Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

Size: px
Start display at page:

Download "Live Webinar : How to be more Successful with your ACQUITY QDa Detector?"

Transcription

1 Live Webinar : How to be more Successful with your ACQUITY QDa Detector? Q&A Transcript Q - How do you generate multiple charges reproductively? A - If you use the same settings on the ACQUITY QDa, generating multiple charged ions is only dependent on the eluent composition. It is therefore important to prepare your eluent in a reproducible way. Q - Is there a database of mass spectra in EMPOWER? A - Yes, there is a MS spectra library functionality within Empower. Q- Detector saturation: will we have a message when we have saturated the detector or do we have to check the spectral data? A - There is no specific message when the detector is saturated. Detector saturation does not always negatively impact your data however if the sample is very concentrated it may not be possible to differentiate between the "main" analyte peak and its isotopes. When this happens the mass assignment may appear to be inaccurate as the software has difficulty establishing where the apex of the spectral peak is. Q - How often should we run the start up with the calibration? A - Normally, you would not need to run the calibration check. The QDa is stable over a long period of time (1 year). If you want to be sure or are working in a regulated environment, every time. Q - Why does the additional pump on the performance model increase sensitivity? A - The performance ACQUITY QDa has besides the bigger pump, also a bigger orifice (the opening of the MS) which allows more ions enter the MS. As we let more ions into the MS we need a bigger pump to maintain the vacuum. As we let in more ions, the sensitivity is increased.

2 Q - Is there a way of knowing whether you have multiple charge species present? A - When you have single-charged ions, the first 13C isotope is 1 mass unit from the mono-isotopic mass (molecular ion). When you have double-charged species, the first 13C isotope is 1/2 mass unit from the mono-isotopic mass. Q - Does switching on the front button also start the pumps? A - Yes. Q - Can Empower do a 3D plot like it can with PDA data? A - Empower can present a 3D plot very much like the PDA data, although the peaks are very much narrower in the m/z dimension than they would be in the wavelength dimension. A contour plot is also produced and is often more useful. Q - Can you scan positive and negative at the same time? A - Yes.

3 Q - We have been using QDa for 2 years and we managed to handle all situations. A new one arose regarding the turbo pump and its maintenance. We contacted the local office and they told that each year Oil cartridge should be exchanged. Can you send me PM? A - Yes, we have informed the local service organisation who will contact you. Q - For probe temperature, what would you recommend increasing it to if you are running at higher flow rates (1 to 2ml/min)? A - The probe temperature should be set to 600 in the Advanced Mode of the instrument method editor. When operating at these high flow rates, especially with largely aqueous mobile phase, be sure that the vent drain is directed sloping down and is flowing freely. If liquid accumulates in the drain tube, it can trigger a vacuum error. Q - Is it possible to switch between LC and waste for not overwhelming the MS with uninteresting matrix? Our SQD can only switch one now? A - The QDa does not have an integral divert valve due to space restrictions. It may be possible to divert some flow via the column manager module depending on your configuration. Q - Does a combination Water/ACN without additive work? A - You can expect to see signal with a purely water/can mobile phase. Electrospray ionization, however, works better with a more conductive mobile phase. A low concentration of a volatile acid or salt will give a larger and more stable signal. Q - I want to increase the M H Signal and lower the fragmentation signals. How can I do this? A - The cone voltage setting has the largest influence on the amount of fragmentation. I would use a setting of 15 volts for my first injection, and then make a series of runs at a series of progressively lower cone voltage in steps of 2 volts. A somewhat higher range is typically used for negative ionization mode. Q - Can we perform low level quantifications? A - This is difficult to answer as the question is what are low levels. With the standard ACQUITY QDa, you are normally more sensitive than the quantification levels you can achieve with a PDA.

4 Q - What is the reason for the detection limit at 1250 Da? Is it a limitation in the software or instrument hardware? A - The m/z range upper limit is set at 1250 to ensure that isotopic resolution of all detectable analytes is achieved at the maximum scanning speed (10,000 Da/s) of the instrument. It's a question of consistently providing quality spectral data. Q - I am using AccQTag Ultra, with the solvents included in the kit. Can these solvents be used with QDa? A - Yes, the eluents provided for the AccQ-Tag Ultra Method is compatible with Electrospray Ionization. Choose a lower cone voltage to minimize fragmentation. Observation of a signal at is the symptom that cone voltage should be lower. Q - What is the difference between standard and performance QDa detector? A - The difference between standard and performance QDa is a bigger orifice (entrance of the MS) and a larger pump (rotary). This leads to a sensitivity increase (5-7). Please be aware that this is compound dependent. Q - How much more sensitivity is there between Standard and Performance? A - In general there is about a factor of 5-7 difference between standard and performance (performance being the more sensitive). Please be aware that this is compound dependent! Q - Do we expect (MSMS) similar to this QDa in future? Just by adding fragmentation cell and a second quadrupole? A - Nothing imminent is planned. Q - Should we leave the unit on even if there is no analysis? A - The ACQUITY QDa has been designed to turn it off when you do not need it. So if you are not using it for a longer period (over night/weekend) then we recommend to turn it off as after about 20 minutes after turning it on you can use it again. (We recommend to use the same off/on process as you would use with a PDA detector).

5 Q - With QDa, is it possible to quantify 2 compounds with the same retention time? A - Yes it is absolutely possible to quantify 2 (or more) compounds with the same retention time. With Single Ion Recording you define the m/z value you want to monitor over which retention time, and that can be the same retention time. The maximum amount of compounds you can quantify at the same retention time will depend on the mass range you selected. Q - How to differentiate between 3 and 4 ions? A - The difference between 3 and 4 ions can be calculated from the 13C isotopic pattern. When you have singly charged ions, the first 13C isotope is 1 mass unit from the mono-isotopic mass. (molecular ion). When you have doubly charged species, the first 13C isotope is 1/2 mass unit from the mono-isotopic mass. If you have triply charged species, the first 13C isotope is 1/3 mass unit from the mono-isotopic mass etc. Q - Is there a recommended solvents/additives mixture for UPC2? A - When using QDa with UPC2, a make-up flow is needed. A good starting point therefore is MeOH with 2% water. You can use the same additives as you normally use when doing UPLC-MS analysis, as has been discussed during the webinar, e.g ammonium formate, ammonium acetate etc. Q - What is the recommended probe temperature when using UPC2? A - The recommended probe temperature is 600 C or higher. Also, using UPC2, a higher gas flow might be useful to get better sensitivity. As this depends on your flow, exact numbers are difficult to give. Q - What is better? To keep the QDa on standby or to shut it down for different periods of time? (over night/weekend etc) A - This is up to user preference and how long you wish to wait. It takes approximately 6 minutes for the QDa to be in a ready state from OFF (20 minutes if you run a calibration check upon start-up). For optimal sensitivity (note, this is not always needed for qualitative work), it is recommended to wait 30 to 45 minutes. The benefit of switching off is obviously the reduction in your electricity bill and lab cooling requirements.

6 Q - By scanning in SIR do we see all the fragments on the desired product? A - The name SIR means single ion recording, this means you are monitoring 1 m/z. If you look at the SIR function with the m/z of interest you will not be able to see fragments. However, you can define a full scan acquisition function parallel to the SIR function(s) you have defined. In the full MS results you might be able to see fragments (depending on the compound, cone voltage etc). Q - You spoke about PDA and MS data. Can the delay between the two detectors be set up to align chromatograms? and where? A - Both the 3D and any extracted or SIR 2D channels can be offset for alignment in the processing method on the Smoothing/Offset tab. Q - How can I use isotopic peaks at large unknown molecules? A - As the ACQUITY QDa has unit resolution, it will be difficult to distinguish isotopic patterns at high charge states. If you need that, the better MS instrument to use would be the Waters high resolution XEVO G2-XS QTOF. Q - How can you identify large unknown molecules? Is there a kind of action plan? A - The question is a bit what do you mean by identify and what is large? As you might recall from the webinar, larger molecules can be identified as long as the m/z ratio is below The ACQUITY QDa is capable of identifying known large molecules, as you normally have standards available, so you know retention time and preferable charge state. If you want to identify unknown large molecules the better instrument to use is a high resolution instrument like the Waters XEVO G2-XS QTOF. Q - How can I be sure there are sulphur and/or silica atoms in the molecules? A - Sulphur has an isotopic pattern of 94 sulphur 32 and 4% sulphur 34. This pattern you can see in your isotopic distribution of your molecular ion ([M+H]+ as the second isotope would be higher than expected on basis of only 13C isotopes. The same applies for Silica, as there you have 3 isotopes with distribution: 28Si:29Si: 30Si = 100:5:3. Q - How can I know where a functional group is placed in a (large) molecule? A - With the ACQUITY QDa you might be able to see a fragment which corresponds to either a functional group of interest and/or the loss of the functional group. If you are interested in structural elucidation, a high resolution mass spectrometer like the WATERS XEVO G2-XS QTOF would be the better choice to use.

7 Q - What is the lowest cone voltage possible? A - The cone voltage can be set as low as 1volt. We have seen a very few analytes where that setting was optimal, but a higher setting gives better sensitivity with little fragmentation. Q - Why is the capillary voltage much lower than in the SQD2 (3kV instead of 0.8kV)? A - As with many of considerations around the source (probe position, gas flows) the capillary voltage has been defaulted and ranges set to values which will provide the best combination of sensitivity and robustness. On instruments such as the SQD2 it is very easy for new users to accidentally set capillary voltage parameters and probe positions which can result in short term sensitivity gains but poor robustness. The QDa has been designed to ensure new users and expert users can achieve the same level of instrument performance consistently Q - If I measure SIR of two masses, there is a different in response. If I measure this in one function together or in two functions separately. Why? A - It may not be possible to answer this question without seeing how the methods have been configured. It could be due to a difference in overlapping time regions and consequently a change in dwell time spent on the transition in the two examples you provided. If you would like further input you could send screenshots of your QDa methods. Q - We use as additive FA (ph=3). Is the ph too low and what is a better additive? A - Formic Acid at PH-3 is perfect for positive ionisation. For negative ionisation a higher PH is better, so the use of a buffer would be advised. Q - When the QDa is in standby what is the recommended flow? A - When the QDa is in standby it is not recommended to run any LC flow into source. If you are not acquiring data on the QDa for an extended period of time it is recommended to disconnect the QDa and have the PDA/TUV eluent go to waste.

8 Q - It is better as you said to use negative mode for acidic compounds (like carboxylic acid), but these compounds are often better eluted on C18 column with acidic additives (HCOOH for example) which so reduces the response, correct? What is the best solution? A - The best solution is to use a post-column addition of base to neutralize the acidic mobile phase. I typically use ammonium hydroxide. I make my post column reagent at 5 times the molar concentration of acid in the mobile phase. The flow rate of the post column pump is then 20% of the flow through the column so that I am mixing equimolar amounts of acid and base into the source. For negative ionization, I will usually use a post-column reagent that is 50% water and 50% isopropanol to further enhance the ionization. Q - How to get a more robust system when running a lot of different applications? A - The experience with the ACQUITY QDa is that the system is very robust. This is mainly due to the patented reversed cone sample aperture. We have customers who use the ACQUITY QDa in open access environment who do not see any robustness issues. Depending on the applications you are running, you might want to consider changing the sample aperture preventively.

Webinar: How to be More Successful with your ACQUITY QDa Detector Eric van Beelen

Webinar: How to be More Successful with your ACQUITY QDa Detector Eric van Beelen Webinar: How to be More Successful with your ACQUITY QDa Detector Eric van Beelen 2015 Waters Corporation 1 We are here to address any Questions you may have That s the benefit of participating to the

More information

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays : Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays Mark Wrona, Jayne Kirk, and Yun Alelyunas Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Ability

More information

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW Paula Hong, John Van Antwerp, and Patricia McConville Waters Corporation, Milford, MA, USA Historically UV detection has been

More information

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity Janet Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS High

More information

UPLC Intact MASS Analysis Application Kit

UPLC Intact MASS Analysis Application Kit (Part Number: 176001519) CONTENTS I. INTACT MASS ANALYSIS APPLICATION KIT COMPONENTS II. INSTALLING THE PRE-COLUMN TUBING INTO MASSPREP MICRO DESALTING COLUMN III. CONNECTING THE COLUMN TO LCT PREMIER

More information

Routine MS Detection for USP Chromatographic Methods

Routine MS Detection for USP Chromatographic Methods Daniel S. Root, Thomas E. Wheat, and Patricia McConville Waters Corporation, Milford, MA, USA APPLICATION BENEFITS This approach enables mass spectral analysis of peaks directly from unmodified USP HPLC

More information

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography Tips and Techniques to Optimize ELSD and MS based Purification Utilizing ELSD and MS as Secondary Detectors for Prep

More information

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS) Liquid Chromatography Mass Spectrometry (LC/MS) Prelab Questions: Questions to be answered before doing the experiment. The answers are due at the beginning of each experiment without exception (the questions

More information

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies APPLICATION BENEFITS Regulatory guidelines and recommendations place a greater emphasis on the detection

More information

Thermo Finnigan LTQ. Specifications

Thermo Finnigan LTQ. Specifications IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Finnigan LTQ Specifications

More information

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION Quantitation by High Resolution Mass Spectrometry: Using Target Enhancement and Tof-MRM to Achieve Femtogram-level On-column Sensitivity for Quantitation of Drugs in Human Plasma Yun W. Alelyunas, Mark

More information

Finnigan LCQ Advantage MAX

Finnigan LCQ Advantage MAX www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Finnigan LCQ Advantage MAX The Finnigan LCQ Advantage MAX ion trap mass spectrometer

More information

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA Use of High Speed/High Resolution Size-Based Chromatographic Separation of Surfactants and Oligomeric Materials with Single Quadrupole Mass Spectrometry Michael O Leary, Jennifer Gough, Tanya Tollifson

More information

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs Application Note Author Harry Prest, Ph.D. Senior Chemist in GC-MS Agilent Technologies, Inc. 5301 Stevens Creek Blvd Santa Clara,

More information

Analysis of Polar Metabolites using Mass Spectrometry

Analysis of Polar Metabolites using Mass Spectrometry Analysis of Polar Metabolites using Mass Spectrometry TransMed Course: Basics in Clinical Proteomics and Metabolomics. Oct 10-19, 2012 dd.mm.yyyy Vidya Velagapudi, Ph.D, Adjunct Professor Head of the Metabolomics

More information

LC-MS Based Metabolomics

LC-MS Based Metabolomics LC-MS Based Metabolomics Analysing the METABOLOME 1. Metabolite Extraction 2. Metabolite detection (with or without separation) 3. Data analysis Metabolite Detection GC-MS: Naturally volatile or made volatile

More information

BEST PRACTICES IN THE ANALYSIS OF RAPIFLUOR-MS LABELED GLYCANS USING THE ACQUITY QDa DETECTOR (PERFORMANCE MODEL)

BEST PRACTICES IN THE ANALYSIS OF RAPIFLUOR-MS LABELED GLYCANS USING THE ACQUITY QDa DETECTOR (PERFORMANCE MODEL) BEST PRACTICES IN THE ANALYSIS OF RAPIFLUOR-MS LABELED GLYCANS USING THE ACQUITY QDa DETECTOR (PERFORMANCE MODEL) The following document provides basic understanding and guidance in the configuration of

More information

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS Marian Twohig, Michael D. Jones, Dominic Moore, Peter Lee, and Robert Plumb Waters Corporation, Milford, MA, USA APPLICATION

More information

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research Heather A Brown, 1 Claudia Rossi, 2 and Lisa J Calton 1 1 Waters Corporation, Wilmslow, UK 2

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment LCQ Deca XP Plus Improved Ion Optics for Greater Sensitivity and Precision

More information

Chromatography and other Separation Methods

Chromatography and other Separation Methods Chromatography and other Separation Methods Probably the most powerful class of modern analytical methods for analyzing mixture of components---and even for detecting a single component in a complex mixture!

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Applied Biosystems QStar Pulsar i Features of the API QSTAR Pulsar i The

More information

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Chemistry Instrumental Analysis Lecture 37. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 37 Most analytes separated by HPLC are thermally stable and non-volatile (liquids) (unlike in GC) so not ionized easily by EI or CI techniques. MS must be at

More information

Supporting Information

Supporting Information 1 Supporting Information 2 3 Discovery and implications of C 2 and C 3 perfluoroalkyl sulfonates in aqueous film forming foams (AFFF) and groundwater 4 Krista A. Barzen-Hanson a and Jennifer A. Field b*

More information

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph SCIETIFIC ISTRUMET EWS 2016 Vol. 7 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation Introduction of Hitachi Chromaster 5610 MS Detector for High Performance

More information

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific LTQ Orbitrap Velos

More information

Multi-residue analysis of pesticides by GC-HRMS

Multi-residue analysis of pesticides by GC-HRMS An Executive Summary Multi-residue analysis of pesticides by GC-HRMS Dr. Hans Mol is senior scientist at RIKILT- Wageningen UR Introduction Regulatory authorities throughout the world set and enforce strict

More information

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD) The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD) Basically all compounds which are less volatile than the mobile phase can be detected. Detection is based on a Universal property of

More information

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany Analysis of Pesticides by Supercritical Fluid Chromatography/ Mass Spectrometry Optimizing the Agilent 126 Infi nity Analytical SFC System in Combination with the Agilent 646 Triple Quadrupole LC/MS Application

More information

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Application Note Clinical Research Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Authors Yanan Yang 1, Victor Mandragon 2, and Peter Stone 1 1

More information

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis Application Note AN89.V.1 Extraction of Methylmalonic Acid from Serum Using ISOLUTE SAX Page 1 Extraction of Methylmalonic Acid from Serum Using ISOLUTE SAX Prior to LC-MS/MS Analysis Sample Preparation

More information

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Application Note Authors Kevin McCann, Sameer Nene, Doug McIntyre, Edmond Neo, Dennis Nagtalon, and

More information

Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate Yu 3

Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate Yu 3 Using Natural Products Application Solution with UNIFI for the Comparison of the Chemical Ingredients of Shuanghuanglian Oral Drink from Two Different Manufacturers Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper,

More information

Potential sources of contamination on Mass Spectrometers and suggested cleaning procedures

Potential sources of contamination on Mass Spectrometers and suggested cleaning procedures Potential sources of contamination on Mass Spectrometers and suggested cleaning procedures Purpose: This document provides information on potential sources of background contamination commonly encountered

More information

Agilent ESI and APCI sources: for polar to non-polar compounds

Agilent ESI and APCI sources: for polar to non-polar compounds 1 Agilent 6400 Series Triple Quadrupole Users Workshop 1 Agilent ESI and APCI sources: for polar to non-polar compounds Nebulizer Pressure Corona current Nebulizer Pressure Vaporizer Vcap Vcap Drying Gas

More information

Agilent All Ions MS/MS

Agilent All Ions MS/MS Agilent All Ions MS/MS Workflow Overview A Determine fragment ions for LC/MS Quant method B Develop final Quant method Develop LC/MS Qualitative Analysis method Process data with Find by Formula Build

More information

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB. Quantitative analysis of small molecules in biological samples 100 Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB % 0 300 500 700 900 1100 1300 1500 1700 m/z Class Overview Introduction

More information

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS Application Note Forensic Toxicology Authors Martin Josefsson, and Markus Roman National Board of Forensic Medicine Linköping,

More information

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry Jon Hao, Rong Ye, and Mason Tao Poochon Scientific, Frederick, Maryland 21701 Abstract Background:

More information

INTRODUCTION OVERVIEW

INTRODUCTION OVERVIEW COMPARISON OF CCS(N 2 ) MEASUREMENTS OBTAINED FROM TWO DIFFERENT T-WAVE ION S WITH DIRECT MEASUREMENTS USING A DRIFT TUBE ION Kevin Giles, Martin Palmer, Keith Richardson, Nick Tomczyk Waters MS Technologies

More information

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC Online Reaction Monitoring of In- Manufacturing Samples by UPLC Tanya Tollifson Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Better throughput, yield, and process understanding are possible

More information

A Case of Pesticide Poisoning: The Use of a Broad-Scope Tof Screening Approach in Wildlife Protection

A Case of Pesticide Poisoning: The Use of a Broad-Scope Tof Screening Approach in Wildlife Protection A Case of Pesticide Poisoning: The Use of a Broad-Scope Tof Screening Approach in Wildlife Protection George Keenan 1, Anna Giela 1, Michael Taylor 1, Jennifer Watson 1, Eleanor Riches 2, James Morphet

More information

Introduction. Chapter 1. Learning Objectives

Introduction. Chapter 1. Learning Objectives Chapter 1 Introduction Learning Objectives To understand the need to interface liquid chromatography and mass spectrometry. To understand the requirements of an interface between liquid chromatography

More information

Q-TOF PREMIER DYNAMIC RANGE ENHANCEMENT AND ACCURATE MASS MEASUREMENT PERFORMANCE

Q-TOF PREMIER DYNAMIC RANGE ENHANCEMENT AND ACCURATE MASS MEASUREMENT PERFORMANCE Q-TOF PREMIER DYAMIC RAGE EHACEMET AD ACCURATE MASS MEASUREMET PERFORMACE Michael McCullagh, Jose Castro-Perez, and Lisa Calton Waters Corporation, Manchester, UK ITRODUCTIO The use of time-of-flight (TOF)

More information

Analysis of Organic Light Emitting Diode Materials by UltraPerformance Convergence Chromatography Coupled with Mass Spectrometry (UPC 2 /MS)

Analysis of Organic Light Emitting Diode Materials by UltraPerformance Convergence Chromatography Coupled with Mass Spectrometry (UPC 2 /MS) Analysis of Organic Light Emitting Diode Materials by UltraPerformance Convergence Chromatography Coupled with Mass Spectrometry (UPC 2 /MS) Michael D. Jones, Andrew Aubin, Peter J. Lee, and Tim Jenkins

More information

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract. Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS Application Food Analysis Authors Sheher Mohsin Agilent Technologies, Inc. Schaumburg, Il USA Michael Zumwalt

More information

Agilent G3212 GC-APCI Source

Agilent G3212 GC-APCI Source Agilent G3212 GC-APCI Source Quick Start Guide Where to find information 2 Getting Started 3 Step 1. Start the Data Acquisition program for the GC and the Q-TOF 3 Step 2. Prepare the GC and Q-TOF for data

More information

LC MS analysis of metabolites. Basis of Chromatography

LC MS analysis of metabolites. Basis of Chromatography GBS 724 LC MS analysis of metabolites Stephen Barnes, PhD 4 7117; sbarnes@uab.edu Basis of Chromatography A moving (mobile) phase passes over an inert, stationary phase The compounds differentially interact

More information

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance Michael D. Jones, Marian Twohig, Karen Haas, and Robert S. Plumb Waters Corporation,

More information

Traditional Herbal Medicine Structural Elucidation using SYNAPT HDMS

Traditional Herbal Medicine Structural Elucidation using SYNAPT HDMS Traditional Herbal Medicine Structural Elucidation using SYNAPT HDMS with Time-Aligned Parallel (TAP) Fragmentation Kate Yu, Jose Castro-Perez, and John Shockcor Waters Corporation, Milford, MA, U.S. INTRODUCTION

More information

Mass Spectrometry in MCAL

Mass Spectrometry in MCAL Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring

More information

Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC)

Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC) 1.0 Equipment: Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC) Thermo Exactive Orbitrap Mass Spectrometer Direct Analysis

More information

C HA R AC T E RIZ AT IO N O F INK J E T P RINT E R C A RT RIDG E INK S USING A CHEMOMETRIC APPROACH

C HA R AC T E RIZ AT IO N O F INK J E T P RINT E R C A RT RIDG E INK S USING A CHEMOMETRIC APPROACH C HA R AC T E RIZ AT I N F INK J E T P RINT E R C A RT RIDG E INK S USING A CHEMMETRIC APPRACH Diana Uría and Paul Silcock Waters Corporation, Manchester, UK AIM To evaluate the potential of UPLC -ToF-MS

More information

Agilent 6400 Series Triple Quadrupole Users Workshop. MH QQQ Users workshop 2/21/2014 1

Agilent 6400 Series Triple Quadrupole Users Workshop. MH QQQ Users workshop 2/21/2014 1 Agilent 6400 Series Triple Quadrupole Users Workshop 1 Agilent 6400 Series Triple Quadrupole Users Workshop QQQ Method Development and Optimization MassHunter Quantitation: Batch and Method setup QC s,

More information

for Acclaim Mixed-Mode HILIC-1 Column

for Acclaim Mixed-Mode HILIC-1 Column for Acclaim Mixed-Mode HILIC-1 Column Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 1 of 17 Product Manual for ACCLAIM Mixed-Mode HILIC-1 Column 5µm, 4.6 x 250mm, P/N 066844 5µm, 4.6 x 150mm, P/N

More information

MassHunter Software Overview

MassHunter Software Overview MassHunter Software Overview 1 Qualitative Analysis Workflows Workflows in Qualitative Analysis allow the user to only see and work with the areas and dialog boxes they need for their specific tasks A

More information

2010 Waters Corporation. Xevo TQ-S for Bioanalysis

2010 Waters Corporation. Xevo TQ-S for Bioanalysis Xevo TQ-S for Bioanalysis Key innovations High Performance Stepwave RADAR Scan Wave Versatility Ion Sources Simplicity of Operation Engineered Simplicity The Xevo TQ family Built on the same foundations

More information

Luminescence transitions. Fluorescence spectroscopy

Luminescence transitions. Fluorescence spectroscopy Luminescence transitions Fluorescence spectroscopy Advantages: High sensitivity (single molecule detection!) Measuring increment in signal against a dark (zero) background Emission is proportional to excitation

More information

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching s by Removal of Isobaric Interferences Differential Mobility Used as a Tool to Address Selectivity Challenges Adrian M. Taylor

More information

Varian Galaxie Chromatography Data System for Preparative HPLC

Varian Galaxie Chromatography Data System for Preparative HPLC Varian Galaxie Chromatography Data System for Preparative HPLC By Gary Burce Varian, Inc. 2700 Mitchell Drive, Walnut Creek, CA 95498 USA Abstract Galaxie is an ideal chromatography data system for the

More information

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS 1. List two advantages of temperature programming in GC. a) Allows separation of solutes with widely varying retention factors in a reasonable

More information

LC-HRMS: Challenges for Routine Implementation

LC-HRMS: Challenges for Routine Implementation Joint EURL pesticides meeting October 23 rd -25th, 2013 Almeria Spain LC-HRMS: Challenges for Routine Implementation Richard J. Fussell Food and Environment Research Agency (York, UK) www.fera.defra.gov.uk

More information

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd.

More information

Performing Peptide Bioanalysis Using High Resolution Mass Spectrometry with Target Enhancement MRM Acquisition

Performing Peptide Bioanalysis Using High Resolution Mass Spectrometry with Target Enhancement MRM Acquisition Performing Peptide Bioanalysis Using High Resolution Mass Spectrometry with Target Enhancement MRM Acquisition Yun Wang Alelyunas, Mark D. Wrona, and Nick Tomczyk Waters Corporation, Milford, MA, USA GOAL

More information

Analyst Software. Peptide and Protein Quantitation Tutorial

Analyst Software. Peptide and Protein Quantitation Tutorial This document is provided to customers who have purchased AB Sciex equipment to use in the operation of such AB Sciex equipment. This document is copyright protected and any reproduction of this document

More information

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis PO-CON1768E Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide ASMS 2017 TP-194 David Baker 1, Christopher Titman 1, Neil Loftus 1, Jonathan Horner

More information

Integrated approach to HPLC method development: Using all the tools in the chromatographer s toolbox

Integrated approach to HPLC method development: Using all the tools in the chromatographer s toolbox Integrated approach to HPLC method development: Using all the tools in the chromatographer s toolbox Rosario LoBrutto Pharmaceutical and Analytical Development Novartis Pharmaceuticals Overall Strategy

More information

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Application Note Doping Control Authors Prof. Maria Kristina Parr Freie Universität Berlin Institute

More information

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems Technical Overview Introduction All Ions MS/MS is a technique that is available for Agilent high resolution

More information

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation Technical Overview The Agilent InfinityLab D-LC Solution with Active Solvent Modulation Achieving Improved Resolution and Sensitivity for Challenging Combinations of Separation Conditions Author Sonja

More information

HPLC solvents and mobile phase additives

HPLC solvents and mobile phase additives HPLC solvents and mobile phase additives The following is intended to be a guide for the choice of solvent and mobile phase additives. Of course the choice of solvents for LC will be dictated primarily

More information

QTOF Training Manual

QTOF Training Manual 1236 Hach Hall 515-294-5544 www.cif.iastate.edu QTOF Training Manual (11/1/211 K.H.) Table of Contents Important Notes p.3 Overview of QTOF Hardware p.5 Overview of LC System p.5 Electrospray Ionization

More information

POST-COLUMN ADDITION AS A TOOL TO ENHANCE PERFORMANCE IN MICROFLOW LC-MS

POST-COLUMN ADDITION AS A TOOL TO ENHANCE PERFORMANCE IN MICROFLOW LC-MS POST-COLUMN ADDITION AS A TOOL TO ENHANCE PERFORMANCE IN MICROFLOW LC-MS Angela Doneanu, Michael Donegan and Jim Murphy The need for improving electrospray ionization (ESI-MS) sensitivity has been the

More information

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate Michael D. Jones, Xiang Jin Song, Robert S. Plumb, Peter J. Lee, and

More information

The Theory of HPLC. Quantitative and Qualitative HPLC

The Theory of HPLC. Quantitative and Qualitative HPLC The Theory of HPLC Quantitative and Qualitative HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases Agilent 129 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract The Agilent

More information

THE IMPLEMENTATION OF A SCREENING WORKFLOW FOR ION MOBILITY QUADRUOPOLE TIME-OF-FLIGHT MASS SPECTROMETRIC ANALYSIS OF PFOS ISOMERS

THE IMPLEMENTATION OF A SCREENING WORKFLOW FOR ION MOBILITY QUADRUOPOLE TIME-OF-FLIGHT MASS SPECTROMETRIC ANALYSIS OF PFOS ISOMERS THE IMPLEMENTATION O A SCREENING WORKLOW OR ION MOBILITY QUADRUOPOLE TIME-O-LIGHT MASS SPECTROMETRIC ANALYSIS O POS ISOMERS Ken Rosnack ken_rosnack@waters.com Ingrid Ericson Jogsten and Bert van Bavel

More information

Does Hydrophilic Polydopamine Coating Enhance Membrane Rejection of

Does Hydrophilic Polydopamine Coating Enhance Membrane Rejection of 1 Supporting Information 2 3 4 Does Hydrophilic Polydopamine Coating Enhance Membrane Rejection of Hydrophobic Endocrine Disrupting Compounds? 5 6 7 Hao Guo, Yu Deng, Zhijia Tao, Zhikan Yao, Jianqiang

More information

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer APPLICATION NOTE 72385 Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer Authors Stephan Meding, Katherine Lovejoy, Martin Ruehl Thermo Fisher Scientific,

More information

Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS Rapid and Simultaneous Analysis of Urinary Catecholamines and Metanephrines Using Mixed-Mode SPE and Hydrophilic Interaction Chromatography (HILIC) for Clinical Research Jonathan P. Danaceau, Erin E. Chambers,

More information

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions GAS CHROMATOGRAPHY MASS SPECTROMETRY Pre-Lab Questions Questions to be answered before doing the experiment. The answers are due at the beginning of each experiment without exception (the questions are

More information

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS Application Note: 56 Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS Yang Shi, Catherine Lafontaine, Matthew Berube, John Fink, François Espourteille Thermo

More information

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish. 1 Supporting Information 2 3 Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish. 4 5 John Moreman, Okhyun

More information

CRL MASS SPECTROMETRY FACILITY USER MANUAL LCT CLASSIC A & B

CRL MASS SPECTROMETRY FACILITY USER MANUAL LCT CLASSIC A & B Mass Spectrometry Instrument Training Guide Page 1 of 18 24/10/2012 CRL MASS SPECTROMETRY FACILITY USER MANUAL LCT CLASSIC A & B 1st Floor Mass Spec Lab: 00.097 This is a guide to using the LCT classic

More information

MassHunter TOF/QTOF Users Meeting

MassHunter TOF/QTOF Users Meeting MassHunter TOF/QTOF Users Meeting 1 Qualitative Analysis Workflows Workflows in Qualitative Analysis allow the user to only see and work with the areas and dialog boxes they need for their specific tasks

More information

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents Marco Karsten, Bas, Dolman, Giovanni Maio, Frank Steiner, Holger Franz, Frank Arnold and Remco Swart LC Packings,

More information

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 129 Infi nity LC, and 646 Triple Quadrupole LC/MS Application Note Clinical Research Authors Linda Côté and Christophe

More information

Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis

Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis Application Note AN772 Introduction This application note describes the extraction of cocaine and a full range

More information

Analyst Software. Automatic Optimization Tutorial

Analyst Software. Automatic Optimization Tutorial This document is provided to customers who have purchased AB Sciex equipment to use in the operation of such AB Sciex equipment. This document is copyright protected and any reproduction of this document

More information

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide What is the MassHunter METLIN Metabolite PCD/PCDL? 2 Where to find more information 2 Kit Content 3 Installation 5 Before you start 5 Install MassHunter

More information

M M e M M H M M H. Ion Sources

M M e M M H M M H. Ion Sources Ion Sources Overview of Various Ion Sources After introducing samples into a mass spectrometer, the next important step is the conversion of neutral molecules or compounds to gas phase ions. The ions could

More information

Pittcon P

Pittcon P PO-CON1701E 76% Increase in Throughput for Determination of Semi-Volatiles Using Narrow-Bore GC Columns and Rapid Data Acquisition with a Highly Sensitive Quadrupole GCMS System Pittcon 2017 1440-14P Brahm

More information

Choosing the metabolomics platform

Choosing the metabolomics platform GBS 748 Choosing the metabolomics platform Stephen Barnes, PhD 4 7117; sbarnes@uab.edu So, I have my samples what s next? You ve collected your samples and you may have extracted them Protein precipitation

More information

Determination of urea in ultrapure water by IC-MS/MS

Determination of urea in ultrapure water by IC-MS/MS APPLICATION NOTE 72482 Determination of urea in ultrapure water by IC-MS/MS Authors Soon Fatt Lee, 1 Fiona Teh Hui Boon, 1 Chris Cheah Hun Teong, 1 and Jeff Rohrer 2 ¹Thermo Fisher Scientific, Singapore

More information

Use of High Speed/High Resolution Size-Based Chromatographic Separation of Polymeric Mixtures with Offline Infrared Detection

Use of High Speed/High Resolution Size-Based Chromatographic Separation of Polymeric Mixtures with Offline Infrared Detection Use of High Speed/High Resolution Size-Based Chromatographic Separation of Polymeric Mixtures with Michael O Leary 1, Jennifer Gough 1, Damian Morrison 1, Alain Creissen 2 1 Waters Corporation, Milford,

More information

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada) Quantitation and Identification of Pharmaceuticals and Personal Care Products (PPCP) in Environmental Samples using Advanced TripleTOF MS/MS Technology André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

Information Dependent Acquisition (IDA) 1

Information Dependent Acquisition (IDA) 1 Information Dependent Acquisition (IDA) Information Dependent Acquisition (IDA) enables on the fly acquisition of MS/MS spectra during a chromatographic run. Analyst Software IDA is optimized to generate

More information

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone Mia Summers and Kenneth J. Fountain Waters Corporation, 34 Maple St.,

More information

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies Page 1 Agilent GC/MS Portfolio Smallest footprint, single GC column 5975VL with 6850GC Industry leading

More information

Powerful Scan Modes of QTRAP System Technology

Powerful Scan Modes of QTRAP System Technology Powerful Scan Modes of QTRAP System Technology Unique Hybrid Triple Quadrupole Linear Ion Trap Technology Provides Powerful Workflows to Answer Complex Questions with No Compromises While there are many

More information