Size: px
Start display at page:

Download ""

Transcription

1 CHAPTER 7: TRANSDUCERS In general terms, the transduction process involves the transformation of one form of energy into another form. This process consists of sensing with specificity the input energy from the measurand by means of a "sensing element" and then transforming it into another form by a "transduction element." The sensor-transduction element combination shown in figure below will henceforth be referred to as the "transducer". Measurand relates to the quantity, property, or state that the transducer seeks to translate into an electrical output. Transducers may be classified as self-generating or externally powered. Self-generating transducers develop their own voltage or current and in the process absorb all the energy needed from the measurand. Externally powered transducers, as the name implies, must have power supplied from an external source, though they may absorb some energy from the measurand. I. TRANSDUCTION MECHANISMS Capacitive Thermoelectric effects (Seebeck and Peltier) Inductive and electromagnetic Resistive Ionization effects Resistive and thermoresistive Photoelectric effect Piezoresistive effect Photoresistive effect Hall effect Photovoltaic effect Lateral effect Acoustooptic effect Extrinsic, interferometric and evanes- Fluorescence and fluorescence quenching cent effects in optical fiber Field effect Magnetoresistive effect Doppler effect Tunneling effect EE Transducers 125

2 II. MEASURANDS Displacement Atomic and surface profiles Position Gas concentration and ph Velocity ph and partial pressure of O 2 and CO 2 Acceleration in blood Force and load Infrared radiation Strain Torque Rotation and encoding Magnetic field Vibrations Acoustic fields Flow Medical imaging Temperature Non-destructive testing Pressure Audio fields and noise Vacuum Rotation and guidance III. SELECTION OF TRANSDUCERS When selecting a transducer, in addition to the question of cost, careful attention must be given to the following Sensitivity Range Physical properties Loading effect and distortion Frequency response Electrical output format Output impedance Power requirements Noise Error or accuracy Calibration Environment EE Transducers 126

3 IV. CLASSIFICATION OF TRANSDUCERS EE Transducers 127

4 Before discussing the transduction principles, we first examine several methods for combining individual transduction principles into one single compound transducer. These composite methods are employed to reduce or even totally eliminate certain restrictions associated with individual transducers. A widespread method is to use 2 identical transducers in a balanced configuration shown in Figure 3.1 below. If the transducers have the same transfer characteristic, the output y of the balanced configuration is: y = f(x) f(-x) Here f(x) may be the non-linear transfer function that we intend to linearize. Let assume that f(x) can be expressed with a Taylor expansion in the following form: Using the balance equation above, we find: f(x)= a 0 +a 1 x+a 2 x 2 + a 3 x 3 + a 4 x 4 + a 5 x 5. y = 2a 1 x + 2a 3 x 3 + 2a 5 x 5 + Evidently, the constant (or offset) a 0 and the even terms a 2 x 2, a 4 x 4, disappear when balancing the two transducers. If the non-linearity of f(x) does not contains any uneven terms, we will obtain a perfectly linear system. The systems then referred to as a difference configuration. However, usually balancing will only improve the linearity of the system over a limited range of the input quantity x. Such system is called a differential configuration. The balanced configuration is not sensitive to external disturbances, since it inherently makes use of parallel compensation. The configuration is immune to additive disturbances when the transducers T and T have the same sensitivity to these disturbances. In order to be immune to multiplicative disturbances, the transducers T and T must have disturbance coefficients with the same magnitude, but opposite signs. A second commonly used configuration is the feedback configuration of 2 transducers T1 and T2 in Figure 3.2 below. The purpose of the system is to convert an input signal x into an electrical output signal y. We could use a single transducer T1 for this. Assume T1 is unsuitable to be used directly, due to an unacceptably large non-linearity and has too large a sensitivity to disturbances. If we have a second transducer, capable of the reverse conversion (conversion of y to x) and this conversion is linear and not susceptible to disturbances, then when we combine both transducers T1 and T2 (with an amplifier to increase the loop gain) in the feedback configuration, we can realize a compound transducer for conversion of measurement signal from x into y, with the same characteristics as the employed reverse transducer. The necessary conditions for achieving this EE Transducers 128

5 are a large gain and a quasi-static dynamic behavior for T1, T2 and A. In practice, though, the dynamic behavior of especially T1 and T2 is often of a higher order and, therefore, the situation may not be as ideal as described above. Finally, the reliability of the transducers is often a problem. Sometimes this can be solved by using several transducers which all measure the same quantity. This redundant configuration greatly improves the reliability of the system at, of course, an increased cost. V. MECHANOELECTRIC TRANSDUCERS These transducers measure mechanical quantities. 1. Displacement transducers Displacement transducers can measure either linear displacement (translation) or angular displacement (rotation). They can also be classified according to the principle of transduction on which they are based. We can distinguish between, for instance, resistive, capacitive, inductive and optical translation or rotation transducers. These mechanical transducers are also referred to as gauges or sensors. a. Resistive displacement transducers The potentiometric transducer is a popular type of displacement transducer. For measuring translation it is, in fact, no more than a sliding potentiometer. For measuring rotation, a rotating potentiometer can be used. Usually the potentiometers are wire-wound to achieve a better accuracy, temperature coefficient, etc. However, a limitation of wire-wound displacement transducers is their finite resolution. The maximal resolution R = x/ax is equal to the number of turns on the potentiometer body. Metal film potentiometers do not have this limitation. A disadvantage of all transducers of the potentiometric type is that mechanical wear and chemical corrosion can change the transfer characteristic over the lifetime of the transducer. Another type of resistive displacement sensor makes use of the fact that the electrical resistance of a conductor depends on the dimensions of the conductor. The resistance R is a function of the cross sectional area of the conductor, its length l and its resistivity ρ. EE Transducers 129

6 R=R(A,l,ρ) If the conductor is mechanically strained or compressed, the parameters A, l, and ρ, and as a consequence R, will change. This enables one to measure very small displacements. Figure 3.4a shows a piece of wire (stain gauge) which is elongated a distance l by applied tensile stresses. Figure 3.4(b) shows the meanders of a metal foil strain gauge are made extra wide when they turn, in order to reduce the sensitivity of the strain gauge to strain perpendicular to the main axis of operation. If, besides the magnitude of the strain, we also wish to measure the direction of the strain, a combination of strain gauges is used, arranged in a certain geometric pattern, for instance 3 strain gauges oriented at 120 degrees angles with respect to one another. This arrangement is known as a strain gauge rosette. b. Capacitive displacement sensors b. Capacitive displacement transducers Capacitance C is a function of distance d between the electrodes of a structure, the surface area A of the electrodes and the permittivity ε of the dielectric. C=C(d,A,ε) Therefore the capacitance can be changed when the three variables are changing. Figure 3.6 shows the displacements sensors using the detection of capacitance and the applicable capacitances for each displacement. It is shows that the variable distance transducer has a non-linear characteristics and can be linearized by using it in a balanced configuration. EE Transducers 130

7 The force F(x) causes the movement of sensor with a constant voltage across the capacitor is given by: c. Inductive displacement sensors It is not only possible to vary the self-inductance of a single coil as a function of the displacement to be measured, it is possible to vary the mutual inductance between two coils as a function of the displacement. An obvious way of varying the inductance of a coil is to vary the effective number of turns, Figure 3.8(a), and vary the magnetic resistance (reluctance) of the yoke by means of an air gap of variable width, Figure 3.8(b). The displacement of the second transducer, based on variable reluctance, is incorporated in a bridge network to obtain a linear transfer characteristic. d. Optical displacement sensors Displacement can also be detected optically, by means of an encoding strip (translation) or a rotary encoder (rotation). Figure 3.10(a) shows an optical displacement sensor which utilizes an encoder strip consisting of rows of alternating transparent and opaque bars. The position of the strip is converted directly to a digital signal with a narrow beam of light and a number of light sensors. The digital code is determined by the position of the transparent and opaque bars on the strip. EE Transducers 131

8 2. Velocity transducers There are two transducers used to measure velocity: translational and angular transducers. The measure of velocity is often converted to a frequency measurement. The conversion is performed by a strip or disc on which a large number of marks (detection elements) have been put at equal distances x. The velocity can be calculated from v= xn/t = xf, in which n is the number of detection element which passes the detector in t seconds and f is the frequency of output signal. The detection can be performed optically, mechanically, inductively or capacitively. a. Measurement of velocity by differentiation or integration: For linear velocity v(t) dx(t) v (t) = dt For angular velocity ω(t): d!(t) "(t) = dt where x(t) is the linear displacement where θ(t) is the rotation angle. From these two equations, it is evident that we can obtain a signal, for the velocity of an object, by calculating the derivative of the output signal of a displacement sensor. It is simple to design an electronic circuit for differentiating an electrical signal. Another possibility for obtaining the velocity is by integrating of the linear acceleration a(t) t v(t) =! a(t)dt + v(0) or of an angular acceleration α(t): The integration can be performed by a relative simple analogue electronic circuit. It should be noted that the differentiator and integrator circuits having a disadvantage of increasing or decreasing very slowly, even when the input is zero, due to leakage or drop. b. Inductive velocity transducers: In this kind of transducers, the velocity of the measurement object is made to give rise to a change of magnetic flux Φ, which induces an electrical potential in a conductor. The induced voltage of Figure 3.11(a) in turn i of the coil of this inductive velocity pick up is given by: For a total of n turns, the coil terminal voltage is: 0 t "(t) =!#(t)dt + "(0) 0 Vi = d$ i dt n EE Transducers V = 132 " d$ i i i= n dt

9 Since ΦI=ΦI(x), in which x is the position of the magnet with respect to the center of the coil, and x=x(t), we find: n d$ n = i dx d$ V " = i i v" = vk(x) i= n dt dt i= 1 dx Thus, the output voltage V is proportional to the velocity v of the magnet for a given value of x. The sensitivity of the transducer is equal to k. Unfortunately, since k=k(x), the transducer is non-linear. Therefore, again this type of transducer is often used in a balanced configuration. Since the magnet is moving here the velocity sensor is referred to as a magneto-dynamic transducer. 3. Acceleration transducers Transducers for measuring acceleration rely on the measurement of the force F required to give a known mass (the seismic mass m) the same acceleration, a, as the measurement object. From the force and the mass, the acceleration is determined: a=f/m. The extra mass has to be kept to a minimum, especially when the measurement object is highly elastic or has a low mass (extra mass influences the measure acceleration). VI. THERMO-ELECTRIC TRANSDUCERS 1. Resistive temperature sensors a. Metal: The electrical resistance of any material depends to a certain extend on the temperature and can be used to convert a temperature measurement into a measurement of resistance. Depending on the material used, we can distinguish two kinds of thermometers: metal thermometers and semiconductor thermometers. The resistive of pure metals can be written in the form of power series: 2 2 R(T) = R(T0 )[1 + #(T T0 ) + &(T T0 ) + %(T T0 ) +...] EE Transducers 133

10 in which R(T) is the resistance of the sensor at temperature T and R(T 0 ) is the resistance at certain reference temperature T 0. If the temperature range is not too large, the first 2 terms of the expansion will suffice, the sensor is approximately linear. The most frequently used metals are platinum and nickel. The measurement range of a platinum sensor runs from 70K to 1000K and nickel from 200K to 500K. At T 0 =273K: α Pt =3.85 x 10-3 K -1 α Nt =6.17 x 10-3 K -1 β Pt =-5.83 x 10-7 K -2 γ Pt =-3.14 x K -3 b. Semiconductors: In both intrinsic and extrinsic semiconductors, this effect is overshadowed by a much stronger effect: the number of free charge carriers depends on the absolute temperature. The higher the temperature, the larger the number of electrons which cross the band gap from the valence band into the conduction band (intrinsic) or the larger the number of activated donor and acceptor atoms (extrinsic). The number of free charge carriers increases according to: n=n 0 e -Eg/2kT in which E g is the energy required for the crossing the band gap and k is Boltzmann s constant ( x J/K). Thus the resistance of a semiconductor will decrease as the temperature increases, the semiconductor has a Negative Temperature Coefficients (NTC-resistor). The resistance of a semiconductor can be expressed as: R(T)=Ae B/T The coefficients A and B also depend on the temperature and therefore a more accurate expression is given by: B(1/ T 1/ T0 ) R(T) = R(T0 )e Obviously, a semiconductor sensor is highly non-linear. The temperature coefficient is given by: In practice, the value of the coefficient B lies between 2700K and 5400K at a temperature of 300K. Thus, at 300K, the temperature coefficient ranges from 3 x 10-2 K -1 to 6 x 10-2 K -1. At 300K, a semiconductor sensor is an order of magnitude more sensitive than a metal sensor. Such a semiconductor temperature-sensing resistor is often referred to as thermistor. More improvement in the making of thermistors yields better linear characteristic. 2. IC temperature sensors 1 dr(t) B #(T) = = R(T) dt 2 T An alternative temperature sensor can be found in the bipolar transistor. Such a sensor makes use of the fundamental band gap voltage of silicon, which depends in the temperature. Two EE Transducers 134

11 bipolar transistors located close together, on the same IC, are biased to different collector current. If the ratio of the (collector) current densities (the transistors may have different areas) is equal to r, the difference between the base-emitter voltages of the two transistors is equal to (kt/q)ln(r). This base-emitter voltage difference is a linear measure of absolute temperature. Addition electronic circuits amplify this voltage to provide a practical output value. Typical IC sensor specifications are: range -55 o C to 150 o C, non-linearity over the entire range of approximately 0.3K, sensitivity 10mV/K or 1uA.K, instability over 1000hr of operation ±0.08K, dissipation 1.5mW to 3mW. 3. Thermal couples When two different metals are brought into atomic contact with each other, an electrical potential difference is generated. This so-called junction potential depends only on the nature of the two metals and on the absolute temperature. The surface area of the junction has no influence. For many combinations of metals the junction potential difference is approximately linearly proportional to the absolute temperature of the junction, provided that the temperature range is not too large. When two junctions are connected in series, as illustrated in Fig. 3.16(a), a net thermo voltage V will result if the two junctions are at different temperatures. This configuration is called a thermocouple. Several characteristics of thermocouples have been plotted in Fig. 3.16(b). The thermovoltage is a measure of the temperature difference between the two junctions. The output voltage of a thermocouple can be written more accurately as a power series of the temperature difference T-To, with To a certain calibration temperature than it can if it is assumed to be proportional to temperature: i.e. As n increases, this expression will describe the behavior of a given thermocouple more and more accurately. Every thermocouple (so every combination of two metals) is characterized by its own series of temperature independent coefficients a i (i = 1,..., n). An inaccuracy of ±l% requires roughly eight coefficients (n=8) for most materials. The coefficient a 1 is referred to as the Seebeck coefficient. Fig. 3.16(c) shows how this coefficient would depend on the temperature if a thermocouple were to be described by only one single coefficient, a 1. EE Transducers 135

12 If we are dealing with a large temperature range, we must use more than one coefficient for reasons of accuracy. There are four physical effects that contribute to the output voltage of the thermocouple. -Seebeck effect. This is the desired effect which is measured when no current is allowed to flow through the thermocouple. It arises from the temperature dependence of the junction potential difference. This junction potential difference originates from the difference in Fermi levels of two dissimilar metals. The higher the temperature, the larger the number of electrons with a higher energy level than the Fermi level. This causes the junction potential difference to become temperature dependent. -Peltier effect. When a current flows through a junction of two dissimilar metals, the temperature of the junction will change. Depending on the direction of the current, the junction will become either warmer or cooler than ambient. This effect is caused by the fact that with every electrical conduction process there is also transportation of heat. In a metal, thermal conduction as well as electrical conduction are caused by free electrons. The Peltier effect is undesirable in a thermocouple since it gives rise to a temperature error. -Thomson effect. If a current is flowing through a uniform metal conductor in the direction in which there is a negative temperature gradient, thermoelectric heat will be generated. If the direction of the current is reversed, heat will be extracted from the conductor. This reversible effect also originates from the fact the electrical conduction process in a metal is accompanied by the transfer of heat and, inversely, heat conduction is accompanied by electrical conduction. This effect also gives rise to errors. -Joule heat. With the last two effects, we assumed that no heat was generated by the dissipation of electrical energy in the electrical resistance of the metals. If the total resistance is R, then, per second, I 2 R joules is dissipated. The thermocouple will therefore heat itself up. Thus, the clear conclusion is that no current may flow through the thermocouple when one is interested in accurate temperature measurements; the measurement circuit must have a high input impedance. An additional source of measurement errors is moisture. Moisture can create a galvanic element with both metals, which generates a galvanic cell voltage many times larger than that of the thermocouple. Therefore, thermocouples are often supplied in a waterproof case. If we wish to measure the absolute temperature as opposed to a temperature difference by means of a thermocouple, we must hold one of the junctions at a fixed known reference temperature. This can be achieved by controlling the temperature of one of the junctions with a thermostat. It is also possible to compensate the temperature of the reference junction, as indicated in Fig The temperature of the reference junction is measured here by a resistance sensor R(T), which is connected to a bridge network. The output voltage of the bridge is connected in series with the thermocouple such that it compensates for the temperature of the reference junction. Of course, the temperature sensitivity of the reference junction must oppose that of the bridge network. The temperature measured by the active junction AB is usually located some distance away from the rest of the circuitry. The metal of the thermocouple junction is usually too EE Transducers 136

13 expensive to use for long interconnections and, therefore, two wires A'B' of a cheaper metal are used for the interconnections. Provided that these two wires A' and B' have the same thermoelectric characteristics as the two wires A and B of the measurement junction, no error will be introduced. It can be seen that when two (other) metal conductors are connected in series with the two junctions of the thermocouple, as in Fig. 3.17, the new junctions will not contribute a net potential difference, as long as the new junctions are kept at the same temperature (isothermal). The extra junction potentials will cancel each other. Temperature differences between these new junctions will cause measurement errors. If the cable which connects the two junctions of the thermocouple has the same thermoelectric characteristics it is referred to as compensation cable. 4. Radiation thermometers A radiation thermometer absorbs a fraction of the infrared radiation emitted by the measurement object. A radiation thermometer for high temperatures is usually called a pyrometer. The radiation is usually focused on the actual thermal detector by means of a concave mirror (as illustrated in Fig. 3.18(a)). If the temperature of the measurement object is lower than that of the detector, it will supply heat energy to the object, causing the detector to cool. The use of lenses is avoided, especially for low temperatures, since lenses which easily transmit heat radiation (diathermanous infra-red lenses) are very expensive. Other pyrometers are based on quantum detectors. This principle of operation relies on the electrons of the material being excited by infrared radiation. This can only occur when the quantum energy E of the radiation quanta is equal to or larger than a certain threshold energy Eo. This threshold energy corresponds to the transition of electrons to a higher energy state. With: EE E hc Transducers 0 = hf0 = 137 0

14 This threshold energy is related to infrared radiation above a certain minimal frequency fo or below a certain maximal wavelength λ 0 Fig. 3.18(b) shows the spectral sensitivity of this kind of quantum detector. Photodiodes are used as quantum detectors for near infrared radiation; photoresistors are used in the far infrared region. Quantum detectors respond quickly, but only measure radiation in a limited band of wavelengths. Pyrometers are used for measuring extremely high temperatures (T>1000K) when no other means are available. They are also used for measuring objects with a high thermal resistance, such as plastics, rock, etc. Besides the described methods for converting a temperature into an electrical signal, there are many more. One is the (quartz) crystal thermometer which is based on the temperature dependence of the resonance frequency of a piezoelectric crystal. The crystal is used to determine the frequency of an oscillator accurately. This frequency can be determined by the enumeration method, resulting in a small inaccuracy of only 0.01K and a very high resolution of 10 4 in the range from -80 C to 250 C. These excellent specifications are realized by cutting the quartz crystal very accurately with respect to the crystal orientation axes. This is done in such a way that the temperature sensitivity of the resonance frequency is maximal and as stable as possible. A calibration curve of temperature versus frequency is recorded and built into the instrument (PROM) to increase the accuracy even further. VII. MAGNETOELECTRIC TRANSDUCERS The induction of a magnetic field can be measured with transducers referred to as magnetometers or magnetic field sensors. The magnetic induction B is expressed in teslas (T). Sometimes, it is also referred to as the magnetic flux density. A unit which is consistent with this terminology is the weber per square meter with, of course, 1T = 1Wb/m2. Sometimes a rotating coil (with area A, n turns and angular frequency ω) is used for measuring the induction B of a static magnetic field. Assuming the coil is so small that the induction is constant across the surface area of the coil and B n is the component of B perpendicular to the axis of rotation, the flux Φ through the coil is equal to Φ = B n A sin θ(t), where (θ) is the instantaneous angle between the coil and B n. With θ(t) =ωt, the induced ac voltage is given by: d$ V = n = nbn"cos("t) dt We can determine B n from this expression. Induction sensors always require a changing flux. The change in magnetic flux can obviously also arise from an alternating magnetic field in a static coil. Another type of magnetometer is based on the influence of a magnetic field on the electrical resistance of a material. As early as 1856 W. Thomson (Lord Kelvin) discovered that if one exposes a current conducting body to a magnetic field the electrical resistance changes. This effect, called the magnetoresistive effect, was only used much later for realizing transducers. It was not until the American physicist E. F. Hall had discovered the so called Hall effect that an explanation for this phenomenon could be given. EE Transducers 138

15 The Hall effect is a result of the Lorentz force, exerted on charge carriers in solids. When a platelet of conducting material is positioned in a magnetic field (as sketched in Fig. 3.19), the charge carriers will be deflected in a direction perpendicular to the direction of motion of these charge carriers and perpendicular to the induction vector B of the magnetic field. The Lorentz force acting on a charge q with velocity v is equal to: F l =q(vb) As the charge carriers are deflected, a transverse charge gradient will build up, resulting in an electrical field E across the plate. This field will subject the charge carriers to an opposing force F e which is given by: F e =qe At a certain point, equilibrium is reached between the Lorentz force and the force produced by the electrical field, so F e = F l and therefore: E=vB Assuming the charge carriers all have approximately the same velocity v, the current density J is equal to nqv, with n the concentration of charge carriers. If, in addition, we assume that B is perpendicular to v, as in Fig. 3.19, then E = JB/nq. The factor 1/nq is called the Hall coefficient and usually denoted by R H. With I=bdJ and V =Eb we find: V = 1 IB nq d = RH For semiconductor materials, in which the majority charge carriers are holes (p-type semiconductors), q is positive and the output voltage of the Hall plate will have the polarity indicated in Fig If the majority charge carriers are electrons (n-type semiconductors) the polarity will be opposite. The Hall coefficient is large for semiconductors, since the concentration n of charge carriers is much smaller in semiconductors than in metals. Obviously, a Hall plate can measure static magnetic fields, without any moving or rotating parts. A Hall plate is also suitable for high-frequency measurements; it has a wide frequency range (up to several GHz). Furthermore, the disturbance to the magnetic field, caused by the Hall plate is very small. In addition to measuring magnetic fields directly, Hall plates are often used for measuring large DC currents and in current probes for oscilloscopes. IB d EE Transducers 139

16 VIII. TRANSDUCER SELECTION The selection of a transducer begins with the specifications of the physical quantity to be measured. The user must ascertain the required accuracy of the measurement, the duration of the test, and may need to consider cyclic behavior or other factors. In addition, consideration needs to be given to the environment in which the transducer is placed. Finally, the calibration procedure should be considered. Each of these categories is discussed as part of the selection criteria for transducers. The general selection criteria listed in this section should be considered representative; additional factors may be necessary for specific transducers. MEASUREMENT REQUIREMENTS The measurement requirements for a transducer are as follows Range: The range is the set of values a transducer is designed to measure. The minimum and maximum values of the transducer's range are called the endpoints. Some transducers can be adjusted to cover a different range by attenuating the measurand -for example, a sensitive radiation transducer can be used if the measured quantity is attenuated using radiation-absorbing filters. It is not always possible to find a single transducer to cover the entire range of measurand values; in these cases transducers with overlapping ranges must be selected Input threshold The input threshold is the smallest detectable value of the measured quantity starting near the zero value of the variable. For an input to be discerned, it must be possible to assign a unique number to the input. The selection of a transducer requires that it respond in a discernible manner to the threshold. Dynamic behavior The dynamic behavior of a transducer specifies how the transducer can respond to a changing input. No transducer could follow an instantaneous change. (For that matter, no measurand can change instantaneously!) The transducer's dynamic performance is usually specified as a frequency response or response time, depending on the type of transducer. The response time is the time required to reach a specified percentage (typically 90% to 99%) of the final value for a given change of the input. The response time is measured in much the same way as the time constant for an RC or RL circuit. (Recall that the time constant is the time required for the output to reach 63% of its final value.) Accuracy and resolution Accuracy is the difference between the measured and accepted value. The accuracy requirements for a particular measurement can greatly affect the total cost of the measurement system. In addition, certain transducers, such as strain gages and pressure transducers, have a fatigue life that can change the accuracy, depending on the duration and cyclic behavior of the measurand. In some cases, the accuracy isn't as important as the ability to detect a small change (resolution), as when quantities are being compared. For example, in underground-tank testing, the interface between the liquid and air can be located by observing the small temperature difference between the air and liquid. In other cases, consistency is the most important criterion. Other accuracy errors include nonlinearities due to a zero shift or drift due to aging, which can affect the long-term repeatability of measurements EE Transducers 140

17 Repeatability and hysteresis errors Repeatability is the maximum difference between consecutive measurements of the same quantity when the measured point is approached each time from the same direction for full-range traverses. It is usually expressed as a percentage difference of the full-scale output. Hysteresis error is the maximum difference between consecutive measurements for the same quantity when the measured point is approached each time from a different direction for full-range traverses. An example is when backlash in gearing causes the readings of a dial to be different, depending on whether the gearing was turned in one direction or the other. OPERATIONAL AND ENVIRONMENTAL CONSIDERATIONS The operational and environmental considerations for a transducer are as follows Natural hazards The transducer, wiring, and connectors must all be able to withstand the effects of exposure to the required environment. Natural hazards include the effects of dust and dirt, high or low temperatures, water (including salt water), and humid conditions. Wire insulation is available that is resistant to these effects as well as solvents, acids, bases, and so forth. In contrast, the transducer should not present a hazard to the environment in which it is placed-including causing electrical problems such as explosion hazard or shock hazard. Human-caused hazards Human-caused hazards include high-radiation environments, corrosive or dangerous chemicals, immersion, abrasion, vibration, and explosive environments, to name a few. The electrical signal from the transducer may be interfered with if the signal cables are routed in an electrically noisy environment, another possible hazard for proper operation, particularly with transducers that have a low output signal. Power requirements Power requirements depend on the type of transducer. Passive transducers, such as photocells, convert some of the incident source energy into electrical energy. Others require a source of excitation, which can be a dc or ac source. If the transducer is being operated in a remote or noisy environment, the power leads become a source of potential problems. Signal-conditioning requirements If the transducer produces a very small signal or it is located at a remote location or in a noisy environment, amplification or other signal conditioning may be required at the transducer. The transducer output may need to be converted into a compatible format for the remainder of the instrumentation system. Physical requirements In some installations, the space available may be limited or the measurand may be over a limited region. If the quantity to be measured is concentrated, such as in the case of a collimated light source, the physical size of the transducer can affect the output. Loading effects Loading effects cause the measurand to be disturbed in some manner by the presence of the transducer. All measurements in some way modify the quantity to be measured. For example, a rotating-vane flowmeter extracts a small amount of the energy from the fluid to turn the vane and EE Transducers 141

18 thus changes the flow from its undisturbed value. Loading can also occur in the electrical-measuring circuit. Electrical loading can occur when a transducer with an equivalent high Thevenin resistance is connected to an amplifier with a finite input impedance. In this case the transducer signal is reduced by the connection to the amplifier. Many transducers have a very high equivalent Thevenin resistance (ph electrodes, for example), so it is important to be aware of the need for compatible amplifiers. Human factors Human factors that need to be considered in the selection of a transducer are the operating skill of the persons installing and using the transducer, the ease of installation, the cost of the transducer, and the required maintenance. CALIBRATION REQUIREMENTS Our final consideration for selecting a transducer is the calibration requirements. The calibration interval and type of calibration necessary need to be considered. The calibration interval is determined by the operating life of the transducer and other factors such as long-term sensitivity shift, zero shift, and the accuracy requirements of the application. The calibration interval, complexity, and need to refer to calibration data can affect the total cost of the transducer. The ideal calibration will precisely predict the response of the transducer in the application setting. This prediction may be difficult, particularly when there are large differences between the calibration process and the application such as different loading, dynamic response, or environmental conditions. Calibration methods vary widely, depending on the transducer, but should be made in a manner that is traceable to NIST. The errors associated with the standard should be much smaller than the transducer that is being calibrated. The calibration can consist of comparing the transducer to a known reference instrument or a physical standard (such as a known mass for a load cell) or using a physical reference (such as the triple point of water for a temperature transducer). Frequently, the physical parameter is varied and the response of the transducer is observed and compared to a known reference. Generally, the specific calibration points should extend over the full range of the measurand to avoid the need for extrapolation. For example, a pressure transducer should be subjected to the same range of pressures in the calibration process that it will be subjected to in its intended application to reveal any nonlinearities or other problems. The data should be taken in both an increasing and decreasing direction to reveal any hysteresis present. Data taken during calibration is called a calibration record. A line connecting the data points is called a calibration curve for the particular transducer. If the calibration of the transducer is not done under the same conditions as the operating conditions, systematic error can result whenever the transducer is used. For example, a radiation transducer that is calibrated with a radio- active source that has a different spectrum than the spectrum that is to be monitored will lead to flawed data if the difference is not accounted for. A calibration performed in a manner that gives the transducer time for the output to settle to a fixed value is called a static calibration. Transducers used in dynamic measurements can be tested for their response with a dynamic calibration. A dynamic calibration is often a comparison of the transducer that is being calibrated with a known reference transducer that is faster. Another dynamic test is called a step-function response test. In it, the transducer is subjected to a rapid change in the input measurand, typically from 10% to 90% of the transducer's range. For example, a temperature transducer is very quickly moved to a much warmer or colder environment and its response is observed. The time required for the output to settle to the new temperature is a measure of the transducer's response time. EE Transducers 142

19 IX. BRIDGE MEASUREMENT Many transducers - such as strain gages, resistance temperature devices, and certain displacement transducers-use resistive transduction principles. Typically, these transducers produce a very small resistance change in response to the measurand. A circuit that can detect these very small changes is the Wheatstone bridge. Because of their great sensitivity and other advantages, bridge circuits are widely used for transducer measurements. A bridge circuit that consists of four resistance arms with a source of energy (called the excitation) and a detector is shown in Figure This circuit is shown with two pairs of leads to the power supply, a common method of ensuring that the voltage at the bridge is stable. Current to the bridge flows on the excitation leads. Any variation in the voltage due to IR drop in the excitation leads is detected by the sense leads and is used to regulate the supply. Notice that the sense leads carry almost no current, so they have almost no IR drop. Figure 13-4 The balance condition can be detected with great sensitivity using an instrumentation amplifier (IA) connected across the differential output, as shown in Figure For general application, the circuit shown is good choice (CMRR is typically greater than or equal to 110 db). The IA is selected for low-drift, high- common-mode rejection and gain stability. Figure 13-5 EE Transducers 143

20 When a Wheatstone bridge is used to detect the change in resistance of a single resistive transducer, the transducer is placed in one arm of the bridge, and the output of the bridge is observed. Frequently, more than one active transducer is used in a bridge to increase the sensitivity of the measurement. It is not always necessary to balance the bridge to determine the unknown resistance; instead, the magnitude of the off-balance condition can be used as an indicator of the resistance or to detect a change in resistance. The output change is not a linear function of the resistance change (it is approximately linear when Rl» R unknown ), but this non-linearity has some advantages in certain measurements. It is possible to construct bridges that tend to linearize the output of nonlinear transducers such as thermistors. Transducer manufacturers have developed a number of techniques for increasing the linearity, sensitivity, and stability of bridge circuits. These include specialized amplifiers, changing the excitation source, and matching the thermal or other characteristic of the transducer with the bridge resistors. For example, strain gages respond to more than just strain; they also respond to temperature changes. To cancel the temperature effects, a "dummy" gage can be placed in the same environment but not subjected to the strain of the measuring strain gage. The dummy gage is placed on the same side of the bridge as the measuring gage. Temperature effects change the resistance of both gages in a like manner, causing no change in the output. Many resistive transducers produce only a tiny change in resistance for a given input change. For sensitive measurements, the detecting instrument must have good common-mode rejection (CMR) because each side of the bridge includes a common-mode signal from the excitation supply. When the detecting instrument is located some distance from the transducer, an even more serious common-mode problem can occur when ground current causes a common-mode source to appear between the bridge ground and the instrument ground. Notice that the output (signal leads) of the bridge in Figures 13-4 and 13-5 are not connected to circuit ground, a condition referred to as a floating output. When a voltmeter is connected as a detecting instrument to the bridge outputs, the voltmeter signal connections should be isolated from circuit ground; however, there is always some high impedance to ground (Figure 13-6(a)). Assume that the impedance to ground is different between the high and low inputs (the usual case). Current from the common-mode ground source finds a path through the voltmeter's leads and generates a differential voltage due to the impedance difference in the return paths. This means that EE Transducers 144

21 a common-mode signal has been converted into a differential-mode signal, and the meter will respond to this potential. One possible solution is shown in Figure 13-6(b). The excitation supply is floating-in other words, its reference is isolated from the circuit ground by using an isolation transformer or a battery. This method can provide CMRRs greater than 160 db. The left side of the bridge is connected to circuit ground and the right side of the bridge is connected to a single-ended, non-inverting op-amp. Another solution is to use a guarded voltmeter. The guard connection is made so that it shunts common-mode current away from the meter's inputs. One of the best ways to do this is to connect the guard lead to a low-impedance point that is at the same potential as the low side of the meter. This is done by adding a low-impedance divider to the bridge, as illustrated in Figure 13-6(c). Most of the common-mode current will flow through the low-impedance path provided by the voltage divider. There are other variations of sensing circuits that are designed to optimize dc offset, voltage or temperature drift, non-linearity, noise performance, or other characteristic. EE Transducers 145

TRANSDUCERS transducer Measurand

TRANSDUCERS transducer Measurand TRANSDUCERS Transduction: transformation of one form of energy into another form. Sensing with specificity the input energy from the measurand by means of a "sensing element" and then transforming it into

More information

I. MEASUREMENT OF TEMPERATURE

I. MEASUREMENT OF TEMPERATURE I. MEASUREMENT OF TEMPERATURE Most frequent measurement and control Direct contact: thermometer, Indirect contact: pyrometer (detect generated heat or sensing optical properties) 1. Definition of temperature

More information

ME 515 Mechatronics. Overview of Computer based Control System

ME 515 Mechatronics. Overview of Computer based Control System ME 515 Mechatronics Introduction to Sensors I Asanga Ratnaweera Department of Faculty of Engineering University of Peradeniya Tel: 081239 (3627) Email: asangar@pdn.ac.lk Overview of Computer based Control

More information

INSTRUMENTATION ECE Fourth Semester. Presented By:- Sumit Grover Lect., Deptt. of ECE

INSTRUMENTATION ECE Fourth Semester. Presented By:- Sumit Grover Lect., Deptt. of ECE INSTRUMENTATION ECE Fourth Semester Presented By:- Sumit Grover Lect., Deptt. of ECE Detailed Contents Objectives Sensors and transducer Classification of transducers Temperature transducers Resistance

More information

Part 2. Sensor and Transducer Instrument Selection Criteria (3 Hour)

Part 2. Sensor and Transducer Instrument Selection Criteria (3 Hour) Part 2 Sensor and Transducer Instrument Selection Criteria (3 Hour) At the end of this chapter, you should be able to: Describe the definition of sensor and transducer Determine the specification of control

More information

Unit 3 Transducers. Lecture_3.1 Introduction to Transducers

Unit 3 Transducers. Lecture_3.1 Introduction to Transducers Unit 3 Transducers Lecture_3.1 Introduction to Transducers Introduction to transducers A transducer is a device that converts one form of energy to other form. It converts the measurand to a usable electrical

More information

Lecture 19. Measurement of Solid-Mechanical Quantities (Chapter 8) Measuring Strain Measuring Displacement Measuring Linear Velocity

Lecture 19. Measurement of Solid-Mechanical Quantities (Chapter 8) Measuring Strain Measuring Displacement Measuring Linear Velocity MECH 373 Instrumentation and Measurements Lecture 19 Measurement of Solid-Mechanical Quantities (Chapter 8) Measuring Strain Measuring Displacement Measuring Linear Velocity Measuring Accepleration and

More information

Transducers. ME 3251 Thermal Fluid Systems

Transducers. ME 3251 Thermal Fluid Systems Transducers ME 3251 Thermal Fluid Systems 1 Transducers Transform values of physical variables into equivalent electrical signals Converts a signal from one form to another form 2 Types of Transducers

More information

Sensors and Transducers. mywbut.com

Sensors and Transducers. mywbut.com Sensors and Transducers 1 Objectives At the end of this chapter, the students should be able to: describe the principle of operation of various sensors and transducers; namely.. Resistive Position Transducers.

More information

SENSORS AND TRANSDUCERS

SENSORS AND TRANSDUCERS Electrical Measurements International Program Department of Electrical Engineering UNIVERSITAS INDONESIA ANDRITTO ABDUL GHAFFAR ANDHIKA ADIEL INSANI Lecturer : Ir. Chairul Hudaya, ST, M.Eng., Ph.D., IPM

More information

e453.eps 1 Change (or the absolute value) in the measured physical variable 2 Change in the sensor property is translated into low-power-level

e453.eps 1 Change (or the absolute value) in the measured physical variable 2 Change in the sensor property is translated into low-power-level 3 Basic Phenomenon in Effect in Sensor Operation Sensors Prof. Dr. M. Zahurul Haq zahurul@me.buet.ac.bd http://teacher.buet.ac.bd/zahurul/ Department of Mechanical Engineering Bangladesh University of

More information

Slide 1. Temperatures Light (Optoelectronics) Magnetic Fields Strain Pressure Displacement and Rotation Acceleration Electronic Sensors

Slide 1. Temperatures Light (Optoelectronics) Magnetic Fields Strain Pressure Displacement and Rotation Acceleration Electronic Sensors Slide 1 Electronic Sensors Electronic sensors can be designed to detect a variety of quantitative aspects of a given physical system. Such quantities include: Temperatures Light (Optoelectronics) Magnetic

More information

Transducers. EEE355 Industrial Electronics

Transducers. EEE355 Industrial Electronics Transducers EEE355 Industrial Electronics 1 Terminology Transducers convert one form of energy into another Sensors/Actuators are input/output transducers Sensors can be passive (e.g. change in resistance)

More information

MCT151: Introduction to Mechatronics Lecture 10: Sensors & Transduction Mechanisms

MCT151: Introduction to Mechatronics Lecture 10: Sensors & Transduction Mechanisms Faculty of Engineering MCT151: Introduction to Mechatronics Lecture 10: Sensors & Transduction Mechanisms Slides are borrowed from Dr. Mohamed Elshiekh lectures Types of sensors Sensors are considered

More information

Transducer. A device to which change or converts physical quantity in a more easily measurable quantity. Transducer. (Input) Sensor.

Transducer. A device to which change or converts physical quantity in a more easily measurable quantity. Transducer. (Input) Sensor. Transducer A device to which change or converts physical quantity in a more easily measurable quantity Transducer (Input) Sensor (Output) Actuator Sensor A device which senses and detects the physical

More information

Overview. Sensors? Commonly Detectable Phenomenon Physical Principles How Sensors Work? Need for Sensors Choosing a Sensor Examples

Overview. Sensors? Commonly Detectable Phenomenon Physical Principles How Sensors Work? Need for Sensors Choosing a Sensor Examples Intro to Sensors Overview Sensors? Commonly Detectable Phenomenon Physical Principles How Sensors Work? Need for Sensors Choosing a Sensor Examples Sensors? American National Standards Institute A device

More information

ECE421: Electronics for Instrumentation MEP382: Design of Applied Measurement Systems Lecture #2: Transduction Mechanisms

ECE421: Electronics for Instrumentation MEP382: Design of Applied Measurement Systems Lecture #2: Transduction Mechanisms ECE421: Electronics for Instrumentation MEP382: Design of Applied Measurement Systems Lecture #2: Transduction Mechanisms Mostafa Soliman, Ph.D. April 28 th 2014 Slides are borrowed from Dr. Moahmed Elshiekh

More information

QUESTION BANK DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING UNIT I - INTRODUCTION SYLLABUS

QUESTION BANK DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING UNIT I - INTRODUCTION SYLLABUS QUESTION BANK DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING YEAR/SEM NAME OF THE SUBJECT NAME OF THE FACULTY : II / IV : EE6404 MEASUREMENTS AND INSTRUMENTATION : K.M.S.MUTHUKUMARA RAJAGURU, AP/EEE

More information

APPLICATIONS OF VIBRATION TRANSDUCERS

APPLICATIONS OF VIBRATION TRANSDUCERS APPLICATIONS OF VIBRATION TRANSDUCERS 1) Measurements on Structures or Machinery Casings: Accelerometers and Velocity Sensors Used in gas turbines, axial compressors, small and mid-size pumps. These sensors

More information

Temperature Measurement

Temperature Measurement MECE 3320 Measurements & Instrumentation Temperature Measurement Dr. Isaac Choutapalli Department of Mechanical Engineering University of Texas Pan American Introduction Temperature is one of the most

More information

Unit 57: Mechatronic System

Unit 57: Mechatronic System Unit 57: Mechatronic System Unit code: F/60/46 QCF level: 4 Credit value: 5 OUTCOME 2 TUTORIAL 2 - SENSOR TECHNOLOGIES 2 Understand electro-mechanical models and components in mechatronic systems and products

More information

Section 7. Temperature Measurement

Section 7. Temperature Measurement Section 7 Temperature Measurement 7/25/2017 Engineering Measurements 7 1 Working Definition Temperature is a measure of the average kinetic energy of the molecules that make of a substance. After time,

More information

Control Engineering BDA30703

Control Engineering BDA30703 Control Engineering BDA30703 Lecture 4: Transducers Prepared by: Ramhuzaini bin Abd. Rahman Expected Outcomes At the end of this lecture, students should be able to; 1) Explain a basic measurement system.

More information

e453.eps 1 Change (or the absolute value) in the measured physical variable 2 Change in the sensor property is translated into low-power-level

e453.eps 1 Change (or the absolute value) in the measured physical variable 2 Change in the sensor property is translated into low-power-level 3 Basic Phenomenon in Effect in Sensor Operation Measurement & Sensors Prof. Dr. M. Zahurul Haq http://teacher.buet.ac.bd/zahurul/ Department of Mechanical Engineering Bangladesh University of Engineering

More information

Set No. 1 1. (a) Differentiate among Desired, Modifying and Interfering inputs. (b) How do you eliminate the effects of interfering and modifying inputs? Explain 2. (a) Define the term Transducer and explain

More information

Lecture 20. Measuring Pressure and Temperature (Chapter 9) Measuring Pressure Measuring Temperature MECH 373. Instrumentation and Measurements

Lecture 20. Measuring Pressure and Temperature (Chapter 9) Measuring Pressure Measuring Temperature MECH 373. Instrumentation and Measurements MECH 373 Instrumentation and Measurements Lecture 20 Measuring Pressure and Temperature (Chapter 9) Measuring Pressure Measuring Temperature 1 Measuring Acceleration and Vibration Accelerometers using

More information

Theory and Design for Mechanical Measurements

Theory and Design for Mechanical Measurements Theory and Design for Mechanical Measurements Third Edition Richard S. Figliola Clemson University Donald E. Beasley Clemson University John Wiley & Sons, Inc. New York / Chichester / Weinheim / Brisbane

More information

ECNG3032 Control and Instrumentation I

ECNG3032 Control and Instrumentation I sensor ECNG3032 Control and Instrumentation I Lecture 1 Temperature Sensors Sensors The sensor is the first element in the measurement system. Measurand Transducer Principle Excitation Signal Interface

More information

PHYSICS (Theory) Time Allowed: 3 hours Maximum Marks: 70

PHYSICS (Theory) Time Allowed: 3 hours Maximum Marks: 70 Sample Paper (CBSE) Series PHY/SP/1D Code No. SP/1-D PHYSICS (Theory) Time Allowed: hours Maximum Marks: 70 General Instructions: (i) (ii) (iii) (iv) (v) All questions are compulsory. This question paper

More information

15. Compare the result with the value you have taken above Compare the calculated pressure value with the actual pressure value that you have

15. Compare the result with the value you have taken above Compare the calculated pressure value with the actual pressure value that you have 105) to convert it to. 15. Compare the result with the value you have taken above. 17. 16. Compare the calculated pressure value with the actual pressure value that you have taken from the, is it the same?

More information

Temperature. 3

Temperature. 3 Temperature In 1848, Sir William Thomson (Lord Kelvin) stated the zero principle of dynamics. This principle enabled him to define thermodynamic temperature and to establish an objective method of measuring

More information

Sensors and transducers

Sensors and transducers Sensors and transducers Measurement is an important subsystem of a mechatronics system. Its main function is to collect the information on system status and to feed it to the micro-processor(s) for controlling

More information

VALLIAMMAI ENGINEERING COLLEGE

VALLIAMMAI ENGINEERING COLLEGE VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur 603 203 DEPARTMENT OF ELECTRONICS AND INSTRUMENTATION ENGINEERING QUESTION BANK V SEMESTER EI6502 -INDUSTRIAL INSTRUMENTATION I Regulation 2013

More information

Module I Module I: traditional test instrumentation and acquisition systems. Prof. Ramat, Stefano

Module I Module I: traditional test instrumentation and acquisition systems. Prof. Ramat, Stefano Preparatory Course (task NA 3.6) Basics of experimental testing and theoretical background Module I Module I: traditional test instrumentation and acquisition systems Prof. Ramat, Stefano Transducers A

More information

Temperature Scales. Temperature, and Temperature Dependent on Physical Properties. Temperature. Temperature Scale

Temperature Scales. Temperature, and Temperature Dependent on Physical Properties. Temperature. Temperature Scale Temperature Scales The Celsius, Fahrenheit, and Kelvin Temperature Scales: Temperature, and Temperature Dependent on Physical Properties Physics Enhancement Programme Dr. M.H. CHAN, HKBU 9 T F T 5 T T

More information

REVISED HIGHER PHYSICS REVISION BOOKLET ELECTRONS AND ENERGY

REVISED HIGHER PHYSICS REVISION BOOKLET ELECTRONS AND ENERGY REVSED HGHER PHYSCS REVSON BOOKLET ELECTRONS AND ENERGY Kinross High School Monitoring and measuring a.c. Alternating current: Mains supply a.c.; batteries/cells supply d.c. Electrons moving back and forth,

More information

Overview. Sensors? Commonly Detectable Phenomenon Physical Principles How Sensors Work? Need for Sensors Choosing a Sensor Examples

Overview. Sensors? Commonly Detectable Phenomenon Physical Principles How Sensors Work? Need for Sensors Choosing a Sensor Examples Intro to Sensors Overview Sensors? Commonly Detectable Phenomenon Physical Principles How Sensors Work? Need for Sensors Choosing a Sensor Examples Sensors? American National Standards Institute A device

More information

Strain, Force, and Pressure

Strain, Force, and Pressure 10-1 10-1 Strain, Force, and Pressure Force is that which results in acceleration (when forces don t cancel). Strain is the change in shape of an object...... usually due to some force. (Force is usually

More information

Conventional Paper I-2010

Conventional Paper I-2010 Conventional Paper I-010 1. (a) Sketch the covalent bonding of Si atoms in a intrinsic Si crystal Illustrate with sketches the formation of bonding in presence of donor and acceptor atoms. Sketch the energy

More information

CHAPTER 5. BRIDGES AND THEIR APPLICATION Resistance Measurements. Dr. Wael Salah

CHAPTER 5. BRIDGES AND THEIR APPLICATION Resistance Measurements. Dr. Wael Salah CHAPTER 5 BRIDGES AND THEIR APPLICATION Resistance Measurements 1 RESISTANCE MEASUREMENTS Conventional Ways of Measuring Resistance:- 1) Using a Ohmmeter Convenient but inaccurate, requires calibration

More information

Measurements in Mechatronic design. Transducers

Measurements in Mechatronic design. Transducers Measurements in Mechatronic design Transducers Quantities Current Voltage Torque Force Magnetic flux Distance Temperature Measurement system Physical quanties Transducer Signal conditioning Measurement

More information

1. Mark the correct statement(s)

1. Mark the correct statement(s) 1. Mark the correct statement(s) Figure to the right shows a mass measurement scale using a spring. 1.1 The span of the scale is a) 16 kg b) 21 kg c) 11 kg d) 5-16 kg 1.2 The range of the scale is a) 16

More information

Force and Displacement Measurement

Force and Displacement Measurement Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 20 Simple ways to measure a force http://scienceblogs.com/dotphysics/200/02/diy_force_probe.php Example: Key Force/Deflection measure

More information

Experimentation. Third Edition. Wheeler. Ahmad R. Ganji. School of Engineering. With Third Edition contributions by. V. V. Krishnan. Brian S.

Experimentation. Third Edition. Wheeler. Ahmad R. Ganji. School of Engineering. With Third Edition contributions by. V. V. Krishnan. Brian S. Introduction to Engineering Experimentation Third Edition Anthony J. Ahmad R. Wheeler Ganji School of Engineering San Francisco State University With Third Edition contributions by V. V. Krishnan San Francisco

More information

2. (a) Differentiate between rare metal thermocouples and base metal thermocouples.

2. (a) Differentiate between rare metal thermocouples and base metal thermocouples. Code No: R05410304 Set No. 1 1. (a) Distinguish between direct and indirect methods of measurement with suitable examples. (b) What are desired, modifying and interfering inputs for an instrumentation

More information

INSTITUTE OF AERONAUTICAL ENGINEERING (Autonomous) Dundigal, Hyderabad

INSTITUTE OF AERONAUTICAL ENGINEERING (Autonomous) Dundigal, Hyderabad INSTITUTE OF AERONAUTICAL ENGINEERING (Autonomous) Dundigal, Hyderabad - 500 04 ELECTRONICS AND COMMUNICATION ENGINEERING Name : Electronic Measurements and Instrumentation Code : A50422 Class : III -

More information

Sensing, Computing, Actuating

Sensing, Computing, Actuating Sensing, Computing, ctuating Sander Stuijk (s.stuijk@tue.nl) Department of Electrical Engineering Electronic Systems 2 THERMOELECTRIC EFFECT (Chapter 5.11) 3 Thermocouple cylinder head temperature (thermocouple)

More information

Strain Measurement. Prof. Yu Qiao. Department of Structural Engineering, UCSD. Strain Measurement

Strain Measurement. Prof. Yu Qiao. Department of Structural Engineering, UCSD. Strain Measurement Strain Measurement Prof. Yu Qiao Department of Structural Engineering, UCSD Strain Measurement The design of load-carrying components for machines and structures requires information about the distribution

More information

Course Name: Sensor and Transducer Course Code: EE 802B Credit: 3

Course Name: Sensor and Transducer Course Code: EE 802B Credit: 3 Course Name: Sensor and Transducer Course Code: EE 802B Credit: 3 Prerequisites: Sl. No. Subject Description Level of Study 01 Basic Electrical & Electronics Engineering 02 Electrical & Electronics Measurement

More information

Resistance and Conductance

Resistance and Conductance 1 2 1 Resistance and Conductance Resistance, R (Ohm ), is the tendency of a material to impede the flow of electric charges through it. The instantaneous voltage across a resistor is directly proportional

More information

ELECTRONIC SENSORS PREAMBLE. This note gives a brief introduction to sensors. The focus is. on sensor mechanisms. It describes in general terms how

ELECTRONIC SENSORS PREAMBLE. This note gives a brief introduction to sensors. The focus is. on sensor mechanisms. It describes in general terms how ELECTRONIC SENSORS PREAMBLE This note gives a brief introduction to sensors. The focus is on sensor mechanisms. It describes in general terms how sensors work. It covers strain gage sensors in detail.

More information

1 P a g e h t t p s : / / w w w. c i e n o t e s. c o m / Physics (A-level)

1 P a g e h t t p s : / / w w w. c i e n o t e s. c o m / Physics (A-level) 1 P a g e h t t p s : / / w w w. c i e n o t e s. c o m / Physics (A-level) Electromagnetic induction (Chapter 23): For a straight wire, the induced current or e.m.f. depends on: The magnitude of the magnetic

More information

COURSE OF Prepared By: MUHAMMAD MOEEN SULTAN Department of Mechanical Engineering UET Lahore, KSK Campus

COURSE OF Prepared By: MUHAMMAD MOEEN SULTAN Department of Mechanical Engineering UET Lahore, KSK Campus COURSE OF Active and passive instruments Null-type and deflection-type instruments Analogue and digital instruments In active instruments, the external power source is usually required to produce an output

More information

INSTITUTE OF AERONAUTICAL ENGINEERING Dundigal, Hyderabad DEPARTMENT OF ECE QUESTION BANK. : G.Lakshminarayana, Asst.

INSTITUTE OF AERONAUTICAL ENGINEERING Dundigal, Hyderabad DEPARTMENT OF ECE QUESTION BANK. : G.Lakshminarayana, Asst. ` INSTITUTE OF AERONAUTICAL ENGINEERING Dundigal, Hyderabad - 500 04 DEPARTMENT OF ECE QUESTION BANK Name Code Class Branch P a g e : Electronic Measurements and Instrumentation : A504 : III - B. Tech

More information

Biosensors and Instrumentation: Tutorial 2

Biosensors and Instrumentation: Tutorial 2 Biosensors and Instrumentation: Tutorial 2. One of the most straightforward methods of monitoring temperature is to use the thermal variation of a resistor... Suggest a possible problem with the use of

More information

Using a Mercury itc with thermocouples

Using a Mercury itc with thermocouples Technical Note Mercury Support Using a Mercury itc with thermocouples Abstract and content description This technical note describes how to make accurate and reliable temperature measurements using an

More information

PANDIAN SARASWATHI YADAV ENGINEERING COLLEGE DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING EE6404-MEASUREMENTS AND INSTRUMENTATION

PANDIAN SARASWATHI YADAV ENGINEERING COLLEGE DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING EE6404-MEASUREMENTS AND INSTRUMENTATION PANDIAN SARASWATHI YADAV ENGINEERING COLLEGE DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING EE6404-MEASUREMENTS AND INSTRUMENTATION ACADEMIC YEAR: 2015-2016 (EVEN SEMESTER) Branch: EEE QUESTION BANK

More information

5. You may use the following values of physical constants wherever necessary. Class XII Physics (042) Sample Question Paper

5. You may use the following values of physical constants wherever necessary. Class XII Physics (042) Sample Question Paper Class XII Physics (04) Sample Question Paper 018-19 Time allowed: hours. Max. Marks: 70 General Instructions: 1. All questions are compulsory. There are 7 questions in all.. This question paper has four

More information

Physical Phenomena Used in Sensors

Physical Phenomena Used in Sensors Physical Phenomena Used in Sensors Tiantian Xie Auburn University Bogdan M. Wilamowski Auburn University. Introduction... -. Piezoresistive Effect... -.3 Thermoelectric Effect...-5.4 Piezoelectric Effect...-5.5

More information

Chapter 6. Fiber Optic Thermometer. Ho Suk Ryou

Chapter 6. Fiber Optic Thermometer. Ho Suk Ryou Chapter 6. Fiber Optic Thermometer Ho Suk Ryou Properties of Optical Fiber Optical Fiber Composed of rod core surrounded by sheath Core: conducts electromagnetic wave Sheath: contains wave within the core

More information

Units (Different systems of units, SI units, fundamental and derived units)

Units (Different systems of units, SI units, fundamental and derived units) Physics: Units & Measurement: Units (Different systems of units, SI units, fundamental and derived units) Dimensional Analysis Precision and significant figures Fundamental measurements in Physics (Vernier

More information

Physics (Theory) There are 30 questions in total. Question Nos. 1 to 8 are very short answer type questions and carry one mark each.

Physics (Theory) There are 30 questions in total. Question Nos. 1 to 8 are very short answer type questions and carry one mark each. Physics (Theory) Time allowed: 3 hours] [Maximum marks:70 General Instructions: (i) All questions are compulsory. (ii) (iii) (iii) (iv) (v) There are 30 questions in total. Question Nos. to 8 are very

More information

Revision Guide for Chapter 2

Revision Guide for Chapter 2 Revision Guide for Chapter 2 Contents Student s Checklist Revision Notes Electric current and charge... 5 Potential difference... 5 Conductance and resistance... 6 Parallel circuit... 7 Series circuit...

More information

SENSORS and TRANSDUCERS

SENSORS and TRANSDUCERS SENSORS and TRANSDUCERS Tadeusz Stepinski, Signaler och system The Thermal Energy Domain Physics» Seebeck effect» Peltier effect» Thomson effect Thermal effects in semiconductors Thermoelectric sensors

More information

STRAIN GAUGES YEDITEPE UNIVERSITY DEPARTMENT OF MECHANICAL ENGINEERING

STRAIN GAUGES YEDITEPE UNIVERSITY DEPARTMENT OF MECHANICAL ENGINEERING STRAIN GAUGES YEDITEPE UNIVERSITY DEPARTMENT OF MECHANICAL ENGINEERING 1 YEDITEPE UNIVERSITY ENGINEERING FACULTY MECHANICAL ENGINEERING LABORATORY 1. Objective: Strain Gauges Know how the change in resistance

More information

UMEÅ UNIVERSITY Department of Physics Agnieszka Iwasiewicz Leif Hassmyr Ludvig Edman SOLID STATE PHYSICS HALL EFFECT

UMEÅ UNIVERSITY Department of Physics Agnieszka Iwasiewicz Leif Hassmyr Ludvig Edman SOLID STATE PHYSICS HALL EFFECT UMEÅ UNIVERSITY Department of Physics 2004-04-06 Agnieszka Iwasiewicz Leif Hassmyr Ludvig Edman SOLID STATE PHYSICS HALL EFFECT 1. THE TASK To measure the electrical conductivity and the Hall voltage for

More information

Temperature Measurements

Temperature Measurements ME 22.302 Mechanical Lab I Temperature Measurements Dr. Peter Avitabile University of Massachusetts Lowell Temperature - 122601-1 Copyright 2001 A transducer is a device that converts some mechanical quantity

More information

10 Measurement of Acceleration, Vibration and Shock Transducers

10 Measurement of Acceleration, Vibration and Shock Transducers Chapter 10: Acceleration, Vibration and Shock Measurement Dr. Lufti Al-Sharif (Revision 1.0, 25/5/2008) 1. Introduction This chapter examines the measurement of acceleration, vibration and shock. It starts

More information

Chapter 27. Current and Resistance

Chapter 27. Current and Resistance Chapter 27 Current and Resistance Electric Current Most practical applications of electricity deal with electric currents. The electric charges move through some region of space. The resistor is a new

More information

COURSE OUTLINE. Introduction Signals and Noise Filtering Sensors: Piezoelectric Force Sensors. Sensors, Signals and Noise 1

COURSE OUTLINE. Introduction Signals and Noise Filtering Sensors: Piezoelectric Force Sensors. Sensors, Signals and Noise 1 Sensors, Signals and Noise 1 COURSE OUTLINE Introduction Signals and Noise Filtering Sensors: Piezoelectric Force Sensors Piezoelectric Force Sensors 2 Piezoelectric Effect and Materials Piezoelectric

More information

BME/EECS Biomedical Instrumentation & Design Matt O Donnell

BME/EECS Biomedical Instrumentation & Design Matt O Donnell BME/EECS 458 - Biomedical Instrumentation & Design Matt O Donnell 2.0 Introduction The front-end of all biomedical instruments includes either a sensor or actuator, or both. In this course we will focus

More information

SENSORS and TRANSDUCERS

SENSORS and TRANSDUCERS SENSORS and TRANSDUCERS Tadeusz Stepinski, Signaler och system The Mechanical Energy Domain Physics Surface acoustic waves Silicon microresonators Variable resistance sensors Piezoelectric sensors Capacitive

More information

INSTITUTE OF AERONAUTICAL ENGINEERING (Autonomous) Dundigal, Hyderabad

INSTITUTE OF AERONAUTICAL ENGINEERING (Autonomous) Dundigal, Hyderabad INSTITUTE OF AERONAUTICAL ENGINEERING (Autonomous) Dundigal, Hyderabad - 500 043 ELECTRONICS AND COMMUNICATION ENGINEERING Course Name : Electronic Measurements and Instrumentation Course Code : A50422

More information

Glossary Innovative Measurement Solutions

Glossary Innovative Measurement Solutions Glossary GLOSSARY OF TERMS FOR TRANSDUCERS, LOAD CELLS AND WEIGH MODULES This purpose of this document is to provide a comprehensive, alphabetical list of terms and definitions commonly employed in the

More information

INC 331 Industrial Process Measurement. Instrument Characteristics

INC 331 Industrial Process Measurement. Instrument Characteristics INC 331 Industrial Process Measurement Instrument Characteristics Introduction Measurement is the experimental process of acquiring any quantitative information. When doing a measurement, we compare the

More information

Module 3 Electrical Fundamentals

Module 3 Electrical Fundamentals 3.1 Electron Theory Structure and distribution of electrical charges within: atoms, molecules, ions, compounds; Molecular structure of conductors, semiconductors and insulators. 3.2 Static Electricity

More information

Electronic Circuits Summary

Electronic Circuits Summary Electronic Circuits Summary Andreas Biri, D-ITET 6.06.4 Constants (@300K) ε 0 = 8.854 0 F m m 0 = 9. 0 3 kg k =.38 0 3 J K = 8.67 0 5 ev/k kt q = 0.059 V, q kt = 38.6, kt = 5.9 mev V Small Signal Equivalent

More information

5. ELECTRIC CURRENTS

5. ELECTRIC CURRENTS 5. ELECTRIC CURRENTS TOPIC OUTLINE Section Recommended Time Giancoli Section 5.1 Potential Difference, Current, Resistance 5.2 Electric Circuits 3h 19.1, 19.2 6.2 Electric Field and Force 6.3 Magnetic

More information

Thermal Sensors and Actuators

Thermal Sensors and Actuators Thermal Sensors and Actuators Part I Fundamentals of heat transfer Heat transfer occurs where there is a temperature gradient until an equilibrium is reached. Four major mechanism Thermal conduction Natural

More information

Objective Type Questions Instrumentation System & Devices (IDS)

Objective Type Questions Instrumentation System & Devices (IDS) 1. A balance beam scale uses which of the following units? a. grams b.pounds c. ounces d. kilograms 2. Which of the following would be about the height of the average doorway? a. 2 meters b. 2 centimeters

More information

Bridge Measurement 2.1 INTRODUCTION Advantages of Bridge Circuit

Bridge Measurement 2.1 INTRODUCTION Advantages of Bridge Circuit 2 Bridge Measurement 2.1 INTRODUCTION Bridges are often used for the precision measurement of component values, like resistance, inductance, capacitance, etc. The simplest form of a bridge circuit consists

More information

1 THE CONCEPT OF TEMPERATURE

1 THE CONCEPT OF TEMPERATURE 1 THE CONCEPT OF TEMPERATURE 1 1.1 Historical Perspective, 2 1.2 Early Definitions of Temperature, 9 1.3 A Simple Qualitative Definition of Temperature, 10 1.4 Units of Temperature for Various Temperature

More information

coil of the circuit. [8+8]

coil of the circuit. [8+8] Code No: R05310202 Set No. 1 III B.Tech I Semester Regular Examinations, November 2008 ELECTRICAL MEASUREMENTS (Electrical & Electronic Engineering) Time: 3 hours Max Marks: 80 Answer any FIVE Questions

More information

Semiconductor Physics

Semiconductor Physics Semiconductor Physics Motivation Is it possible that there might be current flowing in a conductor (or a semiconductor) even when there is no potential difference supplied across its ends? Look at the

More information

PRE-BOARD EXAMINATION STD : XII MARKS : 150

PRE-BOARD EXAMINATION STD : XII MARKS : 150 PRE-BOARD EXAMINATION STD : XII MARKS : 150 SUB : PHYSICS TIME : 3.00 Hrs I.Choose the correct answer: 30x1=30 1.Which of the following quantities not a scalar? a)electric flux b) electric potential c)

More information

Outline of College Physics OpenStax Book

Outline of College Physics OpenStax Book Outline of College Physics OpenStax Book Taken from the online version of the book Dec. 27, 2017 18. Electric Charge and Electric Field 18.1. Static Electricity and Charge: Conservation of Charge Define

More information

ELECTRICITY AND MAGNETISM, A. C. THEORY AND ELECTRONICS, ATOMIC AND NUCLEAR PHYSICS

ELECTRICITY AND MAGNETISM, A. C. THEORY AND ELECTRONICS, ATOMIC AND NUCLEAR PHYSICS UNIT 2: ELECTRICITY AND MAGNETISM, A. C. THEORY AND ELECTRONICS, ATOMIC AND NUCLEAR PHYSICS MODULE 1: ELECTRICITY AND MAGNETISM GENERAL OBJECTIVES On completion of this Module, students should: 1. understand

More information

EC Objective Paper I (Set - D)

EC Objective Paper I (Set - D) EC-Objective Paper-I ESE-5 www.gateforum.com EC Objective Paper I (Set - D). If a system produces frequencies in the output are not present in the input, then the system cannot be Minimum phase system

More information

Module 2. Measurement Systems. Version 2 EE IIT, Kharagpur 1

Module 2. Measurement Systems. Version 2 EE IIT, Kharagpur 1 Module 2 Measurement Systems Version 2 EE IIT, Kharagpur 1 Lesson 8 Measurement of Level, Humidity and ph Version 2 EE IIT, Kharagpur 2 Instructional Objectives At the end of this lesson, the student will

More information

III B.Tech. II Semester Regular Examinations, April/May INSTRUMENTATION & CONTROL SYSTEMS (Mechanical Engineering) Time: 3 Hours Max Marks: 75

III B.Tech. II Semester Regular Examinations, April/May INSTRUMENTATION & CONTROL SYSTEMS (Mechanical Engineering) Time: 3 Hours Max Marks: 75 R10 Set No: 1 1. (a) Distinguish between accuracy and Precision. Which of these is more desirable during the act of measurement and why? (b) Discuss the necessity and importance of dynamic performance

More information

Most matter is electrically neutral; its atoms and molecules have the same number of electrons as protons.

Most matter is electrically neutral; its atoms and molecules have the same number of electrons as protons. Magnetism Electricity Magnetism Magnetic fields are produced by the intrinsic magnetic moments of elementary particles associated with a fundamental quantum property, their spin. -> permanent magnets Magnetic

More information

An Electronic Thermal Transducer

An Electronic Thermal Transducer An Electronic Thermal Transducer PRJ NO: 16 By Okore Daniel F17/969/ Supervisor : Dr. Mwema Examiner: Dr. Abungu INTRODUCTION The aim of this project was to design and implement a temperature sensor in

More information

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626 OPTI510R: Photonics Khanh Kieu College of Optical Sciences, University of Arizona kkieu@optics.arizona.edu Meinel building R.626 Announcements Homework #6 is assigned, due May 1 st Final exam May 8, 10:30-12:30pm

More information

Subject: BT6008 Process Measurement and Control. The General Control System

Subject: BT6008 Process Measurement and Control. The General Control System WALJAT COLLEGES OF APPLIED SCIENCES In academic partnership with BIRLA INSTITUTE OF TECHNOLOGY Question Bank Course: Biotechnology Session: 005-006 Subject: BT6008 Process Measurement and Control Semester:

More information

Solved Problems. Electric Circuits & Components. 1-1 Write the KVL equation for the circuit shown.

Solved Problems. Electric Circuits & Components. 1-1 Write the KVL equation for the circuit shown. Solved Problems Electric Circuits & Components 1-1 Write the KVL equation for the circuit shown. 1-2 Write the KCL equation for the principal node shown. 1-2A In the DC circuit given in Fig. 1, find (i)

More information

CBSE Physics Set I Outer Delhi Board 2012

CBSE Physics Set I Outer Delhi Board 2012 Q28. a) In Young s double slit experiment, derive the condition for (I) constructive interference and (II) destructive interference at a point on the screen. b) A beam of light consisting of two wavelengths,

More information

CBSE_2014_SET_3 Physics

CBSE_2014_SET_3 Physics CBSE_2014_SET_3 Physics 1. A conducting loop is held below a current carrying wire PQ as shown. Predict the direction of the induced current in the loop when the current in the wire is constantly increasing.

More information

Cryogenic Instrumentation I Thermometry OUTLINE Thermometry Pt (pure metal) Temperature Ranges of Thermometer Application Typical Resistive Thermal

Cryogenic Instrumentation I Thermometry OUTLINE Thermometry Pt (pure metal) Temperature Ranges of Thermometer Application Typical Resistive Thermal Cryogenic Instrumentation I 1. Thermometry 2. anges of Application 3. Constant Volume 4. Thermocouples 5. Time esponse Data 6. 4 Terminal esistance Measurement OUTLINE 8. Pt (pure metal) 9. Typical esistive

More information

THERMOCOUPLE CHARACTERISTICS TRAINER

THERMOCOUPLE CHARACTERISTICS TRAINER THERMOCOUPLE CHARACTERISTICS TRAINER (Model No : ) User Manual Version 2.0 Technical Clarification /Suggestion : / Technical Support Division, Vi Microsystems Pvt. Ltd., Plot No :75,Electronics Estate,

More information

Summary Notes ALTERNATING CURRENT AND VOLTAGE

Summary Notes ALTERNATING CURRENT AND VOLTAGE HIGHER CIRCUIT THEORY Wheatstone Bridge Circuit Any method of measuring resistance using an ammeter or voltmeter necessarily involves some error unless the resistances of the meters themselves are taken

More information