2011_ST20_Taylor.pdf. SSPT Document not peer-reviewed by CORESTA
|
|
- Arabella Tate
- 5 years ago
- Views:
Transcription
1 An Ion Chromatographic Method for the Quantitative Determination of Hydrogen Cyanide in Cigarette Smoke using Pulsed Amperometric Detection Dr W Guthery and Dr M J Taylor Filtrona Technology Centre CORESTA Joint Study Group Meeting, Graz 9 13 October 2011
2 Introduction Hydrogen Cyanide has long been considered an important smoke compound due to its toxicity It is considered to be extremely toxic and short-term exposure can lead to headaches, dizziness, nausea and vomiting. Hydrogen Cyanide is thought to be a respiratory irritant and a contributor to smoking related chronic obstructive lung disease and cardiovascular disease The levels of hydrogen cyanide in smoke have been studied for many years and it has been on the list of compounds measured and reported in Canada and Brazil for over ten years More recently it has been included on the emissions list in Taiwan and Thailand and is also on the FDA draft list of toxicants and the second WHO list of nine compounds Hydrogen cyanide is one of seven compounds on the STMA harm reduction index used in China Typical yields of Hydrogen Cyanide in smoke can be up to 500 ppm but the short term WEL in the UK for continuous exposure is 10 ppm The accurate measurement of Hydrogen Cyanide levels in smoke are essential to monitor any reduction in Hydrogen Cyanide levels
3 Background Hydrogen Cyanide in whole smoke has in the past been measured by a range of techniques including, ion selective electrodes, potentiometric titrations, colourmetric titrations, gas chromatography and automated colourmetric systems using continuous flow analysers More recently continuous flow analysers seem to be the method of choice Whole smoke Hydrogen cyanide is trapped in impingers containing sodium hydroxide solution Cyanides are converted to cyanogen chloride by reaction with chloramine-t and then react with a pyridine/pyrazolone solution to form a coloured compound measured at 540 nm The continuous flow analysers can measure 50 samples per hour and has a reasonable linear range 0.08 to 4 µg /ml However, the method has some disadvantages mainly the use of toxic and very odourous pyridine as one of the required reagents and the lower limit of detection can cause problems for cigarettes with yields below 5 µg/cig
4 Continuous Flow Analyser AA3 2.0 ml/min Wash Solution 0.80 ml/min Sample 5 x Turn Coil 0.8 ml/min NaOH Solution 0.32 ml/min Air 0.23 ml/min Air 5 x Turn Coil 0.80 ml/min Phosphate buffer 10 x Turn Coil 20 x Turn Coil Waste Colourimeter 540 nm 0.10 ml/min Resample 0.23 ml/min Chloramine - T 0.80 ml/min Pyridine/Pyrazolone Solution Computer Waste
5 Ion Chromatography Method - Summary Ten cigarettes smoked on rotary smoking machine Whole smoke trapped in two impingers Aliquot of trap solution syringe filtered Analysed using IC with electrochemical detection
6 Dionex Ion Chromatograph
7 Electrochemical Cell
8 Principle of Operation The ED40 amperometry cell is a miniature flow-through cell with a titanium cell body (counter electrode) and a Ag/AgCl reference electrode. Installed directly after the ion chromatography column (suppressor not required) Mobile phase flows in a thin channel (0.2 ml) parallel to the surface of a flat disc electrode. A potentiostat diverts the cell current through the counter electrode Two modes of operation - DC Amperometry, a constant voltage is applied to the working electrode and the resulting current is the detector output - Integrated or Pulsed Amperometry, a repetitive series of potentials is applied to the cell. The integrated current (charge) from the oxidation is measured during a portion of a repeating potential vs. time waveform
9 Silver Working Electrodes Two types of working electrode solid and disposable Disposable electrodes have been reported to have shown comparable or better reproducibility and linearity compared with solid electrodes 1. Advantage of ease of use compared with solid electrodes which require regular cleaning. Disadvantage is the cost 1 Liang, L et al., J.Chromatogr. A, 2005, 1085, 37-41
10 Analytical Conditions Instrument: Columns: Injection: Column Temp: 30 C Eluent: Flow rate: Pressure: Detection: Working electrode: Collection rate: Run time: Dionex ion chromatography system with gradient pump and ED40 electrochemical cell IonPac AS7 (4 x 250mm) anion-exchange column with AG7 (4 x 50mm) guard column and Metal-Free Trap column (MFC-1) 25 ml full loop 0.1M NaOH/ 0.2M NaOAc 1 ml/min ~1800 psi Pulsed Amperometric Detection (PAD) Disposable Ag with Ag/AgCl reference electrode 1.00 Hz 20 mins
11 The Cyanide Optimised Waveform Silver is oxidised in the presence of the cyanide ions and its electrons are released A series of potentials is defined as a waveform A three-potential PAD waveform using E 1, E 2 and E 3 is applied over 1-sec. E 1 is the detection potential. The remaining potentials clean and restore the electrode for subsequent detection Cyanide Waveform Time (sec) Potential vs Ag/AgCl (V) Gain region Integration Ramp Off Off On On On (Start) On On Off (End) On On Off On Off Off On Off Off On
12 Whole Smoke Trapping Method Summary 10 cigs smoked on a Rotary smoking machine 8 mm Teflon insert positioned at the MS head All of the smoke phases trapped into two impingers containing 100 and 20 ml 0.1M NaOH/0.2M NaOAc 55 mm CF pad placed at the syringe head Extracts combined then shaken and filtered through 0.45 mm nylon filter
13 1 2: cyanide Example Chromatogram Silver electrodes are selective to other ions including sulphides, bromides, and thiosulphates. AS7 column demonstrated clear separation from matrix interferences AS7 contains a strong anion exchange (SAX) stationary phase (alkyl quaternary ammonium). It was necessary to use a high ionic strength mobile phase to enable elution with a reasonable turnaround time MAY 2011_ #17 MS 1R5F 1 ED_1 nc min
14 Calibration Data Quantitative analysis was performed using standards made up from certified reference material, CN - standard (1000ppm in OH - ), supplied by SPEX CertiPrep. Seven calibration standards ( mg/ml) prepared in 0.1M NaOH Linear response; Corr. Coeff. (%) cyanide External ED_1 Area [nc*min] Concentration mg/ml
15 Sample Stability Storage Time (at 22C) K3R4F Mean µg/cig SD < 12 hours > 24 hours
16 Reference Cigarette Values Sample Mean Yield µg/cig AA 3 IC K3R4F K1R5F CM SATF 2006 Collaborative Study Mean Yield for all Laboratories for K1R5F = 20.2 µg/cig
17 Comparison of Trapping Techniques K3R4F Cigarettes Sample Mean Yield µg/cig AA 3 IC Impingers Granular Traps Granular Trap Extended 44 mm CF Holder Silica Gel Beads 0.5 to 1.5 mm Diameter Approximately 8 g of beads per trap 3 Cigarettes per trap smoked using linear smoking machine
18 Comparison of Methods Parameter AA 3 IC Range mg/ml 0.08 to to 10.0 Samples/Hour 50 3 Reagents Pyridine Aqueous Overnight Operation No Yes Coefficient of Linearity
19 Conclusions A method for the determination of Hydrogen cyanide in whole smoke using IC/PAD is under development Yield values are comparable with current methods The IC method has lower limits of detection and does not use any organic reagents but is slower More work will be carried out on the long term stability of the method and its applicability to sidestream smoke
20 2011_ST20_Taylor.pdf
Dionex IonPac AS28-Fast-4µm column
CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS-Fast-4µm Columns Product Specifications The Thermo Scientific Dionex IonPac AS-Fast-4µm column is a high-capacity, hydroxide-selective anionexchange column
More informationThermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance
CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS anion-exchange column is designed for the fast analysis of inorganic
More informationBritish American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke
British American Tobacco Group Research & Development Method - Determination of ammonia in mainstream smoke 1 SCOPE OF APPLICATION The method is applicable to quantitative determination of the yields of
More informationHYDROGEN CYANIDE 6017
HYDROGEN CYANIDE 6017 HCN MW: 27.03 CAS: 74-90-8 RTECS: MW6825000 METHOD: 6017, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003 OSHA: 10 ppm (skim) NIOSH: STEL 4.7 ppm ACGIH: C 10 ppm (skin) (1 ppm
More informationHigh-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems
High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems Yan Liu, Zhongqing Lu, and Chris Pohl; Thermo Fisher Scientific, Sunnyvale, CA USA Overview
More informationDetermination of Adsorbable Organic Halogen in Wastewater
Determination of Adsorbable Organic Halogen in Wastewater Jingli Hu, Jeff Rohrer, and Kirk Chassaniol Thermo Fisher Scientific, Sunnyvale, CA The world leader in serving science Outline AOX (Adsorbable
More informationCORESTA Recommended Method No. 83
Cooperation Centre for Scientific Research Relative to Tobacco Special Analytes Sub-Group CORESTA Recommended Method No. 83 DETERMINATION OF AMMONIA IN MAINSTREAM CIGARETTE SMOKE BY ION CHROMATOGRAPHY
More informationNew On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography
New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography Yan Liu, Zhongqing Lu, and Chris Pohl, Thermo Fisher Scientific, Sunnyvale, CA USA Overview Purpose: In this work, new high-pressure
More informationProPac WCX-10 Columns
ProPac WCX-10 Columns Guidance for column use Tips to maximize column lifetime ProPac WCX-10 Column Tips and Tricks This guide provides essential information and invaluable guidelines for mobile phases,
More informationNew Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection
New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection Barbara Shao, Terri Christison, Fei Pang, Cathy Tanner, and Frank Hoefler, Thermo Fisher Scientific, Sunnyvale,
More informationDetermination of trace anions in concentrated hydrofluoric acid
APPLICATION NOTE 78 Determination of trace anions in concentrated hydrofluoric acid Authors Archava Siriraks Thermo Fisher Scientific, Sunnyvale, CA Keywords HF, ICS-5000 +, IonPac AS10, IonPac AC10, ion
More informationDetermination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection
Determination of Tartaric cid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection Suparerk Tukkeeree, Chanita Chantarasukon, and Jeff Rohrer Thermo Fisher Scientific, angkok,
More informationDetermination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production
Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Thunyarat Phesatcha, Suparerk Tukkeeree, Jeff Rohrer 2 Thermo
More informationanion-exchange column
columnsionpac ASA Anion-Exchange Column The IonPac ASA is a carbonateselective anion-exchange column designed for the fast separation of inorganic anions, including fluoride, chlorite, bromate, chloride,
More informationcolumns IonPac SCS 1 Silica Cation Separator
columns IonPac SCS Silica Cation Separator The SCS Silica Cation Separator is designed for use with nonsuppressed conductivity detection, or single-column ion chromatography (SCIC). This column is particularly
More informationCyanide and sulfide analysis using amperometric detection and Metrosep A Supp /4.0
Metrosep A Supp 10-100 /4.0 Branch Environment, Food, Beverages Keywords IC; 850; 858; Metrosep A Supp 10-100/4.0; Cyanide; Sulfide; 2.850.9110; DC Mode Summary The determination of sulfide and cyanide
More informationcolumns IonPac AS17-C Anion Exchange Column
columns IonPac AS-C Anion Exchange Column The IonPac AS-C is a hydroxide-selective anion exchange column designed for fast gradient separation of inorganic anions. The key application for the AS-C column
More informationKeywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000
Evaluation of Various Anion-Exchange Chemistries for the Analysis of Haloacetic Acids in Drinking Water Using -D Matrix Elimination Ion Chromatography and Suppressed Conductivity Detection White Paper
More informationSulfur speciation in mining leachates by ion exchange with direct UV detection, using a perchlorate mobile phase.
direct UV detection, using a perchlorate mobile phase. Branch Branch 15: Mineral resources Branch 2: Water, wastewater, air, environmental protection Keywords IC / Metrohm 944 Professional UV/VIS Detector
More informationION CHROMATOGRAPHY SYSTEM S 150
ION CHROMATOGRAPHY SYSTEM S 150 WATER ANALYSIS ENVIRONMENTAL ANALYSIS ANION & CATION ANALYSIS ION CHROMATOGRAPHY IIon Chromatography is an analytical separation technique based on ionic interactions. Dissolved
More informationDeterminations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection
Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection Terri Christison, Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA Overview
More informationInline sample preparation for the determination of anions in sodium hydroxide
APPLICATION UPDATE 72331 Inline sample preparation for the determination of anions in sodium hydroxide Authors Hua Yang and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords Sample preparation,
More informationThermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water
Thermo Scientific Dionex Ion Chromatography Solutions Global water safety bromate analysis in drinking water Safe drinking water public health assurance As a vital limited resource required for survival,
More informationHigh Performance Liquid Chromatography
High Performance Liquid Chromatography What is HPLC? It is a separation technique that involves: Injection of small volume of liquid sample Into a tube packed with a tiny particles (stationary phase).
More informationMethod Abstract. Flow Solution Fluoride, USGS by ISE and FIA
Scope Summary Interferences Fluoride, USGS by ISE and FIA This method is used for the determination of fluoride in drinking water, surface water, and domestic and industrial wastes according to USGS Method
More informationDetermination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System
Application Update 55 Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free System Introduction Hydrogen peroxide is an essential chemical in the fabrication
More informationUsing a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components
APPLICATION UPDATE 157 Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components Authors Sumate Pengpumkiat, Weerapong Worawirunwong,
More informationFlow Solution FS 3700 Automated Chemistry Analyzer
Method Abstract Total Cyanide by ISO 14403 Document #42600116 Flow Solution FS 3700 Automated Chemistry Analyzer Total Cyanide by In-Line Ultraviolet Digestion, Gas Diffusion, and Photometric Detection,
More informationTime Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples
Application Note 79 Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples INTRODUCTION Cow s milk is of particular dietary value to infants,
More informationThermo Scientific. Anion-Exchange Column
CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS9-µm Anion-Exchange Column Product Specifications The Themo Scientific Dionex IonPac AS9-µm high-capacity, hydroxide-selective, anion-exchange column is
More informationSpeciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry
APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos
More informationGUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY
Page 1 of 13 CIPAC/4105/R GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Prepared for CIPAC by Dr M J Tandy*, P M Clarke and B White (UK) The rapid
More information4. Ion chromatography (IC) 4.1. Educational aims and objectives
4. Ion chromatography (IC) 4.1. Educational aims and objectives Principles of ion chromatography method and instrumentation. Methods for ionic compounds analysis including ion chromatography and its subtypes.
More informationHigh Performance Liquid Chromatography
Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN
More informationHigh Performance Liquid Chromatography
Updated: 3 November 2014 Print version High Performance Liquid Chromatography David Reckhow CEE 772 #18 1 HPLC System David Reckhow CEE 772 #18 2 1 Instrument Basics PUMP INJECTION POINT DETECTOR COLUMN
More informationThe determination of trace anions in concentrated phosphoric acid
TECHNICAL NOTE The determination of trace anions in concentrated phosphoric acid Authors Edward Kaiser and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords Ion chromatography, IC, IonPac
More informationHPLC Praktikum Skript
HPLC Praktikum Skript Assistants: Gianluca Bartolomeo HCI D330, 3 46 68, bartolomeo@org.chem.ethz.ch Sahar Ghiasikhou HCI E330, 2 29 29, ghiasikhou@org.chem.ethz.ch 1. Introduction In chromatographic techniques,
More informationQuantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory
Customer Application Note: 114 Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory Sébastien N. Ronkart, Ph.D.; Société wallonne des eaux, rue de la
More informationThe Determination of Trace Anions in Concentrated Phosphoric Acid
Technical Note The Determination of Trace Anions in Concentrated Phosphoric Acid INTRODUCTION The determination of trace anions in phosphoric acid is hampered by a large excess of phosphate ion. Chloride
More informationAccurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography
Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography Carl Fisher and Linda Lopez Thermo Fisher Scientific, Sunnyvale,
More informationAcclaim Mixed-Mode WCX-1
Acclaim Mixed-Mode WCX-1 Product Manual for the Acclaim Mixed-Mode WCX-1 Column Page 1 of 28 PRODUCT MANUAL for the Acclaim Mixed-Mode WCX-1 Columns 4.6 x 150 mm, P/N (068353) 4.6 x 250 mm, P/N (068352)
More informationIon Chromatography (IC)
Ion Chromatography (IC) Purpose: This module provides an introduction to Ion Chromatography (IC). In this module the basic theory and applications of IC will be presented at a level that assumes a basic
More informationSchool of Chemistry UNIVERSITY OF KWAZULU-NATAL, WESTVILLE CAMPUS JUNE 2009 EXAMINATION CHEM340: INSTRUMENTAL ANALYSIS.
School of Chemistry UNIVERSITY OF KWAZULU-NATAL, WESTVILLE CAMPUS JUNE 2009 EXAMINATION CHEM340: INSTRUMENTAL ANALYSIS DURATION: 3 HOURS TOTAL MARKS: 100 Internal Examiners: Professor A Kindness Dr T Msagati
More informationDetermination of urea in ultrapure water by IC-MS/MS
APPLICATION NOTE 72482 Determination of urea in ultrapure water by IC-MS/MS Authors Soon Fatt Lee, 1 Fiona Teh Hui Boon, 1 Chris Cheah Hun Teong, 1 and Jeff Rohrer 2 ¹Thermo Fisher Scientific, Singapore
More informationCation Exchange HPLC Columns
Cation Exchange PRP-X00 Polymeric cation exchange packing for separation of inorganic cations and organic cations. Easily separate mono or divalent cations. ph stable from to. Use with organic solvent
More informationDetermination of an anionic fluorochemical surfactant in a semiconductor etch bath
APPLICATIN NTE 9 Determination of an anionic fluorochemical surfactant in a semiconductor etch bath Authors Mark Laikhtman and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA Keywords Ion chromatography,
More informationDetermination of Chloride, Fluoride, Bromide, Nitrate, Sulphate and Phosphate in Water Sample...
I J P F A International Science Press Determination of Chloride, Fluoride, Bromide, Nitrate, Sulphate and Phosphate in Water Sample from Different Area of North Gujarat Region by Ion Chromatography Bharat
More informationThermo Scientific Dionex IonPac AS23 Anion-Exchange Column
Chromatography Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS high-capacity, carbonate based anion-exchange column is designed for
More informationAnion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM
IC Application Work AW CH6-1048-012011 Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM Branch: Chemical industry; Water, wastewater, environmental protection,
More informationRapid Post-Blast Inorganic Explosive Analysis by Suppressed Ion Chromatography
Rapid Post-Blast Inorganic Explosive Analysis by Suppressed Ion Chromatography L. Canella, C. Austin, A. Beavis, P. Maynard, M. Dawson, C Roux, and P. Doble Centre for Forensic Science, University of Technology,
More informationTechnical Report. Determination of Nitrite and Nitrate in Smokeless Tobacco Products by Ion Chromatography and Continuous Flow Analysis
Tobacco and Tobacco Products Analytes Sub-Group Technical Report Determination of Nitrite and Nitrate in Smokeless Tobacco Products by Ion Chromatography and Continuous Flow Analysis 2016 Collaborative
More informationfor Acclaim Mixed-Mode WCX-1
for Acclaim Mixed-Mode WCX- Product Manual for the Acclaim Mixed-Mode WCX- Column Page of 8 Product Manual for Acclaim Mixed-Mode WCX- Columns Acclaim Mixed-Mode WCX-, 3µm, Analytical column, 3.0 x 50mm
More informationHints for Strong Ion Exchange Resins
Hints for Strong Ion Exchange Resins Chromatography Application Note AN98 Abstract Ion exchange columns are a powerful means of isolating and purifying compounds, but their use is limited due to lack of
More informationfor Acclaim Mixed-Mode HILIC-1 Column
for Acclaim Mixed-Mode HILIC-1 Column Product Manual for ACCLAIM Mixed-Mode HILIC-1 Page 1 of 17 Product Manual for ACCLAIM Mixed-Mode HILIC-1 Column 5µm, 4.6 x 250mm, P/N 066844 5µm, 4.6 x 150mm, P/N
More informationDetermination of Inorganic Anions in Drinking Water by Ion Chromatography
Determination of Inorganic s in Drinking Water by Ion Chromatography Peter Jackson Thermo Fisher Scientific Inc. Application Note Introduction The determination of common inorganic anions in drinking water
More informationcolumns IonPac AS22 Anion-Exchange Column
columns IonPac AS Anion-Exchange Column The IonPac AS is a carbonate based anion-exchange column designed for the determination of inorganic anions and low-molecular-weight organic acids including fluoride,
More informationANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis
DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes
More informationHigh Performance Liquid Chromatography
STANDARDBASE techniques: High Performance Liquid Chromatography Drenthe College, The Netherlands 1. Introduction HPLC. High Performance Liquid Chromatography High Performance Liquid Chromatography (HPLC)
More informationCh.28 HPLC. Basic types of Liquid Chromatography Partition (LLC) Adsorption (LSC) Ion Exchange (IC) Size Exclusion (SEC or Gel Chromatography)
Ch.28 HPLC 28.1 Basic types of Liquid Chromatography Partition (LLC) Adsorption (LSC) Ion Exchange (IC) Size Exclusion (SEC or Gel Chromatography) High Performance (Pressure) LC Glass column st.steel (high
More informationAnalysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography
Application Note Inorganic Ions, Water Testing, Minerals, Metals, Basic Chemicals Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography Authors Anne Mack, Adam Bivens
More informationMETHOD 7199 DETERMINATION OF HEXAVALENT CHROMIUM IN DRINKING WATER, GROUNDWATER AND INDUSTRIAL WASTEWATER EFFLUENTS BY ION CHROMATOGRAPHY
METHOD 7199 DETERMINATION OF HEXAVALENT CHROMIUM IN DRINKING WATER, GROUNDWATER AND INDUSTRIAL WASTEWATER EFFLUENTS BY ION CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 This method provides procedures for
More informationSeparation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC
TECHNICAL NOTE 122 Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC Authors Terri Christison and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA,
More informationAccurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography
Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography Kirk Chassaniol Manager of IC Technical Sales Support Thermo Fisher Scientific 1 The world leader in serving
More informationDetermination of Trace Cations in Power Plant Waters Containing Morpholine
Application Note 8 Determination of Trace Cations in Power Plant Waters Containing Morpholine INTRODUCTION Morpholine and ammonium are used as additives in power plant waters. Morpholine acts as a corrosion
More informationMETHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY
METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8032 is used to determine trace amounts of acrylamide monomer (CAS No. 79-06-1) in aqueous matrices. This method may be
More informationcolumns IonPac AS18 Anion-Exchange Columns
columns IonPac AS Anion-Exchange Columns The IonPac AS is a hydroxideselective anion-exchange column designed for the determination of inorganic anions and low-molecularweight organic acids including fluoride,
More informationDirect Determination of Small Organic Acids in Sea Water by IC-MS
Direct Determination of Small Organic Acids in Sea Water by IC-MS Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA Application Note Key Words MSQ Plus Single Quadrupole Mass
More informationcolumns Acclaim Mixed-Mode WCX-1 for Separating Basic Molecules
columns Acclaim Mixed-Mode WCX- for Separating Basic Molecules The Acclaim Mixed-Mode WCX- is a novel, high-efficiency, silica-based column specially designed for separating various basic analytes. This
More informationCation Exchange HPLC Columns
Cation Exchange HPLC Columns Hamilton offers seven polymeric packing materials for cation exchange separations. Type Recommended Application(s) PRP-X00 PRP-X00 PRP-X800 HC-0 HC-7 Ca + HC-7 H + HC-7 Pb
More informationCHAPTER CHROMATOGRAPHIC METHODS OF SEPARATIONS
Islamic University in Madinah Department of Chemistry CHAPTER - ----- CHROMATOGRAPHIC METHODS OF SEPARATIONS Prepared By Dr. Khalid Ahmad Shadid Chemistry Department Islamic University in Madinah TRADITIONAL
More informationDepartment of Civil and Environmental Engineering, Stanford University, 473 Via Ortega, Stanford, California 94305, United States
Supporting Information for Capture and Reductive Transformation of Halogenated Pesticides by an Activated Carbon-Based Electrolysis System for Treatment of Runoff Yuanqing Li 1 and William A. Mitch 1,
More informationA New On-line Cyanide Analyzer for Measurement of Cyanide in Hydrometallurgical Processing of Precious Metal Ores
Application Note 37890312 A New On-line Cyanide Analyzer for Measurement of Cyanide in Hydrometallurgical Processing of Precious Metal Ores Keywords CNSolution 9310 On-line Cyanide Analyzer Cyanidation
More informationLC III: HPLC. Originally referred to as High-Pressure Liquid Chromatography. Now more commonly called High Performance Liquid Chromatography
LC III: HPLC What is HPLC? Originally referred to as High-Pressure Liquid Chromatography Now more commonly called High Performance Liquid Chromatography In general: The instrument controlled version of
More informationTrace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography
Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography Harpreet Dhillon and John Statler Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 11 Introduction
More informationThermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes
columns Thermo Scientific Dionex IonPac AS Anion-Exchange Column Isocratic Separation of Sulfur Species and Inorganic Anions Using the Dionex IonPac AS Column. Fluoride. Bromate. Chloride. Nitrite. Bromide.
More informationThermo Scientific Dionex IonPac AS24A Anion-Exchange Column
CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS4A Anion-Exchange Column Product Specifications Thermo Scientific Dionex IonPac AS4A anion-exchange column is designed for the separation of haloacetic
More informationDETERMINATION OF NINE VOLATILE NITROSAMINES AND HYDROXY-NITROSAMINES IN CIGARETTE FILLER AND MAINSTREAM TOBACCO SMOKE
DETERMINATION OF NINE VOLATILE NITROSAMINES AND HYDROXY-NITROSAMINES IN CIGARETTE FILLER AND MAINSTREAM TOBACCO SMOKE Mehran SHARIFI, Peter JOZA, Bill RICKERT 69 th TOBACCO SCIENCE RESEARCH CONFERENCE
More informationFor new orders of the following parts discussed in this manual, please use the updated part numbers listed below.
Errata Product Manual for Dionex IonPac UTAC2 and AC-ER 065376-0 For new orders of the following parts discussed in this manual, please use the updated part numbers listed below. Part Old Part Number in
More informationDetermination of trace anions in high-purity waters by high volume/direct injection ion chromatography
APPLICATION NOTE 113 Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography Author Edward Kaiser Thermo Fisher Scientific, Sunnyvale, CA Keywords Trace ionic
More informationStandard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2
Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State
More informationThermo Scientific Dionex IonPac AS26 Anion-Exchange Column
Chromatography Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS column is a high-capacity, hydroxide-selective, anion-exchange column
More informationDetermination of trace anions in concentrated glycolic acid
TECHNICAL NOTE Determination of trace anions in concentrated glycolic acid Authors Edward Kaiser and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA Keywords Ion chromatography, IC, Dionex ICS- +,
More informationHPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25
High Performance Liquid Chromatography (HPLC) Harris Chapter 25 12/1/2005 Chem 253 - Chapter 25 1 HPLC Separation of nonvolatile or thermally unstable compounds. If the analyte/sample can be found to be
More informationAcclaim Mixed-Mode WAX-1 Columns
User Manual Acclaim Mixed-Mode WAX- Columns 06565 Revision 03 October 05 Product Manual for the Acclaim Mixed-Mode WAX- Column Page of 3 Product Manual for Acclaim Mixed-Mode WAX- Columns 5µm, 0 x 50 mm,
More informationUser Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.
User Manual IonPac NS2 Columns 065501 Revision 02 October 2015 For Research Use Only. Not for use in diagnostic procedures. Product Manual for IonPac NS2 Columns IonPac NS2, 5µm, 4x250mm Analytical Column
More informationfor IonPac AG11-HC IonPac AS11-HC
for IonPac AG-HC IonPac AS-HC IonPac AS-HC Manual Document No. 3333-7 Page of 45 PRODUCT MANUAL for the IONPAC AG-HC GUARD COLUMN (4 x 5 mm, P/N 59) ( x 5 mm, P/N 593) IONPAC AS-HC ANALYTICAL COLUMN (4
More informationExpanded capabilities in ammonium and amine detection
PRODUCT SPECIFICATIONS Thermo Scientifi c Dionex SC-CERS 500 Salt Converter-Cation Electrolytically Regenerated Suppressor Expanded capabilities in ammonium and amine detection Benefits Broadened ammonium
More informationXinyu LIU, Peter JOZA, Andrew MASTERS, Bill RICKERT
Determination of Ethylene Oxide (ETO) in Mainstream Cigarette Smoke Using Hydrobromic Acid Derivatization and Gas Chromatography-Mass Spectrometry Method Xinyu LIU, Peter JOZA, Andrew MASTERS, Bill RICKERT
More informationDetermination of Trace Anions in Concentrated Hydrofluoric Acid
Technical Note Determination of Trace Anions in Concentrated Hydrofluoric Acid INTRODUCTION There is a need for a reliable method to determine trace anions in hydrofluoric acid. The large excess of fluoride
More informationMETHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV)
METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV) 1.0 SCOPE AND APPLICATION 1.1 Method 9012 is used to determine the concentration of inorganic cyanide in an aqueous waste or leachate.
More informationChromatographic Methods of Analysis Section - 4 : Ion Exchange Chrom. Prof. Tarek A. Fayed
Chromatographic Methods of Analysis Section - 4 : Ion Exchange Chrom. Prof. Tarek A. Fayed Ion Exchange Chromatography (IEC) In this type of chromatography, the solid stationary phase )organic resin) is
More informationChapter 3 Electrochemical methods of Analysis-Potentiometry
Chapter 3 Electrochemical methods of Analysis-Potentiometry Electroanalytical chemistry Contents Introduction Galvanic and electrolytic cells Salt bridge Electrode potential and cell potential Indicator
More informationDetermination of Sulfate and Chloride in Ethanol by Ion Chromatography
Application Note 75 Determination of Sulfate and Chloride in Ethanol by Ion Chromatography Ethanol used as a blending agent in gasoline can be contaminated with chloride and sulfate that form plugging
More informationPackings for HPLC. Packings for HPLC
Summary of packings for HPLC In analytical HPLC, packings with particle sizes of 3 to 10 µm are preferred. For preparative separation tasks, also particles with diameters larger than 10 µm are applied.
More informationEluent Preparation for High-Performance Anion-Exchange Chromatography with Pulsed Amperometric Detection
Eluent Preparation for High-Performance Anion-Exchange Chromatography with Pulsed Amperometric Detection Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Technical Note 7 Introduction High-performance
More informationPacked Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents
No. 089 SEPARATION REPORT Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS Table of Contents 1. Introduction 1 2. Column Specification 1 3. Features of Packing Materials
More informationQuantitative determination of ethephon in soluble concentration (SL) by Ion chromatography
Quantitative determination of ethephon in soluble concentration (SL) by Ion chromatography Author : Li Guoping Yu Rong Shui Rong Institute for the Control of Agrochemicals, Ministry of Agriculture, P.
More informationFor new orders of the following parts discussed in this manual, please use the updated part numbers listed below.
Errata Product Manual for Dionex IonPac AS and AG Columns 03479-2 For new orders of the following parts discussed in this manual, please use the updated part numbers listed below. Part Old Part Number
More informationMETHOD #: Recommended for Approval for NPDES (August, 1991)
METHOD #: 218.6 Recommended for Approval for NPDES (August, 1991) TITLE: ANALYTE: INSTRUMENTATION: Determination Of Dissolved Hexavalent Chromium In Drinking Water, Groundwater And Industrial Wastewater
More informationHPLC Columns. HILICpak VT-50 2D MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia
HPLC Columns MANUAL HILICpak VT-50 2D Shodex HPLC Columns Europe, Middle East, Africa, Russia For technical support please use contact details shown below: SHOWA DENKO EUROPE GmbH Shodex Business Konrad-Zuse-Platz
More information