Cyanide Analysis with the CNSolution 3100
|
|
- Clara Shaw
- 5 years ago
- Views:
Transcription
1 Cyanide Analysis with the CNSolution 3100
2 The CNSolution is your solution to cyanide analysis Gas diffusion amperometry methods: Save time Save money Are more accurate Have fewer interferences Are green The CNSolution = the FS3100 Change to a photometric detector and run any colorimetric chemistry
3 Cyanide Methods
4 Cyanide methods measure the various cyanide species Strong Metal Complexes Fe, Co, Pt, Au Total CN Available CN Weak to Moderately Strong Metal Complexes Ag, Cd, Cu, Hg, Ni, Zn Free CN HCN CN -
5 Manual distillation is used to dissociate as HCN Macro Distillation MIDI Distillations
6 Cyanide methods that utilize distillation have significant disadvantages Time Consuming One hour long distillation (does not take into account setup and teardown) CATC requires two, one hour distillations Bulky and Relatively Expensive Glassware Operator-dependent results (technique) Multiple Interferences
7 Gas diffusion eliminates distillation and associated cost Lowest cost Dollars per Year Consumables + labor per year MicroDist Midi Dist CNSolution
8 Samples per 8 hour Shift Eliminating distillation increases laboratory capacity 800 Highest sample capacity Midi Dist Micro Dist CNSolution
9 Aquatic Free Cyanide Analysis
10 ASTM D can be run on a OIA1677 CNSolution analyzer Buffer Same Cartridge as OIA 1677
11 Obtain accurate, cost effective free cyanide results in minutes, not hours
12 Available Cyanide Analysis
13 Ligand Exchange methods measure available cyanide Method Number OIA 1677 ASTM D 6888 Description Measurement Ligand Exchange / Flow Injection Analysis Ligand Exchange / Flow Injection Analysis Gas Diffusion - Amperometry Gas Diffusion - Amperometry GD-amperometry methods do not require distillation
14 Ligand Exchange GD-amperometry methods get better recovery WAD CATC OIA 1677
15 Ligand Exchange GD-amperometry methods are more precise OIA 1677
16 OIA 1677 or ASTM D6888 flow diagram Acid Reagent
17 Ligand Exchange GD-amperometry methods have fewer interferences CATC WAD OIA 1677 N-organics SCN,NH 3,NO 2 Excessive Iron Cyanide Concentration Dependent None S 2 O 3, H 2 O 2 Concentration Dependent
18 Ligand Exchange GD-amperometry methods give you results in minutes CATC WAD OIA 1677 Sample Preparation 2 distillations 2 3 hours 1 distillation 2 3 hours No distillation Analysis 1 2 minutes 1 2 minutes 1 2 minutes Total Time 3 4 hours 3 4 hours 1 2 minutes
19 Samples per hour Ligand Exchange GD-amperometry methods means more samples samples per hour Distillation Diffusion 20 0
20 Ligand Exchange GD-amperometry methods provide the best benefits No distillation (eliminates 1 4 hours preliminary sample treatment) Low MDL (0.5 ppb) No Interferences High Throughput (up to 90 samples per hour) Ease of Operation, very simple chemistry.
21 Total Cyanide Methods Automated gas diffusion distillation and nondistillation methods
22 GD-amperometry provides the safest, easiest, and most accurate technique No pyridine Acid Reagent ASTM D is a green chemistry
23 ASTM D7511 is easy to understand and operate and does not distill. No pyridine Green Chemistry
24 Analytical Throughput per 8 Hour Shift Midi-Dist MICRO DIST CNSolution
25 Dollars Estimated Reagent and Consumable Cost per Analysis for Distillation (US Dollars) Midi-Dist MICRO DIST CNSolution
26 Estimated Labor (in minutes) Required to Distill 20 Samples for Analysis Midi-dist MICRO DIST CNSolution
27 Estimated Annual Labor Cost to Distill and Analyze 40 Samples per Month (Including Overhead) $7, $6, $5, $4, $3, $2, $1, $0.00 Midi Dist MICRO DIST CNSolution Assume tech pay at $15/hour
28 Estimated Annual Cost for Consumables to Distill and Analyze 40 Samples per Month $4, $3, $3, $2, $2, $1, $1, $ $0.00 Midi Dist MICRO DIST CNSolution
29 Estimated Annual Cost for Reagents to Distill and Analyze 40 Samples per Month $2, $2, $1, $1, $ $0.00 Midi Dist MICRO DIST CNSolution
30 Midi Dist MICRO DIST CNSolution Annual Cost for Distillation vs. Non- Distillation Techniques $25, $20, $15, $10, /month 100 / month $5, $0.00
31 Midi Dist MICRO DIST CNSolution Comparison of Cost per Sample by Cyanide Species and Method $35.00 $30.00 $25.00 $20.00 $15.00 $10.00 $5.00 $0.00 Available Total
32 Midi Dist MICRO DIST CNSolution Profit per Sample by Cyanide Species at $40 per Test $40.00 $35.00 $30.00 $25.00 $20.00 $15.00 $10.00 $5.00 $0.00 Available Total
33 Midi Dist MICRO DIST CNSolution Profit per Month by Cyanide Species at $40 per Test and 40 Samples per Month $1, $1, $1, $1, $ $ $ $ $0.00 Available Total
34 Midi Dist MICRO DIST CNSolution Profit per Shift by Cyanide Species at $40 per Test Based on Maximum Capacity $20, $18, $16, $14, $12, $10, $8, $6, $4, $2, $0.00 Available Total
35 Profit per Year Estimated Profit by Number of Samples Analyzed at Maximum Capacity $800,000 $700,000 $600,000 MICRO DIST $500,000 $400,000 Midi Dist $300,000 $200,000 $100,000 $ Samples per Month
36 Thoughts Regarding Break Even Analysis Manual distillation and manual spectrometer versus CNSolution Manual equipment costs 3 x less CNSolution profit 3 x higher per test Break Even = 1000 samples for both! That s 20 samples per week regardless
37 Description of the Analyzer
38 The CNSolution 3100 available cyanide system
39 The CNSolution 3100 total cyanide system UV irradiation
40 A close up of the analytical module Detector Gas Diffusion and Mixing Flow-cell
41 A high quality peristaltic pump Separate module, individual channels, long pump tube life.
42 A random access xyz sampler Integrated circulating wash station 90 or 360 positions Separate standard and QC vials
43 Winflow software provide accurate quantitation of CN results
44 The CNSolution 3100 is accurate, and cost effective Rapidly analyze cyanide Eliminate time consuming error causing distillations Expand capability to colorimetric FIA and SFA
45 Thank You Questions?
Automating Wet Chemical Analysis. William Lipps Market Specialist for Water Analyzer Products December 11, 2008
Automating Wet Chemical Analysis William Lipps Market Specialist for Water Analyzer Products December 11, 2008 The advantages of automating wet chemical analysis. Save time Decrease cost Improve quality
More informationCyanide Methods for Lachat Automated Ion Analyzers. Flow Injection Analysis
Cyanide Methods for Lachat Automated Ion Analyzers Flow Injection Analysis 19 January 2018 ii QuikChem Methods List Use this list to: Identify and select analytical methods for your analyte, range, and
More informationAutomating Wet Chemical Analysis
Automating Wet Chemical Analysis Who Should Automate? Anyone looking to improve laboratory efficiency Environmental Agricultural Food and beverage Pharmaceutical Mining Etc. Advantages of Automating Save
More informationA New On-line Cyanide Analyzer for Measurement of Cyanide in Hydrometallurgical Processing of Precious Metal Ores
Application Note 37890312 A New On-line Cyanide Analyzer for Measurement of Cyanide in Hydrometallurgical Processing of Precious Metal Ores Keywords CNSolution 9310 On-line Cyanide Analyzer Cyanidation
More informationAutomation of Wet Chemical Analysis Methods
Automation of Wet Chemical Analysis Methods Application Note 21150205 Keywords CNSolution FS 3100 DA 3500 Discrete Analysis Ion Chromatography FIA Flow Injection Analysis Flow Solution Segmented Flow Analysis
More informationFlow Solution FS 3700 Automated Chemistry Analyzer
Method Abstract Total Cyanide by ISO 14403 Document #42600116 Flow Solution FS 3700 Automated Chemistry Analyzer Total Cyanide by In-Line Ultraviolet Digestion, Gas Diffusion, and Photometric Detection,
More informationAutomated Photometry. Presented by Scott Tucker. Instruments
Application Breakthroughs in Automated Photometry Presented by Scott Tucker Application Chemist, Lachat Instruments Introduction Advantages of automated photometry FIAversus other techniques Lachat Application
More informationDI 3500 Discrete Aoalvzer
DI 3500 Discrete Aoalvzer Automated Chemistry and Ion Analysis ':t O+Analytical:!, A World ofsolutions. - Discrete Analysis Efficient Microscale Chemistry Increasingly laboratories are being challenged
More informationIntroduction. Terminology
Use of ASTM D 7284-08 Standard Test Method for Total Cyanide in Water by Micro Distillation followed by Flow Injection Analysis with Gas Diffusion Separation and Amperometric Detection Introduction A new
More informationNitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA)
Methodology Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA) (Cartridge Part #A002603) 1.0 Scope and Application 1.1 This method is used to determine the concentration of nitrate
More informationCyanide Analysis of Wastewater Samples from FCC and Hydrocracking Operations
Cyanide Analysis of Wastewater Samples from FCC and Hydrocracking Operations Introduction Fluid catalytic cracking (FCC) is a major unit operation in refineries around the world. FCC is used to convert
More informationAnalysis of Cyanide (Total, Weak Acid Dissociable, and Free) - PBM
Inorganics Revision Date: July 10, 2017 Analysis of Cyanide (Total, Weak Acid Dissociable, and Free) - PBM Parameter Analytical Method Cyanide (Total, Weak Acid Dissociable, Free) Analysis of Cyanide Species
More informationElemental analysis of river sediment using the Agilent 4200 MP-AES
Elemental analysis of river sediment using the Agilent 4200 MP-AES Application note Environmental: Soils, sludges & sediments Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction
More informationAnalysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS
TECHNICAL NOTE 4322 Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS Keywords Argon gas dilution, AGD, High matrix samples, Seawater Goal To critically
More informationFlow Solution FS 3700 Automated Chemistry Analyzer
Method Abstract Total Cyanide by ASTM D7511 Document #43671116 Flow Solution FS 3700 Automated Chemistry Analyzer Total Cyanide by In-Line Ultraviolet Digestion and Amperometric Detection ASTM D7511, Segmented
More informationComparing the MICROLAB 600 to Volumetric Glassware and Air Displacement Pipettes. June 2011
Comparing the MICROLAB 600 to tric Glassware and Air Displacement Pipettes June 2011 tric Glassware and Air Displacement Pipettes Table of Contents ``Abstract...3 ``Introduction...4 ``Method & Results...5
More informationHexavalent Chromium by Flow Injection Analysis (FIA)
Methodology Hexavalent Chromium by Flow Injection Analysis (FIA) (Cartridge Part #A002831) 1.0 Scope and Application 1.1 This method is used for the determination of hexavalent chromium in water, including
More informationHigh-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)
High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) Application Note Environmental Authors Steve Wilbur Agilent Technologies,
More informationBy Mohsin Al-Saleh Clinical Biochemistry Department (SQUH)
Laboratory Automation By Mohsin Al-Saleh Clinical Biochemistry Department (SQUH) Objectives: Define automation and lab automation Advantages of laboratory automation History of laboratory automation Fluidic
More informationComparing the Microlab 600 to Volumetric Glassware and Air Displacement Pipettes. January 2013
Comparing the Microlab 600 to tric Glassware and Air Displacement Pipettes January 2013 White Paper: Comparing the Microlab 600 to tric Table of Contents Abstract... 3 Introduction... 4 Methods & Results...
More informationAnalysis of domestic sludge using the Agilent 4200 MP-AES
Analysis of domestic sludge using the Agilent 4200 MP-AES Application note Environmental Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction Managing the treatment and disposal
More informationICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer
Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer is powerful simultaneous full
More informationBromide by Flow Injection Analysis (FIA)
Methodology Bromide by Flow Injection Analysis (FIA) (Cartridge Part #A002763) 1.0 Scope and Application 1.1 This method is used for the determination of bromide in natural water and wastewater. 1.2 The
More informationReverse Flow Injection Analysis for Determination of Manganese(II) in Natural Water. Jintana Klamtet
NU Science Journal 2006; 2(2): 165 173 Reverse Flow Injection Analysis for Determination of Manganese(II) in Natural Water Jintana Klamtet Department of Chemistry, Faculty of Science, Naresuan University,
More informationGyrolab Protein A Kit Quantification of residual Protein A ligands in the presence of excess amounts of IgG
Quantification of residual Protein A ligands in the presence of excess amounts of IgG Product number P0020456 P0020457 Product Name Gyrolab Protein A Kit for MabSelect SuRe Ligand Gyrolab Protein A Kit
More informationContinuous Flow Analysis Methods David Davey University of South Australia. 21 February
Continuous Flow Analysis Methods David Davey University of South Australia 21 February 2012 1 Flow Analysis Methods 1950s-60s Segmented Flow Methods Clinical Analysis Skeggs 1970s Flow Injection Analysis
More informationHigh-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application
High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application Environmental Author Neal Julien Midwest Research Institute
More informationChloride by Flow Injection Analysis (FIA)
Methodology Chloride by Flow Injection Analysis (FIA) (Cartridge Part #A001568) 1.0 Scope and Application 1.1 This method is used for the determination of chloride in drinking water, surface water, saline
More informationA LAZAR LABORATORY HANDBOOK MEASUREMENT OF NITRIC OXIDE IN BIOLOGICAL SAMPLES
A LAZAR LABORATORY HANDBOOK MEASUREMENT OF NITRIC OXIDE IN BIOLOGICAL SAMPLES Copyright Lazar Research Laboratories, Inc. 2004-1 - QUESTIONS? Call us at 1-800-824-2066 (1-323-931-1204 international), fax
More informationElemental Scientific. Stability. Elemental Scientific
Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dilution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field (field@icpms.com) Brief The prepfast inline
More informationEnhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS
Enhancing the productivity of food sample analysis with the Agilent 77x ICP-MS Application note Foods testing Authors Sebastien Sannac, Jean Pierre Lener and Jerome Darrouzes Agilent Technologies Paris,
More informationToday s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy
Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Evrim Kilicgedik Product Specialist, Atomic Spectroscopy Agilent Technologies 04.11.2011 2011 The Atomic Spectroscopy
More informationAlkalinity, USEPA by Flow Injection Analysis (FIA)
Methodology Alkalinity, USEPA by Flow Injection Analysis (FIA) (Cartridge Part #A002745) 1.0 Scope and Application 1.1 This method is used for the determination of alkalinity in drinking water, surface
More informationThe Agilent 7700x ICP-MS Advantage for Drinking Water Analysis
The Agilent 77x ICP-MS Advantage for Drinking Water Analysis Application Note Environmental Authors Steve Wilbur Agilent Technologies, Inc. 338 146th Place SE, Suite 3, Bellevue Washington, 987 USA Introduction
More informationAutomation of the radiochemical procedures for the sequential separation of radionuclides
LSC2017 - An International Conference on Advances in Liquid Scintillation Spectrometry, Copenhagen Denmark, 1 5 May 2017 Automation of the radiochemical procedures for the sequential separation of radionuclides
More informationDissolution Automation: Basic Questionnaire.
Dissolution Automation: Basic Questionnaire. In order to quote the most suitable automated Dissolution System we would like you to answer this questionnaire: 1. Do you want to work : a) On line (on line
More informationExperiment 1 (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with 1,10- Phenanthroline
Experiment (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with,0- Phenanthroline Background The first step of an analytical spectrophotometric procedure for quantitative determination
More informationSEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW
SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW Paula Hong, John Van Antwerp, and Patricia McConville Waters Corporation, Milford, MA, USA Historically UV detection has been
More informationBECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple
BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple Reliable TOC measurements Stop throwing away your first rep. QbD1200 has virtually eliminated sample-to-sample carryover
More informationCHAPTER CHROMATOGRAPHIC METHODS OF SEPARATIONS
Islamic University in Madinah Department of Chemistry CHAPTER - ----- CHROMATOGRAPHIC METHODS OF SEPARATIONS Prepared By Dr. Khalid Ahmad Shadid Chemistry Department Islamic University in Madinah TRADITIONAL
More informationAnion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM
IC Application Work AW CH6-1048-012011 Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM Branch: Chemical industry; Water, wastewater, environmental protection,
More informationO Sodium Bicarbonate, NaHCO 3 Ammonium Molybdate Tetrahydrate, (NH 4. O 4 Sodium Chloride, NaCl Ascorbic Acid, C 6 H 8 O 6
Scope Summary Interferences Performance Specifications This method is used for the determination of silica in estuarine and coastal waters (seawater) according to USEPA Method 366.0. This method can also
More informationMethod for Nitrite determination on Low-Range Samples
Method for Nitrite determination on Low-Range Samples 1. PRINCIPLE This method is designed for the determination of nitrite in soil, water and other forms of samples. The sample is prepared offline if
More informationOES - Optical Emission Spectrometer 2000
OES - Optical Emission Spectrometer 2000 OES-2000 is used to detect the presence of trace metals in an analyte. The analyte sample is introduced into the OES-2000 as an aerosol that is carried into the
More informationCyanide, Total or Weak Acid Dissociable, by Manual Distillation - PBM
Inorganics Revision Date: March 8, 2017 (Draft) Cyanide, Total or Weak Acid Dissociable, by Manual Distillation - PBM Parameter Analytical Method Introduction Method Summary MDL(s) and EMS Analyte Codes
More information1901 Application of Spectrophotometry
1901 Application of Spectrophotometry Chemical Analysis Problem: 1 Application of Spectroscopy Organic Compounds Organic compounds with single bonds absorb in the UV region because electrons from single
More informationMETHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV)
METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV) 1.0 SCOPE AND APPLICATION 1.1 Method 9012 is used to determine the concentration of inorganic cyanide in an aqueous waste or leachate.
More informationExploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis
Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis A. Caruso, M. Santoro, P. Magni, S. Pelagatti, and R. Facchetti Thermo Fisher Scientific, Milan,
More informationPros and Cons of Water Analysis Methods
Water Lens, LLC 4265 San Felipe, Suite 1100 Houston, Texas 77027 Office: (844) 987-5367 www.waterlensusa.com Pros and Cons of Water Analysis Methods Prepared by: Adam Garland, CTO Water Lens, LLC ICP-MS/OES
More informationMETHOD 9012B TOTAL AND AMENABLE CYANIDE (AUTOMATED COLORIMETRIC, WITH OFF-LINE DISTILLATION)
METHOD 9012B TOTAL AND AMENABLE CYANIDE (AUTOMATED COLORIMETRIC, WITH OFF-LINE DISTILLATION) 1.0 SCOPE AND APPLICATION 1.1 This method is used to determine the concentration of inorganic cyanide (CAS Registry
More informationMETHOD DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY
METHOD 335.4 DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division Revision 1.0 August 1993 ENVIRONMENTAL MONITORING
More informationHYDROGEN CYANIDE 6017
HYDROGEN CYANIDE 6017 HCN MW: 27.03 CAS: 74-90-8 RTECS: MW6825000 METHOD: 6017, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003 OSHA: 10 ppm (skim) NIOSH: STEL 4.7 ppm ACGIH: C 10 ppm (skin) (1 ppm
More informationNitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow
1. Application Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow Parameters and Codes: Nitrogen, ammonia, dissolved, I-2522-90 (mg/l as N): 00608 Nitrogen, ammonia, total-in-bottom-material,
More informationSimple Use Compact and Robust Fast and Efficient
INFECTIOUS DISEASES Simple Use Minimal hands on time Intuitive operator interface and touchscreen Simpli fies training and day-to-day operation Primary sample tubes, aliquot tubes and cups No reagent handling:
More informationMETHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE)
METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE) 1.0 SCOPE AND APPLICATION 1.1 This automated method is applicable to ground water, drinking and surface waters, and domestic and industrial wastes
More informationThis immunoassay kit allows for the in vitro quantitative determination of Aflatoxin M1 concentrations in milk, milk power.
Aflatoxin M1 (AFM1) ELISA Kit This immunoassay kit allows for the in vitro quantitative determination of Aflatoxin M1 concentrations in milk, milk power. This package insert must be read in its entirety
More informationStandard Methods for the Examination of Water and Wastewater
4500-NO 2 NITROGEN (NITRITE)*#(1) 4500-NO 2 A. Introduction 1. Occurrence and Significance For a discussion of the chemical characteristics, sources, and effects of nitrite nitrogen, see Section 4500-N.
More informationEnAlgae SOP sheet: Silicate - SFA. 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2.
EnAlgae SOP sheet: Silicate - SFA 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2. 2. Principle Analysis is performed on a segmented flow analyser (SFA), model AutoAnalyser
More informationDetermination of trace anions in concentrated hydrofluoric acid
APPLICATION NOTE 78 Determination of trace anions in concentrated hydrofluoric acid Authors Archava Siriraks Thermo Fisher Scientific, Sunnyvale, CA Keywords HF, ICS-5000 +, IonPac AS10, IonPac AC10, ion
More informationcobas c 111 analyzer Big analyzer performance in a small package
For more information, contact your Roche Diagnostics representative or call 1-800-852-8766 (option 7). cobas c 111 analyzer Big analyzer performance in a small package www.poc.roche.com COBAS, COBAS B,
More informationUltra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES
Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Application note Geochemistry, metals, mining Authors John Cauduro Agilent Technologies, Mulgrave, Australia Introduction
More informationMyBioSource.com. This package insert must be read in its entirety before using this product.
Tetracyclines ELISA Kit Catalog Number. MBS940077 This immunoassay kit allows for the in vitro quantitative determination of Tetracyclines concentrations in honey, tissue(chicken, pork). This package insert
More informationEPA Office of Water Method Update Rule Final August 28, 2017 William Lipps October 2017
EPA Office of Water Method Update Rule Final August 28, 2017 William Lipps October 2017 1 / 9 The update consists of minor changes, edits, revisions, and 7 new methods 80 70 60 50 40 30 20 10 0 Standard
More informationDetermination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors
Determination of major, minor and trace elements in rice flour using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Application note Food testing Authors John Cauduro Agilent Technologies,
More informationThermo Scientific icap RQ ICP-MS: Typical limits of detection
TECHNICAL NOTE 43427 Thermo Scientific icap RQ ICP-MS: Typical limits of detection Author Tomoko Vincent Keywords BEC, interference removal, KED, LOD Introduction Inductively Coupled Plasma Mass Spectrometry
More informationLow Level Phosphorus: A Method Comparison Study
Low Level Phosphorus: A Method Comparison Study Katie Adams* Stephanie Le, Theresa McBride, Tom Pearson EPA Region 10 Laboratory August 2012 8/29/2012 1 U.S. Environmental Protection Agency Goal: 1 µg/l
More informationSKALAR. your partner in chemistry automation. The San ++ continuous flow analyzer
SKALAR your partner in chemistry automation The San ++ continuous flow analyzer Applications Over the years Skalar has been working and developing applications for a variety of industries. Our comprehensive
More informationIn-Solution Digestion: Multi-Plate v1.0 Quick Start Guide
In-Solution Digestion: Multi-Plate v1.0 Quick Start Guide This guide is for users who have been trained in the proper use of the AssayMAP Bravo Platform and understand the safety guidelines in the Bravo
More informationAir Monitoring. Semi-continuous determination of ambient air quality
Air Monitoring Semi-continuous determination of ambient air quality The Particle Into Liquid Sampler a simple solution for the determination of ions in aerosol particles 02 Combustion of fossil fuels for
More informationMercury Determination in Lichen, BCR-482, Using the QuickTrace M-7600 CVAAS Mercury Analyzer
Application Note - AN1502 Mercury Determination in Lichen, BCR-482, Using the QuickTrace M-7600 CVAAS Mercury Analyzer Page 1 Jeff Forsberg, Product Manager, John McQuatters, Mercury Product Specialist,
More informationPAR Method Method to 2.0 mg/l Pb TNTplus 850
, TNTplus 850, 10216 DOC316.53.01056 PAR Method Method 10216 0.1 to 2.0 mg/l Pb TNTplus 850 Scope and Application: For wastewater and process control Test preparation Before starting the test: FOR DR 2800:
More informationHPLC Sample Preparation
HPLC Sample Preparation Rely on Sartorius solutions and benefit from optimized workflows for clean results mau 400 350 300 250 200 150 100 50 0 50 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0
More informationApplications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry
Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Fundamentals and Applications to the Petrochemical Industry Outline Some background and fundamentals of ICPMS
More informationTotal Kjeldahl Nitrogen by Gas Diffusion and Flow Injection Analysis (FIA)
Methodology Total Kjeldahl Nitrogen by Gas Diffusion and Flow Injection Analysis (FIA) (Cartridge Part #A002380) 1.0 Scope and Application 1.1 This method is used for the determination of Total Kjeldahl
More informationElemental Scientific. Spike Recovery. Elemental Scientific
Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dissolution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field (field@icpms.com) Brief The prepfast
More informationElemental Scientific. Spike Recovery. Elemental Scientific
Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dissolution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field (field@icpms.com) Brief The prepfast
More informationChapter 3 Electrochemical methods of Analysis-Potentiometry
Chapter 3 Electrochemical methods of Analysis-Potentiometry Electroanalytical chemistry Contents Introduction Galvanic and electrolytic cells Salt bridge Electrode potential and cell potential Indicator
More informationChemical Analysis. PIONA+ Analyzer. Characterization of Engine Fuels by Hydrocarbon Group Type. Gas Chromatography. think forward
Chemical Analysis PIONA+ Analyzer Characterization of Engine Fuels by Hydrocarbon Group Type think forward Gas Chromatography Bruker 450 PIONA+ Analyzer Bruker s innovative PIONA+ TM Analyzer provides
More informationLecture 11 Reaction Types and Mechanisms for Inorganic Complexes
2P32 Principles of Inorganic Chemistry Dr. M. Pilkington Lecture 11 Reaction Types and Mechanisms for Inorganic Complexes Variations in reactivity Reaction types substitution, dissociation, addition and
More informationDR/4000 PROCEDURE. IRON, Total
Method 8365 Powder Pillows DR/4000 PROCEDURE FerroMo Method* (0 to 1.800 mg/l) Scope and Application: For cooling water containing molybdate-based treatment; digestion is required for determining total
More informationRapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.
Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS Application Environmental Author Steven Wilbur Agilent Technologies, Inc. 3380 146th Place, SE, Suite 300 Bellevue, WA
More informationChemical Foundations Part 2
Chemical Foundations Part 2 Reading: Downloads: Ch 4 sections 8 11 Ch 5 sections1 7 Periodic table Ion Chart * = important homework question Homework: 4.8 question 44*, 46, 52 4.10 questions 66, 68, 70,
More information1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction
1. Application 1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction Parameters and Codes: EDB and DBCP, whole water recoverable, O-3120-90 Parameter (µg/l)
More information2011_ST20_Taylor.pdf. SSPT Document not peer-reviewed by CORESTA
An Ion Chromatographic Method for the Quantitative Determination of Hydrogen Cyanide in Cigarette Smoke using Pulsed Amperometric Detection Dr W Guthery and Dr M J Taylor Filtrona Technology Centre CORESTA
More informationExercise 2-4. Titration of a Buffer Solution EXERCISE OBJECTIVES
Exercise 2-4 Titration of a Buffer Solution EXERCISE OBJECTIVES To define the terms buffer solution and buffer capacity; To titrate a buffer solution with a weak acid solution; To plot a graph using the
More informationRNA Transport. R preps R preps
RNA Transport R0527-00 5 preps R0527-01 50 preps July 2014 RNA Transport Table of Contents Introduction...2 Kit Contents/Storage and Stability...3 Protocol...4 Storage Procedure...4 Recovery Procedure...5
More informationcostech instruments ECS 4010 Elemental Combustion System CHNS-O
instruments ECS 4010 Elemental Combustion System CHNS-O costech Costech Analytical Tecnologies Inc. 26074 Avenue Hall, Suite 14 Valencia, CA 91355 - USA Phone: (661) 297-2395 Fax: (661) 297-5492 e-mail:
More informationMethod OIA-1677: Available Cyanide by Ligand Exchange and Flow Injection Analysis (FIA)
Methodology Method OIA-1677: Available Cyanide by Ligand Exchange and Flow Injection Analysis (FIA) (Cartridge Part #A001982 or #A002654) 1.0 Scope and Application 1.1 This method is used for the determination
More informationTime Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples
Application Note 79 Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples INTRODUCTION Cow s milk is of particular dietary value to infants,
More informationChem 102H Exam 2 - Spring 2005
Name I.D. # Chem 102H Exam 2 - Spring 2005 PHYSICAL CNSTANTS/CNVERSIN FACTRS Speed of light = 3.00! 10 8 m/s Planck!s const. = 6.63! 10-34 J s Avagadro!s Number = 6.02! 10 23 Electron charge = 1.602! 10-19
More informationZinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures
Villanova University Date: Oct 2011 Page 1 of 9 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP F Zinc Metal Determination Perkin Elmer
More informationFigure 1. The supported liquid extraction process using the ISOLUTE SLE plate (single well shown)
Supported Liquid Extraction: Automate those Tiresome Bioanalytical Protocols L. Williams, H. Lodder, S. Merriman, A. Howells, S. Jordan, J. Labadie, M. Cleeve, C. Desbrow, R. Calverley and M. Burke 1 Argonaut
More informationSpeed and flexibility
Speed and flexibility Automate your environmental, water, detergent, food and beverage analysis Thermo Scientific Gallery and Gallery Plus discrete industrial analyzers Improve productivity, reproducible
More informationChem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS
Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS 1. List two advantages of temperature programming in GC. a) Allows separation of solutes with widely varying retention factors in a reasonable
More informationOverview of X-Ray Fluorescence Analysis
Overview of X-Ray Fluorescence Analysis AMPTEK, INC., Bedford, MA 01730 Ph: +1 781 275 2242 Fax: +1 781 275 3470 sales@amptek.com 1 What is X-Ray Fluorescence (XRF)? A physical process: Emission of characteristic
More informationExperiment 14 - Qualitative Analysis
Introduction Qualitative analysis involves the identification of the substances in a mixture. When chemical methods are used in the identification of mixtures of metal cations, these ions are usually separated
More information881 Compact IC pro 882 Compact IC plus. Compact ion chromatography for intelligent, reliable and precise analysis
881 Compact IC pro 882 Compact IC plus Compact ion chromatography for intelligent, reliable and precise analysis 881 Compact IC pro and 882 Compact IC plus the compact solution for routine analysis 02
More informationAuthors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin
Elemental Scientific prepfast IC: Inline Autodilution Method to Eliminate Species Interconversion for LC-ICPMS Based Applications Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan
More informationResidual Chlorine Monitor
Data Sheet SS/AW00 Issue Residual Chlorine Monitor AW00 Total and Free Residual Chlorine Measurement add potassium iodide to reagent to convert free chlorine measurement to total chlorine Reagentless Operation
More information2Fe 2 O 3 +3H 2 S FeS+FeS x +S+3H 2 O
Elemental analysis of hydrocarbon streams using Dry colorimetry analyzers, a catalyst saviour Bushra Dawood, Application Coordinator C.I. Analytics www.cianalytics.com The Petrochemical industry has refined
More information