1 Villanova University Date: Oct 2011 Page 1 of 9 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP F Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures Author: Water Resources Laboratory Technician Sonali Joshi Approved: VUSP Director Robert G. Traver SOP-F-1
2 Villanova University Date: Oct 2011 Page 2 of 9 Distribution John Komlos Quality Assurance Officer Laboratory Copy, maintained by Quality Assurance Officer Revision Date Responsible Person Description of Change 0 01 Jan 2010 Sonali Joshi Initial 1 01 Oct 2011 Sonali Joshi Minor Update 2 The following laboratory staff have read and been trained on this Manual. A copy of this page will be maintained in the Laboratory. Training is good for one year. Signature Name / Position Date Trainer Robert G. Traver Director Sonali Joshi Laboratory Technician John Komlos Quality Assurance Officer SOP-F-2
3 Villanova University Date: Oct 2011 Page 3 of 9 TABLE OF CONTENTS 1.0 SCOPE AND APPLICATION METHOD SUMMARY DEFINITIONS HEALTH AND SAFETY WARNINGS INTERFERENCES HANDLING AND PRESERVATION EQUIPMENT AND SUPPLIES QUALITY CONTROL PROCEDURES DATA ANALYSIS AND CALCULATIONS WASTE MANAGEMENT REFERENCES 8 SOP-F-3
4 Villanova University Date: Oct 2011 Page 4 of Scope and Application The purpose of this standard operating procedure (SOP) is to show the proper set up, procedures, and operation steps for Zinc in stormwater samples using the Perkin Elmer AAnalyst 800 Flame Atomic Absorption Spectrometer. This document serves as the basis for maintaining accurate and reproducible results, and defines pertinent validation criteria for accepting/rejecting final data. Standards and blanks will be employed as a means of validating data within range of the instrument s detection limits Method Summary The Perkin-Elmer Model AAnalyst 800 is an atomic absorption spectrometer which measures the concentrations of metals in various matrices. For our purposes, it provides integrated readings in absorbances over a period selected by the user from 0.2 to 60 seconds. The liquid sample is sprayed by a pneumatic nebulizer into a flame. Sample particles vaporize and dissociate into free atoms due to the temperature and reducing power of the flame. The concentration of metal element in the sample is approximated in the parts per million range from the absorbance and standard calibration curve Definitions NA 4.0 Health and Safety Warnings 4.1 Close the door on the instrument before turning on the flame. Never look directly at the flame. The UV light can cause serious eye damage. 4.2 Do not use oxygen in the burner. 4.3 Be sure there an adequate ventilation system for the removal of exhaust gases. 4.4 Change the acetylene tank when the pressure falls to 520 kpa (75 psig). Do not set the acetylene outlet pressure above 110 kpa (15 psig). 4.5 If high concentrations of silver, copper, or mercury salts have been nebulized into an acetylene flame, clean the burner chamber immediately after use. SOP-F-4
5 Villanova University Date: Oct 2011 Page 5 of The escape of fuel gas or fuel gas mixture from the drain tube, the burner system, the gas control, or the gas connections results in a serious explosion hazard. 4.7 Locate the acetylene tank away from sources of heat, open flames, or sparks. Make sure cylinder valves are closed when not in use Interferences NONE 6.0. Handling and Preservation All glassware and plastic ware used for collection, transportation, and laboratory analysis of the metal samples are acid-washed using the following procedure. The glassware and plastic ware are first cleaned with a laboratory detergent and rinsed with tap water. Next, the container is rinsed in 1:10 Nitric Acid Solution. The container is then rinsed with distilled water three times and allowed to air dry. The samples are preserved in a 2% nitric acid matrix and refrigerated. They are analyzed within six months of preservation Equipment and Supplies 7.1 Adequate ventilation system 7.2 Supply of compressed air. 7.3 Tank of acetylene 7.4 Zinc Source Lamp 7.5 Sample containers 7.6 Concentrated Nitric Acid 7.7 Stock Zinc Solution SOP-F-5
6 Villanova University Date: Oct 2011 Page 6 of Quality Control A system of standard solutions and blanks are used to check the laboratory techniques for the AA800 Atomic Absorption Spectrometer. The blanks are samples of 2% nitric acid that are treated in the same manner as the samples to be tested. A set of four to five standards ranging from 0.05 to 10 ppm of Zinc is used to attain a linear absorbance-concentration curve, with a correlation coefficient greater than or equal to To ensure accuracy, a check standard must be applied for at least every 10 samples (note start with the lowest standard). The check standard must agree with the initial reading as seen on the calibration curve by 25%. If agreement fails, it is presumed that the readings of the set of samples run before the check standard are unreliable and must be rerun. Trouble shooting begins with checking the instrument and procedure for any obvious flaws. Next, the reagents should be checked and re-prepared, if necessary. Finally, at least one set of duplicate samples are run for every batch analyzed. In this test, one sampling site is chosen at random in which two separate samples are taken and treated/analyzed separately. Both samples should render readings within 10% of each other to be considered valid. If they do not fall within this range, procedures are examined to be sure that all samples were treated in the same manner. SOP-F-6
7 Villanova University Date: Oct 2011 Page 7 of Procedures 9.1 Matrix preparation Sample matrix: 2%v Trace Metal grade nitric acid (HNO 3 ) prepared with MQ water. 9.2 Standards preparation Master stock solutions are available for Zn. These have expiration dates and must be replaced on a regular basis. Four to five standards are prepared by weight from the stock, and diluted using the 2% Nitric Acid as the matrix. 9.3 Switching ON the system Switch the spectrometer ON Set the acetylene tank output pressure to 69 kpa (10 psig) Set the air supply pressure to kpa (80-90psig) Press FLAME ON/OFF to ignite the flame Allow burner head to warm up If necessary, readjust the acetylene pressure to 69 kpa (10 psig) Turn on the computer and start the software by clicking on the icon WinLab32 for AA The computer interface will start communicating with the equipment When the communication is successful, check marks appear on the screen 9.4 Installing and aligning lamps Always wear ultraviolet-absorbing safety glasses when viewing the following sources: the flame; hollow cathode or electrodeless discharge lamps; and deuterium lamps. The ultraviolet radiation emitted by these sources can harm your eyes Install cathode lamp for the element to be analyzed in the AA Spectrophotometer On the computer interface, click LAMP. AA 800 has the capability of adjusting the lamp on its own and there is no need for manual intervention. 9.5 Analysis Once the system is set up, start by running four to five standards of known concentration and ensure that the calibration curve is obtained as per the conditions mentioned in section 8. The nebulizer pulls in 20 µl of the standards and displays the absorbance. Once the calibration curve is obtained, the samples can be analyzed. SOP-F-7
8 Villanova University Date: Oct 2011 Page 8 of Data Analysis and Calculations All data analysis and sample calculations are performed in the AA Laboratory Benchtop Software. The results obtained from the analysis is in the form of print outs. The concentration of the metals calculated using AA spectroscopy is populated into the excel sheet Metals Testing in the WREE folder. Care is taken to include only those concentrations that lie between the lowest and highest calibrant. Concentrations below the lowest calibrant are given a designation ND. Note: The Lab Manager may disregard one excessive outlier as long as the standards bound the data. SOP-F-8
9 Villanova University Date: Oct 2011 Page 9 of 9 Note: Data obtained from flame analysis for Zn is in ppm range (while all other metals are reported in ppb). To maintain consistency in the designated excel sheet where the results are entered, ppm units are converted to ppb units. To do so, the results of Zn analysis from the print out are multiplied by 1000 and entered into the excel sheet Waste Management All waste is disposed of in the Metals Waste Container References Perkin Elmer Corporation (1982) Analytical Methods for Atomic Absorption Spectrophotometry, Norwalk, CT. Perkin Elmer Corporation (1982) Perkin Elmer Atomic Absorption Spectrophotometer: System Description and Maintainence, Norwalk, CT. U.S. Environmental Protection Agency (USEPA). Urban Stormwater BMP Performance Monitoring U.S. Environmental Protection Agency, 2002a. SOP-F-9
Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy
Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy A. Evaluation of Analytical Parameters in Atomic Absorption Spectroscopy Objective The single feature that contributes most to making atomic
More informationBrooklyn College Department of Chemistry
Brooklyn College Department of Chemistry Instrumental Analysis (Chem 42/790) Atomic Absorption Spectroscopy An atomic absorption spectrometer is used in this experiment to analyze a copper-base alloy for
More informationUSGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7
Troy, NY 12180 Date: 03/16/2012 Page 1 of 7 USGS District Laboratory, Troy, NY Inductively Coupled Plasma Optical Emission Spectrometry Standard Operating Procedure 1. Scope and Application 1.1 Analytes
More informationRevision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:
ALLOWAY METHOD OUTLINE Standard Laboratory Method: SM Parameter: Methylene Blue Method: Colorimetric Reporting Level: Reference: 0.05 mg/l Standard Methods for the Examination Of Water and Wastewater;
More informationAtomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon
Atomic Absorption Spectrophotometry Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon Defination In analytical chemistry, Atomic absorption spectroscopy is a
More informationDRAFT EAST AFRICAN STANDARD
ICS 71.100.70 DRAFT EAST AFRICAN STANDARD Oils for cosmetic industry Methods of test Part 16 : Determination of Heavy metal Content EAST AFRICAN COMMUNITY EAS 2015 First Edition 2015 Copyright notice
More informationAtomic Absorption Spectrometer ZEEnit P series
Atomic Absorption Spectrometer ZEEnit P series Technical Data ZEEnit series Update 07/2014 OBue 1/ 5 ZEEnit P series Variable high-end AA Spectrometer with Deuterium and Zeeman Background Correction with
More informationCHM 317H1S Winter 2018 Section E - Flame Atomic Spectrophotometry
CHM 317H1S Winter 2018 Section E - Flame Atomic Spectrophotometry Section E Page 1 1. List of Experiments 1. Sodium in Drinks by Atomic Emission 2. Calcium Magnesium Water Hardness 2. Locker Inventory
More informationSodium Chloride - Analytical Standard
Sodium Chloride - Analytical Standard Determination of Total Mercury Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-106-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes
More informationAnalysis of domestic sludge using the Agilent 4200 MP-AES
Analysis of domestic sludge using the Agilent 4200 MP-AES Application note Environmental Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction Managing the treatment and disposal
More informationStandard Operating Procedure VUSP M
Villanova University Date Nov 2009 Page 1 of 17 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP M Nitrate by Discrete Analysis Systea
More informationExperiment. Quantification of Ascorbic acid by Fluorescence Spectroscopy1
Experiment. Quantification of Ascorbic acid by Fluorescence Spectroscopy Modified 10/2017 Experiment. Quantification of Ascorbic acid by Fluorescence Spectroscopy1 Objective: The goal of this experiment
More informationStandard Operating Procedure VUSP N
Villanova University Date Oct 2011 Page 1 of 14 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP N Nitrite, USEPA by Discrete Analysis
More informationDirect Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry
Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry NOTICE: This method has been developed by the EURL-CEFAO using the facilities available in its laboratories.
More informationCOMPARISON OF ATOMIZERS
COMPARISON OF ATOMIZERS FOR ATOMIC ABSORPTION SPECTROSCOPY Introduction Atomic spectroscopic methods are all based on the interaction of light and analyte atoms in the gas phase. Thus, a common component
More informationUSGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure
Troy, NY 12180 Date: 12/12/2017 Page 1 of 7 USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure 1. Scope and Application 1.1 Analytes Dissolved inorganic
More informationSOUTH AFRICAN NATIONAL STANDARD. Standard test methods for lead in water
ISBN 978-0-626-19882-4 Edition 1 SOUTH AFRICAN NATIONAL STANDARD Standard test methods for lead in water This national standard is the identical implementation of ASTM D 559:200 and is adopted with the
More informationMETHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)
METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE) 1.0 SCOPE AND APPLICATION 1.1 Method 7060 is an atomic absorption procedure approved for determining the concentration of arsenic in wastes,
More informationEXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy
EXPERIMENT 7 Determination of Sodium by Flame Atomic-Emission Spectroscopy USE ONLY DEIONIZED WATER (NOT DISTILLED WATER!) THROUGHOUT THE ENTIRE EXPERIMENT Distilled water actually has too much sodium
More informationANALYSIS OF ZINC IN HAIR USING FLAME ATOMIC ABSORPTION SPECTROSCOPY
ANALYSIS OF ZINC IN HAIR USING FLAME ATOMIC ABSORPTION SPECTROSCOPY Introduction The purpose of this experiment is to determine the concentration of zinc in a sample of hair. You will use both the calibration
More informationMETHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)
683 METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD) NOTE: This method does not include all of the specifications (e.g., equipment and
More information3. Chemical Hygiene Plan: Laboratory Standard Operating Procedures. A. Laboratory Specific Information and Signatures
3. Chemical Hygiene Plan: Laboratory Standard Operating Procedures A. Laboratory Specific Information and Signatures The Chemical Hygiene Plan: Laboratory Standard Operating Procedures (section 3 only),
More informationFLAME PHOTOMETRY AIM INTRODUCTION
FLAME PHOTOMETRY AIM INTRODUCTION Atomic spectroscopy is based on the absorption, emission or fluorescence process of light by atoms or elementary ions. Information for atomic scale is obtained in two
More informationALLOWAY METHOD OUTLINE
ALLOWAY METHOD OUTLINE Standard Laboratory Method SM4500-Cl -G Parameter Residual Chlorine & Free Chlorine Method DPD Colorimetric Test Kit Date Issued Originator: Section Supervisor: QA Manager Date:
More informationStandard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and
Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared by: OEWRI Laboratory Manager
More informationElemental analysis of river sediment using the Agilent 4200 MP-AES
Elemental analysis of river sediment using the Agilent 4200 MP-AES Application note Environmental: Soils, sludges & sediments Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction
More informationFluorescence Spectrophotometry
Chemistry 422L Manual Page 27 I. Introduction Fluorescence Spectrophotometry Ru(bpy) 3 2+, where bpy = 2, 2' bipyridine, has been one of the most widely studied metal complexes in recent years. Interest
More informationCadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls
Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific
More informationMaintenance with OQ. ZEEnit 700. Operational Qualification OQ within the scope of a maintenance for the. Atomic Absorption Spectrometer
Maintenance with OQ Operational Qualification OQ within the scope of a maintenance for the Atomic Absorption Spectrometer ZEEnit 700 Contents 1. GENERAL INFORMATION...4 2. DEFINITIONS...4 2.1 Validation...4
More informationCadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows
Nitrate DOC316.53.01067 Cadmium Reduction Method Method 8192 0.01 to 0.50 mg/l NO 3 N (LR) Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific
More informationTPH (Total Petroleum Hydrocarbons)
TPH (Total Petroleum Hydrocarbons) DOC316.53.01475 Immunoassay 1 Method 10050 Scope and application: For water. 1 This test is semi-quantitative. Results are shown as more or less than the threshold value
More information3 - Atomic Absorption Spectroscopy
3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,
More informationExercise 2-4. Titration of a Buffer Solution EXERCISE OBJECTIVES
Exercise 2-4 Titration of a Buffer Solution EXERCISE OBJECTIVES To define the terms buffer solution and buffer capacity; To titrate a buffer solution with a weak acid solution; To plot a graph using the
More informationPhosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation
Phosphorus, Total DOC316.53.01121 USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method 8190 0.06 to 3.50 mg/l PO 3 4 (0.02 to 1.10 mg/l P) Test N Tube Vials Scope and application: For water,
More informationA Spectrophotometric Analysis of Calcium in Cereal
CHEM 311L Quantitative Analysis Laboratory Revision 1.2 A Spectrophotometric Analysis of Calcium in Cereal In this laboratory exercise, we will determine the amount of Calium in a serving of cereal. We
More informationSodium Chloride - Analytical Standard
Sodium Chloride - Analytical Standard Determination of Arsenic Former numbering: ECSS/CN 312-1982 & ESPA/CN-E-105-1994 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes
More informationGeneva College Hazard Communication Program Presentation
Geneva College Hazard Communication Program Presentation Design 2005, 2012 Zywave, Inc. All rights reserved. Hazard Communication: Agenda In today s session, we will discuss the following: - Our Hazard
More informationDeterminations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission
0 chapter Sodium and Potassium Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission Spectroscopy 67 S. S. Nielsen (ed.), Food Analysis Laboratory Manual Springer
More informationTetraphenylborate Method Method to 7.0 mg/l K Powder Pillows
Potassium DOC316.53.01127 Tetraphenylborate Method Method 8049 0.1 to 7.0 mg/l K Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific information
More informationMETHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY
METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY 1.0 SCOPE AND APPLICATION 1.1 This digestion procedure is used for the preparation
More informationIndividual Laboratory Safety & Chemical Hygiene Plan
APPENDIX A: CHEMICAL HYGIENE PLAN Part B Individual Laboratory Safety & Chemical Hygiene Plan Principal Investigator Date filed Department Building/Floor/Room E-mail address @Pace.edu Phone Facsimile Laboratory
More informationSTANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS
PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
More informationStandard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.
Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared
More informationIntroduction to Spectroscopy: Analysis of Copper Ore
Introduction to Spectroscopy: Analysis of Copper Ore Using a Buret and Volumetric Flask: 2.06 ml of solution delivered 2.47 ml of solution delivered 50.00 ml Volumetric Flask Reading a buret: Burets are
More informationGlobally Harmonized System of Classification and Labeling of Chemicals (GHS) LC-1033p Rev. 06/16
Globally Harmonized System of Classification and Labeling of Chemicals (GHS) Rev. 06/16 Agenda Introduction to Globally Harmonized System of Classifying and Labeling Chemicals Hazard & Precautionary Statements
More informationStandard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1
Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis,
More informationPart II. Cu(OH)2(s) CuO(s)
The Copper Cycle Introduction In this experiment, you will carry out a series of reactions starting with copper metal. This will give you practice handling chemical reagents and making observations. It
More informationExperiment 7A ANALYSIS OF BRASS
Experiment 7A ANALYSIS OF BRASS FV 10/21/10 MATERIALS: Spectronic 20 spectrophotometers, 2 cuvettes, brass sample, 7 M HNO 3, 0.100 M CuSO 4, 2 M NH 3, two 50 ml beakers, 100 ml beaker, two 25 ml volumetric
More informationExperiment #7. Determination of an Equilibrium Constant
Experiment #7. Determination of an Equilibrium Constant Introduction It is frequently assumed that reactions go to completion, that all of the reactants are converted into products. Most chemical reactions
More informationaa + bb cc + dd Equation 1
Experiment: The Determination of K eq for FeSCN 2+ Introduction For any reversible chemical reaction at equilibrium, the concentrations of all reactants and products are constant or stable. There is no
More informationExperiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)
: Absorption Spectroscopy of Cobalt(II) OBJECTIVES In successfully completing this lab you will: prepare a stock solution using a volumetric flask; use a UV/Visible spectrometer to measure an absorption
More informationPlant Indicators for Acids and Bases
SCIENCE EXPERIMENTS ON FILE Revised Edition 5.28-1 Plant Indicators for Acids and Bases Carole R. Goshorn Topic Acid/base indicators Time Preparation by teacher 50 minutes; Part B, 45 minutes! Safety Adult
More informationDetermining the Concentration of a Solution: Beer s Law
Determining the Concentration of a Solution: Beer s Law Vernier Spectrometer 1 The primary objective of this experiment is to determine the concentration of an unknown copper (II) sulfate solution. You
More informationTest Method: CPSC-CH-E
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 10901 DARNESTOWN RD GAITHERSBURG, MD 20878 Test Method: CPSC-CH-E1001-08 Standard Operating Procedure
More informationCadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)
Nitrate, MR DOC316.53.01069 Cadmium Reduction Method Method 8171 0.1 to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Scope and application: For water, wastewater
More informationRATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION
Rate Law Determination of Crystal Violet Hydroxylation Revised 5/22/12 RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION Adapted from "Chemistry with Computers" Vernier Software, Portland OR, 1997
More informationScope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.
Nitrogen, Ammonia DOC316.53.01078 USEPA 1 Nessler Method 2 Method 8038 0.02 to 2.50 mg/l NH 3 N Reagent Solution Scope and application: For water, wastewater and seawater. Distillation is required for
More informationnovaa 800 D Atomic Absorption Spectrometer
Technical Data Atomic Absorption Spectrometer Cpt : +27 (0) 21 905 0476 Jhb : +27 (0) 11 794 Dbn : +27 (0) 31 266 2454 1/7 General The is a compact atomic absorption spectrometer with deuterium background
More informationInstrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.
Barium DOC316.53.01315 Turbidimetric Method 1 Method 10251 2 to 100, 20 to 1000, 200 to 10,000 mg/l Ba (spectrophotometers) Powder Pillows 2 to 80, 20 to 800, 200 to 8000 mg/l Ba (colorimeters) Scope and
More informationChemical Health and Safety General Program
Chemical Health and Safety General Program I. Objective To establish minimum requirements for storage, handling and use of chemicals. II. Scope This process applies to employees and operations involved
More informationASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION
ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION Written by Pamela Doolittle, University of Wisconsin- Madison, pssemrad@wisc.edu 2014 In this experiment you will determine the phosphorus (P) concentration
More informationStandard Operating Procedure
Standard Operating Procedure Procedure Radioimmunoassay with I Department Location SOP Prepared By: Section 1: Purpose Radioimmunoassays are used for detecting the concentration of a specific antigen or
More informationMEASUREMENT: PART II
1 MEASUREMENT: PART II Copyright: Department of Chemistry, University of Idaho, Moscow, ID 83844-2343, 2013. INTRODUCTION Read and/or review Section 1.7 and Figure 7.5 in your textbook. The first part
More informationApplication Submitted
Proposal #6 Application Submitted To: Program: Title of Project: Suggested Timeline: Dr. Arlene Courtney Ch 462 Experimental Chemistry Analysis of Lead leached from Fiesta Ware by use of Atomic Absorption
More informationThe Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction
application Note Atomic Absorption Authors David Bass Senior Product Specialist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT 06484 USA The Determination of Toxic, Trace, and
More informationToday s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy
Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Evrim Kilicgedik Product Specialist, Atomic Spectroscopy Agilent Technologies 04.11.2011 2011 The Atomic Spectroscopy
More informationPowder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)
Aluminum DOC316.53.01002 Aluminon Method 1 Method 8012 0.008 to 0.800 mg/l Al 3+ (spectrophotometers) Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters) Scope and application: For water and wastewater.
More informationSAFETY TRAINING LEAFLET 08 ACETYLENE, CALCIUM CARBIDE, LIME SLUDGE AND PURIFYING MATERIALS
SAFETY TRAINING LEAFLET 08 ACETYLENE, CALCIUM CARBIDE, LIME SLUDGE AND PURIFYING MATERIALS Doc 23.08/18 EUROPEAN INDUSTRIAL GASES ASSOCIATION AISBL AVENUE DES ARTS 3-5 B 1210 BRUSSELS Tel: +32 2 217 70
More informationHach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis
Hach Method 1061 Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Company Method 1061 Revision 1. December 015 Organic Carbon in Finished
More informationAtomic Spectroscopy AA/ICP/ICPMS:
Atomic Spectroscopy AA/ICP/ICPMS: A Comparison of Techniques VA AWWA/VWEA Lab Practices Conference July 25, 2016 Dan Davis Shimadzu Scientific Instruments AA/ICP/ICPMS: A Comparison of Techniques Topics
More informationBy Authority Of THE UNITED STATES OF AMERICA Legally Binding Document
By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By the Authority Vested By Part 5 of the United States Code 552(a) and Part 1 of the Code of Regulations 51 the attached document has
More informationExperimental Procedure Lab 402
Experimental Procedure Lab 402 Overview One set of solutions having known molar concentrations of FeNCS 2+ is prepared for a calibration curve, a plot of absorbance versus concentration. A second set of
More informationProduct Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C
Book 5 - Laboratory Policies and Procedures C03-1 2009-06-15 1 Scope 1.1 This method describes a general procedure for the determination of leachable arsenic, selenium, cadmium, antimony, and barium in
More informationExperiment #9. Atomic Emission Spectroscopy
Introduction Experiment #9. Atomic Emission Spectroscopy Spectroscopy is the study of the interaction of light with matter. This interaction can be in the form of the absorption or the emission of electromagnetic
More informationFacilities Management
Policy Number: 700.20 Title: Chemical Fume Hood Policy Implementation Date: 2002 Last Audited: August, 2017 Last Revised: October 23rd, 2017 Facilities Management Introduction The laboratory chemical fume
More informationExercise 2-2. Titration of a Strong Acid EXERCISE OBJECTIVES
Exercise 2-2 Titration of a Strong Acid EXERCISE OBJECTIVES To describe the effect of a ph variation on a chemical indicator; To titrate water containing a strong base solution with a strong acid solution;
More informationMontgomery County Community College Document Number: MET DeKalb Pike Revision Number: 0
Page 1 of 5 Approvals: Preparer: Jason McMillan Reviewer: Dr. Maggie Bryans Date: 08JAN14 Date: 10JAN14 1. Purpose: Operation of Eppendorf Research Plus through to the P-1000. 2. Scope: Applies to the
More informationEnAlgae SOP sheet: Silicate - SFA. 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2.
EnAlgae SOP sheet: Silicate - SFA 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2. 2. Principle Analysis is performed on a segmented flow analyser (SFA), model AutoAnalyser
More information2101 Atomic Spectroscopy
2101 Atomic Spectroscopy Atomic identification Atomic spectroscopy refers to the absorption and emission of ultraviolet to visible light by atoms and monoatomic ions. It is best used to analyze metals.
More informationDetermination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors
Determination of major, minor and trace elements in rice flour using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Application note Food testing Authors John Cauduro Agilent Technologies,
More informationMicro Volume QuEChERS kit
225-37872 Sep. 2018 Small Capacity Pretreatment Kit Micro Volume QuEChERS kit Instruction Manual Read this manual thoroughly before you use the product. Keep this manual for future reference. This page
More informationAtomic Spectra for Atoms and Ions. Light is made up of different wavelengths
Atomic Spectra for Atoms and Ions What will you be doing in lab next week? Recording the line spectra of several different substances in discharge tubes. Recording the line spectra of several ions from
More informationChloride, Nitrate, Nitrite, Phosphate Determination High Performance Liquid Chromatography (HPLC) Procedures
Villanova University Date Oct: 2008 Page 1 of 21 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP C Chloride, Nitrate, Nitrite, Phosphate
More informationElectrochemical Cells
Electrochemical Cells PURPOSE To see how changes in concentration and ph affect the potential in an electrochemical cell, and confirm the Nernst equation. GOALS To examine how standard reduction potentials
More informationTHE IRON(III) THIOCYANATE REACTION SYSTEM
Experiment 7 THE IRON(III) THIOCYANATE REACTION SYSTEM Prepared by Ross S. Nord, Chemistry Department, Eastern Michigan University PURPOSE To investigate a novel reaction system by utilizing a spectrophotometer.
More informationExperiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013
Absorbance Experiment 13 Dilutions and Data Handling in a Spreadsheet rev 1/2013 GOAL: This lab experiment will provide practice in making dilutions using pipets and introduce basic spreadsheet skills
More informationREVIEW OF LAB TECHNIQUES
Experiment 1 REVIEW OF LAB TECHNIQUES Prepared by Masanobu M. Yamauchi and Ross S. Nord, Eastern Michigan University PURPOSE To review density calculations, Beer s Law and the use of electronic balances,
More informationChemical Equilibrium: Finding a Constant, Kc
Chemical Equilibrium: Finding a Constant, Kc Experiment 20 The purpose of this lab is to experimentally determine the equilibrium constant, K c, for the following chemical reaction: Fe 3+ (aq) + SCN (aq)
More informationSafe Use of Pyrophoric/Water Reactive Materials
Introduction Safe Use of Pyrophoric/Water Reactive Materials Pyrophoric and water reactive materials can ignite spontaneously on contact with air, moisture in the air, or water. Improper handling of these
More informationREVIEW OF LAB TECHNIQUES
Experiment 1 REVIEW OF LAB TECHNIQUES Prepared by Masanobu M. Yamauchi and Ross S. Nord, Eastern Michigan University PURPOSE To review density calculations, Beer s Law and the use of electronic balances,
More informationSafety in the Chemistry Laboratory
Safety in the Chemistry Laboratory CHAPTER1 Safety must be everyone s primary concern in the chemistry lab. Understanding and following all safety rules in the organic chemistry lab is critical to your
More informationChemistry 1215 Experiment #11 Spectrophotometric Analysis of an Unknown Brass Sample
Chemistry 1215 Experiment #11 Spectrophotometric Analysis of an Unknown Brass Sample Objective In this experiment you will use spectrophotometric measurements to determine the copper concentration of a
More informationIntroduction to Spectroscopy: Analysis of Copper Ore
Absorbance Introduction to Spectroscopy: Analysis of Copper Ore Introduction The goal of this lab is to determine the unknown concentration of two different copper solution samples, taken from fictitious
More informationHach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water
Hach Method 1041 Spectrophotometric Measurement of Free Chlorine (Cl ) in Finished Drinking Water Hach Company Method 1041 Revision 1. November 015 Spectrophotometric Measurement of Free Cl in Finished
More informationExperiment 3: Atomic Absorption Determination of Lead in Soil Samples Inspired by Mielke, H. American Scientist 1999, 87,
Experiment 3: Atomic Absorption Determination of Lead in Soil Samples Inspired by Mielke, H. American Scientist 1999, 87, 62-73. Introduction Experimental Work on March 2 and March 3-24 (in Olin-Rice 387)
More informationOxygen Demand, Chemical
Oxygen Demand, Chemical DOC316.53.01104 USEPA Reactor Digestion Method Method 10212 250 to 15,000 mg/l COD (UHR) TNTplus 823 Scope and application: For wastewater and process waters; digestion is required.
More informationSECTION 3 PRODUCT COMPOSITION The test kit is composed of QuSTICK Strep A Reagent Stick, Reagent A, Reagent B, Positive Control, and Negative Control.
SECTION 1 PRODUCT AND COMPANY IDENTIFICATION Product Identification: Ref No. 6000, 6000-025, etc. Test kit contains QuSTICK Strep A Reagent Stick, Reagent A, Reagent B, Positive Control, and Negative Control.
More informationExperiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)
Chem 1B Saddleback College Dr. White 1 Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III) Objectives To use spectroscopy to relate the absorbance of a colored solution to its concentration. To
More informationChemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy
Atomic line widths: Narrow line widths reduce the possibility of spectral overlap and thus interferences. The band width at half height is used to indicate width. This is also sometimes called the effective
More informationSTANDARD OPERATING PROCEDURES
PAGE: 1 of 7 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
More information