isotope at Da. The peaks must be free of interference from other masses. Peak intensities must be below the DDTC (Digital Dead Time Correction) limit. High Mass Precursor Selection Applicable to instruments with 8,000 Da quadrupole only. The low energy MS/MS spectrum of m/z from a solution of 2 μg/μl sodium iodide in 50/50 isopropanol/water will contain only m/z and its fragments. The intensity of the largest fragment ion will be less than 5% of the intensity of the precursor ion. MS/MS data will be acquired over the mass range 100 8,000 Da, with collision energy of 10 ev. MALDI-TOF MODE PERFORMANCE SPECIFICATIONS Mass Resolution in Positive Ion (a) V mode 8,000 measured on the (m + H) + isotope cluster from insulin B chain (m/z ). (b) W mode 13,000 measured on the (m + H) + isotope cluster from insulin B chain (m/z ).

4 Mass Resolution in Negative Ion LOCKMASS REFERENCE SPRAY T-WAV E ION GUIDE DRE LENS PUSHER DETECTOR (a) V mode ANALYTE SPRAY QUADRUPOLE TRAP ION MOBILITY SEPARATION TRANSFER 8,000 measured on the (m - H) - isotope cluster from insulin B chain (m/z ). (b) W mode 1 2 OIL-FREE AIR-COOLED TURBOMOLECULAR PUMPS SCROLL PUMP 5 13,000 measured on the (m - H) - isotope cluster from insulin B chain (m/z ). Positive Ion MS Sensitivity Measured in The peak at m/z from 10 fmol of [Glu] -Fibrinopeptide B will have an intensity of greater than 30,000 counts acquired from an entire line through a sample well. The instrument mass range will be set to 2,000 Da. When this combined acquisition is smoothed (5 window, 1 number Savitzky Golay) and background subtracted the signal-tonoise ratio will be greater than 90:1 when compared to a region of the mass scale between 1,768 and 1,818 Da. Negative Ion MS Sensitivity Measured in The peak at m/z from 100 fmol of [Glu] -Fibrinopeptide B will have an intensity of greater than 30,000 counts acquired from an entire line through a sample well. The instrument mass range will be set to 2,000 Da. When this combined acquisition is smoothed (5 window, 1 number savitzy Golay) and background subtracted the signal-tonoise ratio should be greater than 90:1 when compared to a region of the mass scale between 1,768 and 1,818 Da. Positive Ion MS/MS Sensitivity Measured in V mode, Precursor Mass of The peak at m/z (y 9 ) from 10 fmol of [Glu] -Fibrinopeptide B will have an intensity of greater than 1,000 counts for an entire line through a sample well. The instrument mass range will be set to 1,620 (Precursor+50) and the collision energy and instrument conditions set such that the intensity of this peak will be greater than 40% of the intensity of the precursor peak. When this combined acquisition is smoothed (5 window, 1 number savitzy Golay) and background subtracted the signalto-noise ratio should be greater than 60:1 for the peak when compared to a region of the mass scale between 1,450 and 1,490 Da. Mass Measurement Measured in V mode, Positive Ion The mass measurement accuracy will be better than 2 ppm rms, measured from a mixture of PEG oligomers between 700-2,500 Da; using an internal reference peak and an instrument calibration that covers the same mass range.

5 MOBILIT Y-TOF MODE PERFORMANCE SPECIFICATIONS The Triwave cell is the enabling technology of the SYNAPT HDMS System. It is comprised of three main components: (a) Trap T-Wave Ensures high-efficiency by trapping ions prior to ion mobility separation; the trap can also operate as a collision cell. (b) IMS T-Wave Provides reproducible separation of ions based on their ion mobility. (c) Transfer T-Wave Transfers separated ions to the oa-tof for mass analysis; the transfer T-Wave can also operate as a collision cell. Mobility Functionality Separation of ions based on their mobility (size, shape and charge) Time Aligned Parallel (TAP) fragmentation High duty cycle (HDC) for extended detection limits Drift Time measurements: Record drift times between 33µs and 90 ms Control of Mobility Separation Control of Triwave height and velocity over linear ramps and user defined programs Mobility gas pressure control from ~10-4 mbar to 1 mbar (nitrogen) Provision for multiple mobility separation gas types Automated control of mobility conditions for routine operation Comprehensive manual control of mobility and Triwave operation for research use Waters is a registered trademark of Waters Corporation. HDMS, High Definition MS, Synapt, Triwave, and The Science of What s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners Waters Corporation. Printed in the U.S.A. January EN JH-PDF Waters Corporation 34 Maple Street Milford, MA U.S.A T: F:

Traditional Herbal Medicine Structural Elucidation using SYNAPT HDMS

Traditional Herbal Medicine Structural Elucidation using SYNAPT HDMS Traditional Herbal Medicine Structural Elucidation using SYNAPT HDMS with Time-Aligned Parallel (TAP) Fragmentation Kate Yu, Jose Castro-Perez, and John Shockcor Waters Corporation, Milford, MA, U.S. INTRODUCTION

More information

Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry

Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry Jonathan Williams Waters Corporation, Milford, MA, USA A P P L I C AT

More information

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays : Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays Mark Wrona, Jayne Kirk, and Yun Alelyunas Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Ability

More information

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION Quantitation by High Resolution Mass Spectrometry: Using Target Enhancement and Tof-MRM to Achieve Femtogram-level On-column Sensitivity for Quantitation of Drugs in Human Plasma Yun W. Alelyunas, Mark

More information

MALDI-HDMS E : A Novel Data Independent Acquisition Method for the Enhanced Analysis of 2D-Gel Tryptic Peptide Digests

MALDI-HDMS E : A Novel Data Independent Acquisition Method for the Enhanced Analysis of 2D-Gel Tryptic Peptide Digests -HDMS E : A Novel Data Independent Acquisition Method for the Enhanced Analysis of 2D-Gel Tryptic Peptide Digests Emmanuelle Claude, 1 Mark Towers, 1 and Rachel Craven 2 1 Waters Corporation, Manchester,

More information

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance Michael D. Jones, Marian Twohig, Karen Haas, and Robert S. Plumb Waters Corporation,

More information

Fereshteh Zandkarimi, Samanthi Wickramasekara, Jeff Morre, Jan F. Stevens and Claudia S. Maier

Fereshteh Zandkarimi, Samanthi Wickramasekara, Jeff Morre, Jan F. Stevens and Claudia S. Maier Chapter 2 Electrospray Ionization Traveling Wave Ion Mobility Spectrometry Mass Spectrometry for the Analysis of Plant Phenolics: An Approach for Separation of Regioisomers Fereshteh Zandkarimi, Samanthi

More information

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS Marian Twohig, Michael D. Jones, Dominic Moore, Peter Lee, and Robert Plumb Waters Corporation, Milford, MA, USA APPLICATION

More information

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate Michael D. Jones, Xiang Jin Song, Robert S. Plumb, Peter J. Lee, and

More information

Multi-stage Mass Spectrometry up to MS 4 on a QTof System

Multi-stage Mass Spectrometry up to MS 4 on a QTof System Valerie Koppen, 1 Russell Mortishire-Smith, 2 Filip Cuyckens 1 1 Janssen R&D, Beerse, Belgium 2 Waters Corporation, Wilmslow, United Kingdom APPLICATION BENEFITS This application note illustrates how multi-stage

More information

Q-TOF PREMIER DYNAMIC RANGE ENHANCEMENT AND ACCURATE MASS MEASUREMENT PERFORMANCE

Q-TOF PREMIER DYNAMIC RANGE ENHANCEMENT AND ACCURATE MASS MEASUREMENT PERFORMANCE Q-TOF PREMIER DYAMIC RAGE EHACEMET AD ACCURATE MASS MEASUREMET PERFORMACE Michael McCullagh, Jose Castro-Perez, and Lisa Calton Waters Corporation, Manchester, UK ITRODUCTIO The use of time-of-flight (TOF)

More information

Factors Affecting Mass Measurement Accuracy in a TOF Mass Spectrometer. Curtis D. Cleven, James Little, Eastman Chemical Company, Kingsport, TN

Factors Affecting Mass Measurement Accuracy in a TOF Mass Spectrometer. Curtis D. Cleven, James Little, Eastman Chemical Company, Kingsport, TN Factors Affecting Mass Measurement Accuracy in a TOF Mass Spectrometer Curtis D. Cleven, James Little, Eastman Chemical Company, Kingsport, TN INTRODUCTION 1 Mass measurement accuracy and precision are

More information

APPLICATION SOLUTIONS FOR STRUCTURAL PROTEOMICS

APPLICATION SOLUTIONS FOR STRUCTURAL PROTEOMICS APPLICATION SOLUTIONS FOR STRUCTURAL PROTEOMICS TABLE OF CONTENTS The Application of SYNAPT High Defintion Mass Spectrometry for the Conformational Studies of Protein Complexes... 1 High Definition Mass

More information

Analysis of Labeled and Non-Labeled Proteomic Data Using Progenesis QI for Proteomics

Analysis of Labeled and Non-Labeled Proteomic Data Using Progenesis QI for Proteomics Analysis of Labeled and Non-Labeled Proteomic Data Using Progenesis QI for Proteomics Lee Gethings, Gushinder Atwal, Martin Palmer, Chris Hughes, Hans Vissers, and James Langridge Waters Corporation, Wilmslow,

More information

Improved Ion Mobility Separation of Protein Conformations in the Gas Phase with SYNAPT G2 HDMS

Improved Ion Mobility Separation of Protein Conformations in the Gas Phase with SYNAPT G2 HDMS Improved Ion Mobility Separation of Protein Conformations in the Gas Phase with SYNAPT G2 HDMS St John Skilton, Iain Campuzano, Kevin Giles, Hilary Major, and Weibin Chen Waters Corporation, Milford, MA,

More information

Electron Transfer Dissociation of N-linked Glycopeptides from a Recombinant mab Using SYNAPT G2-S HDMS

Electron Transfer Dissociation of N-linked Glycopeptides from a Recombinant mab Using SYNAPT G2-S HDMS Electron Transfer Dissociation of N-linked Glycopeptides from a Recombinant mab Using SYNAPT G2-S HDMS Jonathan P. Williams, Jeffery M. Brown, Stephane Houel, Ying Qing Yu, and Weibin Chen Waters Corporation,

More information

PosterREPRINT COMPARISON OF PEAK PARKING VERSUS AUTOMATED FRACTION ANALYSIS OF A COMPLEX PROTEIN MIXTURE. Introduction

PosterREPRINT COMPARISON OF PEAK PARKING VERSUS AUTOMATED FRACTION ANALYSIS OF A COMPLEX PROTEIN MIXTURE. Introduction Introduction The study of protein expression allows a greater understanding of biological function and can now routinely be performed using mass spectrometry. However, analysis of the post-translational

More information

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific LTQ Orbitrap Velos

More information

Finnigan LCQ Advantage MAX

Finnigan LCQ Advantage MAX www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Finnigan LCQ Advantage MAX The Finnigan LCQ Advantage MAX ion trap mass spectrometer

More information

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW Paula Hong, John Van Antwerp, and Patricia McConville Waters Corporation, Milford, MA, USA Historically UV detection has been

More information

Thermo Finnigan LTQ. Specifications

Thermo Finnigan LTQ. Specifications IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Finnigan LTQ Specifications

More information

INTRODUCTION OVERVIEW

INTRODUCTION OVERVIEW COMPARISON OF CCS(N 2 ) MEASUREMENTS OBTAINED FROM TWO DIFFERENT T-WAVE ION S WITH DIRECT MEASUREMENTS USING A DRIFT TUBE ION Kevin Giles, Martin Palmer, Keith Richardson, Nick Tomczyk Waters MS Technologies

More information

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada) Quantitation and Identification of Pharmaceuticals and Personal Care Products (PPCP) in Environmental Samples using Advanced TripleTOF MS/MS Technology André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate Yu 3

Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate Yu 3 Using Natural Products Application Solution with UNIFI for the Comparison of the Chemical Ingredients of Shuanghuanglian Oral Drink from Two Different Manufacturers Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper,

More information

A Case of Pesticide Poisoning: The Use of a Broad-Scope Tof Screening Approach in Wildlife Protection

A Case of Pesticide Poisoning: The Use of a Broad-Scope Tof Screening Approach in Wildlife Protection A Case of Pesticide Poisoning: The Use of a Broad-Scope Tof Screening Approach in Wildlife Protection George Keenan 1, Anna Giela 1, Michael Taylor 1, Jennifer Watson 1, Eleanor Riches 2, James Morphet

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment LCQ Deca XP Plus Improved Ion Optics for Greater Sensitivity and Precision

More information

Triwave. More Complete Characterization of Mixtures and Molecules

Triwave. More Complete Characterization of Mixtures and Molecules Triwave More Complete Characterization of Mixtures and Molecules INTRODUCTION This white paper describes: Benefits of T-Wave Ion Mobility Separations (IMS) Triwave design and implementation of T-Wave IMS

More information

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity Janet Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS High

More information

C HA R AC T E RIZ AT IO N O F INK J E T P RINT E R C A RT RIDG E INK S USING A CHEMOMETRIC APPROACH

C HA R AC T E RIZ AT IO N O F INK J E T P RINT E R C A RT RIDG E INK S USING A CHEMOMETRIC APPROACH C HA R AC T E RIZ AT I N F INK J E T P RINT E R C A RT RIDG E INK S USING A CHEMMETRIC APPRACH Diana Uría and Paul Silcock Waters Corporation, Manchester, UK AIM To evaluate the potential of UPLC -ToF-MS

More information

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies APPLICATION BENEFITS Regulatory guidelines and recommendations place a greater emphasis on the detection

More information

MICROMASS Q-Tof micro MASS SPECTROMETER

MICROMASS Q-Tof micro MASS SPECTROMETER www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment MICROMASS Q-Tof micro MASS SPECTROMETER QTOF Micro Performance Specifications

More information

Protein identification is made easy with Micromass optional BioLynx and ProteinLynx software packages.

Protein identification is made easy with Micromass optional BioLynx and ProteinLynx software packages. SPECIFICATION NOTE 126 Q-Tof 2 Q-Tof 2 is a new type of mass spectrometry tool for peptide sequencing, and protein characterisation and metabolite identification, a powerful alternative to the conventional

More information

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Heather A Brown, 1 Claudia Rossi, 2 and Lisa J Calton 1 1 Waters Corporation, Wilmslow, UK 2

More information

ICP-MS. High Resolution ICP-MS.

ICP-MS. High Resolution ICP-MS. ICP-MS attom ES High Resolution ICP-MS www.nu-ins.com Attom ES Enhanced Sensitivity Enhanced Speed Enhanced Selectivity Enhanced Software The Attom ES from Nu Instruments is a double focussing inductively

More information

Analyst Software. Automatic Optimization Tutorial

Analyst Software. Automatic Optimization Tutorial This document is provided to customers who have purchased AB Sciex equipment to use in the operation of such AB Sciex equipment. This document is copyright protected and any reproduction of this document

More information

UPLC Intact MASS Analysis Application Kit

UPLC Intact MASS Analysis Application Kit (Part Number: 176001519) CONTENTS I. INTACT MASS ANALYSIS APPLICATION KIT COMPONENTS II. INSTALLING THE PRE-COLUMN TUBING INTO MASSPREP MICRO DESALTING COLUMN III. CONNECTING THE COLUMN TO LCT PREMIER

More information

Performing Peptide Bioanalysis Using High Resolution Mass Spectrometry with Target Enhancement MRM Acquisition

Performing Peptide Bioanalysis Using High Resolution Mass Spectrometry with Target Enhancement MRM Acquisition Performing Peptide Bioanalysis Using High Resolution Mass Spectrometry with Target Enhancement MRM Acquisition Yun Wang Alelyunas, Mark D. Wrona, and Nick Tomczyk Waters Corporation, Milford, MA, USA GOAL

More information

IDENTIFICATION OF ORGANOMETALLIC COMPOUNDS USING FIELD DESORPTION IONIZATION ON THE GCT

IDENTIFICATION OF ORGANOMETALLIC COMPOUNDS USING FIELD DESORPTION IONIZATION ON THE GCT IDETIFICATIO OF ORGAOMETALLIC COMPOUDS USIG FIELD DESORPTIO IOIZATIO O THE GCT David Douce 1, Michael Jackson 1, Robert Lewis 1, Peter Hancock 1, Martin Green 1 and Stuart Warriner 2 1 Waters Corporation,

More information

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps ION ANALYZERS MASS ANALYSER sample Vacuum pumps Mass analysers - separate the ions according to their mass-to-charge ratio MASS ANALYSER Separate the ions according to their mass-to-charge ratio in space

More information

Metabolic Phenotyping Using Atmospheric Pressure Gas Chromatography-MS

Metabolic Phenotyping Using Atmospheric Pressure Gas Chromatography-MS Metabolic Phenotyping Using tmospheric Pressure Gas hromatography-ms Vladimir Shulaev, 2 Ghaste Manoj, 2,3 Steven Lai, 1 arolina Salazar, 2 Nobuhiro Suzuki, 2 Janna rossley, 2 Feroza Kaneez oudhury, 2

More information

Types of Analyzers: Quadrupole: mass filter -part1

Types of Analyzers: Quadrupole: mass filter -part1 16 Types of Analyzers: Sector or double focusing: magnetic and electric Time-of-flight (TOF) Quadrupole (mass filter) Linear ion trap Quadrupole Ion Trap (3D trap) FTICR fourier transform ion cyclotron

More information

1. Prepare the MALDI sample plate by spotting an angiotensin standard and the test sample(s).

1. Prepare the MALDI sample plate by spotting an angiotensin standard and the test sample(s). Analysis of a Peptide Sequence from a Proteolytic Digest by MALDI-TOF Post-Source Decay (PSD) and Collision-Induced Dissociation (CID) Standard Operating Procedure Purpose: The following procedure may

More information

Introduction to the Q Trap LC/MS/MS System

Introduction to the Q Trap LC/MS/MS System www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap LC/MS/MS Introduction to the Q Trap LC/MS/MS System The Q Trap

More information

for the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York

for the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York Mass Spectrometry for the Novice John Greaves and John Roboz (^>, yc**' CRC Press J Taylor & Francis Group Boca Raton London New York CRC Press is an imprint of the Taylor & Francis Croup, an informa business

More information

WADA Technical Document TD2003IDCR

WADA Technical Document TD2003IDCR IDENTIFICATION CRITERIA FOR QUALITATIVE ASSAYS INCORPORATING CHROMATOGRAPHY AND MASS SPECTROMETRY The appropriate analytical characteristics must be documented for a particular assay. The Laboratory must

More information

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS) Liquid Chromatography Mass Spectrometry (LC/MS) Prelab Questions: Questions to be answered before doing the experiment. The answers are due at the beginning of each experiment without exception (the questions

More information

LC-HRMS: Challenges for Routine Implementation

LC-HRMS: Challenges for Routine Implementation Joint EURL pesticides meeting October 23 rd -25th, 2013 Almeria Spain LC-HRMS: Challenges for Routine Implementation Richard J. Fussell Food and Environment Research Agency (York, UK) www.fera.defra.gov.uk

More information

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching s by Removal of Isobaric Interferences Differential Mobility Used as a Tool to Address Selectivity Challenges Adrian M. Taylor

More information

LC-MS Based Metabolomics

LC-MS Based Metabolomics LC-MS Based Metabolomics Analysing the METABOLOME 1. Metabolite Extraction 2. Metabolite detection (with or without separation) 3. Data analysis Metabolite Detection GC-MS: Naturally volatile or made volatile

More information

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS Back to Basics Section D: Ion Optics CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS TABLE OF CONTENTS QuickGuide...413 Summary...415 Introduction...417 The physical basis of orthogonal TOF....... 419 Pulsedmainbeamsofions...421

More information

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery Metabolite Stability Screening and Hotspot Metabolite Identification by Combining High-Resolution, Accurate-Mass Nonselective and Selective Fragmentation Tim Stratton, Caroline Ding, Yingying Huang, Dan

More information

Stephen McDonald and Mark D. Wrona Waters Corporation, Milford, MA, USA INT RO DU C T ION. Batch-from-Structure Analysis WAT E R S SO LU T IONS

Stephen McDonald and Mark D. Wrona Waters Corporation, Milford, MA, USA INT RO DU C T ION. Batch-from-Structure Analysis WAT E R S SO LU T IONS Chemical Intelligence in UNIFI: Enhancing in-silico Decision Making to Provide Confident and Accurate Results for Metabolite Identification Experiments Stephen McDonald and Mark D. Wrona Waters Corporation,

More information

pmod An Advanced Protein Deconvolution Algorithm with Automated Peak Modeling for Charge Deconvolution of Mass Spectrometry Data

pmod An Advanced Protein Deconvolution Algorithm with Automated Peak Modeling for Charge Deconvolution of Mass Spectrometry Data pmod An Advanced Protein Deconvolution Algorithm with Automated Peak Modeling for Charge Deconvolution of Mass Spectrometry Data Application Note Authors Ning Tang, Xiaoling Wu, and Stephen Madden Agilent

More information

Routine MS Detection for USP Chromatographic Methods

Routine MS Detection for USP Chromatographic Methods Daniel S. Root, Thomas E. Wheat, and Patricia McConville Waters Corporation, Milford, MA, USA APPLICATION BENEFITS This approach enables mass spectral analysis of peaks directly from unmodified USP HPLC

More information

Stephen McDonald, Mark D. Wrona, Jeff Goshawk Waters Corporation, Milford, MA, USA INTRODUCTION

Stephen McDonald, Mark D. Wrona, Jeff Goshawk Waters Corporation, Milford, MA, USA INTRODUCTION Combined with a Deep Understanding of Complex Metabolic Routes to Increase Efficiency in the Metabolite Identification Search Stephen McDonald, Mark D. Wrona, Jeff Goshawk Waters Corporation, Milford,

More information

Profiling of Diferulates (Plant Cell Wall Cross- Linkers) Using Ultrahigh-performance Liquid. Chromatography-Tandem Mass Spectrometry

Profiling of Diferulates (Plant Cell Wall Cross- Linkers) Using Ultrahigh-performance Liquid. Chromatography-Tandem Mass Spectrometry Supporting Information for: Profiling of Diferulates (Plant Cell Wall Cross- Linkers) Using Ultrahigh-performance Liquid Chromatography-Tandem Mass Spectrometry Ramin Vismeh a,b, Fachuang Lu c,d, Shishir

More information

Ion-mobility mass spectrometry of polyoxometalate Keplerate clusters and their supramolecular assemblies

Ion-mobility mass spectrometry of polyoxometalate Keplerate clusters and their supramolecular assemblies SUPPLEMENTARY INFORMATION Table of Contents Summary 1. General materials and methods 2. High-resolution mass spectra 3. Ion-mobility mass spectra 4. Fragmentation of clusters in the TWIMS drift tube 5.

More information

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS) Mass Analyzers

More information

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) The Agilent

More information

Agilent G3212 GC-APCI Source

Agilent G3212 GC-APCI Source Agilent G3212 GC-APCI Source Quick Start Guide Where to find information 2 Getting Started 3 Step 1. Start the Data Acquisition program for the GC and the Q-TOF 3 Step 2. Prepare the GC and Q-TOF for data

More information

Accurate Mass Analysis of Hydraulic Fracturing Waters

Accurate Mass Analysis of Hydraulic Fracturing Waters Application Note Environmental Accurate Mass Analysis of Hydraulic Fracturing Waters Using the Kendrick mass defect with the Agilent LC/Q-TOF MS Authors E. Michael Thurman and Imma Ferrer Department of

More information

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION A purpose-built collection of the best-inclass components for forensic toxicology Whether you re challenged to present reliable and secure forensic sample

More information

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA Use of High Speed/High Resolution Size-Based Chromatographic Separation of Surfactants and Oligomeric Materials with Single Quadrupole Mass Spectrometry Michael O Leary, Jennifer Gough, Tanya Tollifson

More information

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput Yi Zhang 1, Zhiqi Hao 1, Markus Kellmann 2 and Andreas FR. Huhmer

More information

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap Pro LC/MS/MS Accurate, High-Throughput Protein Identification

More information

IRMS. perspective. Isotope Ratio Mass Spectrometry.

IRMS. perspective. Isotope Ratio Mass Spectrometry. IRMS perspective Isotope Ratio Mass Spectrometry www.nu-ins.com Perspective The Nu Instruments Perspective geometry forms the basis for two state of the art IRMS instruments. Designed for flexibility,

More information

Catalysis CAPABILITIES

Catalysis   CAPABILITIES Catalysis www.extrel.com CAPABILITIES Contents Extrel instruments have been recognized for their exceptional performance by the world s leading researchers for more than 50 years. Reliability and flexibility

More information

EXPRESSION MORE PEPTIDES. MORE PROTEINS. MORE REPRODUCIBILITY. MORE QUANTIFICATION.

EXPRESSION MORE PEPTIDES. MORE PROTEINS. MORE REPRODUCIBILITY. MORE QUANTIFICATION. A New Era in Protein Quantification [ systems BIOLOGY Solutions ] [ systems BIOLOGY Solutions ] EXPRESSION MORE PEPTIDES. MORE PROTEINS. MORE REPRODUCIBILITY. MORE QUANTIFICATION. The Waters Expression

More information

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF Application Note Donghui Yi and Christine Miller Jon D. Williams, GlaxoSmithKline Introduction Mass spectrometry (MS) has become a core technology

More information

Supporting Information. Nitroxide Mediated Polymerization of methacrylates at moderate temperature

Supporting Information. Nitroxide Mediated Polymerization of methacrylates at moderate temperature Supporting Information Nitroxide Mediated Polymerization of methacrylates at moderate temperature Christophe Detrembleur, Christine Jérôme, Julien De Winter, Pascal Gerbaux, Jean-Louis Clément, Yohann

More information

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS Mass Spectrometry Hyphenated Techniques GC-MS LC-MS and MS-MS Reasons for Using Chromatography with MS Mixture analysis by MS alone is difficult Fragmentation from ionization (EI or CI) Fragments from

More information

Rapid Distinction of Leucine and Isoleucine in Monoclonal Antibodies Using Nanoflow. LCMS n. Discovery Attribute Sciences

Rapid Distinction of Leucine and Isoleucine in Monoclonal Antibodies Using Nanoflow. LCMS n. Discovery Attribute Sciences Rapid Distinction of Leucine and Isoleucine in Monoclonal Antibodies Using Nanoflow LCMS n Dhanashri Bagal *, Eddie Kast, Ping Cao Discovery Attribute Sciences Amgen, South San Francisco, California, United

More information

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Q&A Transcript ---------------- Q - How do you generate multiple charges reproductively? A - If you use the same settings on the

More information

Analysis of Polar Metabolites using Mass Spectrometry

Analysis of Polar Metabolites using Mass Spectrometry Analysis of Polar Metabolites using Mass Spectrometry TransMed Course: Basics in Clinical Proteomics and Metabolomics. Oct 10-19, 2012 dd.mm.yyyy Vidya Velagapudi, Ph.D, Adjunct Professor Head of the Metabolomics

More information

Choosing the metabolomics platform

Choosing the metabolomics platform GBS 748 Choosing the metabolomics platform Stephen Barnes, PhD 4 7117; sbarnes@uab.edu So, I have my samples what s next? You ve collected your samples and you may have extracted them Protein precipitation

More information

Theory and Practical Consequences.

Theory and Practical Consequences. Theory and Practical Consequences. Number of hits depending on reachable mass accuracy 250 hits @ ±10 ppm 137 hits @ ±5 ppm 78 hits @ ±3 ppm 29 hits @ ±1 ppm 3 hits @ ±0.1 ppm 1 hit @ ± 0.05 ppm Intens.

More information

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction Testing for GHB in hair by GC tandem quadrupole MS Marie Bresson, Vincent Cirimele, Pascal Kintz, Marion Villain; Laboratoire Chemtox, Illkirch, France Timothy Jenkins, Waters Corporation, Manchester,

More information

EGASUS. 4D GCxGC TOFMS. Delivering the Right Results

EGASUS. 4D GCxGC TOFMS. Delivering the Right Results P EGAU 4D GCxGC TFM Delivering the Right Results LEC s P EGAU 4D GCxGC-TFM The Ultimate Analytical Instrument Pioneering GCxGC technology, combined with the Pegasus TFM and ChromaTF software provides today

More information

[ xevo G2 tof ] Your Quantitative and Qualitative Questions Answered. Exactly.

[ xevo G2 tof ] Your Quantitative and Qualitative Questions Answered. Exactly. [ xevo G2 tof ] Your Quantitative and Qualitative Questions Answered. Exactly. Xevo G2 Tof Quantitative and qualitative answers from a single analysis. For scientists who need to confidently identify and

More information

Tandem MS = MS / MS. ESI-MS give information on the mass of a molecule but none on the structure

Tandem MS = MS / MS. ESI-MS give information on the mass of a molecule but none on the structure Tandem MS = MS / MS ESI-MS give information on the mass of a molecule but none on the structure In tandem MS (MSMS) (pseudo-)molecular ions are selected in MS1 and fragmented by collision with gas. collision

More information

Conformational Characterization of Calmodulin by Hydrogen Deuterium Exchange Mass Spectrometry

Conformational Characterization of Calmodulin by Hydrogen Deuterium Exchange Mass Spectrometry Conformational Characterization of Calmodulin by Hydrogen Deuterium Exchange Mass Spectrometry Joomi Ahn, Martha Stapels, Michael Eggertson, Keith Fadgen, and Ying Qing Yu Waters Corporation, Milford,

More information

TANDEM MASS SPECTROSCOPY

TANDEM MASS SPECTROSCOPY TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER

More information

NEW TOOLS FOR FINDING AND IDENTIFYING METABOLITES IN A METABOLOMICS WORKFLOW

NEW TOOLS FOR FINDING AND IDENTIFYING METABOLITES IN A METABOLOMICS WORKFLOW NEW TOOLS FOR FINDING AND IDENTIFYING METABOLITES IN A METABOLOMICS WORKFLOW Julia E. Wingate 1 ; Elliott Jones 2 ; Armin Graber 3 ; Klaus Weinberger 3 1Applied Biosystems, Toronto, Canada; 2Applied Biosystems,

More information

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 DEPARTMENTS OF CHEMISTRY AND BIOCHEMISTRY GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 Mass Analysers Shabaz Mohammed October 20, 2015 High Vacuum System Turbo pumps Diffusion pumps Rough pumps Rotary

More information

Optimization of the sample preparation and extraction methodology

Optimization of the sample preparation and extraction methodology Supporting information Results and discussion Optimization of the sample preparation and extraction methodology Visual comparison of the slopes (curve in standard mixture vs MMC curve with identical concentration

More information

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk Previously: Resolution and other basics MALDI Electrospray 40 Lecture 2 Mass analysers Detectors

More information

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward Bruker Daltonics EASY-nLC Tailored HPLC for nano-lc-ms Proteomics think forward Nano-HPLC World-Class Performance with a Small Footprint Bruker Daltonics presents a nano-lc system, perfectly integrated

More information

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS Application Note EPA Method 535 EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS API 3200 LC/MS/MS System Overview Described here is the

More information

MASS SPECTROMETRY. Topics

MASS SPECTROMETRY. Topics MASS SPECTROMETRY MALDI-TOF AND ESI-MS Topics Principle of Mass Spectrometry MALDI-TOF Determination of Mw of Proteins Structural Information by MS: Primary Sequence of a Protein 1 A. Principles Ionization:

More information

Targeted protein quantification

Targeted protein quantification Targeted Quantitative Proteomics Targeted protein quantification with high-resolution, accurate-mass MS Highly selective Very sensitive Complex samples HR/AM A more complete quantitative proteomics picture

More information

The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications

The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications Doug Stevens Waters Corp. Introduction The range of GC/MS instrument types has expanded

More information

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS Dominic Roberts, 1 Sara Stead, 1 and Jennifer A. Burgess 2 1 Waters Corporation, Manchester, UK 2 Waters Corporation,

More information

Information Dependent Acquisition (IDA) 1

Information Dependent Acquisition (IDA) 1 Information Dependent Acquisition (IDA) Information Dependent Acquisition (IDA) enables on the fly acquisition of MS/MS spectra during a chromatographic run. Analyst Software IDA is optimized to generate

More information

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems Technical Overview Introduction All Ions MS/MS is a technique that is available for Agilent high resolution

More information

Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365

Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365 Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365 Effortless Performance The LCMS-9030 quadrupole time-of-flight (Q-TOF) mass spectrometer integrates the world s fastest

More information

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF Analysis of Biomarkers in Crude Oil Using the Agilent 7 GC/Q-TOF Application Note Petrochemical and Environmental Authors Frank David Research Institute for Chromatography, Kennedypark 6, B-85 Kortrijk,

More information

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs Application Note Author Harry Prest, Ph.D. Senior Chemist in GC-MS Agilent Technologies, Inc. 5301 Stevens Creek Blvd Santa Clara,

More information

[ XEVO G2-XS QTOF ] EXTRAORDINARY POSSIBILITIES FOR YOUR EVERYDAY ANALYSIS

[ XEVO G2-XS QTOF ] EXTRAORDINARY POSSIBILITIES FOR YOUR EVERYDAY ANALYSIS [ XEV G2-XS QTF ] EXTRARDIARY PSSIBILITIES FR YUR EVERYDAY AALYSIS EXTREME SESITIVITY, EXTRARDIARY SELECTIVITY, Whether you re a scientist who wants to truly understand your samples, or a lab manager with

More information

Untargeted Screening on the X500R QTOF

Untargeted Screening on the X500R QTOF Untargeted Screening on the X500R QTOF R. E. Haufler, Paul Winkler, Andrei Schreiber, SCIEX, Concord, ON Canada NEMC August 2016 RUO-MKT-11-4397 Outline X500R technology and performance X500R QTOF Scan

More information

Mass spectrometry gas phase transfer and instrumentation

Mass spectrometry gas phase transfer and instrumentation Objectives of the Lecture spectrometry gas phase transfer and instrumentation Matt Renfrow January 15, 2014 1. Make ions 2. Separate/Analyze 3. Detect ions 4. What is mass resolution and mass accuracy?

More information