Tandem MS = MS / MS. ESI-MS give information on the mass of a molecule but none on the structure
|
|
- Ruby Tyler
- 6 years ago
- Views:
Transcription
1 Tandem MS = MS / MS ESI-MS give information on the mass of a molecule but none on the structure In tandem MS (MSMS) (pseudo-)molecular ions are selected in MS1 and fragmented by collision with gas. collision induced decay CID electron transfer decay ETD (= ECD) The fragment ions are analyzed in a second MS.
2 Quadrupole ESI ion source Entrance optics Mass analyzer Q 0 Q 1 detector MS Separation of primary ions
3 ESI ion source Entrance optics Triple Quadrupole Collision cell Q 2 Q 0 Q 1 Q 3 detector MS ion transfer just a longer flight path, nothing gained Separation of primary ions MS / MS tandem MS Precursor ion selection Difference: Collision energy Separation of fragment ions
4 ESI ion source HPLC Entrance optics Target analysis by MS-MS Collision cell Q 2 Q 0 Q 1 Q 3 detector MS / MS Precursor ion selection In LC-MS: mass selected for compounds eluting in a certain time window Analysis of fragment ions
5 ESI ion source HPLC Entrance optics Target analysis by MS-MS Collision cell Q 2 Q 0 Q 1 Q 3 detector MS / MS Precursor ion selection - fixed in time window Example: dexamethazon m/z 393 Transitions Reaction monitoring SRM / MRM Detection of specific fragment ions Only m/z 147 Or: 121,147,237
6 Tandem MS = MS / MS Target analysis: mass of analyte and mass of fragments are known beforehand. MS1 and MS2 are preset on target masses maximum dwell time, maximum sensitivity In proteomics applicationsnothing is known. Precursor mass is determined by survey scan Precursor mass is selected by operator (off-line) or PC (on-line; data-dependent acquisiton ) according to abundance, charge state and additional information
7 Tandem MS = MS / MS Precursor Ion scan: Fragment masses indicate structural details e.g. 365 reveals glycopeptides Neutral loss scan: Loss of a certain mass by removal of chemical group, e.g. 18 by H 2 O Loss of 98 indicates phosphorylation Requirement for proteomics applications: Resolution of multiply charged isotope clusters, high accuracy of MSMS Q-TOF, ion trap, IT-ICR
8 Hybrid-instruments: Quadrupole-TOF (Q-TOF) entrance lenses MS1 quadrupole Collision cell octapole MS2 TOF rotary vacuum pumps rough pumps turbomolecular pumps for high vacuum inside instrument PC for control and data aquisition Server for databank searches N 2 -generator(and oil-free compressor) Argon (collision gas) Waters Synapt II: R in V-mode: R in W-mode: Bruker Maxis 4G: up to TOF as MS2 allows higher resolution, accuracy and upper mass limit.
9
10 Hybrid-instruments with Orbitrap analyzers Combination of ion trap and Orbitrap analyzer Newest option: Combination of quadrupole with Orbitrap analyzer
11 Applications of MS-MS Hybrid instruments or Trap: Exact mass analysis of unknown compounds over a wide mass. Typical application: peptide identification by MS-MS structural analysis of biochemicals > fast "scan" rate of TOF or Trap
12 Q-TOF in MS Mode entrance lenses MS1 octapole MS2 Primary ions are collected and sent to MS1
13 Q-TOF in MS Mode MS1 MS2 MS1 does not filter, all ions pass through
14 Q-TOF in MS Mode MS1 MS2 collision cell is inactiv (ions are slow) ions pass unaltered
15 Q-TOF in MS Mode MS1 MS2 TOF analyses primary ions
16 Q-TOF in MS/MS Mode entrance lenses MS1 octapole MS2 Primary ions are collected and sent to MS1
17 Q-TOF in MS/MS Mode entrance lenses MS1 Collision cell MS2 MS1 selects parent ionof a certain mass (m/z); Others cannot pass
18 Q-TOF in MS/MS Mode entrance lenses MS1 Collision cell MS2 Collision with gas atoms (e.g.ar) causes fragmentation of ions (collission induced dissocation = CID) Collision energy is controlled by kinetic energy of the analyte ions.
19 Q-TOF in MS/MS Mode entrance lenses MS1 Collision cell MS2 Daughter ions leave the collision cell
20 Q-TOF in MS/MS Mode entrance lenses MS1 Collision cell MS2 MS2 (TOF) analyses fragment ions
21 Proteomics work with ESI-MS/MS De novo sequencing of a peptide of mass from a wasp venom allergen vulgaris PLA MSMS vulgaris_pla_msms MaxEnt 3 68 [Ev4631,It50,En1] (0.040,200.00,0.060, ,2,Cmp) MSMS ES+ AVI Y M A E C IK KI C E A M Y I V A bmax ymax % b y3 y y5 2+ [MH 2] y M/z y y theoret. [MH] AVIYMAECLK + VIYMAECLK IYMAECLK + + YMAECLK + MAECLK + AECLK + ECLK + CLK + LK Doubly charged precursor singly charged fragments
22 100 % Proteomics work with ESI-MS/MS mass 100 % FC+H Y G A C W T T D I S MSMS ES+ 200 M exp : TOF MSMS ES mass I FCISIDTTWCAGYCYTR Doubly charged precursor singly charged fragments
23 Peptide fragmentation Major fragments derived from a peptide (protonated) y 3 y 2 y H + 2 H + 2 H R 1 R 2 R 3 R 4 H N C C N C C N C C N C COOH 2 H O H H O H H O H H a 1 b 1 a 2 b 2 a 3 b 3 Doubly charged precursor singly charged fragments
24 Fragmentation of a singly charged peptide y 6 y 5 y 4 y 2 y 1 NH 3 R 1 R 2 R 3 R 4 R 5 R 6 R H 2 N C C N C C N C C N C C N C C N C C N C COOH O H O H O O H O O H H H H H H H H H H y 3 NH 2 R 1 R 2 R 3 R 4 R 5 R 6 R H 2 N C C N C C N C C N C C N C C N C C N C COOH O H O H O O H O O H H H H H H H H H H H R 1 R 2 H 2 N C C N C C N C C H O H H O H neutral b-fragment R 3 O H H NH 2 R 4 R 5 R 6 R N C C N C C N C C N C COOH H O H H O H O H H H H y 4 -ion
25 Fragmentation of a doubly charged peptide NH 3 R 1 R 2 R 3 R 4 R 5 R 6 R H 3 N C C N C C N C C N C C N C C N C C N C COOH O H O H O O H O O H H H H H H H H H H NH 2 R 1 R 2 R 3 R 4 R 5 R 6 R H 3 N C C N C C N C C N C C N C C N C C N C COOH O H O H O O H O O H H H H H H H H H H H R 1 R 2 H 3 N C C N C C N C C H O H H O H Doubly charged precursor singly charged fragments b 3 -ion R 3 O H H NH 2 R 4 R 5 R 6 R N C C N C C N C C N C COOH H O H H O H O H H H H y 3 -ion peptide mass + H
26 Peptide fragmentation R 1 R 2 H 3 N C C N C C N C C H O H H O H R 3 R 1 R 2 R 3 O H 2 N C C N C C N C C O b 3 -ions peptide mass -17 O O H H H H H acylium ion R 1 R 2 R 3 H 2 N C C N C C N C O O H H H H H a 3 -ion peptide mass -45 R 3 H 2 N C H immonium ion reveal amino acids in peptide
27 Collision energy Collision induced or collison activated dissociation of parent ions (CID or CAD) Triple quads, ion traps, Q-Tofs and similar mass specs can onlyprovide low energy(ev range) fragmentations. can be modulated within certain range (adjusted to mass of peptide) Yields relatively simple fragment spectra. Disadvantage: Leu and Ile cannot be discriminated High energy CID in BE or TOF-TOF instruments.
28 Applications of nano-lc / MS-MS MS BSA 100fmol, Time= min (#339#341), 100%=159692arb MS/MS y5 y6 y9 PBC m/z , all MS b3 y2 y3 y4 b7 b8 y7 b9 y10 b11 b m/z all MS/MS Time [min] 100 fmol BSA injected on column
29 Protein identification by LC / MS-MS Archeal histon Result of the database search of silver stained protein gel spot. Archeal histon was unambiguously found in a Halobacterium salinarum (genome) database using MASCOT TM. Important software packages for protein identification: MASCOT, GPM, SEQUEST. and company derivatives e.g. MassLynx, Proteome Discoverer ProteinScape..
30 Data-dependant aquisition (DDA) At first, the machine works in the MS mode (survey mode) until mass is detected that is: - of sufficient intensity - not in exclude list (background, trypsin, keratin) - doubly or triply charged (-is in include list ) Then, machine switches into MS/MS mode to acquire CID spectrum of this compoundfor e.g. ca. 1 sec Then, this mass is locked for some time to prevent redundancy. Often, the survey mode detects more than one signal MSMS 1, MSMS 2, etc. before switched back to survey.
31 MS/MS specials I.) Dependancy of ion type and collision energy - the larger the more energy required - charges help fragmentation - careful choice of collision energy profile II.) DDA tends to overlook many peptides Solutions: - increase speed of instrument - optimize selection criteria - rerun sample with inclusion list and/or exclusion list Quality criterion: % of identifiable peptide spectra Notorious problem: hybrid spectra (precursor selection not rigorous)
32 MS/MS specials Precursor selection problem turned upside down: MS/MS ALL (in segments: SWATH)
33 MS/MS specials Searching for peptide fragment (and mass) in data banks e.g. Swissprot By MASCOT or SEQUEST and related tools yields list of hits with probability The score depends on: 1.) number of peptides fitting to a particular peptide one hit wonders usually disregarded 2.) number of fragments fitting to theoretical digest of peptide (also: those NOT fitting) 3.) size of peptide (the larger, the better) 4.) size of protein (the smaller the better) 5.) allowance of missed cleavage sites 6.) allowance of modifications the more the worse (search space!) 7.) size of databank (too small is bad!)
34 Sample inlet systems for ESI Syringe pump 5 to 50 µl / min System test Calibration Simple samples Liquid chromatography 50 to 1000 µl / min Most typical LC-MS applications (pharmaceutical, environmental, forensic etc.)
35 Sample inlet systems for ESI Nanospray tips 20 nl /min 40 mm For limited sample amounts in bioscience 1-2 µl of sample give 30 min of analysis time Nanoflow LC 100 to 1000 nl / min For demanding life science applications split
36 Sample inlet systems for ESI Column i.d. Flow rate Technique 4.0 mm 1.0 ml/min Conventional HPLC 2.0 mm 0.25 ml/min Small bore LC 1.0 mm 0,0625 ml/min Micro LC 300 µm 5.6 µl/min Capillary LC 180 µm 2.0 µl/min Capillary LC 75 µm 350 nl/min Nano LC
37 Sample inlet systems for ESI S/N= mm i.d. S/N = ml/min Signal to Noise ratio 75 µm i.d.
38 Sample inlet systems for ESI 4.6 mm i.d 1.0 mm i.d. 300 µm i.d. 75 µm i.d. UV 206 nm ESI has a similar, even stronger concentration dependance 2 pmol digested myoglobin (each column)
39 Chromatographiesäulen im Vergleich 4 mm 0.18 mm 75 µm
40 Nano-Elektrospray
41 Nano-Elektrospray Diameter 1 mm Diameter 0.32 mm Diameter mm Diameter 4 mm, Flow rate 1.6 ml / min Flow rates?
Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer
IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific LTQ Orbitrap Velos
More informationMass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS
Mass Spectrometry Hyphenated Techniques GC-MS LC-MS and MS-MS Reasons for Using Chromatography with MS Mixture analysis by MS alone is difficult Fragmentation from ionization (EI or CI) Fragments from
More informationTANDEM MASS SPECTROSCOPY
TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER
More informationThermo Finnigan LTQ. Specifications
IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Finnigan LTQ Specifications
More informationRapid Distinction of Leucine and Isoleucine in Monoclonal Antibodies Using Nanoflow. LCMS n. Discovery Attribute Sciences
Rapid Distinction of Leucine and Isoleucine in Monoclonal Antibodies Using Nanoflow LCMS n Dhanashri Bagal *, Eddie Kast, Ping Cao Discovery Attribute Sciences Amgen, South San Francisco, California, United
More informationFinnigan LCQ Advantage MAX
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Finnigan LCQ Advantage MAX The Finnigan LCQ Advantage MAX ion trap mass spectrometer
More informationfor the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York
Mass Spectrometry for the Novice John Greaves and John Roboz (^>, yc**' CRC Press J Taylor & Francis Group Boca Raton London New York CRC Press is an imprint of the Taylor & Francis Croup, an informa business
More informationComputational Methods for Mass Spectrometry Proteomics
Computational Methods for Mass Spectrometry Proteomics Eidhammer, Ingvar ISBN-13: 9780470512975 Table of Contents Preface. Acknowledgements. 1 Protein, Proteome, and Proteomics. 1.1 Primary goals for studying
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment LCQ Deca XP Plus Improved Ion Optics for Greater Sensitivity and Precision
More informationInformation Dependent Acquisition (IDA) 1
Information Dependent Acquisition (IDA) Information Dependent Acquisition (IDA) enables on the fly acquisition of MS/MS spectra during a chromatographic run. Analyst Software IDA is optimized to generate
More informationHigh-Throughput Protein Quantitation Using Multiple Reaction Monitoring
High-Throughput Protein Quantitation Using Multiple Reaction Monitoring Application Note Authors Ning Tang, Christine Miller, Joe Roark, Norton Kitagawa and Keith Waddell Agilent Technologies, Inc. Santa
More informationFundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics
Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass
More informationMass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University
Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk Previously: Resolution and other basics MALDI Electrospray 40 Lecture 2 Mass analysers Detectors
More informationProtein Quantitation II: Multiple Reaction Monitoring. Kelly Ruggles New York University
Protein Quantitation II: Multiple Reaction Monitoring Kelly Ruggles kelly@fenyolab.org New York University Traditional Affinity-based proteomics Use antibodies to quantify proteins Western Blot RPPA Immunohistochemistry
More informationMass spectrometry has been used a lot in biology since the late 1950 s. However it really came into play in the late 1980 s once methods were
Mass spectrometry has been used a lot in biology since the late 1950 s. However it really came into play in the late 1980 s once methods were developed to allow the analysis of large intact (bigger than
More informationQuantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry
Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry Jon Hao, Rong Ye, and Mason Tao Poochon Scientific, Frederick, Maryland 21701 Abstract Background:
More informationWorkflow concept. Data goes through the workflow. A Node contains an operation An edge represents data flow The results are brought together in tables
PROTEOME DISCOVERER Workflow concept Data goes through the workflow Spectra Peptides Quantitation A Node contains an operation An edge represents data flow The results are brought together in tables Protein
More informationIntroduction to the Q Trap LC/MS/MS System
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap LC/MS/MS Introduction to the Q Trap LC/MS/MS System The Q Trap
More informationAccurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap Pro LC/MS/MS Accurate, High-Throughput Protein Identification
More informationProtein Quantitation II: Multiple Reaction Monitoring. Kelly Ruggles New York University
Protein Quantitation II: Multiple Reaction Monitoring Kelly Ruggles kelly@fenyolab.org New York University Traditional Affinity-based proteomics Use antibodies to quantify proteins Western Blot Immunohistochemistry
More information(Refer Slide Time 00:09) (Refer Slide Time 00:13)
(Refer Slide Time 00:09) Mass Spectrometry Based Proteomics Professor Sanjeeva Srivastava Department of Biosciences and Bioengineering Indian Institute of Technology, Bombay Mod 02 Lecture Number 09 (Refer
More informationTypes of Analyzers: Quadrupole: mass filter -part1
16 Types of Analyzers: Sector or double focusing: magnetic and electric Time-of-flight (TOF) Quadrupole (mass filter) Linear ion trap Quadrupole Ion Trap (3D trap) FTICR fourier transform ion cyclotron
More informationTraditional Herbal Medicine Structural Elucidation using SYNAPT HDMS
Traditional Herbal Medicine Structural Elucidation using SYNAPT HDMS with Time-Aligned Parallel (TAP) Fragmentation Kate Yu, Jose Castro-Perez, and John Shockcor Waters Corporation, Milford, MA, U.S. INTRODUCTION
More informationChoosing the metabolomics platform
GBS 748 Choosing the metabolomics platform Stephen Barnes, PhD 4 7117; sbarnes@uab.edu So, I have my samples what s next? You ve collected your samples and you may have extracted them Protein precipitation
More informationTranslational Biomarker Core
Translational Biomarker Core Instrumentation Thermo Scientific TSQ Quantum Triple Quadrupole Mass Spectrometers. There are two TSQ Quantum Ultra AM instruments available in the TBC. The TSQ Quantum Ultra
More informationElectron Transfer Dissociation of N-linked Glycopeptides from a Recombinant mab Using SYNAPT G2-S HDMS
Electron Transfer Dissociation of N-linked Glycopeptides from a Recombinant mab Using SYNAPT G2-S HDMS Jonathan P. Williams, Jeffery M. Brown, Stephane Houel, Ying Qing Yu, and Weibin Chen Waters Corporation,
More informationMASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps
ION ANALYZERS MASS ANALYSER sample Vacuum pumps Mass analysers - separate the ions according to their mass-to-charge ratio MASS ANALYSER Separate the ions according to their mass-to-charge ratio in space
More informationSelf-assembling covalent organic frameworks functionalized. magnetic graphene hydrophilic biocomposite as an ultrasensitive
Electronic Supplementary Material (ESI) for Nanoscale. This journal is The Royal Society of Chemistry 2017 Electronic Supporting Information for: Self-assembling covalent organic frameworks functionalized
More informationMass Spectrometry in MCAL
Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring
More informationAnalysis of Polar Metabolites using Mass Spectrometry
Analysis of Polar Metabolites using Mass Spectrometry TransMed Course: Basics in Clinical Proteomics and Metabolomics. Oct 10-19, 2012 dd.mm.yyyy Vidya Velagapudi, Ph.D, Adjunct Professor Head of the Metabolomics
More informationMS-based proteomics to investigate proteins and their modifications
MS-based proteomics to investigate proteins and their modifications Francis Impens VIB Proteomics Core October th 217 Overview Mass spectrometry-based proteomics: general workflow Identification of protein
More informationWADA Technical Document TD2003IDCR
IDENTIFICATION CRITERIA FOR QUALITATIVE ASSAYS INCORPORATING CHROMATOGRAPHY AND MASS SPECTROMETRY The appropriate analytical characteristics must be documented for a particular assay. The Laboratory must
More informationLC-MS Based Metabolomics
LC-MS Based Metabolomics Analysing the METABOLOME 1. Metabolite Extraction 2. Metabolite detection (with or without separation) 3. Data analysis Metabolite Detection GC-MS: Naturally volatile or made volatile
More informationIdentification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate
Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate Michael D. Jones, Xiang Jin Song, Robert S. Plumb, Peter J. Lee, and
More informationSRM assay generation and data analysis in Skyline
in Skyline Preparation 1. Download the example data from www.srmcourse.ch/eupa.html (3 raw files, 1 csv file, 1 sptxt file). 2. The number formats of your computer have to be set to English (United States).
More informationGRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2
DEPARTMENTS OF CHEMISTRY AND BIOCHEMISTRY GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 Mass Analysers Shabaz Mohammed October 20, 2015 High Vacuum System Turbo pumps Diffusion pumps Rough pumps Rotary
More informationMass Spectrometry. Quantitative Mass Spectrometry Chiral Mass Spectrometry
Mass Spectrometry Quantitative Mass Spectrometry Chiral Mass Spectrometry Quantitation by MS Goal is to develop methodology to sensitively, specifically, accurately and rapidly measure one or more compounds
More informationABI 3200 Q TRAP LC/MS/MS System
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL 001.847.913.0777 ABI 3200 Q TRAP LC/MS/MS System The advantages of an ion trap and the performance of a triple quad. All in one system.
More informationBST 226 Statistical Methods for Bioinformatics David M. Rocke. January 22, 2014 BST 226 Statistical Methods for Bioinformatics 1
BST 226 Statistical Methods for Bioinformatics David M. Rocke January 22, 2014 BST 226 Statistical Methods for Bioinformatics 1 Mass Spectrometry Mass spectrometry (mass spec, MS) comprises a set of instrumental
More informationPowerful Scan Modes of QTRAP System Technology
Powerful Scan Modes of QTRAP System Technology Unique Hybrid Triple Quadrupole Linear Ion Trap Technology Provides Powerful Workflows to Answer Complex Questions with No Compromises While there are many
More informationLecture 8: Mass Spectrometry
intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?
More informationWADA Technical Document TD2015IDCR
MINIMUM CRITERIA FOR CHROMATOGRAPHIC-MASS SPECTROMETRIC CONFIRMATION OF THE IDENTITY OF ANALYTES FOR DOPING CONTROL PURPOSES. The ability of a method to identify an analyte is a function of the entire
More informationBruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward
Bruker Daltonics EASY-nLC Tailored HPLC for nano-lc-ms Proteomics think forward Nano-HPLC World-Class Performance with a Small Footprint Bruker Daltonics presents a nano-lc system, perfectly integrated
More informationPosterREPRINT COMPARISON OF PEAK PARKING VERSUS AUTOMATED FRACTION ANALYSIS OF A COMPLEX PROTEIN MIXTURE. Introduction
Introduction The study of protein expression allows a greater understanding of biological function and can now routinely be performed using mass spectrometry. However, analysis of the post-translational
More informationBioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)
Liquid Chromatography Mass Spectrometry (LC/MS) Prelab Questions: Questions to be answered before doing the experiment. The answers are due at the beginning of each experiment without exception (the questions
More informationInstrumentation development for coupling ion/ion reactions and ion mobility in biological mass spectrometry
Graduate Theses and Dissertations Iowa State University Capstones, Theses and Dissertations 2008 Instrumentation development for coupling ion/ion reactions and ion mobility in biological mass spectrometry
More informationPeptideProphet: Validation of Peptide Assignments to MS/MS Spectra. Andrew Keller
PeptideProphet: Validation of Peptide Assignments to MS/MS Spectra Andrew Keller Outline Need to validate peptide assignments to MS/MS spectra Statistical approach to validation Running PeptideProphet
More informationAtomic masses. Atomic masses of elements. Atomic masses of isotopes. Nominal and exact atomic masses. Example: CO, N 2 ja C 2 H 4
High-Resolution Mass spectrometry (HR-MS, HRAM-MS) (FT mass spectrometry) MS that enables identifying elemental compositions (empirical formulas) from accurate m/z data 9.05.2017 1 Atomic masses (atomic
More informationTutorial 1: Setting up your Skyline document
Tutorial 1: Setting up your Skyline document Caution! For using Skyline the number formats of your computer have to be set to English (United States). Open the Control Panel Clock, Language, and Region
More informationYun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION
Quantitation by High Resolution Mass Spectrometry: Using Target Enhancement and Tof-MRM to Achieve Femtogram-level On-column Sensitivity for Quantitation of Drugs in Human Plasma Yun W. Alelyunas, Mark
More informationProteomics. November 13, 2007
Proteomics November 13, 2007 Acknowledgement Slides presented here have been borrowed from presentations by : Dr. Mark A. Knepper (LKEM, NHLBI, NIH) Dr. Nathan Edwards (Center for Bioinformatics and Computational
More informationLecture 8: Mass Spectrometry
intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?
More informationAgilent ESI and APCI sources: for polar to non-polar compounds
1 Agilent 6400 Series Triple Quadrupole Users Workshop 1 Agilent ESI and APCI sources: for polar to non-polar compounds Nebulizer Pressure Corona current Nebulizer Pressure Vaporizer Vcap Vcap Drying Gas
More informationMethods for proteome analysis of obesity (Adipose tissue)
Methods for proteome analysis of obesity (Adipose tissue) I. Sample preparation and liquid chromatography-tandem mass spectrometric analysis Instruments, softwares, and materials AB SCIEX Triple TOF 5600
More informationInterazioni di ioni con elettroni (ECD, ETD) e fotoni (Ion spectroscopy) Gianluca Giorgi. via Aldo Moro Siena
Interazioni di ioni con elettroni (ECD, ETD) e fotoni (Ion spectroscopy) Gianluca Giorgi Università degli Studi di Siena Dipartimento di Biotecnologie, Chimica e Farmacia via Aldo Moro 53100 Siena e-mail:
More informationFast Protein and Peptide Separations Using Monolithic Nanocolumns and Capillary Columns
Application Note 3 Fast Protein and Peptide Separations Using Monolithic Nanocolumns and Capillary Columns INTRODUCTION Polymeric monolithic stationary phases offer an alternative to the classical microparticulate
More informationIdentification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry
Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry Jonathan Williams Waters Corporation, Milford, MA, USA A P P L I C AT
More informationMass Spectrometry Course
Mass Spectrometry Course Árpád Somogyi Mass Spectrometry Laboratory, Department of Chemistry and Biochemistry University of Arizona, Tucson, AZ Eötvös University, Budapest April 11-20, 2012 1 2 UA Chemistry
More informationImproved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 *
Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 * 1 Department of Chemistry, Pomona College, Claremont, California
More informationKey questions of proteomics. Bioinformatics 2. Proteomics. Foundation of proteomics. What proteins are there? Protein digestion
s s Key questions of proteomics What proteins are there? Bioinformatics 2 Lecture 2 roteomics How much is there of each of the proteins? - Absolute quantitation - Stoichiometry What (modification/splice)
More informationProtein analysis using mass spectrometry
Protein analysis using mass spectrometry Michael Stadlmeier 2017/12/18 Literature http://www.carellgroup.de/teaching/master 3 What is Proteomics? The proteome is: the entire set of proteins in a given
More informationThe Power of LC MALDI: Identification of Proteins by LC MALDI MS/MS Using the Applied Biosystems 4700 Proteomics Analyzer with TOF/TOF Optics
APPLICATION NOTE TOF MS The Power of LC MALDI: Identification of Proteins by LC MALDI MS/MS Using the Applied Biosystems 4700 Proteomics Analyzer with TOF/TOF Optics Purpose The Applied Biosystems 4700
More informationToxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.
1 Supporting Information 2 3 Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish. 4 5 John Moreman, Okhyun
More information1. Prepare the MALDI sample plate by spotting an angiotensin standard and the test sample(s).
Analysis of a Peptide Sequence from a Proteolytic Digest by MALDI-TOF Post-Source Decay (PSD) and Collision-Induced Dissociation (CID) Standard Operating Procedure Purpose: The following procedure may
More informationQuattro Micro - How does it work?
Quattro Micro - How does it work? 1 Introduction This document is designed to familiarise you with the principles behind how the Quattro Micro works. The level of this document is designed as Level One
More informationA Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance
A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance Michael D. Jones, Marian Twohig, Karen Haas, and Robert S. Plumb Waters Corporation,
More informationHR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput
HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput Yi Zhang 1, Zhiqi Hao 1, Markus Kellmann 2 and Andreas FR. Huhmer
More informationNew Approaches in the Analysis of Amadori Compounds
7 TH Wartburg Symposium on Flavor Chemistry and Biology Eisenach, April 21 st - April 23 rd, 2004 New Approaches in the Analysis of Amadori Compounds T. Davidek, I. Blank, K. Kraehenhuehl, J. Hau and S.
More informationHigh-Field Orbitrap Creating new possibilities
Thermo Scientific Orbitrap Elite Hybrid Mass Spectrometer High-Field Orbitrap Creating new possibilities Ultrahigh resolution Faster scanning Higher sensitivity Complementary fragmentation The highest
More informationFigure S1. Interaction of PcTS with αsyn. (a) 1 H- 15 N HSQC NMR spectra of 100 µm αsyn in the absence (0:1, black) and increasing equivalent
Figure S1. Interaction of PcTS with αsyn. (a) 1 H- 15 N HSQC NMR spectra of 100 µm αsyn in the absence (0:1, black) and increasing equivalent concentrations of PcTS (100 µm, blue; 500 µm, green; 1.5 mm,
More informationEffective Strategies for Improving Peptide Identification with Tandem Mass Spectrometry
Effective Strategies for Improving Peptide Identification with Tandem Mass Spectrometry by Xi Han A thesis presented to the University of Waterloo in fulfillment of the thesis requirement for the degree
More informationProfiling of Diferulates (Plant Cell Wall Cross- Linkers) Using Ultrahigh-performance Liquid. Chromatography-Tandem Mass Spectrometry
Supporting Information for: Profiling of Diferulates (Plant Cell Wall Cross- Linkers) Using Ultrahigh-performance Liquid Chromatography-Tandem Mass Spectrometry Ramin Vismeh a,b, Fachuang Lu c,d, Shishir
More informationTOMAHAQ Method Construction
TOMAHAQ Method Construction Triggered by offset mass accurate-mass high-resolution accurate quantitation (TOMAHAQ) can be performed in the standard method editor of the instrument, without modifications
More informationOverview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)
Quantitation and Identification of Pharmaceuticals and Personal Care Products (PPCP) in Environmental Samples using Advanced TripleTOF MS/MS Technology André Schreiber AB SCIEX Concord, Ontario (Canada)
More informationLECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile
LECTURE-11 Hybrid MS Configurations HANDOUT PREAMBLE As discussed in our previous lecture, mass spectrometry is by far the most versatile technique used in proteomics. We had also discussed some of the
More informationBruker Daltonics. microtof-q. Cutting Edge Performance with Sub-ppm Confidence. ESI-Qq TOF. think forward
Bruker Daltonics microtof-q Cutting Edge Performance with Sub-ppm Confidence think forward ESI-Qq TOF The advantage of ultimate confidence Do you need maximum confidence in your analytical system? Bruker
More informationKey Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery
Metabolite Stability Screening and Hotspot Metabolite Identification by Combining High-Resolution, Accurate-Mass Nonselective and Selective Fragmentation Tim Stratton, Caroline Ding, Yingying Huang, Dan
More informationChemistry Instrumental Analysis Lecture 37. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 37 Most analytes separated by HPLC are thermally stable and non-volatile (liquids) (unlike in GC) so not ionized easily by EI or CI techniques. MS must be at
More informationMass spectrometry gas phase transfer and instrumentation
Objectives of the Lecture spectrometry gas phase transfer and instrumentation Matt Renfrow January 15, 2014 1. Make ions 2. Separate/Analyze 3. Detect ions 4. What is mass resolution and mass accuracy?
More informationTUTORIAL EXERCISES WITH ANSWERS
TUTORIAL EXERCISES WITH ANSWERS Tutorial 1 Settings 1. What is the exact monoisotopic mass difference for peptides carrying a 13 C (and NO additional 15 N) labelled C-terminal lysine residue? a. 6.020129
More informationQuantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays
: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays Mark Wrona, Jayne Kirk, and Yun Alelyunas Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Ability
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Applied Biosystems QStar Pulsar i Features of the API QSTAR Pulsar i The
More informationAgilent 6400 Series Triple Quadrupole Users Workshop. MH QQQ Users workshop 2/21/2014 1
Agilent 6400 Series Triple Quadrupole Users Workshop 1 Agilent 6400 Series Triple Quadrupole Users Workshop QQQ Method Development and Optimization MassHunter Quantitation: Batch and Method setup QC s,
More information[ instrument specifications ]
Designed for leading researchers working at the limits of conventional mass spectrometry capabilities who need to further characterize and define their samples the Waters SYNAPT High Definition MS (HDMS
More informationProtein identification is made easy with Micromass optional BioLynx and ProteinLynx software packages.
SPECIFICATION NOTE 126 Q-Tof 2 Q-Tof 2 is a new type of mass spectrometry tool for peptide sequencing, and protein characterisation and metabolite identification, a powerful alternative to the conventional
More informationAgilent 6400 Series Triple Quadrupole LC/MS/MS Users Session
Agilent 6400 Series Triple Quadrupole LC/MS/MS Users Session QQQ Method Development and Optimization MassHunter Quant: Method setup Peak detection optimization Quant troubleshooting David Presser Application
More informationImproved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer
Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer Reiko Kiyonami, Mary Blackburn, Andreas FR Hühme: Thermo Fisher
More informationAll Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems
All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems Technical Overview Introduction All Ions MS/MS is a technique that is available for Agilent high resolution
More informationMSc Chemistry Analytical Sciences. Advances in Data Dependent and Data Independent Acquisition for data analysis in proteomic research
MSc Chemistry Analytical Sciences Literature Thesis Advances in Data Dependent and Data Independent Acquisition for data analysis in proteomic research by Florian L. R. Lucas 11198877 September 2016 12
More informationQualitative Proteomics (how to obtain high-confidence high-throughput protein identification!)
Qualitative Proteomics (how to obtain high-confidence high-throughput protein identification!) James A. Mobley, Ph.D. Director of Research in Urology Associate Director of Mass Spectrometry (contact: mobleyja@uab.edu)
More informationEffective desalting and concentration of in-gel digest samples with Vivapure C18 Micro spin columns prior to MALDI-TOF analysis.
Introduction The identification of proteins plays an important role in today s pharmaceutical and proteomics research. Commonly used methods for separating proteins from complex samples are 1D or 2D gels.
More informationMS-MS Analysis Programs
MS-MS Analysis Programs Basic Process Genome - Gives AA sequences of proteins Use this to predict spectra Compare data to prediction Determine degree of correctness Make assignment Did we see the protein?
More informationIntroduction to LC-MS
Wednesday April 5, 2017 10am Introduction to LC-MS Amy Patton, MS Laboratory Manager, Pinpoint Testing, LLC Little Rock, AR DESCRIPTION: Amy Patton, laboratory manager for Pinpoint Testing, will begin
More informationNew Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses
New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses Technical Overview Authors Abstract Peter Stone, Thomas Glauner, Frank Kuhlmann, Tim Schlabach and Ken Miller
More informationIsotopic-Labeling and Mass Spectrometry-Based Quantitative Proteomics
Isotopic-Labeling and Mass Spectrometry-Based Quantitative Proteomics Xiao-jun Li, Ph.D. Current address: Homestead Clinical Day 4 October 19, 2006 Protein Quantification LC-MS/MS Data XLink mzxml file
More informationDIA-Umpire: comprehensive computational framework for data independent acquisition proteomics
DIA-Umpire: comprehensive computational framework for data independent acquisition proteomics Chih-Chiang Tsou 1,2, Dmitry Avtonomov 2, Brett Larsen 3, Monika Tucholska 3, Hyungwon Choi 4 Anne-Claude Gingras
More informationIntroduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph
SCIETIFIC ISTRUMET EWS 2016 Vol. 7 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation Introduction of Hitachi Chromaster 5610 MS Detector for High Performance
More informationSimplified Approaches to Impurity Identification using Accurate Mass UPLC/MS
Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS Marian Twohig, Michael D. Jones, Dominic Moore, Peter Lee, and Robert Plumb Waters Corporation, Milford, MA, USA APPLICATION
More informationSpectrum-to-Spectrum Searching Using a. Proteome-wide Spectral Library
MCP Papers in Press. Published on April 30, 2011 as Manuscript M111.007666 Spectrum-to-Spectrum Searching Using a Proteome-wide Spectral Library Chia-Yu Yen, Stephane Houel, Natalie G. Ahn, and William
More informationMassHunter TOF/QTOF Users Meeting
MassHunter TOF/QTOF Users Meeting 1 Qualitative Analysis Workflows Workflows in Qualitative Analysis allow the user to only see and work with the areas and dialog boxes they need for their specific tasks
More information