1 Villanova University Date: Oct 2011 Page 1 of 9 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP F Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures Author: Water Resources Laboratory Technician Sonali Joshi Approved: VUSP Director Robert G. Traver SOP-F-1

2 Villanova University Date: Oct 2011 Page 2 of 9 Distribution John Komlos Quality Assurance Officer Laboratory Copy, maintained by Quality Assurance Officer Revision Date Responsible Person Description of Change 0 01 Jan 2010 Sonali Joshi Initial 1 01 Oct 2011 Sonali Joshi Minor Update 2 The following laboratory staff have read and been trained on this Manual. A copy of this page will be maintained in the Laboratory. Training is good for one year. Signature Name / Position Date Trainer Robert G. Traver Director Sonali Joshi Laboratory Technician John Komlos Quality Assurance Officer SOP-F-2

3 Villanova University Date: Oct 2011 Page 3 of 9 TABLE OF CONTENTS 1.0 SCOPE AND APPLICATION METHOD SUMMARY DEFINITIONS HEALTH AND SAFETY WARNINGS INTERFERENCES HANDLING AND PRESERVATION EQUIPMENT AND SUPPLIES QUALITY CONTROL PROCEDURES DATA ANALYSIS AND CALCULATIONS WASTE MANAGEMENT REFERENCES 8 SOP-F-3

4 Villanova University Date: Oct 2011 Page 4 of Scope and Application The purpose of this standard operating procedure (SOP) is to show the proper set up, procedures, and operation steps for Zinc in stormwater samples using the Perkin Elmer AAnalyst 800 Flame Atomic Absorption Spectrometer. This document serves as the basis for maintaining accurate and reproducible results, and defines pertinent validation criteria for accepting/rejecting final data. Standards and blanks will be employed as a means of validating data within range of the instrument s detection limits Method Summary The Perkin-Elmer Model AAnalyst 800 is an atomic absorption spectrometer which measures the concentrations of metals in various matrices. For our purposes, it provides integrated readings in absorbances over a period selected by the user from 0.2 to 60 seconds. The liquid sample is sprayed by a pneumatic nebulizer into a flame. Sample particles vaporize and dissociate into free atoms due to the temperature and reducing power of the flame. The concentration of metal element in the sample is approximated in the parts per million range from the absorbance and standard calibration curve Definitions NA 4.0 Health and Safety Warnings 4.1 Close the door on the instrument before turning on the flame. Never look directly at the flame. The UV light can cause serious eye damage. 4.2 Do not use oxygen in the burner. 4.3 Be sure there an adequate ventilation system for the removal of exhaust gases. 4.4 Change the acetylene tank when the pressure falls to 520 kpa (75 psig). Do not set the acetylene outlet pressure above 110 kpa (15 psig). 4.5 If high concentrations of silver, copper, or mercury salts have been nebulized into an acetylene flame, clean the burner chamber immediately after use. SOP-F-4

5 Villanova University Date: Oct 2011 Page 5 of The escape of fuel gas or fuel gas mixture from the drain tube, the burner system, the gas control, or the gas connections results in a serious explosion hazard. 4.7 Locate the acetylene tank away from sources of heat, open flames, or sparks. Make sure cylinder valves are closed when not in use Interferences NONE 6.0. Handling and Preservation All glassware and plastic ware used for collection, transportation, and laboratory analysis of the metal samples are acid-washed using the following procedure. The glassware and plastic ware are first cleaned with a laboratory detergent and rinsed with tap water. Next, the container is rinsed in 1:10 Nitric Acid Solution. The container is then rinsed with distilled water three times and allowed to air dry. The samples are preserved in a 2% nitric acid matrix and refrigerated. They are analyzed within six months of preservation Equipment and Supplies 7.1 Adequate ventilation system 7.2 Supply of compressed air. 7.3 Tank of acetylene 7.4 Zinc Source Lamp 7.5 Sample containers 7.6 Concentrated Nitric Acid 7.7 Stock Zinc Solution SOP-F-5

6 Villanova University Date: Oct 2011 Page 6 of Quality Control A system of standard solutions and blanks are used to check the laboratory techniques for the AA800 Atomic Absorption Spectrometer. The blanks are samples of 2% nitric acid that are treated in the same manner as the samples to be tested. A set of four to five standards ranging from 0.05 to 10 ppm of Zinc is used to attain a linear absorbance-concentration curve, with a correlation coefficient greater than or equal to To ensure accuracy, a check standard must be applied for at least every 10 samples (note start with the lowest standard). The check standard must agree with the initial reading as seen on the calibration curve by 25%. If agreement fails, it is presumed that the readings of the set of samples run before the check standard are unreliable and must be rerun. Trouble shooting begins with checking the instrument and procedure for any obvious flaws. Next, the reagents should be checked and re-prepared, if necessary. Finally, at least one set of duplicate samples are run for every batch analyzed. In this test, one sampling site is chosen at random in which two separate samples are taken and treated/analyzed separately. Both samples should render readings within 10% of each other to be considered valid. If they do not fall within this range, procedures are examined to be sure that all samples were treated in the same manner. SOP-F-6

7 Villanova University Date: Oct 2011 Page 7 of Procedures 9.1 Matrix preparation Sample matrix: 2%v Trace Metal grade nitric acid (HNO 3 ) prepared with MQ water. 9.2 Standards preparation Master stock solutions are available for Zn. These have expiration dates and must be replaced on a regular basis. Four to five standards are prepared by weight from the stock, and diluted using the 2% Nitric Acid as the matrix. 9.3 Switching ON the system Switch the spectrometer ON Set the acetylene tank output pressure to 69 kpa (10 psig) Set the air supply pressure to kpa (80-90psig) Press FLAME ON/OFF to ignite the flame Allow burner head to warm up If necessary, readjust the acetylene pressure to 69 kpa (10 psig) Turn on the computer and start the software by clicking on the icon WinLab32 for AA The computer interface will start communicating with the equipment When the communication is successful, check marks appear on the screen 9.4 Installing and aligning lamps Always wear ultraviolet-absorbing safety glasses when viewing the following sources: the flame; hollow cathode or electrodeless discharge lamps; and deuterium lamps. The ultraviolet radiation emitted by these sources can harm your eyes Install cathode lamp for the element to be analyzed in the AA Spectrophotometer On the computer interface, click LAMP. AA 800 has the capability of adjusting the lamp on its own and there is no need for manual intervention. 9.5 Analysis Once the system is set up, start by running four to five standards of known concentration and ensure that the calibration curve is obtained as per the conditions mentioned in section 8. The nebulizer pulls in 20 µl of the standards and displays the absorbance. Once the calibration curve is obtained, the samples can be analyzed. SOP-F-7

8 Villanova University Date: Oct 2011 Page 8 of Data Analysis and Calculations All data analysis and sample calculations are performed in the AA Laboratory Benchtop Software. The results obtained from the analysis is in the form of print outs. The concentration of the metals calculated using AA spectroscopy is populated into the excel sheet Metals Testing in the WREE folder. Care is taken to include only those concentrations that lie between the lowest and highest calibrant. Concentrations below the lowest calibrant are given a designation ND. Note: The Lab Manager may disregard one excessive outlier as long as the standards bound the data. SOP-F-8

9 Villanova University Date: Oct 2011 Page 9 of 9 Note: Data obtained from flame analysis for Zn is in ppm range (while all other metals are reported in ppb). To maintain consistency in the designated excel sheet where the results are entered, ppm units are converted to ppb units. To do so, the results of Zn analysis from the print out are multiplied by 1000 and entered into the excel sheet Waste Management All waste is disposed of in the Metals Waste Container References Perkin Elmer Corporation (1982) Analytical Methods for Atomic Absorption Spectrophotometry, Norwalk, CT. Perkin Elmer Corporation (1982) Perkin Elmer Atomic Absorption Spectrophotometer: System Description and Maintainence, Norwalk, CT. U.S. Environmental Protection Agency (USEPA). Urban Stormwater BMP Performance Monitoring U.S. Environmental Protection Agency, 2002a. SOP-F-9

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy A. Evaluation of Analytical Parameters in Atomic Absorption Spectroscopy Objective The single feature that contributes most to making atomic

More information

Brooklyn College Department of Chemistry

Brooklyn College Department of Chemistry Brooklyn College Department of Chemistry Instrumental Analysis (Chem 42/790) Atomic Absorption Spectroscopy An atomic absorption spectrometer is used in this experiment to analyze a copper-base alloy for

More information

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7 Troy, NY 12180 Date: 03/16/2012 Page 1 of 7 USGS District Laboratory, Troy, NY Inductively Coupled Plasma Optical Emission Spectrometry Standard Operating Procedure 1. Scope and Application 1.1 Analytes

More information

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method: ALLOWAY METHOD OUTLINE Standard Laboratory Method: SM Parameter: Methylene Blue Method: Colorimetric Reporting Level: Reference: 0.05 mg/l Standard Methods for the Examination Of Water and Wastewater;

More information

Atomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon

Atomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon Atomic Absorption Spectrophotometry Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon Defination In analytical chemistry, Atomic absorption spectroscopy is a

More information

DRAFT EAST AFRICAN STANDARD

DRAFT EAST AFRICAN STANDARD ICS 71.100.70 DRAFT EAST AFRICAN STANDARD Oils for cosmetic industry Methods of test Part 16 : Determination of Heavy metal Content EAST AFRICAN COMMUNITY EAS 2015 First Edition 2015 Copyright notice

More information

Atomic Absorption Spectrometer ZEEnit P series

Atomic Absorption Spectrometer ZEEnit P series Atomic Absorption Spectrometer ZEEnit P series Technical Data ZEEnit series Update 07/2014 OBue 1/ 5 ZEEnit P series Variable high-end AA Spectrometer with Deuterium and Zeeman Background Correction with

More information

CHM 317H1S Winter 2018 Section E - Flame Atomic Spectrophotometry

CHM 317H1S Winter 2018 Section E - Flame Atomic Spectrophotometry CHM 317H1S Winter 2018 Section E - Flame Atomic Spectrophotometry Section E Page 1 1. List of Experiments 1. Sodium in Drinks by Atomic Emission 2. Calcium Magnesium Water Hardness 2. Locker Inventory

More information

Sodium Chloride - Analytical Standard

Sodium Chloride - Analytical Standard Sodium Chloride - Analytical Standard Determination of Total Mercury Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-106-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes

More information

Analysis of domestic sludge using the Agilent 4200 MP-AES

Analysis of domestic sludge using the Agilent 4200 MP-AES Analysis of domestic sludge using the Agilent 4200 MP-AES Application note Environmental Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction Managing the treatment and disposal

More information

Standard Operating Procedure VUSP M

Standard Operating Procedure VUSP M Villanova University Date Nov 2009 Page 1 of 17 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP M Nitrate by Discrete Analysis Systea

More information

Experiment. Quantification of Ascorbic acid by Fluorescence Spectroscopy1

Experiment. Quantification of Ascorbic acid by Fluorescence Spectroscopy1 Experiment. Quantification of Ascorbic acid by Fluorescence Spectroscopy Modified 10/2017 Experiment. Quantification of Ascorbic acid by Fluorescence Spectroscopy1 Objective: The goal of this experiment

More information

Standard Operating Procedure VUSP N

Standard Operating Procedure VUSP N Villanova University Date Oct 2011 Page 1 of 14 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP N Nitrite, USEPA by Discrete Analysis

More information

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry NOTICE: This method has been developed by the EURL-CEFAO using the facilities available in its laboratories.

More information

COMPARISON OF ATOMIZERS

COMPARISON OF ATOMIZERS COMPARISON OF ATOMIZERS FOR ATOMIC ABSORPTION SPECTROSCOPY Introduction Atomic spectroscopic methods are all based on the interaction of light and analyte atoms in the gas phase. Thus, a common component

More information

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure Troy, NY 12180 Date: 12/12/2017 Page 1 of 7 USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure 1. Scope and Application 1.1 Analytes Dissolved inorganic

More information

SOUTH AFRICAN NATIONAL STANDARD. Standard test methods for lead in water

SOUTH AFRICAN NATIONAL STANDARD. Standard test methods for lead in water ISBN 978-0-626-19882-4 Edition 1 SOUTH AFRICAN NATIONAL STANDARD Standard test methods for lead in water This national standard is the identical implementation of ASTM D 559:200 and is adopted with the

More information

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) 1.0 SCOPE AND APPLICATION 1.1 Method 7060 is an atomic absorption procedure approved for determining the concentration of arsenic in wastes,

More information

EXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy

EXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy EXPERIMENT 7 Determination of Sodium by Flame Atomic-Emission Spectroscopy USE ONLY DEIONIZED WATER (NOT DISTILLED WATER!) THROUGHOUT THE ENTIRE EXPERIMENT Distilled water actually has too much sodium

More information

ANALYSIS OF ZINC IN HAIR USING FLAME ATOMIC ABSORPTION SPECTROSCOPY

ANALYSIS OF ZINC IN HAIR USING FLAME ATOMIC ABSORPTION SPECTROSCOPY ANALYSIS OF ZINC IN HAIR USING FLAME ATOMIC ABSORPTION SPECTROSCOPY Introduction The purpose of this experiment is to determine the concentration of zinc in a sample of hair. You will use both the calibration

More information

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD) 683 METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD) NOTE: This method does not include all of the specifications (e.g., equipment and

More information

3. Chemical Hygiene Plan: Laboratory Standard Operating Procedures. A. Laboratory Specific Information and Signatures

3. Chemical Hygiene Plan: Laboratory Standard Operating Procedures. A. Laboratory Specific Information and Signatures 3. Chemical Hygiene Plan: Laboratory Standard Operating Procedures A. Laboratory Specific Information and Signatures The Chemical Hygiene Plan: Laboratory Standard Operating Procedures (section 3 only),

More information

FLAME PHOTOMETRY AIM INTRODUCTION

FLAME PHOTOMETRY AIM INTRODUCTION FLAME PHOTOMETRY AIM INTRODUCTION Atomic spectroscopy is based on the absorption, emission or fluorescence process of light by atoms or elementary ions. Information for atomic scale is obtained in two

More information

ALLOWAY METHOD OUTLINE

ALLOWAY METHOD OUTLINE ALLOWAY METHOD OUTLINE Standard Laboratory Method SM4500-Cl -G Parameter Residual Chlorine & Free Chlorine Method DPD Colorimetric Test Kit Date Issued Originator: Section Supervisor: QA Manager Date:

More information

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared by: OEWRI Laboratory Manager

More information

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Elemental analysis of river sediment using the Agilent 4200 MP-AES Elemental analysis of river sediment using the Agilent 4200 MP-AES Application note Environmental: Soils, sludges & sediments Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction

More information

Fluorescence Spectrophotometry

Fluorescence Spectrophotometry Chemistry 422L Manual Page 27 I. Introduction Fluorescence Spectrophotometry Ru(bpy) 3 2+, where bpy = 2, 2' bipyridine, has been one of the most widely studied metal complexes in recent years. Interest

More information

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Maintenance with OQ. ZEEnit 700. Operational Qualification OQ within the scope of a maintenance for the. Atomic Absorption Spectrometer

Maintenance with OQ. ZEEnit 700. Operational Qualification OQ within the scope of a maintenance for the. Atomic Absorption Spectrometer Maintenance with OQ Operational Qualification OQ within the scope of a maintenance for the Atomic Absorption Spectrometer ZEEnit 700 Contents 1. GENERAL INFORMATION...4 2. DEFINITIONS...4 2.1 Validation...4

More information

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows Nitrate DOC316.53.01067 Cadmium Reduction Method Method 8192 0.01 to 0.50 mg/l NO 3 N (LR) Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

TPH (Total Petroleum Hydrocarbons)

TPH (Total Petroleum Hydrocarbons) TPH (Total Petroleum Hydrocarbons) DOC316.53.01475 Immunoassay 1 Method 10050 Scope and application: For water. 1 This test is semi-quantitative. Results are shown as more or less than the threshold value

More information

3 - Atomic Absorption Spectroscopy

3 - Atomic Absorption Spectroscopy 3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,

More information

Exercise 2-4. Titration of a Buffer Solution EXERCISE OBJECTIVES

Exercise 2-4. Titration of a Buffer Solution EXERCISE OBJECTIVES Exercise 2-4 Titration of a Buffer Solution EXERCISE OBJECTIVES To define the terms buffer solution and buffer capacity; To titrate a buffer solution with a weak acid solution; To plot a graph using the

More information

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation Phosphorus, Total DOC316.53.01121 USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method 8190 0.06 to 3.50 mg/l PO 3 4 (0.02 to 1.10 mg/l P) Test N Tube Vials Scope and application: For water,

More information

A Spectrophotometric Analysis of Calcium in Cereal

A Spectrophotometric Analysis of Calcium in Cereal CHEM 311L Quantitative Analysis Laboratory Revision 1.2 A Spectrophotometric Analysis of Calcium in Cereal In this laboratory exercise, we will determine the amount of Calium in a serving of cereal. We

More information

Sodium Chloride - Analytical Standard

Sodium Chloride - Analytical Standard Sodium Chloride - Analytical Standard Determination of Arsenic Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-105-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes

More information

Geneva College Hazard Communication Program Presentation

Geneva College Hazard Communication Program Presentation Geneva College Hazard Communication Program Presentation Design 2005, 2012 Zywave, Inc. All rights reserved. Hazard Communication: Agenda In today s session, we will discuss the following: - Our Hazard

More information

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission 0 chapter Sodium and Potassium Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission Spectroscopy 67 S. S. Nielsen (ed.), Food Analysis Laboratory Manual Springer

More information

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows Potassium DOC316.53.01127 Tetraphenylborate Method Method 8049 0.1 to 7.0 mg/l K Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific information

More information

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY 1.0 SCOPE AND APPLICATION 1.1 This digestion procedure is used for the preparation

More information

Individual Laboratory Safety & Chemical Hygiene Plan

Individual Laboratory Safety & Chemical Hygiene Plan APPENDIX A: CHEMICAL HYGIENE PLAN Part B Individual Laboratory Safety & Chemical Hygiene Plan Principal Investigator Date filed Department Building/Floor/Room E-mail address @Pace.edu Phone Facsimile Laboratory

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University. Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared

More information

Introduction to Spectroscopy: Analysis of Copper Ore

Introduction to Spectroscopy: Analysis of Copper Ore Introduction to Spectroscopy: Analysis of Copper Ore Using a Buret and Volumetric Flask: 2.06 ml of solution delivered 2.47 ml of solution delivered 50.00 ml Volumetric Flask Reading a buret: Burets are

More information

Globally Harmonized System of Classification and Labeling of Chemicals (GHS) LC-1033p Rev. 06/16

Globally Harmonized System of Classification and Labeling of Chemicals (GHS) LC-1033p Rev. 06/16 Globally Harmonized System of Classification and Labeling of Chemicals (GHS) Rev. 06/16 Agenda Introduction to Globally Harmonized System of Classifying and Labeling Chemicals Hazard & Precautionary Statements

More information

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis,

More information

Part II. Cu(OH)2(s) CuO(s)

Part II. Cu(OH)2(s) CuO(s) The Copper Cycle Introduction In this experiment, you will carry out a series of reactions starting with copper metal. This will give you practice handling chemical reagents and making observations. It

More information

Experiment 7A ANALYSIS OF BRASS

Experiment 7A ANALYSIS OF BRASS Experiment 7A ANALYSIS OF BRASS FV 10/21/10 MATERIALS: Spectronic 20 spectrophotometers, 2 cuvettes, brass sample, 7 M HNO 3, 0.100 M CuSO 4, 2 M NH 3, two 50 ml beakers, 100 ml beaker, two 25 ml volumetric

More information

Experiment #7. Determination of an Equilibrium Constant

Experiment #7. Determination of an Equilibrium Constant Experiment #7. Determination of an Equilibrium Constant Introduction It is frequently assumed that reactions go to completion, that all of the reactants are converted into products. Most chemical reactions

More information

aa + bb cc + dd Equation 1

aa + bb cc + dd Equation 1 Experiment: The Determination of K eq for FeSCN 2+ Introduction For any reversible chemical reaction at equilibrium, the concentrations of all reactants and products are constant or stable. There is no

More information

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II) : Absorption Spectroscopy of Cobalt(II) OBJECTIVES In successfully completing this lab you will: prepare a stock solution using a volumetric flask; use a UV/Visible spectrometer to measure an absorption

More information

Plant Indicators for Acids and Bases

Plant Indicators for Acids and Bases SCIENCE EXPERIMENTS ON FILE Revised Edition 5.28-1 Plant Indicators for Acids and Bases Carole R. Goshorn Topic Acid/base indicators Time Preparation by teacher 50 minutes; Part B, 45 minutes! Safety Adult

More information

Determining the Concentration of a Solution: Beer s Law

Determining the Concentration of a Solution: Beer s Law Determining the Concentration of a Solution: Beer s Law Vernier Spectrometer 1 The primary objective of this experiment is to determine the concentration of an unknown copper (II) sulfate solution. You

More information

Test Method: CPSC-CH-E

Test Method: CPSC-CH-E UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 10901 DARNESTOWN RD GAITHERSBURG, MD 20878 Test Method: CPSC-CH-E1001-08 Standard Operating Procedure

More information

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Nitrate, MR DOC316.53.01069 Cadmium Reduction Method Method 8171 0.1 to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Scope and application: For water, wastewater

More information

RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION

RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION Rate Law Determination of Crystal Violet Hydroxylation Revised 5/22/12 RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION Adapted from "Chemistry with Computers" Vernier Software, Portland OR, 1997

More information

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater. Nitrogen, Ammonia DOC316.53.01078 USEPA 1 Nessler Method 2 Method 8038 0.02 to 2.50 mg/l NH 3 N Reagent Solution Scope and application: For water, wastewater and seawater. Distillation is required for

More information

novaa 800 D Atomic Absorption Spectrometer

novaa 800 D Atomic Absorption Spectrometer Technical Data Atomic Absorption Spectrometer Cpt : +27 (0) 21 905 0476 Jhb : +27 (0) 11 794 Dbn : +27 (0) 31 266 2454 1/7 General The is a compact atomic absorption spectrometer with deuterium background

More information

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right. Barium DOC316.53.01315 Turbidimetric Method 1 Method 10251 2 to 100, 20 to 1000, 200 to 10,000 mg/l Ba (spectrophotometers) Powder Pillows 2 to 80, 20 to 800, 200 to 8000 mg/l Ba (colorimeters) Scope and

More information

Chemical Health and Safety General Program

Chemical Health and Safety General Program Chemical Health and Safety General Program I. Objective To establish minimum requirements for storage, handling and use of chemicals. II. Scope This process applies to employees and operations involved

More information

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION Written by Pamela Doolittle, University of Wisconsin- Madison, pssemrad@wisc.edu 2014 In this experiment you will determine the phosphorus (P) concentration

More information

Standard Operating Procedure

Standard Operating Procedure Standard Operating Procedure Procedure Radioimmunoassay with I Department Location SOP Prepared By: Section 1: Purpose Radioimmunoassays are used for detecting the concentration of a specific antigen or

More information

MEASUREMENT: PART II

MEASUREMENT: PART II 1 MEASUREMENT: PART II Copyright: Department of Chemistry, University of Idaho, Moscow, ID 83844-2343, 2013. INTRODUCTION Read and/or review Section 1.7 and Figure 7.5 in your textbook. The first part

More information

Application Submitted

Application Submitted Proposal #6 Application Submitted To: Program: Title of Project: Suggested Timeline: Dr. Arlene Courtney Ch 462 Experimental Chemistry Analysis of Lead leached from Fiesta Ware by use of Atomic Absorption

More information

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction application Note Atomic Absorption Authors David Bass Senior Product Specialist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT 06484 USA The Determination of Toxic, Trace, and

More information

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Evrim Kilicgedik Product Specialist, Atomic Spectroscopy Agilent Technologies 04.11.2011 2011 The Atomic Spectroscopy

More information

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters) Aluminum DOC316.53.01002 Aluminon Method 1 Method 8012 0.008 to 0.800 mg/l Al 3+ (spectrophotometers) Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters) Scope and application: For water and wastewater.

More information

SAFETY TRAINING LEAFLET 08 ACETYLENE, CALCIUM CARBIDE, LIME SLUDGE AND PURIFYING MATERIALS

SAFETY TRAINING LEAFLET 08 ACETYLENE, CALCIUM CARBIDE, LIME SLUDGE AND PURIFYING MATERIALS SAFETY TRAINING LEAFLET 08 ACETYLENE, CALCIUM CARBIDE, LIME SLUDGE AND PURIFYING MATERIALS Doc 23.08/18 EUROPEAN INDUSTRIAL GASES ASSOCIATION AISBL AVENUE DES ARTS 3-5 B 1210 BRUSSELS Tel: +32 2 217 70

More information

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Method 1061 Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Company Method 1061 Revision 1. December 015 Organic Carbon in Finished

More information

Atomic Spectroscopy AA/ICP/ICPMS:

Atomic Spectroscopy AA/ICP/ICPMS: Atomic Spectroscopy AA/ICP/ICPMS: A Comparison of Techniques VA AWWA/VWEA Lab Practices Conference July 25, 2016 Dan Davis Shimadzu Scientific Instruments AA/ICP/ICPMS: A Comparison of Techniques Topics

More information

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By the Authority Vested By Part 5 of the United States Code 552(a) and Part 1 of the Code of Regulations 51 the attached document has

More information

Experimental Procedure Lab 402

Experimental Procedure Lab 402 Experimental Procedure Lab 402 Overview One set of solutions having known molar concentrations of FeNCS 2+ is prepared for a calibration curve, a plot of absorbance versus concentration. A second set of

More information

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C Book 5 - Laboratory Policies and Procedures C03-1 2009-06-15 1 Scope 1.1 This method describes a general procedure for the determination of leachable arsenic, selenium, cadmium, antimony, and barium in

More information

Experiment #9. Atomic Emission Spectroscopy

Experiment #9. Atomic Emission Spectroscopy Introduction Experiment #9. Atomic Emission Spectroscopy Spectroscopy is the study of the interaction of light with matter. This interaction can be in the form of the absorption or the emission of electromagnetic

More information

Facilities Management

Facilities Management Policy Number: 700.20 Title: Chemical Fume Hood Policy Implementation Date: 2002 Last Audited: August, 2017 Last Revised: October 23rd, 2017 Facilities Management Introduction The laboratory chemical fume

More information

Exercise 2-2. Titration of a Strong Acid EXERCISE OBJECTIVES

Exercise 2-2. Titration of a Strong Acid EXERCISE OBJECTIVES Exercise 2-2 Titration of a Strong Acid EXERCISE OBJECTIVES To describe the effect of a ph variation on a chemical indicator; To titrate water containing a strong base solution with a strong acid solution;

More information

Montgomery County Community College Document Number: MET DeKalb Pike Revision Number: 0

Montgomery County Community College Document Number: MET DeKalb Pike Revision Number: 0 Page 1 of 5 Approvals: Preparer: Jason McMillan Reviewer: Dr. Maggie Bryans Date: 08JAN14 Date: 10JAN14 1. Purpose: Operation of Eppendorf Research Plus through to the P-1000. 2. Scope: Applies to the

More information

EnAlgae SOP sheet: Silicate - SFA. 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2.

EnAlgae SOP sheet: Silicate - SFA. 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2. EnAlgae SOP sheet: Silicate - SFA 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2. 2. Principle Analysis is performed on a segmented flow analyser (SFA), model AutoAnalyser

More information

2101 Atomic Spectroscopy

2101 Atomic Spectroscopy 2101 Atomic Spectroscopy Atomic identification Atomic spectroscopy refers to the absorption and emission of ultraviolet to visible light by atoms and monoatomic ions. It is best used to analyze metals.

More information

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors Determination of major, minor and trace elements in rice flour using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Application note Food testing Authors John Cauduro Agilent Technologies,

More information

Micro Volume QuEChERS kit

Micro Volume QuEChERS kit 225-37872 Sep. 2018 Small Capacity Pretreatment Kit Micro Volume QuEChERS kit Instruction Manual Read this manual thoroughly before you use the product. Keep this manual for future reference. This page

More information

Atomic Spectra for Atoms and Ions. Light is made up of different wavelengths

Atomic Spectra for Atoms and Ions. Light is made up of different wavelengths Atomic Spectra for Atoms and Ions What will you be doing in lab next week? Recording the line spectra of several different substances in discharge tubes. Recording the line spectra of several ions from

More information

Chloride, Nitrate, Nitrite, Phosphate Determination High Performance Liquid Chromatography (HPLC) Procedures

Chloride, Nitrate, Nitrite, Phosphate Determination High Performance Liquid Chromatography (HPLC) Procedures Villanova University Date Oct: 2008 Page 1 of 21 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP C Chloride, Nitrate, Nitrite, Phosphate

More information

Electrochemical Cells

Electrochemical Cells Electrochemical Cells PURPOSE To see how changes in concentration and ph affect the potential in an electrochemical cell, and confirm the Nernst equation. GOALS To examine how standard reduction potentials

More information

THE IRON(III) THIOCYANATE REACTION SYSTEM

THE IRON(III) THIOCYANATE REACTION SYSTEM Experiment 7 THE IRON(III) THIOCYANATE REACTION SYSTEM Prepared by Ross S. Nord, Chemistry Department, Eastern Michigan University PURPOSE To investigate a novel reaction system by utilizing a spectrophotometer.

More information

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013 Absorbance Experiment 13 Dilutions and Data Handling in a Spreadsheet rev 1/2013 GOAL: This lab experiment will provide practice in making dilutions using pipets and introduce basic spreadsheet skills

More information

REVIEW OF LAB TECHNIQUES

REVIEW OF LAB TECHNIQUES Experiment 1 REVIEW OF LAB TECHNIQUES Prepared by Masanobu M. Yamauchi and Ross S. Nord, Eastern Michigan University PURPOSE To review density calculations, Beer s Law and the use of electronic balances,

More information

Chemical Equilibrium: Finding a Constant, Kc

Chemical Equilibrium: Finding a Constant, Kc Chemical Equilibrium: Finding a Constant, Kc Experiment 20 The purpose of this lab is to experimentally determine the equilibrium constant, K c, for the following chemical reaction: Fe 3+ (aq) + SCN (aq)

More information

Safe Use of Pyrophoric/Water Reactive Materials

Safe Use of Pyrophoric/Water Reactive Materials Introduction Safe Use of Pyrophoric/Water Reactive Materials Pyrophoric and water reactive materials can ignite spontaneously on contact with air, moisture in the air, or water. Improper handling of these

More information

REVIEW OF LAB TECHNIQUES

REVIEW OF LAB TECHNIQUES Experiment 1 REVIEW OF LAB TECHNIQUES Prepared by Masanobu M. Yamauchi and Ross S. Nord, Eastern Michigan University PURPOSE To review density calculations, Beer s Law and the use of electronic balances,

More information

Safety in the Chemistry Laboratory

Safety in the Chemistry Laboratory Safety in the Chemistry Laboratory CHAPTER1 Safety must be everyone s primary concern in the chemistry lab. Understanding and following all safety rules in the organic chemistry lab is critical to your

More information

Chemistry 1215 Experiment #11 Spectrophotometric Analysis of an Unknown Brass Sample

Chemistry 1215 Experiment #11 Spectrophotometric Analysis of an Unknown Brass Sample Chemistry 1215 Experiment #11 Spectrophotometric Analysis of an Unknown Brass Sample Objective In this experiment you will use spectrophotometric measurements to determine the copper concentration of a

More information

Introduction to Spectroscopy: Analysis of Copper Ore

Introduction to Spectroscopy: Analysis of Copper Ore Absorbance Introduction to Spectroscopy: Analysis of Copper Ore Introduction The goal of this lab is to determine the unknown concentration of two different copper solution samples, taken from fictitious

More information

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water Hach Method 1041 Spectrophotometric Measurement of Free Chlorine (Cl ) in Finished Drinking Water Hach Company Method 1041 Revision 1. November 015 Spectrophotometric Measurement of Free Cl in Finished

More information

Experiment 3: Atomic Absorption Determination of Lead in Soil Samples Inspired by Mielke, H. American Scientist 1999, 87,

Experiment 3: Atomic Absorption Determination of Lead in Soil Samples Inspired by Mielke, H. American Scientist 1999, 87, Experiment 3: Atomic Absorption Determination of Lead in Soil Samples Inspired by Mielke, H. American Scientist 1999, 87, 62-73. Introduction Experimental Work on March 2 and March 3-24 (in Olin-Rice 387)

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01104 USEPA Reactor Digestion Method Method 10212 250 to 15,000 mg/l COD (UHR) TNTplus 823 Scope and application: For wastewater and process waters; digestion is required.

More information

SECTION 3 PRODUCT COMPOSITION The test kit is composed of QuSTICK Strep A Reagent Stick, Reagent A, Reagent B, Positive Control, and Negative Control.

SECTION 3 PRODUCT COMPOSITION The test kit is composed of QuSTICK Strep A Reagent Stick, Reagent A, Reagent B, Positive Control, and Negative Control. SECTION 1 PRODUCT AND COMPANY IDENTIFICATION Product Identification: Ref No. 6000, 6000-025, etc. Test kit contains QuSTICK Strep A Reagent Stick, Reagent A, Reagent B, Positive Control, and Negative Control.

More information

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Chem 1B Saddleback College Dr. White 1 Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Objectives To use spectroscopy to relate the absorbance of a colored solution to its concentration. To

More information

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy Atomic line widths: Narrow line widths reduce the possibility of spectral overlap and thus interferences. The band width at half height is used to indicate width. This is also sometimes called the effective

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 7 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information