PERFORMANCE VERIFICATION IONSCAN LS

Size: px
Start display at page:

Download "PERFORMANCE VERIFICATION IONSCAN LS"

Transcription

1 Page of 3 PERFORMANCE VERIFICATION IONSCAN LS Smiths Detection 30 Technology Dr. Warren, NJ Copyright Smiths Detection 004 All rights reserved. Reproduction, adaptation or translation without prior written permission is prohibited, except as allowed by copyright laws. Warranty The information contained in this document is subject to change without notice. Smiths Detection - makes no warranty of any kind with regards to this material, including, but not limited to, the implied warranties or merchantability and fitness for a particular purpose. Smiths Detection - shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

2 Page of 3 PRE EXECUTION PROTOCOL APPROVAL SMITHS DETECTION - PROTOCOL APPROVAL I agree that the Performance Verification (PV) procedures, assembled by Smiths Detection, are appropriate for the equipment defined in the table of contents. Name Signature Date CUSTOMER PROTOCOL APPROVAL I agree that the PV procedures in this document are applicable to the equipment defined in the table of contents. Name Signature Date QUALITY PROTOCOL APPROVAL I agree that the PV procedures in this document are applicable to the equipment defined in the table of contents. Name Signature Date POST EXECUTION PROTOCOL APPROVAL SMITHS DETECTION - PROTOCOL APPROVAL I agree that the PV procedures, assembled by Smiths Detection, are appropriate for the equipment defined in the table of contents. Name Signature Date CUSTOMER PROTOCOL APPROVAL I agree that the PV procedures in this document are applicable to the equipment defined in the table of contents. Name Signature Date QUALITY PROTOCOL APPROVAL I agree that the PV procedures in this document are applicable to the equipment defined in the table of contents. Name Signature Date

3 Page 3 of 3 INSTRUMENT TO BE VERIFIED IONSCAN MODEL# Serial Number Firmware Version Operational Time Negative Mode Positive Mode Sample Count TEST LOCATION Instrument underwent Perfromance Verification as of the date this protocol was performed. Company Name Address Department Building Number Room Date Perfromance Verification performed Verified by (Signature) Date

4 Page 4 of 3 TABLE OF CONTENTS PRE EXECUTION PROTOCOL APPROVAL... SMITHS DETECTION - PROTOCOL APPROVAL... CUSTOMER PROTOCOL APPROVAL... QUALITY PROTOCOL APPROVAL... POST EXECUTION PROTOCOL APPROVAL... SMITHS DETECTION - PROTOCOL APPROVAL... CUSTOMER PROTOCOL APPROVAL... QUALITY PROTOCOL APPROVAL... INSTRUMENT TO BE VERIFIED... 3 TEST LOCATION... 3 TABLE OF CONTENTS... 4 PARTS AND MATERIALS... 6 REQUIRED TEST PARTS AND MATERIALS... 6 REQUIRED TEST STANDARDS... 6 GENERAL INSTRUCTIONS... 7 SIGNATURE PROTOCOL... 7 DATE FORMAT... 7 ERROR CORRECTION... 7 NO-ENTRY SPACES... 7 TEST PROCEDURE... 7 DESORBER SEAL CHECK... 8 PURPOSE... 8 BACKGROUND... 8 TEST EQUIPMENT LIST... 8 PROCEDURE... 8 RESULTS... 8 NOISE CHECK... 9 PURPOSE... 9 BACKGROUND... 9 TEST EQUIPMENT LIST... 9 PROCEDURE... 9 RESULTS... 9 CALIBRANT ION CHECK... 0 PURPOSE... 0 BACKGROUND...0 PROCEDURE... 0 RESULTS...

5 Page 5 of 3 REACTANT ION CHECK... PURPOSE... BACKGROUND... PROCEDURE... LIMIT OF DETECTION CHECK... 3 PURPOSE... 3 BACKGROUND...3 TEST EQUIPMENT LIST... 3 PROCEDURE... 3 RESULTS... 3 INJECTION REPEATABILITY CHECK... 4 PURPOSE... 4 BACKGROUND...4 TEST EQUIPMENT LIST... 4 PROCEDURE... 4 RESULTS... 4 DETECTOR LINEARITY CHECK... 5 PURPOSE... 5 BACKGROUND...5 TEST EQUIPMENT LIST... 5 PROCEDURE... 5 RESULTS... 6 CARRYOVER CHECK... 7 PURPOSE... 7 BACKGROUND...7 TEST EQUIPMENT LIST... 7 PROCEDURE... 7 RESULTS... 8 SUMMARY OF PERFORMANCE VERIFICATION... 9 FINAL CONCLUSION... 9 APPENDIX... 0 NOISE ANALYSIS... 0 REGRESSION ANALYSIS... DETERMINATION OF INJECTION REPRODUCIBILITY... REVISION HISTORY... 3

6 Page 6 of 3 PARTS AND MATERIALS The following parts and materials are required for running the Perfromance Verification. REQUIRED TEST PARTS AND MATERIALS Quantity Item.0 µl or 5 µl Syringe Solution Standards Kit pk Teflon Filters (p/n 68340) pk Teflon Membranes (p/n ) Teflon Card, Ring Assembly IM3 Software Manager Version Personal computer running Windows 000 Computer Interface Cable Excel Software Package 00 ml Pesticide Grade or similar hexanes 00 ml Pesticide Grade or similar Acetone REQUIRED TEST STANDARDS Quantity Solution ml 5 pg/ul TNT in hexanes ml 50 pg/µl TNT in hexanes ml 00 pg/µl TNT in hexanes ml 50 pg/µl TNT in hexanes ml 00 pg/µl TNT in hexanes ml 50 pg/µl TNT in hexanes ml 500 pg/µl TNT in hexanes ml 5 ng/ul TNT in hexane ml 50 pg/ul diazepam in acetone ml 00 pg/µl diazepam in acetone ml 50 pg/µl diazepam in acetone ml 00 pg/µl diazepam in acetone ml 50 pg/µl diazepam in acetone ml ng/µl diazepam in acetone ml 5ng/uL diazepam in acetone Solutions prepared by Smiths Detection - using certified analytical standards.,4,6-tnt Diazepam Analyte Concentration Supplier Lot No.

7 Page 7 of 3 GENERAL INSTRUCTIONS SIGNATURE PROTOCOL TESTED BY: Signature and/or initials of Smiths Detection Field Service Engineer, or other personnel performing tests. Date indicates the date test was executed. ACCEPTED BY: Signature and/or initials of customer representative. QA ACCEPTED: Signature and/or initials of customer QA department responsible for reviewing data. DATE FORMAT Unless otherwise specified by the customer, the proper format for the date is: dd/mon/yy. Where the month is the first three letters of the month. eg. 0FEB04 ERROR CORRECTION In the event of an entry error, cross out the incorrect value with a single line and write the correct entry. Initial and date the error. If space does not allow, footnote the error, and correction at the bottom of the page. eg. 0FEB04 NO-ENTRY SPACES Any entry that is not filled in (ie. test not necessary, space for comments not needed, etc.) should be crossed out with one diagonal line. Write a short reason, or N/A on the line along with initials and date. TEST PROCEDURE. Turn instrument on and put into negative ion mode.. Load latest version of instrument default negative mode control parameters. 3. Verify detection algorithm for TNT. 4. Allow instrument to come ready and wait an additional hour before proceeding with test. 5. Perform all checks in negative ion mode. 6. Switch instrument to positive ion mode. 7. Load latest version of instrument default positive mode control parameters. 8. Verify detection algorithm for diazepam. 9. Allow instrument to come ready and wait an additional hour before proceeding with test. 0. Perform all checks in positive ion mode.

8 Page 8 of 3 DESORBER SEAL CHECK PURPOSE To verify that when the IONSCAN-LS performs an analysis and the desorber stage moves up and seals the sample substrate between the desorber and the inlet, a proper seal is obtained. BACKGROUND To analyze a sample, the sample substrate is inserted into the tray on the front of instrument and the slide is moved to the right to introduce the sample into the analyzer. When the analysis cycle starts, the desorber moves up and seals the sample between the desorber heater and the IMS inlet. As the sample is heated the sample carrier gas (clean dry air) sweeps the desorbed vapors into the IMS detector. In order to ensure efficient sample transfer the sample should be properly sealed between the desorber and the inlet to ensure no air escapes out the side. TEST EQUIPMENT LIST. Teflon filter card assembly with standard Teflon filter (p/n 68340) PROCEDURE This test is performed in negative ion mode only.. Allow IONSCAN-LS unit to come ready.. Assemble a new Teflon filter, card and ring. 3. Insert filter card into IONSCAN and analyze. 4. After the analysis, check for seal on bottom of Teflon filter. 5. Record below if the seal on the bottom of Teflon filter is even and centered on the filter. RESULTS Properly Centered Imprint Improperly Centered Imprint Is the imprint of seal on the bottom of Teflon filter (shown in gray) even and centered on the filter as shown in the figure above? Comments

9 Page 9 of 3 NOISE CHECK PURPOSE To verify the IONSCAN-LS detector is operating within the defined acceptance limits for system noise. BACKGROUND Noise is an important characteristic that specifies a detector s performance. Low noise level is especially important when analyzing compounds at low trace levels. IONSCAN detector noise primarily originates from micro-phonic noise (motion of guard grid relative to collector plate) and preamp noise (resistance noise). The root mean square noise is determined by using a built in Ionscan Manager Software function to measure the root mean square noise in a defined drift time interval. RMS Amp RMS (root mean square) amplitude of the signal over all segments of the Seg Range, in the drift time interval specified. P-P Amp Displays the peak-to-peak amplitude of all segments of the range (difference between maximum and minimum of all segments). TEST EQUIPMENT LIST. Instrument Manager Software. (IM Software). PROCEDURE Negative Ion Mode. With slide tray down run three blank samples.. Acquire sample plasmagram with IM software. 3. With built in IM software function, measure RMS and Peak to Peak noise from msec and record in table. Positive Ion Mode. With slide tray down run three blank samples.. Acquire sample plasmagram with IM software. 3. With built in IM software function, measure RMS and Peak to Peak noise from msec and record in table. RESULTS Specified Tolerance Negative Ion Mode Positive Ion Mode Time Interval msec msec Recorded RMS noise < 3 du Peak to Peak noise < 5 du Are noise levels in negative ion mode within specified tolerances? Are noise levels in positive ion mode within specified tolerances?

10 Page 0 of 3 CALIBRANT ION CHECK PURPOSE To verify that a calibrant peak is present in the correct position in both positive and negative modes. BACKGROUND After the IONSCAN unit has warmed up and temperatures have stabilized, the unit searches for the presence of the calibrant peak before the unit goes into READY mode. The IONSCAN-LS unit uses the calibrant peak position to determine the drift times of all analyte peaks which have been programmed in based on the equation below: t anal d K K o = anal t o cal cal d wheret anal d calibrant peak, K anal o and calibrant peaks. is the expected drift time of the analyte ion, t cal and K cal o d is measured drift time of the are the programmed reduced ion mobility values of the analyte Parameters which effect the calibrant ion drift time are drift tube temperature, inlet temperature, drift flow rate, ambient atmospheric pressure and humidity level of the drift gas. All these parameters are either controlled or compensated for. The instrument controls drift tube and inlet temperature and drift flow. An air purification cartridge provides clean dry air for the instrument to use. An absolute pressure transducer determines the absolute pressure that is used to determine the expected position of calibrant. The IONSCAN-LS is designed to operate only when the calibrant peak is identified by its internal algorithm. This test will verify that the calibrant algorithm is properly programmed in and the IONSCAN-LS is detecting it within manufacturer s specifications. PROCEDURE Negative Ion Mode. Record in the table below if the instrument is in READY mode.. Note calibrant installed in instrument. 3. From configuration file, note calibrant Ko. 4. Run a blank sample and record the FWHM and Max Amplitude and delta of the calibrant. Positive Ion Mode. Record in the table below if the instrument is in READY mode.. Note calibrant installed in instrument. 3. From configuration file, note calibrant Ko. 4. Run a blank sample and record the FWHM and Max Amplitude and delta of the calibrant.

11 Page of 3 RESULTS READY after hour? Calibrant Calibrant Ko Calibrant FWHM Calibrant Max Amp Calibrant delta Negative Ion Mode Positive Ion Mode Ready after hour? Calibrant Ko within expected range (See table)? Calibrant FWHM within expected range (See teble)? Calibrant Max Amp greater than threshold listed (See table)? Calibrant delta within + 60? Negative Ion Mode Positive Ion Mode Calibrant Ko Range FWHM Range Max Amp Threshold Isobutyramide Nicotinamide nitrobenzonitrile

12 Page of 3 REACTANT ION CHECK PURPOSE To verify that reactant peaks are present in the correct position in the negative mode. BACKGROUND In the negative mode, a reactant is added via a diffusion source in the sample path to aid in the ionization process. Since the IONSCAN-LS does not actively monitor the reactant level, as it does the calibrant, it is necessary to verify the level to establish that the instrument is operating within specifications. This is done by evaluating the reactant level at the beginning and end of a series of consecutive runs. The IONSCAN-LS must maintain a minimum level over the course of the analyses. PROCEDURE Negative Ion Mode Only. Run a blank sample with no substrate present.. Scroll through segments and using the gaussian fit function, find the maximum amplitude of the reactant peak nearest 7.0ms. 3. Run 5 consecutive analyses (starting each analysis as soon as the IONSCAN-LS comes READY) with no substrate present. 4. On the last sample, scroll through segments and using the gaussian fit function, find the maximum amplitude of the reactant peak nearest 7.0ms. Initial Reactant Amplitude Reactant Amplitude after Sample series. Maximum Reactant Amplitude Amplitude Threshold 600 du 00 du Max Amp above threshold?

13 Page 3 of 3 LIMIT OF DETECTION CHECK PURPOSE To verify the IONSCAN-LS is able to detect a minimum quantity of TNT and diazepam. BACKGROUND To demonstrate IONSCAN-LS units ability to detect 5 pg TNT in negative ion mode and 50 pg diazepam in positive ion mode. TEST EQUIPMENT LIST. Teflon membrane (p/n ). 5 pg/µl trinitrotoluene solution in hexane pg/µl diazepam solution in acetone 4. 5 or 0 µl syringe PROCEDURE Negative Ion Mode Set up:. Place a Teflon membrane in the swipe tray and close tray. Run three blank samples. 3. Deposit a µl aliquot of 5 pg/µl TNT solution in the middle of the Teflon membrane. 4. Allow aliquot to evaporate for 5 seconds then analyze the sample. 5. Record if TNT was detected. 6. Repeat steps 3 to 5 two more times. Positive Ion Mode Set up:. Place a Teflon filter in the swipe tray and close tray.. Run three blank samples. 3. Deposit a µl aliquot of 50 pg/µl diazepam solution in the middle of the Teflon membrane. 4. Allow aliquot to evaporate for 5 seconds then analyze the sample. 5. Record if diazepam was detected. 6. Repeat steps 3 to 5 two more times. RESULTS Run Run Run 3 Was TNT detected? Was diazepam detected? Was 5 pg of TNT detected? Was 50 pg of diazepam detected?

14 Page 4 of 3 INJECTION REPEATABILITY CHECK PURPOSE To verify IONSCAN-LS reproducibility associated with the analysis of sample aliquots is within acceptable limits. BACKGROUND The injection volume precision is based on a sequence of runs where a TNT standard is used in negative mode and diazepam standard is used in positive ion mode. These standards are injected 5 times, always with the same volume. The average cumulative amplitudes are determined for each sample and the relative standard deviation of the average is calculated for positive and negative mode. TEST EQUIPMENT LIST. Teflon membrane (p/n ). 0.5 ng/µl diazepam in acetone ng/µl trinitrotoluene in hexane 4. 5 or 0µL syringe PROCEDURE Negative Ion Mode. Place a Teflon membrane in the swipe tray and close tray.. Run 3 blank samples. 3. Deposit a µl aliquot of the TNT solution on the Teflon membrane. 4. Allow aliquot to evaporate for 5 seconds then analyze the sample. 5. Record Cumulative Amplitude in table below. 6. Repeat steps 3-5 four more times. 7. Calculate Relative Standard Deviation. Positive Ion Mode. Place a Teflon membrane in the swipe tray and close tray.. Run 3 blank samples. 3. Deposit a µl aliquot of the diazepam solution on of the Teflon membrane. 4. Allow aliquot to evaporate for 5 seconds then analyze the sample. 5. Record Cumulative Amplitude in table below. 6. Repeat steps 3-5 four more times. 7. Calculate Relative Standard Deviation. RESULTS Sample Sample Sample 3 Sample 4 Sample 5 Average Std Dev Relative Standard Deviation (RSD) Is RSD < 0 % 0.5 ng TNT Negative Ion Mode(Cum Amp) 0.5 ng Diazepam Positive Ion Mode(Cum Amp)

15 Page 5 of 3 DETECTOR LINEARITY CHECK PURPOSE To verify the detector shows a linear response over a working range. BACKGROUND Detectors become non-linear when the amount of the test compound increases above a certain limit. The linearity test is based on a sequence of runs where a TNT solution or Diazepam solution is injected several times from vials containing different concentrations using the same injection volume. When the cumulative amplitudes associated with each peak are plotted against the concentrations, the points should lie on a straight line. The correlation of this calibration curve is used as one criterion to measure detector linearity. Five standards will be analyzed to evaluate linearity. TEST EQUIPMENT LIST. Teflon membrane (p/n ). 50,00,50, 00,50 pg/µl diazepam solutions in acetone 3. 50,00, 50, 00, 50 pg/µl TNT solutions in hexane 4. 5 or 0 µl syringe PROCEDURE Negative Ion Mode. Place a Teflon membrane in the swipe tray and close tray.. Run three blank samples. For each TNT solution: 3. Deposit a µl aliquot onto a Teflon membrane. 4. Allow aliquot to evaporate for 5-7 seconds then analyze the sample. 5. Record the CumAmp in the table below. 6. Repeat steps 3-5 two more times. 7. Use a suitable program to plot concentration versus average cumulative amplitude and to calculate the r value based on the equation y = ax + b. Positive Ion Mode. Place a Teflon membrane in the swipe tray and close tray.. Run three blank samples. For each diazepam solution: 3. Deposit a µl aliquot onto a Teflon membrane. 4. Allow aliquot to evaporate for 5-7 seconds then analyze the sample. 5. Record the CumAmp in the table below. 6. Repeat steps 3-5 two more times. 7. Use a suitable program to plot concentration versus average cumulative amplitude and to calculate the r value based on the equation y = ax + b.

16 Page 6 of 3 RESULTS Sample Conc. pg/µl Sol 00 Sol 00 Sol Sol Sol Slope = Intercept = r = Is r > 0.98? TNT Linearity Results Diazepam Linearity Results Run Run Run 3 Average Run Run Run 3 Average

17 Page 7 of 3 CARRYOVER CHECK PURPOSE After a detection, one PASS should be obtained before analyzing the next sample. The purpose of this test is to determine the amount of carryover that occurs after analyzing a large sample. BACKGROUND To test carryover when an injection is done from a vial filled with a sample of high concentration. TEST EQUIPMENT LIST. 500 pg/µl TNT solution. ng diazepam solution 3. Teflon membrane (p/n ) 4. 5 or 0 µl syringe PROCEDURE Negative Ion Mode. Place a Teflon membrane in the swipe tray and close tray.. Run three blank samples. 3. Deposit a µl aliquot of the ng/µl TNT solution on the Teflon membrane. 4. Allow aliquot to evaporate for 5-7 seconds then analyze the sample. 5. Record TNT Cumulative Amplitude in the table below. 6. Run blank samples and record Cumulative Amplitude of TNT peak until a PASS is obtained. 7. Repeat steps 3-6 two more times. If first blank was clean skip this step. Positive Ion Mode. Place a Teflon membrane in the swipe tray and close tray.. Run three blank samples. 3. Deposit a µl aliquot of the ng/µl diazepam solution on the Teflon membrane. 4. Allow aliquot to evaporate for 5-7 seconds then analyze the sample. 5. Record diazepam Cumulative Amplitude in the table below. 6. Run blank samples and record Cumulative Amplitude of diazepam peak until a PASS is obtained 7. Repeat steps 3-6 two more times. If first blank was clean skip this step.

18 Page 8 of 3 RESULTS TNT Carry Over Check Sample 500 pg TNT (TNT Blank. (TNT CumAmp) CumAmp) Blank Blank Blank 3 3 Average Is % carryover at 500 pg level < %? Sum of TNT CumAmp Response for Blanks % Carry Over: (Sum of Blank/ ng response) x 00 Diazepam Carry Over Check Sample.0 ng diazepam (CumAmp) 3 Average Blank. (diazepam CumAmp) Blank Blank Blank 3 Sum of diazepam CumAmp Response for Blanks % Carry Over: (Sum of Blank/ 0 ng response) x 00 Is % carryover at ng level < %?

19 Page 9 of 3 SUMMARY OF PERFORMANCE VERIFICATION Check DESORBER SEAL CHECK Negative ION Mode Positive Ion Mode NOISE CHECK CALIBRANT ION CHECK REACTANT ION CHECK LIMIT OF DETECTION INJECTION REPEATABILITY CHECK DETECTOR LINEARITY CHECK CARRYOVER CHECK FINAL CONCLUSION Did Instrument pass all of the tests in the above table? Yes (PASSED PERFORMANCE VERIFICATION) No (FAILED PERFORMANCE VERIFICATION) Comments:

20 Page 0 of 3 APPENDIX In this appendix the documentation of mathematical or statistical formulas used in the tests described in the previous chapters are discussed. NOISE ANALYSIS During an analysis, multiple scans are co-added to define a segment of analysis and the total number of segments defined by the analysis time. For example a typical scan period is 0 msec, 0 scans are co-added to define segment of analysis (0.4 sec) and the analysis time is 8 seconds giving 0 segments of analysis. The IM3 software noise analysis feature calculates signal and peak-to-peak amplitude data in the specified drift time interval. A drift time interval at the beginning of the analysis is selected which is free of peaks. The following data is then displayed in the top left corner of the plasmagram: Seg Range: RMS Amp: P-P Amp: Max P-P Seg: Max P-P Amp: Range of segments used in the calculations. RMS (root mean square) amplitude of the signal over all segments of the Seg Range, in the drift time interval specified. Displays the peak-to-peak amplitude of all segments of the range (difference between maximum and minimum of all segments). Specifies the segment with the highest peak-to-peak amplitude. Shows the peak-to-peak amplitude of the Max P-P Seg. 3D Plasm agram 40 Peak Height (du) Drift Time (ms) IN THE ABOVE FIGURE NOISE WOULD BE ANALYZED OVER THE DRIFT TIME INTERVAL FROM 5.0 TO 5.3 MSEC. EACH OF THE LINES IN THE FIGURE IS A SEGMENT OF ANALYSIS. DATA SET OVER WHICH NOISE ANALYSIS IS PERFORMED: aij signal (du) for segment j at point i in drift time s j Number of scans per segment for segment j n, n Range of indices ( i ) over which RMS and Peak-to-Peak noise is to be calculated k, k Range of segments ( j ) over which RMS and Peak-to-Peak noise is to be calculated

21 Page of 3 Define: u b b j = ( n n + ) u i= u a ij Min( a u j, min ij j ) = For n i n = Max( aij u ) For n i n j, max j σ j = u [ aij u j ] i= u ( n n + ) Therefore, the RMS Amp= σ = k j= k σ j s j k j= k s j Max P-P Amp: Maximum Peak-to-Peak Amplitude: b = Max b [ b ] P P j, max j,max For j= k, k Max P-P Seg: Segment associated with Max P-P amplitude kmax = Value of j on which b p-p based Further: min Min( b j, min ) max Max( b j, max b = b = Therefore, P-P Amp= b abs = ( b max bmin ) )

22 Page of 3 REGRESSION ANALYSIS The linearity test uses regression analysis during the evaluation. Linearity is derived from the linear function: Let: N = number of discrete observations X i = independent variable, i th observation Y i = dependent variable, i th observation Determination of coefficients: X Y b N = i = N = i = Y X i i N { ( X i X )( Yi Y )} i= = N ( X i X ) i= a = Y bx Regression Coefficient: r = Y X = a + bx N { ( X i X )( Yi Y )} N N ( X ) ( ) i X Yi Y i= i= i= DETERMINATION OF INJECTION REPRODUCIBILITY N = number of replicate samples A = Measured value associated with i th sample i Average A Standard Deviation SD = N i= = N A i N ( Ai A) i= N (Results are a small sampling) Relative Standard Deviation RSD SD 00 = A

23 Page 3 of 3 REVISION HISTORY. Revision.00, Effective Date, August 9, 00. Revision.00, Effective Date, July 7, 003 Removed Calibrant and Reactant Ion Check. Removed Data Acquisition tables from Test Procedure Added option to use Teflon membrane filters 3. Revision 3.00, Effective Date, March 4, 004 Modifications based on Version 6.40, Installation and Operation Qualification of IONSCAN-LS, Effective Date, March 4, Revision 4.00, Effective Date, August 0, 0 Updated method to be similar to PV documentation for IONSCAN-LS/HPI

PERFORMANCE VERIFICATION: IONSCAN LS with Autosampler Module

PERFORMANCE VERIFICATION: IONSCAN LS with Autosampler Module Version 5.0 Page 1 of 8 PERFORMANCE VERIFICATION: IONSCAN LS Smiths Detection 30 Technology Dr. Warren, NJ 07059 908--9100 Copyright Smiths Detection 007 All rights reserved. Reproduction, adaptation or

More information

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide What is the MassHunter METLIN Metabolite PCD/PCDL? 2 Where to find more information 2 Kit Content 3 Installation 5 Before you start 5 Install MassHunter

More information

Analyst Software. Automatic Optimization Tutorial

Analyst Software. Automatic Optimization Tutorial This document is provided to customers who have purchased AB Sciex equipment to use in the operation of such AB Sciex equipment. This document is copyright protected and any reproduction of this document

More information

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note Abstract Area and retention time are the two primary measurements in gas chromatography. The precision with which

More information

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd.

More information

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing Automated Preparation of Simulated Distillation Standards and Samples for ASTM Methods D2887, D7213, D7398 and D6352 using the 7693A System with Easy SamplePrep Software Author Roger L. Firor Agilent Technologies,

More information

SCIEX Cation Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide

SCIEX Cation Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems A49109AE May 2015 AB Sciex Pte. Ltd and its affiliates disclaims all warranties with respect to this document, expressed or implied, including

More information

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Application Note Food Testing and Agriculture Author Chen-Hao (Andy) Zhai and Rong-jie Fu Agilent Technologies (Shanghai)

More information

Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput

Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput JSB is an authorised partner of Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput Application Note - Environmental #129 Author Anne Jurek Applications Chemist

More information

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis Application Note AN89.V.1 Extraction of Methylmalonic Acid from Serum Using ISOLUTE SAX Page 1 Extraction of Methylmalonic Acid from Serum Using ISOLUTE SAX Prior to LC-MS/MS Analysis Sample Preparation

More information

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 66 Triple Quadrupole LC/MS System Application Note Environmental Authors Don Noot, Matthew Noestheden, and Ralph Hindle Vogon

More information

LAP-013. Procedure Title: Date: 6/28/96. Author: Raymond Ruiz and Tina Ehrman ISSUE DATE: 8/26/96 SUPERSEDES: 11/01/94.

LAP-013. Procedure Title: Date: 6/28/96. Author: Raymond Ruiz and Tina Ehrman ISSUE DATE: 8/26/96 SUPERSEDES: 11/01/94. Ethanol Project Chemical Analysis and Testing Task Laboratory Analytical Procedure Procedure Title: LAP-013 HPLC Analysis of Liquid Fractions of Process Samples for Monomeric Sugars and Cellobiose Author:

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 12 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 77A mini TD and 70A GC Application Note Environmental Authors Tingting Bu, Xiaohua Li Agilent Technologies (Shanghai) Co.,

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (11) 04 QUALIFICATION OF EQUIPMENT ANNEX 1: QUALIFICATION OF HPLC EQUIPMENT Full document title and reference Document type

More information

PTA-427 BAROMETRIC PRESSURE TRANSDUCER OPERATOR S MANUAL

PTA-427 BAROMETRIC PRESSURE TRANSDUCER OPERATOR S MANUAL PTA-427 BAROMETRIC PRESSURE TRANSDUCER OPERATOR S MANUAL REVISION: 4/95 COPYRIGHT 1991, 1995 CAMPBELL SCIENTIFIC, INC. WARRANTY AND ASSISTANCE The PTA-427 Barometric Pressure Transducer is warranted by

More information

Signal, Noise, and Detection Limits in Mass Spectrometry

Signal, Noise, and Detection Limits in Mass Spectrometry Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville

More information

MassLynx 4.1 Peak Integration and Quantitation Algorithm Guide

MassLynx 4.1 Peak Integration and Quantitation Algorithm Guide MassLynx 4.1 Peak Integration and Quantitation Algorithm Guide 71500122009/Revision A Copyright Waters Corporation 2005. All rights reserved. Copyright notice 2005 WATERS CORPORATION. PRINTED IN THE UNITED

More information

Precision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System

Precision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System Precision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System Keywords: Hydrodynamic radius, diffusion coefficient, sizing, Taylor dispersion analysis, protein, antibodies,

More information

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor Application Note Pesticides Authors Charlie Spanjers and Paul Dauenhauer University of Minnesota, Twin Cities

More information

Product Data Sheet PD-0036-B. 3M PC Boardmount Plug and Receptacle, 2 mm 95X Series

Product Data Sheet PD-0036-B. 3M PC Boardmount Plug and Receptacle, 2 mm 95X Series Product Data Sheet PD-0036-B 3M PC Boardmount Plug and Receptacle, 2 mm 95X Series Page: 1 of 7 1.0 SCOPE...2 2.0 PRODUCT TESTED...2 3.0 GENERAL CONDITIONS...2 3.1 Test Specimens...2 3.2 Standard Test

More information

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies MS n Analysis With Fast Polarity Switching in the Agilent Series LC/MSD Trap SL Application Note Christine Miller Introduction During mass spectrometric analysis, many compounds respond better to a particular

More information

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application Static Headspace Blood Alcohol Analysis with the G Network Headspace Sampler Application Forensics Author Roger L. Firor and Chin-Kai Meng Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-0

More information

Application. Gas Chromatography February Introduction

Application. Gas Chromatography February Introduction Ambient Headspace Analysis with the Agilent 7683 Automatic Liquid Sampler Application Gas Chromatography February 1998 Authors Matthew S. Klee and Chin Kai Meng Agilent Technologies, Inc. 2850 Centerville

More information

Information on the test material EDS-TM002 and the BAM software package EDX Spectrometer Test for determination of the spectrometer performance

Information on the test material EDS-TM002 and the BAM software package EDX Spectrometer Test for determination of the spectrometer performance BAM 6.8 8.5.213 Information on the test material EDS-TM2 and the BAM software package EDX Spectrometer Test for determination of the spectrometer performance 1. Introduction Energy dispersive spectrometers

More information

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Sensitive Detection of 2-MIB and Geosmin in Drinking Water Sensitive Detection of -MIB and Geosmin in Drinking Water Application Note Environmental Author Yean-Woong You Agilent Technologies, Inc. Seoul, Korea Abstract An automated SPME extraction method for easy

More information

Gyrolab Protein A Kit Quantification of residual Protein A ligands in the presence of excess amounts of IgG

Gyrolab Protein A Kit Quantification of residual Protein A ligands in the presence of excess amounts of IgG Quantification of residual Protein A ligands in the presence of excess amounts of IgG Product number P0020456 P0020457 Product Name Gyrolab Protein A Kit for MabSelect SuRe Ligand Gyrolab Protein A Kit

More information

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION METHOD 80 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION 1.0 SCOPE AND APPLICATION 1.1 Method 80 may be used to determine the concentration of acetonitrile (CAS No. 75-05-8) in aqueous

More information

WeatherHub2 Quick Start Guide

WeatherHub2 Quick Start Guide WeatherHub2 Quick Start Guide Table of Contents 1 Introduction... 1 2 Packing List... 1 3 Connections... 1 4 IP Addressing... 2 5 Browser Access... 3 6 System Info... 3 7 Weather Station Settings... 4

More information

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source Application Note Environmental Authors Peter Gautschi and Harry Prest Senior Application Scientist

More information

Micro Volume QuEChERS kit

Micro Volume QuEChERS kit 225-37872 Sep. 2018 Small Capacity Pretreatment Kit Micro Volume QuEChERS kit Instruction Manual Read this manual thoroughly before you use the product. Keep this manual for future reference. This page

More information

i.mx 6 Temperature Sensor Module

i.mx 6 Temperature Sensor Module NXP Semiconductors Document Number: AN5215 Application Note Rev. 1, 03/2017 i.mx 6 Temperature Sensor Module 1. Introduction All the i.mx6 series application processors use the same temperature sensor

More information

DRI Ecstasy Assay Application BECKMAN COULTER UniCel DxC and Synchron System(s)

DRI Ecstasy Assay Application BECKMAN COULTER UniCel DxC and Synchron System(s) DRI Ecstasy Assay Application BECKMAN COULTER UniCel DxC and Synchron System(s) Beckman Coulter Reorder Number A39935 Homogeneous enzyme immunoassay for the qualitative or semi-quantitative determination

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

Medium Pressure Sensor Analog Output

Medium Pressure Sensor Analog Output Medium Pressure Sensor Analog Output SM6844-015-A-B-5-S FEATURES Analog pressure calibrated and temperature compensated output Amplified analog output Compensated temperature range: 0 to 85oC Absolute

More information

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction User s Guide Waterproof Palm ph Meter Model PH220 Introduction Congratulations on your purchase of the Extech PH220 waterproof ph/mv meter. This microprocessor-based device with tactile buttons is battery

More information

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry The end of mass-speculation Pre-cleaned and certified vials for mass spectrometry Bonded, ultra-pure closures LC/MS and GC/MS Manufacturing the Ultimate Vial for Mass Spectrometry More than a quarter century

More information

UNIVERSITY GENT FACULTY PHARMACEUTICAL SCIENCES Department of Pharmaceutical Analysis Drug Quality and Registration (DruQuaR)

UNIVERSITY GENT FACULTY PHARMACEUTICAL SCIENCES Department of Pharmaceutical Analysis Drug Quality and Registration (DruQuaR) UNIVERSITY GENT FACULTY PHARMACEUTICAL SCIENCES Department of Pharmaceutical Analysis Drug Quality and Registration (DruQuaR) Academical year 2009-2010 DEVELOPMENT OF A NEW HIGH-THROUGHPUT METHOD: ION

More information

Low Pressure Sensor Amplified Analog Output SM6295-BCM-S

Low Pressure Sensor Amplified Analog Output SM6295-BCM-S Low Pressure Sensor Amplified Analog Output SM6295-BCM-S-040-000 FEATURES Pressure range from 0 to 40 cmh 2 O 5.0 V operation Amplified analog output (10 to 90%Vdd) Compensated temperature range: 0 to

More information

Resistivity vs. Temperature of a Superconductor WARNING

Resistivity vs. Temperature of a Superconductor WARNING Resistivity vs. Temperature of a Superconductor 1.0 Expected Learning Outcomes Calibrate a transducer (silicon diode) for low-temperature measurements and measure temperatures down to that of boiling liquid

More information

Scott Anderson, Kathy Lawrence Dennis McCreary, Melissa Wilcox. Grace Davison Discovery Sciences

Scott Anderson, Kathy Lawrence Dennis McCreary, Melissa Wilcox. Grace Davison Discovery Sciences Minimizing Purification Bottlenecks Using a Multi Detector Approach to Flash Chromatography While Achieving Maximum Sample Recovery in Post Synthetic Workups Scott Anderson, Kathy Lawrence Dennis McCreary,

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Requirements of an Automated Sample Delivery System in Today s Realm of Ever Increasing Sensitivity Demands Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek Abstract

More information

TETRACHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC15010

TETRACHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC15010 TETRACHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC15010 BIOCHEMISTRY The Tetrachlorethylene or perchlorethylene, is an organic halide. Its structure is similar to that of a molecule of ethene which

More information

HI-POWER / LED MICRO-08

HI-POWER / LED MICRO-08 HI-POWER / LED MICRO-08 FEATURES + Smallest form factor with up to 1.5W (500mA) output power + Ultra wide illumination field angle (FWHM 155 degree) + Low thermal resistance:

More information

3M Pak 50 Boardmount Socket, P50 series P50-XXXS-XXX-EA. 3M Pak 50 Boardmount Plug, P50 series P50-XXXP-XXX-EA

3M Pak 50 Boardmount Socket, P50 series P50-XXXS-XXX-EA. 3M Pak 50 Boardmount Plug, P50 series P50-XXXP-XXX-EA Page of 8 3M Pak 50 Boardmount Socket, P50 series P50-XXXS-XXX-EA 3M Pak 50 Boardmount Plug, P50 series P50-XXXP-XXX-EA Product Specification 78-50-0-7 Revised 08-09- Page of 8 Table of Contents Title

More information

Technical Procedure for General Laboratory Equipment

Technical Procedure for General Laboratory Equipment Technical Procedure for General Laboratory Equipment 1.0 Purpose - This procedure specifies the required elements for the use of general laboratory equipment. 2.0 Scope - This procedure applies to all

More information

Training Path FNT IT Infrastruktur Management

Training Path FNT IT Infrastruktur Management Training Path FNT IT Infrastruktur Management // TRAINING PATH: FNT IT INFRASTRUCTURE MANAGEMENT Training Path: FNT IT Infrastructure Management 2 9 // FNT COMMAND BASIC COURSE FNT Command Basic Course

More information

3mm Advanced Super Flux LEDs B84-YSC-A1T1U1DH-AM

3mm Advanced Super Flux LEDs B84-YSC-A1T1U1DH-AM 3mm Advanced Super Flux LEDs Feature RoHS compliant. Piranha package. Colorless clear resin. Low thermal resistance Packaged in tubes for automatic insertion equipment. Total flux: 7150 to 14250 mlm at

More information

Simplifying the Process: Automated USP 643 / EP

Simplifying the Process: Automated USP 643 / EP Simplifying the Process: Automated USP 643 / EP 2.2.44 Purified Water and Water For Injection Testing Using A Next Generation TOC Analyzer Application Note By: Brian Wallace and Stephen Lawson Background

More information

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology Gas Chromatography/Mass Spectrometry Author Harry Prest 1601 California Avenue

More information

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) Application Note Environmental Authors Steve Wilbur Agilent Technologies,

More information

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS Application Note Multi-detector Splitter Authors Charlie Spanjers and Andrew Jones Activated Research Company 7561

More information

ELISA QUALITY ASSURANCE Analytical Phase

ELISA QUALITY ASSURANCE Analytical Phase LOGO ELISA QUALITY ASSURANCE Analytical Phase - DR. ALI MIRJALILI QUALITY ASSURANCE Dept. PISHTAZ TEB DIAGNOSTICS 01/02/92 Quality improvement congress Definitions: Quality Assurance Quality Control:-

More information

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539) Determination of ormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell P Column (EPA 539) Application Note Environmental Authors Rong-jie Fu and Chen-ao (Andy) Zhai Agilent Technologies

More information

Author. Abstract. Introduction

Author. Abstract. Introduction Improved Performance for the Analysis of Aromatics in Gasoline by ASTM Method D5769 Using the Agilent 5973 inert Gas Chromatography/Mass Spectrometry System Application Author James D. McCurry Agilent

More information

STATIC GAS MONITOR Type SGM/DEW Revision A of 20 aprile 2016

STATIC GAS MONITOR Type SGM/DEW Revision A of 20 aprile 2016 APPLICATIONS Moisture monitoring of air or gas (SF6) Suitable for indoor or outdoor Industrial, medical or aerospace fields HIGHLIGHTS High voltage circuit breakers commonly used for distribution and transmission

More information

Power Resistor Thick Film Technology

Power Resistor Thick Film Technology Power Resistor Thick Film Technology DESIGN SUPPORT TOOLS click logo to get started FEATURES 50 W at 25 C case temperature heatsink mounted Direct mounting ceramic on heatsink Broad resistance range: 0.010

More information

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay... Table of Contents 1. Principle of Assay... 3 2. Specimens... 3 3. Reagents and Materials (HPLC grade)... 3 4. Standards, Controls, and Solutions... 4 5. Equipment and Special Supplies... 5 6. Instrumentation...

More information

Creating Empirical Calibrations

Creating Empirical Calibrations 030.0023.01.0 Spreadsheet Manual Save Date: December 1, 2010 Table of Contents 1. Overview... 3 2. Enable S1 Calibration Macro... 4 3. Getting Ready... 4 4. Measuring the New Sample... 5 5. Adding New

More information

Testing for particles in injectable products

Testing for particles in injectable products Testing for particles in injectable products 20-Nov-2018 MONITORING PHARMACEUTICALS Liquid sample technology that is compliant with USP must be part of any testing strategy into winning the war on

More information

Determination of Polymer Modifier in Asphalt

Determination of Polymer Modifier in Asphalt Standard Method of Test for Determination of Polymer Modifier in Asphalt AASHTO Designation: T xxx-xx (2005) 1. SCOPE 1.1. This method of test is used to determine the polymer content of an asphalt sample.

More information

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description WHO/SIT/19.R4 TEMEPHOS TECHNICAL TECHNICAL TEMEPHOS 1. Specification 1.1 Description Full specification WHO/SIT/19.R4 Revised 10 December 1999 The material shall consist of temephos together with related

More information

3M Pak 10 Plug, PK10 series PK10-XXXPX-X-DA. 3M Pak 10 Socket, PK10 series PK10-XXXSX-X-DA

3M Pak 10 Plug, PK10 series PK10-XXXPX-X-DA. 3M Pak 10 Socket, PK10 series PK10-XXXSX-X-DA Page of 8 M Pak 0 Plug, PK0 series PK0-XXXPX-X-DA M Pak 0 Socket, PK0 series PK0-XXXSX-X-DA Product Specification Revised 08-0- Page of 8 Table of Contents Title Page.............................................................................

More information

CR10XTCR THERMOCOUPLE REFERENCE

CR10XTCR THERMOCOUPLE REFERENCE CR10XTCR THERMOCOUPLE REFERENCE REVISION: 01/01/03 COPYRIGHT (c) 1995-2003 CAMPBELL SCIENTIFIC, INC. This is a blank page. WARRANTY AND ASSISTANCE The CR10XTCR THERMOCOUPLE REFERENCE is warranted by CAMPBELL

More information

TRICHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC16110

TRICHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC16110 TRICHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC16110 BIOCHEMISTRY Trichlorethylene, marketed under the name of thinner, is used as a degreaser of metals, preparation of paints and cleaning of fabrics.

More information

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS Application ote Food Testing & Agriculture Author Andy Zhai Agilent Technologies, Inc. Shanghai Co. Ltd.

More information

V. LAB REPORT. PART I. ICP-AES (section IVA)

V. LAB REPORT. PART I. ICP-AES (section IVA) V. LAB REPORT The lab report should include an abstract and responses to the following items. All materials should be submitted by each individual, not one copy for the group. The goal for this part of

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek

More information

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Tran T. Nham Introduction

More information

EVERLIGHT ELECTRONICS CO., LTD.

EVERLIGHT ELECTRONICS CO., LTD. Lead (Pb) Free Product - RoHS Compliant Feature Piranha package. Colorless clear resin. Low thermal resistance Packaged in tubes for automatic insertion equipment. Total flux: 7150 to 14250 mlm at 70mA.

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

Select the appropriate PM to be done and then perform the checklist under that section.

Select the appropriate PM to be done and then perform the checklist under that section. Agilent Preventive Maintenance provides factory recommended service for your analytical systems to ensure reliable operation and the accuracy of your results. Delivered by highly-trained and certified

More information

BIOO FOOD AND FEED SAFETY. Histamine Enzymatic Assay Kit Manual. Catalog #: Reference #:

BIOO FOOD AND FEED SAFETY. Histamine Enzymatic Assay Kit Manual. Catalog #: Reference #: BIOO FOOD AND FEED SAFETY Histamine Enzymatic Assay Kit Manual Catalog #: 1032-05 Reference #: 1032-05 BIOO Scientific Corp. 2010 TABLE OF CONTENTS GENERAL INFORMATION... 1 Product Description... 1 Procedure

More information

Application Note. Abstract. Authors. Introduction

Application Note. Abstract. Authors. Introduction Multiple Headspace Extraction for the Quantitative Determination of Residual Monomer and Solvents in Polystyrene Pellets Using the Agilent 7697A Headspace Sampler Application Note Authors Wenwen Shen and

More information

P-LED EALP03SXARA0-AM. Lead (Pb) Free Product - RoHS Compliant

P-LED EALP03SXARA0-AM. Lead (Pb) Free Product - RoHS Compliant Lead (Pb) Free Product - RoHS Compliant Feature Piranha package. Colorless clear resin. Low thermal resistance Packaged in tubes for automatic insertion equipment. Total flux: 7150 to 14250 mlm at 70mA.

More information

HEAVY DUTY SWING-OUT VALVE

HEAVY DUTY SWING-OUT VALVE ISO 900 REGISTERED COMPANY Akron Brass Company. 20 All rights reserved. No portion of this can be reproduced without the express written consent of Akron Brass Company. We will not be responsible for:

More information

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai)

More information

Power Resistor for Mounting onto a Heatsink Thick Film Technology

Power Resistor for Mounting onto a Heatsink Thick Film Technology Power Resistor for Mounting onto a Heatsink Thick Film Technology DESIGN SUPPORT TOOLS click logo to get started FEATURES 300 W at 85 C bottom case temperature Wide resistance range: 0.3 to 900 k E24 series

More information

Natural Rock Sample Manual

Natural Rock Sample Manual Natural Rock Sample Manual Revision 2014-06-12 DURRIDGE Company Inc. 524 Boston Road Billerica, MA 01821 Tel: (978) 667-9556 Fax: (978) 667-9557 service@durridge.com www.durridge.com 2014, DURRIDGE Company

More information

ISOCON-6. 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER

ISOCON-6. 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER ISOCON-6 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER Whilst every effort has been taken to ensure the accuracy of this document, we accept no responsibility for damage, injury, loss or expense resulting

More information

The DialPath Solution an Easier Way to Analyze Liquids By FTIR. Yanqia Wang, Ph.D. Application Engineer - FTIR Agilent Technologies May 7, 2015

The DialPath Solution an Easier Way to Analyze Liquids By FTIR. Yanqia Wang, Ph.D. Application Engineer - FTIR Agilent Technologies May 7, 2015 The DialPath Solution an Easier Way to Analyze Liquids By FTIR Yanqia Wang, Ph.D. Application Engineer - FTIR Agilent Technologies May 7, 2015 Outline Background of FTIR Transmission Technology Traditional

More information

ISSP User Guide CY3207ISSP. Revision C

ISSP User Guide CY3207ISSP. Revision C CY3207ISSP ISSP User Guide Revision C Cypress Semiconductor 198 Champion Court San Jose, CA 95134-1709 Phone (USA): 800.858.1810 Phone (Intnl): 408.943.2600 http://www.cypress.com Copyrights Copyrights

More information

Power Resistor Thick Film Technology

Power Resistor Thick Film Technology Power Resistor Thick Film Technology LTO series are the extension of RTO types. We used the direct ceramic mounting design (no metal tab) of our RCH power resistors applied to semiconductor packages. FEATURES

More information

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-exyl Columns to Separate Estrogens Application Note Pharmaceutical, Environmental Authors John W. enderson Jr. and William J. Long Agilent

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Dehydroepiandrosterone 3-sulfate, Primary Standard D-065 Page 1 of 6 Catalog Number: D-065 Lot: H 3 C Expiration: Description: February 2019 Dehydroepiandrosterone

More information

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF Analysis of Biomarkers in Crude Oil Using the Agilent 7 GC/Q-TOF Application Note Petrochemical and Environmental Authors Frank David Research Institute for Chromatography, Kennedypark 6, B-85 Kortrijk,

More information

AMP + MaxSpec Kit. Item No Customer Service Technical Support

AMP + MaxSpec Kit.   Item No Customer Service Technical Support AMP + MaxSpec Kit Item No. 710000 www.caymanchem.com Customer Service 800.364.9897 Technical Support 888.526.5351 1180 E. Ellsworth Rd Ann Arbor, MI USA TABLE OF CONTENTS GENERAL INFORMATION 3 Materials

More information

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS Agilent Technologies/Wasson ECE Monomer Analyzer Application Authors Fred Feyerherm 119 Forest Cove Dr. Kingwood, TX

More information

HEAVY DUTY SWING-OUT VALVE

HEAVY DUTY SWING-OUT VALVE ISO 900 REGISTERED COMPANY HEAVY DUTY SWING-OUT VALVE 52 53 23 5 2 2 3 28 39 NOT USED WITH SZ HANDLE 2 NOT USED WITH SZ HANDLE 5 2 3 7 9 9 Premier Farnell Corporation. 20 All rights reserved. No portion

More information

Conversion Kit for: 7620 and 7820 Swing-Out Valves ft-lbs ft-lbs 8835/ ft-lbs

Conversion Kit for: 7620 and 7820 Swing-Out Valves ft-lbs ft-lbs 8835/ ft-lbs ISO 900 REGISTERED COMPANY HEAVY DUTY SWING-OUT VALVE 52 53 23 5 2 2 3 28 39 NOT USED WITH SZ HANDLE 2 NOT USED WITH SZ HANDLE 5 2 3 7 9 9 Premier Farnell Corporation. 20 All rights reserved. No portion

More information

Field Pro Operation and Installation Manual

Field Pro Operation and Installation Manual Field Pro Operation and Installation Manual Xtreme Research Corporation SkyScan Division P.O. Box 336, Port Richey, FL 34673-0336 1-800-732-0665 Table of Contents 1. Installation 3 2. Operation 5 3. Care

More information

Advanced Test Equipment Rentals ATEC (2832) Agilent 86130A BitAlyzer Error Analysis Getting Started

Advanced Test Equipment Rentals ATEC (2832) Agilent 86130A BitAlyzer Error Analysis Getting Started Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Agilent 86130A BitAlyzer Error Analysis Getting Started Copyright Agilent Technologies 2001 All Rights Reserved. Reproduction,

More information

METHYLISOBUTYLKETONE (4-methyl-2 pentanone) IN URINE BY GC/MS Headspace Code GC10510

METHYLISOBUTYLKETONE (4-methyl-2 pentanone) IN URINE BY GC/MS Headspace Code GC10510 METHYLISOBUTYLKETONE (4-methyl-2 pentanone) IN URINE BY GC/MS Headspace Code GC10510 BIOCHEMISTRY The methyl isobutyl ketone is a colorless and flammable liquid that is used as organic solvent in the laboratory

More information

CR10TCR THERMOCOUPLE REFERENCE

CR10TCR THERMOCOUPLE REFERENCE CR10TCR THERMOCOUPLE REFERENCE REVISION: 12/96 COPYRIGHT (c) 1995-1996 CAMPBELL SCIENTIFIC, INC. This is a blank page. WARRANTY AND ASSISTANCE The CR10TCR THERMOCOUPLE REFERENCE is warranted by CAMPBELL

More information

MODEL 107 TEMPERATURE PROBE INSTRUCTION MANUAL

MODEL 107 TEMPERATURE PROBE INSTRUCTION MANUAL MODEL 107 TEMPERATURE PROBE INSTRUCTION MANUAL REVISION: 6/00 COPYRIGHT (c) 1983-2000 CAMPBELL SCIENTIFIC, INC. This is a blank page. Warranty and Assistance The MODEL 107 TEMPERATURE PROBE is warranted

More information

Solid Tantalum SMD Capacitors TANTAMOUNT, Hi-Rel COTS, Low ESR, Metal Case

Solid Tantalum SMD Capacitors TANTAMOUNT, Hi-Rel COTS, Low ESR, Metal Case Solid Tantalum SMD Capacitors TANTAMOUNT, Hi-Rel COTS, Low ESR, Metal Case PERFORMANCE CHARACTERISTICS Operating Temperature: -55 C to +125 C (above 85 C, voltage derating is required) Capacitance Range:

More information

COMOSAR E-FIELD PROBE CALIBRATION REPORT

COMOSAR E-FIELD PROBE CALIBRATION REPORT Page: 1/26 Issue: A Date: 2008/10/06 COMOSAR E-FIELD PROBE CALIBRATION REPORT Prepared By: Project Description: Prepared For (End User): LUC Jérôme, SATIMO SAR TEST BENCH Shenzhen Morlab Communication

More information

Power Resistors for Mounting Onto a Heatsink Thick Film Technology

Power Resistors for Mounting Onto a Heatsink Thick Film Technology Power Resistors for Mounting Onto a Heatsink Thick Film Technology DESIGN SUPPORT TOOLS click logo to get started FEATURES 1% tolerance available High power rating = 200 W Wide ohmic value range = 0.046

More information

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70 Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70 Application Note Abstract Geosmin and 2-Methylisoborneol are organic compounds that have a distinct scent. These compounds

More information