Precision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System

Size: px
Start display at page:

Download "Precision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System"

Transcription

1 Precision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System Keywords: Hydrodynamic radius, diffusion coefficient, sizing, Taylor dispersion analysis, protein, antibodies, quality control, protein folding Summary There is an increasing need within the biopharmaceutical industry for techniques which complement or enhance currently available techniques such as dynamic light scattering (DLS), size exclusion chromatography (SEC) and transmission electron microscopy (TEM) for monitoring the amount and extent of aggregation of therapeutic proteins and monoclonal antibodies. All of the techniques previously mentioned have limitations, e.g. it is not possible to determine size for proteins without dilution, they cannot be used in conjunction with in-line systems, e.g. GE AKTA explorers and also have limited application in the analysis of membrane proteins. This application note outlines results with a test protein, myoglobin, using a technique developed by Paraytec for determining the hydrodynamic radius of species in solution. Repeatability and intermediate precision are reported, and the key factor influencing the accuracy of the method is documented. The technique measures the extent of Taylor dispersion of a plug of sample moving through a buffer solution. The extent of dispersion correlates directly with the hydrodynamic radius of the species. Theory behind Taylor Dispersion Analysis Band broadening due to Taylor dispersion has been well characterized in previous literature. 1,2,3 In the method developed by Paraytec, UV area imaging is used to detect the peak of a slug of sample as it migrates through a fused silica capillary past two detection windows. The absorbance versus time data is processed to obtain the peak centre times at the first and second window, t 1 and t 2 respectively, and the corresponding standard deviations, 1 and 2, and variances, 1 2 and 2 2. These values are used to calculate the hydrodynamic radius, R h, using Equation 1 R 4k T r t t h B Equation 1 where k B is the Boltzmann constant, T the absolute temperature, the viscosity of the solution, and r the capillary radius. For dilute solutions used in these experiments, the viscosity of the solution may be assumed to be that of water at that temperature. Further details concerning Equation 1 and data analysis are available separately. 4 The ActiPix software automatically locates the peaks and uses Equation 1 to calculate the hydrodynamic radius of the sample as seen in Figure Paraytec Ltd 1

2 Figure 1. Example of experimental data from Taylor dispersion analysis. Output is using Gaussian fitting function available with Paraytec software. Summary of Technique Paraytec have developed a system for rapid determination of protein radii known as the ActiPix TDA200 Nano-Sizing System (Figure 2). This technology has previously been applied to determine its applicability for rapid quality control of quantum dots. 5 For the purposes of this application note, a brief summary of the technique is given below. The ActiPix Nano-Sizing System is a high precision nano-sizing system consisting of a precision nano-injector, autosampler and detector. Samples are typically stored in the autosampler prior to injection of a few nanolitres of each sample into a fused capillary. A plug of the sample, typically nl, is injected at the capillary inlet of a specially designed sizing cartridge (see Figure 3) and driven by application of external pressure along the capillary. UV absorption of the sample is recorded in the first and the second detection window using the integrated UV area imaging detector (ActiPix D100, Paraytec Ltd). Whilst the area of the peak is the same, the widths of both peaks are different: the signal from the second window has a greater width and lower amplitude due to Taylor dispersion. The peaks are fitted with an appropriate peak fitting function using software supplied with the system. The area under the peak corresponds to the amount of the sample injected. The standard deviations are used to calculate the hydrodynamic radius of the sample Paraytec Ltd 2

3 Figure 2. ActiPix TDA200 Nano-Sizing System. Figure 3. ActiPix Nano-Sizing Cartridge. Experimental: Instrumentation and Materials ActiPix TDA200 HT Nano-Sizing System with ActiPix NanoSizing software (Paraytec Ltd) ActiPix cartridge for sizing application (Paraytec Ltd) 214 nm wavelength filter (Paraytec Ltd) Capillary dimensions: 75 m nominal ID, 360 m OD (Polymicro Technologies) Capillary length: cm total length, 44.7 cm to 1st window, 40.1 cm between windows Protein standard: myoglobin (Sigma Aldrich) Buffer: phosphate buffered saline (PBS), ph 7.4 (Sigma Aldrich) Sample solutions: 1 mg/ml solution prepared by dissolving a weighed amount of protein in the buffer 0.1 mg/ml solution prepared by ten-fold dilution of the 1 mg/ml stock solution into PBS Run sequence Rinse: PBS; 1000 mbar, 5 min Inject: sample; 65 mbar, 12 s Dip: PBS; 0 mbar, 12 s Run: PBS; 65 mbar, 24 min Wash: 0.1 M NaOH; 1000 mbar, 5 min 2010 Paraytec Ltd 3

4 An alkaline wash was applied after every run to ensure the internal surface of the capillary was clean. A dip step was applied to remove any trace of sample on the outside of the capillary. The capillary was re-equilibrated with buffer prior to starting data collection. Stock solutions of myoglobin standard were prepared at 1 mg/ml and 0.1 mg/ml. Using the protocol described in the previous section (Run sequence), 9 consecutive runs with injection of the 1 mg/ml solution were made on each of 4 consecutive days. This enabled repeatability and intermediate precision to be determined for Paraytec s Taylor dispersion analysis method. Drift and limit of detection were also calculated using the data obtained. On one of the days, 9 consecutive runs with injection of the 0.1 mg/ml solution were carried out to evaluate repeatability at 0.1 mg/ml. Results Repeatability Figure 4 below is an overlay of primary data obtained from 9 consecutive runs with injection of 1 mg/ ml myoglobin. Excellent concordance is seen between consecutive runs. Figure 4. Overlay of 9 consecutive runs with injection of 1 mg/ml Paraytec Ltd 4

5 Hydrodynamic radius - repeatability and intermediate precision Table 1 below gives a summary of hydrodynamic radii obtained from 9 consecutive runs with injection of 1 mg/ml myoglobin. The mean of the average radii was ± nm with an RSD of 0.88%. Very good repeatability and intermediate precision from different lots of sample on consecutive days was seen, demonstrating the robustness of this technique. Hydrodynamic radius (nm) Run Day 1 Day 2 Day 3 Day AVERAGE STD DEV Peak area reproducibility RSD (%) Table 1. Repeatability and intermediate precision using Taylor dispersion analysis. A unique feature of determining hydrodynamic radius using Taylor dispersion analysis with UV area imaging in the Paraytec TDA200 instrument is that absolute peak areas are generated. Since peak areas are proportional to the amount of protein injected, this readily enables quantification of the proteins of interest. Table 2 below shows peak area for the data set obtained for Day 2. Peak area repeatability for this data set is concordant with the previous data set, and enables a peak area reproducibility of <1% to be specified. c1 Fit area (mau s) c2 Fit area (mau s) c1area/ c2area Filename Day Day Day Day Day Day Day Day Day Table 2. Peak area repeatability using Taylor dispersion analysis. Average SD RSD (%) Paraytec Ltd 5

6 Noise Figure 5 below shows a 6 minute sample of baseline to enable calculation of system baseline noise. A peak to peak value for noise of mau was determined at 214 nm using a 2 s time constant. Figure 5. Noise performance for TDA200 instrument. Analysis down to 0.1 mg/ml myoglobin Figure 6. Primary data file for 0.1 mg/ml myoglobin obtained using TDA200 software Paraytec Ltd 6

7 The analysis report shown in Figure 6 shows data with no filtering. The TDA200 software has smoothing functionality. In Figure 7 below, the same data is processed with a 2 s time constant. Figure 7. Primary raw data file from Figure 7 after reprocessing with a 2 s filter. Note that the net result of the filtering function is greatly reduced baseline noise. However, the value reported for the hydrodynamic radii are the same before ( nm) and after filtering ( nm). Hydrodynamic radius: repeatability at 1 mg/ml and 0.1 mg/ml 9 repeats each of 1 mg/ml and 0.1 mg/ml of myoglobin standard were run on the first day of the repeatability study. For the 1 mg/ml sample, a hydrodynamic radius value of ± nm was obtained with RSD = 0.79%. For the 0.1 mg/ml sample, a hydrodynamic radius value of ± nm was obtained with RSD = 2.16%. Good concordance is seen for the radii at both concentrations, demonstrating that the hydrodynamic radius of myoglobin is independent of concentration over the concentration range mg/ml. Accuracy of Taylor dispersion technique Taylor dispersion analysis is an absolute method for determination of hydrodynamic radius. The factor most likely to be the source of any systematic error in application of Equation 1 is the capillary radius, r. The capillary diameter (2r) is specified by the manufacturer as 75 ± 3 µm. This leads to a potential systematic error of 8% in the measured hydrodynamic radius. The actual radius of the capillary was determined using a gravimetric method, by measuring the weight changes in both the inlet and outlet vials after pressure driven flow of water at 1000 mbar for 20 minutes. These weight differences enabled determination of the actual radius to be 77.6 µm. Full details of the method and the calculation will be provided in a separate technical note. Table 3 gives recalculated hydrodynamic radii using the actual capillary radius of 77.6 µm. The accurate value of the hydrodynamic radius, using data obtained over three days, is ± nm. This shows that for accurate measurement of the hydrodynamic radius it is critical to know the actual radius of the capillary used. In using the nominal value for capillary radius, a systematic error of 7% is present in the hydrodynamic radius Paraytec Ltd 7

8 Day 1 Day 2 Day Average SD RSD Average SD RSD Table 3. Hydrodynamic radii re-calculated for myoglobin at 1 mg/ml using actual measured capillary radius of 77.6 µm, and data for radii reported to 5 significant figures. Limit of quantification Using raw data obtained for myoglobin at 1 mg/ml (for day 2, n=9), the peak amplitudes at each windows were found to be: First window: Second window: ± mau ± mau By combining these values with the measured peak to peak noise of 0.08 mau, the signal to noise ratios are S/N > 16/0.08 i.e. > 200. Thus quantification and size determination is readily measurable, assuming S/N = 10 down to 0.05 mg/ml. This is below the limit of 0.1 mg/ml normally required as the lower bound to measurements when carrying out protein sizing experiments. Drift Drift was determined using one of the data sets for myoglobin at 1 mg/ml. Using day 2 (n=9) data, drift was calculated at 0.3 ± 0.1 mau/hour. Conclusions Our study demonstrates very good repeatability and intermediate precision in determining the hydrodynamic radius of myoglobin (RSDs better than 1% at protein concentration 1 mg/ml) using Taylor dispersion analysis. Measurement of hydrodynamic radius is readily achieved at 0.1 mg/ml (RSD 2.2%), suggesting radius determination with acceptable precision down to at least 0.05 mg/ml. There is also very good repeatability in peak area. Taylor dispersion analysis shows that both protein concentration and protein radius can be determined in a single run using Paraytec s TDA200 Nano-Sizing System Paraytec Ltd 8

9 References 1. Bello, M. S.; Rezzonico, R.; Righetti, P. G., Science 1994, 266, Cottet, H.; Martin, M.; Papillaud, A.; Souaïd, E.; Collet, H. Commeyras, A., Biomacromolecules 2007, 8, Cottet, H.; Biron J.-P.; Martin, M., Anal. Chem. 2007, 79, Paraytec Technical Note TN001: Hydrodynamic radius using Taylor dispersion. 5. Paraytec Application Note AN005: Rapid sizing of quantum dots and nanoparticles Paraytec Ltd 9

Rapid sizing of quantum dots and nanoparticles

Rapid sizing of quantum dots and nanoparticles Rapid sizing of quantum dots and nanoparticles Keywords: Hydrodynamic radius, diffusion coefficient, sizing, Taylor dispersion analysis, non-organic, quantum dots, quality control, gold nanoparticles Summary

More information

Novel instrumentation to gain insight into the aggregation state of proteins by Taylor Dispersion Analysis: a comparison

Novel instrumentation to gain insight into the aggregation state of proteins by Taylor Dispersion Analysis: a comparison Novel instrumentation to gain insight into the aggregation state of proteins by Taylor Dispersion Analysis: a comparison with Dynamic Light Scattering Wendy Louise Hulse *, Rob Forbes, David Goodall, Alex

More information

ActiPix SDI300 Surface Dissolution Imaging System

ActiPix SDI300 Surface Dissolution Imaging System Product Note: PN002B Figure 1. SDI300 Surface Dissolution Imaging System Introduction The ActiPix SDI300 is a powerful UV area imaging system which enables quantitative imaging of surface phenomena for

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (11) 04 QUALIFICATION OF EQUIPMENT ANNEX 1: QUALIFICATION OF HPLC EQUIPMENT Full document title and reference Document type

More information

Physisorption of Antibodies using BioReady Bare Nanoparticles

Physisorption of Antibodies using BioReady Bare Nanoparticles TECHNICAL RESOURCE Lateral Flow Immunoassays Physisorption of Antibodies using BioReady Bare Nanoparticles Introduction For more than 20 years, lateral flow immunoassay diagnostic tests have provided a

More information

Measuring NP Aggregation Propensities

Measuring NP Aggregation Propensities Project: Measuring NP Aggregation Propensities Measuring the NP stability against aggregation as a function of time and buffers with batch-mode DLS AUTHORED BY: DATE: Luigi Calzolai 22-09-2015 REVIEWED

More information

Sulfotepp impurities in Chlorpyrifos EC formulations

Sulfotepp impurities in Chlorpyrifos EC formulations Page 1 of 16 Method DAS-AM-01-058 Sulfotepp impurities in Chlorpyrifos EC formulations A. ABSTRACT Method DAS-AM-01-058 has been validated for the analysis of the relevant impurity sulfotepp in chlorpyrifos

More information

Key Distance Learning Module I: Getting to Know Your Capillary

Key Distance Learning Module I: Getting to Know Your Capillary Key Distance Learning Module I: Getting to Know Your Capillary Electrophoresis System C.M. White, K.M. Hanson, L.A. Holland*, C. Eugene Bennett Department of Chemistry, West Virginia University, Morgantown,

More information

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION Andrew J. Aubin and Tanya L. Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION Benzocaine (4-Aminobenzoic

More information

The power of multi-detector SEC in the analysis of antibodies

The power of multi-detector SEC in the analysis of antibodies The power of multi-detector SEC in the analysis of antibodies How OMNISEC can measure antibody absolute MW independent of size and structural changes MOLECULAR SIZE MOLECULAR STRUCTURE MOLECULAR WEIGHT

More information

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity Agilent 12 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity A case study with degraded omeprazole drug product Application Note Small Molecule

More information

Supporting Information

Supporting Information Supporting Information Self-cleaning membranes from comb-shaped copolymers with photoresponsive side groups Papatya Kaner 1, Xiaoran Hu 2, Samuel W. Thomas III 2, Ayse Asatekin 1,* 1 Department of Chemical

More information

Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006

Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006 Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006 Introduction: Capillary electrophoresis (CE) is a relatively new, but rapidly growing separation

More information

3.1. Preparation of Buffer Solutions

3.1. Preparation of Buffer Solutions 3.1. Preparation of Buffer Solutions Experimental Part A volume of 50 ml of each buffer solution covering the ph range (pk a ±2) of the compound to be analyzed was prepared. For the classical method of

More information

Simultaneous analysis of the major metal cations and ammonium by CZE

Simultaneous analysis of the major metal cations and ammonium by CZE Simultaneous analysis of the major metal cations and ammonium by CZE This method has been published. You can download a copy of the PDF. Cite this method as: Warren CR, Adams MA (2004) Capillary electrophoresis

More information

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos

More information

Measuring Lysozyme Monomer at 0.1 mg/ml Concentration. Equipment used : Sample Preparation and Measurement :

Measuring Lysozyme Monomer at 0.1 mg/ml Concentration. Equipment used : Sample Preparation and Measurement : Application Report #001 Measuring Lysozyme Monomer at 0.1 mg/ml Concentration Equipment used : ALV-NIBS / HPPS High Sensitivity Version, Lysozyme (MERCK), 0.1 molar Sodium-Acetate buffer (ph 4.25), syringe

More information

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 126 Infinity Binary LC system with UV detection Application Note Food Author Srividya Kailasam Agilent Technologies,

More information

Organic Acids Analysis Kit

Organic Acids Analysis Kit Capillary Electrophoresis Organic Acids Analysis Kit PN 5063-6510 2 Upon receipt, please verify that all kit contents listed on page 4 are included. If any part is missing, contact your Agilent Technologies

More information

CONFOCHECK. Innovation with Integrity. Infrared Protein Analysis FT-IR

CONFOCHECK. Innovation with Integrity. Infrared Protein Analysis FT-IR CONFOCHECK Infrared Protein Analysis Innovation with Integrity FT-IR CONFOCHECK: FT-IR System for Protein Analytics FT-IR Protein Analysis Infrared spectroscopy measures molecular vibrations due to the

More information

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System Application Note Residual Solvent Analysis of USP Method Residual Solvents on the Agilent 889 GC System Author Lukas Wieder, Jie Pan, and Rebecca Veeneman Agilent Technologies, Inc. 8 Centerville Road

More information

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC Technical Procedure for Concentration Determination of 1.0 Purpose This procedure specifies the required elements for the preparation and use of the Agilent 1100/1200 series High Performance Liquid Chromatograph

More information

MEASURING PROTEIN AGGREGATION WITH THE VISCOTEK SEC-MALS 20

MEASURING PROTEIN AGGREGATION WITH THE VISCOTEK SEC-MALS 20 MEASURING PROTEIN AGGREGATION WITH THE VISCOTEK SEC-MALS 20 Introduction Protein aggregation is recognised as a major issue in the biopharmaceutical industry. Proteins have a tendency to aggregate over

More information

TA Instruments Application Note

TA Instruments Application Note Q ( µj) TA Instruments Application Note Isothermal Titration Calorimetry (ITC) with Reduced Cell Volumes: A Comparison of the TA Instruments Nano ITC-Low Volume with the GE Healthcare Auto-iTC 200. Colette

More information

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC USP Method Transfer and Routine Use Analysis of Tablets from HPLC to UPLC Aparna Chavali, Tanya Jenkins and Patricia McConville Waters Corporation, 34 Maple Street, Milford, MA USA APPLICATION BENEFITS

More information

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection APPLICATION NOTE 72686 Determination of underivatized aflatoxins,, G2, and in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection Authors Sylvia Grosse, Mauro De Pra, Frank

More information

8. FORMULATION OF LANSOPRAZOLE NANOPARTICLES

8. FORMULATION OF LANSOPRAZOLE NANOPARTICLES 8. FORMULATION OF LANSOPRAZOLE NANOPARTICLES FORMULATION OF LANSOPRAZOLE NANOPARTICLES Preparation of capsule of modified solubility to protect the drug from degradation To protect the drug from degradation

More information

Protein assay of SpectroArt 200

Protein assay of SpectroArt 200 Technical Bulletin 14 SpectroArt 200 12/01/2008 Protein assay of SpectroArt 200 MATERIAL BSA: Albumin, bovine serum (Sigma) PBS: BupH TM Phosphate Buffered Saline packs (PIERCE) Bradford assay: Bio-Rad

More information

Experimental Design and Data Collection

Experimental Design and Data Collection A. Checks to run to make sure the instrument is in good working condition B. Decide which experiment type is most appropriate, equilibrium or velocity C. Buffer considerations D. Speed selection and length

More information

Field-Flow Fractionation of Macromolecules and Structures That Cannot be Characterized by Conventional GPC/SEC Techniques

Field-Flow Fractionation of Macromolecules and Structures That Cannot be Characterized by Conventional GPC/SEC Techniques The Field-Flow Fractionation Platform Field-Flow Fractionation of Macromolecules and Structures That Cannot be Characterized by Conventional GPC/SEC Techniques Trevor Havard, Evelin Moldenhaur, Soheyl

More information

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump Performance characteristics of the Agilent 129 Infinity Quaternary Pump Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract This Technical Overview presents Proof

More information

SCIEX Cation Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide

SCIEX Cation Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems A49109AE May 2015 AB Sciex Pte. Ltd and its affiliates disclaims all warranties with respect to this document, expressed or implied, including

More information

Advantages of Agilent AdvanceBio SEC Columns for Biopharmaceutical Analysis

Advantages of Agilent AdvanceBio SEC Columns for Biopharmaceutical Analysis Advantages of Agilent AdvanceBio SEC Columns for Biopharmaceutical Analysis Comparing Columns from Different Vendors to Improve Data Quality Technical Overview Introduction Size exclusion chromatography

More information

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China Rapid Deteration of Eight Related Aromatic Acids in the p-phthalic Acid Mother Liquid Using an Agilent 126 Infinity LC System and an Agilent Poroshell 12 SB-C18 Column Application Note Author Xinlei Yang

More information

Advances in particle concentration measurements

Advances in particle concentration measurements Advances in particle concentration measurements Dr Hanna Jankevics Jones Principal Applications Scientist Company Confidential Outline Advances in particle concentration from Malvern Panalytical Capture

More information

BSA pegylation PG Journal of chromatographic A, 1147 (2007)

BSA pegylation PG Journal of chromatographic A, 1147 (2007) BSA pegylation BSA 5mg/mL, pi= 4.7, Sigma A796, 65-7% free sulfyhydryl, cysteine 34 of BSA is not paired with another cysteine in the protein structure 1. PEG 1kD, kd, 3 kd 1mg/mL PEG:BSA=1.2:1 (molar

More information

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection Trade Science Inc. September 2009 Volume 8 Issue 3 ACAIJ, 8(3) 2009 [346-349] Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization

More information

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform -xt SYSTEM Prep and Load Platform -xt Specifications for PAL-xt Systems Valid for PAL-xt System models only Revised March 2013 V5 PAL is a registered trademark of CTC Analytics AG Switzerland -xt SYSTEM

More information

Supplementary information 1. INSTRUCTIONS FOR STUDENTS

Supplementary information 1. INSTRUCTIONS FOR STUDENTS 1 2 Supplementary information 1. INSTRUCTIONS FOR STUDENTS 1.- SAFETY AND WASTE DISPOSAL Reactant CAS Pictogram * Hazards L-ascorbic acid 50-81-7 - Not a dangerous substance Hexadimethrine bromide 28728-55-4

More information

Separation Sciences. 1. Introduction: Fundamentals of Distribution Equilibrium. 2. Gas Chromatography (Chapter 2 & 3)

Separation Sciences. 1. Introduction: Fundamentals of Distribution Equilibrium. 2. Gas Chromatography (Chapter 2 & 3) Separation Sciences 1. Introduction: Fundamentals of Distribution Equilibrium 2. Gas Chromatography (Chapter 2 & 3) 3. Liquid Chromatography (Chapter 4 & 5) 4. Other Analytical Separations (Chapter 6-8)

More information

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD) Application Note 19010803 A Newly Approved ASTM Standard For Analysis of in Using a Pulsed Flame Photometric Detector (PFPD) Keywords ASTM Standard D4735-96 FPD Flame Photometric Detector Model 5380 PFPD

More information

Supplementary information

Supplementary information Supplementary information Self-assembled light-driven photosynthetic-respiratory electron transport chain hybrid proton pump David Hvasanov, a Joshua R. Peterson a and Pall Thordarson *a a School of Chemistry,

More information

Meet Stunner: The one-shot protein concentration and sizing combo

Meet Stunner: The one-shot protein concentration and sizing combo TECH NOTE Meet Stunner: The one-shot protein concentration and sizing combo Introduction What if you could get a better read on the quality of your biologics and use less sample at the same time? Stunner

More information

Appendix II- Bioanalytical Method Development and Validation

Appendix II- Bioanalytical Method Development and Validation A2. Bioanalytical method development 1. Optimization of chromatographic conditions Method development and optimization of chromatographic parameters is of utmost important for validating a method in biological

More information

ProPac WCX-10 Columns

ProPac WCX-10 Columns ProPac WCX-10 Columns Guidance for column use Tips to maximize column lifetime ProPac WCX-10 Column Tips and Tricks This guide provides essential information and invaluable guidelines for mobile phases,

More information

Protocol for Coating QD-COOH on glass slides Chris Ochs 19/09/12 Modified by Kathy Lu 2/27/2013

Protocol for Coating QD-COOH on glass slides Chris Ochs 19/09/12 Modified by Kathy Lu 2/27/2013 Protocol for Coating QD-COOH on glass slides cjochs@smart.mit.edu Chris Ochs 19/09/12 Modified by Kathy Lu 2/27/2013 kalu@ucsd.edu Cleaning glass slides prior to coupling and Amination with APTS (Aminopropyl

More information

Optimization of an Adapta Kinase Assay for CAMK1

Optimization of an Adapta Kinase Assay for CAMK1 Overview Optimization of an Adapta Kinase Assay for CAMK1 This protocol describes how to perform an Adapta assay with the kinase CAMK1. To maximize the ability of the assay to detect ATP-competitive inhibitors,

More information

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC. 7. Stability indicating analytical method development and validation of and in capsule dosage form by HPLC. 7.1 INSTRUMENTS AND MATERIALS USED 7.1.1 INSTRUMENTS 1. Shimadzu LC-2010 CHT with liquid chromatograph

More information

Supporting Information

Supporting Information This journal is (c) The Royal Society of Chemistry 21 Zeta potential based Colorimetric Immunoassay for the direct detection of Diabetic marker HbA1c using Gold Nanoprobes Nishima Wangoo, a,b Jyotsna Kaushal,

More information

Monitoring Protein PEGylation with Ion Exchange Chromatography

Monitoring Protein PEGylation with Ion Exchange Chromatography Monitoring Protein PEGylation with Ion Exchange Chromatography Peter Yu, Deanna Hurum, Leo Wang, Terry Zhang, and Jeffrey Rohrer 245 th ACS National Meeting & Exposition April 9, 213 OT7618_E 4/13S 1 The

More information

DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR IN-VIVO STUDY OF DICLOFENAC POTASSIUM

DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR IN-VIVO STUDY OF DICLOFENAC POTASSIUM IJPSR (2013), Vol. 4, Issue 2 (Research Article) Received on 28 September, 2012; received in revised form, 24 November, 2012; accepted, 23 January, 2013 DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR

More information

PYRIPROXYFEN TECHNICAL

PYRIPROXYFEN TECHNICAL WHO/IS/TC/715/2001 TECHNICAL TECHNICAL 1. Specification 1.1 Description Interim specification WHO/IS/TC/715/2001 The material shall consist of pyriproxyfen together with related manufacturing impurities.

More information

Multiphase Oscillatory Flow Strategy for in Situ Measurement and Screening of Partition Coefficients

Multiphase Oscillatory Flow Strategy for in Situ Measurement and Screening of Partition Coefficients Supporting Information Multiphase Oscillatory Flow Strategy for in Situ Measurement and Screening of Partition Coefficients Milad Abolhasani, Connor W. Coley, and Klavs F. Jensen * Department of Chemical

More information

Zetasizer Nano ZSP: A Perfect Tool For Life Science Applications

Zetasizer Nano ZSP: A Perfect Tool For Life Science Applications Zetasizer Nano ZSP: A Perfect Tool For Life Science Applications Dr Mike Kaszuba Technical Support Manager E-mail: michael.kaszuba@malvern.com Contents Zetasizer Nano ZSP Software Enhancements Protein

More information

RESOLUTION OENO 33/2004 DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION

RESOLUTION OENO 33/2004 DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION The GENERAL ASSEMBLY, Considering Article paragraph iv of the agreement establishing the International organisation of vine and wine Upon

More information

BENZYLPIPERAZINE Latest Revision: June 1, 2005

BENZYLPIPERAZINE Latest Revision: June 1, 2005 BENZYLPIPERAZINE Latest Revision: June 1, 2005 HN N 1. SYNONYMS CFR: Schedule I CAS #: 2759-28-6 Stride II: Other Names: N-Benzylpiperazine 1-Benzylpiperazine N-Benzylpiperazine 1-(phenylmethyl)piperazine

More information

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID Asian Journal of Chemistry Vol. 21, No. 3 (2009), 1739-1746 Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID PRAVEEN KUMAR BALIYAN*, R.P.

More information

SRT -10/10C (10 µm ) Fast Protein Purification For FPLC, HPLC and Prep LC. Sepax Technologies, Inc. Preparative Size Exclusion Chromatography

SRT -10/10C (10 µm ) Fast Protein Purification For FPLC, HPLC and Prep LC. Sepax Technologies, Inc. Preparative Size Exclusion Chromatography Sepax Technologies, Inc. Preparative Size Exclusion Chromatography SRT -0/0C (0 µm ) Fast Protein Purification For FPLC, HPLC and Prep LC Better Surface Chemistry for Better Separation SRT -0/0C Size Exclusion

More information

The Diffusion Barrier Technique, Practical Aspects and Data interpretation

The Diffusion Barrier Technique, Practical Aspects and Data interpretation The Diffusion Barrier Technique, Practical Aspects and Data interpretation Introduction The diffusion barrier technique is a method for reliably measuring the electrophoretic mobility of proteins by minimising

More information

Introduction to Dynamic Light Scattering with Applications. Onofrio Annunziata Department of Chemistry Texas Christian University Fort Worth, TX, USA

Introduction to Dynamic Light Scattering with Applications. Onofrio Annunziata Department of Chemistry Texas Christian University Fort Worth, TX, USA Introduction to Dynamic Light Scattering with Applications Onofrio Annunziata Department of Chemistry Texas Christian University Fort Worth, TX, USA Outline Introduction to dynamic light scattering Particle

More information

LC Technical Information

LC Technical Information LC Technical Information Method Transfer to Accucore.6 μm Columns Containing solid core particles, which are engineered to a diameter of.6μm and a very narrow particle size distribution; Accucore HPLC

More information

Chirascan 6-Cell Peltier Cell Holder: Rapid Optimisation of Buffer Conditions for Stabilising Protein Therapeutics

Chirascan 6-Cell Peltier Cell Holder: Rapid Optimisation of Buffer Conditions for Stabilising Protein Therapeutics CHIRASCAN SERIES APPLICATION NOTE Chirascan 6-Cell Peltier Cell Holder: Rapid Optimisation of Buffer Conditions for Stabilising Protein Therapeutics Abstract: The selection of buffer conditions that maximise

More information

DEVELOPING A BIOFORMULATION STABILITY PROFILE

DEVELOPING A BIOFORMULATION STABILITY PROFILE DEVELOPING A BIOFORMULATION STABILITY PROFILE Light Scattering and Micro-Capillary Viscometry A Malvern Instruments Bioscience Development Initiative Executive Summary: When used in combination, light

More information

Electronic supplementary information

Electronic supplementary information Electronic supplementary information Surface plasmon resonance enhanced upconversion luminescence in aqueous media for TNT selective detection Nina Tu and Leyu Wang* State Key Laboratory of Chemical Resource

More information

Polymer analysis by GPC-SEC. Technical Note. Introduction

Polymer analysis by GPC-SEC. Technical Note. Introduction Polymer analysis by GPC-SEC Technical Note Introduction Gel Permeation Chromatography (GPC), also referred to as Size Exclusion Chromatography (SEC) is a mode of liquid chromatography in which the components

More information

ITC Expert User s Manual

ITC Expert User s Manual ITC Expert User s Manual 1 Section 1: ITC Expert Background... 3 Minimal Heats and Injections... 3 C Parameter... 3 C Limitations... 4 High C... 4 Low C... 6 Concentrations Ratio... 6 Section 2: ITC Expert

More information

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Application note Geochemistry, metals, mining Authors John Cauduro Agilent Technologies, Mulgrave, Australia Introduction

More information

Hybrid Gold Superstructures: Synthesis and. Specific Cell Surface Protein Imaging Applications

Hybrid Gold Superstructures: Synthesis and. Specific Cell Surface Protein Imaging Applications Supporting Information Hybrid Gold Nanocube@Silica@Graphene-Quantum-Dot Superstructures: Synthesis and Specific Cell Surface Protein Imaging Applications Liu Deng, Ling Liu, Chengzhou Zhu, Dan Li and Shaojun

More information

1-(2-METHOXYPHENYL)PIPERAZINE Latest revision: June 27, 2005

1-(2-METHOXYPHENYL)PIPERAZINE Latest revision: June 27, 2005 1-(2-METHOXYPHENYL)PIPERAZINE Latest revision: June 27, 2005 HN O N 1. SYNONYMS CFR: N/A CAS #: Base: 35386-24-4 Hydrochloride: 5464-78-8 Other Names: 2-MeOPP 2-Methoxyphenylpiperazine 2. CHEMICAL AND

More information

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Figure 1. Structures for Vitamin B2 and Vitamin B1. CH 461 & CH 461H F 18 Name Experiment 2C Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART C. HIGH PERFORMANCE LIQUID CHROMATOGRAPHY The purpose of this experiment

More information

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD TO DETERMINE CINITAPRIDE HYDROGEN TARTARATE IN BULK AND PHARMACEUTICAL FORMULATION

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD TO DETERMINE CINITAPRIDE HYDROGEN TARTARATE IN BULK AND PHARMACEUTICAL FORMULATION Research Article ISSN:2230-7346 Journal of Global Trends in Pharmaceutical Sciences Vol.3, Issue 2, pp -619-627, April June 2012 DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD TO DETERMINE CINITAPRIDE HYDROGEN

More information

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Application Note Food Testing and Agriculture Author Chen-Hao (Andy) Zhai and Rong-jie Fu Agilent Technologies (Shanghai)

More information

MALVERN ADVANCED GPC/SEC DETECTORS SETTING THE STANDARD MOLECULAR WEIGHT MOLECULAR SIZE MOLECULAR STRUCTURE

MALVERN ADVANCED GPC/SEC DETECTORS SETTING THE STANDARD MOLECULAR WEIGHT MOLECULAR SIZE MOLECULAR STRUCTURE MOLECULAR WEIGHT MOLECULAR SIZE MOLECULAR STRUCTURE MALVERN ADVANCED GPC/SEC DETECTORS SETTING THE STANDARD Setting the Standard WHAT CAN ADVANCED GPC/SEC DETECTION OFFER YOU? A modern research environment

More information

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography Technical Overview 2D-LC Conventional 1D-LC Abstract This Technical Overview presents

More information

ENV/JM/MONO(2015)17/PART1/ANN2

ENV/JM/MONO(2015)17/PART1/ANN2 Unclassified ENV/JM/MONO(2015)17/PART1/ANN2 ENV/JM/MONO(2015)17/PART1/ANN2 Unclassified Organisation de Coopération et de Développement Économiques Organisation for Economic Co-operation and Development

More information

Supporting Materials Ultra-small Sub-10 nm Near Infrared Fluorescent Mesoporous Silica Nanoparticles

Supporting Materials Ultra-small Sub-10 nm Near Infrared Fluorescent Mesoporous Silica Nanoparticles Supporting Materials Ultra-small Sub-10 nm Near Infrared Fluorescent Mesoporous Silica Nanoparticles Kai Ma, Hiroaki Sai and Ulrich Wiesner* Department of Materials Science and Engineering, Cornell University,

More information

637. Thiamethoxam. HPLC method

637. Thiamethoxam. HPLC method 637. Thiamethoxam HPLC method CIPAC Collaborative Trial according to CIPAC Information Sheet N o 293 Dr. Sven Adolph Syngenta Crop Protection Münchwilen AG CH-4333 Münchwilen Switzerland May 212 page 1

More information

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM IC Application Work AW CH6-1048-012011 Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM Branch: Chemical industry; Water, wastewater, environmental protection,

More information

Highly automated protein formulation development: a case study

Highly automated protein formulation development: a case study Application Note Highly automated protein formulation development: a case study Introduction and background Therapeutic proteins can be inherently prone to degradation and instability, and they often pose

More information

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Andrea Caruso and Massimo Santoro, Thermo Fisher Scientific, Milan, Italy Application Note 1316 Key

More information

Using runt Disposable Cuvettes

Using runt Disposable Cuvettes Tech Note Guide Table of Contents Ease of Use... 3 Sensitivity... 4 Resolution... 5 Accuracy... 6 Faster Sample Prep... 7 Measuring Molecular Weight... 8 Observing Protein Aggregation... 9 Characterize

More information

Using High Speed/High Resolution Size Exclusion Chromatography Separation of Polymeric Materials with Light Scattering Detection

Using High Speed/High Resolution Size Exclusion Chromatography Separation of Polymeric Materials with Light Scattering Detection Using High Speed/High Resolution Size Exclusion Chromatography Separation of Polymeric Materials with Light Scattering Detection Jennifer Gough 1, Michael Jones 1, Damian Morrison 1, and John Stenson 2

More information

PERFORMANCE VERIFICATION IONSCAN LS

PERFORMANCE VERIFICATION IONSCAN LS Page of 3 PERFORMANCE VERIFICATION IONSCAN LS Smiths Detection 30 Technology Dr. Warren, NJ 07059 908--900 Copyright Smiths Detection 004 All rights reserved. Reproduction, adaptation or translation without

More information

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Figure 1. Structures for Vitamin B2 and Vitamin B1. CH 461 & CH 461H F 15 Name Experiment 2C Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART C. HIGH PERFORMANCE LIQUID CHROMATOGRAPHY The purpose of this experiment

More information

Research Article METHOD DEVELOPMENT AND VALIDATION FOR SIMULTANEOUS ESTIMATION OF ELBASVIR AND GRAZOPREVIR BY RP-HPLC

Research Article METHOD DEVELOPMENT AND VALIDATION FOR SIMULTANEOUS ESTIMATION OF ELBASVIR AND GRAZOPREVIR BY RP-HPLC ISSN 2395-3411 Available online at www.ijpacr.com 248 Research Article METHOD DEVELOPMENT AND VALIDATION FOR SIMULTANEOUS ESTIMATION OF ELBASVIR AND GRAZOPREVIR BY RP-HPLC Mantha Vebkatesh* and A. Yasodha

More information

The appearance of phenolphthalein is measured spectrophotometrically at 540nm (2).

The appearance of phenolphthalein is measured spectrophotometrically at 540nm (2). Document Title β-glucuronidase Page 1 of 5 Originating Department QC Approval Departments QA, QC Approval Date 23 rd May 2017 Effective Date 8 th June 2017 1.0 PRODUCT DETAILS 1.1 Enzyme Name: β-glucuronidase

More information

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc. Agilent Application Solution Transfer of a USP method for tolazamide from normal phase HPLC to SFC using the Agilent 126 Infinity Hybrid SFC/UHPLC System Improving peak shape and providing wider UV selectivity

More information

Determination of Polymer Modifier in Asphalt

Determination of Polymer Modifier in Asphalt Standard Method of Test for Determination of Polymer Modifier in Asphalt AASHTO Designation: T xxx-xx (2005) 1. SCOPE 1.1. This method of test is used to determine the polymer content of an asphalt sample.

More information

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd.

More information

Fast ph-assisted functionalization of silver nanoparticles with monothiolated DNA

Fast ph-assisted functionalization of silver nanoparticles with monothiolated DNA Supporting Information for Fast ph-assisted functionalization of silver nanoparticles with monothiolated DNA Xu Zhang ab, Mark R. Servos b, and Juewen Liu* a a Department of Chemistry and Waterloo Institute

More information

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis High-Resolution Sampling D-LC for Pharmaceutical Impurity Analysis Detection of Impurities Hidden Under the API Peak at Relevant Levels Application Note Small Molecule Pharmaceuticals and Generics Authors

More information

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application Pharmaceuticals Author Cliff Woodward and Ronald Majors Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808

More information

Meet Our Uncle: 12 Stability Applications on One Platform

Meet Our Uncle: 12 Stability Applications on One Platform Tech Note Meet Our Uncle: 12 Stability Applications on One Platform Uncle is an all-in-one stability platform that enables twelve different applications with one instrument. Fluorescence, static light

More information

Data sheet. UV-Tracer TM Biotin-Maleimide. For Labeling of Thiol-groups with UV-detectable Biotin CLK-B Description

Data sheet. UV-Tracer TM Biotin-Maleimide. For Labeling of Thiol-groups with UV-detectable Biotin CLK-B Description Cat. No. CLK-B105-10 CLK-B105-100 Amount 10 mg 100 mg For in vitro use only! Quality guaranteed for 12 months Store at -20 C 1. Description UV-Tracer TM Biotin Maleimide for biotinylation reactions of

More information

Strikingly different miscibility of n-octanol in highly-confined and quasi-confined water

Strikingly different miscibility of n-octanol in highly-confined and quasi-confined water Electronic Supplementary Material (ESI) for ChemComm. This journal is The Royal Society of Chemistry 2015 Strikingly different miscibility of n-octanol in highly-confined and quasi-confined water Aparajita

More information

ELECTRONIC SUPPLEMENTARY INFORMATION (ESI) variable light emission created via direct ultrasonic exfoliation of

ELECTRONIC SUPPLEMENTARY INFORMATION (ESI) variable light emission created via direct ultrasonic exfoliation of Electronic Supplementary Material (ESI) for RSC Advances. This journal is The Royal Society of Chemistry 2015 ELECTRONIC SUPPLEMENTARY INFORMATION (ESI) High quantum-yield luminescent MoS 2 quantum dots

More information

Exploiting the Interaction of Pyranine-3 with Poly(L-Lysine) to Mediate Nanoparticle Assembly: Fabrication of Dynamic ph-responsive Nanocontainers

Exploiting the Interaction of Pyranine-3 with Poly(L-Lysine) to Mediate Nanoparticle Assembly: Fabrication of Dynamic ph-responsive Nanocontainers Exploiting the Interaction of Pyranine-3 with Poly(L-Lysine) to Mediate Nanoparticle Assembly: Fabrication of Dynamic ph-responsive Nanocontainers Arlin Jose Amali, a Shashi Singh, b Nandini Rangaraj,

More information

The ultimate in desktop particle characterization

The ultimate in desktop particle characterization Particle size Zeta potential Molecular weight ZS The ultimate in desktop particle characterization detailed specification from www.malvern.com/zetasizernano The Zetasizer Nano ZS brings you the practicality

More information

Make or Buy? The Economics of Gold Nanoparticle Manufacturing for Lateral Flow Assays

Make or Buy? The Economics of Gold Nanoparticle Manufacturing for Lateral Flow Assays TECHNICAL RESOURCE Lateral Flow Immunoassays Make or Buy? The Economics of Gold Nanoparticle Manufacturing for Lateral Flow Assays Introduction Price is an important factor in the commercial success of

More information

Peptide Isolation Using the Prep 150 LC System

Peptide Isolation Using the Prep 150 LC System Jo-Ann M. Jablonski and Andrew J. Aubin Waters Corporation, Milford, MA, USA APPLICATION BENEFITS The Prep 150 LC System, an affordable, highly reliable system for preparative chromatography, is suitable

More information