Using 11.7eV PID Lamps To Accurately Measure VOCs With Ionization Energy Above 10.6eV
|
|
- Eleanore Conley
- 5 years ago
- Views:
Transcription
1 Using 11.7eV PID Lamps To Accurately Measure VOCs With Ionization Energy Above 10.6eV Summary Honeywell RAE Systems strongly recommends using 100 ppm propane as calibration gas for instruments with 11.7eV lamps when the purpose is to measure VOCs with IE (ionization energy) above 10.6eV. The statement that the best policy is Calibration gas = Measurement gas is still valid and is preferred. Honeywell RAE Systems is well-established manufacturer of photoionization detector (PID) lamps with nominal values of 9.8eV, 10.6eV, 11.7eV and different sizes, including 1/4", 3/8, and 1/2. The most widely used is the 10.6eV lamp, with its long life and broad detection range. The 11.7eV analog is demanded by customers for its ability to measure VOC with high ionization energy (IE) up to 11.7eV. In this document, we discuss the main 11.7eV lamp properties and provide recommendations to better calibrate instruments so that measurements with this lamp type are much more accurate and reliable. Lamp Design Typical PID lamp design comprises a glass bulb, crystal window, and glue to attach crystal to glass. The volume inside the lamp is filled with one or more noble gases. Once a high electric field is applied to the lamp, the gas inside is ionized, creating plasma. Plasma emits photons with specific energy, defined by the gas. A crystal window is used so that photons with the desired energy pass through the window, reach the VOC, and ionize it. When an electric field is applied to the PID sensor electrodes, ions migrate to them and create an electric current registered by the instrument. Lamp design is shown in the figure below. Crystal Glue Glass Bulb The crystal used in 11.7eV lamp design is transparent to the photons with energy up to 11.7eV, which allows measuring VOCs with ionization energy up to 11.7eV. To provide photons with such high energy, the lamp is filled with the gas, which has emission spectral lines at 11.83eV and 11.62eV and relative intensity of 1000 and 500, respectively 1. Crystal Transmission Spectra As mentioned before, two main factors affect lamp efficiency: emission lines of the gas (their energy and intensity) and crystal window transmission spectra that allows photons pass through the crystal and reach VOC in the PID compartment. 1
2 Below is an example of modeled transmission spectra of crystal with around 0.5 mm thickness. Transmission, % Crystal transmission spectra vs. photon energy 120 Noble gas lines >T <T Photon Energy, ev The blue curve in the figure is transmission vs. photon energy. The two red vertical lines represent emission lines of the gas that fill the 11.7eV lamp. The higher energy line (11.83eV) coincides with the sharp cut off slope of transmission crystal spectra, and as a result its intensity is very sensitive to minor changes in the crystal s optical properties. The lower energy gas emission line is on the more shallow part of the crystal s spectra and less vulnerable to changes in the crystal s properties. Even though there is no quantitative relationship between VOC sensitivity from photon energy and IE of VOC, the rule of thumb is that the greater the difference between the lamp photon energy and IE of VOC, the more sensitive VOC in the instrument measurement. That is, its CF (correction factor) relative to calibration gas is lower. Note: Because the higher-energy emitted line of the gas inside the lamp is in the same range as the sharp slope of the transmission curve, the 11.7eV lamp output s spectra are very vulnerable to any crystal properties change. There is another intrinsic property of the crystal window that affects the crystal s performance. It is the ability of the transmission curve vs. temperature to shift the cut-off 2 position. Two arrows on the figure show the direction of the curve shift vs. temperature. Once temperature goes up (red arrow), the curve shifts left. That is, the transmission of the higherenergy emission line in the crystal may drop significantly, up to several times. If the temperature decreases, the curve shifts to the right (blue arrow), and in turn, transmission of crystal in the range of higher emitted gas line increases. Changing the intensity of the higher-energy gas emitted line passing through the crystal window relative to the lower one changes the lamp energy spectra and leads to modified VOC sensitivity and its CF relative to calibration gas. Note: This is the intrinsic property of crystal, and we can only mitigate it by calibrating the instrument at about the same ambient temperature as the conditions in the field. VOCs Correction Factor Drift Crystal properties change as the lamp matures. Most likely, changing lamp-emitted spectra leads to changes in the sensitivity of different VOCs to be ionized, depending on IE. For example, VOCs with low IE like isobutylene (IBE) are less sensitive to a change in lamp output from 11.7eV to 11.6eV. At the same time, VOCs with high IE, close to 11eV and above, may be significantly affected by reducing lamp output energy in the high-energy part of spectra. It means real CFs of VOCs with high IE change more significantly as an 11.7eV lamp matures relative to calibration gas with low IE like IBE. The greatest change takes place at the beginning of new a lamp s use and then gradually continues to change. The figure below includes test data from three different batches of 11.7eV lamps. Batch 1 and 2 were made at the same time using the same technological process. Batch 3 is from the newest adjusted and locked-in process of manufacturing. 2
3 Lamps were tested in MiniRAE 3000 instruments for their response to three gases: 100 ppm IBE, 100 ppm propane, and 10 ppm chlorine. Records were done in raw counts in the instruments diagnostic mode. Measurements were performed with fresh lamps, after lamps sat for 5 days on the shelf, and after 8 hours in which the lamps were burning. Raw-count data were converted to CF of Propane vs. IBE (left column), Chlorine vs. IBE (middle column), and Chlorine vs. Propane (right column). Lamp Batch Test CF Pr/IBE CF Cl2/IBE CF Cl2/Pr Batch 1 Initial data days on shelf hours burning Batch 2 Initial data days on shelf hours burning Batch 3 Initial data days on shelf hours burning As shown in the table, CFs of propane and chlorine vs. IBE significantly increase during time sitting on the shelf and in the first 8 hours burning. At the same time, CF of chlorine vs. propane stays constant (batches 1 and 2) or change just slightly (batch 3). Does the CF for chlorine vs. propane change as the lamp is used? The answer to this question is given in the figure below eV lamp, VOCs CFs change CFs CFPr/ibe CFCl2/ibe CFCl2/Pr Time, Hrs Batch 3 was tested in a MiniRAE 3000 for over 300 hours, which is equivalent to more than 30 days of lamp operation running for 8 hours per day and 7 days a week. (That is, the time exceeds lamp warranty life.) Regardless, CFs of propane and Cl 2 vs. IBE changed significantly, while the CF of Cl 2 vs. propane is nearly constant (see table below). Time, hours in test CF Pr/IBE CF Cl 2/IBE CF Cl 2/IBE
4 In addition, comparison measurements of the effect of maturing 11.7eV lamps on CFs were done for the most commonly used VOCs (see table below). Column 2 presents the data from Honeywell RAE Systems Technical Note TN-106 for the VOC CFs developed with new 11.7eV lamps, and column 3 comprises data for newly developed CFs with mature 11.7eV lamps. In both columns 2 and 3, VOCs were calibrated with IBE. Column 4 shows CF changes from mature lamps. The change is significant and varies by factors of 3.6 to 9 times, depending on the VOC. It shows that if the IE of a VOC is below 10.6eV, this change is less drastic (for example, ammonia and ethylene oxide in the table), but it is still noticeable. So, there are no benefits to using an 11.7eV lamp to measure VOCs with IE below 10.6eV. The 10.6eV lamp allows higher accuracy and more consistency. The situation with CFs variability using an 11.7eV lamp looks better when using CFs relative to propane (that is, using propane as a calibration gas). Data in column 7 of the table show that CFs vary, depending on the VOC in the range of 0.7 to 1.8. That is much narrower than if the same VOCs were calibrated to IBE and CFs were used, relative to IBE. To obtain these data, CFs from TN-106 were recalculated using the formula explained below the Table and placed in column 5. Data in column 6 are newly developed CFs of the VOCs most often used together with an 11.7eV lamp and calibrated to 100 ppm propane. So, keeping in mind everything discussed above, if VOCs with a high IE are measured with an 11.7eV lamp calibrated to propane, the existing CFs in TN-106 still could be used, if requirements for the measurement accuracies aren t high. CFs from TN-106 could simply be recalculated to propane as calibration gas, using the formula below the table. The recalculated CFs can be used in the instruments as CFs for the Custom Gases. This option exists in the instrument menu and is described in the instrument manual. For the accurate measurements, CFs developed for the most-often used VOCs in combination with 11.7eV lamps calibrated to propane can be taken from column 6 of the table below. They also can be saved in the instrument as a Custom Gas option. Accurate customer CFs can be developed on request and per the settled procedure in TN Note: Based on the data above, Honeywell RAE Systems strongly recommends using 100 ppm propane as calibration gas for instruments with 11.7eV lamps when the purpose is to measure VOCs with IE above 10.6eV. The statement that the best policy is Calibration gas = Measurement gas is still valid and is preferred. Ionization Energy, ev TN-106 New Lamp CFs to IBE Mature Mature vs. Lamp New lamp TN-106* New lamp CFs to Propane Mature Mature vs. Lamp New lamp Gas Chlorine Propane Methanol Formaldehyde ,1-Dichlorethane ,1,2-11 Trichloroethane Chloroform Ammonia Ethylene oxide
5 * CFs of VOCs relative to propane (column 5 of the table) were calculated based on the CFs of the VOCs relative to isobutylene from TN-106 data, using the formula CF Pr = CF 1 IBE /CF 2 IBE, where CF 1 is the CF of VOC relative to isobutylene and CF 2 is CF of propane relative to isobutylene. All CFs in TN-106 were developed for the freshly made 11.7eV lamps and placed into the column 2 of the table. Column 3 has newly measured CF data for mature 11.7eV lamps for the same VOCs calibrated to IBE. Columns 4 and 7 shows CFs changing of VOCs, measured for new and mature 11.7eV lamps. Column 4 comprises data for the VOCs calibrated with 100 ppm IBE (isobutylene). Column 7 comprises data for the VOCs calibrated with 100 ppm propane. Column 6 includes newly developed CFs of the VOCs calibrated with 100 ppm propane and most often used together with mature 11.7eV lamps. Lamp Maturing Behavior of 11.7eV lamp vs. time was monitored with the lamps from batch 3, the current manufacturing process. Data are plotted in the figure below. Raw counts eV lamp sensitivity drop vs. time Exposure gases concentrations normalized IBE Pr Cl Hrs in test The test was running for over 500 hours in MiniRAE 3000 instruments. This corresponds with approximately 2 months usage, 8 hours per day, running continuously. Lamps show the typical behavior when intensity of the lamp output drops significantly during the burning period and the first hours of work, and then stabilizes at some level while lamps continue running. The end of a lamp s usable life is considered when the lamp cannot be turned on or it fails calibration. Lamp Maintenance For proper lamp function, regular maintenance such as periodic lamp cleaning with IPA (isopropyl alcohol) included in the RAE Systems PID cleaning kit (p/n ) together with other actions, is a good practice. Typical procedures for maintenance are outlined in the instrument manuals and in Technical Note TN A special precaution for 11.7eV lamps is avoiding high humidity or keeping the instruments in such conditions as little as possible. Do not use acetone for lamp cleaning. Never remove depositions from the lamp with just mechanical rubbing, which created scratches on the crystal surfaces and decreases lamp efficiency and can even permanently damage the lamp. Keeping in mind the rather short life of 11.7eV lamps, turn off instruments when they are not in use. And, always store instruments in a dry area. 5
6 References 1 The PID Handbook, p.7, Honeywell RAE Systems Factors-for-Volatile-Compounds-with-RAE-Systems-Instruments_12-13.pdf 4 Moisture-Sensitivity_08-15.pdf 6
Questions, Myths and Misconceptions about Using Photoionization Detectors
Questions, Myths and Misconceptions about Using Photoionization Detectors Solvent, fuel and other VOC vapours are pervasively common in many workplace environments. Increased awareness of the toxicity
More informationRS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers
RS DYNAMICS ECOPROBE 5 Portable IR/PID Gas Analyzer PID + IR PID and IR Analyzers General ECOPROBE 5 has two autonomous analyzers in one case. The combination of analyzers provides a set of data designed
More informationRoHS. Specification CUD8DF1A. Drawn Approval Approval. 서식 Rev: 00
Specification RoHS CUD8DF1A SVC Customer Drawn Approval Approval 1 [ Contents ] 1. Description 2. Outline dimensions 3. Characteristics of CUD8DF1A 4. Characteristic diagrams 5. Binning & Labeling 6. Reel
More informationVolatile Organic Compounds (VOC) Probes
Volatile Organic Compounds (VOC) Probes New Plug and Play accessory probes Compatible with the new 9565 VelociCalc and 7575 Qtrak Also compatible with the 9555 VelociCalc and 7565 Qtrak Firmware must be
More informationUV3000. Accurate, precise, and portable ambient gas point analyzer
UV3000 Accurate, precise, and portable ambient gas point analyzer The Cerex UV3000 is a multifunction analyzer designed to detect part per billion (ppb) to percent level concentrations of multiple gases
More informationSpecification SPW08F0D
Specification SPW08F0D Drawn SSC Approval Customer Approval SPW08F0D 1. Description 2. Absolute Maximum Ratings 3. Electro Optical Characteristics 4. Characteristic Diagram 5. Reliability 6. CIE Chromaticity
More informationRoHS. Specification CUD8AF1C. 서식 Rev: 00
Specification RoHS CUD8AF1C 1 [ Contents ] 1. Description 2. Outline dimensions 3. Characteristics of CUD8AF1C 4. Characteristic diagrams 5. Binning & Labeling 6. Reel packing 7. Recommended solder pad
More informationExperiment 4 Radiation in the Visible Spectrum
Experiment 4 Radiation in the Visible Spectrum Emission spectra can be a unique fingerprint of an atom or molecule. The photon energies and wavelengths are directly related to the allowed quantum energy
More information(PT-A6) Visible Light Detector for Security Lighting
Version: July 4, 2017 (PT-A6) Visible Light Detector for Security Lighting Token Electronics Industry Co., Ltd. Web: www.token.com.tw Email: rfq@token.com.tw Taiwan: No.137, Sec. 1, Zhongxing Rd., Wugu
More information(PT-A6) Visible Light Detector for Security Lighting
Version: July 31, 2017 Electronics Tech. (PT-A6) Visible Light Detector for Security Lighting Web: www.direct-token.com Email: rfq@direct-token.com Direct Electronics Industry Co., Ltd. China: 12F, Zhong
More informationph electrode Instruction Manual For use with the RAH-210 & RPH-250
ph electrode Instruction Manual For use with the RAH-210 & RPH-250 The information in this manual was current at the time of printing. The most current versions of all Hydro Instruments manuals can be
More informationAchim J. Lilienthal Erik Schaffernicht Achim J. Lilienthal. AASS, Örebro University
Achim J. Lilienthal Erik Schaffernicht Mobile Achim J. Achim J. Lilienthal Room T1222 Robotics and Lilienthal Olfaction Room T1211 Lab, Erik Schaffernicht achim.lilienthal@oru.se AASS, Örebro University
More informationTechnical/Application Article 01 Version 1.2
Technical/Application Article 01 Version 1.2 th 25 January 2017 WRH What is a PID? Introduction PID is the abbreviation for Photo-Ionization Detector. A PID is a hand-held, personal, or fixed wallmounted
More informationTechnical Data Sheet. Pb Free. Specification GR101 SSC. Customer. Rev. 02 January 서식번호 : SSC- QP (Rev.0.
Specification GR101 Pb Free Drawn SSC Approval Customer Approval [ Contents ] 1. Description 2. Absolute maximum ratings 3. Electro-Optical characteristics 4. Characteristic diagrams 5. Reliability result
More informationSpecification KWT803-S
Specification KWT803-S Drawn SSC Approval CUSTOMER Approval Contents 1. Description 2. Absolute Maximum Ratings 3. Electro-Optical Characteristics 4. Characteristic Graph 5. Reliability Test Item and Condition
More informationSurface Mount UV LED. NUVA33 Series PART NUMBERING SYSTEM. WAVELENGTH CODES Code Nominal Wavelength
FEATURES SURFACE MOUNT 3.4mm x 3.4mm x 2.37mm WAVELENGTH 365 ~ 45nm FOR UV CURING, PHOTO CATALYST & SENSOR LIGHTING RoHS COMPLIANT COMPATIBLE WITH REFLOW SOLDERING TAPE AND REEL PACKAGING SPECIFICATIONS
More informationThis sensor is now also available with a sturdier flat electrode which enables testing even smaller samples than the conventional electrode.
ph sensor with electrode ENPH-A016 ph sensor with flat electrode ENPH-F052 The ph sensor replaces the traditional ph meter adding powerful capabilities such as collecting the ph data and displaying the
More informationINSTRUCTION MANUAL. Type: ORP500 C, ORP500 CD, ORP222 CD, ORP223 CD EN Introduction Helpful Operating Tips. 2
ORP/Reference Electrodes INSTRUCTION MANUAL Type: ORP500 C, ORP500 CD, ORP222 CD, ORP223 CD EN 02-05 Table of Contents 1. Introduction. 2 2. Helpful Operating Tips. 2 3. Calibration Procedure 3 3.1. Laboratory
More informationSERVOTOUGH SpectraScan
SERVOTOUGH SpectraScan Continuous, real-time analysis of light hydrocarbons Jan Hordijk - Servomex Hydrocarbon stream analysis 5 major technologies used in HydroCarbon land for gas: IR (Infra Red) Laser
More informationSIFT-MS. technology overview
SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY technology overview contents 1. An overview of SIFT-MS 2. Principles of SIFT-MS a. Why chemical ionization? b. Why three reagent ions? c. How does SIFT-MS
More informationCombustible Gas Catalytic Bead (0-100 %LEL) Part No FM Performance Certified 1,4 FM Performance Certified 1
Sensor Data Sheet Document No. 365-2211-31 (Rev D) Combustible Gas Catalytic Bead (0-100 %LEL) Part No. 823-0211-31 FM Performance Certified 1,4 FM Performance Certified 1 Minimum Indicated Concentration...
More informationSolar irradiance measurement up to 2500nm with the Arcoptix FT-NIR
Application note Solar irradiance measurement up to 2500nm with the Arcoptix FT-NIR Introduction Applications that use spectrometers to measure the light energy of radiant sources require an irradiance-calibration,
More informationSodium Sealed Electrode Sensor Bundle Product Number: ENSOD051
imagine explore learn Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051 Overview Sodium is highly abundant in the lithosphere (Earth s outer shell) where it occurs as rock salt deposits as
More informationReprint. Project Summary. Side by Side Evaluation of ChemScan UV-4100 and Hach (Dr. Lange) Optiquant at Littleton-Englewood Colorado
ChemScan Process Analyzer Reprint Project Summary Side by Side Evaluation of ChemScan UV-4100 and Hach (Dr. Lange) Optiquant at Littleton-Englewood Colorado Published by: Applied Spectrometry Associates,
More informationONYX -MCE MULTI-CHANNEL OPTICAL FIBER PYROMETERS WITH ACTIVE EMISSIVITY COMPENSATION PRECISION TEMPERATURE MEASUREMENT FOR DEMANDING INDUSTRIAL
ONYX -MCE MULTI-CHANNEL OPTICAL FIBER PYROMETERS WITH ACTIVE EMISSIVITY COMPENSATION PRECISION TEMPERATURE MEASUREMENT FOR DEMANDING INDUSTRIAL APPLICATIONS Accurate, repeatable, and reliable temperature
More informationclose to performances of the turbine gas meters designed for custody transfer measurements.
QUANTOMETER DESIGN AND FUNCTION CPT Quantometers have been designed in order to provide our customers with reliable and inexpensive measuring instruments for secondary flow measurements. Our great experience
More informationSIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY TECHNOLOGY OVERVIEW \ SYFT TECHNOLOGIES
SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY TECHNOLOGY OVERVIEW \ SYFT TECHNOLOGIES CONTENTS 1. An overview of SIFT-MS 2. Principles of SIFT-MS a. Why chemical ionization? b. Why multiple reagent
More informationIntroduction to Infrared Thermometry
TS-104 Introduction to Infrared Thermometry Fig. 1 - Blackbody Radiation Characteristics General Infrared thermometers have the ability to measure temperature without physical contact. The ability to accomplish
More informationEAHP1215WA2. Features. Description. Applications
Features LM-80 Certified High Power COB & High CRI LED Multi-Chip Solution Dimension:12 mm x 15 mm x 1.6 mm Main Parameters: Luminous Flux, Forward Voltage, Chromaticity and Color Rendering Index RoHS
More informationFAIMS Technology at a Glance
FAIMS Technology at a Glance Field asymmetric ion mobility spectrometry (FAIMS), also known as differential mobility spectrometry (DMS), is a gas detection technology that separates and identifies chemical
More information[Agilent Technologies] [Electrochemistry Products] Quick Reference
Overview Using I9121 Ammonia Gas Sensing Probe on 3200I/M This document demonstrates how to use I 9121 Ammonia Gas Sensing Probe on 3200I or 3200 M to measure pn or the concentration of [NH 4 + ]. Applied
More informationAdvanced Sensor Technologies, Inc. TEL: TOLL FREE: WOW-ASTI ( ) FAX:
IMPORTANT NOTES FOR FLUORIDE ISE MEASUREMENT SYSTEM WITH 56 ANALYZER IMPORTANT NOTE # 1: This addendum ONLY applies to the 2 nd generation Model 56 ISE Analyzers with the 2.19 software loaded. This addendum
More informationSTATIC GAS MONITOR Type SGM/DEW Revision A of 20 aprile 2016
APPLICATIONS Moisture monitoring of air or gas (SF6) Suitable for indoor or outdoor Industrial, medical or aerospace fields HIGHLIGHTS High voltage circuit breakers commonly used for distribution and transmission
More informationAmplified Pyranometer SP-212 & 215
Amplified Pyranometer SP-212 & 215 Amplified Pyranometer Sensor This sensor is calibrated to measure total shortwave radiation. The evaporation of water from soil and the transpiration of water from plant
More informationRapid RH Portable ph Meter. User's Manual
Rapid RH Portable ph Meter User's Manual This page intentionally left blank. 2 Contents 1. Introduction... 4 2. Included with the Rapid RH Portable ph Meter... 4 3. Understanding ph... 5 4. Calibrating
More informationANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis
DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes
More informationAASS, Örebro University
Mobile Robotics Achim J. and Lilienthal Olfaction Lab, AASS, Örebro University 1 Contents 1. "Classical" Electronic Nose 2. Further Gas Sensing Technologies for Mobile Robots 3. (Signal Processing in Electronic
More informationTraining Coarse ECOPROBE 5 CHEMISTRY RS DYNAMICS
Training Coarse ECOPROBE 5 CHEMISTRY Contents Ecoprobe analytical principles Photoionization Infrared spectroscopy Oxygen measurement Pollutants Types and responses Head space New application of Ecoprobe
More informationph Probe Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth 1 99 C
V 2.8 Revised 1/16/18 ph Probe Silver / silver chloride Reads ph Range 0 14 Resolution Response time Max pressure Max depth Temperature range C Cable length Internal temperature sensor Time before recalibration
More informationPC100 Handheld ph/cond./tds/sal. Meter. Instruction Manual
PC100 Handheld ph/cond./tds/sal. Meter Instruction Manual Table of Contents 1. Introduction ------------------------------------------------------------------------------------------------------------------------
More informationExtrel is widely respected for the quality of mass spectrometer systems that are
Extrel is widely respected for the quality of mass spectrometer systems that are available to the world's top research scientists. In response to increasing requests for complete turn-key systems built
More informationTLRE1002A(T02), TLSE1002A(T02), TLOE1002A(T02), TLYE1002A(T02), TLPYE1002A(T02), TLGE1002A(T02), TLFGE1002A(T02), TLPGE1002A(T02)
TL(RE,SE,OE,YE,PYE,GE,FGE,PGE)02A(T02) TOSHIBA LED Lamps TLRE02A(T02), TLSE02A(T02), TLOE02A(T02), TLYE02A(T02), TLPYE02A(T02), TLGE02A(T02), TLFGE02A(T02), TLPGE02A(T02) Panel Circuit Indicator Unit:
More informationTLRU1002A(T02), TLSU1002A(T02), TLOU1002A(T02), TLAU1002A(T02),TLYU1002A(T02), TLGU1002A(T02), TLPGU1002A(T02)
TL(RU,SU,OU,AU,YU,GU,PGU)02A(T02) TOSHIBA LED Lamps TLRU02A(T02), TLSU02A(T02), TLOU02A(T02), TLAU02A(T02),TLYU02A(T02), TLGU02A(T02), TLPGU02A(T02) Panel Circuit Indicator Unit: mm Surface-mount devices
More informationOXEA - Online Elemental Analyzer
02 25 08 OXEA - Online Elemental Analyzer OXEA (Online X-ray Elemental Analyzer) is based on the X-ray fluorescence technology (XRF) which is well known in the laboratory field. With the aid of a patented
More informationPAR Meter Instruction Manual. Product # Product # Year Warranty
PAR Meter Instruction Manual Product #748200 Product #748205 1 Year Warranty Measuring PPF with Quantum Meter Photosynthesis is driven by the number of photons between 400 and 700 nanometers (nm). This
More informationCerno Application Note Extending the Limits of Mass Spectrometry
Creation of Accurate Mass Library for NIST Database Search Novel MS calibration has been shown to enable accurate mass and elemental composition determination on quadrupole GC/MS systems for either molecular
More informationIR F ( C) ENGLISH INSTRUCTION MANUAL. Thermometer
INSTRUCTION MANUAL Thermometer DUAL-LASER TARGETING 12:1 DISTANCE- TO-SPOT RATIO AUTO-SCAN MAX/MIN/AVG/DIFF HIGH/LOW ALARM BACKLIT DISPLAY AUTO POWER-OFF ADJUSTABLE EMISSIVITY IR5 2m -22 752 F (-30 400
More informationAMOLEDS 3020 LED AB - WH302009AZ-c Datasheet
AMOLEDS 3020 LED AB - WH302009AZ-c Datasheet AMOLEDS LEDs is designed for high current and high luminous power application fields. Superior thermal conductivity and high luminous efficiency based on ceramic
More informationReview: Light and Spectra. Absorption and Emission Lines
1 Review: Light and Spectra Light is a wave It undergoes diffraction and other wave phenomena. But light also is made of particles Energy is carried by photons 1 Wavelength energy of each photon Computer
More informationAPCPCWM_ :WP_ WP_ RoHS. Specification. SSC-SZ5-P series. Document No. : SSC- QP
Specification RoHS SSC-SZ5-P series 1 Technical Z-Power LED Data X10490 Sheet Description SZ5-P series The Z-Power series is designed for high current operation and high flux output applications. It incorporates
More informationOwlstone Nanotech Inc
Owlstone Nanotech Inc PIDplus Technical Paper USA POC: Mark Brennan mark.brennan@owlstonenanotech.com 400 Rella Boulevard Suite 160 Suffern, NY 10901 Tel: +1 845-533-4225 Fax: +1 845-533-4232 www.owlstonenanotech.com
More informationShort-Period Air Sampling by Direct-reading Instruments
Short-Period Air Sampling by Direct-reading Instruments OHAO Meeting, Toronto, October 13, 2011 Ian Drummond and Paul Bozek, Dalla Lana School of Public Health, University of Toronto Program for today's
More informationFar UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers
Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N O T E Summary The far UV region (generally covering
More informationBIOO FOOD AND FEED SAFETY. Histamine Enzymatic Assay Kit Manual. Catalog #: Reference #:
BIOO FOOD AND FEED SAFETY Histamine Enzymatic Assay Kit Manual Catalog #: 1032-05 Reference #: 1032-05 BIOO Scientific Corp. 2010 TABLE OF CONTENTS GENERAL INFORMATION... 1 Product Description... 1 Procedure
More informationAgilent Cary 630 FTIR sample interfaces combining high performance and versatility
Agilent Cary 630 FTIR sample interfaces combining high performance and versatility Technical Overview Introduction The Agilent Cary 630 is a robust, easy to use, superior performing FTIR in a compact package.
More informationSUNLIKE DISPLAY. Model No: SC2004C GENERAL SPECIFICATION
GENERAL SPECIFICATION ITEM Product No SC2004CSLB-XA-LB-G DESCRIPTION STN Gray Positive STN Yellow Green Positive STN Blue Negative LCD Type TN Negative TN Positive FSTN Negative White & Black FSTN Positive
More informationPhysics 1C OPTICAL SPECTROSCOPY Rev. 2-AH. Introduction
Introduction In this lab you will use a diffraction grating to split up light into its various colors (like a rainbow). You will assemble a spectrometer, incorporating the diffraction grating. A spectrometer
More informationNational Optical & Scientific Instrument Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210)
National Optical & Scientific Instrument Inc. 6508 Tri-County Parkway Schertz, Texas 78154 Phone (210) 590-9010 Fax (210) 590-1104 INSTRUCTIONS FOR SHOP MICROSCOPES MODEL NUMBERS 186 187 188 189 National
More informationLecture 6. Rapid Thermal Processing. Reading: Chapter 6
Lecture 6 Rapid Thermal Processing Reading: Chapter 6 (Chapter 6) Categories: Rapid Thermal Anneal (RTA) Rapid Thermal Oxidation (RTO) Rapid Thermal Nitridation (RTN) (and oxynitrides) Rapid Thermal Diffusion
More information15 E OWNER S MANUAL PIRATE SPYGLASS BRING EM NEAR PIRATE SPYGLASS TELESCOPE 5 W 30 W 60 W 45 W
3 0 15 15 E OWNER S MANUAL PIRATE SPYGLASS BRING EM NEAR PIRATE SPYGLASS TELESCOPE 5 W 60 W 30 W 45 W Zhumell customers know that there are plenty of ways to experience the world. They also understand
More informationSpecification YGHR411-H
Specification HR411H Pb Free Drawn SSC Approval Customer Approval [ Contents ] 1. Description 2. Absolute maximum ratings 3. ElectroOptical characteristics 4. Characteristic diagrams 5. Reliability result
More informationDHS1900 DHS3000 High Temperature Hand Held Infrareds User Manual
DHS1900 DHS3000 High Temperature Hand Held Infrareds User Manual WD1063 Rev B Revised 02/23/15 Wahl Instruments Inc. 234 Old Weaverville Road Asheville, NC 28804 Toll Free: 800-421-2853 Phone: 828-658-3131
More informationPREFERRED RELIABILITY PRACTICES. Practice:
PREFERRED RELIABILITY PRACTICES Practice No. PD-ED-1239 Page 1 of 6 October 1995 SPACECRAFT THERMAL CONTROL COATINGS DESIGN AND APPLICATION Practice: Select and apply thermal coatings for control of spacecraft
More informationThe Claviature of Gas Analysis
The Claviature of Gas Analysis Mass spectrometric gas analysis for quality assurance of gases and gas mixtures High purity and special gases are becoming increasingly important in the development of modern
More informationNatural Rock Sample Manual
Natural Rock Sample Manual Revision 2014-06-12 DURRIDGE Company Inc. 524 Boston Road Billerica, MA 01821 Tel: (978) 667-9556 Fax: (978) 667-9557 service@durridge.com www.durridge.com 2014, DURRIDGE Company
More informationSpecification KWT803-S
Specification KWT803-S SSC Customer Drawn Approval Approval CONTENTS 1. Description 2. Absolute Maximum Ratings 3. Electro-Optical Characteristics 4. Optical characteristics 5. Reliability Test 6. Color
More informationMultilayer Ceramic Chip Capacitors
HIGH VOLTAGE SERIES JARO high voltage series Multilayer Ceramic Capacitors are constructed by depositing alternative layers of ceramic dielectric materials and internal metallic electrodes, by using advanced
More informationSpecification STW8Q2PA
Specification STW8Q2PA SSC CUSTOMER Drawn Approval Approval CONTENTS 1. Description 2. Absolute Maximum Ratings 3. Electro-Optical Characteristics 4. Optical characteristics 5. Reliability Test 6. Color
More informationSpecification STW8T36B
Specification STW8T36B SSC CUSTOMER Drawn Approval Approval CONTENTS 1. Description 2. Absolute Maximum Ratings 3. Electro-Optical Characteristics 4. Optical characteristics 5. Reliability Test 6. Color
More informationEppendorf UVette. Instructions for use
Instructions for use Instructions Eppendorf see Fig. Tab. p. p. UVette for use Copyright 2014 Eppendorf AG, Hamburg, Germany. All rights reserved, including graphics and images. No part of this publication
More informationFast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates
APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Lee Marotta Tom Kwoka David Scott Miles Snow Toby Astill PerkinElmer, Inc. Shelton, CT Fast, Quantitative Analysis of Residual Solvents in
More informationACTIVITY 3 Introducing Energy Diagrams for Atoms
Name: Class: SOLIDS & Visual Quantum Mechanics LIGHT ACTIVITY 3 Introducing Energy Diagrams for Atoms Goal Now that we have explored spectral properties of LEDs, incandescent lamps, and gas lamps, we will
More informationFar UV Absorbance Detector
Far UV Absorbance Detector Theory Most organic and inorganic species absorb strongly in the far UV. Notable exceptions are the inert gases, helium and nitrogen which absorb very weakly in this region.
More informationFischer 1508BTH-45 5" Brass Barometer with Temperature & Humidity User Manual
Fischer 1508BTH-45 5" Brass Barometer with Temperature & Humidity User Manual Table of Contents 1. Introduction... 2 2. Care and Cleaning... 2 3. Barometer Operation... 2 3.1 How the aneroid barometer
More informationAppendices towards the Operation Manual of Ultrasonic milk analyzer Lactoscan
APPENDIX 7 PH MEASURING 1. General information PH probe is a unit, measuring the solution acidity or alkalinity degree. It is measured on scale of 0 to 14. The term ph is derived from "p", the mathematical
More informationph Probe Spear Tip Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth
V 2.4 Revised 3/1/18 Spear Tip ph Probe Silver / silver chloride Reads ph Range 0 14 Resolution Response time Max pressure Max depth Temperature range C Cable length Internal temperature sensor Time before
More informationUnderstanding catalytic LEL combustible gas sensor performance
: Understanding catalytic LEL combustible gas sensor performance These four conditions are frequently diagrammed as the "Fire Tetrahedron". If any side of the tetrahedron is missing, incomplete or insubstantial;
More informationSpectrometer User s Guide
Spectrometer User s Guide (Order Codes: V-SPEC, SPRT-VIS, SP-VIS, SP-UV-VIS, ESRT-VIS) The spectrometer is a portable light spectrophotometer, combining a spectrometer and a light source/cuvette holder.
More informationSpecification LW514. Drawn Approval Approval. Rev. 02 December 서식번호 : SSC-QP (Rev.00)
Specification LW514 SSC CUSTOMER Drawn Approval Approval CONTENTS 1. Full code of Lamp LED series 2. Feature & Application 3. Absolute Maximum Ratings 4. Electro Characteristics 5. Outline Dimension 6.
More informationSlide 1. Slide 2. Slide 3. Choosing and Using the Best Sensor Technologies for Marine Chemistry Applications
Slide 1 Choosing and Using the Best Sensor Technologies for Marine Chemistry Applications Bob Henderson, BS, MBA GFG Instrumentation Inc. Ann Arbor, MI Toll free: 800-959-0329 Mobile: 734-355-1291 E-mail:
More informationMARINE BINOCULARS ZHUMELL 7X50 MARINE BINOCULARS WITH COMPASS AND RETICLE
0 OWNER S MANUAL MARINE BINOCULARS ZHUMELL 7X50 MARINE BINOCULARS WITH COMPASS AND RETICLE 75 W 60 W 30 W 45 W Zhumell customers know that there are plenty of ways to experience the world. They also understand
More informationMultilayer Nuclear Track Detectors for Retrospective Radon Dosimetry
Multilayer Nuclear Track Detectors for Retrospective Radon Dosimetry V. V. Bastrikov 1, M. V. Zhukovsky 2 1 Experimental Physics Department, Ural State Technical University, Mira St., 19/5, 620002, Ekaterinburg,
More informationPYCKO SCIENTIFIC LIMITED
PYCKO SCIENTIFIC LIMITED 31 London Road, Grantham, NG31 6EX Telephone 01476 401992 bill@pycko.co.uk www.pycko.co.uk Your Alternative To The Obvious RayMon10 TM Detect, measure, identify and analyse radionuclides
More informationTraceable UV-Vis Reference Materials
UV-Vis Spectroscopy Traceable UV-Vis Reference Materials For Regulatory Compliance Introduction Many laboratories, particularly in the pharmaceutical industry, are under increasing pressure to show proof
More informationLED SPECIFICATION 530MY8C
LED SPECIFICATION 53MY8C Features Single color High bright output Low power consumption High reliability and long life Descriptions Dice material Emitting Color Super Bright Amber (Yellow) Device Outline
More informationNATURAL ROCK SAMPLE. Performance Monitoring Accessory for the RAD7 User Manual
NATURAL ROCK SAMPLE Performance Monitoring Accessory for the RAD7 User Manual TABLE OF CONTENTS TABLE OF CONTENTS 2 1 INTRODUCTION 3 2 RAD7 RADON SENSITIVITY CHECK 4 2.1 Purge the RAD7 5 2.2 Hook up to
More informationGXUAN V/3W V4 series
GXUAN 1313 36V/3W V4 series 烜 Introduction Everlight`s COB Series is an aluminum substrate based LED achieving high efficiency while maintaining high CRI at Energy Star / ANSI color temperature ranges.
More informationSPECIFICATION. Checked By Checked By Checked By Checked By Approval. Proposed By Checked By Checked By Checked By Approval
SPECIFICATION 1 Product : 5mm WHITE LAMP 2 Part No : 3 Customer : 4 Date : 2008. 06. 10 Ver.1.1 Customer : Checked By Checked By Checked By Checked By Approval Manufacturer : ITSWELL Co., LTD Proposed
More informationHL-800K Infrared Laser Thermometer. -50 C to +800 C (13:1 ratio) User Manual
HL-800K Infrared Laser Thermometer -50 C to +800 C (13:1 ratio) User Manual TABLE OF CONTENTS INTRODUCTION... 3 FEATURES... 3 WIDE APPLICATION RANGE... 3 SAFETY... 3 DISTANCE & SPOT SIZE... 4 SPECIFICATIONS...
More informationSpecification LY530. Drawn Approval Approval. Rev. 08 December 서식번호 : SSC-QP (Rev.00)
Specification LY530 SSC CUSTOMER Drawn Approval Approval CONTENTS 1. Full code of Lamp LED series 2. Feature & Application 3. Absolute Maximum Ratings 4. Electro Characteristics 5. Outline Dimension 6.
More informationThermography Pocket Guide. Theory Practice Tips & Tricks
Thermography Pocket Guide Theory Practice Tips & Tricks Copyrights, warranty and liability The information compiled in this Pocket Guide is protected by copyright. All rights belong exclusively to Testo
More informationRadiation Detection. 15 th Annual OSC Readiness Training Program.
Radiation Detection 15 th Annual OSC Readiness Training Program www.oscreadiness.org GM Detectors 15 th Annual OSC Readiness Training Program www.oscreadiness.org 1 A closer look 15 th Annual OSC Readiness
More informationGeneral Overview of Gas Filled Detectors
GAS-FILLED DETECTOR General Overview of Gas Filled Detectors Gas-Filled Detectors Ion chamber Proportional counter G-M (Geiger-Miller) counter Diagram of a Generic Gas-Filled Detector A Anode High-voltage
More informationDATA SHEET ES V-XX-XXX-XX CONTENTS
The Brightest LED DATA SHEET ES2835-023V-XX-XXX-XX CONTENTS 1. Introduction and features 2. Applications 3. Package dimensions 4. Absolute maximum ratings 7. Characteristic and curves 8. Reflow solder
More informationSPECIFICATION. Topview 5550 SMD LED IWS 505-UG-K3. Product : Topview 5550 SMD LED Part No : IWS-505-UG-K3 Customer : Date : Ver.1.
SPECIFICATION Product : Part No : Customer : Date : 2008. 06. 10 Ver.1.0 Customer : Approval Manufacturer : ITSWELL Co., LTD Proposed By Approval Comment Suwon Company : 442-190, 802 Uman Industrial Comples,
More informationIV. DIRECT MEASUREMENT LABS. Introduction
IV. DIRECT MEASUREMENT LABS VI-1 I. Purpose Introduction An industrial hygienist frequently needs to evaluate airborne gas and vapor concentrations "on the spot". In other cases, he needs to track down
More informationUser s Guide. Waterproof Palm ph Meter Model PH220. Introduction
User s Guide Waterproof Palm ph Meter Model PH220 Introduction Congratulations on your purchase of the Extech PH220 waterproof ph/mv meter. This microprocessor-based device with tactile buttons is battery
More informationImmediate Free Chlorine Correction for Fluctuating Flow & Changes in Water Quality. Sadiq Khan May 25th, 2016
Immediate Free Chlorine Correction for Fluctuating Flow & Changes in Water Quality Sadiq Khan May 25th, 2016 3 Main Current Methods Of Dosing Chlorine Manual dosing with adjustment based on sampling Online
More informationFischer Banjo Weather Station with Thermometer, Hygrometer, Barometer User Manual
Fischer 4673-22 Banjo Weather Station with Thermometer, Hygrometer, Barometer User Manual Table of Contents 1. Introduction... 2 2. Care and Cleaning... 2 3. Barometer Operation... 2 3.1 How the aneroid
More informationHebei I.T. (Shanghai) Co., Ltd
Hebei I.T. (Shanghai) Co., Ltd LED SPECIFICATION Part No.13PG2C Features/ Single color/ High bright output/ t Low power consumption/ High reliability and long life/ Descriptions/ Dice material/ingan Emitting
More information