Characterization of plasma-treated surfaces by X-ray Photoelectron Spectroscopy (XPS)

Size: px
Start display at page:

Download "Characterization of plasma-treated surfaces by X-ray Photoelectron Spectroscopy (XPS)"

Transcription

1 Characterization of plasma-treated surfaces by X-ray Photoelectron Spectroscopy (XPS) Teresa de los Arcos September 13, 2010 X-Ray photoelectron spectroscopy (XPS) is a technique used to determine the elemental composition and chemical state (among other things) of surfaces. The goal of this practical course is to use XPS in order to gain information about the changes induced in a surface due to dierent plasma treatments. Due to time constrictions, the samples necessary for the execution of this experiment must have been previously prepared in other plasma-related experiments. At the moment, this experiment can be combined with the following f-praktika: Oberächen Modikation mit Mikroplasmen (Dr. Volker Schulz von der Gathen) 402 Anwendungsorientierte Plasmaphysik (Dr. Marc Böke) 1

2 1 Theory X-ray photoelectron spectroscopy (XPS) is an extremely powerful technique for the analysis of the chemical composition and electronic characteristics of surfaces. The technique is based in the photoelectric eect, where electrons are emitted by a material after irradiation with photons. The kinetic energy of the emitted electrons can be related to the energy of the photons by the following formula: hν = E kin + E b + Φ (1) Where hν is the energy of the incoming photon; E kin is the kinetic energy of the emitted electron, which we measure; Φ is the work function of the sample (the extra energy that the electron needs in order to physically leave the solid); and E b is the binding energy of the electron in the material, which is what we want to determine. Therefore, if we irradiate the sample with photons of known energy, we can determine the original binding energy of the electron by simply measuring the kinetic energy. Since each atom has its own distribution of core energy levels, it is possible to use XPS to determine the elemental composition of a sample surface (Si, C, O, etc). Furthermore, the precise binding energy position is also strongly inuenced by the chemical environment of the atom, and therefore, XPS can be used also to determine quantitatively the precise chemical composition of each element (Si, SiO 2, etc.) Photoelectron spectroscopy is a Ultra-High-Vacuum (UHV) methode, and we will operate within a pressure range of Pa. The low pressure is neccesary in order to: avoid scattering of the photoelectrons between the sample surface and the detector, avoid arcing between the high voltage components of the electronic, keep the sample surface clean. The mean free path of the photoelectrons within the sample is relatively short. This means that, although we will produce photoelectrons at several microns depth within the sample, only those electrons originating from the top few atomic layers will be able to actually leave the sample without being reabsorbed. This fact makes XPS extremely surface sensitive, and we will only be able to investigate the chemical composition of approximately the rst 5 nm of the surface. This means that if the sample surface is covered with a contaminant layer of adsorbed species (for example, from contact with the air), we will always measure this contaminant layer. If this layer is thicker than 5 nm we will not be able to see the surface of interest. XPS is therefore a unique surface sensitive technique for chemical analysis. Determination of a binding energy reference: We have seen that according to equation 1, we need to know the value of the work function of the sample in order to determine the binding energy of the electrons. However, from an experimental point of 2

3 view, it is very dicult to determine simultaneously a binding energy reference and the work function of the sample. The solution employed in modern spectrometers is to use as a reference, not the work function of the sample under study, but the work function of the spectrometer. What we do is to put the sample in good electric contact with the spectrometer, so that the two Fermi levels will be aligned. Our value for binding energy zero will be the position of the Fermi Level of both sample and spectrometer. The problem of the determination of the surface work function is also solved, since we can now use in our formula the known constant value of the work function of the spectrometer. The computer will perform this calculation automatically. We will obtain in our computer screen already the converted binding energy values of the photoelectrons. You can see an illustration if this in Figure 1. E kin Ekin Efot Vac Level sample spect Fermi Level Ebin Figure 1: Illustration of Fermi-level alignment of a metallic sample and the spectrometer. In certain cases it is not possible to have a good electric contact between the sample and the spectrometer. This is going to be the case when the samples are not good conductors. In this case there are two eects that we need to take into account: sample charging and a oating Fermi Level for the sample with respect to the spectrometer. Sample charging: A non conducting sample does not have sucient delocalized conduction band electrons available in order to neutralize the positive holes left behind during the photoeect. As a result, a positive potential builds near the sample surface, which retards the outgoing electrons, making them lose some of their original kinetic energy. This 3

4 retardation appears in the spectrum as a positive shift towards higher binding energies. This eect does not occur in good conducting samples or in narrow gap semiconductors, because at room temperature these systems have enough conduction electrons to neutralize the positive charge at the surface. The solution to this problem is to neutralize the positive charge externally. In order to do this, we will irradiate the sample simultaneously with a shower of low energy electrons (to neutralize the positive charge) and low energy ions (to improve the homogeneity of the neutralization). Failure to compensate the charge in this way results not only in extreme shifts in binding energy positions, but also in deformation of the peak shape. Fermi Level alignment: A serious problem is the establishment of a Fermi Level position, or zero reference position for the energy scala. This problem is inherent to the way in which modern spectrometers perform the measurement, and cannot be solved by the neutralization procedure described above. As we can see in the Figure 1, in the case of metallic samples the establishment of a reference energy level is a problem because the Fermi levels of sample and spectrometer align themselves automatically. In the case of a non-conducting sample this alignment does not take place, becuase the sample and the spectrometer are not in electrical contact. In these cases, the best procedure is to establish an arbitrary reference that will serve to x the scala. In the case of polymer samples, we will do this by xing the C1s core level of C-C bonding at a binding energy of exactly ev. In the general case, we will use the C1s signal corresponding to adventitious carbon (an undened carbon-containing layer that appears on the surface on every sample exposed to the air due to adsorption of dierent molecules), which we will x at an arbitrary value of ev. 1.1 The spectrometer The spectrometer that we will use for this experiment is a Versaprobe model from Physical Electronics, whose schema we can see in the Figure 2. It is basically divided in three parts: X-ray source, spherical analyzer and detector. X-ray source: An electron beam with an energy of 15 kv impinges onto an Al anode. There is production of X-ray radiation due to the excitation and radiative relaxation of internal Al core levels. This X-ray radiation passes through a monochromator. Finally we obtain a monochromatized beam of photons with energy of ev, corresponding to the Al K α transition. The diameter of the X-ray beam is 100 µm. Spherical Analyzer: The core of the spectrometer is the spherical analyzer, which performs electron separation according to their kinetic energy. The analyzer is composed of two concentric half spheres, separated by a thin space. The entrance slit is at one side, and the detector is located at the other side. The electrons emitted from the sample have to travel the thin space between the hemispheres in order to reach the detector. By applying a potential dierence between the two hemispheres we create a radial electric 4

5 Energy Analizer Electron Source #1 15kV Monochromator Al K x-rays eV Multi-channel detector Photoelectrons Al Anode Sample Figure 2: Schema of the spectrometer. eld that changes the initial straight path of the electrons into a curved one. Only electrons with a very specic kinetic energy will be able to reach the detector without either colliding with the outer hemisphere (if they are too fast) or the inner (if they are too slow). This particular energy which allows the electrons to reach the detector is called the Pass Energy (E 0 ). The relationship between energy resolution ( E) and pass energy for our spectrometer is given by: E = 0.015E 0. A higher pass energy value corresponds to higher amounts of electrons going through, which means higher intensity in our spectrum. High intensity, however, comes at the cost of spectral resolution. We will use high pass energy values when we want to have an overview of all possible elements present in the sample; we will use low pass energy values when it is necessary to perform detailed analysis of particular core level peaks. Detector: The electrons are nally gathered in a multi-channel detector. It is composed by two parallel plates through which there are 16 channels. Each channel acts as an electron multiplier. Between the two plates we establish a voltage of 2 kv that accelerates the electrons going through the rst plate into the second plate. 2 Procedure In the case of a follow-up of the f-praktikum Oberächen Modikation mit Mikroplasmen (Dr. Volker Schulz von der Gathen) you will investigate the changes induced by the 5

6 plasma treatment on polymers, or on metal surfaces. In this case, we will investigate one un-treated sample and one treated sample in order to establish the chemical changes induced in the surface due to the plasma treatment. In the case of a follow-up to the f-praktikum 402 Anwendungsorientierte Plasmaphysik (Dr. Marc Böke) you will perform a comparative analysis of the bare silicon surface and of the deposited lm. The experimental procedure will be as follows: 1. Sample preparation for the measurement. 2. Low resolution measurement (survey spectrum). 3. High resolution measurements of the relevant core levels. 4. Mathematical analysis of the data and discussion. 2.1 Introduction of the sample in the spectrometer It is important that the samples are never handled with bare hands, because of two reasons: to avoid covering the surface of interest with a fat-layer, and to avoid the degassing of fat molecules in the UHV environment. After cutting the sample to an appropriate size, you will mount it on a sample holder and introduce the sample holder in the interchange lock. You need to use the computer in order to vent this interchange lock. The hardware of the spectrometer is controlled by a single program: PHI SUMMIT. When you click into this icon for the rst time, three windows will open: Watcher, Acquisition settings, and Image. Watcher will allow you to control the transfer of the sample and the venting and pumping down procedures. Image will allow you to control the hardware necessary for the measurement: sample position, neutralizer, ion gun, X-ray power, etc. Acquisition setting will allow you to dene your measurement conditions. How to introduce the sample: First you will introduce your sample in the interchange lock (labelled intro in the window Watcher): 1. Open the N 2 bottle. In the window Watcher, press backll intro. This will vent the small chamber on the side of the spectrometer. CAUTION: The top cover of this chamber is not xed. It can be blown away if the venting pressure is set too high! As a precaution, always vent with your hand on top of this cover. 2. After mounting the sample in the fork of the manipulator, cover the chamber again and press pump intro. You will have to wait ve minutes for the pressure to reach a good value before transferring the sample to the UHV chamber. 6

7 3. In window Image, go to Stage and press the buttom: intro. This will move the sample stage holder (in the UHV chamber) to the appropriate position for transferring the sample. 4. In window Watcher, press transfer sample. Follow the instructions in the screen. Once the sample holder is secured in the stage within the UHV chamber, remove carefully the manipulator until the end (when you reach the end position, valve V1 will close automatically.) 2.2 Spectrometer start-up SECURITY CONSIDERATIONS. For your security: Be careful. You are going to operate a machine which produces X-ray radiation by means of accelerating electron beams with high voltage (15kV). The electron detector operates also with a voltage of 2kV. For the security of the machine: The XPS spectrometer operates in ultra high vacuum (UHV). It is important that you always monitor the pressure before starting any procedure that will switch on the X-ray tube. If the pressure is higher than Pa do not proceed with the measurement. Once your sample is within the UHV chamber, you need to nd the right position for the measurement. 1. In window Image, go to Stage and select a value for the z-position of 16.5 mm. Press move. This will move the sample to a good starting position. Now you have to ne-tune the z-position. 2. By pressing the arrow-buttons in the small box in the screen, move the sample holder and select with help of the camera the position where you want to measure. 3a. If you are measuring conducting samples: In window Image press auto-z. The software will nd the right height for the best measurement. Then go to Acquisition Settings. 3b. If you are measuring polymers or other non-conducting samples (such as an oxide layer): you will need to use neutralizing hardware. This implies preparing a source of low energy electrons (with the neutralizer) and a source of low energy Ar ions (with the ion gun): 1. First, we prepare the neutralizer. In window Image, go to Neutralizer. If the pressure is in the range of 10 7 Pa, press Standby. 7

8 2. Now we prepare the ion gun. (Note that this is also the procedure you have to follow if you want to use the ion gun in order to sputter the sample). In window Watcher press Di vlv Open. This will open Valve V4 and close valve V3. VERY IMPORTANT! Never open the valve that supplies Ar to the ion gun until you are sure that valve V4 is open. The software will not warn you if you make this mistake! If valve V4 is closed when you put Ar to the ion gun, the Ar will enter into the UHV chamber and the pressure will increase without control until eventually the security measures of the machine will kick in and all the electronics will be shut o. 3. In window Image, go to Ion gun. Press Standby. The program will ask you if Valve V4 is open. If you have opened it in the previous step, say Ok (if not, open it now). Now select the extractor pressure box. 4. Open slowly the manual valve (green color) of the ion Gun, controlling simultaneously the main pressure and the extractor pressure. Open the valve until the extractor pressure shows a value of 30 mpa. At the same time, the pressure in the chamber should not be higher than Pa. If the pressure in the main chamber increases over this value without stabilizing, interrupt the procedure immediately and close the valve. 5. Switch on the valve controller RVG050C (in the rack) with the on/o black lever. Now put the white lever limit/set point in the set point position. Wait until the extractor pressure stabilizes at 10 mpa. You can now decide if you want to use the ion gun in the sputtering mode or in the neutralize mode. Deselect now extractor pressure (the continuous measurement of this pressure is not really necessary and it takes too many resources from the computer). 6. After preparing the neutralizing software, go to window Image, SXI/Auto-Z. Select for Mode [AutoZ] the boxes Neutralizer and Ion Gun Neut.. Go to the upper right side of the window Image and press auto-z. The software will nd the right height for the best measurement. Now you can go to Acquisition Settings and program your measurement. 2.3 Programming the measurement The programming of the experiment is performed in the window Acquisition Setting. First, select the folder where your measurement will be saved, and give a name to your measurement. Following measurements will add a number automatically. Survey spectrum. Perform always rst a low resolution measurement with broad energy range. Use the following parameters: Range -2, 1200 ev. 8

9 Pass energy: ev. Energy step 0.5 ev. Repeats: 1. Total Cycles: 5. In the case of conducting samples: Neutralize OFF. If the case of non-conducting samples: Neutralize ON. When everything is ready, press Acquire. Core level lines. After identication of the dierent elements present in the sample, you will measure one core level peak of each element (choose the most intense according to the Handbook of Photoelectron Spectroscopy [1]). Measurement parameters: Range: depends on the region. Select the appropriate range with help of the Element Table in the window Acquisition Setting. Pass energy: 23.5 ev. Energy step 0.05 ev. Repeats: Will depend on the intensity of the chosen peaks. For the most intense signals in the spectrum, 10 scans are usually ok. Total Cycles: As many as are necessary. Note: you can measure several regions in the same measurement. In each cycle, the program will measure each region as many times as you determine with repeats. If you have conducting samples: Neutralize OFF. If you have non-conducting samples: Neutralize ON. When everything is ready, press Acquire. 2.4 Data interpretation In general, XPS spectra are featured as a plot of intensity vs decreasing binding energy (BE). A typical spectrum consists of well dened narrow lines (core level lines) over a electronic background that increases with increasing BE. The background is formed by electrons that have suered inelastic scattering processes after their emission from their original atom. The background is continuous because the energy loss processes are random and multiple. 9

10 2.4.1 Types of lines Several types of lines are observed in a XPS spectra. Some appear always, some depend on the particular physical and chemical characteristics of the sample, and some are instrumental in nature. Here we show a selection of the eects relevant to this experiment: Photoelectron lines These are the narrowest lines observed in the spectra. They are generally symmetric, although pure metals can exhibit considerable asymmetry due to interaction of photoelectrons with conduction electrons. Auger lines The relling of the inner shell vacancy left by the emission of an electron can be done by radiative (uorescence) or non radiative(auger) processes. However, in the energy range where we will work ( 1 kev), Auger will be the predominant lling mechanism. Radiative processes compete with Auger only for transition energies of 10 kev. Therefore, in addition to photoelectrons emitted through the photoeect, Auger electrons will be emitted because of the relaxation of the excited ions remaining after photoemission. In an Auger process (which occurs roughly seconds after the photoelectric event), an outer electron falls into the inner orbital vacancy, and a second electron is simultaneously emitted, carrying o the excess energy. The kinetic energy of the Auger electron is equal to the dierence between the energy of the initial ion and the doubly charged ion, and it is independent of the mode of the initial ionization. There are four main Auger series observable in XPS. They are the KLL, LMM, MNN and NOO series. Spin-orbit splitting Due to spin-orbit interactions (j=l±s), the p,d, and f levels split during ionization, leading to vacancies in the p 1/2 and p 3/2 (l=1), d 3/2 and d 5/2 (l=2) orbitals and so on. The relative intensities of the doublet components depend on the multiplicity of the states, given by the available amount of dierent magnetic states (dierent possible values of the quantum number m z : 2j + 1 dierent possible numbers). Therefore, the spin-orbit splitting ratio is 1:2 for p levels, 2:3 for d levels and 3:4 for f levels. Energy loss lines With some materials, there is the possibility of losing certain amount of energy due to interaction between the photoelectron and other electrons in the surface region of the sample. In insulators these lines are not very intense, and can be rather broad. However, in metals the eect is much more dramatic. Energy loss to the conduction electrons occurs in well dened quanta characteristic of each metal. These plasmons arise from group oscillations of the conduction electrons. The photoelectron line is mirrored at intervals of higher BE with reduced intensity. The energy interval between the main peak and the loss peak is called the plasmon energy. The so-called bulk plasmons are the more prominent of these lines. A second series, the surface plasmons, exists at energy intervals determined approximately by dividing the energy of the bulk plasmon by 2. For insulators, usually only the rst plasmon peak is seen. 10

11 Valence bands These signals are produced by photoelectron emission from molecular orbitals and from solid state energy bands. Dierences between insulators and conductors are specially noted by the absence of presence of electron from conduction bands at the Fermi level Chemical Shift The core electron of an element has a unique binding energy, like a ngerprint. Thus almost all elements except for hydrogen and helium can be identied by measuring the binding energy of its core electron. Furthermore, the binding energy of core electrons is sensitive to the chemical environment of element. The inuence of the chemical environment can be seen as a shift of the binding energy of the corresponding XPS peak, ranging from 0.1 ev to 10 ev. This eect is called chemical shift. The chemical shift is therefore the dierence between the energy of a photoelectron line for an element in a specic compound and the corresponding energy for the element in its pure state. The chemical shift is frequently reported in publications and handbooks. As an example you can see in Figure 3 the C1s core level of a PET (Polyethylene terephthalate) sample. The carbon atoms of the polymer see a dierent chemical environment depending on their position within the polymer chain. Dierent chemical environments result in dierent chemical shifts, which we can analyze also quantitatively. This fact is the essence of the utility of XPS for chemical analysis, and the reason why it is also known as electron spectroscopy for chemical analysis (ESCA). Note: you might nd apparent shifts in binding energy positions that are not due to a chemical shift! This refers to the two eects discussed previously: oating Fermi energy level of the sample, and sample charging (see Section 1). 2.5 Mathematical treatment of the data: Fitting procedure and chemical state identication The identication of chemical states depends primarily on the accurate determination of line energies. Once we have eectively determined the line positions, we can start digging into the chemical composition determination problem by looking at the provided literature. In order to get a precise value for the binding energy position of each peak, we will t the experimental data with a mathematical model. The peak shape for XPS spectra is best described by a convolution of the natural line width, the width of the x-ray line which created the photoelectron line, and the instrumental contribution. Additionally, in some cases we have to consider an asymmetry factor (in the case of metallic components.). With the Multipak program, we will however simulate the line prole with a product of a Gaussian and a Lorentzian lines. The Gaussian component accounts for the instrumental broadening, chemical disorder, charging, etc; the Lorentzian function accounts for the nite lifetime of the core-hole in the photoionization process. The mathematical analysis 11

12 Figure 3: 1s core level peak of a PET (Polyethylene terephthalate) polymer sample. with Multipak will give you the following information (necessary for the discussion of the data) BE position, FWHM, percentage of Lorentz value the relative intensities of the peaks you have tted (with Anotate). Now you can try to determine the chemical composition of the sample, by comparing the BE energies that you have determined, with BE found in literature (use handbook and if necessary, additional literature). Compare the situation that you see before and after plasma treatment. 12

13 Quantitative analysis Since the number of photoelectron of an element is dependent upon the atomic concentration of that element in the sample, XPS is used to not only identify the elements but also quantify the chemical composition. In many cases it is important to be able to quantify the relative concentrations of the various constituents of the sample surface. In order to do this we will use the peak areas together with tabulated sensitivity factors. The program Multipak will perform this quantitative analysis automatically, when you press the % symbol on the top right area of the window. For detailed information on the procedure behind this simple mouse click, you can read the rest of this section. For a sample homogeneous in the analysis volume, the number I of photoelectrons per second in a specic spectra peak is given by: I = nfσθyλat n: Number of atoms of the element per cm 3 of the sample. f: X-ray ux in photons/cm 2 s σ: Photoelectric cross section for the atomic orbital of interest in cm 2. θ: Angular eciency factor for the instrumental arrangement based on the angle between the photon path and the detected electron. y: Eciency in the photoelectric process for formation of photoelectrons. λ: Mean free path of the photoelectrons in the sample A: Area of the sample from which photoelectrons are detected T: Detection eciency for the electrons emitted from the sample We are interested in n, which we can write as: I n = fσθyλat I S. The factors contained in S are either a characteristic of the spectrometer, or have a weak dependence from sample to sample, so that, if we have an homogeneous sample, we can establish the following relationships between peaks from dierent elements: n 1 n 2 = I 1S 1 I 2 S 2 where, for each spectrometer, it is possible to determine the values of S for all elements. Note that these concrete values will be applicable only to a particular spectrometer, since they also need to be corrected by a transmission function typical of each machine. 13

14 So now, a general expression for determining the atom fraction of any constituent in a sample C x can be written simply as: 2.6 Report C x = n x ni = I x/s x Ix /S x For the report, discuss the following points about your measurements: Identify all the features observed in the survey spectra. Using the core level spectra, discuss quantitatively the chemical composition of the sample before and after the plasma treatment. Apart from the discussion relevant to the samples that you have investigated, you can use the following set of questions to help you shape your report: What is the photoeect? What is the Auger-Eect? What kind of information can we get from XPS? What features do we expect to see in a XPS spectrum? What is the origin of these structures? What is the approximate range of detectable BEs of all detectable lines by Al radiation? Why does the background level increase abruptly after each line? What is the so-called chemical shift in XPS? What is the origin of charging problems in XPS measurements? What can we do against it? What are the diculties involved in the determination of a binding energy reference in the case of non conducting samples? References [1] J F Moulder, W F Stickle, P E Sobol, K D Bomben. Handbook of X-ray Photoelectron Spectroscopy. Published by ULVAC-PHI,

Advanced Lab Course. X-Ray Photoelectron Spectroscopy 1 INTRODUCTION 1 2 BASICS 1 3 EXPERIMENT Qualitative analysis Chemical Shifts 7

Advanced Lab Course. X-Ray Photoelectron Spectroscopy 1 INTRODUCTION 1 2 BASICS 1 3 EXPERIMENT Qualitative analysis Chemical Shifts 7 Advanced Lab Course X-Ray Photoelectron Spectroscopy M210 As of: 2015-04-01 Aim: Chemical analysis of surfaces. Content 1 INTRODUCTION 1 2 BASICS 1 3 EXPERIMENT 3 3.1 Qualitative analysis 6 3.2 Chemical

More information

Photoemission Spectroscopy

Photoemission Spectroscopy FY13 Experimental Physics - Auger Electron Spectroscopy Photoemission Spectroscopy Supervisor: Per Morgen SDU, Institute of Physics Campusvej 55 DK - 5250 Odense S Ulrik Robenhagen,

More information

X-Ray Photoelectron Spectroscopy (XPS)-2

X-Ray Photoelectron Spectroscopy (XPS)-2 X-Ray Photoelectron Spectroscopy (XPS)-2 Louis Scudiero http://www.wsu.edu/~pchemlab ; 5-2669 Fulmer 261A Electron Spectroscopy for Chemical Analysis (ESCA) The 3 step model: 1.Optical excitation 2.Transport

More information

Introduction to X-ray Photoelectron Spectroscopy (XPS) XPS which makes use of the photoelectric effect, was developed in the mid-1960

Introduction to X-ray Photoelectron Spectroscopy (XPS) XPS which makes use of the photoelectric effect, was developed in the mid-1960 Introduction to X-ray Photoelectron Spectroscopy (XPS) X-ray Photoelectron Spectroscopy (XPS), also known as Electron Spectroscopy for Chemical Analysis (ESCA) is a widely used technique to investigate

More information

X-Ray Photoelectron Spectroscopy (XPS)-2

X-Ray Photoelectron Spectroscopy (XPS)-2 X-Ray Photoelectron Spectroscopy (XPS)-2 Louis Scudiero http://www.wsu.edu/~scudiero; 5-2669 Fulmer 261A Electron Spectroscopy for Chemical Analysis (ESCA) The 3 step model: 1.Optical excitation 2.Transport

More information

Electron Spectroscopy

Electron Spectroscopy Electron Spectroscopy Photoelectron spectroscopy is based upon a single photon in/electron out process. The energy of a photon is given by the Einstein relation : E = h ν where h - Planck constant ( 6.62

More information

Lecture 5. X-ray Photoemission Spectroscopy (XPS)

Lecture 5. X-ray Photoemission Spectroscopy (XPS) Lecture 5 X-ray Photoemission Spectroscopy (XPS) 5. Photoemission Spectroscopy (XPS) 5. Principles 5.2 Interpretation 5.3 Instrumentation 5.4 XPS vs UV Photoelectron Spectroscopy (UPS) 5.5 Auger Electron

More information

X- ray Photoelectron Spectroscopy and its application in phase- switching device study

X- ray Photoelectron Spectroscopy and its application in phase- switching device study X- ray Photoelectron Spectroscopy and its application in phase- switching device study Xinyuan Wang A53073806 I. Background X- ray photoelectron spectroscopy is of great importance in modern chemical and

More information

X-ray Photoelectron Spectroscopy (XPS)

X-ray Photoelectron Spectroscopy (XPS) X-ray Photoelectron Spectroscopy (XPS) As part of the course Characterization of Catalysts and Surfaces Prof. Dr. Markus Ammann Paul Scherrer Institut markus.ammann@psi.ch Resource for further reading:

More information

PHI 5000 Versaprobe-II Focus X-ray Photo-electron Spectroscopy

PHI 5000 Versaprobe-II Focus X-ray Photo-electron Spectroscopy PHI 5000 Versaprobe-II Focus X-ray Photo-electron Spectroscopy The very basic theory of XPS XPS theroy Surface Analysis Ultra High Vacuum (UHV) XPS Theory XPS = X-ray Photo-electron Spectroscopy X-ray

More information

X-Ray Photoelectron Spectroscopy (XPS) Auger Electron Spectroscopy (AES)

X-Ray Photoelectron Spectroscopy (XPS) Auger Electron Spectroscopy (AES) X-Ray Photoelectron Spectroscopy (XPS) Auger Electron Spectroscopy (AES) XPS X-ray photoelectron spectroscopy (XPS) is one of the most used techniques to chemically characterize the surface. Also known

More information

Birck Nanotechnology Center XPS: X-ray Photoelectron Spectroscopy ESCA: Electron Spectrometer for Chemical Analysis

Birck Nanotechnology Center XPS: X-ray Photoelectron Spectroscopy ESCA: Electron Spectrometer for Chemical Analysis Birck Nanotechnology Center XPS: X-ray Photoelectron Spectroscopy ESCA: Electron Spectrometer for Chemical Analysis Dmitry Zemlyanov Birck Nanotechnology Center, Purdue University Outline Introduction

More information

X-Ray Photoelectron Spectroscopy (XPS)

X-Ray Photoelectron Spectroscopy (XPS) X-Ray Photoelectron Spectroscopy (XPS) Louis Scudiero http://www.wsu.edu/~scudiero; 5-2669 Electron Spectroscopy for Chemical Analysis (ESCA) The basic principle of the photoelectric effect was enunciated

More information

The Benefit of Wide Energy Range Spectrum Acquisition During Sputter Depth Profile Measurements

The Benefit of Wide Energy Range Spectrum Acquisition During Sputter Depth Profile Measurements The Benefit of Wide Energy Range Spectrum Acquisition During Sputter Depth Profile Measurements Uwe Scheithauer, 82008 Unterhaching, Germany E-Mail: scht.uhg@googlemail.com Internet: orcid.org/0000-0002-4776-0678;

More information

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu X-ray Photoelectron Spectroscopy Introduction Qualitative analysis Quantitative analysis Charging compensation Small area analysis and XPS imaging

More information

X-Ray Photoelectron Spectroscopy (XPS)

X-Ray Photoelectron Spectroscopy (XPS) X-Ray Photoelectron Spectroscopy (XPS) Louis Scudiero http://www.wsu.edu/~scudiero; 5-2669 Fulmer 261A Electron Spectroscopy for Chemical Analysis (ESCA) The basic principle of the photoelectric effect

More information

Auger Electron Spectroscopy *

Auger Electron Spectroscopy * OpenStax-CNX module: m43546 1 Auger Electron Spectroscopy * Amanda M. Goodman Andrew R. Barron This work is produced by OpenStax-CNX and licensed under the Creative Commons Attribution License 3.0 1 Basic

More information

X-ray spectroscopy: Experimental studies of Moseley s law (K-line x-ray fluorescence) and x-ray material s composition determination

X-ray spectroscopy: Experimental studies of Moseley s law (K-line x-ray fluorescence) and x-ray material s composition determination Uppsala University Department of Physics and Astronomy Laboratory exercise X-ray spectroscopy: Experimental studies of Moseley s law (K-line x-ray fluorescence) and x-ray material s composition determination

More information

Methods of surface analysis

Methods of surface analysis Methods of surface analysis Nanomaterials characterisation I RNDr. Věra Vodičková, PhD. Surface of solid matter: last monoatomic layer + absorbed monolayer physical properties are effected (crystal lattice

More information

5) Surface photoelectron spectroscopy. For MChem, Spring, Dr. Qiao Chen (room 3R506) University of Sussex.

5) Surface photoelectron spectroscopy. For MChem, Spring, Dr. Qiao Chen (room 3R506) University of Sussex. For MChem, Spring, 2009 5) Surface photoelectron spectroscopy Dr. Qiao Chen (room 3R506) http://www.sussex.ac.uk/users/qc25/ University of Sussex Today s topics 1. Element analysis with XPS Binding energy,

More information

Lecture 5-8 Instrumentation

Lecture 5-8 Instrumentation Lecture 5-8 Instrumentation Requirements 1. Vacuum Mean Free Path Contamination Sticking probability UHV Materials Strength Stability Permeation Design considerations Pumping speed Virtual leaks Leaking

More information

Auger Electron Spectroscopy (AES)

Auger Electron Spectroscopy (AES) 1. Introduction Auger Electron Spectroscopy (AES) Silvia Natividad, Gabriel Gonzalez and Arena Holguin Auger Electron Spectroscopy (Auger spectroscopy or AES) was developed in the late 1960's, deriving

More information

IV. Surface analysis for chemical state, chemical composition

IV. Surface analysis for chemical state, chemical composition IV. Surface analysis for chemical state, chemical composition Probe beam Detect XPS Photon (X-ray) Photoelectron(core level electron) UPS Photon (UV) Photoelectron(valence level electron) AES electron

More information

QUESTIONS AND ANSWERS

QUESTIONS AND ANSWERS QUESTIONS AND ANSWERS (1) For a ground - state neutral atom with 13 protons, describe (a) Which element this is (b) The quantum numbers, n, and l of the inner two core electrons (c) The stationary state

More information

Atomic Spectra HISTORY AND THEORY

Atomic Spectra HISTORY AND THEORY Atomic Spectra HISTORY AND THEORY When atoms of a gas are excited (by high voltage, for instance) they will give off light. Each element (in fact, each isotope) gives off a characteristic atomic spectrum,

More information

5.8 Auger Electron Spectroscopy (AES)

5.8 Auger Electron Spectroscopy (AES) 5.8 Auger Electron Spectroscopy (AES) 5.8.1 The Auger Process X-ray and high energy electron bombardment of atom can create core hole Core hole will eventually decay via either (i) photon emission (x-ray

More information

Auger Electron Spectroscopy (AES) Prof. Paul K. Chu

Auger Electron Spectroscopy (AES) Prof. Paul K. Chu Auger Electron Spectroscopy (AES) Prof. Paul K. Chu Auger Electron Spectroscopy Introduction Principles Instrumentation Qualitative analysis Quantitative analysis Depth profiling Mapping Examples The Auger

More information

X-ray Photoemission Spectroscopy (XPS - Ma4)

X-ray Photoemission Spectroscopy (XPS - Ma4) Master Laboratory Report X-ray Photoemission Spectroscopy (XPS - Ma4) Supervisor: Andrew Britton Students: Dachi Meurmishvili, Muhammad Khurram Riaz and Martin Borchert Date: November 17th 2016 1 Contents

More information

Ma5: Auger- and Electron Energy Loss Spectroscopy

Ma5: Auger- and Electron Energy Loss Spectroscopy Ma5: Auger- and Electron Energy Loss Spectroscopy 1 Introduction Electron spectroscopies, namely Auger electron- and electron energy loss spectroscopy are utilized to determine the KLL spectrum and the

More information

Photoelectron spectroscopy Instrumentation. Nanomaterials characterization 2

Photoelectron spectroscopy Instrumentation. Nanomaterials characterization 2 Photoelectron spectroscopy Instrumentation Nanomaterials characterization 2 RNDr. Věra V Vodičkov ková,, PhD. Photoelectron Spectroscopy general scheme Impact of X-ray emitted from source to the sample

More information

The Use of Synchrotron Radiation in Modern Research

The Use of Synchrotron Radiation in Modern Research The Use of Synchrotron Radiation in Modern Research Physics Chemistry Structural Biology Materials Science Geochemical and Environmental Science Atoms, molecules, liquids, solids. Electronic and geometric

More information

Name: (a) What core levels are responsible for the three photoelectron peaks in Fig. 1?

Name: (a) What core levels are responsible for the three photoelectron peaks in Fig. 1? Physics 243A--Surface Physics of Materials: Spectroscopy Final Examination December 16, 2014 (3 problems, 100 points total, open book, open notes and handouts) Name: [1] (50 points), including Figures

More information

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous? 2. Briefly discuss dead time in a detector. What factors are important

More information

Auger Electron Spectroscopy Overview

Auger Electron Spectroscopy Overview Auger Electron Spectroscopy Overview Also known as: AES, Auger, SAM 1 Auger Electron Spectroscopy E KLL = E K - E L - E L AES Spectra of Cu EdN(E)/dE Auger Electron E N(E) x 5 E KLL Cu MNN Cu LMM E f E

More information

X-RAY SPECTRA. Theory:

X-RAY SPECTRA. Theory: 12 Oct 18 X-ray.1 X-RAY SPECTRA In this experiment, a number of measurements involving x-rays will be made. The spectrum of x-rays emitted from a molybdenum target will be measured, and the experimental

More information

X-ray Photoelectron Spectroscopy/ Electron spectroscopy for chemical analysis (ESCA), By Francis Chindeka

X-ray Photoelectron Spectroscopy/ Electron spectroscopy for chemical analysis (ESCA), By Francis Chindeka X-ray Photoelectron Spectroscopy/ Electron spectroscopy for chemical analysis (ESCA), By Francis Chindeka X-ray photoelectron spectroscopy (XPS) or Electron spectroscopy for chemical analysis (ESCA), Surface

More information

Photon Interaction. Spectroscopy

Photon Interaction. Spectroscopy Photon Interaction Incident photon interacts with electrons Core and Valence Cross Sections Photon is Adsorbed Elastic Scattered Inelastic Scattered Electron is Emitted Excitated Dexcitated Stöhr, NEXAPS

More information

Multi-technique photoelectron spectrometer for micro-area spectroscopy and imaging

Multi-technique photoelectron spectrometer for micro-area spectroscopy and imaging Multi-technique photoelectron spectrometer for micro-area spectroscopy and imaging U. Manju, M. Sreemony and A. K. Chakraborty In this note we present the new multipurpose photoelectron spectroscopy facility

More information

A DIVISION OF ULVAC-PHI. Quantera II. Scanning XPS Microprobe

A DIVISION OF ULVAC-PHI. Quantera II. Scanning XPS Microprobe A DIVISION OF ULVAC-PHI Quantera II Scanning XPS Microprobe X-ray Photoelectron Spectroscopy (XPS/ESCA) is the most widely used surface analysis technique and has many well established industrial and

More information

Low Energy Electrons and Surface Chemistry

Low Energy Electrons and Surface Chemistry G. Ertl, J. Küppers Low Energy Electrons and Surface Chemistry VCH 1 Basic concepts 1 1.1 Introduction 1 1.2 Principles of ultrahigh vacuum techniques 2 1.2.1 Why is UHV necessary? 2 1.2.2 Production of

More information

Lecture 17 Auger Electron Spectroscopy

Lecture 17 Auger Electron Spectroscopy Lecture 17 Auger Electron Spectroscopy Auger history cloud chamber Although Auger emission is intense, it was not used until 1950 s. Evolution of vacuum technology and the application of Auger Spectroscopy

More information

Electron Spettroscopies

Electron Spettroscopies Electron Spettroscopies Spettroscopy allows to characterize a material from the point of view of: chemical composition, electronic states and magnetism, electronic, roto-vibrational and magnetic excitations.

More information

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy Experiment VI Gamma Ray Spectroscopy 1. GAMMA RAY INTERACTIONS WITH MATTER In order for gammas to be detected, they must lose energy in the detector. Since gammas are electromagnetic radiation, we must

More information

Auger Electron Spectrometry. EMSE-515 F. Ernst

Auger Electron Spectrometry. EMSE-515 F. Ernst Auger Electron Spectrometry EMSE-515 F. Ernst 1 Principle of AES electron or photon in, electron out radiation-less transition Auger electron electron energy properties of atom 2 Brief History of Auger

More information

Thermo Scientific K-Alpha + XPS Spectrometer. Fast, powerful and accessible chemical analysis for surface and thin film characterization

Thermo Scientific K-Alpha + XPS Spectrometer. Fast, powerful and accessible chemical analysis for surface and thin film characterization Thermo Scientific K-Alpha + XPS Spectrometer Fast, powerful and accessible chemical analysis for surface and thin film characterization X-ray Photoelectron Spectroscopy Quantitative, chemical identification

More information

An Introduction to Auger Electron Spectroscopy

An Introduction to Auger Electron Spectroscopy An Introduction to Auger Electron Spectroscopy Spyros Diplas MENA3100 SINTEF Materials & Chemistry, Department of Materials Physics & Centre of Materials Science and Nanotechnology, Department of Chemistry,

More information

Photoelectron Peak Intensities in Solids

Photoelectron Peak Intensities in Solids Photoelectron Peak Intensities in Solids Electronic structure of solids Photoelectron emission through solid Inelastic scattering Other excitations Intrinsic and extrinsic Shake-up, shake-down and shake-off

More information

X-ray photoelectron spectroscopy - An introduction

X-ray photoelectron spectroscopy - An introduction X-ray photoelectron spectroscopy - An introduction Spyros Diplas spyros.diplas@sintef.no spyros.diplas@smn.uio.no SINTEF Materials & Chemistry, Materials Physics -Oslo & Centre of Materials Science and

More information

Lecture 22 Ion Beam Techniques

Lecture 22 Ion Beam Techniques Lecture 22 Ion Beam Techniques Schroder: Chapter 11.3 1/44 Announcements Homework 6/6: Will be online on later today. Due Wednesday June 6th at 10:00am. I will return it at the final exam (14 th June).

More information

A DIVISION OF ULVAC-PHI

A DIVISION OF ULVAC-PHI A DIVISION OF ULVAC-PHI X-ray photoelectron spectroscopy (XPS/ESCA) is the most widely used surface analysis technique and has many well established industrial and research applications. XPS provides

More information

Lecture 20 Auger Electron Spectroscopy

Lecture 20 Auger Electron Spectroscopy Lecture 20 Auger Electron Spectroscopy Auger history cloud chamber Although Auger emission is intense, it was not used until 1950 s. Evolution of vacuum technology and the application of Auger Spectroscopy

More information

Bonds in molecules are formed by the interactions between electrons.

Bonds in molecules are formed by the interactions between electrons. CHEM 2060 Lecture 6: Electrostatic Interactions L6-1 PART TWO: Electrostatic Interactions In the first section of this course, we were more concerned with structural aspects of molecules. In this section

More information

and Environmental Science Centre

and Environmental Science Centre 1. Purpose The purpose of this document is to familiarize the user with the mode of function of the FT-Raman available at the facility, and to describe the sampling procedure. 2. Introduction Raman spectroscopy

More information

Auger Electron Spectroscopy

Auger Electron Spectroscopy Auger Electron Spectroscopy Auger Electron Spectroscopy is an analytical technique that provides compositional information on the top few monolayers of material. Detect all elements above He Detection

More information

Lab 5 - ELECTRON CHARGE-TO-MASS RATIO

Lab 5 - ELECTRON CHARGE-TO-MASS RATIO 81 Name Date Partners Lab 5 - ELECTRON CHARGE-TO-MASS RATIO OBJECTIVES To understand how electric and magnetic fields impact an electron beam To experimentally determine the electron charge-to-mass ratio

More information

Energy Spectroscopy. Excitation by means of a probe

Energy Spectroscopy. Excitation by means of a probe Energy Spectroscopy Excitation by means of a probe Energy spectral analysis of the in coming particles -> XAS or Energy spectral analysis of the out coming particles Different probes are possible: Auger

More information

EMISSION SPECTROSCOPY

EMISSION SPECTROSCOPY IFM The Department of Physics, Chemistry and Biology LAB 57 EMISSION SPECTROSCOPY NAME PERSONAL NUMBER DATE APPROVED I. OBJECTIVES - Understand the principle of atomic emission spectra. - Know how to acquire

More information

4. Inelastic Scattering

4. Inelastic Scattering 1 4. Inelastic Scattering Some inelastic scattering processes A vast range of inelastic scattering processes can occur during illumination of a specimen with a highenergy electron beam. In principle, many

More information

Setting The motor that rotates the sample about an axis normal to the diffraction plane is called (or ).

Setting The motor that rotates the sample about an axis normal to the diffraction plane is called (or ). X-Ray Diffraction X-ray diffraction geometry A simple X-ray diffraction (XRD) experiment might be set up as shown below. We need a parallel X-ray source, which is usually an X-ray tube in a fixed position

More information

Inelastic soft x-ray scattering, fluorescence and elastic radiation

Inelastic soft x-ray scattering, fluorescence and elastic radiation Inelastic soft x-ray scattering, fluorescence and elastic radiation What happens to the emission (or fluorescence) when the energy of the exciting photons changes? The emission spectra (can) change. One

More information

Table 1: Residence time (τ) in seconds for adsorbed molecules

Table 1: Residence time (τ) in seconds for adsorbed molecules 1 Surfaces We got our first hint of the importance of surface processes in the mass spectrum of a high vacuum environment. The spectrum was dominated by water and carbon monoxide, species that represent

More information

Lab 5 - ELECTRON CHARGE-TO-MASS RATIO

Lab 5 - ELECTRON CHARGE-TO-MASS RATIO 79 Name Date Partners OBJECTIVES OVERVIEW Lab 5 - ELECTRON CHARGE-TO-MASS RATIO To understand how electric and magnetic fields impact an electron beam To experimentally determine the electron charge-to-mass

More information

A Beginners Guide to XPS

A Beginners Guide to XPS A Beginners Guide to XPS XPS Instrumentation Figure 1: Schematic of an XPS instrument. Photoemission occurs when photon energy is transferred to electrons within bound-states of atoms causing the electron

More information

EEE4106Z Radiation Interactions & Detection

EEE4106Z Radiation Interactions & Detection EEE4106Z Radiation Interactions & Detection 2. Radiation Detection Dr. Steve Peterson 5.14 RW James Department of Physics University of Cape Town steve.peterson@uct.ac.za May 06, 2015 EEE4106Z :: Radiation

More information

Lab 6 - ELECTRON CHARGE-TO-MASS RATIO

Lab 6 - ELECTRON CHARGE-TO-MASS RATIO 101 Name Date Partners OBJECTIVES OVERVIEW Lab 6 - ELECTRON CHARGE-TO-MASS RATIO To understand how electric and magnetic fields impact an electron beam To experimentally determine the electron charge-to-mass

More information

MSE 321 Structural Characterization

MSE 321 Structural Characterization Auger Spectroscopy Auger Electron Spectroscopy (AES) Scanning Auger Microscopy (SAM) Incident Electron Ejected Electron Auger Electron Initial State Intermediate State Final State Physical Electronics

More information

Scintillation Detector

Scintillation Detector Scintillation Detector Introduction The detection of ionizing radiation by the scintillation light produced in certain materials is one of the oldest techniques on record. In Geiger and Marsden s famous

More information

Secondary Ion Mass Spectrometry (SIMS)

Secondary Ion Mass Spectrometry (SIMS) CHEM53200: Lecture 10 Secondary Ion Mass Spectrometry (SIMS) Major reference: Surface Analysis Edited by J. C. Vickerman (1997). 1 Primary particles may be: Secondary particles can be e s, neutral species

More information

4. How can fragmentation be useful in identifying compounds? Permits identification of branching not observed in soft ionization.

4. How can fragmentation be useful in identifying compounds? Permits identification of branching not observed in soft ionization. Homework 9: Chapters 20-21 Assigned 12 April; Due 17 April 2006; Quiz on 19 April 2006 Chap. 20 (Molecular Mass Spectroscopy) Chap. 21 (Surface Analysis) 1. What are the types of ion sources in molecular

More information

PHOTOELECTRON SPECTROSCOPY IN AIR (PESA)

PHOTOELECTRON SPECTROSCOPY IN AIR (PESA) PHOTOELECTRON SPECTROSCOPY IN AIR (PESA) LEADERS IN GAS DETECTION Since 1977 Model AC-3 Features: Atmospheric pressure operation (unique in the world) Estimate work function, ionization potential, density

More information

Ion sputtering yield coefficients from In thin films bombarded by different energy Ar + ions

Ion sputtering yield coefficients from In thin films bombarded by different energy Ar + ions Ion sputtering yield coefficients from thin films bombarded by different energy Ar + ions MJ Madito, H Swart and JJ Terblans 1 Department of Physics, University of the Free State, P.. Box 339, Bloemfontein,

More information

Lab 1: Determination of e/m for the electron

Lab 1: Determination of e/m for the electron Lab 1: Determination of e/m for the electron Background Reading: Tipler, Llewellyn pp. 125 130; this covers the original method of Thomson which is somewhat different from that used in this experiment

More information

Practical Surface Analysis

Practical Surface Analysis Practical Surface Analysis SECOND EDITION Volume 1 Auger and X-ray Photoelectron Spectroscopy Edited by D. BRIGGS ICI PLC, Wilton Materials Research Centre, Wilton, Middlesbrough, Cleveland, UK and M.

More information

Introduction to X-ray Photoelectron Spectroscopy (XPS) Introduction to X-ray Photoelectron Spectroscopy (XPS) Comparison of Sensitivities

Introduction to X-ray Photoelectron Spectroscopy (XPS) Introduction to X-ray Photoelectron Spectroscopy (XPS) Comparison of Sensitivities Introduction to X-ray Photoelectron Spectroscopy (XPS) Sources of Information Principles of XPS and Auger How to prepare samples for XPS Instrumentation, X rays, Photoelectron detection Data acquisition

More information

X-ray Energy Spectroscopy (XES).

X-ray Energy Spectroscopy (XES). X-ray Energy Spectroscopy (XES). X-ray fluorescence as an analytical tool for element analysis is based on 3 fundamental parameters: A. Specificity: In determining an x-ray emission energy E certainty

More information

X-ray Spectroscopy. Interaction of X-rays with matter XANES and EXAFS XANES analysis Pre-edge analysis EXAFS analysis

X-ray Spectroscopy. Interaction of X-rays with matter XANES and EXAFS XANES analysis Pre-edge analysis EXAFS analysis X-ray Spectroscopy Interaction of X-rays with matter XANES and EXAFS XANES analysis Pre-edge analysis EXAFS analysis Element specific Sensitive to low concentrations (0.01-0.1 %) Why XAS? Applicable under

More information

Probing Matter: Diffraction, Spectroscopy and Photoemission

Probing Matter: Diffraction, Spectroscopy and Photoemission Probing Matter: Diffraction, Spectroscopy and Photoemission Anders Nilsson Stanford Synchrotron Radiation Laboratory Why X-rays? VUV? What can we hope to learn? 1 Photon Interaction Incident photon interacts

More information

PHOTOELECTRON SPECTROSCOPY (PES)

PHOTOELECTRON SPECTROSCOPY (PES) PHOTOELECTRON SPECTROSCOPY (PES) NTRODUCTON Law of Photoelectric effect Albert Einstein, Nobel Prize 1921 Kaiser-Wilhelm-nstitut (now Max-Planck- nstitut) für Physik Berlin, Germany High-resolution electron

More information

This experiment is included in the XRP 4.0 X-ray solid state, XRS 4.0 X-ray structural analysis, and XRC 4.0 X-ray characteristics upgrade sets.

This experiment is included in the XRP 4.0 X-ray solid state, XRS 4.0 X-ray structural analysis, and XRC 4.0 X-ray characteristics upgrade sets. The intensity of characteristic X-rays as a TEP Related topics Characteristic X-radiation, energy levels, Bragg s law, and intensity of characteristic X-rays Principle The X-ray spectrum of an X-ray tube

More information

Lab 6 - Electron Charge-To-Mass Ratio

Lab 6 - Electron Charge-To-Mass Ratio Lab 6 Electron Charge-To-Mass Ratio L6-1 Name Date Partners Lab 6 - Electron Charge-To-Mass Ratio OBJECTIVES To understand how electric and magnetic fields impact an electron beam To experimentally determine

More information

Group Members: Your Name In Class Exercise #6. Photon A. Energy B

Group Members: Your Name In Class Exercise #6. Photon A. Energy B Group Members: Your Name In Class Exercise #6 Shell Structure of Atoms Part II Photoelectron Spectroscopy Photoelectron spectroscopy is closely related to the photoelectric effect. When high energy photons

More information

AES - Auger Electron Spectrosopy

AES - Auger Electron Spectrosopy Advanced Materials - Lab Intermediate Physics Ulm University Institute of Solid State Physics AES - Auger Electron Spectrosopy Sebastian Schnurr March 13, 2013 Safety Precautions MAKE SURE THAT YOU UNDERSTAND

More information

Chemical Analysis in TEM: XEDS, EELS and EFTEM. HRTEM PhD course Lecture 5

Chemical Analysis in TEM: XEDS, EELS and EFTEM. HRTEM PhD course Lecture 5 Chemical Analysis in TEM: XEDS, EELS and EFTEM HRTEM PhD course Lecture 5 1 Part IV Subject Chapter Prio x-ray spectrometry 32 1 Spectra and mapping 33 2 Qualitative XEDS 34 1 Quantitative XEDS 35.1-35.4

More information

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency. Light We can use different terms to describe light: Color Wavelength Frequency Light is composed of electromagnetic waves that travel through some medium. The properties of the medium determine how light

More information

Ultraviolet Photoelectron Spectroscopy (UPS)

Ultraviolet Photoelectron Spectroscopy (UPS) Ultraviolet Photoelectron Spectroscopy (UPS) Louis Scudiero http://www.wsu.edu/~scudiero www.wsu.edu/~scudiero; ; 5-26695 scudiero@wsu.edu Photoemission from Valence Bands Photoelectron spectroscopy is

More information

Ma4: X-ray photoelectron spectroscopy (XPS)

Ma4: X-ray photoelectron spectroscopy (XPS) Ma4: X-ray photoelectron spectroscopy (XPS) Contents 1. Overview on the topic and summary of the objectives............. 3 2. Theoretical Fundamentals............................ 3 2.1 Characteristic excitation

More information

X-ray uorescence, X-ray powder diraction and Raman spectrosopy

X-ray uorescence, X-ray powder diraction and Raman spectrosopy X-ray uorescence, X-ray powder diraction and Raman spectrosopy Wubulikasimu Yibulayin, Lovro Pavleti, Kuerbannisa Muhetaer 12.05.2016. Abstract In this report we report about the experiments that were

More information

LASER. Light Amplification by Stimulated Emission of Radiation

LASER. Light Amplification by Stimulated Emission of Radiation LASER Light Amplification by Stimulated Emission of Radiation Laser Fundamentals The light emitted from a laser is monochromatic, that is, it is of one color/wavelength. In contrast, ordinary white light

More information

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 19 Chapter 12 There are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry X-ray Techniques include:

More information

Atomic and nuclear physics

Atomic and nuclear physics Atomic and nuclear physics X-ray physics Attenuation of x-rays LEYBOLD Physics Leaflets P6.3.2.2 Investigating the wavelength dependency of the coefficient of attenuation Objects of the experiment To measure

More information

Supporting Information. Re-Investigation of the Alleged Formation of CoSi Nanoparticles on Silica. Van An Du, Silvia Gross and Ulrich Schubert

Supporting Information. Re-Investigation of the Alleged Formation of CoSi Nanoparticles on Silica. Van An Du, Silvia Gross and Ulrich Schubert Supporting Information Re-Investigation of the Alleged Formation of CoSi Nanoparticles on Silica Van An Du, Silvia Gross and Ulrich Schubert Experimental All manipulations were carried out under an atmosphere

More information

Fig 1: Auger Electron Generation (a) Step 1 and (b) Step 2

Fig 1: Auger Electron Generation (a) Step 1 and (b) Step 2 Auger Electron Spectroscopy (AES) Physics of AES: Auger Electrons were discovered in 1925 but were used in surface analysis technique in 1968. Auger Electron Spectroscopy (AES) is a very effective method

More information

Practical 1P4 Energy Levels and Band Gaps

Practical 1P4 Energy Levels and Band Gaps Practical 1P4 Energy Levels and Band Gaps What you should learn from this practical Science This practical illustrates some of the points from the lecture course on Elementary Quantum Mechanics and Bonding

More information

Determining Chemical Composition. Of Sputtered Uranium Oxide Thin Films. through X-Ray Photoelectron Spectroscopy

Determining Chemical Composition. Of Sputtered Uranium Oxide Thin Films. through X-Ray Photoelectron Spectroscopy Determining Chemical Composition Of Sputtered Uranium Oxide Thin Films through X-Ray Photoelectron Spectroscopy Kristi Adamson Department of Physics & Astronomy Brigham Young University April 2004 Submitted

More information

Spectroscopy of Nanostructures. Angle-resolved Photoemission (ARPES, UPS)

Spectroscopy of Nanostructures. Angle-resolved Photoemission (ARPES, UPS) Spectroscopy of Nanostructures Angle-resolved Photoemission (ARPES, UPS) Measures all quantum numbers of an electron in a solid. E, k x,y, z, point group, spin E kin, ϑ,ϕ, hν, polarization, spin Electron

More information

X-ray practical: Crystallography

X-ray practical: Crystallography X-ray practical: Crystallography Aim: To familiarise oneself with the operation of Tex-X-Ometer spectrometer and to use it to determine the lattice spacing in NaCl and LiF single crystals. Background:

More information

tip conducting surface

tip conducting surface PhysicsAndMathsTutor.com 1 1. The diagram shows the tip of a scanning tunnelling microscope (STM) above a conducting surface. The tip is at a potential of 1.0 V relative to the surface. If the tip is sufficiently

More information

X-Ray Photoelectron Spectroscopy: Theory and Practice

X-Ray Photoelectron Spectroscopy: Theory and Practice X-Ray Photoelectron Spectroscopy: Theory and Practice PHYS-581 (Fall 2010) Contact Information for EMS in RRC-East Alan Nicholls, PhD Director of Research Service Facility - Electron Microscopy Research

More information

Surface Analysis - The Principal Techniques

Surface Analysis - The Principal Techniques Surface Analysis - The Principal Techniques Edited by John C. Vickerman Surface Analysis Research Centre, Department of Chemistry UMIST, Manchester, UK JOHN WILEY & SONS Chichester New York Weinheim Brisbane

More information

Homework 2: Forces on Charged Particles

Homework 2: Forces on Charged Particles Homework 2: Forces on Charged Particles 1. In the arrangement shown below, 2 C of positive charge is moved from plate S, which is at a potential of 250 V, to plate T, which is at a potential of 750 V.

More information