Continuous Flow Analysis Methods David Davey University of South Australia. 21 February

Size: px
Start display at page:

Download "Continuous Flow Analysis Methods David Davey University of South Australia. 21 February"

Transcription

1 Continuous Flow Analysis Methods David Davey University of South Australia 21 February

2 Flow Analysis Methods 1950s-60s Segmented Flow Methods Clinical Analysis Skeggs 1970s Flow Injection Analysis Chemical Analysis Ruzicka and Hansen, Pungor Tecator Lachat GlobalFIA FIAlab 1980s Sequential Injection Marshall, Ruzicka Lab on a valve 1990s MicroTotal Analysis Systems Lab on a chip 21 February

3 21 February

4 Continuous Flow Analysis Methods David Davey University of South Australia 21 February

5 Reagent Carrier Flow Cell Sample Pump Sample Sensors WASTE 21 February GND

6 Lab on a Chip 21 February

7 Segmented Flow Analysis Automated systems in many older laboratories 3 AIR Pump 5 Detector Mixing coil 1 2 Sample Reagent Basic Layout of SFA instrument 21 February

8 SFA Bolus= Volume of Sample and Reagents which remains largely intact in the flow system Manifold= The collection of tubing and connectors which: transports sample and reagent, creates the bolus, promotes mixing, and delivers the reaction product to a detector 21 February

9 Segmented Flow Analysis AIR 21 February

10 Some SFA Features Samples are pumped into the manifold Air separates sample boluses where reaction occurs Air (and some sample) must be removed before the detector Steady state signals are sought, and dispersion is thus small Manifold tube diameters are usually 2-3 mm. Flow is controlled and turbulent Sample throughput is 2-5 per minute. 21 February

11 Flow Injection Analysis Basic Layout of FIA instrument 1 Pump Sample valve 3 Detector 4 2 Mixing coil 1 4 Carrier or Reagent 1 Reagent 2 21 February

12 Some FIA Features Samples are injected/intercalated into the manifold Air bubbles are not wanted! Steady state signals are not required Manifold tubes are less than 1 mm in diameter Flow is controlled and mostly laminar Sample throughput can be very fast, and is often < 1 sample per minute 21 February

13 Flow Analysis Methods Segmented Flow SFA Tubing 2 mm ID Air segmented Sample 1-2 ml Flow Injection Analysis FIA Tubing < 1 mm ID Sample µL Microconduits Channels 0.5 mm 21 February

14 Flow Analysis Methods Segmented Flow SFA Tubing 2 mm ID Air segmented Sample 1-2 ml Flow Injection Analysis FIA Sequential Injection Analysis SIA Tubing < 1 mm ID Sample µL Microconduits Channels 0.5 mm 21 February

15 Flow Analysis Methods Segmented Flow SFA Flow Injection Analysis FIA Sequential Injection SIA Holding & Mixing Coil Channels < 1 mm ID Multdirectional Use of Syringe Pumps Samples10µL-100 µl 21 February

16 Sequential Injection Analysis,SIA SIA uses the same chemistry as FIA and the reaction manifold is similar. A holding and mixing coil is a key component of the design. Samples and Reagents are loaded using:- a multi-directional valve minipumps or multistep syringes under PC control 21 February

17 SIA setup Peristaltic Pump or Multi-step Piston Wash Reagent Detector Holding Coil Reaction Coil Light source Standard Sample 21 February

18 1. Flow Injection Analysis Automated systems in most laboratories small samples are introduced into a continuous liquid stream for controlled reaction and delivery to a detector FIA requires excellent time and volume control an important analytical tool in determining large numbers of samples rapidly and precisely 21 February

19 FIA Physical Needs Multichannel Pump Sample Injection Valve Flow Through Detector 21 February

20 Flow Injection Analysis Basic Layout of FIA instrument 1 Pump Sample valve 3 Detector 4 2 Mixing coil 1 4 Carrier or Reagent 1 Reagent 2 21 February

21 Flow Injection Analysis continued 1 Carrier The carrier may provide some medium control, but not form a product with the analyte. The carrier could control interfering species through ph control By complexation of an interfering compound 21 February

22 Flow Injection Analysis continued 1 4 Reagent(s) FIA relies on carefully optimised conditions and selective chemistry and/or a selective detector to pick out a key compound in a sample. Colorimetric or Fluorimetric Analysis are dominant methods employed. UV-visible detector measures the absorbance of the product of reaction with the analyte, or Fluorescing product gives out light to a detector Chemiluminescent product gives out light to a detector 21 February

23 Methods characterised by Dispersion FIA & SIA Dispersion is the process which is utilised in FIA (& SIA) to create unique analysis methods Dispersion Factor, Df defines the amoiunt of mixing employed 21 February

24 Dispersion is measured relative to a Steady State result D f =1 A Steady State in FIA is a signal which is the maximum achievable for that manifold. The observed response could be decided by sample volume kinetics mixing characteristics 21 February

25 Low dispersion Flow Injection Analysis sample delivery to electrodes, optodes, atomic spectrometers etc with minimal treatment Medium dispersion sample reaction in multipath manifolds with processes in-stream such as preconcentration, sample-reagent mixing, heating and time delays to achieve a particular sensitivity before a product reaches the detector High dispersion methods using enhanced concentration gradients eg in-stream titrations 21 February

26 Dispersion Coefficient D f = C o /C = H o /H Flow Injection Analysis D f range Low dispersion 1 3 Medium dispersion >3 6 (Ruzicka -10) High dispersion > February

27 Low dispersion Flow Injection Analysis Samples are delivered to a detector with minimal change Applications include:- Electrochemical Methods Atomic absorption Some adjustment may be made to the carrier stream TISAB, buffer Low concentration of the analyte 21 February

28 Examples ph, pf, pca ph 4.6 Pump Sample valve Mixing coil Water TISAB February

29 Medium dispersion Urea in serum ml/min Flow Injection Analysis Sample 30 µl Urease 0.5 Hypochlorite nm Phenol 1.6 Standards 4-25 mm Samples 21 February

30 Other Examples Phosphate in water or soil via vanadomolybdate polyphosphate complexation Sulfur dioxide in beverages forms a dyestuff with para-rosaniline at 550 nm. Ammonia or Chloramine in water via indophenol reaction Nitrite in food via azo-dye formation 21 February

31 High Dispersion Peak Width Measurements FIA provides extraordinary control of mixing processes. Thus in-stream reactions have found peak width as the parameter useful In-stream titration Concentration zones EDTA reagent Ca electrode 21 February

32 Information in a peak The response curve is a repository of information peak height or area = k [analyte] peak width = k log [analyte] 21 February

33 Flow Injection Analysis High dispersion - In-stream titrations EDTA reagent Ca electrode Ca electrode EDTA reagent Analyte Ca February

34 Flow Injection Analysis High dispersion - In-stream titrations t equivalence = k log[ca ++ ] t equivalence 21 February

35 Some Questions to try 21 February

36 Examine the literature to find details for: Phosphate in water or soil via vanadomolybdate polyphosphate complexation under acid conditions Ammonia or Chloramine in water via indophenol reaction Nitrite in food via azo-dye formation with napthylamine 21 February

37 Ques1. Design 2 flow manifolds for N speciation Nitrogen speciation is a key measure in determining water quality. Measure ammonia using the NH 3 electrode Nitrate is first reduced on line to form nitrite which then reacts with napthylamine reaction to form an azo dye before a colorimetric end-point 21 February

38 Design a manifold for SO2 in Wine Ques 2. Sulfur dioxide is a key preservative in beverages. It also has regulated levels under export rules. SO 2 is determined after p-rosaniline reaction under acid conditions, and the response read colorimetrically at 550 nm Total sulfur dioxide requires bound material, e.g. the bisulfite addition complex with aldehyde, be broken up first. 21 February

39 Examine the literature to find details for: Phosphate in water or soil via vanadomolybdate polyphosphate complexation under acid conditions Ammonia or Chloramine in water via indophenol reaction Nitrite in food via azo-dye formation with napthylamine 21 February

40 Selective determination of orthophosphate and total inorganic phosphates in detergents by flow injection photometric method Liu Jing-fu and Jiang Gui-bin, Talanta Vol 52, June 2000, S C HC CC DB 30µL RC D R 1.2 ml/min Pump 620 nm 21 February

41 Questions on orthophosphate and total inorganic phosphate Manifold Assume HC is 100 cm of 1 mm tube. What is its volume (π R 2 L assume π =3)? How long will fresh sample take to travel through the coil? If RC has the same volume as HC, what is the approximate time taken by a new sample to reach the detector D? What do HC, CC and RC represent? Let s allow C and R to flush the detector cell (volume 0.2 ml) twice between introducing sample for reaction. For one sample, how many repeat analyses per minute might be possible? 21 February

42 Questions & Answers on orthophosphate and total inorganic phosphate Manifold Assume HC is 100 cm of 1 mm Internal Diameter tube. What is its volume (π R 2 L assume π =3)? How long will fresh sample take to travel through the coil? Volume = π x(0.05^2)x100 ml = approx 0.8 ml 60 sec x 0.8/1.2 = 40 s If RC has the same volume as HC, what is the approximate time taken by a new sample to reach the detector D? HC 40 s ; RC 20 s (flow rate 2.4) gives Total 60 s 21 February

43 Questions on orthophosphate and total inorganic phosphate Manifold cont d What do HC, CC and RC and DB represent? HC heated coil; CC cooling coil; RC analysis reaction coil; DB debubbler. Detergents are poly phosphates, and reaction is needed to break them down to PO 3-4. Frothing and thus bubblers need to be removed at DB. Let s allow C and R to flush the detector cell (volume 0.2 ml) twice between introducing sample for reaction. For one sample, how many repeat analyses per minute might be possible? The question is best seen as having A sample loaded to the injection valve. C and R flush the detector twice, volume 2 x 0.2 ml. Time 60 x 0.4/2.4 s=10 s Note that the injection valve has easily been rinsed and filled in this time. 60 x 0.03/1.2=1.5 s. So Repeat analyses (peak responses) would be possible 6 times per minute, 360 per hour. In 21 February reality 2012 rates such as per hour are achieved. 43

44 Ammonia in Water NH3 + HOCl NH2 Cl + PhenylOH NH2 Cl + H2O NH2 Phenyl OH 0.3 S A B µL RC1 RC2 RC3 D C 1.0 Pump 660 nm ml/min 21 February

45 Ammonia Manifold Questions Assume 1 mm tubes throughout. RC1 is 10 cm long. RC2 and 3 are 50 cm long. If these are the major volume elements, how long would a sample take to reach the detector? What would you suggest A, B and C are? Let s allow reagents to flush the whole manifold between introducing fresh sample for reaction. For one sample type, how many repeat analyses per hour might be possible? This is called the through put. Sample carryover implies reaction of one sample bolus with another. Suggest a maximum repeat sample injection for this system. 21 February

46 Ammonia Manifold Questions Assume 1 mm tubes throughout. RC1 is 10 cm long. RC2 and 3 are 50 cm long. If these are the major volume elements, how long would a sample take to reach the detector? From previous question, volumes are = 0.88 ml Each manifold component has its own flow rate so times are 4.8, 12 and 8 s total time = 25 s What would you suggest A, B and C are? A good answer would be: A ph control; B HOCl; C phenol Let s allow reagents to flush the whole manifold between introducing fresh sample for reaction. For one sample type, how many repeat analyses per hour might be possible? This is called the through put. Since 25 seconds are needed per sample, throughput= 60 x 60/25 =144 h February

47 Ammonia Manifold Questions Sample carryover implies reaction of one sample bolus with another. Suggest a maximum repeat sample injection for this system. I interpret this to mean how many 100 µl samples could squeeze into the manifold. It is roughly 1 in RC1; 4 in both RC2 and RC3. 4 in RC3 at 3 ml/min would imply a sample every 8/4 s or 1800 h -1! However if the volume D were 0.2 ml, 2 samples would end up in the detector cell at once! A mixed signal (carryover would result). The detector volume (and response time) is the ultimate limit on the analysis speed. 21 February

Automating Wet Chemical Analysis

Automating Wet Chemical Analysis Automating Wet Chemical Analysis Who Should Automate? Anyone looking to improve laboratory efficiency Environmental Agricultural Food and beverage Pharmaceutical Mining Etc. Advantages of Automating Save

More information

Automating Wet Chemical Analysis. William Lipps Market Specialist for Water Analyzer Products December 11, 2008

Automating Wet Chemical Analysis. William Lipps Market Specialist for Water Analyzer Products December 11, 2008 Automating Wet Chemical Analysis William Lipps Market Specialist for Water Analyzer Products December 11, 2008 The advantages of automating wet chemical analysis. Save time Decrease cost Improve quality

More information

Method for Nitrite determination on Low-Range Samples

Method for Nitrite determination on Low-Range Samples Method for Nitrite determination on Low-Range Samples 1. PRINCIPLE This method is designed for the determination of nitrite in soil, water and other forms of samples. The sample is prepared offline if

More information

By Mohsin Al-Saleh Clinical Biochemistry Department (SQUH)

By Mohsin Al-Saleh Clinical Biochemistry Department (SQUH) Laboratory Automation By Mohsin Al-Saleh Clinical Biochemistry Department (SQUH) Objectives: Define automation and lab automation Advantages of laboratory automation History of laboratory automation Fluidic

More information

Automation of Wet Chemical Analysis Methods

Automation of Wet Chemical Analysis Methods Automation of Wet Chemical Analysis Methods Application Note 21150205 Keywords CNSolution FS 3100 DA 3500 Discrete Analysis Ion Chromatography FIA Flow Injection Analysis Flow Solution Segmented Flow Analysis

More information

TOUR OF SOIL CHEMISTRY LABS (09/22/07)

TOUR OF SOIL CHEMISTRY LABS (09/22/07) Department of Crop and Soil Sciences Commemorating Our Centennial, 1907 2007 TOUR OF SOIL CHEMISTRY LABS (09/22/07) WELCOME TO THE TOUR OF SOIL CHEMISTRY LABS Instrument: Lachat QuikChem FIA + 8000 Analyzer

More information

DBP Control: Chloramine Chemistry. Chris Griffin Hach Company

DBP Control: Chloramine Chemistry. Chris Griffin Hach Company DBP Control: Chloramine Chemistry Chris Griffin Hach Company 1 BEFORE WE BEGIN 2 Who currently Uses Chlorine only? Before we begin. Uses Chloramination at their water plant or in distribution? Uses Chloramination

More information

Automated Photometry. Presented by Scott Tucker. Instruments

Automated Photometry. Presented by Scott Tucker. Instruments Application Breakthroughs in Automated Photometry Presented by Scott Tucker Application Chemist, Lachat Instruments Introduction Advantages of automated photometry FIAversus other techniques Lachat Application

More information

Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow

Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow 1. Application Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow Parameters and Codes: Nitrogen, ammonia, dissolved, I-2522-90 (mg/l as N): 00608 Nitrogen, ammonia, total-in-bottom-material,

More information

DI 3500 Discrete Aoalvzer

DI 3500 Discrete Aoalvzer DI 3500 Discrete Aoalvzer Automated Chemistry and Ion Analysis ':t O+Analytical:!, A World ofsolutions. - Discrete Analysis Efficient Microscale Chemistry Increasingly laboratories are being challenged

More information

If you like us, please share us on social media. The latest UCD Hyperlibrary newsletter is now complete, check it out.

If you like us, please share us on social media. The latest UCD Hyperlibrary newsletter is now complete, check it out. Sign In Forgot Password Register username username password password Sign In If you like us, please share us on social media. The latest UCD Hyperlibrary newsletter is now complete, check it out. ChemWiki

More information

RECENT ADVANCES AND FUTURE TRENDS IN FLOW INJECTION AND THEIR APPLICATION TO CHEMICAL OCEANOGRAPHY

RECENT ADVANCES AND FUTURE TRENDS IN FLOW INJECTION AND THEIR APPLICATION TO CHEMICAL OCEANOGRAPHY RECENT ADVANCES AND FUTURE TRENDS IN FLOW INJECTION AND THEIR APPLICATION TO CHEMICAL OCEANOGRAPHY António O. S. S. Rangel 1, Raquel B. R. Mesquita 1, Collaborative on Oceanographic Chemical Analysis (COCA)

More information

Reverse Flow Injection Analysis for Determination of Manganese(II) in Natural Water. Jintana Klamtet

Reverse Flow Injection Analysis for Determination of Manganese(II) in Natural Water. Jintana Klamtet NU Science Journal 2006; 2(2): 165 173 Reverse Flow Injection Analysis for Determination of Manganese(II) in Natural Water Jintana Klamtet Department of Chemistry, Faculty of Science, Naresuan University,

More information

Determination of the Scavenging Capacity Against Reactive Nitrogen Species by Automatic Flow Injection-Based Methodologies

Determination of the Scavenging Capacity Against Reactive Nitrogen Species by Automatic Flow Injection-Based Methodologies Chapter 8 Determination of the Scavenging Capacity Against Reactive Nitrogen Species by Automatic Flow Injection-Based Methodologies Marcela A. Segundo, Luís M. Magalhães, Joana P.N. Ribeiro, Marlene Lúcio,

More information

Analytical Chemistry-I

Analytical Chemistry-I MODEL ANSWER M. Sc. (First Semester) 2013, CHEMISTRY (AS-2142) Analytical Chemistry-I SECTION: A (2 marks) 1. (i) UV-visible spectrophotometer (ii) Atomic absorption spectrometer (iii) Nuclear magnetic

More information

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds Atmospheric Analysis Gases Sampling and analysis of gaseous compounds Introduction - External environment (ambient air) ; global warming, acid rain, introduction of pollutants, etc - Internal environment

More information

Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA)

Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA) Methodology Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA) (Cartridge Part #A002603) 1.0 Scope and Application 1.1 This method is used to determine the concentration of nitrate

More information

in Lab-on-Valve System

in Lab-on-Valve System J. Flow Injection Anal., Vol. 19, No.1 (2002) PSI: Optimization of Reagent ased Chloride Assay in Lab-on-Valve System.. Syouhei ish hi ha ma,^^^'* Louis S~arn~avia,~ and Jaromir ~uzicka' ~e~artrnent of

More information

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Figure 1. Structures for Vitamin B2 and Vitamin B1. CH 461 & CH 461H F 15 Name Experiment 2C Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART C. HIGH PERFORMANCE LIQUID CHROMATOGRAPHY The purpose of this experiment

More information

Chapter 3 Electrochemical methods of Analysis-Potentiometry

Chapter 3 Electrochemical methods of Analysis-Potentiometry Chapter 3 Electrochemical methods of Analysis-Potentiometry Electroanalytical chemistry Contents Introduction Galvanic and electrolytic cells Salt bridge Electrode potential and cell potential Indicator

More information

Speed and flexibility

Speed and flexibility Speed and flexibility Automate your environmental, water, detergent, food and beverage analysis Thermo Scientific Gallery and Gallery Plus discrete industrial analyzers Improve productivity, reproducible

More information

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Figure 1. Structures for Vitamin B2 and Vitamin B1. CH 461 & CH 461H F 18 Name Experiment 2C Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART C. HIGH PERFORMANCE LIQUID CHROMATOGRAPHY The purpose of this experiment

More information

Reagents for Chemical Test Kits

Reagents for Chemical Test Kits for Chemical Test Kits Acidity, methyl orange method ; 0 to 500 mg/l HI 3820-100 Reagent kit for 110 tests (CaCO 3 ) Acidity, olive oil, titration with hydroxide 0.00 to 1.00% HI 3897 Reagent kit for 6

More information

A New On-line Cyanide Analyzer for Measurement of Cyanide in Hydrometallurgical Processing of Precious Metal Ores

A New On-line Cyanide Analyzer for Measurement of Cyanide in Hydrometallurgical Processing of Precious Metal Ores Application Note 37890312 A New On-line Cyanide Analyzer for Measurement of Cyanide in Hydrometallurgical Processing of Precious Metal Ores Keywords CNSolution 9310 On-line Cyanide Analyzer Cyanidation

More information

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV)

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV) METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV) 1.0 SCOPE AND APPLICATION 1.1 Method 9012 is used to determine the concentration of inorganic cyanide in an aqueous waste or leachate.

More information

METHOD 9012B TOTAL AND AMENABLE CYANIDE (AUTOMATED COLORIMETRIC, WITH OFF-LINE DISTILLATION)

METHOD 9012B TOTAL AND AMENABLE CYANIDE (AUTOMATED COLORIMETRIC, WITH OFF-LINE DISTILLATION) METHOD 9012B TOTAL AND AMENABLE CYANIDE (AUTOMATED COLORIMETRIC, WITH OFF-LINE DISTILLATION) 1.0 SCOPE AND APPLICATION 1.1 This method is used to determine the concentration of inorganic cyanide (CAS Registry

More information

SECTION D.2 AMMONIA NITROGEN

SECTION D.2 AMMONIA NITROGEN SECTION D.2 AMMONIA NITROGEN CEDR Method Code: NH4F L01 a) Scope and Application i) This method describes the determination of low-level ammonia nitrogen concentrations in filtered samples taken from fresh

More information

Cyanide Analysis with the CNSolution 3100

Cyanide Analysis with the CNSolution 3100 Cyanide Analysis with the CNSolution 3100 The CNSolution is your solution to cyanide analysis Gas diffusion amperometry methods: Save time Save money Are more accurate Have fewer interferences Are green

More information

SKALAR. your partner in chemistry automation. The San ++ continuous flow analyzer

SKALAR. your partner in chemistry automation. The San ++ continuous flow analyzer SKALAR your partner in chemistry automation The San ++ continuous flow analyzer Applications Over the years Skalar has been working and developing applications for a variety of industries. Our comprehensive

More information

A LAZAR LABORATORY HANDBOOK MEASUREMENT OF NITRIC OXIDE IN BIOLOGICAL SAMPLES

A LAZAR LABORATORY HANDBOOK MEASUREMENT OF NITRIC OXIDE IN BIOLOGICAL SAMPLES A LAZAR LABORATORY HANDBOOK MEASUREMENT OF NITRIC OXIDE IN BIOLOGICAL SAMPLES Copyright Lazar Research Laboratories, Inc. 2004-1 - QUESTIONS? Call us at 1-800-824-2066 (1-323-931-1204 international), fax

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009. Translated English of Chinese Standard: GB5009.17-2014 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF GB THE PEOPLE S REPUBLIC OF CHINA National Food Safety Standard-Determination

More information

HPLC Praktikum Skript

HPLC Praktikum Skript HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques,

More information

FIA Coupled to Capillary Electrophoresis - A Tool for Process Monitoring?

FIA Coupled to Capillary Electrophoresis - A Tool for Process Monitoring? (Review) J. Flow Injection Anal., Vol. 17, No. 1 (2000) FIA Coupled to Capillary Electrophoresis - A Tool for Process Monitoring? Bo Karlberg*, Petr Kuban Department of Analytical Chemistry, Stockholm

More information

CHLORAMINATION AND CHLORAMINE ANALYSIS SW AWWA 2014

CHLORAMINATION AND CHLORAMINE ANALYSIS SW AWWA 2014 CHLORAMINATION AND CHLORAMINE ANALYSIS SW AWWA 2014 COMBINED CHLORINE - CHLORAMINATION Chlorine (HOCl and OCl - ) reacts with ammonia to form chloramines, commonly referred to as combined chlorine The

More information

Flow Injection Analysis for the Determination of Ammonia in Exhaust Gas of Thermal Power Plant

Flow Injection Analysis for the Determination of Ammonia in Exhaust Gas of Thermal Power Plant J. Flow Injection Anal., Vol. 17, No. 1 (2000) Flow Injection Analysis for the Determination of Ammonia in Exhaust Gas of Thermal Power Plant Tomonori subo oil*, Yoshio ~iranol, Mitsuko 0shima2 and Shoji

More information

O Sodium Bicarbonate, NaHCO 3 Ammonium Molybdate Tetrahydrate, (NH 4. O 4 Sodium Chloride, NaCl Ascorbic Acid, C 6 H 8 O 6

O Sodium Bicarbonate, NaHCO 3 Ammonium Molybdate Tetrahydrate, (NH 4. O 4 Sodium Chloride, NaCl Ascorbic Acid, C 6 H 8 O 6 Scope Summary Interferences Performance Specifications This method is used for the determination of silica in estuarine and coastal waters (seawater) according to USEPA Method 366.0. This method can also

More information

Cyanide, colorimetric, pyridine-pyrazolone

Cyanide, colorimetric, pyridine-pyrazolone Cyanide, colorimetric, pyridine-pyrazolone Parameters and Codes: Cyanide, dissolved, I-1300-85 mg/l as CN): 00723 Cyanide, total, I-3300-85 (mgll as CN): 00720 Cyanide, total-in-bottom-material, dry wt,

More information

DETERMINATION OF NITRATE AND NITRITE BY ZONE FLUIDICS

DETERMINATION OF NITRATE AND NITRITE BY ZONE FLUIDICS DETERMINATION OF NITRATE AND NITRITE BY ZONE FLUIDICS SPECTROPHOTOMETRIC DETECTION WRITTEN BY GRAHAM MARSHALL Revision: - Date: April 19, 2014 met_nox_-.docx Page 1 DETERMINATION OF NITRATE AND NITRITE

More information

Chemical Oxidation Oxidizing agents

Chemical Oxidation Oxidizing agents Chemical Oxidation CENG 4710 Environmental Control Chemical oxidation is used to detoxify waste by adding an oxidizing agent to chemically transform waste compounds. It is capable of destroying a wide

More information

Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006

Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006 Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006 Introduction: Capillary electrophoresis (CE) is a relatively new, but rapidly growing separation

More information

Analysis of Free Ammonia in Chloramination Applications Using Lab Method and the APA6000

Analysis of Free Ammonia in Chloramination Applications Using Lab Method and the APA6000 Analysis of Free Ammonia in Chloramination Applications Using Lab Method 10200 and the APA6000 Chloramination Chemistry Knowledge of chloramination chemistry is required to understand the rationale behind

More information

EnAlgae SOP sheet: Silicate - SFA. 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2.

EnAlgae SOP sheet: Silicate - SFA. 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2. EnAlgae SOP sheet: Silicate - SFA 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2. 2. Principle Analysis is performed on a segmented flow analyser (SFA), model AutoAnalyser

More information

Sampling for Gases and Vapors

Sampling for Gases and Vapors Sampling for Gases and Vapors EOH 466B Evaluating the Occupational Environment Spring 2008 Gases and Vapors Gases: don't exist in liquid state at STP e.g.: nitrous oxides, ozone, carbon monoxide Vapors:

More information

Name: Date: AP Chemistry. Titrations - Volumetric Analysis. Steps for Solving Titration Problems

Name: Date: AP Chemistry. Titrations - Volumetric Analysis. Steps for Solving Titration Problems Name: Date: AP Chemistry Titrations - Volumetric Analysis Term Volumetric analysis Burette Pipette titrate titre aliquot end point equivalence point indicator primary standard standardisation secondary

More information

Paper 4.3. Introduction

Paper 4.3. Introduction Paper 4.3 Removal of free and combined chlorine at GAC surfaces and impact on pool water quality Bertram Skibinski, PhD student, Susanne Müller, PhD student and Wolfgang Uhl, Chairholder, Water Supply

More information

Astoria-Pacific proudly offers our rapid-t software for the ChemWell-T an easy-to-use, compact automated chemistry analyzer:

Astoria-Pacific proudly offers our rapid-t software for the ChemWell-T an easy-to-use, compact automated chemistry analyzer: ASTORIA-PACIFIC rapid-t with Astoria-Pacific proudly offers our rapid-t software for the ChemWell-T an easy-to-use, compact automated chemistry analyzer: Durable. Accurate. Precise. Economical Open software

More information

Factors affecting zone penetration

Factors affecting zone penetration Chapter 3 Factors affecting zone penetration 3.1 INTRODUCTION There is a tendency by many users of flow-based techniques to apply empirical techniques in the choice of operating conditions and design of

More information

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II) : Absorption Spectroscopy of Cobalt(II) OBJECTIVES In successfully completing this lab you will: prepare a stock solution using a volumetric flask; use a UV/Visible spectrometer to measure an absorption

More information

Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1

Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1 Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson

More information

Orderly Arranged Fluorescence Dyes as a Highly Efficient Chemiluminescence Resonance Energy Transfer Probe for Peroxynitrite

Orderly Arranged Fluorescence Dyes as a Highly Efficient Chemiluminescence Resonance Energy Transfer Probe for Peroxynitrite Supporting Information: Orderly Arranged Fluorescence Dyes as a Highly Efficient Chemiluminescence Resonance Energy Transfer Probe for Peroxynitrite Zhihua Wang, Xu Teng and Chao Lu* State Key Laboratory

More information

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed Date 2017.

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed Date 2017. Final Report Title Characterisation of Job No 2016/11/12-34 AS No. 1234A Client Contact Sample Author report Signed Date 2017 Easy Reach Report 2017 v2.docx 1 of 33 Contents 1. Study Summary Page 3 2.

More information

Automation of Routine Microplate Pipetting and Dispensing

Automation of Routine Microplate Pipetting and Dispensing Automation of Routine Microplate Pipetting and Dispensing Quantitative assays of all sorts are common to many applications in basic science and clinical research. Most of these assays require the movement,

More information

Total Kjeldahl Nitrogen by Gas Diffusion and Flow Injection Analysis (FIA)

Total Kjeldahl Nitrogen by Gas Diffusion and Flow Injection Analysis (FIA) Methodology Total Kjeldahl Nitrogen by Gas Diffusion and Flow Injection Analysis (FIA) (Cartridge Part #A002380) 1.0 Scope and Application 1.1 This method is used for the determination of Total Kjeldahl

More information

Understanding Chlorine Measurement. Rebecca Luedee Environmental Sales

Understanding Chlorine Measurement. Rebecca Luedee Environmental Sales Understanding Chlorine Measurement Rebecca Luedee Environmental Sales Introduction to Chlorine Reaction with Water Forms hydrochloric (HCl) and hypochlorous (HOCl) acids: Cl 2 + H 2 O HOCl + HCl HOCl dissociates

More information

Single Method On-line Analyzers. ADI 2016 Titrolyzer ADI 2018 Ion-Analyzer ADI 2019 Process Colorimeter

Single Method On-line Analyzers. ADI 2016 Titrolyzer ADI 2018 Ion-Analyzer ADI 2019 Process Colorimeter Single Method On-line Analyzers ADI 2016 Titrolyzer ADI 2018 Ion-Analyzer ADI 2019 Process Colorimeter Automatic and dependable on-line ion analysis 02 The single method process analyzers from Metrohm

More information

Water Hardness and Softening (Bring a water sample from home) Minneapolis Community and Technical College Principles of Chemistry II, C1152 v.2.

Water Hardness and Softening (Bring a water sample from home) Minneapolis Community and Technical College Principles of Chemistry II, C1152 v.2. Water Hardness and Softening (Bring a water sample from home) Minneapolis Community and Technical College Principles of Chemistry II, C1152 v.2.16 I. Introduction Hard Water and Water Softening Water that

More information

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System Application Note Authors Kevin McCann, Sameer Nene, Doug McIntyre, Edmond Neo, Dennis Nagtalon, and

More information

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation Phosphorus, Total DOC316.53.01121 USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method 8190 0.06 to 3.50 mg/l PO 3 4 (0.02 to 1.10 mg/l P) Test N Tube Vials Scope and application: For water,

More information

Principles of micro Sequential Injection Analysis in the Lab-on-Valve format and its Introduction into a Teaching Laboratory

Principles of micro Sequential Injection Analysis in the Lab-on-Valve format and its Introduction into a Teaching Laboratory J. Flow Injection Anal., Vol. 24, No. 2 (2007) 103 108 Principles of micro Sequential Injection Analysis in the Lab-on-Valve format and its Introduction into a Teaching Laboratory Michael S. Decuir, Hannah

More information

Figure Small ByT-FAS raw data plots of a quinine tracer injected into the sample carrier stream.

Figure Small ByT-FAS raw data plots of a quinine tracer injected into the sample carrier stream. Figure 3.23. Small ByT-FAS raw data plots of a quinine tracer injected into the sample carrier stream. Quinine sample was injected into the sample side of the system with different sample injection loop

More information

METHOD STATEMENT. Determinand: Ammonia, Chloride, Nitrate, Nitrite, Orthophosphate, Sulphate & TON. Matrix: Leachates, effluents and wastewaters

METHOD STATEMENT. Determinand: Ammonia, Chloride, Nitrate, Nitrite, Orthophosphate, Sulphate & TON. Matrix: Leachates, effluents and wastewaters : Ammonia,, Nitrate,, Orthophosphate, & TON Matrix: Leachates, effluents and wastewaters Principle of Method: Certain analytes, when reacted with specific reagents, will form coloured complexes. The intensity

More information

1225 Lab # 11 Fluoride Speciation and Analysis by Flow Injection using Ion-Selective Electrode: Measurement of Total Fluoride in Water

1225 Lab # 11 Fluoride Speciation and Analysis by Flow Injection using Ion-Selective Electrode: Measurement of Total Fluoride in Water 1225 Lab # 11 Fluoride Speciation and Analysis by Flow Injection using Ion-Selective Electrode: Measurement of Total Fluoride in Water Introduction Fluoride is recognized world- wide to help build stronger

More information

Deep Desulfurization of Diesel using a Single-Phase Micro-reactor

Deep Desulfurization of Diesel using a Single-Phase Micro-reactor Excerpt from the Proceedings of the COMSOL Conference 2009 Boston Deep Desulfurization of Diesel using a Single-Phase Micro-reactor Jake Jones, 1 Alex Yokochi, 1 and Goran Jovanovic *1 1 School of Chemical,

More information

Exercise 6: Determination of Hardness of Water

Exercise 6: Determination of Hardness of Water Fundamentals of Analytical Chemistry, CHC014011L Exercise 6: Determination of Hardness of Water Introduction: Hardness in water is generally caused by the presence of dissolved calcium and magnesium carbonates

More information

Exercise 4-4. Titration of a Buffer Solution EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Buffer solutions

Exercise 4-4. Titration of a Buffer Solution EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Buffer solutions Exercise 4-4 Titration of a Buffer Solution EXERCISE OBJECTIVE Titrate a buffer solution, plot a graph using the titration data, and analyze the titration curve. DISCUSSION OUTLINE The Discussion of this

More information

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography Yan Liu, Zhongqing Lu, and Chris Pohl, Thermo Fisher Scientific, Sunnyvale, CA USA Overview Purpose: In this work, new high-pressure

More information

Fig. Electrochemical Cell/ Potentiometric Titration

Fig. Electrochemical Cell/ Potentiometric Titration Fig. Electrochemical Cell/ Potentiometric Titration The accurate, precise and effective potentiometric measurements can be made with the help of the following two types of electrodes namely : REFERENCE

More information

Investigation of Petasis and Ugi Reactions in Series in an Automated Microreactor System

Investigation of Petasis and Ugi Reactions in Series in an Automated Microreactor System Electronic Supplementary Material (ESI) for RSC Advances. This journal is The Royal Society of Chemistry 2014 Supplementary Information Investigation of Petasis and Ugi Reactions in Series in an Automated

More information

Medicine Concentration in the AutoStart Burette following an Automatic Flow Restart Event

Medicine Concentration in the AutoStart Burette following an Automatic Flow Restart Event www.analyticamedical.com ANALYTICA LTD ABN 12 006 464 866 Medicine Concentration in the AutoStart Burette following an Automatic Flow Restart Event Geoff Daly, Operations Manager, Analytica Ltd. 29 October

More information

Flow Solution FS 3700 Automated Chemistry Analyzer

Flow Solution FS 3700 Automated Chemistry Analyzer Method Abstract Total Cyanide by ISO 14403 Document #42600116 Flow Solution FS 3700 Automated Chemistry Analyzer Total Cyanide by In-Line Ultraviolet Digestion, Gas Diffusion, and Photometric Detection,

More information

Nutrient Analysis, Nitrate, Nitrite, Silicate, Phosphate and Ammonium

Nutrient Analysis, Nitrate, Nitrite, Silicate, Phosphate and Ammonium Nutrient Analysis, Nitrate, Nitrite, Silicate, Phosphate and Ammonium SUMMARY: The phytoplankton macro nutrients nitrate, nitrite, silicate, phosphate and ammonium are analyzed in seawater using a colorimetric

More information

Total Kjeldahl Nitrogen (TKN), USEPA by Flow Injection Analysis (FIA)

Total Kjeldahl Nitrogen (TKN), USEPA by Flow Injection Analysis (FIA) Methodology Total Kjeldahl Nitrogen (TKN), USEPA by Flow Injection Analysis (FIA) (Cartridge Part #A002054) 1.0 Scope and Application 1.1 This method is used for the determination of Total Kjeldahl Nitrogen

More information

Spectrophotometric Determination of Anionic Surfactants in River Water with Cationic Azo Dye by Solvent Extraction- Flow Injection Analysis

Spectrophotometric Determination of Anionic Surfactants in River Water with Cationic Azo Dye by Solvent Extraction- Flow Injection Analysis ANALYTICAL SCIENCES JUNE 1987, VOL. 3 265 Spectrophotometric Determination of Anionic Surfactants in River Water with Cationic Azo Dye by Solvent Extraction- Flow Injection Analysis Shoji MOTOMIZU, Yoshito

More information

SYNTHESIS OF AN AZO DYE revisited (1 or 2 credits)

SYNTHESIS OF AN AZO DYE revisited (1 or 2 credits) SYNTHESIS OF AN AZO DYE revisited (1 or 2 credits) This lab you can revisit the fist experiment of this quarter and synthesize more azo dyes of your choice. The old procedure is given below followed by

More information

Nitrogen in All Its Forms. Assoc. Prof. Kozet YAPSAKLI

Nitrogen in All Its Forms. Assoc. Prof. Kozet YAPSAKLI Nitrogen in All Its Forms Assoc. Prof. Kozet YAPSAKLI Nitrogen is in the Nonmetals Group OXIDATION STATES OF NITROGEN N has 5 electrons in valence shell a7 oxidation states from 3 to +5 Increasing oxidation

More information

Shorten your Time-to-Result

Shorten your Time-to-Result Shorten your Time-to-Result Effective quality and process control requires the shortest time-to-result possible. This is the time needed from sampling to the final quantitative result. Any advantage results

More information

Disinfection. What is disinfection? Why to disinfect?

Disinfection. What is disinfection? Why to disinfect? Disinfection What is disinfection? Why to disinfect? Disinfection is a process which inactivate or remove the pathogenic microorganisms in system. Treated (including disinfection) water should ensure the

More information

N-NITROSODIETHANOLAMINE (NDELA) IN COSMETICS

N-NITROSODIETHANOLAMINE (NDELA) IN COSMETICS As of: February, 2011 N-NITROSODIETHANOLAMINE (NDELA) IN COSMETICS Cosmetics, particularly mascara and skin care products, partly show very high concentration of the carciogenic compound N-Nitrosodiethanolamine

More information

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples A P P L I C A T I O N N ot e Atomic Absorption Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples Introduction The Flow Injection Mercury System (FIMS) is a dedicated system

More information

Reprint. Project Summary. Side by Side Evaluation of ChemScan UV-4100 and Hach (Dr. Lange) Optiquant at Littleton-Englewood Colorado

Reprint. Project Summary. Side by Side Evaluation of ChemScan UV-4100 and Hach (Dr. Lange) Optiquant at Littleton-Englewood Colorado ChemScan Process Analyzer Reprint Project Summary Side by Side Evaluation of ChemScan UV-4100 and Hach (Dr. Lange) Optiquant at Littleton-Englewood Colorado Published by: Applied Spectrometry Associates,

More information

SKALAR. your partner in chemistry automation. The San ++ continuous flow analyzer

SKALAR. your partner in chemistry automation. The San ++ continuous flow analyzer SKALAR your partner in chemistry automation The San ++ continuous flow analyzer Applications Over the years Skalar has been working and developing applications for a variety of industries. Our comprehensive

More information

Disinfection. Disinfection is used to treat both domestic water and wastewater.

Disinfection. Disinfection is used to treat both domestic water and wastewater. Disinfection Disinfection is the selective destruction of disease causing organisms (viruses, bacteria, protozoans). It destroys most recognized pathogenic microorganisms, but not necessarily all microbial

More information

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows Chlorine Free, Indophenol, 10241 Indophenol 1 1 Patent pending. DOC316.53.01256 Method 10241 (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows Scope and Application: For determining residual free chlorine levels

More information

Development of an Automatic Sampling Module to Monitor Concentration of Liquid-Borne Nanoparticles

Development of an Automatic Sampling Module to Monitor Concentration of Liquid-Borne Nanoparticles Sensors and Materials, Vol. 28, No. 6 (2016) 625 630 MYU Tokyo 625 S & M 1217 Development of an Automatic Sampling Module to Monitor Concentration of Liquid-Borne Nanoparticles Yu-Shan Yeh, * Kaoru Kondo,

More information

HPLC. GRATE Chromatography Lab Course. Dr. Johannes Ranke. September 2003

HPLC. GRATE Chromatography Lab Course. Dr. Johannes Ranke. September 2003 HPLC GRATE Chromatography Lab Course Dr. Johannes Ranke Organisation The groups Start at 9:00 am End at 18:00 pm at the latest Friday, 19th we will finish at 2:00 pm Thursday, 11th: Lecture at 08:15 am

More information

Suppression in anion chromatography. More sensitive analysis of anions and organic acids

Suppression in anion chromatography. More sensitive analysis of anions and organic acids Suppression in anion chromatography More sensitive analysis of anions and organic acids What is suppression? 02 Suppression plays a key role in the analysis of anions and organic acids using ion-exchange

More information

Automation of the radiochemical procedures for the sequential separation of radionuclides

Automation of the radiochemical procedures for the sequential separation of radionuclides LSC2017 - An International Conference on Advances in Liquid Scintillation Spectrometry, Copenhagen Denmark, 1 5 May 2017 Automation of the radiochemical procedures for the sequential separation of radionuclides

More information

Lab #12: Determination of a Chemical Equilibrium Constant

Lab #12: Determination of a Chemical Equilibrium Constant Lab #12: Determination of a Chemical Equilibrium Constant Objectives: 1. Determine the equilibrium constant of the formation of the thiocyanatoiron (III) ions. 2. Understand the application of using a

More information

Generation and absorption of CO2 gas

Generation and absorption of CO2 gas Generation and absorption of CO2 gas CO2 is generated by dissolving carbonates in in hydrochloric acid according to the following equation: CaCO3(s) + 2 HCl(l) = CaCl2(aq) + CO2(g) + H2O(l) One convenient

More information

2011_ST20_Taylor.pdf. SSPT Document not peer-reviewed by CORESTA

2011_ST20_Taylor.pdf. SSPT Document not peer-reviewed by CORESTA An Ion Chromatographic Method for the Quantitative Determination of Hydrogen Cyanide in Cigarette Smoke using Pulsed Amperometric Detection Dr W Guthery and Dr M J Taylor Filtrona Technology Centre CORESTA

More information

Examples of Strong Acids: Strong Acid Formula Common Source Hydrochloric Acid HCl Stomach Acid

Examples of Strong Acids: Strong Acid Formula Common Source Hydrochloric Acid HCl Stomach Acid ACIDS AND BASES: PH AND BUFFERS PURPOSE: To determine the ph of common acids and bases using a ph meter, ph paper, and red cabbage indicator. To test the effect of adding an acid or base to a buffer solution.

More information

CHEMISTRY DEPARTMENT

CHEMISTRY DEPARTMENT POINT LEPREAU GENERATING STATION GADOLINIUM ARSENAZO III Issued By: Date: APPROVED BY: of 6 REVISION RECORD AUTHOR REV DESCRIPTION DATE 0 Initial Issue W. Mawhinney Not Tracked Oct./9 W. T. Underhill 2

More information

Alcohol dehydrogenase Assay Kit

Alcohol dehydrogenase Assay Kit Alcohol dehydrogenase Assay Kit Catalog Number KA3785 100 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 General

More information

J. Flow Injection Anal., Vol. 13, No. 2 (1996)

J. Flow Injection Anal., Vol. 13, No. 2 (1996) J. Flow Injection Anal., Vol. 13, No. 2 (1996) Influence of the light-path of the flow cell on the spectrophotometric measurements in an FIA assembly. Spectrophotometric determination of several sulfonamides

More information

Particle Charge Titration Zeta Potential Size Distribution

Particle Charge Titration Zeta Potential Size Distribution Particle Charge Titration Zeta Potential Size Distribution Stabino NANO-flex ZETA-check (without Titration) Stabino ZETA-check Rapid Particle Charge Titrations and Size Distribution In many applications,

More information

ASTORIA PACIFIC. Automated Wet-Chemistry Analyzers and Diagnostic Reagents. rapid-t

ASTORIA PACIFIC. Automated Wet-Chemistry Analyzers and Diagnostic Reagents. rapid-t ASTORIA PACIFIC Automated Wet-Chemistry Analyzers and Diagnostic Reagents rapid-t rapid-t System ASTORIA-PACIFIC PROUDLY OFFERS OUR rapid-t SYSTEM: Durable. Accurate. Precise. Economical. Open software

More information

SMARTCHEM 300 STATE-OF-THE-ART DIRECT READ DISCRETE ANALYZER FOR AUTOMATED CHEMISTRIES

SMARTCHEM 300 STATE-OF-THE-ART DIRECT READ DISCRETE ANALYZER FOR AUTOMATED CHEMISTRIES SMARTCHEM 300 STATE-OF-THE-ART DIRECT READ DISCRETE ANALYZER FOR AUTOMATED CHEMISTRIES SMARTCHEM 300 FULLY AUTOMATED, DIRECT READ, DISCRETE ANALYZER TO MAXIMIZE PRODUCTIVITY FOR ROUTINE AND SPECIAL CHEMISTRIES,

More information

2035 Process Analyzer. Customized online monitoring of industrial processes and wastewater

2035 Process Analyzer. Customized online monitoring of industrial processes and wastewater 2035 Process Analyzer Customized online monitoring of industrial processes and wastewater The 2035 Process Analyzer at a glance 02 The 2035 Process Analyzer is the latest of Metrohm Process Analytics solutions

More information

PRELIMINARY ACTIVITY FOR

PRELIMINARY ACTIVITY FOR PRELIMINARY ACTIVITY FOR Beer s Law Investigations Guided Inquiry Version Experiment 11 The primary objective of this Preliminary Activity is to determine the concentration of an unknown copper (II) sulfate

More information

salicylic acid and acetylsalicylic acid in pharmaceuticals

salicylic acid and acetylsalicylic acid in pharmaceuticals Automatic Chemistry, Vol. 12, No. 6 (November-December 1990), pp. 263-266 Automatic continuous on-line monitoring of salicylic acid and acetylsalicylic acid in pharmaceuticals j. M. L6pez Fernindez,f M.

More information