Eppendorf UVette. Instructions for use
|
|
- Solomon Dorsey
- 6 years ago
- Views:
Transcription
1 Instructions for use
2 Instructions Eppendorf see Fig. Tab. p. p. UVette for use Copyright 2014 Eppendorf AG, Hamburg, Germany. All rights reserved, including graphics and images. No part of this publication may be reproduced without the prior permission of the copyright owner. Eppendorf and the Eppendorf logo are registered trademarks of Eppendorf AG, Hamburg, Germany. UVette is a registered trademark of Eppendorf AG, Hamburg, Germany. Registered trademarks and protected trademarks are not marked in all cases with or in this manual. U.S. Patent No. 6,249,345 English Instructions (EN) for use /032014
3 Operating instructions 3 1 Operating instructions 1.1 Using this manual Before using the consumables for the first time, read these instructions for use and the operating manual of the device that you use the consumables with. You can find the current version of the operating manual on the Internet at These instructions for use do not replace the device operating manual. 2 Product description 2.1 Features The Eppendorf UVette is a plastic cuvette for single use. For measurements in the UV/Vis range at wavelengths from 220 nm to 1600 nm. For volumes of 50 μl or more. The funnel-shaped design prevents a capillary effect: liquid remains in the center of the measuring window, even with small volumes. 4 optical surfaces 2 optical path lengths in one cuvette: 2 mm and 10 mm. The 10 mm optical path length is marked by an arrow. Rotating the cuvette 90 allows you to select between a path length of 10 mm or 2 mm. Abb. 1:Eppendorf UVette optical path lengths 90 Optical path length 10 mm 2 mm Light path height 8.5 mm Fig. 1: Eppendorf UVette optical path lengths Compatible photometer/spectrophotometer The Eppendorf UVette fits in the cuvette shaft of all Eppendorf BioPhotometers and Eppendorf BioSpectrometers. Adapters for various light path heights are available at Eppendorf for filter photometers and spectrophotometers from other manufacturers.
4 4 Product description 2.2 Sales units Eppendorf UVette, individually packaged Eppendorf UVette, routine pack Starter set Individually blister-packaged Protection from contamination and dust For measurements with the highest cuvette purity standards Valuable samples can be recovered after the measurement For routine applications Fast, easy cuvette access Safe storage in a resealable box 80 individually packaged Eppendorf UVettes and one universal adapter for using the Eppendorf UVette in devices from other manufacturers 2.3 Eppendorf purity grades Eppendorf Quality PCR clean Protein-free Batch testing (certified) for the following purity criteria Free of human DNA DNase-free RNase-free Free of PCR inhibitors Protein-free Eppendorf UVette Eppendorf UVette, individually packaged Eppendorf UVette, routine pack Starter set
5 Safety 5 3 Safety 3.1 Intended use The Eppendorf UVette is a UV-transparent plastic cuvette for single use. It is used for photometric analysis of liquids in the UV range and the entire VIS range. The Eppendorf UVette is used in training, routine and research laboratories, and in the life sciences, industry, and chemistry. In-vivo applications (in or on the human body) are not allowed Operation requires skilled personnel who have been trained for the above mentioned areas. 3.2 Warnings for intended use WARNING! Damages to health due to infectious liquids and pathogenic germs. When handling infectious liquids and pathogenic germs, observe the national regulations, the biological security level of your laboratory, the material safety data sheets, and the manufacturer's application notes. Wear personal protective equipment. For full instructions regarding the handling of germs or biological material of risk group II or higher, please refer to the "Laboratory Biosafety Manual" (Source: World Health Organization, current edition of the Laboratory Biosafety Manual). 4 Operation 4.1 Inserting the cuvette Prerequisites The cuvette is free from contamination by dust or fingerprints and free from scratches. The measuring volume in the cuvette is sufficient. Ensure that the minimum measuring volume has been reached. The measuring solution is free of particles and bubbles. NOTICE! Material damage due to improper handling The cuvette may fall if it is not handled properly. Liquid may spray out. Store the cuvette in a cuvette stand. Do not touch the optical surfaces in the lower part of the Eppendorf UVette. Do not touch the corners of the cuvette (protective gloves may become damaged).
6 6 Operation Eppendorf BioPhotometer/Eppendorf BioSpectrometer 1. Position the cuvette so the required optical path length points toward the light path. 2. When inserting the cuvette, press it completely to the bottom against the slight resistance Inserting the cuvette and adapter in photometers from other manufacturers Function of the adapter Adjust the height of the light beam in the corresponding photometer. Diaphragm function: Hides the light bundle on the cross-section of the Eppendorf UVette measuring window. Reduces stray light Like measurements with commercial, quartz glass ultra-micro cuvettes, hiding the light bundle reduces the upper limit of the photometric measuring range. Therefore, the usable, linear measuring range may be lower than the value specified for the photometer. To calculate the highest absorbance values that can be measured with the adapter/eppendorf UVette, determine the measuring range by measuring a dilution series of the desired wavelength. Additional information is available from the website Precise and reproducible positioning of the adapter and cuvette in the cuvette holder is a decisive factor for the reproducibility of the measurement. 1. Select the adapter that matches the light path height of the device. 2. Insert the Eppendorf UVette in the adapter. 3. Insert the adapter and Eppendorf UVette in the cuvette holder of the photometer. Firmly press the Eppendorf UVette downward. The Eppendorf UVette and adapter must not have any lateral play in the cuvette shaft of the device. 4.2 Measuring blanks and samples NOTICE! Faulty measured values due to incorrect orientation of the cuvette. Always insert the cuvette in the same direction in a series of measurements. Do not rotate the cuvette 180 during a measurement series.
7 Operation 7 NOTICE! Faulty measured values due to repeated use of the cuvette. If the cuvette is used several times, material changes can lead to incorrect measured values. Each cuvette may only be used for a single measuring procedure with one blank value and one sample value. The cuvette blank must be determined again for each cuvette. Limited tempering efficiency The tempering efficiency is limited by the contour of the UVette. If temperature control is required, check to see if the UVette is suitable for your applications. 1. Set the parameter for the cuvette path length in the device. 2. Determine the blank value of the cuvette. Abb. 2:Absorption of the water-filled Eppendorf UVette Absorbance Wavelength (nm) Fig. 2: Absorption of the water-filled Eppendorf UVette 3. Fill the cuvette with the sample. Make sure that liquid does not spray out when solutions are mixed in the cuvette. 4. Measure the sample. Do not use the cuvette at wavelengths below 220 nm or above nm.
8 8 Troubleshooting 5 Troubleshooting 5.1 Error search Symptom/ Cause message Measured values are imprecise. Measured values are not plausible. The blank value and sample value have been measured in different cuvettes. The position of the Eppendorf UVette in the device is not correct. Eppendorf UVette was inserted incorrectly. The Eppendorf UVette was used for more than one measurement. Remedy Measure the blank value and sample value in the same Eppendorf UVette. Press the Eppendorf UVette firmly onto the base of the cuvette shaft or adapter. The Eppendorf UVette must be inserted into the device in the same direction for the blank measurement and sample measurement. The Eppendorf UVette may not be rotated 180. Each Eppendorf UVette may only be used for one measurement (blank measurement or sample measurement). The light path height of the Use the adapter (not with the Eppendorf UVette does not Eppendorf BioPhotometer or match the light path height Eppendorf BioSpectrometer). of the device. The adapter may only be used with the light path height that matches the photometer. The optical path length of Insert the Eppendorf UVette in the the Eppendorf UVette does device so the required optical path not match the value set in length (2 mm/10 mm) is penetrated. the device. If the optical path length has been changed, the blank value also needs to be updated. Set the optical path length parameter in the device (2 mm/10 mm). The wavelength is below The Eppendorf UVette may only be 220 nm or above 1600 nm. used for wavelengths from 220 nm to 1600 nm.
9 Technical data 9 6 Technical data Tab. 1: Dimensions Width Depth Height Filling volume 12,5 mm (0.49 in) 12,5 mm (0.49 in) 36,0 mm (1.42 in) 50 μl 2000 μl Tab. 2: Photometric properties Path length 2 mm, 10 mm Light path height 8,5 mm Wavelength range 220 nm 1600 nm Cuvette blank 0,5 A at 260 nm (absorbance of the water-filled cuvette against air) Cuvette blank scattering s 0,015 A at 260 nm Tab. 3: Shelf life and storage Storage before use Shelf life Protect from sunlight and UV rays. Store dry at room temperature. 5 years The Eppendorf UVette is resistant to Acids Acetic acid 96% Perchloric acid 10% Nitric acid 65% Hydrochloric acid 36% Sulfuric acid 40% Trichloroacetic acid 40% Alkaline solutions Ammonia solution 25% Sodium hydroxide solution 50% Alcohols Ethanol Isopropanol Methanol Other reagents Acetone Acetonitrile DMSO 10% Formaldehyde 40% Water-saturated phenol
10 10 Ordering information The Eppendorf UVette is not resistant to Non-polar solvents Fuel Chloroform Diethyl ether Heptane Hexane Oleic acid n-pentane Petroleum ether Carbon tetrachloride Toluol All substances were tested for a period of 24 hours at an ambient temperature of 23 C ±2 C. 7 Ordering information Order no. (International) Order no. (North America) Description Eppendorf UVette 220 nm nm Original Eppendorf plastic cuvette, PCR clean, Protein-free μl, 80 pieces, individually packaged Eppendorf UVette routine pack 220 nm nm Eppendorf Quality μl, 200 pieces, reclosable box Cuvette stand for 16 cuvettes Adapter for photometers/spectrophotometer with light path height mm mm mm mm Eppendorf UVette starter set pieces of UVette, 1 universal adapter for light path height 15 mm, convertible to 8,5 mm Information on additional adapters can be found online at:
11
12 Evaluate your manual Give us your feedback. Your local distributor: Eppendorf AG Hamburg Germany
Eppendorf Vis Cuvettes. Instructions for use
Eppendorf Vis Cuvettes Instructions for use Copyright 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the copyright owner. Eppendorf, the Eppendorf
More informationstance et Helper Eppendorf Pipet Helper Chemical Resistance
stance et Helper Resistance N) Chemical Resistance Copyright 2017 Eppendorf AG, Germany. All rights reserved, including graphics and images. No part of this publication may be reproduced without the prior
More informationN) manual. Biomaster Operating manual
830 ual N) manual Operating manual Copyright 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the copyright owner. Trademarks Eppendorf and the
More information1 Product description. Material. 3 Warnings for intended use. Instructions for use ept.i.p.s. Version B. English (EN)
Instructions ept.i.p.s. see Fig. Tab. p. p. for Version use B 1 Product description The pipette tips ept.i.p.s. are single-use pipette tips for aspirating and dispensing liquids with air cushion pipettes.
More informationept.i.p.s. pipette tips are available with various degrees of purity, with and without filters, and as special tips.
Instructions ept.i.p.s. see Fig. Tab. p. p. for General use Lab Product 1 Product description The pipette tips are single-use pipette tips for aspirating and dispensing liquids with air cushion pipettes.
More informationEppendorf SmartBlock. Instructions for use
Instructions for use Instructions Eppendorf see Fig. Tab. p. p. SmartBlock for use Copyright 2016 Eppendorf AG, Germany. All rights reserved, including graphics and images. No part of this publication
More informationept.i.p.s. Instructions for use Version B
r N) s use for use Instructions for use Version B Copyright 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the copyright owner. Eppendorf and
More informationEppendorf ept.i.p.s. LoRetention Comparison of low retention pipette tips via simple and fast absorbance measurements in the BioPhotometer plus
APPLICATION NOTE No. 218 I November 2009 Eppendorf ept.i.p.s. LoRetention Comparison of low retention pipette tips via simple and fast absorbance measurements in the BioPhotometer plus Natascha Weiß 1,
More informationApplications Note 228 September 2010
Applications Note 228 September 2010 Reproducible photometric determination of DNA concentrations using the Eppendorf UVette in the Eppendorf BioPhotometer plus Martin Armbrecht, Eppendorf AG, Hamburg,
More informationThe Plus for your lab!
BioPhotometer plus NEW! analysis system The Plus for your lab! The easy, rapid and reliable BioPhotometer plus analysis system See it clearly! UVette Precision in plastics The patented* 1 UVette is the
More informationThe Pure Truth. Eppendorf Forensic DNA Grade according to ISO 18385
The Pure Truth Eppendorf Forensic DNA Grade according to ISO 18385 2 Eppendorf Forensic DNA Grade»Eppendorf Forensic DNA Grade, for results you can trust.«the prevention of contamination is one of the
More informationstance 4 Repeater M4 Register your instrument! Multipette M4 Repeater M4 Chemical Resistance
stance 4 Repeater M4 Resistance N) Register your instrument! www.eppendorf.com/myeppendorf Chemical Resistance Copyright 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without
More informationHigh-throughput Isolation of Genomic DNA From Buccal Swab on the Eppendorf epmotion 5075 VAC
APPLICATION NOTE No. 378 I August 2016 High-throughput Isolation of Genomic DNA From Buccal Swab on the Eppendorf epmotion 5075 VAC Hatim Boughanem, Manuel Macias-Gonzalez, BIOEASY SL, Spain Abstract The
More informationElectrical safety testing Checklist
ety al Check testing Check - Checklist EN) ctrical Electrical safety testing Checklist SOP - Electrical Check Copyright 2013 Eppendorf AG. All rights reserved, including graphics and images. No part of
More informationEppendorf BioSpectrometer 6135/6136/6137
nal Spectrometer Qualification 6135/6136/6137 - English N) rational Qualification - English Register your instrument! www.eppendorf.com/myeppendorf Eppendorf BioSpectrometer 6135/6136/6137 SOP - Operational
More informationNo 022 December Application-Specific Requirements for Consumables in the Laboratory
Userguide No 022 December 2011 Application-Specific Requirements for Consumables in the Laboratory Kornelia Ewald, Eppendorf AG, Hamburg, Germany Abstract The demands placed on consumables made of plastic
More informationNow also in Eppendorf LoBind. Conical Competence. Eppendorf Conical Tubes 15 ml and 50 ml
Now also in Eppendorf LoBind Conical Competence Eppendorf Conical Tubes 15 ml and 50 ml 2 Eppendorf Conical Tubes The Eppendorf Tubes Best Performance for Your Samples The Eppendorf Conical Tubes 15 ml
More informationept.i.p.s. Version A
ept.i.p.s. Fig. Tab. p. Instructions see p. for for use use ept.i.p.s. Instructions for use Version A 0012 580.460 Copyright 2012 Eppendorf AG, Hamburg. No part of this publication may be reproduced without
More informationWell within Reach. The Eppendorf Pipette Holder System
Well within Reach The Eppendorf Pipette Holder System 3 Eppendorf Pipette Holder System Pipette Stand 2 and Charger Stand 2 Protect your precious liquid handling device from possible damage, contamination
More informationEvaluating the functionality of Eppendorf BioPhotometer and Eppendorf BioSpectrometer using a Secondary UV-VIS Filter Set
USERGUIDE No. 10 I May 2015 Evaluating the functionality of Eppendorf BioPhotometer and Eppendorf BioSpectrometer using a Secondary UV-VIS Filter Set Martin Armbrecht-Ihle, Lars Borrmann Eppendorf AG,
More informationNEW: 5.0 ml screw cap tube. The Missing Link. Eppendorf Tubes 5.0 ml Discover a new sample handling system
NEW: 5.0 ml screw cap tube The Missing Link 5.0 ml Discover a new sample handling system 5mL_Tube_BRO_GB_2015.indd 1 27.11.15 14:02 2 Combined with the comprehensive line of accessories, 5.0 ml are the
More informationNEW! ep Dualfilter T.I.P.S. Peak of perfection
NEW! ep Dualfilter T.I.P.S. Peak of perfection A generation of tips to set standards Eppendorf pipette tips an investment in quality Eppendorf Totally Integrated Pipetting System Eppendorf Totally Integrated
More informationN) justment. Register your instrument! Eppendorf Research plus. Factory Adjustment
search ment plus N) justment Register your instrument! www.eppendorf.com/myeppendorf Factory Adjustment Copyright 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior
More informationInstrumental Technique: Cuvette. Md Rabiul Islam
Instrumental Technique: Cuvette Md Rabiul Islam 16-7-2016 What is cuvette? A cuvette is a small tube of circular or square cross section, sealed at one end, made of plastic, glass, or fused quartz (for
More informationMicro Volume QuEChERS kit
225-37872 Sep. 2018 Small Capacity Pretreatment Kit Micro Volume QuEChERS kit Instruction Manual Read this manual thoroughly before you use the product. Keep this manual for future reference. This page
More informationEppendorf Xplorer Eppendorf Xplorer plus. Ordering information. Register your instrument!
mation orer /Eppendorf Xplorer plus N) nformation Register your instrument! www.eppendorf.com/myeppendorf Eppendorf Xplorer Eppendorf Xplorer plus Ordering information Copyright 2015 Eppendorf AG, Germany.
More informationChromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials
Nitrate, HR DOC316.53.01068 Chromotropic Acid Method Method 10020 0.2 to 30.0 mg/l NO 3 N (HR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific information
More informationNEW! Your bacteria s favorite! Eppendorf Eporator
NEW! Your bacteria s favorite! Eppendorf Eporator Small footprint great results! Eppendorf Eporator Fast and easy: simple one-button operation! Welcome the Eppendorf Eporator, the solution for efficiently
More informationAutomated purification of high quality genomic DNA
APPLICATION NOTE No. AA267 I October 2012 Automated purification of high quality genomic DNA from various tissues using the Eppendorf MagSep Tissue gdna Kit on the Eppendorf epmotion M5073 Ulrich Wilkening,
More informationTPH (Total Petroleum Hydrocarbons)
TPH (Total Petroleum Hydrocarbons) DOC316.53.01475 Immunoassay 1 Method 10050 Scope and application: For water. 1 This test is semi-quantitative. Results are shown as more or less than the threshold value
More informationAutomated Illumina TruSeq DNA PCR-Free library construction with the epmotion 5075t/TMX
SHORT PROTOCOL No. 15 I August 2016 Automated Illumina TruSeq DNA PCR-Free library construction with the epmotion 5075t/TMX Introduction This protocol describes the configuration and preprogrammed methods
More informationEppendorf Temperature Verification System. Operating manual /0909
Eppendorf Temperature Verification System 6328 900.066-00/0909 Copyright 2009 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the copyright owner. eppendorf
More informationTetraphenylborate Method Method to 7.0 mg/l K Powder Pillows
Potassium DOC316.53.01127 Tetraphenylborate Method Method 8049 0.1 to 7.0 mg/l K Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific information
More informationCadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows
Nitrate DOC316.53.01067 Cadmium Reduction Method Method 8192 0.01 to 0.50 mg/l NO 3 N (LR) Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific
More informationNukEx Nucleic Acid Release Reagent
Instruction for Use NukEx Nucleic Acid Release Reagent For general laboratory use. For in vitro use only. Reagent for the enzymatic release of nucleic acid from tissue samples, ticks, insects and swabs.
More informationPackage Insert for FASafe/AciSafe
SāfTest Inc. PeroxySafe Package Insert Package Insert for FASafe/AciSafe FASafe/AciSafe Kit Package Insert & Control Package Insert Introduction FASafe standard test kit is designed to measure free fatty
More informationIndividual Laboratory Safety & Chemical Hygiene Plan
APPENDIX A: CHEMICAL HYGIENE PLAN Part B Individual Laboratory Safety & Chemical Hygiene Plan Principal Investigator Date filed Department Building/Floor/Room E-mail address @Pace.edu Phone Facsimile Laboratory
More informationOperation Manual. SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION
Operation Manual SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC. 4.18 Foreword
More information1. Preliminary qualitative analysis of unknown substances (liquid or solid).
Name of Procedure: Ultraviolet Spectroscopy Suggested Uses: 1. Preliminary qualitative analysis of unknown substances (liquid or solid). 2. Quantitative analysis of known compounds. Apparatus Used to Perform
More informationTechnical data and operating instructions. Vivaspin 4. Vivaspin 4 10K device for in vitro diagnostic use
Technical data and operating instructions Vivaspin 4 Vivaspin 4 10K device for in vitro diagnostic use 85030-516-16 85030-516-16 Vivaspin 4 ml Introduction Storage conditions shelf life Vivaspin 4 ultrafiltration
More informationThe best material for original Eppendorf Tubes and Plates: Properties and chemical resistance of polypropylene
APPLICATION NOTE No. 056 I June 2013 The best material for original Eppendorf Tubes and Plates: Properties and chemical resistance of polypropylene Natascha Weiß, Susanne Pruszkowski, Eppendorf AG, Hamburg,
More informationSpa & Wellness for Pipettes
epservices for premium performance Spa & Wellness for Pipettes Pipette Performance Plans Treat your pipettes well with premium services from the experts! Total approach philosophy For over 60 years Eppendorf
More informationMercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution
Chloride DOC316.53.01017 Mercuric Thiocyanate Method Method 8113 0.1 to 25.0 mg/l Cl Reagent Solution Scope and application: For water and wastewater. Test preparation Instrument-specific information Table
More informationPhysioCare Concept. Make your Lab a better place
PhysioCare Concept Make your Lab a better place 2 Eppendorf PhysioCare Concept The Eppendorf PhysioCare Concept The mission of Eppendorf has always been to improve the living conditions of their customers.
More informationCompact Orifice FlowMeter
Steam Data Reference Tables IM/OM/SDT Issue 1 Compact Orifice FlowMeter ABB The Company We are an established world force in the design and manufacture of instrumentation for industrial process control,
More informationMETHANOL REFERENCE SPECIFICATIONS
IMPCA INTERNATIONAL METHANOL PRODUCERS & CONSUMERS ASSOCIATION Avenue de Tervuren 149 1150 Brussels, Belgium Tel : (32/2) 7418683 Fax : (32/2) 7418684 e-mail : impca@epca.be IMPCA METHANOL REFERENCE SPECIFICATIONS
More informationApplications Note 202 October 2008
Applications Note 202 October 2008 Technical Report Eppendorf Polypropylene Microplates Fast and secure identification of samples Natascha Weiß 1, Wolf Wente 2, Sophie Freitag 2, Daniel Wehrhahn 1 1 Eppendorf
More informationSOIL ORGANIC CONTENT USING UV-VIS METHOD
Test Procedure for SOIL ORGANIC CONTENT USING UV-VIS METHOD TxDOT Designation: Tex-148-E Effective Date: March 2016 1. SCOPE 1.1 This method determines the soil organic content based on the amount of humic
More informationPowder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)
Aluminum DOC316.53.01002 Aluminon Method 1 Method 8012 0.008 to 0.800 mg/l Al 3+ (spectrophotometers) Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters) Scope and application: For water and wastewater.
More informationMOHAWK COLLEGE OF APPLIED ARTS AND TECHNOLOGY CHEMICAL AND ENVIRONMENTAL TECHNOLOGY DEPARTMENT. Lab Report ROOM NO: FE E309
MOHAWK COLLEGE OF APPLIED ARTS AND TECHNOLOGY CHEMICAL AND ENVIRONMENTAL TECHNOLOGY DEPARTMENT Lab Report ROOM NO: FE E309 EXPERIMENT NO : 4 TITLE : UV Spectrophotometric Analysis of DNA Submitted by Class
More informationComparative Analysis of Protein Recovery Rates in Eppendorf LoBind and Other Low Binding Tubes
APPLICATION NOTE No. 382 I October 2016 Comparative Analysis of Protein Recovery Rates in Eppendorf LoBind and Other Low Binding Tubes Rafal Grzeskowiak¹, Sandrine Hamels², Eric Gancarek 2 ¹Eppendorf AG,
More informationProcedures for Preparing Reagents and Media used in Firearm and Tool Mark Examinations
Procedures for Preparing Reagents and Media used in Firearm and Tool Mark Examinations North Carolina State Bureau of Investigation Firearm and Tool Mark Section July 10, 1996 Revised February 23, 1998
More information96 Well? Done! Eppendorf epmotion 96 semi-automated electronic pipette for fast and precise 96 channel microplate processing
96 Well? Done! Eppendorf epmotion 96 semi-automated electronic pipette for fast and precise 96 channel microplate processing Try it and you will see how easy and fast your 96-well and 384-well plates get
More informationTOP:018.7 U.S. Fish and Wildlife Service TECHNICAL OPERATING PROCEDURE
TOP:018.7 January 31, 2014 U.S. Fish and Wildlife Service Marquette Biological Station 3090 Wright Street Marquette, Michigan 49855 U.S.A. and U.S. Fish and Wildlife Service Ludington Biological Station
More informationThe New Legend. Eppendorf Reference 2 Pipette
The New Legend Eppendorf Reference 2 Pipette 2 Eppendorf Reference 2 Eppendorf Reference 2 3»The new Reference class for modern liquid handling tools.«the name»reference«stands for extraordinary precision
More informationHepatitis B virus IgM ELISA Kit
Hepatitis B virus IgM ELISA Kit Catalog Number KA0289 96 assays Version: 05 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay...
More informationPlease use only the valid version of the package insert provided with the kit. This kit is intended for Research Use Only.
Please use only the valid version of the package insert provided with the kit This kit is intended for Research Use Only. Not for use in diagnostic procedures. INTENDED USE The Vitamin Niacin test is a
More informationAutomated Illumina TruSeq Stranded Total RNA library construction with the epmotion 5075t/TMX
SHORT PROTOCOL No. 01 I November 2014 Automated Illumina TruSeq Stranded Total RNA library construction with the epmotion 5075t/TMX Introduction This protocol describes the configuration and preprogrammed
More informationTube O Reactor. (Micro: Cat. # K, K, K) (Medi: Cat. # K, K, K)
453PR G-Biosciences 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name Tube O Reactor (Micro: Cat. # 786 024 4K, 786 024 8K, 786 024 15K) (Medi: Cat. # 786
More informationTaKaRa BCA Protein Assay Kit
Cat. # T9300A For Research Use TaKaRa BCA Protein Assay Kit Product Manual Table of Contents I. Description... 3 II. Components... 3 III. Storage... 3 IV. Materials Required by not Provided... 3 V. Precautions
More informationHigh purity acids and bases
High purity acids and bases acids and bases Ultrapur acids The life science business of Merck KGaA, Darmstadt, Germany operates as MilliporeSigma in the U.S. and Canada. High purity acids and bases, and
More informationTryptic Digest of Bovine Serum Albumin
CARE products are designed to support our worldwide customers with high-quality consumables, accessories and dedicated kits. The CARE product range is specifically optimized and certified for use with
More informationSuperior well-to-well Consistency with Eppendorf Cell Culture Plates
APPLICATION NOTE No. 384 Superior well-to-well Consistency with Eppendorf Cell Culture Plates Ines Hartmann 1, Aurélie Tacheny 2 1 Eppendorf AG, Hamburg, Germany; 2 Eppendorf Application Technologies SA,
More informationmylab: Chemical Safety Module Last Updated: January 19, 2018
: Chemical Safety Module Contents Introduction... 1 Getting started... 1 Login... 1 Receiving Items from MMP Order... 3 Inventory... 4 Show me Chemicals where... 4 Items Received on... 5 All Items... 5
More informationFar UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers
Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N O T E Summary The far UV region (generally covering
More informationPrecise and Accurate whole Blood Dispensing with Multipette E3x and Combitips advanced
APPLICATION NOTE No. 386 Precise and Accurate whole Blood Dispensing with Multipette E3x and Combitips advanced Vincent Dufey¹, Muriel Art¹, Hanaë Henke² ¹Eppendorf Application Technologies S.A., Namur,
More informationCadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls
Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific
More informationWHAT YOU NEED. Visible Spectrophotometer. Micropipettes. Cuvettes. Parafilm. Timer µL
WHAT YOU NEED Visible Spectrophotometer Micropipettes 10-100µL 100-1000µL Cuvettes Parafilm Timer WHAT TESTS CAN BE DONE Acetic Acid Amino Acid Nitrogen Ammonia Citric Acid Glucose & Fructose Lactic Acid
More informationOptical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology
APPLICATION NOTE Optical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology Introduction UV/VIS spectrophotometers and microplate readers differ fundamentally in
More informationIRDye 800CW Protein Labeling Kit Low MW
IRDye 800CW Protein Labeling Kit Low MW Developed for: Aerius, and Odyssey Family of Imagers Please refer to your manual to confirm that this protocol is appropriate for the applications compatible with
More informationSample Preparation LCW 909. Total Kjeldahl Nitrogen, TKN NB: NEW! LCW 909 GB
Principle Organically bound nitrogen is digested by hot fuming sulphuric acid in the presence of catalysts and converted to ammonium, which, after steam distillation, can be determined as so-called Total
More informationU.S. Patent No. 9,051,563 and other pending patents. Ver
INSTRUCTION MANUAL Direct-zol 96 RNA Catalog Nos. R2054, R2055, R2056 & R2057 Highlights Quick, 96-well purification of high-quality (DNA-free) total RNA directly from TRIzol, TRI Reagent and all other
More informationGluten (Gliadin) Lateral Flow IIR
V.1/May/2012 Gluten (Gliadin) Lateral Flow IIR For the quick detection of wheat protein contaminations in foods and food-processing equipments 20 Test Sticks For Laboratory Use Only Please read full descriptions
More informationPhysioCare Concept improved. The feeling of a new philosophy
PhysioCare Concept improved The feeling of a new philosophy Intuitive operation Eppendorf Reference (fix), Reference (adjustable-volume) What's new? TÜV approved user-friendliness Optimized pipetting force
More informationApplications Note 164 October 2007
Applications Note 164 October 2007 Technical Report Eppendorf Safe-Lock Tubes Outstanding performance for centrifugation and incubation of samples in the laboratory Natascha Weiß 1, Sophie Freitag 2, Jürgen
More informationSearching the Chemical Literature (Assignment #1a)
Searching the Chemical Literature (Assignment #1a) Introduction The object of this laboratory is to introduce you to the various methods of locating information in the literature concerning the properties
More informationPhysioCare Concept improved. The feeling of a new philosophy
PhysioCare Concept improved The feeling of a new philosophy Intuitive operation Eppendorf Reference (fix), Reference (adjustable-volume) What's new? TÜV approved user-friendliness Optimized pipetting force
More informationZetasizer Nano-ZS User Instructions
Zetasizer Nano-ZS User Instructions 1. Activate the instrument computer by logging in to CORAL. If needed, log in to the local instrument computer Username: zetasizer. Password: zetasizer. 2. Instrument
More informationThe solution for all of your
The solution for all of your nanoparticle sizing and zeta potential needs. DelsaNano Series Blood Banking Capillary Electrophoresis Cell Analysis Centrifugation Genomics Lab Automation Lab Tools Particle
More informationCHEMICAL PROTECTIVE GLOVES. 2 Manufacturer s name and address and legal representative (surname, first name):
Testing and Certification of Personal Protective Equipment Category III (PPE) CHEMICAL PROTECTIVE GLOVES ATTACHMENT 1 to the order of placed with Institut für Arbeitsschutz der DGUV (IFA) Alte Heerstrasse
More informationab83360 Ammonia Assay Kit
Version 8 Last updated 18 December 2017 ab83360 Ammonia Assay Kit For the measurement of total ammonia and ammonium levels in various samples This product is for research use only and is not intended for
More informationColorimetric Method Method to 0.70 mg/l Ag Powder Pillows
Silver DOC316.53.01134 Colorimetric Method Method 8120 0.02 to 0.70 mg/l Ag Powder Pillows Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows
More informationN6,6 Posidyne Membranes
Rev. Date: 20-Jun-2010 Ultipor Nylon 6,6 and N6,6 Posidyne Membrane Description Pall provides a complete range of Nylon 6,6 membranes for fluid filtration in the pharmaceutical, biotechnology, electronics,
More informationChloramine (Mono) and Nitrogen, Free Ammonia
Chloramine (Mono) and Nitrogen, Free Ammonia DOC316.53.01016 Indophenol Method 1 Method 10200 0.04 to 4.50 mg/l Cl 2 0.01 to 0.50 mg/l NH 3 N Powder Pillows Scope and application: For the determination
More informationSCIEX Cation Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide
For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems A49109AE May 2015 AB Sciex Pte. Ltd and its affiliates disclaims all warranties with respect to this document, expressed or implied, including
More informationsymphony Reference Electrodes
DOC022.53.80322 symphony Reference Electrodes 07/2012, Edition 1 User Manual Table of contents Safety information on page 3 Preparation for use on page 4 Specifications on page 3 Measurement on page 5
More informationRichter Optica. Instructions for Model: S850 Stereo Zoom Microscope
Richter Optica info@richter-optica.com Instructions for Model: S850 Stereo Zoom Microscope Microscope Head Focus Knob Stage Clips Zoom Knob C-Mount Adapter Eyepiece Focusing Holder Focusing Holder Set
More informationTrioMol Isolation Reagent
TrioMol Isolation Reagent Technical Manual No. 0242 Version 06142007 I Description... 1 II Key Features... 1 III Storage..... 1 IV General Protocol Using Triomol Isolation Reagent 1 V Troubleshooting.
More informationSafe Installation and Use of the NanoRam Immersion Probe
Safe Installation and Use of the NanoRam Immersion Probe Description: NanoRam Immersion Probe - 12 inch length; Immersion Probe Disposable Protective Sleeve Model Number: NR2-IMP; NR2-IMPS Part Number:
More informationTrioMol Isolation Reagent
TrioMol Isolation Reagent Technical Manual No. 0242 Version 06142007 I Description... 1 II Key Features... 1 III Storage..... 1 IV General Protocol Using Triomol Isolation Reagent 1 V Troubleshooting.
More informationPREVENTING HAZARDOUS WASTE FINES
University of Medicine and Dentistry of New Jersey EOHSS FACTSHEET PREVENTING HAZARDOUS WASTE FINES In the last few years institutions such as MIT, University of New Hampshire, Yale, Stanford, and Boston
More informationLarger Display Easier to Use. Generation E. Repeater E3 and E3x: Dispense without compromising accuracy or precision
Larger Display Easier to Use Generation E Repeater E3 and E3x: Dispense without compromising accuracy or precision 2 Eppendorf Repeater E3 / E3x > > Repeater E3 and Repeater E3x are the experts for long
More informationAutomated Illumina TruSeq Stranded mrna library construction with the epmotion 5075t/TMX
SHORT PROTOCOL No. 02 I November 2014 Automated Illumina TruSeq Stranded mrna library construction with the epmotion 5075t/TMX Introduction For the MiSeq and HiSeq next generation sequencing (NGS) systems,
More informationP E R O X Y G E N S A sector group of Cefic
P E R O X Y G E N S A sector group of Cefic CEFIC PEROXYGENS H 2 O 2 AM-7160 Hydrogen Peroxide for industrial use Determination of apparent ph (pha) Potentiometric method March 2003 Notice This document
More informationLatest generation testo air velocity & IAQ probes with cable handle. Instruction manual
Latest generation testo air velocity & IAQ probes with cable handle Instruction manual Contents Contents 1 About this document... 5 2 Safety and disposal... 5 3 System description... 6 4 Description of
More informationHBeAg and HBeAg Ab ELISA Kit
HBeAg and HBeAg Ab ELISA Kit Cat. No.:DEIA3532 Pkg.Size:96T Intended use This kit is designed for detection of HBeAg and HBeAg Ab. General Description The solid phase of HBEAG AND HBEAG ANTIBODY ELISA
More informationSodium Chloride - Analytical Standard
Sodium Chloride - Analytical Standard Determination of Arsenic Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-105-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes
More informationAvery Dennison LW Series Retroreflective Window Sticker Film Issued: July 2013
LW series window stickers are designed to be variable printed for use inside or outside the vehicle. Reflective permit stickers provide easy verification for authenticity. Performance: L-S-300C Table IV
More informationAnnex 2 Formaldehyde
Annex 2 Formaldehyde The 2 methods are analytical procedures for the determination of formaldehyde CH2O, PM/Ref. No 17260, and hexamethylenetetramine (HMTA) C6H12N4, PM/Ref. No 18670, in food simulants
More informationGlobally Harmonized System of Classification and Labeling of Chemicals (GHS) LC-1033p Rev. 06/16
Globally Harmonized System of Classification and Labeling of Chemicals (GHS) Rev. 06/16 Agenda Introduction to Globally Harmonized System of Classifying and Labeling Chemicals Hazard & Precautionary Statements
More information