Front End Electronics Internal Noise Source Characterization
|
|
- Barry Armstrong
- 6 years ago
- Views:
Transcription
1 Front End Electronics Internal Noise Source Characterization Joe Brendler and R.H. Tillman May 14, 2015 Contents 1 Introduction 2 2 System Model 2 3 Measurement Process G r Measurement S 2 Measurement S 1 Measurement Measurement Stability 5 5 Data Processing 6 6 Example Result 7 7 Appendix IntSourceCharFEERev3.m IntTimeCalc.m Virginia Tech. brendler@vt.edu Virginia Tech. hanktillman@vt.edu 1
2 1 Introduction The Front End Electronics (FEE) used in this interferometer 1 uses a 3-State switching scheme for calibration of system gain and noise temperature. The calibration procedure is described in [1]. The FEE uses an SM4 Noise Diode (SM4) as its internal noise source. For accurate calibration, the noise temperature of this source must be well known. This document provides a detailed description of the methodology used to measure the noise temperature of the SM4 Noise Diode on a given FEE. Section 2 presents the system model. Section 3 discusses the measurement process. Section 4 discusses measurement stability for a given integration time used. Section 5 explains the MATLAB script used to process the data. Section 6 presents the results of the measurements. 2 System Model The power measured by a spectrum analyzer is given by S = k(t DUT + T SA )B r (1) where k is Boltzmann s constant, T DUT is the noise temperature of the DUT, T SA is the noise temperature of the spectrum analyzer, and B r is the resolution bandwidth (spectrum analyzer setting). For this measurement, simply using the SM4 noise diode as the DUT does not work, because it s output is not strong enough to dominate over the noise floor of the spectrum analyzer, i.e. T SA >> T DUT. For this reason, gain must be added in front of the spectrum analyzer to make it more sensitive. This gain also adds additional noise to the measurement, which must be measured and removed to obtain the noise temperature of the SM4. For this reason this gain is referred to as the characterized receiver for this measurement. Figure 1 shows a block diagram of the system model for the test setup. Attenuators are used between each gain stage, because the following cascade analysis assumes that stages are impedance matched. The amplifiers used are not perfect devices, and therefore reflections arise between them. The attenuators mitigate the effects of this reflection by reducing the strength of the reflected signal, thus improving the impedance match. The noise diode is an even poorer match to 50Ω, so additional attenuation is used between it and the characterized receiver. With this model, the Power Spectral Density (PSD) measured by the spectrum analyzer is: S 1 = kg r (T ON + T r )B r (2) where k is Boltzmann s constant, gain Gr and noise temperatures T ON and T r are as shown in the block diagram, and B r is the resolution bandwidth. G r can be measured directly. T r can be measured by connecting a matched load to the input of the receiver chain (see Figure 2), measuring its PSD, and solving for T r given that S 2 = kg r (T 0 + T r )B r (3) where T 0 is the known ambient temperature. Figure 2 shows a block diagram of the test setup for the S 2 measurement. The 10 db attenuator and 50Ω resistor are assumed to be a matched load, thus the input temperature in this setup is the ambient temperature, T 0, which was measured to be approximately 296 K
3 Figure 1: Block diagram of the system model for the S 1 measurement Figure 2: Block diagram of the system model for the S 2 measurement 3 Measurement Process The high-level procedure for using the setup described in the previous section to obtain a measurement of T ON, the noise temperature of the SM4 Noise Diode, is as follows: ˆ Measure G r ˆ Measure S 2 and use (3) to solve for T r ˆ Measure S 1 and use (2) to solve for T ON The following sections describe in detail the process that was used to obtain the results presented in Section G r Measurement The G r measurement is done using the FSH3 spectrum analyzer in scalar tracking generator mode. The detector must be set to RMS (the default is Sample) for measurements to have uncertainty as low as desired for this work. The characterized receiver includes everything from the cable connecting the SM4 noise diode to the 20 db attenuator to the cable connecting the ZJL-3G to the FSH3 as shown in Figure 1. Therefore, these cables must be included in the G r gain measurement. First, set the span of the FSH3 to 100 MHz with a 50 MHz center frequency. Set the Resolution Bandwidth to 1 MHz and the Sweep Time to 1 second. Set the FSH3 to tracking generator by selecting Meas, Tracking Generator. Before proceeding, assure that the detector is set 3
4 to RMS. Then, calibrate by selecting Transmission Cal. The display will say to replace the DUT with a through connection. To calibrate the spectrum analyzer, connect a cable from the RF Generator Output directly to the RF Input with an appropriate adapter and select Continue. This will calibrate the FSH3 to remove the loss of the cable. Disconnect the cable from the RF Input and connect it through a 15 db attenuator to the input of the characterized receiver (the cable before the 20 db attenuator). The 15 db attenuator is to assure that the amplifiers operate in their linear region. Connect the output of the ZJL-3G to the FSH3 RF Input with another cable. Power on the amplifiers with appropriate DC supply levels. The gain should be approximately a constant 4.5 db above 10 MHz. Linearity can be confirmed by adding an additional attenuator and observing that the gain decreases by the loss of the attenuator. In order to obtain measurements with extremely low levels of uncertainty, which is desirable for this work, multiple sweeps must be averaged. The FSH View software package 2 is used to pull multiple sweeps from the FSH3 and save them on a computer as.csv files. Connect the spectrum analyzer to a computer running FSH View and select the Multiple Transfers button in the toolbar. Select a folder to save the files to, and select a filename. This information will be needed later for the MATLAB scripts that process the data. For the G r measurement, 30 sweeps were found to be sufficient. Going beyond that surpasses the limit of measurement stability explained in Section 4. Set 1 second between sweeps to assure that each file contains a completely independent sweep. The loss of the 15 db attenuator used to assure linearity must be measured and removed from the G r measurement. This is done by re-calibrating the FSH3 to remove the loss of both the cable attached to the RF Generator Out port AND the RF Input port, so that the attenuator can be inserted and measured directly. Use the same FSH View multiple file procedure to obtain 30 sweeps of the attenuator gain, which should be approximately -15 db across the whole spectrum. 3.2 S 2 Measurement The setup for the S 2 measurement is shown in Figure 2. The 50Ω resistor is implemented with a coaxial load, which can be found in the box of adapters. The same DC power supply settings as used in Section 3.1 must be used for the amplifiers. For both this measurement and S 1, the FSH3 must be set in Analyzer mode. The reference level should be set to -20 dbm with the same Span, Center Frequency, Resolution Bandwidth, and Sweep Time as in Section 3.1. When the characterized receiver is powered on, the noise floor of the FSH3 should visibly increase by an order of magnitude. This is the measurement to be recorded as S 2. The same procedure as in Section 3.1 should be used to obtain the data files from FSH View, except this time many more files are required. The signal levels for the power measurements S 2 and S 1 are significantly lower than that of the gain measurement, and therefore the signal has more uncertainty due to noise. The results shown later in this document used 4000 files, providing uncertainty much less than 1%. This many files requires about 2 and a half hours of recording. 3.3 S 1 Measurement The setup for the S 1 measurement is shown in Figure 1. The FEE must be biased in State 3 (requires a supply voltage > 20 V DC ) so that the noise diode is turned on. First, the characterized 2 Available at 4
5 Figure 3: Lab Bench Test Setup receiver should be powered on. The noise floor should jump as before. Then, the FEE should be powered on and biased to State 3. The signal on the FSH3 should then jump another order of magnitude from the addition of the noise from the SM4 noise diode. The rest of the data collection procedure is the same as with the S 2 measurement. For the S 1 measurement, a temperature data logger should be used to measure the ambient temperature next to the noise diode on the FEE. This device can be configured using OnSet s HoboWare software 3. Figure 3 shows a picture of the test setup on the lab bench with the characterized receiver connected to the spectrum analyzer as in the S 1 and S 2 measurements. The input to the receiver is shown unconnected. This would be connected to the SM4 Noise Diode for S 1 measurements, and to the matched load for S 2 measurements. 4 Measurement Stability This section explains the importance of selecting an appropriate number of files to save for averaging for each measurement. For the final T ON measurement, it is desirable to obtain much less than 1% uncertainty (± 0.5%) between the measurement and a polynomial fit. That requires a certain integration time for each measurement used to compute T ON, such that each individual measurement s own uncertainty is much less than 1%. In Analyzer Mode using the RMS Detector, the FSH3 measures the PSD of a signal at 301 points across the spectrum. The integration time per point per sweep is therefore given by Sweep Time/301 seconds. The variance between a measurement and its polynomial fit decreases as integration time increases. There is a limit to that improvement in variance due to digitizing hardware. The integration time used should come before that limit so that measurements are limited by noise in the analog system and not the digitizer. The spectrum analyzer was set up to continuously create data files to be averaged, each containing a one second sweep, or 1/301 s of integration. This method was used to measure S 2. Figure 4 shows a plot of the variance of the measurement vs. the 3 Available at 5
6 Figure 4: Plot of variance vs. integration time for the S 2 measurement. integration time. The variance is shown to continue decreasing as integration time increases, meaning the limit has not been reached. For this reason, longer measurements could be still be taken to reduce error if deemed necessary. 5 Data Processing The MATLAB script, IntSourceCharFeeRev3.m (included in the Appendix in Section 7) can be used to produce plots of all the measurements made in this process along with polynomial fits and percent error bars to show the uncertainty of the measurements. It has a few constants that must be set each time the script is used. The resolution bandwidth variables BrS2 and BrS1 must be set to the resolution bandwidths used for the S 2 and S 1 measurements, respectively. Based on the directions given above, these should be set to 1e6 for 1 MHz RBW. If the RBW ever changes for future measurements, these variables must be changed accordingly. The variable polyfitorder, which sets the order of the polynomial fit to the curves is currently set to 27, but should be changed if the measurements appear to require a different order polynomial (discussed further at the end of this section). Under the Parse Data cell, the same directory and filename as used in the FSH View multiple file procedure should be set for each measurement 4. These measurements, in order, are S 2, S 1, G r, and the measured 15 db attenuator. The following is a summary of what the IntSourceCharFeeRev3.m script does ˆ Each measurement is parsed so that the columns (power or gain) are placed into an array where rows correspond to frequency points and each column is one 1 second sweep. 4 The files will actually have longer names than written under filename that include timestamps, but this script only requires that the given filename be used. 6
7 Figure 5: Plot of the S 2 measurement ˆ The span is truncated to MHz, the frequency range of interest for this work. All of the data is recorded in db scale, so it is converted to linear units to work with the equations presented above. ˆ All of the columns are averaged into one column that contains the mean of all sweeps. ˆ The attenuator gain is removed from the G r measurement, and T r and T ON are computed using (3) and (2). ˆ A polynomial is fit to each measurement. ˆ Plots of each measurement with the polynomial fit are produced, showing both the actual measurement and the percent error for the measurement. The variance vs. integration time plot shown in Figure 4 was obtained using the IntTime- Calc.m MATLAB script (included in the Appendix in Section 7). The same procedure is used to give the script the relevant filename and directory as with the IntSourceCharFeeRev3.m script, as well as setting the order of the polynomial fit. This script should be used for each measurement to show that the measurement has not been limited by the digitizing hardware of the FSH3. It also shows a plot of the measurement with its polynomial fit, which is useful to assure that the order of polynomial chosen is appropriate. 6 Example Result Figures 5 through 11 show example results for one run of the procedure described in this document. 7
8 Figure 6: Plot of the S 1 measurement Figure 7: Plot of the G r measurement without the attenuation removed. 8
9 Figure 8: Plot of the measured attenuation used to assure the characterized receiver operated linearly during the G r measurement. Figure 9: Plot of the G r measurement with the attenuation removed. 9
10 Figure 10: Plot of the receiver noise temperature, T r. Figure 11: Plot of the SM4 Noise Diode temperature, T ON. 10
11 References [1] R. Tillman, Design and Evaluation of a HELA-10 Based FEE with 3-State Switched Calibration, Virginia Tech, Available: Tech. Rep., April 2013, [Online]. 7 Appendix 7.1 IntSourceCharFEERev3.m Following this page is the IntSourceCharFEERev3.m MATLAB script described in this document. 11
12 % Filename : IntSourceCharFEErev3.m % Author : Joe Brendler % Date Created : February 4, 2015 % Description : Produces a plot of a set of measurements as well as the % + or - 1% uncertainty limits for the average of the set. Used to confirm % individual measurements are within 1% uncertainty of the average of the % dataset. Used to characterize the internal noise source for FEE rev 3.0 % % ======================================================================= % close all; clear all; warning('off','matlab:axes:negativedatainlogaxis'); figindex = 1; BrS2 = 1e6; BrS1 = 1e6; k = *(10^(-23)); To = 296; polyfitorder = 27; % Resolution Bandwidth [Hz] for S2 measurement % Resolution Bandwidth [Hz] for S1 measurement % Boltzmann's Constant % Ambient Temperature [K] % Order of the polynomial fit to the result %% Parse Data directory = 'C:\Users\Joe\Documents\MATLAB\FSHMultiSweeps\150422_S2_NewRx'; % S2 file directory filenames2 = 'S2'; [freq, DATAS2, filelists2] = FSHMultiFiles(filenameS2, directory, 34, 334); % S2 files directory = 'C:\Users\Joe\Documents\MATLAB\FSHMultiSweeps\150422_S1_NewRx'; % S1 file directory filenames1 = 'S1'; [freq, DATAS1, filelists1] = FSHMultiFiles(filenameS1, directory, 34, 334); % S1 files directory = 'C:\Users\Joe\Documents\MATLAB\FSHMultiSweeps\150507_Gr_NewRx'; % Gr file directory filenamegr = 'Gr'; [freq, DATAGr, filelistgr] = FSHMultiFiles(filenameGr, directory, 33, 333); % Gr files directory = 'C:\Users\Joe\Documents\MATLAB\FSHMultiSweeps\150403_attn'; % attn file directory filenameattn = 'attn'; [freq, DATAattn, filelistattn] = FSHMultiFiles(filenameattn, directory, 33, 333); % attn files freq = freq(91:241,1); freq = freq.*10^-6; % Frequency axis in MHz from MHz %% Parse ARX Data %% Convert to Linear Units 12
13 DATAS2 = 10.^(DATAS2./10)*.001; DATAS1 = 10.^(DATAS1./10)*.001; DATAGr = 10.^(DATAGr./10); DATAattn = 10.^(DATAattn./10); % From dbm % From db %% Compute Mean S2 Curve MeanS2 = mean(datas2(91:241,1:(length(datas2(1,:)))),2); files % Mean of all S2 %% Compute Mean S1 Curve MeanS1 = mean(datas1(91:241,1:(length(datas1(1,:)))),2); files % Mean of all S1 %% Compute Mean Gr Curve MeanGr = mean(datagr(91:241,1:(length(datagr(1,:)))),2); files % Mean of all Gr %% Compute Mean attn Curve Meanattn = mean(dataattn(91:241,1:(length(dataattn(1,:)))),2); all attn files % Mean of %% Remove attenuation from Gr Gr_attn_removed = MeanGr./Meanattn; %% Noise Temp Computation % Tr Tr = (MeanS2./(k.*Gr_attn_removed.*BrS2)) - To; % Ton Ton = (MeanS1./(k.*Gr_attn_removed.*BrS1)) - Tr; %% Polynomial Fit for Ton pton = polyfit(freq, Ton, polyfitorder); Ton fit_ton = polyval(pton, freq); varhighton = fit_ton + abs(0.005.*fit_ton); varlowton = fit_ton - abs(0.005.*fit_ton); % Polynomial coefficients for % Polynomial fit to Ton % High 1% uncertainty limit % Low 1% uncertainty limit %% Polynomial Fit for Tr ptr = polyfit(freq, Tr, polyfitorder); fit_tr = polyval(ptr, freq); varhightr = fit_tr + abs(0.005.*fit_tr); % Polynomial coefficients for Tr % Polynomial fit to Tr % High 1% uncertainty limit 13
14 varlowtr = fit_tr - abs(0.005.*fit_tr); % Low 1% uncertainty limit %% Polynomial Fit for Gr (including attn) pgr = polyfit(freq, MeanGr, polyfitorder); fit_gr = polyval(pgr, freq); varhighgr = fit_gr + abs(0.005.*fit_gr); varlowgr = fit_gr - abs(0.005.*fit_gr); % Polynomial fit to Gr % High 1% uncertainty limit % Low 1% uncertainty limit %% Polynomial Fit for Gr (NOT including attn) pgr_attn_removed = polyfit(freq, Gr_attn_removed, polyfitorder); % Polynomial coefficients for Gr fit_attn_removed = polyval(pgr_attn_removed, freq); % Polynomial fit to Gr varhighgr_zfl = fit_attn_removed + abs(0.005.*fit_attn_removed); 1% uncertainty limit varlowgr_zfl = fit_attn_removed - abs(0.005.*fit_attn_removed); 1% uncertainty limit % High % Low %% Polynomial Fit for attn pattn = polyfit(freq, Meanattn, polyfitorder); for Gr fit_attn = polyval(pattn, freq); varhighattn = fit_attn + abs(0.005.*fit_attn); limit varlowattn = fit_attn - abs(0.005.*fit_attn); limit % Polynomial coefficients % Polynomial fit to Gr % High 1% uncertainty % Low 1% uncertainty %% Polynomial Fit for S1 ps1 = polyfit(freq, MeanS1, polyfitorder); Gr fit_s1 = polyval(ps1, freq); varhighs1 = fit_s1 + abs(0.005.*fit_s1); varlows1 = fit_s1 - abs(0.005.*fit_s1); % Polynomial coefficients for % Polynomial fit to Gr % High 1% uncertainty limit % Low 1% uncertainty limit %% Polynomial Fit for S2 ps2 = polyfit(freq, MeanS2, polyfitorder); Gr fit_s2 = polyval(ps2, freq); varhighs2 = fit_s2 + abs(0.005.*fit_s2); varlows2 = fit_s2 - abs(0.005.*fit_s2); % Polynomial coefficients for % Polynomial fit to Gr % High 1% uncertainty limit % Low 1% uncertainty limit %% PSD Plot S1 14
15 figure(figindex); subplot(2,1,1); h1_s1 = plot(freq,varhighs1,'--r',freq,varlows1,'--r',freq,means1,'-b'); % Plots the + or - 1% uncertainty limits h2_s1 = plot(freq, fit_s1, '--g'); ylabel('s1 (W/RBW)'); legend([h1_s1(3) h2_s1],{'mean', 'Fit'}); pererrhigh(1:length(freq)) = 0.5; pererrlow(1:length(freq)) = -0.5; subplot(2,1,2); plot(freq, 100.*(MeanS1 - fit_s1)./fit_s1, freq, pererrhigh, '--r', freq, pererrlow, '--r'); ylabel('percent Error (%)'); figindex = figindex + 1; %% PSD Plot S2 figure(figindex); subplot(2,1,1); h1_s2 = plot(freq,varhighs2,'--r',freq,varlows2,'--r',freq,means2,'-b'); % Plots the + or - 1% uncertainty limits h2_s2 = plot(freq, fit_s2, '--g'); ylabel('s2 (W/RBW)'); legend([h1_s2(3) h2_s2],{'mean', 'Fit'}); pererrhigh(1:length(freq)) = 0.5; pererrlow(1:length(freq)) = -0.5; subplot(2,1,2); plot(freq, 100.*(MeanS2 - fit_s2)./fit_s2, freq, pererrhigh, '--r', freq, pererrlow, '--r'); ylabel('percent Error (%)'); figindex = figindex + 1; %% Receiver Gain Plot Gr (including attn) figure(figindex); subplot(2,1,1); h1_gr = plot(freq,varhighgr,'--r',freq,varlowgr,'--r',freq,meangr,'-b'); % Plots the + or - 1% uncertainty limits 15
16 h2_gr = plot(freq, fit_gr, '--g'); ylabel('receiver Gain Gr'); legend([h1_gr(3) h2_gr],{'mean', 'Fit'}); pererrhigh(1:length(freq)) = 0.5; pererrlow(1:length(freq)) = -0.5; subplot(2,1,2); plot(freq, 100.*(MeanGr - fit_gr)./fit_gr); ylabel('percent Error (%)'); figindex = figindex + 1; %% Receiver Gain Plot Gr (NOT including attn) figure(figindex); subplot(2,1,1); h1_gr = plot(freq,varhighgr_zfl,'--r',freq,varlowgr_zfl,'-- r',freq,gr_attn_removed,'-b'); % Plots the + or - 1% uncertainty limits h2_gr = plot(freq, fit_attn_removed, '--g'); ylabel('gr (attenuation removed)'); legend([h1_gr(3) h2_gr],{'mean', 'Fit'}); pererrhigh(1:length(freq)) = 0.5; pererrlow(1:length(freq)) = -0.5; subplot(2,1,2); plot(freq, 100.*(Gr_attn_removed - fit_attn_removed)./fit_attn_removed); ylabel('percent Error (%)'); figindex = figindex + 1; %% Attenuator Plot for Attn used to measure Gr figure(figindex); subplot(2,1,1); h1_attn = plot(freq,varhighattn,'--r',freq,varlowattn,'--r',freq,meanattn,'- b'); % Plots the + or - 1% uncertainty limits h2_attn = plot(freq, fit_attn, '--g'); ylabel('attenuator Gain'); legend([h1_attn(3) h2_attn],{'mean', 'Fit'}); 16
17 pererrhigh(1:length(freq)) = 0.5; pererrlow(1:length(freq)) = -0.5; subplot(2,1,2); plot(freq, 100.*(Meanattn - fit_attn)./fit_attn); ylabel('percent Error (%)'); figindex = figindex + 1; %% Noise Temp Plot Tr figure(figindex); subplot(2,1,1); h1_tr = plot(freq,varhightr,'--r',freq,varlowtr,'--r',freq,tr,'-b'); % Plots the + or - 1% uncertainty limits h2_tr = plot(freq, fit_tr, '--g'); ylabel('receiver Noise Temperature (K)'); legend([h1_tr(3) h2_tr],{'t_r', 'Fit'}); pererrhigh(1:length(freq)) = 0.5; pererrlow(1:length(freq)) = -0.5; subplot(2,1,2); plot(freq, 100.*(Tr - fit_tr)./fit_tr, freq, pererrhigh, '--r', freq, pererrlow, '--r'); ylabel('percent Error (%)'); figindex = figindex + 1; %% Noise Temp Plot Ton figure(figindex); semilogy(freq, Ton, freq, fit_ton); ylabel('noise Temperature (K)'); legend('t_o_n', 'Fit'); figindex = figindex + 1; %% Polyfit Plot Ton figure(figindex); subplot(2,1,1); 17
18 h1_ton = plot(freq,varhighton,'--r',freq,varlowton,'--r',freq,ton,'-b'); % Plots the + or - 1% uncertainty limits h2_ton = plot(freq, fit_ton, '--g'); ylabel('noise Diode Noise Temperature (K)'); legend([h1_ton(3) h2_ton],{'t_o_n', 'Fit'}); pererrhigh(1:length(freq)) = 0.5; pererrlow(1:length(freq)) = -0.5; subplot(2,1,2); plot(freq, 100.*(Ton - fit_ton)./fit_ton, freq, pererrhigh, '--r', freq, pererrlow, '--r'); ylabel('percent Error (%)'); figindex = figindex + 1; 18
19 7.2 IntTimeCalc.m Following this page is the IntTimeCalc.m MATLAB script described in this document. 19
20 % Filename : IntTimeCalc.m % Author : Joe Brendler % Date Created : January 29, 2014 % Description : Produces a plot of mean-squared error vs. integration % time for spectrum analyzer measurements. Can be used to determine the % point at which further increasing the integration time no longer improves % the variance in a measurement. % % ======================================================================= % close all; clear all; warning('off','matlab:axes:negativedatainlogaxis'); polyfitorder = 27; % Order of the polynomial fit to the result %% Parse Data directory = 'C:\Users\Joe\Documents\MATLAB\FSHMultiSweeps\150422_S2_NewRX'; % Files' Directory filename = 'S2'; [freq, DATA, filelist] = FSHMultiFiles(filename, directory, 34, 334); % Parse data from the sweeps freq = freq(91:241,1); freq = freq.*10^-6; % Frequency axis in MHz from MHz AVG = zeros(length(freq), length(filelist)); matrix holding averages from MHz % preallocate array size for DATA = 10.^(DATA./10); %% Polynomial fit to total average TotalMean = mean(data(91:241,1:(length(data(1,:)))),2); % Mean of all files p = polyfit(freq, TotalMean, polyfitorder); % Polynomial coefficients for MHz fit = polyval(p, freq); % Polynomial fit to the curve %% Compute Averages numaverages = 1; % Start by averaging each file together to make a point % on the plot. Then once the next decade is reached, % average 10 files, 100 for the next decade, and so % on, to evenly distribute points throughout the plot. % Generates a matrix where each column is the average of "i" (number of columns) data % files. count = 1; for i = 1:(length(DATA(1,:))) 20
21 AVG(:,i) = mean(data(91:241,1:i),2); % Returns one column that contains the average % for MHz of i columns % Compute variance and integration time if mod(i,numaverages) == 0 % Compute Variance Var(count) = sum((avg(:,i) - fit).^2)/(i); % Variance b/w curve & fit % Var(count) = var(avg(:,i) - fit); % Using MATLAB's variance function % Compute Integration Time Tau(count) = i*( *10^(-3)); % 3.32 ms/pixel per second, 1s sweep per file % If the current value of tau is in the next decade, multiply the % spacing between points by 10 (to get evenly distributed points) if count == 1 currentdecade = log10(tau(count)); else if log10(tau(count)) >= (currentdecade+1) numaverages = numaverages*10; currentdecade = currentdecade + 1; end end end end count = count + 1; %% Plotting loglog(tau, Var, '-x'); xlabel('integration Time (s)'); ylabel('variance'); %% Polyfit Test % plot of polyfit test. figure(2); plot(freq, TotalMean, freq, fit); ylabel('power [arb db]'); legend('measurement', 'Polyfit'); 21
Quantifying and Verifying trace stability of vector network analyzers at frequencies below 1GHz
1 Quantifying and Verifying trace stability of vector network analyzers at frequencies below 1GHz Hamdi Mani, Chris Groppi, Judd Bowman Abstract we describe measurements done to check the low frequency
More informationTemperature measurement
Luleå University of Technology Johan Carlson Last revision: July 22, 2009 Measurement Technology and Uncertainty Analysis - E7021E Lab 3 Temperature measurement Introduction In this lab you are given a
More informationPhysical Noise Sources
AppendixA Physical Noise Sources Contents A.1 Physical Noise Sources................ A-2 A.1.1 Thermal Noise................ A-3 A.1.2 Nyquist s Formula.............. A-5 A.1.3 Shot Noise..................
More information2.004 Dynamics and Control II Spring 2008
MIT OpenCourseWare http://ocw.mit.edu 2.004 Dynamics and Control II Spring 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms. Massachusetts Institute
More informationOld Dominion University Physics 112N/227N/232N Lab Manual, 13 th Edition
RC Circuits Experiment PH06_Todd OBJECTIVE To investigate how the voltage across a capacitor varies as it charges. To find the capacitive time constant. EQUIPMENT NEEDED Computer: Personal Computer with
More informationECE 1270 Pseudo-Lab Report: Design of an LED Driver
ECE 1270 Pseudo-Lab Report: Design of an LED Driver N. E. Cotter ECE 1270 Abstract-This report presents measurements of a diode and models based on linear regression and downhill simplex search to find
More informationMeasurement of S-Parameters. Transfer of the Reference Plane. Power Waves. Graphic Representation of Waves in Circuits
Lecture 6 RF Amplifier Design Johan Wernehag Electrical and Information Technology Lecture 6 Amplifier Design Toughest week in the course, hang S-Parameters in there Definitions Power Waves Applications
More informationLab manual version 0
Electrical fluctuations of a resistor in thermal equilibrium: measurement of Boltzmann s constant, and the Centigrade value of Absolute zero. Lab manual version 0 Theory of Johnson, or Nyquist noise: Please
More informationCALYS 80P. Field pressure calibrator with high accuracy 0.01%
Field pressure calibrator with high accuracy 0.01% CALYS 80P is a field pressure calibrator within CALYS range with high accuracy of 0.01%. Description CALYS 80P is a field pressure calibrator within CALYS
More informationAdvancements in mm-wave On-Wafer Measurements: A Commercial Multi-Line TRL Calibration Author: Leonard Hayden Presenter: Gavin Fisher
Advancements in mm-wave On-Wafer Measurements: A Commercial Multi-Line TRL Calibration Author: Leonard Hayden Presenter: Gavin Fisher The title of this section is A Commercial Multi-Line TRL Calibration
More informationTemperature Coefficients for DC Resistance of Match and Reference Attenuators
Temperature Coefficients for DC Resistance of Match and Reference Attenuators Raul Monsalve SESE, Arizona State University October 2, 2013 Description This report presents the measurements of DC resistance
More informationKeysight Technologies Measurement Uncertainty of VNA Based TDR/TDT Measurement. Application Note
Keysight Technologies Measurement Uncertainty of VNA Based TDR/TDT Measurement Application Note Table of Contents Introduction... 3 S-parameter measurement uncertainty of VNA... 4 Simplification of systematic
More informationRotary Motion Servo Plant: SRV02. Rotary Experiment #01: Modeling. SRV02 Modeling using QuaRC. Student Manual
Rotary Motion Servo Plant: SRV02 Rotary Experiment #01: Modeling SRV02 Modeling using QuaRC Student Manual SRV02 Modeling Laboratory Student Manual Table of Contents 1. INTRODUCTION...1 2. PREREQUISITES...1
More informationTime Varying Circuit Analysis
MAS.836 Sensor Systems for Interactive Environments th Distributed: Tuesday February 16, 2010 Due: Tuesday February 23, 2010 Problem Set # 2 Time Varying Circuit Analysis The purpose of this problem set
More informationPerformance of Antenna Measurement Setup
Performance of Antenna Measurement Setup Raul Monsalve SESE, Arizona State University November 27, 213 2 Description This report is divided into two parts: PART 1: Check of recovered instrumental S-parameters
More informationENSC387: Introduction to Electromechanical Sensors and Actuators LAB 3: USING STRAIN GAUGES TO FIND POISSON S RATIO AND YOUNG S MODULUS
ENSC387: Introduction to Electromechanical Sensors and Actuators LAB 3: USING STRAIN GAUGES TO FIND POISSON S RATIO AND YOUNG S MODULUS 1 Introduction... 3 2 Objective... 3 3 Supplies... 3 4 Theory...
More informationLab 3: Quanser Hardware and Proportional Control
Lab 3: Quanser Hardware and Proportional Control The worst wheel of the cart makes the most noise. Benjamin Franklin 1 Objectives The goal of this lab is to: 1. familiarize you with Quanser s QuaRC tools
More informationBackground The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law
Phys316 Exploration 2: Verifying Stefan-Boltzmann Relationship Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law Where A is the effective radiating area,
More informationFATEK. FBs-1HLC. Precision Load Cell Module. Operation manual V1.1 06/07/2017 FATEK AUTOMATION CORP.
FATEK FBs-1HLC Precision Load Cell Module Operation manual V1.1 06/07/2017 FATEK AUTOMATION CORP. 1 Contents Chapter 1 1HLC module introduction... 3 1.1 Module specification... 3 1.2 Module appearance
More informationHomework Assignment 11
Homework Assignment Question State and then explain in 2 3 sentences, the advantage of switched capacitor filters compared to continuous-time active filters. (3 points) Continuous time filters use resistors
More informationESE319 Introduction to Microelectronics Bode Plot Review High Frequency BJT Model
Bode Plot Review High Frequency BJT Model 1 Logarithmic Frequency Response Plots (Bode Plots) Generic form of frequency response rational polynomial, where we substitute jω for s: H s=k sm a m 1 s m 1
More informationCapacitor ESR Measurement with Bode 100 and B-WIC
Page 1 of 9 Capacitor ESR Measurement with Bode 100 and B-WIC by Florian Hämmerle 2010 Omicron Lab V1.0 Visit www.omicron-lab.com for more information. Contact support@omicron-lab.com for technical support.
More informationCALYS 120 IS. ATEX documenting multifunction calibrator with high accuracy 0.006%
ATEX documenting multifunction calibrator with high accuracy 0.006% is the top instrument of our range of ATEX documenting multifunction calibrators. Description is the top instrument of our range of ATEX
More informationLecture 4, Noise. Noise and distortion
Lecture 4, Noise Noise and distortion What did we do last time? Operational amplifiers Circuit-level aspects Simulation aspects Some terminology Some practical concerns Limited current Limited bandwidth
More informationWide Area Monitoring System Enhancement
Wide Area onitoring System Enhancement Wide Area onitoring System Enhancement Prepared for Entergy Transmission Principal Authors Steven. Rovnyak, Guang Li, Kejun ei, Louisiana Tech University September
More informationUNIVERSITY OF CALIFORNIA College of Engineering Department of Electrical Engineering and Computer Sciences
UNIVERSITY OF CALIFORNIA College of Engineering Department of Electrical Engineering and Computer Sciences E. Alon Final EECS 240 Monday, May 19, 2008 SPRING 2008 You should write your results on the exam
More informationCore Technology Group Application Note 3 AN-3
Measuring Capacitor Impedance and ESR. John F. Iannuzzi Introduction In power system design, capacitors are used extensively for improving noise rejection, lowering power system impedance and power supply
More informationFor the electronic measurement of current: DC, AC, pulsed..., with galvanic separation between the primary and the secondary circuit.
Current Transducer LDSR 0.3-TP/SP1 I P R N = 300 ma For the electronic measurement of current: DC, AC, pulsed..., with galvanic separation between the primary and the secondary circuit. Features Closed
More informationImpedance Spectroscopy Fortgeschrittenen Praktikum I / II
VP Unibas Impedance Spectroscopy Fortgeschrittenen Praktikum I / II VP Unibas August 4, 204 Abstract This experiment examines the impedance of a single RC-circuit and two RC-circuits in series. Dierent
More informationDepartment of Mechanical and Aerospace Engineering. MAE334 - Introduction to Instrumentation and Computers. Final Examination.
Name: Number: Department of Mechanical and Aerospace Engineering MAE334 - Introduction to Instrumentation and Computers Final Examination December 12, 2003 Closed Book and Notes 1. Be sure to fill in your
More informationDigital Control Semester Project
Digital Control Semester Project Part I: Transform-Based Design 1 Introduction For this project you will be designing a digital controller for a system which consists of a DC motor driving a shaft with
More informationBE 3600 BIOMEDICAL INSTRUMENTATION (LAB) - WEEK 2
BE 3600 BIOMEDICAL INSTRUMENTATION (LAB) - WEEK 2 Principles of Biosensors OBJECTIVE: Learn the various modalities for construction of sensors utilizing electrical and optical measurement techniques. EXPERIMENT
More informationSPARQ S-Parameter Measurements with Impulse Response Time Limiting
SPARQ S-Parameter Measurements with Impulse Response Time Limiting Dr. Alan Blankman, Product Manager Summary Beginning with version 6.5.0.2 of the SPARQ software, users have the option of limiting the
More informationDevelopment of a Pulse Shape Discrimination IC
Development of a Pulse Shape Discrimination IC Michael Hall Southern Illinois University Edwardsville VLSI Design Research Laboratory October 20, 2006 Design Team Southern Illinois University Edwardsville:
More informationS-Parameter Calibration of Two-Port Setup: Gavin Fisher Cascade Microtech
-Parameter Calibration of Two-Port etup: How to choose the optimal calibration method? Gavin Fisher Cascade Microtech Content Error Modeling of a two-port setup Calibration methods OLT elf-calibration
More informationDouble Inverted Pendulum (DBIP)
Linear Motion Servo Plant: IP01_2 Linear Experiment #15: LQR Control Double Inverted Pendulum (DBIP) All of Quanser s systems have an inherent open architecture design. It should be noted that the following
More informationError Correction in Vector Network Analyzers
Error Correction in Vector Network Analyzers Thomas C. Baier, DG8SAQ May 19, 2009 Abstract This article describes systematic errors encountered in vector network analysis and how they can be mathematically
More informationPHOENIX CONTACT - 04/2016. Features
Signal conditioner Data sheet 100238_de_06 1 Description PHOENIX CONTACT - 04/2016 Features The MCR-C-UI-UI(-450)-DCI(-NC) 3-way isolation amplifier is used to electrically isolate and convert analog signals.
More informationCounting Photons to Calibrate a Photometer for Stellar Intensity Interferometry
Counting Photons to Calibrate a Photometer for Stellar Intensity Interferometry A Senior Project Presented to the Department of Physics California Polytechnic State University, San Luis Obispo In Partial
More informationLast Revision: August,
A3-1 HALL EFFECT Last Revision: August, 21 2007 QUESTION TO BE INVESTIGATED How to individual charge carriers behave in an external magnetic field that is perpendicular to their motion? INTRODUCTION The
More informationExperiment # 5 5. Coupled Water Tanks
Experiment # 5 5. Coupled Water Tanks 5.. Objectives The Coupled-Tank plant is a Two-Tank module consisting of a pump with a water basin and two tanks. The two tanks are mounted on the front plate such
More informationDOAS measurements of Atmospheric Species
Practical Environmental Measurement Techniques: DOAS measurements of Atmospheric Species Last change of document: April 14, 2014 Supervisor: Dr. Andreas Richter, room U2090, tel 62103, and Dr. Folkard
More informationAccuracy Estimations of S-Parameter Measurements at the Wafer-Level
WORKSHOP WSF13 (EuMC/EuMIC) Determining Accuracy of Measurements at High Frequencies from Error to Uncertainty Accuracy Estimations of S-Parameter Measurements at the Wafer-Level H. Heuermann*, A. Rumiantsev**
More information1. Prepare the MALDI sample plate by spotting an angiotensin standard and the test sample(s).
Analysis of a Peptide Sequence from a Proteolytic Digest by MALDI-TOF Post-Source Decay (PSD) and Collision-Induced Dissociation (CID) Standard Operating Procedure Purpose: The following procedure may
More informationConservation of Momentum Using PASCO TM Carts and Track to Study Collisions in One Dimension
14 Conservation of Conservation of Using PASCO TM Carts and Track to Study s in One Dimension When two objects collide momentum is transferred between them. p is defined as the product of mass and velocity
More information5730A High Performance Multifunction Calibrator. Extended specifications
730A High Performance Multifunction Calibrator Extended specifications General Specifications Warm-Up Time... Twice the time since last warmed up, to a maximum of 30 minutes. System Installation... Rack
More informationNonlinear Vector Network Analyzer Applications
Nonlinear Vector Network Analyzer Applications presented by: Loren Betts and David Root Agilent Technologies Presentation Outline Nonlinear Vector Network Analyzer Applications (What does it do?) Device
More informationRotary Motion Servo Plant: SRV02. Rotary Experiment #11: 1-DOF Torsion. 1-DOF Torsion Position Control using QuaRC. Student Manual
Rotary Motion Servo Plant: SRV02 Rotary Experiment #11: 1-DOF Torsion 1-DOF Torsion Position Control using QuaRC Student Manual Table of Contents 1. INTRODUCTION...1 2. PREREQUISITES...1 3. OVERVIEW OF
More informationRipple Method for Evaluating Residual Errors
Ripple Method for Evaluating Residual Errors H. Heuermann Univ. of Applied Sciences Aachen, Institute of High Frequency Tech., Eupener Str. 7, D-5266 Aachen, Germany Heuermann@FH-Aachen.de Heuermann HF-Technik
More informationThis section reviews the basic theory of accuracy enhancement for one-port networks.
Vector measurements require both magnitude and phase data. Some typical examples are the complex reflection coefficient, the magnitude and phase of the transfer function, and the group delay. The seminar
More informationDetermining the Optimum High Voltage for PMT s
Determining the Optimum High Voltage for PMT s Bertrand H. J. Biritz April 4, 006 In any experiment which uses photo-multiplier tubes, one needs to determine the optimum high voltage for each tube. Optimum
More informationA Nonlinear Dynamic S/H-ADC Device Model Based on a Modified Volterra Series: Identification Procedure and Commercial CAD Tool Implementation
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 52, NO. 4, AUGUST 2003 1129 A Nonlinear Dynamic S/H-ADC Device Model Based on a Modified Volterra Series: Identification Procedure and Commercial
More informationMicrowave Impedance Measurement for Nanoelectronics
276 M. RNDUS, K. HOFFMNN, MICROWVE IMPEDNCE MESUREMEN FOR NNOELECRONICS Microwave Impedance Measurement for Nanoelectronics Martin RNDUS, Karel HOFFMNN Dept. of Electromagnetic Field, Czech echnical University
More informationTransient Response of Transmission Lines and TDR/TDT
Transient Response of Transmission Lines and TDR/TDT Tzong-Lin Wu, Ph.D. EMC Lab. Department of Electrical Engineering National Sun Yat-sen University Outlines Why do we learn the transient response of
More information3.14 mv ma. Objectives. Overview
Phys 3 Lab 7 Ch 0 Simple DC and RC Circuits Equipment: power supply, banana cables, circuit board, switch, 0, 70, 460, & 30, k,two multi-meters, differential voltage probe, Phys 3 experiment kits: batteries
More informationESS 265 Spring Quarter 2005 Time Series Analysis: Linear Regression
ESS 265 Spring Quarter 2005 Time Series Analysis: Linear Regression Lecture 11 May 10, 2005 Multivariant Regression A multi-variant relation between a dependent variable y and several independent variables
More informationDeveloping a Scientific Theory
Name Date Developing a Scientific Theory Equipment Needed Qty Equipment Needed Qty Photogate/Pulley System (ME-6838) 1 String (SE-8050) 1 Mass and Hanger Set (ME-8967) 1 Universal Table Clamp (ME-9376B)
More informationLight Work Memo 4 rev 3: System Temperature Measurements July 7 1
Light Work Memo 4 rev 3: System Temperature Measurements - 2015 July 7 1 Subject: System Temperature Measurements set 4 Memo: 4, rev 3 From: Glen Langston Date: 2015 July 7 This note describes a Radio
More informationPackage Lead Code Identification SINGLE 3 SERIES 3 UNCONNECTED PAIR RING QUAD
Surface Mount RF Schottky Barrier Diodes Technical Data HSMS-28XX Series Features Surface Mount SOT-2/SOT- 4 Package Low Turn-On Voltage (As Low as 4 V at ma) Low FIT (Failure in Time) Rate* Six-sigma
More informationObservation of neutral hydrogen using FFT spectrometer Argos on a 5m telescope
Research Collection Report Observation of neutral hydrogen using FFT spectrometer Argos on a 5m telescope Author(s): Monstein, Christian; Meyer, Hansueli Publication Date: 2006 Permanent Link: https://doi.org/10.3929/ethz-a-005228693
More informationPascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure
DATASHEET Pascal ET Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure - electrical signals (ma, mv, V, ) - temperature (TC and
More informationRAPID growth in satellite-communications and mobile-communications
Calculate The Uncertainty Of N Measurements Simple modifications to the basic noise-figure equations can help in predicting uncertainties associated with test equipment. Duncan Boyd Senior Hardware Development
More informationISOCON-6. 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER
ISOCON-6 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER Whilst every effort has been taken to ensure the accuracy of this document, we accept no responsibility for damage, injury, loss or expense resulting
More informationSERIALLY PROGRAMMABLE CLOCK SOURCE. Features
DATASHEET ICS307-03 Description The ICS307-03 is a dynamic, serially programmable clock source which is flexible and takes up minimal board space. Output frequencies are programmed via a 3-wire SPI port.
More informationDocument Number: SPIRE-UCF-DOC Issue 1.0. November Matt Griffin
Sensitivity of the SPIRE Detectors to Operating Parameters Document Number: SPIRE-UCF-DOC-9 Issue. November 4 7 Matt Griffin Contents. Introduction.... List of symbols... 3. Principles of semiconductor
More informationThermal Conductivity and Electrical Loss of Thin Wall Millimeter Wave Stainless Steel Waveguides
Thermal Conductivity and Electrical Loss of Thin Wall Millimeter Wave Stainless Steel Waveguides J.D. Gallego, I. Malo, I. López, C. Diez González IT-CDT 015-14 Apdo. 148 19080 Guadalajara SPAIN Phone:
More informationEXPERIMENT 7: ANGULAR KINEMATICS AND TORQUE (V_3)
TA name Lab section Date TA Initials (on completion) Name UW Student ID # Lab Partner(s) EXPERIMENT 7: ANGULAR KINEMATICS AND TORQUE (V_3) 121 Textbook Reference: Knight, Chapter 13.1-3, 6. SYNOPSIS In
More informationThree Axis Magnetic Field Transducer x-h3x-xx_e3d-2.5khz-1-0.2t
DESCRIPTION: TYPICAL APPLICATIONS: The is a high accuracy magnetic flux density-toanalogue-voltage transducer with a highlevel and temperature compensated output signal for each of the three components
More informationInformation on the test material EDS-TM002 and the BAM software package EDX Spectrometer Test for determination of the spectrometer performance
BAM 6.8 8.5.213 Information on the test material EDS-TM2 and the BAM software package EDX Spectrometer Test for determination of the spectrometer performance 1. Introduction Energy dispersive spectrometers
More informationHow we wanted to revolutionize X-ray radiography, and how we then "accidentally" discovered single-photon CMOS imaging
How we wanted to revolutionize X-ray radiography, and how we then "accidentally" discovered single-photon CMOS imaging Stanford University EE Computer Systems Colloquium February 23 rd, 2011 EE380 Peter
More informationDIGITAL COMMUNICATIONS. IAGlover and P M Grant. Prentice Hall 1997 PROBLEM SOLUTIONS CHAPTER 6
DIGITAL COMMUNICATIONS IAGlover and P M Grant Prentice Hall 997 PROBLEM SOLUTIONS CHAPTER 6 6. P e erf V σ erf. 5 +. 5 0.705 [ erf (. 009)] [ 0. 999 979 ]. 0 0 5 The optimum DC level is zero. For equiprobable
More informationDesigning Information Devices and Systems I Fall 2018 Homework 5
Last Updated: 08-09-9 0:6 EECS 6A Designing Information Devices and Systems I Fall 08 Homework 5 This homework is due September 8, 08, at 3:59. Self-grades are due October, 08, at 3:59. Submission Format
More informationExperiment 4. RC Circuits. Observe and qualitatively describe the charging and discharging (decay) of the voltage on a capacitor.
Experiment 4 RC Circuits 4.1 Objectives Observe and qualitatively describe the charging and discharging (decay) of the voltage on a capacitor. Graphically determine the time constant τ for the decay. 4.2
More informationDC-motor modelling and parameter identification
DC-motor modelling and parameter identification This version: November 1, 2017 Name: LERTEKNIK REG P-number: Date: AU T O MA RO TI C C O N T L Passed: LINKÖPING Chapter 1 Introduction The purpose of this
More informationSUPPLEMENTARY INFORMATION
Systematic shift caused by trap asymmetry The major systematic correction in the reported cyclotron frequency ratio comparison of an antiproton at ν c, p and a negatively charged hydrogen ion (H ) at ν
More informationBasic Electronics. Introductory Lecture Course for. Technology and Instrumentation in Particle Physics Chicago, Illinois June 9-14, 2011
Basic Electronics Introductory Lecture Course for Technology and Instrumentation in Particle Physics 2011 Chicago, Illinois June 9-14, 2011 Presented By Gary Drake Argonne National Laboratory Session 2
More informationCHAPTER.6 :TRANSISTOR FREQUENCY RESPONSE
CHAPTER.6 :TRANSISTOR FREQUENCY RESPONSE To understand Decibels, log scale, general frequency considerations of an amplifier. low frequency analysis - Bode plot low frequency response BJT amplifier Miller
More informationSingle-Time-Constant (STC) Circuits This lecture is given as a background that will be needed to determine the frequency response of the amplifiers.
Single-Time-Constant (STC) Circuits This lecture is given as a background that will be needed to determine the frequency response of the amplifiers. Objectives To analyze and understand STC circuits with
More informationExperiment 9 Equivalent Circuits
Experiment 9 Equivalent Circuits Name: Jason Johnson Course/Section: ENGR 361-04 Date Performed: November 15, 2001 Date Submitted: November 29, 2001 In keeping with the honor code of the School of Engineering,
More informationOptoelectronic Applications. Injection Locked Oscillators. Injection Locked Oscillators. Q 2, ω 2. Q 1, ω 1
Injection Locked Oscillators Injection Locked Oscillators Optoelectronic Applications Q, ω Q, ω E. Shumakher, J. Lasri,, B. Sheinman, G. Eisenstein, D. Ritter Electrical Engineering Dept. TECHNION Haifa
More informationUV-Vis optical fiber assisted spectroscopy in thin films and solutions
UV-Vis optical fiber assisted spectroscopy in thin films and solutions Description UV-Visible absorption and transmission spectra provide fundamental information for all experiments related to the attenuation
More informationLABORATORY NUMBER 9 STATISTICAL ANALYSIS OF DATA
LABORATORY NUMBER 9 STATISTICAL ANALYSIS OF DATA 1.0 INTRODUCTION The purpose of this laboratory is to introduce the student to the use of statistics to analyze data. Using the data acquisition system
More informationAdvanced Current Mirrors and Opamps
Advanced Current Mirrors and Opamps David Johns and Ken Martin (johns@eecg.toronto.edu) (martin@eecg.toronto.edu) slide 1 of 26 Wide-Swing Current Mirrors I bias I V I in out out = I in V W L bias ------------
More informationStatic and Kinetic Friction
Experiment Static and Kinetic Friction Prelab Questions 1. Examine the Force vs. time graph and the Position vs. time graph below. The horizontal time scales are the same. In Region I, explain how an object
More informationCalibration Uncertainty Estimation. at the Wafer Level
Calibration Uncertainty Estimation for the S-Parameter S Measurements at the Wafer Level A. Rumiantsev SUSS MicroTec Test Systems GmbH Sussstr. 1, Sacka, D-01561, Germany a.rumiantsev@ieee.org Outline
More informationHooke s Law. Equipment. Introduction and Theory
Hooke s Law Objective to test Hooke s Law by measuring the spring constants of different springs and spring systems to test whether all elastic objects obey Hooke s Law Equipment two nearly identical springs,
More informationStatic and Kinetic Friction
Ryerson University - PCS 120 Introduction Static and Kinetic Friction In this lab we study the effect of friction on objects. We often refer to it as a frictional force yet it doesn t exactly behave as
More informationSCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION
SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION OBJECTIVE The primary objective of this experiment is to use an NaI(Tl) detector, photomultiplier tube and multichannel analyzer software system
More informationVNA Tools II S-parameter uncertainty calculation
VA Tools II S-parameter uncertainty calculation Michael Wollensack METAS 5. May 0 Michael Wollensack METAS Outline Introduction VA Measurement Model Database Uncertainty Visualization Results Michael Wollensack
More informationImpulse and Change in Momentum
Activity 15 PS-2826 Impulse and Change in Momentum Mechanics: momentum, impulse, change in momentum GLX setup file: impulse Qty Equipment and Materials Part Number 1 PASPORT Xplorer GLX PS-2002 1 PASPORT
More informationEVLA Memo 143: Improving the frequency resolution of the default atmospheric opacity model
EVLA Memo 143: Improving the frequency resolution of the default atmospheric opacity model J. Marvil NRAO April 6, 2010 ABSTRACT In consideration of the wider bandwidths and improved frequency accessibility
More informationPHYSICS 122 Lab EXPERIMENT NO. 6 AC CIRCUITS
PHYSICS 122 Lab EXPERIMENT NO. 6 AC CIRCUITS The first purpose of this laboratory is to observe voltages as a function of time in an RC circuit and compare it to its expected time behavior. In the second
More informationCircular Motion and Centripetal Force
[For International Campus Lab ONLY] Objective Measure the centripetal force with the radius, mass, and speed of a particle in uniform circular motion. Theory ----------------------------- Reference --------------------------
More informationFinal Report August 2010
Bilateral Comparison of 100 pf Capacitance Standards (ongoing BIPM key comparison BIPM.EM-K14.b) between the CMI, Czech Republic and the BIPM, January-July 2009 J. Streit** and N. Fletcher* *Bureau International
More informationLab 01: Harmonic Motion I. Theory: Three experiments. The first we measure the oscillation of a spring, the second of a rubber band (non-linear).
Dr. W. Pezzaglia Physics 8C Lab, Spring 04 Page Las Positas College Lab # Harmonic Motion 04Jan3 Lab 0: Harmonic Motion I. Theory: Three experiments. The first we measure the oscillation of a spring, the
More informationGHz 6-Bit Digital Phase Shifter
180 90 45 22.5 11.25 5.625 ASL 2004P7 3.1 3.5 GHz 6-Bit Digital Phase Shifter Features Functional Diagram Frequency Range: 3.1 to 3.5 GHz RMS Error < 2 deg. 5 db Insertion Loss TTL Control Inputs 0.5-um
More informationHomework Assignment 08
Homework Assignment 08 Question 1 (Short Takes) Two points each unless otherwise indicated. 1. Give one phrase/sentence that describes the primary advantage of an active load. Answer: Large effective resistance
More informationINTEGRATED CIRCUITS. For a complete data sheet, please also download:
INTEGRATED CIRCUITS DATA SHEET F a complete data sheet, please also download: The IC06 74HC/HCT/HCU/HCMOS Logic Family Specifications The IC06 74HC/HCT/HCU/HCMOS Logic Package Infmation The IC06 74HC/HCT/HCU/HCMOS
More informationSection 4. Nonlinear Circuits
Section 4 Nonlinear Circuits 1 ) Voltage Comparators V P < V N : V o = V ol V P > V N : V o = V oh One bit A/D converter, Practical gain : 10 3 10 6 V OH and V OL should be far apart enough Response Time:
More informationProduct Data. Brüel & Kjær B. Sound Intensity Calibrator Type 3541
Product Data Sound Intensity Calibrator Type 3541 USES: Sound intensity and particle velocity calibrations Sound pressure calibration Measurement of pressure-residual intensity index FEATURES: Intensity
More information